Skip to main content
International Journal of Emergency Medicine logoLink to International Journal of Emergency Medicine
. 2026 Jul 21;19:186. doi: 10.1186/s12245-026-01312-z

An exploratory feasibility study of a novel portable mainstream capnograph in a prehospital environment

Niccolò Pedrotti 1,2,✉, Anders Larsson 4, Malin Jonsson Fagerlund 5,6, David Smekal 2,3
PMCID: PMC13401322  PMID: 42481937

Abstract

Background

Monitoring end-tidal carbon dioxide (EtCO2) and respiratory rate (RR) is a non-invasive and reliable, continuous breathing assessment method in intubated and spontaneously breathing patients. The prehospital spreading of this technique in awake patients has been limited by mainstream capnographs´ dimensions and costs. Yet the potential prehospital applications of a wearable mainstream capnograph are many. This study aims to assess the ability to measure EtCO2 of the novel mainstream MARIE capnograph in normal ambulance transport conditions.

Methods

A single-centre, observational, exploratory feasibility study conducted in Region Uppsala between December 2024 and January 2025. First 20 healthy adult volunteers were transported approximately 10 min by ambulance, having EtCO2 and RR continuously measured by the mainstream device. Three measurements were performed (in the beginning, in the middle and at the end of the transport). A fourth value was noted on arrival in the Emergency Department (ED) with a reference capnograph for comparison with the third value by the mainstream device. The same measurement procedure was then repeated on 30 clinically stable adult patients transported by ambulance to the ED.

Results

A total of 20 volunteers and 29 patients were included in the final analysis. The comparison between the mainstream and reference capnograph showed that median EtCO2 values were 4.6 [4.1–5.2] vs. 4.9 [4.4–5.2] kPa respectively (p = 0.35). Bland Altman analysis showed a mean difference between the mainstream capnograph and reference device of 0.16 kPa (95% CI -0.07-0.41), with moderate correlation between measurements. The median values of breaths per minute were 17 [14–21] vs. 19 [15–22] respectively (p < 0.01).

Conclusion

This study suggests that the mainstream capnograph MARIE is able to measure credible EtCO2 values in standard ambulance transport conditions.

Clinical trial registration

ClinicalTrials.gov, TRN NCT06905613, Registration date 16th December 2024.

Supplementary Information

The online version contains supplementary material available at 10.1186/s12245-026-01312-z.

Keywords: Capnography, Mainstream, Monitoring, Ambulance, EMS, Prehospital

Introduction

Capnography is a non-invasive method to continuously monitor breathing by measuring end-tidal carbon dioxide (EtCO2) and creating a waveform. Often combined with respiratory rate (RR), capnography is today basic requirement during anaesthesia and intensive care to assess ventilation in patients requiring assisted or controlled ventilation under sedation or general anaesthesia [1–4]. In spontaneously breathing healthy subjects, EtCO2 provides reliable surrogate values for the partial CO2 tension in arterial blood [5, 6]. The same relationship is however not seen in patients with lung pathologies such as chronic obstructive pulmonary disease (COPD) or immature lungs [7, 8], in these groups correct EtCO2 waveform evaluation is crucial [9]. Several reviews have recommended the use of capnography in critically ill, mechanically ventilated patients, during airway management, and during cardiopulmonary resuscitation (CPR), because capnography provides valuable information about gas exchange and dead space in acute lung injury, confirms correct endotracheal intubation, and serves as a surrogate indicator of circulatory status [9–12]. The consensus for routine use of capnography during sedation, anaesthesia and CPR is also supported by many international guidelines [13–19]. Capnography in prehospital emergency care has mainly been used as a tool to guide and optimize CPR, for confirmation of correct position of the endotracheal tube and during transfer of intubated, critically ill patients [11, 12]. This technology has also been studied in the diagnosis of pulmonary embolism, but its adoption in the prehospital setting among spontaneously breathing patients has been limited by the lack of affordable, reliable instruments [20, 21]. Capnographs intended for spontaneously breathing patients traditionally use side-stream technology with a suction pump that draws a small flow of sample gas from the patient to the monitor where the gas analysis sensor is located. Side-stream technology implies the use of relatively long tubing system (susceptible to kinking and obstruction), a water trap and a suctioning pump. Mainstream capnographs have the gas analysis sensors close to the patient, producing an undistorted waveform even at high RR but they are often not perfectly designed to carry at the patients’ side in the prehospital setting. Mainstream technology is considered superior for confirmation of proper endotracheal intubation [22], since the gas analysis sensor is positioned directly within the path of gas flow, as opposed to the side-stream sensor located off the main path of flow in the airway circuit. This superiority has been seen also among postoperative non-intubated patients and in ICU patients with non-invasive ventilation (NIV) following extubation [23, 24]. There are many potential applications of a small, wearable, battery-driven, mainstream capnograph in a prehospital emergency care environment where access to injured patients is both challenging and crucial.

