Abstract
Optoacoustic tomography (OAT) is a promising modality for breast cancer diagnosis because tumor angiogenesis and, potentially, hypoxia can be visualized using quantitative OAT (qOAT) techniques. Clinically meaningful inference generally requires accurate image reconstruction, which depends on measurement quality and the imager design. Virtual imaging offers a cost-effective alternative to experimental prototyping for system design evaluation and supports computational method development. This work presents a comprehensive virtual imaging framework for breast qOAT, extending a stochastic numerical breast phantom (NBP) generator by incorporating skin tone variation and both benign and malignant lesions. It enables end-to-end simulation, modeling transducer spatial and electro-acoustic impulse responses. Its utility is demonstrated through a case study comparing two system designs. A total of 1020 NBPs and associated measurement data have been made publicly available to accelerate research in optoacoustic and optical imaging. The framework provides a versatile platform for advancing computational methods and guiding system design optimization.
Keywords: Optoacoustic tomography, Photoacoustic computed tomography, Virtual imaging, In silico imaging, Numerical breast phantoms, Breast imaging
1. Introduction
Three-dimensional (3D) optoacoustic tomography (OA-T), also known as photoacoustic computed tomography (PACT), enables volumetric imaging of the absorbed optical energy density in biological tissues using short, non-ionizing laser pulses, typically at near-infrared (NIR) wavelengths [1], [2], [3], [4], [5], [6]. Optical illumination induces transient thermoelastic expansion, which generates initial pressure distributions through the photoacoustic effect [4], [5]. Ultrasonic transducers surrounding the tissue detect the resulting acoustic waves, which propagates through the tissue [1], [2], [4], [7]. These measurements are subsequently used for image reconstruction [8], [9], [10], [11]. On a macroscopic scale, OAT achieves higher spatial resolution at greater tissue depths than purely optical imaging techniques because acoustic waves scatter substantially less than light in biological tissues [2], [4], [12]. In addition, OAT exploits strong optical contrast from the spectral absorption of endogenous chromophores, primarily oxy- and deoxy-hemoglobin, yielding greater contrast than conventional ultrasound imaging [2], [5], [6].
In OAT, the induced initial pressure distribution is reconstructed by solving an acoustic source inverse problem using measured data [8], [9], [10], [11], [13]. This enables visualization of blood vasculature, which is highly absorptive due to its rich hemoglobin content. In principle, functional parameters, such as blood oxygen saturation, and molecular composition can be estimated from acoustic measurements at multiple light excitation wavelengths using techniques known as quantitative OAT (qOAT), also referred to as quantitative PACT [5], [6], [13], [14]. These estimates are obtained by solving a multiphysics inverse problem that involves both photon transport and acoustic wave propagation [13], [14]. By delineating vascular anatomy and quantifying functional parameters, optoacoustic imaging provides both structural and functional insights. These capabilities make OAT particularly well suited for diagnosis of breast cancer characterized by features such as tumor angiogenesis and hypoxia [1], [2], [4], [5], [15], [16], [17].
Accurate and robust image reconstruction is generally important for extracting clinically meaningful information from OAT images, particularly under diverse clinical conditions. In general, the quality of reconstructed images is fundamentally limited by the quality of measured data, which depends on the design of the OAT imaging system. Unlike many other imaging modalities, 3D OAT imagers offer considerable design flexibility, as each consists of a light delivery subsystem and an acoustic data acquisition subsystem [2], [4], [15]. In the acoustic data acquisition subsystem, detector characteristics such as sensitivity and bandwidth determine the achievable signal-to-noise ratio (SNR) and spatial resolution [2], [12]. Measurement geometry, defined by the spatial arrangement of detectors, and temporal sampling, determined by the acquisition rate, together establish the data completeness, a critical factor for image reconstruction [8], [9]. The configuration of the light delivery subsystem directly determines the initial pressure distribution generated within the tissue, and thus, data quality [6]. The broad range of possible design choices, each of which may significantly affect image quality, necessitates careful evaluation and optimization. This process has been challenging due to the need to construct physical prototypes and well-characterized phantoms, acquire experimental data, and perform comparative validations [5], [10], [18].
To address these challenges, virtual imaging provides a cost-effective and ethical alternative for systematically exploring the extensive parameter space inherent in OAT system design, facilitating efficient optimization at a fraction of the cost and effort required for experimental imaging [5], [10], [18], [19]. Virtual imaging enables the generation of simulated measurement data with complete control over physical factors that affect data quality, which enables the analysis of competing imaging system designs. Through the specification of the to-be-imaged object, virtual imaging also allows for systematic simulation of a wide range of physiological and pathological conditions, supporting diagnostic task-based assessment of image quality [20], [21], [22], [23]. However, the practical value of these advantages depends critically on how realistically the virtual objects can be modeled in physiological, optical, and acoustic terms, as well as on the fidelity with which the relevant imaging physics are represented.
Previous virtual imaging studies of OAT have often relied on oversimplified numerical phantoms [24], [25], [26]. Validation of computational methods or imaging system designs employing such phantoms may result in misleading conclusions [27]. In several studies [7], [28], anatomically realistic datasets of objects were generated by segmenting clinical images acquired from other modalities. However, the number of images in the reported datasets is generally limited. Additionally, in most previous studies, optical properties were directly assigned at a few preselected wavelengths without modeling their dependence on chromophore concentrations, limiting their utility for functional imaging [7], [24], [25], [26], [28], [29]. To overcome these limitations and ensure that virtual imaging studies are applicable in practice, virtual objects must reflect anatomical and physiological variability and model tissue optics by linking chromophore concentrations to optical absorption coefficients [5], [14]. Stochastic object models [30] represent a means to capture the diversity needed for robust method development and system optimization [5], [9], [14], [31]. Flexible wavelength selection is also desirable for greater versatility.
To fully benefit from virtual imaging, the computational model must closely approximate the physics of real-world imaging. Without high model fidelity, insights gained from simulation may not translate to real-world performance on experimental data. Light sources are frequently approximated with alternative setups because matching experimental illumination conditions is not always feasible [5], [7], suggesting opportunities to extend current simulation capabilities. Additionally, incorporating detailed transducer response modeling and accurate measurement geometry into large-scale 3D simulations is impractical due to computational constraints [32]. Implementation-level improvements to current simulation tools may enable scalable simulations and enhance acoustic modeling fidelity.
A software tool for generating 3D numerical breast phantoms (NBPs) for use in OAT imaging, referred to as the stochastic optoacoustic NBP (SOA-NBP) [5], has been developed. Phantoms generated using SOA-NBP exhibit stochastic, anatomically and optoacoustically realistic spatial distributions of breast tissue and malignant lesions, enabling virtual imaging studies that reflect clinically relevant variability. By incorporating a wavelength-dependent tissue optics model parameterized by underlying physiological quantities, the framework enables the generation of phantoms at arbitrary wavelengths within the NIR range, which is required for their use in qOAT virtual imaging studies. Eighty-four NBP datasets have been released, each providing an anatomical map as well as functional, optical, and acoustic property maps that collectively define the phantom. Each dataset also includes the corresponding optical fluence maps, initial pressure maps, and pressure measurements, all simulated at three wavelengths using a virtual imaging system configuration that emulates an existing experimental setup [3]. These datasets have already been employed in studies within the optoascoustic (photoacoustic) and optical imaging communities [9], [10], [11], [14], [18], [27], [31], [33], [34], [35], [36], [37], [38].
In this work, a comprehensive virtual imaging framework for qOAT of the breast is proposed. This framework builds on the previously developed SOA-NBP software [39], which stochastically generates 3D NBPs. The proposed framework extends the SOA-NBP by incorporating a detailed skin model that accounts for skin-tone-dependent optical absorption and by introducing models of benign lesions in addition to malignant tumors. These enhancements enable the framework to represent the anatomical and physiological diversity of the target population and to guide system designs toward robust performance across such variability. The simulation components have also been advanced to more faithfully model the entire imaging process. Specifically, the framework incorporates the transducers’ spatial impulse response (SIR) in large-scale 3D simulations and their electro-acoustic impulse response (EIR) modeled from experimental data. These extensions establish a versatile virtual imaging environment that can facilitate a variety of research applications. The datasets generated in this study using the proposed framework have been made publicly available (see Data Availability for details).
The contributions of this work are fourfold. First, an extensive review of virtual imaging in 3D OAT, including commonly employed tools, is presented. Second, building upon previous work, a publicly available collection of 1020 anatomically and optoacoustically realistic NBPs is provided, together with the corresponding optical fluence maps, initial pressure maps, and pressure measurements to support and accelerate research within the broader optical and optoacoustic imaging communities. The diversity spans breast density types, skin tones, and lesion phenotypes. This dataset is particularly well suited for the development and evaluation of data-driven techniques beyond image reconstruction, including a wide range of image processing and image analysis tasks. Third, a reproducible and modular simulation framework, incorporating extended functionalities of existing tools, provides a practical foundation for modeling realistic OAT imaging scenarios. Fourth, the utility of the proposed virtual imaging framework is demonstrated through a case study involving a comparative evaluation of OAT imaging system designs.
The remainder of the paper is organized as follows. Section 2 provides reviews on SOA-NBP and virtual imaging in OAT. The proposed virtual imaging framework is detailed in Section 3. Section 4 presents example NBPs and the corresponding OAT images. An application of the framework for evaluating example OAT imaging system designs is demonstrated in Section 5. The paper concludes in Section 6.
2. Background
SOA-NBP, a key component in the considered 3D virtual imaging framework, is reviewed in Section 2.1. Section 2.2 describes the governing physics and forward model of OAT. A brief review of an existing virtual imaging framework for optoacoustic imaging is provided in Section 2.3.
2.1. SOA-NBP: Stochastic Optoacoustic Numerical Breast Phantom
SOA-NBP [5] is a software tool for generating ensembles of anatomically and physiologically realistic 3D NBPs for use in OAT imaging. This tool was developed by substantially modifying and extending the phantom generation software originally created for digital mammography and digital breast tomosynthesis as part of the U.S. FDA’s virtual imaging clinical trials for regulatory evaluation (VICTRE) project [21]. As in the original FDA phantom, breast shape, size, and tissue composition are user-controllable. Four breast types, as defined in the breast imaging reporting and data system (BI-RADS) [40], are supported: (A) almost entirely fatty breasts, (B) breasts with scattered areas of fibroglandular density, (C) heterogeneously dense breasts, and (D) extremely dense breasts. The NBP consists of various tissue types: fat, skin, glandular tissue, nipple, muscle, ligament, terminal duct lobular unit, duct, artery, and vein.
SOA-NBP addresses the specific requirements of OAT virtual imaging by incorporating an improved blood vasculature model and an advanced malignant lesion model. The lesion model explicitly represents internal tumor heterogeneity, including a viable tumor cell (VTC) region, a necrotic core, and a peritumoral angiogenesis region. To ensure physiological consistency, the optical absorption coefficient is computed based on the local concentrations of major chromophores: oxy- and deoxy-hemoglobin, water, fat, and melanosomes. Specifically, the optical absorption coefficient in each tissue is calculated as a weighted sum of the absorption coefficients of these pure chromophores, with the weights corresponding to their respective volume fractions [41], [42]. This approach supports flexible generation of tissue optical property maps across the NIR range of 700 to 1100 nm, which is commonly used in OAT [1], [2], [3], [4], [15]. Furthermore, the blood oxygen saturation distribution is modeled to represent smooth, biologically realistic transitions between tissue regions.
