Abstract
Background
Data on long-term outcomes and durability of the ACURATE neo valve are limited. This study aimed to obtain long-term outcome and echocardiographic data among ACURATE neo recipients.
Methods
Consecutive patients undergoing transfemoral transcatheter aortic valve implantation for severe native aortic stenosis from 4 centers using the ACURATE neo valve between 2012 and 2018 (n = 961) were included. Exclusion criteria for the durability analysis were 30-day mortality, valve-in-valve implantation or surgical aortic valve replacement within 30 days, and lack of echocardiographic follow-up (FU) beyond 30 days, leaving 758 patients included in the durability cohort. Primary endpoints were the cumulative incidence of late bioprosthetic valve failure (BVF), structural valve deterioration (SVD), and Kaplan–Meier estimates of all-cause mortality.
Results
In the overall cohort (median age of 82 years, 61.3% females, median and maximum FU time of 5.0 and 11.3 years), estimates of all-cause mortality at 1, 5, and 8 years were 13.6, 43.1, and 67.3%, respectively. In the durability cohort, the cumulative incidence was 3.5% for moderate SVD, 2.1% for severe SVD, and 3.2% for BVF at 8 years. Paired FU echocardiography beyond 5 years (n = 243, median FU time of 6.4 years) showed mean aortic valve gradients and effective orifice area of 9 mmHg and 1.8 cm2 post-transcatheter aortic valve implantation and 6 mmHg and 1.7 cm2 at final FU echo.
Conclusions
In this multicenter observational cohort treated with the ACURATE neo platform, 8-year survival was 32.7%, and the cumulative incidence of severe SVD (2.1%) and BVF (3.2%) was limited. These findings provide long-term durability estimates in an elderly, real-world population.
Keywords: Degeneration, Durability, Long-term outcome, Structural valve deterioration
Graphical abstract
Graphical Abstract: ACURATE neo long-term outcomes. Mortality was analyzed in the survival cohort (n = 961), SVD and BVF were assessed in the durability cohort (n = 758), and valve performance ≥5 years was determined in a subset of patients with paired echocardiography follow up exceeding 5 years (n = 243). Abbreviations: AS, aortic stenosis; BVF, bioprosthetic valve failure; DP mean, mean transaortic gradient; EOA, effective orifice area; SVD, structural valve deterioration; TF, transfemoral.

Highlights
-
•
The Kaplan–Meier estimate of all-cause mortality at 8 years was 67.3%.
-
•
The cumulative incidence at 8 years was 2.1% for severe structural valve deterioration and 3.2% for bioprosthetic valve failure.
-
•
Valvular aortic regurgitation was the most common mode of failure.
-
•
Long-term valve performance after a median of 6.4 years was preserved.
-
•
These findings may be helpful in the management of patients that were treated with this device.
Introduction
Transcatheter aortic valve implantation (TAVI) has evolved as the standard treatment option for patients with severe aortic stenosis. Among elderly patients, TAVI is indicated irrespective of the operative risk, whereas in younger patients, TAVI may be considered in higher-risk situations. According to European guidelines, the age threshold for TAVI irrespective of operative risk is 75 years.1 In recent years, outcomes have improved substantially, and there is already a shift toward the treatment of younger and lower-risk cohorts, even though this trend is not supported due to the lack of long-term durability data of transcatheter heart valves (THVs). Current data on long-term THV performance have several limitations, including follow-up (FU) periods rarely exceeding 5 years, incomplete serial echocardiographic valve assessment, relatively small sample sizes, and the competing risk of death among elderly study patients and its appropriate statistical appraisal. The history of surgical bioprostheses suggests that structural valve deterioration (SVD) or bioprosthetic valve failure (BVF) as clinical correlate may increasingly occur beyond a period of 5 years.2 Finally, long-term data are not available for all THV types, and in the absence of a class effect, long-term data of a specific THV type may not necessarily be extrapolated to others.
The ACURATE platform (neo, neo2, and Prime) represents a self-expanding valve with supra-annular design that is characterized by low gradients, low rates of conduction disturbances, low risk of coronary obstruction, and easy coronary access.3,4 The first iteration received CE mark in 2011, but so far only midterm echocardiographic FU data were available.5 Following unfavorable outcomes in the Investigational Device Exemption trial, the ACURATE platform was withdrawn from the market in May 2025, mainly due to unproportional regulatory demands by the European notified body. However, although the device is no longer available for clinical use, reporting long-term clinical and echocardiographic data remains clinically relevant given approximately 100,000 implantations worldwide. In particular, insights into the timing and predominant mode of failure would be valuable for the ongoing management of patients treated with this platform.
This study aimed to analyze long-term outcomes and echocardiographic data among ACURATE neo recipients to determine rates of SVD and BVF.
Methods
Study Cohort
Consecutive patients undergoing transfemoral TAVI for severe native aortic stenosis from 4 centers (Kerckhoff Heart Center, Bad Nauheim, Justus-Liebig University of Giessen, Giessen, Germany; Heart Center Lucerne, Lucerne, Switzerland; Instituto Dante Pazzanese de Cardiologia, São Paulo, Brazil) using the ACURATE neo between 2012 and 2018 (n = 980) were screened for study inclusion.
The characteristics and implantation technique of the ACURATE neo have been described previously.3 All patients were discussed within the Heart Team of each center in adherence to existing guidelines.6
The flowchart in Figure 1 shows the different subsets of patients with specific inclusion and exclusion criteria according to the endpoint. All patients with available vital status information were included in the survival analysis (overall cohort). For the durability analysis, patients with 30-day mortality, with valve-in-valve implantation or surgical aortic valve replacement within 30 days, or without echocardiographic FU beyond 30 days were excluded, resulting in a durability cohort of 758 patients.