The primary objective of this study was to assess whether a novel, portable, mainstream capnograph called MARIE could measure and show credible EtCO2 values under standard ambulance transport conditions. Secondary objectives were to measure eventual differences in EtCO2 values between MARIE and side-stream reference devices and to assess the ability to measure RR of the mainstream capnograph.

Methods

Study design

This is a single-centre, observational, prospective, exploratory feasibility study. The “STARD” reporting guideline was used to guide the reporting of our study, where applicable [25].

Setting

A novel, mainstream capnograph was tested in ambulances in Region Uppsala, Sweden. This region has an area of 8209 km2, around 400.000 inhabitants and is served by one university hospital and one regional hospital. The Emergency Medical Service (EMS) system in this region has four ambulance stations and one single regional Emergency Medical Dispatch (EMD). Every ambulance is staffed by either two nurses or one nurse and an emergency medical technician. In our setting, as in most parts of Sweden, ambulance capnography is only used during CPR, while ventilation in spontaneously breathing patients is mostly assessed through clinical assessment and rarely through electrocardiogram-based RR recorders.

Capnography devices and measurements

The mainstream capnograph assessed in the study was called MARIE, third prototype version: P3 (Fig. 1). It has three main components: a disposable nasal cannula, a cabled mainstream infra-red sensor and a battery-driven monitor. EtCO2 measurements, a capnography waveform and RR values appear directly on the monitor screen without calibration delays. Extra oxygen can be delivered through the device but this function was not tested in this trial. The monitor´s dimensions are 70 × 43 × 20 mm and the total weight is 57.5 g (56 g for the monitor + 1.5 g for the sensor cable). The accuracy of the MARIE device regarding EtCO2 is stated by the producer to be ± 0.5 kPa (calculated assuming an EtCO2 value of 5 kPa) and ± 1 breaths per minute regarding RR. Measurements performed with this mainstream capnograph are labelled with the letter “M”.

Fig. 1.

Fig. 1

The mainstream capnograph, MARIE

Reference EtCO2 measurements were carried out with capnographs placed in the ED. These measures were performed in the ED’s ambulance garage as soon as the patient was leaving the ambulance in order to minimize the time between the last measure performed with the mainstream capnograph and the reference capnograph. The devices used were either Medtronic Capnostream 35 or Philips IntelliVue X3. Both devices measure EtCO2 through a disposable nasal cannula using side-stream technology, show a capnography waveform and RR values on their screens and have a weight of 1 kg vs. 1.4 kg respectively. Both devices have an accuracy of ± 0.26 kPa for EtCO2 (calculated assuming an EtCO2 value of 5 kPa) and ± 1 breaths per minute for RR. They share the same underlying technology, the X3 being an evolution of the Capnostream 35. Measurements obtained with these side-stream capnographs are denoted by the letter “S”.

To reach the primary objective of the study, EtCO2 values obtained with the mainstream capnograph should fall within the normal physiological range of 4.5–6.0 kPa. A secondary objective was that the difference in EtCO2 between the mainstream device (M3) and the side-stream reference device (S1) should remain within the predefined acceptance interval based on accuracy specifications. At 5 kPa, this acceptance interval was 0.76 kPa, requiring the difference between M3 and S1 to be less than 0.8 kPa. This threshold is consistent with the minimum accuracy requirement for EtCO2 specified in the technical standard for respiratory gas monitors (± 0.43 kPa + 8% of the measured gas level = ± 0.83 kPa at 5 kPa).

