Abstract
Cardiovascular biomarkers are central to the diagnosis and management of a range of acute and chronic disease states, yet current approaches remain dependent on episodic venous blood sampling. Dermal interstitial fluid (ISF), a dynamic and accessible biofluid, offers a promising alternative for continuous and minimally invasive biomarker monitoring. In this Review, we explore the emerging convergence of ISF biosensing technologies with cardiovascular diagnostics. We discuss the physiological basis of ISF sampling and highlight engineering advances in microneedle arrays, hydrogel implants and electrokinetic extraction methods. We examine state-of-the-art biosensing platforms—ranging from enzymatic electrochemical sensors to aptamer- and antibody-based systems—and evaluate their applicability to key cardiovascular biomarkers, including troponin, natriuretic peptides, CRP and microRNAs. Although early studies demonstrate proof-of-concept for ISF-based detection of several cardiac biomarkers, clinical translation remains limited. While still in proof of concept stage, continuous ISF monitoring could disrupt traditional models of care by enabling earlier detection, personalised risk stratification and remote disease management. However, it also presents challenges related to clinical workflow integration, data interpretation and regulatory oversight. Bridging these gaps will require multidisciplinary innovation across materials science, device engineering and clinical validation. By transitioning from episodic blood sampling to continuous dermal monitoring, this technology promises to redefine cardiovascular care; shifting the paradigm from reactive acute management to proactive, personalised, and predictive precision medicine.
Keywords: biomarkers, digital algorithms, precision medicine, sensing, wearables
1. Introduction
Cardiovascular disease (CVD) is the leading cause of mortality worldwide with an estimated 19.8 million deaths in 2022 alone (1). Reducing CVD mortality has been a key objective of healthcare organisations globally, with the World Health Organisation (WHO) targeting early disease detection as an area for improvement (2). To address this global burden, clinical focus has shifted towards improved diagnostic precision. Early detection enables the prevention and optimal management of CVD to reduce morbidity and mortality. Despite the utility of venous blood sampling, current diagnostic workflows face significant limitations, particularly in the ‘indeterminate’ patient group. In the context of acute coronary syndromes, a significant proportion (15%–40%) of patients present with troponin levels that are neither clearly normal nor diagnostic of infarction, necessitating prolonged hospital observation and repeat blood draws (3, 4). This diagnostic uncertainty places a substantial economic and logistical burden on healthcare systems (5, 6). Continuous monitoring of biomarker trends in dermal ISF is hypothesised to offer a potential solution to this ‘clinical grey zone,’ providing high-resolution temporal data that could differentiate acute pathological release from chronic background elevation earlier than episodic venous sampling.
Among the key strategies for early detection are techniques based on biomarker monitoring. Biomarkers are measurable indicators of biological processes, pathological states, or responses to interventions (7) and provide objective and quantifiable information about cardiovascular health. They can be molecules such as proteins, metabolites, or nucleic acids found in blood, interstitial fluid, saliva, or other biological fluids. In the context of health monitoring and diagnostics, biomarkers provide critical insights into the state and dynamics of a person's health. There are several well-established cardiac biomarkers that enable clinicians to detect and monitor disease. They are used in the acute (troponin, BNP), sub-acute (troponin, BNP, CRP, glucose,) and chronic (BNP, lipids, HbA1c, iron studies) setting. Monitoring of these biomarkers play a central role in international clinical guidelines (8). Biomarkers are typically detected from venous blood samples, requiring the associated healthcare resources (9, 10). However, biomarkers are detectable in a wide range of biofluids including dermal interstitial fluid (ISF). Comparative to venous sampling, skin is readily accessible, and sampling is minimally invasive for patients.
The use of continuous glucose monitoring via ISF in the management of diabetes has been an exemplar that has demonstrated validity with gold standard measurements and improved patient outcomes in clinical practice (11). Recent studies show that understanding the biochemical relationship between serum and dermal ISF is critical for advancing minimally invasive diagnostics and demonstrated that establish dermal ISF as a promising matrix for clinical application (12). There is now opportunity to apply these advantages to CVD detection and management.
To date, there are no clinical studies evaluating the use of continuous subcutaneous monitoring of cardiac biomarkers in human subjects. This novel sampling route represents a new area of research for CVD. There are many potential advantages over traditional blood-based biomarker measurement continuous sampling and detection will be explored in this review. Dermal ISF cardiac biomarker measurement offers the opportunity to improve diagnostic accuracy, widen diagnostic windows, elicit novel biological insights and generate larger data sets with greater temporal resolution for machine learning algorithms; though these benefits will need to be clinically validated.
In this review we interrogate the current state of research for sampling and detecting cardiac biomarkers in dermal ISF with the aim of exploring the next steps towards the development of continuous wearable sensors for clinical applications.
We review current sensor technology and the most viable sampling and measurement methods. Specific cardiac biomarker characteristics are summarised with a focus on the potential application of current sampling and monitoring technologies. Finally, we will review how continuous sensors might revolutionise cardiac care in the future.
1.1. Composition and role of dermal interstitial fluid
Interstitial fluid (ISF) is present in all extracellular environments in the human body. It has traditionally been characterised as a fluid reserve for blood volume functioning passively. However, there is an increasing body of evidence that indicates it is in fact a dynamic biofluid involved in the regulation of multisystem processes.
ISF is the largest volume fluid compartment and the most accessible biofluid in the human body (13). For the average 70 kg man, 60% of the total body weight is comprised of water, equalling 42 L; out of the extracellular fluid volume, approx. 75% or 10.5L of the volume is present in the interstitial space, and approx. 25% of that water is in the plasma, equivalent to 3.5L (14). ISF is found in visceral organs, the central nervous system and the lymphatic system. It is most abundant and accessible in subcutaneous tissues. As summarised in Table 1, when compared to other non-invasive biofluids such as saliva or urine, dermal ISF may offer the optimal balance of continuous wearable compatibility and biomarker concentration fidelity. Notably, while saliva and urine successfully harbour cardiovascular biomarkers like NT-proBNP and Troponin, their concentrations are significantly lower or heavily dependent on filtration rates, making them less suitable for the real-time, high-resolution monitoring required in acute cardiovascular care.
Table 1.
A comparison of human biofluid sampling characteristics for key cardiovascular biomarkers.
| Biofluid | CV biomarkers detected | Relative concentration (vs. venous blood) | Continuous wearable compatibility | Translational readiness |
|---|---|---|---|---|
| Venous Blood | Troponin, NPs, CRP, cMyBP-C, Lactate | Baseline reference concentration | Low (Requires invasive venepuncture; episodic) | Established Clinical Gold Standard |
| Dermal ISF (15–21) | Troponin I, Lactate, Lipids, microRNAs | Size-dependent: Small molecules (e.g., lactate) ≈ 100% of blood; Proteins (e.g., Troponin, CRP) are significantly attenuated. Larger proteins subject to lag times. | High (Microneedle arrays, transdermal optical) | Emerging (Human pilots for Lactate and Troponin) |
| Saliva (18, 22–27) | NT-proBNP, CRP, Troponin, microRNAs | Significantly lower (e.g., >1000-fold lower for NT-proBNP) | Moderate (Intraoral sensors emerging, but prone to fouling) | Research/Episodic screening |
| Urine (18, 28–30) | NT-proBNP, Troponin, microRNAs | Variable (Highly dependent on renal filtration rates) | Low (Not physically suitable for continuous real-time sensing) | Validated for episodic community screening |
| Sweat (20, 31) | Lactate, Sodium (20, 31) | Variable (Poor correlation with blood for large proteins) | High (Epidermal patches) | Validated for metabolites; unproven for CV proteins |
The dermis and hypodermis constitute the thickest layers of skin and sit beneath the avascular epidermis, as illustrated in Figure 1. The vascular and lymphatic systems sit within this layer interacting through ISF that permeates the gel like extracellular matrix (ECM) (see magnified boxes in Figure 1 (32). The ECM is made up of a 3D mesh of specialised cells, collagens, laminins and proteoglycans which can regulate ISF contents (33). ISF functions as a dynamic fluid medium through which nutrients, waste, immune mediators and hormones are transported through the interstitial ECM system. ISF content is dependent on whole body physiology and is replenished every ∼24–48hrs (32, 34). Importantly, ISF contains no coagulation factors. This makes in-situ continuous sampling and detection attractive – a key advantage over blood sampling.
Figure 1.
A schematic diagram of the blood-interstitium interface. Vascular and lymphatic systems interact in the dermis via interstitial fluid which runs through the mesh network of the extracellular matrix. The interface is regulated by physiological flow and permeability of the vascular endothelium. The schematic reflects the dermal layers with magnifications of the dermis detailing the gelatinous nature of the extracellular matrix (upper) and the capillary endothelium interface regulating dermal interstitial fluid and blood transport. Created in https://BioRender.com.
The volume of ISF follows the Starling mechanism and is determined by the balance of transcapillary hydrostatic and osmotic pressure, and the permeability of capillary cells and lymphatic drainage (34). The transport of molecules into the ISF is determined by both diffusion and fluid flow (13, 35). Small to medium molecules up to ∼70 kDa readily transfer through paracellular routes. Thus, small molecule and ion concentrations are comparable in ISF and intravascular fluid. Several comparative metabolite and protein studies have demonstrated minimum constituent differences between ISF and vascular serum, albeit at lower relative concentrations (36–39). Therefore, many established and emerging cardiovascular biomarkers of small molecular size are likely to be present in ISF and are reviewed in detail later.
Larger molecules >100 kDa are dependent on transcellular processes but remain subject to gradients in concentration and fluid pressure. It is important to note that permeability of the ISF-blood barrier is not fixed and can be affected by inflammation, paracrine mediators and mechanical factors. Transport of these molecules into the ISF is governed by size-dependent paracellular or transcellular mechanisms, establishing ISF as a dynamic reservoir with a composition comparable to blood. The likely mechanisms of transport of key cardiovascular biomarkers is summarised in the conceptual model diagram (see Figure 2).
Figure 2.
A schematic diagram summarising cardiac biomarker transport to the interstitium. The diagram shows the conceptual model of transport of proteins across the vascular endothelium to dermal interstitial fluid which is largely dependent on biomarker size characteristics. Transport is either paracellular or transcellular, with some biomarkers using both pathways. Small metabolites such as lactate (0.09 kDa) are able to diffuse freely across the membrane; medium sized molecules such as troponin (18-24 kDa) and C-reactive protein (CRP)(115 kDa) are able to transport across the membrane through paracellular or transcellular transporters driven by concentration gradients. Larger molecules such as Myosin-Binding Protein C (MyBP-C)(140 kDa) require more complex membrane transporters. proNT-BNP, N-terminal pro-B-type natriuretic peptide; kDA, kilodalton. Created in https://BioRender.com.
2. Methods for sampling dermal interstitial fluid
2.1. ISF extraction methods
There are several methods of ISF extraction described in the literature and available commercially (see Figure 3). Advantages of dermal ISF sampling include minimal invasiveness, simple use, continuous sampling and coupling to microchips and sensors. Key challenges to extraction are low volume acquisition, target protein concentration, extracellular matrix effects and tissue disruption. ISF extraction has been reviewed in detail by others (40), and a relevant summary for this review is outlined below:
Figure 3.
A schematic diagram of dermal interstitial fluid sampling methods. The diagram summarises 3 common methods of dermal interstitial fluid sampling. Hollow microneedle array sampling (top), filament/wick sampling (middle), reverse iontophoresis (bottom). ISF, interstitial fluid. Created in https://BioRender.com.
2.2. Microneedles
This method is a minimally invasive technique that utilises arrays of micron-scale projections, typically ranging from 150 to 1500 µm in length with tip diameters <20 µm, to penetrate the stratum corneum and access the dermal ISF without stimulating deep nociceptors (see Figure 4) (12, 38, 41–43). Microneedles for ISF (44) should fulfil certain well specified features when applied against the skin, such as penetration depth, insertion and fracture force, mainly related to their material mechanical properties, while remaining biocompatible; they are typically made of metal, silicon or polymers, forming body interfaces of adhesive patches that enable continuous sampling of dermal ISF. Polymers are claimed to provide better solutions for safer and longer-term biomarker monitoring and, in some cases, they come with transient self-dissolving properties. The application is typically painless provided certain conditions for the diameter of the individual needle are respected (45). It is worth noting that there are multiple types of microneedle architectures and strategies to access the ISF: (i) hollow microneedles (15), essentially aiming to extract small volumes of ISF for analysis with sensors operating outside the body, (ii) porous microneedles utilizing capillary-driven channels to extract ISF (46), and, (iii) solid-state microneedles with embedded sensory layers, enabling measurements in ISF directly under the skin, removing the need of ISF extraction (47). Despite many recent advances, microneedle patch technology is mostly used for research purposes. Further development of manufacturing processes and biosensor coupling is needed to reach clinical and commercial applications.
Figure 4.
Photomicrographs and photographs showing technological applications of hollow microneedles for interstitial fluid extraction. (A) Self-extracting dextran-based hydrogel microneedle arrays microneedles. Reproduced from “Digital images for (A)” by Bastien Darmau, Marta Sacchi, Isabelle Texier and Andrew J. Gross, licensed under CC BY 4.0. (B) Ascilion's silicon microneedles. Adapted from “Photos of manufactured sharp hollow microneedles in monocrystalline silicon. (a) Scanning electron microscope picture of an individual microneedle” by Markus Renlund, Laurenz Kopp Fernandes, Pelle Rangsten, Mikael Hillmering, Sara Mosel, Ziad Issa, Volkmar Falk, Alexander Meyer and Felix Schoenrath, licensed under CC BY. (C) Xsensio's magnetic wristband cassette holder. Reproduced from “Xsensio's dermal interstitial fluid (dISF) collector. (a) Base component of the dISF collector, serving as a support structure with a wristband on both sides to apply pressure and embedded magnets to attach to the collection component” by Yann Sprunger, Johan Longo, Ali Saeidi, and Adrian M. Ionescu, licensed under CC BY. (D) Xsensio's metal microneedle cassette. Reproduced from “Xsensio's dermal interstitial fluid (dISF) collector. (b) Collection component featuring BDTM 31G microneedles connected to TygonTM tubes for dISF extraction“ by Yann Sprunger, Johan Longo, Ali Saeidi, and Adrian M. Ionescu, licensed under CC BY.
2.3. Electroosmotic/reverse iontophoresis
Reverse iontophoresis is a non-invasive collection method that draws ISF to a collection system by applying a driving current across two nodes (48). This method has been utilised in past commercially available continuous glucose monitors (49). The method has also been utilised to collect troponin and microRNA samples (16). A key limitation of this technique is the exclusion of larger biomarkers and potential interference from sweat, and the effects of, skin temperature and hydration status.