Figure 1.
Study flow chart.
Abbreviations: AS, aortic stenosis; FU, follow-up, SAVR, structural valve deterioration; TF, transfemoral; VinV, valve-in-valve implantation.
At Kerckhoff Heart Center, patients who were alive without recent FU echocardiography were contacted and asked for participation in the study with written informed consent. Those who agreed to participate were scheduled either for an outpatient visit in our hospital or, if unable to travel, a home visit to obtain transthoracic echocardiography examinations. Patients that declined study participation were asked for written permission to request medical records of the most recent examination (which included echocardiograms, patient history, and medication) by their cardiologist or family doctor. At the remaining centers, the FU visits and echocardiography were part of the clinical routine and were analyzed retrospectively.
Patients who could not be contacted and of whom the alive status could not be determined were declared as lost-to-FU. Among patients who died, we attempted to obtain data on the cause of death and records on the most recent echocardiography before death by contacting relatives, referring physicians, or medical reports. The census of FU was February 2024.
Baseline data including demographics, comorbidities, risk scores, and echocardiography results were recorded and consolidated in a joint database. Any inconsistencies of the data were resolved by direct communication with the investigator of each participating center. The study was conducted in adherence to the Declaration of Helsinki and was approved by the local ethics committees.
Imaging Workup
Echocardiography assessment included baseline transthoracic echocardiography, post-TAVI (either in-hospital before discharge or at 30-day FU), and the last available FU echocardiography. Long-term valve performance was assessed among patients with available paired FU echocardiography beyond 5 years.
In addition to standard measurements of aortic root dimensions on computed tomography, the cover index (%) was determined as 100 × ([prosthesis size (mm) – perimeter derived annulus diameter (mm)]/prosthesis size [mm]). The aortic valve calcium score was measured according to the Agatston method using noncontrast-enhanced multidetector computed tomography (scans).
Outcomes of Interest
The primary outcome was the cumulative incidence of late (>30 days after the index procedure) BVF, encompassing (1) valve-related death, (2) severe hemodynamic SVD, or (3) aortic valve reintervention. Valve-related death was assumed when the date of death was within 30 days following either aortic valve reintervention or diagnosis of SVD. Coprimary outcomes were valve performance beyond 5-year FU and the cumulative incidence of SVD. Moderate hemodynamic SVD was defined as (1) the mean gradient ≥20 and < 40 mmHg and/or ≥10 and < 20 mmHg change from baseline (before discharge or within 30 days of valve implantation) and/or (2) moderate new or worsening (>1+/4+) intraprosthetic aortic regurgitation (AR). Severe hemodynamic SVD was defined as (1) the mean gradient ≥40 mmHg and/or ≥20 mmHg change from baseline (before discharge or within 30 days of valve implantation) and/or (ii) severe new or worsening (>1+/4+) intraprosthetic AR. Morphological SVD was defined as abnormalities of the following domains: leaflet integrity (i.e., torn or flail causing intraframe regurgitation), leaflet structure (i.e., pathological thickening and/or calcification causing valvular stenosis or central regurgitation), leaflet function (i.e., impaired mobility resulting in stenosis and/or central regurgitation), and strut/frame (i.e., fracture or failure).7 Secondary outcomes were time-to-event rates of all-cause mortality, stroke, and hospitalization (valve-related or worsening heart failure) in the total cohort.
Statistical Analysis
Continuous data are presented as median and interquartile range (IQR). Comparison of groups was performed using the Mann-Whitney U test and the Chi-squared or Fisher exact test, as appropriate. The reverse Kaplan–Meier method was used to calculate the median FU time. The estimates of late BVF and SVD over time were assessed with the Kaplan–Meier method (actuarial analysis) and the cumulative incidence method according to Aalen–Johansen (actual analysis), accounting for the competing risk of death. In this elderly population with high competing mortality, cumulative incidence estimates more appropriately reflect the risk of experiencing valve-related events and avoid overestimation inherent to the Kaplan–Meier method, which treats death as a censoring event. Furthermore, incidence rates of BVF and SVD per 100 patient-years were calculated. Cox regression analyses were performed to determine predictors of all-cause mortality beyond 30 days, SVD, and late BVF; for this purpose, all variables with p values <0.10 in the univariate analysis or that were deemed clinically relevant were entered into the multivariable analysis. Proportional hazards assumption was tested using Schoenfeld residuals, and collinearity was assessed using variance inflation factors. Continuous variables were assessed for linearity using Martingale residuals and visual inspection of LOWESS-smoothed plots (locally weighted scatterplot smoothing). For all analyses, a 2-sided p value <0.05 was considered to indicate statistical significance. All analyses were performed with STATA IC (version 18.0 BE; StataCorp LCC, College Station, Texas).
Results
Study Cohort
The study flowchart is shown in Figure 1. The median time to death or last FU in the overall cohort was 5.0 years (IQR 2.3–6.4) with a maximum FU time of 11.3 years. At the time of the census, there were 409 survivors (42.6%) with a median FU time of 6.2 years (IQR 5.2–7.3). The number of patients with available paired FU echocardiography beyond 30 days was 758 (78.9%), whereas 243 (25.3%) patients had a FU echo beyond 5 years with a median FU time of 6.4 years (IQR 5.9–7.4) and a maximal FU time of 11.0 years. Baseline characteristics are summarized in Table 1. The median age was 82.0 years (IQR 78.8–85.4), 61.3% were females, and the median EuroScore II was 3.3% (IQR .3–5.6). Procedural details and in-hospital outcomes are provided in Table 2.
Table 1.
