Abstract
Background
During ischemic cardiogenic shock (CS), increased afterload overburdens the damaged myocardium and may worsen outcomes. The measure, effective arterial elastance (Ea), integrates systemic vascular resistance, arterial compliance, and heart rate, and reflects left ventricular (LV) afterload. While Ea is traditionally derived from invasive LV pressure–volume (PV) analysis, a simplified bedside formula (0.9 × systolic arterial pressure/stroke volume) has been validated in stable intensive care patients. Its accuracy in CS, characterized by rapidly evolving hemodynamics, remains unknown.
Methods
In total, 20 female Danish Landrace pigs were instrumented with a femoral arterial catheter, pulmonary artery catheter (PAC), and LV PV catheter. CS was induced by microsphere embolization of the left main coronary artery. Ea was estimated from femoral arterial pressure and stroke volume measured with the PAC and compared with PV-derived Ea as a reference. Ea was measured at healthy baseline, CS onset, 60 minutes of untreated CS and after a dobutamine infusion. Agreement was evaluated with Pearson correlation and Bland–Altman analysis.
Results
At baseline, onset of CS, and at established CS, correlations between the estimated and the PV-derived Ea remained strong (r = 0.68–0.74, all P ≤ 0.002). During dobutamine infusion, correlation persisted (r = 0.63, P = 0.020). Bland–Altman analysis showed negligible bias at healthy baseline and CS onset, slight underestimation during established CS, and modest overestimation during dobutamine-treated CS, with the widest limits of agreement observed during established untreated CS.
Conclusions
Estimated Ea reliably reproduced PV-derived Ea during disease states. This simple bedside measure may enable continuous monitoring of afterload in ischemic CS without LV catheterization.
Keywords: acute coronary syndromes, afterload, arterial elastance, cardiogenic shock, hemodynamics, pressure–volume loops
Background
Cardiogenic shock (CS) following myocardial ischemia remains a syndrome with high in-hospital mortality. CS is commonly caused by coronary occlusion, resulting in abrupt loss of left ventricular (LV) contractile force and stroke volume (SV). To maintain end-organ perfusion, reflex sympathetic vasoconstriction occurs. This increases systemic vascular resistance (SVR) (1) but at the cost of increased LV afterload and diminished forward blood flow. These hemodynamic abnormalities are complex and need to be reliably characterized and monitored for appropriate hemodynamic supportive treatment. In CS, elevated afterload overburdens damaged myocardium and is associated with adverse outcomes, increased oxygen consumption, and aggravation of heart failure (2). Furthermore, vasoactive drugs used postcardiac arrest or during treatment of CS may alter LV afterload. Inopressors such as dopamine or norepinephrine may increase afterload, which may further aggravate cardiac function (3), while inodilators such as dobutamine or milrinone may decrease afterload. Thus, reliable estimation and monitoring of afterload hold potential to improve patient care.
For the assessment of afterload, the effective arterial elastance (Ea) is of particular interest (4, 5) as it incorporates SVR, arterial compliance, and heart rate (HR). Ea is traditionally measured using an invasive pressure–volume (PV) catheter in the LV and is considered a more precise reflection of LV afterload compared with arterial pressure or SVR (6, 7). However, LV catheterization is still highly specialized and not yet utilized in clinical practice, and thus, cannot be routinely used for assessment of Ea. Consequently, other methods for Ea estimation are needed. Ea can be approximated as the LV end-systolic pressure (ESP) to SV ratio, and thus, needs a reliable method of estimation of ESP. Therefore, translating the Ea concept to the bedside may suffer from approximations, including the estimation of ESP from arterial pressure. The literature lacks validation of such an estimation of Ea in patients with CS; however, recently, a bedside formula has been validated in hemodynamically stable intensive care patients using femoral arterial pressure (8). Yet the performance of this formula under the extreme, rapidly evolving hemodynamics of ischemic CS and vasoactive treatment is unknown. Hence, a knowledge gap exists regarding the value of bedside estimated Ea (Eaestimated) as a real-time surrogate of afterload during the precipitous decline from normal hemodynamics to overt CS when SV, SVR, and HR all change simultaneously.
