Relationship between Voxel Value of Cone Beam CT and Hounsfield Unit of CT: Systematic Review and Meta-Analysis

Daryush Goodarzipour 1, Ahmad Reza Shamshiri 2, Zahra Khodadadi 3,*
PMCID: PMC13428513  PMID: 42542793

Abstract

This systematic review aims to investigate the relationship between voxel value obtained from Cone Beam Computed Tomography (CBCT) in studies compared to Hounsfield of Multidetector Computed Tomography (MDCT) in homogeneous and heterogeneous samples. A literature search was carried out in the databases PubMed, Scopus, Embase, and Web of Science searching for relevant literature until February 2022 (updated at July 2023). A risk of bias assessment of the studies was performed using a modified checklist based on the Cochrane Collaboration’s tool and the Journal of Biomedical Informatics. The software version 20.104 of MedCalc was used to conduct the meta-analysis of correlation coefficients. Out of 4750 articles in the initial search, 13 met the eligibility criteria. Out of the articles, eight studies were included in the meta-analysis. Both heterogeneous and homogenous samples showed a strong correlation between the voxel value of CBCT and Hounsfield Unit (HU), with high heterogeneity (r=0.900 and 0.998 respectively and I2>70%). Two other meta-analyses were conducted for kVp and voxel size, which showed a high correlation. The 95% confidence interval was used to present the estimated pooled correlation. The strong correlation of voxel value and HU indicates the possible potential of CBCT in radiographic bone density measurement. However, further research is needed to obtain an accurate conversion equation for translating voxel values of CBCT to HU. Calibration of voxel values within each scan using a reference object and consideration of both linear and non-linear regression could improve accuracy.

Key Words: Cone-Beam Computed Tomography, Hounsfield Unit, Voxel Value, Gray Scale

Introduction

Cone Beam Computed Tomography (CBCT) is widely used in dentistry as a complementary imaging technique to conventional methods. CBCT scan is an important imaging technique for implant treatment (evaluation of bone quantity before the surgery and examination of the adjacent vital structures such as the inferior alveolar canal and maxillary sinus), assessment of the maxillofacial lesions, trauma, and osteomyelitis. In addition, the voxel value of CBCT has been shown to aid in determining the type of bone for implant insertion, and prediction of implant stability. [1, 2]

The standard reference for radiographic measurement of Bone Material Density (BMD) is the Multidetector Computed Tomography (MDCT). BMD is used to evaluate the success of implant treatment and anchorage tools in orthodontics. [3-5] Hounsfield Unit (HU) of MDCT is a quantitative measurement of BMD and is proportional to the amount of X-ray attenuation and is assigned to each pixel to display an image. Materials with higher attenuation coefficient (higher density) have higher HUs. The standard formula for calculating the HU is as follows [6]:

HUmaterial =1000×μmaterial -μwater μwater

In CBCT, the attenuation rate is indicated by gray scale or voxel value. Although CBCT software manufacturers refer to the gray scale as HU, it is essential to note that these HU measurements are not necessarily accurate [1, 2].

CBCT, in addition to lower dose and cost and less image acquisition time, has better spatial resolution, contrast range, and gray scales than MDCT. However, due to the artifact associated with beam hardening, pyramid shape of X-ray, the lack of a standard method for the CBCT scanner, and the exposure parameters, it is unclear whether the voxel value can represent HU [7, 8].

MDCT units typically operate at 120 kV, while dental CBCT units operate within a nominal kV range of 70 to 120 kV. These different operating voltages result in varying effective energies between the two types of scanners, with CBCT being associated with higher levels of scatter radiation and artifacts compared to MDCT. This can make estimating bone density using CBCT more challenging. Despite these limitations, some studies have identified a linear relationship between the Hounsfield Units (HU) obtained from MDCT and the voxel values derived from CBCT. Moreover, authors have suggested that it may be possible to estimate bone density using the voxel values of CBCT by applying a standard conversion formula that relates gray values (GV) to HU [1, 7-12]. On the other hand, some other studies are concerned about the possibility of converting gray value to HU [6, 13, 14]. There are fundamental differences between MDCT and CBCT, which complicate the use of quantitative voxel value (e.g., size of field size (FOV), relatively high scatter radiation and limited reconstruction algorithms) [15].

There is no consensus about the use of the voxel value of CBCT for BMD assessment. This systematic review aims to investigate the relationship between voxel value obtained from CBCT in studies compared to the HU of MDCT in two groups of homogeneous and heterogeneous samples.

MATERIALS AND METHODS

This systematic review was done according to the Preferred Reporting Items for Systematic reviews and Meta-Analyses (PRISMA) checklist.

Study design:

This review involves in vitro and in vivo studies according to the following PICO: (P) Population: homogenous and heterogeneous samples, (I) Intervention: GV of CBCT scan, (C) Comparison: HU of MDCT, (O) Outcome: correlation of the GV and HU.

The homogenous samples include almost uniform-density materials, such as iodine solution. The heterogeneous group consists of samples with various densities throughout the material and subsequently various densities in the ROI.

Inclusion and exclusion criteria:

All the studies conducted on heterogeneous and homogeneous samples focusing on comparing the voxel value and HU are included. The included studies must use a gold standard to determine the HU: phantom or material with known HU or density, or MDCT scan. Additionally, they must report the correlation coefficient (r-value) or the coefficient of determination (R2-value) of GV of CBCT and HU.

Exclusion criteria are review studies, letter to editor, personal opinion, book chapter, conference paper, and studies with no gold standard or with a different aim.

Search strategy and Study selection:

PubMed, Scopus, Embase, and Web of Science databases were searched for relevant literature until February 25, 2022. The search was subsequently updated as of July 11, 2023. Also, a google scholar search was done for gray literature. The search strategy is mentioned in Supplementary 1. No date or language limit was applied.

