Abstract
Introduction
Robotic magnetic navigation (RMN) allows for catheter ablation in the presence of complex cardiac anatomy, such as that seen in congenital heart disease (CHD). The objective of this study was to evaluate the feasibility, efficacy and safety of this procedure in a wide range of CHD conditions.
Methods
Patients with CHD who underwent catheter ablation for cardiac arrhythmias using RMN were included in this single-centre observational study. Patients’ baseline characteristics, procedural data, periprocedural complications and acute success rates were obtained. The efficacy of the ablation was evaluated during follow-up in the outpatient clinic using 24-h Holter monitoring.
Results
The study population consisted of 38 patients (26 males) with a median age of 36.8 (IQR 30.8-39.6) years. Surgically corrected transposition of the great arteries was the most common CHD, observed in 14 patients (36.8%). Nineteen patients (50.0%) had more than one congenital defect. Intra-atrial re-entrant tachycardia was the most common arrhythmia ablated, accounting for 21 procedures (55.3%).
Median procedural time was 232.5 min (IQR 186.3-281.3); the median radiation dose and radiation time were 2.4 mGy m2 (IQR 1.3-4.4) and 14.0 min (IQR 9.5-30.4), respectively. Acute arrhythmia cessation was observed in 81.6% of procedures. Major complications were noted in 2 cases (5.3%). The median time from the ablation to follow-up was 5.4 months (IQR 3.4-9.3). Arrhythmia recurrence was observed in 7 cases (18.4%).
Conclusions
RMN is a valuable tool for catheter ablation of arrhythmias in patients with congenital heart disease and complex anatomy.
Keywords: Congenital heart disease, Robotic magnetic navigation, Catheter ablation
Graphical abstract
Highlights
-
•
Ablation using RMN in CHD patients appears safe and feasible for many arrhythmias.
-
•
Post-Mustard/Senning ablation using RMN avoided transbaffle puncture.
-
•
Catheter ablation using RMN was safely performed in patients with CHD anatomy.
1. Introduction
Cardiac arrhythmias are one of the most common causes of morbidity and mortality in adults with congenital heart disease (CHD) [1]. Radiofrequency (RF) catheter ablation is commonly used for the treatment of cardiac arrhythmias [2]. Understanding congenital anatomy and reviewing the details of prior cardiac surgery are essential for obtaining proper access during the ablation procedure. Nevertheless, reaching the arrhythmia substrate area and performing consistent lesions with the ablation catheter can be challenging [3,4].
Robotic magnetic navigation (RMN) systems allow catheter ablation to be manoeuvred with the use of a magnetic field and remote control [5]. They provide less exposure to X-ray radiation and decrease the risk of tissue perforation, although they may extend the procedure time compared to manual methods [6]. RMN constitutes a valuable tool to achieve access to specific arrhythmia substrate sites in patients with CHD, especially those who have undergone reparative surgery for the most complex defects, in whom conventional manually steered catheters may be unable to reach the target sites reliably [7]. The flexibility of the RMN catheter may be particularly advantageous in patients with tortuous vascular pathways, conduits, and baffles, including Mustard or Senning operations for dextro-transposition of the great arteries (d-TGA).
1.1. Aim
The aim of this study was to evaluate the efficacy, feasibility, and safety of catheter ablation in patients with CHD using the new Genesis RMN system (Stereotaxis, St. Louis, Missouri, USA).
1.2. Study oversight and support
This research received no commercial support for its conduct or the preparation of this report. In accordance with Polish law, it was determined that ethics committee approval was not required.
2. Methods
All consecutive patients with CHD who underwent catheter RF ablation for cardiac arrhythmias using RMN were included from April 2022 to September 2024. Anti-arrhythmic drugs were discontinued at least five half-lives prior to the procedure. Procedures were performed using the Genesis Robotic Magnetic Navigation System (Stereotaxis, St. Louis, Missouri, USA) and a magnetic-guided 8 Fr RF ablation catheter (THERMOCOOL® RMT Catheter, Biosense Webster, Irvine, CA, USA).
Indications for RMN included a difficult-to-access arrhythmogenic substrate site using manual techniques or suspected difficulties in maintaining catheter stability during RF energy delivery. A CT scan and echocardiography were performed prior to ablation to visualise cardiac anatomy and function. CHD was classified as mild, moderate, or severe according to a consensus statement to reflect the complexity of the disease [8].
Patients’ baseline characteristics, procedural data, such as procedure duration, radiation dose, type of vascular access, radiofrequency application parameters, and periprocedural complications were obtained. The efficacy of the ablation was evaluated during follow-up in the outpatient clinic with the use of 24-h Holter monitoring. Ablation efficacy was assessed during outpatient follow-up through clinical evaluation and rhythm monitoring. Follow-up included patient interviews regarding recurrent arrhythmia symptoms and review of any out-of-hospital electrocardiographic (ECG) recordings, together with at least one objective rhythm assessment performed ≥3 months after the procedure, including 12-lead ECG, 24-h Holter monitoring, or interrogation of an implantable cardiac device, when applicable. Acute procedural success was defined as arrhythmia termination and non-inducibility at the end of the procedure. Long-term success was defined as freedom from documented recurrence of the clinical arrhythmia, with recurrence defined as at least one episode of sustained clinical arrhythmia documented during follow-up.