Baseline characteristics
| Variable | Overall cohort n = 961 | Durability cohort n = 758 | Excluded cohort n = 203 | p |
|---|---|---|---|---|
| Age, y | 82.0 [78.8–85.4] | 81.7 [78.5–84.9] | 83.7 [80.2–87.1] | <0.001 |
| Female sex | 589 (61.3%) | 461 (60.8%) | 128 (63.1%) | 0.561 |
| Body mass index, kg/m2 | 26.9 [24.1–30.4] | 27.2 [24.3–30.5] | 25.8 [23.7–9.0] | 0.003 |
| EuroSCORE II, % | 3.3 [2.3–5.6] | 3.3 [2.2–5.6] | 3.4 [2.5–5.6] | 0.282 |
| eGFR, ml/min/1.73 m2 | 60.0 [43.0–80.0] | 60.4 [44.0–80.0] | 57.0 [38.0–79.0] | 0.129 |
| Coronary artery disease | 549 (57.1%) | 435 (57.4%) | 114 (56.2%) | 0.753 |
| Prior stroke | 107 (11.1%) | 80 (10.6%) | 27 (13.3%) | 0.269 |
| Atrial fibrillation | 347 (36.2%) | 260 (34.4%) | 87 (42.9%) | 0.027 |
| Prior PPI | 100 (10.4%) | 76 (10.0%) | 24 (11.8%) | 0.457 |
| COPD | 182 (18.9%) | 143 (18.9%) | 39 (19.2%) | 0.911 |
| Ejection fraction, % | 65 [55–65] | 65 [55–65] | 64 [55–65] | 0.288 |
| Mean gradient, mmHg | 42 [32–53] | 43 [33–53] | 41 [29–53] | 0.050 |
| AVA, cm2 | 0.7 [0.6–0.8] | 0.7 [0.6–0.8] | 0.7 [0.6–0.8] | 0.306 |
| Perimeter-derived annulus, mm | 23.8 [22.6–25.1] | 23.8 [22.6–25.0] | 23.9 [22.6–25.5] | 0.322 |
| Cover index, % | 4.8 [2.8–7.2] | 4.9 [2.9–7.3] | 4.5 [2.2–7.1] | 0.065 |
| AVCS, AU | 2252 [1570–3056] | 2228 [1559–3037] | 2292 [1628–3193] | 0.354 |
Values denote median [IQR] or n (%).
Abbreviations: AU, arbitrary unit; AVA, aortic valve area; AVCS, aortic valve calcium score; COPD, chronic obstructive pulmonary disease; eGFR, estimated glomerular filtration rate; IQR, interquartile range; PPI, permanent pacemaker implantation.
Table 2.
Procedural outcomes and complications
| Variable | Overall cohort n = 961 | Durability cohort n = 758 | Excluded cohort n = 203 | p |
|---|---|---|---|---|
| Procedural data | ||||
| Prosthesis size | 0.093 | |||
| S, 23 mm | 244 (25.4%) | 104 (24.3%) | 60 (29.6%) | |
| M, 25 mm | 408 (42.5%) | 335 (44.2%) | 73 (36.0%) | |
| L, 27 mm | 309 (32.2%) | 239 (31.5%) | 70 (34.5%) | |
| Procedural duration, min | 41 [32–56] | 40 [32–56] | 42 [34–57] | 0.266 |
| Fluoroscopy time, min | 9.6 [7.4–12.5] | 9.3 [7.2–12.0] | 10.4 [7.6–14.5] | 0.001 |
| Contrast agent, ml | 94 [70–120] | 95 [72–120] | 89 [70–120] | 0.433 |
| Predilatation | 768 (79.9%) | 609 (80.3%) | 159 (78.3%) | 0.524 |
| Postdilatation | 346 (36.0%) | 270 (35.6%) | 76 (37.4%) | 0.632 |
| Ejection fraction post, % | 65.0 [60.0–65.0] | 65.0 [60.0–65.0] | 65.0 [55.0–65.0] | 0.455 |
| Mean gradient post, mmHg | 8 [5–11] | 8 [6–11] | 7 [5–10] | 0.005 |
| EOA post, cm2 | 1.7 [1.5–2.0] | 1.7 [1.5–2.0] | 1.7 [1.4–1.9] | 0.390 |
| Implantation depth NCC, mm | 5.0 [4.0–6.0] | 5.0 [4.0–6.0] | 5.0 [4.0–7.0] | 0.679 |
| Technical success (VARC-3) | 899 (93.5%) | 732 (96.6%) | 167 (82.3%) | <0.001 |
| Device success (VARC-3) | 799 (83.1%) | 674 (88.9%) | 125 (61.6%) | <0.001 |
| Complications | ||||
| PVL post-TAVI | 0.004 | |||
| None/trace | 309 (32.5%) | 237 (31.3%) | 72 (37.3%) | |
| Mild | 601 (63.2%) | 495 (65.3%) | 106 (54.9%) | |
| Moderate | 41 (4.3%) | 26 (3.4%) | 15 (7.8%) | |
| Severe PPM | 24/520 (4.6%) | 21/408 (5.1%) | 3/112 (2.7%) | 0.270 |
| Major cardiac structural complication | 23 (2.4%) | 10 (1.3%) | 13 (6.4%) | <0.001 |
| Major vascular complication | 99 (10.5%) | 68 (9.0%) | 31 (15.3%) | 0.009 |
| Bleeding BARC 2-4 | 159 (16.5%) | 103 (13.6%) | 56 (27.6%) | <0.001 |
| Any stroke | 25 (2.6%) | 11 (1.5%) | 14 (6.9%) | <0.001 |
| Delirium | 43/876 (4.9%) | 20/678 (2.9%) | 23/198 (11.6%) | <0.001 |
| Acute kidney injury St. 2-4 | 71/808 (8.8%) | 40/623 (6.4%) | 31/185 (16.8%) | <0.001 |
| Pacemaker implantation | 85 (8.8%) | 56 (7.4%) | 29 (14.3%) | 0.002 |
Values denote median [IQR] or n (%).
