. 2024 Apr 2;89(5):2747–2760. doi: 10.1111/1750-3841.17043

Simple detection of gluten in commercial gluten‐containing samples with a novel nanoflower electrosensor made of molybdenum disulfide with comparison of the ELISA method

Fırat Salman 1, Adem Zengin 2, Hilal Çelik Kazici 2,✉
PMCID: PMC13281137  PMID: 38563096

Abstract

In this study, a new electrochemical sensor based on molybdenum disulfide (MoS2) nanoflowers/glassy carbon electrode (GCE was created for the sensitive detection of gluten. The prepared nanocatalysts were characterized using scanning electron microscopy with energy dispersive spectroscopy, x‐ray diffraction, and x‐ray photoelectron spectroscopy. The effects of the prepared nanocatalysts, pH value, and dropping amounts on the results were examined in detail. The electrochemical performance of the developed sensor (MoS2 nanoflowers/GCE) was then evaluated using differential pulse voltammetry, and the sensor was found to have significant electrochemical activity against gluten. A substantial linear connection was observed in the range of 0.5–100 ppm of gluten concentration under optimum experimental circumstances, and the detection limit between peak current and gluten concentration was determined as 1.16 ppm. The findings showed that the MoS2 nanoflowers/GCE gluten sensor has exceptional selectivity and stability. Finally, the generated electrochemical sensor was effectively utilized for gluten detection in commercial gluten‐containing materials with a detection limit of 0.1652 ppm. Thus, the developed MoS2 nanoflowers/GCE sensor offers a potential method for the detection of other molecules and is a promising candidate for gluten detection in commercial samples.

Keywords: differential pulse voltammetry, electrochemical sensor, gluten, MoS2 nanoflowers

1. INTRODUCTION

Celiac disease (CD) is a multi‐systemic autoimmune disorder caused by gluten protein components present in grains such as wheat, oats, rye, and barley in genetically predisposed individuals. The symptoms of this disease were first described in 1887 as growth retardation, diarrhea, and exhaustion. In 1953, it was determined that grains such as wheat, oats, rye, and barley cause celiac disease (Van Bergeijk et al., 1993). Gluten proteins (which can be found in wheat, rye, and barley) that cause celiac disease are classified as gluten, gliadin, hordein, and secalin (Niewinski, 2008). Proteases in the upper section of the digestive tract are unable to completely break down gluten proteins, thus staying intact in the intestinal lumen. In celiac patients, the enzyme transglutaminase deaminates gluten protein fractions in the upper small intestine. This structure causes villous damage, as well as the release of other tissue‐damaging factors. As a result, the surface epithelium is harmed, and the villi (sucking feathers) become flattened (Bethune & Khosla, 2008; Catassi & Fasano, 2008). When villi are damaged, the absorption and digestion of foods cannot be properly completed (malabsorption).

For celiac and other gluten‐related diseases, the only treatment method is a gluten‐free diet. The exact amount of gluten that individuals can tolerate can vary. However, even today, gluten cannot be completely eliminated in products labeled as “gluten‐free.” For this reason, it is very important to determine the gluten limits in gluten‐free products so that they do not cause harmful effects on celiac patients (Walter et al., 2014). The safe limit that will not cause possible discomfort is a daily gluten intake of less than 10 mg (Catassi et al., 2007). The Turkish Food Codex Communiqué on meals suitable for people with gluten intolerance (Communiqué No: 2012/4) states that 20 ppm of gluten is the maximum amount that can be present in gluten‐free meals (Eksin et al., 2015). The only reason the applied treatment fails is that gluten is not completely removed from the diet. Failure to fully comply with this treatment causes an increase in the morbidity and mortality rates (Biagi et al., 2012; Niewinski, 2008). Patients who have not been diagnosed, or who do not comply with a gluten‐free diet are prone to many complications (osteoporosis, infertility, autoimmune diseases, etc.). The rise in cancer mortality among celiac disease patients and their family is proof to the condition's clinical importance (Brousse & Meijer, 2005). To guarantee that celiac patients’ diets are safe, it is crucial to assess the purity of gluten‐free items. Verification of specificity, reproducibility, and reliability are necessary for a viable analytical approach to measure the gluten levels in foods (Scherf & Poms, 2016). For this purpose, certain analytical methods for detecting gluten were developed and are documented in the literature. Enzyme‐linked immunosorbent assays (ELISA) based on R5 and G12 antibodies are advised under the present regulations (Alvarez et al., 2013). ELISAs are widely used because of their sensitivity and specificity and their suitability for continuous analysis, as well as the absence of any other method that can be considered as an independent reference for gluten analysis. The development of novel immunosensors using techniques like aptamers, microarrays, and multi‐analyte profiling can offer an alternative to ELISA and several other analytical techniques (such as proteomics, mass spectrometry, and genomics). Some criteria in analytical methodologies are sought for the accurate measurement of gluten levels in foods. These criteria must have a high enough sensitivity, specificity, repeatability, robustness, and ability to be verified by experiments in multiple independent laboratories (Slot et al., 2015).

The sensitivity limit that must be met in the currently used analytical methods in gluten‐free product analysis should be 20 mg gluten/kg as reported by the Codex. Many of the ELISAs used are frequently preferred due to their specific structure, high sensitivity, and allowing routine analysis (Amigo & Popping, 2012). ELISAs continue to have several limitations, however, including influences from food matrix and processing history and unpredictability of the findings in current ELISA test kits for diverse assays, despite the introduction of novel specific antibodies for use in gluten analysis and several enhanced extraction procedures. An ideal combination of extraction operation and testing equipment should consider the challenges and factors specific to highly heat‐treated and hydrolyzed gluten. Polymerase Chain Reaction (PCR)‐based tests have also been known to have these complications (Gomaa & Boye, 2015; Sandberg et al., 2003). DNA‐based approaches can potentially provide better findings than ELISAs in terms of sensitivity and specificity, however they are not suited for gluten detection in severely processed or hydrolyzed materials with significant DNA degradation (Scharf et al., 2013). The most promising methods for accurately quantifying residues of gluten to assure the safety of gluten‐free diets for CD patients are liquid chromatography–mass spectrometry (LC–MS)/mass spectrometry (MS) separation methods because of their high selectivity, sensitivity, adaptability, and extensive application for heated and hydrolyzed gluten (Fiedler et al., 2014; Haraszi et al., 2011). However, due to the high cost of the equipment and the need for specialists, these methods are not commonly used in conventional analysis, but are instead used in specialized laboratories and research institutions. New methods aim to improve gluten analysis with simple, rapid, and in situ detection procedures that provide lower cross‐reactivity as well as better sensitivity with related species (Scherf & Poms, 2016). Electrochemical techniques are inexpensive, portable, easy to operate, and fast; therefore, they are often preferred over the other analytical techniques and have been shown to be useful tools for food detection (Amrutha et al., 2019; Edwin et al., 2021; Eksin et al., 2015; Manjunatha et al., 2009; Prinith et al., 2019; Raril et al., 2018).

