Diagnostic Accuracy of a Simple and Rapid Diagnostic Test RLDT for the Detection of Cholera in Bangladesh

Kamrul Islam 1, Ashraful Islam Khan 2, M D Taufiqur Rahman Bhuiyan 3, Fan He 4, Zahid Hasan Khan 5, Md Taufiqul Islam 6, Firdausi Qadri 7,1, Subhra Chakraborty 8,1,✉,3
PMCID: PMC13218445  PMID: 42221374

Abstract

Background

The rising incidence and case fatality rates of cholera outbreaks underscore the need for better diagnostic tests for early detection and confirmation of the outbreak to enable rapid responses and obtain reliable surveillance data. We evaluated the Rapid LAMP-based Diagnostic Test (RLDT) for detecting cholera to determine if this test has the potential to fill the gap.

Methods

This study included training and evaluation of RLDT in the cholera reference laboratory in Bangladesh. Stool samples collected from suspected cholera patients seeking care at the icddr,b hospital, Dhaka, were tested using cholera RLDT, microbiological culture, Rapid Diagnostic Test (RDT crystal VC O1, O139), and PCR and compared.

Results

Out of 205 cases analyzed, 136 (66%) were positive for cholera by RLDT, 131 (64%) by PCR, 92 (45%) by culture, and 90 (44%) by RDT. The RLDT was highly specific, as all the culture positives were positive by RLDT, except 1 (sensitivity of 99%), and could detect all the PCR positives except 2 (sensitivity of 97%). Almost perfect agreement was found between RLDT and PCR (kappa 0.9). Rapid LAMP-based Diagnostic Test being more sensitive than culture and RDT, the agreement was moderate (kappa 0.6 and 0.5, respectively). Rapid Diagnostic Test missed detection of cholera in 11 samples, positive by other tests. The accuracy of RLDT was 100% in detecting cholera from water samples.

Conclusions

Rapid LAMP-based Diagnostic Test was specific and sensitive, detecting 21% more cholera-positive cases than culture or RDT. Rapid LAMP-based Diagnostic Test has the potential to improve the rapid detection of cholera in outbreak situations and for surveillance data.

Keywords: cholera, diagnostic test, diarrhea, RDT, RLDT


Rapid LAMP-based Diagnostic Test could be implemented in a reference laboratory in low and middle-income where cholera is endemic. Rapid LAMP-based Diagnostic Test is highly specific and more sensitive compared to culture and Rapid Diagnostic Test for the detection of cholera. Rapid LAMP-based Diagnostic Test could be scaled up for cholera detection.


Cholera is a severe dehydrating diarrheal disease caused by Vibrio cholerae serogroup O1 and sporadically by V. cholerae O139. Globally, 3–5 million cases and over 120 000 deaths occur annually due to cholera [1]. Approximately 1.3 billion people are at risk of cholera around the world [2, 3].

Stool culture remains the gold standard for laboratory detection and surveillance of cholera, although stool culture is not 100% sensitive [4, 5]. Culturing stool or rectal swab on the selective media either directly or after enrichment in alkaline peptone water (APW), followed by biochemical analysis, serotyping, and bio-typing with antisera [6–8], takes 2–4 days and requires competent laboratory support and technical skills, which are not always available where cholera is endemic.

Several PCR methods have been developed, but a standard, validated technique is not available [9]. PCR requires trained technicians, a well-equipped laboratory, and a supply of cold chain–maintained reagents that are not often available outside the central laboratories.

Several cholera Rapid Diagnostic Tests (RDTs) have been developed. The commonly used RDTs are based on the lateral flow that detects the lipopolysaccharide of V. cholerae O1 and O139 by immunochromatographic assay [10]. Although RDTs are well-suited to meet the demand for cholera diagnosis where culture or PCR-based methods are not feasible, these RDTs suffer from limitations. The most evaluated RDT, Crystal VC (Arkray Healthcare Pvt. Ltd, Gujarat, India), has demonstrated a wide variation of sensitivity and specificity (ranging from 58% to 100% respectively) [4, 11–18]. A systematic review of diagnostic tests for cholera demonstrated a notable absence of evidence to support the use of these rapid tests [11] as they did not meet the expected minimal performance of a sensitivity of at least 90% and a specificity of at least 85% recommended by the Global Task Force for Cholera Control for the Target Product Profile of RDT [19]. Cholera surveillance and disease burden estimation also rely on the accuracy of the diagnostic assay.

