Abstract
Stereotactic body radiotherapy (SBRT) enables high-precision, high-dose treatment of liver lesions with acceptable toxicity. With the introduction of flattening filter-free (FFF) beams and the use of different photon energies, there is potential for improved dosimetry and treatment efficiency; however, the optimal selection of these parameters remains uncertain. This systematic review evaluated the impact of photon energy (6 MV versus 10 MV) and beam modality (flattened versus FFF) on dosimetric quality, organ-at-risk (OAR) sparing, treatment efficiency, and clinical feasibility in liver SBRT. A comprehensive search of PubMed, Embase, and Scopus databases from January 2010 to April 2025 identified dosimetric and clinical studies comparing FFF and flattened beams as well as different photon energies. Outcomes of interest included target coverage, conformity index (CI), homogeneity index (HI), beam-on time (BOT), monitor units (MUs), and OAR doses. Risk of bias was assessed using the Newcastle-Ottawa Scale.
A total of 23 studies met the inclusion criteria with adherence to Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) guidelines. FFF beams consistently reduced beam-on time by approximately 30-60%, improved dose conformity, and enhanced OAR sparing compared with flattened beams. Among FFF modalities, 6FFF demonstrated superior conformity and steeper dose gradients, particularly for small and centrally located lesions, whereas 10FFF provided improved skin sparing and was more suitable for larger or deeper tumors. Although FFF plans required higher MUs, this was offset by significantly faster treatment delivery. Fractionation strategies were influenced by tumor size, anatomical location, and baseline liver function, with a greater number of fractions generally preferred for larger or centrally located lesions.
Overall, FFF beams, particularly 6FFF and 10FFF, offer improved treatment efficiency and favorable dosimetric outcomes in liver SBRT. The choice of beam energy and modality should be individualized based on lesion characteristics, including size, depth, and proximity to critical structures. FFF-based SBRT represents a practical and widely implementable approach, although further prospective studies are needed to support standardization of clinical practice.
Keywords: dosimetry, flattened photon beams, flattening filter-free (fff) beams, linac, liver sbrt, photon energy, stereotactic body radiotherapy
Introduction and background
Stereotactic Body Radiotherapy (SBRT) has emerged as an advanced radiotherapeutic technique for the precise and high-dose treatment of various malignancies, including primary and metastatic liver tumors [1,2]. SBRT delivers ablative radiation doses over a limited number of fractions, using steep dose gradients to spare adjacent normal tissues [3-5]. This approach has shown promise for hepatocellular carcinoma (HCC) and liver metastases, especially in patients who are not surgical candidates or for whom other local therapies such as radiofrequency ablation (RFA) or transarterial chemoembolization (TACE) are unsuitable [6-9].
Unlike conventional radiotherapy, SBRT requires exceptional accuracy in target localization, dose delivery, and sparing of normal liver parenchyma due to the inherent radiosensitivity and regenerative capacity of the liver. The dosimetric quality of SBRT depends on several factors, including treatment planning algorithms, motion management strategies, fractionation regimens, and critically, the choice of beam modality and energy [10,11].
Traditionally, photon beams used in linear accelerators include a flattening filter (FF) to produce a uniform dose distribution across the field. However, the development of flattening filter-free (FFF) technology has revolutionized modern SBRT. FFF beams allow for higher dose rates, significantly reducing beam-on time (BOT), which in turn minimizes the risk of intra-fraction motion, an especially important consideration for liver SBRT, given respiratory-induced organ motion [12]. Moreover, FFF beams offer a sharper dose falloff and reduced peripheral dose, which may enhance target conformity and OAR sparing. However, their non-uniform beam profile and potential increase in monitor units (MUs) raise concerns regarding dose homogeneity and delivery efficiency [13].
In parallel, the choice of photon energy typically between 6 MV and 10 MV has implications for depth dose characteristics, skin sparing, and the magnitude of lateral scatter, all of which may impact both tumor control and toxicity. Lower energy beams (e.g., 6 MV) often exhibit sharper dose gradients and are preferred in small, central-seated targets, while higher energies (e.g., 10 MV) may be beneficial in reducing surface dose but pose a higher risk of neutron contamination and increased transmission through tissues [14].
Recent advances in treatment planning systems (TPS) and delivery platforms, such as Volumetric Modulated Arc Therapy (VMAT), Dynamic Conformal Arc (DCA), and Intensity-Modulated Radiotherapy (IMRT), have enabled precise execution of SBRT with either FF or FFF beams and variable photon energies [15]. However, there remains a lack of consensus in the clinical community regarding the optimal combination of beam type and energy, particularly in the setting of liver SBRT.
Several dosimetric studies and clinical reports have compared FFF vs. FF beams and 6 MV vs. 10 MV photon energy in various anatomical sites, with lung SBRT receiving the most attention [16]. Although SBRT has been successfully applied across several anatomical sites such as the lung, spine, and liver, each presents unique physical and biological challenges that demand site-specific optimization [17]. Lung SBRT focuses on managing respiratory motion and tissue heterogeneity at air-tissue interfaces, requiring robust image guidance, motion management, and dose conformity to avoid pulmonary toxicity [18]. Spine SBRT, in contrast, emphasizes steep dose gradients near the spinal cord to prevent radiation myelopathy, necessitating sub millimeter precision, rigid immobilization, and advanced image-guided verification [19]. Liver SBRT combines elements of both, requiring management of complex respiratory and cardiac motion, while ensuring sparing of radiosensitive hepatic tissue and nearby gastrointestinal organs. The liver’s limited regenerative capacity and vascular anatomy make normal-tissue protection particularly critical [20].
However, liver SBRT poses unique challenges due to the organ’s mobility, heterogeneous composition, and proximity to radiosensitive structures such as the stomach, bowel, and kidneys [6-9]. Thus, findings from other sites cannot be directly extrapolated to the liver.
Despite increasing use of FFF beams, no prior systematic review has specifically evaluated both photon energy selection and beam modality in liver SBRT. Hence, this systematic review aims to critically analyze the existing literature on SBRT for liver lesions, with a focus on the impact of photon energy selection and beam modality (FF vs. FFF) on treatment quality. Specifically, we evaluate outcomes related to target coverage, dose conformity, OAR sparing, beam-on time, and overall treatment efficiency, and provide clinical insights into the selection of the most appropriate beam configuration based on the evidence to date.
