Abstract
Volcanic halogen emissions play a critical role in atmospheric chemistry and serve as valuable indicators of volcanic activity. However, direct in situ quantification and speciation of reactive halogen species (RHS) remain analytically challenging, particularly under the constraints of UAV-based sampling. In this study, we present a suite of miniaturised and field-deployable methods for halogen speciation in volcanic plumes, optimised for unmanned aerial vehicle (UAV) platforms. Two types of coated syringe filters, based on cis-stilbene and 1,3,5-trimethoxybenzene (TMB), enable in situ derivatisation and selective detection of molecular halogens and interhalogens (e.g. Cl2, BrCl, Br2, I2). An alkaline trap system (“Bubbler”) was also developed for total halogen and sulphur quantification. Method performance was validated using a custom-designed chlorine permeation source in controlled laboratory and dilution chamber experiments. Detection limits were sub-ng for GC-HRMS (gas chromatography–high-resolution mass spectrometry)-based speciation and sub-µg for ion chromatography of total halogens. These results demonstrate the feasibility of quantitative, UAV-based halogen speciation in volcanic plumes, offering new possibilities to enhance our understanding in volcanic gas chemistry and atmospheric trace gas analysis.
Graphical abstract

Supplementary Information
The online version contains supplementary material available at 10.1007/s00216-026-06403-7.
Keywords: Volcanic gases, Molecular halogens, Interhalogens, Bromine chloride, Alkaline trap, GC–MS
Introduction
Volcanic emissions represent a significant natural source of atmospheric trace gases, influencing both regional air quality and global climate. While major volcanic gas constituents such as H2O, CO2, and SO2 dominate in terms of abundance, reactive halogen-containing trace gases strongly influence atmospheric oxidation capacity, ozone chemistry, and aerosol formation, even at trace levels [1–3]. Moreover, the composition of volcanic plumes provides valuable information on magmatic processes and can serve as an early warning indicator of volcanic unrest. Thermodynamic modelling of volcanic gas emissions further supports the key role of halogens in trace element transport, indicating that many metals are likely present as chloride or mixed hydroxy-chloro species [4].
Hydrogen halides (HF, HCl, HBr, HI) are typically the most abundant halogen species released [5, 6]. These acidic gases, along with their oxidised counterparts, such as molecular halogens (Cl2, Br2, I2) and interhalogens (BrCl, ICl), contribute to complex multiphase chemistry in the troposphere. The detection of bromine monoxide (BrO) in volcanic plumes has highlighted the importance of in-plume oxidation pathways, driven by interactions with oxidants entrained from the atmosphere [3, 7–10]. Such processes can initiate autocatalytic halogen activation cycles, with implications for ozone depletion [11, 12]. However, reactive halogen species might also be used as volcanic activity proxies [13, 14] similar to traditional gas ratios like CO2/SO2 or SO2/HCl [15–18]. Although bromine chemistry has received most of the attention, iodine and chlorine also contribute to reactive halogen chemistry in volcanic plumes. For example, IO was detected via satellite in the plume of Mt. Kasatochi [19], and other species like OClO [20, 21], ClO [22] or Cl2 [23] have also been measured.
While remote sensing techniques—such as Differential Optical Absorption Spectroscopy (DOAS)—have been widely employed to quantify selected halogen species (e.g. BrO, IO) [24, 25], many key reactive intermediates remain undetectable with current optical methods although they have been predicted and incorporated into chemistry models [10, 21, 26, 27]. Moreover, remote sensing techniques, particularly optical spectroscopy, do not generally provide access to the full halogen budget. While HF is essentially chemically inert and thus a reliable measure of total fluorine, and HCl often serves as a reasonable proxy for total chlorine, this is not the case for bromine and iodine. For these elements, the primary emitted HBr and HI cannot be detected by field-deployable remote sensing instruments and rapid plume chemistry redistributes the emitted halogens among multiple reactive species. Ground-based direct sampling studies have long provided detailed insights into volcanic gas compositions and sampling strategies, as summarised in comprehensive reviews of fumarolic gas investigations [28]. Building up on these foundation, direct sampling approaches, such as alkaline traps and denuder systems [29–34], offer complementary capabilities by enabling quantitative collection of both reduced and oxidised halogen species.
Recent advances in UAV technology now allow aerial sampling in hazardous or logistically challenging environments. This study presents two complementary UAV-compatible sampling systems: (1) a miniaturised alkaline trap (“Bubbler”) for quantifying total halogen and sulphur content, and (2) in situ derivatisation filters coated with either cis-stilbene or TMB, enabling selective detection of specific RHS based on electrophilic addition or electrophilic aromatic substitution. The combination of field-deployable sampling and high-resolution analytical techniques (gas chromatography–high-resolution mass spectrometry (GC-HRMS) and ion chromatography (IC)) offers improved sensitivity and speciation capability for halogen analysis in volcanic plumes. This approach extends previous chromatographic developments, as comprehensively reviewed in [35], highlighting the use of GC and LC methods in volcanic gas surveillance. To characterise filter reactivity and validate method performance under controlled conditions, a novel chlorine permeation source was developed.
The purpose of this paper is to present the development and validation of drone-based analytical methods for the in situ speciation of gaseous halogens in volcanic plumes, enabling spatial and temporal characterisation of reactive halogen compounds under challenging field conditions. These tools extend the analytical limits of volcanic gas monitoring, providing both higher sensitivity and greater chemical resolution in airborne halogen measurements.