Each NBP produced using SOA-NBP consists of anatomical, functional, optical, and acoustic property distributions. The functional properties correspond to total hemoglobin concentration (M), blood oxygen saturation (%), and the volume fractions of blood , water , fat , and melanosome (%). The optical properties include optical absorption coefficient (mm−1), scattering coefficient (mm−1), scattering anisotropy , and refractive index . The acoustic properties are sound speed (mm/), density (g/mm), and attenuation coefficient (dB/MHz) with power law exponent . All properties are assigned to each tissue and are stochastically prescribed within physiologically realistic ranges determined through a comprehensive literature survey [5].
The spatial distribution of is modeled by solving a diffusion–reaction partial differential equation [43] using the finite element library FEniCS [44]. This formulation yields smooth transitions between tumor subregions and surrounding healthy tissue, mimicking physiological spatial variability. For cases incorporating tumor angiogenesis, a spatially varying local increase in the distribution is introduced within the VTC and peripheral angiogenesis regions to represent sub-resolution microvasculature. Newly sprouted capillaries have diameters smaller than the spatial resolution of the discretized NBPs and therefore cannot be geometrically resolved. Their contribution is captured through locally increased blood volume fractions, thereby accounting for partial-volume effects at the voxel level. Further methodological details are provided in [5].
Although each tissue is assigned physiologically realistic optical and acoustic properties, the overall spatial composition of breast tissues determines whether the resulting breast volume maintains physiologically consistent global characteristics. To evaluate this, the effective optical attenuation coefficient, spatially averaged over the entire breast volume, was selected as a representative global property because it reflects the combined effects of independently sampled and assigned optical properties within the same underlying anatomy. Estimates derived from the produced NBPs at three illumination wavelengths (757, 800, and 850 nm) were compared with values derived from the measurements reported in the literature, and strong agreement was observed [5]. Furthermore, the volumetric fractions of vasculature in the breast, computed from both clinical OAT breast images and the generated NBPs, also showed comparable mean values [5].
NBP datasets produced using SOA-NBP have been deployed in studies that involve the estimation of the initial pressure distribution by solving the acoustic inverse problem [9], [10], [31], [38], joint estimation of the initial pressure and sound speed distributions [11], [34], [35], [37], estimation of functional quantities from multi-wavelength acoustic measurements through multiphysics inversion [14], [27], and the comparison of light delivery subsystem designs in a 3D OAT breast system [18]. Specifically, the datasets have supported the development of learning-based methods [9], [14], [27] as well as the validation of existing and newly proposed methods [10], [11], [14], [31], [34], [35], [36], [37], [38]. Notably, a method developed using this dataset demonstrated robust generalizability to experimental data [9]. These applications demonstrate the practical utility of the datasets for the optoacoustic (photoacoustic) imaging community. Furthermore, the optical property distributions included in these datasets have also been utilized in research on diffuse optical tomography beyond OAT [33], highlighting their versatility and value for the broader optical imaging community.
2.2. Physics and forward model of OAT
OAT virtual imaging refers to the computational simulation of the OAT forward process using numerical object models. This process consists of two stages: (1) simulation of the optoacoustically induced initial pressure field, and (2) simulation of its propagation through tissue and detection by ultrasonic transducers surrounding the tissue [5], [13].
In the first stage, the initial pressure distribution at position and time , induced by optical illumination at wavelength , is given by:
| (1) |
where is the absorbed optical energy density, is the optical fluence, and is the dimensionless Grüneisen parameter quantifying the conversion efficiency from absorbed optical energy to initial pressure via thermoelastic expansion. For soft tissue, is typically assumed to be constant and is often set to 1 for simplicity [5], [24]. In this case, is equivalent to .
In OAT, light transport is described by the radiative transfer equation (RTE), assuming conservation of photon energy during collisions and a constant refractive index within the medium [13]. Because direct solutions of the RTE are computationally demanding, two approaches are commonly used for practical computation of : the diffusion approximation (DA) [13] and the Monte Carlo (MC) method [45]. In the DA, the RTE is approximated by the diffusion equation under the assumption that the medium’s reduced scattering coefficient, , greatly exceeds its absorption coefficient, [13]. This assumption does not hold in media containing non-scattering regions, such as water used for acoustic coupling in OAT imaging, resulting in inaccurate estimates. In contrast, the MC method, widely regarded as the gold standard, accurately models photon propagation in scattering and absorbing media without relying on the simplifying assumptions required by the DA [13]. MC simulations use random sampling to stochastically model the trajectories of numerous photons and estimate from the resulting photon distribution.
In the second stage, acoustic wave propagation is governed by the acoustic wave equation, and the pressure at detection locations is obtained by solving this equation. Acoustic attenuation in biological soft tissues is typically described by a frequency power law [46]:
| (2) |
where is the acoustic attenuation coefficient, is the frequency-independent attenuation coefficient, is the angular temporal frequency, and is the power law exponent. In practical numerical implementations, is typically assumed to be constant [47]. When soft tissue is modeled as an acoustically heterogeneous, lossy medium, the optoacoustically induced pressure wavefield , at position , time , and optical excitation wavelength , can be described by a coupled system of first-order partial differential equations [47], [48]:
subject to the initial conditions in Eq. (1) and . Here, is a vector-valued function that represents the acoustic particle velocity field, denotes the acoustic absorption coefficient, and is the dispersion proportionality coefficient. The -space pseudo-spectral time-domain method is commonly employed to discretize and solve these coupled equations, as it offers high accuracy, low numerical dispersion, stability, and computational efficiency [48].
2.3. Existing virtual imaging frameworks for optoacoustic imaging
For virtual imaging studies in optoacoustic imaging, several tools are available for simulating photon transport and acoustic wave propagation (see Appendix A). The most widely used are MCX [49], [50], a GPU-accelerated photon transport simulator based on the MC method, and k-Wave [32], an acoustic wave simulator that solves the acoustic wave equation using the above-mentioned -space pseudo-spectral time-domain method. To streamline simulation workflows, a toolkit called SIMPA [51] has been developed. SIMPA is a Python-based software package that operates at the configuration layer, serving as a user-friendly adapter to connect simulation modules and manage data flow between them. It also enables the creation of simple tissue phantoms with shapes such as layers, spheres, elliptical tubes, cuboids, parallelepipeds, and vessel trees, and supports system configurations for commercial imagers, specifically the MSOT Acuity Echo, InVision 256-TF [52], and RSOM Xplorer P50 from iThera Medical (Munich, Germany). The developers of SIMPA have also applied it in comparative studies of two-dimensional (2D) OAT image reconstruction methods [52], demonstrating its utility for image quality assessment studies.
3. Virtual imaging framework for 3D qOAT of the breast
This section presents a comprehensive virtual imaging framework for 3D qOAT of the breast. As illustrated in Fig. 1, the framework consists of three primary steps: (1) generation of optoacoustic NBPs, (2) simulation of the optoacoustically induced initial pressure field via photon transport, and (3) simulation of optoacoustic measurements through acoustic wave propagation.
Fig. 1.
Virtual imaging framework for 3D qOAT of the breast. Anatomical, functional, optical, and acoustic NBPs are generated using the SOA-NBP tool [5]. The generated optical NBPs are used to compute the optoacoustically induced initial pressure field. The resulting distribution and the acoustic NBPs are then used to simulate acoustic wave propagation and optoacoustic measurements, accounting for transducer response characteristics. Compared with the previously proposed SOA-NBP phantom generation tool, the present framework introduces a two-layer skin model and benign lesion models in Step (1), and extends the workflow to virtual imaging by incorporating additional flexibility in light source modeling in Step (2) and explicit modeling of both the SIR and the EIR of the transducers in Step (3).
Optoacoustic NBPs were generated using the previously proposed SOA-NBP tool, which produces anatomically realistic healthy breast tissue maps with embedded subcutaneous blood vasculature. In this study, the tool was extended to include a two-layer skin model that accounts for skin tone variations, as well as an expanded set of lesion types that can optionally be inserted. Functional, optical, and acoustic properties were then stochastically assigned to each tissue type, yielding the complete NBP set. The simulation of the optoacoustically induced initial pressure field involves computing the optical fluence distribution and subsequently deriving the initial pressure distribution, both based on the optical NBPs. The fluence computation was implemented with additional light source customization. Optoacoustic measurements were obtained by simulating acoustic wave propagation and detection, followed by conversion of the detected pressure signals into voltage signals. To faithfully replicate imaging physics under real-world system configurations, the SIR and EIR of transducers were incorporated.
Details of the extended SOA-NBP are provided in Section 3.1. The two simulation steps are described in Sections 3.2, 3.3, respectively, with emphasis on the extensions introduced in this study.
3.1. Extended SOA-NBP
The previous SOA-NBP implementation [5] focused on a single skin tone and malignant tumors, without accounting for variability across other skin tones or lesion types. To broaden its applicability, the extended version models various skin tones and additionally incorporated two benign lesion models. These features are described in Sections 3.1.1, 3.1.2, respectively.
3.1.1. Modeling skin tones
Melanin is a chromophore that exhibits a higher optical absorption coefficient than oxy- and deoxy-hemoglobin in the NIR range of 700 to 1100 nm [41]. Melanin is distributed exclusively in the epidermis, the most superficial layer of the skin, and its concentration determines skin tone [41]. Variations in skin tone, therefore, directly affect the amount of light absorbed in the superficial skin layer during an OAT scan [5], [53]. Darker skin tones have higher melanin concentrations, resulting in a reduction of the amount of light that penetrates to deeper tissues [18], [54], [55]. This reduction in optical fluence, together with depth-dependent optical attenuation, further decreases the optoacoustic contrast in deeper regions and thereby lowering the SNR. Despite this important effect, the extent to which skin tone affects OAT imaging remains relatively unexplored. Only recently have a few studies [18], [54], [55] begun to investigate this effect. Incorporating physiologically realistic skin tone variability into virtual imaging frameworks enables the generation of synthetic optoacoustic data that capture skin pigmentation differences, thereby supporting the systematic investigation of skin tone’s impact on signal detectability and diagnostic performance.
To represent skin tone variability in NBPs, a two-layer skin model was introduced, consisting of a dermis-only layer and an epidermis-included layer. Total skin thickness, as measured in mammographic studies, ranges from 0.5 to 3.1 mm, including both the epidermis and dermis [56], [57], with the epidermis accounting for approximately 3.7–16.8% of the total thickness [57]. Because the spatial resolution currently achievable with OAT (a few hundred microns) [3], [4], [15] is insufficient to resolve the epidermis, the epidermis-included layer in the anatomical phantom was defined as a one-voxel-thick region along the outer skin surface encompassing the entire epidermis and part of the dermis. The thicknesses of the total skin and epidermis are user-configurable. In this study, they were set to 1.5 mm and 0.02 mm, respectively, as an example. Voxels labeled as skin in NBPs generated by the previously proposed SOA-NBP were assigned to the dermis-only layer. The epidermis-included layer was defined as the set of dermis voxels directly adjacent to the background in any of the , , or directions. The definitions of newly introduced or reassigned tissue type labels are provided in Appendix B.
Skin tone is commonly classified using the Fitzpatrick phototype scale [58], which ranges from type I (always burns, never tans) to type VI (never burns), based on the skin’s response to ultraviolet exposure. In phototype I, melanin contributes minimally to optical absorption, resulting in limited contrast in OAT images, whereas phototype VI contains four to eight times more melanin, leading to substantially higher optical absorption within the epidermis [53], [59]. Melanin concentration is represented by , the volume fraction of melanosomes, which are the organelles that synthesize melanin. In SOA-NBP, is defined as one of the functional properties, together with the volume fractions of blood , water , and fat , total hemoglobin concentration , and blood oxygen saturation . For phototypes I to VI, was sampled from uniform distributions within the ranges specified in Table 1 [53]. Because melanosomes are absent in the dermis, the dermis-only layer was assigned . For the epidermis-included layer, the effective was calculated as the arithmetic average based on the configured epidermal thickness and voxel size. The remaining functional, optical, and acoustic properties were assigned to each tissue type as described in [5].
