Direct visualization–assisted enteral stent placement in malignant colorectal obstruction: a case report

Tianjie Han 1,†, Junshan Li 2,*,†
PMCID: PMC13440149  PMID: 42559123

Abstract

Background

Malignant bowel obstruction caused by colorectal cancer is a common and serious clinical condition. Enteral stent placement is widely used as a palliative treatment or as a bridge to surgery. Accurate guidewire placement across stenotic segments remains technically challenging.

Case summary

A 39-year-old woman with sigmoid colon adenocarcinoma presented with abdominal pain and cessation of flatus and defecation for more than 20 days. Abdominal computed tomography demonstrated sigmoid colon cancer with intestinal obstruction. Colonoscopy revealed circumferential stenosis approximately 25 cm from the anus. A direct visualization system (EyeMax, Micro-Tech, Nanjing, China) was used through the biopsy channel and advanced across the stenotic segment under direct visualization. A Jagwire was successfully passed through the stricture via the EyeMax working channel. After withdrawal of the EyeMax, a self-expandable metallic stent was deployed over the guidewire under endoscopic visualization without fluoroscopy. The procedure was successful with immediate symptom relief. A delayed fecalith obstruction occurred and was effectively treated with endoscopic fragmentation. Following the procedure, the patient resumed passage of flatus and stool, and abdominal pain was significantly relieved. One month after stent placement, the patient underwent radical resection for sigmoid colon cancer without perioperative complications. During 1-year postoperative follow-up, the patient remained asymptomatic, with no evidence of recurrent intestinal obstruction, or tumor recurrence.

Conclusion

The EyeMax system enables real-time intraluminal visualization, facilitating safe and accurate guidewire placement in complex colorectal strictures. This technique may reduce reliance on blind manipulation and improve procedural safety.

Keywords: case report, colorectal cancer, enteral stent, EyeMax, malignant bowel obstruction

Introduction

Malignant bowel obstruction is a frequent complication of advanced colorectal cancer and often leads to significant morbidity. Endoscopic placement of self-expandable metallic stents (SEMS) has become an established method for relieving obstruction and improving quality of life. This technique is also widely used as a bridge to elective surgery in selected patients (1, 2).

In conventional practice, guidewire passage is often performed under fluoroscopic or indirect endoscopic guidance, which may involve blind manipulation in near-complete obstruction. Ultra-slim endoscopy represents an alternative, but its maneuverability and visualization may be limited. Severe luminal stenosis may hinder the advancement of the guidewire, increasing the risk of complications such as perforation and bleeding (3, 4). The EyeMax system (9F, 3 mm, Micro-Tech, Nanjing, China) is a miniature direct visualization platform that can be introduced through the working channel of a standard endoscope, allowing real-time intraluminal observation and instrument guidance. However, its application in colorectal stent placement has rarely been reported. In this report, we describe the successful application of the EyeMax system in enteral stent placement for malignant bowel obstruction.

Case presentation

Chief complaints

A 39-year-old woman presented with abdominal pain and cessation of flatus and defecation for more than 20 days.

History of present illness

In the 20 days before admission, she developed progressive abdominal distension and pain accompanied by cessation of flatus and defecation.

History of past illness

The patient had been diagnosed with sigmoid colon adenocarcinoma with pelvic lymph node metastasis 3 months prior to admission. She received four cycles of chemotherapy consisting of cetuximab combined with the FOLFOX regimen.

Personal and family history

The patient had no history of substance abuse, consanguineous marriage with parental relatives, or familial genetic disorders.

Physical examination

On physical examination, the vital signs were as follows: body temperature, 36.7 °C; blood pressure, 117/84 mmHg; heart rate, 76 bpm; respiratory rate, 18 breaths/min.

Laboratory examinations

On the day of admission, laboratory tests revealed several abnormalities: routine blood tests showed a hemoglobin level of 95 g/L (normal: 115–150 g/L), blood biochemistry indicated decreased levels of albumin (39 g/L; normal: 40–55 g/L). Other parameters, including liver and kidney function, blood glucose, electrolytes, and thyroid function, were within normal ranges.

Imaging examinations

Abdominal computed tomography demonstrated sigmoid colon cancer complicated with intestinal obstruction (Figures 1A, B). Colonoscopy (CF-HQ290, Olympus, Japan with a 3.2 mm working channel) performed revealed circumferential stenosis approximately 25 cm from the anus with mucosal erosion (Figure 2A). The distal lumen diameter at the stenotic segment was approximately 3 mm. The EyeMax direct visualization system (outer diameter 3.0 mm, 1.2 mm working channel, Micro-Tech, Nanjing, China) was introduced through the biopsy channel of the colonoscope and advanced beyond the stenotic segment. The length of stenosis was measured to be approximately 3.5 cm.