Materials and methods
Reagents and standards
All experiments were conducted using analytical-grade reagents. Sodium hydroxide monohydrate (99.996%), sodium chloride (99.99%), sodium bromide (> 99%), sodium fluoride (97%), and sodium sulphate (99%) were dissolved in ultrapure water (18.2 MΩ·cm) to prepare calibration standards and absorbing solutions for the alkaline trap. Aqueous SO2 solution (5%) was used for evaluating SO2 collection, and hydrogen peroxide (> 30%) together with manganese dioxide (99%) served during the post-sampling oxidation procedure. A protonated cation exchange resin (AmberLite Mac-3H) was regenerated with dilute nitric acid (0.01 M) and used for neutralising alkaline solutions prior to ion chromatographic analysis. Standards for GC-HRMS quantification of derivatised halogens included Cl-TMB, Br-TMB, I-TMB, stilbene dichloride (S-Cl2), stilbene dibromide (S-Br2), and stilbene bromochloride (S-BrCl). All were dissolved in acetone (GC grade) and stored in amber vials at 5 °C.
Alkaline trap and operation
The alkaline trap used in this study was designed as a miniaturised alternative to classical impingers with the goal of reducing sample volume and weight during UAV deployment. Each trap consisted of a borosilicate vial equipped with a screw cap through which a 4 µm polyethylene frit (HPLC solvent filter, Techlab) was inserted. The frit was connected to the inlet line. For each experiment, two bubblers were filled with 4 mL of 0.5 M NaOH, while a third bubbler—connected downstream—remained empty to act as a safety trap in case of liquid entrainment. The setup of the alkaline trap (“Bubbler”) is shown with some technical details in Fig. 1. Sampling in the laboratory was typically performed at 250–500 mL min−1 for 15–30 min and supplied by Sensidyne GilAir Plus pumps for all laboratory experiments. For UAV-based field deployments, lightweight membrane pumps were employed to minimise payload and power consumption while maintaining sufficient and stable flow rates for gas sampling [36]. During operation, the frit remained fully submerged to ensure fine bubble formation and efficient gas–liquid transfer. After sampling, 1 mL of the alkaline solution was oxidised by addition of 2 µL of hydrogen peroxide for 24 h at room temperature. Excess peroxide was removed by adding a small amount of MnO2 and allowing the suspension to react for 48 h. Finally, samples were neutralised by cation exchange resin that had been prewashed until anion-free.
Fig. 1.
Schematic representation of the alkaline trap (“Bubbler”) with some technical details
Ion chromatographic measurements were performed using a Dionex ICS-1100 equipped with an AS14A analytical column and precolumn, 4 mm AERS suppressor (55 mA), a 100 µL injection loop, and an autosampler. The eluent consisted of a carbonate/bicarbonate mixture (1 mM NaHCO3 + 8 mM Na2CO3) delivered at 1.0 mL min−1. The detector was set at 40 °C, and all samples and standards were analysed at least in duplicate. Chromeleon 7.1.2 was used for peak integration.
Chlorine permeation source
A reliable Cl2 standard was essential for filter characterisation and a new source had to be developed, since the techniques used for Br2 and I2 gas generation could not be applied to chlorine [37]. Therefore, two PTFE permeation sources were constructed (Fig. 2) and filled with pre-cooled liquefied chlorine. Chlorine gas was obtained by adding concentrated hydrochloric acid dropwise to potassium permanganate in a glass flask. The evolving gas was continuously flushed by a gentle nitrogen stream into a glass collection tube placed in a −70 °C cooling bath, where the gas condensed as a yellow liquid. After filling, both permeation tubes were sealed with PTFE plugs. Their permeation behaviour was characterised gravimetrically. Each device was placed in a temperature-controlled chamber at 30 °C and continuously flushed with nitrogen at 50 mL min−1. The mass was recorded at weekly intervals for several months to determine both the initial equilibration phase and the long-term stable output rate.
Fig. 2.
Chlorine permeation tubes, Source 1 (left) and Source 2 (right), all dimensions in mm, PTFE source (hatched area), chlorine reservoir (yellow), active permeation area (red)
Safety considerations
The preparation and handling of chlorine permeation sources involving condensed Cl₂ were performed exclusively in a certified chemical fume hood. Chlorine generation, condensation, filling, storage, and characterisation were carried out under continuous ventilation. Personal protective equipment included safety goggles, laboratory coat, chemically resistant gloves, and insulated gloves for cryogenic handling. Leak integrity was verified gravimetrically by weighing each permeation source immediately after filling and again after 24 h. No abnormal mass loss was observed. During the entire study, the sources were stored in a fume hood. For gravimetric measurements, the sources were briefly removed. Potential chlorine release during this step is negligible due to low permeation rates and short handling times.
All procedures complied with institutional laboratory safety regulations. No external transport of liquefied chlorine or special permits beyond standard laboratory safety training were required.
Filter preparation, measurement, and chamber experiments
Syringe filters (Macherey–Nagel GF-100/25, 25 mm diameter, 1 µm pore size) were used as substrates for the derivatisation coatings. For cis-stilbene filters, 150 µL of a 30 mM solution of cis-stilbene (0.81 mg ± 2%, defined by pipetting precision) in acetone was applied to each filter. The solvent was allowed to evaporate under a gentle stream of laboratory air. The filters were used immediately or sealed and stored at 5 °C for no longer than 3 days to prevent isomerisation. To explicitly assess the shelf life and potential cis-to-trans isomerisation of cis-stilbene on the filter material, a dedicated storage experiment was performed. A cis-stilbene-loaded filter was stored for three days at room temperature, subsequently extracted and analysed by GC, and compared to a freshly prepared filter. Within the analytical precision of the method, no statistically significant loss of the cis-stilbene signal was observed, indicating sufficient short-term stability on the filter under ambient conditions (see data repository). TMB filters were prepared analogously using 30 mM TMB (0.76 mg ± 2%) in acetone; in contrast to cis-stilbene, TMB is not prone to isomerisation and could therefore be stored at room temperature without detectable changes. After sampling, filters were placed in glass vials and extracted four times with 300 µL of acetone (for stilbene filters) or toluene (for TMB filters). Combined extracts were evaporated under nitrogen at 35 °C using a TurboVap system and reconstituted in 200 µL acetone before analysis. The extracts were sonicated twice for 5 min to ensure complete dissolution.