Here, we present the first systematic validation of the bedside Ea estimation in a well-established large-animal CS model. We compare the estimated Ea calculated from femoral arterial pressure and SV measurement with the reference Ea derived from simultaneous high-fidelity LV PV measurements in animals in a healthy state and following ischemic CS.
Methods
Study design
In total, 20 female Danish Landrace pigs were examined. The study was conducted after obtaining the necessary authorization from the Danish National Animal Experiment Inspectorate. This study was conducted using animals that were part of a previously established research protocol in our laboratory at Aarhus University Hospital, ensuring maximum statistical power while adhering to the 3R principles aimed at minimizing animal use in research (9–11). The anesthesia protocol is described elsewhere (9–11). Ischemic CS was induced through repeated injections of polyvinyl microspheres (Contour, Boston Scientific, USA) into the left main coronary artery. CS was considered present with a 30% reduction in cardiac output (CO), measured using thermodilution, or with a 30% reduction in the perfusion marker mixed venous oxygen saturation (SvO2), measured using a blood gas analyzer.
Hemodynamic monitoring
Briefly, cardiac output was obtained by the standard bolus thermodilution technique using a pulmonary artery catheter with three 10 mL boluses of cold saline injection into the right atrium, and the thermistor at the distal tip of the catheter recorded the resulting temperature-time curve. The average of three consecutive measurements with less than 10% variation was used. These cutoffs were chosen as they result in stable LV CS with marked contractile dysfunction, elevated afterload, elevated filling pressures, hypotension, and hypoperfusion without the need for immediate treatment escalation (1). The animals were instrumented prior to the induction of CS (Figure 1). Femoral artery pressure was continuously monitored. SV was estimated as cardiac output divided by heart rate. SVR was calculated using the standard formula. Pulmonary artery wedge pressure was also measured.
Figure 1.
Animal instrumentation. Animals were instrumented prior to CS induction with a pulmonary artery catheter, a left ventricle pressure–volume admittance catheter, and a femoral arterial catheter. Arterial elastance (Ea) was calculated as described.
Pressure–volume measurements
An LV PV admittance catheter (Transsonic, USA) was inserted through a carotid artery and used to obtain the standard reference Ea, as described elsewhere (1, 10). Briefly, the admittance-based PV catheter recorded instantaneous LV pressure and volume at high temporal resolution throughout each cardiac cycle. Ea was computed automatically using the Labchart 8 Pro software (AdInstruments, Australia). The automated software works as follows. ESP was identified as the LV pressure at the point of minimal LV volume, corresponding to the upper-left corner of the PV loop. SV was derived as the difference between the end-diastolic and end-systolic volumes. Reference Ea was then computed as ESP/SV, representing the slope of the line connecting the end-diastolic PV point to the end-systolic PV point and reflecting the net arterial load imposed on the LV per unit of ejected volume. LV end-systolic elastance (Ees), reflecting LV chamber contractility, was calculated from the PV-loop data as LVESP/(ESV−V0), with V0 derived from baseline PV-loop calibration and kept constant across serial measurements. Ventricular–arterial coupling (VAC) was then calculated as Ea/Ees, thereby relating arterial load to LV contractile function. LV end-diastolic pressure was also measured.