Figure 1.

CT scan panel with two grayscale axial slices labeled A and B; panel A displays the pelvic region showing pelvic bones and soft tissues, while panel B shows the abdomen with visible vertebra and internal organs.

(A, B) Abdominal computed tomography showing sigmoid colon cancer with intestinal obstruction.

Figure 2.

Panel A shows an endoscopic view of a bleeding lesion with visible blood and tissue irregularity. Panel B depicts a constricted yellowish lumen, severely stenotic colorectal lumen under direct visualization, with an inset of nodular mucosa. Panel C presents a deployed stent within an inflamed and erythematous tissue, suggesting post-procedural placement. Panel D displays a stent within a greenish, bile-filled lumen along with an inset image of a similar endoscopic field.

Colonoscopic findings and procedure of stent placement under EyeMax guidance. (A) Circumferential luminal stenosis located approximately 25 cm from the anus with mucosal erosion. (B) The EyeMax system advanced beyond the stenotic segment and placement of Jagwire into distal lumen. (C) Deployment of enteral metallic stent along the guidewire. (D) Confirmation of satisfactory stent position.

Final diagnosis

Malignant bowel obstruction secondary to sigmoid colon adenocarcinoma.

Treatment

Under direct endoscopic visualization, the EyeMax system was carefully advanced through the stenotic segment, allowing clear identification of the distal lumen. A 0.035-inch Jagwire (Boston Scientific, USA) was then inserted through the working channel of the EyeMax system and successfully passed across the stricture under continuous visualization (Figure 2B). After successful guidewire placement, the EyeMax system was withdrawn while maintaining the guidewire position. A 20-mm diameter, 8-cm long self-expandable metallic enteral stent (Micro-Tech) was successfully deployed across the stenotic segment over the guidewire into the colon (Figure 2C), the stent was placed as a bridge to surgery. The stent was deployed across the stenotic segment under direct endoscopic visualization without fluoroscopic guidance. Correct positioning and full expansion of the stent were confirmed endoscopically (Figure 2D). Symptoms initially improved. Later, recurrent abdominal pain occurred. CT showed fecal impaction in the rectum and stent lumen (Figures 3A–C). Enema treatment was unsuccessful (Supplementary Video 1).

Figure 3.

Panel A shows a coronal CT scan of the pelvis with an obstructive mass in the rectum. Panel B provides an axial CT image confirming the location. Panel C is another coronal CT view post-treatment. Panel D is an endoscopic photograph of a large, yellowish mass in the rectal lumen. Panel E depicts endoscopic removal of the mass. Panel F shows the emptied rectal cavity after extraction.

Computed tomography (CT) and colonoscopic findings of fecalith obstruction. (A, B) CT images show fecalith formation within the enteral stent. (C) Fecalith impaction within the rectum. (D) Colonoscopy demonstrates fecalith obstruction within the rectum. (E) Endoscopic snare fragmentation of the fecalith was performed. (F) Restoration of stent patency with visible passage of stool.

The procedure was successful with immediate symptom relief. However, 2 days after stent placement, the patient had recurrent abdominal pain and distension with no flatus or defecation. Colonoscopy confirmed fecalith obstruction (Figure 3D). A single fecalith measuring approximately 15 × 10 mm was located at the distal end of the stent. Fragmentation was performed using a 25 mm oval snare (Micro-Tech). The snare was gradually tightened around the fecalith under direct visualization, and mechanical fragmentation was achieved by repeated opening-closing. The procedure was technically straightforward without evidence of stent migration, perforation, or bleeding. No immediate or delayed procedure-related complications occurred. Complete clearance of fragmented debris was confirmed by post-fragmentation colonoscopy (Figures 3E, F).

Outcome and follow-up

After the procedure, the patient resumed passage of flatus and stool. The enteral stent successfully served as a bridge to surgery. One month after stent placement, the patient underwent radical resection for sigmoid colon cancer without perioperative complications. During 1-year postoperative follow-up, the patient remained asymptomatic, with no evidence of recurrent intestinal obstruction, or tumor recurrence.

Discussion

Self-expandable metallic stents have become an effective treatment for malignant bowel obstruction and are widely used as either palliative therapy or a bridge to surgery (3, 5). Compared with emergency surgery, SEMS placement can reduce morbidity, shorten hospital stay, and improve quality of life (6, 7).

