Abstract
Osseointegration is critical for the long-term success of orthopedic and dental implants. This review highlights the biological mechanisms underlying bone-implant integration and recent advances in bioactive materials, drug delivery systems, and surface modification strategies. Implant design features—such as geometry, porosity, and coating technologies—are discussed alongside in vitro and in vivo evaluation models. Clinical applications span orthopedics, dentistry, prosthetics, and maxillofacial reconstruction. Emerging approaches, including nitric oxide-releasing coatings and controlled BMP-2 delivery, offer promising solutions to complications such as infection and ectopic ossification. These integrated strategies aim to enhance implant stability, functionality, and long-term clinical outcomes.
Keywords: biomedical engineering, bone-implant interface, implant surface modification, orthopedic and dental implants, osseointegration
1 ∣. Introduction
Orthopedic implants have been crucial in restoring mobility and enhancing the quality of life for patients with bone defects, fractures, and joint degeneration. Since their development, these implants have evolved significantly in both materials and design. The global orthopedic devices market is expected to reach nearly $72 billion by 2026, accounting for 12% of the projected global medical devices market [1]. In the United States alone, approximately 1.6 million people live with an amputation, with peripheral artery disease as the leading cause, followed by severe injury, tumors, and infections [2]. As the population ages and the prevalence of these conditions increases, the demand for implants that offer long-term biological stability has never been higher.
Osseointegration, the direct bonding and bone formation between an artificial implant and living bone without intervening fibrous tissue, is essential for the success of these implants. First described by Per-Ingvar Brånemark in the 1960s, osseointegration has transformed the field of orthopedic surgery and has become widely applied in dental implants, orthopedic reconstructions, and limb prosthetics [3]. Advances in osseointegration have driven the development of implant materials and surface modifications to enhance implant performance, integration, and longevity [4]. These modifications, including physical methods like grit blasting and plasma spraying, chemical techniques such as anodization and acid etching, and biological approaches like cell seeding and protein coating, aim to improve surface roughness, chemistry, and topography to promote cell adhesion, bone growth, and overall implant success [5]. However, the field is now shifting from these “static” 2D modifications toward functional 3D nanofiber constructs and biomimetic scaffolds that replicate the complex architecture of the natural extracellular matrix (ECM) to accelerate healing [6].
Bioactive materials, which can chemically bond with bone and promote cellular activity, have become a focus in advancing osseointegration. They include metals like titanium and stainless steel, ceramics such as hydroxyapatite (HA) and zirconia, polymers, and resorbable polyesters. This diverse range of biomaterials has its advantages and limitations in terms of biocompatibility, mechanical properties, and corrosion resistance [7]. The use of bioactive materials has expanded beyond orthopedics, finding applications in dentistry and maxillofacial prosthetics as well. Recent breakthroughs, such as the development of high-strength and tough Janus bionic periosteum, offer a solution by providing materials that match the dynamic mechanical and biological requirements of growing bone [8]. Furthermore, to combat the persistent threat of implant-associated infections, multifunctional coatings—such as gentamicin-loaded polydopamine on magnesium alloys—are being developed to induce bone regeneration while providing localized antibacterial properties [9].
Clinically, osseointegration has revolutionized multiple fields. Orthopedic implants, particularly in percutaneous osseointe-grated prostheses (POP), joint replacements and limb prostheses have greatly benefited from these advances, allowing for better patient outcomes and longer-lasting solutions [10]. Yet, even with successful bone bonding, the “skin-implant interface” remains a vulnerable point for infection migration. Innovations in craniofacial tissue engineering are now focusing on scalable bone regeneration to address these complex anatomical challenges [11]. For irregular bone defects that defy standard implant geometries, the emergence of kneadable dough-type hydrogels—which can transform from a dynamic to a rigid network in situ—represents a paradigm shift in reconstructive flexibility [12]. In dentistry, osseointegration has enabled the successful implantation of dental prosthetics, while in maxillofacial surgery, it has provided innovative options for facial reconstruction [13]. The potential of osseointegration continues to grow as it addresses the complexities of different medical conditions and injuries.
This review provides a comprehensive and critical overview of osseointegration, encompassing the underlying biological mechanisms, the evolution of implant materials and designs, surface modification strategies, preclinical evaluation models, and clinical applications in orthopedics, dentistry, and maxillofacial prosthetics. This work synthesizes recent advancements while emphasizing unresolved challenges and technical limitations, such as the management of medication-related osteonecrosis of the jaw (MRONJ) and the development of Janus bionic periosteum for pediatric growth [14]. Relevant literature was identified through PubMed, Scopus, and Web of Science databases using targeted search terms such as “osseointegration”, “implant biomaterials”, “biodegradable metals”, “surface modification”, “additive manufacturing”, “bone-implant interface”, “BMP-2 delivery”, “bioactive coatings”, “Janus bionic membranes”, and “dough-type hydrogels”. Priority was given to peer-reviewed studies that not only demonstrated innovation in materials and surface technologies but also offered mechanistic insights, translational relevance, or critical discussion of current barriers. This review aims to go beyond descriptive synthesis by performing a horizontal evaluation of state-of-the-art approaches—bridging the translational gap between preclinical success and long-term clinical survival—to highlight knowledge gaps and emerging strategies that can inform next-generation implant development and clinical translation.
2 ∣. Mechanism of Osseointegration
Osseointegration is the process in which bone integrates with an implant, typically a dental or orthopedic implant, forming a stable and functional connection between the bone and the artificial material. This process involves sequential key steps: (1) Initial Implant Contact, (2) Inflammation and Cellular Response, (3) Bone formation, and (4) Maturation and Remodeling.
2.1. ∣. Initial Implant Contact
When an implant is placed on the bony sockets, the implant surface is introduced to cells, including red blood cells, platelets, and inflammatory cells such as polymorphonuclear granulocytes and monocytes, that migrate from post-capillary venules into the surrounding tissue (Figure 1A) [15]. Platelets play a crucial role by adhering to the implant surface, forming a blood clot, and releasing growth factors and cytokines from their alphagranules. These include PDGF, TGF-β, and IGF-1, which initiate the healing and osseointegration process. Additionally, platelets interact with subendothelial matrix proteins via integrins, which are essential for adhesion and aggregation [16]. These interactions stabilize the blood clot and activate signaling pathways essential for subsequent cellular responses [17]. Furthermore, platelets significantly contribute to angiogenesis, which is crucial for bone healing by releasing VEGF and angiogenin.
FIGURE 1 ∣.

Mechanism of Osseointegration (A) Initial implant contact: The implant surface comes into contact with blood cells, including monocytes. Platelets adhere to the implant surface. (B) Inflammation and cellular response in fracture healing: Immune cells, including T-helper cells and macrophages, are recruited to the gap between the implant and bone. Angiogenesis begins, and mesenchymal stem cells proliferate and differentiate to initiate bone formation. (C) Bone formation. Osteoblasts, derived from mesenchymal stem cells, form a woven bone structure. (D) Maturation and remodeling. The initially fragile woven bone gradually transforms into a stronger lamellar bone.
2.2 ∣. Inflammation and Cellular Response in Fracture Healing
Immediately following a bone fracture, the inflammatory phase begins, marked by an influx of immune cells and the formation of a hematoma. This hematoma, composed of cellular debris, immune cells, cytokines, and platelets, forms at the fracture site to initiate the healing process. Within a few days of fracture, the proliferative phase begins, marked by the recruitment of mesenchymal stem cells (MSCs) that undergo proliferation and differentiation into preosteoblasts [18]. It is during this stage that angiogenesis begins with the recruitment of endothelial progenitors to form new vasculatures. Once vasculature is established, intramembranous bone formation begins, followed by the initiation of endochondral ossification. In intramembranous ossification, undifferentiated mesenchymal stem cells (MSCs) and committed osteoprogenitors from the periosteum directly form bone, leading to the development of a hard callus over the fracture gap (Figure 1B). In contrast, endochondral ossification involves the proliferation and differentiation of MSCs into cartilage, which later undergoes calcification and forms bone. Around 2 weeks after a fracture, the soft callus begins to resorb as mineralization initiates, leading to the formation of woven bone [19]. Advances in bone healing research have demonstrated the direct and critical role of immune cells in coordinating angiogenesis and osteogenesis in fracture healing. In mice lacking both T and B lymphocytes, fracture healing is accelerated as well in mice lacking B lymphocytes alone [20]. Conversely, γ/δ T cell-deficient mice have impaired fracture healing. Furthermore, Th17 cells drive osteoclastogenesis while regulatory T cells suppress osteoclast formation.
Immune cells play diverse roles in regulating angiogenesis. For example, Th1 cells secrete interleukin (IL)-12 and interferon (IFN)-γ, which exhibit anti-angiogenic and anti-proliferative effects on endothelial cells. In contrast, Th2 cytokines such as IL-6 and IL-17 from Th17 cells promote angiogenesis and stimulate endothelial cell proliferation. Under normal physiological conditions, bone marrow plasma cells also increase vasculature by releasing soluble factors such as vascular endothelial growth factor (VEGF) and insulin-like growth factor-1. During the phase corresponding to chondrocytes becoming hypertrophic and forming bone, there is a significant increase in immature myeloid cells (IMCs) at the fracture area, and they stay elevated until the fracture begins to mineralize [21].
After a bone fracture, there is extensive tissue damage that leads to a secretion of growth factors and cytokines such as SDF-1, BMP-2, and VEGF. The IMC isolated from the fracture area was positive for the major chemotactic factors; CXCR4, Flk-1 (VEGFR-2), and BMPR-1a [21]. These factors likely play a crucial role in mobilizing and enhancing the proliferation of IMCs and other cells, including osteoprogenitors, following a fracture.
2.3 ∣. Bone Formation
As described in previous sections, new bone formation following fractures involves a complex network of events involving major cell types: immune cells, endothelial cells, osteoblasts, and osteoclasts (Figure 1C). Immediately after injury, an inflammatory reaction occurs in the microenvironment, leading to the initiation of repair. The inflammatory milieu subsides within a couple of days to trigger angiogenic events, the failure of which leads to delayed union or non-union fractures [22]. In order to establish neovasculature at the site of hematoma, expression of angiogenic factors occurs within the fracture of the hematoma. The establishment of neovasculature leads to the deposition of hypertrophic chondrocytes, and when blood vessels reestablish within the callus, woven bone formation begins. The key proteins involved in new bone formation are bone morphogenetic proteins (BMPs), which belong to the transforming growth factor-beta (TGF-β) family that induces signal transduction through type I and type II receptors [23]. TGF-β signals via Smad proteins, which function as transcription factors for activating osteogenic gene expression. Out of the up to 20 types of BMPs, BMP-2 and BMP-7 are more potent enhancers of osseointegration [24].
2.4 ∣. Maturation and Remodeling
Initially formed woven bone is structurally weak due to the collagen fibrils being laid down in a randomly positioned manner [25]. Over time, the fragile woven bone gradually transitions into stronger and more organized lamellar bone by angiogenesis, osteoblast activity, and mineralization, enhancing its strength and stability (Figure 1D) [26]. This process, known as bone maturation, stabilizes the bone structure.