Participants and study protocol

In the first part of the study, after signing an informed consent form, twenty healthy adult volunteers were transported by ambulance between Uppsala ambulance station and Uppsala University Hospital ED. End-tidal CO2 and RR were measured continuously by the mainstream capnograph (M), and the values were noted at the beginning, in the middle and at the end of the transport at arrival to the ED (timing labels: M1, M2 and M3 respectively) on a paper CRF. Directly after arrival at the ED the nasal canula was shifted to another canula connected to a reference side-stream capnograph (S) and after about 20 s an EtCO2 value and a RR were noted (timing label: S1) in the study CRF (Fig. 2). For logistical reasons, and to minimize the time interval between M3 and S1, the same ambulance was used for all transports of healthy volunteers. All measurements in this group were performed on the same day, in this cohort, the healthy volunteers served as their own case-controls, being measured with two different devices. As no adverse events occurred during this first part of the study, we proceeded to the second part. Thirty, consecutive, adult non-critical patients transported to ED by ambulance in Region Uppsala were included. To minimize interference with the measurements, and due to patient safety concerns raised by the ethical authorities monitoring the study, patients were excluded if they needed supplemental oxygen, if their condition justified a transport with “lights and sirens” and in case of known pregnancy. After providing informed consent, patients were connected to the mainstream capnograph. EtCO2 and RR were recorded at the beginning and midpoint of transport (M1 and M2), and a final measurement (M3) was obtained with the mainstream capnograph just before the patient was removed from the ambulance. The reference measurement (S1) was then performed with the side-stream capnograph immediately afterward, in the ambulance garage directly connected to the ED. All measurements in the patient group were performed at the bedside by the EMS personnel simultaneously providing patient care; all EMS personnel were instructed to give full priority to patient care. Unexpected adverse events, capnograph malfunction, and unplanned deviations from the study protocol were also recorded in the case report form (CRF) and reported as part of the results.

Fig. 2.

Fig. 2

Study timeline

Analysis

To detect a clinically relevant deviation with sufficient statistical power, the required sample size was calculated to be at least 30 subjects. To minimize risk to patients, the study began with 20 healthy volunteers; combined with the subsequent patient cohort, this exceeded the total of 30 subjects required for adequate statistical power. Characteristics were described as number (percentage) for categorical variables and mean (± standard deviation, SD) or median [interquartile range, IQR] for continuous variables as appropriate. To assess differences between normally distributed data t-tests were used and Mann-Whitney tests were used for non-normally distributed data. Deming regression was used to assess correlation between the new mainstream capnograph and reference capnograph, while Bland-Altman plot was used to assess eventual systematic bias between the two methods. A pre-defined significance level of p < 0.05 was used in all cases. Data analysis was performed with “Excel” (Microsoft Excel for Microsoft 365, v2108) and “R” version 4.3.2 [26].

Results

Demographics

Between December 18th 2024 and January 25th 2025, 20 healthy subjects and 91 patients were considered eligible for inclusion, of these, all 20 healthy subjects and 29 patients were included in the final analysis of the study. Participants with missing data and/or missing signed informed consent were excluded from the study (Fig. 3). The most common reasons for the ambulance dispatch were chest pain, fainting, dizziness, trauma and decreased general condition (Table 1 of the supplemental material). Demographic and baseline data of study population were as presented in Table 1. Among the subjects with a comorbidity, 5 patients had asthma and 3 patients had COPD.

Fig. 3.

Fig. 3

Study flow diagram

Table 1.