In order to accurately and selectively detect gluten, Erdem et al. developed an electrochemical disposable pencil graphite electrode (PGE) in conjunction with the differential pulse voltammetry (DPV) technique. The linear concentration range for gluten was determined as between 20 and 100 µg/mL under ideal experimental circumstances, and the detection limit was 7.11 µg/mL (Eksin et al., 2015). Recently, a promising method for finding gluten in flours was reported, combining surface‐modified nanoparticles with electrochemical detection. So, utilizing the molecularly imprinted polymers (MIP) method, which enables the construction of a complex polymeric cavity with functional monomers using a template, a gluten‐sensitive receptor was developed. In this study, magnetic MIPs were employed to allow an external magnetic field to analytically detect an electrochemical signal. A detection limit of 8.5 mg/kg was determined after testing several wheat samples for gluten‐magnetic molecularly imprinted polymers (MMIP) (Limthin et al., 2019).

Electrodes play a very important role in electroanalytical techniques as they provide the interface between the analyte and the electrochemical cell. Modified electrodes have been applied in various fields such as energy production and analytical chemistry. Modified electrodes have been particularly successful in the electrochemical determination of trace levels of amino acids, peptides, proteins, alcohols, and sugars, but have also been used for the study of inorganic ions. Electrodes modified with nanomaterials have significantly increased the sensitivity of electroanalytical techniques because nanomaterials support the electrocatalysis process and increase the surface area of the electrode (Cox et al., 1996).

In this study, MoS2 nanoparticles, a promising two‐dimensional material with a wide range of potential applications due to its superior electrochemical, mechanical, and optical characteristics (Pietrzyk et al., 2018; Tian et al., 2014; Zhou et al., 2016), were synthesized using chemical synthesis. Detailed characterization (x‐ray diffraction [XRD], scanning electron microscopy [SEM], and x‐ray photoelectron spectroscopy [XPS]) was performed and the developed modified MoS2 nanoflowers/GCE electrode was subjected to the gluten test with electrochemical measurements and DPV (Tigari et al., 2020). Finally, MoS2 nanoflowers/GCE was successfully applied to a commercial sample and method validation was achieved with the ELISA method.

2. MATERIALS AND METHODS

2.1. Reagents and instrumentation

Sigma‐Aldrich supplied the high‐purity sodium molybdate dihydrate (Na2MoO4·2H2O, ≥99 %) as a source for molybdenum and the l‐cysteine (C3H7NO2S, 97 %) as a source for sulfur, which were both used in the preparation of the MoS2 nanoparticles. Additionally, gluten, ethanol, and deionized water, among other things, were purchased from Sigma‐Aldrich (St. Louis, MO, USA). All reagents were used without further purification.

Gluten ELISA kits were also acquired from Sigma‐Aldrich and the assay was carried out according to the manufacturer guidelines.

The crystal structure was observed with XRD (PANalytical, the Netherlands). The morphology of MoS2 was observed with a scanning electron microscopy with energy dispersive spectroscopy (SEM‐EDS, ZEISS GeminiSEM) and XPS (SPECS). Electrochemical measurements were made using a BioLogic SP‐50 potentiostat (BioLogic Company France).

2.2. Synthesis of MoS2 nanoparticles and electrochemical measurements

MoS2 nanoflowers were synthesized with a one‐step hydrothermal method (Liang et al., 2018). Briefly, 50 mg of Na2MoO4·2H2O and 100 mg of l‐cysteine were added to 25 mL of distilled water and stirred for 30 min at room temperature to form a homogenous suspension. A stainless‐steel autoclave coated with Teflon was then used to heat the suspension for 24 h at 220°C. The black material was then separated using centrifugation, washed with plenty of water and ethanol, and allowed to dry at 70°C overnight.

To form a catalyst ink, 5 mg of catalyst was dispersed in 1 mL of water–ethanol solution (volume ratio of 1:1) with 20 µL of Nafion solution (5 wt%) to form a homogeneous ink by sonication for 5 min. Before each experiment, the GCE was refined with slurries of smaller alumina sizes (1, 0.3, and 0.05 µm) followed by sonication for 5 min in purified water. The obtained modified MoS2 nanoflowers/GCE was tested with the DPV method by applying 20 mV pulse width conditions with 0 to 1.2 V scanning range and 50 mV/s scan rate for gluten detection by electrochemical analysis. For DPV measurements, a glassy carbon electrode with a diameter of 2 mm as a working electrode and a Pt wire and an Ag | AgCl electrode was used as reference electrode in 3 M KCl. In this study, determination of gluten was tested in vinegar and method validation was carried out with the ELISA method.

3. RESULTS AND DISCUSSION

3.1. Characterization

The structure and morphology of MoS2 nanoflowers were investigated using SEM and the micrographs of the sample are shown in Figure 1a. The samples have mean diameters of 230 nm and the high magnification SEM image showed that the nanoflowers comprise thin nanosheets with a mean diameter of 18 nm. In addition, SEM‐elemental mapping of nanoflowers showed a homogenous distribution of Mo and S elements and also recorded the same elements as the energy dispersive spectroscopy (EDS) spectrum of the sample (Figure 1b) indicating the successful synthesis of MoS2 nanoflowers.

FIGURE 1.

FIGURE 1

(a) Scanning electron microscopy (SEM) micrograph of MoS2 nanoflowers with low (left) and high (right) magnifications and (b) SEM‐elemental mapping of Mo and S elements and energy dispersive spectroscopy spectrum of MoS2 nanoflowers.

XRD was used to determine the crystallinity and crystal structure of MoS2 nanoflowers. The XRD pattern of the MoS2 nanoflowers (Figure 2) shows the main diffraction peaks recorded at 14.3°, 33.6°, 43.1°, and 58.6 °, which can be indexed as (002), (100), (103), and (110), respectively, which matched well with the hexagonal phase of MoS2 (JCPDS: 98‐011‐8125) (Liu et al., 2019).

