Abstract
Background and aims
Vascular complications occur in a non-negligible proportion of transfemoral transcatheter aortic valve implantation (Tf-TAVI) procedures. There is only limited evidence regarding the efficacy and safety of covered balloon-expandable (CBE) stents in the management of Tf-TAVI related vascular complications. We aimed to investigate the efficacy and safety of CBE stent implantation to treat access-related vascular complications in patients undergoing Tf-TAVI.
Methods
The present retrospective analysis included patients undergoing CBE stent implantation following Tf-TAVI from April 2012 to January 2023 at our centre. The primary endpoint was technical success defined as successful device delivery and implantation at the intended location with angiographic confirmation of vessel patency and absence of residual bleeding. Clinical outcomes and color Doppler ultrasonography findings were evaluated at discharge, 30 days and longest available follow-up.
Results
Among 3331 Tf-TAVI procedures, 93 patients (2.8%) required covered stent implantation for the treatment of access related vascular complications. Technical success was achieved in 92 patients (98.9%). BeGraft and Atrium Advanta V12 CBE stents were implanted in 76 (81.7%) and 17 (18.3%) patients, respectively. Median clinical follow-up was 455 [304; 798] days, with both in-hospital and 30-days mortality equaling 4.3% and 1-year mortality 15.1%. Pre-discharge Doppler ultrasonography was performed in 98.9% patients, with normal findings in 79.8% and minor findings (such as pseudoaneurysm, non-flow-limiting dissection, arterio-venous fistula) in the remaining patients. Clinical follow-up was available in 96.2% patients; no cases of new-onset claudication, need for repeat surgical or transcatheter vascular interventions were recorded. Follow-up Doppler ultrasonography (54.4% patients) showed persistent patency and no signs of stent failure in all patients.
Conclusions
CBE stent implantation for the treatment of access site related vascular complications after Tf-TAVI is associated with excellent technical success rates, optimal short- to mid-term patency rates as well as satisfactory long-term clinical outcomes.
Supplementary Information
The online version contains supplementary material available at 10.1007/s00392-025-02651-2.
Keywords: Covered balloon-expandable stent, Vascular complication, Transcatheter aortic valve implantation, Transfemoral access
Introduction
A transfemoral (Tf) approach currently represents the access of choice in patients undergoing transcatheter aortic valve implantation (TAVI) for severe aortic valve stenosis (AVS) [1, 2]. Although continuous device iterations, increasing operator experience and expansion of TAVI indications towards lower risk populations have led to a drastic reduction in the rate of access-related major vascular complications, the latter still occur in a non-negligible proportion of procedures and have a negative impact on patient outcomes [3–5].
Prompt recognition as well as management of large-bore access-related vascular complications is of paramount importance, with available treatment options ranging from prolonged manual compression, endovascular balloon inflation, covered stent implantation or surgical vascular repair [6]. Since the vast majority of vascular complications are recognized at the end of the procedure with a contralateral arterial access already available, an endovascular approach represents an attractive treatment strategy, which might even have potential advantages compared to conventional surgical repair [7]; indeed, considering the advanced age and comorbidities of patients undergoing TAVI procedures, the latter option portends a non-negligible morbidity and mortality risk.
If implantation of covered stents is required, both covered self-expanding (CSE) and covered balloon-expandable (CBE) stent platforms are available with the choice being usually dictated by operator’s expertise and preferences [8]. Currently, the evidence supporting use of covered stent platforms in the management of large-bore access-related vascular complications in the setting of TAVI is limited to small retrospective registries, mostly reporting short- to mid-term outcomes following CSE stents [9–12], while no direct comparative studies evaluating efficacy and safety of CSE and CBE stent platforms have been performed so far. Against this background, we aimed to investigate the efficacy and safety of CBE stent implantation for the treatment of access-related vascular complications in patients undergoing Tf-TAVI in a large, all-comers retrospective patient cohort.
Methods
Study population
This is a retrospective, observational analysis including all consecutive patients undergoing implantation of covered stents for the treatment of access site related vascular complications in the setting of Tf-TAVI between April 2012 and January 2023 at the Department of Cardiology of the German Heart Centre Munich, Germany.
Patient clinical and procedural information was extracted from a dedicated, prospective institutional database. During the study period, 3331 Tf-TAVI procedures were performed at our centre using commercially available transcatheter heart valve (THV) systems. Indication for TAVI was discussed in a multidisciplinary heart team, after careful consideration of patient age and clinical characteristics as well as findings of transthoracic echocardiography, multi-slice computed tomography (MSCT) and coronary angiography.
The study was performed in accordance with the principles of the Declaration of Helsinki and all patients provided written informed consent for the procedure. Ethical approval was obtained from the Ethics Committee of the Technical University of Munich under the OBSERVTAVI Registry (#525/17).
Computed tomography analysis
MSCT was analyzed by two cardiologists with extensive experience in the evaluation of MSCT. Based on MSCT reconstruction, qualitative and quantitative measures of valve size and peripheral vessel size as well as presence and distribution of calcification were performed by multi-planar reconstruction (MPR) using the 3-Mensio software (Pie Medical Imaging, Maastricht, The Netherlands). Vessel calcification was graded based on the degree of its circumferential extension for each ilio-femoral segment. Vessel angulation was measured at the most acute angle in the abdominal aorta and the iliac artery segments; tangent angle function was applied between two neighbouring vascular segments including the acute angle. Lumen vascular diameter was measured for every aorto-iliac segment and plaque characteristics were also evaluated.
