Abstract
Objectives
To propose and validate a personalized iodine delivery rate (IDR) based injection protocol for coronary CT angiography (cCTA) on photon-counting CT.
Materials and methods
First, ideal IDR (IDRIDEAL) was retrospectively calculated for a HU target of 500 in 55 keV image reconstructions. Next, linear regression analysis was performed with IDRIDEAL and demographic parameters to derive a candidate IDR formula. This was implemented in two validation groups characterized by injection rate (3.5 and 5.0 mL/s). Here, coronary enhancement was quantified and equivalence assessed for a predefined HU range (500 ± 50). Additionally, a reader study assessed perceived enhancement of the coronary tree.
Results
This IRB-approved study used a retrospective cohort of 162 patients (group A: 58 ± 12 years; 81 men) and two prospective cohorts of 51 patients (group B1: 60 ± 13 years; 22 men and group B2: 59 ± 12 years; 30 men). IDRIDEAL correlated best with fat-free mass (FFM) (r = 0.67) and was integrated for contrast personalization. Prospectively, mean coronary enhancement was 533 ± 97 HU and 528 ± 68 HU for both groups (p = 0.79) with mean IDR values of 1.01 ± 0.11 gI/s and 1.07 ± 0.14 gI/s. However, distribution variances were significantly different (p = 0.015). Subjective scoring showed no differences between the two groups on overall and per-vessel level (p > 0.05).
Conclusion
A personalized, IDR-based injection protocol for cCTA was proposed and validated. FFM was best for IDRIDEAL prediction. Higher injection rates provided more precise coronary enhancement.
Key Points
Question Virtual mono-energetic images lead to low iodine delivery rate settings, but the impact on coronary enhancement is not clear.
Findings An iodine delivery rate-based injection protocol implementing personalized contrast volume dilutions with set injection duration and high injection rate improved coronary enhancement.
Clinical relevance Iodine delivery rate fine-tuning is a promising approach for coronary CT angiography with low iodine volumes. High injection rates provide more accurate and precise coronary enhancement.
Graphical Abstract
Keywords: Computed tomography angiography, Contrast media, Coronary artery disease, Coronary vessels
Introduction
Photon-counting CT (PCCT) is becoming an increasingly important innovation in the field of cardiovascular imaging, promising high-quality cardiac examinations with the possibility to capture spectral information as part of the clinical routine in coronary CT angiography (cCTA) and reconstruct CT images as if they were acquired with mono-energetic beams. Compared to other scanners, there is also no compromise in scan time or temporal resolution, and PCCT is also particularly effective at low dose settings [1, 2]. The opportunity to select virtual mono-energetic images (VMI) allows for the production of images optimized for a specified clinical task [3–6]. In cCTA, optimal energies must be selected for both vessel enhancement and reliable quantification of atherosclerotic plaques.
The injection of iodinated contrast media (ICM) is essential for visualizing the coronary arteries and should ideally lead to an enhancement of 350 to 500 Hounsfield Units (HU) [7, 8]. Contrary to polychromatic images, where vessel enhancement is dependent on set tube voltage [9, 10], dual-energy CT systems allow to use optimized VMI levels [11–13]. The possibility of reducing iodine load using low VMI levels has already been studied for more than 10 years, including cardiovascular imaging [14–17]. With the introduction of PCCT, ICM injection protocols should be reinvestigated, and more in particular if a set HU range is aimed for in the entire patient cohort.
Personalized ICM injection protocols should lead to appropriate vessel enhancement, while keeping the total iodine dose (TID) minimal. Previous studies have already demonstrated the association of various body habitus metrics with vessel enhancement in cCTA scans [9, 18–29]. However, in addition to body habitus, ICM characteristics and injection protocol settings cannot be ignored. In practice, selecting optimal injection protocol parameters is a multi-parameter challenge [10, 26].
The present study focused on iodine delivery rate (IDR), which represents the amount of iodine injected per second. Several publications have already shown the utility of IDR to target appropriate vascular enhancement in clinical practice: using a set injection time, a fixed IDR can be reached by adjusting the injection rate and iodine volume [30–34]. However, the impact of varying injection rate on coronary enhancement in cCTA is less well documented and could potentially introduce negative implications when too low, provided that VMIs enable to reduce these IDR values considerably [35].
The aim of this study is twofold: (1) to propose a personalized IDR-based contrast formula by using easily obtainable body metrics for cCTA on a PCCT, and (2) to investigate the impact of injection rate on coronary enhancement, implementing personalized contrast volume dilutions in a fixed injection time setting. The ultimate aim is to propose a validated injection protocol to reach a set coronary enhancement throughout the entire patient population.
Materials and methods
Study design and patient population
This mono-centric study was evaluated and approved by the local IRB of the University Hospitals Leuven, Belgium (S58042 and S70047). All included patients were referred for a clinically indicated cCTA scan to exclude coronary artery disease (CAD). Clinical indications comprised atypical thoracic pain, dyslipidemia, familial history of CAD and coexistence of multiple risk factors [36]. Inclusion criteria were: age older than 18 years, scanning performed with the high-pitch scan mode using predefined scan and injection timings. Renal dysfunction (eGFR < 30 mL/min/1.73 m2), known hypersensitivity to ICM and pregnancy were the predefined exclusion criteria.