The case fatality rate is often highest at the beginning of an outbreak or epidemic, and delayed recognition of a cholera outbreak results in a delayed public health response, resulting in high mortality [4]. Therefore, there is a crucial need for a rapid cholera diagnostic test with better accuracy, which can confirm the outbreak early and at the outbreak site to initiate rapid response and also obtain reliable surveillance data to identify the priority areas for multisectoral intervention.

Cholera is a water-borne infectious disease. Methods for identifying culturable V. cholerae from water and sewage include culture and PCR techniques [20–22]. We previously adapted the Crystal VC dipstick assay and developed a simple technique for identifying cholera in water [23]. However, this assay requires incubation in APW for 24 hours. A rapid but sensitive assay would be valuable to quickly identify if water sources are contaminated with cholera.

To address these challenges in the detection of cholera, Chakraborty et al [10] developed a simple, sensitive, and rapid detection test, “Rapid LAMP (Loop-mediated Isothermal Amplification)–based Diagnostic Test (RLDT),” for the detection of cholera from both fecal and water samples, which is cold-chain and almost electricity-free. The RLDT detects cholera directly from stool in less than an hour, with the lowest detection limit similar to quantitative PCR [10]. In this study, we evaluated the performance of the RLDT test in comparison with conventional PCR, culture, and the commercial RDT, Crystal VC, in the reference laboratory of a cholera-endemic country, Bangladesh.

METHODS

Ethics Statement

This study was approved by the Research Review and the Ethical Review Committees of the International Centre for Diarrhoeal Disease Research, Bangladesh (icddr,b) and the Institutional Review Board of the Johns Hopkins University. Written informed consent was obtained from the caregiver of the children, assent from children those 11–17 years of age, and adults gave their own consent.

Study Participants and Specimen Collection

We collected stool specimens from 210 adults and children who were seeking care for diarrhea at the Dhaka Hospital of the icddr,b, Bangladesh, during October 2021 to April 2022. Of note, from March 1 to April 10, one of the worst cholera outbreaks occurred in Dhaka, and a huge number of patients attended the icddr,b hospital for treatment [24]. Thus, we had the opportunity to test the RLDT during a cholera outbreak in this study.

Diagnostic Tests to Detect Cholera From Stool

Microbiological culture was performed by culturing stool directly on taurocholate-tellurite gelatin agar plates or after enrichment in APW (1% peptone, 1% NaCl; pH 8.5) for overnight, followed by confirmation of the V. cholerae isolates by slide agglutination with monoclonal antibodies specific to V. cholerae serovar O1 (Ogawa or Inaba) and O139.

The Crystal VC RDT (includes both O1 and O139) was performed on fresh stool samples according to the manufacturer's instructions. In brief, liquid stool was mixed with diluent and the test strip was dipped into the tube followed by recording the results.

We conducted the PCR from the dried stool on filter paper as follows. The stool spot was cut into small pieces and placed into a microfuge tube, washed, and 150 µL of nuclease-free deionized water was added, followed by heat lysis at 100°C for 10 minutes. The lysate was used for multiplex PCR on a Thermo cycler C-1000 instrument (Bio-Rad) targeting V. cholerae O1-rfb and ctxA genes [25]. PCR products were analyzed on a 2% agarose gel using Gel Red (BioTium, USA) stain. To confirm the discordant results between RLDT and PCR, we used another set of primers targeting the gene of outer membrane protein (OmpW) [26].

Rapid LAMP-based Diagnostic Test assay was performed directly from fresh stool specimens using the RLDT kit as previously described [10]. In short, a stool sample was added to the sample processing tube containing lysis buffer, processed by heat lysis, and the lysate was added to the RLDT reaction strips, which were transferred to a battery-operated handheld reader for reading the result in 40 minutes. The RLDT strips contain all the reagents needed for amplification, already lyophilized, targeting O1 rfb and ctxA genes, and an assay inhibitor control. The positive or negative result was interpreted automatically by the reader [10].

