Anionic surfactant reformed carbon composite voltammetric sensor for the efficient and sensitive detection of carmoisine in food samples

Battira M Sharmila 1, Jamballi G Manjunatha 1,✉, Karnayana P Moulya 1, Thiago C Canevari 2, Gaber E Eldesoky 3
PMCID: PMC13328487  PMID: 42321345

Abstract

Carmoisine (CM), an azo dye, imparts vibrant red color when added to the food, beverage, cosmetic and pharmaceutical products. The excessive intake of CM causes various health effects and hence it is essential to develop appropriate methods for the determination and quantification of CM in food samples. In this current work, a rapid detection and analysis of CM was performed via cyclic voltammetry (CV) and differential pulse voltammetry (DPV) techniques by harnessing sodium dodecyl sulfate modified composite sensor (SDSMCS). The morphological features and elemental composition of bare carbon nanotube graphite composite sensor (BCNGCS) and SDSMCS surfaces were compared and the difference between the working sensors was assessed. The active surface area was computed as 0.031 cm2 for BCNGCS and 0.049 cm2 for SDSMCS surfaces by employing CV technique. Key insights into the charge transfer process involved in the oxidation of CM was obtained by electrochemical impedance spectroscopy (EIS). The modified sensor demonstrated improved electrochemical features at optimal experimental conditions. The analysis validated that the electro oxidation of CM proceeds through adsorption-controlled kinetics at the surface of SDSMCS and exhibited a significantly low limit of detection (LOD) of 0.009 µM by CV and 0.073 µM by DPV technique, in the linear range of 0.1 µM to 1.1 µM and 0.1 µM to 0.9 µM, respectively. The attributes like reproducibility and repeatability of the developed SDSMCS were evaluated and the respective relative standard deviation (RSD) of 4.95% and 4.90% verified that the fabricated sensor is reliable for detecting CM in various commercial food samples with a recovery rate exceeding 97%. Surfactant- reformed carbon-based sensors provide a cost-effective and reliable platform that aligns with the evolving trends in field of electrochemical sensing.

Keywords: Anionic surfactant, Azo dye, Carbon nanotubes, Carmoisine, Voltammetric sensor

Subject terms: Chemistry, Environmental sciences, Materials science, Nanoscience and technology

Introduction

Dyes or colorants refer to the compounds that give color to materials like polymers, textiles, food stuff etc. Natural dyes which are extracted from natural sources are eco-friendly, affordable, non- toxic whereas synthetic colorants pose threat to human health and environment along with certain advantages like fastness, being available in various color range and user friendliness1. However, the significant stability of the latter towards light, pH and oxygen has made them an unfailing constituent in the food, textile, paper, pharmacological, agrochemical and cosmetic industries2. Chemically, CM is a disodium salt with two naphthalene subunits as shown in Fig. 1. It is also known as azorubine and acid red 14 with Europe number (E number) of E122 and available in red or maroon powdered form3,4. Obtained by the azo coupling reaction, CM finds its wide application in food, drug and cosmetic industries5,6. Studies have reported that these colorants impose ill effects on human health due to their potential toxic nature7–9. The daily intake of CM has been limited to 4 mg per kilogram of a person’s body weight, as per Food and Agriculture Organization (FAO) and World Health Organization (WHO) regulations10. Over consumption of CM causes variations in renal and hepatic parameters, thus inducing the formation of free radicals causing oxidative stress11. Recent reports suggest that CM, when ingested even in low doses, can change the biochemical markers and cause adverse effects on vital organs. The effect becomes more pronounced at high doses and also on long term exposure to low doses12. Several methods like spectrophotometry, chromatography, flow injection analysis and chemiluminescence have been developed so far to detect the presence of CM in food items13–17. However, the dominant demerits like expensive instrumentation and complex procedures restrict the use of these sophisticated techniques. Electrochemical analysis has emerged as a promising, low- cost and hassle -free approach and it involves redox reactions occurring at the interface of the sensor and the electrolyte18. Voltammetry is the widely used technique for the determination of electrochemical features of electrode systems modified in various ways19. This technique is highly sensitive, precise and depends on the voltage current for the detection and quantification of electroactive species20,21.

Fig. 1.

Fig. 1

Structure of CM.

Surfactants are amphiphilic compounds with a hydrophobic head and a hydrophilic tail22,23. These readily get adsorbed on the surface of the sensor and augment its detection efficiency at the solution and the sensor interface24,25. SDS is an anionic surface-active agent which when adsorbed on the sensor surface, enhances the electrochemical performance of the sensor by significantly interacting with the electro active species present in the solution. The micelle forming feature of SDS molecules form stable aggregates on the sensor surface. The electrostatic force of attraction between these micelles and the electro active species increases the concentration of the analyte in the vicinity of the sensor surface which boosts the electron transfer rate, thus intensifying the selective and sensitive detection of the analytes26,27. The sensors fabricated by electrode materials like graphite powder (GP) and carbon nanotubes (CNT) display a high surface area with dynamic range of porosity, thus increasing the accessibility of the electrolytes to the internal surface area28. GP possesses numerous positive features such as great conductivity, affordability, chemical inertness and hence serves as a versatile electrode material by offering low background current and wide potential windows in voltammetric analysis. The CNT exhibit exceptional qualities such as high chemical stability, excellent conductivity and remarkable catalytic properties as the tube matrices encourage the electron transfer process to a significant extent29–31. In this work, CV and DPV techniques have been employed due to their rapid response, simplicity and the ability to convey adequate information about the kinetics of the electrochemical processes32,33. This novel approach to detect CM in food samples using a sensor comprising of composite electrode materials reformed with surfactant has provided excellent outcome proving the reliability of this fabricated sensor in availing the relevant applications.