The study comprised of (1) a retrospective study (group A) to propose a personalized formula to calculate an injection volume and (2) a prospective study to validate latter formula and to verify the impact of a standardized injection rate, 3.5 and 5.0 mL/s, in two study groups (group B1 and B2 respectively) (Fig. 1). The retrospective cohort comprised of eligible cCTA scans throughout 3 months from May 2024 to July 2024. For the prospective cohort, 127 consecutive patients were enrolled from October 2024 to December 2024, of which 25 patients were excluded because the high-pitch scan mode was not preferable. In both retrospective and prospective study groups, no extravasation events were recorded.
Fig. 1.
Flowchart of the study. cCTA, coronary CT angiography
CT scan acquisition
All scans were performed on a first-generation dual-source PCCT scanner (NAEOTOM Alpha.Peak, Siemens Healthineers) equipped with two cadmium telluride detectors (collimation: 144 × 0.4 mm; gantry rotation speed: 0.25 s). Patients received 5 mg of sublingual nitroglycerin and, if resting heart rate exceeded 70 beats per minute, intravenous beta-blockers were administered prior to scanning. The scan protocol was performed in the electrocardiogram-gated high-pitch spectral acquisition mode (QuantumPlus scan mode, Siemens Healthineers) during one heart cycle, with a tube voltage of 140 kVp, an image quality (IQ) level chosen as a function of the calcium score (limited calcium load, IQ level = 50; else, IQ level = 110) and automated tube current modulation (CareDose4D, Siemens Healthineers) [37]. Contrast timing was defined using the bolus-tracking technique, with a region of interest (ROI) in the descending aorta using a threshold of 140 HU at 90 kVp. A scan delay of 7 s was set after the HU limit was reached.
Contrast injection protocol
For both study parts, a biphasic injection protocol was performed with a dual-head power injector (Dual Shot alpha 7, Nemoto Kyorindo Co.) (Table 1). A 17 s ICM bolus (Visipaque 320 (iodixanol 320 mgI/mL) and Omnipaque 350 (iohexol 350 mgI/mL)) was injected into, preferably, the antecubital vein through an 18-G intravenous catheter. In the retrospective cohort, this bolus was undiluted and characterized by an injection rate (3.0–4.0 mL/s) and volume (51–68 mL) determined with a simple in-house weight-based rule. In the prospective cohort, this bolus was characterized by a personalized IDR (based on the retrospectively proposed personalized formula (Eq. 10)) and a set injection rate (group B1 = 3.5 mL/s and group B2 = 5.0 mL/s). Next, ICM dilution (ICM:saline) was calculated based on the personalized iodine concentration, derived from the personalized IDR, and the base ICM concentration (Eq. 1). This bolus was followed by a saline chasing bolus of 30 mL at an injection rate matched with the primary bolus.
| 1 |
Table 1.
Injection protocol characteristics
| Retrospective group A | Group B1 Injection rate = 3.5 mL/s |
Group B2 Injection rate = 5.0 mL/s |
|
|---|---|---|---|
| ICM bolus | |||
| IDR selection | Weight-based | FFMD-based* | FFMD-based* |
| Injection time (s) | 17 | 17 | 17 |
| Injection rate (mL/s) | 3.0–4.0 | 3.5 | 5.0 |
| Injection volume (mL) | 51–68 | 60 | 85 |
| ICM dilution† | No | Yes, if neededˠ | Yes, if neededˠ |
| ICM brand˜ | Visipaque 320, Omnipaque 350 | Omnipaque 350 | Omnipaque 350 |
| Base concentration (mgI/mL) | 320, 350 | 350 | 350 |
| Saline chasing bolus | |||
| Injection rate (mL/s)‡ | 3.0–4.0 | 3.5 | 5.0 |
| Injection volume (mL) | 30 | 30 | 30 |
IDR iodine delivery rate, FFMD, fat-free mass based on the Deurenberg formula, ICM iodinated contrast media
* Using Eq. 10, derived from the retrospective part of this study
ˠ Variable, depending on the predicted personalized ICM concentration and base concentration (Eq. 1)
† ICM bolus was diluted by injector settings, using saline
˜ In the retrospective group, 80% received Visipaque 320 and 20% received Omnipaque 350
‡ Matched with set injection rate of ICM bolus
Proposing a formula for ideal IDR prediction using retrospective data
For each scan in the retrospective cohort, the retrospective IDR (IDRRP) was calculated using the documented injection rate and iodine concentration (Eq. 2). Next, coronary enhancement was defined as the averaged HU value (HURP) from HU measurements in the ascending aorta (at the level of the ostium of the left main coronary artery) and in the proximal part of the right coronary artery. For these measurements, a circular ROI was drawn in the vessel lumen, obviating the coronary vessel wall. All images were reconstructed in an identical manner, using a Bv44 kernel and slice thickness of 0.4 mm, and enhancement measurements were performed on a VMI level of 55 keV.
| 2 |
In this study, a HU value of 500 in the 55 keV VMI reconstructed images was defined as the ideal HU enhancement (HUIDEAL). This relatively high target was chosen to guarantee good coronary enhancement when increasing the VMI keV levels, as this could impact stenosis grading [3–5]. However, at this moment of the study, the ideal keV VMI image choice for lumen and plaque quantification was not yet fixed, and therefore the guidelines from the vendor were applied. Figure 2 showcases an example of these different VMI reconstructions in a patient with ideal coronary enhancement.
Fig. 2.