FIGURE 2.

FIGURE 2

X‐ray diffraction pattern of MoS2 nanoflowers.

The chemical composition and oxidation states of the elements in MoS2 nanoflowers were determined by XPS. As displayed in Figure 3a, in the survey scan of MoS2 nanoflowers, the appearance of Mo and S clearly confirmed the presence of Mo and S. The appearance of weak C 1s can be attributed to the adsorption of carbonaceous contaminants during the XPS operation (Pu et al., 2015). In the core‐level spectra of Mo 3d (Figure 3b), the presence of Mo 3d doublet at about 228.1 eV (3d 5/2) and 231.2 eV (3d 3/2) indicate the typical 4+ oxidation state of Mo (Chang et al., 2013). In addition, the two different peaks of Mo6+ (231.9 eV for 3d 5/2 and 235.1 eV for 3d 3/2) were also observed, implying slight oxidation of MoS2 (Hai et al., 2016). In the core‐level spectra of S 2p (Figure 3c), distinct peaks were observed at 161.5 eV (S 2p 3/2) and 162.8 eV (S 2p 1/2), respectively, which supported S2− binding energies in 2H‐MoS2 (Oakes et al., 2016). Moreover, the core level of O 1s spectra (Figure 3d) can be deconvoluted into two peaks at 531.7 and 532.9 eV, attributed to O2− species which are most probably due to the absorption of substances containing oxygen, such as water (Huang et al., 2020).

FIGURE 3.

FIGURE 3

(a) X‐ray photoelectron spectroscopy (XPS) survey scan of MoS2 nanoflowers and XPS core‐level spectra of MoS2 nanoflowers for (b) Mo 3d, (c) S 2p, and (d) O 1s.

3.2. Electrochemical characterization of gluten

The thickness of a flat catalyst layer on a solid substrate determines its diffusion state (Du et al., 2017). Therefore, to investigate the impact of diffusion limitation in catalysts, electrodes must be produced with various catalyst layer thicknesses. The MoS2 catalyst slurry was obtained from 5 mg MoS2 and 1 mL (Nafion + ethanol). The effect of catalyst loading on the electrode (MoS2 nanoflowers/GCE) surface in the presence of 20 µg/mL (ppm) gluten dissolved in dimethyl sulfoxide (DMSO) in the medium was transferred to 1, 2, 3, 4, and 5 µL and examined in 0.1 M phosphate buffered saline (PBS) buffer (pH 4.5). The gluten detection performance of the catalyst loading was studied using the DPV method at 20 mV pulse width and 50 mV/s scan rate (Figure 4a) and the results are plotted on a bar graph (Figure 4b). Instead of cyclic voltammetry, DPV was used, as this method offers better sensitivity and signal‐to‐noise ratio (Salman et al., 2020). At 1–2 µL thickness, the gluten current is slightly increased (from 15 to 16 µA). With a thickness of 3 µL, the detection current increased to 19 µA. In addition, the above selectivity was reduced to 10 µA in the case of a thicker catalyst layer (4–5 µL).

FIGURE 4.

FIGURE 4

(a) Differential pulse voltammetry of the MoS2 loading amount on the GCE surface. (b) Column graph of the catalyst loading amount (reaction conditions; pulse width: 20 mV, scan rate: 50 mV/s, 20 ppm gluten, 5 mg MoS2 in 0.1 M PBS, pH 4.5 at room temperature).

The scan rate controls how fast the applied potential is scanned, and faster scan rates lead to a reduction in the size of the diffusion layer, resulting in higher currents being observed.

In our study, the active surface area of the electrode produced using the cyclic voltammetry (CV approach was determined. For electrochemically reversible electron transfer processes involving freely diffusing redox species, the Randles–Sevcik equation (Equation 1) describes how the peak current ip (A) increases linearly with the square root of the scan rate υ (V/s).

For this purpose, the electrochemical active surface area (A) in a 5 mM [Fe(CN)6]3−/4− solution containing 0.1 M KCl was measured by CV tests at different scanning speeds.

As shown in Figure 5a, both anode and cathode peak currents increase with the increase of scanning speed, indicating that the electrode reaction of the process is diffusion controlled. In a study examining the methane selectivity with cobalt‐based monolithic catalysts in the literature, monolithic catalysts with layers greater than 50 µm were reported to exhibit a significantly reduced methane selectivity and were observed to suffer from diffusion limitations (Guettel et al., 2008).

ip=(2.69×105)n3/2D1/2A·1/2c (1)

where “c” is the concentration of K3[Fe(CN)6] (c = 5.0 mM) and “A“ represents the electrochemically active substance area, “n” is the number of electrons, “ʋ” is the scan rate, and “D” is the diffusion coefficient (6.7 × 106 cm2/s1). In Figure 5, the electrochemical active surface area of MoS2 nanoflowers/GCE is 0.082 cm2 (0.053 cm2), which is almost twice that of the bare electrode. It is responsible for the enhanced detection sensitivity.

FIGURE 5.

FIGURE 5

(a) CV of MoS2/GCE in 5 mM [Fe(CN)6]3−/4− containing 0.1 M KCl at different scan rates. (b)The linear relationship between the positive and negative peak current and the square root of the scanning rate. (c) The relationship between the positive and negative potential of the peak and the logarithm of the scanning rate.

Electron transfer between the electrolyte and electrode surfaces was elucidated by the electrochemical impedance spectroscopy (EIS) method. A typical EIS spectrum contains a higher frequency semicircle and a lower frequency linear region. The semicircular part corresponds to the electron transfer control process in the spectrum, and the linear part corresponds to the diffusion control process. The diameter of the semicircle is equivalent to the electron transfer resistance value (Rct) of the electrode surface, and generally Rct is smaller and the interfacial electron transfer rate is higher for the smaller semicircle diameter. The EIS results for the bare electrode and MoS2/GCE are shown in Figure 6. MoS2/GCE showed the smallest Rct, and this finding indicates that the electrode fabricated with MoS2 has a strong electron conduction path between the electrode and the electrolyte.

FIGURE 6.

FIGURE 6

Electrochemical impedance spectroscopy spectra recorded on bare GCE and MoS2/GCE in 5.0 mM [Fe (CN)6]3−/4− solution. Scan rate: 50 mV/s.

