Abstract
Accurately mounting dental casts on an articulator is an essential step in prosthetic treatments. In digital dentistry, virtual articulator mounting procedures rely on virtual facebow records. However, virtual facebow records usually require devices like face scanners or jaw motion tracking systems that are not commonly available in most dental practices. The present technique report describes a straightforward intraoral scanner‐based virtual facebow transfer approach. In this technique, a reference facebow joint support was first scanned and aligned with a virtual articulator. Then, a patient's facebow joint support and bite fork assembly were scanned chairside with an intraoral scanner and aligned with the virtual articulator by matching common features with the reference facebow joint support. After aligning the patient's intraoral scans with the patient's bite fork scan that was already superimposed on the virtual articulator, a virtual mounting process was achieved. Once the corresponding reference facebow joint supports have been generated, this technique can be easily implemented with most facebow systems and be seamlessly integrated into daily clinical practice as only an intraoral scanner and a conventional facebow were required.
Keywords: digital dentistry, intraoral scanner, virtual articulator, virtual facebow, virtual mounting
Mounting casts on a virtual articulator are essential for prosthetic treatment planning and restoration fabrication. 1 , 2 , 3 Conventionally, the digital mounting is accomplished by scanning physically mounted casts using a laboratory scanner with an articulator scan function. 4 , 5 , 6 With the increasing integration of digital intraoral scanning techniques in clinical practice, the orientation of intraoral scans on a virtual articulator in the same spatial relationship as in the patient's mouth has become an essential step. 7 , 8
Currently, digital intraoral scan mounting techniques can be classified into two main workflows: the “indirect workflow,” where intraoral scans are 3D‐printed and physically mounted on a mechanical articulator and then transferred to the virtual articulator by using a laboratory scanner with an articulator scan function; 6 , 9 , 10 and the “direct workflow,” where virtual facebow techniques or digital jaw motion tracking systems are used to directly mount intraoral scans on the virtual articulator. 3 , 4 , 7 , 8 , 11 , 12 , 13 , 14 , 15 The indirect workflow is both time‐consuming and error‐prone due to its multi‐step nature, 7 , 15 while the direct workflow presents challenges for being integrated into daily practice, as it generally requires specific and expensive devices or software programs. 8 , 14 , 15
The purpose of this article was to introduce a straightforward intraoral scanner‐based virtual cast mounting approach. By scanning a patient's facebow joint support and bite fork assembly with an intraoral scanner and aligning it with the virtual articulator using a scanned reference facebow joint support, the patient's intraoral scan can be directly mounted on the virtual articulator. Once the digital reference facebow joint support has been generated, this technique can be performed in most dental practices with only an intraoral scanner and a conventional facebow being required.
TECHNIQUE
An Artex articulator (Artex CR Adjustable; Amann Girrbach, Charlotte, NC) was used to demonstrate the present virtual transferring protocol. Figure 1 shows the general workflow of this technique.
FIGURE 1.

General intraoral scanner‐based virtual facebow transfer workflow described in this technique. IOS, intraoral scanner.
Generate reference facebow joint support
-
1.
Mount a facebow joint support (Joint Support; Amann Girrbach, Charlotte, NC) on the corresponding mechanical articulator (Artex CR Adjustable; Amann Girrbach, Charlotte, NC) using a facebow (Artex Facebow; Amann Girrbach, Charlotte, NC) with a bite fork (Bite Fork; Amann Girrbach, Charlotte, NC) and a transfer table (Transfer Table; Amann Girrbach, Charlotte, NC) (Figure 2a,b), to obtain a reference spatial relationship between the facebow joint support and the mechanical articulator.
-
2.
Scan the mounted facebow joint support with the mechanical articulator using an industrial scanner (FreeScan UE Pro; Shining 3D, Hangzhou, China) with the “Laser Scan” function. The accuracy of the scanner is up to 0.02 mm under the “Laser Scan” function according to the manufacturer. Ensure the connection between the facebow joint support and the facebow frame is well captured, as it will serve as a common feature for the alignment with the patient's facebow joint support (Figure 2c,d). Export the obtained scan data as a standard tessellation language (STL) file (Supplementary file 1).
-
3.
Load the STL file into a dental computer‐aided design (CAD) software program (Exocad; exocad GmbH, Darmstadt, Germany). Launch the virtual articulator function and align the STL file with the virtual articulator by matching the common features using the best‐fit algorithm (Figure 3a‐c). A reference facebow joint support that bridges the physical facebow and the virtual articulator is now generated.
FIGURE 2.