Experimental methods

Instrumentation

The analysis was performed with CHI-6038E (electrochemical analyzer from CH Instruments, USA) consisting of working electrodes (BCNGCS and SDSMCS), a counter electrode (platinum wire) and a reference electrode (calomel electrode). The morphology and elemental composition of the working electrodes was assessed by scanning electron microscopy (SEM) and energy-dispersive X-ray spectroscopy (EDXS) at Vigyan Bhavan, University of Mysore, India. The required buffer solutions were prepared using a digital pH meter (EQ-610) procured from Equiptronics, India. Double distilled water (DW) for the preparation of the various solutions was obtained from VITSIL-VBSD/ VBDD water distillation unit.

Reagents and chemicals

CM (C20H12N2Na2O7S2, 98% pure) and SDS (C12H25NaSO4, 97% pure) were procured from Otto Chemie Pvt. Ltd, Mumbai and Isochem laboratories, Kerala, respectively. CNT (diameter of 110–170 nm, length of 5–9 micron and 90% assay) were obtained from Sigma - Aldrich chemicals. 90% pure silicone oil was procured from Molychem Chemical Manufacturers. 90% pure GP, 99.5% pure potassium chloride (KCl), 98.5% pure potassium ferrocyanide K4[Fe(CN)6], 99% pure monosodium dihydrogen phosphate (NaH2PO4) and disodium hydrogen phosphate (Na2HPO4) were obtained from Nice Chemicals located in Kerala. Chemicals and reagents used in the experimental procedures were of analytical quality and were utilized without any additional purification.

Preparation of the working sensors

A quantified amount of GP and CNT were blended with silicone oil (acts as a binder) in the ratio of 70:30 to attain a uniform paste. The 3 mm cavity of the teflon tube was methodically filled with the blended paste. To obtain an even surface, the filled end of the teflon tube was gently wiped on a tissue paper. Electrical connection was established by implanting a copper wire through the teflon tube. Throughout the analysis, this fabricated electrode is used as BCNGCS. To achieve the modified sensor for the electro analysis, BCNGCS surface was treated with SDS by drop coating technique and allowed to stand for 5 min for the adequate adsorption of SDS on the sensor surface. The reformed SDSMCS surface was then rinsed with DW to eliminate the excess SDS if any.

Real sample preparation

Food samples containing CM such as candies and food color were purchased from local grocer to utilize in the analysis. The food samples were mixed thoroughly in distilled water and allowed to settle for 30 min for the complete dissolution of the dye. The supernatant solution was filtered and mixed with appropriate amount of phosphate buffer solution (PBS) for the analysis.

Results and discussions

Morphological interpretation of working sensors

The morphological features of the sensor surfaces were assessed by SEM as an effort to correlate the voltammetric performance of the sensor to the surface modification. The images depicted in Fig. 2(a) and Fig. 2(b) exhibits disparity between the two functioning sensors. BCNGCS displays the rough surface which serves as a platform for the adequate deposition of the surfactant during modification. SDSMCS reveals the uniform accumulation of SDS on the surface and the accretion of the surfactant aids in elevating the electrochemical performance of the sensor resulting in the sensitive detection of CM.

Fig. 2.

Fig. 2

(a) SEM image of BCNGCS, (b) SEM image of SDSMCS.

Furthermore, the elemental composition of BCNGCS and SDSMCS were assessed by EDXS. As illustrated in Fig. 3(a) the presence of 72.5% of carbon, 14.4% of oxygen, 12.2% of silicon in BCNGCS is confirmed. Similarly, Fig. 3(b) discloses the existence of 70.5% of carbon, 16% of oxygen, 10.6% of silicon, 1.1% of sodium and 1.2% of sulphur in SDSMCS. The disparity in the elemental composition of BCNGCS and SDSMCS confirms the deposition of SDS on the modified sensor surface.

Fig. 3.

Fig. 3

EDXS images (a) BCNGCS and (b) SDSMCS depicting the elemental composition.

EIS study and surface area calculation

CV analysis for K4[Fe(CN)6] (1.0 mM) in 0.1 M KCl (supporting electrolyte) was performed at the surface of various electrode materials and the anodic peak current responses obtained is displayed in Fig. 4(a) (curve ‘a’, ‘b’, ‘c’ and ‘d’ for GP, CNT, BCNGCS and SDSMCS, respectively).

Fig. 4.

Fig. 4

(a) CV response for GP (curve a), CNT (curve b), BCNGCS (curve c) and SDSMCS (curve d) (b) Corresponding Nquist plot with equivalent circuit.