2.4. Filament/wick/needle
This group of methods work by collecting ISF through a hollow tube that sits within the dermis. Depending on the method, ISF is drawn through a tube by capillary action, skin suction or external pressure. This method allows for in-situ detection and has been shown to work in humans for several days without need for renewal. This method is currently used in a commercially available continuous glucose sensor (50–52).
2.5. Emerging extraction techniques
There are several other methods of ISF sampling described in the literature that require further development to be adopted into wearable or clinically used devices. Suction blister extraction, for example, uses negative pressure and heat to separate the epidermis from the dermis, creating a blister filled with ISF. While effective for obtaining larger volumes for proteomic analysis (36, 37), it is generally too invasive and time-consuming for continuous wearable applications. Other emerging techniques include microdialysis and hollow micro-capillary loops, which allow for continuous perfusion and sampling but currently require external pump systems that may limit ambulatory use (53–55).
3. Biorecognition-layer-driven approaches to interstitial fluid sensing
Detection of biomarkers in the dermal ISF requires modification of current technologies to meet the requirements of repeated in-situ testing. ISF offers the advantages of being an easily accessible and non-coagulable biofluid; this eliminates most pre-processing required for laboratory blood tests. However, detection methods must overcome low concentration levels, extracellular matrix effects and repeat detection stability.
Both electrical and optical modalities have demonstrated significant potential for the non-invasive monitoring of biomarkers in dermal interstitial fluid (ISF) within wearable continuous monitoring platforms. Optical approaches, benefiting from high intrinsic sensitivity, can in principle achieve lower detection limits. However, their integration into miniaturized, low-power wearable form factors remains challenging due to constraints in optical component miniaturization, power consumption, and signal stability under dynamic, real-world conditions. Consequently, optical methods are currently more suited to point-of-care (PoC) applications, where compactness and continuous operation are less critical.
In contrast, electrical detection technologies have matured to a stage where they can be readily implemented in scalable, low-power wearable systems taking advantage of silicon CMOS technologies. These methods rely on monitoring electrical signals—either direct current (DC) or alternating current (AC) responses—in passive elements or active devices such as field-effect transistors with integrated functional layers. In all cases, specificity arises from a functional biorecognition layer that transduces the binding of the target analyte into a quantifiable electrical signal proportional to its concentration. Electrical ISF sensors utilise a diverse range of transduction mechanisms to convert biorecognition events into quantifiable signals. While early wearable platforms predominantly utilised amperometric enzymatic detection, recent advances have expanded the sensor landscape significantly. Current architectures include, but are not limited to: (a) enzyme-based sensors, which exploit catalytic turnover for signal amplification; (b) ion-selective electrodes (ISEs), which respond to changes in ionic activity; (c) affinity-based sensors utilizing (c) aptamers or (d) antibodies for detection. Other technologies such as field-effect transistors (FETs), which offer label-free, high-sensitivity detection by modulating channel conductance upon analyte binding have been described (56). Furthermore, emerging platforms are increasingly integrating impedance-based and CRISPR-mediated sensing modalities to enhance specificity for complex nucleic acid targets in ISF (57).
3.1. Enzyme linked sensors
Most of the CE-marked/FDA approved or commercially available wearable ISF monitors use enzyme linked sensors. These sensors use immobilised enzymes that catalyse reactions with the target analyte. The enzyme can be immobilised to a molecular ‘wire’ or redox polymer which acts to facilitate electron transfer from the enzyme active site to the electrode generating a signal proportional to analyte concentration (50, 58). This method of detection is the predominant technology found in commercially available dermal sensors that have been clinically validated (glucose) or are under clinical evaluation (lactate) (58–66). This method relies on analyte-enzyme specificity which is not available for larger molecules such as troponin or BNP. This may be overcome by antibody-enzyme conjugation, similar in principle to widely used enzyme linked immunosorbent assay (ELISA) based tests. However, these types of sensors have been mainly demonstrated for biomarkers having high concentration levels (metabolites).
3.2. Ion-sensitive electrodes (ISEs) and membranes (ISMs)
ISEs can measure pH and specific ion concentrations in ISF. They work using an ion selective membrane that generates a potential difference detected by an electrode (67). This sensor type is capable of detecting physiologically important ion concentrations continuously (22, 31, 68, 69). ISE use is limited to ion concentrations in the dermal ISF, which is relevant in heart failure patients. Peripheral oedema is a relatively late clinical sign of heart failure. It is caused by an expansion and change in ionic composition of ISF (70). ISE based sensors may therefore offer a more sensitive and early detection of haemodynamic changes driving these factors. It is also possible to couple thermistors and monometers that can measure dermal temperatures and pressures; these may offer further insight into heart failure related oedema mechanisms. The ISMs with specific ionophores are particularly suited for multi-modal detection of electrolytes in almost any types of biofluids, including ISF, being particularly scalable in terms of size and low volumes to be analysed. For instance, the multiplexed detection of Na+, K+, Ca2+ and pH has been demonstrated by Zhao et al. in a fully integrated CMOS chip with back-end-of-line integration of ISMs on metal electrodes achieving excellent sensitivity and selectivity together with ultra-low power consumption. Similar works combining metabolite and ion detection demonstrated the superiority of electrical methods combining ISE/ISM and enzymatic sensors. Translating this to a wearable format, Parrilla et al. developed a wearable, all-solid-state potentiometric microneedle patch specifically for intradermal potassium detection, demonstrating that ISEs can be successfully miniaturised for direct ISF electrolyte monitoring without fluid extraction (47).
3.3. Aptasensors
The aptamer-based sensor (Aptasensor) is an emerging technology that has several advantages for application in wearable monitors. Aptasensors utilise short oligonucleotide structures (aptamers) that have high specificity and affinity for a range of potential biomarkers including small molecules, ions, proteins and pathogens (71–74). These versatile aptamers are synthetically manufactured and can bind to any desired target. Analyte binding generates a conformation change in the aptamer that can be measured electrically, optically or by mass, compatible with existing detection methods. A typical example of electrical detection based on the functionalization of graphene electrodes with aptamers and detection with a subsequent field effect transistor has been reported for the detection of a small molecule like cortisol (75).
A key advantage of this technology is that it relies on the intrinsic aptamer-target binding properties that can enable label-free detection (76). This approach design avoids the need for additional tagging to enzymes, fluorescent dyes or nanoparticles for signal generation. In addition, nucleic acids are less immunogenic and have improved stability over antibodies.
These properties make aptasensors ideal for use in wearable sensors that require repeat analyte interactions over longer periods of time. However a key challenge is the potentially reduced analyte-aptamer interaction in the complex ECM which may interfere with binding. A solution to non-specific response is the use of a dual-sensor aptasensor architecture that subtracts the signal from a reference sensor operating on the same substrate and functionalized with a mixed aptamer, which allows for non-specific binding and drift corrections (75).
In the cardiovascular domain, aptamer technology has successfully been used to detect troponin, natriuretic peptides, myoglobin and CK (77–87). However these methods remain experimental and have mostly been applied to serum samples ex-vivo. The application and performance of these cardiovascular biomarker DNA aptamers is relatively untested in-vivo.
3.4. Antibody sensors
The antibody-based sensor (immunosensor) is a mature and widely used technology that leverages the high specificity and affinity of antibodies for their target antigens. Antibodies are Y-shaped proteins produced by the immune system that can bind a wide range of analytes, including proteins, peptides, small molecules, and pathogens, with remarkable selectivity. In biosensing applications, antibodies (or engineered antibody fragments, sometimes called nanobodies) are immobilized on a transducer surface, and analyte binding is transduced into an electrical, optical, or mass change signal, compatible with various detection platforms.
A major advantage of antibody-based sensors is their broad availability and well-established biofunctionalization chemistry, enabling their integration into a wide variety of substrates and device architectures. Antibody fragments, in particular, can enhance sensor performance by reducing the Nerst limit and improving binding kinetics. This makes immunosensors well-suited for point-of-care diagnostics and potentially for wearable continuous monitoring, provided that surface regeneration strategies are developed to allow repeated analyte binding cycles. A representative example of electrical immunosensing is the label-free detection of C-reactive protein (CRP) using a double-gate silicon nanowire field-effect transistor (SiNW FET) sensor array functionalized with antibody fragments (88). In this work, the authors exploited a pseudo-super-Nernstian back-gate operation mode to achieve highly sensitive detection of CRP over clinically relevant concentration ranges without the need for labels or secondary amplification steps.
However, compared to aptamers, antibodies are more susceptible to denaturation under varying temperature, pH, or mechanical stress, and they require careful storage conditions to maintain functionality. In complex biofluids such as interstitial fluid, non-specific adsorption can affect accuracy, therefore, approaches such as subtracted reference sensing channels or antifouling surface coatings are commonly implemented.
4. Current commercial biosensing platforms
While the continuous monitoring of cardiac biomarkers represents a novel frontier, the fundamental technology for harvesting and sensing dermal ISF is already a mature clinical reality in metabolic medicine. The widespread adoption of Continuous Glucose Monitoring (CGM) has validated the ‘lab-on-skin’ concept, proving that ISF biomarkers can be reliably tracked over weeks to months with high clinical accuracy (31, 59–63, 89, 90).
As detailed in Table 2, the current commercial landscape is dominated by amperometric enzymatic sensors. Devices such as the Abbott FreeStyle Libre, Dexcom G-series, and Medtronic Guardian utilise a transcutaneous filament functionalised with oxidases (e.g., glucose oxidase) to generate an electrical current proportional to analyte concentration. More recently, fully implantable platforms like the Senseonics Eversense and Profusa Lumee have introduced optical detection methods utilising fluorescence and phosphorescence quenching, respectively—to extend sensor longevity up to 180 days.
Table 2.
Commercially available dermal interstitial fluid devices with continuous monitoring.
| Biomarker | Device/platform | Sampling interface | Detection method | Development status | Key references |
|---|---|---|---|---|---|
| Glucose | Abbott FreeStyle Libre 2/3 | Subcutaneous Filament | Wired Enzyme Amperometry (Osmium mediator) | FDA/CE Approved | (50, 63) |
| Dexcom G6/G7 | Subcutaneous Wire | Enzymatic Amperometric (Glucose Oxidase) | FDA/CE Approved | (61, 90) | |
| Eversense E3 (Senseonics) | Subcutaneous Implant | Fluorescence (Boronic Acid Quenching) | FDA/CE Approved | (62, 89) | |
| Medtronic Guardian Connect | Subcutaneous Implanted electrode | Enzymatic Electrochemical | FDA/CE Approved | (60) | |
| Tissue Oxygen | Profusa Lumee Oxygen | Hydrogel Implant | Phosphorescence Quenching | CE Marked | (91–93) |
| Ketones | Sibionics KS3 | Transcutaneous enzymatic wire | Enzymatic amperometric (beta-hydroxybutyrate) | CE Marked (Consumer/Wellness) | (94, 95) |
Critically, the commercial expansion of these platforms from glucose to other analytes is already underway. Regulatory bodies have recently granted CE marks for continuous ketone monitoring (e.g., Sibionics KS3) (94, 95) and tissue oxygen sensing (91–93), demonstrating the versatility of ISF as a diagnostic matrix. These FDA-approved and CE-marked devices serve as the foundational hardware upon which future cardiac-specific sensors (discussed in Section 5) will likely be modelled.
5. Cardiovascular biomarkers of interest
As discussed earlier, the transition of analytes from the bloodstream into the dermal ISF is mediated by the vascular endothelium and the extracellular matrix (ECM). The molecular weight and structure of each cardiovascular biomarker dictate its specific transport mechanism and, consequently, its likelihood of detection. As illustrated in Figure 2, smaller molecules largely permeate via passive paracellular diffusion, whereas larger protein complexes may rely on transcellular pathways. Understanding these transport kinetics is critical for interpreting the concentration gradients and time-lags inherent to ISF-based sensing. Here we review the literature relevant to how cardiac biomarkers might be detected in ISF and how continuous monitoring impacts clinical care.
5.1. Troponin
5.1.1. Troponin structure and function
Troponin I (TnI) and Troponin T (TnT) are structural proteins of the cardiac sarcomere (24 kDa and 37 kDa, respectively) released into circulation following cardiomyocyte necrosis. Their relatively small molecular weight facilitates paracellular diffusion across the endothelial barrier, making them viable candidates for detection in the interstitial compartment (96, 97).
5.1.2. Troponin in clinical practice
The detection of myocardial specific troponin subunits isoforms in blood serum underpins accurate diagnosis of acute myocardial infarction (AMI) (8). Since the development of the first troponin specific immunoassays nearly 50 years ago (98), much progress has been made to improve immunoassay sensitivity down to picogram per millilitre concentrations (99).
The development of high sensitivity troponin assays has enabled the rapid diagnosis of AMI. A disadvantage of high sensitivity assays is the detection of troponin levels that are of indeterminate clinical significance. AMI diagnostic algorithms make use of serial troponin testing to address this issue (8). There is emerging work looking at the clinical significance of troponin concentration trends in cardiac patients to further stratify risk (100–103). Repeated invasive troponin spot sampling is difficult, resource intensive and undesirable to patients. If successfully translated to clinical practice, continuous monitoring of troponin may therefore offer novel insight into personalised patient risk in the AMI group as well as wider cardiac diseases.
5.1.3. Dermal interstitial fluid troponin
Troponin subunit molecular size ranges from 18 to 37 kDa, which is small enough to pass through the capillary membrane to ISF by passive paracellular route. Protein studies of ISF in tumour cell lines have identified many proteins of similar molecular size present in detectable quantities including tropomyosin α subunits (39). In a study looking at a novel electroosmotic method, TnI was detectable in the ISF of AMI rat models at pg/ml concentrations peaking 16 h after infarction (16). Interestingly the same study showed ISF concentrations of micro-RNAs peak much earlier which will be discussed later. A recent study has also shown detectable troponin in dermal ISF using a microneedle capacitive sensor. The platform utilised a 5 × 5 conical microneedle array functionalized with anti-cTnI antibodies to achieve label-free detection of cardiac troponin I directly in the interstitial compartment. The system demonstrated a dynamic range of 10 pg/mL to 10 ng/mL with a limit of detection (LOD) of 3.27 pg/mL and a rapid response time of under 15 min. Crucially, the study successfully validated the sensor's performance through spike-and-recovery tests in human serum and real-time monitoring in animal models, showing a high correlation with systemic blood levels (17). Troponin complex proteins are notably detectable in saliva, urine and amniotic fluid suggesting that the troponin subunits freely pass membrane barriers in the salivary glands, glomerulus and placenta (23, 28, 29). There is even evidence that troponin subunits can pass the highly regulated blood brain barrier (104). A US group has recently demonstrated the clinical feasibility of a wrist-worn transdermal optical sensor in identifying elevated troponin-I levels in patients with acute myocardial infarction, achieving an area under the curve (AUC) of 0.90 in validation cohorts (105). This proof-of-concept study supports the hypothesis that troponin permeates the dermal interstitium in detectable quantities during acute ischemic events.