Procedural technique
Vessel access, obtained through a standard percutaneous Seldinger technique in all patients, was followed by the placement of one or two suture-based closure systems (Perclose ProGlide—Abbott Vascular). Following THV implantation, the sheath was retrieved and the suture-based closure systems were tightened; in case of persistent residual bleeding, plug-based closure systems (6 F or 8 F) were applied at the discretion of the operator. In case of major vascular complication not manageable by other means (uncontrollable bleeding and/or vessel stenosis/occlusion/dissection) (Fig. 1A), a cross-over technique was used to gain access to the superficial femoral artery and advance a 0.035 inch supportive guidewire as well as a 7 F/8 F destination guiding sheath in the superficial/common femoral artery of the main access site, respectively, followed by the implantation of covered stents under fluoroscopic guidance (Fig. 1B and C). The size (length and diameter) of the covered stent was chosen based on both CT and angiographic measurements. Finally, a digital subtraction angiography was performed to confirm vessel patency and absence of residual bleeding and/or vascular complications (Fig. 1D).
Fig. 1.
A Bleeding complication at the main access vessel following closure device failure; B Balloon-expandable covered stent (Bentley 8 × 37 mm) positioning at the bleeding level; C Covered stent implantation; D Final angiographic result
Clinical and Doppler ultrasound follow-up
Patients were monitored for the occurrence of adverse events during the hospital stay and a vascular color Doppler ultrasound was recommended before discharge in all patients to exclude silent vascular injury, such as significant vascular stenosis or residual dissections. Clinical follow-up was performed by office visit, phone contact or structured follow-up letter. Doppler ultrasonographic follow-up was recommended for all patients at 1 year.
Endpoints
The primary endpoint of the study was represented by technical success, defined as successful delivery and implantation of the covered stent at the intended vascular location with angiographic confirmation of vessel patency and absence of residual bleeding, need for bailout surgery or intra-procedural death. In-hospital mortality, new onset claudication, acute limb ischemia or amputation of the ipsilateral extremity, need for repeat percutaneous or surgical vascular intervention as well as evidence of stent occlusion or fracture were also assessed. Vascular complications were evaluated and reported in accordance with the Valve Academic Research Consortium (VARC)−3 criteria [13]. Doppler ultrasonographic data obtained before discharge or during follow-up were recorded in all patients.
Statistical analysis
Categorical variables are shown as counts (%). Normality of distribution of continuous data was assessed using the Shapiro–Wilk test. Continuous data are shown as mean ± standard deviation (SD) or median with [25 th–75 th] percentiles depending on the distribution pattern. Statistical analysis was performed using SPSS Statistics Version 29.0.
Results
Patient population
Between April 2012 and January 2023, 3331 Tf-TAVI procedures were performed at our centre. Minor and/or major vascular complications occurred in 387 patients (11.6%), with covered stent implantation to treat access-site related vascular complications being required in 93 patients (2.8%) (Fig. 2). Patient baseline clinical and echocardiographic characteristics are shown in Table 1. The median [25 th–75 th percentile] age was 81 [77; 85] years and 71% of patients (n = 66) had concomitant coronary artery disease. Left ventricular ejection fraction was 56% [45%; 60%] and the majority of patients had a moderate-to-high surgical mortality risk (EuroScore I: 12.64 [8.14; 23.46]; EuroScore II 4.07 [2.49; 7.27]).
Fig. 2.
Flowchart of the study population
Table 1.
Baseline clinical and echocardiographic characteristics
| N = 93 | |
|---|---|
| Age, years | 81 [77; 85] |
| Sex, female | 59 (63.4) |
| Body mass index, kg/m2 | 26.9 [24.1; 29.1] |
| Body surface area, m2 | 1.80 [1.68; 2.0] |
| New York Heart Association Class ≥ III | 57 (61.3) |
| Arterial hypertension | 86 (92.5) |
| Diabetes mellitus | 35 (37.6) |
| Dyslipidemia | 73 (78.5) |
| Smoking history | 9 (9.7) |
| Atrial fibrillation | 51 (54.8) |
| Chronic obstructive pulmonary disease | 10 (10.8) |
| Peripheral artery disease | 25 (26.9) |
| Previous pacemaker | 13 (14) |
| Coronary artery disease | 66 (71) |
| Previous percutaneous coronary intervention | 36 (38.7) |
| Previous myocardial infarction | 9 (9.7) |
| Previous coronary artery bypass grafting | 8 (8.6) |
| Previous aortic valve replacement | 10 (10.7) |
| Surgical | 7 (7.5) |
| TAVI | 3 (3.2) |
| History of cancer | 19 (20.4) |
| Serum creatinine, mg/dL | 1.08 [0.89; 1.43] |
| Estimated glomerular filtration rate, mL/min | 57 [40; 71] |
| Chronic dialysis | 2 (2.2) |
| Previous stroke/transient ischemic attack | 8 (8.6) |
| Left ventricular ejection fraction, % | 56 [45; 60] |
| Aortic valve area, cm2 | 0.71 [0.55; 0.82] |
| Mean transvalvular pressure gradient, mmHg | 39.8 ± 15.3 |
| Systolic pulmonary artery pressure, mmHg | 46.1 ± 14.1 |
| EuroScore I | 12.64 [8.14; 23.46] |
| EuroScore II | 4.07 [2.49; 7.27] |
Data are shown as counts (%), mean ± SD (standard deviation) or median [25th–75 th percentiles]
TAVI transcatheter aortic valve implantation
CT analysis
CT characteristics are reported in Table 2. The mean most acute angle on the main access side was 90 ± 27°. Minimal lumen diameter (MLD) at the main access side was 7.2 ± 1.8 mm for the common iliac artery (CIA), 6.1 ± 1.5 mm for the external iliac artery (EIA) and 6.05 ± 1.4 mm for the common femoral artery (CFA). Circumferential vessel calcification (> 180°) at the main access side was observed in 38 patients (40.9%) for the CIA, 14 patients (15.1%) for the EIA and 15 patients (16.1%) for the CFA. Aortic valve characteristics are shown in Supplementary Table 1. Aortic valve mean diameter was 23.9 ± 2.4 mm, while severe aortic valve calcification was present in 24.1% of patients.