The bone tissue in contact with the implant undergoes morphological remodeling in response to mechanical stimulation, such as weight-bearing or movement-induced pressure [27]. Cortical bone forms along the fixture surface, several millimeters thick, with canaliculi that facilitate communication and nutrient exchange. A network of collagen bundles surrounds osteocytes and integrates into a glycoprotein layer, organizing the Haversian bone into osteons, which contribute to the density and homogeneity of the calcified bone. By the 4th week, the bone establishes complete binding along the implant surface. By the 12th week, the newly formed bone integrated into the lamellar bone, achieving a successful osseointegration with the surface of the implant [28].
Osseointegration begins with platelet activation and hematoma formation, followed by an inflammatory phase where immune cells and growth factors initiate angiogenesis. Mesenchymal stem cells then differentiate to form initial woven bone, which matures into organized lamellar bone through mechanical stimulation. This sequential process—spanning from early cellular response to long-term cortical remodeling—replaces fragile bone with dense osteons, ensuring a stable, functional interface and the structural integrity of the implant [29].
3 ∣. Bioactive Materials and Drug Delivery Systems
Achieving optimal osseointegration requires a strategic selection of materials that balance mechanical stability with biological signaling. Rather than viewing metals, ceramics, and polymers as isolated categories, modern implant design focuses on hybridizing these materials to overcome their individual limitations. This section evaluates the clinical tradeoffs between various biomaterials and examines drug delivery systems aimed at synchronizing bone formation with material presence.
3.1 ∣. Metals and Ceramics
The choice of biomaterial dictates initial cell-material interactions and determines long-term mechanical survival. Although traditional load-bearing applications have relied on the high strength of permanent metals, a critical analytical shift is occurring toward materials that offer bio-synchronous behavior [30, 31].
Metals are valued for their high strength and durability, making them ideal for load-bearing applications. Titanium (Ti) and its alloys offer a superior strength-to-weight ratio and low elastic modulus compared to stainless steel, reducing the risk of stress shielding [32]. However, a significant clinical drawback of these permanent metals is their inability to participate in the natural bone remodeling cycle. To address this, biodegradable metals like magnesium (Mg) and zinc (Zn) are emerging as alternatives [33]. As demonstrated by Li et al. Zn ions promote osteogenic differentiation and angiogenesis [34], ensuring that the gradual degradation of the implant is synchronized with the biological maturation and replacement of new bone tissue.
In contrast to the ductile nature of metals, ceramics provide high bioactivity but are limited by inherent brittleness [35]. Bioinert ceramics like Zirconia are favored in dental applications for esthetics, yet the true clinical value lies in bioactive and bioresorbable categories like hydroxyapatite (HA) and betatricalcium phosphate (β-TCP). A critical comparison reveals a clinical trade-off (Table 1): HA provides chemical stability but slows degradation, whereas β-TCP offers rapid resorption and higher osteoinductive capacity, often serving as a promising alternative to autografts [38].
TABLE 1.
Comparison of Metallic and Ceramic Implant Materials
| Material Class | Materials | Clinical Strength | Osseointegration Mechanism |
Clinical Limitation |
|---|---|---|---|---|
| Permanent Metals | Ti6Al4V, CoCr 32 | Exceptional fatigue resistance for high-load sites. | Mechanical interlocking; requires surface texturing. | Bio-inert; risk of stress shielding. |
| Biodegradable Metals | Mg alloys, Zn-metals 33 34 | Eliminates secondary surgery; releases bioactive ions. | Active biological stimulation of osteoblasts. | Difficult to control corrosion rate; early mechanical loss. |
| Bio-inert Ceramics | Zirconia, Aluminium Oxide37 | High wear resistance; excellent dental esthetics. | Direct bone-to-implant contact (BIC); chemical stability. | Brittle; potential for low-temp degradation (Zirconia). |
| Bioactive Ceramics | HA, β-TCP, OCP 38 36 | Mimicry of bone mineral; high osteoconductivity. | Chemical bonding; formation of a biological apatite layer. | Low tensile strength; brittle; cannot be standalone load-bearing. |
3.2 ∣. Polymers and Composite Scaffolds
Where metals offer strength and ceramics offer bioactivity, polymers provide the highest degree of structural and degradative tunability. Natural polymers (collagen, silk) provide native bioinformatic guidance that synthetic versions (PCL, PLA) lack [39]. However, a critical limitation is their lower mechanical strength and potential inflammatory responses from acidic degradation by-products [40].
High-performance polymers like PEEK match the elastic modulus of cortical bone but are inherently bio-inert. Consequently, as detailed by Su et al. [41], surface modification or the creation of composite scaffolds is required to initiate biological activity. These hybrids attempt to mimic the natural composite of bone by combining the toughness of polymers with the osteoconductivity of mineral phases (Table 2).
TABLE 2.
Comparison of Natural vs. Synthetic Polymers and Composites
| Material Type | Materials | Structural Adaptability |
Biocompatibility / Signaling |
Clinical Limitations |
|---|---|---|---|---|
| Natural Polymers | Collagen, Silk39 | Low (difficult to control architecture). | High: Provides native cues for cell adhesion. | Risk of immunogenicity; rapid degradation. |
| Synthetic Polymers | PCL, PLA, PLGA39 | High: Ideal for 3D printing and tunable degradation. | Low: Lacks cell-recognition motifs. | Acidic by-products can trigger inflammation. |
| Composites | PCL/HA, PEEK/Ceramic40 41 | Moderate: Balances toughness with bioactivity. | Moderate: Combines support with osteogenic minerals. | Achieving a stable interface between phases. |
3.3 ∣. Bioactive Coatings and Drug Delivery Strategies
Although the material scaffold provides the foundation, drug delivery systems (DDS) actively instruct the biological environment. These systems are generally categorized by their therapeutic goal: Osteoinduction (recruiting bone cells), angiogenesis (vascular support), or remodeling modulation (preventing bone loss). Currently available delivery systems range from simple physical adsorption of growth factors to sophisticated hydrogel coatings designed for cell-mediated release. Although these bioactive coatings aim to synchronize molecular signaling with the body's natural healing timeline, achieving precise control over release kinetics remains a significant hurdle. (Table 3) compares these primary delivery strategies, their biological targets, and the complications that arise when these systems fail to maintain long-term stability.
TABLE 3.
Comparison of Growth Factors and Drug Delivery Strategies
| Strategy | Key Bioactive Agents |
Biological Target (Section 2) |
Clinical Advantage | Long term Limitations |
|---|---|---|---|---|
| Osteoinductive Delivery | BMP-2, BMP-742 | Sec 2.3: Bone Formation | Potent induction of bone in large or non-union defects. | Safety: “Burst release” leads to ectopic bone and inflammation. |
| Angiogenic Signaling | VEGF42 | Sec 2.2:Vascularization | Establishes nutrient supply via capillary formation. | Stability: Short protein half-life; difficult to maintain. |
| Anti-Resorptive Delivery | Bisphosphonates43 | Sec 2.4:Remodeling | Maintains peri-implant density; critical for osteoporosis. | Remodeling: Risk of “frozen bone” and necrosis of the jaw. |
The “translational gap” identified here—specifically regarding the risk of BMP-2 burst release—highlights the limitations of current preclinical models. Many drug delivery strategies show exceptional results in 8–12 week animal studies, where the rapid metabolism of rodents masks the potential for chronic inflammation. In human clinical applications, success depends on synchronizing the release kinetics (Table 3) with the human bone remodeling cycle (Section 2.4). Failure to achieve this “bio-synchronicity” is a leading cause of late-stage implant failure, necessitating further exploration into hybrid systems that synergistically optimize both mechanical and biological performance [42, 43].
4 ∣. Emerging Innovations in Implant Surface Engineering and Osseointegration
Optimal implant design is crucial for achieving successful osseointegration and long-term stability. It necessitates careful consideration of factors such as physical shape, geometry, and porosity. To further enhance bioactivity and facilitate tissue integration, various surface modification techniques are employed [44] (Figure 2). For metallic implants, these include chemical treatments like acid etching, fluoride treatment, and ion implantation, which create surface topographies conducive to bone growth. Additionally, bioactive coatings, such as HA and tricalcium phosphate, are utilized to improve bone integration. Notably, the application of fibronectin fragments has shown promise in promoting integrin clustering and enhancing osteogenic signaling. Furthermore, advancements in additive manufacturing, particularly 3D printing, enable the creation of customized implants tailored to individual patient anatomy, optimizing load distribution and minimizing stress concentrations. Ultimately, a holistic approach that integrates these design and surface modification strategies is essential for maximizing implant performance and ensuring superior clinical outcomes [46].
FIGURE 2 ∣.

Different types of implant surface modifications to enhance osseointegration. Various surface modification techniques for Ti dental implants, including surface treatment (Sandblasting, Acid Etching, Anodization, Laser Radiation) and coating (Plasma Spraying, Nano Spray Drying, Sol–Gel, Hydrothermal, Self-Assembly, 3D Printed), along with considerations of physical shape and geometry. These techniques enhance bioactivity and promote long-term stability by manipulating surface topography, porosity, and biological modifications. Reproduced with permission from Sun X-D, et al. ACS Biomaterial Sci Eng. 2023; (9) 2023 American Chemical Society. Reference [45].
4.1 ∣. Physical Shape and Geometry
Primary stability, the initial mechanical stability of an implant upon placement, is largely determined by its macro-geometry and is essential for successful osseointegration [47]. Implant macrogeometry refers to the size, thread pattern, and shape of the implant. Increasing bone-implant contact (BIC) enhances stability and promotes more favorable osseointegration outcomes [48]. Increases in diameter and length of implants promote stability and decrease stress on both the bone and implant [49]. However, it has been shown that changes in length have less impact compared to changes in diameter [49]. The use of platform switching, when an abutment has a smaller diameter than the implant platform, can lead to reduced bone loss over time. Typical thread types are square, V, buttress/reverse buttress, and spiral [48]. The changes in the pitch and the depth of the thread, leading to more surface area, improve the primary stability of the implant. The last factor of implant geometry is the overall shape. Tapered or conical-shaped implants have been shown to have a higher success rate than cylindrical implants [47]. They are characterized by more positively distributed compressive forces on the implant, higher implant torque values, and less trauma during surgery. Overall, optimizing macro-geometric properties in implants is a crucial first step toward achieving long-term success in osseointegration.
4.2 ∣. Surface Topography
Implant surface topography plays a crucial role in osseointegration and long-term implant success. It can be categorized into three main levels: macro-roughness (> 100 μm), microroughness (1–10 μm), and nano-roughness (1–100 nm) [50]. Macro-level features, such as implant geometry and porous structures, enhance primary stability and long-term fixation by facilitating mechanical interlocking with bone. Microroughness promotes bone growth and interlocking at the implant interface, improving osteoblast adhesion, proliferation, and differentiation [51]. Nano-level modifications increase surface roughness, wettability, and surface-free energy, facilitating protein adsorption and cell adhesion. Research suggests that implants with moderately rough surfaces may provide optimal conditions for osseointegration, with an optimal range between 20 and 25 μm for titanium-based implants [50].