Abbreviations: BARC, Bleeding Academic Research Consortium; EOA, effective orifice area; IQR, interquartile range; NCC, noncoronary cusp; PPM, prosthesis-patient mismatch; PVL, paravalvular leakage; TAVI, transcatheter aortic valve implantation; VARC, Valve Academic Research Consortium.
Long-Term Survival
In the overall cohort, the Kaplan–Meier estimates of all-cause mortality at 1, 5, and 8 years were 13.6% (95% CI 11.6–15.9), 43.1% (95% CI 39.9–46.4), and 67.3% (95% CI 63.4–71.1), respectively (Figure 2). Higher age, atrial fibrillation, renal failure, lower mean aortic valve gradient at baseline and postprocedure, acute kidney injury, and delirium were independent predictors of all-cause mortality beyond 30 days (Table 3).
Figure 2.
Survival estimates in the overall cohort.
Table 3.
Cox regression analysis for all-cause mortality beyond 30 days
| Covariate | Hazard ratio (95% CI) | p | Hazard ratioadj (95% CI) | p |
|---|---|---|---|---|
| Age, per year | 1.05 (1.03–1.07) | <0.001 | 1.03 (1.01–1.05) | <0.001 |
| Female sex | 1.04 (0.87–1.25) | 0.629 | ||
| Coronary artery disease | 1.11 (0.94–1.33) | 0.211 | ||
| Atrial fibrillation | 1.80 (1.51–2.15) | <0.001 | 1.22 (1.00–1.50) | 0.049 |
| eGFR, per ml/min | 0.99 (0.98–0.99) | <0.001 | 0.99 (0.99–0.99) | 0.007 |
| Pmeanpre, per mmHg | 0.98 (0.97–0.99) | <0.001 | 0.99 (0.98–0.99) | 0.041 |
| AVCS, per AU | 0.99 (0.99–1.00) | 0.796 | ||
| EuroSCORE II, per % | 1.03 (1.01–1.05) | 0.001 | 1.01 (0.98–1.03) | 0.517 |
| New LBBB | 0.74 (0.43–1.26) | 0.270 | ||
| PPI | 1.56 (1.19–2.05) | 0.001 | 1.33 (0.99–1.77) | 0.053 |
| Bleeding BARC 2-4 | 1.40 (1.12–1.75) | 0.003 | 1.12 (0.87–1.43) | 0.371 |
| Major vasc complication | 1.27 (0.96–1.68) | 0.087 | ||
| Stroke | 1.37 (0.75–2.49) | 0.304 | 0.94 (0.46–1.90) | 0.858 |
| Delirium | 2.32 (1.60–3.36) | <0.001 | 1.80 (1.20–2.71) | 0.004 |
| AKI St. 2-4 | 1.09 (0.94–1.84) | 0.110 | 1.48 (1.02–2.14) | 0.036 |
| Pmean post | 0.93 (0.91–0.95) | <0.001 | 0.95 (0.93–0.97) | <0.001 |
| Severe PPM | 1.13 (0.68–1.86) | 0.639 | ||
| PVL ≥ moderate | 1.30 (0.89–1.91) | 0.167 | 1.11 (0.74–1.68) | 0.594 |
Significant p-values are highlighted in bold.
Abbreviations: AKI, acute kidney injury; AU, arbitrary unit; AVCS, aortic valve calcium score; BARC, Bleeding Academic Research Consortium; eGFR, estimated glomerular filtration rate; LBBB, left bundle branch block; Pmean, mean aortic gradient; PPI, permanent pacemaker implantation; PPM, prosthesis-patient mismatch; PVL, paravalvular leakage; vasc, vascular.
Valve Performance
The valve performance in the SVD cohort and in the subset of patients with echocardiography FU beyond 5 years is shown in Figure 3. In the latter group, the mean aortic valve gradients and effective orifice area were 45 mmHg (IQR 37–54) and 0.7 cm2 (IQR 0.6–0.8) at baseline, 9 mmHg (IQR 7–13) and 1.8 cm2 (1.5–2.0) post-TAVI, and 6 mmHg (IQR 4–8) and 1.7 cm2 (IQR 1.4–2.0) at the final FU echocardiography. Paravalvular leakage (PVL) ≥ moderate was 2.8% (7/243) post-TAVI and 4.6% (11/242) at long-term FU. Of the 7 cases with PVL ≥ moderate post-TAVI, 4 were classified as PVL ≤ mild at final FU echo. Meanwhile, of the 11 cases with PVL ≥ moderate at final FU echocardiography, 8 had been classified as ≤ mild at post-TAVI. Supplemental Table 1 compares patients with echocardiography FU ≥ 5 years versus those without within the SVD cohort.
Figure 3.
Long-term valve performance. (A to C) Valve performance at last FU in the durability cohort (n = 758) and (D to F) the subset of patients with echocardiography FU beyond 5 years (n = 243).
Abbreviations: DP mean, mean aortic gradient; EOA, effective orifice area; FU, follow-up; TAVI, transcatheter aortic valve implantation.