It is apparent from the analysis that the composite electrode material and surfactant modification has enhanced the peak current significantly due to the advancement in the electro active surface area and hence the surface-active sites. With a purpose to validate the same, the electro active surface area of GP, CNT, BCNGCS and SDSMCS were calculated to be 0.015 cm2, 0.022 cm2, 0.031 cm2 and 0.049 cm2, respectively by substituting the obtained anodic peak current values in Randles - Sevcik equation,

graphic file with name d33e481.gif 1

Here, Ip.a., A, n, C0, ν and D designates anodic peak current, electro active surface area, number of electrons (here n = 1), concentration of K4[Fe(CN)6, scan rate and diffusion co-efficient (7.6 × 10− 6 cm2s− 1), respectively. The variation in active surface area of the electrode materials validates that the SDS deposition has caused raise in the active surface area of SDSMCS which has apparently increased the active sites causing an enhanced electrochemical activity.

EIS provides mechanistic and kinetic data of the electrochemical processes which is essential to interpret the efficiency of the fabricated sensor34. As a redox probe, 1.0 mM K4[Fe(CN)6] in 0.1 M KCl was subjected to EIS at the functioning sensor surfaces and the achieved outcomes are represented in Fig. 4(b). The EIS data were fitted to a Randles equivalent circuit in which Rs, R1, C1, Rct, W and CPE signify the solution resistance, resistance, capacitance, charge transfer resistance, Warburg impedance and constant phase element. The Nquist plot displays semi-circles at higher frequencies and linear portions (due to the presence of W) at the lower frequencies. The semi-circles are of dissimilar radii for GP (curve a), CNT (curve b), BCNGCS (curve c) and SDSMCS (curve d) and the Warburg element confirms diffusion of the redox probe. The Rct values observed for GP, CNT, BCNGCS and SDSMCS were 307 Ω, 221 Ω, 222 Ω and 187 Ω, respectively. The variance in the Rct values of different electrode materials reflects the difference in their electron transfer kinetics. The smaller semi-circle with lower Rct value (187 Ω) obtained for SDSMCS validates the influence of the modifier (SDS) in the enhancement of the electron transfer kinetics due to amplified electro- active surface area resulting in increased conductivity.

Optimization of the sensor modifier

The electrochemical performance of the SDSMCS altered when different surface modifiers were used in the analysis. The optimization of the modifier in order to achieve maximum voltammetric response was performed using CV technique. The current responses of 1.0 mM CM at the three different BCNGCS surfaces modified with surfactants (using drop coating technique) such as cetyltrimethylammonium bromide (CTAB), Triton X-100 (TX-100) and SDS were recorded and compared as depicted in Fig. 5(a). Among these, SDS modified sensor responded with a high peak current response of 96.13 µA for CM in PBS (0.2 M) at a potential window of 0.3–0.9 V. The maximum current response can be accredited to the favorable interface offered by SDS for the CM oxidation at the sensor surface. In addition, the appropriate amount of surfactant for drop coating method in order to attain maximum electrochemical activity was evaluated. CV was recorded for 1.0 mM CM with BCNGCS modified with SDS in four different volumes of 5 µL, 10 µL, 15 µL and 20 µL. The observed current responses are displayed in Fig. 5(b) which discloses that 10 µL of SDS drop coated on BCNGCS yields a stable peak with significantly high peak current value of 88.22 µA. The oxidation peak current augmented when the amount of SDS was increased from 5 µL to 10 µL and further increase in the volume (15 µL and 20 µL) resulted in the diminishing of the peak current responses. This variation is due to the immobilization of the sensor surface in the presence of excess SDS35. Hence, for further investigation of CM, 10 µL of SDS was considered for the sensor modification.

Fig. 5.

Fig. 5

CV representation of (a) 1.0 mM CM at BCNGCS surfaces modified with CTAB, TX-100 and SDS, (b) 1.0 mM CM at BCNGCS surfaces modified with different volume of SDS.

Optimization of accumulation time and potential

The time required for the analyte to get adsorbed on the surface of the sensor greatly influences the sensitivity of the peak current. As represented in Fig. 6(a), CV was recorded for 1.0 mM CM at SDSMCS surface at different time intervals (0 s to 120 s) at 0.1 V/s scan rate in PBS (0.2 M). It is apparent that the 20 s exhibits an oxidation peak with higher peak current value, which signifies that the adequate accumulation of CM on the sensor surface occurs at 20 s. Hence, during the analysis the accumulation time of 20 s was maintained to achieve efficient analytical results. Furthermore, the potential window was optimized by recording CV in various potential gaps as represented in Fig. 6(b). Evidently, the peak current value is high at potential window of 0.3–0.9 V in comparison with the other potential ranges, owing to the elevated electrochemical activity of CM at SDSMCS surface. Hence, the potential gap of 0.3 –0.9 V was considered for the voltammetric measurements in the analysis.

Fig. 6.

Fig. 6

(a) CV depiction of 1.0 mM CM at SDSMCS surface at different time intervals, (b) anodic peak current responses 1.0 mM CM at SDSMCS surface at different potential window.

Electrochemical behavior of CM at the functional sensor surfaces

The electrochemical activity of CM at BCNGCS and SDSMCS surfaces was inspected via CV technique. As demonstrated in Fig. 7, the voltammetric measurements were made for 1.0 mM CM at 0.1 V/s scan rate in PBS (0.2 M) at BCNGCS and SDSMCS surfaces (curve ‘a’ and curve ‘c’, respectively). The curve ‘b’ corresponds to the CV response of 0.2 M PBS in the absence of CM. The anodic current response at BCNGCS is feeble (10.63 µA at an anodic peak potential of 0.7917 V which is in converse with the significant current response of CM (96.13 µA at a peak potential of 0.5762 V) at SDSMCS surface. An obvious variation in the oxidation current responses of BCNGCS and SDSMCS is observed and the enhancement in the electrochemical activity can be attributed to the surfactant modification of the sensor.