(a) Mount reference facebow joint support on mechanical articulator using facebow with bite fork and transfer table. (b) Reference facebow joint support mounted on articulator. (c) Scanned reference facebow joint support and articulator. (d) Lateral view.
FIGURE 3.

(a) Align scanned reference facebow joint support and articulator assembly with virtual articulator by matching common features. (b) Alignment completed. (c) Lateral view.
Obtain patient facebow record and digitalize it chairside by using an intraoral scanner
-
4.
Make a conventional facebow record on the patient (Figure 4a). Remove the patient's facebow joint support and bite fork assembly from the facebow frame. To facilitate subsequent scanning with the intraoral scanner, wrap cloth tape (Steam indicator tape; 3M, Maplewood, MN) around the shining metal surface of the facebow joint support (Figure 4b). Ensure that the connection between the facebow joint support and the facebow frame is not covered for an easier alignment with the reference facebow joint support.
-
5.
Digitalize the facebow joint support and bite fork assembly with an intraoral scanner (TRIOS 3; 3Shape A/S, Copenhagen, Denmark). To increase the scan accuracy, two separate scans were made on the facebow joint support and the bite fork and then were registered together as shown in Figure 4. First, scan the bite fork along with the maxillary occlusion registration using the “Lower Jaw Scan” function of the scanner (Figure 4c). Then, scan the facebow joint support using the “Upper Jaw Scan” function (Figure 4d). Lastly, scan the common connection part between the facebow joint support and the bite fork using the “Bite Scan” function to combine the two scans (Figure 4e). Export the resulting scan data as STL files.
FIGURE 4.

(a) Facebow placed on the patient following standard transferring procedures. (b) Facebow joint support and bite fork assembly covered with cloth tapes. Ensure exposure of connection part between facebow joint support and facebow frame for subsequent alignment. (c) Facebow bite fork with maxillary occlusion registration scanned with “Lower Jaw Scan” function. (d) Facebow joint support scanned with “Upper Jaw Scan” function. (e) Scans aligned by scanning common features using “Bite Scan” function. (f) Scanned arches at maximal intercuspal position.
Patient intraoral scan acquisition
-
6.
Create a new scanning order and scan the maxillary and mandibular arches as well as the maximal intercuspal position (MIP) of the patient with the intraoral scanner. Export the obtained scan data as STL files (Figure 4f).
Virtual articulator mounting
-
7.
Load the obtained patient's facebow joint support and bite fork assembly scan into the dental CAD software program (Exocad; exocad GmbH, Darmstadt, Germany) scene with the reference facebow joint support mounted on the virtual articulator. Align the patient's facebow joint support and bite fork assembly with the reference facebow joint support by matching the common connection part between the facebow joint support and the facebow frame. To enhance the alignment accuracy, uncommon features were marked and excluded from the best‐fit alignment process (Figure 5a,b). Once the alignment was achieved, the patient's facebow joint support and bite fork assembly were mounted on the virtual articulator (Figure 5c).
-
8.
Load the patient's intraoral scans into the software program (Exocad; exocad GmbH, Darmstadt, Germany). Align the maxillary arch scan with the mounted facebow joint support and bite fork assembly by matching the occlusion registration on the bite fork (Figure 6a,b). Then align the mandibular arch scan using the MIP scan as a reference (Figure 6c). The virtual mounting of intraoral scans is now complete (Figure 6d).
FIGURE 5.

(a) Align the patient's facebow joint support and bite fork assembly with reference facebow joint support by matching the common connection part between facebow joint support and facebow frame. Uncommon features marked (blue areas) and excluded from best‐fit alignment to enhance accuracy. (b) Best‐fit alignment completed. (c) Patient's facebow joint support and bite fork assembly mounted on virtual articulator.
FIGURE 6.

(a) Align maxillary arch scan with bite fork by matching occlusion registration. (b) Alignment completed. (c) Mandibular arch scan aligned by using maximal intercuspal position scan data. (d) Intraoral scans mounted on virtual articulator.
Mounting accuracy verification
-
9.
To assess the accuracy of the present technique, the patient's bite fork was mounted on the mechanical articulator (Figure 7a) and then transferred to the same virtual articulator by using the articulator scanning function with a laboratory scanner (DS‐MIX; Shining 3D, Hangzhou, China). The patient's maxillary arch scan was aligned with the scanned bite fork (Figure 7b). Once aligned, two maxillary arch scans mounted by the present technique and the conventional method were obtained (Figure 7c) and imported into a 3D analysis software program (Geomagic Wrap; 3D systems, Roch Hill, SC) for deviation calculation. Most distances between the two scans were less than 1 mm as assessed with the software program (Figure 7d), which is consistent with other virtual facebow techniques and is clinically acceptable. 9 , 16 , 17
FIGURE 7.