Analyses were performed using a Thermo TRACE 1610 gas chromatograph coupled to an Orbitrap Exploris high-resolution mass spectrometer operating in electron-impact mode (70 eV). A programmed-temperature-vaporisation (PTV) injector was employed minimising thermal decomposition. For the cis-stilbene products, the injector temperature was initially held at 70 °C and then ramped to 200 °C at 14.5 °C min−1. The oven was operated according to the following programme: 40 °C for 2 min, 30 °C min−1 to 160 °C, then 4 °C min−1 to 190 °C (3 min hold), and finally 80 °C min−1 to 260 °C (5 min). TMB derivatisation products were analysed using a separate programme starting at 90 °C (3 min), followed by 10 °C min−1 to 210 °C, 7 °C min−1 to 235 °C, and 20 °C min−1 to 250 °C (5 min hold). Quantification was carried out in SIM mode using characteristic fragment ions listed in Table 1. For stilbene derivatives, response factors of anti-isomers were used for quantifying syn-isomers in the absence of authentic standards.
Table 1.
Characteristic ion fragments and retention times for reaction products of halogens with cis-stilbene and TMB
| Filter | Analyte | SIM (selected ion monitoring) | m/z | Retention time (min) |
|---|---|---|---|---|
| Cis-stilbene | S-Cl2 | C7H6Cl | 125.0153 | 14.4 and 14.9 |
| S-BrCl | C14H12Cl | 215.0622 | 16.5 and 16.8 | |
| S-Br2 | C14H12Br | 259.0117 | 17.5 and 17.7 | |
| TMB | Cl-TMB | C9H11O3Cl | 202.0391 | 13.0 |
| Br-TMB | C9H11O3Br | 245.9886 | 14.0 | |
| I-TMB | C9H11O3I | 293.9747 | 15.0 |
Collection efficiency experiments were performed in a 5 L dilution chamber through which halogen gases were introduced either in nitrogen (0% RH) or in laboratory air (95 ± 5% RH). Chlorine was introduced using the developed permeation tubes, while bromine and iodine sources followed previous studies [37]; bromine chloride was generated in situ by simultaneous activation of bromine and chlorine sources in the dilution chamber.
Three filters or bubblers were connected in series, and the amount of reaction product on each stage was quantified to evaluate breakthrough behaviour (Fig. 3). Collection efficiencies (CE) were calculated using Eq. 1. Flow rates of 250, 500, and 750 mL min−1 were tested. Additional experiments evaluated the influence of humidity, oxygen, and sampling duration. Recovery tests were conducted by applying known amounts of calibration standards directly onto filters, followed by the same extraction and analysis procedure as for sampled filters. The resulting recovery factors were incorporated into the determination of detection and quantification limits.
| 1 |
Fig. 3.
Experimental setup for collection efficiency measurements of alkaline traps and coated filters
Any potential coating loss during preparation, laboratory handling, or storage is inherently accounted for by the experimental design. This includes shelf life aspects such as storage at room temperature or at reduced temperature, depending on the coating material. Filters were prepared, stored and handled in the same manner as for field deployment and subsequently subjected to collection efficiency experiments under deliberately extreme laboratory conditions with respect to flow rate, sampling duration and temperature.
Determination of detection and quantification limits
Limits of detection (LOD) and quantification (LOQ) were determined from calibration statistics. LODs were calculated as 3.3 times the standard deviation of the y-intercept of the calibration curve divided by the slope. LOQs were defined as three times the corresponding LOD. Calibration data used for LOD and LOQ estimation were restricted to the low-concentration range relevant for trace-level detection. This ensures that the derived limits reflect instrumental noise and variability close to the detection threshold rather than signal behavior at higher concentrations. Details on the calibration ranges used for LOD and LOQ determination are provided in “S2”, data repository.
Drone setup
The described sampling system can be integrated into multi-rotor UAV platforms with sufficient payload capacity and flight time, such as the DJI Matrice 350 RTK. A compact sensor module equipped with a gas sensor (e.g. SO2) provides real-time information to the operator and can serve as a trigger for activating the sampling sequence [33]. The sampling unit is based on three independent pumps, each of which is responsible for maintaining a constant flow through one of the collection devices (two coated filters, e.g. cis-stilbene and TMB, and one alkaline trap). Pump operation can be remotely controlled, allowing selective activation during the sampling period. The complete system (Fig. 4), sensor unit, control electronics, and pumps was designed, built and modified by Niklas Karbach [36] and with the bubblers and filters has a total weight of approximately 600 g. The sampling system operates autonomously using dedicated batteries and does not draw electrical power from the UAV, any impact on flight time is therefore limited to the additional payload mass. The present study focuses on the development and laboratory validation of UAV-compatible sampling systems. All experiments described here were conducted under laboratory and chamber conditions. No UAV-based field sampling in volcanic plumes was performed within the scope of this work. Consequently, no airspace permissions or field-operation approvals were required for the experiments presented.
Fig. 4.
Sampling system with electronics, sensors, communication, pumps, and bubblers (disconnected) mounted on a drone platform. Syringe filters not included
Laboratory results and discussion
Performance and analytical stability of the alkaline trap
The alkaline trap was developed to achieve quantitative absorption of acidic volcanic gases while minimising weight and sampling volume for UAV-based applications. In controlled dilution chamber experiments (Fig. 3), the system achieved high absorption efficiencies for both weak and strong acidic gases. The collection efficiency tests were performed by evaporating an aqueous SO2 solution (5%, 0.43 mL, corresponding to 22 mg SO2) into an ambient air stream containing 550 ± 100 ppm CO2 and calculated using Eq. 1. As summarised in Table 2, SO2 was captured with an efficiency of 100 ± 1.0%, while CO2 exhibited an efficiency of 98 ± 5%. These values confirm that even weakly acidic gases are effectively absorbed, indicating that hydrogen halides (HF, HCl, HBr) will be collected with near-complete efficiency.