To date, there are no dedicated studies to determine human concentrations of troponin subunits in dermal ISF by direct sampling. This remains an area for further study – of particular interest will be the trend of troponin over time in AMI patients and the temporal relationship with clinical events and symptoms. The physiology of ISF dictates that protein concentrations are likely to peak later than blood serum concentrations. However, this potential disadvantage may be overcome by high sensitivity detection methods already well developed for troponin. Additionally, continuous monitoring may offer novel trend insights over spot testing, which is discussed later.
5.2. B-type natriuretic peptide (BNP) and NT-proBNP
5.2.1. Structure and function of natriuretic peptides
B-type Natriuretic Peptide (BNP, 3.5 kDa) and its N-terminal prohormone (NT-proBNP, 8.5 kDa) are small peptide hormones secreted in response to ventricular stretch (106, 107). Their low molecular mass allows for rapid equilibration between the vascular and interstitial compartments, theoretically minimising the ‘lag time’ to detection in the ISF often observed with larger protein biomarkers (97–99). To evaluate the feasibility of continuous cardiovascular monitoring, biomarkers must be prioritised based on their transport kinetics and existing validation data. Table 3 summarises these characteristics, highlighting the relationship between molecular weight and the primary mechanism of translocation into the dermal ISF. While small molecules like lactate and peptides like BNP exhibit favorable diffusion profiles, larger structural proteins such as cMyBP-C present significant transcellular transport challenges that may necessitate more sensitive sensing modalities.
Table 3.
Translational readiness and transport characteristics of cardiovascular biomarkers in dermal ISF.
| Biomarker | Molecular weight (kDa) | Primary transport mechanism | Current ISF evidence level | Priority application |
|---|---|---|---|---|
| Lactate | 0.09 | Passive Diffusion | Validated (Human research/Pilot) (20, 64–66) | Acute Shock; Ischaemia monitoring |
| BNP | 3.5 | Paracellular Diffusion | Research (Human Urine/Saliva proxy) (24, 30, 36, 108, 109) | Heart Failure Monitoring/Treatment Titration |
| NT-proBNP | 8.5 | Paracellular Diffusion | Research (Human Urine/Saliva proxy) (24, 30, 36, 108, 109) | Heart Failure Monitoring/Treatment Titration |
| microRNAs | ∼7–8 | Paracellular (Vesicular/Protein-bound) | Research (Animal models, Human proxy fluids) (16, 18) | Novel Ischaemia Detection |
| Troponin I | 24 | Paracellular | Human (Transdermal optical detection, proxy fluids)/Animal models (16, 17, 28, 29, 39, 104, 105) | Acute coronary syndromes |
| Troponin T | 37 | Paracellular | Human (Transdermal optical detection, proxy fluids)/Animal models (16, 17, 28, 29, 39, 104, 105) | Acute coronary syndromes |
| CRP | 115 | Transcellular/Transcytosis | Research (Human proxy fluids) (25, 26, 110–118) | Plaque stability; Chronic inflammation |
| cMyBP-C | 140 | Transcellular/Transcytosis | Theoretical | Early AMI detection (<3 h) |
5.2.2. The use of natriuretic peptides in clinical practice
The release of NPs by ventricular myocytes is in response to myocardial stretch - triggering a neurohumoral response to maintain cardiac output. This process is central to the pathophysiology of heart failure. BNP/NT-proBNP are a reliable diagnostic marker of heart failure with >90% sensitivity, but decreasing specificity with age (119, 120). NPs are cited in both the European Society of Cardiology and American College of Cardiology/American Heart Association practice guidelines for diagnosis. Notably the European guidance utilise NPs for exclusion, whilst American guidelines advocate NPs as part of inclusion criteria – reflecting the differing interpretations of specificity (121, 122).
The use of NPs expand to monitoring of cardiac remodelling, disease progression, response to treatment, prognostication, risk stratification and mortality (123–128). Some groups advocate serial testing of NPs in HF, although this remains an area under review. Nonetheless, current North American guidelines support the use of NP testing at the start and end of hospital admission as a minimum (122). The evidence therefore suggests that the use of serial NP testing in heart failure enhances diagnosis and management; thus continuous monitoring may lead to further improvements in these areas.
5.2.3. Natriuretic peptides in biofluids
NPs are small peptides that are readily detectable in blood serum with several commercially available assays able to detect in the pg/ml range. NT-proBNP is a linear stable peptide with a longer half-life than BNP making it an ideal biomarker for detection in extravascular fluids (129). BNP and NT-proBNP are small peptides at 3.5 kDa and 8.5 kDa respectively meaning they can readily pass through the capillary membrane-dermal ISF barrier via the paracellular route. Their small size is likely to reduce the ISF matrix effects making detection in this fluid more viable. It is also important to note that extravascular volume (and therefore dermal interstitial fluid volume) is typically increased in heart failure pathophysiology leading to clinical oedema. This will increase the volume of ISF available for sample but the effect on peptide concentration is unknown.
To date there are no dedicated publications looking at the concentration or detection of NPs in dermal interstitial fluid specifically. However, numerous experiments characterising the proteomic contents of dermal ISF have found many proteins of similar molecular size and structure including proteins such as angiotensinogen relevant to heart failure (36). The absence of NPs from these studies, likely reflects the lack of cardiac disease in the population samples rather than absence of NPs in ISF specifically.
Nonetheless, there is good evidence that NPs are found in other extravascular fluid filtrates. Bellagambi and colleagues performed NT-proBNP analysis in saliva samples of acute heart failure patients (24). They showed a comparable relative decrease in saliva vs. blood NT-proBNP at discharge, reflecting successful treatment. Although the absolute concentration of NT-proBNP was 103 lower in saliva, the trend of change was maintained; suggesting continuous monitoring of extravascular fluid yields results reflective of blood serum concentration changes.
Numerous studies have shown that NPs can be reliably detected in urine and correlates temporally with blood serum concentration trends (30, 108, 130). Non invasive detection has been adopted to improve community screening highlighting the wider applicability of the non-invasive biomarker approach (109).
5.3. C-reactive protein (CRP)
5.3.1. Structure and function of CRP
C-reactive protein (CRP) is a highly conserved pentameric protein of the Pentraxin group of proteins. Human CRP is formed from 5 identical ∼23 kDa subunits forming a horseshoe structure of 115 kDa (131, 132). CRP is an acute phase protein produced predominantly by hepatocytes in response to interleukin activation and acute inflammation. Its function as an immune mediator and biomarker of inflammation is well characterised (110). The concentration of CRP is closely correlated to inflammation with acute rises seen in infection, malignancy and myocardial infarction.
5.3.2. The use of CRP in clinical practice
CRP is a marker and mediator of myocardial infarction through inflammatory effects within the atherosclerotic coronary plaque. Temporal variability and lower sensitivity/specificity limit its use (on its own) as a diagnostic marker of acute myocardial infarction (AMI) (126, 133–135).
However, the development of high sensitivity CRP assays has enabled further understanding of chronic inflammation states relevant to atherosclerosis and plaque development. Thus, there is increasing evidence that CRP interpretation can aid cardiovascular disease prognostication with CRP correlating to mortality, infarction size and risk of heart failure (136–139).
These observations have led to clinical trials testing CRP guided treatment strategies. The JUPITER trial treated patients with elevated hs-CRP but normal LDL using rosuvastatin – demonstrating reduced major cardiovascular events (140). The CANTOS trial demonstrated reduced cardiovascular events in patients treated with the monoclonal antibody targeting interleukin-1β; the trial recruited previous AMI patients with raised CRP - supporting the inflammatory plaque hypothesis (141). Colchicine had been shown to reduce events in AMI patients, but this was not replicated in other trials (142, 143). There remains a lack of consensus on how to use CRP in clinical management and further study is required. Nonetheless continuous CRP monitoring (perhaps in combination with other analytes) offers the potential to monitor real time clinical risk and therapeutic response in cardiac patients.
5.3.3. CRP in biofluids
Transport to ISF is dependent on both transcellular and paracellular diffusion making detection more difficult. However, the ISF membrane barrier becomes more permeable in inflammatory conditions which may facilitate higher ISF concentrations.
There is evidence that CRP is detectable in a range of biofluids supporting the hypothesis that CRP crosses the ISF blood barrier. CRP has been detected in pericardial fluid and correlates to myocardial infarction (111, 112). CRP is detectable in saliva and has been used as a non invasive test to screen for cardiovascular disease, periodontal disease and rheumatoid arthritis (25, 26). Several studies have looked at the detection of CRP in drain fluid - CRP is detectable in abdominal ISF and correlates to risk of anastomotic leaks after bowel resection (113). Chen and colleagues have shown CRP detectable in pleural ISF allowing the identification of empyema from parapneumonic effusions (144). There is further evidence that CRP is detectable in synovial fluid, amniotic fluid and highly regulated cerebrospinal fluid (114–118).
Although these studies show CRP is present in extravascular biofluids - there are no dedicated studies looking at the detection of CRP in dermal ISF. Furthermore, the detection of CRP in these fluids may reflect localised CRP mediated inflammation rather than systemic CRP concentration. Continuous ISF CRP detection may therefore have greater utility in non-cardiac conditions relating to infection, inflammation and malignancy.
6. Emerging biomarkers
There are many cardiovascular biomarkers in the research domain that are not in clinical use, but relevant to this review given their future potential. These biomarkers offer the potential of further diagnostic and pathophysiological understanding of cardiac disease; especially where there is uncertainty about optimal treatment such as – percutaneous coronary intervention (PCI) in non-ST elevation MI, PCI in chronic stable angina, management of chronic total occlusions, drug therapy for heart failure with preserved ejection fraction, asymptomatic severe aortic stenosis.
6.1. microRNAs
MicroRNAs (miRs) are short single stranded non-coding RNA molecules ∼22 nucleotides in length (145). They function as post-translational modulators of gene expression by partial or complete complementarity with target messenger RNAs (mRNAs) (146). There are over 2,500 human miRs catalogued with several hundred of interest as both biomarkers of cardiac disease and therapeutic targets (147). Study of miR expression in cardiac disease states have revealed novel mechanistic insights (148, 149). Of note miR-21 expression has been shown to mediate a number of cellular processes in ischaemic heart disease, cardiac hypertrophy and heart failure (150).
A linear miR of 22 nucleotides has an average molecular weight of ∼7–8 kDa. However, to prevent degradation miR are encapsulated within extracellular vesicles or bound to RNA-binding proteins in biofluids (145). This makes diffusion and transport to ISF dependant on cellular mechanisms. As discussed earlier, miR-499-5p is detectable in the dermal ISF, saliva and urine of rats (16). There is evidence that miR can be found in nearly all human biofluids including, amniotic/pleural/peritoneal/cerebrospinal fluid, tears and breast milk (18). Interestingly there is a largely equal distribution and quantity of miRs across these biofluids, that may reflect systemic concentrations of miR. Current detection methods rely on purification and amplification steps. A challenge of dermal ISF detection will be lower concentration and the lack of in-situ detection and quantification technology. Further elucidation of the relationship between cardio-specific miRs and disease is required as well study on dermal ISF miR quantities in cardiac patients.
6.2. cMyBP-C
Cardiac Myosin-Binding Protein C (cMyBP-C) is cardiac muscle protein that regulates interaction between myosin and actin to modulate muscle contraction force and velocity (151). It has a similar sensitivity and specificity for AMI diagnosis to troponin with better performance in early presenters <3 h of chest pain (152). cMyBP-C is a relatively large protein (140 kDa) with complex structure of 11 domains. Blood concentration after AMI ranges from 120 to 900 ng/L; intact dermal ISF concentrations are likely to be limited by size dependant transcellular diffusion. To date, there are no studies looking at non vascular biofluid detection of cMyBP-C and this is an area for future study.
6.3. Lactate
Lactate is small molecule (0.09 kDa) and well known biomarker used in several acute clinical settings. It has important metabolic and physiological functions (153). Within the dermal compartment, lactate enters the interstitial fluid through both systemic diffusion across capillaries making detection much less technically challenging. Human detection technologies have transitioned from invasive microdialysis to minimally invasive, wearable microneedle biosensors that provide continuous, real-time measurements (19, 20, 64–66). Clinical studies validate that these sensors accurately track venous lactate concentrations with high patient tolerance and a minimal lag time of approximately five to ten minutes (66). Elevated lactate levels in cardiovascular patients serve as powerful independent prognostic indicators for serious clinical outcomes, including in heart failure and mortality following myocardial infarction (154–157). The relative ease of lactate detection in the dermal ISF is reflected in the maturity of devices described in the literature. Although no device is yet approved for medical use, lactate dermal ISF detection is arguably the closest to clinical translation.
7. Clinical applications and potential future impact
The development of wearable continuous sensor technology coincides with a significant advance in the analysis and handling of large data sets through machine learning. These technologies are converging with the potential to produce a paradigm shift in clinical diagnostic approaches, with a change from spot testing to continuous monitoring of known analytes. There is growing evidence that traditionally well understood clinical tests, can provide further novel clinical insights when analysed in large data sets through machine learning algorithms (158–161). This approach may help understand key areas of uncertainty in cardiovascular disease.
7.1. Coronary care
Although much progress has been made in the diagnosis of coronary disease using troponin and imaging tests; there remains diagnostic uncertainty in several areas that continuous monitoring may address.
In acute coronary syndromes, troponin is a key clinical test. Current diagnostic troponin algorithms perform well when patients have presented an hour after symptoms with a troponin level above or below a set concentration (8). However, many patients present outside this time frame with troponin levels between cut of values. These patients do not meet diagnostic criteria but have 4–5x higher risk of death than patients with normal troponin level (162). Current practice advocates further observation, testing and expert clinical review (8). However, this can result in delay of appropriate diagnosis and treatment requiring significant health resources. Even for patients with troponin concentrations below diagnostic cut off values, there is a 0.5% risk of 30 day MACE (162). Given a total 1.39 million patients presented with chest pain to the National Health Service (NHS) England in 2023/24 (163), these groups represent a significant diagnostic challenge to health systems.