Various surface treatment methods are employed to achieve desired surface characteristics on the implant surface, with each method offering unique advantages and disadvantages. (1) Sandblasting and Acid Etching: This subtractive method, often used commercially, creates a multi-level rough surface that promotes both osteoblast adhesion and proliferation [52]. While highly effective for osseointegration, rougher surfaces can provide more attachment sites for bacteria, posing a risk for peri-implant infections. (2) Plasma Spraying: A thermal, additive process that deposits materials like HA or titanium dioxide (TiO2) [52]. It offers rapid deposition and low cost, but can suffer from phase impurities, poor adhesive strength, and high residual stress that risks coating delamination. (3) Physical Vapor Deposition (PVD): A versatile vacuum-based method for creating dense and uniform thin-film coatings with strong adhesion and enhanced bioactivity [52]. (4) Anodization and Microarc Oxidation (MAO): Electrochemical processes that create bioactive oxide films on the titanium surface with micro-nano pores [52]. These methods are highly effective at enhancing osteoblast adhesion and proliferation by creating a large specific surface area and improving wettability. (5) Laser Treatments: Laser-based techniques are a rapidly evolving method that offer precision and non-contact manufacturing for both texturing and coating applications [52, 53]. Laser treatments can create unique microstructures that improve osseointegration, enhance wettability, and produce a thick, contamination-free oxide layer. Studies show that specific laser-created structures can also reduce bacterial adhesion and biofilm formation, offering a dual benefit of improved osseointegration and antimicrobial properties [53, 54]. However, it's important to note that increased surface roughness may also be associated with increased transcription of reactive oxygen species (ROS)-related and metal-related cellular stress genes, potentially affecting long-term implant performance [43].
Successful osseointegration relies on a combination of implant design factors, including optimal geometry, porosity, and surface modifications. Through careful manipulation of these variables, such as increasing bone-implant contact, promoting tissue growth, and tailoring surface topography, implants can be designed to enhance stability and longevity. Ongoing research and innovation in these areas are crucial for improving clinical success and long-term functionality of orthopedic implants. Advancements in material science and manufacturing techniques continue to enable the development of implants that mimic natural bone structures more closely, ultimately optimizing the healing process.
4.3 ∣. Porosity
The primary role of porosity in orthopedic metal implants is to enhance tissue adhesion, growth, and vascularization by increasing the surface area available for cell attachment. The additive manufacturing method, a computer-controlled process that builds three-dimensional structures layer by layer, is widely used to introduce porosity into implants [55]. Additionally, techniques such as plasma spraying, sandblasting, acid etching, micro-arc oxidation (MAO), and sol–gel coating are commonly employed to create porous surfaces. The optimal porosity of implants varies depending on both the material and the specific type of bone. For instance, Ti-6Al-4 V alloy structures exhibit an effective modulus comparable to that of human cortical bone when their porosity ranges from 23% to 32% by volume [56]. In titanium implants, a porosity of 60%–70% has been identified as optimal for maximizing osseointegration. Similarly, trabecular bone-mimicking tantalum scaffolds with 70%–80% porosity have demonstrated superior bone ingrowth capabilities [57].
4.4 ∣. Bioactive Coatings and Modifications
Biological modifications include a range of strategies aimed at promoting implant integration by directly influencing cellular behavior and tissue response. (1) Bioactive Coatings: These coatings, composed of materials like hydroxyapatite (HA) or bioactive glass, promote osseointegration by mimicking the natural bone environment and stimulating bone cell growth. HA coatings function as a scaffold, facilitating bone cell attachment, growth, and mineral deposition, while bioactive glass forms an HA-like layer that enhances bone cell activity [44]. (2) Bioactive Molecules: The incorporation of extracellular matrix (ECM) proteins (e.g., fibronectin and collagen), bioactive peptides (e.g., RGD sequences), and growth factors (e.g., BMPs and VEGF) enhances cell adhesion, proliferation, and signaling. Growth factors, in particular, can be incorporated into coatings to stimulate osteoblast differentiation and vascularization, enhancing bone formation and remodeling [58, 59]. (3) Strontium-Based Coatings: Strontium (Sr) has emerged as a particularly promising element for implant coatings, especially for patients with systemic conditions like osteoporosis [60]. A recent systematic review and meta-analysis of animal studies found that Sr-coated titanium implants significantly improved osseointegration in osteoporotic models [60]. The positive effect is attributed to strontium's unique dual mechanism of action, which simultaneously increases osteoblast activity and decreases osteoclast activity [60]. (4) Immunomodulatory Strategies: Implant surfaces can also be modified to modulate the host immune response and create a pro-healing environment [44]. By controlling macrophage polarization and reducing the inflammatory response, these surfaces can promote a constructive healing process rather than a destructive one. Strategies include incorporating anti-inflammatory molecules or designing surface topographies that guide immune cell behavior to favor tissue regeneration.
4.5 ∣. Three-Dimensionally Printed Implant
A 3D-printed implant is one of the significant developments in medical technology, offering improved integration with the patient's bone tissue. Compared to traditional implants, they provide several advantages. The ability to customize implants and materials to match each patient's specific anatomy enhances functionality and patient outcomes by allowing for more natural load distribution and reducing stress concentrations. The layer-by-layer manufacturing process enables exceptional precision in creating intricate structures, whereas rapid prototyping accelerates design iteration and refinement [61].
A key feature of 3D-printed implants is the incorporation of lattice structures, which mimic the natural architecture of bone and promote osseointegration. These structures enhance bone ingrowth, improve implant stability, and reduce the risk of implant loosening by providing a scaffold for new bone formation. Additionally, they help distribute mechanical stress more evenly, minimizing stress shielding effects and improving longterm durability. The ability to fabricate such intricate lattice structures is a major advantage of 3D printing over traditional manufacturing methods [62].
Furthermore, 3D printing has enabled the development of multi-material implants, which combine varied materials within a single device to optimize performance. This approach enhances wear resistance in articulating surfaces, improves biocompatibility in areas that contact bone tissue, and allows for tailored mechanical properties throughout the implant. For example, a single implant may incorporate cobalt chrome for its wear-resistant properties while utilizing titanium for its superior biocompatibility. By integrating these material combinations, 3D-printed implants achieve greater durability, functionality, and patient-specific adaptability.
Another emerging trend is the integration of artificial intelligence (AI) with 3D-printing workflows. AI-driven design automation allows for the rapid generation and optimization of patient-specific implants using imaging data such as MRI or CT scans. The development of smart implants equipped with embedded sensors, along with 5D/6D printing technologies for adaptive, sensor-integrated, or stimuli-responsive devices, is expanding the scope of clinical applications. In addition, the use of hybrid and bioresorbable materials, such as polycaprolactone (PCL) and titanium alloys, offers improved biocompatibility and functional versatility. These advancements shorten production timelines from weeks to days, facilitating faster and less invasive surgeries with higher precision. As a result, modern 3D-printed implants not only increase patient safety and reduce recovery times but also lay the foundation for the next generation of intelligent, responsive implant systems [63-65].
Through a strategic combination of mechanical, chemical, and biological surface modification techniques, it is possible to significantly enhance implant osseointegration, leading to improved long-term implant success. Each modification method offers unique advantages, and the optimal approach depends on the specific requirements of the implant and the patient. Ongoing research continues to refine these techniques, bringing us closer to creating implants that seamlessly integrate with living tissue.
4.6 ∣. Clinical and Experimental Validation of Implant Design Modifications
A diverse range of experimental and clinical studies further validates the practical application of the aforementioned material and bioactive modifications. As summarized in (Table 4), various macro-design and surface alterations—ranging from optimized thread geometries to porous structural scaffolds—consistently demonstrate superior primary stability compared to conventional, non-modified implant designs. These studies, conducted across various animal models and human clinical trials, highlight that the strategic integration of these modifications leads to significantly higher bone-to-implant contact (BIC) and improved load distribution. By enhancing the immediate mechanical interlocking and long-term biological response, these advancements provide the empirical foundation for current osseointegration protocols.
TABLE 4.
Comparison of Implant Designs and Primary Stability:
| Study | Methods used in the study | Implant Designs and modifications |
Outcomes |
|---|---|---|---|
| Kuroshima et al.66 | Titanium alloy implants were placed in the proximal tibial metaphysis of rabbits and repetitive mechanical loading was applied | −60° and +60° grooves around the neck | +60° grooves significantly increased bone-to-implant contact, bone mass, and bone mineral density with mechanical loading |
| Vandamme et al.67 | Implants positioned in a bone chamber of a rabbit proximal tibia are compared through four hundred loading cycles and histological analysis | Cylindrical and screw implants under stress and screw implant without stress as control | Both bone-to-implant and osteoid-to-implant contact was highest on the stress loaded screw implants |
| Muktadar et al.68 | Hisological comparison of implants placed in rabbit right distal condyle of the femur | Self-cutting V-thread and power thread designs with 0.4 and 0.6 mm thread depth with high and low insertion torque | Significantly higher bone-to-implant contact is observed in the high torque power-shaped threads and high new bone formation shown in high torque V-shaped threads |
| Hsu et al.69 | Using microcomputed tomography constructed 3D implant models implanted in rigid cellular polyurethan blocks to determine BIC area | Diameter of 3.75, 4, 5, and 6mm | The 6 mm diameter implants produced the highest bone-to-implant contact area |
| Tardelli at al.70 | Implants were evaluated by insertion torque, pullout testing, and photoelasticity on polyurethane blocks | 3.5mm diameter conical implants with either double thread or double thread/progressive thread and 2.9mm cylindrical implant with double threads | Both conical implants present higher primary stability and better stress distribution |
| Hong et al.71 | Histometric analysis, removal torque testing, topographical analysis, and resonance frequency analysis was completed on samples from the mandibles of beagles | Threaded titanium implant with a core diameter of 3.25mm or a core diameter of 1.25mm and a 1mm width porous scaffold housed between the threads | The addition of the porous titanium structure between the threads of the implant resulted in significant increases in bone-to-implant contact of the total implant area and the apical portion of the test group |
| Gehrke et al.72 | Implant samples from sheep mandibles underwent insertion torque testing, histological analysis, and marginal bone level measurement | Conical design implants with trapezoidal threads or trapezoidal threads with “healing chambers” distributed between the threads inserted at different torque values with or without a fresh socket | Higher primary bone-to-implant contact was found in the implants inserted at lower torque values with “healing chambers” in both conditions |
| Fanali et al.73 | Periotest and insertion torque values measured for implants inserted into high-density rigid polyurethane blocks | Five implant groups with rounded or beveled apex, troncoconical or V thread, wide or narrow furrows/thread pitch, and differing diameters | Implants with a round apex and wide V-thread pitch showed higher primary stability |
| Cao et al.74 | Cadaveric femora were used to virtually simulate and measure osseointegrated prosthesis implantation of different femoral length amputations | Implant diameter difference of ± 1mm I shot, medium, and long amputation lengths | Medium and long amputation lengths lead to significantly higher bone-to-implant contact than the short amputation length with an increase of implant diameter size in the short length resulting in higher bone-to-implant contact |
| Makary et al.75 | Min implants were used in a randomized controlled trial in the human posterior maxilla and underwent histomorphometric analysis | Nanostructured calcium-incorporated surface titanium plants with or without plasma surface treatment | Plasma surface treated implants showed a high bone-to-implant contact level, presence of newly formed bone, and high levels of osteoconduction |
4.7 ∣. Challenges and Outlook
Despite significant progress, several limitations and future research directions remain. While advanced surface modifications like nano-structured surfaces and bioactive coatings show promise in promoting early-stage osseointegration and preventing bacterial adhesion, long-term studies validating their clinical durability, infection resistance, and effects on immune modulation are scarce [44, 52, 60]. The interaction between these advanced coatings (e.g., NO-releasing, peptide-modified) and the host immune response remains poorly characterized, presenting a major barrier to widespread adoption.