With the purpose of further understanding the electrochemical reaction kinetics of the sensor, the charge transfer coefficient (α) was calculated using Laviron's formula (Equation 2):

lgKaKc=lgα1−α (2)

In the above equations, Kα and Kc are the slopes of the linear relationship between the peak potential and the logarithm of the scan rate in Figure 5c. According to theoretical calculation, α is 0.99. The higher electron mobility proves the outstanding electrical conductivity and excellent sensing potential of MoS2/GCE.

In electroanalysis, electrode–solution coupling is of vital importance because the maintenance of a diffusion curve causes analytes to migrate from the entire solution approach to the electrode surface in the absence of applied voltage, and the concentration of the analyte across the surface of the electrode can influence a number of electrochemical variables (Kazici et al., 2019).

In the current study, while examining the effect of accumulation times between 0 and 70 s at various time points to investigate the effect of concentration in the region, it was concluded that analyte saturation was reached at this time, as 10 s provided the highest peak current. This optimized time tactic was applied in subsequent trials (Figure 7).

FIGURE 7.

FIGURE 7

Amount of accumulating time's effect on the 20 ppm Glu electrochemical study at a scan rate of 50 mV/s at various time intervals between 0 and 70 s.

The current team kept working on improving the pH of the supporting electrolyte and developing an analytical method for gluten detection. The effect of pH values of working solutions was investigated in the range from 3 to 10.0 using 0.1 M PBS solution. Experimental results are given by the DPV method (Figure 8a) and bar graph (Figure 8b). In the present study, a significant decrease in gluten detection current was recorded after pH 4.5. Gluten is a protein family found in wheat, rye, oats, and barley, which can cause health problems in people who are sensitive to it (Scherf & Poms, 2016). While mild pH (4.5) treatment provides high detection current by causing higher protein solubility, higher pH (10) may have reduced the detection current by causing damage to the structure and denaturation of proteins (Shan et al., 2012). In addition, the pH value was examined in the range of 3–10, as seen in Figure 8a,b, but the range of 4.5–10 was examined to examine the linear change of its change with potential. It is obvious that the current decreases at lower pH values. As pH increases, the peak potential (Epa) decreases proportionally and becomes linear as shown in Figure 8c. This is attributed to protons participating in the electrochemical reaction process.

FIGURE 8.

FIGURE 8

(a) Differential pulse voltammetry gluten detection on the MoS2 nanoflowers/GCE surface in different pH solutions. (b) Column graph of pH. (c) The peak potential versus pH value (reaction conditions; pulse width: 20 mV, scan rate: 50 mV/s, 20 ppm, 5 mg MoS2, 3 µL MoS2 slurry on GCE in 0.1 M PBS, pH 3–10 at room temperature).

The linear regression equation between Epa and pH is Epa (V) = 1.0006 – 0.072 pH (R 2 = 0.98); from here the absolute value of the slope (0.072 V/pH) is close to the theoretical Nernstian value: 0.059 V/pH means a fairly equal number of electrons and protons involved in the redox reaction process.

In electroanalysis, electrode–solution coupling is of vital importance because the maintenance of a diffusion curve causes analytes to migrate from the entire solution approach to the electrode surface in the absence of applied voltage, and the concentration of the analyte across the surface of the electrode can influence a number of electrochemical variables (Shufang et al., 2022).

3.3. DPV‐based non‐enzymatic gluten detection by using the MoS2 nanoflowers/GCE electrode

To evaluate the electroanalytical efficacy of the MoS2 nanoflowers/GCE, gluten detection was carried out under the optimized conditions mentioned above. Again, the DPV method was used, as it is a useful tool for the swift and precise detection of analytes in electroanalytical chemistry due to structural high current sensitivity and low charge contribution to background current (Li et al., 2019).

Figure 9a depicts the DPV curves of MoS2 nanoflowers/GCE for determining different concentrations of gluten in PBS solution (0.1 M, pH = 4.5). Gluten concentrations were 0.5, 1, 5, 10, 20, 40, 60, 80, and 100 ppm, respectively. The scan range was between 0 and 1.2 V, amplitude 50 mV, pulse duration 0.5 s, and potential increase 4 mV. As shown in the figure, when the analyte concentration increases, the peak value of the response currents increases, indicating that an efficient electron transfer takes place between the gluten and modified electrode (MoS2 nanoflowers/GCE). The oxidation peak potentials of gluten on these MoS2 nanoflowers/GCE sensor fabricated at various gluten concentrations were stable at around 0.6 V. The linear relationship between gluten (C) concentration and response currents (I) is displayed in Figure 7b. I(μA)=0.75346Cgluten(ppm)+0.74086,R2=0.99 was the linear equation obtained for 0.5–100 ppm. The quantity limit (LOQ) and the detection limit (LOD) of the current assay were computed as follows: detection limit (LOD) = 3σ/S and quantity limit (LOQ) = 10σ/S; where σ is the standard deviation and S is the slope of the calibration curve. LOD and LOQ values were found as 0.16 and 0.57 ppm, respectively. According to the Turkish Food Codex regulation, the gluten LOD is substantially lower (less than 20 ppm) than the permissible amount for gluten‐free branded items (Eksin et al., 2015). In terms of linear range and LOD, the suggested sensor's responses were also contrasted with those of the previous gluten sensors. The findings are given in Table 1. The sensor in this study has the lowest detection limit among the other electrochemical gluten sensors, a satisfactory linear range among highly limited electrochemical gluten sensors in the literature, as well as superior properties in comparison to the other expensive methods in terms of cheapness and ease of use.

FIGURE 9.

FIGURE 9

(a) Differential pulse voltammetry curves of the MoS2 nanoflowers/GCE electrodes fabricated with different gluten concentrations (from 0.5 to 100 ppm) in pH 4.5 PBS. (b) Relationship between the peak currents and concentrations.

TABLE 1.

Overview of gluten analysis methods designed to detect gluten values of 20 mg/kg or less, sorted first by method type and then by linear volume range and sensitivity and reference number.