Eaestimated was calculated using the formula 0.9 × systolic femoral pressure/SV (8). The agreement between estimated and reference Ea was evaluated at the following four timepoints: (i) in a healthy state, (ii) immediately following onset of ischemic CS, (iii) following 60 min of untreated ischemic CS, and (iv) after 15 min of an intravenous infusion of dobutamine 5 µg/kg/min after 3 h of untreated CS (Figure 2). Data analysis and graphical representations were performed using R software (version 4.2.1, RStudio, PBC) with Pearson correlation [95% confidence intervals (CIs)] and illustrated by Bland–Altman analysis stratified by predefined experimental timepoint separately at healthy baseline, CS onset, established untreated CS, and dobutamine-treated CS, reporting mean bias and 95% limits of agreement (LoA) after the exclusion of observations lying beyond 1.5 × IQR. Two-sided P-values <0.05 were deemed significant. Data are summarized as mean ± SD.
Figure 2.
Study design. Hemodynamic measurements were first obtained at a healthy baseline. CS was then induced by repeated injections of microspheres into the left main coronary artery until either CO or SvO₂ had decreased by 30%. Measurements were repeated immediately after successful induction of CS, again after 60 min of untreated CS, and finally after 15 min of dobutamine infusion following an additional 180 min of untreated CS.
Results
Hemodynamic response to coronary embolization
Coronary embolization resulted in a marked hemodynamic deterioration consistent with ischemic CS (Table 1). Cardiac output fell from 4.2 ± 1.0 to 2.6 ± 0.4 L·min⁻1 (P < 0.001), driven primarily by a reduction in stroke volume from 71 ± 20 to 39 ± 9 mL (P < 0.001), with only a modest compensatory increase in heart rate. Ea increased while Ees decreased, resulting in increased VAC and indicating marked ventricular–arterial uncoupling during acute CS. Mean arterial pressure declined from 93 ± 13 to 71 ± 12 mmHg (P < 0.001), accompanied by a rise in LV filling pressures reflecting combined forward failure and backward congestion. Systemic vascular resistance increased markedly from 1,756 ± 534 to 2,928 ± 517 dyn·s·cm⁻⁵ (P < 0.05), indicative of compensatory sympathetic vasoconstriction. Mixed venous oxygen saturation fell from 58 ± 8% to 39 ± 7% (P < 0.001), confirming tissue hypoperfusion. Following 60 min of untreated CS, hemodynamic parameters remained largely unchanged from CS onset, with the exception of a partial reduction in SVR and increased heart rate. During dobutamine-treated CS, cardiac output increased and SVR decreased, while VAC improved but remained elevated compared with baseline.
Table 1.
Hemodynamic values.
| N = 20 | Healthy | Onset of CS | Established untreated CS | Dobutamine-treated CS |
|---|---|---|---|---|
| Cardiac output (L min−1) | 4.2 ± 1.0 | 2.6 ± 0.4‡ | 2.6 ± 0.6‡ | 5.1 ± 1.1 |
| Stroke volume (mL) | 71 ± 20 | 39 ± 9‡ | 38 ± 11† | 54 ± 13* |
| Heart rate (bpm) | 62 ± 18 | 69 ± 22* | 75 ± 27† | 99 ± 29‡ |
| Mean arterial blood pressure (mm Hg) | 93 ± 13 | 71 ± 12‡ | 70 ± 12‡ | 77 ± 13† |
| Systolic arterial blood pressure (mm Hg) | 125 ± 15 | 95 ± 14‡ | 95 ± 15‡ | 112 ± 24* |
| Diastolic arterial blood pressure (mmHg) | 75 ± 14 | 58 ± 13‡ | 58 ± 12‡ | 58 ± 9† |
| Pulmonary artery wedge pressure (mmHg) | 7 ± 2 | 12 ± 3‡ | 13 ± 4‡ | 13 ± 3‡ |
| Systemic vascular resistance (dyn s−1 cm5) | 1,756 ± 534 | 2,928 ± 517* | 2,060 ± 677* | 1,175 ± 329* |
| Mixed venous oxygen saturation (%) | 58 ± 8 | 39 ± 7‡ | 34 ± 6‡ | 56 ± 10 |
| LV end-diastolic pressure (mmHg) | 13 ± 5 | 18 ± 6‡ | 19 ± 6‡ | 20 ± 8‡ |
| Ees (mmHg mL−1) | 2.12 ± 0.56 | 1.11 ± 0.43‡ | 1.52 ± 0.89* | 1.45 ± 0.54† |
| VAC | 0.82 ± 0.27 | 2.39 ± 1.23‡ | 2.05 ± 1.24‡ | 1.32 ± 0.6† |
| Eaestimated (mmHg mL−1) | 1.72 ± 0.60 | 2.25 ± 0.61‡ | 2.25 ± 0.62† | 1.95 ± 0.54 |
| Eareference (mmHg mL−1) | 1.73 ± 0.76 | 2.24 ± 0.89† | 2.36 ± 1.02† | 1.74 ± 0.69 |
Paired t-tests comparing hemodynamic values in the healthy state with the onset of CS, following 60 minutes of untreated CS (i.e. “established untreated CS”) and following dobutamine treatment respectively. ‡P < 0.001, †P < 0.01, *P < 0.05.