Structural Valve Deterioration and Bioprosthetic Valve Failure
Estimates of late BVF and SVD are shown in Table 4 and Figure 4. The cumulative incidence of moderate SVD was 3.5% (95% CI 1.5–6.7). The cumulative incidence of severe SVD was 0.5% (95% CI 0.01–0.1) and 2.1% (95% CI 0.6–5.3) at 5 and 8 years, respectively. During 2313.9 patient-years of FU, the incidence rate of SVD was 64.8 (95% CI 39.1–107.5) per 100 patient-years. The cumulative incidence of late BVF was 0.9% (95% CI 0.04–0.2) and 3.2% (95% CI 1.6–5.7) at 5 and 8 years, respectively. During 3639.7 patient-years of FU, the incidence rate of late BVF was 35.7 (95% CI 20.7–61.5) per 100 patient-years.
Table 4.
Durability outcomes
| Outcome (n = 758) | Actual analysis 8-y Cumulative incidence |
Actuarial analysis 8-y Kaplan–Meier | Incidence rate per 100 patient-y |
|---|---|---|---|
| Moderate SVD, % | 3.5 (1.5–6.7) | 7.6 (3.5–15.9) | 43.2 (23.3–80.3)∗ |
| Severe SVD, % | 2.1 (0.6–5.3) | 4.6 (1.5–13.6) | 21.6 (8.9–51.9)∗ |
| BVF, % | 3.2 (5.7–1.6) | 6.5 (3.1–13.3) | 46.7 (20.7–61.5)† |
Values are presented as % (95% CI).
The overall median FU time in the durability cohort was 5.5 y (IQR 3.4–6.7) and 6.2 y (IQR 5.4–7.3) among survivors.
Abbreviations: BVF, bioprosthetic valve failure; FU, follow-up; IQR, interquartile range; SVD, structural valve deterioration.
2313.9 patient-year.
3639.7 patient-year.
Figure 4.
Durability outcomes:cumulative incidence of (A) severe SVD and (B) late BVF.
Abbreviations: BVF, bioprosthetic valve failure; SVD, structural valve deterioration.
Table 5 lists all BVF and SVD cases including the failure mode, time to failure, and vital status. Tables 6 and 7 show Cox regression analyses for any SVD and BVF. Independent predictors of any SVD were left ventricular ejection fraction at baseline (adjusted hazard ratio 1.13 [95% CI 1.02; 1.25]; p = 0.019) and PVL ≥ moderate at discharge (adjusted hazard ratio 7.27 [95% CI 1.42; 37.12]; p = 0.017). Independent predictors of BVF were younger age (adjusted hazard ratio per year 0.91 [95% CI 0.84; 0.98]; p = 0.021) and PVL ≥ moderate at discharge (adjusted hazard ratio 7.98 [95% CI 2.18; 29.22]; p = 0.002).
Table 5.
Overview of cases with BVF or SVD
| Case | BVF | SVD | Failure mode/reason for intervention | Time to BVF or SVD, d | Vital status at last FU | Time to last FU, d |
|---|---|---|---|---|---|---|
| 1 | Redo-TAVI | Severe SVD | Valvular AR | 2752 | Alive | 3129 |
| 2 | Redo-TAVI | Severe SVD | Valvular AR | 1751 | Dead | 3347 |
| 3 | Moderate SVD | Valvular AR | 2672 | Alive | 2921 | |
| 4 | Redo-TAVI | Moderate SVD | Valvular AR | 2787 | Dead | 2830 |
| 5 | Redo-TAVI | Severe SVD | Valvular AR | 1718 | Alive | 2837 |
| 6 | SAVR | Endocarditis | 2864 | Alive | 2864 | |
| 7 | Moderate SVD | High gradient | 2012 | Alive | 2012 | |
| 8 | Moderate SVD | Valvular AR | 2624 | Dead | 2655 | |
| 9 | Severe SVD | Severe SVD | High gradient | 2723 | Alive | 2723 |
| 10 | Severe SVD | Severe SVD | Valvular AR | 223 | Dead | 667 |
| 11 | Moderate SVD | High gradient | 450 | Dead | 2064 | |
| 12 | SAVR Valve-related death |
Endocarditis | 50 | Dead | 51 | |
| 13 | Moderate SVD | High gradient | 426 | Alive | 1539 | |
| 14 | Redo-TAVI | PVL | 114 | Alive | 2447 | |
| 15 | Moderate SVD | Valvular AR High gradient |
2226 | Alive | 2226 | |
| 16 | Redo-TAVI | Hemolysis | 481 | Alive | 2205 | |
| 17 | Redo-TAVI | Moderate SVD | Valvular AR | 1485 | Alive | 2116 |
| 18 | Moderate SVD | High gradient | 1090 | Alive | 1090 | |
| 19 | Redo-TAVI | Moderate SVD | Valvular AR High gradient |
1913 | Dead | 2157 |
| 20 | Valve-related death | Leaflet thrombosis | 1850 | Dead | 1850 |
Abbreviations: AR, aortic regurgitation; BVF, bioprosthetic valve failure; FU, follow-up; PVL, paravalvular leakage; SAVR, surgical aortic valve replacement; SVD, structural valve deterioration; TAVI, transcatheter aortic valve implantation.
Table 6.
Cox regression analysis for any SVD
| Covariate | Hazard ratio (95% CI) | p | Hazard ratioadj (95% CI) | p |
|---|---|---|---|---|
| Age, per year | 0.95 (0.86–1.03) | 0.211 | 0.93 (0.85–1.01) | 0.104 |
| Female sex | 0.67 (0.23–1.98) | 0.476 | ||
| Hypertension | 0.00 (0.00–0.00) | 1.000 | ||
| Diabetes | 0.99 (0.34–2.91) | 0.995 | ||
| Hyperlipidemia | 0.53 (0.16–1.76) | 0.303 | ||
| eGFR, per ml/min | 1.00 (0.99–1.02) | 0.319 | ||
| Pmeanpre, per mmHg | 1.1 (0.98–1.05) | 0.294 | ||
| AVCS, per AU | 0.99 (0.99–1.00) | 0.820 | ||
| Cover index, per % | 0.88 (0.76–1.04) | 0.140 | 0.88 (0.75–1.03) | 0.118 |
| Predilatation | 0.50 (0.16–1.60) | 0.244 | ||
| Postdilatation | 1.43 (0.52–3.97) | 0.489 | ||
| LVEF post, per % | 1.10 (1.00–1.20) | 0.044 | 1.13 (1.02–1.25) | 0.019 |
| Pmean post | 1.03 (0.92–1.14) | 0.618 | ||
| Moderate or severe PPM | 0.70 (0.14–3.51) | 0.672 | ||
| PVL ≥ moderate | 4.25 (0.96–18.91) | 0.057 | 7.27 (1.42–37.12) | 0.017 |
Significant p-values are highlighted in bold.
