Cardiac remodelling and functional status after cardiac resynchronization therapy: comparison between de-novo implantation and upgrade from right ventricular pacing

Henrik Laurits Bjerre 1,2,✉, Emil Anton Frandsen 3, Anders Sommer 4, Charlotte Stephansen 5, Niels Risum 6, Steen Hvitfeldt Poulsen 7,8, Jens Cosedis Nielsen 9,10, Mads Brix Kronborg 11,12
PMCID: PMC13344856  PMID: 42361293

Abstract

Introduction

Patients with heart failure induced by chronic right ventricular (RV) pacing commonly undergo upgrade to cardiac resynchronization therapy (CRT). We aimed to compare the effect size of improvement in cardiac function and functional capacity between patients with RV pacing and intrinsic conduction at baseline.

Methods

We included patients from the ImagingCRT, ElectroCRT and DANISH-CRT trials aged ≥40 years with left ventricular ejection fraction (LVEF) ≤35% and prolonged QRS. The primary endpoint was change in left ventricular end-systolic volume (LVESV) from baseline to 6-month follow-up. Secondary endpoints included change in QRS duration, echocardiographic measures (LVEF, left ventricular end-diastolic volume, left ventricular mass index, and left atrial volume index), N-terminal prohormone of brain natriuretic peptide, six-minute walk test, quality of life, New York Heart Association functional class and loop diuretic use.

Results

We included 592 patients [mean age 70.7 ± 9 years; 24.8% women; mean LVEF 27 ± 6%; 97 (16.4%) with RV pacing and 495 (83.6%) with intrinsic conduction at baseline]. We found significant relative reduction in LVESV in patients with RV pacing (−34 ± 25%) and intrinsic conduction (−32 ± 26%) with no difference between the groups [mean difference −2%, 95% CI (−8; 4), P = .469]. Absolute LVEF improvement was 14 ± 9% and 13 ± 10%, respectively. Reduction in QRS duration was larger with RV pacing compared with intrinsic conduction (−38 ± 26 vs −26 ± 24 ms). We found significant improvements in the remaining secondary endpoints, with no difference between the groups.

Conclusion

Patients with RV paced QRS morphology who are upgraded to CRT derive substantial improvement in cardiac function and functional capacity, comparable to patients with intrinsic conduction undergoing CRT.

Keywords: Cardiac device, Cardiac resynchronization therapy, CRT, Biventricular pacing, RV pacing, Upgrade

Graphical Abstract

Graphical Abstract.

For image description, please refer to the figure legend and surrounding text.

Effect of biventricular cardiac resynchronization therapy (CRT) in patients with intrinsic conduction and right ventricular (RV) pacing at baseline. Patients with RV pacing at baseline who were upgraded to CRT derived substantial improvement in cardiac remodelling and functional capacity from baseline to 6 months follow-up. The effect size was comparable to patients with intrinsic conduction at baseline undergoing CRT.

Introduction

Chronic right ventricular (RV) pacing can lead to worsening of left ventricular (LV) function and induce symptomatic heart failure (HF).1–3 Patients with pacing-induced HF may be upgraded to a biventricular cardiac resynchronization therapy (CRT) device by adding an epicardial LV lead in a coronary sinus tributary vein and replacing the generator.4 Most evidence on the beneficial effects from CRT originates from patients with intrinsic conduction undergoing de novo implantation, while less research has investigated efficacy of CRT in patients who are upgraded from a previous pacemaker or implantable cardioverter defibrillator (ICD) due to HF considered induced by RV pacing.5 A recent meta-analysis6 with six (n = 161) small randomized controlled trials (RCT) and 47 observational studies (n = 2,644) found improved LVEF and functional capacity following upgrade from a pacemaker or ICD to a biventricular CRT device. Likewise, the BUDAPEST-CRT7 trial showed that upgrade to a CRT defibrillator (CRT-D) was superior to an ICD on the primary composite endpoint of all-cause mortality, HF hospitalization and cardiac remodelling. Although this demonstrates a beneficial effect of upgrade procedures, studies have reported conflicting results on whether the size of this effect is comparable to that seen in de novo procedures.8–12 The interpretation of these results is made additionally complex by several advancements over the past 25 years, including different patient selection and technological improvements such as quadripolar leads, optimization algorithms and targeting strategies for lead positioning, as well as evolution in pharmacological HF treatment. In a contemporary setting, it therefore remains unclear whether patients with an RV paced QRS who are upgraded to CRT derive similar short-term benefits as patients with intrinsic conduction who undergo a de novo CRT procedure. In a large contemporary CRT cohort, this study aimed to evaluate the effect size of six-month benefits in terms of cardiac remodelling and functional capacity in patients with RV paced QRS and intrinsic conduction (Graphical abstract).

Methods

Design and study population

This region-wide prospective cohort study comprised patients enrolled in three RCTs; ‘Multimodality Imaging-guided Left Ventricular Lead Placement in Cardiac Resynchronization Therapy’ (ImagingCRT),13 ‘Electrically vs Imaging-guided Implant of the LV Lead in Cardiac Resynchronization Therapy’ (ElectroCRT)14 and ‘Does Electric Targeted LV Lead Positioning Improve Outcome in Patients With Heart Failure and Prolonged QRS’ (DANISH-CRT).15 All trials were approved by The Central Denmark Regional Committee on Health Research Ethics (files no. 20100245, 1-10-72-230-14 and 1-10-72-330-16, respectively) and the Danish Data Protection Agency (files no. 2011-41-6017, 1-16-02-622-14 and 1-16-02-660-16, respectively) and conformed to the declaration of Helsinki. All patients provided written informed consent. Patients were referred from one of five hospitals in The Central Denmark Region, comprising a population of 1.3 million, and all implantations were performed at Aarhus University Hospital, Denmark. Patients were enrolled in the respective trials from 2011–2014, 2015–2017 and 2018–2024. The inclusion criteria have been previously described.15–17 Briefly, the trials enrolled patients aged ≥40 years with symptomatic HF corresponding to New York Heart Association (NYHA) functional class II-IVa with LVEF ≤35% on echocardiography despite optimal medical therapy. Eligibility with respect to QRS duration varied slightly between trials: left bundle branch block (LBBB) with QRS ≥120 ms or RV paced QRS ≥180 ms in ImagingCRT and ElectroCRT, and true LBBB with QRS ≥130 ms, LBBB-like, right bundle branch block (RBBB) or interventricular conduction delay (IVCD) all with QRS≥150 ms or RV paced QRS with >50% RV pacing in the DANISH-CRT trial. Trial participants were excluded from this pooled cohort study if echocardiograms were not available at baseline and follow-up. Patients were followed from baseline to 6 months follow-up. The LV lead was guided to the site of latest mechanical activation, latest electrical activation or according to the operator’s preference depending on the assigned randomization arm in each respective trial. The RV lead was aimed towards a mid-septal position, with post-implant optimization of AV- and VV-delays according to trial-specific protocols using echocardiographic and/or electrocardiographic guidance. Optimization protocols were identical for patients with RV pacing and intrinsic conduction at baseline.15–17 The highest achievable degree of biventricular pacing was prioritized rather than fusion with intrinsic conduction (i.e. LV pacing only or triple fusion).