The selection process of the studies was performed independently by two authors (DG, ZK): first, by title and abstract, followed by a review of the full text. Disagreements were resolved by a third opinion (AS).

Data collection process and data items:

All the included studies were evaluated according to the inclusion and exclusion criteria, and the key features were extracted: author(s), publication year, record number(s) (The sample size was considered the number of records used for r or R2 calculation, which is not the same as the patient or phantom’s number.), sample type (homogeneous or heterogeneous), CBCT machine brand, exposure parameters, r or R2-value(s) (determined by Pearson correlation test). If available, mean GV and HU were also noted.

Risk of bias assessment:

A modified checklist was designed based on the Cochrane Collaboration’s tool for assessing the risk of bias (RoB) in randomized trials and the Journal of Biomedical Informatics (JBI) to evaluate the risk of bias in the included studies. Since both checklists have applicable and non-applicable items for the purpose of the present study, the most relevant items were chosen. The modified checklist consists of five sections: performance of the study (blinding), reporting of outcomes, similarity of measurement method between the index test and standard reference, reliability of the measurement, and reproducibility of procedure (see Supplementary 2). The studies with 2 or 3 unclear scores in any section considered as unclear RoB and 4 or 5 unclear scores considered as high RoB. High score in any section indicated a high RoB.

Two authors (ZK and DG) cooperatively assessed Rob of the included studies. Any disagreements were resolved through the involvement of a third opinion (AS).

Statistical method:

The current study aimed to conduct a systematic review and meta-analysis of studies reporting r or R2-values and record’s number. Rather than using sample size, reported number of the records was extracted from tables, figures, or text for the correlation meta-analysis. In studies reported R2-values, the square root was computed and employed as the R-value for analysis purposes.

The present study involved the classification of data into two distinct groups, namely heterogeneous and homogeneous samples. Moreover, subgroup analyses were used to determine the potential sources of heterogeneity. By combining the Pearson correlation coefficients of the included studies, a forest plot was produced. Pearson correlation coefficients were classified as weak (0.1 to 0.3), moderate (0.4 to 0.7), and strong (0.8 to 1) based on their interpretation. This study utilized Cochran's Q test to determine the presence of statistical heterogeneity, with a P-value<0.10 indicating statistically significant heterogeneity. Additionally, the magnitude of statistical heterogeneity between studies was assessed through the use of I2, where values of 25%, 50%, and 75% were considered low, moderate, and high heterogeneity levels, respectively [   16 ]. The software version 20.104 of MedCalc was used to conduct the meta-analysis of correlation coefficients. The 95% confidence interval (CI) was used to present the estimated pooled correlation.

Results

Study selection:

The initial search on the database identified 4750 papers. After removing the duplication, 2511 citations remained. Following the title and abstract evaluation, 58 were selected for full-text assessment. One paper [17] was included through cross-reference of the included papers, and another paper [18] included by the search update. Finally, 13 papers [7,11,17-27] met the eligibility criteria and were included in the systematic review. A flowchart summarizing this process is shown in Figure 1.

Fig 1.

Fig 1

Preferred Reporting Items for Systematic Reviews (PRISMA) flow diagram for study selection

Study characteristics:

The included studies were conducted from 2010 to 2022, evaluating the correlation between CBCT voxel value and HU. They were evaluated in two groups of homogenous and heterogeneous samples, which were determined by the density distribution throughout the sample. The homogenous group consists of four papers [17,21,25,26] reporting 53 different scan protocols with the corresponding correlation and one study [22] reporting one R2-value for four scan protocols (Table 1). The heterogeneous group contains eight studies [7,11,18-20,23,24,27] and 21 different scan protocols (Table 2).

Table 1.

Summary of articles with homogeneous samples

Author CBCT Reference sample Number of records1 Correlation
(r- value)
Device KVp mAs Voxel size (μm) FOV (mm)
Chindasombatjaroen
(2011) [16]
Alphard Vega 3030 80 102 390 200×200 MDCT
(LightSpeed QX/i)
Iopamidol
(at 8 different concentrations)
8 0.9998
80 119 390 200×200 8 0.9998
80 153 390 200×200 8 0.9998
80 170 390 200×200 8 0.9998
80 204 390 200×200 8 0.9998
100 102 390 200×200 8 0.9998
100 119 390 200×200 8 0.9998
100 153 390 200×200 8 0.9998
100 170 390 200×200 8 0.9998
100 204 390 200×200 8 0.9999
Nomura
(2010) [17]
3D Accuitomo MDCT
(Somatom)
Iodine with different concentrations 12 0.9909
Pauwels
(2013) [18]
3D Accuitomo 170 90 155 250 17×12 MDCT
(Somatom)
PMMA2, air, aluminum, HA3 50 mg/cm3, HA100, and HA 200 5 0.9984
90 90 250 17×12 5 0.9982
90 155 80 6×6 5 0.9967
90 90 80 6×6 5 0.9969
3D Accuitomo XYZ 80 72 125 4×3 5 0.7014
GALILEOS Comfort 85 28 290 15×15 5 0.9965
i-CATH NG 120 36.5 300 23×16 5 0.9951
120 18.5 300 23×16 5 0.9948
120 36.5 250 16×13 5 0.9972
120 20 250 16×13 5 0.9885
120 18.5 400 16×13 5 0.998
120 10 400 16×13 5 0.9905
Kodak 9000 3D 70 110 76 5×3.6 5 0.7997
Kodak 9500 90 110 300 20×18 5 0.9959
90 110 200 14.5×8.3 5 0.9991
NewTom Vgi 110 180 240 12×8 5 0.9983
110 40 240 12×8 5 0.9982
PaX-Uni3D 85 120 200 5×5 5 0.9595
Picasso Trio 85 115.2 200 12×7 5 0.9957
Table 1 cont’d
85 72 300 12×7 5 0.9936
ProMax 3D 84 168 160 8×8 5 0.9942
84 168 320 8×8 5 0.9941
SCANORAH 3D 85 57 200 10×7.5 5 0.9967
85 30.4 200 10×7.5 5 0.9996
85 30.4 300 10×7.5 5 0.9982
SkyView 90 97.5 340 17×17 5 0.9958
90 52 340 17×17 5 0.9957
90 58.5 340 17×17 5 0.9947
Veraviewepocs 3D 70 51 125 8×8 5 0.7777
Pauwels
(2011) [19]
GALILEOS Comfort 85 28 300 150×150 MDCT
(GE Prospeed, SOMATOM)
PMMA, air, aluminum, HA 50 mg/cm3, HA100, and HA 200 5 0.998
i-CAT Classic High-dose 120 35 200 160×80 5 0.9973
i-CAT Classic Low-dose 120 10 400 160×80 5 0.987
ILUMA Elite 120 76 200 210×140 5 0.9946
ProMax 3D High-dose 84 168 160 80×80 5 0.9996
ProMax 3D Low-dose 84 20 320 80×80 5 0.9993
SCANORA 3D High-dose 85 36 130 60×60 5 0.9988
SCANORA 3D Low-dose 85 24 200 60×60 5 0.9993
SkyView High-dose 90 96 340 173×173 5 0.999
SkyView Low-dose 90 52 340 173×173 5 0.9986
Veraviewepocs 3D 70 51 130 80×80 5 0.6864
Haghanifar
(2017) [20]
SORDEX 89 75.6 NA 6×7 MDCT
(Somatom)
Iodine with different concentrations NA 0.9969
NEWTOM 110 6.912 NA 18×16 NA 0.9787
1