Cholera Detection From the Water Samples

From the cholera patients enrolled in this study, the first 50 patients who gave consent to collect drinking water from their households were chosen. The water samples were collected from either source water (eg, tap water) or stored water. After 400 mL of water samples were filtered through a 0.22 μm GV membrane (GVWP02500, Durapore, Merck Millipore, Ireland), the membrane was cut in half, and one half was used for RLDT directly [10]. The remaining half was placed in APW to enrich for overnight at 37°C, followed by culture on Taurocholate-Tellurite-Gelatin Agar (TTGA) and serotyping. For RLDT, the first half of the membrane was cut into small pieces and heat lysed using lysis buffer, followed by process as described for stool [10] and also tested from the enriched samples.

Statistical Analysis

Cohen's kappa was used to describe the agreement between a pair of detection methods. GraphPad Prism7 was used for analysis and graphical presentation of data. Sensitivity and specificity of different diagnostic tests were calculated using the R version 4.5.1. Although RLDT detects both O1 and ctxA genes, since the latter alone could be confirmatory of cholera, the results shown in this analysis are based primarily on the positivity of ctxA gene for both RLDT and PCR, unless otherwise noted.

Interpretation of Cohen's Kappa Test

Cohen's kappa is a reliable statistical measure used to assess both interrater and intrarater agreement. Like correlation coefficients, its values range from −1 to +1, where 0 reflects agreement expected by chance and 1 indicates perfect agreement between raters. Interpretation of kappa values is as follows: values ≤0 indicate no agreement; 0.01–0.20 suggest none to slight agreement; 0.21–0.40 indicate fair agreement; 0.41–0.60 represent moderate agreement; 0.61–0.80 indicate substantial agreement; and 0.81–1.00 reflect almost perfect agreement.

RESULTS

Clinical Characteristics of Study Participants

Clinical and demographic characteristics of all study participants are summarized in Table 1. Of the 210 suspected cholera patients, the majority were adults and had severe dehydration. Prior to the hospital visit, 35% patients had taken antibiotics. Patients with moderate or severe dehydration were treated with recommended antibiotics along with rehydration with oral rehydration solution (ORS) or intravenous fluid as required.

Table 1.

Characteristics of the Patients Enrolled in the Study

Characteristics Values (%)
No. of male patients 119 (57)
Median age in year (25th and 75th centiles) 30 (22 and 38)
Children, ≤18 y of age 38 (18)
Adult patients 172 (82)
Patients with severe dehydration 125 (60)
Patients with moderate dehydration 72 (34)
Patients with no dehydration 13 (6)
Antibiotic taken before coming to the study hospital 73 (35)

Cholera RLDT Set Up and Training in Bangladesh

Rapid LAMP-based Diagnostic Test kits were shipped to the site. A team from Johns Hopkins University, visited the cholera reference laboratory, Dhaka, Bangladesh, to train the laboratory staff at icddr,b. In brief, a copy of the English test kit manual containing illustrations on the test procedure was provided along with a PowerPoint Presentation, followed by a hands-on training session on the first day. On the second day, the lab personnel performed the test under the observation of the trainers, and from the third day, they conducted the tests independently.

Comparison of Cholera RLDT Assay With Other Assays

Out of 205 suspected cholera cases analyzed, 136 (66%) were positive for cholera by RLDT, 131 (64%) by PCR, 92 (45%) by direct or enriched culture, and 90 (44%) by RDT (Figure 1). Among the culture positives, 11 (11.96%) were O1 Ogawa, and the rest were O1 Inaba. Since 5 of the samples gave discrepant results between the 2 different PCR assays used and we are uncertain of the correct PCR results, we excluded these samples from the analysis.

Figure 1.

This histogram shows the percentages of cholera-positive samples detected by 4 different diagnostic methods. The methods are Culture, RDT, PCR, and RLDT from left to right. The y-axis shows the percentage of positives from 0% to 100%. This figure indicates that RLDT detected cholera in 66%, PCR detected 64%, Culture detected 45%, and RDT detected 44% of suspected cases.