However, successful stent placement depends on accurate guidewire passage through stenotic lesions. In conventional practice, SEMS placement is performed under endoscopic and fluoroscopic guidance. However, blind or semi-blind guidewire manipulation in tight strictures increases the risk of complications such as perforation and bleeding (4). Previous studies have identified guidewire-related perforation as a major procedural risk, particularly in complex or near-complete obstructions. Blind manipulation may increase the risk of perforation or bleeding (8, 9).

Although pediatric nasogastroscopes or ultra-slim endoscopes have been reported as alternatives for difficult colorectal stent placement (10, 11), but they have limitations in maneuverability, working channel size, and stability in severely angulated or tight malignant obstructions (12, 13). Instead, a dedicated direct-visualization system was selected for its smaller diameter, enhanced visualization, and guidewire compatibility. However, the lack of comparison with an ultra-slim endoscope is a limitation.

The EyeMax system (9F, Micro-Tech, Nanjing, China) is a miniature, direct visualization platform with an outer diameter of 3.0 mm and a 1.2 mm working channel, designed to be introduced through the biopsy channel of a standard colonoscope. The EyeMax direct visualization system, similar to Spyglass, has been utilized for the diagnosis of pancreatic and biliary tract diseases (14). The system used shares conceptual features with cholangioscopy, but its small caliber and direct visualization facilitate guidewire advancement when fluoroscopy alone is insufficient. Recent studies have reported the application of the EyeMax system in pancreaticobiliary diseases, as well as in the management of colonic diverticular bleeding and appendicitis (15–17). Recent reports have shown the value of choledochoscopy in challenging settings, such as diagnosing biliary metastasis from gastric cancer (18). Inspired by these, our case extends direct visualization to malignant colorectal obstruction during enteral stent placement. The EyeMax system allows real-time visualization of the stenotic lumen, enabling precise identification of the true lumen and controlled guidewire advancement. This may reduce procedural risk and improve technical success, particularly in near-complete obstructions. The EyeMax system offers a unique advantage by being passed through the biopsy channel of a standard colonoscope, eliminating the need for a second endoscopic insertion. The most important finding of this case is the feasibility of direct visualization–assisted guidewire placement using the EyeMax system (Figure 4).

Figure 4.

Illustrated stepwise diagram shows EyeMax-assisted enteral stent placement for malignant stenosis, with six numbered panels depicting colonoscope insertion, EyeMax deployment, jagwire passage, stent delivery, and final stent expansion confirmed endoscopically. Below, five endoscopic images labeled A to E correspond to each procedural step, visually documenting progression from severe stenosis, system advancement, jagwire passage, stent deployment, to fully expanded stent apposed to mucosa.

Schematic workflow and endoscopic images of EyeMax-assisted enteral stent placement.

However, this technique also has limitations. First, this is a single case report, and the safety and efficacy of this approach require validation in larger studies. Second, the additional device may increase procedural complexity and cost. Third, the applicability of the EyeMax system may be limited in extremely tortuous anatomy.

Another important aspect highlighted by this case is delayed stent obstruction caused by fecalith formation. Although less common than tumor-related stent dysfunction, fecalith obstruction represents a distinct and increasingly recognized cause of stent failure. In a personal series of 100 patients with stage IV obstructing colorectal cancer, fecal impaction necessitated repeat colonoscopic intervention in 19% of cases during follow-up (19). Similarly, a large multinational registry of 167 patients reported fecal impaction-related stent occlusion in 1.2% of cases (20). In comparison, tumor growth accounted for 8% of recurrent obstructions in the same series, highlighting that fecalith obstruction is a comparably frequent phenomenon requiring vigilant post-stenting management. The pathophysiology of fecalith formation is likely multifactorial. Contributing factors include reduced colonic motility, altered stool consistency, partial stent expansion, and tumor-related luminal irregularities. These factors may promote stool stasis within the stent, ultimately leading to impaction and recurrent obstruction (8, 21). Furthermore, the patient's prior chemotherapy with cetuximab and FOLFOX may have contributed to altered stool consistency and reduced colonic motility, a known side effect of certain chemotherapeutic agents (22). Furthermore, treatment strategies for stent obstruction depend on the underlying cause. While tumor-related obstruction may require repeat stenting or surgical intervention, fecalith obstruction can often be managed using minimally invasive endoscopic techniques (9, 23). In the present case, snare-assisted fragmentation effectively restored luminal patency without the need for re-intervention or surgery. This case also underscores the importance of post-stenting management. Preventive strategies, including bowel regimen optimization, stool softeners, and patient education, may help reduce the risk of fecal impaction.