Representative VMI reconstructions (including maximum intensity projection (MIP) at 55 keV) of a cCTA scan from the prospective group, characterized by an injection rate of 5.0 mL/s, which reached the ideal coronary enhancement of 500 HU at 55 keV (WW: 600, WL: 200). Contrast enhancement reduces with increasing keV. cCTA, coronary CT angiography; VMI, virtual mono-energetic image
From IDRRP, HURP and HUIDEAL the ideal IDR (IDRIDEAL) was then calculated for all patients based on the basic principle that IDR and HU enhancement relate linearly in a first approach (Eq. 3) [29].
| 3 |
Age, gender, and various body metrics were collected and correlated with the calculated IDRIDEAL. Based on literature, the following body metrics were selected: length (L), body weight (BW), water equivalent diameter (WED) [38], body mass index (BMI), body surface area (BSA), lean body weight (LBW) and fat-free mass, the latter calculated with the Deurenberg formula (FFMD) [39]. Equations 4—9 show these metrics.
| 4 |
| 5 |
| 6 |
| 7 |
| 8 |
| 9 |
Next, linear regression analysis was performed with the calculated IDRIDEAL and the different patient parameters. The body metric with the highest correlation coefficient was selected as our best candidate for the personalized injection formula, and the resulting regression formula would then be used to predict the IDRIDEAL for routine use in cCTA.
Prospective validation of the proposed personalized injection formula
The prospective validation study comprised two groups, characterized by a predefined injection rate (group B1 = 3.5 mL/s and group B2 = 5.0 mL/s). Keeping injection time and scan delay identical to what was used in the retrospective cases, the retrospectively derived candidate formula for IDRIDEAL was used to calculate the dilution rate of the iodine volume, taking into account the predefined injection rate. The resulting coronary enhancement was measured in an identical manner as defined in the retrospective cohort. Additionally, signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were quantified using a circular ROI in the ascending aorta (at the level of the ostium of the left main coronary artery) and in the epicardial adipose tissue [40].
Subjective quality assessment of coronary artery enhancement
Coronary enhancement of all prospective scans was independently scored by a radiologist (V.S.) with 5 years of experience in cardiovascular radiology and a board-certified radiologist with more than 20 years of experience in cardiovascular imaging (S.D.). Both readers were blinded to the used injection rate and scored all cases from the 55 keV VMI images with identical reconstruction settings. Window width and level (WW:600, WL:200) were set identical for each case, yet no restrictions were applied to adjust these parameters. Coronary enhancement was scored for overall and per-vessel enhancement using a five-point Likert scale characterized by: 5 = excellent, 4 = good, 3 = moderate, 2 = poor and 1 = nondiagnostic.
Statistical analysis
All statistical analyses were performed using GraphPad Prism (version 5.04, Dotmatics). Descriptive statistics was performed in terms of mean and standard deviation. Normality of the data was assessed with the D’Agostino and Pearson omnibus normality test. Differences between all 3 groups were assessed using an ANOVA or Kruskal–Wallis test, for normal and non-normally distributed datasets, respectively. In the retrospective cohort, correlations were calculated using the Pearson correlation or non-parametric Spearman correlation test, for normal and non-normal datasets, respectively. The correlation coefficients were interpreted using the following categories: 0.00–0.19, very weak; 0.20–0.39, weak; 0.40–0.59, moderate; 0.60–0.79, strong; 0.80–1.00, very strong. Categorical distributions were reported as percentages and evaluated using a Chi-squared test. For the prospective validation cohorts, sample sizes were calculated based on the following null-hypothesis: the difference between the mean HU of the group with personalized ICM injection protocol and the target HU value of 500 HU is outside the equivalence interval. We assumed a power of 0.80, a significance level of 0.05, an anticipated standard deviation of 90 HU and an equivalence limit of 50 HU. The calculated sample size was 51 participants. Equivalence of the measured HU enhancement with the set HU enhancement of 500 ± 50 HU was determined with a two one-sided t-test. Differences between both prospective cohorts were assessed using either an unpaired t-test or Mann–Whitney test, for normal and non-normally distributed datasets. Variance of coronary enhancement was verified with an F-test. Subjective enhancement scores for both prospective cohorts were compared using a Mann–Whitney test. The significance level throughout the study was set to 0.05.
Results
Patient population
In this study, 264 patients were included, subdivided into a retrospective cohort (162 patients) and two prospective cohorts (51 patients each) (Fig. 1). All patient demographics and body metrics are listed in Table 2. There were no significant differences observed in length, weight, age, WED, BMI, BSA, LBW and FFMD among all three groups (p > 0.05). Furthermore, there was no significant difference (p = 0.44) observed in the CAD-RADS distributions between all three groups.
Table 2.