Endpoints

The primary endpoint was relative change in LV end-systolic volume (LVESV) from baseline to 6-month follow-up. Secondary endpoints included change in LV end-diastolic volumes (LVEDV), LVEF, left atrial volume index (LAVI), LV mass index (LVMI), QRS duration on surface ECG, distance on six-minute walk test (6MWT), N-terminal prohormone of brain natriuretic peptide (NT-proBNP), NYHA functional class, quality of life (QoL) and use of loop diuretics. We also evaluated conventional measures of CRT response defined as ≥15% relative reduction in LVESV, ≥5% absolute increase in LVEF or ≥1 reduction in NYHA functional class. Echocardiograms were analysed in core-lab/single-centre setup and blinded for other echocardiograms from the same patient. All echocardiograms were performed on General Electric (GE) healthcare scanners (Vivid E9 in ImagingCRT and ElectroCRT and Vivid E95 in DANISH-CRT) and analysed using identical software (EchoPac, GE healthcare). LVESV, LVEDV, and LVEF were assessed using the Simpson’s biplane method. LAVI was measured using the method of discs, and LVMI was measured using the American Society of Echocardiography modified equation, both indexed to the body surface area (Du Bois’ formula). LAVI and LVMI measurements were only available for patients enrolled in DANISH-CRT while NT-proBNP at follow-up and 6MWT at baseline only were measured in ImagingCRT and ElectroCRT. QoL was assessed with the 21-item Minnesota Living with Heart Failure questionnaire18 (range 0–105, lower values indicate better QoL) in ImagingCRT and ElectroCRT and with the 12-item Kansas City Cardiomyopathy Questionnaire19 (range 0–100, higher values indicate better QoL) in DANISH-CRT. To enable comparison across the instruments, we used the standardized ΔZ-score relative to the baseline standard deviation (SD) with positive values reflecting improvement irrespective of questionnaires.

Statistics

All demographic, procedural and medical data were collected and stored in electronic web-based case report forms (eCRF). Continuous variables were reported as means and standard deviations (SD) in case of normal distribution and as medians and interquartile range (IQR) in absence of normal distribution. Normality was assessed by visual inspection of histograms and Q-Q plots. Categorical variables were presented as absolute numbers and percentages. Endpoints were compared using unpaired t-test for continuous variables in case of normal distribution, Wilcoxon rank-sum test in absence of normal distribution and Fisher’s Exact test for categorical variables. Endpoints were presented as mean difference with 95% confidence intervals (CI) or as median ratios after back-transformation from log-transformation. The primary endpoint was evaluated using a multivariable linear regression model to explore the influence of measured potential confounders. Covariates were selected a priori using directed acyclic graphs (DAG). Potential effect-modification by trial was explored by testing the interaction between trial and baseline QRS morphology (RV pacing vs intrinsic conduction). We used box plots with medians, IQR and upper and lower adjacent values to visualize distribution of continuous endpoint variables. For the reported P-values in the box plots we used Wilcoxon rank-sum test for between-group differences and Wilcoxon signed rank test for paired analysis of changes from baseline to 6-month follow-up. We performed sensitivity analysis for all endpoints with exclusion of (i) patients with intrinsic non-LBBB QRS morphology and (ii) patients LVEF >35% at baseline. For the comparison of RV pacing, intrinsic LBBB and intrinsic non-LBBB we used one-way ANOVA and subsequent pairwise comparison with Bonferroni correction, or Kruskal–Wallis test pending normality. Lastly, we conducted hypothesis-generating subgroup analyses according to pre-existing device type (pacemaker or ICD) and duration of RV pacing (earlier or later than the median duration from the patient’s first device to CRT upgrade). Statistical analysis was performed using Stata (version 18; StataCorp).

Results

We identified 727 patients enrolled from April 2011 to June 2024. We excluded 135 patients who did not have available echocardiograms from both baseline and 6-month follow-up. Of the total 592 patients included, 97 (16.4%) were upgraded from a pre-existing pacemaker or ICD and had RV paced QRS complex at baseline. The remaining 495 (83.6%) were de novo recipients or patients who were upgraded from a pre-existing device but had intrinsic conduction at baseline. In patients with intrinsic conduction, 476 (96%) had LBBB QRS morphology while only 19 (4%) had LBBB-like morphology, right bundle branch block (RBBB) or IVCD. Patients with RV paced QRS at baseline were upgraded from a pacemaker (81%) more often than an ICD (19%). In contrast, those with a pre-existing device, but who had intrinsic conduction, were upgraded from an ICD (78%) more often than a pacemaker (22%). Patients with RV paced QRS were older and more likely to have ischaemic heart disease and atrial fibrillation compared with patients with intrinsic conduction (Table 1). Both patient groups had substantial use of guideline-directed medical therapy (GDMT) with prescription rates of 85–88% for betablockers and 94–96% for angiotensin converting enzyme inhibitors (ACEi), angiotensin-II-receptor blockers (ARB) or neprilysin inhibitors (ARNi), with 27–36% of patients achieving target dose. However, more patients with intrinsic conduction than with RV pacing were treated with spironolactone or eplerenone (Table 2). At baseline, patients with RV paced QRS had longer QRS duration but better cardiac function (lower LVESV, lower LVEDV and slightly higher LVEF) compared with patients with intrinsic conduction (Table 3). They also had lower functional capacity with higher NYHA functional class and shorter distance on the 6MWT. Patients with RV paced QRS had greater reduction in QRS duration than patients with intrinsic conduction (Table 3).

Table 1.