Table 2.
Results of collection efficiency measurements for alkaline traps
| Method | Analyte | Flow (mL min−1) | Sampling time (min) | Humidity (%) | Collection efficiency (%) |
|---|---|---|---|---|---|
| Alkaline trap | SO2 (22 mg) | 500 ± 25 | 150 | 65 ± 10 | 100 ± 1 |
| CO2 (550 ± 100 ppm, ambient) | 98 ± 5 |
A notable observation was the formation of a storage time dependent impurity signal at 3.7 min in the IC chromatogram (Fig. 5), attributed to slow interactions between NaOH and the plastic storage vessel. Glass containers were not considered as an alternative, as concentrated alkaline solutions are known to attack glass surfaces, potentially releasing silicates and introducing additional artefacts relevant for anion analysis.
Fig. 5.
Standard solution of NaF and NaCl in 0.5 M NaOH after sample treatment showing an additional signal at 3.7 min, probably caused by a reaction of hydroxide with the plastic container after two days of storage
This impurity did not interfere with fluoride or chloride unless alkaline solutions were stored for extended periods. Storing the alkaline solution at 5 °C significantly slowed the formation of impurities, indicating that the underlying reaction could be reduced but not completely prevented. Consequently, rapid analysis of collected samples is recommended to minimise storage-related contamination. This demonstrates that the alkaline trap provides stable analytical performance under field-relevant storage times.
Recovery experiments confirmed that oxidation (H2O2) and neutralisation using cation exchange resin do not compromise analytical accuracy. A reference solution and an identical alkaline-treated solution yielded recovery rates of 95 ± 5% for all target analytes. Together with high collection efficiency, this enables accurate quantification of total halogen and total sulphur species, which are critical parameters for deriving ratios such as BrCl/Brtotal or Brtotal/Stotal during volcanic monitoring.
The limits of detection and quantification calculated from calibration standards, sample volume, flow rate, recovery and collection efficiency, are listed in Table 3. Resulting LODs ranged from 0.012 µg (HBr) to 1.7 µg (SO2), corresponding to sub-ppb to mid-ppb mixing ratios under typical UAV sampling conditions (250 mL min−1, 15 min). These values are well suited for near-vent measurements at volcanoes such as Mt. Etna.
Table 3.
Alkaline trap: LODs and LOQs in absolute amounts and in their mixing ratios at normal conditions using typical sampling settings, flow 250 mL min−1, sampling time 15 min
| Limit | Alkaline trap “Bubbler” | |||
|---|---|---|---|---|
| HF | HCl | HBr | SO2 | |
| LOD (µg) | 0.022 | 0.62 | 0.011 | 0.51 |
| LOD (ppbv) | 6.6 | 100 | 0.84 | 47 |
| LOQ (µg) | 0.066 | 1.9 | 0.034 | 1.5 |
| LOQ (ppbv) | 20 | 310 | 2.5 | 140 |
Stability and output characteristics of the chlorine permeation source
For characterisation the new chlorine source was stored at 30 °C and weighed once a week. Mass loss curves recorded over several months showed a clear transition from an initial non-linear permeation phase to a stable, linear regime (see Fig. S2, Supplementary Information). Steady-state output was reached after ~ 6 weeks (Tube 1) and ~ 9 weeks (Tube 2) and remained constant for 4–8 months, depending on the amount of liquid chlorine initially loaded.
Interestingly, both tubes—despite different wall thicknesses—exhibited identical output rates (0.245 and 0.246 mg h−1). This is consistent with their nearly identical permeation-area-to-wall-thickness ratios (A/L), the dominant factor controlling permeation [38, 39]. Permeability coefficients derived from the steady-state regime agreed well with literature values (Table 4), confirming that the PTFE devices behave predictably according to established permeation theory [40].
Table 4.
Permeability of Cl2 in PTFE Membranes; Pe = [m3(STP) m/(m2hbar)] × 107
| A/L (m) | Output rate (mg/h) | Pe (this study) at 30 °C, 8.9 bar | Pe [40] at 30 °C, 1–3 bar | |
|---|---|---|---|---|
| Source 1 | 0.37 | 0.246 | 0.26 | 0.13 |
| Source 2 | 0.36 | 0.245 | 0.26 |
Performance of cis-stilbene-coated filters
Cis-stilbene reacts with Cl2, Br2 and BrCl via electrophilic addition (Fig. 6), generating distinct dihalogenated stilbene products. Collection efficiencies determined in the dilution chamber (Table 5) exceeded 99.9% for all tested analytes under extreme conditions (750 mL min−1, 30 min). This confirms that cis-stilbene is highly reactive toward molecular halogens and interhalogens.
Fig. 6.
Reaction of cis-stilbene with chlorine, analogous reaction for bromine and bromine chloride
Table 5.