Patient demographics and clinical results
| Retrospective group A | Group B1 Injection rate = 3.5 mL/s |
Group B2 Injection rate = 5.0 mL/s |
p-value | |
|---|---|---|---|---|
| No. of patients | 162 | 51 | 51 | |
| Patient demographics | ||||
| Length (cm) | 172 ± 10 | 170 ± 9 | 172 ± 10 | p = 0.40 |
| Weight (kg) | 78 ± 16 | 74 ± 14 | 78 ± 15 | p = 0.26 |
| Age (years) | 58 ± 12 | 60 ± 13 | 59 ± 12 | p = 0.50 |
| WED (cm) | 26 ± 3 | 26 ± 3 | 26 ± 4 | p = 0.25 |
| BMI (kg/m2) | 26 ± 5 | 25 ± 4 | 26 ± 4 | p = 0.42 |
| BSA (m2) | 1.9 ± 0.2 | 1.9 ± 0.2 | 1.9 ± 0.2 | p = 0.23 |
| LBW (kg) | 53 ± 9 | 51 ± 8 | 53 ± 9 | p = 0.23 |
| FFMD (kg) | 51 ± 11 | 48 ± 10 | 52 ± 11 | p = 0.14 |
| Gender (% male) | 50 | 43 | 59 | p = 0.28 |
| Clinical results* | ||||
| CAD-RADS 0 (%) | 48 | 51 | 43 | |
| CAD-RADS 1 (%) | 23 | 20 | 27 | |
| CAD-RADS 2 (%) | 16 | 12 | 20 | |
| CAD-RADS 3 (%) | 3 | 8 | 8 | |
| CAD-RADS 4A (%) | 5 | 4 | 2 | |
| CAD-RADS 4B (%) | 3 | 0 | 0 | |
| CAD-RADS 5 (%) | 2 | 6 | 0 | |
All data are summarized as mean ± standard deviation, unless indicated otherwise
WED water equivalent diameter, BMI body mass index, BSA body surface area, LBW lean body weight, FFMD fat-free mass based on the Deurenberg formula, CAD-RADS coronary artery disease reporting and data system
* CAD-RADS distributions were tested using a Chi-squared test and showed no significant differences (p = 0.44)
A retrospectively derived cCTA-specific formula to predict the ideal IDR
Coronary enhancement in the retrospective study was 548 ± 90 HU and followed a normal distribution. For each patient, IDRIDEAL was calculated and showed in 59% of patients a possibility for reduction when compared to the IDRRP. Only in 22% of cases, the IDRIDEAL showed an increase compared to IDRRP. Mean IDRIDEAL was 1.08 ± 0.21 gI/s compared to the observed IDRRP of 1.15 ± 0.13 gI/s. Next, correlation of IDRIDEAL was significant (p < 0.05) with all body metrics. Patient age showed a weak negative correlation with IDRIDEAL (r = −0.25). BMI (r = 0.31) and WED (r = 0.38) showed a weak positive correlation, while patient length (r = 0.56) showed a moderate positive correlation. Most importantly, patient weight (r = 0.60), BSA (r = 0.65), LBW (r = 0.67) and FFMD (r = 0.67) showed all strong positive correlations. Linear regressions of the four highest correlating body metrics are shown in Fig. 3. Based on the additional integration of the age parameter in the calculation and slightly better R2 value, FFMD was selected over LBW as predictor for IDRIDEAL. Therefore, the linear regression of IDRIDEAL as a function of FFMD was adopted to predict IDRIDEAL (Eq. 10).
| 10 |
Fig. 3.
Linear regression analysis of the four highest correlating body metrics IDRIDEAL. BSA, body surface area; FFMD, fat-free mass based on the Deurenberg formula; IDR, iodine delivery rate; LBW, lean body weight
Prospective validation of injection protocols integrating IDRIDEAL
In the prospective study part that used Eq. 10 to determine the personalized IDR, coronary enhancement of neither group B1 (characterized by an injection rate of 3.5 mL/s) nor group B2 (characterized by an injection rate of 5.0 mL/s) showed a significant correlation with any body metric (p > 0.05). All correlation coefficients showed very weak to weak correlations, ranging from 0.019 to 0.27 and 0.016 to 0.20 for group B1 and B2, respectively.
Mean IDRIDEAL was 1.01 ± 0.11 gI/s for group B1 and 1.07 ± 0.14 gI/s for group B2. Mean coronary enhancements were 533 ± 97 HU and 528 ± 68 HU in group B1 and group B2, respectively, and followed in both groups a normal distribution. Mean coronary enhancement between both groups showed no significant difference (p = 0.79); however, the variance in HU enhancement in group B2 was significantly lower (p = 0.015) compared to group B1. This observation is present in the histograms representing HU distribution in Fig. 4: there is a well-defined HU peak around the 500–550 HU bins in group B2 compared to a wider bell-shaped curve in group B1. Furthermore, only the coronary enhancement in group B2 (mean HU [90% CI]; 528 HU [512–544 HU]) was deemed equivalent with the set target value of 500 ± 50 HU, while this was not the case for group B1 (mean HU [90% CI]; 533 HU [510–556 HU]). Additional information on TID, set IDR, injected ICM volume, heart rate, time to scan, z-axis coverage, scan time, chosen IQ level, radiation dose, SNR and CNR are listed in Table 3.
Fig. 4.
Mean coronary enhancement distributions in both prospective groups
Table 3.