Overall
Both estimated and reference Ea increased from the healthy state to the CS states (Table 1). After pooling all 80 paired observations, the time-specific Bland–Altman analysis showed that agreement was preserved across the experimental states, but with state-dependent variation in bias and LoA (Figure 3). Bias was negligible at the healthy baseline and at CS onset, shifted toward slight underestimation during established untreated CS, and toward modest overestimation during dobutamine-treated CS. Estimated Ea reproduced the reference Ea with a negligible bias (2.04 ± 0.63 for Eaestimated vs. 2.03 ± 0.88 mm Hg mL⁻1 for Eareference) and a strong linear relationship (Pearson r = 0.72, 95% CI 0.58–0.82, P < 0.001) (Figures 4A,B).
Figure 3.
Time-specific Bland–Altman analysis of femoral pressure-derived Ea vs. PV-derived reference Ea. The Bland–Altman plots compare femoral pressure-derived Eaestimated with pressure–volume-derived Eareference at the healthy baseline, onset of cardiogenic shock (CS), established untreated CS, and dobutamine-treated CS. The y-axis shows the difference between methods (Eaestimated−Eareference), and the x-axis shows the mean of the two methods. Solid horizontal lines indicate mean bias, dashed lines indicate the upper and lower 95% limits of agreement, and dotted lines indicate zero difference.
Figure 4.
Linear regressions of estimated and reference Ea. (A) The linear regression analysis for the pooled observations for the estimated and reference Ea. (B) The linear regression of the estimated and reference Ea based on timepoints.
Healthy baseline
Before CS, Ea was nearly identical when calculated by the two methods (1.72 ± 0.60 vs. 1.73 ± 0.76 mm Hg mL⁻1), with an excellent correlation (r = 0.74, 95% CI 0.44–0.89, P < 0.001) and no systematic bias on Bland–Altman analysis (bias −0.006 mm Hg·mL⁻1; LoA −1.014 to 1.002), indicating that the estimate is accurate under normal ventricular–arterial coupling.
Onset of ischemic cardiogenic shock
Ea increased immediately after coronary embolization. Ea was nearly identical when calculated by the two methods (2.25 ± 0.61 vs. 2.24 ± 0.89 mm Hg mL⁻1), with a still-robust correlation (r = 0.68, 95% CI 0.32–0.87, P = 0.001). The slight overestimation bias (+0.01 mm Hg mL⁻1) remained within the predefined LoA (+0.009 mm Hg·mL⁻1; limits of agreement −1.274 to 1.292).
Established untreated shock
At the end of the 60-min period of untreated CS, the model retained good fidelity (2.25 ± 0.62 vs. 2.36 ± 1.02 mm Hg mL⁻1); correlation strength was maintained (r = 0.69, 95% CI 0.32–0.87, P = 0.002) with a modest underestimation bias (−0.106 mm Hg·mL⁻1). The LoA were widest at this timepoint (−1.567 to 1.354 mm Hg·mL⁻1), indicating greater individual-level variability during established untreated CS.