Method Linear range Sensitivity Ref.
ELISA – Antibodies; G12 (LOQ: 2.9 ± 0.3 µg/mL), 2D4 (LOQ: 2.8 ± 0.2 µg/mL), Skerritt (LOQ: 0.85 ± 0.04 µg/mL) Panda et al. (2017)
ELISA – LOQ: 5 mg/kg of gluten Lacorn et al. (2019)
Genomics, PCR – LOD: 0.9 mg/kg of gluten Ahmed and Meng (2019)
Genomics, PCR – LOD: 1–5 mg/kg of gluten García‐García et al. (2019)
Sensor, aptamers (GE‐Cys‐PAMAM‐Gli1) – LOD: 5 mg/kg of gluten Malvano et al. (2017)
Sensor, aptamers, and G12 and barley antibodies – LOD: 0.1–1.0 µg/mL of gluten White et al. (2018)
DNA aptamers 0.075–100 µM LOD: 0.5 µg/mL Amaya‐Gonzalez et al. (2014)
Iridium complex Phosphorescent Probe 5–200 µg/mL LOD: 2.6 µg/mL Huang et al. (2018)

Gold nanoparticles‐light

Scattering

1.5–15 µg/mL LOD: 0.8 µg/mL Molina‐Delgado et al. (2011)
Sensor, MMIP 20–1000 µg/mL LOD: 8.5 mg/kg of gluten‐ MMIP Limthin et al. (2019)
Sensor, PGE 20–100 µg/mL LOD: 7.11 µg/mL of gluten Eksin et al. (2015)
Sensor, MoS2 nanoflowers/GCE 0.5–100 µg/mL LOD:0.16 and LOQ:0.57 µg/mL of gluten This study

Abbreviations: GE, gold electrode; MMIP, magnetic molecularly imprinted polymers; PGE, pencil graphite electrode.

The sensitivity of the sensor was evaluated as 17.92 µA mL/µg cm2 (slope divided by the electrode surface area).

Recently, different ELISA and enzyme‐linked oligonucleotide assay (ELONA) formats that may identify gluten proteins have been developed to test the gluten content of foods (Arendt et al., 2008). All sensors have molecules (such as antibodies–aptamers) that act as receptors fixed on the surface of the substrate. Following antigens binding, an enzymatic process occurs that produces fluorescent or electrochemical signals. For the amperometric sensor, a gold‐layered working electrode, Ag/AgCl‐layered reference electrode, and counter electrode were attached to a chip surface. With this study, we can report that the functionality of these chip systems is promising.

3.4. Specificity study

In addition, interference experiments were also conducted to determine the effect of impurities such as organic acids on the current density of gluten (Figure 10). According to the literature, gluten can be oxidized simultaneously with some substances (such acetic acid [Ac], uric acid [UA], ascorbic acid [AA], dopamine [DA], citric acid [CA], benzoic acid [BA], dextrose [DX], and lactose [LC]) and contribute to the interference signals (Su et al., 2011). Figure 10 presents the selectivity test results for 10 ppm of gluten with 10 ppm UA, AA, CA, Ac, BA, DA, DX, and LC, respectively. It was found that all these elements produced insignificant current responses and therefore, the MoS2 nanoflowers/GCE had excellent specificity and anti‐interference ability for some interfering species.

FIGURE 10.

FIGURE 10

Interfering effect of 10 ppm (uric acid, ascorbic acid, citric acid, acetic acid, benzoic acid, and dopamine) on the bar graph of differential pulse voltammetry response of MoS2 nanoflowers/GCE upon the successive addition of 10 ppm of gluten in 0.1 mM PBS, pH 4.5.

3.5. Reproducibility, stability, repeatability, and durability assay

Reproducibility was also investigated as shown in Figure 11a. Six sensors prepared in the same batch were examined in pH 4.5 PBS individually. The relative standard deviation (RSD) of the response currents was 2.03%, demonstrating that the improved sensor showed excellent performance in terms of reproducibility.

FIGURE 11.

FIGURE 11

(a) The reproducibility of MoS2 nanoflowers/GCE electrode for gluten detection. (b) Peak current of CV versus storing time for MoS2 nanoflowers/GCE in 20 ppm of gluten in 0.1 mM PBS, pH 4.5 at 50 mV/s. (c) Differential pulse voltammetry peak current versus number of assay in five 10 ppm of gluten in 0.1 mM PBS, pH 4.5 at 50 mV/s using one MoS2 nanoflowers/GCE electrode.

In addition, stability of the MoS2 nanoflowers/GCE sensor was investigated to determine the activity of the sensor in practical use (Figure 11b). The fabricated electrode was stored in an airtight container for 30 days to ensure optimum reliability and stability over time at 4°C for, and as a result of the measurement at the end of the 12th day, 94% remained and at the end of the 30th day, 83.8% remained. This indicates that the prepared MoS2 nanoflowers/GCE sensor has good stability (Figure 11b).

A daily evaluation was performed to verify that measurements were repeatable and prepared MoS2 nanoflowers/GCE electrode was used for four parallel detections, and the relative standard deviation RSD was 1.54%, indicating that the repeatability of the MoS2 nanoflowers / GCE sensor was very good (Figure 11c)

In addition, the durability of the sensing system was examined by measuring the electrochemical response of gluten at the modified electrode in the presence of 20 ppm gluten for 15 cycles (Figure 12a). From these consecutive cyclic runs, the % RSD of peak current was calculated as 4.32%. Up to the first 10 cycles, the current change was 5%. The current change percentage of the developed electrode decreased to 1.5% after 10 cycles and stabilized (Figure 12b).

FIGURE 12.

FIGURE 12

 (a) Effect of the number of cycles on the differential pulse voltammetry gluten detection on the MoS2 nanoflowers/GCE surface. (b) Column graph of cycle number against the percentage of current change (reaction conditions; pulse width: 20 mV, scan rate: 50 mV/s, 20 ppm gluten, 5 mg MoS2, 3 µL MoS2 slurry on GCE in 0.1 M PBS, pH 4.5 at room temperature).

3.6. Determination of gluten in commercial sample

To further investigate the practical application of the developed gluten sensor, different gluten concentrations were detected in vinegar samples. The reliability of the produced electrode was examined by a recovery test and for this purpose, the resulting vinegar samples had gluten added for detection in 0.01 M PBS (pH = 4.5) medium. Each sample was tested in triplicate and the average recoveries are given in Table 2. The results demonstrate that the gluten concentration is very close to the concentration of the gluten standard solution in the real sample and the recovery range and RSD values are 98%–99.7% and 4.13%–5.03%, respectively, with the proposed method. To confirm the amount of gluten with the proposed test, the gluten concentration in the vinegar sample described above was obtained using ELISA analysis (Panda et al., 2017). The gluten concentration obtained in the real sample showed good agreement between the recommended and ELISA methods, for both recovery (99.5%–100.5%) and standard error (2.10%–3.64%) values. Recovery rates and RSD values indicate that the proposed sensor (MoS2 nanoflowers/GCE) may be a promising candidate for quantitative detection of gluten.

TABLE 2.