Sampling conditions (T = 303.25 K) and collection efficiencies for cis-stilbene and TMB-coated filters
| Method | Analyte/concentration | Flow (mL min−1) | Sampling time (min) | Humidity (%) | Collection efficiency for all analytes (%) |
|---|---|---|---|---|---|
| Cis-Stilbene |
Cl2 (5.2 ppmv) Br2 (60 ppbv) |
250 ± 12.5 | 30 | 0 (pure N2) | 100 ± 1 |
|
Cl2 (2.6 ppmv) Br2 (30 ppbv) |
500 ± 25 | 30 | 0 (pure N2) | 100 ± 1 | |
|
Cl2 (1.7 ppmv) Br2 (20 ppbv) |
750 ± 37.5 | 30 | 95 ± 5 (laboratory air) | 100 ± 1 | |
|
Cl2 (2.6 ppmv) Br2 (30 ppbv) |
500 ± 25.0 | 60 | 0 (pure N2) |
48 ± 1 (Cl2) 65 ± 1 (BrCl) 91 ± 1 (Br2) |
|
| TMB |
Cl2 (1.7 ppmv) Br2 (20 ppbv) I2 (14 ppbv) |
750 ± 37.5 | 60 | 95 ± 5 (laboratory air) | 100 ± 1 |
At prolonged sampling (60 min) and increased flow (500 mL min−1), breakthrough was observed (Cl2: 48%, BrCl: 65%), with significantly better retention of Br2 (91%). The difference likely reflects both stilbene volatility (leading to partial loss of coating) and the intrinsic reactivity of the analytes. These conditions represent extreme laboratory scenarios; field sampling durations rarely exceed 10–20 min on UAV platforms, ensuring complete collection. Furthermore, the possible reaction of cis-stilbene with hydrogen halides (HCl, HBr, HI), present in significant quantities in volcanic plumes, was tested. Aqueous solutions of these halides (1 mL, 1%) were individually evaporated under a nitrogen stream and passed through cis-stilbene-coated filters. No corresponding reaction products were detected. HF has not been experimentally investigated because working with HF involves considerable experimental and safety restrictions. Given that no addition was observed for HCl, HBr, or HI, and considering the well-known decrease in reactivity from HI to HF in electrophilic additions to alkenes, an addition of HF can be excluded under the investigated conditions.
Experiments with I2, ICl, and IBr yielded no detectable products. This observation is consistent with literature reports describing iodine-catalysed cis-to-trans isomerisation of stilbenes, in which thermally labile iodine addition intermediates rapidly decompose to form trans-stilbene [41]. Oxidants such as ozone may affect plume composition upstream of sampling but are not expected to react directly with the coating materials on the timescales relevant for filter sampling. The brominated stilbene products showed pronounced thermal sensitivity and degraded rapidly under conventional split/splitless injection. A PTV injection mode was therefore essential. This ensures that solvent is evaporated before analyte volatilisation, minimising thermal decomposition. Recovery rates of S-Cl2, S-BrCl, and S-Br2 were 87%, 92%, and 97%, respectively, confirming robust extraction and quantification across all derivatisation products.
Performance of TMB-coated filters
TMB selectively undergoes electrophilic aromatic substitution with halogen species in oxidation states 0 and +1 (Fig. 7). Collection efficiencies were consistently > 99.9% for Cl2, Br2, BrCl, and I2 across various conditions, including low humidity (0% RH in N2) and high humidity (95% RH in laboratory air). Table 5 summarises the full efficiency dataset. Recovery rates were 87% (Cl-TMB), 90% (Br-TMB), and 95% (I-TMB).
Fig. 7.
Reaction of TMB with halogen species (X) of oxidation state 0 and +1 yields the corresponding X–TMB adducts
Possible reactions of TMB with halides were investigated in the same way as with cis-stilbene-coated filters, but no reaction products were detected. In connection with halogen studies in volcanic plumes, the question arises as to what extent reactions with halogen radicals (Br·, BrO·) may also be responsible for the observed product formation. Therefore, TMB filters were exposed to elemental bromine under daylight and dark conditions. Under dark conditions (i.e. without bromine photolysis forming bromine radicals), higher Br-TMB signals were obtained than under photolysis conditions, which rules out major contributions from radical reactions. In addition, it is generally expected that radical substitution on TMB would occur at one of the methyl groups, resulting in the formation of an R-CH2Br derivative. However, no such products were observed. The same applies to the reaction of BrO with TMB, which cannot be easily reproduced in the laboratory due to the difficulty of generating defined BrO concentrations. Moreover, a hypothetical electrophilic substitution of BrO on the aromatic ring would require the formation of a hydroxyl radical as the leaving group, making such a reaction mechanistically unfavourable. We therefore rule out any significant contribution from radical bromine compounds (Br· or BrO·) to the measurement results obtained using the method presented.
Detection limits and mixing ratios for all filter systems
Detection limits for both filter types are summarised in Table 6. For stilbene filters, LODs ranged from 0.12 ng (BrCl) to 0.30 ng (Cl2); TMB filters achieved similar sensitivities with LODs between 0.060 ng (I2) and 0.16 ng (Br2).
Table 6.
Coated filters: determined LODs and LOQs in absolute amounts and in their mixing ratios at normal conditions using typical sampling conditions, flow 250 mL min−1, sampling time 15 min at 1 atm
| Limit | Stilbene filter | TMB filter | ||||
|---|---|---|---|---|---|---|
| Cl2 | BrCl | Br2 | Cl2 | Br2 | I2 | |
| LOD (ng) | 0.30 | 0.12 | 0.19 | 0.094 | 0.16 | 0.060 |
| LOD (pptv) | 7.1 | 2.7 | 3.3 | 2.2 | 3.7 | 1.1 |
| LOQ (ng) | 0.90 | 0.35 | 0.56 | 0.28 | 0.48 | 0.18 |
| LOQ (pptv) | 21 | 8.0 | 9.8 | 6.7 | 11 | 3.2 |
|
For typical UAV sampling (250 mL min-1, 15 min), these values correspond to: Cl2: 7.1–21 pptv BrCl: 2.7–8.0 pptv Br2: 3.3–9.8 pptv I2: 1.1–3.2 pptv | ||||||
These detection limits represent a 1–2 orders of magnitude improvement over previously published denuder systems [33] and enable detection of reactive halogen species at concentrations typical of ageing volcanic plumes. Field measurements of reactive bromine species using TMB-based sampling at Mt Etna have reported mixing ratios in the low-ppbv range [33]. The sub-pptv detection limits achieved here are therefore well below concentrations observed close to the vent and demonstrate sufficient sensitivity for near-source plume conditions.