Fig. 7.

Fig. 7

CV representation of 1.0 mM CM in 0.2 M PBS at BCNGCS (curve a), SDSMCS (curve c) surfaces, only PBS at SDSMCS surface (curve b).

Study of the impact of electrolytic pH

The pH of the supporting electrolyte greatly influences the electrochemical activity of CM at SDSMCS surface and it is essential to optimize the pH for further analyses. CV was documented for 1.0 mM CM in 0.2 M PBS of varied pH (5.5 to 8.0) and as illustrated in Fig. 8(a), it was observed that the oxidation peak current fluctuates with varying pH. It is evident from the Fig. 8 (b) (anodic peak current, Ip.a. versus pH plot) that at pH of 6.5, SDSMCS exhibits high peak current value of 96.38 µA and at other pH values, the current response subsides. The optimum pH value in which the maximum electrochemical activity is observed is 6.5 and for further subsequent analyses, the PBS of pH 6.5 was considered. The negative linear shift observed in the plot of Ep.a. (anodic peak potential) versus pH has been represented in Fig. 8(c). The linearity can be expressed in the linear regression equation (LRE),

graphic file with name d33e624.gif 2

Fig. 8.

Fig. 8

(a) CV of CM (1.0 mM) in PBS (0.2 M) of varied pH, (b) Ip.a. vs. pH plot and (c) Ep.a. vs. pH plot.

The slope of 0.069 (approximately nearer to the theoretical value 0.059) in Eq. (2) conveys that electrochemical oxidation of CM at SDSMCS, involves same number of electrons and protons.

Effect of scan rate

CV was employed to assess the effect of scan rate on oxidation peak current and oxidation peak potential during the CM detection. The analysis was performed for 1.0 mM CM in 0.2 M PBS at varied scan rate ranging from 0.125 to 0.375 V/s and the outcome of the CV analysis has been illustrated in Fig. 9(a). As the scan rate is augmented from 0.125 to 0.375 V/s, oxidation peak current shift towards positive potential as depicted in the Fig. 9(b), indicating that the oxidation process is kinetically controlled. The absence of cathodic peak signifies that the process is irreversible36. A fair linearity was observed between the plot of Ip.a. vs. scan rate which can be expressed in LRE as:

graphic file with name d33e673.gif 3

Fig. 9.

Fig. 9

(a) CV response of CM in 0.2 M PBS (pH 6.5) at varied scan rate, (b) plot of Ip.a. vs. scan rate (ν), (c) graph of log Ip.a. versus log ν and (d) plot of Ep.a. versus log ν.

Further, the linearity obtained from the plot of log of Ip.a. vs. log of scan rate (ν) as represented in Fig. 9(c), can be expressed in LRE as below:

graphic file with name d33e684.gif 4

The linearity observed in Eq. (3) with regression coefficient R2 = 0.9952 and a slope of 0.96 (≈ 1) obtained in Eq. (4) validates that CM oxidation progresses through adsorption-controlled kinetics at SDSMCS surface. The number of electrons involved in the electro oxidation process was evaluated by obtaining necessary data from plot of Ep.a. vs. log of scan rate as portrayed in Fig. 9(d). The LRE can be represented as –

graphic file with name d33e704.gif 5

Substituting the slope from Eq. (5) in Laviron’s equation, the number of electrons was enumerated as 1.18 (≈ 1).

graphic file with name d33e713.gif 6

Here, ‘R’, ‘T’, ‘α’, ‘n’, ‘F’, ‘k0’ signifies universal gas constant, absolute temperature, charge transfer co-efficient, number of electrons, Faraday’s constant and heterogeneous rate constant, respectively.

The feasible mechanism of adsorption-controlled electro-oxidation of CM at SDSMCS surface occurring with the transference of one electron has been illustrated in Fig. 10. At the surface of SDSMCS, the naphthol moiety of CM undergoes single electron oxidation to an unstable phenoxy radical intermediate. Furthermore, this intermediate undergoes rearrangement (C-N bond) and oxidation (azo group) leading to the oxidative destruction of the molecule. This mechanism accounts for the occurrence of single irreversible oxidation peak in the CV analysis37–39.

Fig. 10.

Fig. 10

CM oxidation on SDSMCS surface.

Impact of CM concentration on oxidation peak current

With a purpose to assess the sensitivity of the fabricated SDSMCS in detection and quantification of CM, techniques such as CV and DPV were performed for various concentrations of CM in 0.2 M PBS. From Fig. 11(a) and 11(c), the linear upsurge of the anodic peak current with the varied CM concentration (0.1 µM to 1.1 µM in CV and 0.1 µM to 0.9 µM in DPV) can be observed. The slope values obtained from the calibration plots Fig. 11(b) and 11(d) were applied in the expressions, LOQ = 10 S/N and LOD=3 S/N to get respective LOQ and LOD values. Here, ‘S’ and ‘N’ denote the standard deviation of blank and slope of calibration plots. Accordingly, the values obtained were LOD = 0.009 µM and LOQ = 0.031 µM by CV technique and LOD = 0.073 µM and LOQ = 0.244 µM by DPV technique. When these values were compared with the values reported in earlier works (as illustrated in Table 1), the present work provided a considerably low LOD and LOQ, substantiating the superior sensitive attribute of the fabricated SDSMCS.