Study population demographics. 49 subjects were included. Number of patients (%), median [Q1, Q3], EtCO2 and RR are presented as median values of M1, M2 and M3 measurements performed with MARIE capnograph

Whole population        n = 49 (100%) Healthy volunteers n = 20 (41%) Patients n = 29 (59%)
Female sex 22 (45%) 6 (30%) 16 (55%)
Age (years) 53 [37, 82] 41 [34, 46] 76 [49, 83]
Past medical history 12 (24%) 4 (20%) 8 (27%)
EtCO2 (kPa) 4.6 [4.1, 5.2] 4.8 [4.3, 4.9] 4.5 [3.9, 5.4]
Respiratory Rate (bpm) 17 [14–21] 17 [15, 22] 17 [14, 18]

Main results: end-tidal CO2 values

The comparison between measurement values at M3 with the mainstream capnograph and S1 measurements with the reference capnograph showed a mean EtCO2 value of 4.5 (± 1.0) kPa vs. 4.7 (± 0.7) kPa respectively, p = 0.35. The median time in minutes between M3 and S1 was 3 [2–6] in the whole study population, while it was 2 [1–2] vs. 5 [4–7] in heathy volunteers and in the patients group respectively. When aggregating M1, M2 and M3 the median EtCO2 value was 4.6 [4.1–5.2] kPa and the coefficient of variability was 22%.

Secondary analyses

Bland-Altman analysis showed a mean difference between the mainstream capnograph and reference device of 0.16 kPa (95% CI -0.07-0.41, Fig. 4) with moderate correlation (Pearson’s r = 0.55, p = 0.17) between measurements (Fig. 1 of the supplemental material). One mainstream capnograph showed a suspected drift in the measurements with increasing values over time. A post-hoc exploratory data analysis was repeated without data from the drifting experimental capnograph (a total of 5 cases), Bland-Altman analysis showed a mean difference between the mainstream capnograph and reference device of 0.3 kPa (95% CI 0.1–0.5, Fig. 5), with still a moderate correlation (Pearson’s r = 0.66, p < 0.01) between measurements (Fig. 2 of the supplemental material).

Fig. 4.

Fig. 4

Bland-Altman plot comparing mainstream and side-stream methods (all devices n = 49)

Fig. 5.

Fig. 5

Bland-Altman plot comparing mainstream and side-stream methods (post-hoc exploratory analysis without one mainstream device with suspected drift in measures n = 44)

A difference was noted in the comparison between median RR values between the values at M3 with the mainstream capnograph and the S1 measurement with reference capnograph, 17[14-21] vs. 19[15-22] breaths per minute respectively, p < 0.01. When aggregating M1, M2 and M3 values for RR with the mainstream capnograph the median value was 17[14-21] breaths per minute.

In all but one case, EtCO2 values were measured and presented during the whole transport. One mainstream device was in fact disconnected during two minutes from a patient due to motion sickness and vomiting. Neither adverse events, nor technical alarms or alarms for high/low EtCO2 or RR were noted.

Discussion

Median of all EtCO2 values measured with the novel mainstream MARIE capnograph was 4.6 [4.1–5.2] kPa and no statistically significant difference was seen when comparing EtCO2 values measured by the mainstream and the reference side-stream capnographs, 4.5 vs. 4.7 kPa respectively (p = 0.35). These results suggests that the novel mainstream capnograph is able to measure credible EtCO2 values in standard ambulance transport conditions. Furthermore, the measured values did not statistically differ from values measured with the reference side-stream capnographs showing promising preliminary agreement. Previous studies comparing side-stream and mainstream capnography devices have reported differences in measurements ranging from 0.4 to 0.49 kPa [8]. In our study, the observed difference was smaller. However, direct comparison with cited literature is difficult because we did not measure arterial partial pressure of CO₂, which limits our ability to assess the absolute accuracy of either device.

A difference of 0.16 kPa and a moderate correlation between the two methods (r = 0.55, p = 0.17) was seen. One of the mainstream devices demonstrated a suspected drift in the measurements (higher measured EtCO2 values over time were noticed by the device producers in an internal analysis). A post-hoc exploratory analysis was therefore repeated without the device with the suspected drift and a difference of 0.3 kPa was seen, but the moderate correlation between the methods remained (r = 0.66, p < 0.01). This result should be interpreted with caution since it derives from a post-hoc analysis and a presumed drift. Higher EtCO2 values measured by the side-stream device in our group represent a surprising finding, given that S1 RR values were simultaneously higher, suggesting a higher minute ventilation. Pekdemir et al. reported a similar finding, with higher EtCO2 values measured by the side-stream device in subjects both with and without pulmonary pathology; however, this result warrants further investigation in future studies [8]. This feasibility study was carried out with a prototype mainstream capnograph called MARIE P3 that might need calibration or additional software development. Further studies of a final prototype are warranted to secure that the presumed issue with drift in measurements has been eradicated. However, the EtCO2 difference between mainstream and side-stream capnographs was well below the pre-defined accuracy acceptance cut-off value of 0.8 kPa in both Bland-Altman analyses.