The hypothetical application of a wearable continuous troponin sensor offers several potential advantages in these indeterminate groups: a) repeat troponin testing is performed non-invasively and continuously providing real time trend based data as a surrogate of infarction, b) in-situ sensor analysis removes the need for venepuncture, lab analysis and associated healthcare resources [‘lab-on-chip approach’], c) wearable sensors will enable ambulatory management of patients with sufficiently low immediate risk.
Continuous monitoring in the established acute coronary syndrome group may also offer further opportunity. Despite several landmark studies, there remains uncertainty in several areas of management for NSTEMI: invasive vs. conservative, early invasive vs. late invasive, image guided, single vs. multivessel (164–171). So far, a blanket management approach for NSTEMI patients has failed to yield convincing results, suggesting this group requires more detailed stratification that may be facilitated by data from continuous monitoring.
7.2. Heart failure
A key advantage of the wearable sensor is the ‘lab on chip’ concept that enables monitoring outside of hospital environments, especially for chronic conditions such as heart failure. NPs are a well-established marker of disease and potentially more informative when repeated over time (172, 173). Thus heart failure management through the use of a BNP or NT-proBNP wearable continuous sensor has many potential applications.
In stable chronic heart failure patients’ serial NP testing results in better drug optimisation and reduction of heart failure related events at least (174–176). Several studies have shown that NP directed therapy can improve outcomes (172). All of these studies have used laboratory NP testing requiring venepuncture and healthcare settings. It is hypothesised that non-invasive continuous NP monitoring will offer superior data trends, which may subsequently inform better management decisions, although robust prospective trials are required to confirm this assumption.
As discussed already, North American heart failure guidance suggests at least 2 NP tests for acute heart failure patients admitted to hospital (122). NP monitoring enhances clinician decision making in the acute setting in combination with conventional measures (fluid balance, oedema status, electrolytes). This combined approach is already effective with spot NP testing and may be further improved with continuous NP data; although there is equivocal data regarding the effect of serial NP monitoring on all-cause mortality (174–176).
Acute heart failure management in the ambulatory setting is now well established and effective (177–180). Remote systems are increasingly being used to aid more effective management with success depending on meaningful monitoring targets (181, 182). Wearable sensors can easily be linked to wireless microchips enabling mobile monitoring systems and smart phone integration. These features will enable healthcare systems to manage more patients outside of hospital as the ‘internet of things [IoT] concept takes shape globally (183).
7.3. Data interpretation, digital algorithms, and AI integration
The transition from episodic spot testing to high-frequency interstitial monitoring necessitates a fundamental shift in clinical data interpretation. A primary challenge in this paradigm is the significant data volume of continuous biomarker monitoring that could overwhelm traditional clinical workflows. To translate these raw data streams into actionable insights, robust digital algorithms and machine learning (ML) frameworks are essential.
There is rapidly growing evidence showing that multimodal AI integration that combines physiological data such as heart rate variability, motion sensors and electrocardiogram (ECG) signals can significantly enhance diagnostic accuracy and patient outcomes in cardiovascular care (184–192). Such systems move beyond static, population-wide thresholds by utilising ML to establish personalised baselines, identifying subtle deviations from a patient's unique physiological norm. The use of AI models to manage and contextualise large continuous biomarker datasets will therefore be key to the practical implementation of wearable sensors in the clinical setting.
A key challenge will be the potential high false-positive rates inherent to sensitive monitoring systems. In the intensive care setting, up to 99% of alarms have been identified as non-actionable, contributing to significant clinician desensitisation and alarm fatigue (193). AI-based management interventions offer a solution to this burden; recent systematic reviews indicate that intelligent algorithms can suppress false alarms by up to 92.25% and reduce total caregiver notifications by over 99% without compromising patient safety (194, 195). By filtering movement artifacts and accounting for the physiological time-lag between blood and ISF compartments, these clinical decision support tools can triage high-risk patients for urgent review while effectively monitoring lower-risk individuals in ambulatory settings.
8. Challenges and limitations
At present, no cardiovascular specific analyte sensor has been developed and deployed for clinical use. Although several promising technologies have been reviewed here, they will need to overcome some remaining issues for clinical acceptance. Broadly these can be split into technical, ISF-specific and clinical challenges.
8.1. Technical challenges
There are several ISF sampling methods available and under investigation. Each method has its own specific advantage profile. However, all methods are dependent on local factors at the area of sampling (pressure, moisture, skin/soft tissue injury) and are therefore susceptible to sampling errors. All dermal ISF methods are only able to collect microlitre quantities, of which the constituent concentrations are in the nanogram to picogram per millilitre range. Some sampling methods may also disrupt localised mechanisms and structures causing biofouling errors. Nonetheless, several sampling methods have been shown to be robust for clinical and research use, especially for small molecules.
Detection methods remain a key area for development. Of the methods reviewed here, enzymatic sensors are the most widely adopted technology in clinical use but may be limited to small molecule detection given the lack of cardiac biomarker specific enzymes. Although this may be overcome using antibody linked enzymes. Enzymatic-electrochemical detection is a proven technology and can be miniaturised to wearable sensors. Biofouling, calibration requirements and enzyme denaturation may potentially limited frequent sampling rates required for wearable sensors.
ISE detection may have some limited utility in cardiac patients but ultimately lacks specificity for more complex biomarkers. The technology is being used in elite sports monitoring with some established wearables commercially, but medical use may be limited.
Aptamer based sensors offer promising specificity advantages for the future. Further development is required to understand if the underlying challenges of this technology can be overcome. Nuclease degradation, lower affinity binding in complex fluids and immobilisation are the key issues from experiments thus far. Further human studies of complex analyte detection are required to fully evaluate aptasensors.
8.2. Biological challenges
Dermal ISF is a large volume fluid dispersed across the complex ECM. ISF is infused over tissues delivered by capillary blood flow and collected through venous and lymphatic capillary systems. ISF fluid is under negative pressure in normal physiological conditions. These features pose several inherent challenges: a) Sample volumes and constituents are dependent on local tissue factors and may not reflect global physiological status, b) localised ISF sample volumes are small and are not replenished, c) the complex ECM may i) trap or interact with analytes of interest making detection more difficult, ii) interfere with sensor probes causing biofouling and detection errors, d) there is an inherent time lag of blood biomarker transfer to ISF which may lead to reduce diagnostic value.
Given the array of biosensing technologies emerging, the most significant translational hurdle is likely to be whether dermal ISF is currently a sufficiently validated biological substrate for acute cardiovascular monitoring. Reliable continuous ISF collection and robust biological validation remain major bottlenecks.
Crucially, while the blood-ISF correlation is well-established and mathematically predictable for small metabolites like glucose and lactate, robust human data demonstrating consistent correlations for larger cardiovascular proteins (such as troponin and CRP) remain severely limited. The complex ECM may trap or interact with these larger analytes, exacerbating the inherent pharmacokinetic time-lag between blood and ISF. Until large-scale, prospective clinical studies definitively map these concentration gradients and prove that ISF fluctuations reliably mirror venous blood during acute cardiac events, the clinical utility of ISF as a standalone diagnostic substrate for cardiovascular care remains unproven.
8.3. Clinical/regulatory challenges
While this review highlights promising advances in dermal interstitial fluid (ISF) biomarker detection, there remains a significant translational gap between technological potential and demonstrated clinical benefit. To date, few studies have directly evaluated whether continuous biomarker monitoring improves diagnostic accuracy or patient outcomes in cardiovascular care.
Several studies included in this review support the potential of continuous monitoring. For example, the detection of NT-proBNP trends in saliva and urine has shown correlation with blood levels and response to heart failure treatment, suggesting that continuous monitoring could enhance disease tracking and prognostication. Similarly, troponin monitoring in ISF has demonstrated delayed but detectable trends in animal models of myocardial infarction. These findings support the feasibility of non-invasive monitoring and the potential for capturing temporal biomarker profiles beyond the capability of spot blood sampling.
However, these technologies have yet to be validated in real-world clinical environments. The shift from traditional point-in-time testing to continuous, real-time biomarker surveillance will require several paradigm changes in healthcare delivery. First, clinical teams must develop new diagnostic frameworks that incorporate biomarker trends over time rather than static thresholds. Second, robust evidence must be generated to demonstrate that real-time monitoring provides actionable data that improves clinical decision-making and patient outcomes.
Importantly, continuous monitoring may prove disruptive to existing healthcare systems. The increased volume of data generated from wearable biosensors could overwhelm clinical workflows and require substantial investment in infrastructure, data analysis tools, and clinician training. Healthcare providers may face challenges in interpreting continuous streams of biomarker data, establishing clinical significance, and avoiding false alarms.
Ultimately, demonstrating clinical value will require prospective trials to evaluate diagnostic accuracy, cost-effectiveness, and impact on outcomes. Until then, the promise of continuous cardiovascular biomarker monitoring remains compelling—but unproven.
9. Conclusion
The detection of cardiovascular biomarkers in dermal interstitial fluid (ISF) represents a promising frontier in the early diagnosis and management of cardiovascular disease. The minimally invasive nature and continuous sampling capabilities of dermal ISF provide distinct advantages over traditional blood-based biomarker detection, potentially improving diagnostic accuracy, broadening diagnostic windows, and enabling novel biological insights. These benefits are particularly relevant for the monitoring of biomarkers such as troponin, BNP/NT-proBNP, and emerging indicators like microRNAs, which hold significant diagnostic and prognostic value in acute coronary syndromes and heart failure.
Advances in microneedle arrays, reverse iontophoresis, and filament-based extraction methods have demonstrated the feasibility of continuous ISF sampling. Coupled with promising detection technologies—including enzyme-linked sensors, ion-sensitive electrodes, and aptamer-based sensors—there is substantial potential to develop wearable devices capable of real-time cardiovascular monitoring. The small molecule sensors (glucose, lactate, sodium, potassium) are the technically most feasible and mature technologies ready for clinical translation; although they lack specificity for cardiovascular disease detection and monitoring. More complex biomarkers such as troponin are emerging as technically detectable in the ISF but require further development in human studies. These innovations could significantly enhance patient care by facilitating early diagnosis, personalised treatment strategies, and timely interventions, ultimately improving clinical outcomes.
9.1. Barriers to clinical adoption
However, technical and inherent challenges remain. The low analyte concentrations in ISF, matrix effects of the extracellular environment, and the time lag in biomarker diffusion from blood to ISF present obstacles to sensor sensitivity and specificity. Additionally, biofouling, calibration stability, and localised sampling inconsistencies must be addressed to ensure reliable and reproducible measurements. Overcoming these challenges will require continued multidisciplinary research integrating materials science, bioengineering, and clinical cardiology.
9.2. Future direction
Future research should prioritise the validation of ISF biomarker detection in clinical populations, the refinement of sensor technologies for robust and long-term performance, and the integration of these systems with data analytics and machine learning for predictive health insights. This will require clinical trials evaluating continuous monitoring against current practice and clinical outcomes. By addressing these gaps, dermal ISF biomarker detection could revolutionise cardiovascular disease management, transitioning from reactive to proactive, data-driven patient care.
In summary, while challenges remain, the convergence of advanced sampling techniques and innovative biosensing technologies positions dermal ISF as a transformative medium for continuous cardiovascular biomarker monitoring. Realising this potential could fundamentally shift the paradigm of cardiovascular diagnostics and patient management, paving the way for more precise and timely healthcare interventions.
Funding Statement
The author(s) declared that financial support was received for this work and/or its publication. This work forms part of the research portfolio of the National Institute for Health Research Barts Biomedical Research Centre (NIHR203330). YB acknowledges funding support from The Dowager Countess Eleanor Peel Trust.
Footnotes
Edited by: Juan Manuel Velázquez-Enríquez, Universidad del Istmo (Unistmo), Mexico
Reviewed by: Alessandro Mengozzi, University of Pisa, Italy
Xue Jiang, Spear Bio, United States
Author contributions
YB: Writing – original draft. AY: Writing – review & editing. DJ: Writing – review & editing. DL: Writing – review & editing. PV: Writing – review & editing, Methodology. AI: Writing – review & editing, Methodology. AM: Methodology, Visualization, Supervision, Funding acquisition, Conceptualization, Resources, Writing – original draft.
Conflict of interest
AM reports receipt of honoraria from Abbott; a funding award from Abbott for a clinical study on a topic related to the subject of this manuscript. The remaining author(s) declared that this work was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
The author DL declared that they were an editorial board member of Frontiers, at the time of submission. This had no impact on the peer review process and the final decision.
Generative AI statement
The author(s) declared that generative AI was not used in the creation of this manuscript.
Any alternative text (alt text) provided alongside figures in this article has been generated by Frontiers with the support of artificial intelligence and reasonable efforts have been made to ensure accuracy, including review by the authors wherever possible. If you identify any issues, please contact us.
Publisher's note
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.