Baseline characteristics

Total (n = 592) RV paced (n = 97) Intrinsic (n = 495) P-value
Age, years 73 (64–78) 75 (71–80) 72 (63–78) <.001
Sex, male 445 (75) 79 (81) 366 (74) .118
BMI, kg/cm2 27 (5) 28 (5) 27 (5) .405
Ischaemic heart disease 271 (46) 55 (57) 216 (44) .017
Hypertension 304 (52) 58 (61) 246 (50) .055
Diabetes 164 (28) 25 (26) 139 (28) .642
Chronic obstructive pulmonary disease 57 (10) 6 (6) 51 (10) .205
Atrial fibrillation <.001
 Persistent or permanent 92 (16) 35 (36) 57 (12)
 Paroxysmal 119 (20) 25 (26) 94 (19)
Previous device 161 (27) 97 (100) 64 (13) <.001
 Pacemaker 93 (58) 79 (81) 14 (22) <.001
  Original indication
   AV block 48 (52) 43 (54) 5 (36)
   Sinus node dysfunction 26 (28) 19 (24) 7 (50)
   Permanent atrial fibrillation 11 (12) 11 (14) 0 (0)
   Other/unknown 8 (9) 6 (6) 2 (14)
 ICD 68 (42) 18 (19) 50 (78) <.001
  Original indication
   Primary prevention 38 (56) 10 (56) 28 (56)
   Secondary prevention 30 (44) 8 (44) 22 (44)
 RV-pace% on last interrogation, % 92 (1–99) 99 (90–100) 1 (0–8)
 Time since previous device, years 4.0 (1.8–6.7) 3.8 (1.7–6.7) 4.5 (2.5–6.8) .167
 Time since first device, years 5.0 (2.3–8.9) 4.5 (1.9–7.6) 5.7 (3.1–9.4) .071
Systolic blood pressure, mmHg 122 (110–135) 123 (113–137) 122 (110–134) .218
Diastolic blood pressure, mmHg 71 (65–79) 71 (64–80) 71 (65–78) .440
Creatinine 95 (80–121) 98 (84–120) 95 (79–122) .085
eGFR .155
 <29 27 (5) 8 (8) 19 (4)
 30–60 237 (40) 38 (40) 199 (40)
 >60 325 (55) 50 (52) 275 (56)
Haemoglobin 8 (8–9) 8 (8–9) 8 (8–9) .892
Procedure time, min 90 (68–104) 90 (64–105) 90 (68–103) .694
X-ray time, min 21 (11–25) 21 (11–25) 21 (11–25) .617

Data are presented median (IQR) for continuous variables and n (%) for categorical measures. Between-group differences are assessed using the Wilcoxon rank-sum test for continuous variables and Pearson’s chi-square test for categorical variables. BBB, bundle branch block; BMI, body mass index; eGFR, estimated glomerular filtration rate; IQR, interquartile range; RV, right ventricular.

Table 2.

Medication at baseline

Total (n = 592) RV paced (n = 97) Intrinsic (n = 495) P-value
Beta blocker 524 (89) 82 (85) 442 (89) .221
 <50% of TD 191 (32) 35 (36) 156 (32)
 50–99% of TD 157 (27) 21 (22) 136 (27)
 ≥100% of TD 176 (30) 26 (27) 150 (30) .317
ACEi, ARB or ARNi 555 (94) 93 (96) 462 (93) .491
 <50% of TD 194 (33) 37 (38) 157 (32)
 50–99% of TD 151 (26) 26 (27) 125 (25)
 ≥100% of TD 208 (35) 30 (31) 178 (36) .504
Spironolactone/eplerenone 346 (59) 44 (45) 302 (61) .005
 <50% of TD 39 (7) 6 (6) 33 (7)
 50–99% of TD 268 (45) 32 (33) 236 (48)
 ≥100% of TD 39 (7) 6 (6) 33 (7) .031
SGLT2 inhibitor 88 (15) 14 (15) 74 (15) 1.000
Loop diuretic 409 (69) 68 (71) 341 (69) .809
 Furosemide equipotent dose 60 (40–120) 60 (40–120) 60 (40–120) .922
Thiazide diuretic 23 (4) 5 (5) 18 (4) .562
Calcium channel blocker 43 (7) 14 (15) 29 (6) .008
Digoxin 32 (5) 5 (5) 27 (5) 1.000
Amiodarone 65 (11) 10 (10) 55 (11) 1.000
Antiplatelet drug 318 (54) 43 (44) 275 (56) .045
Oral anticoagulation 207 (35) 52 (54) 155 (31) <.001

Categorical data are presented as n (%) and between-group differences were assessed using Fisher’s exact test. Continuous variable is presented as median (IQR) and between-group difference was assessed using Wilcoxon rank-sum test. ACEi, angiotensin converting enzyme inhibitor; ARB, angiotensin-II-receptor blocker, SGLT2, sodium-glucose-linked-transporter-2; TD, target dose.

Table 3.

Endpoints at baseline and follow-up and change from baseline to follow-up for patients with RV pacing and intrinsic conduction and between-group difference with 95% confidence intervals