Our study showed a difference in measured RR between the mainstream capnograph and the side-stream reference device, 17 vs. 19 breaths per minute respectively, p < 0.01. Respiratory rate can fluctuate over short time intervals and it may as well be consciously altered by study subjects [27]. Our study design may have introduced confounding factors, such as variable timing of in-ambulance measurements (M1, M2, M3) and patient speech which likely influenced the observed RR values. However, the underlying cause of the observed lack of agreement in RR measurements between devices remains unresolved and warrants further investigation. Future studies are planned to evaluate the accuracy and reliability of the novel mainstream capnograph for RR measurement under more controlled conditions.

The mainstream capnograph showed no alarms and functioned during the whole study without the need for charging. Furthermore, both the device and the nasal cannula were easy to apply and no subject had to interrupt the study due to device discomfort. Considering these feasibility results during standard ambulance transport conditions, its small dimensions and weight, the mainstream capnograph showed potential for future prehospital implementation.

Limitations

To the best of our knowledge, no previous research has compared mainstream and side-stream capnographs in spontaneously breathing patients in the prehospital environment. Our results might potentially contribute to the development of a portable and reliable capnograph that could be used to better monitor breathing in the largest group of patients transported by ambulance: those who are awake and spontaneously breathing. Although conducted at a single centre, the study involved all ambulance stations within Region Uppsala and two receiving hospitals, enabling rapid recruitment of both healthy volunteers and patients. To ensure adherence to Good Clinical Practice (GCP) and to maintain data integrity, an independent external monitor was appointed to oversee study conduct, verify source documentation, and ensure compliance with ethical and regulatory standards.

However, this study is not without limitations. Of 91 eligible patients, 7 were screened but not included, without documented reason. Furthermore, three patients were initially enrolled, but their data were excluded from analysis due to concerns regarding the validity of their informed consent signatures (Fig. 3). This decision was made to ensure ethical compliance and protect study integrity. Additionally, this was a pragmatic feasibility study: subjects were permitted to speak during measurements, and the timing of measurements was determined by EMS personnel based on clinical workflow. While this design minimized disruption of patient care, it may have introduced a variability affecting the quality and consistency of the measurements.

Conclusion

Our study suggests that the portable mainstream capnograph MARIE can provide clinically credible end-tidal CO₂ measurements under standard ambulance transport conditions, provided that no measurement drift is present. Further studies are warranted to evaluate the final version of this device with respect to its ability to accurately measure both end-tidal CO₂ and respiratory rate even during delivery of oxygen.

Supplementary Information

Below is the link to the electronic supplementary material.

Supplementary Material 1 (124.4KB, jpeg)
Supplementary Material 2 (113.5KB, jpeg)
Supplementary Material 3 (13.7KB, docx)
Supplementary Material 4 (254KB, docx)

Acknowledgements

EMS in Uppsala County for making this study possible and for their tireless effort to provide high-quality care throughout the years. Research nurses in the department of Anesthesia and Intensive Care for their kind and professional help.

Abbrevations

EtCO2

End-tidal carbon dioxide

RR

Respiratory rate

COPD

Chronic obstructive pulmonary disease

CPR

Cardiopulmonary resuscitation

EMD

Emergency medical dispatch

EMS

Emergency medical service

ED

Emergency department

CRF

Case report form

NIV

Non-invasive ventilation

PaCO2

Arterial partial carbon dioxide pressure

GCP

Good Clinical Practice

Author contributions

NP and DS designed the study, searched and obtained ethical approval. NP and DS produced the necessary documentation to allow for external study monitoring. NP and DS recruited and conducted study measures on healthy volunteers and partially on patients. NP and DS conducted data analysis and wrote the manuscript. NP created all pictures and tables in either “Biorender” https:/app.biorender.com), “Drawio” (https:/www.drawio.com) or “R” [26]. AL conducted data analysis and was a major contributor in writing the manuscript. MJF contributed in writing the manuscript. All authors read and approved the final manuscript.

