Estimation of single channel kinetic parameters from data subject to limited time resolution.

R K Milne 1, G F Yeo 1, B W Madsen 1, R O Edeson 1
PMCID: PMC1330550  PMID: 2470426

Abstract

The limited responsiveness of single-channel recording systems results in some brief events not being detected, and if this is ignored parameter estimation from the observed data will be biased. Statistical methods of correcting for this limited time resolution in a two-state Markov model have been proposed by Neher (1983. J. Physiol. (Lond.). 339:663-678) and by Colquhoun and Sigworth (1983. Single Channel Recording. 191-263). However, a numerical study by Blatz and Magleby (1986. Biophys. J. 49:967-980) indicated differences of 3-40% in the corrected values given by the two techniques. Here we explain why Neher's method produces biased results and the Colquhoun and Sigworth approach, which is no more difficult, provides reasonably accurate estimates.

Full text

PDF
673

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Ball F., Sansom M. Temporal clustering of ion channel openings incorporating time interval omission. IMA J Math Appl Med Biol. 1987;4(4):333–361. doi: 10.1093/imammb/4.4.333. [DOI] [PubMed] [Google Scholar]
  2. Blatz A. L., Magleby K. L. Correcting single channel data for missed events. Biophys J. 1986 May;49(5):967–980. doi: 10.1016/S0006-3495(86)83725-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
  3. Milne R. K., Yeo G. F., Edeson R. O., Madsen B. W. Stochastic modelling of a single ion channel: an alternating renewal approach with application to limited time resolution. Proc R Soc Lond B Biol Sci. 1988 Apr 22;233(1272):247–292. doi: 10.1098/rspb.1988.0022. [DOI] [PubMed] [Google Scholar]
  4. Neher E. The charge carried by single-channel currents of rat cultured muscle cells in the presence of local anaesthetics. J Physiol. 1983 Jun;339:663–678. doi: 10.1113/jphysiol.1983.sp014741. [DOI] [PMC free article] [PubMed] [Google Scholar]
  5. Prod'hom B., Pietrobon D., Hess P. Direct measurement of proton transfer rates to a group controlling the dihydropyridine-sensitive Ca2+ channel. Nature. 1987 Sep 17;329(6136):243–246. doi: 10.1038/329243a0. [DOI] [PubMed] [Google Scholar]
  6. Roux B., Sauvé R. A general solution to the time interval omission problem applied to single channel analysis. Biophys J. 1985 Jul;48(1):149–158. doi: 10.1016/S0006-3495(85)83768-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
  7. Sachs F., Auerbach A. Single-channel electrophysiology: use of the patch clamp. Methods Enzymol. 1983;103:147–176. doi: 10.1016/s0076-6879(83)03011-6. [DOI] [PubMed] [Google Scholar]
  8. Sigworth F. J., Sine S. M. Data transformations for improved display and fitting of single-channel dwell time histograms. Biophys J. 1987 Dec;52(6):1047–1054. doi: 10.1016/S0006-3495(87)83298-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  9. Wilson D. L., Brown A. M. Effect of limited interval resolution on single channel measurements with application to Ca channels. IEEE Trans Biomed Eng. 1985 Oct;32(10):786–797. doi: 10.1109/TBME.1985.325494. [DOI] [PubMed] [Google Scholar]
  10. Yeo G. F., Milne R. K., Edeson R. O., Madsen B. W. Statistical inference from single channel records: two-state Markov model with limited time resolution. Proc R Soc Lond B Biol Sci. 1988 Oct 22;235(1278):63–94. doi: 10.1098/rspb.1988.0063. [DOI] [PubMed] [Google Scholar]

Articles from Biophysical Journal are provided here courtesy of The Biophysical Society