References
- 1.Mensah GA, Fuster V, Murray CJL, Roth GA; Global Burden of Cardiovascular Diseases and Risks Collaborators. Global burden of cardiovascular diseases and risks, 1990-2022. J Am Coll Cardiol. (2023) 82(25):2350–473. 10.1016/J.JACC.2023.11.007 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.World Health Organisation (WHO). Cardiovascular Diseases Fact Sheet (2021). Available online at: https://www.who.int/news-room/fact-sheets/detail/cardiovascular-diseases-(cvds) (Accessed January 15, 2025)
- 3.Westwood M, Van Asselt T, Ramaekers B, Whiting P, Thokala P, Joore M, et al. High-sensitivity troponin assays for the early rule-out or diagnosis of acute myocardial infarction in people with acute chest pain: a systematic review and cost-effectiveness analysis. Health Technol Assess (Rockv). (2015) 19(44):1–234. 10.3310/HTA19440 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Anantha Narayanan M, Garcia S. Role of high-sensitivity cardiac troponin in acute coronary syndrome. US Cardiol Rev. (2019) 13(1):5–10. 10.15420/USC.2018.16.1 [DOI] [Google Scholar]
- 5.Goodacre S, Cross E, Arnold J, Angelini K, Capewell S, Nicholl J. The health care burden of acute chest pain. Heart. (2005) 91(2):229–30. 10.1136/HRT.2003.027599 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Bhatti Y, Stevenson A, Weerasuriya S, Khan S. Reducing avoidable chest pain admissions and implementing high-sensitivity troponin testing. BMJ Open Qual. (2019) 8(4):629. 10.1136/BMJOQ-2019-000629 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Califf RM. Biomarker definitions and their applications. Exp Biol Med. (2018) 243(3):213–21. 10.1177/1535370217750088/PDF [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Thygesen K, Alpert JS, Jaffe AS, Chaitman BR, Bax JJ, Morrow DA, et al. Fourth universal definition of myocardial infarction (2018). Eur Heart J. (2019) 40(3):237–69. 10.1093/EURHEARTJ/EHY462 [DOI] [PubMed] [Google Scholar]
- 9.Rifai N, Gillette MA, Carr SA. Protein biomarker discovery and validation: the long and uncertain path to clinical utility. Nat Biotechnol. (2006) 24(8):971–83. 10.1038/nbt1235 [DOI] [PubMed] [Google Scholar]
- 10.Anderson NL, Anderson NG. The human plasma proteome. Mol Cell Proteomics. (2002) 1(11):845–67. 10.1074/mcp.r200007-mcp200 [DOI] [PubMed] [Google Scholar]
- 11.Lind M, Polonsky W, Hirsch IB, Heise T, Bolinder J, Dahlqvist S, et al. Continuous glucose monitoring vs conventional therapy for glycemic control in adults with type 1 diabetes treated with multiple daily insulin injections: the GOLD randomized clinical trial. JAMA. (2017) 317(4):379–87. 10.1001/JAMA.2016.19976 [DOI] [PubMed] [Google Scholar]
- 12.Sprunger Y, Longo J, Saeidi A, Ionescu AM. Bridging blood and skin: biomarker profiling in dermal interstitial fluid (dISF) for minimally invasive diagnostics. Biosensors (Basel). (2025) 15(5):301. 10.3390/BIOS15050301/S1 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Aukland K, Nicolaysen G. Interstitial fluid volume: local regulatory mechanisms. Physiol Rev. (1981) 61(3):556–643. 10.1152/physrev.1981.61.3.556 [DOI] [PubMed] [Google Scholar]
- 14.Tobias A, Ballard BD, Mohiuddin SS. Physiology, Water Balance. Florida: StatPearls; (2022). Available online at: https://www.ncbi.nlm.nih.gov/books/NBK541059/ (Accessed August 5, 2025). [PubMed] [Google Scholar]
- 15.Renlund M, Kopp Fernandes L, Rangsten P, Hillmering M, Mosel S, Issa Z, et al. Use of a silicon microneedle chip-based device for the extraction and subsequent analysis of dermal interstitial fluid in heart failure patients. Diagnostics. (2025) 15(8):989. 10.3390/DIAGNOSTICS15080989 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Xiong H, Zhu C, Dai C, Ye X, Li Y, Li P, et al. An alternating current electroosmotic flow-based ultrasensitive electrochemiluminescence microfluidic system for ultrafast monitoring, detection of proteins/miRNAs in unprocessed samples. Adv Sci. (2024) 11(6):e2307840. 10.1002/advs.202307840 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Mirzajani H, Zolfaghari P, Koca BY, Urey H. Minimally invasive and in situ capacitive sensing of cardiac troponin I from interstitial fluid. ACS Sens. (2025) 10(12):9335. 10.1021/acssensors.5c01691 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Weber JA, Baxter DH, Zhang S, Huang DY, How Huang K, Jen Lee M, et al. The MicroRNA Spectrum in 12 body fluids. Clin Chem. (2010) 56(11):1733–41. 10.1373/clinchem.2010.147405 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Tehrani F, Teymourian H, Wuerstle B, Kavner J, Patel R, Furmidge A, et al. An integrated wearable microneedle array for the continuous monitoring of multiple biomarkers in interstitial fluid. Nat Biomed Eng. (2022) 6(11):1214–24. 10.1038/S41551-022-00887-1 [DOI] [PubMed] [Google Scholar]
- 20.Boscarino T, Djassemi O, Khan MI, Chaiya N, Moonla C, Choi B, et al. Monitoring blood lactate dynamics through sweat and interstitial fluid biofluids. Talanta. (2026) 298(Pt B):128985. 10.1016/j.talanta.2025.128985 [DOI] [PubMed] [Google Scholar]
- 21.Nilsson AK, Sjöbom U, Christenson K, Hellström A. Lipid profiling of suction blister fluid: comparison of lipids in interstitial fluid and plasma. Lipids Health Dis. (2019) 18(1):164. 10.1186/s12944-019-1107-3 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Liu J, Tang Y, Cheng Y, Huang W, Xiang L. Electrochemical biosensors based on saliva electrolytes for rapid detection and diagnosis. J Mater Chem B. (2023) 11(1):33–54. 10.1039/d2tb02031a [DOI] [PubMed] [Google Scholar]
- 23.Mirzaii-Dizgah I, Riahi E. Salivary troponin I as an indicator of myocardial infarction. Indian J Med Res. (2013) 138(6):861. Available online at: https://pmc.ncbi.nlm.nih.gov/articles/PMC3978973/ (Accessed October 18, 2024). [PMC free article] [PubMed] [Google Scholar]
- 24.Bellagambi FG, Petersen C, Salvo P, Ghimenti S, Franzini M, Biagini D, et al. Determination and stability of N-terminal pro-brain natriuretic peptide in saliva samples for monitoring heart failure. Sci Rep. (2021) 11(1):13088. 10.1038/s41598-021-92488-2 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Wetterö J, von Löhneysen S, Cobar F, Kristenson M, Garvin P, Sjöwall C. Pronounced diurnal pattern of salivary C-reactive protein (CRP) with modest associations to circulating CRP levels. Front Immunol. (2021) 11:607166. 10.3389/fimmu.2020.607166 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Bedi GN, Acharya S, Kumar S, Mapari SA. Salivary high-sensitivity C-reactive protein and its clinical relevance in modern medicine: a comprehensive review. Cureus. (2024) 16(4):e58165. 10.7759/cureus.58165 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Mishra V, Patil R, Khanna V, Tripathi A, Singh V, Pandey S, et al. Evaluation of salivary cardiac troponin-I as potential marker for detection of acute myocardial infarction. J Clin Diagn Res. (2018) 12(7):ZC44. 10.7860/JCDR/2018/32109.11791 [DOI] [Google Scholar]
- 28.Pervan P, Svaguša T, Prkačin I, Savuk A, Bakos M, Perkov S. Urine high sensitive troponin I measuring in patients with hypertension. Signa Vitae. (2017) 13(8):62–4. 10.22514/SV133.062017.13 [DOI] [Google Scholar]
- 29.Chen JY, Lee SY, Li YH, Lin CY, Shieh MD, Ciou DS. Urine high-sensitivity troponin I predict incident cardiovascular events in patients with diabetes Mellitus. J Clin Med. (2020) 9(12):3917. 10.3390/JCM9123917 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Cortés R, Portolés M, Salvador A, Bertomeu V, de Burgos FG, Martínez-Dolz L, et al. Diagnostic and prognostic value of urine NT-proBNP levels in heart failure patients. Eur J Heart Fail. (2006) 8(6):621–7. 10.1016/J.EJHEART.2005.11.011 [DOI] [PubMed] [Google Scholar]
- 31.Mazzaracchio V, Serani A, Fiore L, Moscone D, Arduini F. All-solid state ion-selective carbon black-modified printed electrode for sodium detection in sweat. Electrochim Acta. (2021) 394:139050. 10.1016/j.electacta.2021.139050 [DOI] [Google Scholar]
- 32.Aukland K, Reed RK. Interstitial-lymphatic mechanisms in the control of extracellular fluid volume. Physiol Rev. (1993) 73(1):1–78. 10.1152/physrev.1993.73.1.1 [DOI] [PubMed] [Google Scholar]
- 33.Griffith LG, Swartz MA. Capturing complex 3D tissue physiology in vitro. Nat Rev Mol Cell Biol. (2006) 7(3):211–24. 10.1038/nrm1858 [DOI] [PubMed] [Google Scholar]
- 34.Reed RK, Rubin K. Transcapillary exchange: role and importance of the interstitial fluid pressure and the extracellular matrix. Cardiovasc Res. (2010) 87(2):211–7. 10.1093/cvr/cvq143 [DOI] [PubMed] [Google Scholar]
- 35.Swartz MA, Fleury ME. Interstitial flow and its effects in soft tissues. Annu Rev Biomed Eng. (2007) 9:229–56. 10.1146/ANNUREV.BIOENG.9.060906.151850 [DOI] [PubMed] [Google Scholar]
- 36.Müller AC, Breitwieser FP, Fischer H, Schuster C, Brandt O, Colinge J, et al. A comparative proteomic study of human skin suction blister fluid from healthy individuals using immunodepletion and iTRAQ labeling. J Proteome Res. (2012) 11(7):3715–27. 10.1021/pr3002035 [DOI] [PubMed] [Google Scholar]
- 37.Niedzwiecki MM, Samant P, Walker DI, Tran V, Jones DP, Prausnitz MR, et al. Human suction blister fluid composition determined using high-resolution metabolomics. Anal Chem. (2018) 90(6):3786–92. 10.1021/acs.analchem.7b04073 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Kashaninejad N, Munaz A, Moghadas H, Yadav S, Umer M, Nguyen NT. Microneedle arrays for sampling and sensing skin interstitial fluid. Chemosensors. (2021) 9(4):83. 10.3390/chemosensors9040083 [DOI] [Google Scholar]
- 39.Celis JE, Gromov P, Cabezón T, Moreira JM, Ambartsumian N, Sandelin K, et al. Proteomic characterization of the interstitial fluid perfusing the breast tumor microenvironment: a novel resource for biomarker and therapeutic target discovery. Mol Cell Proteomics. (2004) 3(4):327–44. 10.1074/mcp.M400009-MCP200 [DOI] [PubMed] [Google Scholar]
- 40.Friedel M, Thompson IAP, Kasting G, Polsky R, Cunningham D, Soh HT, et al. Opportunities and challenges in the diagnostic utility of dermal interstitial fluid. Nat Biomed Eng. (2023) 7(12):1541–55. 10.1038/s41551-022-00998-9 [DOI] [PubMed] [Google Scholar]
- 41.Laszlo E, De Crescenzo G, Nieto-Argüello A, Banquy X, Brambilla D. Superswelling microneedle arrays for dermal interstitial fluid (prote)omics. Adv Funct Mater. (2021) 31(46):2106061. 10.1002/adfm.202106061 [DOI] [Google Scholar]
- 42.Razzaghi M, Seyfoori A, Pagan E, Askari E, Hassani Najafabadi A, Akbari M. 3D printed hydrogel microneedle arrays for interstitial fluid biomarker extraction and colorimetric detection. Polymers (Basel). (2023) 15(6):1389. 10.3390/polym15061389 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43.Darmau B, Sacchi M, Texier I, Gross AJ. Self-extracting dextran-based hydrogel microneedle arrays with an interpenetrating bioelectroenzymatic sensor for transdermal monitoring with matrix protection. Adv Healthc Mater. (2025) 14(2):e2403209. 10.1002/ADHM.202403209 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44.Kim G, Ahn H, Chaj Ulloa J, Gao W. Microneedle sensors for dermal interstitial fluid analysis. Med-X. (2024) 2(1):15. 10.1007/S44258-024-00028-0 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45.Prausnitz MR. Microneedles for transdermal drug delivery. Adv Drug Deliv Rev. (2004) 56(5):581–7. 10.1016/J.ADDR.2003.10.023 [DOI] [PubMed] [Google Scholar]
- 46.Gao G, Zhang L, Li Z, Ma S, Ma F. Porous microneedles for therapy and diagnosis: fabrication and challenges. ACS Biomater Sci Eng. (2023) 9(1):85–105. 10.1021/ACSBIOMATERIALS.2C01123 [DOI] [PubMed] [Google Scholar]
- 47.Parrilla M, Cuartero M, Padrell Sánchez S, Rajabi M, Roxhed N, Niklaus F, et al. Wearable all-solid-state potentiometric microneedle patch for intradermal potassium detection. Anal Chem. (2019) 91(2):1578–86. 10.1021/ACS.ANALCHEM.8B04877 [DOI] [PubMed] [Google Scholar]
- 48.Giri TK, Chakrabarty S, Ghosh B. Transdermal reverse iontophoresis: a novel technique for therapeutic drug monitoring. J Controlled Release. (2017) 246:30–8. 10.1016/j.jconrel.2016.12.007 [DOI] [PubMed] [Google Scholar]
- 49.Tierney MJ, Tamada JA, Potts RO, Jovanovic L, Garg S. Clinical evaluation of the GlucoWatch® biographer: a continual, non-invasive glucose monitor for patients with diabetes. Biosens Bioelectron. (2001) 16(9):621–9. 10.1016/S0956-5663(01)00189-0 [DOI] [PubMed] [Google Scholar]
- 50.Bailey T, Bode BW, Christiansen MP, Klaff LJ, Alva S. The performance and usability of a factory-calibrated flash glucose monitoring system. Diabetes Technol Ther. (2015) 17(11):787–94. 10.1089/dia.2014.0378 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 51.Home | FreeStyle Libre | Abbott. Available online at: https://www.freestyle.abbott/uk-en/home.html (Accessed August 6, 2025).