Fig. 11.

Fig. 11

CV and DPV depiction of oxidation peak current enhancement with varying concentration (Fig. 11(a) and 11(c), respectively), Calibration plots representing the linear relationship of oxidation current and diverse CM concentration (Fig. 11(b) and 11(d), respectively for CV and DPV).

Table 1.

LOD values of related works.

Technique Electrode Linear range LOD Reference
DPV GO-Fe3O4-G4 PAMAM/ILCPE 0.1–170 0.02 µM 40
SQV CPE/1-M-3BIBr/NiO/CNTs 70–650 0.02 µM 41
Spectrophotometry - 0.02–3.50 µg mL− 1 0.017 µg mL− 1 42
CV GCE 0.05–0.5 mg/L 0.02 mg/L 43
CV GCE modified with CTAB 0.2–2.0 mg/dm3 0.1 mg/dm3 44
DPV MoS2 NSs/ SPE 0.1–400 µM 0.03 µM 45
DPV ZnO HQSs - SPGE 0.08–190 µM 0.02 µM 46
ASV HMDE 0.03–0.9 µM 0.004 µM 47
DPV g-C3N4/ILs/CPE 0.4–125 µM 0.1 µM 48
DPV CPT- BDDE 0.059–1.31 µmolL− 1 7.0 nmolL− 1 49
CV/ DPV SDSMCS

0.1 µM to 1.1 µM /

0.1 µM to 0.9 µM

0.009 µM

/0.073 µM

Present work

GO-Fe3O4-G4 PAMAM/ILCPE - graphene oxide Fe3O4 fourth generation poly (amidoamine) ionic liquid modified carbon paste electrode.

SQV- Square wave voltammetry.

CPE/1-M-3BIBr/NiO/CNTs - carbon paste and NiO/CNTs composite modified with 1-methyl-3-butylimidazolium bromide.

GCE - glassy carbon electrode.

CTAB - cetyltrimethylammonium bromide.

MoS2 NSs/ SPE - screen printed electrode modified with molybdenum disulfide nanosheets.

ZnO HQSs - SPGE - screen printed electrode modified with zinc oxide hollow quasi- spheres.

ASV - adsorptive stripping voltammetry.

HMDE - hanging mercury drop electrode.

g-C3N4/ILs/CPE - graphitic carbon nitride/1-butyl-3-methylimidazolium hexafluorophosphate ionic liquids modified carbon paste electrode.

CPT- BDDE - Cathodically pretreated boron - doped diamond electrode.

Applicability of SDSMCS in real samples

The application of the developed sensor in practical samples was investigated via CV technique. The formulated solution of food color and gems were subjected to the analysis and the current responses for the concentration ranging from 0.1 µM to 0.4 µM and 0.6 µM to 0.9 µM (for candies and food color are demonstrated in Fig. 12(a) and 12(b), correspondingly.

Fig. 12.

Fig. 12

CV responses of (a) candies and (b) food color in 0.2 M PBS at a scan rate of 0.1 V/s.

From the data gained from the analysis, the recovery rates were calculated and the outcome was > 97% as illustrated in Table 2. From this analysis, it is apparent that the developed SDSMCS is highly efficient in detecting CM in food samples with considerable recovery rates. Additionally, this substantial outcome can be accredited to the highly selective and reliable voltammetric technique employed in the analysis. In comparison with chromatographic and spectrophotometric techniques, CV is less vulnerable to matrix effects as the detection is based on the selectivity limited to the electro- active species undergoing redox reactions at the applied potential, rather than retention time or absorbance50.

Table 2.

Recovery rates of food samples.

Sample Added (µM) Found (µM) Recovery %
Candies 0.1 0.097 97.10
0.2 0.195 97.59
0.3 0.296 98.81
0.4 0.389 97.34
Food color 0.6 0.588 98.04
0.7 0.690 98.59
0.8 0.779 97.33
0.9 0.882 97.96

Effect of interferents on oxidation peak potential

In the course of analysis, the presence of interferents might disrupt the current signals leading to inaccurate results. To analyze the effect of interfering species on the efficacy of SDSMCS, CV responses were documented for CM (1.0 mM) in PBS in the presence of interferents such as Na+, K+, Ca2+, Pb2+, Zn2+ and Mg2+ ions. In addition to these metal ions, the CV analysis was performed in the presence of simple sugars like fructose (Fr) and galactose (Ga) and dyes such as indigo carmine (IC) and methyl orange (MO). The observed results have been represented in Fig. 13 which demonstrates that the presence of interferents caused a slight variation of less than 5% in the peak potential, but no marked difference was noticed in the oxidation peak current values. This analysis authenticates the sensitivity of SDSMCS even in the existence of interfering metal ions, hence proving it as a reliable tool for the analysis of CM.

Fig. 13.

Fig. 13

Graphical representation of potential variation in the presence of interferents in CV analysis of CM in PBS.