Histogram of the percentages of samples positive for cholera by each method of detection.

Similarities and Differences Between the Four Diagnostic Assays

Overall, 78 (60%) of the suspected cholera cases were positive by all 4 assays (Figure 2). A total of 92 stool samples were positive by culture, of which 89 were positive by direct culture, and an additional 3 samples were positive after enrichment in APW for overnight. Among these 92 culture positives, 90 were also positive by RLDT. Of the 2 samples that were negative, one was only positive for O1, and the other was negative for both genes by RLDT. Of the culture positives, 91 were positive by PCR, and only 81 were positive by the RDT. Rapid LAMP-based Diagnostic Test detected 46 (22.44%) and 48 (23.41%), and PCR detected 40 (19.51%) and 42 (20.49%) more cholera cases than culture and RDT, respectively. A total of 127 cases were positive by both the PCR and RLDT. Rapid LAMP-based Diagnostic Test detected an additional 8 cases, and PCR detected an additional 2 cases, which were negative by the other 3 assays. In RDT, 2 of the samples were positive for both O1 and O139.

Figure 2.

This Venn diagram shows the overlap of positive test results among 4 diagnostic assays: RLDT, RDT, PCR, and Culture. Seventy-eight (60%) cases were positive by all 4 methods. RLDT detected 46 (22.44%) and 48 (23.41%), and PCR detected 40 (19.51%) and 42 (20.49%) more cholera cases than culture and RDT, respectively. This diagram illustrates the higher sensitivity of RLDT and PCR compared to Culture and RDT.

A Venn diagram that illustrates the overlap of the positive test results from 4 diagnostic methods.

Pairwise Comparison Between the RLDT and Three Diagnostic Assays

The Cohen's kappa test revealed almost perfect agreement between RLDT and PCR (kappa 0.86). The agreement between RLDT and RDT and RLDT and culture was similar, with a near substantial agreement of kappa 0.55 and kappa 0.53, respectively (Figure 3).

Figure 3.

This figure shows the pairwise comparison between RLDT and the other 3 diagnostic methods, Culture, RDT, and PCR. Cohen's kappa values are annotated, showing that RLDT and PCR have the highest agreement, at 0.86. RLDT with Culture and RDT are showing moderate agreement, 0.55 and 0.53, respectively.

Pairwise comparison between RLDT, Culture, RDT, and PCR.

We also analyzed the agreement between assays among the patients who received antibiotics before coming to the hospital and those who did not (Figure 4A and 4B). The agreement between RLDT and PCR increased to 0.91 among those with prior antibiotics, while it decreased to 0.83 among those who did not. The agreement between RLDT and culture was decreased to kappa 0.49 among those who had antibiotics, while the agreement between RLDT and RDT remained unchanged between the 2 groups.

Figure 4.

Two bar charts (A and B) showing pairwise agreement between diagnostic assays differentiated by prior antibiotic use. (A) Includes patients who had taken antibiotics before hospital visit, showing the RLDT-PCR agreement as 0.91, while RLDT-Culture and RLDT-RDT agreements are 0.49 and 0.54. (B) Includes patients without prior antibiotic use, where RLDT-PCR agreement was 0.83, and RLDT-Culture remained higher than in the antibiotic group, at 0.57. RLDT-RDT agreement remained stable across both groups.

Pairwise comparison between the diagnostic assays among the patients based on prior antibiotic consumption. A, Illustrates the agreements between assays among patients who received antibiotics before seeking care at the icddr,b hospital, and (B) shows the agreements among patients who did not receive prior antibiotics.

Although none of the diagnostic assays used in this study could be considered the gold standard, and a comparison of RLDT with another assay as the gold standard may not be justified, we still analyzed the sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) of RLDT compared to other assays as the gold standard (Supplementary Table 1). Overall, the sensitivity of RLDT was over 90% across the comparison with all other assays, while the specificity widely varied (58.26%–97.73%) since RLDT and PCR are much more sensitive than culture and RDT. Considering ctxA as the target gene, compared to PCR, RLDT had a sensitivity of 96.95% and a specificity of 87.84%. Considering that when either O1 or ctxA is positive, the sample is positive for cholera, and the sensitivity and specificity of RLDT were 97.73% and 86.3%, respectively. The sensitivity of RLDT varied between 90.22% and 97.83% and 93.33% and 98.89% compared to culture and RDT, respectively.