Comparison of the proposed method and enzyme‐containing immunosorbent assay method for recovery of gluten in spiked vinegar samples (n = 3) and dilution factor (1:10).

Proposed method ELISA method
Added gluten (µg/mL) Found (µg/mL) Recovery a (%) RSD% Found (µg/mL) Recovery a (%) RSD%
0 – – – – – –
2 1.96 ± 0.081 98.0 4.13 1.99 ± 0.055 99.5 2.76
4 3.94 ± 0.178 98.5 4.52 4.02 ± 0.121 100.5 2.76
6 5.94 ± 0.299 99.0 5.03 6.01 ± 0.211 100.2 3.51
8 7.97 ± 0.336 99.6 4.22 8.02 ± 0.292 100.3 3.64
10 9.97 ± 0.479 99.7 4.80 10.1 ± 0.212 100.1 2.10
a

Recovery (%) = [(Total)—(the measured concentration in real sample)] / spiked × 100.

4. CONCLUSIONS

Herein, we present a simple and precise enzyme‐free electrochemical gluten sensor based on the GCE electrode modified with MoS2 nanoflower particles. First, we performed the surface characterization of MoS2 nanoflower particles, and then we completed various optimization studies (such as the number of droplets on the surface and pH) for gluten detection using catalysts obtained by dropping these particles onto the GCE surface with an electrochemical technique. The application of gluten determination was tested with the DPV method, and the detection limit was obtained as 0.16 ppm which is in accordance with the official limits described in the Codex Standard. Also, this technique is highly selective with a wide concentration range of 0.5–100 ppm. In addition, the ability of this sensor to detect a mixture was studied, and to this end, different interventions, including dopamine, show that the sensor's ability to detect gluten in complex matrices is excellent. This electrochemical detection protocol provides fast, inexpensive, direct, sensitive, and selective determination of gluten without any external separation or using expensive biomolecules such as antibodies, DNA, or aptamers. This technique is a candidate to be a potential approach for other molecular detection methods.

AUTHOR CONTRIBUTIONS

Fırat Salman: Formal analysis; data curation; validation. Adem Zengin: Conceptualization; writing—review and editing; investigation; supervision. Hilal Çelik Kazici: Conceptualization; investigation; funding acquisition; writing—original draft; methodology; writing—review and editing; supervision.

CONFLICT OF INTEREST STATEMENT

The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.

ACKNOWLEDGMENTS

We would like to thank Prof. Dr. Eylem Turan (Ankara Medipol University Faculty of Pharmacy), Assoc. Dr. Mustafa Bilici (Van Yüzüncü Yıl University, Vocational School of Security), and Burçin Taneri for their comments and contributions that improved the article.