Stability, repeatability and reproducibility of SDSMCS

The efficiency and reliability of the prepared SDSMCS were further evaluated by analyzing the attributes like repeatability, reproducibility and stability. CM (1.0 mM) was subjected to CV under optimum experimental conditions at the surface of five different SDSMCS. The current responses were observed to be almost similar with RSD of 4.95%, authenticating the high reproducibility of SDSMCS. Additionally, the current responses were recorded via CV at a single SDSMCS surface for five different solutions of 1.0 mM CM in PBS. The outcome demonstrated a great repeatability with RSD of 4.90%. Further, the current responses were documented with a gap of eight days (first day and the last day of work). The current maintenance noticed was 98.8%, hence substantiating the stability of SDSMCS for a significant time period.

Conclusion

The fabrication of SDSMCS and its implementation in the assessment of CM analysis has been presented in this work. The morphological study of the bare and the modified sensors were performed by SEM which provided a clear distinction in the surfaces to verify the deposition of SDS on the sensor surface. At pH of 6.5, the anodic peak current of CM at SDSMCS was recorded as 96.38 µA at anodic peak potential of 0.5762 V. The adsorption-controlled oxidation process was found to be maximum with transfer of equal number of electrons and protons. The CV and DPV measurements demonstrated a steady increase of the anodic peak current with the varied CM concentration at optimal experimental conditions. The relevant data from the calibration plot provided a LOD and LOQ value of 0.009 µM and 0.031 µM (by CV) and 0.073 µM and 0.244 µM (by DPV), correspondingly. The practical application of SDSMCS was gauged and the outcome was remarkable with recovery rate exceeding 97% for food samples. The analysis substantiated the superior anti-interference detecting capacity of SDSMCS along with significant repeatability, reproducibility and stability attributes. Certain developments such as incorporation of nano materials and the use of bio-based surfactants in the analysis could further advance the performance and sustainability of the current work.

Acknowledgements

J. G. Manjunatha thankfully acknowledges the financial support from VGST, Bangalore under the Research project no. KFIST(L2)/GRD-1020/2021-22/430.Gaber E. Eldesoky thankfully acknowledges the support by the Ongoing Research Funding Program (ORF- 2026-161), King Saud University, Riyadh, Saudi Arabia.

Author contributions

All authors have contributed to the study, conception and design of the current work. Conceptualization, material preparation, data collection, analysis and manuscript writing were performed by Battira M. Sharmila, Karnayana P. Moulya and Jamballi G. Manjunatha. Revision, editing and supervision of all phases of the analysis were completed by Jamballi G Manjunatha, Thiago C. Canevari and Gaber E. Eldesoky .

Data availability

The data that support the findings of this study are available on request from the corresponding author.

Declarations

Competing interests

The authors declare no competing interests.

Ethical issues

No ethical issues for this work.