Conclusion

This case demonstrates that the EyeMax system is a feasible tool for direct visualization–assisted guidewire placement in malignant colorectal obstruction. This technique may be useful both in palliative settings and as a bridge to surgery in selected patients with malignant colorectal obstruction.

Funding Statement

The author(s) declared that financial support was not received for this work and/or its publication.

Footnotes

Edited by: Angel Lanas, University of Zaragoza, Spain

Reviewed by: Takahiro Hiratsuka, Oita University, Japan

Hamidreza Bolhasani, DataBioX Research, Iran

Data availability statement

The original contributions presented in the study are included in the article/Supplementary material, further inquiries can be directed to the corresponding author.

Ethics statement

The studies involving humans were approved by Shandong Provincial Third Hospital. The studies were conducted in accordance with the local legislation and institutional requirements. The participants provided their written informed consent to participate in this study. Written informed consent was obtained from the individual(s) for the publication of any potentially identifiable images or data included in this article.

Author contributions

TH: Writing – original draft. JL: Writing – review & editing.

Conflict of interest

The author(s) declared that this work was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Correction note

This article has been corrected with minor changes. These changes do not impact the scientific content of the article.

Generative AI statement

The author(s) declared that Generative AI was not used in the creation of this manuscript.

Any alternative text (alt text) provided alongside figures in this article has been generated by Frontiers with the support of artificial intelligence and reasonable efforts have been made to ensure accuracy, including review by the authors wherever possible. If you identify any issues, please contact us.

Publisher's note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Supplementary material

The Supplementary Material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fmed.2026.1848674/full#supplementary-material

Table_1.docx (12.8KB, docx)
Download video file (24.5MB, mp4)