Evaluation of RLDT in Water Samples

A total of 50 drinking water samples were tested. After overnight enrichment in APW, 5 (10%) samples were positive by both the RDT and culture. While 2 (4%) were positive by the direct RLDT without enrichment, all 5 were positive by the RLDT when tested after enrichment. Of the positive samples, 1 was O1 Inaba, and the rest were O1 Ogawa; the serotypes matched between the V. cholerae isolates from clinical and water samples.

DISCUSSION

This is the first evaluation of cholera RLDT in a low and middle-income (LMIC) country. This study demonstrated that lab staff members in a laboratory in LMICs can be trained in cholera RLDT, and the assay could be successfully implemented. We also showed that RLDT is highly sensitive and has substantial agreement with other assays.

As populations in LMICs increasingly coalesce in overcrowded mega-cities with poor sanitation, and global mobility accelerates coupled with more frequent extreme weather events, new and more virulent strains of V. cholerae are expected to spread more quickly and thus outbreaks are likely to rise in vulnerable regions [27–29]. This underscores the urgent need for stronger prevention measures and faster response strategies. Achieving this depends on having rapid, accurate diagnostic tools. Diagnosing cholera early at the onset of an outbreak, at the field level, should allow for a more timely response and mitigate the size, scope, and duration of the outbreak and subsequent spread of the illness. In this regard, as described before, culture and PCR are inadequate due to their complexity, being lengthy, requiring resources, and skilled personnel. Thus, timely diagnosis calls for RDTs. However, field and laboratory evaluations of the commercially available RDTs demonstrated variability in performance [11–18].

Rapid LAMP-based Diagnostic Test is a simple and rapid molecular test and has the potential to fill this void. In this study, the laboratory staff members were trained in RLDT in Bangladesh, where cholera is endemic, and the assay could be successfully implemented. We demonstrated that the sensitivity and specificity of RLDT were 97% and 88%, respectively, compared to PCR, with 90% agreement between these 2 assays. Thus, our study shows that while RLDT, in contrast to PCR, is simple, rapid (<1 hour), cold-chain, and almost electricity-free, requires minimum equipment, but on the other hand, has a similar sensitivity as PCR.

This study also demonstrated that RLDT was highly specific, with a sensitivity of 97% and 99% when compared to PCR and culture, respectively. In addition, RLDT could detect 24% more cholera than culture, of which 19% was confirmed by PCR. Like culture, the crystal VC RDT was also less sensitive than RLDT, and the latter detected 23% more cholera cases than the RDT, of which 20% were also positive by PCR. Other than early outbreak detection, the high sensitivity of the RLDT also has an advantage as a surveillance tool to estimate the real burden of cholera, as well as to estimate the efficacy of the oral cholera vaccine (OCV), since the shedding of V. cholerae in the stool sample could be reduced due to protection from the OCV.

Of note, in this study at the cholera reference laboratory, the RDT missed 11cholera cases, which were positive by the other 3 assays. As shown before, the accuracy of RDT compared to culture could be improved by enrichment in 1% APW for ∼6 hours; however, the sensitivity compared to PCR remained moderate [30, 31]. Notably, extending the RDT test incubation by 6 hours may impose an additional burden on the already strained health facilities at LMICs, as this may not align with the fixed working hours of health facility staff members.

As described before, the positivity of cholera by culture could be affected by prior antibiotic consumption [32]. Although RLDT is a molecular test, interestingly, the agreement between RLDT and PCR was higher among the prior antibiotic group, while, as expected, it was lower for the agreement between culture and RLDT. The agreement between RLDT and RDT was not affected by antibiotics, as reported before [32]. Antibiotic may have affected the PCR results, which needs to be further investigated.