Dobutamine-treated cardiogenic shock
Following inotropic support with dobutamine, both Eaestimated and Eareference declined. The correlation between the methods remained statistically significant (r = 0.63, 95% CI 0.12–0.88, P = 0.020), though slightly attenuated compared to earlier timepoints with modest overestimation (bias +0.180 mm Hg·mL⁻1; limits of agreement −0.849 to 1.209).
Discussion
Elevated afterload is a well-established aggravating factor in ischemic cardiogenic shock, as it increases myocardial oxygen demand and further compromises forward flow in the failing ventricle (2, 3). Vasoactive agents commonly used in shock management may either exacerbate or alleviate this burden depending on their pharmacological profile. Thus, the ability to monitor afterload continuously and with minimal invasiveness may provide important prognostic and therapeutic guidance. Our findings support that femoral pressure-derived Ea may serve as a reliable surrogate for invasively measured Ea in this context, potentially enabling clinicians to detect detrimental increases in afterload in real time and tailor therapy accordingly. Additional markers, such as Ees and VAC, further clarify the hemodynamic phenotype of ischemic CS. While Ea reflects the arterial load imposed on the ventricle, Ees reflects LV chamber contractility; VAC integrates these two components. Thus, an increase in Ea may be tolerated if matched by preserved or increased Ees, whereas a simultaneous increase in Ea and reduction in Ees produces ventricular–arterial uncoupling. In the present study, acute CS was characterized not only by increased Ea but also by a marked reduction in Ees and a pronounced increase in VAC. This supports the physiological interpretation that the failing LV became mismatched with the arterial system. Dobutamine increased cardiac output and reduced SVR, but VAC remained higher than in the healthy state, suggesting only partial restoration of ventricular–arterial coupling despite improved forward flow.
Several approaches have been proposed for estimating Ea and VAC at the bedside. Invasive PV-loop analysis remains the reference method because it directly relates LV end-systolic pressure to SV and allows simultaneous assessment of Ees. However, this approach is impractical in routine clinical shock management. Non-invasive and minimally invasive alternatives include echocardiographic estimation of SV combined with an arterial pressure-derived estimate of LVESP (12), MAP/SV-based approximations (13), arterial tonometry (14), and single-beat echocardiographic methods for Ees (15). These approaches increase clinical feasibility but introduce assumptions related to pressure site, LVESP approximation, volume measurement, and loading conditions. The results of this study demonstrate that the femoral pressure estimate of Ea remains reliable under healthy conditions, through the full dynamic evolution of acute CS, and during vasoactive treatment, when compared with direct Ea measurement using PV loops. These findings extend the previous validation of Ea estimation in hemodynamically stable patients (8) to the more extreme setting of an acute failing LV, confirming that the simple 0.9 × systolic femoral pressure/SV formula remains applicable even when SV decreases and SVR increases in concert. The use of femoral arterial pressure is also physiologically and clinically relevant. Radial arterial pressure may differ substantially from more central arterial pressure during shock and vasopressor therapy, and radial measurements may underestimate central pressure in critically ill patients (16, 17).
Because the estimated Ea correlated with the reference Ea with only small state-dependent shifts in bias, clinicians could use a single femoral artery tracing and thermodilution-derived SV to follow afterload continuously during the critical phase of developing CS, without the need for left heart catheterization. Importantly, the transition from negligible bias at baseline and CS onset to slight underestimation during established untreated CS and modest overestimation during dobutamine treatment suggests that acute changes in arterial compliance or wave reflection may alter the proportionality constant.