Table 1.
Melanosome volume fraction in epidermis [53].
| Fitzpatrick skin phototype | range (%) |
|---|---|
| I | [1.3, 3.8) |
| II | [3.8, 6.3) |
| III | [6.3, 9.3) |
| IV | [9.3, 14.3) |
| V | [14.3, 21.3) |
| VI | [21.3, 30.3] |
3.1.2. Modeling lesion types
In VICTRE NBP [21], only cancerous masses, for example as shown in Fig. 2(a), and calcifications were modeled. The previous SOA-NBP added two malignant tumor subcomponents, a necrotic core and a peripheral angiogenesis region, resulting in a heterogeneous malignant tumor model, as shown in Fig. 2(b). To enable the use of a broader range of breast lesion models, SOA-NBP has been augmented to include benign lesions with functional, optical, and acoustic characteristics distinct from malignant tumors [16], [17], [60], [61]. Specifically, fibroadenomas, the most common solid masses in women aged 14–35 [60], and simple cysts, the most prevalent breast masses in women aged 35–50 [62], were modeled, as illustrated in Fig. 2(c) and (d), respectively.
Fig. 2.
Anatomical numerical lesion phantoms (NLPs): (a) malignant tumor with homogeneous composition (viable tumor cells only), (b) malignant tumor with heterogeneous composition (viable tumor cells with a necrotic core and a peripheral angiogenesis region), (c) fibroadenoma, and (d) simple cyst. In (b), the lesion volume is clipped along a central plane to reveal the internal composition, whereas the full volumes are displayed in (a), (c), and (d). Volume rendering was performed using ParaView [63].
Both of the considered benign lesions are generally round or oval in shape [60], [64], [65]. Fibroadenomas were modeled with slight surface lobulations to represent their dual composition of glandular and connective tissue, as shown in Fig. 2(c) [60], [64]. In contrast, simple cysts were modeled with smooth contours, consistent with their fluid-filled nature, as presented in Fig. 2(d) [65]. Anatomical numerical lesion phantoms (NLPs) for both benign lesions were generated using the VICTRE tool [21], with spicule formation disabled by setting the parameters spicule.meanInitial and spicule.stdInitial to zero in the configuration file. The resulting anatomical NLPs can be inserted at user-specified locations or at candidate sites automatically selected by the VICTRE tool, while avoiding overlap with previously inserted lesions, skin, nipple, or muscle. The definitions of newly introduced or reassigned tissue type labels are described in Appendix B, and specific lesion shape parameters used in this study are provided in Appendix C.
For fibroadenomas, the functional properties and resulting optical absorption coefficient are similar to those of glandular tissue [66]. Thus, the functional property values were sampled from the predefined probability distributions for glandular tissue [5] and assigned to the voxels corresponding to fibroadenomas. The optical absorption coefficient was computed from these functional property values following the model described in Eq. (1) of [5]. Due to the connective tissue component, which exhibits relatively high optical scattering, the optical scattering coefficient was calculated using the power law model described in Eqs. (2) and (3) of [5] with , , and nm [41]. The scattering anisotropy and were assumed identical to those of glandular tissue [5]. For acoustic modeling, the sound speed was drawn from a normal distribution with mean mm/ and standard deviation mm/s [67]; both values were slightly higher than those of glandular tissue. The remaining acoustic properties were assigned the same values as those for glandular tissue.
Simple cysts consist of nearly pure water, with negligible fat content and no hemoglobin. Accordingly, the water volume fraction was set to 100%, and the volume fractions of other chromophores, , , and , were set to 0%. For both optical and acoustic NLPs, the voxels corresponding to this tissue type were assigned the respective properties of water [5], [42], [68], [69], [70], [71].
3.2. Simulation of the optoacoustically induced initial pressure field
The induced initial pressure distribution was computed according to Eq. (1) as the elementwise product of the Grüneisen parameter , the optical absorption coefficient distribution , and the optical fluence distribution . In this study, was set to 1, as often assumed for soft tissue [5], [24]. The fluence distribution was simulated with the GPU-accelerated MCX [49], [50]. The optical NBPs were used to specify the optical properties of the medium in the simulation. Most configuration steps were aligned with those in previous studies employing MCX; however, this framework introduced two novel features. First, a new light source type was introduced to simulate linear segment illuminators used in several 3D OAT imaging systems [1], [72], [73]. Second, a computationally efficient strategy was adopted to reduce simulation time. Each feature is described below.
Light sources should be configured to closely reflect the light delivery subsystem of the target OAT imager. For systems employing a bottom-up illuminator with divergent emission [35], [75], illumination can be modeled as a cone beam configuration, available as a built-in option in MCX. In this case, the beam position and orientation, together with the cone half-angle determined by the numerical aperture of the illuminator, should be specified in the simulation. As an extension of the proposed framework, systems equipped with linear fiber-optic illuminators [1], [72], [73] were modeled using a newly implemented source type, a line beam with conical angular emission. In this source type, photon launch positions are sampled uniformly along a predefined optical line illuminator, and their initial propagation directions are sampled from a cone-shaped angular distribution around the specified source axis. The cone half-angle determines the divergence of the emitted photons. The angular distribution is chosen to be uniform in polar angle between 0 and the specified half-angle, with azimuthal angle uniformly distributed in . This model approximates the behavior of a densely packed linear emitter array by treating it as a continuous line emitter with controlled divergence. An example optical simulation result from this configuration is shown in Fig. 3.
Fig. 3.
Optical fluence distribution simulated using a newly implemented linear segment illuminator, approximated as a continuous line emitter with a conical angular distribution, shown from two viewing angles. A 20 mm-long illuminator with a numerical aperture of 0.89 was placed on the surface of the water tank. Volume rendering was performed using ParaView [63], with logarithmic scaling and value-dependent opacity adjustment applied for visualization.
In the standard approach, is simulated on the same grid as . For stable computation, the number of photons must greatly exceed the number of voxels in the simulation domain [49]. Consequently, the computation time depends on both the voxel size and domain size. To reduce the computational cost, a coarse-grid photon transport simulation with upsampling was employed. Specifically, the optical property distributions and were first downsampled to a coarser grid using trilinear interpolation prior to the MCX simulation. The resulting was then upsampled back to the original resolution, also via trilinear interpolation, after which was obtained by multiplying this upsampled with on the final grid. This process introduces a smoothing effect similar to that applied in the acoustic simulation stage, and therefore the approximation has limited impact on the simulated detected pressure signals. The reduction in computational cost, along with the resulting error in the simulated pressure data relative to the standard approach for the example setting considered in this study, are presented in Appendix D.
3.3. Simulation of optoacoustic measurements
The simulation of optoacoustic measurements was carried out using both the initial pressure distribution obtained in Section 3.2 and the acoustic NBPs generated with the SOA-NBP tool [5] as inputs. As outlined in Fig. 1, the first step involves acoustic wave propagation and detection, in which pressure waves from the initial pressure distribution are propagated through the acoustic medium defined by the acoustic NBPs and recorded on finite-sized detection surfaces of the transducers while accounting for their SIR. In the second step, these detected pressure signals are converted into optoacoustic measurements, i.e., voltage signals, by incorporating the EIR of the transducers.
For the simulation of acoustic wave propagation and detection, a custom Python library was implemented by extending j-Wave [76]. Two key features that are currently unavailable in j-Wave were incorporated. First, acoustic attenuation described by a frequency power law was implemented. Second, for finite-sized transducers, numerical integration of the simulated pressure field over each transducer’s detection surface [77], [78] at each time step was incorporated. By accounting for the transducer’s SIR, this extension enabled finite-sized transducer modeling and enhanced the realism of the simulated detected pressure signals, albeit within inherent approximations.
The simulated detected pressure signals were convolved with the EIR of the transducers to obtain synthetic optoacoustic measurements. Any EIR can be incorporated into the framework, whether directly measured or modeled. The EIR can be experimentally measured using a delta-pulse source, where a short laser pulse irradiates a highly absorbing planar surface and generates a pressure pulse [79], [80]. However, the measured EIR may include artifacts introduced by the measurement environment, and thus may not fully represent the intrinsic transducer response. To address this, the transducer response was parameterized using the complex exponential method [74], which captures the damped resonant behavior of piezoelectric ultrasonic transducers. The model parameters were estimated from the measured EIR using least squares, and the model order was selected using Akaike’s information criterion (AIC) [74]. Further details of the parameterization are provided in Appendix E. Fig. 4 shows an example of an experimentally measured EIR and the corresponding modeled EIR.
Fig. 4.
Experimentally measured (black solid line) and modeled EIR (red dashed line) of a piezoelectric transducer (Imasonic, France). The time-domain response is shown at the top and the normalized frequency-domain magnitude at the bottom. The measured EIR (black solid line) was acquired by TomoWave (Houston, TX), and the modeled EIR (red dashed line) was obtained employing the complex exponential method [74], with the model order selected based on AIC [74] (see Appendix E for details).
4. Examples of generated NBPs and corresponding OAT images
To demonstrate the capabilities of the proposed framework, example NBPs are presented in Section 4.1. Using the virtual imaging system configurations described in Section 4.2 and the simulation settings detailed in Section 4.3, OAT measurements were simulated from these example NBPs. The corresponding synthetic OAT images are presented in Section 4.4. While Section 3 focused on the extended features introduced in the proposed framework, Section 4.3 specifies the remaining conventional simulation parameters used in this example, providing more details of the simulation setup to ensure reproducibility.
Although incorporating the transducers’ SIR and EIR enables high-fidelity physical modeling in virtual imaging studies, these extended features remain optional depending on research objectives. The examples presented in this section highlight the framework’s flexibility by demonstrating datasets generated with both point-like and finite-size transducer assumptions. All computations were performed on a system equipped with a 16-core Intel Xeon Gold 6130 CPU, 256 GB RAM, and an NVIDIA Tesla V100 GPU with 32 GB of memory.
4.1. Example NBPs
Fig. 5 presents example NBPs, including two naturally shaped breasts of different sizes (first two rows) and one breast constrained in shape by the use of a virtual breast stabilizer cup (bottom row). All NBPs were discretized at a voxel size of 0.125 mm. They incorporate variations across BI-RADS breast density types [40] (A, D, and C), skin phototypes (III, V, and I), and lesion types. The NBPs in the first two rows contain malignant tumors with heterogeneous tissue composition, whereas the NBP in the bottom row includes a malignant tumor with homogeneous tissue composition, a malignant tumor with heterogeneous tissue composition, a fibroadenoma, and a simple cyst. Representative distributions of functional, optical, and acoustic properties are illustrated rather than the complete property sets defined in the NBPs. The additional steps introduced in the extended SOA-NBP framework required approximately 17 min for two-layer skin insertion, 5 min for benign lesion generation, and 2 min for lesion insertion. The computation time varied with the phantom volume dimensions. These additional steps accounted for approximately 8.3% of the total computation time for a single NBP construction.
Fig. 5.
Representative distributions of functional, optical, and acoustic properties: (a) tissue labels, (b) oxygen saturation , (c) blood volume fraction , (d) melanosome volume fraction , (e) optical absorption coefficient , and (f) sound speed of the type A, D, and C breasts (top to bottom). The type A breast (top) is naturally shaped, has skin phototype III ( 8.575%), and contains one malignant tumor with heterogeneous tissue composition. The type D breast (center) is naturally shaped, has skin phototype V ( 15.25%), and contains two malignant tumors with heterogeneous tissue composition. The type C breast (bottom) has a shape constrained by a hemiellipsoidal stabilizer cup and skin phototype I ( 1.9%) and includes four lesions: a malignant tumor with homogeneous tissue composition, a malignant tumor with heterogeneous tissue composition, a fibroadenoma, and a simple cyst. All lesions were positioned at 0, except for the two malignant tumors in the type C breast. To visualize internal structures, only half of each breast volume is shown. The malignant tumor with heterogeneous tissue composition in the type C breast lies outside the displayed field of view. In panel (a), lesion tissues are illustrated with the same colors as in Fig. 2, while healthy tissues are shown in yellow. ParaView [63] was used for volume rendering.