Funding

Open access funding provided by Uppsala University. This research was funded by Vinnova, Sweden´s Innovation Agency (Grant number 2023–04188). The capnographs used in this study and the disposable nasal cannulas were provided by Oxlantic Medical AB, Uppsala, Sweden. This is an academic study and Oxlantic AB was not involved in the conduction of the study or analyzing of results.

Data availability

The datasets generated, used and analysed during the current study are available from the corresponding author on reasonable request.

Declarations

Ethics approval and consent to participate

The study was approved by the “Swedish Medical Products Agency” and the “Swedish Ethical Review Authority” (Dnr 5.1-2024-76464). Written, signed informed consent was collected before inclusion in accordance with “The Helsinki Declaration of the World Medical Association” [28].

Consent for publication

Not applicable.

Competing interests

The authors declare no competing interests.

Footnotes

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.Owens B, Hall C. Application of End-Tidal CO2 Monitoring to ICU Management. Crit Care Nurs Q. 2024;47(2):157–62. [DOI] [PubMed] [Google Scholar]
  • 2.Riley CM. Continuous Capnography in Pediatric Intensive Care. Crit Care Nurs Clin North Am. 2017;29(2):251–8. [DOI] [PubMed] [Google Scholar]
  • 3.Song Q, Pan Y, Kanazawa T, Morimatsu H. Relationship of Intraoperative SpO2 and ETCO2 Values with Postoperative Hypoxemia in Elderly Patients after Non-Cardiac Surgery. Acta Med Okayama. 2023;77(5):537–43. [DOI] [PubMed] [Google Scholar]
  • 4.Mehta JH, Williams GW 2nd, Harvey BC, Grewal NK, George EE. The relationship between minute ventilation and end tidal CO2 in intubated and spontaneously breathing patients undergoing procedural sedation. PLoS ONE. 2017;12(6):e0180187. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Takano Y, Sakamoto O, Kiyofuji C, Ito K. A comparison of the end-tidal CO2 measured by portable capnometer and the arterial PCO2 in spontaneously breathing patients. Respir Med. 2003;97(5):476–81. [DOI] [PubMed] [Google Scholar]
  • 6.Cinar O, Acar YA, Arziman I, Kilic E, Eyi YE, Ocal R. Can mainstream end-tidal carbon dioxide measurement accurately predict the arterial carbon dioxide level of patients with acute dyspnea in ED. Am J Emerg Med. 2012;30(2):358–61. [DOI] [PubMed] [Google Scholar]
  • 7.Doğan N, Şener A, Günaydın GP, İçme F, Çelik GK, Kavaklı H, et al. The accuracy of mainstream end-tidal carbon dioxide levels to predict the severity of chronic obstructive pulmonary disease exacerbations presented to the ED. Am J Emerg Med. 2014;32(5):408–11. [DOI] [PubMed] [Google Scholar]
  • 8.Pekdemir M, Cinar O, Yilmaz S, Yaka E, Yuksel M. Disparity between mainstream and sidestream end-tidal carbon dioxide values and arterial carbon dioxide levels. Respir Care. 2013;58(7):1152–6. [DOI] [PubMed] [Google Scholar]
  • 9.Nassar BS, Schmidt GA. Capnography Dur Crit Illn Chest. 2016;149(2):576–85. [DOI] [PubMed] [Google Scholar]
  • 10.Gattinoni L, Carlesso E, Cressoni M. Assessing gas exchange in acute lung injury/acute respiratory distress syndrome: diagnostic techniques and prognostic relevance. Curr Opin Crit Care. 2011;17(1):18–23. [DOI] [PubMed] [Google Scholar]
  • 11.Whitaker DK, Benson JP. Capnography standards for outside the operating room. Curr Opin Anaesthesiol. 2016;29(4):485–92. [DOI] [PubMed] [Google Scholar]
  • 12.Heradstveit BE, Sunde K, Sunde GA, Wentzel-Larsen T, Heltne JK. Factors complicating interpretation of capnography during advanced life support in cardiac arrest–a clinical retrospective study in 575 patients. Resuscitation. 2012;83(7):813–8. [DOI] [PubMed] [Google Scholar]