Future research should focus on the development of multifunctional surfaces that seamlessly integrate both antimicrobial and osteoinductive properties without triggering adverse inflammation [60]. Additionally, the standardization of preclinical models for evaluating these features is urgently needed to facilitate more reliable comparisons across studies. Furthermore, robust clinical validation in large animal models and long-term human trials is a crucial next step for many of these novel materials and advanced delivery systems [44, 53]. This comprehensive approach will be essential for ensuring the long-term success and predictability of next-generation implant materials.
5 ∣. Feasibility Evaluation Using In Vitro and In Vivo Models
In vitro and in vivo models are both necessary tools for pre-clinical research into osseointegration (Figure 3). Culturing osteoblasts for in vitro adhesion, proliferation, and matrix mineralization has been found to be successful in modeling bone formation. These typical models are done by seeding the osteoblast cell cultures onto different implant materials and/or surface properties [77]. Osteoblast cultures on implant surfaces are commonly assessed using adhesion assays, proliferation assays, calcium quantification, Northern blot analysis, and scanning electron microscopy to evaluate osseointegration [78]. However, in vitro testing has limitations, as it cannot fully replicate the complex environment in which the implant will function. This has led to increased interest in ex vivo or explant models, which offer a more accurate representation of osseointegration outside of animal studies [79]. Bone specimens are harvested from long bones of animals and cultured alongside different implant materials. This can be done in a bioreactor or other structured models [80]. These models, particularly when samples are harvested from animals used in unrelated research or from a single animal, help reduce the need for invasive animal studies while enhancing knowledge and understanding of osseointegration [79].
FIGURE 3 ∣.

In vitro and in vivo results of implant osseointegration. (I) Fluorescence microscopy images of human osteoblast cells after incubation with the 10 min eluates of a titanium implant (a) and a zirconium implant (b) after staining with fluorescein diacetate (FDA) and propidium iodide. The presence of green color without red indicates only living cells. (II) Close-up SEM images of a titanium implant (a) and a zirconium implant (b) seeded with human osteoblast cells (magnification 1000×). (III) Acid fuchsin and methylene blue staining on the 12th week after surgery. Larger direct contact between the new bone and the hydroxyapatite (HA) coated implant could be observed in groups A and B than in groups C and D. The upper part of the picture (black) is the implant. The intermediate part (dark gray) is the HA coating. The lower part is the newly formed bone. * The white arrows indicate the interface between the new bone and the HA-coated implant. Image I & II by Möller B et.al. from Int J Oral Maxillofac Surg. 2012 May; 41 (5) licensed under CC BY [35]. Image IV: Reproduced with permission from K H, et.al. ACS Biomaterial Sci Eng. 2023; (9) 2023 American Chemical Society [76].
After in vitro and ex vivo modeling is complete, the results must be confirmed in animal experiments. In vivo modeling for osteointegration is done in the long bones of rats, rabbits, dogs, goats, sheep, and primates [81]. During implantation surgery, intraoperative precautions need to be followed to minimize tissue damage that would lead to reduced bone healing. These include avoiding high drill speeds, gradually increasing drill size, reducing the implant-bone gap, and keeping a low internal temperature of the bone [76]. Typically, a mature bone-to-implant connection is established after 12 weeks; however, within just 16 days, the implant surface becomes covered with mineralized tissue [28]. Testing after sample collection includes radiographic analysis, histological analysis, and mechanical assessment. Histological analysis provides knowledge of bone quality, contact percentages, type of bone, osteocyte characteristics, immune response, and vascularization. Radiographical analysis is used as a less invasive way to quantify osteointegration. Analyses include tomography, electron microscopy, Micro-CT, energy dispersive X-ray spectroscopy, and others [82]. Lastly, mechanical testing can be done to assess the physical properties of the bone-implant interface. These tests include push-out or pull-out test, applying lateral load, and the reverse torque test at the implant fixture [83]. Resonance frequency analysis has also been used to determine the stiffness of the bone-implant interface; the stiffer the interface, the higher the presence of dense bone formation [84].
5.1 ∣. The Translational Gap
A critical challenge in implant research is the translational gap, where promising results in preclinical models often fail to manifest as long-term clinical success in humans. This discrepancy is primarily rooted in the physiological divergence between small animal models and human patients. As discussed in Section 2.4, the human bone remodeling cycle is a protracted process typically spanning 6 months, encompassing phases of resorption, reversal, and formation [85]. In contrast, small animal models, such as rats or rabbits, exhibit a significantly accelerated metabolism, with a mature bone-to-implant connection often established within just 12 weeks [86]. While these models are effective for evaluating short-term success such as initial cell adhesion and early-stage woven bone formation, they fail to accurately replicate the long-term remodeling-driven failure that can occur in humans over years of functional loading. The integration of large animal models, such as dogs, sheep, and pigs, provides a more representative simulation of human clinical conditions due to their comparable bone macrostructure and load-bearing profiles. However, even in these larger models, differences in bone composition and healing rates persist, as they often demonstrate faster turnover rates that can mask late-stage complications like stress shielding or aseptic loosening. Consequently, current preclinical evaluations may overestimate the long-term stability of bioactive coatings or drug delivery systems by focusing on a window that corresponds only to the initial phase of the human integration timeline. Bridging this gap requires a fundamental shift toward longitudinal studies in both small and large animal models that account for the synchronicity between implant degradation or release kinetics and the significantly slower human bone turnover rate.
6 ∣. Clinical Applications
6.1 ∣. Orthopedic Applications
Orthopedic implants have been utilized for decades, and their demand is expected to rise due to an aging population, increased physical activity, and longer life expectancy [87]. These implants are commonly used in total joint arthroplasty (TJA), fracture fixation, and spinal surgery. TJA is one of the most frequently performed surgical procedures worldwide, with implant fixation achieved through cemented, cementless, or hybrid methods [88].
Cementless fixation depends on biological integration. It provides initial mechanical stability through press-fitting (Table 4) and is followed by secondary stabilization via osseointegration, the process in which bone grows into the implant surface to form a durable bond [89]. In total hip arthroplasty (THA), cementless fixation is widely preferred, especially for younger patients. This is due to several advantages, such as lowering all revision rates, elimination of bone cement implantation syndrome, and shorter operative times. However, it does carry a higher risk of periprosthetic fractures [90].
In Total Knee Replacement (TKA), cemented fixation has long been accepted as the gold standard due to its reliability, broad applicability, and ability to compensate for minor bone imperfections. However, as TKA procedures become more common among younger, more active, and obese patients, and as cementless implant designs improve, cementless TKA is gaining attention. Similar to THA, it offers shorter operative times and may reduce the risk of osteolysis due to third-body wear [91]. While some studies show comparable survivorship and outcomes for both methods, others indicate a higher risk of aseptic loosening with cementless fixation.
For fracture fixation, the primary goal is to stabilize the fracture, facilitate bone healing, and restore function. This can be achieved using intramedullary (IM) nails, plates, and screws, with the choice depending on the type and location of the fracture, as well as patient-specific factors. IM nails are commonly used for long-bone diaphyseal fractures because their load-sharing design allows for early mobilization and weight-bearing [92]. Plates and screws are crucial in fracture fixation, with two commonly used types: compression plates and locking plates. The versatility of these tools is particularly beneficial in fracture fixation. Compression plates provide absolute stability and are primarily used for periarticular fractures, where precise anatomical healing is essential. In contrast, locking plates offer relative stability and are particularly useful when perfect alignment is not critical for good function, such as in diaphyseal fractures, especially in osteoporotic bone [93].
Similar principles of implant stability and bone integration apply in spine surgery, where osseointegrated implants like interbody fusion cages, pedicle screw systems, and rods play a crucial role in achieving spinal stability and promoting fusion [94]. Interbody fusion cages are placed in the intervertebral disc space after disc removal, restoring disc height, widening the intervertebral foramen, and facilitating bone integration [95]. Pedicle screws provide firm fixation, stabilizing fractures, correcting deformities, and supporting fusion procedures by restoring spinal alignment. Spinal rods, used across all spinal segments, help maintain physiological curvature and enhance spinal stability [96]. In addition to implants for structural support, bone grafts and bone substitutes play a crucial role in enhancing bone healing and integration. These materials are especially valuable in cases where natural healing is insufficient or when bone defects are present. They are widely used in oncologic surgery, trauma, revision TJA, and spinal procedures to support bone formation. Among the available options, autografts are considered the gold standard in orthopedics and are typically harvested from the iliac crest or fibula. Autogenous bone grafts possess essential properties for bone formation, including osteoconductivity, osteoinductivity, and osteogenic potential. However, their use is associated with potential complications such as blood loss, hematoma formation, and neurovascular injury [97].
6.2 ∣. Prosthetic Applications
The clinical success of limb reconstruction is defined by the stability of the bone-implant interface and the long-term management of the percutaneous interface (skin-penetrating opening). Currently, three major systems—OPRA (Osseointegrated Prostheses for the Rehabilitation of Amputees), ILP (Integral Leg Prosthesis), and OPL (Osseointegrated Prosthetic Limb)—dominate the clinical landscape, each presenting unique design features and specific clinical trade-offs [98]. Additionally, Percutaneous Osseointegrated Prostheses (POPs) have been developed to provide a direct connection between the residual limb and the prosthetic device, specifically designed to avoid the stress shielding of distal bone that often leads to resorption over time [99, 100].
6.2.1 ∣. Comparison of Fixation Logic: Screw-Fit vs. Press-Fit
A horizontal evaluation reveals a fundamental divergence in fixation logic: Screw-fit (biological emphasis) versus Press-fit (mechanical emphasis) (Table 5). (1) Screw-Fit (OPRA): As the most extensively studied system, the OPRA device utilizes a threaded, screw-type implant and a conservative two-stage surgical protocol [101]. While this method demonstrates a survival rate of 83% at 10 years and 72% at 15 years, its primary limitation is mechanical. The long healing period (up to 6 months for full weight-bearing) and the stress concentrations on the threads lead to a higher risk of fixture or abutment breakage compared to other designs [101]. (2) Press-Fit (ILP and OPL): These systems employ a friction-fit mechanism similar to modern hip arthroplasty, allowing for significantly faster rehabilitation. ILP patients generally achieve full weight-bearing within 2 to 3 months [102]. The OPL, a modified version of the ILP, offers a one-stage surgical procedure and unrestricted weight-bearing, though it may take patients over a year to fully regain proprioception [103]. Although press-fit designs reduce the risk of mechanical failure, they introduce a potential risk of stress shielding due to high stem stiffness, a challenge the POP system specifically attempts to mitigate [99, 101].
Table 5.