Table 2.
Computed tomography analysis of peripheral vessels
| N = 93 | |
|---|---|
| Most acute iliac vessel angulation, degrees | |
| Main access | 90.7 ± 27.0 |
| Right | 89.7 ± 28.5 |
| Left | 92.1 ± 27.7 |
| Main access, MLD, mm | 5.5 ± 1.4 |
| CIA minimal diameter, mm | 7.2 ± 1.8 |
| CIA reference diameter, mm | 9.2 ± 1.9 |
| CIA index | 0.81 [0.72; 0.94] |
| EIA minimal diameter, mm | 6.1 ± 1.5 |
| EIA reference diameter, mm | 7.4 ± 1.6 |
| EIA index | 0.87 [0.80; 0.96] |
| CFA minimal diameter, mm | 6.05 ± 1.4 |
| CFA reference diameter, mm | 7.4 ± 1.4 |
| CFA index | 0.84 [0.72; 0.95] |
| Circumferential degree of calcification of peripheral vessels | |
| Main access | |
| Common iliac artery | |
| No calcification | 4 (4.3) |
| 0 to 180 degrees | 51 (54.8) |
| More than 180 degrees | 38 (40.9) |
| External iliac artery | |
| No calcification | 34 (36.6) |
| 0 to 180 degrees | 45 (48.4) |
| More than 180 degrees | 14 (15.1) |
| Common femoral artery | |
| No calcification | 15 (16.1) |
| 0 to 180 degrees | 63 (67.8) |
| More than 180 degrees | 15 (16.1) |
Data are shown as counts (%), mean ± SD (standard deviation) or median [25th–75 th percentiles]
CIA common iliac artery, CFA common femoral artery, EIA external iliac artery, MLD minimum lumen diameter, mm millimeters
Procedural characteristics and in-hospital outcomes
Procedural characteristics are shown in Table 3. The main vascular access was achieved via right or left CFA in 74 (79.6%) and 19 (20.4%) patients, respectively. The minimal lumen diameter was 5.5 ± 1.4 mm, with a sheath to artery ratio of 0.86 [0.71; 1.04]. Sheath size was 14 F in 82 patients (88.2%) and 16 F in 11 patients (11.8%). An upfront percutaneous transluminal angioplasty (PTA) of the iliofemoral vessels to facilitate sheath insertion was required in 10 patients (10.8%). A balloon-expandable THV was implanted in 63.8% of patients while a self-expanding THV was used in the remaining patients. Vascular complications requiring covered stent implantation were located at the main access site in 87 (93.5%) patients and at the contralateral access site in 6 (6.5%) patients. CBE stents were used in all cases, with BeGraft (Bentley InnoMed GmbH, Hechingen, Germany) stents being implanted in 76 (81.7%) and Atrium Advanta V12 (Getinge) stents in 17 (18.3%) patients, respectively. Seventy-eight patients (83.9%) received one covered stent, whereas 2 or more CBE stents were required in 15 (16.1%) patients.
Table 3.