RV paced (n = 97) Intrinsic (n = 495) Difference (95% CI) P-value
Biventricular pacing, %
 Follow-up 99 (97–99) 99 (96–99) — .092
QRS duration, ms
 Baseline 185 ± 21 166 ± 18 18 (15; 23) <.001
 Follow-up 146 ± 22 140 ± 20 6 (2; 11) .006
 Absolute change −38 ± 26 −26 ± 24 −12 (−17; −6) <.001
LVESV, mL
 Baseline 131 ± 52 157 ± 69 −26 (−40; −11) .001
 Follow-up 82 ± 37 104 ± 58 −21 (−33; −9) .001
 Absolute change −48 ± 43 −51 ± 51 5 (−6; 15) .402
 Relative change (%) −34 ± 25 −32 ± 26 −2 (−8; 4) .469
LVEDV, mL
 Baseline 182 ± 63 211 ± 82 −29 (−47; −12) .001
 Follow-up 141 ± 50 165 ± 70 −24 (−39; −10) .001
 Absolute change −41 ± 50 −46 ± 60 5 (−7; 18) .400
 Relative change (%) −20 ± 23 −19 ± 25 −1 (−6; 5) .797
LVEF, %
 Baseline 29 ± 6 27 ± 6 2 (1; 3) .004
 Follow-up 43 ± 9 40 ± 10 3 (1; 5) .007
 Absolute change 14 ± 9 13 ± 10 1 (−1; 3) .356
LV mass indexa, g/m2
 Baseline 121 ± 25 126 ± 34 −5 (−15; 5) .322
 Follow-up 108 ± 26 111 ± 34 −3 (−13; 7) .525
 Absolute change −13 ± 25 −14 ± 31 2 (−7; 10) .697
LA volume indexa, mL/m2
 Baseline 42 ± 19 37 ± 17 5 (0; 11) .069
 Follow-up 43 ± 21 36 ± 16 7 (2; 13) .010
 Absolute change 1 ± 12 −1 ± 11 2 (−1; 6) .195
NT-proBNPb,c, ng/L
 Baseline 1773 (984–3049) 1319 (637–3115) 1.29 (0.99; 1.67) .053
 Follow-up 1163 (786–1901) 687 (283–1602) 1.74 (1.10; 2.75) .019
 Relative change (GMR) 0.56 (0.43–0.72) 0. 48 (0.42–0.54) 1.17 (0.84; 1.64) .353
6MWT, m
 Baselineb 342 ± 143 395 ± 111 −53 (−90; −11) .014
 Follow-up 401 ± 100 438 ± 100 −37 (−62; −13) .003
 Absolute change 58 ± 65 41 ± 65 17 (−7; 42) .160
NYHA functional class
 Baseline I/II/III or IVa, n (%) 0(0)/42(43)/55(57) 0(0)/255(52)/240(48) — .299
 Follow-up I/II/III or IVa, n (%) 25(26)/53(56)/17(18) 165(33)/270(55)/59(12) — .297
 ≥1 improvement, n (%) 57 (60) 320 (65) — .414
Quality of Life
MLWHF b
 Baseline 36 ± 21 34 ± 22 2 (−5; 10) .536
 Follow-up 21 ± 19 20 ± 19 1 (−5; 8) .706
 Absolute change −16 ± 25 −14 ± 18 −2 (−8; 5) .553
KCCQ12 a
 Baseline 54 ± 24 57 ± 19 −4 (−10; 2) .227
 Follow-up 68 ± 20 72 ± 19 −4 (−10; 1) .138
 Absolute change 15 ± 22 14 ± 18 1 (−4; 7) .645
Standardized change, Δ z-score 0.74 ± 1.12 0.65 ± 0.86 0.09 (−0.11; 0.30) .387
Loop diuretics, mg
 Baseline 60 (40–120) 60 (40–80) —
 Follow-up 40 (40–100) 40 (40–80) —
 Dose reduction, n (%) 23 (33) 102 (29) — .473

Data are presented as mean (SD) or median (IQR) for continuous variables pending normal distribution and n (%) for categorical measures. Between-group differences are assessed using unpaired t-test in case of normal distribution, the Wilcoxon rank-sum test in absence of normal distribution and Pearson’s chi-square test for categorical variables. Within-group differences are assessed using the paired t-test in case of normal distribution and the Wilcoxon signed rank test in absence of normal distribution.

aData only available from the DANISH-CRT trial (n = 300).

bData only available from the Imaging CRT and Electro CRT trials (n = 292).

cThe median ratio was calculated as the geometrical mean ratio = the mean difference on the log scale. The median ratio may differ from the ratio of raw medians due to distributional asymmetry. QoL was assessed with MLWHF (range 0–105, lower scores reflect better QoL) and KCCQ12 (range 0–100, higher values indicate better QoL) and standardized Δ z-score using the baseline standard deviation (positive values indicate improvement) was used to enable comparison across instruments. 6MWT, Six-minute walk test; KCCQ12, Kansas City Cardiomyopathy Questionaire 12; LVEDV, left ventricular end-diastolic volume; LVEF, left ventricular ejection fraction; LVESV, left ventricular end-systolic volume; MLWHF, Minnesota Living with Heart Failure; NT-proBNP, N-terminal pro-Brain Natriuretic Peptide; NYHA, New York Heart Association.

Regarding the primary endpoint (Figure 1, Table 3), we found a significant relative reduction in LVESV from baseline to 6-month follow-up for patients with RV pacing and intrinsic conduction (−34 ± 25% and −32 ± 26%, respectively) with no significant between-group difference [mean difference −2%, 95% CI (−8; 4), P = .469]. This was consistent in the multivariable linear regression model adjusting for age, sex, ischaemic heart disease, atrial fibrillation, QRS duration at baseline and trial [mean difference −2%, 95% CI (−8; 5), P = .633, Supplementary Table S1]. There was no evidence of effect-modification by trial (P = .65). Likewise, we found a significant reduction in LVEDV, LVMI, NT-proBNP and use of loop diuretics, and increase in LVEF and 6MWT distance, with no significant between-group differences. LAVI did not change from baseline to 6-month follow-up in any groups (Table 3). QoL improved significantly (mean standardized ΔZ-score 0.74 ± 1.11 and 0.65 ± 0.86 times the SD at baseline for RV pacing and intrinsic conduction, respectively) with no significant difference between the groups (Table 3). We found no differences between the groups in improvement on NYHA functional class (Figure 2). In terms of conventional definitions of CRT response, we observed slightly higher rates of echocardiographic response (84.5 vs 76.9%, P = .078 and 86.6 vs 78.7%, P = .095, for relative reduction in LVESV ≥15% and absolute improvement in LVEF ≥5%, respectively) and slightly lower response in ≥1 improvement in NYHA functional class (58.8% vs 64.6%, P = .299) with RV pacing compared with intrinsic conduction but none reached statistical significance (Figure 3).

Figure 1.

Boxplots showing that left ventricular end-systolic volume, left ventricular end-diastolic volume and left ventricular ejection fraction differs at baseline and at 6-month follow-up, but the change from baseline to 6-month follow-up following cardiac resynchronization therapy is similar between patients with right ventricular pacing and intrinsic conduction

Boxplots showing median and IQR with upper and lower adjacent values representing the maximum value within 1.5 times the IQR at baseline and 6-month follow-up (left panel) and change from baseline to follow-up (right panel) for LVESV (upper panel), LVEDV, (middle panel) and LVEF (lower panel) for patients with RV pacing and intrinsic conduction. Values outside the upper and lower adjacent values are omitted due to data sensitivity (n < 5). Between-group differences were tested using the Wilcoxon rank-sum test and paired change from baseline to follow-up was tested using the Wilcoxon signed-rank test. IQR, interquartile range; LVEDV, left ventricular end-diastolic volume; LVEF, left ventricular ejection fraction; LVESV, left ventricular end-systolic volume

Figure 2.