References

  • 1.van Hooft JE, Veld JV, Arnold D, Beets-Tan RGH, Everett S, Götz M, et al. Self-expandable metal stents for obstructing colonic and extracolonic cancer: european society of gastrointestinal endoscopy (ESGE) guideline - update 2020. Endoscopy. (2020) 52:389–407. doi: 10.1055/a-1140-3017 [DOI] [PubMed] [Google Scholar]
  • 2.McHugh FT, Ryan ÉJ, Ryan OK, Tan J, Boland PA, Whelan MC, et al. Management strategies for malignant left-sided colonic obstruction: a systematic review and network meta-analysis of randomized controlled trials and propensity score matching studies. Dis Colon Rectum. (2024) 67:878–94. doi: 10.1097/DCR.0000000000003256 [DOI] [PubMed] [Google Scholar]
  • 3.Arezzo A, Forcignanò E, Bonino MA, Balagué C, Targarona E, Borghi F, et al. Long-term oncologic results after stenting as a bridge to surgery vs. emergency surgery for malignant left-sided colonic obstruction: a multicenter randomized controlled trial (ESCO trial). Ann Surg. (2020) 272:703–8. doi: 10.1097/SLA.0000000000004324 [DOI] [PubMed] [Google Scholar]
  • 4.Mauro A, Scalvini D, Borgetto S, Fugazzola P, Mazza S, Perretti I, et al. Malignant acute colonic obstruction: multidisciplinary approach for endoscopic management. Cancers. (2024) 16:821. doi: 10.3390/cancers16040821 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Tao Y, Li GB, Zhang HY, Cao K, Wang ZJ, Han JG. Current practice for the treatment of obstructive left-sided colon cancer in China: a nationwide cross-sectional survey. Surg Today. (2023) 53:459–69. doi: 10.1007/s00595-022-02623-3 [DOI] [PubMed] [Google Scholar]
  • 6.Yamamoto M, Osugi N, Nakamatsu D, Matsumoto K, Fukui K, Nishida T. Short-term efficacy and long-term limitations of self-expandable metallic stent placement for colorectal obstruction due to extracolonic malignancies. DEN OPEN. (2026) 6:e70234. doi: 10.1002/deo2.70234 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Belle S. Endoscopic decompression in colonic distension. Visc Med. (2021) 37:142–8. doi: 10.1159/000514799 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Han S-H, Lee JH. Colonic stent-related complications and their management. Clin Endosc. (2014) 47:415–9. doi: 10.5946/ce.2014.47.5.415 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Ahmed O, Lee JH, Thompson CC, Faulx A, AGA. clinical practice update on the optimal management of the malignant alimentary tract obstruction: expert review. Clin Gastroenterol Hepatol. (2021) 19:1780–8. doi: 10.1016/j.cgh.2021.03.046 [DOI] [PubMed] [Google Scholar]
  • 10.Yuan X, Zhang Z, Gao H, Bao Z, Wang W, Zheng T, et al. Colorectal stent placement assisted by a slim gastroscope: technique and outcomes. Surg Endosc. (2021) 35:2398–402. doi: 10.1007/s00464-021-08334-z [DOI] [PubMed] [Google Scholar]
  • 11.Lamazza A, Fiori E, Schillaci A, Sterpetti AV, A. new technique for placement of a self-expanding metallic stent (SEMS) in patients with colon rectal obstruction: a prospective study of 43 patients. Surg Endosc. (2013) 27:1045–8. doi: 10.1007/s00464-012-2522-y [DOI] [PubMed] [Google Scholar]
  • 12.Lamazza A, Fiori E, Schillaci A, Sterpetti AV, A. pediatric nasogastroscope facilitates colorectal endoscopic stenting. Surg Laparosc Endosc Percutan Tech. (2018) 28:e109–12. doi: 10.1097/SLE.0000000000000574 [DOI] [PubMed] [Google Scholar]
  • 13.Yang Z, Liu W, Yu S, Liu C. Underwater ultra-slim endoscopy for radiation-free intestinal stenting: overcoming difficult strictures. Endoscopy. (2025) 57:E1224–5. doi: 10.1055/a-2724-7196 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Robles-Medranda C, Alcivar-Vasquez J, Raijman I, Kahaleh M, Puga-Tejada M, Del Valle R, et al. Accurate and safe diagnosis and treatment of neoplastic biliary lesions using a novel 9F and 11F digital single-operator cholangioscope. Endosc Int OPEN. (2024) 12:E498–506. doi: 10.1055/a-2282-6678 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Zheng J, Sun Y, Xu B. Endoscopic diagnosis of appendiceal hemorrhage with a disposable imaging catheter (EyeMax): a case report. Front Med. (2026) 13:1734727. doi: 10.3389/fmed.2026.1734727 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Han T-J, Li J-S. Utilizing the EyeMax direct visualization system for the treatment of colonic diverticular bleeding: a case report. World J Gastroenterol. (2025) 31:111906. doi: 10.3748/wjg.v31.i43.111906 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Lin D, Su M, Guo Z, Hu J, Zhong Q, Chen Y, et al. Direct visualization endoscopic retrograde appendicitis therapy for treatment of acute uncomplicated appendicitis. Surg Endosc. (2024) 38:6156–60. doi: 10.1007/s00464-024-11112-2 [DOI] [PubMed] [Google Scholar]
  • 18.Li C-K, Cao R-R, Su D-S, Ming J, Li Y-C, Shao X-D, et al. Diagnosis of bile duct metastasis from gastric cancer by endoscopic retrograde cholangiopancreatography combined with choledochoscopy: a case report. World J Gastrointest Surg. (2026) 18:112416. doi: 10.4240/wjgs.v18.i1.112416 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Lamazza A, Fiori E, Schillaci A, DeMasi E, Pontone S, Sterpetti AV. Self-expandable metallic stents in patients with stage IV obstructing colorectal cancer. World J Surg. (2012) 36:2931–6. doi: 10.1007/s00268-012-1767-x [DOI] [PubMed] [Google Scholar]
  • 20.Jiménez-Pérez J, Casellas J, García-Cano J, Vandervoort J, García-Escribano OR, Barcenilla J, et al. Colonic stenting as a bridge to surgery in malignant large-bowel obstruction: a report from two large multinational registries. Am J Gastroenterol. (2011) 106:2174–80. doi: 10.1038/ajg.2011.360 [DOI] [PubMed] [Google Scholar]
  • 21.Alhassan N, Helmi H, Alzamil A, Batais S, Alkhayyal Y, Alshammari S, et al. Clinical outcomes of endoscopic stent in curative and palliative management of malignant colonic obstruction: a retrospective cohort study. BMC Gastroenterol. (2025) 25:628. doi: 10.1186/s12876-025-04246-7 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Chen M, Li Y, Chen P. Restore intestinal steady-state: new advances in the clinical management of chemotherapy-associated diarrhea and constipation. J Mol Histol. (2025) 56:101. doi: 10.1007/s10735-025-10367-w [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Seo SY, Kim SW. Endoscopic management of malignant colonic obstruction. Clin Endosc. (2020) 53:9–17. doi: 10.5946/ce.2019.051 [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Table_1.docx (12.8KB, docx)
Download video file (24.5MB, mp4)

Data Availability Statement

The original contributions presented in the study are included in the article/Supplementary material, further inquiries can be directed to the corresponding author.


Articles from Frontiers in Medicine are provided here courtesy of Frontiers Media SA

RESOURCES