REFERENCES

  1. Ahmed, N. , & Meng, M. (2019). Detection of gluten‐rich cereals in processed foods with enhanced sensitivity by targeting mitochondrial DNA using PCR. Food Analytical Methods, 12(3), 811–825. [Google Scholar]
  2. Alvarez, P. A. , Mongeon, V. J. , & Boye, J. I. (2013). Characterization of a gluten reference material: Wheat‐contaminated oats. Journal of Cereal Science, 57(3), 418–423. [Google Scholar]
  3. Amaya‐Gonzalez, S. , de‐Los‐Santos‐Álvarez, N. , Miranda‐Ordieres, A. J. , & Lobo‐Castanon, M. J. (2014). Aptamer binding to celiac disease‐triggering hydrophobic proteins: A sensitive gluten detection approach. Analytical Chemistry, 86(5), 2733–2739. [DOI] [PubMed] [Google Scholar]
  4. Amigo, C. D. , & Popping, B. (2012). Labeling regulations, detection methods, and assay validation. Journal of AOAC International, 95(2), 337–348. [DOI] [PubMed] [Google Scholar]
  5. Amruthaa, B. M. , Manjunathaa, J. G. , Bhattb, A. S. , Rarila, C. , & Pushpanjal, P. A. (2019). Electrochemical sensor for the determination of alizarin red‐S at non‐ionic surfactant modified carbon nanotube paste electrode. Physical Chemistry Research, 7(3), 523–533. [Google Scholar]
  6. Arendt, E. K. , Morrissey, A. , Moore, M. M. , & Dal Bello, F. (2008). Gluten‐free breads. In Arendt E. K. & Bello F. (Eds.), Gluten‐free cereal products and beverages. pp. 289–319. Oxford, UK: Academic Print Elsevier. [Google Scholar]
  7. Bethune, M. T. , & Khosla, C. (2008). Parallels between pathogens and gluten peptides in celiac sprue. PLoS Pathogens, 4(2), e34. [DOI] [PMC free article] [PubMed] [Google Scholar]
  8. Biagi, F. , Bianchi, P. I. , Marchese, A. , Trotta, L. , Vattiato, C. , Balduzzi, D. , Brusco, G. , Andrealli, A. , Cisarò, F. , Astegiano, M. , Pellegrino, S. , Magazzù, G. , Klersy, C. , & Corazza, G. R. (2012). A score that verifies adherence to a gluten‐free diet: A cross‐sectional, multicentre validation in real clinical life. British Journal of Nutrition, 108(10), 1884–1888. [DOI] [PubMed] [Google Scholar]
  9. Brousse, N. , & Meijer, J. (2005). Malignant complications of coeliac disease. Best Practice & Research Clinical Gastroenterology, 19(3), 401–412. [DOI] [PubMed] [Google Scholar]
  10. Catassi, C. , Fabiani, E. , Iacono, G. , D'Agate, C. , Francavilla, R. , Biagi, F. , Volta, U. , Accomando, S. , Picarelli, A. , De Vitis, I. , Pianelli, G. , Gesuita, R. , Carle, F. , Mandolesi, A. , Bearzi, I. , & Fasano, A. (2007). A prospective, double‐blind, placebo‐controlled trial to establish a safe gluten threshold for patients with celiac disease. The American Journal of Clinical Nutrition, 85(1), 160–166. [DOI] [PubMed] [Google Scholar]
  11. Catassi, C. , & Fasano, A. (2008). Celiac disease. Current Opinion in Gastroenterology, 24(6), 687–691. 10.1097/MOG.0b013e32830edc1e [DOI] [PubMed] [Google Scholar]
  12. Chang, Y. H. , Lin, C. T. , Chen, T. Y. , Hsu, C. L. , Lee, Y. H. , Zhang, W. , & Li, L. J. (2013). Highly efficient electrocatalytic hydrogen production by MoSx grown on graphene‐protected 3D Ni foams. Advanced Materials, 25(5), 755–755. [DOI] [PubMed] [Google Scholar]
  13. Cox, J. A. , Tess, M. E. , & Cummings, T. E. (1996). Electroanalytical methods based on modified electrodes: A review of recent advances. Reviews in Analytical Chemistry, 15, 173–223. [Google Scholar]
  14. Du, Z. , Wood, D. L. , Daniel, C. , Kalnaus, S. , & Li, J. (2017). Understanding limiting factors in thick electrode performance as applied to high energy density Li‐ion batteries. Journal of Applied Electrochemistry, 47(3), 405–415. [Google Scholar]
  15. Edwin, D. S. S. , Manjunatha, J. G. , Raril, C. , Girish, T. , Ravishankar, D. K. , & Arpitha, H. J. (2021). Electrochemical analysis of indigo carmine using polyarginine modified carbon paste electrode. Journal of Electrochemical Science and Engineering, 11(2), 87–96. [Google Scholar]
  16. Eksin, E. , Congur, G. , & Erdem, A. (2015). Electrochemical assay for determination of gluten in flour samples. Food Chemistry, 184, 183–187. [DOI] [PubMed] [Google Scholar]
  17. Fiedler, K. L. , McGrath, S. C. , Callahan, J. H. , & Ross, M. M. (2014). Characterization of grain‐specific peptide markers for the detection of gluten by mass spectrometry. Journal of Agricultural and Food Chemistry, 62(25), 5835–5844. [DOI] [PubMed] [Google Scholar]
  18. García‐García, A. , Madrid, R. , Sohrabi, H. , de la Cruz, S. , García, T. , Martín, R. , & González, I. (2019). A sensitive and specific real‐time PCR targeting DNA from wheat, barley and rye to track gluten contamination in marketed foods. LWT, 114, 108378. [Google Scholar]
  19. Gomaa, A. , & Boye, J. (2015). Simultaneous detection of multi‐allergens in an incurred food matrix using ELISA, multiplex flow cytometry and liquid chromatography mass spectrometry (LC–MS). Food Chemistry, 175, 585–592. [DOI] [PubMed] [Google Scholar]
  20. Guettel, R. , Knochen, J. , Kunz, U. , Kassing, M. , & Turek, T. (2008). Preparation and catalytic evaluation of cobalt‐based monolithic and powder catalysts for Fischer−Tropsch synthesis. Industrial & Engineering Chemistry Research, 47(17), 6589–6597. [Google Scholar]
  21. Hai, X. , Chang, K. , Pang, H. , Li, M. , Li, P. , Liu, H. , Shi, L. , & Ye, J. (2016). Engineering the edges of MoS2 (WS2) crystals for direct exfoliation into monolayers in polar micromolecular solvents. Journal of the American Chemical Society, 138(45), 14962–14969. [DOI] [PubMed] [Google Scholar]
  22. Haraszi, R. , Chassaigne, H. , Maquet, A. , & Ulberth, F. (2011). Analytical methods for detection of gluten in food—Method developments in support of food labeling legislation. Journal of AOAC International, 94(4), 1006–1025. [PubMed] [Google Scholar]
  23. Huang, C. , Sheth, S. , Li, M. , Ran, G. , & Song, Q. (2018). Rapid and selective luminescent sensing of allergenic gluten by highly phosphorescent switch‐on probe. Talanta, 190, 292–297. [DOI] [PubMed] [Google Scholar]
  24. Huang, S. , You, Z. , Jiang, Y. , Zhang, F. , Liu, K. , Liu, Y. , Chen, X. , & Lv, Y. (2020). Fabrication of ultrathin MoS2 nanosheets and application on adsorption of organic pollutants and heavy metals. Processes, 8(5), 504. [Google Scholar]
  25. Kazıcı, H. C. , Yayla, M. , Ulaş, B. , Aktaş, N. , & Kivrak, H. (2019). Development of nonenzymatic benzoic acid detection on PdSn/GCE/Vulcan XC‐72R prepared via polyol method. Electroanalysis, 31(6), 1118–1124. [Google Scholar]
  26. Lacorn, M. , Weiss, T. , Wehling, P. , Arlinghaus, M. , & Scherf, K. (2019). Quantification of wheat, rye, and barley gluten in oat and oat products by ELISA RIDASCREEN® total gluten: Collaborative study, first action 2018.15. Journal of AOAC International, 102(5), 1535–1543. [DOI] [PubMed] [Google Scholar]
  27. Li, F. , Yu, Z. , Han, X. , & Lai, R. Y. (2019). Electrochemical aptamer‐based sensors for food and water analysis: A review. Analytica Chimica Acta, 1051, 1–23. [DOI] [PubMed] [Google Scholar]