Notably, the correlation between the estimated Ea and the reference Ea remained significant even during inotropic support. However, the slightly elevated overestimation bias suggested that, while the model retained utility under pharmacological intervention, Ea estimation using femoral artery pressure recordings may be partially influenced by treatment-induced hemodynamic changes. The time-specific Bland–Altman plots also indicate that the agreement was the least precise during established untreated CS, where the LoA were widest despite preserved correlation. This suggests that femoral pressure-derived Ea may be reliable for tracking group-level and directional changes in afterload, while caution is warranted when interpreting single absolute values in prolonged untreated shock. A contribution from time-dependent PV catheter signal drift or calibration instability cannot be excluded, as PV catheter measurements are sensitive to signal drift and volume calibration procedures (18–20). However, such drift would not fully explain the shift toward overestimation during dobutamine treatment, suggesting that pharmacologically induced changes in vascular tone, wave reflection, or femoral-to-central pressure amplification may also have contributed.
The hemodynamic profiles at CS onset and established untreated CS were notably similar, with near-identical Ea values at both timepoints, but differences in HR and SVR. This possibly reflects the integrative nature of Ea, as, although Ea is largely load-independent, the marked rise in SVR at CS onset and the compensatory tachycardia at established CS have opposing effects on Ea, yielding comparable aggregate values despite distinct underlying hemodynamic compositions.
Limitations
This study has several limitations. First, the experiment was performed in a controlled porcine model of ischemic CS, and the findings may not fully translate to human CS, where age, comorbidities, chronic vascular disease, and pre-existing reductions in arterial compliance may influence arterial load and pressure transmission. Second, Eaestimated was derived from femoral systolic pressure and thermodilution-derived SV, and therefore still requires invasive arterial and pulmonary artery catheterization. Thus, the present method is less invasive than LV PV catheterization but is not fully non-invasive. Third, femoral pressure remains a peripheral surrogate of LV end-systolic pressure, and changes in vascular tone, wave reflection, and arterial compliance during CS or vasoactive therapy may alter the relationship between femoral systolic pressure and true LV end-systolic pressure. Furthermore, dobutamine-treated CS was assessed at a single dose and timepoint, and the performance of femoral pressure-derived Ea during other vasoactive regimens, mechanical circulatory support, or more prolonged shock states remains unknown.
Future research should test whether a dynamic scaling factor could further narrow the limits of agreement and whether non-invasive SV and blood pressure measurements could substitute for values measured using invasive procedures. Nonetheless, by demonstrating robust agreement across disease states, our study supports incorporating bedside estimated Ea using arterial pressure measurement into real-time assessment and therapeutic algorithms for CS.
Acknowledgments
We thank Alexander Møller Larsen, Kasper Lykke Wethelund, Halvor Guldbrandsen, and Lasse Juul Christensen from the Department of Cardiology, Aarhus University Hospital, Mark Stoltenberg Ellegaard and Buse Bor from the Department of Anesthesiology and Intensive Care, AUH, and Selma Elisabeth Fogh Rubenius from Aarhus University for their voluntary contributions to data acquisition.
Funding Statement
The author(s) declared that financial support was received for this work and/or its publication. OKH is supported the Graduate School of Health at Aarhus University Faculty. OKH and KBH is supported by the Danish Cardiovascular Academy, which is funded by the Novo Nordisk Foundation (grant number NNF20SA0067242) and The Danish Heart Foundation. The project received funding from the Sophus Jacobsen Foundation (Copenhagen, Denmark), Aase & Ejnar Danielsens Foundation (Kongens Lyngby, Denmark). The funders was not involved in the study design, collection, analysis, interpretation of data, the writing of this article, or the decision to submit it for publication.
Footnotes
Edited by: Filip Konecny, McMaster University, Canada
Reviewed by: Vasilios E. Papaioannou, Democritus University of Thrace, Greece
Ana Ochagavía, Bellvitge University Hospital, Spain
Data availability statement
The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.
Ethics statement
This animal study was approved by the Danish National Animal Experiment Inspectorate. (Hemodynamic Optimization with Carbonates, Permit no: 2023-15-0201-01466, issued on 19/06-2023). This study was conducted in accordance with the local legislation and institutional requirements.