Procedural characteristics
| N = 93 | |
|---|---|
| Elective procedure | 84 (90.3) |
| Procedure time, min | 90 [70; 111] |
| Fluoroscopy time, min | 20.3 [15.3; 29.3] |
| Fluoroscopy dose, cGy*cm2 | 4648.1 [2257.5; 7057.6] |
| Contrast medium, mL | 244 [200; 320] |
| Main access | |
| Right | 74 (79.6) |
| Left | 19 (20.4) |
| Minimal lumen diameter, mm | 5.5 ± 1.4 |
| Sheath to artery ratio | 0.86 [0.71; 1.04] |
| Sheath size (Fr) | |
| 14 | 82 (88.2) |
| 16 | 11 (11.8) |
| Upfront PTA to facilitate sheath insertion | 10 (10.8) |
| Location of vascular complication | |
| Main access | 87 (93.5) |
| Contralateral access | 6 (6.5) |
| Complication management | |
| Covered balloon-expandable stent | 93 (100) |
| Number of covered balloon-expandable stents | |
| One | 78 (83.9) |
| Two or more | 15 (16.1) |
| Type of covered balloon-expandable stent | |
| BeGraft | 76 (81.7) |
| Atrium Advanta V12 | 17 (18.3) |
| Diameter of the stent, mm | 9 [8; 9] |
| 6 mm | 1 (1.1%) |
| 7 mm | 7 (7.5%) |
| 8 mm | 35 (37.6%) |
| 9 mm | 32 (34.4%) |
| 10 mm | 16 (17.2%) |
| Unknown | 2 (2.2%) |
| Length of the stent, mm | 37 [37; 37] |
| 27 mm | 8 (8.6%) |
| 28 mm | 2 (2.2%) |
| 30 mm | 1 (1.1%) |
| 37 mm | 61 (65.6%) |
| 38 mm | 13 (14%) |
| 39 mm | 1 (1.1%) |
| 57 mm | 2 (2.2%) |
| 59 mm | 1 (1.1%) |
| Successful transcatheter heart valve delivery | 91 (97.8) |
| Successful transcatheter heart valve implantation | 91 (97.8) |
| Multiple valves | 0 (0) |
| Aortic valve pre-dilation | 58/91 (63.7) |
| Aortic valve post-dilation | 37/91 (40.7) |
| Transcatheter heart valve implanted | 91/93 (97.8) |
| Sapien XT | 2/91 (2.2) |
| Sapien 3 | 19/91 (20.9) |
| Sapien 3 Ultra | 37/91 (40.7) |
| Acurate neo | 6/91 (6.6) |
| Acurate neo2 | 19/91 (20.9) |
| Evolut R | 8/91 (8.8) |
| Size of implanted transcatheter heart valve | |
| 23 mm | 34/91 (37.4) |
| 25 mm | 12/91 (13.2) |
| 26 mm | 32/91 (35.2) |
| 27 mm | 3/91 (3.2) |
| 29 mm | 10/91 (11) |
Data are shown as counts (%), mean ± SD (standard deviation) or median [25th–75 th percentiles]
PTA percutaneous transluminal angioplasty
Technical success, defined as successful delivery and implantation of the covered stent in the absence of residual bleeding, need for bailout surgery or intra-procedural death was achieved in 92 patients (98.9%). Successful THV delivery and implantation was achieved in 91 patients (97.8%). THV delivery was unsuccessful in 2 (2.2%) patients; in one patient, after combined upfront intravascular lithotripsy (IVL) and non-compliant (NC) balloon angioplasty of the ilio-femoral vessels and successful sheath placement, repeated advancement attempts of a balloon‐expandable THV led to partial extroversion of the valve frame, with ensuing vessel laceration. Despite placement of a stopping balloon in the abdominal aorta and implantation of several covered stents, refractory hemorrhagic shock leading to intraprocedural death occurred. In the second patient, despite successful sheath placement, THV valve advancement proved impossible and, following sheath removal, percutaneous placement of a covered stent was required to achieve hemostasis at the access site; the patient died on the 3rd postprocedural day, due to cardiogenic shock. Albeit very rare, these examples underscore the importance of achieving complete haemostasis in an effective and timely manner and illustrate the significant morbidity and mortality related to a failure in achieving these goals.
In-hospital clinical and echocardiographic findings are reported in Table 4. All-cause mortality was observed in four patients (4.3%). Besides the two aforementioned patients, one patient died due to septic shock (1.1%) and the other patient due to hemorrhagic shock (1.1%).
Table 4.
In-hospital clinical and echocardiographic outcomes
| Intra-procedural death | 1 (1.1) |
| In hospital all-cause mortality | 4 (4.3) |
| Cardiogenic shock | 1 (1.1) |
| Hemorrhagic shock | 2 (2.2) |
| Septic shock | 1 (1.1) |
| Stroke | |
| Major | 1 (1.1) |
| Minor | 2 (2.2) |
| Periprocedural acute myocardial infarction | 1 (1.1) |
| New permanent pacemaker | 7 (7.5) |
| Acute kidney injurya | 14 (15.1) |
| Red blood cell units transfused | 0 [0; 2] |
| Intensive care unit length of stay, days | 1 [1; 3] |
| Degree of paravalvular regurgitation | |
| Angiographic PVL ≥ 2 | 4 (4.3) |
| Echocardiographic PVL ≥ 2 | 2 (2.2) |
| Mean trans-prosthetic gradient, mmHg | 11 [8; 15] |
| Maximal trans-prosthetic gradient, mmHg | 20.6 [15.4; 28.7] |
| Post-procedural systolic pulmonary artery pressure, mmHg | 39.9 ± 15.5 |
| Minimal post-procedural hemoglobin value, gr/dL | 8.4 [7.5; 10.2] |
Data are shown as counts (%), mean ± SD (standard deviation) or median [25th–75 th percentiles]
PVL paravalvular leak
aAcute kidney injury was defined according to RIFLE criteria
Access site and valve-related complications
The access-site and cardiac structural- and technical valve-related complications are shown in Table 5. Closure device failure was observed in 37 patients (39.8%), puncture-related bleeding in 21 patients (22.6%) and dissection or stenosis in 20 patients (21.5%). No conversion to surgery or valve embolization/migration was observed. One patient (1.1%) had an embolic coronary obstruction that was successfully managed through percutaneous transcatheter intervention.