4.2. Virtual imaging system configurations
The NBPs in Fig. 5 were virtually imaged using two imaging system configurations. Both systems employed an identical light delivery subsystem but differed in their OAT data acquisition subsystems, with one using point-like transducers and the other finite-size transducers.
The virtual light delivery subsystem was configured with 20 arc-shaped illuminators, each with a radius of 145 mm and a central angle of 80°, uniformly distributed in azimuth. As shown in Fig. 6(a), five linear fiber-optic segments were mounted on the surface of each illuminator. A single segment is illustrated in Fig. 6(b). Each segment emitted a line beam with a conical angular distribution characterized by a half-angle of 12.5°, directed along the axis perpendicular to the segment and oriented toward the origin of the coordinate system, as depicted in Fig. 6(a).
Fig. 6.
Virtual light delivery subsystem (a, b) and OAT data acquisition subsystems of System A (c) and System B (d). Both systems employed the identical virtual light delivery subsystem, with five linear fiber-optic segments (red lines) attached to the surface of each arc-shaped illuminator, which emits line beams with a conical angular distribution. A total of 20 illuminator arcs were used, resulting in 100 line beams illuminating the breast. Panels (c) and (d) show the measurement geometries of the virtual OAT data acquisition subsystems in Systems A and B, respectively, including all transducer elements and tomographic views. Each transducer element (blue dot) was oriented toward the center of curvature of the spherical aperture.
The virtual OAT data acquisition subsystem of System A was equipped with a hemispherical imaging bowl with a radius of 85 mm, filled with water for acoustic coupling. Acoustic reflections from the imaging bowl boundary were not included in the simulations, and a non-reflecting boundary was assumed at the bowl surface. As shown in Fig. 6(c), a rotating arc-shaped transducer array was positioned on the bowl surface, consisting of 108 point-like transducer elements evenly distributed along an arc aperture with a central angle of 80° and a polar angular spacing of approximately 0.75°. The rotating arc array completed 480 tomographic views with an azimuthal angular step of 0.75°. Each virtual transducer element recorded 3720 time samples at a sampling frequency of 20 MHz. The modeled EIR shown in Fig. 4 was assumed for all elements.
In System B, the virtual OAT data acquisition subsystem employed a spherical-cap imaging bowl with a radius of 85 mm and a height of 56 mm, as shown in Fig. 6(d). As in System A, a non-reflecting boundary was assumed at the bowl surface. A rotating arc-shaped transducer array was positioned on the bowl surface, consisting of 82 transducer elements with a size of 1.1 mm 1.1 mm, evenly distributed along an arc aperture with a central angle of 70° and a polar angular spacing of approximately 0.86°. The rotating arc array completed 320 tomographic views with an azimuthal angular step of 1.125°. Each virtual transducer element recorded 2267 time samples at a sampling frequency of 20 MHz. The same modeled EIR as in System A was assumed.
4.3. Simulation configuration
The MCX simulation was performed with the light delivery subsystem configuration described in Section 4.2. The custom line source type implemented in this study was employed to mimic the beams produced by each fiber bundle segment. The optical wavelengths assumed were 757, 800 (isosbestic points of deoxy- and oxy-hemoglobin), and 850 nm. Spatially heterogeneous distributions of and were prescribed in MCX using the muamus_float optical medium format (option 100). The anisotropy factor and refractive index were fixed at constant values, using their average values within the breast, since they do not vary significantly across the tissues that constitute most of the breast volume [5].
Following the coarse-grid photon transport simulation with upsampling introduced in Section 3.2, the optical fluence distribution was simulated on a grid with a voxel size of 0.5 mm using downsampled optical NBPs, resulting in a simulation domain of 340 × 340 × 170 voxels. Each of the 100 line beams was simulated with 108 photons for a duration of 50 ns. The resulting was then interpolated to the 680 × 680 × 340 voxel grid (voxel size of 0.25 mm) to compute the initial pressure distribution . The computational cost is provided in Appendix D.
The optoacoustic measurement simulation for the type A and D breasts was performed using the data acquisition subsystem of System A, whereas the type C breast was simulated with System B (see Section 4.2). The acoustic NBPs were downsampled from a voxel size of 0.125 mm to 0.25 mm using trilinear interpolation, and the resulting resolution was used in the simulation. For System A, point-like transducer elements were positioned on a grid of size 700 × 700 × 350 voxels. After removing duplicate positions introduced by discretization of Cartesian coordinates with the given grid spacing (voxel size of 0.25 mm), 51,472 unique transducer locations remained. This example corresponds to an on-grid approximation of transducer positions for computational simplicity, although the framework is flexible and can also accommodate arbitrary off-grid positions in a manner similar to finite-size transducer modeling [77]. For System B, the SIR of finite-sized transducers was incorporated as described in Section 3.3, using a simulation grid of size 696 × 696 × 288 voxels. To prevent undesired acoustic reflections at the simulation boundaries, a perfectly matched layer was applied. For both systems, the EIR of the transducers was modeled as described in Section 3.3. The simulation of acoustic wave propagation and detection required 12.42 min and 21.09 GB of GPU memory for System A, and 10.54 min and GPU memory of 18.59 GB for System B.
Electronic noise was modeled as independent and identically distributed Gaussian noise with a mean value of zero. The variance was set based on the measurement data simulated using the type C NBP to yield an SNR of 5 dB. The same noise model was applied to generate the OAT measurements of the other example NBPs. While larger transducers generally provide higher SNR due to spatial averaging of the pressure field, in these examples, however, the same noise level was assumed for both systems as a modeling choice.
4.4. Simulated initial pressure distributions and OAT images
Fig. 7(a)–(c) show the simulated initial pressure distribution of the type A, D, and C NBPs in Fig. 5 at an illumination wavelength of 800, 850, and 757 nm, respectively. Fig. 7(d) and (e) present the ratios within blood vasculature and lesions of the type C breast, calculated between the illumination wavelengths of 800 and 757 nm and between 850 and 800 nm, respectively. These simulated distributions were then used as inputs to generate detected pressure, from which the corresponding noisy OAT measurements were acquired using the virtual imaging systems described in Section 4.2.
Fig. 7.
Simulated initial pressure distribution . Panels (a)–(c) show the distributions of the type A, D, and C breasts from Fig. 5 at illumination wavelengths of 850, 800, and 757 nm, respectively. Panels (d) and (e) show ratios within blood vasculature and lesions of the type C breast, calculated for the wavelength pairs (, ) (800, 757) nm and (850, 800) nm, respectively. Volume rendering was performed using ParaView [63].
For image reconstruction, Wiener deconvolution [81], [82] was first applied to the simulated OAT measurements to compensate for the transducers’ EIR, using the experimentally measured response shown in Fig. 4. From the resulting deconvolved data, 3D distributions of were reconstructed with a voxel size of 0.5 mm using the universal backprojection method (UBP) [83], under the assumption of a two-region, piecewise-constant sound speed model. The sound speed values for water () and breast tissue () were selected via a 2D grid search over mm/ in increments of 0.005 mm/, choosing the pair that minimized the mean squared error (MSE) between the UBP reconstruction and the true distribution. It should be noted that this strategy is not applicable in practice, since the true distribution is unavailable. Here, this approach was adopted solely for the purpose of presenting example OAT images. Fig. 8(a) and (b) show the reconstructed images of the type A breast ( mm/ and mm/) and the type D breast ( mm/ and mm/), respectively. An example OAT image of the type C breast is presented as part of the case study in Section 5.
Fig. 8.
Reconstructed OAT images of (a) the type A and (b) the type D breasts from Fig. 5, obtained using UBP with a two-region sound speed model. The corresponding true distributions are shown in Fig. 7(a) and (b), respectively. Volume rendering was performed using ParaView [63], with the color map adjusted for improved visibility. In panel (a), the type A breast with skin phototype III shows a skin boundary recognizable through subcutaneous vasculature, whereas in panel (b), the type D breast with skin phototype V, which has higher optical absorption by melanin than phototype III, exhibits a more distinct skin boundary itself. Because point-like transducers were employed in virtual imaging, no blurring effects related to transducer size were observed in both images.
5. Case study: Evaluation of OAT imaging system designs
To demonstrate the utility of the proposed virtual imaging framework, a case study was conducted to evaluate and compare three OAT imaging system designs employing transducers of varying dimensions. System B, which employed transducer elements measuring 1.1 mm 1.1 mm as described above, served as the baseline design. Two additional variants of System B were considered, with enlarged element sizes of 1.1 mm 3.3 mm and 1.1 mm 6.6 mm. The simulation of acoustic wave propagation and detection required 10.61 min and 19.48 GB of GPU memory for the 1.1 mm 3.3 mm configuration, and 10.75 min and 20.94 GB of GPU memory for the 1.1 mm 6.6 mm configuration.
For each system configuration, virtual imaging was performed using the type C NBP shown in Fig. 5 with the simulation setup described in Section 4.3. For the systems with 1.1 mm 3.3 mm and 1.1 mm 6.6 mm elements, the variance of the Gaussian noise was scaled to be inversely proportional to the transducer detection surface area. This scaling reflects area averaging of uncorrelated noise at larger apertures. The resulting SNRs of the simulated pressure data were 8.34 dB and 9.54 dB, respectively.
Two image reconstruction methods were employed: the UBP [83] with a two-region sound speed model and a parameterized joint reconstruction (JR) approach [11], [84]. In the JR, both the initial pressure distribution and the breast sound speed were estimated simultaneously, with the sound speed restricted to a two-region parameterization [84]. A single constant value was assigned to the breast region, while the water sound speed was fixed at mm/. The JR problem was formulated as an optimization-based inverse problem [11], with the wave equation in Eq. (3) modeling the forward propagation of acoustic waves. The model was numerically implemented using the -space pseudo-spectral time-domain method, with the transducers’ SIR explicitly incorporated. Non-negativity and total variation constraints were imposed on the distribution. During reconstruction, the acoustic attenuation parameters ( and ) and density were fixed to known constants for water, whereas for the breast region they were determined based on the generated NBP ensemble. The optimization algorithm, including the constraints and stopping rule for the iterations, followed the approach in [11] and was combined with the two-region parameterization in [84], both adapted from 2D to the present 3D setting [37]. For comparison, in the two-region UBP, was set to the parameterized JR estimate (approximately 1.498 mm/).
The reconstructed images had a voxel size of 0.5 mm, which was larger than the grid spacing of 0.25 mm employed in the OAT measurement simulations. This choice was made to avoid the so-called inverse crime [85], that is, the use of the same discretization in both the forward model and the reconstruction. The comparison between the two-region UBP and the parameterized JR aimed to investigate the necessity of accounting for the transducer’s SIR during reconstruction under the assumed system designs. Although the transducer’s EIR could be incorporated into the imaging model at each iteration of the parameterized JR, it was instead compensated identically for both methods in this study by preprocessing all simulated measurements with Wiener deconvolution, as described in Section 4.4.
Fig. 9(a) and (b) present the reconstruction results obtained with the two-region UBP and the parameterized JR, respectively. As the transducer element size increased, the UBP results exhibited pronounced blurring, as shown in Fig. 9(a). In contrast, the parameterized JR maintained relatively consistent spatial resolution across all three system designs, as presented in Fig. 9(b). These results indicate that, under the given measurement geometry, analytic reconstruction methods that do not account for the SIR suffer noticeable resolution degradation when applied to data acquired with transducer elements larger than 1.1 mm 3.3 mm. In such cases, incorporating the transducer’s SIR during image reconstruction is necessary for achieving high spatial resolution.