- 52.Alva S, Brazg R, Castorino K, Kipnes M, Liljenquist DR, Liu H. Accuracy of the third generation of a 14-day continuous glucose monitoring system. Diabetes Ther. (2023) 14(4):767. 10.1007/S13300-023-01385-6 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 53.Woodley D, Sauder D, Talley MJ, Silver M, Grotendorst G, Qwarnstrom E. Localization of basement membrane components after dermal-epidermal junction separation. J Invest Dermatol. (1983) 81(2):149–53. 10.1111/1523-1747.ep12543517 [DOI] [PubMed] [Google Scholar]
- 54.Chaurasia CS, Müller M, Bashaw ED, Benfeldt E, Bolinder J, Bullock R, et al. AAPS-FDA workshop white paper: microdialysis principles, application and regulatory perspectives. Pharm Res. (2007) 24(5):1014–25. 10.1007/s11095-006-9206-z [DOI] [PubMed] [Google Scholar]
- 55.Korf J, Huinink KD, Posthuma-Trumpie GA. Ultraslow microdialysis and microfiltration for in-line, on-line and off-line monitoring. Trends Biotechnol. (2010) 28(3):150–8. 10.1016/j.tibtech.2009.12.005 [DOI] [PubMed] [Google Scholar]
- 56.Guo L, Zhang F, Wang L, Zhou G, Zheng Y, Ren L, et al. Autonomous all-in-one patch for closed-loop diagnosis–therapy of melanoma. ACS Sens. (2025) 10(6):4506–21. 10.1021/ACSSENSORS.5C00791 [DOI] [PubMed] [Google Scholar]
- 57.Hao X, Qin C, He C, Yuan Q, Yang P, Wen Y, et al. In situ detection of bacteria from skin interstitial fluid via CRISPR microneedles: an amplification-free platform for point-of-care diagnostics. Biosens Bioelectron. (2026) 293:118123. 10.1016/J.BIOS.2025.118123 [DOI] [PubMed] [Google Scholar]
- 58.Govindaraj M, Srivastava A, Muthukumaran MK, Tsai PC, Lin YC, Raja BK, et al. Current advancements and prospects of enzymatic and non-enzymatic electrochemical glucose sensors. Int J Biol Macromol. (2023) 253:126680. 10.1016/j.ijbiomac.2023.126680 [DOI] [PubMed] [Google Scholar]
- 59.U.S. Food, Administration D. 510(k) Premarket Notification K233373: Lingo Glucose System (2024). Available online at: https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?ID=K233373 (Accessed January 7, 2026)
- 60.U.S. Food, Administration D. 510(k) Premarket Notification K222447: Guardian 4 Glucose Sensor (2023). Available online at: https://www.accessdata.fda.gov/cdrh_docs/reviews/K222447.pdf (Accessed January 7, 2026)
- 61.U.S. Food, Administration D. 510(k) Premarket Notification K221128: Dexcom G7 Continuous Glucose Monitoring System (2022). Available online at: https://www.accessdata.fda.gov/cdrh_docs/pdf22/K221128.pdf (Accessed January 7, 2026)
- 62.U.S. Food, Administration D. Premarket Approval (PMA) P160048/S016: Eversense E3 Continuous Glucose Monitoring System (2022). Available online at: https://www.accessdata.fda.gov/cdrh_docs/pdf16/P160048S016B.pdf (Accessed January 7, 2026)
- 63.U.S. Food, Administration D. 510(k) Premarket Notification K213996: FreeStyle Libre 3 Continuous Glucose Monitoring System (2022). Available online at: https://www.accessdata.fda.gov/cdrh_docs/reviews/K213996.pdf (Accessed January 7, 2026)
- 64.Bollella P, Sharma S, Cass AEG, Antiochia R. Microneedle-based biosensor for minimally-invasive lactate detection. Biosens Bioelectron. (2019) 123:152–9. 10.1016/j.bios.2018.08.010 [DOI] [PubMed] [Google Scholar]
- 65.Wang Q, Molinero-Fernandez Á, Wei Q, Xuan X, Konradsson-Geuken Å, Cuartero M, et al. Intradermal lactate monitoring based on a microneedle sensor patch for enhanced in vivo accuracy. ACS Sens. (2024) 9(6):3115–25. 10.1021/ACSSENSORS.4C00337 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 66.Ming DK, Jangam S, Gowers SA, Wilson R, Freeman DM, Boutelle MG, et al. Real-time continuous measurement of lactate through a minimally invasive microneedle patch: a phase I clinical study. BMJ Innov. (2022) 8(2):87. 10.1136/BMJINNOV-2021-000864 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 67.Wang Y, Ma S, Hu L, Fan Z, Lin Y. Wearable and printable devices for electrolytes sensing. Nano Futures. (2023) 7(3):032002. 10.1088/2399-1984/ACE40E [DOI] [Google Scholar]
- 68.Nyein HY, Gao W, Shahpar Z, Emaminejad S, Challa S, Chen K, et al. A wearable electrochemical platform for noninvasive simultaneous monitoring of Ca2+and pH. ACS Nano. (2016) 10(7):7216–24. 10.1021/acsnano.6b04005 [DOI] [PubMed] [Google Scholar]
- 69.Zhao F, Liu Y, Dong H, Feng S, Shi G, Lin L, et al. An electrochemophysiological microarray for real-time monitoring and quantification of multiple ions in the brain of a freely moving rat. Angew Chem. (2020) 132(26):10512–6. 10.1002/ange.202002417 [DOI] [PubMed] [Google Scholar]
- 70.Abassi Z, Khoury EE, Karram T, Aronson D. Edema formation in congestive heart failure and the underlying mechanisms. Front Cardiovasc Med. (2022) 9:933215. 10.3389/fcvm.2022.933215 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 71.Wang L, Wang R, Wei H, Li Y. Selection of aptamers against pathogenic bacteria and their diagnostics application. World J Microbiol Biotechnol. (2018) 34(10):149. 10.1007/S11274-018-2528-2 [DOI] [PubMed] [Google Scholar]
- 72.Woo HM, Kim KS, Lee JM, Shim HS, Cho SJ, Lee WK, et al. Single-stranded DNA aptamer that specifically binds to the influenza virus NS1 protein suppresses interferon antagonism. Antiviral Res. (2013) 100(2):337–45. 10.1016/J.ANTIVIRAL.2013.09.004 [DOI] [PubMed] [Google Scholar]
- 73.Wang ZJ, Chen EN, Yang G, Zhao XY, Qu F. Research advances of aptamers selection for small molecule targets. Chin J Anal Chem. (2020) 48(5):573–82. 10.1016/S1872-2040(20)60013-5 [DOI] [Google Scholar]
- 74.Xiao X, Li H, Zhao L, Zhang Y, Liu Z. Oligonucleotide aptamers: recent advances in their screening, molecular conformation and therapeutic applications. Biomed Pharmacother. (2021) 143:112232. 10.1016/J.BIOPHA.2021.112232 [DOI] [PubMed] [Google Scholar]
- 75.Gilani A, Saeidi A, Longo J, Sprunger Y, Kamaei S, Ameti A, et al. AC-Driven Aptamer-Decorated graphene FET for cortisol detection. In: Proceedings of IEEE Sensors (2024), 1–4. 10.1109/SENSORS60989.2024.10785129 [DOI] [Google Scholar]
- 76.Zhou J, Rossi J. Aptamers as targeted therapeutics: current potential and challenges. Nat Rev Drug Discov. (2017) 16(3):181–202. 10.1038/NRD.2016.199 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 77.Zhang J, Lv X, Feng W, Li X, Li K, Deng Y. Aptamer-based fluorometric lateral flow assay for creatine kinase MB. Microchim Acta. (2018) 185(8):364. 10.1007/s00604-018-2905-4 [DOI] [PubMed] [Google Scholar]
- 78.Li S, Zheng Y, Zou Q, Liao G, Liu X, Zou L, et al. Engineering and application of a myoglobin binding split aptamer. Anal Chem. (2020) 92(21):14576–81. 10.1021/acs.analchem.0c02869 [DOI] [PubMed] [Google Scholar]
- 79.Wang Q, Liu W, Xing Y, Yang X, Wang K, Jiang R, et al. Screening of DNA aptamers against myoglobin using a positive and negative selection units integrated microfluidic chip and its biosensing application. Anal Chem. (2014) 86(13):6572–9. 10.1021/ac501088q [DOI] [PubMed] [Google Scholar]
- 80.Zhao J, Zhu ZZ, Huang X, Hu X, Chen H. Magnetic gold nanocomposite and aptamer assisted triple recognition electrochemical immunoassay for determination of brain natriuretic peptide. Microchim Acta. (2020) 187(4):231. 10.1007/s00604-020-4221-z [DOI] [PubMed] [Google Scholar]
- 81.Lin MC, Nawarak J, Chen TY, Tsai HY, Hsieh JF, Sinchaikul S, et al. Rapid detection of natriuretic peptides by a microfluidic LabChip analyzer with DNA aptamers: application of natriuretic peptide detection. Biomicrofluidics. (2009) 3(3):034101. 10.1063/1.3194283 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 82.Grabowska I, Sharma N, Vasilescu A, Iancu M, Badea G, Boukherroub R, et al. Electrochemical aptamer-based biosensors for the detection of cardiac biomarkers. ACS Omega. (2018) 3(9):12010–8. 10.1021/acsomega.8b01558 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 83.Sharma A, Bhardwaj J, Jang J. Label-free, highly sensitive electrochemical aptasensors using polymer-modified reduced graphene oxide for cardiac biomarker detection. ACS Omega. (2020) 5(8):3924–31. 10.1021/acsomega.9b03368 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 84.Surya SG, Majhi SM, Agarwal DK, Lahcen AA, Yuvaraja S, Chappanda KN, et al. A label-free aptasensor FET based on au nanoparticle decorated Co3O4 nanorods and a SWCNT layer for detection of cardiac troponin T protein. J Mater Chem B. (2019) 8(1):18–26. 10.1039/c9tb01989h [DOI] [PubMed] [Google Scholar]
- 85.Agarwal DK, Kandpal M, Surya SG. Characterization and detection of cardiac troponin-T protein by using ‘aptamer’ mediated biofunctionalization of ZnO thin-film transistor. Appl Surf Sci. (2019) 466:874–81. 10.1016/j.apsusc.2018.10.086 [DOI] [Google Scholar]
- 86.Ara MN, Hyodo M, Ohga N, Akiyama K, Hida K, Hida Y, et al. Identification and expression of troponin T, a new marker on the surface of cultured tumor endothelial cells by aptamer ligand. Cancer Med. (2014) 3(4):825–34. 10.1002/cam4.260 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 87.Yang X, Zhao Y, Sun L, Qi H, Gao Q, Zhang C. Electrogenerated chemiluminescence biosensor array for the detection of multiple AMI biomarkers. Sens Actuators B Chem. (2018) 257:60–7. 10.1016/j.snb.2017.10.108 [DOI] [Google Scholar]
- 88.Capua L, Sprunger Y, Elettro H, Risch F, Grammoustianou A, Midahuen R, et al. Label-free C-reactive protein si nanowire FET sensor arrays with super-nernstian back-gate operation. IEEE Trans Electron Devices. (2022) 69(4):2159–65. 10.1109/TED.2022.3144108 [DOI] [Google Scholar]
- 89.Kropff J, Choudhary P, Neupane S, Barnard K, Bain SC, Kapitza C, et al. Accuracy and longevity of an implantable continuous glucose sensor in the PRECISE study: a 180-day, prospective, multicenter, pivotal trial. Diabetes Care. (2017) 40(1):63–8. 10.2337/DC16-1525 [DOI] [PubMed] [Google Scholar]
- 90.Wadwa RP, Laffel LM, Shah VN, Garg SK. Accuracy of a factory-calibrated, real-time continuous glucose monitoring system during 10 days of use in youth and adults with diabetes. Diabetes Technol Ther. (2018) 20(6):395–402. 10.1089/DIA.2018.0150 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 91.U.S. Food, Administration D. 510(k) Premarket Notification K193096: Lumee Oxygen Platform (2019). Available online at: https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?ID=K193096 (Accessed January 7, 2026)
- 92.Wisniewski NA, Nichols SP, Gamsey SJ, Pullins S, Au-Yeung KY, Klitzman B, et al. Tissue-integrating oxygen sensors: continuous tracking of tissue hypoxia. Adv Exp Med Biol. (2017) 977:377. 10.1007/978-3-319-55231-6_49 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 93.Montero-Baker MF, Au-Yeung KY, Wisniewski NA, Gamsey S, Morelli-Alvarez L, Mills Sr JL, et al. The first-in-man “Si se puede” study for the use of micro-oxygen sensors (MOXYs) to determine dynamic relative oxygen indices in the feet of patients with limb-threatening ischemia during endovascular therapy. J Vasc Surg. (2015) 61(6):1501–1510.e1. 10.1016/J.JVS.2014.12.060 [DOI] [PubMed] [Google Scholar]
- 94.Sabelli N, Haidar A, Majdpour D, Pasqua MR, Tsoukas M, Perz N, et al. 229-OR: the accuracy of a novel continuous ketone monitoring system (CKM) and metrics of ketosis during ketogenic diets in type 1 diabetes (T1D). Diabetes. (2025) 74(Supplement_1):229-OR. 10.2337/db25-229-OR [DOI] [Google Scholar]
- 95.SiBio KS3 Continuous Ketone Monitor - 14 Days Real-Time Tracking. Available online at: https://www.sibiosensor.com/products/sibio-ks3-ckm-continuous-ketone-monitoring-system?variant=46479712551126 (Accessed January 7, 2026).