Footnotes

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.Negi, A. Natural Dyes and Pigments: Sustainable Applications and Future Scope. Sustain. Chem.6, 23 (2025). [Google Scholar]
  • 2.Kumari, R. et al. Highly efficient and reliable voltammetry food sensor for Tartrazine dye using a nanocomposite reformed electrode. Microchem J.196, 109583 (2024). [Google Scholar]
  • 3.Barrera, H. et al. Electro-oxidation–plasma treatment for azo dye carmoisine (acid red 14) in an aqueous solution. Materials13, 1463 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Basu, A. & Suresh Kumar, G. Minor groove binding of the food colorant carmoisine to DNA: Spectroscopic and calorimetric characterization studies. J. Agric. Food Chem.62, 317–326 (2014). [DOI] [PubMed] [Google Scholar]
  • 5.Amin, K. A., Hameid, A., Abd Elsttar, A. H. & H. & Effect of food azo dyes tartrazine and carmoisine on biochemical parameters related to renal, hepatic function and oxidative stress biomarkers in young male rats. Food Chem. Toxicol.48, 2994–2999 (2010). [DOI] [PubMed] [Google Scholar]
  • 6.Lipskikh, O. I., Korotkova, E. I., Dorozhko, E. V., Derina, K. V. & Voronova, O. A. Voltammetric determination of carmoisine in soft drinks. Inorg. Mater.53, 1427–1431 (2017). [Google Scholar]
  • 7.Sohrabi, M. R., Khavaran, A., Shariati, S. & Shariati, S. Removal of carmoisine edible dye by Fenton and photo Fenton processes using Taguchi orthogonal array design. Arab. J. Chem.10, S3523–S3531 (2017). [Google Scholar]
  • 8.Mahale, S. R., Rajendrachari, S., Shamanth, V. & Vinaykumar, R. Voltammetric determination of various food azo dyes using different modified carbon paste electrodes. Biointerface Res. Appl. Chem.12, 4557–4566 (2022). [Google Scholar]
  • 9.Ozdemir, Y. & Akkan, A. A. Determination of Patent Blue V and carmoisine in gelatine desserts by derivative spectrophotometry. Turk. J. Chem.23, 221–229 (1999). [Google Scholar]
  • 10.Aguilar, F. et al. Scientific opinion on the re-evaluation of azorubine/carmoisine (E 122) as a food additive. EFSA J.7, 1332 (2009). [Google Scholar]
  • 11.Mitra, M., Dehghan, G., Hosseinpourfeizi, M. A. & Moosavi-Movahedi, A. A. Spectroscopic and electrochemical studies on the interaction of carmoisine food additive with native calf thymus DNA. Spectrosc. Lett.46, 250–256 (2013). [Google Scholar]
  • 12.Sasaki, Y. F. The comet assay with 8 mouse organs: results with 39 currently used food additives. Mutat. Res. Genet. Toxicol. Environ. Mutagen.519, 103–119 (2002). [DOI] [PubMed] [Google Scholar]
  • 13.Novais, C. Natural food colorants and preservatives: A review, a demand, and a challenge. J. Agric. Food Chem.70, 2789–2805 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Sarvestani, J., Reza, M. & Doroudi, Z. Determination of Ultra Trace Amounts of Carmoisine in Food Specimens by Ultrasound-assisted Surfactant-enhanced Emulsification Microextraction Method Coupled with UV-Visible Spectrophotometry. J Chem. Health Risks12, 409-419 (2022).
  • 15.Karatepe, A., Akalin, C. & Soylak, M. Spectrophotometric determination of carmoisine after cloud point extraction using Triton X-114. Turk. J. Chem.41, 256–262 (2017). [Google Scholar]
  • 16.Abdullah, N. S., Hassan, M. A. & Hassan, R. O. A Novel Flow Injection Method with Chemiluminescence Detection for the Determination of Carmoisine in Gelatin Desserts. J. Fluoresc. 35, 3405–3413 (2025). [DOI] [PubMed] [Google Scholar]
  • 17.Xiaoling, L. et al. Colloidal gold-based immunochromatographic assay for the detection of carmoisine in soft drinks. Food Biosci.66, 106220 (2025). [Google Scholar]
  • 18.El Boumlasy, A. M., Bouchta, S., Choukairi, M. & D. & Development and applications of carbon paste and Sonogel-Carbon electrodes modified with nanomaterials: Perspectives in pharmaceutical, biological, environmental and food analysis: A review. Trends Anal. Chem.194, 118502 (2026). [Google Scholar]
  • 19.Trojanowicz, M. Applications of gold nanoparticles in electroanalysis in Comprehensive Analytical Chemistry (eds. Valcárcel, M. & López-Lorente, Á. I) 429–476 ( Elsevier, 2014).
  • 20.Sharmila, B. M. An electrochemically modified high-performance sensor for the voltammetric analysis and quantification of diabetes inducing alloxan in food samples. J. Mater. Sci. : Mater. Electron.36, 557 (2025). [Google Scholar]
  • 21.Mayet, A. M. Molecularly imprinted polymers for biosensing of hormones in food safety and biomedical analysis: Progress and perspectives. Mater. Today Chem.35, 101899 (2024). [Google Scholar]
  • 22.Mulligan, C. N. Rhamnolipid biosurfactants: Solubility and environmental issues in Thermodynamics, Solubility and Environmental Issues (ed. Letcher, T. M) 279–298 (Elsevier, 2007).
  • 23.Bulian, F. & Graystone, J. A. Raw materials for wood coatings (2) – solvents, additives and colorants in Wood Coatings (eds. Bulian, F. & Graystone, J. A) 95–135 (Elsevier, 2009).
  • 24.Prinith, N. S., Manjunatha, J. G. & Hareesha, N. Electrochemical validation of L-tyrosine with dopamine using composite surfactant modified carbon nanotube electrode. J. Iran. Chem. Soc.18, 3493–3503 (2021). [Google Scholar]
  • 25.Amrutha, B. M., Manjunatha, J. G., Bhatt, A. S., Chenthattil, R. & Pushpanjali, P. A. Electrochemical sensor for the determination of alizarin red-S at non-ionic surfactant modified carbon nanotube paste electrode. Phys. Chem. Res.7, 523–533 (2019). [Google Scholar]
  • 26.Zhu, M. et al. Differential pulse stripping voltammetric determination of metronidazole with graphene-sodium dodecyl sulfate modified carbon paste electrode. Int. J. Electrochem. Sci.13, 4100–4114 (2018). [Google Scholar]