For other molecular halogen species targeted by the cis-stilbene filters, quantitative concentration data in volcanic plumes are generally scarce, one study of Cl2 reporting ppmv concentration [23]. The achieved detection limits indicate that these species can be addressed across a wide range of plume conditions, although interpretation of absolute mixing ratios remains challenging due to strong spatial and temporal variability and plume dilution during UAV-based sampling.
Conclusions
The presented methods were evaluated for their performance and optimised for rapid analyte enrichment, minimal weight, short sampling times, and high analytical sensitivity. The alkaline trap represents a miniaturised alternative (optimised weight, sample volume and sample treatment) over existing techniques, in which the strengths of the established approaches are combined, specifically designed for UAV-based deployment, still enabling the quantification of total halogen and sulphur species for realistic plume conditions (e.g. Mt Etna volcanic plume). Absolute detection and quantification limits were 0.022 and 0.066 µg for HF, 0.62 and 1.9 µg for HCl, 0.011 and 0.034 µg for HBr, and 0.51 and 1.5 µg for SO2.
Cis-stilbene-coated syringe filters allow selective quantification of chlorine, bromine, and bromine chloride, with absolute detection and quantification limits of 0.30 and 0.90 ng for Cl2, 0.12 and 0.35 ng for BrCl, and 0.19 and 0.56 ng for Br2. TMB-coated filters, adapted from recently developed denuder systems and combined with GC-HRMS analysis, provide improved detection limits for reactive halogen species in oxidation states 0 and + 1. Absolute detection and quantification limits for TMB filters were 0.094 and 0.28 ng for chlorine, 0.16 and 0.48 ng for bromine, and 0.060 and 0.18 ng for iodine.
By combining total halogen quantification using the alkaline trap with selective speciation of reactive halogen species using derivatisation filters, the presented UAV-compatible approach enables direct and simultaneous assessment of total and oxidised halogen fractions in volcanic plumes. This provides the methodological basis for determining widely used volcanological ratios such as Br(total)/S(total), Br(reactive)/Br(total), BrCl/Br(total), or (Br2 + BrCl)/Br(total) under near-vent conditions, which are key parameters for interpreting halogen activation, plume chemistry, and volcanic degassing processes.
Plume sampling can be hazardous, and certain locations, in particular aged plumes, (minutes after the gas emissions) are usually inaccessible. UAV-based sampling offers a viable solution although it is limited by payload capacity and sampling time. Consequently, short sampling durations and weight-optimised devices are highly advantageous. The study of gas- and particle-phase processes in volcanic plumes, where rapid chemical transformations occur, would be virtually impossible without UAVs. In the future, autonomous drone missions with onboard sensors could enable remote volcano monitoring and direct measurements.
Several limitations of the methods presented should be considered. Brominated stilbene reaction products exhibit pronounced thermal sensitivity, which necessitates mild injection conditions such as programmed-temperature vaporisation and represents an analytical constraint. The coated filter systems are selective and differ in their chemical scope. Cis-stilbene selectively traps molecular halogens and interhalogens while other halogen species are not retained. TMB reacts with halogen species in oxidation states 0 and +1; however, the exact reaction pathways and potential contributions from other reactive halogen species could not be independently verified. Radical halogen species are not actively trapped by either coating and may therefore contribute to the overall halogen budget without being reflected in the speciation results. Furthermore, UAV-based sampling is inherently limited in sampling duration and spatial coverage. Volcanic plumes are often highly heterogeneous in space and time, and the measured concentrations therefore represent local plume snapshots rather than fully integrated plume compositions. The present study provides a laboratory-based benchmark for UAV-compatible halogen sampling. Application under real volcanic plume conditions represents the next step to assess performance under atmospheric complexity beyond controlled environments.
Future research should explore improved analytical techniques for detecting inorganic halogens in basic solutions that can achieve detection limits comparable to the presented filter methods. This is particularly relevant because bromine concentrations in volcanic plumes are often close to or below the detection limits of conventional ion chromatography, and iodine, which occurs at even lower levels, is practically undetectable by IC in alkaline solution. Successful determinations of bromine and iodine by ICP–MS have already been demonstrated, providing a sensitive approach for total halogen quantification. Alternatively, conversion of inorganic halogens into organic derivatives suitable for GC–MS analysis could offer an efficient complementary method.
Supplementary Information
Below is the link to the electronic supplementary material.
Acknowledgements
The authors would like to thank Terrestrial Magmatic Systems (TeMaS), a collaborative effort that coordinates interdisciplinary research on magmatic processes between the universities of Mainz, Frankfurt, and Heidelberg, for financial support. We also thank the German Research Foundation (DFG) for further financial support through TRR 301 Project ID 428312742 and HO 1748/24-1, HO 1748/23-1 and HO 1748/19-1. We would also like to thank the Istituto Nazionale di Geofisica e Vulcanologia, Italy, grant “Progetto INGV Pianeta Dinamico (MUSUNGU)” grant - code CUP D53J19000170001 - funded by the Italian Ministry MIUR (“Fondo Finalizzato al rilancio degli investimenti delle amministrazioni centrali dello Stato e allo sviluppo del Paese”, legge 145/2018). The authors thank Niklas Karbach for developing a prototype sensor system for UAV-based gas sampling. B.G. would like to express his sincere thanks to the ANAKON 2025, for awarding the Best Poster Prize, sponsored by Analytical and Bioanalytical Chemistry, based on the preliminary results of this study.