(a) Patient's facebow joint support and bite fork assembly mounted on the transfer table. (b) Patient's bite fork transferred to the virtual articulator with maxillary arch scan aligned. (c) Maxillary casts mounted by conventional approach (red) and present technique (white). (d) Color map showing deviations between two casts were less than 1 mm.
DISCUSSION
This technique described a chairside virtual facebow transfer approach using an intraoral scanner and a conventional facebow. It requires a reference scan of the facebow joint support on a physical articulator that can be directly downloaded from Supplemental File 1. This reference scan applies to all cases using the same brand of dental articulators. For different articulator systems, a new reference scan is necessary. After mounting the reference facebow scan on the virtual articulator, a spatial connection between the physical facebow and the virtual articulator is established. This connection serves as a reference for aligning a scanned patient's facebow joint support and bite fork assembly with the virtual articulator, facilitating the chairside intraoral scan mounting process.
The connection between the facebow joint support and facebow frame served as a consistent feature for aligning the patient's facebow scan with the reference facebow scan in this technique, as its spatial position remains unchanged across different transfers (Figure 8a,b). Once the reference facebow was aligned with the virtual articulator, the spatial relation between the connection part and the virtual articulator was recorded. By aligning the patient's facebow joint support scan with the reference facebow scan through the unchanged connection part, the patient's facebow joint support scan was virtually mounted on the virtual articulator. An Artex facebow and articulator system was used in this article. However, this technique could be integrated with most facebow systems on the market as this connection part between the facebow joint support and the facebow frame is typically designed (Figure 8c), which could be scanned and used as a stable reference to achieve the virtual mounting process.
FIGURE 8.