Rapid LAMP-based Diagnostic Test could also detect cholera in drinking water, as found in this study; RLDT, after APW enrichment, was able to detect all 5 samples, which were positive by the enrichment RDT and culture. However, although RLDT showed an excellent limit of detection (LOD) in detecting cholera directly from spiked water samples in the laboratory [10], in this study, it failed to detect 3 of the 5 cholera-positive drinking water samples without enrichment.

Rapid LAMP-based Diagnostic Test detects both ctxA and O1rfb. In this evaluation, ctxA target was more sensitive than the O1rfb. In the future, only the ctxA target of the RLDT could be used to detect cholera, since ctxA alone could be confirmatory for cholera, and in case of the emergence of a new cholera serogroup (as happened with O139), the RLDT ctxA will still be able to detect the strain, which RDT will miss. Using only one target will reduce the cost, allowing more samples to be tested at a time. The RLDT ctxA positive samples could be tested later for O1 rfb to determine the serogroup if needed.

This study has limitations. In this study, for the convenience of the lab staff members, enrichment of water samples for overnight was used for the 3 assays. RLDT, however, had shown to have a LOD 10 colony forming unit (CFU) after 6 hours of enrichment (Chakraborty personal communication), which needs to be tested in the field in the future. Although the RLDT and PCR are more sensitive compared to culture, culture-based diagnostic methods remain crucial for downstream applications like antibiotic resistance monitoring, serotyping, and whole genome sequencing. To mitigate this challenge, since RLDT is rapid (<1 hour), the RLDT positive stool samples could be sent to the reference laboratory of the country for culture.

In conclusion, our results suggest that RLDT could be implemented in a reference laboratory in LMIC where cholera is endemic. Rapid LAMP-based Diagnostic Test was found to be highly specific and more sensitive compared to culture and RDT for the detection of cholera from stool. Rapid LAMP-based Diagnostic Test could also accurately detect cholera from drinking water samples. Rapid LAMP-based Diagnostic Test assures broader application as a culture-independent, simple, rapid, and cheap diagnostic test for cholera detection in resource-poor settings, where such a simple and field-adapted assay has a substantial impact on controlling cholera outbreaks or epidemic emergency situations. Testing of cholera RLDT in health care settings is underway.

Supplementary Material

ofag292_Supplementary_Data

Notes

Acknowledgment. We thank the laboratory staff members at icddr,b and at Johns Hopkins University for their contribution in the procurement of the RLDT kits and testing the samples in this study. We also thank the patients who participated in this study. We acknowledge the contribution of icddr,b's core donors, including the Government of Bangladesh, and Canada, for their continuous support and commitment to icddr,b's research efforts.

Data availability. All the data are included in the manuscript and Supplementary material. Any additional data can be obtained from the corresponding author upon reasonable request to irboffice@jhu.edu.

Author contributions. Conceptualization: S. C. and F. Q.; methodology: S. C., F. Q., A. I. K., K. I., and M. T. R. B.; investigation: K. I., A. I. K., M. T. R. B., F. H., Z. H. K., and M. T. I.; writing—original draft preparation: K. I. and S. C.; writing, review, and editing: K. I., A. I. K., M. T. R. B., F. H., Z. H. K., M. T. I., S. C., and F. Q.; funding acquisition: S. C. All authors have reviewed and agreed to the published version of the manuscript.

Financial support. This study was funded by the Wellcome Trust.

Contributor Information

Kamrul Islam, International Centre for Diarrhoeal Disease Research, Bangladesh (icddr,b), Dhaka, Bangladesh.

Ashraful Islam Khan, International Centre for Diarrhoeal Disease Research, Bangladesh (icddr,b), Dhaka, Bangladesh.

M D Taufiqur Rahman Bhuiyan, International Centre for Diarrhoeal Disease Research, Bangladesh (icddr,b), Dhaka, Bangladesh.

Fan He, Department of International Health, Johns Hopkins Bloomberg School of Public Health, Johns Hopkins University, Baltimore, Maryland, USA.

Zahid Hasan Khan, International Centre for Diarrhoeal Disease Research, Bangladesh (icddr,b), Dhaka, Bangladesh.

Md Taufiqul Islam, International Centre for Diarrhoeal Disease Research, Bangladesh (icddr,b), Dhaka, Bangladesh.