Review
Methods
Protocol and Registration
This systematic review was conducted in adherence with the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) guidelines [21]. The review protocol was prospectively registered in the International Prospective Register of Systematic Reviews (PROSPERO; CRD420251060403; https://www.crd.york.ac.uk/PROSPERO/view/CRD420251060403).
Objectives
The primary objective of this systematic review was to evaluate the dosimetric advantages of FFF photon beams compared with conventional FF photon beams in SBRT for liver lesions. Specifically, the analysis focused on key parameters including target coverage, OAR sparing, treatment delivery efficiency as measured by BOT, and the number of MUs required for treatment delivery. These parameters were selected to comprehensively assess both plan quality and treatment efficiency.
The secondary objective was to compare different photon energies, particularly 6 MV versus higher energies such as 10 MV and others, in terms of their impact on the same dosimetric and delivery-related outcomes. This comparison aimed to determine whether photon energy selection influences treatment quality, normal tissue protection, and delivery efficiency in liver SBRT.
Eligibility Criteria
Eligibility criteria were defined to include studies involving patients with liver tumors treated using photon-based SBRT, with explicit comparisons between FF and FFF beams or between different photon energies. Eligible studies were required to report relevant dosimetric or clinical outcomes, including planning target volume (PTV) coverage, conformity index (CI), homogeneity index (HI), OAR doses, BOT, and MUs. Only studies published in English between January 2010 and April 2025 were considered. Exclusion criteria included non-SBRT studies, non-English publications, conference abstracts, and review articles.
A comprehensive literature search was conducted across PubMed, Embase, and Scopus databases using combinations of keywords related to “SBRT,” “liver,” “photon energy,” and “flattening filter-free” (Appendix A). The search spanned from January 2010 to April 2025. Study selection was performed independently by two reviewers based on predefined inclusion and exclusion criteria, with disagreements resolved through consensus or consultation with a third reviewer. The selection process was documented using a PRISMA flow diagram.
Data Extraction and Analysis
Data extraction was carried out using a standardized form to ensure consistency. Extracted variables included study characteristics, treatment techniques, photon energy, beam type, dose calculation algorithms, and reported outcomes. Key dosimetric indices were either directly extracted or calculated where necessary. The CI was defined as the ratio of the volume receiving the reference dose to the target volume (VIR/TV), while the homogeneity index (HI) was calculated as the ratio of the maximum isodose within the target to the reference isodose (Imax/RI). The dose gradient index (GI) was defined as the ratio of the volume receiving the prescription isodose to the volume receiving half of the prescription isodose (DVPI/DVHPI). Additionally, D700cc was defined as the dose received by 700 cc of normal liver tissue.
The risk of bias in included studies was assessed using appropriate tools based on study design. The Cochrane Risk of Bias 2.0 tool was applied for randomized controlled trials [22], while the Newcastle-Ottawa Scale (NOS) was used for observational studies [23]. The risk of missing evidence was evaluated using the Risk Of Bias due to Missing Evidence (ROB-ME) framework [24]. Furthermore, the certainty of evidence for each outcome was assessed using the Grading of Recommendations Assessment, Development and Evaluation (GRADE) approach [25].
Given the anticipated clinical and methodological heterogeneity among the included studies, a narrative synthesis was primarily conducted. Where sufficient homogeneity existed, quantitative synthesis was performed using a random-effects model, and statistical heterogeneity was evaluated using the I² statistic.
Results
Study Selection and Characteristics
A total of 23 studies were included after screening 412 initial records (Figure 1).
Figure 1. PRISMA flow diagram of the study collection.
PRISMA: Preferred Reporting Items for Systematic Reviews and Meta-Analyses [21].
These comprised planning studies and clinical evaluations involving VMAT, IMRT, DCA, and 3 Dimensional Conformal Radiation Therapy (3DCRT) for liver SBRT. Both HCC and metastatic liver lesions were analyzed (Table 1).
Table 1. Primary study characteristics and results .
#: number of fractions or sessions over which radiotherapy will be delivered; SRS: Stereotactic Radiosurgery; PTV: Planning Target Volume; SBRT: Stereotactic Body Radiotherapy; DCA: Dynamic Conformal Arc; DCAT: Dynamic Conformal Arc Therapy; FFF: Flattening Filter-Free; FF: Flattening Filter; CI: Conformity Index; HI: Homogeneity Index; OAR: Organ-At-Risk; BOT: Beam-On Time; MU: Monitor Units; VMAT: Volumetric Modulated Arc Therapy; IMRT: Intensity Modulated Arc Therapy; IMPT: Intensity Modulated Proton Therapy; CK: Cyber Knife; Sparc: Spot Scanning Proton Arc; AAA: Anisotropic Analytical Algorithm; MC: Monte Carlo; CC: Collapsed Cone; RTOG: Radiation Therapy Oncology Group; MLC: Multi Leaf Collimator; AXB-Acuros XB.