Bar chart showing a similar improvement in New York Heart Association functional class from baseline to 6-month follow-up following cardiac resynchronization therapy between patients with RV pacing and intrinsic conduction

Distribution of NYHA functional class at baseline and follow-up for patients with RV pacing and intrinsic conduction. NYHA, New York Heart Association, RV, right ventricular

Figure 3.

Bar chart showing response rate following cardiac resynchronization therapy between patients with RV pacing and intrinsic conduction according to conventional definitions of treatment response which are a 15% or larger relative reduction in left ventricular end-systolic volume, a 5% or larger absolute increase in left ventricular ejection fraction, a 1 or more improvement in New York Heart Association functional class, respectively, or any of the three

Percentage of patients classified as responders according to conventional definitions of CRT response including relative reduction in LVESV ≥15% (upper left), absolute increase in LVEF ≥5% (upper right), ≥1 improvement in NYHA functional class (lower left) or any of the previous parameters (lower right)

Patients with non-LBBB intrinsic conduction had significantly less improvement from baseline to 6-month follow-up compared with both intrinsic LBBB and RV paced QRS on the primary (ΔLVESV −13 ± 24% vs −34 ± 25 vs −33 ± 26, respectively, P < .001) and secondary endpoints (Supplementary Table S2). Sensitivity analysis was consistent with the primary analysis on all endpoints when excluding patients with non-LBBB QRS morphology (n = 19, Supplementary Table S2) and patients with LVEF >35% at baseline (n = 30, Supplementary Table S3), respectively. Subgroup analyses suggested that patients with a pre-existing pacemaker experienced more extensive cardiac remodelling than patients with a previous ICD, with no differences in functional capacity (Supplementary Table S4). We found no differences in cardiac remodelling or functional capacity pending the duration of RV pacing (Supplementary Table S5).

Discussion

In this prospective cohort study, we found that the 6-month effect size following CRT implantation in terms of improved cardiac function and functional capacity was similar for patients with RV pacing and patients with intrinsic conduction at baseline. These results were robust in sensitivity analyses where only patients with LBBB and with LVEF ≤35% were included. We also found that patients with LBBB-like, RBBB and IVCD seem to benefit less compared with patients with true LBBB and RV pacing even though they had significant QRS prolongation ≥150 ms. These findings are concordant with those of Wokhlu et al.20 (ΔLVEF was 7.7 ± 10.7 for RV pacing, 7.9 ± 10.2 for LBBB, 3.1 ± 8.5 for RBBB and 2.7 ± 7.3% for IVCD) although the effect size in our current study is larger. This difference likely reflects different patient selection, improved HF medication and CRT procedural development over the past 15 years. This includes lead types and positioning, device optimization algorithms and AV and VV delay optimization, which has been widely adopted by operators despite lack of decisive evidence from large trials. For example, AV and VV delay optimization was used in 50–60% of cases according to the CRT Survey II.5 Rickard et al.12 also found similar echocardiographic response following CRT between RV pacing and LBBB at baseline (ΔLVEF 9.6 ± 10.6 vs. 11.6 ± 12.2%) and Khursid et al.11 found an effect size in LVEF improvement similar to ours for patients with pacing-induced HF (LVEF increased from 29.3% to 45.3%). In contrast Vamos et al.9 reported a smaller improvement in cardiac function with RV pacing compared with LBBB (ΔLVEF 2.9 ± 9.0 vs 6.7 ± 9.4%, ΔLVEDV 0.0 ± 12.2 vs −3.5 ± 6.7 mL). The authors pointed to several possible explanations for this difference, including upgrade procedures being performed too late. However, in our subgroup analysis we found no difference in cardiac remodelling or functional capacity in patients with shorter compared with longer duration between the first devices to CRT upgrade. The authors also invoked higher procedural risk for upgrade procedures and that the ventricular activation pattern induced by RV pacing is inherently different than intrinsic LBBB. Although RV pacing resembles LBBB on surface ECG, studies have demonstrated differences between intrinsic LBBB and RV paced QRS morphology, including wider QRS complex, longer transseptal activation time, later activation of the LV free wall and changes in the site of the latest electrical activation.21–24 The LV activation pattern may also differ depending on the RV lead position however, the RV lead position does not seem to influence the risk of developing pacing-induced HF25,26 but determining the exact lead position solely based on fluoroscopy is notoriously difficult.27 Although RV pacing induces variable LV activation patterns which are distinctive from intrinsic LBBB, we found similar improvements in both conventional echocardiographic parameters and functional capacity. The majority of patients with RV pacing at baseline were upgraded from a pre-existing pacemaker rather than an ICD and patients upgraded from a pacemaker showed more extensive remodelling than those upgraded from an ICD. This likely reflects that in pacemaker recipients, who typically have preserved LVEF at the time of implantation and in whom dyssynchrony induced by RV pacing is the key element eliciting HF, LV function can be expected to recover substantially. In contrast, in patients that had an ICD implanted in already established HF with reduced LVEF, and subsequently have developed conduction disorder, RV pacing can be an innocent bystander, and reverse remodelling can be absent or limited to the level seen prior to development of the conduction disorder. To investigate this question, however, requires serial evaluations of LV functions over time, which was not available, and thus constitutes a limitation to this study. We also considered endpoint measurements which have not previously been reported in this setting, with similar reductions in both LVMI, NT-proBNP and use of loop diuretics. Interestingly we found no reduction in LAVI from baseline to 6-month follow-up, consistent with a previous study that utilized cardiac computed tomography for the measurement of atrial volumes and found a very small reduction in LAVI and no association between LAVI reduction and CRT response.28

Patients with RV pacing were older and had a higher burden of comorbidities at the time of CRT implantation and there was a tendency towards slightly better echocardiographic response and slightly worse improvement in NYHA functional class, although not statistically significant. This likely reflects that the comorbidities in patients with RV pacing limit their functional capacity despite prominent improvement in cardiac function. This consideration is supported by the higher NYHA functional class and shorter distance on 6MWT at follow-up. Importantly, NYHA functional has poor correlation with cardiac function and long-term outcomes and it is unknown if the similar short-term effect of CRT translates to similar long-term survival and freedom from HF hospitalization, as previous studies have reported both better,10 similar29 and worse12,30 long-term survival in patients undergoing upgrade compared with de novo procedures. In terms of direct clinical implications, our findings suggest that patients with RV paced QRS who undergo CRT upgrade can be informed to expect a benefit comparable to those receiving de novo CRT. This is important given that upgrade procedures are often considered more complex procedures, with higher risk of procedural risks and lead-related complications.31–34 For example, the rate of any complication in the BUDAPEST-CRT trial upgrade group was 25/211 (12.3%). The magnitude of this added procedural risk is uncertain6 and may be mitigated by procedural strategies.35 Ultimately, the clinical decision to upgrade a patient from RV pacing to CRT must balance the anticipated physiological improvement against the additional procedural risk, a question that should ideally be tested in a parallel design RCT.