Abbreviations: AU, arbitrary unit; AVCS, aortic valve calcium score; eGFR, estimated glomerular filtration rate; LVEF, left ventricular ejection fraction; Pmean, mean aortic gradient; PPM, prosthesis-patient mismatch; PVL, paravalvular leakage; SVD, structural valve deterioration.
Table 7.
Cox regression analysis for BVF
| Covariate | Hazard ratio (95% CI) | p | Hazard ratioadj(95% CI) | p |
|---|---|---|---|---|
| Age, per year | 0.92 (0.85–1.00) | 0.058 | 0.91 (0.84–0.98) | 0.021 |
| Female sex | 1.83 (0.68–4.87) | 0.228 | ||
| Hypertension | 0.63 (0.14–2.79) | 0.545 | ||
| Diabetes | 0.73 (0.23–2.27) | 0.592 | ||
| Hyperlipidemia | 1.83 (0.56–6.01) | 0.318 | ||
| eGFR, per ml/min | 0.99 (0.98–1.01) | 0.631 | ||
| Pmeanpre, per mmHg | 0.98 (0.95–1.02) | 0.395 | ||
| AVCS, per AU | 1.00 (0.99–1.00) | 0.836 | ||
| Predilatation | 0.39 (0.12–1.33) | 0.134 | ||
| Postdilatation | 1.53 (0.51–4.58) | 0.443 | ||
| Pmean post | 0.96 (0.85–1.08) | 0.482 | ||
| Moderate or severe PPM | 0.59 (0.12–2.84) | 0.511 | ||
| PVL ≥ moderate | 5.96 (1.69–21.01) | 0.005 | 7.98 (2.18–29.22) | 0.002 |
Significant p-values are highlighted in bold.
Abbreviations: AU, arbitrary unit; AVCS, aortic valve calcium score; BVF, bioprosthetic valve failure; eGFR, estimated glomerular filtration rate; Pmean, mean transaortic gradient; PPM, prosthesis-patient mismatch; PVL, paravalvular leakage.
Discussion
We present the first long-term outcomes including echocardiography after transfemoral TAVI using the ACURATE neo THV from 4 international centers. The main findings of the present study are as follows.
-
(i)
The Kaplan–Meier estimates of all-cause mortality at 5 and 8 years were 43.1 and 67.3%, respectively.
-
(ii)
At 8 years, the cumulative incidence was 3.5% for moderate SVD, 2.1% for severe SVD, and 3.2% for BVF.
-
(iii)
Long-term valve performance after a median of 6.4 years was preserved.
Long-Term Mortality
The all-cause mortality we observed in the present study is consistent to previous reports on long-term survival in intermediate-to high-risk TAVI cohorts. In the Neo2 CE-mark study, 5-year all-cause mortality was 44%,8 whereas in a meta-analysis by Shimamura et al.9 involving various early-generation prosthesis types, 5-year all-cause mortality was 53%. Estimates of 8-year all-cause mortality in intermediate-to high-risk populations vary between 70 and 80%.10, 11, 12, 13 The fact that the 8-year all-cause mortality is in a lower range is notable, given that the Achilles’ heel of the ACURATE neo was PVL ≥ moderate, which has been reported at approximately 10% in the randomized SCOPE I and II trials and in some observational studies.14,15 In the SCOPE I trial, the failure of the ACURATE neo to meet noninferiority versus the balloon-expandable SAPIEN 3 prosthesis was mainly driven by higher rates of acute kidney injury and PVL ≥ moderate (9 vs. 3%).14 Nevertheless, at 3-year FU of the SCOPE I study, all-cause mortality was similar between the 2 groups, even showing a crossing of the survival curves.16 Likewise, in the present ACURATE neo cohort, long-term mortality was not majorly affected by PVL ≥ moderate, which is corroborated by our finding that PVL ≥ moderate was not associated with long-term all-cause mortality in the Cox regression analysis (Table 4). Indeed, similar observations have been made in previous investigations with other THVs.10,11,17
The finding that lower post-procedural mean transvalvular gradients were independently associated with all-cause mortality over time seems to be counterintuitive but may be explained by low-flow aortic stenosis indicating left ventricular damage.
Long-Term Valve Performance
In the subset of patients with long-term echocardiography FU, hemodynamic performance was preserved over a median FU time of 6.4 years (Figure 3). Interestingly, there was a slight increase in PVL ≥ moderate from 2.9 to 4.6% but a decrease in mild PVL from 66.5 to 37.2%. Regarding valvular AR, the rate of at least mild regurgitation increased over time from 2.5% to almost 10%. However, most cases were mild and not clinically relevant; only 3% had moderate or severe valvular AR.