The number of records used by the author to calculate r-value;

2

Polymethyl methacrylate;

3

Hydroxyapatite

Table 2.

Summary of articles with heterogeneous samples

Ref.1 CBCT Reference Sample Records2 VV3 HU Diff.4 R5
Device KVp Voxel size
(μm)
FOV (mm)
[27] Soredex SCANORA 90 250 13 × 14.5 MSCT
(SOMATOM)
Cortical-
cancellous bone
(dry mandible)
100 1,053.31 744.35 308.96 0.977
Cancellous bone
(dry mandible)
100 816.62 572.45 244.17 0.931
Cortical bone
(dry mandible)
100 1505.26 1354 151.26 0.978
[24] Planmeca 90 NA NA MDCT
(Aquilion PRIME)
Head phantom 32 1098.162 1265.5 -167.34 0.911
[23] NewTom GIANO NA NA NA MDCT
(SOMATOM)
Bone blocks of cow ribs 52 1084.5 805.9 278.60 0.978
[7] Newtom 5G 110 150 120×80 MDCT
(Philips)
Cadaver
(human mandible)
80 773.28 348.25 425.03 0.968
[11] Accuitomo 170 90 80 40×40 MSCT
(Philips)
Cadaver
(human mandible)
20 377.49 222.85 154.64 0.89
[19] KaVo 3D eXam 120 400 NA MSCT
(SOMATOM)
Dens axis (Patient) 19 309.9 538.2 -228.30 0.812
Right maxilla (Patient) 19 318.4 468.7 -150.30 0.830
Left maxilla (Patient) 19 262.4 420 -157.60 0.848
Right mandible (Patient) 19 192 217.7 -25.70 0.648
Left mandible (Patient) 19 237.6 221.2 16.40 0.734
Right sphenoid (Patient) 19 -6.3 206 -212.30 0.842
Left sphenoid (Patient) 19 22 233.6 -211.60 0.818
Right mastoid (Patient) 19 -7.9 204.2 -212.10 0.8779
Left mastoid (Patient) 19 -89.9 6.8 -96.70 0.888
Table 2 cont’d
Right zygomatic bone (Patient) 19 488 705.6 -217.60 0.877
Left zygomatic bone (Patient) 19 242.6 556 -313.40 0.469
[20] Newtom VGi 110 300 NA MSCT
(SOMATOM)
Soft tissue (patient) NA NA NA NA 0.92
Hard tissue (patient) NA NA NA NA 0.86
General (patient) NA NA NA NA 0.95
[18] NewTom 90 NA 110×80 MSCT
(SOMATOM)
Goat head NA 355.14 257.25 97.89 0.496
1

Reference number;

2

The number of records used by the author to calculate r-value;

3

Voxel value;

4

Difference;

5

Correlation (r-value)

Risk of bias assessment:

Most of the included studies showed a low risk of bias for all evaluated scopes (selective reporting, similarity of the measurement methods, reference standard, reproducibility, and reliability of method), except the blindness, which was not mentioned in the studies (Table 3). The records number of Haghanifar et al.’s study [25] and Razi et al.’s study [20] and exposure factors of Khavid et al. [23] and Nomura et al.’s study[22] was not mentioned in the paper.

Table 3.