  28. Liang, X. , Zhang, X. J. , You, T. T. , Yang, N. , Wang, G. S. , & Yin, P. G. (2018). Three‐dimensional MoS2‐NS@ Au‐NPs hybrids as SERS sensor for quantitative and ultrasensitive detection of melamine in milk. Journal of Raman Spectroscopy, 49(2), 245–255. [Google Scholar]
  29. Limthin, D. , Chattrairat, K. , Leepheng, P. , Wisutthipat, S. , Gansa, P. , Klamchuen, A. , & Phromyothin, D. (2019). Methyl methacrylate magnetic molecularly imprinted polymer for gluten determination. Journal of Metals, Materials and Minerals, 29(1), 42–48. [Google Scholar]
  30. Liu, X. , Liu, L. , Wu, Y. , Wang, Y. , Yang, J. , & Wang, Z. (2019). Rosette‐like MoS2 nanoflowers as highly active and stable electrodes for hydrogen evolution reactions and supercapacitors. RSC Advances, 9(24), 13820–13828. [DOI] [PMC free article] [PubMed] [Google Scholar]
  31. Malvano, F. , Albanese, D. , Pilloton, R. , & Di Matteo, M. (2017). A new label‐free impedimetric aptasensor for gluten detection. Food Control, 79, 200–206. [Google Scholar]
  32. Manjunatha, J. G. , Kumara Swamy, B. E. , Mamatha, G. P. , Gilbert, O. , Shreenivas, M. T. , & Sherigara, B. S. (2009). Electrocatalytic response of dopamine at mannitol and triton X‐100 modified carbon paste electrode: A cyclic voltammetric study. International Journal of Electrochemical Science, 4, 1706–1718. [Google Scholar]
  33. Molina‐Delgado, M. , Aguilar‐Caballos, M. , & Gómez‐Hens, A. (2011). Usefulness of gold nanoparticles as labels for the determination of gliadins by immunoaffinity chromatography with light scattering detection. Talanta, 85(5), 2391–2396. [DOI] [PubMed] [Google Scholar]
  34. Niewinski, M. M. (2008). Advances in celiac disease and gluten‐free diet. Journal of the American Dietetic Association, 108(4), 661–672. [DOI] [PubMed] [Google Scholar]
  35. Oakes, L. , Carter, R. , Hanken, T. , Cohn, A. P. , Share, K. , Schmidt, B. , & Pint, C. L. (2016). Interface strain in vertically stacked two‐dimensional heterostructured carbon‐MoS2 nanosheets controls electrochemical reactivity. Nature Communications, 7(1), 1–7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  36. Panda, R. , Boyer, M. , & Garber, E. A. (2017). A multiplex competitive ELISA for the detection and characterization of gluten in fermented‐hydrolyzed foods. Analytical and Bioanalytical Chemistry, 409(30), 6959–6973. [DOI] [PubMed] [Google Scholar]
  37. Pietrzyk, B. , Miszczak, S. , Kaczmarek, Ł. , & Klich, M. (2018). Low friction nanocomposite aluminum oxide/MoS2 coatings prepared by sol‐gel method. Ceramics International, 44(7), 8534–8539. [Google Scholar]
  38. Prinith, N. S. , Jamballi, G. M. , & Chenthattil, R. (2019). Electrocatalytic analysis of dopamine, uric acid and ascorbic acid at poly(adenine) modified carbon nanotube paste electrode: A cyclic voltammetric study. Anaytical and Bioanalytical Electrochemistry, 11(6), 742–756. [Google Scholar]
  39. Pu, Z. , Liu, Q. , Asiri, A. M. , Luo, Y. , Sun, X. , & He, Y. (2015). 3D macroporous MoS2 thin film: In situ hydrothermal preparation and application as a highly active hydrogen evolution electrocatalyst at all pH values. Electrochimica Acta, 168, 133–138. [Google Scholar]
  40. Raril, C. , & Manjunatha, J. G. (2018). Sensitive electrochemical analysis of resorcinol using polymer modified carbon paste electrode: A cyclic voltammetric study. Analytical and Bioanalytical Electrochemistry, 10(4), 488–498. [Google Scholar]
  41. Salman, F. , Zengin, A. , & Kazici, C. H. (2020). Synthesis and characterization of Fe3O4‐supported metal–organic framework MIL‐101 (Fe) for a highly selective and sensitive hydrogen peroxide electrochemical sensor. Ionics, 26(10), 5221–5232. [Google Scholar]
  42. Sandberg, M. , Lundberg, L. , Ferm, M. , & Yman, I. M. (2003). Real time PCR for the detection and discrimination of cereal contamination in gluten free foods. European Food Research and Technology, 217(4), 344–349. [Google Scholar]
  43. Scharf, A. , Kasel, U. , Wichmann, G. , & Besler, M. (2013). Performance of ELISA and PCR methods for the determination of allergens in food: An evaluation of six years of proficiency testing for soy (Glycine max L.) and wheat gluten (Triticum aestivum L.). Journal of Agricultural and Food Chemistry, 61(43), 10261–10272. [DOI] [PubMed] [Google Scholar]
  44. Scherf, K. A. , & Poms, R. E. (2016). Recent developments in analytical methods for tracing gluten. Journal of Cereal Science, 67, 112–122. [Google Scholar]
  45. Shan, Y. , Ma, M. , Huang, X. , Guo, Y. , Jin, G. , & Jin, Y. (2012). Simple pH treatment as an effective tool to improve the functional properties of ovomucin. Journal of Food Science, 77(7), C740–C745. [DOI] [PubMed] [Google Scholar]
  46. Shufang, R. , Runyan, F. , Shounian, C. , Liqiu, H. , Qingtao, W. , & Zhixiang, Z. (2022). Construction of a sensitive electrochemical sensor based on hybrid 1 T/2H MoS2 nanoflowers anchoring on rGO nanosheets for the voltammetric determination of acetaminophen. Microchemical Journal, 175, 107129. [Google Scholar]
  47. Slot, I. D. B. , Bremer, M. G. , van der Fels‐Klerx, I. , & Hamer, R. J. (2015). Evaluating the performance of gluten ELISA test kits: The numbers do not tell the tale. Cereal Chemistry, 92(5), 513–521. [Google Scholar]
  48. Su, Y. , Hu, R. , Huang, W. , & Hu, K. (2011). A novel anti‐interference and pH‐modulation device: Application to enzyme‐free glucose detection. Microchimica Acta, 173(1), 19–26. [Google Scholar]
  49. Tian, Y. , Ge, L. , Wang, K. , & Chai, Y. (2014). Synthesis of novel MoS2/g‐C3N4 heterojunction photocatalysts with enhanced hydrogen evolution activity. Materials Characterization, 87, 70–73. [Google Scholar]
  50. Tigari, G. , & Manjunatha, J. G. (2020). Optimized voltammetric experiment for the determination of phloroglucinol at surfactant modified carbon nanotube paste electrode. Instruments and Experimental Techniques, 63(5), 750–757. [Google Scholar]
  51. Van Bergeijk, J. , Mulder, C. , & Thies, J. (1993). Coeliac disease. Three cases of delayed diagnosis after a sojourn in the tropics. The Netherlands Journal of Medicine, 43(5‐6), 222–226. [PubMed] [Google Scholar]
  52. Walter, T. , Wieser, H. , & Koehler, P. (2014). Production of gluten‐free wheat starch by peptidase treatment. Journal of Cereal Science, 60(1), 202–209. [Google Scholar]
  53. White, S. P. , Frisbie, C. D. , & Dorfman, K. D. (2018). Detection and sourcing of gluten in grain with multiple floating‐gate transistor biosensors. ACS Sensors, 3(2), 395–402. [DOI] [PubMed] [Google Scholar]
  54. Zhou, J. , Qin, J. , Zhao, N. , Shi, C. , Liu, E.‐Z. , He, F. , Li, J. , & He, C. (2016). Salt‐template‐assisted synthesis of robust 3D honeycomb‐like structured MoS2 and its application as a lithium‐ion battery anode. Journal of Materials Chemistry A, 4(22), 8734–8741. [Google Scholar]

Articles from Journal of Food Science are provided here courtesy of Wiley