  • 13.Standards for basic anesthetic monitoring: american society of anesthesiologists. 2020. Available from: https://www.asahq.org/standards-and-practice-parameters/standards-for-basic-anesthetic-monitoring.
  • 14.Checketts MR, Alladi R, Ferguson K, Gemmell L, Handy JM, Klein AA, et al. Recommendations for standards of monitoring during anaesthesia and recovery 2015: Association of Anaesthetists of Great Britain and Ireland. Anaesthesia. 2016;71(1):85–93. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Care, AaNZCoAaI. Medicine. Australasian Anaesthesia. 2015.
  • 16.Society UIC. Update of the UK Intensive Care Society. 2010.
  • 17.Merry AF, Cooper JB, Soyannwo O, Wilson IH, Eichhorn JH. International Standards for a Safe Practice of Anesthesia 2010. Can J Anaesth. 2010;57(11):1027–34. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Practice guidelines for moderate procedural sedation. - American Society of Anesthesiology. 2018.
  • 19.Highlights of the 2020 American Heart Association: Guidelines for CPR and ECC. 2020.
  • 20.Wahlen BM, Bey T, Wolke BB. Measurement of end-tidal carbon dioxide in spontaneously breathing patients in the pre-hospital setting. A prospective evaluation of 350 patients. Resuscitation. 2003;56(1):35–40. [DOI] [PubMed] [Google Scholar]
  • 21.Rumpf TH, Krizmaric M, Grmec S. Capnometry in suspected pulmonary embolism with positive D-dimer in the field. Crit Care. 2009;13(6):R196. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Dyett JF, Moser MS, Tobin AE. Prospective observational study of emergency airway management in the critical care environment of a tertiary hospital in Melbourne. Anaesth Intensive Care. 2015;43(5):577–86. [DOI] [PubMed] [Google Scholar]
  • 23.Kasuya Y, Akca O, Sessler DI, Ozaki M, Komatsu R. Accuracy of postoperative end-tidal Pco2 measurements with mainstream and sidestream capnography in non-obese patients and in obese patients with and without obstructive sleep apnea. Anesthesiology. 2009;111(3):609–15. [DOI] [PubMed] [Google Scholar]
  • 24.Sakuraya M, Douno E, Iwata W, Takaba A, Hadama K, Kawamura N, et al. Accuracy evaluation of mainstream and sidestream end-tidal carbon dioxide monitoring during noninvasive ventilation: a randomized crossover trial (MASCAT-NIV trial). J Intensive Care. 2022;10(1):17. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Bossuyt PM, Reitsma JB, Bruns DE, Gatsonis CA, Glasziou PP, Irwig L, et al. STARD 2015: an updated list of essential items for reporting diagnostic accuracy studies. BMJ. 2015;351:h5527. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.R Core Team. R: A language and environment for statistical computing [Internet]. Vienna, Austria: R Foundation for Statistical Computing; 2023. Available from: https://www.R-project.org/.
  • 27.Natarajan A, Su HW, Heneghan C, Blunt L, O’Connor C, Niehaus L. Measurement of respiratory rate using wearable devices and applications to COVID-19 detection. NPJ Digit Med. 2021;4(1):136. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.World Medical A. World Medical Association Declaration of Helsinki: Ethical Principles for Medical Research Involving Human Participants. JAMA. 2025;333(1):71–4. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supplementary Material 1 (124.4KB, jpeg)
Supplementary Material 2 (113.5KB, jpeg)
Supplementary Material 3 (13.7KB, docx)
Supplementary Material 4 (254KB, docx)

Data Availability Statement

The datasets generated, used and analysed during the current study are available from the corresponding author on reasonable request.


Articles from International Journal of Emergency Medicine are provided here courtesy of Springer-Verlag

RESOURCES