Lastly, recent years have seen the introduction and widespread clinical adaptation of conduction system pacing (CSP), that utilizes His-Bundle pacing (HBP) or Left bundle branch area pacing (LBBAP) to pace the innate conduction system thereby narrowing the QRS complex and restoring synchronous ventricular contraction. Comparison of RV pacing, CRT and CSP for various patient groups are warranted and ongoing (NCT05815745, NCT05650658). Likely, CRT as biventricular pacing will remain an important tool, also combined with CSP, should this pacing mode prove non-inferior or superior to CRT in selected patients.36 Importantly, our study demonstrates a prominent effect size of contemporary CRT on a wide range of endpoints in both patients with RV paced and intrinsic LBBB QRS morphology at baseline who were well-selected and well-treated with GDMT, and this serves as an important reference point for future comparisons with other pacing modalities.

Limitations

Although prospective, this study is subject to limitations pertaining to its observational design, including risk of confounding and inability to demonstrate causality. The results apply only to patients fulfilling the RCTs’ eligibility criteria and may not translate to other CRT populations. Minor differences in eligibility criteria between the RCTs, most notably in the definition of upgrade procedures, could result in some heterogeneity. Procedural risks and lead-related complications were not investigated, as the sample size was insufficient for robust analysis. Such data have been reported in the original publications of two of the three RCTs.13,14 Subgroup analysis on pre-existing device type and duration of RV pacing also comprised a limited number of patients and should be interpreted cautiously. Still, this prospective study comprises a large cohort of patients from the modern CRT era with high-quality data from RCTs on a wide range of outcomes and the effect size was prominent in both groups.

Conclusion

Patients with RV paced QRS morphology who are upgraded to a CRT device derive substantial improvement in cardiac function and functional capacity, comparable to patients with intrinsic conduction undergoing CRT.

Supplementary Material

xvag183_Supplementary_Data

Contributor Information

Henrik Laurits Bjerre, Department of Cardiology, Aarhus University Hospital, Palle Juul-Jensens Boulevard 99, 8200 Aarhus N, Denmark; Department of Clinical Medicine, Aarhus University, Palle Juul-Jensens Boulevard 82, 8200 Aarhus N, Denmark.

Emil Anton Frandsen, Department of Cardiology, Copenhagen University Hospital—Rigshospitalet, Blegdamsvej 9, 2100 København, Denmark.

Anders Sommer, Department of Cardiology, Aarhus University Hospital, Palle Juul-Jensens Boulevard 99, 8200 Aarhus N, Denmark.

Charlotte Stephansen, Department of Cardiology, Aalborg University Hospital, Hobrovej 18-22, 9000 Aalborg, Denmark.

Niels Risum, Department of Cardiology, Copenhagen University Hospital—Rigshospitalet, Blegdamsvej 9, 2100 København, Denmark.

Steen Hvitfeldt Poulsen, Department of Cardiology, Aarhus University Hospital, Palle Juul-Jensens Boulevard 99, 8200 Aarhus N, Denmark; Department of Clinical Medicine, Aarhus University, Palle Juul-Jensens Boulevard 82, 8200 Aarhus N, Denmark.

Jens Cosedis Nielsen, Department of Cardiology, Aarhus University Hospital, Palle Juul-Jensens Boulevard 99, 8200 Aarhus N, Denmark; Department of Clinical Medicine, Aarhus University, Palle Juul-Jensens Boulevard 82, 8200 Aarhus N, Denmark.

Mads Brix Kronborg, Department of Cardiology, Aarhus University Hospital, Palle Juul-Jensens Boulevard 99, 8200 Aarhus N, Denmark; Department of Clinical Medicine, Aarhus University, Palle Juul-Jensens Boulevard 82, 8200 Aarhus N, Denmark.

Supplementary data

Supplementary data are available at ESC Heart Failure online.

Declarations

Disclosure of Interest

J.C.N. is executive editor for Europace, no other conflicts of interest. M.B.K. received speakers fee from Biosense Webster, Abbott, Novartis, Bayer, Novo Nordisk. Remaining authors have no conflicts of interest.

Data Availability

No data were generated or analysed for this manuscript.

Funding

H.L.B. was supported by Danish Cardiovascular Academy (grant number PhD2021011-HF), which is funded by the Novo Nordisk Foundation; grant number NNF20SA0067242; and the Danish Heart Foundation. J.C.N. was supported by grants from the Novo Nordisk Foundation (grant numbers NNF16OC0018658, NNF17OC0029148).

Ethical Approval

Ethical Approval was not required.

Pre-registered Clinical Trial Number

None supplied.