Table 5.
Access-site and cardiac structural- and technical valve-related complications
| N = 93 | |
|---|---|
| Procedural and technical valve-related complications | |
| Conversion to surgery | 0 (0) |
| Valve embolization | 0 (0) |
| Valve migration | 0 (0) |
| Ectopic valve deployment | 0 (0) |
| Cardiac structural complications | |
| Cardiac perforation | 0 (0) |
| Cardiac tamponade | 0 (0) |
| Coronary obstruction | 1 (1.1) |
| Access-site related complications | |
| Closure device failure | 37 (39.8) |
| Puncture related bleeding | 21 (22.6) |
| Dissection/stenosis | 20 (21.5) |
| Thrombosis | 5 (5.4) |
| Perforation | 7 (7.5) |
| Pseudoaneurysm | 3 (3.2) |
Data are shown as counts (%)
Clinical and Doppler ultrasonographic follow-up
Median clinical follow-up was 455 [304; 798] days, with both in-hospital and 30-days mortality equaling 4.3%; importantly, no additional deaths and no repeat surgical or transcatheter peripheral vascular interventions were required during the critical time window comprised between patient discharge and 30-days follow-up. All-cause mortality equaled 15.1% at one year and 23.7% at the longest available clinical follow-up (Table 6). In-hospital and follow-up Doppler ultrasonographic data are shown in Table 6. In-hospital Doppler ultrasonographic examination was performed in 88/89 (98.9%) patients, with normal findings in 71 patients (79.8%), main access-related pseudoaneurysm in 2 patients (2.2%), contralateral access-related pseudoaneurysm in 5 patients (5.6%), main access related dissection in one patient (1.1%), main access arterio-venous fistula in one patient (1.1%), contralateral access arterio-venous fistula in one patient (1.1%) and a monophasic flow in two patients with advanced peripheral artery disease (2.2%) (Fig. 3). Clinical follow-up was available in 76/79 (96.2%) patients. Three patients were lost to follow-up. No cases of new-onset claudication, need for repeat surgical or transcatheter vascular interventions were recorded. Doppler ultrasonographic follow-up (92 [46; 101] days) was available in 43 (54.4%) patients and showed persistent stent patency in all patients with no signs of stent failure or additional pathological findings. Two patients with pre-existing advanced peripheral artery disease had a monophasic flow related to the underlying vascular disease but no signs of stent failure.
Table 6.
Clinical and Doppler ultrasound follow-up
| N = 93 | |
|---|---|
| Follow-up, days | 455 [304; 798] |
| In-hospital all-cause mortality | 4 (4.3) |
| 30-days all-cause mortality | 4 (4.3) |
| Six-months all-cause mortality | 11 (11.8) |
| One-year all-cause mortality | 14 (15.1) |
| Overall all-cause mortality total | 22 (23.7) |
| Pre-discharge Doppler ultrasound available | 88/89 (98.9) |
| Pre-discharge Doppler ultrasound findings | |
| Normal | 71 (79.8) |
| Main access vessel pseudoaneurysm | 2 (2.2) |
| Contralateral access vessel pseudoaneurysm | 5 (5.6) |
| Main access dissection | 1 (1.1) |
| Main access arterio-venous fistula | 1 (1.1) |
| Contralateral arterio-venous fistula | 1 (1.1) |
| Monophasic signal | 2 (2.2) |
| Clinical follow-up available | 76/79 (96.2) |
| Lost to follow-up | 3/89 (3.4) |
| Doppler ultrasound control follow-up, days | 92 [46; 101] |
| Follow-up Doppler ultrasound available | 43/79 (54.4) |
| Normal | 41/43 (95.3) |
| Monophasic signal | 2/43 (4.7) |
Data are shown as counts (%), mean ± SD (standard deviation) or median [25th–75 th percentiles]
Fig. 3.
Findings of pre-discharge vascular Doppler ultrasonographic examinations in patients undergoing implantation of covered balloon-expandable stents for the treatment of access-site related vascular complications in the setting of transfemoral transcatheter aortic valve implantation
Discussion
The present study sought to investigate immediate and mid- to long-term efficacy and safety of CBE stent implantation used for the management of access-related vascular complications in the setting of Tf-TAVI procedures. The major findings of the study can be summarized as follows: i) CBE stent implantation is highly effective in achieving complete hemostasis as well as vessel patency, with a technical success rate of 98.9%; ii) mid-term Doppler ultrasonographic and long-term clinical follow-up support the persistent efficacy and safety of CBE stent implantation in iliofemoral vessels, with no requirement for repeat transcatheter or surgical peripheral vascular interventions or clinically apparent stent failure during the available follow-up. Additionally, this is the first report confirming the efficacy and safety of the new generation single layer polytetrafluoroethylene (PTFE) BeGraft (Bentley InnoMed GmbH, Hechingen, Germany) CBE stent, used in the vast majority of procedures in the present registry, in the management of access-related vascular complications in the setting of Tf-TAVI procedures.