Fig. 9.
Reconstructed OAT images of the type C breast from Fig. 5 using (a) two-region UBP and (b) parameterized JR method that incorporates the SIR of the transducer. For both methods, the transducer’s EIR was compensated identically by preprocessing all simulated measurements using Wiener deconvolution. From left to right, results are shown for System B with transducer element sizes of 1.1 mm 1.1 mm, 1.1 mm 3.3 mm, and 1.1 mm 6.6 mm. The corresponding true initial pressure distribution is provided in Fig. 7(c). Volume rendering was performed using ParaView [63], with the color map adjusted but kept identical across all images for consistent visualization. The UBP reconstruction results in (a) exhibit greater resolution degradation as the transducer element size increased, whereas the images estimated using parameterized JR in (b) maintain relatively consistent spatial resolution across all three system designs. These results suggest that, under the given measurement geometry, system designs with transducer elements larger than 1.1 mm 3.3 mm require incorporating the transducer’s SIR during image reconstruction to achieve high spatial resolution.
A comprehensive evaluation of imaging system design requires systematic investigations that extend beyond this illustrative case study. Potential directions include: (1) assessing the trade-off between SNR and spatial resolution when enlarging the transducer detection surface, (2) quantifying estimation accuracy and imaging depth when applying JR methods that account for acoustic heterogeneity and attenuation, and (3) evaluating figures of merit for specific clinical tasks, such as signal detection and lesion localization, rather than relying solely on conventional image quality measures. These investigations can be effectively pursued by use of the proposed virtual imaging framework.
6. Discussion and conclusion
In this work, a comprehensive framework was established to enable virtual imaging studies of 3D qOAT of the breast. Building upon the previously proposed SOA-NBP tool, the framework incorporates a two-layer skin model that captures variations in skin tone. It extends the lesion modeling by including two benign lesion types, fibroadenomas and simple cysts, in addition to malignant tumors. By use of stochastically generated NBPs, the framework enables end-to-end simulation of the entire imaging process, from optoacoustically induced initial pressure generation to acoustic wave propagation and optoacoustic measurement acquisition. Additional flexibility in light source modeling was incorporated, and the transducer characteristics, represented by the SIR and the EIR, were explicitly modeled to faithfully replicate imaging physics under realistic system conditions. To accelerate further research, the datasets generated in this study will be made available on request. The framework is broadly applicable, with potential use cases extending beyond OAT. For instance, NBPs generated employing the SOA-NBP tool have already been utilized in optical imaging studies [33]. The present framework further provides a platform for systematic investigations of system designs and reconstruction methods in both optical and acoustic inversion.
Previous breast phantoms used in OAT typically exhibited oversimplified anatomical structures and did not support stochastic generation to represent population-level variability [7], [24], [25], [26], [28]. The previous SOA-NBP, on which the present work builds, focused on generating anatomically and physiologically plausible breast phantoms through stochastic modeling. Unlike previous breast phantoms that independently assigned functional properties and optical absorption coefficients, the SOA-NBP defined the optical absorption coefficient as a function of the concentrations of major chromophores, thereby enabling qOAT studies under physiologically realistic conditions. A detailed comparison between the SOA-NBP and previous breast phantoms is provided in [5]. The present work extends the SOA-NBP framework beyond object modeling. In addition to introducing further object features, considerable effort was directed toward enhancing the simulation pipeline, as described in Sections 3.2, 3.3. These improvements enable more realistic and scalable full-scale 3D OAT simulations, broadening the applicability of the framework for virtual imaging studies.
While the existing simulation toolkit SIMPA [51] provides a general-purpose simulation environment for optoacoustic imaging and supports object models constructed from basic geometric primitives, as described in Section 2.3, the object modeling component included in the proposed framework was specifically designed for breast OAT and incorporates dedicated functionality for generating anatomically and physiologically plausible NBPs. SIMPA also provides predefined parameter sets as representative configurations of existing imaging systems and interfaces with existing simulation tools, i.e., MCX and k-Wave. In contrast, the proposed framework incorporates several extensions to the simulation pipeline. Specifically, it introduces additional flexibility in light source modeling beyond what is available in the original MCX implementation. Moreover, the acoustic simulation component incorporates a more memory-efficient implementation of the transducer SIR, enabling SIR-based simulations for large whole-breast domains to be performed in a single run, whereas the corresponding k-Wave implementation may require multiple runs under GPU memory constraints. In addition, explicit modeling of the EIR of the transducer is incorporated.
The tissue property values used in the SOA-NBP are based on measurements reported in the literature. As more clinical or experimental data on the functional, optical, and acoustic properties of breast tissues become available, they can be incorporated within the proposed framework to further refine the realism of the NBPs, constrain the sampling process, and better capture correlations among tissue properties within each tissue type. This study demonstrated virtual imaging using two example spherical-type (hemispherical or partial cap) tomographic system configurations. Nevertheless, the proposed framework is not limited to these configurations and can be applied to other system designs with different illumination conditions and measurement geometries, provided that the system specifications are available. As an example, the framework can be applied to handheld probe-based imaging systems employing a linear transducer array, for which the ability to account for the SIR of individual transducer elements is particularly relevant. Moreover, elevation-focused transducer characteristics can be modeled by incorporating lens-induced focusing effects using the approach described in [86].
Experimental validation of the proposed virtual imaging pipeline using physical phantoms is currently underway as part of an ongoing study, with preliminary results presented in a recent publication [87]. The current NBPs do not yet incorporate lesion-dependent vascular features, such as the distinct spatial distribution and connectivity of blood vessels typically observed in OAT images around malignant lesions compared with benign ones. In addition, some existing breast OAT systems employ slight breast compression, which is not currently modeled in the framework. Extending the framework to incorporate these features is an area for future work.
CRediT authorship contribution statement
Seonyeong Park: Writing – original draft, Visualization, Software, Methodology, Investigation, Formal analysis, Data curation, Conceptualization. Gangwon Jeong: Software. Umberto Villa: Writing – review & editing, Software, Funding acquisition. Mark A. Anastasio: Writing – review & editing, Supervision, Resources, Funding acquisition, Conceptualization.
Declaration of competing interest
The authors declare the following financial interests/personal relationships which may be considered as potential competing interests: Mark A. Anastasio reports financial support was provided by National Institutes of Health. Umberto Villa reports financial support was provided by National Institutes of Health. If there are other authors, they declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.
Acknowledgments
This work was supported in part by the National Institutes of Health, United States grants EB031585 and EB034261. This work used the Delta system at the National Center for Supercomputing Applications through allocation MDE230007 from the Advanced Cyberinfrastructure Coordination Ecosystem: Services & Support (ACCESS) program, which is supported by U.S. National Science Foundation grants #2138259, #2138286, #2138307, #2137603, and #2138296. The authors would like to thank Drs. Mohammad Eghtedari and André Kajdacsy-Balla for their consultation on the numerical modeling of lesions to ensure realistic representation of their pathological and functional properties. They are also grateful to Dr. Alexander A. Oraevsky for insightful discussions on EIR modeling and for providing EIR measurement data, and to Dr. Richard Su for collecting the EIR measurement data.
Biographies

Seonyeong Park received the B.S. degree in Electronics, Computer, and Telecommunication Engineering and the M.S. degree in Information and Communications Engineering from Pukyong National University, Busan, Korea, in 2011 and 2013, respectively. She earned the Ph.D. degree in Electrical and Computer Engineering from Virginia Commonwealth University, Richmond, VA, USA, in 2017. She is currently a Research Assistant Professor in the Department of Bioengineering at the University of Illinois Urbana-Champaign, Urbana, IL, USA. Her research interests include photoacoustic computed tomography, numerical modeling of biomedical objects, image reconstruction, and inverse problems.

Gangwon Jeong is currently a Postdoctoral Research Associate in the Mallinckrodt Institute of Radiology at Washington University in St. Louis. He received his Ph.D. in Bioengineering from the University of Illinois at Urbana-Champaign. His research focuses on medical image reconstruction, particularly in ultrasound and photoacoustic computed tomography.

Umberto Villa received the B.S. and M.S. degrees in Mathematical Engineering from the Politecnico di Milano, Milan, Italy, in 2005 and 2007, respectively, and the Ph.D. degree in Mathematics from Emory University, Atlanta, GA, USA, in 2012. He is an Assistant Professor of Biomedical Engineering at The University of Texas at Austin, Austin, TX, USA. His research interests lie in the computational and mathematical aspects of large-scale inverse problems, imaging science, and uncertainty quantification.

Mark A. Anastasio received the Ph.D. degree from The University of Chicago, Chicago, IL, USA, in 2001. He is currently the Mallinckrodt Endowed Professor of Imaging Sciences, Siteman Cancer Center Pedal the Cause Professor, Vice Chair of Imaging Sciences and AI Research, and Director of both the Computational Imaging Research Center and the Center for Computationally and AI-Enabled Imaging Sciences, Washington University in St. Louis, St. Louis, MO, USA. His research broadly addresses computational image science, inverse problems in imaging, and machine learning for imaging applications. He has contributed broadly to emerging biomedical imaging technologies that include photoacoustic computed tomography, ultrasound computed tomography, and X-ray phase-contrast imaging. His work has been supported by numerous NIH grants and he served for two years as the Chair of NIH EITA Study Section. Dr. Anastasio is a fellow of the Society of Photo-Optical Instrumentation Engineers (SPIE), American Institute for Medical and Biological Engineering (AIMBE), the International Academy of Medical and Biological Engineering (IAMBE), and the Institute of Electrical and Electronics Engineers (IEEE).
Footnotes
This article is part of a Special issue entitled: ‘Photoacoustic Phantoms’ published in Photoacoustics.
Given their role as Co-Guest Editor, Mark Anastasio had no involvement in the peer review of this article and has no access to information regarding its peer-review. Full responsibility for the editorial process for this article was delegated to another journal editor.
Contributor Information
Seonyeong Park, Email: sp33@illinois.edu.
Gangwon Jeong, Email: gangwon@wustl.edu.
Umberto Villa, Email: uvilla@austin.utexas.edu.
Mark A. Anastasio, Email: anastasio@wustl.edu, https://anastasiolab.wustl.edu/.
Appendix A. Existing simulation tools for virtual optoacoustic imaging
For high-fidelity simulation of photon transport in biological tissue, several MC simulation tools are available, including GPU-accelerated simulators such as MCX and MMC (implemented in C++ and CUDA) and MATLAB toolboxes such as MCmatlab and Valo MC. MCX and MMC also provide MATLAB interfaces, known as MCXLAB [49], [50] and MMCLAB [88], respectively. MCX [49], [50] and MCmatlab [89] use Cartesian grids, enabling simulation of light transport in media with voxel-level heterogeneity. MMC [88] and Valo MC [90] employ finite tetrahedral elements, supporting modeling of element-level heterogeneity. These tools provide various light source models. ValoMC includes direct, cosine, Gaussian, isotropic, and pencil light source types. MCmatlab supports pencil, isotropic (line or point), infinite plane wave, Laguerre-Gaussian LG01, radial-factorizable (e.g., Gaussian), and X/Y factorizable (e.g., rectangular LED emitter) source models. MCX and MMC provide pencil, isotropic, uniform cone, collimated Gaussian, one-sheeted hyperboloid Gaussian, uniform planar, uniform disk, and patterned source types.