  • 27.Švancara, I. & Sýs, M. Carbon Paste Electrodes Surface-Modified with Surfactants: Principles of Surface Interactions at the Interface between Two Immiscible Liquid Phases. Sensors23, 9891 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Selva Kumar, A. K. & Compton, R. G. Understanding carbon nanotube voltammetry: Distinguishing adsorptive and thin layer effects via single-entity electrochemistry. J. Phys. Chem. Lett.13, 5557–5562 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Umasankar, Y., Unnikrishnan, B., Chen, S. & Ting, T. Effective determination of acetaminophen present in pharmaceutical drug using functionalized multi-walled carbon nanotube film. Int. J. Electrochem. Sci.7, 484–498 (2012). [Google Scholar]
  • 30.Quintero-Jaime, A. F., Conzuelo, F., Schuhmann, W., Cazorla-Amorós, D. & Morallón, E. Multi-wall carbon nanotubes electrochemically modified with phosphorus and nitrogen functionalities as a basis for bioelectrodes with improved performance. Electrochim. Acta. 387, 138530 (2021). [Google Scholar]
  • 31.Lu, Y., Li, X. & Compton, R. G. Electro-oxidation of amino-functionalized multiwalled carbon nanotubes. Chem. Sci.13, 1355–1366 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Moulya, K. P., Manjunatha, J. G., Hareesha, N., Habila, M. A. & Sillanpää, M. Electrochemically polymerized dopamine activated carbon paste sensor for selective and sensitive detection of rutin in the presence of riboflavin. Microchem J.201, 110561 (2024). [Google Scholar]
  • 33.Hoyos-Arbeláez, J., Vázquez, M. V. & Contreras-Calderón, J. Electrochemical methods as a tool for determining the antioxidant capacity of food and beverages: A review. Food Chem.221, 1371–1381 (2017). [DOI] [PubMed] [Google Scholar]
  • 34.Achache, M. et. al. A novel carbon paste electrode modified with neodymium nanocomposites for the electrochemical detection of caffeic acid in coffee and fruit juice. Microchem J.217, 115066 (2025). [Google Scholar]
  • 35.Kozak, J., Tyszczuk-Rotko, K., Wójciak, M. & Sowa, I. Electrochemically activated screen-printed carbon sensor modified with anionic surfactant (aSPCE/SDS) for simultaneous determination of paracetamol, diclofenac and tramadol. Materials14, 3581 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Lebègue, E., Bard, A. J., Faulkner, L. R. & White, H. S. Electrochemical Methods: Fundamentals and Applications, 3rd edition. Transit. Met. Chem.48, 433–436 (2023).
  • 37.Chebotarev, A. N., Pliuta, K. V. & Snigur, D. V. Determination of Carmoisine onto Carbon-Paste Electrode Modified by Silica Impregnated with Cetylpyridinium Chloride. ChemistrySelect5, 3688–3693 (2020). [Google Scholar]
  • 38.Sarvestani, J., Mohammad, R. & Doroudi, Z. A Comprehensive Review on Electroanalytical Methodologies for the Determination of Carmoisine (E122). Food Anal. Methods. 15, 1–10 (2022). [Google Scholar]
  • 39.Shashank, K. R. et al. An Electrochemically Polymerized Glycine-Modified Carbon Paste Sensor for the Analysis of Carmoisine in the Presence of Environmentally Significant Dyes. J. Electron. Mater.54, 9811–9825 (2025). [Google Scholar]
  • 40.Nejad, F. G., Sheikhshoaie, I. & Beitollahi, H. Simultaneous detection of carmoisine and tartrazine in food samples using GO-Fe₃O₄-PAMAM and ionic liquid based electrochemical sensor. Food Chem. Toxicol.162, 112864 (2022). [DOI] [PubMed] [Google Scholar]
  • 41.Bijad, M., Karimi-Maleh, H., Farsi, M. & Shahidi, S. An electrochemical-amplified platform based on the nanostructure voltammetric sensor for the determination of carmoisine in the presence of tartrazine in dried fruit and soft drink samples. J. Food Meas. Charact.12, 634–640 (2018). [Google Scholar]
  • 42.Pourreza, N. & Ghomi, M. Simultaneous cloud point extraction and spectrophotometric determination of carmoisine and brilliant blue FCF in food samples. Talanta84, 0–243 (2011). [DOI] [PubMed] [Google Scholar]
  • 43.Lipskikh, O. I. et al. Voltammetric determination of carmoisine in soft drinks. Inorg. Mater.53, 1427–1431 (2017). [Google Scholar]
  • 44.Kropacheva, T. & Elmashev, T. A. Voltammetric determination of carmoisine in beverages using surfactant-modified electrodes. Her Technol. Univ.27, 24–28 (2024). [Google Scholar]
  • 45.Abboud, H. J. et al. MoS₂ nanosheets-modified screen-printed electrode for the simultaneous detection of carmoisine and tartrazine. J. Electrochem. Sci. Eng.15, 2620 (2025). [Google Scholar]
  • 46.Mohammadi, S. Z., Tajik, S., Mousazadeh, F., Baghadam-Narouei, E. & Nejad, F. G. ZnO hollow quasi-spheres modified screen-printed graphite electrode for determination of carmoisine. Micromachines14, 1433 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 47.Nhan, H. et al. Determination of carmoisine in food by adsorptive stripping voltammetry. VNU J. Sci. : Nat. Sci. Technol.41, 3 (2025). [Google Scholar]
  • 48.Mohammadi, S. Z. & Mousazadeh, F. A sensitive sensor based on graphitic carbon nitride/1-butyl-3-methylimidazolium hexafluorophosphate ionic liquids modified carbon paste electrode for sensing of carmoisine dye in food samples. J. Electrochem. Sci. Eng.12, 1143–1152 (2022). [Google Scholar]
  • 49.Micheletti, L., Coldibeli, B., Salamanca-Neto, C., Almeida, L. C. & Sartori, E. R. Assessment of the use of boron-doped diamond electrode for highly sensitive voltammetric determination of the azo-dye carmoisine E – 122 in food and environmental matrices. Talanta220, 121417 (2020). [DOI] [PubMed] [Google Scholar]
  • 50.Kalcher, K. et al. Sensors based on carbon paste in electrochemical analysis: A review with particular emphasis on the period 1990–1993. Electroanalysis7, 5–22 (1995). [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

The data that support the findings of this study are available on request from the corresponding author.