Author contributions
OH: Conceptualization, Investigation, Project administration, Validation, Visualization, Writing – original draft, Writing – review & editing. NG: Supervision, Writing – original draft, Writing – review & editing. KB-H: Investigation, Methodology, Supervision, Writing – original draft, Writing – review & editing. RN: Conceptualization, Investigation, Supervision, Writing – original draft, Writing – review & editing.
Conflict of interest
The author(s) declared that this work was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Generative AI statement
The author(s) declared that generative AI was not used in the creation of this manuscript.
Any alternative text (alt text) provided alongside figures in this article has been generated by Frontiers with the support of artificial intelligence and reasonable efforts have been made to ensure accuracy, including review by the authors wherever possible. If you identify any issues, please contact us.
Publisher's note
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.
Supplementary material
The Supplementary Material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fcvm.2026.1815870/full#supplementary-material
References
- 1.Hørsdal OK, Frederiksen PH, Helgestad OKL, Ravn HB, Møller JE, Wiggers H, et al. The immediate cardiovascular and mitochondrial response in ischemic cardiogenic shock. J Cardiovasc Transl Res. (2025) 18(5):1204–17. 10.1007/s12265-025-10647-6 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Vahdatpour C, Collins D, Goldberg S. Cardiogenic shock. J Am Heart Assoc. (2019) 8:e011991. 10.1161/JAHA.119.011991 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Peberdy MA, Callaway CW, Neumar RW, Geocadin RG, Zimmerman JL, Donnino M, et al. Part 9: post–cardiac arrest care. Circulation. (2010) 122(18 Suppl 3):S768–86. 10.1161/CIRCULATIONAHA.110.971002 [DOI] [PubMed] [Google Scholar]
- 4.Saxena A, Garan AR, Kapur NK, O’Neill WW, Lindenfeld J, Pinney SP, et al. Value of hemodynamic monitoring in patients with cardiogenic shock undergoing mechanical circulatory support. Circulation. (2020) 141:1184–97. 10.1161/CIRCULATIONAHA.119.043080 [DOI] [PubMed] [Google Scholar]
- 5.Guinot P-G, Longrois D, Kamel S, Lorne E, Dupont H. Ventriculo-arterial coupling analysis predicts the hemodynamic response to norepinephrine in hypotensive postoperative patients: a prospective observational study. Crit Care Med. (2018) 46:e17–25. 10.1097/CCM.0000000000002772 [DOI] [PubMed] [Google Scholar]
- 6.Monge García MI, Saludes Orduña P, Cecconi M. Understanding arterial load. Intensive Care Med. (2016) 42:1625–7. 10.1007/s00134-016-4212-z [DOI] [PubMed] [Google Scholar]
- 7.Kelly RP, Ting CT, Yang TM, Liu CP, Maughan WL, Chang MS, et al. Effective arterial elastance as index of arterial vascular load in humans. Circulation. (1992) 86:513–21. 10.1161/01.CIR.86.2.513 [DOI] [PubMed] [Google Scholar]
- 8.Jozwiak M, Millasseau S, Richard C, Monnet X, Mercado P, Dépret F, et al. Validation and critical evaluation of the effective arterial elastance in critically ill patients. Crit Care Med. (2019) 47:e317–24. 10.1097/CCM.0000000000003645 [DOI] [PubMed] [Google Scholar]