| S. No | Study | Beam modalities | Treatment technique | Type of liver lesions | No of patients | Algorithm used for optimization | Total dose/fraction | Favorable beam modality as per the specified criteria |
| 1 | Kaur SP et al., 2025 [26] | 6FFF | VMAT | - | 11 | Not mentioned | Not mentioned | 6FFF (Better Dose Distribution and Fewer MU’s) |
| 2 | Bonù et al., 2024 [27] | 6FF, 6FFF, Proton | VMAT | Both | 213 | Not mentioned | Range: 50–60 Gy in 3–5 fractions | Proton (spares more healthy liver) VMAT FFF (spares more healthy liver) |
| 3 | Brekner et al., 2023 [28] | 6FF, 10FF, 6 FFF, 10FFF | VMAT | HCC | 17 | CC and MC | Not mentioned | 6FFF, 10FFF (remarkable reduction of the BOT) 10FFF (Improved GTV coverage) 6FFF, 10FFF (Higher D98% of PTV) 6FFF, 10FFF (D700cc of liver is less) 10FFF (Low skin dose) |
| 4 | Suresh et al., 2022 [29] | 6FF, 10FF, 6 FFF, 10FFF | 3DCRT, DCA, VMAT | Both | 20 | AAA | 40Gy/5# | VMAT FFF (Better PTV Coverage and OAR Dose) VMAT FFF (Highly Conformal Plans) VMAT FFF (Faster Treatment delivery) |
| 5 | Park et al., 2020 [30] | 10 FFF | VMAT | HCC | 290 | Not mentioned | Range: 30–60 Gy in 3–4 fractions | SBRT 10FFF (High local control rates in patients) |
| 6 | Moon et al., 2020 [31] | 6FF | DCAT, VMAT | Both | 25 | MC | Range: 48–60 Gy in 4–5 fractions | DCAT (shorter calculation times and delivery times with smaller MU) DCAT (should be considered preferential to VMAT in patients with poor breathing or poor coordination) VMAT (Better OAR dose) VMAT (Considered when multiple lesions) |
| 7 | Thaper et al., 2020 [32] | 10FF | DCA_SSO_ VDR, VMAT | - | 25 | MC | Range: 25–50 Gy in 5-15 fractions | VMAT (Less OAR dose) DCA_SSO_VDR (Better CI and HI Value) DCA_SSO_VDR (Less MUs) |
| 8 | Stathakis et al., 2019 [33] | 6FFF | DCAT, VMAT | - | 4 | MC | Range: 45–55 Gy in 3-9 fractions | DCAT (shorter calculation times and delivery times with smaller MU) DCAT (Showed comparable plan quality compared to the VMAT plans) DCAT (dosimetrically equivalent and an efficient alternative to VMAT plans) |
| 9 | Munirathinam et al., 2018 [34] | 6FF, 6FFF | VMAT | Mets | 8 | Not mentioned | 50Gy/10# | 6FFF (BOT Reduction) 6FFF (Better dose distribution) |
| 10 | Ogata et al., 2017 [35] | 10FFF, 10FF | 3DCRT | Both | 10 | AAA | 48Gy/4# | FFF and FFF (Similar target coverage and sparing of normal tissue) FFF (Better HI value) |
| 11 | Vieillevigne et al., 2016 [36] | 6FF, 6FFF, 10FF, 10FFF | DCA, VMAT | - | Phantom | AAA | 60Gy/3# | VMAT FFF (Better sparing of normal tissue) VMAT FFF (BOT Reduction) |
| 12 | Kim GJ et al., 2016 [37] | 10FF, 10FFF | VMAT | - | 2 | Not mentioned | 45Gy/3# | 10 FFF (BOT Reduction) |
| 13 | Jin et al., 2016 [38] | 6FFF | IMRT, VMAT | HCC | 9 | Not mentioned | 50 or 32.5/5# | Cyberknife (Lower Mean dose for critical structures and better CI) Varian Linac (Less treatment delivery time) |
| 14 | Kim JY et al., 2015 [39] | Proton, Photon | PBT, H- IMRT, VMAT | HCC | 30 | Not mentioned | 60Gy/10# | Proton (spares more healthy liver) |
| 15 | Stieb et al., 2015 [40] | 6FFF or 10FFF | VMAT | Both | 10 | Not mentioned | Range: 35–60 Gy in 4–10 fractions | FFF (No unexpected toxicity occurred) FFF (time efficient and appears to be safe) |
| 16 | Esposito et al., 2015 [41] | Multiple techniques, Beam energies | 3DCRT, IMRT, VMAT | Mets | 5 | Not mentioned | 75Gy/3# | VMAT FFF (BOT Reduction) |
| 17 | Gandhi et al., 2015 [42] | Photon, Proton | VMAT, PBT | HCC | 10 | Not mentioned | 50Gy/5# | Proton (Better Normal Liver Sparing) Proton (Less radiation toxicity) |
| 18 | Mancosu et al., 2012 [43] | FF, FFF | VMAT | Both | 55 | AAA (Analytical Anisotropic Algorithm) | 75Gy/3# | FFF (BOT Reduction) |
| 19 | Prendergast et al., 2012 [44] | FF, FFF | VMAT, IMRT | Both | 99 | Not mentioned | Range: 6–20 Gy in 3–5 fractions | FFF (Improved Treatment Delivery Time) FFF (BOT Reduction) |
| 20 | Reggiori et al., 2012 [45] | 10FF, 10FFF | VMAT | Mets | 10 | Not mentioned | 75Gy/3# | 10 FFF (BOT Reduction) |
| 21 | Lang et al., 2012 [46] | 6FFF, 10FFF | VMAT | Both | Total 26 and 4 for liver | AAA | Range: 40–50 Gy in 4–12 fractions | VMAT FFF (BOT Reduction) VMAT FFF (Minimal patient and target shift during treatment) |
| 22 | Scorsetti et al., 2011 [47] | 6FFF or 10FFF | VMAT | Mets | 10 | AAA | 75Gy/3# | FFF (BOT Reduction) FFF (Promising tumor local control) |
| 23 | Petersen et al., 2011 [48] | Photon, Proton | IMPT, IMRT | Mets | 10 | AAA, PB | Range: 37.5– 50.25 Gy in 3 fractions | IMPT (Better treatment plans and dose coverage less OAR dose) |
In view of heterogeneity of studies, a narrative synthesis was undertaken. The included studies demonstrated an overall moderate risk of bias, primarily driven by non-randomized study designs, small sample sizes in several analyses, and substantial confounding due to technological and patient-related heterogeneity (Appendix B).The summary table for key outcomes such as BOT reduction, CI/HI values is given as Appendix C.
Beam Modalities: FF vs. FFF Beams
FFF beams consistently reduced BOT (by 30-60%) and improved conformity and OAR sparing compared to FF beams. Improvements in GTV coverage, liver sparing, and reduced skin dose were commonly reported. FFF beams also enabled better utilization of high dose per fraction regimens. FFF delivery allowed for faster treatment sessions, which is clinically advantageous for minimizing intra-fraction motion and improving patient throughput.
FFF beams provided equal or superior target coverage in terms of D95%, D98%, and PTV coverage. Liver dose constraints (e.g., D700cc, mean liver dose) were better achieved using FFF beams in many studies. FFF delivery reduced the dose to healthy liver tissue, bowel, stomach, and skin in most cases and improved spatial and temporal accuracy, reducing patient movement and positional drift during long treatments.
Photon Energy Comparison (6 MV vs. 10 MV)
6 MV FFF: Several studies favored 6FFF for its sharper penumbra, better conformity, and enhanced sparing of critical structures, particularly in small-to-medium-sized lesions, and found that 6FFF reduced D700cc of normal liver and provided improved dose gradients around PTV.
10 MV FFF: 10FFF was advantageous in specific contexts, particularly for reducing skin dose and for larger or deep-seated lesions, where improved beam penetration was needed. 10FFF provided superior GTV coverage and reduced skin dose, especially when used with VMAT.
Mixed beam use: A few studies, utilized both 6FFF and 10FFF in the same cohort. Their findings suggested that the choice of energy should be tailored based on PTV size, depth, and proximity to sensitive structures. Larger lesions may benefit from the deeper penetration of 10FFF, while smaller or superficial lesions may be best treated with 6FFF.