Scan timing, injector settings, coronary enhancement, heart rate and radiation dose estimates
| Group B1 Injection rate = 3.5 mL/s |
Group B2 Injection rate = 5.0 mL/s |
p-value | |
|---|---|---|---|
| No. of patients | 51 | 51 | |
| Time to scan (s) | 30 ± 3 | 28 ± 3 | p < 0.01 |
| Z-axis coverage (cm) | 20 ± 1 | 20 ± 1 | p = 0.79 |
| Scan time (s) | 0.3 ± 0.02 | 0.3 ± 0.02 | p = 0.79 |
| TID (gI) | 17.4 ± 1.9 | 18.2 ± 2.5 | p = 0.06 |
| ICM volume (mL)* | 50 ± 5 | 52 ± 7 | p = 0.06 |
| IDR (gI/s) | 1.01 ± 0.11 | 1.07 ± 0.14 | p = 0.03 |
| Heart rate (bpm) | 60 ± 7 | 61 ± 8 | p = 0.40 |
| Mean coronary enhancement (HU) | 533 ± 97 | 528 ± 68 | p = 0.79 |
| Aorta ascendens (HU) | 531 ± 98 | 529 ± 68 | p = 0.91 |
| Proximal RCA (HU) | 535 ± 98 | 527 ± 71 | p = 0.68 |
| IQ level 50 (%) | 76 | 49 | p < 0.01 |
| IQ level 110 (%) | 24 | 51 | p < 0.01 |
| CTDIVOL (mGy) | 2.7 ± 1.5 | 3.7 ± 2.0 | p < 0.01 |
| DLP (mGy*cm) | 53 ± 30 | 72 ± 41 | p < 0.01 |
| SNR | 18 ± 5 | 19 ± 5 | p = 0.85 |
| CNR | 22 ± 5 | 23 ± 5 | p = 0.95 |
All data are summarized as mean ± standard deviation
TID total iodine dose, ICM iodinated contrast media, IDR iodine delivery rate, RCA right coronary artery, CTDIVOL volume computed tomography dose index, DLP dose length product, SNR signal-to-noise ratio, CNR contrast-to-noise ratio, IQ image quality
* Utilizing an ICM with a base concentration of 350 mgI/mL
Visual grading analysis of coronary enhancement
Overall coronary enhancement was scored excellent by both readers and median [Q1–Q3]) were: 5 [4, 5] vs 5 [4, 5] and 5 [4, 5] vs 5 [5] for group B1 and B2, respectively. For both readers, no significant difference was observed between the two groups (p = 0.84 and 0.28). Additionally, per-vessel enhancement scores showed no significant differences between the two groups for both readers (p > 0.05). Subjective enhancement scores are listed in Table 4. The objective analysis in terms of SNR and CNR did not show any significant differences either (p = 0.85 and 0.95).
Table 4.
Visual grading analysis of coronary enhancement
| Group B1 Injection rate = 3.5 mL/s |
Group B2 Injection rate = 5.0 mL/s |
|||
|---|---|---|---|---|
| Overall coronary enhancement | 5 [4, 5] | 5 [4, 5] | 5 [4, 5] | 5 [5] |
| Left main | 5 [4, 5] | 5 [5] | 5 [4, 5] | 5 [5] |
| Proximal RCA | 5 [4, 5] | 5 [5] | 5 [4, 5] | 5 [5] |
| Mid RCA | 5 [4, 5] | 5 [5] | 5 [4, 5] | 5 [4, 5] |
| Distal RCA | 5 [4, 5] | 4 [3–5] | 5 [4, 5] | 4 [3–5] |
| Proximal LAD | 5 [4, 5] | 5 [5] | 5 [4, 5] | 5 [5] |
| Mid LAD | 5 [4, 5] | 5 [4, 5] | 5 [4, 5] | 5 [4, 5] |
| Distal LAD | 5 [4, 5] | 4 [3–5] | 5 [4, 5] | 4 [3–5] |
| Proximal LCx | 5 [4, 5] | 5 [5] | 5 [4, 5] | 5 [5] |
| Mid LCx | 5 [4, 5] | 5 [3–5] | 5 [4, 5] | 4 [4, 5] |
| Distal LCx | 5 [4, 5] | 4 [2–4] | 5 [4, 5] | 3 [2–4] |
All data are summarized as median [Q1–Q3]
RCA right coronary artery, LAD left anterior descending artery, LCx left circumflex artery
Discussion
Our study demonstrated that an IDR-based injection protocol with a constant injection rate of 5.0 mL/s resulted in a more precisely targeted coronary enhancement, compared to a lower injection rate of 3.5 mL/s. The injection protocol implemented a personalized IDR, estimated with the calculated FFMD of the patient, using a fixed injection time and, consequently, a calculated contrast agent dilution percentage.
FFMD and LBW were identified as the best candidates for the prediction of the IDR that would achieve the target enhancement, followed by BSA, patient weight and length. Patient age, BMI and WED showed weak correlations. These results are in agreement with findings of previous work [9, 10, 20–22, 28]. We decided to use FFMD, which takes into account both the fat-free body composition and patient’s sex, to predict IDRIDEAL (Eq. 10). These parameters seem to be best in representing the volumetric important measures of the cardiac system and vasculature for the application of cCTA.
The integration of IDR in injection protocols is an interesting concept, as it captures the impact of both injection rate and iodine concentration on coronary enhancement. Various phantom and clinical studies have already shown the utility of this parameter, obtaining similar enhancement levels when using different contrast media concentrations [32, 33, 41, 42]. In agreement with these previous studies, we found that the mean coronary enhancement in both prospective study groups was not significantly different. Additionally, our results showed that coronary enhancement and the body metrics were no longer significantly correlated in the prospective groups, proving that our proposed formula for personalized IDR (Eq. 10) successfully takes into account body habitus. This is what was aimed for.
With the introduction of PCCT, the need for high iodine concentration injections is history as VMIs of lower energies are standard-of-care and could be calculated for every patient. As determined in this study, an IDR ranging from 0.85 to 1.30 gI/s is not exceptional. Similar IDR values were also obtained by Emrich et al [35]. Implementing these low IDR values, for example, with a fixed iodine concentration of 350 mgI/mL, would result in injection rates ranging from 2.4 to 3.7 mL/s. It should be noted that the implementation of these lower injection rates should be well considered, as our data indicate that the injection rate has an impact on coronary enhancement distribution throughout the patient population. In this study, the prospective group, characterized by 3.5 mL/s injection rate, showed a significantly higher variance in coronary enhancement, compared to the group characterized by 5.0 mL/s injection rate. We hypothesize that these differences can be (in part) explained by (1) the total volume and (2) the velocity of the injected bolus. In our study, the injected bolus volumes were fixed per injection rate, with however saline replacing iodine to tune the IDR. Total volume of the injected bolus in the 3.5 mL/s group was 60 mL, compared to 85 mL in the 5 mL/s group. We assume that injecting a higher volume using a higher injection rate allows less dispersion of the contrast bolus in the blood, compared to a smaller and more slowly injected bolus and is therefore less susceptible to underlying patient-specific factors beyond body habitus.