Besides continuous device iterations and increasing operator experience, which have already led to significant reductions in the rate of access-related major vascular complications, another promising option in preventing their occurrence is represented by ultrasound-guided puncture. By providing real-time images of important vessel anatomical features, such as location of the femoral artery bifurcation as well as presence and extension of atherosclerotic plaque and/or calcification, this technique facilitates the selection of an optimal puncture site. Although prospective, randomized studies supporting its use in the setting of Tf-TAVI are still lacking, there is growing evidence showing that use of this approach is associated with a significant reduction of vascular and bleeding complications associated with procedures requiring a femoral access [14].
Due to the lack of direct comparative studies, no clear recommendations are available regarding the optimal management strategy of vascular complications in the setting of TAVI procedures. Although surgical repair has represented in many centers the time-honored management strategy, due to the advanced age and comorbidities of patients undergoing TAVI procedures, it portends a non-negligible morbidity and mortality risk [15]. By allowing a rapid and effective vascular repair in the setting of large bore vascular access, endovascular approaches have become an attractive treatment strategy, which might even display potential advantages compared to conventional surgical repair [7].
Based on their release mechanism as well as engineering and design, covered stents can be classified into CSE and CBE. These two families display significant differences in terms of intrinsic physical properties, implantation characteristics as well as interaction with native vessel anatomy, which might translate into advantages and/or drawbacks in specific vascular territories. CBE stents display higher rigidity and radial stiffness and allow a more precise implantation and lesser foreshortening compared to CSE stents, which might be of particular interest in the proximity of large bifurcations. On the other hand, CSE stents are significantly more elastic and flexible, offer superior resistance to axial fatigue, which might translate into superior long-term mechanical stent integrity; the latter characteristic, might be particularly important in vascular territories subjected to cyclic flexion–extension movements, such as the common femoral artery. For this reason, the majority of previous reports investigating the use of covered stents in the setting of TAVI-related vascular complications have concentrated on CSE stents [9–12]. Taken together, these studies have reported high technical success rates [11] and excellent long-term patency rates of CSE stents [12]. On the other hand, comparative studies investigating efficacy and safety of balloon-expandable and self-expanding uncovered stents for the treatment of ilio-femoral lesions have shown somehow conflicting results [16, 17]. The Society for Cardiovascular Angiography and Interventions (SCAI) guidelines on device selection, give stronger recommendations to BE as compared to SE uncovered stents in aorto‐iliac arterial interventions [18], whereas the ESC guidelines give no specific recommendations regarding stent type [19]. The number of investigations evaluating clinical outcomes following CBE stent implantation is limited [20, 21], with the largest registry to date by Maurina et al. [20] retrospectively evaluating the outcomes of 78 patients undergoing implantation of the Atrium Advanta V12 (Getinge) CBE stents. The authors reported excellent technical success rates (96.2% as compared to 98.9% in the present registry) with bailout surgery required in only two patients; analogously to the present report, at a median follow-up of 429 (89–994) days [455 (304–798) days in the present registry], no cases of acute ipsilateral lower limb ischemia or need for repeat intervention were reported, thereby confirming the persistent excellent efficacy and safety of CBE stents in this setting (Supplementary Table 2).
Importantly, in the present registry, the vast majority (81.7%) of the patients underwent implantation of a BeGraft (Bentley) CBE stent; while there is a reasonable amount of data supporting the persistent efficacy and favorable long-term patency rate of the Advanta V12 (Getinge) CBE stent in the treatment of aorto-iliac occlusive disease [22], the present report is, to the best of our knowledge, the first to provide data regarding procedural efficacy and mid- to long-term clinical efficacy and safety following implantation of this novel CBE stent in ilio-femoral vessels in the setting of access-site-related vascular complications of Tf-TAVI procedures. The BeGraft (Bentley InnoMed GmbH, Hechingen, Germany) CBE stent is a Cobalt-Chrome (L-605), open-cell platform covered with a single-layer microporous ePTFE membrane (203 ± 25 μm) clamped at the proximal and distal stent ends, which has already shown encouraging results in terms of both clinical and angiographic efficacy as well as of safety profile in the management of perforations located in coronary vessels [23] as well as of iatrogenic access site related arterial side-branch injury [24]. The results of the present study extend these previous findings and confirm the excellent efficacy and safety of this novel stent platform for the treatment of access related vascular complications in the setting of TAVI procedures.
Study limitations
The present report has a number of limitations. Despite its moderately large sample size, this is a single center, retrospective, observational study and a selection bias cannot be rule out. Moreover, although the reasonably long and thoroughly conducted clinical follow-up is reassuring regarding the absence of hard clinical endpoints, only 54.4% of the patient collective underwent a Doppler ultrasonographic follow-up and the occurrence of asymptomatic stent failure (stent fracture or restenosis) might be underestimated.
Conclusions
CBE stent implantation for the treatment of access site-related vascular complications after Tf-TAVI is associated with excellent technical success rates, optimal short- to mid-term patency rates as well as satisfactory long-term clinical outcomes. Due to the combination of optimal acute as well as mid- to long-term efficacy and safety, the results of the present study, the largest to evaluate patient clinical outcomes following the use of CBE stents to date, support the use of CBE stents for the management of access-related vascular complications in the setting of TAVI procedures.
Supplementary Information
Below is the link to the electronic supplementary material.