- 9.Hørsdal OK, Ellegaard MS, Larsen AM, Guldbrandsen H, Moeslund N, Møller JE, et al. Lactate infusion improves cardiac function in a porcine model of ischemic cardiogenic shock. Crit Care. (2025) 29:113. 10.1186/s13054-025-05346-2 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Hørsdal OK, Larsen AM, Wethelund KL, Dalsgaard FF, Seefeldt JM, Helgestad OKL, et al. The ketone body 3-hydroxybutyrate increases cardiac output and cardiac contractility in a porcine model of cardiogenic shock: a randomized, blinded, crossover trial. Basic Res Cardiol. (2025) 120(3):579–96. 10.1007/s00395-025-01103-2 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Hørsdal OK, Wethelund KL, Gopalasingam N, Lyhne MD, Ellegaard MS, Møller-Helgestad OK, et al. Cardiovascular effects of increasing positive end-expiratory pressure in a model of left ventricular cardiogenic shock in female pigs. Anesthesiology. (2024) 141(6):1105–18. 10.1097/ALN.0000000000005201 [DOI] [PubMed] [Google Scholar]
- 12.Fazzini L, Padkins MR, Berg-Hansen K, Gori M, Kane GC, Hillerson DB, et al. Left ventricular-arterial coupling and mortality in the cardiac intensive care unit. Eur Heart J Acute Cardiovasc Care. (2025) 14:394–402. 10.1093/ehjacc/zuaf038 [DOI] [PubMed] [Google Scholar]
- 13.Monge Garcia MI, Jian Z, Settels JJ, Hatib F, Cecconi M, Pinsky MR. Reliability of effective arterial elastance using peripheral arterial pressure as surrogate for left ventricular end-systolic pressure. J Clin Monit Comput. (2019) 33:803–13. 10.1007/s10877-018-0236-y [DOI] [PubMed] [Google Scholar]
- 14.Gayat E, Mor-Avi V, Weinert L, Yodwut C, Lang RM. Noninvasive quantification of left ventricular elastance and ventricular-arterial coupling using three-dimensional echocardiography and arterial tonometry. Am J Physiol Heart Circ Physiol. (2011) 301:H1916–23. 10.1152/ajpheart.00760.2011 [DOI] [PubMed] [Google Scholar]
- 15.Chen C-H, Fetics B, Nevo E, Rochitte CE, Chiou K-R, Ding P-A, et al. Noninvasive single-beat determination of left ventricular end-systolic elastance in humans. J Am Coll Cardiol. (2001) 38:2028–34. 10.1016/S0735-1097(01)01651-5 [DOI] [PubMed] [Google Scholar]
- 16.Kim WY, Jun JH, Huh JW, Hong SB, Lim C-M, Koh Y. Radial to femoral arterial blood pressure differences in septic shock patients receiving high-dose norepinephrine therapy. Shock. (2013) 40:527–31. 10.1097/SHK.0000000000000064 [DOI] [PubMed] [Google Scholar]
- 17.Dorman T, Breslow MJ, Lipsett PA, Rosenberg JM, Balser JR, Almog Y, et al. Radial artery pressure monitoring underestimates central arterial pressure during vasopressor therapy in critically ill surgical patients. Crit Care Med. (1998) 26:1646–9. 10.1097/00003246-199810000-00014 [DOI] [PubMed] [Google Scholar]
- 18.Hørsdal OK, Berg-Hansen K, Gopalasingam N, Møller Larsen A, Christensen LJ, Ellegaard MS, et al. Fixed volume-intercept approach for serial contractility assessment with closed-chest pressure-volume loops in a porcine model. J Vis Exp. (2026) 230:e70220. 10.3791/70220 [DOI] [PubMed] [Google Scholar]
- 19.Pacher P, Nagayama T, Mukhopadhyay P, Bátkai S, Kass DA. Measurement of cardiac function using pressure–volume conductance catheter technique in mice and rats. Nat Protoc. (2008) 3:1422–34. 10.1038/nprot.2008.138 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Lyhne MD, Schultz JG, Dragsbaek SJ, Hansen JV, Mortensen CS, Kramer A, et al. Closed chest biventricular pressure-volume loop recordings with admittance catheters in a porcine model. J Vis Exp. (2021) 2021(171):e62661. 10.3791/62661 [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.