Risk of bias assessment of the included studies

Author Blinding Reporting Measurement Reliability Reproducibility
Chindasombatjaroen (2011) (26) U1 L2 L L L
Nomura (2010) (22) U L L L L
Pauwels (2013) (21) U L L L L
Pauwels (2011) (17) U L L L L
Cassetta (2013) (27) U L L L L
Kamaruddin (2016) (24) U L L L L
Khavid (2021) (23) U L L L L
Parsa (2012) (7) U L L L L
Parsa (2013) (11) U L L L L
Rostetter (2015) (19) U L L L L
Razi (2019) (20) U L L L L
Haghanifar (2017) (25) U L U U L
Gaur (2022) (18) U L U L L
1

Unclear;

2

Low

Synthesis of results

The included studies were grouped based on their samples, which were categorized as homogeneous or heterogeneous. In the homogeneous group, the maximum number of records was reported by Pauwels et al. (2013) [21] , while the minimum was reported by Nomura et al. [22] (145 and 12, respectively). In the heterogeneous group, Cassetta et al. [27] reported the maximum number of records, whereas Kamaruddin et al. reported the minimum (300 and 32, respectively). Regarding the r-values, Pauwels et al. (2013) [21] reported the minimum value (0.7014 for 3D Accuitomo XYZ, small FOV), whereas r-values ≥ 0.9990 were reported by Pauwels et al. (2011) [17] and Chindasombatjaroen et al [26] . In the heterogeneous group, Rostetter et al. [19] reported the minimum r-value (r-value = 0.469), while the maximum was reported by Cassetta et al. [27] and Khavid et al. [23] (r-value ≥ 0.97). Unfortunately, we were unable to determine the number of records in Haghanifar et al., Gaur et al. and Razi et al.'s paper [18,20,25] and the exposure parameters for Khavid et al. and Nomura et al.’s papers [22,23] either from the reported results or by corresponding with the authors. Therefore, meta-analysis was conducted with eight studies [11,17,19,21,24,26-28] .

The lowest correlation in the included studies was related to evaluation of left zygomatic bone of 19 patients (heterogeneous group) which was considerably different from its right counterpart (r-value= 0.877 (right side) compared to 0.469 (left side)).

Although the robust measurement method of the study, which involved merging DICOM datasets and using the mean of 10 dots as the final radiodensity of the anatomical area [25], this r-value seemed an outlier data and no explanation was found in the paper. Thus, it was excluded from the present meta-analysis.

Meta-analysis

The included studies that report kVp, mA, number of records and R2/r-value were used to conduct the meta-analysis. The homogeneous group exhibited a wide spectrum of tube voltage ranging between 70-120 kVp.

Consequently, this group was further partitioned into two subgroups based on tube voltage; low voltage (70-89kVp) and high voltage (90-120kVp), for more robust meta-analytical investigations. Due to the high heterogeneity of the studies, a random effect model was used. The result showed a high correlation in each group: 0.998 (95% CI: 0.996 to 0.999) for homogeneous samples, 0.997 (95% CI: 0.992 to 0.999) for homogeneous samples with 70-89kVp, 0.999 (95% CI: 0.997 to 0.999) for homogeneous samples with 90-120kVp, 0.900 (95% CI: 0.843 to 0.937) for heterogeneous samples, and 0.993 (95% CI: 0.989 to 0.996) for all studies. All groups were found to be heterogeneous, with significant Q-test result (P<0.10) and inconsistency (I2) values ranging from 41.48%to 86.95%. With the exception of the high voltage homogeneous sample group with an I2 value of 41.48%, the inconsistency test results for all groups were high (I2>64%) (Table 4).

Table 4.

Correlation meta-analysis of voxel value of CBCT and HU

Subgroup variable Number of records Correlation coefficient
(random effect)
95% CI Significance level
(p-Value)
Inconsistency
(I2)
Sample type
Heterogeneous samples 622 0.900 0.843 to 0.937 <0.001 86.95%
Homogeneous samples 280 0.998 0.996 to 0.999 <0.001 64.72%
Homogeneous samples (70-89kVp) 130 0.997 0.992 to 0.999 <0.001 76.35%
Homogeneous samples (90-120kVp) 150 0.999 0.997 to 0.999 <0.001 41.48%
Voltage (kVp)
70-89 130 0.997 0.992 to 0.999 <0.001 76.35%
90-120 772 0.988 0.979 to 0.993 <0.001 89.70%
Total 902 0.993 0.988 to 0.996 <0.001 89.85%
Voxel size
75-250µm 515 0.988 0.979 to 0.993 <0.001 75.87%
250<µm 355 0.995 0.987 to 0.998 <0.001 93.05%
Total 870 0.993 0.989 to 0.996 <0.001 89.85%

Additionally, two other meta-analyses were conducted for kVp and voxel size. The included studies were grouped based on the scan voltage: 70-89, and 90-120kVp, and due to the high heterogeneity of the data, random effect meta-analyses were used. The correlation coefficient was 0.997 (95% CI: 0.992 to 0.999), and 0.988 (95% CI: 0.979 to 0.993), respectively. Also, correlation meta-analyses of voxel size were 0.988 (95% CI: 0.979 to 0.993) for 75-250µm voxel size and 0.995 (95% CI: 0.987 to 0.998) for 250<µm voxel size (Table 4).

Forest plot for correlations of GV and HU in homogeneous samples with 70-89kVp and 90-120kVp is shown in Figure 2 and 3, respectively. Forest plot for correlations of GV and HU in heterogeneous samples is shown in Figure 4. The plot size is proportional to the precision of the estimated effect sizes (r-values), and the bars indicate the corresponding 95 % CIs. The diamond is placed on the summary correlation coefficient of the observational studies.

Fig 2.

Fig 2

Forest plot of homogeneous sample with 70-89kVp

Fig 3.

Fig 3

Forest plot of homogeneous sample with 90-120kVp

The overall summary estimate suggests a moderately strong positive correlation between GV and HU. However, the wide confidence interval around this estimate indicates that there is substantial uncertainty about the true underlying effect size. This may be attributed to the high heterogeneity in the studies and small sample size.