Fractionation Schedules and Interaction With Beam Type
Among the included studies, fractionation schedules for liver SBRT ranged from single-fraction (25-30 Gy) to multi-fraction regimens of three to five fractions (total 30-60 Gy). Most dosimetric comparisons involving FFF and FF beams used three- to five-fraction protocols, consistent with institutional liver SBRT practice. Some of the studies reported that the advantage of FFF beams, particularly reducing BOT by up to 60%, was most evident in single- or few-fraction schedules, where intrafraction motion is a greater concern. In contrast, for five-fraction regimens, FFF and FF beams produced similar CIs and HIs, though treatment efficiency still improved. These findings indicate that the dosimetric benefit of FFF delivery is amplified with higher per-fraction doses, aligning with the motion-sensitive nature of liver SBRT.
Discussion
This systematic review comprehensively evaluated the impact of photon beam characteristics specifically FFF versus flattened FF beams and varying photon energies on the quality and efficiency of SBRT for liver lesions. Our findings indicate a consistent dosimetric advantage of using FFF beams, particularly 6FFF and 10FFF, across a broad range of clinical and planning contexts.
Clinical Relevance of FFF Beams in Liver SBRT
The FFF beams affirms that BOT reductions and shorter treatment durations decrease the likelihood of intra-fraction motion, which in turn improves geometric accuracy and target localization. Moreover, the potential of FFF beams to improve plan conformity and steeper dose gradients is well-supported. A more conformal dose distribution is especially important in the liver due to the organ’s proximity to radiosensitive structures like the stomach, bowel, kidneys, and spinal cord. Many studies included in our review, particularly those using VMAT-based FFF plans, reported improved CIs and lower doses to OARs, thereby enhancing the therapeutic ratio [26,27,35,49].
Photon Energy Considerations: 6FFF vs. 10FFF
Across included studies, lesion size and depth definitions were heterogeneous. However, planning data generally classified small and superficial lesions as those with PTV <100 cm³ or located within 5 cm from the liver surface, and large or deep lesions as those with PTV >100-150 cm³ or situated >5 cm below the capsule. These thresholds, although not universal, provide practical guidance for energy selection in clinical planning.
A key objective of this review was to investigate whether the choice of photon energy (typically 6 MV or 10 MV) influences dosimetric quality in liver SBRT. The results suggest that 6FFF beams are generally preferable for small, centrally located, or shallow lesions. This is primarily due to their sharper penumbra and enhanced dose conformity, which facilitate more precise dose sculpting and improved sparing of healthy liver tissue [33,46,48].
10FFF beams offer benefits in select situations, particularly when deeper tissue penetration is needed, such as in large or deeply located lesions. They also demonstrate improved skin sparing due to the beam's higher energy. However, these benefits are sometimes offset by a broader penumbra and reduced conformity compared to 6FFF. Hence, the choice between 6FFF and 10FFF should be individualized based on tumor characteristics, anatomical location, and institutional planning capabilities [42,50].
Dosimetric Trade-offs and Planning Complexity
While FFF beams were occasionally associated with slightly higher MUs, this increase was not clinically concerning and was often offset by the significantly reduced BOT. Although FFF beams offer clear advantages in terms of reduced BOT, improved treatment efficiency, and sharper dose gradients, certain trade-offs should be acknowledged. Several studies reported a modest increase in MUs with FFF delivery, particularly for large or complex target geometries. This increase, however, did not translate into clinically significant differences in treatment time or patient dose due to the higher dose rate of FFF beams [37,43]. Additionally, dose homogeneity may be slightly reduced in large or asymmetric fields because of the non-uniform intensity profile inherent to FFF beams [13]. Nevertheless, modern planning algorithms (Analytical Anisotropic Algorithm (AAA), Monte Carlo) and VMAT optimization effectively mitigate these effects, maintaining acceptable HIs and CIs [51]. Therefore, while FFF beams are not universally superior in all scenarios, their clinical efficiency and motion-management advantages outweigh these minor limitations in most liver SBRT contexts. Importantly, treatment techniques such as VMAT and IMRT can exploit the high dose rates of FFF beams without compromising dosimetric quality. Additionally, several studies reported that FFF beams offered reduced peripheral dose and improved sparing of non-target liver volume (e.g., D700cc), a parameter increasingly recognized for predicting hepatic toxicity in SBRT. This makes FFF particularly valuable in patients with pre-existing liver dysfunction or cirrhosis common comorbidities in hepatocellular carcinoma [31,43,45,52].
Fractionation Schedules
Across the included studies, fractionation schedules ranged from single-fraction high-dose treatments (e.g., 25-30 Gy × 1) to more conventional regimens of three to five fractions totalling 30-60 Gy. The benefit of FFF beams was most pronounced in single- and few-fraction regimens, where reduced BOT directly minimized intra-fraction motion and improved treatment efficiency. For multi-fraction schedules (three to five fractions), FFF delivery continued to enhance workflow and patient comfort but contributed less to dosimetric differences compared with flattened beams. Some of the included studies specifically noted greater BOT reduction and improved conformity with 6FFF in hypo fractionated regimens [28,43,53]. These findings suggest that the clinical advantage of FFF delivery is fractionation-dependent, being most impactful when per-fraction doses and intrafraction motion risks are highest.
Heterogeneity and Limitations
There was notable heterogeneity across the included studies in terms of planning systems and optimization algorithms, fractionation regimens (ranging from one to five fractions), target sizes and anatomical locations, and respiratory motion management strategies. Although a meta-analysis was not feasible due to this variability, the consistent trends observed across multiple studies strengthen the validity of our conclusions.
A key limitation of the current review is that most included studies focused exclusively on dosimetric parameters, with limited or absent reporting of clinical outcomes such as local control, hepatic toxicity, or overall survival. Consequently, while the dosimetric advantages of FFF beams are evident, their translation into tangible clinical benefit remains to be validated. Future prospective trials should integrate both dosimetric and clinical endpoints to determine whether improved conformity, lower organ doses, and shorter BOTs correlate with enhanced tumor control and reduced hepatic or gastrointestinal toxicity in real-world settings. Publication bias could not be formally assessed due to the heterogeneity and predominance of dosimetric studies; however, the possibility of selective reporting of favorable outcomes cannot be excluded. Comparison across different fractionation regimens introduces radiobiological variability, which may influence OAR tolerance and limits direct dosimetric comparability.