The most widely used acoustic wave simulator in the field of optoacoustic imaging is the MATLAB toolbox k-Wave [32], which solves the acoustic wave equation using the -space pseudo-spectral time-domain method. It supports arbitrarily specified (i.e., off-grid) transducer positions to reduce mismatch with actual transducer placement [77] and allows modeling of finite-sized transducers by incorporating their SIR. A corresponding Python library, j-Wave [76], has also been developed as a JAX-based differentiable acoustic simulator, but it currently lacks acoustic attenuation modeling and does not support either off-grid transducer positioning or finite-sized transducer modeling.
Appendix B. Tissue label definitions for SOA-NBP
In anatomical NBPs, tissue types were labeled with unsigned 8-bit integers. In both the prior SOA-NBP and the present extension, tissue types corresponding to those in the VICTRE NBP [21] retained their original label values, whereas newly introduced tissue types were assigned values not used in VICTRE to ensure compatibility with the existing label definitions.
Voxels labeled as skin with a value of 2 in VICTRE and the previous SOA-NBP were assigned to the dermis-only layer. The epidermis-included layer was re-labeled to a value of 3, which is unused in VICTRE. In VICTRE, the tissue label value 200 denotes the cancerous mass; this value was retained for the VTC region in the malignant tumor models of SOA-NBP. The necrotic core, peripheral angiogenesis region, fibroadenoma, and simple cyst were assigned label values of 210, 190, 180, and 160, respectively. These values are unused in VICTRE [21]. Tissue type labels are summarized in Table 2, with newly introduced types in SOA-NBP shown in bold.
Table 2.
Tissue type labels.
| Tissue type | Label |
|---|---|
| Fat | 1 |
| Dermis-only | 2 |
| Epidermis-included | 3 |
| Glandular | 29 |
| Nipple | 33 |
| Ligament | 88 |
| Terminal duct lobular unit | 95 |
| Duct | 125 |
| Artery | 150 |
| Simple cyst | 160 |
| Fibroadenoma | 180 |
| Peripheral angiogenesis | 190 |
| Viable tumor cell | 200 |
| Necrotic core | 210 |
| Vein | 225 |
Table 3.
Lesion shape parameters.
| Parameter | VTC region | Fibroadenoma | Cyst |
|---|---|---|---|
| base.complexity | 1 | 0.8 | 0.5 |
| mass.lMax | 4 | 7 | 5 |
| mass.meanSigma2 | 0.31 | 0.2 | 0.03 |
| mass.stdSigma2 | 0.04 | 0.04 | 0.01 |
| mass.powerLaw | 4 | 5 | 5 |
| mass.meanLF | 611.2 | 30 | 10 |
| mass.stdLF | 70.6 | 8 | 3 |
| mass.meanShape | 0.36 | 0.75 | 1 |
| mass.stdShape | 0.48 | 0.8 | 0 |
| mass.meanLFRad | 0.299 | 0.2 | 0 |
| mass.stdLFRad | 0.073 | 0.06 | 0 |
| mass.meanLFLen | 0.113 | 0.1 | 0 |
| mass.stdLFLen | 0.021 | 0.025 | 0 |
| mass.meanFuzzAlpha | 0.015 | 0.01 | 0 |
| mass.stdFuzzAlpha | 0 | 0 | 0 |
| spicule.meanInitial | 100 | 0 | 0 |
| spicule.stdInitial | 365 | 0 | 0 |
| spicule.meanNeigh | 8.98 | NA | NA |
| spicule.stdNeigh | 1.89 | NA | NA |
| spicule.meanInitRad | 0.1 | NA | NA |
| spicule.stdInitRad | 0.2 | NA | NA |
| spicule.meanRadDec | 0.33 | NA | NA |
| spicule.stdRadDec | 0.25 | NA | NA |
| spicule.meanInitLen | 0.2 | NA | NA |
| spicule.stdInitLen | 0.2 | NA | NA |
| spicule.meanLenDec | 0.4 | NA | NA |
| spicule.stdLenDec | 0.3 | NA | NA |
| spicule.meanContProb | 0.717 | NA | NA |
| spicule.stdContProb | 0.057 | NA | NA |
| spicule.meanBranchAng | 6.55 | NA | NA |
| spicule.stdBranchAng | 0.62 | NA | NA |
Appendix C. Example parameters for lesion shape modeling
Anatomical NLPs for modeling malignant tumors, fibroadenomas, and simple cysts can be generated using the VICTRE tool with a configuration file [21]. In this file, the voxel size parameter imgRes (mm) and the lesion radius parameter alpha (mm) are specified by users. Table 3 lists example VICTRE tool parameter values used in this study to model the VTC region of malignant tumors, fibroadenomas, and simple cysts. These values control the overall geometry, boundary irregularity, and surface characteristics of each lesion type. For the VTC region, the following parameters were modified from the default VICTRE NLP configuration: spicule.meanInitial, spicule.stdInitial, spicule.meanInitRad, spicule.stdInit- Rad, spicule. meanRadDec, spicule.stdRadDec, spicule.mean- InitLen, spicule.stdInitLen, and spicule.meanLenDec. All other parameters remain at their default values.
Appendix D. Comparison between coarse-grid and fine-grid photon transport simulations
To evaluate the computational efficiency of the coarse-grid photon transport simulation with upsampling described in Section 3.2 and its impact on the simulated acoustic pressure data, a comparison with the standard fine-grid simulation was conducted. Simulations were performed using System A described in Section 4.2 and the setup in Section 4.3 at an illumination wavelength of 757 nm. Photon transport simulations were conducted on an NVIDIA Tesla V100 GPU with 32 GB of memory.
As summarized in Table 4, the coarse-grid photon transport simulation with upsampling required approximately 37.85 min per scan and 868 MB of GPU memory per light source simulation, whereas the fine-grid simulation required about 817.48 min and 2442 MB of GPU memory. In both cases, the number of photons was chosen to be significantly higher than the number of voxels in the corresponding simulation domain to obtain statistically converged Monte Carlo estimates [49]. To assess the impact of the proposed approach on the simulated pressure data, acoustic pressure signals were generated from the resulting initial pressure distributions. EIR modeling and noise addition were not included. A normalized root-mean-square error (NRMSE) was computed across all transducer locations and time samples, with normalization performed using the root-mean-square of the reference pressure signal from the fine-grid photon transport simulation. The coarse-grid simulation with upsampling achieved a 21-fold reduction in computation time while introducing an NRMSE of only 3.62%. This corresponds to an equivalent error SNR of 28.84 dB, which exceeds the pressure-data SNRs used to model electronic noise in the numerical studies (5 dB).
Table 4.
Comparison between the coarse-grid photon transport simulation with upsampling and fine-grid photon transport simulation.
| Parameter | Coarse-grid photon transport simulation with upsampling | Fine-grid photon transport simulation (reference) |
|---|---|---|
| Voxel size (mm) | 0.5 | 0.25 |
| Domain dimension (voxels) | 340 × 340 × 170 | 680 × 680 × 340 |
| Number of photons per light source | 108 | 109 |
| GPU memory (MB) | 868 | 2442 |
| Computation time (min) | 37.85 | 817.48 |
Appendix E. Complex exponential parameterization of the EIR
The EIR was modeled employing the complex exponential method [74], which represents transient and deterministic signals as a finite sum of complex exponential functions. The model is parameterized by , where is the number of damped cosine components, and , , , and denote the amplitude, damping ratio, frequency, and phase of the th component, respectively. Using this parameterization, the modeled EIR, denoted as and represented as a discretized time-domain vector of length with sampling interval , is expressed as [74]:
| (4) |
The parameter set was estimated from the measured EIR using least squares [74]. The model order was selected using AIC to balance model complexity with accuracy. In this study, was chosen as the value of within the range that minimized AIC [74].
Data availability
Data will be made available on request.
References
- 1.Oraevsky A.A., Karabutov A.A., Solomatin S.V., Savateeva E.V., Andreev V.A., Gatalica Z., Singh H., Fleming R.D. In: Biomedical Optoacoustics II. Oraevsky A.A., editor. Vol. 4256. SPIE, International Society for Optics and Photonics; 2001. Laser optoacoustic imaging of breast cancer in vivo; pp. 6–15. [DOI] [Google Scholar]
- 2.Oraevsky A., Clingman B., Zalev J., Stavros A., Yang W., Parikh J. Clinical optoacoustic imaging combined with ultrasound for coregistered functional and anatomical mapping of breast tumors. Photoacoustics. 2018;12:30–45. doi: 10.1016/j.pacs.2018.08.003. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Oraevsky A., Su R., Nguyen H., Moore J., Lou Y., Bhadra S., Forte L., Anastasio M., Yang W. In: Photons Plus Ultrasound: Imaging and Sensing 2018. Oraevsky A.A., Wang L.V., editors. Vol. 10494. SPIE, International Society for Optics and Photonics; 2018. Full-view 3D imaging system for functional and anatomical screening of the breast; pp. 217–226. [DOI] [Google Scholar]
- 4.Lin L., Hu P., Shi J., Appleton C.M., Maslov K., Li L., Zhang R., Wang L.V. Single-breath-hold photoacoustic computed tomography of the breast. Nat. Commun. 2018;9(2352):1–9. doi: 10.1038/s41467-018-04576-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Park S., Villa U., Li F., Cam R.M., Oraevsky A.A., Anastasio M.A. Stochastic three-dimensional numerical phantoms to enable computational studies in quantitative optoacoustic computed tomography of breast cancer. J. Biomed. Opt. 2023;28(6) doi: 10.1117/1.JBO.28.6.066002. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Park S., Brooks F.J., Villa U., Su R., Anastasio M.A., Oraevsky A.A. Normalization of optical fluence distribution for three-dimensional functional optoacoustic tomography of the breast. J. Biomed. Opt. 2022;27(3) doi: 10.1117/1.JBO.27.3.036001. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Lou Y., Zhou W., Matthews T.P., Appleton C.M., Anastasio M.A. Generation of anatomically realistic numerical phantoms for photoacoustic and ultrasonic breast imaging. J. Biomed. Opt. 2017;22(4):1–13. doi: 10.1117/1.JBO.22.4.041015. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Poudel J., Lou Y., Anastasio M.A. A survey of computational frameworks for solving the acoustic inverse problem in three-dimensional photoacoustic computed tomography. Phys. Med. Biol. 2019;64(14):14TR01. doi: 10.1088/1361-6560/ab2017. [DOI] [PubMed] [Google Scholar]
- 9.Chen P., Park S., Cam R.M., Huang H.-K., Villa U., Anastasio M.A. In: Photons Plus Ultrasound: Imaging and Sensing 2024. Oraevsky A.A., Wang L.V., editors. Vol. 12842. SPIE, International Society for Optics and Photonics; 2024. Learning a semi-analytic reconstruction method for photoacoustic computed tomography with hemispherical measurement geometries. [DOI] [Google Scholar]