The subgroup analyses conducted suggest that sample type, low tube voltage and tube current may be sources of heterogeneity, as evidenced by the variation in Q and I2 values across different groups, which could explain some of the observed variability in effect sizes across studies. Tube voltage at 70kVp was found to be particularly problematic, resulting in low correlation in some exposure protocols with homogeneous samples (r-value < 0.8) [17,21]. When groups with identical tube voltage (90-120kVp) were compared for their heterogeneity, including homogeneous and heterogeneous samples as well as the total, it was found that the homogeneous samples exhibited the lowest inconsistency. (Table 4) Voxel size was also investigated; however, the findings remained inconclusive due to the presence of both protocol scans with lower correlation for voxel sizes less than 130 µm and those with a 400 µm voxel size. Certainly, there are other sources of heterogeneity that cannot be investigated thoroughly in this meta-analysis, such as differences in measurement methods, other parameters of exposure, variations in study conditions, and discrepancies between CBCT units from similar and different commercial brands.

Fig 4.

Fig 4

Forest plot of heterogeneous sample

Discussion

CBCT is a widely used modality in dentistry, especially for preoperative implant planning. However, the reliability of the GV of CBCT has been questioned due to increased noise level, inherent scattering artifacts of the scanning technology, limitation of currently applied reconstruction algorithms that leads to lower contrast resolution than MDCT [6,7]. This systematic review and meta-analysis investigated the available evidence on the correlation of GV of CBCT and HU of MDCT as the reference standard, which demonstrated a moderately strong correlation (r=0.993). In addition to the overall meta-analysis, data were analyzed in subgroups of heterogeneity/ homogeneity sample and some exposure parameters.

Tube voltage is one of the most important exposure parameters that affect GV. In a literature review, El-Tabarany et al. classified this parameter as a significant factor [29]. In the present review, kVp of the studies were evaluated in two groups (70-89, 90-120kVp), and meta-analysis of the data revealed an almost similar correlation in both groups (r-value= 0.997, and 0.988, respectively). Also, it was found that the tube voltage served as a source of heterogeneity in the dataset of this study and was found to be associated with lower correlation in some exposure protocols (specifically, at 70kVp) [21]. In their study, Chindasombatjaroen et al. [26] found that increasing the tube voltage in MDCT led to a decrease in HU value, while tube current did not have a significant effect on density measurement. On the other hand, both tube current and tube voltage affect GV in CBCT scans, with an increase in either resulting in a decrease in GV. Additionally, Vasconcelos et al. demonstrated that higher kVp and the Metal Artifact Reduction (MAR) algorithm can reduce metal artifacts, resulting in less deviated GV [30]. Similarly, CBCT unites with high filtration and high kVp may decrease the risk of beam hardening artifacts, ultimately leading to more accurate measurements of attenuation coefficients and therefore GV [31].

Voxel size is another element that may influence the CBCT gray value, contributing to spatial resolution. Smaller voxel size enables CBCT images to exhibit fine details but at the cost of higher image noise, which indiscriminately affects the mean voxel value [11]. Voxel size is usually attached to the other exposure parameters, such as FOV and kVp. Pauwels et al. [21], evaluating the impact of different voxel sizes of 13 CBCT devices (ranging from 0.125 to 0.4mm) noted that this factor did not affect the accuracy of CBCT voxel value. Due to the variability of scan protocols (kVp, mA, FOV, etc.) and methodologies across the enrolled studies, this paper could not draw conclusive findings on voxel size.

Regarding the region of interest (ROI), fusing the MDCT and CBCT can reduce the possible human errors in the measurement of voxel value. Among the twelve studies included in this review, five employed observer-independent software to ensure accurate matching of CBCT and MDCT data, including four studies with heterogeneous samples [7,19,24,27] and one study with homogeneous samples [26]. Two other heterogeneous sample studies [11,20] used reference points to merge CBCT and MDCT images. When evaluating anatomical structures of patients who may have different head positions during scanning, determining the ROI is particularly crucial due to increased susceptibility to metal and motion artifacts. The studies with heterogeneous samples included evaluations of cadaver [7,11], head phantom [24], cow bone block [23], patient [19,20], and dry mandible [27]. However, these sample types have limitations that may deviated from the clinical situation and should be interpreted with caution; for instance, dry bone samples lack soft tissue and bone moisture, and the homogeneity of different samples in head phantoms while the physiological tissue are heterogeneous [7,20]. Homogeneous sample group evaluated different concentrations of media. In two studies [17,21], in addition to three different concentrations of hydroxyapatite, aluminum and air were evaluated as object with very high and very low density, respectively, which were associated with higher correlation of CBCT voxel value and HU. On the other hand, voxel value of object with medium density showed larger error value and lower correlation. In another study [22], the samples included aluminum, but its correlation was not evaluated separately. In the homogeneous sample group, two studies using the same phantom, filled the sample container with polymethyl methacrylate (PMMA) and another study [22] used a water phantom to simulate a human head.

To the best of our knowledge, this is the first systematic review conducting a meta-analysis of the correlation of CBCT voxel value and MDCT HU in two groups of homogenous and heterogeneous samples. The heterogeneous samples revealed a lower correlation between CBCT voxel value and HU with higher heterogeneity of the studies (r= 0.900, I2=86.95%). Furthermore, when comparing heterogeneous samples with high voltage homogeneous samples (where the tube voltage for both groups was identical and between 90-120 kVp), it was observed that the second group exhibited less heterogeneity in the studies (I2 = 41.48% compared to 86.95%). These findings suggest that type of sample could be a potential source of heterogeneity in the studies, which has an impact on the correlation of GV and HU. It can be attributed to the presence of different densities along the FOV, such as soft tissue and adjacent anatomical structures, which makes the sample more similar to the clinical situation. Also, different densities such as bony lesions or anatomic structures adjacent to implants cause random effect of the noise on the measured gray value [21]. Furthermore, it must be considered that reconstruction algorithms assign gray values based on the amount and distribution of densities in the FOV to achieve an optimized contrast. To put it in another way, the presence of higher densities shifts the histogram, leading to a decrease the gray values in the reconstructed image [21]. For this problem, using a reference object in the FOV with known density could make it possible to calibrate the histogram, which results in a more accurate gray value [21,32]. Additionally, Komori et al. investigating ossicular chain and inner ear soft tissue, proposed to use fixed gray value thresholds for each purpose exclusively to standardized volume-rendering of CBCT images [33].