Future Directions
Prospective studies with clinical endpoints, cost-effectiveness analyses, and individualized beam selection strategies are warranted to optimize liver SBRT outcomes.
Conclusions
This review shows that FFF photon beams, particularly 6FFF and 10FFF, offer notable dosimetric and treatment efficiency benefits for liver SBRT. They reduce treatment time, improve conformity, and enhance OAR sparing. 6FFF is advantageous for small, central lesions, while 10FFF suits larger or deep-seated tumors. The choice of beam energy should be patient- and lesion-specific. In summary, the findings of this review support the integration of FFF beam technology, particularly 6FFF and 10FFF, into routine SBRT planning for liver tumors. The selection of beam energy and configuration should be individualized, guided by tumor size, depth, anatomical proximity to critical organs, and available technology. Future prospective trials reporting both dosimetric and clinical outcomes are essential to validate these observations and develop consensus-based clinical guidelines.
Appendices
Appendix A
Figure 2. Search strategy with MeSH terms.
Appendix B
Table 2. Risk of bias assessment for the included studies.
| S. No | Study | Design | Sample size | Selection bias | Performance bias | Detection bias | Confounding | Key limitations | Overall RoB |
| 1 | Kaur et al., 2025 [26] | Retrospective planning | 33 | Moderate | Low | Low | Moderate | Mixed sites, retrospective design | Moderate |
| 2 | Bonù et al., 2024 [27] | Multicenter retrospective | 213 lesions | Moderate | Moderate | Low | High | Technique selection bias | Moderate |
| 3 | Brekner et al., 2023 [28] | Retrospective planning | 44 | Moderate | Low | Low | Moderate | Lesion heterogeneity | Moderate |
| 4 | Suresh et al., 2022 [29] | Planning study | 20 | Moderate | Low | Low | Moderate | Multiple techniques compared | Moderate |
| 5 | Park et al., 2020 [30] | Retrospective clinical | 290 | Moderate | Moderate | Low | High | Clinical heterogeneity | Moderate–High |
| 6 | Moon et al., 2020 [31] | Retrospective planning | 25 | Moderate | Low | Low | Moderate | Planner variability | Moderate |
| 7 | Thaper et al., 2020 [32] | Retrospective planning | 25 | Moderate | Moderate | Low | Moderate | Technique heterogeneity | Moderate |
| 8 | Stathakis et al., 2019 [33] | Planning study | 4 | Moderate | Low | Low | Low–Moderate | Very small sample | Moderate |
| 9 | Munirathinam et al., 2018 [34] | Planning study | 8 | High | Low | Low | Moderate | Very small sample | Moderate–High |
| 10 | Ogata et al., 2017 [35] | Planning study | 10 | Moderate | Low | Low | Low–Moderate | Small sample | Moderate |
| 11 | Vieillevigne et al., 2016 [36] | Planning (phantom + clinical) | 24 PTVs | Moderate | Low | Low | Moderate | Mixed datasets | Moderate |
| 12 | Kim et al., 2016 [37] | Planning study | 5 | High | Low | Low | Moderate | Extremely small sample | High |
| 13 | Jin et al., 2016 [38] | Planning study | 9 | Moderate | Moderate | Low | High | Platform comparison bias | Moderate–High |
| 14 | Kim et al., 2015 [39] | Planning study | 30 | Moderate | Low | Low | High | Proton vs photon bias | Moderate |
| 15 | Stieb et al., 2015 [40] | Retrospective clinical | 84 | Moderate | Moderate | Low | High | Mixed tumor sites, short FU | Moderate |
| 16 | Esposito et al., 2015 [41] | Multicenter planning | 5 | High | High | Low | High | Inter-institution variability | High |
| 17 | Gandhi et al., 2015 [42] | Model + validation | 10 | Moderate | Moderate | Low | High | Modeling assumptions | Moderate–High |
| 18 | Mancosu et al., 2012 [43] | Prospective cohort | 55 | Low–Moderate | Low | Low | Moderate | No comparator arm | Moderate |
| 19 | Prendergast et al., 2012 [44] | Retrospective cohort | 99 | Moderate | Moderate | Low | High | Temporal bias | Moderate–High |
| 20 | Reggiori et al., 2012 [45] | Planning study | 10 | Moderate | Low | Low | Moderate | Small cohort | Moderate |
| 21 | Lang et al., 2012 [46] | Clinical + planning | 26 | Moderate | Low | Low | Moderate | Mixed tumor types | Moderate |
| 22 | Scorsetti et al., 2011 [47] | Prospective cohort | 10 | Low–Moderate | Low | Low | Moderate | No comparator arm | Moderate |
| 23 | Petersen et al., 2011 [48] | Planning study | 10 | High | Low | Low | High | Proton vs photon bias | High |
Appendix C
Table 3. Summary table of BOT reduction, CI, and HI .
BOT: beam-on time; CI: conformity index; HI: homogeneity index; FF: flattening filter; FFF: flattening filter free.