Author contribution
B.G. carried out the majority of the laboratory work and data processing, with support from N.B. and T.H. during the field experiments. The evaluation and interpretation of the results were performed jointly by all three authors. All authors contributed to the writing and revision of the manuscript and approved the final version.
Funding
Open Access funding enabled and organized by Projekt DEAL. Progetto INGV Pianeta Dinamico (MUSUNGU), CUP D53J19000170001, Terrestical Magmatic Systems TeMaS, Deutsche Forschungsgemeinschaft, HO 1748/19-1, HO 1748/23-1, HO 1748/24-1, TRR 301 Project ID 428312742
Data availability
The raw data supporting this study, including raw data with mass spectra for each stilbene and TMB reaction product, low level calibration files, raw data that support storage behaviour of cis-stilbene filters and permeation mass loss time series are available under: 10.5281/zenodo.18417414. Additional mass spectra for each product were added in the Supplemental Information.
Declarations
Conflict of interest
The authors declare no competing interests.
Footnotes
Parts of this work have been presented at Anakon 2025 in Leipzig and have been awarded with an ABC Best Poster Award.
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Symonds RB, Rose WI, Reed MH. Contribution of C1- and F-bearing gases to the atmosphere by volcanoes. Nature. 1988. 10.1038/334415a0.2834647 [Google Scholar]
- 2.Saiz-Lopez A, von Glasow R. Reactive halogen chemistry in the troposphere. Chem Soc Rev. 2012. 10.1039/C2CS35208G. [DOI] [PubMed] [Google Scholar]
- 3.von Glasow R. Atmospheric chemistry in volcanic plumes. Proc Natl Acad Sci U S A. 2010. 10.1073/pnas.0913164107. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Mandon CL, Kaasalainen H, Calabrese S, Shock EL, Prapaipong P, Tassi F, Gunnarsson I, Gunnarsson-Robin J, Stefánsson A. Trace element transport by volcanic gases at Vulcano (Sicily, Italy) – speciation, deposition and fluxes. J Volcanol Geotherm Res. 2025. 10.1016/j.jvolgeores.2024.108235. [Google Scholar]
- 5.Cadle RD. Volcanic emissions of halides and sulfur compounds to the troposphere and stratosphere. J Geophys Res. 1975. 10.1029/JC080i012p01650. [Google Scholar]
- 6.Cadle RD. A comparison of volcanic with other fluxes of atmospheric trace gas constituents. Rev Geophys. 1980. 10.1029/RG018i004p00746. [Google Scholar]
- 7.Bobrowski N, Hönninger G, Galle B, Platt U. Detection of bromine monoxide in a volcanic plume. Nature. 2003. 10.1038/nature01625. [DOI] [PubMed] [Google Scholar]
- 8.Oppenheimer C, Tsanev VI, Braban CF, Cox RA, Adams JW, Aiuppa A, et al. BrO formation in volcanic plumes. Geochim Cosmochim Acta. 2006. 10.1016/j.gca.2006.04.001.
- 9.Gerlach TM. Volcanic sources of tropospheric ozone‐depleting trace gases. Geochem Geophys Geosyst. 2004. 10.1029/2004GC000747. [Google Scholar]
- 10.Roberts TJ, Braban CF, Martin RS, Oppenheimer C, Adams JW, Cox RA, Jones RL, Griffiths PT. Modelling reactive halogen formation and ozone depletion in volcanic plumes. Chem Geol. 2009. 10.1016/j.chemgeo.2008.11.012. [Google Scholar]
- 11.Daniel JS, Solomon S, Portmann RW, Garcia RR. Stratospheric ozone destruction: the importance of bromine relative to chlorine. J Geophys Res. 1999. 10.1029/1999JD900381. [Google Scholar]
- 12.Von Glasow R, Crutzen PJ. Model study of multiphase DMS oxidation with a focus on halogens. Atmospheric Chemistry and Physics. 2004. 10.5194/acp-4-589-2004.
- 13.Bobrowski N, Giuffrida G. Bromine monoxide / sulphur dioxide ratios in relation to volcanological observations at Mt. Etna 2006–2009. 2012. Solid Earth. 10.5194/se-3-433-2012 .
- 14.Lübcke P, Bobrowski N, Arellano S, Galle B, Garzón G, Vogel L, et al. BrO/SO 2 molar ratios from scanning DOAS measurements in the NOVAC network. Solid Earth. 2014. 10.5194/se-5-409-2014.
- 15.Noguchi K, Kamiya H. Prediction of volcanic eruption by measuring the chemical composition and amounts of gases. Bull Volcanol. 1963. 10.1007/BF02597298. [Google Scholar]
- 16.Pennisi M, Le Cloarec M-F. Variations of Cl, F, and S in Mount Etna’s plume, Italy, between 1992 and 1995. J Geophys Res. 1998. 10.1029/97JB03011.
- 17.Aiuppa A, Federico C. Anomalous magmatic degassing prior to the 5th April 2003 paroxysm on Stromboli. Geophys Res Lett. 2004. 10.1029/2004GL020458. [Google Scholar]
- 18.Carroll MR, editor. Volatiles in magmas: Volatiles in Magmas Course … Napa Valley Sheraton Hotel in California, Dec. 2 - 4, 1994. Reviews in mineralogy, vol. 30. Washington, DC: Mineralogical Society of America; 1994.
- 19.Schönhardt A, Richter A, Theys N, Burrows JP. Space-based observation of volcanic iodine monoxide. Atmos Chem Phys. 2017. 10.5194/acp-17-4857-2017.