There have been two previous systematic reviews conducted on this topic. A recent study evaluating these two variables showed a positive correlation (r=0.669). The present study searched more databases and used a larger sample from the included papers to perform a meta-analysis. Our result is consistent with Selvaraj et al. [34] who found a good correlation between CBCT voxel value and HU, although we found a higher correlation (r=0.993), which may be attributed to a larger sample size and also, involving homogeneous samples in this study which showed a stronger correlation. In addition, Eguren et al. systematically reviewed the possibility of GV conversion to HU. They defined three steps: equipment calibration, correlation, prediction equation model, and standard formula. This study mentioned a lack of CBCT standardization and calibration as an obstacle to voxel value conversion to HU [5]. Most of the studies included in this review revealed a strong correlation and linear relationship between HU and GV across different densities and in both homogeneous and heterogeneous samples. However, Rostetter et al. [19] evaluating different anatomical structures of patients, reported a linear relationship, but noted that the strength of this relationship depends on the anatomical position in the head and its corresponding density. In addition, Nomura et al. [22] found that both linear and non-linear regression models demonstrated high r-value, but non-linear regression provided a better fit to the plots.

In the terms of comparing the mean values of HU and GV, most of the studies included in this review reported higher HU values than GV [7,19,22.24.26], while three studies [18,23,25,27] reported lower HU values in comparison to GVs. However, only two of these studies which belong to the first group (HU values > GV), had the same exposure factors in MDCT and CBCT scans. Rostetter et al. [19] used the same tube voltage for both CBCT and MDCT scans, while Chindasombatjaroen et al. [26] compared scan protocols with the same tube current and tube voltage. It could be attributed to multiple factors, including elevated scatter radiation levels, noise, and artifacts of CBCT scan technology. When reconstructing images, scatter radiation can lead to an underestimation of absorption, which is dependent on the object and proportional to the amount of scatter present. To minimize radiation exposure, CBCT machines use lower tube currents than medical CT, leading to a lower signal-to-noise ratio and increased noise levels, resulting in more inconsistency and larger standard deviations in GVs. Furthermore, as CBCT acquires a larger volume compared to highly collimated fan beam medical CT, artifacts have a more pronounced effect on image quality [35].

This systematic review aimed to evaluate the correlation of GV and HU, and for this purpose, articles reporting the correlation coefficient (r-value) and coefficient of determination (R2-value) were included. However, it is essential to distinguish between them. The correlation coefficient (r-value) indicates the strength and the direction of a linear relationship between two variables ranging between -1 and 1. The coefficient of determination (R2-value) assesses how well a model predicts an outcome ranging between 0 and 1. In other words, less scattered data point higher R2 value [6] . Although the r or R2 value of the literature regarding the correlation of GV and HU is typically high (Tables 1 and 2), it should be kept in mind that this is not necessarily proof of the accuracy of GV and must be interpreted with caution. Despite the high correlation found in this meta-analysis, significant deviation (25.7-313.40 numerical value in the included studies, Table 2) in GV makes it unreliable for clinical use. Only at R2 values close to 1 (>0.999), these deviations are small enough to consider clinical use [6]. Eguren et al.'s systematic review [5] found a high correlation between actual and expected GV but with significant variability which was 500 GV numerical value with a deviation of 20%.

Bone density assessment is crucial in implant placement and the ultimate success of the procedure. A classification for jaw bone density first proposed by Misch is based on the HU and defined in five groups: D1: >1250, D2: 850–1250, D3: 350–850, D4: 350-0, and D5<0 HU. [36] Considering these ranges, in the case of a slight deviation of voxel value below the proposed ranges, it would be possible to use CBCT for implant planning. Pauwels et al. [21] have investigated 13 different CBCT machines and stated that some exposure protocols demonstrate reasonable accuracy for bone density estimation. However, the authors did not recommend differential diagnosis of lesions by quantitative assessment of CBCT voxel value. It is in line with Etoz et al. [37], that found no relationship between the voxel value of CBCT and pathological diagnosis of apical cyst and granuloma.

The main limitation of this study was the heterogeneity of the methods, samples, and data. Additionally, all the information was not mentioned in some studies, such as the interval between CBCT and MDCT, the conditions under which viewers assessed bone density, and whether these modifications could affect radiologic bone density. Furthermore, CBCT devices are not identical, and there are noticeable differences in exposure protocols, software, and hardware.

CONCLUSION

In conclusion, the strong correlation between voxel values and HU indicates the potential for CBCT in measuring radiographic bone density. However, further research is needed to obtain an accurate conversion equation for translating voxel values of CBCT to HU. Calibration of voxel values within each scan using a reference object and consideration of both linear and non-linear regression could improve the accuracy of voxel value measurement. Despite existing variations, some trends can be observed, although the translation of GV to HU remains debatable and requires more consistent scientific evidence.

Acknowledgment

This study was supported by Tehran University of Medical Sciences, School of Dentistry.

Notes:

Cite this article as: Goodarzipour D, Shamshiri AR, Khodadadi Z. Relationship between Voxel Value of Cone Beam CT and Hounsfield Unit of CT: Systematic Review and Meta-Analysis. Front Dent. 2026:23:05. http://doi.org/10.18502/fid.v23i5.20888

CONFLICT OF INTEREST STATEMENT

None declared.