| Study | BOT Reduction / BOT Data | Conformity Index (CI) | Homogeneity Index (HI) |
| Kaur et al., 2025 [26] | Not reported | Not reported | Not reported |
| Bonù et al., 2024 [27] | Not reported | 0.74–0.8 | Not reported |
| Brekner et al., 2023 [28] | 78.4% BOT reduction (FFF) | 1.18 | Minor improvement |
| Suresh et al., 2022 [29] | ≈ 75% BOT reduction (FFF) | 1.19 | 0.08 ± 0.03 |
| Park et al., 2020 [30] | No BOT/CI/HI data | — | — |
| Moon et al., 2020 [31] | ~20% reduction (DCAT faster) | VMAT 1.03, DCAT 1.05 | VMAT 1.07, DCAT 1.08 |
| Thaper et al., 2020 [32] | ≈ 70% reduction (DCA SSO VDR faster) | VMAT ≈ 1.05, DCA SSO VDR ≈ 1.10 | — |
| Stathakis et al., 2019 [33] | ≈ 59–60% reduction (DCAT faster) | VMAT ≈ 1.33, DCAT ≈ 1.42 | — |
| Munirathinam et al., 2018 [34] | ~30–40% reduction (FFF faster) | FFF 0.91, FB 0.79 | ≈ 1.06 |
| Ogata et al., 2017 [35] | 70–75% reduction (FFF faster) | FFF ≈ 1.05, FF ≈ 1.02 | — |
| Vieillevigne et al., 2016 [36] | 55–75% reduction (FFF faster) | ≈ 1.0–1.2 | Nearly identical |
| Kim et al., 2016 [37] | 50–70% reduction (FFF faster) | FFF CI 0.95, FF CI 0.94 | FFF HI 1.12, FF HI 1.11 |
| Jin et al., 2016 [38] | 50% reduction (VMAT FFF) | CK CI 1.07, Varian Linac 1.19 | Nearly identical |
| Kim et al., 2015 [39] | Not quantified | ≈ 1.45 | ≈ 1.03 |
| Stieb et al., 2015 [40] | 40–50% reduction (FFF faster) | FFF CI 1.15, FF CI 1.13 | FFF HI 1.38, FF HI 1.3 |
| Esposito et al., 2015 [41] | 2.4–57 min → up to 90% reduction (VMAT FFF) | 1.10–1.25 | 1.07–1.53 (mean ≈ 1.16 ± 0.27) |
| Gandhi et al., 2015 [42] | Not reported | — | — |
| Mancosu et al., 2012 [43] | 45% reduction (FFF faster) | FFF CI 1.14, FF CI 1.15 | FFF HI 1.1, FF HI 1.09 |
| Prendergast et al., 2012 [44] | Not reported | — | — |
| Reggiori et al., 2012 [45] | 70% reduction (FFF faster) | FFF CI 1.19, FF CI 1.04 | Nearly identical |
| Lang et al., 2012 [46] | ≈ 70% reduction (FFF faster) | 1.10 | Not reported |
| Scorsetti et al., 2011 [47] | ≈ 70% reduction (FFF faster) | 1.20 | Not reported |
| Petersen et al., 2011 [48] | Not reported | — | — |
Disclosures
Conflicts of interest: In compliance with the ICMJE uniform disclosure form, all authors declare the following:
Payment/services info: All authors have declared that no financial support was received from any organization for the submitted work.
Financial relationships: All authors have declared that they have no financial relationships at present or within the previous three years with any organizations that might have an interest in the submitted work.
Other relationships: All authors have declared that there are no other relationships or activities that could appear to have influenced the submitted work.
Author Contributions
Concept and design: Ajay Kumar Kondeti, Kesava Ramgopal Adavikolanu, Chandramouli R, Soorej Balan Kaliyath, Thirumal M, Abhishek R. K S, NagaSai Divya Kari, Pragnahitha Neerudu, Ambedkar Yadala, Madhuri Palla
Acquisition, analysis, or interpretation of data: Ajay Kumar Kondeti, Kesava Ramgopal Adavikolanu, Chandramouli R, Soorej Balan Kaliyath, Thirumal M, Abhishek R. K S, NagaSai Divya Kari, Pragnahitha Neerudu, Ambedkar Yadala, Madhuri Palla
Drafting of the manuscript: Ajay Kumar Kondeti, Kesava Ramgopal Adavikolanu, Chandramouli R, Soorej Balan Kaliyath, Thirumal M, Abhishek R. K S, NagaSai Divya Kari, Pragnahitha Neerudu, Ambedkar Yadala, Madhuri Palla
Critical review of the manuscript for important intellectual content: Ajay Kumar Kondeti, Kesava Ramgopal Adavikolanu, Chandramouli R, Soorej Balan Kaliyath, Thirumal M, Abhishek R. K S, NagaSai Divya Kari, Pragnahitha Neerudu, Ambedkar Yadala, Madhuri Palla
Supervision: Ajay Kumar Kondeti, Kesava Ramgopal Adavikolanu
References
- 1.Stereotactic body radiation therapy in multiple organ sites. Timmerman RD, Kavanagh BD, Cho LC, Papiez L, Xing L. J Clin Oncol. 2007;25:947–952. doi: 10.1200/JCO.2006.09.7469. [DOI] [PubMed] [Google Scholar]
- 2.Stereotactic radiotherapy of malignancies in the abdomen: methodological aspects. Lax I, Blomgren H, Näslund I, Svanström R. Acta Oncol. 1994;33:677–683. doi: 10.3109/02841869409121782. [DOI] [PubMed] [Google Scholar]
- 3.Stereotactic body radiation therapy: the report of AAPM Task Group 101. Benedict SH, Yenice KM, Followill D, et al. Med Phys. 2010;37:4078–4101. doi: 10.1118/1.3438081. [DOI] [PubMed] [Google Scholar]
- 4.American Society for Therapeutic Radiology and Oncology (ASTRO) and American College of Radiology (ACR) practice guideline for the performance of stereotactic body radiation therapy. Potters L, Kavanagh B, Galvin JM, et al. Int J Radiat Oncol Biol Phys. 2010;76:326–332. doi: 10.1016/j.ijrobp.2009.09.042. [DOI] [PubMed] [Google Scholar]
- 5.Definition and quality requirements for stereotactic radiotherapy: consensus statement from the DEGRO/DGMP Working Group Stereotactic Radiotherapy and Radiosurgery. Guckenberger M, Baus WW, Blanck O, et al. Strahlenther Onkol. 2020;196:417–420. doi: 10.1007/s00066-020-01603-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.The challenge of inoperable hepatocellular carcinoma (HCC): results of a single-institutional experience on stereotactic body radiation therapy (SBRT) Scorsetti M, Comito T, Cozzi L, et al. J Cancer Res Clin Oncol. 2015;141:1301–1309. doi: 10.1007/s00432-015-1929-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Sequential phase I and II trials of stereotactic body radiotherapy for locally advanced hepatocellular carcinoma. Bujold A, Massey CA, Kim JJ, et al. J Clin Oncol. 2013;31:1631–1639. doi: 10.1200/JCO.2012.44.1659. [DOI] [PubMed] [Google Scholar]
- 8.Multi-institutional phase I/II trial of stereotactic body radiation therapy for liver metastases. Rusthoven KE, Kavanagh BD, Cardenes H, et al. J Clin Oncol. 2009;27:1572–1578. doi: 10.1200/JCO.2008.19.6329. [DOI] [PubMed] [Google Scholar]