Comparison of Leading Clinical Prosthetic Implant Systems
| System | Fixation Logic | Surgical Protocol | Rehab to Weight-Bearing | Clinical Limitation |
|---|---|---|---|---|
| OPRA103 | Screw-Fit (Threaded) | Two-Stage | ~6 Months | High mechanical breakage risk; long rehabilitation. |
| ILP104 | Press-Fit(Smooth/Coated) | One or Two-Stage | 2–3 Months | Risk of stress shielding, infection. |
| OPL105 | Press-Fit (Bio-contoured) | One-Stage | Immediate/Unrestricted | Slow proprioceptive recovery (>1 year). |
| POP101,102 | Specialized (Press-fit) | Variable | Variable | Avoids stress shielding; limited long-term data. |
6.2.2 ∣. Long-Term Clinical Outcomes
Clinical data indicates a recurring disparity between the high functional success rates reported in animal models or short-term (1-year) human trials and the complications that emerge during long-term follow-up. While osseointegration at the bone level is typically successful, the implant-skin interface remains the primary point of clinical failure. This vulnerability is evidenced by the 15-year survival drop in the OPRA system—from 83% to 72%—and the persistent incidence of infection in ILP cohorts, highlighting that the transcutaneous exit site is the most significant barrier to long-term success [101, 102]. Because short-term experiments often fail to capture the late-stage bacterial migration from the skin to the bone, which can lead to osteomyelitis, next-generation prosthetics must prioritize the establishment of a robust “biological seal” at the skin level to match the integration achieved within the bone.
6.3 ∣. Dental Implant Applications
Dental implants are the gold standard for tooth replacement, offering durability, function, and esthetics. Unlike traditional crown and bridge prosthetics, implants integrate with bone, preserving structure and supporting restorations. Advances in implantology have led to various implant types suiting different clinical needs: (1) Endosteal Implants: The most common implants, surgically placed in the jawbone, serve as artificial roots for crowns, bridges, or full-arch restorations. Their high success rate and versatility make them widely preferred. (2) Subperiosteal Implants: Placed above the jawbone but beneath the gingiva, these were historically used for severe bone resorption cases. However, their use has declined due to advancements in bone grafting and more stable implant options. (3) Zygomatic Implants: Anchored in the cheekbone, these implants are used in severe maxillary bone loss cases where conventional implants are unfeasible. Additionally, mini-implants are used for temporary anchorage, overdenture stabilization, and orthodontics. Their minimally invasive placement is advantageous, though they cannot support single crowns or extensive restorations [104].
Success depends on bone quality and loading protocol. Adequate bone volume and density are essential for osseointegration. Bone grafting may be required in cases of deficiency. Implant loading can be immediate or delayed, depending on bone stability and the need to minimize micromotion. Despite high success rates, complications such as peri-implantitis and implant failure occur [105].
6.3.1 ∣. Complications and Consideration
(1)Peri-implantitis: This inflammatory condition leads to bone loss and implant failure if untreated. Risk factors include poor hygiene, smoking, and systemic conditions like diabetes. A 2022 study found a 15.9% peri-implantitis rate among 132 patients, with smokers having a 72.7% incidence [106]. Preventive measures, such as professional maintenance and patient education, are crucial. (2) Implant failure with systemic conditions: One pilot study examines the impact of osteoporosis and diabetes on dental implant stability over 12 months. It found that osteoporosis patients had higher initial stability but negatively impacted long-term success. Diabetes affects osseointegration due to delayed healing and inflammation, and increased infection risk makes implant failure more common in diabetic conditions. Personalized treatment strategies are necessary to enhance implant success [107]. Dental implants provide an effective solution for tooth replacement, but their long-term success relies on effective osseointegration, which depends on bone quality, patient health, and proper maintenance. Osseointegration is influenced by factors such as systemic conditions, bone remodeling capacity, and implant surface modifications. Complications like peri-implantitis and conditions like diabetes and osteoporosis highlight the need for individualized treatment approaches and close monitoring. Advances in biomaterials, surface treatments, and loading protocols continue to enhance osseointegration, leading to improved implant longevity and patient outcomes.
6.4 ∣. Maxillofacial Prosthetic Applications
Bone defects in the craniomaxillofacial region can be derived from trauma, malignant tumors, severe infections, oncologic surgery, or congenital disorders. Surgical reconstruction may have higher success rates in replacing the lost tissue over a prosthetic. However, depending on the availability of donor tissue or the psychophysical condition of the patients, maxillofacial prosthetics might be a better approach. For example, in the case of cancer, the use of prostheses may be beneficial if tumor surveillance is desired to prevent the recurrence of the disease [108].
7 ∣. Future Perspectives
Despite significant advancements in osseointegration, several challenges persist, notably infection, aseptic loosening, and stress shielding. The most common complication is a superficial skin infection, though more severe cases like osteomyelitis can lead to implant removal in 9% of cases [109]. To counter this, new methods are emerging, such as nitric oxide (NO)-releasing coatings, which show promise in dramatically reducing biomaterial-associated infections [110]. While traditional antibiotic-loaded bone cement is limited by its short-lived effects [111], new approaches use nanoparticles and hydrogels to provide a sustained release of therapeutic compounds like BMP-2 [112], ensuring better biological activity and absorption by osteoprogenitor cells [113].
Aseptic loosening, the second most common issue, results from the failure of initial fixation [114], leading to wear particles that trigger inflammation and bone resorption [115]. Over 10% of patients require revision surgery within 15 years for this reason [116]. Strategies to prevent it include using longer, thicker stems and porous coatings to enhance stability. Stress shielding occurs when a rigid implant prevents a natural load distribution on the bone, leading to localized bone resorption [117]. Mitigation involves designing implants with features like hollow stems and using flexible, isoelastic materials [118].
Emerging technologies are also being developed. Immunomodulation seeks to control the immune response by adjusting specific pathways to prevent chronic inflammation, which can lead to fibrous encapsulation and implant loosening [113]. Researchers are exploring ways to sequentially promote M1 then M2 polarization or release specific ions to “educate” immune cells toward tissue-supportive roles [119, 120]. Additionally, AI-aided design is revolutionizing implant planning. By integrating data from 3D imaging, AI algorithms can automatically segment anatomical structures, suggest ideal implant placement, and generate surgical guides [121, 122]. Beyond static anatomical fitting, the next frontier lies in “Personalized Synchronicity.” Future AI frameworks will move toward predicting a patient's unique bone remodeling cycle—influenced by age, metabolic activity, and mechanical loading—rather than relying on generalized data. By integrating these predictive insights with advanced 3D printing, we can fabricate stimuli-responsive, patient-specific implants. These next-generation devices will be engineered to synchronize their degradation or drug-release kinetics with the host's biological clock, ensuring that structural support and bio-inductive cues are provided in tandem with the human 6-month remodeling phases.
Achieving widespread clinical use for these innovations requires close interdisciplinary collaboration. This includes partnerships between biomaterials scientists, surgeons, and veterinary specialists for clinically relevant preclinical studies. For regulatory readiness, collaboration with regulatory scientists is essential to ensure compliance with standards like Good Laboratory Practice (GLP) and Good Manufacturing Practice (GMP). Finally, engaging health economists is vital to ensure new technologies are not only clinically effective but also economically viable for widespread adoption. This cross-disciplinary approach will accelerate the journey from lab discovery to safe and effective clinical solutions.
8 ∣. Conclusion
This review outlines key strategies for enhancing osseointegration, emphasizing the biological mechanisms, material innovations in orthopedics, implant dentistry, and clinical approaches that underpin successful bone-implant integration. We highlight the sequential cellular events that govern boneimplant integration and detail advancements in bioactive materials, drug delivery systems, and surface engineering strategies designed to enhance osseointegration. Implant design features, such as geometry and porosity, along with evaluation methods used in vitro, ex vivo, and in vivo models, are also discussed to underscore their importance in preclinical validation. Clinically, these technologies are increasingly applied across orthopedics, prosthetics, dentistry, and maxillofacial reconstruction. To address ongoing challenges such as infection and ectopic ossification, emerging strategies—including nitric oxide (NO)-releasing coatings for their anti-inflammatory and antibacterial effects, and controlled delivery of osteoinductive factors like BMP-2 offer promising solutions [123]. Together, these multidisciplinary approaches are shaping the future of implant science to achieve more reliable, long-lasting, and patient-specific outcomes.
Acknowledgments
This work was supported by National Institutes of Health (NIH) grants R01HL163802 and Alabama Research and Development Enhancement Fund: 1ARDEF22 09 (PI: Ho-Wook Jun) as well as Alabama Research and Development Enhancement Fund: 1ARDEF23 02 and NIH R21DE034497 (PI: Kyounga Cheon).
Funding
This work was supported by National Institutes of Health (R01HL163802, R21DE034497), Alabama Research and Development Enhancement Fund (1ARDEF22 09, 1ARDEF23 02).
Footnotes
Conflicts of Interest
The authors declare no conflicts of interest.
Data Availability Statement
The data that support the findings of this study are available from the corresponding author upon reasonable request.