Firdausi Qadri, International Centre for Diarrhoeal Disease Research, Bangladesh (icddr,b), Dhaka, Bangladesh.

Subhra Chakraborty, Department of International Health, Johns Hopkins Bloomberg School of Public Health, Johns Hopkins University, Baltimore, Maryland, USA.

Supplementary Data

Supplementary materials are available at Open Forum Infectious Diseases online. Consisting of data provided by the authors to benefit the reader, the posted materials are not copyedited and are the sole responsibility of the authors, so questions or comments should be addressed to the corresponding author.

References

  • 1. Harris  JB, LaRocque  RC, Qadri  F, Ryan  ET, Calderwood  SB. Cholera. Lancet  2012; 379:2466–76. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2. Chowdhury  F, Ross  AG, Islam  MT, McMillan  NAJ, Qadri  F. Diagnosis, management, and future control of cholera. Clin Microbiol Rev  2022;35:e00211–21. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3. Shears  P. Recent developments in cholera. Curr Opin Infect Dis  2001; 14:553. [DOI] [PubMed] [Google Scholar]
  • 4. Harris  JR, Cavallaro  EC, De Nóbrega  AA, et al.  Field evaluation of Crystal VC® Rapid Dipstick test for cholera during a cholera outbreak in Guinea-Bissau. Trop Med Int Health  2009; 14:1117–21. [DOI] [PubMed] [Google Scholar]
  • 5. Nato  F, Boutonnier  A, Rajerison  M, et al.  One-step immunochromatographic dipstick tests for rapid detection of Vibrio cholerae O1 and O139 in stool samples. Clin Diagn Lab Immunol  2003; 10:476–8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6. Bopp  CA, Ries  AA, Wells  JG. Laboratory methods for the diagnosis of epidemic dysentery and cholera. 1999. Available at: https://stacks.cdc.gov. Accessed September 14, 2025.
  • 7. Telesmanich  NR, Lomov  IM. [Laboratory diagnosis of cholera: analysis and prospects for improvement]. Klin Lab Diagn  2009:51–5. [PubMed] [Google Scholar]
  • 8. Ramamurthy  T, Das  B, Chakraborty  S, Mukhopadhyay  AK, Sack  DA. Diagnostic techniques for rapid detection of Vibrio cholerae O1/O139. Vaccine  2020; 38:A73–82. [DOI] [PubMed] [Google Scholar]
  • 9. Gubala  AJ. Multiplex real-time PCR detection of Vibrio cholerae. J Microbiol Methods  2006; 65:278–93. [DOI] [PubMed] [Google Scholar]
  • 10. Chakraborty  S, Velagic  M, Connor  S. Development of a simple, rapid, and sensitive molecular diagnostic assay for cholera. PLoS Negl Trop Dis  2023; 17:e0011113. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11. Dick  MH, Guillerm  M, Moussy  F, Chaignat  CL. Review of two decades of cholera diagnostics – how far have we really Come?  PLoS Negl Trop Dis  2012; 6:e1845. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12. Kalluri  P, Naheed  A, Rahman  S, et al.  Evaluation of three rapid diagnostic tests for cholera: does the skill level of the technician matter?  Trop Med Int Health  2006; 11:49–55. [DOI] [PubMed] [Google Scholar]
  • 13. Mukherjee  P, Ghosh  S, Ramamurthy  T, et al.  Evaluation of a rapid immunochromatographic dipstick kit for diagnosis of cholera emphasizes its outbreak utility. Jpn J Infect Dis  2010; 63:234–8. [PubMed] [Google Scholar]
  • 14. Page  AL, Alberti  KP, Mondonge  V, Rauzier  J, Quilici  ML, Guerin  PJ. Evaluation of a rapid test for the diagnosis of cholera in the absence of a gold standard. PLoS One  2012; 7:e37360. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15. Ley  B, Khatib  AM, Thriemer  K, et al.  Evaluation of a rapid dipstick (crystal VC) for the diagnosis of cholera in zanzibar and a comparison with previous studies. PLoS One  2012; 7:e36930. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16. Sinha  A, SenGupta  S, Ghosh  S, et al.  Evaluation of a rapid dipstick test for identifying cholera cases during the outbreak. Indian J Med Res  2012; 135:523–8. [PMC free article] [PubMed] [Google Scholar]