This study is limited in that it is a mono-centric study with only a limited number of patients included. Second, it should be noted that the calculated IDRIDEAL was selected as an approximation for the ideal coronary enhancement. We calculated IDRIDEAL in the retrospective group based on the assumption that coronary enhancement and iodine concentration relate linearly. Furthermore, the coronary enhancement was only measured in the ascending aorta and proximal right coronary artery, which was defined as an indication for total coronary enhancement. Additionally, in the retrospective cohort, injection rate varied from 3.0 to 4.0 mL/s. As discussed in this study, the HU value distribution throughout the included patient population could be affected, and therefore, coronary enhancement does not depend only on the calculated IDR. Third, cardiac output was not implemented in this study, and therefore, the impact was not assessed.
Notwithstanding the limitations, these findings show the advantage of maintaining a sufficiently high injection rate in injection protocols for cCTA, even if low IDR rates can be used. Future work could include the potential of VMI-based injection protocols, ultimately tailoring contrast injections both to the patient and clinical task [43, 44]. If a higher or lower HU would be aimed for, the dilution rates would have to be increased or lowered linearly, at least in a first approximation. For example, a target HU of 425 in 55 keV reconstructed images would already allow a further TID reduction of 15%.
In conclusion, this study proposed and validated a personalized formula for an IDR-based injection protocol in cCTA. The fat-free mass body metric was derived as the best candidate for the prediction of ideal IDR. Consequently, the impact of injection rate was assessed and showed that a higher injection rate resulted in more precisely targeted coronary enhancement throughout the patient population.
Abbreviations
- BMI
Body mass index
- BSA
Body surface area
- CAD
Coronary artery disease
- cCTA
Coronary computed tomography angiography
- CNR
Contrast-to-noise ratio
- FFM
Fat-free mass
- ICM
Iodinated contrast media
- IDR
Iodine delivery rate
- LBW
Lean body weight
- PCCT
Photon-counting computed tomography
- ROI
Region of interest
- SNR
Signal-to-noise ratio
- TID
Total iodine dose
- VMI
Virtual mono-energetic image
- WED
Water equivalent diameter
Funding
AI-POD has received funding from the European Union’s Horizon Europe research and innovation programme under grant agreement 101080302, and from the Swiss State Secretariat for Education, Research and Innovation (SERI) under contract number 23.00174. Views and opinions expressed are, however, those of the author(s) only and do not necessarily reflect those of the European Union or HaDEA. Neither the European Union nor the granting authority can be held responsible for them.
Compliance with ethical standards
Guarantor
The scientific guarantor of this publication is Prof. Hilde Bosmans.
Conflict of interest
The authors of this manuscript declare no relationships with any companies, whose products or services may be related to the subject matter of the article.
Statistics and biometry
No complex statistical methods were necessary for this paper.
Informed consent
Written informed consent was obtained from all prospectively included subjects (patients) in this study.
Ethical approval
Institutional Review Board approval was obtained.
Study subjects or cohorts overlap
None.
Methodology
Prospective and retrospective study
Experimental study
Performed at one institution
Footnotes
Publisher’s Note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Flohr T, Schmidt B (2023) Technical basics and clinical benefits of photon-counting CT. Invest Radiol 58:441–450 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Flohr T, Schmidt B, Ulzheimer S, Alkadhi H (2023) Cardiac imaging with photon counting CT. Br J Radiol. 10.1259/bjr.20230407 [DOI] [PMC free article] [PubMed]
- 3.Wolf EV, Halfmann MC, Varga-Szemes A et al (2024) Photon-counting detector CT virtual monoenergetic images for coronary artery stenosis quantification: phantom and in vivo evaluation. AJR Am J Roentgenol 222:e2330481. 10.2214/AJR.23.30481 [DOI] [PubMed] [Google Scholar]
- 4.Symons R, Choi Y, Cork TE et al (2018) Optimized energy of spectral coronary CT angiography for coronary plaque detection and quantification. J Cardiovasc Comput Tomogr 12:108–114. 10.1016/j.jcct.2018.01.006 [DOI] [PubMed] [Google Scholar]
- 5.Sartoretti T, Moser LJ, Rusek S et al (2024) Photon-counting detector coronary CT angiography: defining the optimal monoenergetic level for grading of calcified coronary stenosis. J Cardiovasc Comput Tomogr 18:616–617 [DOI] [PubMed] [Google Scholar]
- 6.Pan Y, Gao Y, Wang Z et al (2024) Effects of low-tube voltage coronary CT angiography on plaque and pericoronary fat assessment: intraindividual comparison. Eur Radiol 34:5713–5723. 10.1007/s00330-024-10648-0 [DOI] [PubMed] [Google Scholar]