(a) Spatial position between connection part and facebow frame remains unchanged across different facebow transfers. (b) Spatial position between connection part and articulator remains unchanged across different facebow transfers. By aligning patient's facebow joint support with the reference facebow through this unchanged connection part, a virtual mounting is achieved. (c) An unchanged connection part can be found on most facebow systems, which can be scanned and used as a reference to achieve the virtual mounting procedures proposed in this technique.
To enhance the accuracy of this technique, the standard scanning protocol endorsed by the manufacturer should be followed during the intraoral scanning process. 16 , 17 Meanwhile, it is crucial to cover the patient's metal facebow joint support and bite fork assembly with cloth tape to facilitate scanning, as the shining metal surface is hard to be scanned with an intraoral scanner. 18 , 19 During the alignment of the patient's facebow joint support with the reference facebow joint support, uncommon features should be excluded to increase the best‐fit alignment accuracy. 20
Limitations of this technique include that a mechanical articulator with a reference facebow joint support must be first scanned with an industrial scanner; this can be solved by using the scanning services provided by industrial scanning companies or by downloading the attached supplemental file if the same type of articulator is being used. Meanwhile, the scan only needs to be performed once for each type of facebow and articulator. Once generated, the reference facebow joint support can be imported into the dental CAD software program like any other reference tool and used repeatedly. In this article, only 1 patient was assessed. Further research with a large sample size is needed to verify the accuracy of the present technique.
SUMMARY
This technique described a chairside virtual facebow transfer approach using an intraoral scanner and a conventional facebow. This technique is adaptable and can be integrated into daily clinical practice once the corresponding reference facebow joint support is established.
CONFLICT OF INTEREST STATEMENT
The authors declare that there are no conflicts of interest in this article.
Supporting information
Supporting Information
REFERENCES
- 1. Koralakunte PR, Aljanakh M. The role of virtual articulator in prosthetic and restorative dentistry. J Clin Diagn Res. 2014;8(7):ZE25–ZE258. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2. Kordass B, Gärtner C, Söhnel A, Bisler A, Voss G, Bockholt U, et al. The virtual articulator in dentistry: concept and development. Dent Clin North Am. 2002;46(3):493–506. [DOI] [PubMed] [Google Scholar]
- 3. Lepidi L, Galli M, Mastrangelo F, Venezia P, Joda T, Wang HL, et al. Virtual articulators and virtual mounting procedures: where do we stand? J Prosthodont. 2021;30(1):24–35. [DOI] [PubMed] [Google Scholar]
- 4. Kim JE, Kim SJ, Kwon DH, Shim JS, Kim JH. Mounting casts on a mechanical articulator by using digital multisource data: a dental technique. J Prosthet Dent. 2021;125(1):41–45. [DOI] [PubMed] [Google Scholar]
- 5. Úry E, Fornai C, Weber GW. Accuracy of transferring analog dental casts to a virtual articulator. J Prosthet Dent. 2020;123(2):305–313. [DOI] [PubMed] [Google Scholar]
- 6. Inoue N, Scialabba R, Lee JD, Lee SJ. A comparison of virtually mounted dental casts from traditional facebow records, average values, and 3D facial scans. J Prosthet Dent. 2024;131(1):136–143. [DOI] [PubMed] [Google Scholar]
- 7. Lepidi L, Chen Z, Ravida A, Lan T, Wang HL, Li J. A full‐digital technique to mount a maxillary arch scan on a virtual articulator. J Prosthodont. 2019;28(3):335–338. [DOI] [PubMed] [Google Scholar]
- 8. Yang S, Feng N, Li D, Wu Y, Yue L, Yuan Q. A novel technique to align the intraoral scans to the virtual articulator and set the patient‐specific sagittal condylar inclination. J Prosthodont. 2022;31(1):79–84. [DOI] [PubMed] [Google Scholar]
- 9. Petre A, Drafta S, Oancea L. A new technique to transfer the upper maxillary arch position using a facebow, a transfer table, and a reference block with a CAD application. J Prosthodont. 2024;33(2):195–200. [DOI] [PubMed] [Google Scholar]
- 10. Lustig LL, Williams W, Gozalo‐Diaz D. Printing and mounting digital casts in the centric relation position for use on an analog articulator: a dental technique. J Prosthet Dent. 2021;125(4):581–584. [DOI] [PubMed] [Google Scholar]
- 11. Yang S, Dong B, Zhang Q, Li J, Yuan Q, Yue L. An indirect digital technique to transfer 3D printed casts to a mechanical articulator with individual sagittal condylar inclination settings using CBCT and intraoral scans. J Prosthodont. 2022;31(9):822–827. [DOI] [PubMed] [Google Scholar]
- 12. Solaberrieta E, Garmendia A, Minguez R, Brizuela A, Pradies G. Virtual facebow technique. J Prosthet Dent. 2015;114(6):751–755. [DOI] [PubMed] [Google Scholar]
- 13. Petre A, Drafta S, Stefanescu C, Oancea L. Virtual facebow technique using standardized background images. J Prosthet Dent. 2019;121(5):724–728. [DOI] [PubMed] [Google Scholar]
- 14. Li Q, Bi M, Yang K, Liu W. The creation of a virtual dental patient with dynamic occlusion and its application in esthetic dentistry. J Prosthet Dent. 2021;126(1):14–18. [DOI] [PubMed] [Google Scholar]
- 15. Lin H, Pan Y, Wei X, Wang Y, Yu H, Cheng H. Comparison of the performance of various virtual articulator mounting procedures: a self‐controlled clinical study. Clin Oral Investig. 2023;27(7):4017–4028. [DOI] [PubMed] [Google Scholar]
- 16. Solaberrieta E, Mínguez R, Barrenetxea L, Otegi JR, Szentpétery A. Comparison of the accuracy of a 3‐dimensional virtual method and the conventional method for transferring the maxillary cast to a virtual articulator. J Prosthet Dent. 2015;113(3):191–197. [DOI] [PubMed] [Google Scholar]
- 17. Lam WYH, Hsung RTC, Choi WWS, Luk HWK, Cheng LYY, Pow EHN. A clinical technique for virtual articulator mounting with natural head position by using calibrated stereophotogrammetry. J Prosthet Dent. 2018;119(6):902–908. [DOI] [PubMed] [Google Scholar]
- 18. Revilla‐León M, Kois DE, Kois JC. A guide for maximizing the accuracy of intraoral digital scans. Part 1: operator factors. J Esthet Restor Dent. 2023;35(1):230–240. [DOI] [PubMed] [Google Scholar]
- 19. Revilla‐León M, Kois DE, Kois JC. A guide for maximizing the accuracy of intraoral digital scans: part 2‐Patient factors. J Esthet Restor Dent. 2023;35(1):241–249. [DOI] [PubMed] [Google Scholar]
- 20. Revilla‐León M, Gohil A, Barmak AB, Zandinejad A, Raigrodski AJ, Alonso Pérez‐Barquero J. Best‐fit algorithm influences on virtual casts' alignment discrepancies. J Prosthodont. 2023;32(4):331–339. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Supporting Information