- 10.Chen P., Park S., Jeong G., Cam R.M., Huang H.-K., Villa U., Anastasio M.A. In: Photons Plus Ultrasound: Imaging and Sensing 2025. Oraevsky A.A., Wang L.V., editors. Vol. 13319. SPIE, International Society for Optics and Photonics; 2025. Benchmarking deep learning-based reconstruction in photoacoustic computed tomography with clinically relevant synthetic datasets. [DOI] [Google Scholar]
- 11.Jeong G., Villa U., Anastasio M.A. Revisiting the joint estimation of initial pressure and speed-of-sound distributions in photoacoustic computed tomography with consideration of canonical object constraints. Photoacoustics. 2025;43 doi: 10.1016/j.pacs.2025.100700. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Yao J., Xia J., Wang L.V. Multiscale functional and molecular photoacoustic tomography. Ultrason. Imaging. 2016;38(1):44–62. doi: 10.1177/0161734615584312. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Tarvainen T., Cox B. Quantitative photoacoustic tomography: modeling and inverse problems. J. Biomed. Opt. 2023;29(S1):S11509. doi: 10.1117/1.JBO.29.S1.S11509. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Cam R.M., Park S., Villa U., Anastasio M.A. In: Photons Plus Ultrasound: Imaging and Sensing 2024. Oraevsky A.A., Wang L.V., editors. Vol. 12842. SPIE, International Society for Optics and Photonics; 2024. Investigation of a learned image reconstruction method for three-dimensional quantitative photoacoustic tomography of the breast; p. 128420M. [DOI] [Google Scholar]
- 15.Toi M., Asao Y., Matsumoto Y., Sekiguchi H., Yoshikawa A., Takada M., Kataoka M., Endo T., Kawaguchi-Sakita N., Kawashima M., Fakhrejahani E., Kanao S., Yamaga I., Nakayama Y., Tokiwa M., Torii M., Yagi T., Sakurai T., Togashi K., Shiina T. Visualization of tumor-related blood vessels in human breast by photoacoustic imaging system with a hemispherical detector array. Sci. Rep. 2017;7(41970):1–11. doi: 10.1038/srep41970. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Neuschler E.I., Butler R., Young C.A., Barke L.D., Bertrand M.L., Böhm-Vélez M., Destounis S., Donlan P., Grobmyer S.R., Katzen J., Kist K.A., Lavin P.T., Makariou E.V., Parris T.M., Schilling K.J., Tucker F.L., Dogan B.E. A pivotal study of optoacoustic imaging to diagnose benign and malignant breast masses: A new evaluation tool for radiologists. Radiology. 2018;287(2):398–412. doi: 10.1148/radiol.2017172228. PMID: 29178816. [DOI] [PubMed] [Google Scholar]
- 17.Dogan B.E., Menezes G.L.G., Butler R.S., Neuschler E.I., Aitchison R., Lavin P.T., Tucker F.L., Grobmyer S.R., Otto P.M., Stavros A.T. Optoacoustic imaging and gray-scale US features of breast cancers: Correlation with molecular subtypes. Radiology. 2019;292(3):564–572. doi: 10.1148/radiol.2019182071. PMID: 31287388. [DOI] [PubMed] [Google Scholar]
- 18.Park S., Villa U., Oraevsky A., Anastasio M. In: Photons Plus Ultrasound: Imaging and Sensing 2023. Oraevsky A.A., Wang L.V., editors. Vol. PC12379. SPIE, International Society for Optics and Photonics; 2023. Numerical investigation of impact of skin phototype on three-dimensional optoacoustic tomography of the breast. [DOI] [Google Scholar]
- 19.Li F., Villa U., Park S., Anastasio M.A. 3-D stochastic numerical breast phantoms for enabling virtual imaging trials of ultrasound computed tomography. IEEE Trans. Ultrason. Ferroelectr. Freq. Control. 2022;69(1):135–146. doi: 10.1109/TUFFC.2021.3112544. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Abadi E., Segars W.P., Tsui B.M.W., Kinahan P.E., Bottenus N., Frangi A.F., Maidment A., Lo J., Samei E. Virtual clinical trials in medical imaging: a review. J. Med. Imaging. 2020;7(4) doi: 10.1117/1.JMI.7.4.042805. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Badano A., Graff C.G., Badal A., Sharma D., Zeng R., Samuelson F.W., Glick S.J., Myers K.J. Evaluation of digital breast tomosynthesis as replacement of Full-Field digital mammography using an in silico imaging trial. JAMA Netw. Open. 2018;1(7) doi: 10.1001/jamanetworkopen.2018.5474. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Myers K.J. In: Medical Imaging 2023: Physics of Medical Imaging. Yu L., Fahrig R., Sabol J.M., editors. Vol. 12463. SPIE, International Society for Optics and Photonics; 2023. State of the art in the task-based assessment of medical imaging systems. [DOI] [Google Scholar]
- 23.Abadi E., Barufaldi B., Lago M., Badal A., Mello-Thoms C., Bottenus N., Wangerin K.A., Goldburgh M., Tarbox L., Beaucage-Gauvreau E., Frangi A.F., Maidment A., Kinahan P.E., Bosmans H., Samei E. Toward widespread use of virtual trials in medical imaging innovation and regulatory science. Med. Phys. 2024;51(12):9394–9404. doi: 10.1002/mp.17442. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Akhlaghi N., Pfefer T.J., Wear K.A., Garra B.S., Vogt W.C. Multidomain computational modeling of photoacoustic imaging: Verification, validation, and image quality prediction. J. Biomed. Opt. 2019;24(12) doi: 10.1117/1.JBO.24.12.121910. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Fadden C., Kothapalli S.R. A single simulation platform for hybrid photoacoustic and RF-Acoustic computed tomography. Appl. Sci. (Basel) 2018;8(9):1568. doi: 10.3390/app8091568. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Sowmiya C., Thittai A.K. Medical Imaging 2017: Biomedical Applications in Molecular, Structural, and Functional Imaging. Vol. 10137. 2017. Simulation of photoacoustic tomography (PAT) system in COMSOL(r) and comparison of two popular reconstruction techniques; p. 101371O. (Proc. SPIE). [DOI] [Google Scholar]
- 27.Cam R.M., Park S., Villa U., Anastasio M.A. Application of a virtual imaging framework for investigating a deep learning-based reconstruction method for 3D quantitative photoacoustic computed tomography. Photoacoustics. 2026;48 doi: 10.1016/j.pacs.2025.100792. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Ma Y., Yang C., Zhang J., Wang Y., Gao F., Gao F. Human breast numerical model generation based on deep learning for photoacoustic imaging. 2020 42nd Annual International Conference of the IEEE Engineering in Medicine & Biology Society; EMBC; 2020. pp. 1919–1922. [DOI] [PubMed] [Google Scholar]
- 29.Bao Y., Deng H., Wang X., Zuo H., Ma C. Development of a digital breast phantom for photoacoustic computed tomography. Biomed. Opt. Express. 2021;12(3):1391–1406. doi: 10.1364/BOE.416406. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Zhou W., Bhadra S., Brooks F.J., Li H., Anastasio M.A. Learning stochastic object models from medical imaging measurements by use of advanced ambient generative adversarial networks. J. Med. Imaging. 2022;9(1) doi: 10.1117/1.JMI.9.1.015503. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Yang K., Park S., Huang H.-K., Villa U., Anastasio M.A. In: Photons Plus Ultrasound: Imaging and Sensing 2025. Oraevsky A.A., Wang L.V., editors. Vol. 13319. SPIE, International Society for Optics and Photonics; 2025. Compensating for the spatial impulse response in 3D photoacoustic computed tomography using a learned data space restoration method. [DOI] [Google Scholar]
- 32.Treeby B.E., Cox B.T. K-Wave: MATLAB toolbox for the simulation and reconstruction of photoacoustic wave fields. J. Biomed. Opt. 2010;15(2):1–12. doi: 10.1117/1.3360308. [DOI] [PubMed] [Google Scholar]
- 33.Hossain M.I., Xue M., Wang L., Zhu Q. DOT-AE-GAN: a hybrid autoencoder–GAN model for enhanced ultrasound-guided diffuse optical tomography reconstruction. J. Biomed. Opt. 2025;30(7) doi: 10.1117/1.JBO.30.7.076003. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Suhonen M., Lucka F., Pulkkinen A., Arridge S., Cox B., Tarvainen T. Reconstructing multiple initial pressure and speed of sound distributions simultaneously in photoacoustic tomography. Photoacoustics. 2025;45 doi: 10.1016/j.pacs.2025.100748. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Song Y., Song Q., Tian C. In: Fourth Conference on Biomedical Photonics and Cross-Fusion (BPC 2025) Qu J., Wei X., editors. Vol. 13785. SPIE, International Society for Optics and Photonics; 2025. A multi-grid method for joint reconstruction of initial pressure and speed of sound in photoacoustic tomography. [DOI] [Google Scholar]
- 36.Bader M., Longo A., Jüstel D., Ntziachristos V. Low frequency detection in clinical multispectral optoacoustic tomography. Photoacoustics. 2025;46 doi: 10.1016/j.pacs.2025.100785. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Jeong G., Park S., Villa U., Anastasio M.A. In: Photons Plus Ultrasound: Imaging and Sensing 2026. Oraevsky A.A., Wang L.V., editors. Vol. 13851. SPIE, Interational Society for Optics and Photonics; 2026. Joint reconstruction of initial pressure and sound speed distributions in 3D photoacoustic computed tomography using object constraints. [DOI] [Google Scholar]
- 38.Yang K., Park S., Jeong G., Liu C.Z., Tong X., Cui M., Ku G., Villa U.E., Wang L.V., Anastasio M.A. In: Photons Plus Ultrasound: Imaging and Sensing 2026. Oraevsky A.A., Wang L.V., editors. Vol. 13851. SPIE, International Society for Optics and Photonics; 2026. Application of a learned SIR compensation method to improve elevational resolution in ring-array-based photoacoustic computed tomography. [DOI] [Google Scholar]
- 39.Park S., Villa U., Li F., Cam R.M., Oraevsky A.A., Anastasio M.A. 2023. Stochastic optoacoustic numerical breast phantom (SOA-NBP): A python library to generate anatomically and physiologically realistic numerical breast phantoms for OAT virtual imaging studies. URL https://github.com/comp-imaging-sci/soa-nbp. [Google Scholar]
- 40.American College of Radiology . American College of Radiology; Reston, VA, USA: 2013. ACR BI-RADS® Atlas, Breast Imaging Reporting and Data System. [Google Scholar]
- 41.Jacques S.L. Optical properties of biological tissues: A review. Phys. Med. Biol. 2013;58(11):R37–61. doi: 10.1088/0031-9155/58/11/R37. [DOI] [PubMed] [Google Scholar]
- 42.S.A. Prahl, Tabulated Molar Extinction Coefficient for Hemoglobin in Water, URL http://omlc.org/spectra/hemoglobin/summary.html.