References

  • 1.Razi T, Niknami M, Alavi Ghazani F. Relationship between Hounsfield Unit in CT Scan and Gray Scale in CBCT. J Dent Res Dent Clin Dent Prospects. 2014 Spring;8(2):107–10. doi: 10.5681/joddd.2014.019. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Mallya S, Lam E. St. Louis, United states: Elsevier; 2019. White and Pharoah's Oral radiology: principles and interpretation; 659 pp. [Google Scholar]
  • 3.Shapurian T, Damoulis PD, Reiser GM, Griffin TJ, Rand WM. Quantitative evaluation of bone density using the Hounsfield index. Int J Oral Maxillofac Implants. 2006 Mar-Apr;21(2):290–7. [PubMed] [Google Scholar]
  • 4.Shahlaie M, Gantes B, Schulz E, Riggs M, Crigger M. Bone density assessments of dental implant sites: 1. Quantitative computed tomography. Int J Oral Maxillofac Implants. 2003 Mar-Apr;18(2):224–31. [PubMed] [Google Scholar]
  • 5.Eguren M, Holguin A, Diaz K, Vidalon J, Linan C, Pacheco-Pereira C, et al. Can gray values be converted to Hounsfield units? A systematic review. Dentomaxillofac Radiol. 2022 Jan 1;51(1):20210140. doi: 10.1259/dmfr.20210140. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Pauwels R, Jacobs R, Singer SR, Mupparapu M. CBCT-based bone quality assessment: are Hounsfield units applicable? . Dentomaxillofac Radiol. 2015;44(1):20140238. doi: 10.1259/dmfr.20140238. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Parsa A, Ibrahim N, Hassan B, Motroni A, van der Stelt P, Wismeijer D. Reliability of voxel gray values in cone beam computed tomography for preoperative implant planning assessment. Int J Oral Maxillofac Implants. 2012 Nov-Dec;27(6):1438–42. [PubMed] [Google Scholar]
  • 8.Valiyaparambil JV, Yamany I, Ortiz D, Shafer DM, Pendrys D, Freilich M, et al. Bone quality evaluation: comparison of cone beam computed tomography and subjective surgical assessment. Int J Oral Maxillofac Implants. 2012 Sep-Oct;27(5):1271–7. [PubMed] [Google Scholar]
  • 9.Mah P, Reeves TE, McDavid WD. Deriving Hounsfield units using grey levels in cone beam computed tomography. Dentomaxillofac Radiol. 2010 Sep;39(6):323–35. doi: 10.1259/dmfr/19603304. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Aranyarachkul P, Caruso J, Gantes B, Schulz E, Riggs M, Dus I, et al. Bone density assessments of dental implant sites: 2. Quantitative cone-beam computerized tomography. Int J Oral Maxillofac Implants. 2005 May-Jun;20(3):416–24. [PubMed] [Google Scholar]
  • 11.Parsa A, Ibrahim N, Hassan B, van der Stelt P, Wismeijer D. Bone quality evaluation at dental implant site using multislice CT, micro-CT, and cone beam CT. Clin Oral Implants Res. 2015;26(1):e1–7. doi: 10.1111/clr.12315. [DOI] [PubMed] [Google Scholar]
  • 12.Dahiya K, Kumar N, Bajaj P, Sharma A, Sikka R, Dahiya S. Qualitative Assessment of Reliability of Cone-beam Computed Tomography in evaluating Bone Density at Posterior Mandibular Implant Site. J Contemp Dent Pract. 2018 Apr 1;19(4):426–430. [PubMed] [Google Scholar]
  • 13.Scarfe WC, Angelopoulos C. Maxillofacial cone beam computed tomography: principles, techniques and clinical applications. Switzerland: Springer; 2018. [Google Scholar]
  • 14.Silva IM, Freitas DQ, Ambrosano GM, Bóscolo FN, Almeida SM. Bone density: comparative evaluation of Hounsfield units in multislice and cone-beam computed tomography. Braz Oral Res. 2012 Nov-Dec;26(6):550–6. doi: 10.1590/s1806-83242012000600011. [DOI] [PubMed] [Google Scholar]
  • 15.González-García R, Monje F. The reliability of cone-beam computed tomography to assess bone density at dental implant recipient sites: a histomorphometric analysis by micro-CT. Clin Oral Implants Res. 2013 Aug;24(8):871–9. doi: 10.1111/j.1600-0501.2011.02390.x. [DOI] [PubMed] [Google Scholar]
  • 16.Borenstein M, Hedges LV, Higgins JP, Rothstein HR. Introduction to meta-analysis. West Sussex, United Kingdom: John Wiley & Sons Ltd. Chichester; 2009. 421 pp. [Google Scholar]
  • 17.Pauwels R, Stamatakis H, Manousaridis G, Walker A, Michielsen K, Bosmans H, et al. Development and applicability of a quality control phantom for dental cone-beam CT. J Appl Clin Med Phys. 2011 Nov 15;12(4):3478. doi: 10.1120/jacmp.v12i4.3478. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Gaur A, Dhillon M, Puri N, Sethi Ahuja U, Rathore A. Questionable accuracy of CBCT in determining bone density: A comparative CBCT-CT in vitro study. Dent Med Probl. 2022 Jul-Sep;59(3):413–419. doi: 10.17219/dmp/143504. [DOI] [PubMed] [Google Scholar]
  • 19.Rostetter C, Metzler P, Schenkel JS, Seifert B, Luebbers HT. Comparison of in vivo cone-beam and multidetector computed tomographic scans by three-dimensional merging software. Br J Oral Maxillofac Surg. 2015 Dec;53(10):1021–6. doi: 10.1016/j.bjoms.2015.09.030. [DOI] [PubMed] [Google Scholar]