References
- 1.Nadeem M and Weiss AC, “Medical Product Development Part 1: Idea, Partnerships, and Commercialization,” Journal of Hand Surgery. American Volume 46 (2021): 703–708, 10.1016/j.jhsa.2021.04.020. [DOI] [PubMed] [Google Scholar]
- 2.Farrokhi S, Perez K, Eskridge S, and Clouser M, “major Deployment-Related Amputations of Lower and Upper Limbs, Active and Reserve Components, U.S. Armed Forces, 2001–2017,” Military Surgeon 25 (2018): 10–16. [PubMed] [Google Scholar]
- 3.Agarwal S, Mistry L, Mistry S, et al. , “The Per-Ingvar Branemark Era (1929-2014): Evolution of a no Compromise Prosthetic Dental Replacement,” Cureus 16 (2024): e71708, 10.7759/cureus.71708. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Liang W, Zhou C, Bai J, et al. , “Current Developments and Future Perspectives of Nanotechnology in Orthopedic Implants: An Updated Review,” Frontiers in Bioengineering and Biotechnology 12 (2024): 1342340, 10.3389/fbioe.2024.1342340. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Zhu G, Wang G, and Li JJ, “Advances in Implant Surface Modifications to Improve Osseointegration,” Materials Advances 2 (2021): 6901–6927, 10.1039/d1ma00675d. [DOI] [Google Scholar]
- 6.Zhang Z, Li H, Qian M, et al. , “Up IGF-I via High-Toughness Adaptive Hydrogels for Remodeling Growth Plate of Children,” Regenerative Biomaterials 12 (2025): 4, 10.1093/rb/rbaf004. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Schierano G, Mussano F, Faga MG, et al. , “An Alumina Toughened Zirconia Composite for Dental Implant Application: In Vivo Animal Results,” BioMed Research International 2015 (2015): 157360, 10.1155/2015/157360. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Elumalai K, Srinivasan S, and Shanmugam A, “Review of the Efficacy of Nanoparticle-Based Drug Delivery Systems for Cancer Treatment,” Biomedical Technology 5 (2024): 1, 10.1016/j.bmt.2023.09.001. [DOI] [Google Scholar]
- 9.Wang N, Chen J, Chen Y, et al. , “Kneadable Dough-Type Hydrogel Transforming From Dynamic to Rigid Network to Repair Irregular Bone Defects,” Bioactive Materials 40 (2024): 21, 10.1016/j.bioactmat.2024.06.021. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Thesleff A, Branemark R, Hakansson B, and Ortiz-Catalan M, “Biomechanical Characterisation of Bone-Anchored Implant Systems for Amputation Limb Prostheses: A Systematic Review,” Annals of Biomedical Engineering 46 (2018): 377–391, 10.1007/s10439-017-1976-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Vignolo SM, Roth DM, Wu L, Cosgrove J, and Bertassoni LE, “Strategies for Craniofacial Tissue Engineering: Innovations for Scalable Bone Regeneration,” Plastic and Aesthetic Research 12 (2025): 9, 10.20517/2347-9264.2025.09. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Kamaraj M, Moghimi N, Chen J, et al. , “New Dimensions of Electrospun Nanofiber Material Designs for Biotechnological Uses,” Trends in Biotechnology 42 (2023): 8, 10.1016/j.tibtech.2023.11.008. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Bohara S and Suthakorn J, “Surface Coating of Orthopedic Implant to Enhance the Osseointegration and Reduction of Bacterial Colonization: A Review,” Biomaterials Research 26 (2022): 26, 10.1186/s40824-022-00269-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Patel PA, Nassar SI, Nguyen SA, et al. , “The Use of Bone Morphogenic Protein and Related Therapies in the Management of Medication-Related Osteonecrosis of the Jaw: A Scoping Review,” Plastic and Aesthetic Research 12 (2025): 13, 10.20517/2347-9264.2024.153. [DOI] [Google Scholar]
- 15.Mavrogenis AF, Dimitriou R, Parvizi J, and Babis GC, “Biology of Implant Osseointegration,” Journal of Musculoskeletal & Neuronal Interactions 9 (2009): 61–71. [PubMed] [Google Scholar]
- 16.Pandey C, Rokaya D, and Bhattarai BP, “Contemporary Concepts in Osseointegration of Dental Implants: A Review,” BioMed Research International 2022 (2022): 6170452, 10.1155/2022/6170452. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Park JY and Davies JE, “Red Blood Cell and Platelet Interactions With Titanium Implant Surfaces,” Clinical Oral Implants Research 11 (2000): 530–539, 10.1034/j.1600-0501.2000.011006530.x. [DOI] [PubMed] [Google Scholar]
- 18.Wang X, Wang Y, Gou W, Lu Q, Peng J, and Lu S, “Role of Mesenchymal Stem Cells in Bone Regeneration and Fracture Repair: A Review,” International Orthopaedics 37 (2013): 2491–2498, 10.1007/s00264-013-2059-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Dimitriou R, Tsiridis E, and Giannoudis PV, “Current Concepts of Molecular Aspects of Bone Healing,” Injury 36 (2005): 1392–1404, 10.1016/j.injury.2005.07.019. [DOI] [PubMed] [Google Scholar]
- 20.Toben D, Schroeder I, El Khassawna T, et al. , “Fracture Healing Is Accelerated in the Absence of the Adaptive Immune System,” Journal of Bone and Mineral Research 26 (2011): 113–124, 10.1002/jbmr.185. [DOI] [PubMed] [Google Scholar]
- 21.Levy S, Feduska JM, Sawant A, Gilbert SR, Hensel JA, and Ponnazhagan S, “Immature Myeloid Cells Are Critical for Enhancing Bone Fracture Healing Through Angiogenic Cascade,” Bone 93 (2016): 113–124, 10.1016/j.bone.2016.09.018. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Schmidt-Bleek K, Schell H, Lienau J, et al. , “Initial Immune Reaction and Angiogenesis in Bone Healing,” Journal of Tissue Engineering and Regenerative Medicine 8 (2014): 120–130, 10.1002/term.1505. [DOI] [PubMed] [Google Scholar]
- 23.Poynton AR and Lane JM, “Safety Profile for the Clinical Use of Bone Morphogenetic Proteins in the Spine,” Spine 27 (2002): S40–S48, 10.1097/00007632-200208151-00010. [DOI] [PubMed] [Google Scholar]
- 24.Dolanmaz D, Saglam M, Inan O, et al. , “Monitoring Bone Morphogenetic Protein-2 and −7, Soluble Receptor Activator of Nuclear Factor-kappaB Ligand and Osteoprotegerin Levels in the Peri-Implant Sulcular Fluid During the Osseointegration of Hydrophilic-Modified Sandblasted Acid-Etched and Sandblasted Acid-Etched Surface Dental Implants,” Journal of Periodontal Research 50 (2015): 62–73, 10.1111/jre.12182. [DOI] [PubMed] [Google Scholar]
- 25.Clarke B, “Normal Bone Anatomy and Physiology,” Clinical Journal of the American Society of Nephrology 3 (2008): S131–S139, 10.2215/CJN.04151206. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Shapiro F and Wu JY, “Woven Bone Overview: Structural Classification Based on Its Integral Role in Developmental, Repair and Pathological Bone Formation Throughout Vertebrate Groups,” European Cells and Materials 38 (2019): 137–167, 10.22203/eCM.v038a11. [DOI] [PubMed] [Google Scholar]
- 27.Schulte FA, Ruffoni D, Lambers FM, et al. , “Local Mechanical Stimuli Regulate Bone Formation and Resorption in Mice at the Tissue Level,” PLoS One 8 (2013): e62172, 10.1371/journal.pone.0062172. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Depprich R, Zipprich H, Ommerborn M, et al. , “Osseointegration of Zirconia Implants: An SEM Observation of the Bone-Implant Interface,” Head & Face Medicine 4 (2008): 25, 10.1186/1746-160X-4-25. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Hadjidakis DJ and Androulakis II, “Bone Remodeling,” Annals of the New York Academy of Sciences 1092 (2006): 385–396, 10.1196/annals.1365.035. [DOI] [PubMed] [Google Scholar]
- 30.Anderson JM, Patterson JL, Vines JB, Javed A, Gilbert SR, and Jun H-W, “Biphasic Peptide Amphiphile Nanomatrix Embedded With Hydroxyapatite Nanoparticles for Stimulated Osteoinductive Response,” 5 (2011): 9463, 10.1021/nn203247. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Chen J, Deng L, Porter C, et al. , “Angiogenic and Osteogenic Synergy of Human Mesenchymal Stem Cells and Human Umbilical Vein Endothelial Cells Cocultured on a Nanomatrix,” Scientific Reports 8 (2018): 2, 10.1038/s41598-018-34033-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Lewallen EA, Riester SM, and Bonin CA, “Biological Strategies for Improved Osseointegration and Osteoinduction of Porous Metal Orthopedic Implants,” Tissue Engineering Part B: Reviews 21 (2015): 218–230, 10.1089/ten.teb.2014.0333. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Istrate B, Cojocaru F-D, Henea M-E, et al. , “In Vitro and In Vivo Analysis of the Mg-Ca-Zn Biodegradable Alloys,” Journal of Functional Biomaterials 15 (2024): 166, 10.3390/jfb15060166. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Li P, Dai J, and Li Y, “Zinc Based Biodegradable Metals for Bone Repair and Regeneration: Bioactivity and Molecular Mechanisms,” Materials Today Bio 25 (2023): 100932, 10.1016/j.mtbio.2023.100932. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Möller B, Terheyden H, Açil Y, et al. , “A Comparison of Biocompatibility and Osseointegration of Ceramic and Titanium Implants: An In Vivo and In Vitro Study,” International Journal of Oral and Maxillofacial Surgery 41 (2012): 638–645, 10.1016/j.ijom.2012.02.004. [DOI] [PubMed] [Google Scholar]
- 36.Thomé G, Sandgren R, Bernardes S, et al. , “Osseointegration of a Novel Injection Molded 2-Piece Ceramic Dental Implant: A Study in Minipigs,” Clinical Oral Investigations 25 (2021): 603–615, 10.1007/s00784-020-03513-z. [DOI] [PubMed] [Google Scholar]
- 37.Brochu BM, Sturm SR, Kawase JA De Queiroz Goncalves, et al. , “Advances in Bioceramics for Bone Regeneration: A Narrative Review,” Biomimetics 9 (2024): 690, 10.3390/biomimetics9110690. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Shan J, Wang S, Xu H, et al. , “Incorporation of Cerium Oxide Into Zirconia Toughened Alumina Ceramic Promotes Osteogenic Differentiation and Osseointegration,” Journal of Biomaterials Applications 36 (2022): 36535, 10.1177/08853282211036535. [DOI] [PubMed] [Google Scholar]
- 39.Ye B, Wu B, Su Y, Sun T, and Guo X, “Recent Advances in the Application of Natural and Synthetic Polymer-Based Scaffolds in Musculoskeletal Regeneration,” Polymers 14 (2022): 4566, 10.3390/polym14214566. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Ershad-Langroudi A, Babazadeh N, Alizadegan F, Mousaei SM, and Moradi G, “Polymers for Implantable Devices,” Journal of Industrial and Engineering Chemistry 137 (2024): 30, 10.1016/j.jiec.2024.03.030. [DOI] [Google Scholar]
- 41.Buck E, Li H, and Cerruti M, “Surface Modification Strategies to Improve the Osseointegration of Poly(Etheretherketone) and Its Composites,” Macromolecular Bioscience 20 (2020): e1900271, 10.1002/mabi.201900271. [DOI] [PubMed] [Google Scholar]
- 42.Hunziker EB, Enggist L, Küffer A, Buser D, and Liu Y, “Osseointegration: The Slow Delivery of BMP-2 Enhances Osteoinductivity,” Bone 51, no. 1 (2012): 98–106, 10.1016/j.bone.2012.04.004. [DOI] [PubMed] [Google Scholar]