Abbreviations
- CBE
Covered balloon-expandable
- CSE
Covered self-expanding
- CT
Computed tomography
- IVL
Intravascular lithotripsy
- PTA
Percutaneous transluminal angioplasty
- TAVI
Transcatheter aortic valve implantation
- Tf
Transfemoral
- THV
Transcatheter heart valve
- VARC-3
Valve Academic Research Consortium-3
Funding
Open Access funding enabled and organized by Projekt DEAL.
Declarations
Conflict of interest
Hector A. Alvarez-Covarrubias received lecture fees from SIS Medical AG, LifeTech and Edward Lifesciences, proctor fees from LifeTech and financial support for attending meetings or travel expenses from Abbott Vascular and LifeTech, not related to the current work. Salvatore Cassese reports consulting fees from SIS Medical, lecture fees from Abiomed, Astra Zeneca, SIS Medical and Teleflex, and financial support for attending meetings or travel expenses from SIS Medical, not related to the current work. Tobias Rheude received lecture fees from SIS Medical AG and AstraZeneca, not related to the current work. Sebastian Kufner reports speaker and consulting fees from AstraZeneca, Bristol Myers Squibb, and Bentley and speaker fees from Abbott, Boehringer-Ingelheim and Translumina and a research grant from Bentley, not related the current work. Heribert Schunkert reports personal fees from AMGEN, Daiichi-Sankyo, MSD Sharp&Dohme, Astra-Zeneca, Bazer Vital, Boehringer-Ingelheim, Novartis, Servier, Sanofi Aventis and Synlab not related to the current work. Michael Joner reports personal fees from Abbott, Alchimedics S.A.S., Astra Zeneca, Biotronik, Medtronic, Recor, Shockwave, TriCares, Veryan, grants and personal fees from Boston Scientific, Cardiac Dimensions, Edwards, grant from Infraredx outside the submitted work. Erion Xhepa reports consulting services for Edwards Lifesciences, lecture fees and honoraria from AstraZeneca, Boston Scientific, SIS Medical and Translumina not related to the current work; proctor fees from Abbott Vascular and financial support for attending meetings and/or travel expenses from Abbott Vascular and SIS Medical. The remaining authors have no conflicts of interest to declare.
References
- 1.Vahanian A, Beyersdorf F, Praz F, Milojevic M, Baldus S, Bauersachs J et al (2022) 2021 ESC/EACTS guidelines for the management of valvular heart disease. Eur Heart J 43(7):561–632 [DOI] [PubMed] [Google Scholar]
- 2.Otto CM, Nishimura RA, Bonow RO, Carabello BA, Erwin JP 3rd, Gentile F et al (2021) 2020 ACC/AHA guideline for the management of patients with valvular heart disease: a report of the American college of cardiology/American heart association joint committee on clinical practice guidelines. Circulation 143(5):e72–e227 [DOI] [PubMed] [Google Scholar]
- 3.Sherwood MW, Xiang K, Matsouaka R, Li Z, Vemulapalli S, Vora AN et al (2020) Incidence, temporal trends, and associated outcomes of vascular and bleeding complications in patients undergoing transfemoral transcatheter aortic valve replacement: insights from the society of thoracic surgeons/American college of cardiology transcatheter valve therapies registry. Circ Cardiovasc Interv 13(1):e008227 [DOI] [PubMed] [Google Scholar]
- 4.Abdel-Wahab M, Hartung P, Dumpies O, Obradovic D, Wilde J, Majunke N et al (2022) Comparison of a pure plug-based versus a primary suture-based vascular closure device strategy for transfemoral transcatheter aortic valve replacement: the CHOICE-CLOSURE randomized clinical trial. Circulation 145(3):170–183 [DOI] [PubMed] [Google Scholar]
- 5.Rheude T, Ruge H, Altaner N, Pellegrini C, Alvarez Covarrubias H, Mayr NP et al (2024) Comparison of strategies for vascular ACCESS closure after transcatheter aortic valve implantation: the ACCESS-TAVI randomized trial. Eur Heart J. 10.1093/eurheartj/ehae784 [DOI] [PubMed] [Google Scholar]
- 6.Toggweiler S, Leipsic J, Binder RK, Freeman M, Barbanti M, Heijmen RH et al (2013) Management of vascular access in transcatheter aortic valve replacement: part 2: vascular complications. JACC Cardiovasc Interv 6(8):767–776 [DOI] [PubMed] [Google Scholar]
- 7.Meertens M, Wegner M, Fischnaler C, Wienemann H, Macherey S, Lee S et al (2023) Surgical versus interventional treatment of major access site complications during transfemoral TAVI procedures at a large volume center. J Endovasc Ther. 10.1177/15266028231204291 [DOI] [PubMed] [Google Scholar]
- 8.Kufner S, Cassese S, Groha P, Byrne RA, Schunkert H, Kastrati A et al (2015) Covered stents for endovascular repair of iatrogenic injuries of iliac and femoral arteries. Cardiovasc Revasc Med 16(3):156–162 [DOI] [PubMed] [Google Scholar]
- 9.Ben Abdallah I, Urena M, Sutter W, Bezard C, Pellenc Q, Cerceau P et al (2021) Covered stents as a first-line treatment for vascular access complications during transfemoral transcatheter aortic valve implantation: eight-year experience from a single center. Angiology 72(1):70–77 [DOI] [PubMed] [Google Scholar]