- 7.Cademartiri F, Mollet NR, Lemos PA et al (2006) Higher intracoronary attenuation improves diagnostic accuracy in MDCT coronary angiography. AJR Am J Roentgenol. 10.2214/AJR.05.1406 [DOI] [PubMed]
- 8.Fei X, Du X, Yang Q et al (2008) 64-MDCT coronary angiography: phantom study of effects of vascular attenuation on detection of coronary stenosis. AJR Am J Roentgenol 191:43–49. 10.2214/AJR.07.2653 [DOI] [PubMed] [Google Scholar]
- 9.Van Gompel G, Delombaerde L, Zanca F et al (2023) A patient- and acquisition-tailored injection approach for improving consistency of CT enhancement towards a target CT value in coronary CT angiography. J Appl Clin Med Phys. 10.1002/acm2.13867 [DOI] [PMC free article] [PubMed]
- 10.Bae KT (2010) Intravenous contrast medium administration and scan timing at CT: considerations and approaches. Radiology 256:32–61. 10.1148/radiol.10090908 [DOI] [PubMed] [Google Scholar]
- 11.Cundari G, Deilmann P, Mergen V et al (2023) Saving contrast media in coronary CT angiography with photon-counting detector CT. Acad Radiol. 10.1016/j.acra.2023.06.025 [DOI] [PubMed]
- 12.Higashigaito K, Mergen V, Eberhard M et al (2023) CT angiography of the aorta using photon-counting detector CT with reduced contrast media volume. Radiol Cardiothorac Imaging. 10.1148/ryct.220140 [DOI] [PMC free article] [PubMed]
- 13.Stammen L, Hoeijmakers EJI, Flohr TG et al (2025) Reducing contrast media dosage for pulmonary embolism CTPA in PCD-CT: a comparative study of EID-CT and PCD-CT in the era of individualized protocolling. Eur Radiol. 10.1007/s00330-025-12054-6 [DOI] [PMC free article] [PubMed]
- 14.Dubourg B, Caudron J, Lestrat JP et al (2014) Single-source dual-energy CT angiography with reduced iodine load in patients referred for aortoiliofemoral evaluation before transcatheter aortic valve implantation: impact on image quality and radiation dose. Eur Radiol 24:2659–2668. 10.1007/s00330-014-3263-1 [DOI] [PubMed]
- 15.Rotzinger DC, Si-Mohamed SA, Yerly J et al (2021) Reduced-iodine-dose dual-energy coronary CT angiography: qualitative and quantitative comparison between virtual monochromatic and polychromatic CT images. Eur Radiol 31:7132–7142. 10.1007/s00330-021-07809-w/Published [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Raju R, Thompson AG, Lee K et al (2014) Reduced iodine load with CT coronary angiography using dual-energy imaging: a prospective randomized trial compared with standard coronary CT angiography. J Cardiovasc Comput Tomogr 8:282–288. 10.1016/j.jcct.2014.06.003 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Popp D, Siedlecki M, Friedrich L et al (2025) Potential of photon-counting detector CT technology for contrast medium reduction in portal venous phase thoracoabdominal CT. Eur Radiol 35:4635–4648. 10.1007/s00330-025-11409-3 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Abbara S, Blanke P, Maroules CD et al (2016) SCCT guidelines for the performance and acquisition of coronary computed tomographic angiography: a report of the Society of Cardiovascular Computed Tomography Guidelines Committee: endorsed by the North American Society for Cardiovascular Imaging (NASCI). J Cardiovasc Comput Tomogr 10:435–449. 10.1016/j.jcct.2016.10.002 [DOI] [PubMed] [Google Scholar]
- 19.ACR Committee on Drugs and Contrast Media (2017) ACR manual on contrast media. Version 10.3 American College of Radiology
- 20.Yin WH, Yu YT, Zhang Y et al (2020) Contrast medium injection protocols for coronary CT angiography: should contrast medium volumes be tailored to body weight or body surface area? Clin Radiol 75:395.e17–395.e24. 10.1016/j.crad.2019.12.005 [DOI] [PubMed] [Google Scholar]
- 21.Mihl C, Kok M, Altintas S et al (2016) Evaluation of individually body weight adapted contrast media injection in coronary CT-angiography. Eur J Radiol 85:830–836. 10.1016/j.ejrad.2015.12.031 [DOI] [PubMed] [Google Scholar]
- 22.Nakaura T, Awai K, Yauaga Y et al (2008) Contrast injection protocols for coronary computed tomography angiography using a 64-detector scanner: comparison between patient weight-adjusted-and fixed iodine-dose protocols. Invest Radiol 43:512–519. 10.1097/RLI.0b013e3181727505 [DOI] [PubMed] [Google Scholar]
- 23.Masuda T, Nakaura T, Funama Y et al (2017) Aortic and hepatic contrast enhancement during hepatic-arterial and portal venous phase computed tomography scanning: multivariate linear regression analysis using age, sex, total body weight, height, and cardiac output. J Comput Assist Tomogr 41:309–314. 10.1097/RCT.0000000000000513 [DOI] [PubMed] [Google Scholar]
- 24.Bae KT, Seeck BA, Hildebolt CF et al (2008) Contrast enhancement in cardiovascular MDCT: effect of body weight, height, body surface area, body mass index, and obesity. AJR Am J Roentgenol 190:777–784. 10.2214/AJR.07.2765 [DOI] [PubMed] [Google Scholar]
- 25.Sandfort V, Choi Y, Symons R, Chen MY, Bluemke DA (2020) An optimized test bolus contrast injection protocol for consistent coronary artery luminal enhancement for coronary CT angiography. Acad Radiol 27:371–380. 10.1016/j.acra.2019.05.003 [DOI] [PubMed] [Google Scholar]