- 43.Lewallen EA, Trousdale WH, and Thaler R, “Surface Roughness of Titanium Orthopedic Implants Alters the Biological Phenotype of Human Mesenchymal Stromal Cells,” Tissue Engineering: Part A 27 (2021): 1503–1516, 10.1089/ten.TEA.2020.0369. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44.Shayeb MA, Elfadil S, Abutayyem H, et al. , “Bioactive Surface Modifications on Dental Implants: A Systematic Review and Meta-Analysis of Osseointegration and Longevity,” Clinical Oral Investigations 28 (2024): 592, 10.1007/s00784-024-05958-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45.Sun X-D, Liu T-T, Wang Q-Q, Zhang J, and Cao M-S, “Surface Modification and Functionalities for Titanium Dental Implants,” ACS Biomaterials Science & Engineering 9 (2023): 4442–4461, 10.1021/acsbiomaterials.3c00183. [DOI] [PubMed] [Google Scholar]
- 46.Agarwal R and García AJ, “Biomaterial Strategies for Engineering Implants for Enhanced Osseointegration and Bone Repair,” Advanced Drug Delivery Reviews 94 (2015): 53–62, 10.1016/j.addr.2015.03.013. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 47.Herrero-Climent M, Lemos BF, Herrero-Climent F, et al. , “Influence of Implant Design and Under-Preparation of the Implant Site on Implant Primary Stability. An In Vitro Study,” International Journal of Environmental Research and Public Health 17 (2020): 4436. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 48.Ibrahim A, Heitzer M, Bock A, et al. , “Relationship Between Implant Geometry and Primary Stability in Different Bony Defects and Variant Bone Densities: An In Vitro Study,” Materials 13 (2020): 13194349, 10.3390/ma13194349. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 49.Georgiopoulos B, Kalioras K, Provatidis C, Manda M, and Koidis P, “The Effects of Implant Length and Diameter Prior to and After Osseointegration: A 2-D Finite Element Analysis,” Journal of Oral Implantology 33 (2007): 243–256, 10.1563/1548-1336(2007)33[243:Teoila]2.0.Co;2. [DOI] [PubMed] [Google Scholar]
- 50.Barfeie A, Wilson J, and Rees J, “Implant Surface Characteristics and Their Effect on Osseointegration,” British Dental Journal 218 (2015): E9, 10.1038/sj.bdj.2015.171. [DOI] [PubMed] [Google Scholar]
- 51.Shalabi MM, Gortemaker A, Van't Hof MA, Jansen JA, and Creugers NH, “Implant Surface Roughness and Bone Healing: A Systematic Review,” Journal of Dental Research 85 (2006): 496–500, 10.1177/154405910608500603. [DOI] [PubMed] [Google Scholar]
- 52.Bakitian FA, “A Comprehensive Review of the Contemporary Methods for Enhancing Osseointegration and the Antimicrobial Properties of Titanium Dental Implants,” Cureus 16 (2024): e68720, 10.7759/cureus.68720. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 53.Kolarovszki B, Ficsor S, Frank D, Katona K, Soos B, and Turzo K, “Unlocking the Potential: Laser Surface Modifications for Titanium Dental Implants,” Lasers in Medical Science 39 (2024): 162, 10.1007/s10103-024-04076-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 54.Gittens RA, Olivares-Navarrete R, Schwartz Z, and Boyan BD, “Implant Osseointegration and the Role of Microroughness and Nanostructures: Lessons for Spine Implants,” Acta Biomaterialia 10 (2014): 3363–3371, 10.1016/j.actbio.2014.03.037. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 55.Song C, Liu L, Deng Z, et al. , “Research Progress on the Design and Performance of Porous Titanium Alloy Bone Implants,” Journal of Materials Research and Technology 23 (2023): 2626–2641, 10.1016/j.jmrt.2023.01.155. [DOI] [Google Scholar]
- 56.Bandyopadhyay A, Espana F, Balla VK, Bose S, Ohgami Y, and Davies NM, “Influence of Porosity on Mechanical Properties and In Vivo Response of Ti6Al4V Implants,” Acta Biomaterialia 6 (2010): 1640–1648, 10.1016/j.actbio.2009.11.011. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 57.Jiao J, Hong Q, Zhang D, et al. , “Influence of Porosity on Osteogenesis, Bone Growth and Osteointegration in Trabecular Tantalum Scaffolds Fabricated by Additive Manufacturing,” Frontiers in Bioengineering and Biotechnology 11 (2023): 1117954, 10.3389/fbioe.2023.1117954. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 58.Anderson JM, Vines JB, Patterson JL, Chen H, Javed A, and Jun H-W, “Osteogenic Differentiation of Human Mesenchymal Stem Cells Synergistically Enhanced by Biomimetic Peptide Amphiphiles Combined With Conditioned Medium,” Acta Biomaterialia 7 (2011): 16, 10.1016/j.actbio.2010.08.016. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 59.Anderson JM, Kushwaha M, Tambralli A, Bellis SL, Camata RP, and Jun H-W, “Osteogenic Differentiation of Human Mesenchymal Stem Cells Directed by Extracellular Matrix-Mimicking Ligands in a Biomimetic Self-Assembled Peptide Amphiphile Nanomatrix,” Biomacromolecules 10 (2009): 2935–2944, 10.1021/bm9007452. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 60.Sayed O, Abdalla MM, Elsayed A, El-Mahallawy Y, and Al-Mahalawy H, “Does Strontium Coated Titanium Implants Enhance the Osseointegration in Animal Models Under Osteoporotic Condition? A Systematic Review and Meta-Analysis,” BDJ Open 10 (2024): 69, 10.1038/s41405-024-00220-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 61.Meng M, Wang J, Huang H, Liu X, Zhang J, and Li Z, “3D Printing Metal Implants in Orthopedic Surgery: Methods, Applications and Future Prospects,” Journal of Orthopaedic Translation 42 (2023): 94–112, 10.1016/j.jot.2023.08.004. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 62.Park JW, Park H, Kim JH, Kim HM, Yoo CH, and Kang HG, “Fabrication of a Lattice Structure With Periodic Open Pores Through Three-Dimensional Printing for Bone Ingrowth,” Scientific Reports 12 (2022): 17223, 10.1038/s41598-022-22292-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 63.Mobarak MH, Islam MA, Hossain N, et al. , “Recent Advances of Additive Manufacturing in Implant Fabrication—A Review,” Applied Surface Science Advances 18 (2023): 100462, 10.1016/j.apsadv.2023.100462. [DOI] [Google Scholar]
- 64.Goldsmith I and Goldsmith I, “3D Bioprinting the Human Chest Wall: Fiction or Fact,” Npj Biomedical Innovations 1 (2024): 2, 10.1038/s44385-024-00002-w. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 65.Cong B and Zhang H, “Innovative 3D Printing Technologies and Advanced Materials Revolutionizing Orthopedic Surgery: Current Applications and Future Directions,” Frontiers in Bioengineering and Biotechnology 13 (2025): 1542179, 10.3389/fbioe.2025.1542179. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 66.Kuroshima S, Nakano T, and Ishimoto T, “Optimally Oriented Grooves on Dental Implants Improve Bone Quality Around Implants Under Repetitive Mechanical Loading,” Acta Biomaterialia 48 (2017): 21, 10.1016/j.actbio.2016.11.021. [DOI] [PubMed] [Google Scholar]
- 67.Vandamme K, Naert I, and Geris L, “Influence of Controlled Immediate Loading and Implant Design on Peri-Implant Bone Formation,” Journal of Clinical Periodontology 34 (2007): 172, 10.1111/j.1600-051X.2006.01014.x. [DOI] [PubMed] [Google Scholar]
- 68.Muktadar AK, Gangaiah M, and Chrcanovic BR, “Evaluation of the Effect of Self-Cutting and Nonself-Cutting Thread Designed Implant With Different Thread Depth on Variable Insertion Torques: An Histomorphometric Analysis in Rabbit,” Clinical Implant Dentistry and Related Research 20 (2018): 12611, 10.1111/cid.12611. [DOI] [PubMed] [Google Scholar]
- 69.Hsu JT, Shen YW, and Kuo CW, “Impacts of 3D Bone-To-Implant Contact and Implant Diameter on Primary Stability of Dental Implant,” Journal of the Formosan Medical Association 116 (2017): 5, 10.1016/j.jfma.2017.05.005. [DOI] [PubMed] [Google Scholar]
- 70.Tardelli JDC, Valente MLC, Macedo AP, and Reis AC, “Evaluation of Biomechanical and Stress Distribution of Different Dental Implant Designs: Primary Stability and Photoelastic Analysis,” IRBM 43 (2022): 3, 10.1016/j.irbm.2021.01.003. [DOI] [Google Scholar]
- 71.Hong JY, Ko SY, and Lee W, “Enhancement of Bone Ingrowth Into a Porous Titanium Structure to Improve Osseointegration of Dental Implants: A Pilot Study in the Canine Model,” Materials 13 (2020): 13143061, 10.3390/ma13143061. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 72.Gehrke SA, Júnior JA, and Treichel TL, “Effects of Insertion Torque Values on the Marginal Bone Loss of Dental Implants Installed in Sheep Mandibles,” Scientific Reports 12 (2022): 5, 10.1038/s41598-021-04313-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 73.Fanali S, Tumedei M, Pignatelli P, Petrini M, Piattelli A, and Iezzi G, “The Effect of Threads Geometry on Insertion Torque (IT) and Periotest Implant Primary Stability: A High-Density Polyurethane Simulation for the Anterior Mandible,” Crystals 11 (2021): 308. [Google Scholar]
- 74.Cao M, Doherty DB, and He H, “Osseointegration for Transfemoral Amputees: Influence of Femur Length and Implant Sizing on BoneImplant Contact and Alignment,” Injury 56 (2025): 112162, 10.1016/j.injury.2025.112162. [DOI] [PubMed] [Google Scholar]
- 75.Makary C, Menhall A, and Lahoud P, “Bone-To-Implant Contact in Implants With Plasma-Treated Nanostructured Calcium-Incorporated Surface (XPEEDActive) Compared to Non-Plasma-Treated Implants (XPEED): A Human Histologic Study at 4 Weeks,” Materials 17 (2024): 2331, 10.3390/ma17102331. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 76.Huang K, Wu T, Lou J, et al. , “Impact of Bone-Implant Gap Size on the Interfacial Osseointegration: An In Vivo Study,” BMC Musculoskeletal Disorders 24 (2023): 115, 10.1186/s12891-023-06215-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 77.Ahmad M, McCarthy M, and Gronowicz G, “An In Vitro Model for Mineralization of Human Osteoblast-Like Cells on Implant Materials,” Biomaterials 20 (1999): 211–220, 10.1016/S0142-9612(98)00152-5. [DOI] [PubMed] [Google Scholar]
- 78.Dee KC, Rueger DC, Andersen TT, and Bizios R, “Conditions Which Promote Mineralization at the Bone-Implant Interface: A Model In Vitro Study,” Biomaterials 17 (1996): 209–215, 10.1016/0142-9612(96)85765-6. [DOI] [PubMed] [Google Scholar]
- 79.Veronesi F, Torricelli P, Martini L, et al. , “An Alternative Ex Vivo Method to Evaluate the Osseointegration of Ti-6Al-4V Alloy Also Combined With Collagen,” Biomedical Materials 16 (2021): 025007, 10.1088/1748-605X/abdbda. [DOI] [PubMed] [Google Scholar]
- 80.Hall TAG, Theodoridis K, Kohli N, Cegla F, and van Arkel RJ, “Active Osseointegration in an Ex Vivo Porcine Bone Model,” Frontiers in Bioengineering and Biotechnology 12 (2024): 12, 10.3389/fbioe.2024.1360669. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 81.Sivolella S, Brunello G, Ferroni L, et al. , “A Novel In Vitro Technique for Assessing Dental Implant Osseointegration,” Tissue Engineering, Part C, Methods 22 (2016): 132–141, 10.1089/ten.tec.2015.0158. [DOI] [PubMed] [Google Scholar]
- 82.Shah FA, Stenlund P, Martinelli A, Thomsen P, and Palmquist A, “Direct Communication Between Osteocytes and Acid-Etched Titanium Implants With a Sub-Micron Topography,” Journal of Materials Science. Materials in Medicine 27 (2016): 167, 10.1007/s10856-016-5779-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 83.Frosch S, Nüsse V, Frosch K-H, Lehmann W, and Buchhorn G, “Osseointegration of 3D Porous and Solid Ti–6Al–4V Implants—Narrow Gap Push-Out Testing and Experimental Setup Considerations,” Journal of the Mechanical Behavior of Biomedical Materials 115 (2021): 104282, 10.1016/j.jmbbm.2020.104282. [DOI] [PubMed] [Google Scholar]