- 9.Stereotactic body radiation therapy for liver metastases. Scorsetti M, Clerici E, Comito T. J Gastrointest Oncol. 2014;5:190–197. doi: 10.3978/j.issn.2078-6891.2014.039. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Review and uses of stereotactic body radiation therapy for oligometastases. Alongi F, Arcangeli S, Filippi AR, Ricardi U, Scorsetti M. Oncologist. 2012;17:1100–1107. doi: 10.1634/theoncologist.2012-0092. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Image-guided stereotactic body radiotherapy (SBRT) with enhanced visualization of tumor and hepatic parenchyma in patients with primary and metastatic liver malignancies. Kirichenko AV, Lee D, Wagner P, et al. Cancers (Basel) 2025;17:1088. doi: 10.3390/cancers17071088. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Flattening-filter-free intensity modulated breath-hold image-guided SABR (Stereotactic ABlative Radiotherapy) can be applied in a 15-min treatment slot. Boda-Heggemann J, Mai S, Fleckenstein J, et al. Radiother Oncol. 2013;109:505–509. doi: 10.1016/j.radonc.2013.09.014. [DOI] [PubMed] [Google Scholar]
- 13.Flattening filter-free accelerators: a report from the AAPM Therapy Emerging Technology Assessment Work Group. Xiao Y, Kry SF, Popple R, et al. J Appl Clin Med Phys. 2015;16:5219. doi: 10.1120/jacmp.v16i3.5219. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.The dosimetric significance of using 10 MV photons for volumetric modulated arc therapy for post-prostatectomy irradiation of the prostate bed. Kleiner H, Podgorsak MB. Radiol Oncol. 2016;50:232–237. doi: 10.1515/raon-2016-0012. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Comparison of volumetric-modulated arc therapy and dynamic conformal arc treatment planning for cranial stereotactic radiosurgery. Molinier J, Kerr C, Simeon S, Ailleres N, Charissoux M, Azria D, Fenoglietto P. J Appl Clin Med Phys. 2016;17:92–101. doi: 10.1120/jacmp.v17i1.5677. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Stereotactic body radiotherapy for early-stage lung cancer: a systematic review on the choice of photon energy and linac flattened/unflattened beams. Gill A, Hirst AL, Rowshanfarzad P, Gill S, Bucknell N, Dass J, Sabet M. Radiat Oncol. 2024;19:1. doi: 10.1186/s13014-023-02392-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Stereotactic body radiation therapy for lung, spine and oligometastatic disease: current evidence and future directions. Dunne EM, Fraser IM, Liu M. Ann Transl Med. 2018;6:283. doi: 10.21037/atm.2018.06.40. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Application of respiratory motion management technology for patients with lung cancer treated with stereotactic body radiotherapy (Review) Shao K, Chen W, Xu Y, Yang S. Oncol Lett. 2025;30:415. doi: 10.3892/ol.2025.15161. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.The role of CT and MR imaging in stereotactic body radiotherapy of the spine: from patient selection and treatment planning to post-treatment monitoring. Azadbakht J, Condos A, Haynor D, et al. Cancers (Basel) 2024;16:3692. doi: 10.3390/cancers16213692. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Liver metastases and SBRT: a new paradigm? Comito T, Clerici E, Tozzi A, D'Agostino G. Rep Pract Oncol Radiother. 2015;20:464–471. doi: 10.1016/j.rpor.2014.10.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Preferred reporting items for systematic reviews and meta-analyses: the PRISMA statement. Moher D, Liberati A, Tetzlaff J, Altman DG. PLoS Med. 2009;6:0. [PMC free article] [PubMed] [Google Scholar]
- 22.RoB 2: a revised tool for assessing risk of bias in randomised trials. Sterne JAC, Savović J, Page MJ, et al. BMJ. 2019;28:4898. doi: 10.1136/bmj.l4898. [DOI] [PubMed] [Google Scholar]
- 23.Wells G, Shea B, O’Connell D, Peterson J, Welch V, Losos M. The Newcastle-Ottawa Scale (NOS) for assessing the quality of nonrandomized studies in meta-analyses. [ Feb; 2026 ]. 2000. http://www.ohri.ca/programs/clinical_epidemiology/oxford.asp http://www.ohri.ca/programs/clinical_epidemiology/oxford.asp
- 24.ROB-ME: a tool for assessing risk of bias due to missing evidence in systematic reviews with meta-analysis. Page MJ, Sterne JA, Boutron I, et al. BMJ. 2023;383:0. doi: 10.1136/bmj-2023-076754. [DOI] [PubMed] [Google Scholar]
- 25.What is "quality of evidence" and why is it important to clinicians? Guyatt GH, Oxman AD, Kunz R, Vist GE, Falck-Ytter Y, Schünemann HJ. BMJ. 2008;336:995–998. doi: 10.1136/bmj.39490.551019.BE. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Effect of minimum segment width on SRT/SBRT volumetric arc therapy plans for flattening filter free beams. Kaur SP, Shivanjli Shivanjli, Oinam AS, et al. http://dx.doi.org/10.31557/apjcb.2025.10.2.309-313 Asian Pac J Canc Biol. 2025;10:309–313. [Google Scholar]
- 27.High dose proton and photon-based radiation therapy for 213 liver lesions: a multi-institutional dosimetric comparison with a clinical perspective. Bonù ML, Nicosia L, Turkaj A, et al. Radiol Med. 2024;129:497–506. doi: 10.1007/s11547-024-01788-w. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Stereotactic body radiotherapy with volumetric intensity-modulated arc therapy and flattening filter-free beams: dosimetric considerations. Brekner MC, Imhoff D, Rödel C, et al. Strahlenther Onkol. 2024;200:346–357. doi: 10.1007/s00066-023-02181-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.The dosimetric comparison of different SBRT techniques for treatment of liver cancer using flattened and flattening filter free beam. Suresh T, Madeswaran S. Iran J Med Phys. 2022;19:371–382. [Google Scholar]
- 30.Clinical outcomes of stereotactic body radiation therapy for small hepatocellular carcinoma. Park S, Jung J, Cho B, et al. J Gastroenterol Hepatol. 2020;35:1953–1959. doi: 10.1111/jgh.15011. [DOI] [PubMed] [Google Scholar]