References

  • 1. Wilkoff  BL, Cook  JR, Epstein  AE, Greene  HL, Hallstrom  AP, Hsia  H, et al.  Dual-chamber pacing or ventricular backup pacing in patients with an implantable defibrillator: the Dual Chamber and VVI Implantable Defibrillator (DAVID) Trial. JAMA  2002;288:3115–23. 10.1001/jama.288.24.3115 [DOI] [PubMed] [Google Scholar]
  • 2. Lamas  GA, Lee  KL, Sweeney  MO, Silverman  R, Leon  A, Yee  R, et al.  Ventricular pacing or dual-chamber pacing for sinus-node dysfunction. N Engl J Med  2002;346:1854–62. 10.1056/NEJMoa013040 [DOI] [PubMed] [Google Scholar]
  • 3. Sweeney  MO, Hellkamp  AS, Ellenbogen  KA, Greenspon  AJ, Freedman  RA, Lee  KL, et al.  Adverse effect of ventricular pacing on heart failure and atrial fibrillation among patients with normal baseline QRS duration in a clinical trial of pacemaker therapy for sinus node dysfunction. Circulation  2003;107:2932–7. 10.1161/01.CIR.0000072769.17295.B1 [DOI] [PubMed] [Google Scholar]
  • 4. Glikson  M, Nielsen  JC, Kronborg  MB, Michowitz  Y, Auricchio  A, Barbash  IM, et al.  2021 ESC guidelines on cardiac pacing and cardiac resynchronization therapy: developed by the task force on cardiac pacing and cardiac resynchronization therapy of the European Society of Cardiology (ESC) with the special contribution of the European Heart Rhythm Association (EHRA). Eur Heart J  2021;42:3427–520. 10.1093/eurheartj/ehab364 [DOI] [PubMed] [Google Scholar]
  • 5. Linde  CM, Normand  C, Bogale  N, Auricchio  A, Sterlinski  M, Marinskis  G, et al.  Upgrades from a previous device compared to de novo cardiac resynchronization therapy in the European Society of Cardiology CRT Survey II. Eur J Heart Fail  2018;20:1457–68. 10.1002/ejhf.1235 [DOI] [PubMed] [Google Scholar]
  • 6. Kaza  N, Htun  V, Miyazawa  A, Simader  F, Porter  B, Howard  JP, et al.  Upgrading right ventricular pacemakers to biventricular pacing or conduction system pacing: a systematic review and meta-analysis. Europace  2023;25:1077–86. 10.1093/europace/euac188 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7. Merkely  B, Hatala  R, Wranicz  JK, Duray  G, Földesi  C, Som  Z, et al.  Upgrade of right ventricular pacing to cardiac resynchronization therapy in heart failure: a randomized trial. Eur Heart J  2023;44:4259–69. 10.1093/eurheartj/ehad591 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8. Gage  RM, Burns  KV, Bank  AJ. Echocardiographic and clinical response to cardiac resynchronization therapy in heart failure patients with and without previous right ventricular pacing. Eur J Heart Fail  2014;16:1199–205. 10.1002/ejhf.143 [DOI] [PubMed] [Google Scholar]
  • 9. Vamos  M, Erath  JW, Bari  Z, Vagany  D, Linzbach  SP, Burmistrava  T, et al.  Effects of upgrade versus de novo cardiac resynchronization therapy on clinical response and long-term survival: results from a multicenter study. Circ Arrhythm Electrophysiol  2017;10:e004471. 10.1161/CIRCEP.116.004471 [DOI] [PubMed] [Google Scholar]
  • 10. Jędrzejczyk-Patej  E, Mazurek  M, Kotalczyk  A, Kowalska  W, Konieczny-Kozielska  A, Kozielski  J, et al.  Upgrade from implantable cardioverter-defibrillator vs. de novo implantation of cardiac resynchronization therapy: long-term outcomes. Europace  2021;23:113–22. 10.1093/europace/euaa339 [DOI] [PubMed] [Google Scholar]
  • 11. Khurshid  S, Obeng-Gyimah  E, Supple  GE, Schaller  R, Lin  D, Owens  AT, et al.  Reversal of pacing-induced cardiomyopathy following cardiac resynchronization therapy. JACC Clin Electrophysiol  2018;4:168–77. 10.1016/j.jacep.2017.10.002 [DOI] [PubMed] [Google Scholar]
  • 12. Rickard  J, Gorodeski  EZ, Patel  D, Hussein  A, Moennich  LA, Cantillon  D, et al.  Long term outcomes in patients with chronic right ventricular pacing upgraded to cardiac resynchronization therapy. J Cardiovasc Electrophysiol  2019;30:1979–83. 10.1111/jce.14038 [DOI] [PubMed] [Google Scholar]
  • 13. Sommer  A, Kronborg  MB, Nørgaard  BL, Poulsen  SH, Bouchelouche  K, Böttcher  M, et al.  Multimodality imaging-guided left ventricular lead placement in cardiac resynchronization therapy: a randomized controlled trial. Eur J Heart Fail  2016;18:1365–74. 10.1002/ejhf.530 [DOI] [PubMed] [Google Scholar]
  • 14. Stephansen  C, Sommer  A, Kronborg  MB, Jensen  JM, Nørgaard  BL, Gerdes  C, et al.  Electrically vs. Imaging-guided left ventricular lead placement in cardiac resynchronization therapy: a randomized controlled trial. Europace  2019;21:1369–77. 10.1093/europace/euz184 [DOI] [PubMed] [Google Scholar]
  • 15. Kronborg  MB, Frausing  M, Svendsen  JH, Johansen  JB, Riahi  S, Haarbo  J, et al.  Does targeted positioning of the left ventricular pacing lead towards the latest local electrical activation in cardiac resynchronization therapy reduce the incidence of death or hospitalization for heart failure?  Am Heart J  2023;263:112–22. 10.1016/j.ahj.2023.05.011 [DOI] [PubMed] [Google Scholar]
  • 16. Sommer  A, Kronborg  MB, Poulsen  SH, Böttcher  M, Nørgaard  BL, Bouchelouche  K, et al.  Empiric versus imaging guided left ventricular lead placement in cardiac resynchronization therapy (ImagingCRT): study protocol for a randomized controlled trial. Trials  2013;14:113. 10.1186/1745-6215-14-113 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17. Stephansen  C, Sommer  A, Kronborg  MB, Jensen  JM, Bouchelouche  K, Nielsen  JC. Electrically guided versus imaging-guided implant of the left ventricular lead in cardiac resynchronization therapy: a study protocol for a double-blinded randomized controlled clinical trial (ElectroCRT). Trials  2018;19:600. 10.1186/s13063-018-2930-y [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18. Rector  TS, Cohn  JN. Assessment of patient outcome with the Minnesota Living with Heart Failure questionnaire: reliability and validity during a randomized, double-blind, placebo-controlled trial of pimobendan. Pimobendan Multicenter Research Group. Am Heart J  1992;124:1017–25. 10.1016/0002-8703(92)90986-6 [DOI] [PubMed] [Google Scholar]