- 84.Meredith N, Alleyne D, and Cawley P, “Quantitative Determination of the Stability of the Implant-Tissue Interface Using Resonance Frequency Analysis,” Clinical Oral Implants Research 7 (1996): 261–267, 10.1034/j.1600-0501.1996.070308.x. [DOI] [PubMed] [Google Scholar]
- 85.Kenkre J, Bassett J, and JS Kenkre JB, “The Bone Remodelling Cycle,” Annals of Clinical Biochemistry 55 (2018): 9371, 10.1177/0004563218759371. [DOI] [PubMed] [Google Scholar]
- 86.Scarano A, Khater AGA, Gehrke SA, Inchingolo F, and Tari SR, “Animal Models for Investigating Osseointegration: An Overview of Implant Research Over the Last Three Decades,” Journal of Functional Biomaterials 15 (2024): 83, 10.3390/jfb15040083. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 87.Sloan M, Premkumar A, and Sheth NP, “Projected Volume of Primary Total Joint Arthroplasty in the U.S., 2014 to 2030,” Journal of Bone and Joint Surgery. American Volume 100 (2018): 1455–1460, 10.2106/JBJS.17.01617. [DOI] [PubMed] [Google Scholar]
- 88.Sharma M, “Cementless Total Hip Arthroplasty,” in Hip Arthroplasty: Current and Future Directions, ed. Sharma M(Springer Nature Singapore, 2023), 61–91. [Google Scholar]
- 89.Kheir MM, Drayer NJ, and Chen AF, “An Update on Cementless Femoral Fixation in Total Hip Arthroplasty,” Journal of Bone and Joint Surgery. American Volume 102 (2020): 1646–1661, 10.2106/JBJS.19.01397. [DOI] [PubMed] [Google Scholar]
- 90.Li J, Zhang M, Yao J, Shao L, Fang C, and Cheng CK, “Risk Factors for Periprosthetic Femoral Fractures After Cementless Total Hip Arthroplasty,” Journal of Arthroplasty 39 (2024): 2547–2554, 10.1016/j.arth.2024.06.005. [DOI] [PubMed] [Google Scholar]
- 91.Schwabe MT and Hannon CP, “The Evolution, Current Indications and Outcomes of Cementless Total Knee Arthroplasty,” Journal of Clinical Medicine 11, no. 22 (2022): 6608, 10.3390/jcm11226608. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 92.Bong MR, Kummer FJ, Koval KJ, and Egol KA, “Intramedullary Nailing of the Lower Extremity: Biomechanics and Biology,” Journal of the American Academy of Orthopaedic Surgeons 15 (2007): 97–106, 10.5435/00124635-200702000-00004. [DOI] [PubMed] [Google Scholar]
- 93.Egol KA, Kubiak EN, Fulkerson E, Kummer FJ, and Koval KJ, “Biomechanics of Locked Plates and Screws,” Journal of Orthopaedic Trauma 18 (2004): 488–493, 10.1097/00005131-200409000-00003. [DOI] [PubMed] [Google Scholar]
- 94.Warburton A, Girdler SJ, Mikhail CM, Ahn A, and Cho SK, “Biomaterials in Spinal Implants: A Review,” Neurospine 17 (2020): 101–110, 10.14245/ns.1938296.148. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 95.Fogel G, Martin N, Williams GM, et al. , “Choice of Spinal Interbody Fusion Cage Material and Design Influences Subsidence and Osseointegration Performance,” World Neurosurgery 162 (2022): e626–e634, 10.1016/j.wneu.2022.03.087. [DOI] [PubMed] [Google Scholar]
- 96.Costachescu B, Niculescu AG, Grumezescu AM, and Teleanu DM, “Screw Osteointegration-Increasing Biomechanical Resistance to Pull-Out Effect,” Materials 16 (2023): 5582, 10.3390/ma16165582. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 97.Campana V, Milano G, Pagano E, et al. , “Bone Substitutes in Orthopaedic Surgery: From Basic Science to Clinical Practice,” Journal of Materials Science. Materials in Medicine 25 (2014): 2445–2461, 10.1007/s10856-014-5240-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 98.Hoellwarth JS, Tetsworth K, Akhtar MA, and Al Muderis M, “The Clinical History and Basic Science Origins of Transcutaneous Osseointegration for Amputees,” Advances in Orthopedics 2022 (2022): 7960559, 10.1155/2022/7960559. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 99.Jeyapalina S, Beck JP, Bachus KN, Williams DL, and Bloebaum RD, “Efficacy of a Porous-Structured Titanium Subdermal Barrier for Preventing Infection in Percutaneous Osseointegrated Prostheses,” Journal of Orthopaedic Research 30 (2012): 1304–1311, 10.1002/jor.22081. [DOI] [PubMed] [Google Scholar]
- 100.Miller A, Jeyapalina S, Agarwal JP, and Beck JP, “Association Between Blood Markers and the Progression of Osseointegration in Percutaneous Prostheses Patients-A Pilot Study,” Journal of Biomedical Materials Research. Part B, Applied Biomaterials 112 (2024): e35398, 10.1002/jbm.b.35398. [DOI] [PubMed] [Google Scholar]
- 101.Hagberg K, Ghasemi Jahani SA, Omar O, and Thomsen P, “Osseointegrated Prostheses for the Rehabilitation of Patients With Transfemoral Amputations: A Prospective Ten-Year Cohort Study of Patient-Reported Outcomes and Complications,” Journal of Orthopaedic Translation 38 (2023): 56–64, 10.1016/j.jot.2022.09.004. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 102.Reetz D, Atallah R, Mohamed J, van de Meent H, Frolke JPM, and Leijendekkers R, “Safety and Performance of Bone-Anchored Prostheses in Persons With a Transfemoral Amputation: A 5-Year Follow-Up Study,” Journal of Bone and Joint Surgery. American Volume 102 (2020): 1329–1335, 10.2106/JBJS.19.01169. [DOI] [PubMed] [Google Scholar]
- 103.Al Muderis M, Lu W, and Li JJ, “Osseointegrated Prosthetic Limb for the Treatment of Lower Limb Amputations: Experience and Outcomes,” Der Unfallchirurg 120 (2017): 306–311, 10.1007/s00113-016-0296-8. [DOI] [PubMed] [Google Scholar]
- 104.Buser D, Sennerby L, and Bruyn HD, “Modern Implant Dentistry Based on Osseointegration: 50 Years of Progress, Current Trends and Open Questions,” Periodontology 2000 73 (2017): 12185, 10.1111/prd.12185. [DOI] [PubMed] [Google Scholar]
- 105.Insua A, Galindo-Moreno P, and Miron RJ, “Emerging Factors Affecting Peri-Implant Bone Metabolism,” Periodontology 2000 94 (2024): 12532, 10.1111/prd.12532. [DOI] [PubMed] [Google Scholar]
- 106.Astolfi V, Ríos-Carrasco B, Gil-Mur FJ, et al. , “Incidence of Peri-Implantitis and Relationship With Different Conditions: A Retrospective Study,” International Journal of Environmental Research and Public Health 19 (2022): 4147, 10.3390/ijerph19074147. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 107.Sachelarie L, Scrobota I, and Cioara F, “The Influence of Osteoporosis and Diabetes on Dental Implant Stability: A Pilot Study,” Medicina 61 (2025): 74, 10.3390/medicina61010074. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 108.Verbist M, Vandevelde AL, Geusens J, Sun Y, Shaheen E, and Willaert R, “Reconstruction of Craniomaxillofacial Bone Defects With 3D-Printed Bioceramic Implants: Scoping Review and Clinical Case Series,” Journal of Clinical Medicine 13 (2024): 13, 10.3390/jcm13102805. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 109.Nablo BJ, Prichard HL, Butler RD, Klitzman B, and Schoenfisch MH, “Inhibition of Implant-Associated Infections via Nitric Oxide Release,” Biomaterials 26 (2005): 6984–6990, 10.1016/j.biomaterials.2005.05.017. [DOI] [PubMed] [Google Scholar]
- 110.Thomson AL, Suhardi VJ, Niu Y, et al. , “A Translational Murine Model of Aseptic Loosening With Osseointegration Failure,” Journal of Orthopaedic Research 42 (2024): 2525–2534, 10.1002/jor.25915. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 111.Hwang P, Shin CM, Sherwood JA, et al. , “A Multi-Targeting Bionanomatrix Coating to Reduce Capsular Contracture Development on Silicone Implants,” Biomaterials Research 27 (2023): 34, 10.1186/s40824-023-00378-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 112.Lee D, Wufuer M, Kim I, et al. , “Sequential Dual-Drug Delivery of BMP-2 and Alendronate From Hydroxyapatite-Collagen Scaffolds for Enhanced Bone Regeneration,” Scientific Reports 11 (2021): 746, 10.1038/s41598-020-80608-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 113.Vantucci CE, Krishan L, Cheng A, Prather A, Roy K, and Guldberg RE, “BMP-2 Delivery Strategy Modulates Local Bone Regeneration and Systemic Immune Responses to Complex Extremity Trauma,” Biomaterials Science 9 (2021): 1668–1682, 10.1039/d0bm01728k. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 114.Drees P, Eckardt A, Gay RE, Gay S, and Huber LC, “Mechanisms of Disease: Molecular Insights Into Aseptic Loosening of Orthopedic Implants,” Nature Clinical Practice. Rheumatology 3 (2007): 165–171, 10.1038/ncprheum0428. [DOI] [PubMed] [Google Scholar]
- 115.Overmann AL and Forsberg JA, “The State of the Art of Osseointegration for Limb Prosthesis,” Biomedical Engineering Letters 10 (2020): 5–16, 10.1007/s13534-019-00133-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 116.Liverani E, Rogati G, Pagani S, Brogini S, Fortunato A, and Caravaggi P, “Mechanical Interaction Between Additive-Manufactured Metal Lattice Structures and Bone in Compression: Implications for Stress Shielding of Orthopaedic Implants,” Journal of the Mechanical Behavior of Biomedical Materials 121 (2021): 104608, 10.1016/j.jmbbm.2021.104608. [DOI] [PubMed] [Google Scholar]
- 117.Xi W, Hegde V, Zoller SD, et al. , “Point-Of-Care Antimicrobial Coating Protects Orthopaedic Implants From Bacterial Challenge,” Nature Communications 12 (2021): 5473, 10.1038/s41467-021-25383-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 118.Koons GL, Kontoyiannis PD, Diba M, Chim LK, Scott DW, and Mikos AG, “Effect of 3D Printing Temperature on Bioactivity of Bone Morphogenetic Protein-2 Released From Polymeric Constructs,” Annals of Biomedical Engineering 49 (2021): 2114–2125, 10.1007/s10439-021-02736-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 119.Huo S, Liu S, Liu Q, et al. , “Copper–Zinc-Doped Bilayer Bioactive Glasses Loaded Hydrogel With Spatiotemporal Immunomodulation Supports MRSA-Infected Wound Healing,” Advanced Science 11 (2024): 2302674, 10.1002/advs.202302674. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 120.Xiang E, Vaquette C, Liu S, et al. , “Biomimetic Surface Nanoengineering of Biodegradable Zn-Based Orthopedic Implants for Enhanced Biocompatibility and Immunomodulation,” Advanced Functional Materials 34 (2024): 34, 10.1002/adfm.202410033. [DOI] [Google Scholar]
- 121.Schubert O, Schweiger J, and Stimmelmayr M, “Digital Implant Planning and Guided Implant Surgery—Workflow and Reliability,” British Dental Journal 226 (2019): 44, 10.1038/sj.bdj.2019.44. [DOI] [PubMed] [Google Scholar]
- 122.Elgarba BM, Fontenele RC, Du X, et al. , “Artificial Intelligence Versus Human Intelligence in Presurgical Implant Planning: A Preclinical Validation,” Clinical Oral Implants Research 36 (2025): 835–845, 10.1111/clr.14429. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 123.Moon C-Y, Nam OH, Kim M, et al. , “Effects of the Nitric Oxide Releasing Biomimetic Nanomatrix Gel on Pulp-Dentin Regeneration: Pilot Study,” PLoS One 13 (2018): e0205534, 10.1371/journal.pone.0205534. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
The data that support the findings of this study are available from the corresponding author upon reasonable request.