- 31.Which is better for liver SBRT: dosimetric comparison between DCAT and VMAT for liver tumors. Moon YM, Jeon W, Yu T, Bae SI, Kim JY, Kang JK, Choi CW. Front Oncol. 2020;10:1170. doi: 10.3389/fonc.2020.01170. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Dosimetric comparison of dynamic conformal arc integrated with segment shape optimization and variable dose rate versus volumetric modulated arc therapy for liver SBRT. Thaper D, Kamal R, Singh G, et al. Rep Pract Oncol Radiother. 2020;25:667–677. doi: 10.1016/j.rpor.2020.04.017. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.A dosimetric comparison between volumetric-modulated arc therapy and dynamic conformal arc therapy in SBRT. Stathakis S, Narayanasamy G, Licon AL, et al. https://www.jbuon.com/archive/24-2-838.pdf. J BUON. 2019;24:838–843. [PubMed] [Google Scholar]
- 34.Dosimetric comparison of flattened and flattening filter-free beams for liver stereotactic body irradiation in deep inspiration breath hold, and free breathing conditions. Munirathinam N, Pawaskar PN. J Radiother Pract [Internet. 2019;18:169–110. [Google Scholar]
- 35.Identification of the suitable leaf margin for liver stereotactic body radiotherapy with flattening filter-free beams. Ogata T, Nishimura H, Mayahara H, Uehara K, Okayama T. Med Dosim. 2017;42:268–272. doi: 10.1016/j.meddos.2017.06.002. [DOI] [PubMed] [Google Scholar]
- 36.Dosimetric comparison of flattened and unflattened beams for stereotactic body radiation therapy: Impact of the size of the PTV on dynamic conformal arc and volumetric modulated arc therapy. Vieillevigne L, Bessieres S, Ouali M, Lanaspeze C. Phys Med. 2016;32:1405–1414. doi: 10.1016/j.ejmp.2016.10.007. [DOI] [PubMed] [Google Scholar]
- 37.Evaluation on radiation treatment planning according to use of flattening filter (FF) or flattening filter-free (FFF) in high-energy X-ray equipment. Kim GJ, Choi JK, Hong DH, Kim HG, Beak GM. Indian J Sci Technol. 2016;9:1–5. [Google Scholar]
- 38.Dosimetric and delivery efficiency investigation for treating hepatic lesions with a MLC-equipped robotic radiosurgery-radiotherapy combined system. Jin L, Price RA, Wang L, Meyer J, Fan JJ, Ma CM. Med Phys. 2016;43:727–733. doi: 10.1118/1.4939259. [DOI] [PubMed] [Google Scholar]
- 39.Normal liver sparing by proton beam therapy for hepatocellular carcinoma: Comparison with helical intensity modulated radiotherapy and volumetric modulated arc therapy. Kim JY, Lim YK, Kim TH, et al. Acta Oncol. 2015;54:1827–1832. doi: 10.3109/0284186X.2015.1009637. [DOI] [PubMed] [Google Scholar]
- 40.Safety of high-dose-rate stereotactic body radiotherapy. Stieb S, Lang S, Linsenmeier C, Graydon S, Riesterer O. Radiat Oncol. 2015;10:27. doi: 10.1186/s13014-014-0317-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.Multicentre treatment planning inter-comparison in a national context: the liver stereotactic ablative radiotherapy case. Esposito M, Maggi G, Marino C, et al. Phys Med. 2016;32:277–283. doi: 10.1016/j.ejmp.2015.09.009. [DOI] [PubMed] [Google Scholar]
- 42.Clinical decision tool for optimal delivery of liver stereotactic body radiation therapy: Photons versus protons. Gandhi SJ, Liang X, Ding X, et al. Pract Radiat Oncol. 2015;5:209–218. doi: 10.1016/j.prro.2015.01.004. [DOI] [PubMed] [Google Scholar]
- 43.Stereotactic body radiation therapy for liver tumours using flattening filter free beam: dosimetric and technical considerations. Mancosu P, Castiglioni S, Reggiori G, et al. Radiat Oncol. 2012;7:16. doi: 10.1186/1748-717X-7-16. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44.Flattening filter-free linac improves treatment delivery efficiency in stereotactic body radiation therapy. Prendergast BM, Fiveash JB, Popple RA, et al. J Appl Clin Med Phys. 2013;14:4126. doi: 10.1120/jacmp.v14i3.4126. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45.Can volumetric modulated arc therapy with flattening filter free beams play a role in stereotactic body radiotherapy for liver lesions? A volume-based analysis. Reggiori G, Mancosu P, Castiglioni S, et al. Med Phys. 2012;39:1112–1118. doi: 10.1118/1.3679858. [DOI] [PubMed] [Google Scholar]
- 46.Clinical application of flattening filter free beams for extracranial stereotactic radiotherapy. Lang S, Shrestha B, Graydon S, et al. Radiother Oncol. 2013;106:255–259. doi: 10.1016/j.radonc.2012.12.012. [DOI] [PubMed] [Google Scholar]
- 47.Feasibility and early clinical assessment of flattening filter free (FFF) based stereotactic body radiotherapy (SBRT) treatments. Scorsetti M, Alongi F, Castiglioni S, et al. Radiat Oncol. 2011;6:113. doi: 10.1186/1748-717X-6-113. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 48.Normal liver tissue sparing by intensity-modulated proton stereotactic body radiotherapy for solitary liver tumours. Petersen JB, Lassen Y, Hansen AT, Muren LP, Grau C, Høyer M. Acta Oncol. 2011;50:823–828. doi: 10.3109/0284186X.2011.590526. [DOI] [PubMed] [Google Scholar]
- 49.Dosimetric differences in flattened and flattening filter-free beam treatment plans. Yan Y, Yadav P, Bassetti M, Du K, Saenz D, Harari P, Paliwal BR. J Med Phys. 2016;41:92–99. doi: 10.4103/0971-6203.181636. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 50.A dosimetric comparison of different radiotherapy modalities for non-resected oligometastatic liver disease. Dejonckheere CS, Bilski M, Nour Y, et al. Clin Transl Radiat Oncol. 2025;52:100947. doi: 10.1016/j.ctro.2025.100947. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 51.Dosimetric impact of statistical uncertainty on Monte Carlo dose calculation algorithm in volumetric modulated arc therapy using Monaco TPS for three different clinical cases. Palanisamy M, David K, Durai M, Bhalla N, Puri A. Rep Pract Oncol Radiother. 2019;24:188–199. doi: 10.1016/j.rpor.2019.01.005. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 52.Efficacy of flattening-filter-free beam in stereotactic body radiation therapy planning and treatment: a systematic review with meta-analysis. Dang TM, Peters MJ, Hickey B, Semciw A. J Med Imaging Radiat Oncol. 2017;61:379–387. doi: 10.1111/1754-9485.12583. [DOI] [PubMed] [Google Scholar]
- 53.Using FFF beams to improve the therapeutic ratio of lung SBRT. Vassiliev ON, Peterson CB, Chang JY, Mohan R. J Radiother Pract. 2021;20:419–425. doi: 10.1017/s1460396920000576. [DOI] [PMC free article] [PubMed] [Google Scholar]