  • 17. Boncy  J, Rossignol  E, Dahourou  G, et al.  Performance and utility of a rapid diagnostic test for cholera: notes from Haiti. Diagn Microbiol Infect Dis  2013; 76:521–3. [DOI] [PubMed] [Google Scholar]
  • 18. Muzembo  BA, Kitahara  K, Ohno  A, Debnath  A, Okamoto  K, Miyoshi  SI. Cholera rapid diagnostic tests for the detection of Vibrio cholerae O1: an updated meta-analysis. Diagnostics (Basel)  2021; 11:2095. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19. GTFCC . Target product profile for a rapid diagnostic test for surveillance of cholera. Available at: https://www.gtfcc.org/resources/target-product-profile-for-a-rapid-diagnostic-test-for-surveillance-of-cholera-outbreaks/. Accessed September 15, 2025.
  • 20. Barrett  TJ, Blake  PA, Morris  GK, Puhr  ND, Bradford  HB, Wells  JG. Use of Moore swabs for isolating Vibrio cholerae from sewage. J Clin Microbiol  1980; 11:385–8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21. Fields  PI, Popovic  T, Wachsmuth  K, Olsvik  O. Use of polymerase chain reaction for detection of toxigenic Vibrio cholerae O1 strains from the Latin American cholera epidemic. J Clin Microbiol  1992; 30:2118–21. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22. Alam  M, Hasan  NA, Sultana  M, et al.  Diagnostic limitations to accurate diagnosis of cholera. J Clin Microbiol  2010; 48:3918–22. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23. Chakraborty  S, Alam  M, Scobie  HM, Sack  DA. Adaptation of a simple dipstick test for detection of Vibrio cholerae O1 and O139 in environmental water. Front Microbiol  2013; 4:320. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24. Monir  MM, Islam  MT, Mazumder  R, et al.  Genomic attributes of Vibrio cholerae O1 responsible for 2022 massive cholera outbreak in Bangladesh. Nat Commun  2023; 14:1154. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25. Sayeed  MA, Islam  K, Hossain  M, et al.  Development of a new dipstick (Cholkit) for rapid detection of Vibrio cholerae O1 in acute watery diarrheal stools. PLoS Negl Trop Dis  2018; 12:e0006286. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26. Nandi  B, Nandy  RK, Mukhopadhyay  S, Nair  GB, Shimada  T, Ghose  AC. Rapid method for Species-specific identification of Vibrio cholerae using primers targeted to the gene of outer membrane protein OmpW. J Clin Microbiol  2000; 38:4145–51. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27. Zwane  AP, Kremer  M. What works in fighting diarrheal diseases in developing countries? A critical review. World Bank Res Obs  2007; 22:1–24. [Google Scholar]
  • 28. Cariri  FAMO, Costa  APR, Melo  CC, et al.  Characterization of potentially virulent non-O1/non-O139 Vibrio cholerae strains isolated from human patients. Clin Microbiol Infect  2010; 16:62–7. [DOI] [PubMed] [Google Scholar]
  • 29. Lipp  EK, Huq  A, Colwell  RR. Effects of global climate on infectious disease: the cholera model. Clin Microbiol Rev  2002; 15:757–70. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30. Ontweka  LN, Deng  LO, Rauzier  J, et al.  Cholera rapid test with enrichment step has diagnostic performance equivalent to culture. PLoS One  2016; 11:e0168257. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31. Bwire  G, Orach  CG, Abdallah  D, et al.  Alkaline peptone water enrichment with a dipstick test to quickly detect and monitor cholera outbreaks. BMC Infect Dis  2017; 17:726. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32. Perez-Saez  J, Bhuiyan  TR, Hegde  ST, et al.  Rethinking cholera diagnostic test performance, interpretation and evaluation: a field-based latent-class analysis in Bangladesh. Lancet Microbe  2025; 6:101170. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

ofag292_Supplementary_Data

Articles from Open Forum Infectious Diseases are provided here courtesy of Oxford University Press

RESOURCES