- 20.Kern C, Lyons JJ. Spatial distribution of halogen oxides in the plume of Mount Pagan Volcano, Mariana Islands. Geophys Res Lett. 2018. 10.1029/2018GL079245. [Google Scholar]
- 21.Bobrowski N, von Glasow R, Aiuppa A, Inguaggiato S, Louban I, Ibrahim OW, Platt U. Reactive halogen chemistry in volcanic plumes. J Geophys Res. 2007. 10.1029/2006JD007206. [Google Scholar]
- 22.Lee C, Kim YJ, Tanimoto H, Bobrowski N, Platt U, Mori T, Yamamoto K, Hong CS. High ClO and ozone depletion observed in the plume of Sakurajima volcano, Japan. Geophys Res Lett. 2005. 10.1029/2005GL023785. [Google Scholar]
- 23.Zelenski M, Taran Y. Volcanic emissions of molecular chlorine. Geochim Cosmochim Acta. 2012. 10.1016/j.gca.2012.03.034. [Google Scholar]
- 24.Platt U, Bobrowski N, Butz A. Ground-based remote sensing and imaging of volcanic gases and quantitative determination of multi-species emission fluxes. Geosciences. 2018. 10.3390/geosciences8020044. [Google Scholar]
- 25.Gonzalez Abad G, Souri AH, Bak J, Chance K, Flynn LE, Krotkov NA, Lamsal L, Li C, Liu X, Miller CC, Nowlan CR, Suleiman R, Wang H. Five decades observing Earth’s atmospheric trace gases using ultraviolet and visible backscatter solar radiation from space. J Quant Spectrosc Radiat Transf. 2019. 10.1016/j.jqsrt.2019.04.030. [Google Scholar]
- 26.Marécal V, Voisin-Plessis R, Roberts TJ, Aiuppa A, Narivelo H, Hamer PD, et al. Halogen chemistry in volcanic plumes: a 1D framework based on MOCAGE 1D (version R1.18.1) preparing 3D global chemistry modelling. 2023. Geosci Model Dev. 10.5194/gmd-16-2873-2023.
- 27.Narivelo H, Hamer PD, Marécal V, Surl L, Roberts T, Pelletier S, et al. A regional modelling study of halogen chemistry within a volcanic plume of Mt Etna’s Christmas 2018 eruption. 2023. Atmos Chem Phys. 10.5194/acp-23-10533-2023.
- 28.Vaselli O, Tassi F, Montegrossi G, Capaccioni B. Sampling and analysis of volcanic gases. Acta Vulcanologica. 2006. 10.1400/93821.
- 29.Giggenbach WF. A simple method for the collection and analysis of volcanic gas samples. Bull Volcanol. 1975. 10.1007/BF02596953. [Google Scholar]
- 30.Finnegan DL, Kotra JP, Hermann DM, Zoller WH. The use of7LiOH-impregnated filters for the collection of acidic gases and analysis by instrumental neutron activation analysis. Bull Volcanol. 1989. 10.1007/BF01081977. [Google Scholar]
- 31.Gutmann A, Bobrowski N, Liotta M, Hoffmann T. Bromine speciation in volcanic plumes: new in situ derivatization LC-MS method for the determination of gaseous hydrogen bromide by gas diffusion denuder sampling. Atmos Meas Tech. 2021. 10.5194/amt-14-6395-2021.
- 32.Huang R-J, Hoffmann T. Development of a coupled diffusion denuder system combined with gas chromatography/mass spectrometry for the separation and quantification of molecular iodine and the activated iodine compounds iodine monochloride and hypoiodous acid in the marine atmosphere. Anal Chem. 2009. 10.1021/ac801839v. [DOI] [PubMed] [Google Scholar]
- 33.Rüdiger J, Bobrowski N, Liotta M, Hoffmann T. Development and application of a sampling method for the determination of reactive halogen species in volcanic gas emissions. Anal Bioanal Chem. 2017. 10.1007/s00216-017-0525-1. [DOI] [PubMed] [Google Scholar]
- 34.Wittmer J, Bobrowski N, Liotta M, Giuffrida G, Calabrese S, Platt U. Active alkaline traps to determine acidic‐gas ratios in volcanic plumes: sampling techniques and analytical methods. Geochem Geophys Geosyst. 2014. 10.1002/2013GC005133. [Google Scholar]
- 35.Asensio-Ramos M. Chromatographic techniques in volcanic surveillance: from gases to dissolved species. J Chromatogr Open. 2025. 10.1016/j.jcoa.2025.100270.
- 36.Karbach N, Bobrowski N, Hoffmann T. Observing volcanoes with drones: studies of volcanic plume chemistry with ultralight sensor systems. Sci Rep. 2022. 10.1038/s41598-022-21935-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Huang R-J, Hoffmann T. Diffusion technique for the generation of gaseous halogen standards. J Chromatogr A. 2010. 10.1016/j.chroma.2010.01.078. [DOI] [PubMed] [Google Scholar]
- 38.Flaconneche B, Martin J, Klopffer MH. Transport properties of gases in polymers: experimental methods. Oil & Gas Science and Technology - Rev IFP. 2001. 10.2516/ogst:2001022. [Google Scholar]
- 39.Crank J. The mathematics of diffusion: Part 4: diffusion in a plane sheet. 2nd ed. Oxford: Clarendon Press; 1975. [Google Scholar]
- 40.Hägg M-B. Membrane purification of Cl2 gas. J Membr Sci. 2000. 10.1016/S0376-7388(00)00455-5. [Google Scholar]
- 41.Muizebelt WJ, Nivard RJF. The mechanism of the iodine-catalyzed thermal isomerization of substituted ethylenes. Chem Commun (London). 1965. 10.1039/C19650000148. [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
The raw data supporting this study, including raw data with mass spectra for each stilbene and TMB reaction product, low level calibration files, raw data that support storage behaviour of cis-stilbene filters and permeation mass loss time series are available under: 10.5281/zenodo.18417414. Additional mass spectra for each product were added in the Supplemental Information.