- 26.Oda S, Utsunomiya D, Nakaura T et al (2018) Basic concepts of contrast injection protocols for coronary computed tomography angiography. Curr Cardiol Rev 15:24–29. 10.2174/1573403x14666180918102031 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Masuda T, Nakaura T, Funama Y et al (2018) Development and validation of generalized linear regression models to predict vessel enhancement on coronary CT angiography. Korean J Radiol 19:1021–1030. 10.3348/kjr.2018.19.6.1021 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Matsumoto Y, Higaki T, Masuda T et al (2019) Minimizing individual variations in arterial enhancement on coronary CT angiographs using “contrast enhancement optimizer”: a prospective randomized single-center study. Eur Radiol 29:2998–3005. 10.1007/s00330-018-5823-2 [DOI] [PubMed] [Google Scholar]
- 29.Coudyzer W, Cockmartin L, Miseur B et al (2025) Personalized contrast agent volumes in abdominal CT: bridging theory with practice. J Belg Soc Radiol 109:1–8. 10.5334/jbsr.3906 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Du Y, Wang YN, Wang Q et al (2023) A comparison of the use of contrast media with different iodine concentrations for enhanced computed tomography. Front Physiol. 10.3389/fphys.2023.1141135 [DOI] [PMC free article] [PubMed]
- 31.Rengo M, Dharampal A, Lubbers M et al (2019) Impact of iodine concentration and iodine delivery rate on contrast enhancement in coronary CT angiography: a randomized multicenter trial (CT-CON). Eur Radiol 29:6109–6118. 10.1007/s00330-019-06196-7 [DOI] [PubMed]
- 32.Mihl C, Wildberger JE, Jurencak T et al (2013) Intravascular enhancement with identical iodine delivery rate using different iodine contrast media in a circulation phantom. Invest Radiol 48:813–818. 10.1097/RLI.0b013e31829979e8 [DOI] [PubMed] [Google Scholar]
- 33.Kok M, Mihl C, Seehofnerová A et al (2015) Automated tube voltage selection for radiation dose reduction in CT angiography using different contrast media concentrations and a constant iodine delivery rate. AJR Am J Roentgenol 205:1332–1338. 10.2214/AJR.14.13957 [DOI] [PubMed] [Google Scholar]
- 34.Kristiansen CH, Thomas O, Nyquist AB et al (2025) A randomised controlled trial comparing three clinical administration strategies in spectral detector CT pulmonary angiography with low contrast medium dose. Eur Radiol 35:4649–4660. 10.1007/s00330-025-11420-8 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Emrich T, O’Doherty J, Schoepf UJ et al (2023) Reduced iodinated contrast media administration in coronary CT angiography on a clinical photon-counting detector CT system: a phantom study using a dynamic circulation model. Invest Radiol 58:148–155. 10.1097/RLI.0000000000000911 [DOI] [PubMed] [Google Scholar]
- 36.Budaj A, Buechel RR, Alfonso Chiariello G et al (2024) 2024 ESC guidelines for the management of chronic coronary syndromes. Eur Heart J 45:3415–3537. 10.1093/eurheartj/ehae177 [DOI] [PubMed] [Google Scholar]
- 37.Mergen V, Rusek S, Civaia F et al (2024) Virtual calcium removal in calcified coronary arteries with photon-counting detector CT—first in-vivo experience. Front Cardiovasc Med. 10.3389/fcvm.2024.1367463 [DOI] [PMC free article] [PubMed]
- 38.McCollough C, Bakalyar DM, Bostani M et al (2014) Use of water equivalent diameter for calculating patient size and size-specific dose estimates (SSDE) in CT: the report of AAPM Task Group 220. AAPM Rep 2014:6–23 [PMC free article] [PubMed]
- 39.Deurenberg P, Andreoli A, Borg P et al (2021) The validity of predicted body fat percentage from body mass index and from impedance in samples of five European populations. Eur J Clin Nutr 55:973–979. 10.1038/sj.ejcn.1601254 [DOI] [PubMed] [Google Scholar]
- 40.Mergen V, Sartoretti T, Baer-Beck M et al (2022) Ultra-high-resolution coronary CT angiography with photon-counting detector CT: feasibility and image characterization. Invest Radiol 57:780–788. 10.1097/RLI.0000000000000897 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.Lubbers MM, Kock M, Niezen A et al (2018) Iodixanol versus iopromide at coronary CT angiography: lumen opacification and effect on heart rhythm—the randomized IsoCOR trial. Radiology 286:71–80. 10.1148/radiol.2017162779 [DOI] [PubMed] [Google Scholar]
- 42.Kok M, Mihl C, Hendriks BMF et al (2016) Patient comfort during contrast media injection in coronary computed tomographic angiography using varying contrast media concentrations and flow rates results from the EICAR trial. Invest Radiol 51:810–815. 10.1097/RLI.0000000000000284 [DOI] [PubMed] [Google Scholar]
- 43.Klambauer K, Flohr T, Moser LJ et al (2025) Contrast media and radiation dose optimization with task-based automatic keV selection: a proof-of-concept study with photon-counting detector CT. Eur Radiol. 10.1007/s00330-025-11738-3 [DOI] [PMC free article] [PubMed]
- 44.Higashigaito K, Euler A, Eberhard M, Flohr TG, Schmidt B, Alkadhi H (2022) Contrast-enhanced abdominal CT with clinical photon-counting detector CT: assessment of image quality and comparison with energy-integrating detector CT. Acad Radiol 29:689–697. 10.1016/j.acra.2021.06.018 [DOI] [PubMed] [Google Scholar]