- 96.Katrukha IA. Human cardiac troponin complex. Structure and functions. Biochemistry (Moscow). (2013) 78(13):1447–65. 10.1134/S0006297913130063/METRICS [DOI] [PubMed] [Google Scholar]
- 97.Marston S, Zamora JE. Troponin structure and function: a view of recent progress. J Muscle Res Cell Motil. (2020) 41(1):71–89. 10.1007/S10974-019-09513-1/FIGURES/15 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 98.Katus HA, Remppis A, Looser S, Hallermeier K, Scheffold T, Kübler W. Enzyme linked immuno assay of cardiac troponin T for the detection of acute myocardial infarction in patients. J Mol Cell Cardiol. (1989) 21(12):1349–53. 10.1016/0022-2828(89)90680-9 [DOI] [PubMed] [Google Scholar]
- 99.Apple FS, Collinson PO; Biomarkers for the ITF on CA of C. Analytical characteristics of high-sensitivity cardiac troponin assays. Clin Chem. (2012) 58(1):54–61. 10.1373/CLINCHEM.2011.165795 [DOI] [PubMed] [Google Scholar]
- 100.Wassie M, Lee MS, Sun BC, Wu YL, Baecker AS, Redberg RF, et al. Single vs serial measurements of cardiac troponin level in the evaluation of patients in the emergency department with suspected acute myocardial infarction. JAMA Netw Open. (2021) 4(2):e2037930. 10.1001/JAMANETWORKOPEN.2020.37930 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 101.Arslan M, Dedic A, Boersma E, Dubois EA. Serial high-sensitivity cardiac troponin T measurements to rule out acute myocardial infarction and a single high baseline measurement for swift rule-in: a systematic review and meta-analysis. Eur Heart J Acute Cardiovasc Care. (2020) 9(1):14–22. 10.1177/2048872618819421 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 102.Ohtake H, Terasawa T, Zhelev Z, Iwata M, Rogers M, Peters JL, et al. Serial high-sensitivity cardiac troponin testing for the diagnosis of myocardial infarction: a scoping review. BMJ Open. (2022) 12(11):e066429. 10.1136/bmjopen-2022-066429 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 103.Doudesis D, Lee KK, Boeddinghaus J, Ferry AV, Tuck C, Lowry MT, et al. Machine learning for diagnosis of myocardial infarction using cardiac troponin concentrations. Nat Med. (2023) 29(5):1201–10. 10.1038/s41591-023-02325-4 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 104.Zahid T, Eskander N, Emamy M, Ryad R, Jahan N. Cardiac troponin elevation and outcome in subarachnoid hemorrhage. Cureus. (2020) 12(8):e9792. 10.7759/CUREUS.9792 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 105.Sengupta S, Biswal S, Titus J, Burman A, Reddy K, Fulwani MC, et al. A novel breakthrough in wrist-worn transdermal troponin-I-sensor assessment for acute myocardial infarction. Eur Heart J Digit Health. (2023) 4(3):145–54. 10.1093/EHJDH/ZTAD015 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 106.Nakagawa Y, Nishikimi T, Kuwahara K. Atrial and brain natriuretic peptides: hormones secreted from the heart. Peptides (NY). (2019) 111:18–25. 10.1016/J.PEPTIDES.2018.05.012 [DOI] [PubMed] [Google Scholar]
- 107.Goetze JP. Biochemistry of pro-B-type natriuretic peptide-derived peptides: the endocrine heart revisited. Clin Chem. (2004) 50(9):1503–10. 10.1373/CLINCHEM.2004.034272 [DOI] [PubMed] [Google Scholar]
- 108.Michielsen ECHJ, Bakker JA, Van Kimmenade RRJ, Pinto YM, Van Dieijen-Visser MP. The diagnostic value of serum and urinary NT-proBNP for heart failure. Ann Clin Biochem. (2008) 45(4):389–94. 10.1258/acb.2007.007069 [DOI] [PubMed] [Google Scholar]
- 109.Ng LL, Loke IW, Davies JE, Geeranavar S, Khunti K, Stone MA, et al. Community screening for left ventricular systolic dysfunction using plasma and urinary natriuretic peptides. J Am Coll Cardiol. (2005) 45(7):1043–50. 10.1016/j.jacc.2004.12.058 [DOI] [PubMed] [Google Scholar]
- 110.Zhou HH, Tang YL, Xu TH, Cheng B. C-reactive protein: structure, function, regulation, and role in clinical diseases. Front Immunol. (2024) 15:1425168. 10.3389/FIMMU.2024.1425168/BIBTEX [DOI] [PMC free article] [PubMed] [Google Scholar]
- 111.Ege T, Us MH, Cikirikcioglu M, Arar C, Duran E. Analysis of C-reactive protein and biochemical parameters in pericardial fluid. Yonsei Med J. (2006) 47(3):372–6. 10.3349/ymj.2006.47.3.372 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 112.Schrag B, Iglesias K, Mangin P, Palmiere C. Procalcitonin and C-reactive protein in pericardial fluid for postmortem diagnosis of sepsis. Int J Legal Med. (2012) 126(4):567–72. 10.1007/s00414-012-0692-8 [DOI] [PubMed] [Google Scholar]
- 113.Bilgin IA, Hatipoglu E, Aghayeva A, Arikan AE, Incir S, Mamal Torun M, et al. Predicting value of serum procalcitonin, C-reactive protein, drain fluid culture, drain fluid interleukin-6, and tumor necrosis factor-α levels in anastomotic leakage after rectal resection. Surg Infect (Larchmt). (2017) 18(3):350–6. 10.1089/sur.2016.222 [DOI] [PubMed] [Google Scholar]
- 114.Marchocki Z, Vinturache A, Collins K, O’ Reilly P, O’Donoghue K. Amniotic fluid C-reactive protein as a predictor of infection in caesarean section: a feasibility study. Sci Rep. (2018) 8(1):6372. 10.1038/s41598-018-24569-8 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 115.Borna S, Abdollahi A, Mirzaei F. Predictive value of mid-trimester amniotic fluid high-sensitive C-reactive protein, ferritin, and lactate dehydrogenase for fetal growth restriction. Indian J Pathol Microbiol. (2009) 52(4):498. 10.4103/0377-4929.56136 [DOI] [PubMed] [Google Scholar]
- 116.Gray BM, Simmons DR, Mason H, Barnum S, Volanakis JE. Quantitative levels of C-reactive protein in cerebrospinal fluid in patients with bacterial meningitis and other conditions. J Pediatr. (1986) 108(5, Part 1):665–70. 10.1016/S0022-3476(86)81038-1 [DOI] [PubMed] [Google Scholar]
- 117.Coccaro EF, Lee R, Coussons-Read M. Cerebrospinal fluid and plasma C-reactive protein and aggression in personality-disordered subjects: a pilot study. J Neural Transm. (2015) 122(2):321–6. 10.1007/s00702-014-1263-6 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 118.Ettinger M, Savov P, Calliess T, Windhagen H, Lichtinghagen R, Lukasz A, et al. Improved diagnostic accuracy with the classification tree method for diagnosing low-grade periprosthetic joint infections by quantitative measurement of synovial fluid alpha-defensin and C-reactive protein. Int Orthop. (2020) 44(1):31–8. 10.1007/s00264-019-04338-6 [DOI] [PubMed] [Google Scholar]
- 119.Zaphiriou A, Robb S, Murray-Thomas T, Mendez G, Fox K, McDonagh T, et al. The diagnostic accuracy of plasma BNP and NTproBNP in patients referred from primary care with suspected heart failure: results of the UK natriuretic peptide study. Eur J Heart Fail. (2005) 7(4):537–41. 10.1016/J.EJHEART.2005.01.022 [DOI] [PubMed] [Google Scholar]
- 120.Maisel AS, Krishnaswamy P, Nowak RM, McCord J, Hollander JE, Duc P, et al. Rapid measurement of B-type natriuretic peptide in the emergency diagnosis of heart failure. N Engl J Med. (2002) 347(3):161–7. 10.1056/NEJMOA020233 [DOI] [PubMed] [Google Scholar]
- 121.McDonagh TA, Metra M, Adamo M, Gardner RS, Baumbach A, Böhm M, et al. 2021 ESC guidelines for the diagnosis and treatment of acute and chronic heart failure: developed by the task force for the diagnosis and treatment of acute and chronic heart failure of the European Society of Cardiology (ESC) with the special contribution of the heart failure association (HFA) of the ESC. Eur Heart J. (2021) 42(36):3599–726. 10.1093/eurheartj/ehab368 [DOI] [PubMed] [Google Scholar]
- 122.Heidenreich PA, Bozkurt B, Aguilar D, Allen LA, Byun JJ, Colvin MM, et al. 2022 AHA/ACC/HFSA guideline for the management of heart failure: a report of the American College of Cardiology/American Heart Association joint committee on clinical practice guidelines. Circulation. (2022) 145(18):e895–e1032. 10.1161/CIR.0000000000001063 [DOI] [PubMed] [Google Scholar]
- 123.Fonarow GC, Peacock WF, Phillips CO, Givertz MM, Lopatin M. Admission B-type natriuretic peptide levels and in-hospital mortality in acute decompensated heart failure. J Am Coll Cardiol. (2007) 49(19):1943–50. 10.1016/j.jacc.2007.02.037 [DOI] [PubMed] [Google Scholar]
- 124.Januzzi JL, van Kimmenade R, Lainchbury J, Bayes-Genis A, Ordonez-Llanos J, Santalo-Bel M, et al. NT-proBNP testing for diagnosis and short-term prognosis in acute destabilized heart failure: an international pooled analysis of 1256 patients: the international collaborative of NT-proBNP study. Eur Heart J. (2006) 27(3):330–7. 10.1093/eurheartj/ehi631 [DOI] [PubMed] [Google Scholar]
- 125.Vergaro G, Januzzi JL, Jr, Cohen Solal A, Aimo Alberto, Arzilli Chiara, Zyw Luc, et al. NT-proBNP prognostic value is maintained in elderly and very elderly patients with chronic systolic heart failure. Int J Cardiol. (2018) 271:324–30. 10.1016/j.ijcard.2018.04.006 [DOI] [PubMed] [Google Scholar]
- 126.Cohen-Solal A, Logeart D, Huang B, Cai D, Nieminen MS, Mebazaa A. Lowered B-type natriuretic peptide in response to levosimendan or dobutamine treatment is associated with improved survival in patients with severe acutely decompensated heart failure. J Am Coll Cardiol. (2009) 53(25):2343–8. 10.1016/j.jacc.2009.02.058 [DOI] [PubMed] [Google Scholar]
- 127.Januzzi JL, Jr, Sakhuja R, O’Donoghue M, Baggish AL, Anwaruddin S, Chae CU, et al. Utility of amino-terminal pro–brain natriuretic peptide testing for prediction of 1-year mortality in patients with dyspnea treated in the emergency department. Arch Intern Med. (2006) 166(3):315–20. 10.1001/archinte.166.3.315 [DOI] [PubMed] [Google Scholar]
- 128.Aimo A, Gaggin HK, Barison A, Emdin M, Januzzi JL. Imaging, biomarker, and clinical predictors of cardiac remodeling in heart failure with reduced ejection fraction. JACC Heart Fail. (2019) 7(9):782–94. 10.1016/j.jchf.2019.06.004 [DOI] [PubMed] [Google Scholar]
- 129.Wu AHB, Smith A, Wieczorek S, Mather JF, Duncan B, White CM, et al. Biological variation for N-terminal pro- and B-type natriuretic peptides and implications for therapeutic monitoring of patients with congestive heart failure. Am J Cardiol. (2003) 92(5):628–31. 10.1016/S0002-9149(03)00741-0 [DOI] [PubMed] [Google Scholar]
- 130.Ng LL, Geeranavar S, Jennings SC, Loke I, O'Brien RJ. Diagnosis of heart failure using urinary natriuretic peptides. Clin Sci. (2004) 106(2):129–33. 10.1042/CS20030234 [DOI] [PubMed] [Google Scholar]
- 131.Ramadan MAM, Shrive AK, Holden D, Myles DAA, Volanakis JE, DeLucas LJ, et al. The three-dimensional structure of calcium-depleted human C-reactive protein from perfectly twinned crystals. Acta Crystallogr D. (2002) 58(6 Part 2):992–1001. 10.1107/S0907444902005693 [DOI] [PubMed] [Google Scholar]
- 132.Thompson D, Pepys MB, Wood SP. The physiological structure of human C-reactive protein and its complex with phosphocholine. Structure. (1999) 7(2):169–77. 10.1016/S0969-2126(99)80023-9 [DOI] [PubMed] [Google Scholar]
- 133.Cristell N, Cianflone D, Durante A, Ammirati E, Vanuzzo D, Banfi M, et al. High-sensitivity C-reactive protein is within normal levels at the very onset of first ST-segment elevation acute myocardial infarction in 41% of cases. J Am Coll Cardiol. (2011) 58(25):2654–61. 10.1016/j.jacc.2011.08.055 [DOI] [PubMed] [Google Scholar]
- 134.Tello-Montoliu A, Marín F, Roldán V, Tello-Montoliu A, Mainar L, López MT, et al. A multimarker risk stratification approach to non-ST elevation acute coronary syndrome: implications of troponin T, CRP, NT pro-BNP and fibrin D-dimer levels. J Intern Med. (2007) 262(6):651–8. 10.1111/j.1365-2796.2007.01871.x [DOI] [PubMed] [Google Scholar]
- 135.James SK, Armstrong P, Barnathan E, Califf R, Lindahl B, Siegbahn A, et al. Troponin and C-reactive protein have different relations to subsequent mortality and myocardial infarction after acute coronary syndrome: a GUSTO-IV substudy. J Am Coll Cardiol. (2003) 41(6):916–24. 10.1016/S0735-1097(02)02969-8 [DOI] [PubMed] [Google Scholar]
- 136.Al Aseri ZA, Habib SS, Marzouk A. Predictive value of high sensitivity C-reactive protein on progression to heart failure occurring after the first myocardial infarction. Vasc Health Risk Manag. (2019) 15:221–7. 10.2147/VHRM.S198452 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 137.Mayr A, Klug G, Schocke M, Trieb T, Mair J, Pedarnig K, et al. Late microvascular obstruction after acute myocardial infarction: relation with cardiac and inflammatory markers. Int J Cardiol. (2012) 157(3):391–6. 10.1016/j.ijcard.2010.12.090 [DOI] [PubMed] [Google Scholar]
- 138.Ohlmann P, Jaquemin L, Morel O, El Behlgiti R, Faure A, Michotey M-O, et al. Prognostic value of C-reactive protein and cardiac troponin I in primary percutaneous interventions for ST-elevation myocardial infarction. Am Heart J. (2006) 152(6):1161–7. 10.1016/j.ahj.2006.07.016 [DOI] [PubMed] [Google Scholar]
- 139.Karadeniz M, Duran M, Akyel A, Yarlıoğlueş M, Öcek AH, Çelik İE, et al. High sensitive CRP level is associated with intermediate and high syntax score in patients with acute coronary syndrome. Int Heart J. (2015) 56(4):377–80. 10.1536/ihj.14-299 [DOI] [PubMed] [Google Scholar]
- 140.Ridker PM, Danielson E, Fonseca FAH, Genest J, Gotto AM, Kastelein JJP, et al. Rosuvastatin to prevent vascular events in men and women with elevated C-reactive protein. N Engl J Med. (2008) 359(21):2195–207. 10.1056/NEJMoa0807646 [DOI] [PubMed] [Google Scholar]
- 141.Ridker PM, Everett BM, Thuren T, MacFadyen JG, Chang WH, Ballantyne C, et al. Antiinflammatory therapy with canakinumab for atherosclerotic disease. N Engl J Med. (2017) 377(12):1119–31. 10.1056/NEJMoa1707914 [DOI] [PubMed] [Google Scholar]
- 142.Tardif JC, Simon K, Waters DD, Kouz S, Bertrand OF, Diaz R, et al. Efficacy and safety of low-dose colchicine after myocardial infarction. N Engl J Med. (2019) 381(26):2497–505. 10.1056/NEJMoa1912388 [DOI] [PubMed] [Google Scholar]
- 143.Jolly SS, d’Entremont MA, Lee SF, Mian R, Tyrwhitt J, Kedev S, et al. Colchicine in acute myocardial infarction. N Engl J Med. (2025) 392(7):633–42. 10.1056/NEJMoa2405922 [DOI] [PubMed] [Google Scholar]