  • 20.Razi T, Emamverdizadeh P, Nilavar N, Razi S. Comparison of the Hounsfield unit in CT scan with the gray level in cone-beam CT. J Dent Res Dent Clin Dent Prospects. 2019 Summer;13(3):177–182. doi: 10.15171/joddd.2019.028. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Pauwels R, Nackaerts O, Bellaiche N, Stamatakis H, Tsiklakis K, Walker A, et al. Variability of dental cone beam CT grey values for density estimations. Br J Radiol. 2013 Jan;86(1021):20120135. doi: 10.1259/bjr.20120135. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Nomura Y, Watanabe H, Honda E, Kurabayashi T. Reliability of voxel values from cone-beam computed tomography for dental use in evaluating bone mineral density. Clin Oral Implants Res. 2010 May;21(5):558–62. doi: 10.1111/j.1600-0501.2009.01896.x. [DOI] [PubMed] [Google Scholar]
  • 23.Khavid A, Sametzadeh M, Godiny M, Moarrefpour MM. Comparison of the hounsfield unit values obtained from cone-beam computed tomography (CBCT) and multidetector computed tomography (MDCT) images for different bone densities. J Contemp Med Sci. 2021 Mar;7(2):92–5. [Google Scholar]
  • 24.Kamaruddin N, Rajion ZA, Yusof A, Aziz ME. Relationship between Hounsfield unit in CT scan and gray scale in CBCT. AIP Conf Proc. 2016 December 19 ;1791(1):020005. [Google Scholar]
  • 25.Haghanifar S, Yousefi S, Moudi E, Abesi F, Bijani A, Moghadamnia AA, et al. Accuracy of densitometry of two cone beam computed tomography equipment in comparison with computed tomography. Electron Physician. 2017 May 25;9(5):4384–4390. doi: 10.19082/4384. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Chindasombatjaroen J, Kakimoto N, Shimamoto H, Murakami S, Furukawa S. Correlation between pixel values in a cone-beam computed tomographic scanner and the computed tomographic values in a multidetector row computed tomographic scanner. J Comput Assist Tomogr. 2011 Sep-Oct;35(5):662–5. doi: 10.1097/RCT.0b013e31822d9725. [DOI] [PubMed] [Google Scholar]
  • 27.Cassetta M, Stefanelli LV, Pacifici A, Pacifici L, Barbato E. How accurate is CBCT in measuring bone density? A comparative CBCT-CT in vitro study. Clin Implant Dent Relat Res. 2014 Aug;16(4):471–8. doi: 10.1111/cid.12027. [DOI] [PubMed] [Google Scholar]
  • 28.Parsa A, Ibrahim N, Hassan B, Motroni A, van der Stelt P, Wismeijer D. Influence of cone beam CT scanning parameters on grey value measurements at an implant site. Dentomaxillofac Radiol. 2013;42(3):79884780. doi: 10.1259/dmfr/79884780. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.El-Tabarany KS, Aboulfotouh MM, Ashmawy MS. VALIDITY OF CONE BEAM COMPUTED TOMOGRAPHY VOXEL DENSITY VALUES: A REVIEW. Int J Adv. Res. 2018 Sep;6(9):103–121. [Google Scholar]
  • 30.Vasconcelos TV, Leandro Nascimento EH, Bechara BB, Freitas DQ, Noujeim M. Influence of Cone Beam Computed Tomography Settings on Implant Artifact Production: Zirconia and Titanium. Int J Oral Maxillofac Implants. 2019 Sep/Oct;34(5):1114–1120. doi: 10.11607/jomi.7129. [DOI] [PubMed] [Google Scholar]
  • 31.Molteni R. Prospects and challenges of rendering tissue density in Hounsfield units for cone beam computed tomography. Oral Surg Oral Med Oral Pathol Oral Radiol. 2013 Jul;116(1):105–19. doi: 10.1016/j.oooo.2013.04.013. [DOI] [PubMed] [Google Scholar]
  • 32.Reeves TE, Mah P, McDavid WD. Deriving Hounsfield units using grey levels in cone beam CT: a clinical application. Dentomaxillofac Radiol. 2012 Sep;41(6):500–8. doi: 10.1259/dmfr/31640433. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Komori M, Miuchi S, Hyodo J, Kobayashi T, Hyodo M. The gray scale value of ear tissues undergoing volume-rendering high-resolution cone-beam computed tomography. Auris Nasus Larynx. 2018 Oct;45(5):971–979. doi: 10.1016/j.anl.2018.01.012. [DOI] [PubMed] [Google Scholar]
  • 34.Selvaraj A, Jain RK, Nagi R, Balasubramaniam A. Correlation between gray values of cone-beam computed tomograms and Hounsfield units of computed tomograms: A systematic review and meta-analysis. Imaging Sci Dent. 2022 Jun;52(2):133–140. doi: 10.5624/isd.20210274. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Schulze R, Heil U, Gross D, Bruellmann DD, Dranischnikow E, Schwanecke U, et al. Artefacts in CBCT: a review. Dentomaxillofac Radiol. 2011 Jul;40(5):265–73. doi: 10.1259/dmfr/30642039. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Resnik R. Misch's contemporary implant dentistry. 4th ed. Vol. 2021. St. Louis, United states: Elsevier; 2020 . 1241 pp. [Google Scholar]
  • 37.Etöz M, Amuk M, Avcı F, Yabacı A. Investigation of the effectiveness of CBCT and gray scale values in the differential diagnosis of apical cysts and granulomas. Oral Radiol. 2021 Jan;37(1):109–117. doi: 10.1007/s11282-020-00459-6. [DOI] [PubMed] [Google Scholar]

Articles from Frontiers in Dentistry are provided here courtesy of Tehran University of Medical Sciences

RESOURCES