- 43.Popel A.S. Theory of oxygen transport to tissue. Crit. Rev. Biomed. Eng. 1989;17(3):257–321. [PMC free article] [PubMed] [Google Scholar]
- 44.Logg A., Mardal K.-A., Wells G. Automated solution of differential equations by the finite element method: The FEniCS book. Springer Science & Business Media; 2012. [Google Scholar]
- 45.Wang L., Jacques S.L., Zheng L. MCML—Monte Carlo modeling of light transport in multi-layered tissues. Comput. Methods Programs Biomed. 1995;47(2):131–146. doi: 10.1016/0169-2607(95)01640-F. [DOI] [PubMed] [Google Scholar]
- 46.Szabo T.L. Academic Press; 2013. Diagnostic Ultrasound Imaging: Inside Out. [Google Scholar]
- 47.Treeby B.E., Cox B.T. Modeling power law absorption and dispersion for acoustic propagation using the fractional Laplacian. J. Acoust. Soc. Am. 2010;127(5):2741–2748. doi: 10.1121/1.3377056. [DOI] [PubMed] [Google Scholar]
- 48.Tabei M., Mast T.D., Waag R.C. A k-space method for coupled first-order acoustic propagation equations. J. Acoust. Soc. Am. 2002;111(1):53–63. doi: 10.1121/1.1421344. [DOI] [PubMed] [Google Scholar]
- 49.Fang Q., Boas D.A. Monte Carlo simulation of photon migration in 3D turbid media accelerated by graphics processing units. Opt. Express. 2009;17(22):20178–20190. doi: 10.1364/OE.17.020178. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 50.Yu L., Nina-Paravecino F., Kaeli D., Fang Q. Scalable and massively parallel Monte Carlo photon transport simulations for heterogeneous computing platforms. J. Biomed. Opt. 2018;23(1):1–4. doi: 10.1117/1.Jbo.23.1.010504. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 51.Gröhl J., Dreher K.K., Schellenberg M., Rix T., Holzwarth N., Vieten P., Ayala L., Bohndiek S.E., Seitel A., Maier-Hein L. SIMPA: An open-source toolkit for simulation and image processing for photonics and acoustics. J. Biomed. Opt. 2022;27(8) doi: 10.1117/1.JBO.27.8.083010. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 52.Gröhl J., Kunyansky L., Poimala J., Else T.R., Di Cecio F., Bohndiek S.E., Cox B.T., Hauptmann A. Digital twins enable full-reference quality assessment of photoacoustic image reconstructions. J. Acoust. Soc. Am. 2025;158(1):590–601. doi: 10.1121/10.0037188. [DOI] [PubMed] [Google Scholar]
- 53.Jacques S.L. 1998. Skin optics. Oregon Medical Laser Center News, January 1998, https://omlc.org/news/jan98/skinoptics.html. [Google Scholar]
- 54.Else T.R., Hacker L., Gröhl J., Bunce E.V., Tao R., Bohndiek S.E. Effects of skin tone on photoacoustic imaging and oximetry. J. Biomed. Opt. 2023;29(S1):S11506. doi: 10.1117/1.JBO.29.S1.S11506. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 55.Rasquinha R.D., Gubbi M.R., Bell M.A.L. Impact of skin tone on target size detectability in photoacoustic breast imaging. Biophotonics Discov. 2024;2(1) doi: 10.1117/1.BIOS.2.1.012502. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 56.Ulger H., Erdogan N., Kumanlioglu S., Unur E. Effect of age, breast size, menopausal and hormonal status on mammographic skin thickness. Ski. Res. Technol. 2003;9(3):284–289. doi: 10.1034/j.1600-0846.2003.00027.x. [DOI] [PubMed] [Google Scholar]
- 57.Huang S.-Y., Boone J.M., Yang K., Kwan A.L.C., Packard N.J. The effect of skin thickness determined using breast CT on mammographic dosimetry. Med. Phys. 2008;35(4):1199–1206. doi: 10.1118/1.2841938. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 58.Fitzpatrick T.B. The validity and practicality of sun-reactive skin types I through VI. Arch. Dermatol. 1988;124(6):869–871. doi: 10.1001/archderm.1988.01670060015008. [DOI] [PubMed] [Google Scholar]
- 59.Tseng S.-H., Bargo P., Durkin A., Kollias N. Chromophore concentrations, absorption and scattering properties of human skin in-vivo. Opt. Express. 2009;17(17):14599–14617. doi: 10.1364/OE.17.014599. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 60.Ramala S.R., Chandak S., Chandak M.S., Annareddy S. A comprehensive review of breast fibroadenoma: Correlating clinical and pathological findings. Cureus. 2023;15(12) doi: 10.7759/cureus.49948. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 61.Zhu Q., Cronin E.B., Currier A.A., Vine H.S., Huang M., Chen N., Xu C. Benign versus malignant breast masses: Optical differentiation with US-guided optical imaging reconstruction. Radiology. 2005;237(1):57–66. doi: 10.1148/radiol.2371041236. PMID: 16183924. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 62.Berg W.A., Sechtin A.G., Marques H., Zhang Z. Cystic breast masses and the ACRIN 6666 experience. Radiol. Clin. North Am. 2010;48(5):931–987. doi: 10.1016/j.rcl.2010.06.007. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 63.Ahrens J., Geveci B., Law C. In: Visualization Handbook. Hansen C.D., Johnson C.R., editors. Elsevier Butterworth-Heinemann; Burlington, MA, USA: 2005. ParaView: An End-User tool for large data visualization; pp. 717–731. [DOI] [Google Scholar]
- 64.Santi B.D., Lucka F., Bulthuis R.F., Veltman J., van der Schaaf M., Messal H., Brinkhuis M., Cox B., Manohar S. In: Photons Plus Ultrasound: Imaging and Sensing 2025. Oraevsky A.A., Wang L.V., editors. Vol. 13319. SPIE, International Society for Optics and Photonics; 2025. Image analysis of patient data from the hybrid photoacoustic-ultrasound tomography system PAM3. [DOI] [Google Scholar]
- 65.Hines N., Slanetz P.J., Eisenberg R.L. Cystic masses of the breast. Am. J. Roentgenol. 2010;194(2):W122–W133. doi: 10.2214/AJR.09.3688. PMID: 20093563. [DOI] [PubMed] [Google Scholar]
- 66.Choe R., Konecky S.D., Corlu A., Lee K., Durduran T., Jr. D.R.B., Pathak S., Czerniecki B.J., Tchou J.C., Fraker D.L., DeMichele A., Chance B., Arridge S.R., Schweiger M., Culver J.P., Schnall M.D., Putt M.E., Rosen M.A., Yodh A.G. Differentiation of benign and malignant breast tumors by in-vivo three-dimensional parallel-plate diffuse optical tomography. J. Biomed. Opt. 2009;14(2) doi: 10.1117/1.3103325. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 67.Li C., Duric N., Littrup P., Huang L. In vivo breast sound-speed imaging with ultrasound tomography. Ultrasound Med. Biol. 2009;35(10):1615–1628. doi: 10.1016/j.ultrasmedbio.2009.05.011. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 68.Hale G.M., Querry M.R. Optical constants of water in the 200-nm to 200-m wavelength region. Appl. Opt. 1973;12(3):555–563. doi: 10.1364/AO.12.000555. [DOI] [PubMed] [Google Scholar]
- 69.Buiteveld H., Hakvoort J.H.M., Donze M. Optical properties of pure water. Ocean Opt. XII. 1994;2258:174–183. doi: 10.1117/12.190060. [DOI] [Google Scholar]
- 70.ToolBox E. 2004. Water - speed of sound vs. Temperature. URL https://www.engineeringtoolbox.com/sound-speed-water-d_598.html. [Google Scholar]
- 71.Hasgall P.A. 2018. IT’IS database for thermal and electromagnetic parameters of biological tissues. URL https://www.itis.swiss/database, Version 4.0. [DOI] [Google Scholar]
- 72.Liopo A., Su R., Oraevsky A.A. Melanin nanoparticles as a novel contrast agent for optoacoustic tomography. Photoacoustics. 2015;3(1):35–43. doi: 10.1016/j.pacs.2015.02.001. Special Issue: Contrast Agents for optoacoustic imaging and sensing: Design and biomedical applications. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 73.Dumani D.S., Yu A., Thompson W., Brecht H.-P., Ivanov V., Anastasio M.A., Cook J., Ermilov S.A., Emelianov S.Y. In: Photons Plus Ultrasound: Imaging and Sensing 2019. Oraevsky A.A., Wang L.V., editors. Vol. 10878. SPIE, International Society for Optics and Photonics; 2019. Preclinical small animal imaging platform providing co-registered 3D maps of photoacoustic response and fluorescence; pp. 396–404. [DOI] [Google Scholar]
- 74.Tallavo F., Cascante G., Pandey M. Ultrasonic transducers characterisation for evaluation of stiff geomaterials. Géotechnique. 2011;61(6):501–510. doi: 10.1680/geot.9.P.087. [DOI] [Google Scholar]
- 75.Kruger R.A., Kuzmiak C.M., Lam R.B., Reinecke D.R., Del Rio S.P., Steed D. Dedicated 3D photoacoustic breast imaging. Med. Phys. 2013;40(11) doi: 10.1118/1.4824317. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 76.Stanziola A., Arridge S.R., Cox B.T., Treeby B.E. J-wave: An open-source differentiable wave simulator. SoftwareX. 2023;22 doi: 10.1016/j.softx.2023.101338. [DOI] [Google Scholar]
- 77.Wise E.S., Cox B.T., Jaros J., Treeby B.E. Representing arbitrary acoustic source and sensor distributions in Fourier collocation methods) J. Acoust. Soc. Am. 2019;146(1):278–288. doi: 10.1121/1.5116132. [DOI] [PubMed] [Google Scholar]
- 78.Javaherian A., Setarehdan S.K. 2025. Full-waveform approximation of finite-sized acoustic Apertures: Forward and adjoint wavefields. [Google Scholar]
- 79.Conjusteau A., Ermilov S.A., Su R., Brecht H.-P., Fronheiser M.P., Oraevsky A.A. Measurement of the spectral directivity of optoacoustic and ultrasonic transducers with a laser ultrasonic source. Rev. Sci. Instrum. 2009;80(9) doi: 10.1063/1.3227836. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 80.Conjusteau A., Ermilov S.A., Brecht H.-P., Oraevsky A.A. In: Photons Plus Ultrasound: Imaging and Sensing 2010. Oraevsky A.A., Wang L.V., editors. Vol. 7564. SPIE, International Society for Optics and Photonics; 2010. Effects of optical energy distribution on the generation of acoustic plane waves. [DOI] [Google Scholar]
- 81.Wiener N. John Wiley & Sons; New York: 1949. Extrapolation, Interpolation, and Smoothing of Stationary Time Series, with Engineering Applications. [Google Scholar]
- 82.Van de Sompel D., Sasportas L.S., Jokerst J.V., Gambhir S.S. Comparison of deconvolution filters for photoacoustic tomography. PLoS One. 2016;11(3):1–28. doi: 10.1371/journal.pone.0152597. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 83.Xu M., Wang L.V. Universal back-projection algorithm for photoacoustic computed tomography. Phys. Rev. E. 2005;71 doi: 10.1103/PhysRevE.71.016706. [DOI] [PubMed] [Google Scholar]
- 84.Matthews T.P., Poudel J., Li L., Wang L.V., Anastasio M.A. Parameterized joint reconstruction of the initial pressure and sound speed distributions for photoacoustic computed tomography. SIAM J. Imaging Sci. 2018;11(2):1560–1588. doi: 10.1137/17M1153649. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 85.Kaipio J., Somersalo E. Statistical inverse problems: discretization, model reduction and inverse crimes. J. Comput. Appl. Math. 2007;198(2):493–504. [Google Scholar]
- 86.Li F., Villa U., Duric N., Anastasio M.A. A forward model incorporating elevation-focused transducer properties for 3-D full-waveform inversion in ultrasound computed tomography. IEEE Trans. Ultrason. Ferroelectr. Freq. Control. 2023;70(10):1339–1354. doi: 10.1109/TUFFC.2023.3313549. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 87.Park S., Su R., Luo H., Jeong G., Oraevsky A.A., Anastasio M.A. In: Photons Plus Ultrasound: Imaging and Sensing 2026. Oraevsky A.A., Wang L.V., editors. Vol. 13851. SPIE, International Society for Optics and Photonics; 2026. Experimental validation of a virtual imaging framework for three-dimensional functional optoacoustic tomography. [DOI] [Google Scholar]
- 88.Fang Q., Yan S. Graphics processing unit-accelerated mesh-based Monte Carlo photon transport simulations. J. Biomed. Opt. 2019;24(11) doi: 10.1117/1.JBO.24.11.115002. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 89.Marti D., Aasbjerg R.N.N., Andersen P.E.E., Hansen A.K.K. MCmatlab: An open-source, user-friendly, MATLAB-integrated three-dimensional Monte Carlo light transport solver with heat diffusion and tissue damage. J. Biomed. Opt. 2018;23(12) doi: 10.1117/1.JBO.23.12.121622. [DOI] [PubMed] [Google Scholar]
- 90.Leino A.A., Pulkkinen A., Tarvainen T. ValoMC: A Monte Carlo software and MATLAB toolbox for simulating light transport in biological tissue. OSA Contin. 2019;2(3):957–972. doi: 10.1364/OSAC.2.000957. [DOI] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
Data will be made available on request.