- 144.Chen SC, Chen W, Hsu WH, Yu YH, Shih CM. Role of pleural fluid C-reactive protein concentration in discriminating uncomplicated parapneumonic pleural effusions from complicated parapneumonic effusion and empyema. Lung. (2006) 184(3):141–5. 10.1007/s00408-005-2573-0 [DOI] [PubMed] [Google Scholar]
- 145.Bartel DP. MicroRNAs: genomics, biogenesis, mechanism, and function. Cell. (2004) 116(2):281–97. 10.1016/S0092-8674(04)00045-5 [DOI] [PubMed] [Google Scholar]
- 146.Bartel DP. MicroRNAs: target recognition and regulatory functions. Cell. (2009) 136(2):215–33. 10.1016/j.cell.2009.01.002 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 147.Kozomara A, Birgaoanu M, Griffiths-Jones S. miRBase: from microRNA sequences to function. Nucleic Acids Res. (2019) 47(D1):D155–62. 10.1093/nar/gky1141 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 148.van Rooij E, Olson EN. MicroRNA therapeutics for cardiovascular disease: opportunities and obstacles. Nat Rev Drug Discov. (2012) 11(11):860–72. 10.1038/nrd3864 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 149.Small EM, Frost RJA, Olson EN. MicroRNAs add a new dimension to cardiovascular disease. Circulation. (2010) 121(8):1022–32. 10.1161/CIRCULATIONAHA.109.889048 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 150.Cheng Y, Zhang C. MicroRNA-21 in cardiovascular disease. J Cardiovasc Transl Res. (2010) 3(3):251–5. 10.1007/s12265-010-9169-7 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 151.Offer G, Moos C, Starr R. A new protein of the thick filaments of vertebrate skeletal myofibrils: extraction, purification and characterization. J Mol Biol. (1973) 74(4):653–76. 10.1016/0022-2836(73)90055-7 [DOI] [PubMed] [Google Scholar]
- 152.Kaier TE, Twerenbold R, Puelacher C, Marjot J, Imambaccus N, Boeddinghaus J, et al. Direct comparison of cardiac myosin-binding protein C with cardiac troponins for the early diagnosis of acute myocardial infarction. Circulation. (2017) 136(16):1495–508. 10.1161/CIRCULATIONAHA.117.028084 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 153.Lee TY. Lactate: a multifunctional signaling molecule. Yeungnam Univ J Med. (2021) 38(3):183. 10.12701/YUJM.2020.00892 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 154.Klemm G, Markart S, Hermann A, Staudinger T, Hengstenberg C, Heinz G, et al. Lactate as a predictor of 30-day mortality in cardiogenic shock. J Clin Med. (2024) 13(7):1932. 10.3390/JCM13071932/S1 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 155.Raj Kulshrestha M, Tiwari V, Tripathi P. Serum lactate level as a potent biomarker in acute myocardial infarction. Int J Clin Biochem Res. (2025) 6(3):353–5. 10.18231/J.IJCBR.2019.077 [DOI] [Google Scholar]
- 156.Zymliński R, Biegus J, Sokolski M, Siwołowski P, Nawrocka-Millward S, Todd J, et al. Increased blood lactate is prevalent and identifies poor prognosis in patients with acute heart failure without overt peripheral hypoperfusion. Eur J Heart Fail. (2018) 20(6):1011–8. 10.1002/ejhf.1156 [DOI] [PubMed] [Google Scholar]
- 157.Vermeulen RP, Hoekstra M, Nijsten MWN, van der Horst IC, van Pelt LJ, Jessurun GA, et al. Clinical correlates of arterial lactate levels in patients with ST-segment elevation myocardial infarction at admission: a descriptive study. Crit Care. (2010) 14(5):R164. 10.1186/cc9253 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 158.De Fauw J, Ledsam JR, Romera-Paredes B, Nikolov S, Tomasev N, Blackwell S, et al. Clinically applicable deep learning for diagnosis and referral in retinal disease. Nat Med. (2018) 24(9):1342–50. 10.1038/s41591-018-0107-6 [DOI] [PubMed] [Google Scholar]
- 159.Driban M, Yan A, Selvam A, Ong J, Vupparaboina KK, Chhablani J. Artificial intelligence in chorioretinal pathology through fundoscopy: a comprehensive review. Int J Retina Vitreous. (2024) 10(1):1–12. 10.1186/S40942-024-00554-4 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 160.Martínez-Sellés M, Marina-Breysse M. Current and future use of artificial intelligence in electrocardiography. J Cardiovasc Dev Dis. (2023) 10(4):175. 10.3390/JCDD10040175 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 161.Ose B, Sattar Z, Gupta A, Toquica C, Harvey C, Noheria A. Artificial intelligence interpretation of the electrocardiogram: a state-of-the-art review. Curr Cardiol Rep. (2024) 26(6):561–80. 10.1007/S11886-024-02062-1/FIGURES/1 [DOI] [PubMed] [Google Scholar]
- 162.Mills NL, Lee KK, McAllister DA, Churchhouse AMD, MacLeod M, Stoddart M, et al. Implications of lowering threshold of plasma troponin concentration in diagnosis of myocardial infarction: cohort study. Br Med J. (2012) 344(7852):e1533. 10.1136/BMJ.E1533 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 163.NHS England. Hospital Accident & Emergency Activity, 2023-24, AE_2324_ECDS_open_data_csv (2024). Available online at: https://digital.nhs.uk/data-and-information/publications/statistical/hospital-accident-emergency-activity/2023-24 (Accessed August 3, 2025)
- 164.FRagmin and Fast Revascularisation during InStability in Coronary artery disease (FRISC II) Investigators. Invasive compared with non-invasive treatment in unstable coronary-artery disease: fRISC II prospective randomised multicentre study. Lancet. (1999) 354(9180):708–15. 10.1016/S0140-6736(99)07349-3 [DOI] [PubMed] [Google Scholar]
- 165.Kofoed KF, Kelbæk H, Hansen PR, Torp-Pedersen C, Høfsten D, Kløvgaard L, et al. Early versus standard care invasive examination and treatment of patients with non-ST-segment elevation acute coronary syndrome. Circulation. (2018) 138(24):2741–50. 10.1161/CIRCULATIONAHA.118.037152 [DOI] [PubMed] [Google Scholar]
- 166.Rezkalla SH, Kloner RA. Invasive versus conservative management in coronary artery disease. Clin Med Res. (2023) 21(2):95–104. 10.3121/cmr.2023.1806 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 167.Cannon CP, Weintraub WS, Demopoulos LA, Vicari R, Frey MJ, Lakkis N, et al. Comparison of early invasive and conservative strategies in patients with unstable coronary syndromes treated with the glycoprotein IIb/IIIa inhibitor tirofiban. N Engl J Med. (2001) 344(25):1879–87. 10.1056/NEJM200106213442501 [DOI] [PubMed] [Google Scholar]
- 168.Neumann FJ, Kastrati A, Pogatsa-Murray G, Mehilli J, Bollwein H, Bestehorn H-P, et al. Evaluation of prolonged antithrombotic pretreatment (“cooling-off” strategy) before intervention in patients with unstable coronary syndromes: a randomized controlled trial. JAMA. (2003) 290(12):1593–9. 10.1001/jama.290.12.1593 [DOI] [PubMed] [Google Scholar]
- 169.Fox KAA, Poole-Wilson PA, Henderson RA, Clayton TC, Chamberlain DA, Shaw TRD, et al. Interventional versus conservative treatment for patients with unstable angina or non-ST-elevation myocardial infarction: the British heart foundation RITA 3 randomised trial. Lancet. (2002) 360(9335):743–51. 10.1016/S0140-6736(02)09894-X [DOI] [PubMed] [Google Scholar]
- 170.Boden WE, O’Rourke RA, Crawford MH, O'Rourke RA, Blaustein AS, Deedwania PC, et al. Outcomes in patients with acute non–Q-wave myocardial infarction randomly assigned to an invasive as compared with a conservative management strategy. N Engl J Med. (1998) 338(25):1785–92. 10.1056/NEJM199806183382501 [DOI] [PubMed] [Google Scholar]
- 171.de Winter RJ, Windhausen F, Cornel JH, Dunselman PHJM, Janus CL, Bendermacher PEF, et al. Early invasive versus selectively invasive management for acute coronary syndromes. N Engl J Med. (2005) 353(11):1095–104. 10.1056/NEJMoa044259 [DOI] [PubMed] [Google Scholar]
- 172.Masson S, Latini R, Anand IS, Vago T, Angelici L, Barlera S, et al. Direct comparison of B-type natriuretic peptide (BNP) and amino-terminal proBNP in a large population of patients with chronic and symptomatic heart failure: the valsartan heart failure (val-HeFT) data. Clin Chem. (2006) 52(8):1528–38. 10.1373/CLINCHEM.2006.069575 [DOI] [PubMed] [Google Scholar]
- 173.Fuery MA, Leifer ES, Samsky MD, Sen S, O’Connor CM, Fiuzat M, et al. Prognostic impact of repeated NT-proBNP measurements in patients with heart failure with reduced ejection fraction. Heart Fail. (2024) 12(3):479–87. 10.1016/J.JCHF.2023.11.007 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 174.Li P, Luo Y, Chen YM. B-type natriuretic peptide-guided chronic heart failure therapy: a meta-analysis of 11 randomised controlled trials. Heart Lung Circ. (2013) 22(10):852–60. 10.1016/j.hlc.2013.03.077 [DOI] [PubMed] [Google Scholar]
- 175.Porapakkham P, Porapakkham P, Zimmet H, Billah B, Krum H. B-type natriuretic peptide-guided heart failure therapy: a meta-analysis. Arch Intern Med. (2010) 170(6):507–14. 10.1001/ARCHINTERNMED.2010.35 [DOI] [PubMed] [Google Scholar]
- 176.Felker GM, Hasselblad V, Hernandez AF, O'Connor CM. Biomarker-guided therapy in chronic heart failure: a meta-analysis of randomized controlled trials. Am Heart J. (2009) 158(3):422–30. 10.1016/j.ahj.2009.06.018 [DOI] [PubMed] [Google Scholar]
- 177.Ahmed FZ, Taylor JK, John AV, Khan MA, Zaidi AM, Mamas MA, et al. Ambulatory intravenous furosemide for decompensated heart failure: safe, feasible, and effective. ESC Heart Fail. (2021) 8(5):3906–16. 10.1002/ehf2.13368 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 178.Yancy CW, Burnett JC, Fonarow GC, Silver MA. Decompensated heart failure: is there a role for the outpatient use of nesiritide? Congest Heart Fail. (2004) 10(5):230–6. 10.1111/j.1527-5299.2004.03910.x [DOI] [PubMed] [Google Scholar]
- 179.Leiva O, Bhatt AS, Vaduganathan M. Innovation in ambulatory care of heart failure in the era of coronavirus disease 2019. Heart Fail Clin. (2020) 16(4):433–40. 10.1016/j.hfc.2020.06.004 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 180.Health Quality Ontario. Community-based care for the specialized management of heart failure: an evidence-based analysis. Ont Health Technol Assess Ser. (2009) 9(17):1–42. Available online at: http://www.ncbi.nlm.nih.gov/pubmed/23074521 (Accessed January 13, 2025). [PMC free article] [PubMed] [Google Scholar]
- 181.Ong MK, Romano PS, Edgington S, Aronow HU, Auerbach AD, Black JT, et al. Effectiveness of remote patient monitoring after discharge of hospitalized patients with heart failure: the better effectiveness after transition – heart failure (BEAT-HF) randomized clinical trial. JAMA Intern Med. (2016) 176(3):310–8. 10.1001/JAMAINTERNMED.2015.7712 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 182.Adamson PB, Ginn G, Anker SD, Bourge RC, Abraham WT. Remote haemodynamic-guided care for patients with chronic heart failure: a meta-analysis of completed trials. Eur J Heart Fail. (2017) 19(3):426–33. 10.1002/EJHF.638 [DOI] [PubMed] [Google Scholar]
- 183.Manickam P, Mariappan SA, Murugesan SM, Hansda S, Kaushik A, Shinde R, et al. Artificial intelligence (AI) and internet of medical things (IoMT) assisted biomedical systems for intelligent healthcare. Biosensors (Basel). (2022) 12(8):562. 10.3390/BIOS12080562 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 184.Lee MS, Shin TG, Lee Y, Kim DH, Choi SH, Cho H, et al. Artificial intelligence applied to electrocardiogram to rule out acute myocardial infarction: the ROMIAE multicentre study. Eur Heart J. (2025) 46(20):1917–29. 10.1093/EURHEARTJ/EHAF004 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 185.Hempel P, Bender T, Spicher N. Enhancing explainability in ECG analysis through evidence-based AI interpretability. In: Proceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBS (2024). 10.1109/EMBC53108.2024.10782420 [DOI] [PubMed] [Google Scholar]
- 186.Kolhar M, Kazi RNA, Mohapatra H, Al Rajeh AM. AI-driven real-time classification of ECG signals for cardiac monitoring using i-AlexNet architecture. Diagnostics. (2024) 14(13):134. 10.3390/DIAGNOSTICS14131344 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 187.Bartusik-Aebisher D, Rogóż K, Aebisher D. Artificial intelligence and ECG: a new frontier in cardiac diagnostics and prevention. Biomedicines. (2025) 13(7):1685. 10.3390/BIOMEDICINES13071685 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 188.Ukil A, Bandyopadhyay S. Automated cardiac health screening using smartphone and wearable sensors through anomaly analytics. In: EAI/Springer Innovations in Communication and Computing (2019), 145–72. 10.1007/978-3-319-93491-4_8 [DOI] [Google Scholar]
- 189.Ergen O. Robust atrial fibrillation monitoring utilizing graphene aerogel-based nano-tattoo. Mater Lett. (2021) 291:129525. 10.1016/J.MATLET.2021.129525 [DOI] [Google Scholar]
- 190.Hannan A, Cheema SM, Pires IM. Machine learning-based smart wearable system for cardiac arrest monitoring using hybrid computing. Biomed Signal Process Control. (2024) 87:105519. 10.1016/J.BSPC.2023.105519 [DOI] [Google Scholar]
- 191.Abedi A, Verma A, Jain D, Kaetheeswaran J, Chui C, Lankarany M, et al. AI-driven real-time monitoring of cardiovascular conditions with wearable devices: scoping review. JMIR Mhealth Uhealth. (2025) 13(1):e73846. 10.2196/73846 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 192.Islam MR, Kabir MM, Mridha MF, Alfarhood S, Safran M, Che D. Deep learning-based IoT system for remote monitoring and early detection of health issues in real-time. Sensors. (2023) 23(11):5204. 10.3390/S23115204 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 193.Andrade-Méndez B, Arias-Torres DO, Gómez-Tovar LO. Alarm fatigue in the intensive care unit: relevance and response time. Enferm Intensiva. (2020) 31(3):147–53. 10.1016/J.ENFI.2019.11.002 [DOI] [PubMed] [Google Scholar]
- 194.Li B, Yue L, Nie H, Cao Z, Chai X, Peng B, et al. The effect of intelligent management interventions in intensive care units to reduce false alarms: an integrative review. Int J Nurs Sci. (2023) 11(1):133. 10.1016/J.IJNSS.2023.12.008 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 195.Ma X, Wang L, Meng S, Xu R, Chen R, Zhang X, et al. A retrospective cross-sectional study showing wearable smartwatches enhance patient safety and efficiency in the intensive care unit. Commun Med. (2025) 5(1):341. 10.1038/S43856-025-01072-6 [DOI] [PMC free article] [PubMed] [Google Scholar]