Durability Data
As this is the first report of THV durability outcomes for the ACURATE platform, no previous comparative data are available. The cumulative incidence of moderate SVD (3.5%) and severe SVD (2.1%) and that of BVF (3.2%) of the ACURATE neo THV at 8 years is in the range of durability data of other THV types beyond 5 years, although most existing studies are limited to retrospective analyses predominantly from single centers with relatively small sample sizes and incomplete echocardiography assessments. Another major limitation that may preclude comparability is that, in some studies, the denominators of SVD and BVF analyses may not have been well defined. As the definition of both SVD and BVF requires echocardiography, studies with incomplete echocardiography FU may have overestimated the denominator by inclusion of patients without recent echocardiography data.10, 11, 12, 13 The only durability report from a randomized trial originates from the NOTION study, which compared TAVI using the self-expanding CoreValve THV versus surgical aortic valve replacement, although the TAVI arm included only 145 patients. At 8 years, the cumulative incidence of SVD in the TAVI arm was 13.9%, and the cumulative incidence of BVF was 8.7%.17 In a meta-analysis by Shimamura et al.9, freedom from SVD was 95.5% at 5 years and 15.1% at 8 years. Several other retrospective registries analyzing durability data beyond 5-year FU report inconsistent BVF rates between 4.6 and 18.0% using actuarial analysis10,12,13 and between 2.5 and 8.7% using actual analysis,10,12,13,17 whereas rates of SVD vary according to the definition and applied method (moderate SVD: 3.6 to 14.9%; severe SVD: 2.4 to 5.9%).12,13,17
The finding that PVL ≥ moderate was an independent predictor of both BVF and SVD (Tables 6 and 7) is difficult to explain, but it likely reflects an indirect pathophysiologically plausible association, as PVL may adversely affect hemodynamic flow and increase turbulence and stress on the leaflets and thereby accelerate valve degeneration. Regarding BVF, redo-TAVI for relevant PVL may additionally contribute to this observed association.
Failure Mode
The primary failure mode of the ACURATE neo THV seems to be valvular AR. In our series, two-thirds of SVD cases exhibited valvular AR (Table 5), and valve reinterventions were commonly performed for valvular AR. Even though a direct comparison is not appropriate, previous data with other THV types suggest a lower proportion of valvular AR as failure mode. In the 8-year FU of the NOTION trial, the rate of SVD after TAVI with the CoreValve prosthesis was 13.9%, of which 3.7% had valvular AR.17 Of 10 cases of SVD in another CoreValve long-term FU series, 4 had relevant valvular AR and another 4 had both valvular AR and stenosis as mode of failure.13 Long-term FU data from a cohort with both self-expanding and balloon-expandable THV showed that among 7 cases with SVD beyond 30 days, 4 had stenosis, 2 had relevant valvular AR, and 1 had mixed mode of failure.12
Limitations
The present study represents a retrospective multicenter cohort analysis with inherent limitations. FU strategies, including the timing and availability of echocardiographic examinations, differed between participating centers and were largely driven by routine clinical practice. Consequently, serial echocardiography at regular annual intervals was available in only a minority of patients, precluding a detailed assessment of gradual temporal changes in valve performance. Long-term echocardiographic FU beyond 5 years was available only in a subset of surviving patients, introducing potential survivor bias and limiting the generalizability of long-term valve performance findings.
Echocardiographic assessments were performed locally without adjudication by an independent core laboratory. Although core laboratory adjudication improves standardization and reduces interobserver variability, it does not eliminate measurement uncertainty. Given that durability endpoints were primarily driven by predefined hemodynamic criteria or clinical valve failure, any potential misclassification is unlikely to have materially affected the main conclusions. Regression analyses must be considered exploratory and hypothesis-generating. Finally, given the advanced age and risk profile of the study population, the findings should not be extrapolated to younger or lower-risk patients or to contemporary THV platforms.
Conclusions
In this multicenter observational cohort of patients treated with the ACURATE neo platform with extended FU, survival rate was 32.7%, and the cumulative incidence for severe SVD (2.1%) and BVF (3.2%) through 8 years was low. Within the limitations of the present analysis, these findings provide long-term durability estimates in an elderly, real-world population and should be interpreted in this clinical context.
Funding
This work received financial support from Boston Scientific.
Disclosure Statement
Won-Keun Kim reports receipt of proctor/speaker honoraria from Abbott, Anteris, Boston Scientific, Cardiawave, Edwards Lifesciences, HID Imaging, Jenavalve, Medtronic, Meril Life Sciences, and P & F. Helge Möllmann reports receipt of honoraria or consultation fees from Abbott, Biotronik, Boston Scientific, Edwards Lifesciences, Medtronic, and SMT. Matthias Renker reports receipt of proctor honoraria from Boston Scientific and Medtronic. Efstratios Charitos reports receipt of proctor honoraria from Boston Scientific. Yeong-Hoon Choi reports receipt of speaker/proctor fees from Edwards Lifesciences, Cytosorbents, and Getinge. Stefan Toggweiler reports receipt of honoraria from Medtronic, Boston Scientific, Biosensors, Edwards Lifesciences, Hi-D Imaging, Abbott Vascular, Medira, Shockwave, Teleflex, atHeart Medical, Cardiac Dimensions, Polares Medical, Amarin, Sanofi, AstraZeneca, ReCor Medical, Daiichi Sankyo, Bayer, and Armira; reports receipt of institutional research grants from Edwards Lifesciences, Abbott Vascular, Boston Scientific, Fumedica, Novartis, Boehringer Ingelheim, and Polares Medical; is the founder and owner of CardiaNova GmbH; holds equity in Hi-D Imaging; and has filed a patent application for a balloon catheter (10 2015 114 609.8).
The other authors had no conflicts to declare.
Footnotes
Supplemental data for this article can be accessed on the https://doi.org/10.1016/j.shj.2026.100841.