  • 19. Spertus  JA, Jones  PG. Development and validation of a short version of the Kansas City cardiomyopathy questionnaire. Circ Cardiovasc Qual Outcomes  2015;8:469–76. 10.1161/CIRCOUTCOMES.115.001958 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20. Wokhlu  A, Rea  RF, Asirvatham  SJ, Webster  T, Brooke  K, Hodge  DO, et al.  Upgrade and de novo cardiac resynchronization therapy: impact of paced or intrinsic QRS morphology on outcomes and survival. Heart Rhythm  2009;6:1439–47. 10.1016/j.hrthm.2009.07.009 [DOI] [PubMed] [Google Scholar]
  • 21. Varma  N. Left ventricular conduction delays induced by right ventricular apical pacing: effect of left ventricular dysfunction and bundle branch block. J Cardiovasc Electrophysiol  2008;19:114–22. 10.1111/j.1540-8167.2007.00995.x [DOI] [PubMed] [Google Scholar]
  • 22. Mafi Rad  M, Blaauw  Y, Dinh  T, Pison  L, Crijns  HJ, Prinzen  FW, et al.  Different regions of latest electrical activation during left bundle-branch block and right ventricular pacing in cardiac resynchronization therapy patients determined by coronary venous electro-anatomic mapping. Eur J Heart Fail  2014;16:1214–22. 10.1002/ejhf.178 [DOI] [PubMed] [Google Scholar]
  • 23. Varma  N. Left ventricular electrical activation during right ventricular pacing in heart failure patients with LBBB: visualization by electrocardiographic imaging and implications for cardiac resynchronization therapy. J Electrocardiol  2015;48:53–61. 10.1016/j.jelectrocard.2014.09.002 [DOI] [PubMed] [Google Scholar]
  • 24. Bjerre  HL, Nielsen  JC, Pedersen  ALD, Gerdes  C, Kristensen  J, Jensen  JM, et al.  Agreement on site of latest electrical activation between right ventricular pacing and intrinsic conduction as target for left ventricular lead position in cardiac resynchronization therapy. Heart Rhythm O2  2025;6:566–75. 10.1016/j.hroo.2025.02.006 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25. Kaye  GC, Linker  NJ, Marwick  TH, Pollock  L, Graham  L, Pouliot  E, et al.  Effect of right ventricular pacing lead site on left ventricular function in patients with high-grade atrioventricular block: results of the Protect-Pace study. Eur Heart J  2015;36:856–62. 10.1093/eurheartj/ehu304 [DOI] [PubMed] [Google Scholar]
  • 26. Galand  V, Martins  RP, Donal  E, Behar  N, Crocq  C, Soulié  GG, et al.  Septal versus apical pacing sites in permanent right ventricular pacing: the multicentre prospective SEPTAL-PM study. Arch Cardiovasc Dis  2022;115:288–94. 10.1016/j.acvd.2021.12.007 [DOI] [PubMed] [Google Scholar]
  • 27. Fyenbo  DB, Sommer  A, Stephansen  C, Nørgaard  BL, Kronborg  MB, Kristensen  J, et al.  Cardiac computed tomography-verified right ventricular lead position and outcomes in cardiac resynchronization therapy. J Interv Card Electrophysiol  2022;64:783–92. 10.1007/s10840-022-01193-1 [DOI] [PubMed] [Google Scholar]
  • 28. Hansen  PB, Sommer  A, Nørgaard  BL, Kronborg  MB, Nielsen  JC. Left atrial size and function as assessed by computed tomography in cardiac resynchronization therapy: association to echocardiographic and clinical outcome. Int J Cardiovasc Imaging  2017;33:917–25. 10.1007/s10554-017-1070-x [DOI] [PubMed] [Google Scholar]
  • 29. Brandão  M, Almeida  JG, Fonseca  P, Monteiro  J, Santos  E, Rosas  F, et al.  Long-term clinical outcomes after upgrade to resynchronization therapy: a propensity score-matched analysis. Heart Rhythm O2  2021;2:671–9. 10.1016/j.hroo.2021.06.009 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30. Trenson  S, Voros  G, Martens  P, Ingelaere  S, Betschart  P, Voigt  JU, et al.  Long-term outcome after upgrade to cardiac resynchronization therapy: a propensity score-matched analysis. Eur J Heart Fail  2024;26:511–20. 10.1002/ejhf.3073 [DOI] [PubMed] [Google Scholar]
  • 31. Kirkfeldt  RE, Johansen  JB, Nohr  EA, Jørgensen  OD, Nielsen  JC. Complications after cardiac implantable electronic device implantations: an analysis of a complete, nationwide cohort in Denmark. Eur Heart J  2014;35:1186–94. 10.1093/eurheartj/eht511 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32. Schwertner  WR, Behon  A, Merkel  ED, Tokodi  M, Kovács  A, Zima  E, et al.  Long-term survival following upgrade compared with de novo cardiac resynchronization therapy implantation: a single-centre, high-volume experience. Europace  2021;23:1310–8. 10.1093/europace/euab059 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33. Cheung  JW, Ip  JE, Markowitz  SM, Liu  CF, Thomas  G, Feldman  DN, et al.  Trends and outcomes of cardiac resynchronization therapy upgrade procedures: a comparative analysis using a United States National Database 2003–2013. Heart Rhythm  2017;14:1043–50. 10.1016/j.hrthm.2017.02.017 [DOI] [PubMed] [Google Scholar]
  • 34. Nemer  DM, Patel  DR, Madden  RA, Wilkoff  BL, Rickard  JW, Tarakji  KG, et al.  Comparative analysis of procedural outcomes and complications between de novo and upgraded cardiac resynchronization therapy. JACC Clin Electrophysiol  2021;7:62–72. 10.1016/j.jacep.2020.07.022 [DOI] [PubMed] [Google Scholar]
  • 35. Frausing  MHJP, Kronborg  MB, Johansen  JB, Nielsen  JC. Avoiding implant complications in cardiac implantable electronic devices: what works?  Europace  2020;23:163–73. 10.1093/europace/euaa221 [DOI] [PubMed] [Google Scholar]
  • 36. Glikson  M, Burri  H, Abdin  A, Cano  O, Curila  K, De Pooter  J, et al.  European Society of Cardiology (ESC) clinical consensus statement on indications for conduction system pacing, with special contribution of the European Heart Rhythm Association of the ESC and endorsed by the Asia Pacific Heart Rhythm Society, the Canadian Heart Rhythm Society, the Heart Rhythm Society, and the Latin American Heart Rhythm Society. Europace  2025;27:euaf050. 10.1093/europace/euaf050 [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

xvag183_Supplementary_Data

Data Availability Statement

No data were generated or analysed for this manuscript.