- 10.Perrin N, Ellenberger C, Licker M, Hachulla AL, Cikirikcioglu M, Frei A et al (2015) Management of vascular complications following transcatheter aortic valve implantation. Arch Cardiovasc Dis 108(10):491–501 [DOI] [PubMed] [Google Scholar]
- 11.Sedaghat A, Neumann N, Schahab N, Sinning JM, Hammerstingl C, Pingel S et al (2016) Routine endovascular treatment with a stent graft for access-site and access-related vascular injury in transfemoral transcatheter aortic valve implantation. Circ Cardiovasc Interv. 10.1161/CIRCINTERVENTIONS.116.003834 [DOI] [PubMed] [Google Scholar]
- 12.Sedaghat A, Hansen KL, Schahab N, May MC, Weber M, Stundl A et al (2019) Long-term follow-up after stent graft placement for access-site and access-related vascular injury during TAVI-the Bonn-Copenhagen experience. Int J Cardiol 281:42–46 [DOI] [PubMed] [Google Scholar]
- 13.Varc-3 Writing C, Genereux P, Piazza N, Alu MC, Nazif T, Hahn RT et al (2021) Valve Academic Research Consortium 3: updated endpoint definitions for aortic valve clinical research. Eur Heart J 42(19):1825–1857 [DOI] [PubMed] [Google Scholar]
- 14.Sanguettoli F, Pavasini R, Nobile G, Pricoco F, Tasneem S, Biscaglia S et al (2024) “Safe femoral” artery approach in endovascular procedures: a systematic review of its safety and feasibility. Curr Probl Cardiol 49(1 Pt C):102114 [DOI] [PubMed] [Google Scholar]
- 15.Sinclair N, Mordhorst A, Yang GK, MacDonald PS, Sidhu R, Reid JDS (2021) Vascular access complications and clinical outcomes of vascular surgical repairs following transcatheter aortic valve implantation (TAVI). Ann Vasc Surg 74:258–263 [DOI] [PubMed] [Google Scholar]
- 16.Krankenberg H, Zeller T, Ingwersen M, Schmalstieg J, Gissler HM, Nikol S et al (2017) Self-expanding versus balloon-expandable stents for iliac artery occlusive disease: the randomized ICE trial. JACC Cardiovasc Interv 10(16):1694–1704 [DOI] [PubMed] [Google Scholar]
- 17.Choi WG, Rha SW, Choi BG, Park S, Kim JB, Kang DO et al (2024) Balloon-expandable cobalt chromium stent versus self-expandable nitinol stent for the atherosclerotic iliac arterial disease (SENS-ILIAC trial) trial: a randomized controlled trial. Heart Vessels. 10.1007/s00380-024-02431-4 [DOI] [PubMed] [Google Scholar]
- 18.Feldman DN, Armstrong EJ, Aronow HD, Banerjee S, Diaz-Sandoval LJ, Jaff MR et al (2020) SCAI guidelines on device selection in aorto-Iliac arterial interventions. Catheter Cardiovasc Interv 96(4):915–929 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Aboyans V, Ricco JB, Bartelink MEL, Bjorck M, Brodmann M, Cohnert T et al (2018) 2017 ESC guidelines on the diagnosis and treatment of peripheral arterial diseases, in collaboration with the European society for vascular surgery (ESVS): document covering atherosclerotic disease of extracranial carotid and vertebral, mesenteric, renal, upper and lower extremity arteries endorsed by: the European stroke organization (ESO)the task force for the diagnosis and treatment of peripheral arterial diseases of the European society of cardiology (ESC) and of the European society for vascular surgery (ESVS). Eur Heart J 39(9):763–816 [DOI] [PubMed] [Google Scholar]
- 20.Maurina M, Condello F, Mangieri A, Sanz-Sanchez J, Stefanini GG, Bongiovanni D et al (2022) Long term follow-up after balloon expandable covered stents implantation for management of transcatheter aortic valve replacement related vascular access complications. Catheter Cardiovasc Interv 100(5):903–909 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Drouelle E, De Poli F, Couppie P, Uhry S, Heyer H, Morel O et al (2019) Common femoral artery bailout stenting with covered stent graft due to TAVR vascular complication: clinical long term follow-up. Ann Cardiol Angeiol (Paris) 68(5):316–324 [DOI] [PubMed] [Google Scholar]
- 22.Mwipatayi BP, Ouriel K, Anwari T, Wong J, Ducasse E, Panneton JM et al (2020) A systematic review of covered balloon-expandable stents for treating aortoiliac occlusive disease. J Vasc Surg 72(4):1473–86.e2 [DOI] [PubMed] [Google Scholar]
- 23.Voll F, Koch T, Tolg R, Lenz T, Schroeter M, Lenders G et al (2023) Clinical safety and efficacy of new-generation single-layer polytetrafluorethylene covered coronary stents. Cardiovasc Revasc Med 52:30–36 [DOI] [PubMed] [Google Scholar]
- 24.Strauss L, Gibello L, Voll F, Alvarez-Covarrubias HA, Lenz T, Cassese S et al (2024) New-generation single-layer PTFE-covered coronary stent for endovascular repair of iatrogenic arterial side-branch injury in non-coronary lesions for the RECOVER (REsults after percutaneous interventions with COVERed stents) Investigators. Cardiovasc Revasc Med. 10.1016/j.carrev.2024.09.018 [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.