Supplementary Material
References
- 1.Vahanian A., Beyersdorf F., Praz F., et al. 2021 ESC/EACTS guidelines for the management of valvular heart disease. Eur Heart J. 2022;43(7):561–632. doi: 10.1093/eurheartj/ehab395. [DOI] [PubMed] [Google Scholar]
- 2.Johnston D.R., Soltesz E.G., Vakil N., et al. Long-term durability of bioprosthetic aortic valves: implications from 12,569 implants. Ann Thorac Surg. 2015;99(4):1239–1247. doi: 10.1016/j.athoracsur.2014.10.070. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Kim W.K., Hengstenberg C., Hilker M., et al. Transcatheter aortic valve implantation with the ACURATE neo valve: indications, procedural aspects and clinical outcomes. EuroIntervention. 2020;15(18):e1571–e1579. doi: 10.4244/EIJ-D-19-00908. [DOI] [PubMed] [Google Scholar]
- 4.Kim W.K., Mollmann H., Montorfano M., et al. Outcomes and performance of the ACURATE neo2 transcatheter heart valve in clinical practice: one-yearresults of the ACURATE neo2 PMCF study. EuroIntervention. 2024;20(1):85–94. doi: 10.4244/EIJ-D-23-00823. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Ruck A., Shahim B., Manouras A., et al. Midterm durability of the ACURATE transcatheter aortic valve implantation system based on VARC-3 definitions. EuroIntervention. 2024;20(12):e781–e782. doi: 10.4244/EIJ-D-23-01018. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Vahanian A., Alfieri O., Andreotti F., et al. Guidelines on the management of valvular heart disease (version 2012) Eur Heart J. 2012;33(19):2451–2496. doi: 10.1093/eurheartj/ehs109. [DOI] [PubMed] [Google Scholar]
- 7.Capodanno D., Petronio A.S., Prendergast B., et al. Standardized definitions of structural deterioration and valve failure in assessing long-term durability of transcatheter and surgical aortic bioprosthetic valves: a consensus statement from the European association of percutaneous cardiovascular interventions (EAPCI) endorsed by the European society of cardiology (ESC) and the European association for cardio-thoracic surgery (EACTS) Eur Heart J. 2017 doi: 10.1093/eurheartj/ehx303. [DOI] [PubMed] [Google Scholar]
- 8.Mollmann H., Noack T., Hilker M., et al. Five-year outcomes of transcatheter aortic valve implantation with the ACURATE neo2 valve system. EuroIntervention. 2024;20(15):e951–e953. doi: 10.4244/EIJ-D-23-00852. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Shimamura J., Takemoto S., Fukuhara S., et al. Long-term outcomes after transcatheter aortic valve replacement: meta-analysis of Kaplan-Meier-derived data. Catheter Cardiovasc Interv. 2023;102(7):1291–1300. doi: 10.1002/ccd.30854. [DOI] [PubMed] [Google Scholar]
- 10.Holy E.W., Kebernik J., Abdelghani M., et al. Long-term durability and haemodynamic performance of a self-expanding transcatheter heart valve beyond five years after implantation: a prospective observational study applying the standardised definitions of structural deterioration and valve failure. EuroIntervention. 2018;14(4):e390–e396. doi: 10.4244/EIJ-D-18-00041. [DOI] [PubMed] [Google Scholar]
- 11.Deutsch M.A., Erlebach M., Burri M., et al. Beyond the five-year horizon: long-term outcome of high-risk and inoperable patients undergoing TAVR with first-generation devices. EuroIntervention. 2018;14(1):41–49. doi: 10.4244/EIJ-D-17-00603. [DOI] [PubMed] [Google Scholar]
- 12.Barbanti M., Costa G., Zappulla P., et al. Incidence of long-term structural valve dysfunction and bioprosthetic valve failure after transcatheter aortic valve replacement. J Am Heart Assoc. 2018;7(15) doi: 10.1161/JAHA.117.008440. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Panico R.A., Giannini C., De Carlo M., et al. Long-term results and durability of the CoreValve transcatheter aortic bioprosthesis: outcomes beyond five years. EuroIntervention. 2019;14(16):1639–1647. doi: 10.4244/EIJ-D-18-00779. [DOI] [PubMed] [Google Scholar]
- 14.Lanz J., Kim W.K., Walther T., et al. Safety and efficacy of a self-expanding versus a balloon-expandable bioprosthesis for transcatheter aortic valve replacement in patients with symptomatic severe aortic stenosis: a randomised non-inferiority trial. Lancet. 2019;394(10209):1619–1628. doi: 10.1016/S0140-6736(19)32220-2. [DOI] [PubMed] [Google Scholar]
- 15.Tamburino C., Bleiziffer S., Thiele H., et al. Comparison of self-expanding bioprostheses for transcatheter aortic valve replacement in patients with symptomatic severe aortic stenosis: SCOPE 2 randomized clinical trial. Circulation. 2020;142(25):2431–2442. doi: 10.1161/CIRCULATIONAHA.120.051547. [DOI] [PubMed] [Google Scholar]
- 16.Lanz J., Mollmann H., Kim W.K., et al. Final 3-Year outcomes of a randomized trial comparing a self-expanding to a balloon-expandable transcatheter aortic valve. Circ Cardiovasc Interv. 2023;16(7) doi: 10.1161/CIRCINTERVENTIONS.123.012873. [DOI] [PubMed] [Google Scholar]
- 17.Jorgensen T.H., Thyregod H.G.H., Ihlemann N., et al. Eight-year outcomes for patients with aortic valve stenosis at low surgical risk randomized to transcatheter vs. surgical aortic valve replacement. Eur Heart J. 2021;42(30):2912–2919. doi: 10.1093/eurheartj/ehab375. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.