2.1. Statistical analyses
Categorical data were presented as counts and percentages. Continuous variables were reported as medians and interquartile ranges (IQR – Q1 to Q3) for non-normally distributed data, or as means and standard deviations (SD) for normally distributed data. Patients were categorized by CHD complexity (simple, moderate, or complex). Exploratory pairwise comparisons between complexity groups were performed using Fisher's exact test for categorical variables, Welch's two-sample t-test for normally distributed continuous variables, and the Mann–Whitney U test for non-normally distributed continuous variables.
The analysis involved pairwise comparisons between the three categories. No adjustment was made for multiple comparisons. Statistical analysis was performed using R version 4.1.2 and RStudio software. A p-value <0.05 was considered statistically significant.
3. Results
The study group comprised 35 patients, undergoing a total of 38 catheter ablation procedures (26 in males), with a median patient age of 36.8 years (IQR 30.8–39.6). Dextro-transposition of the great arteries (d-TGA) was the most common CHD in the study population – 11 patients after Mustard or Senning procedures, 1 after a Rastelli procedure, and 1 after a Blalock–Hanlon procedure. Moreover, 19 patients (50.0%) had more than one congenital defect (Fig. 1).
Fig. 1.
A. Congenital heart diseases in described population. B. History of CHD correction. Abbreviations: ASD – Atrial septal defect; AVSD – atrioventricular septal defect; BAV – Bicuspid aortic valve; CoA – Coarctation of aorta; PDA – Patent ductus arteriosus; TGA – Transposition of great arteries; VSD – Ventricular septal defect.
Two patients (5.3%) had concomitant dilated cardiomyopathy. Heart failure was diagnosed in 25 patients (65.8%), with NYHA class I in most cases, 14 (36.8%), class II in 8 (21%), and class III in 3 (5.3%). Two patients had undergone tricuspid valve replacement.
Intra-atrial re-entrant tachycardia (IART), defined as a regular atrial arrhythmia with a confirmed macro-re-entry circuit during the procedure, was the most common arrhythmia – accounting for 27 procedures (71.1%) (Supplementary Material, Fig. S1). One patient underwent two ablations of an accessory pathway, and another underwent three IART ablation procedures.
Median procedural time was 232.5 min (IQR 186.3-281.3). The median radiation dose and radiation time were 2.4 mGy m2 (IQR 1.3–4.4) and 14.0 min (IQR 9.5–30.4), respectively.
Periprocedural data are presented in Table 1.
Table 1.
Baseline patients’ characteristics, types of ablated arrhythmias, procedural/periprocedural data and follow-up according to CHD complexity. *-Simple-moderate,**- simple-severe, and ***-moderate-severe p-values.
| Characteristic | Total (n = 38) | Simple (n = 7) | Moderate (n = 21) | Severe (n = 10) | P-value |
|---|---|---|---|---|---|
| Baseline characteristics | |||||
| Male sex | 26 (68.4%) | 4 (57.1%) | 15 (71.4%) | 7 (70.0%) | 0.647* |
| 0.644** | |||||
| 1.000*** | |||||
| Age [years] | 36.8 (30.8-39.6) | 37.7 (37.1-44.2) | 37.0 (32.1-41.9) | 28.6 (19.9-33.9) | 0.499* |
| 0.005** | |||||
| 0.006*** | |||||
| Heart failure | 25 (65.8%) | 5 (71.4%) | 15 (71.4%) | 5 (50.0%) | 1.000* |
| 0.622** | |||||
| 0.423*** | |||||
| Cardiac pacemaker | 7 (18.4%) | 1 (14.3%) | 5 (23.8%) | 1 (10.0%) | 1.000* |
| 1.000** | |||||
| 0.634*** | |||||
| CRT-D | 2 (5.3%) | 1 (14.3%) | 1 (4.8%) | 0 (0.0%) | 0.444* |
| 0.412** | |||||
| 1.000*** | |||||
| Beta-blocker | 32 (84.2%) | 6 (85.7%) | 16 (76.2%) | 10 (100.0%) | 1.000* |
| 0.412** | |||||
| 0.147*** | |||||
| Anti-arrhythmic drug | 10 (26.3%) | 2 (28.6%) | 3 (14.3%) | 5 (50.0%) | 0.574* |
| 0.622** | |||||
| 0.074*** | |||||
| Right systemic ventricle | 12 (31.6%) | 1 (14.3%) | 9 (42.9%) | 2 (20.0%) | 0.364* |
| 1.000** | |||||
| 0.262*** | |||||
| Common systemic ventricle | 4 (10.5%) | 0 (0.0%) | 0 (0.0%) | 4 (40.0%) | 1.000* |
| 0.103** | |||||
| 0.007*** | |||||
| Severely reduced contractility of the systemic ventricle | 3 (7.9%) | 1 (14.3%) | 1 (4.8%) | 1 (10.0%) | 0.444* |
| 1.000** | |||||
| 1.000*** | |||||
| Atrial fibrillation | 2 (5.3%) | 0 (0.0%) | 2 (9.5%) | 0 (0.0%) | 1.000* |
| 1.000** | |||||
| 1.000*** | |||||
| Intra-atrial re-entrant tachycardia | 27 (71.1%) | 4 (57.1%) | 17 (81.0%) | 6 (60.0%) | 0.318* |
| 1.000** | |||||
| 0.381*** | |||||
| Focal atrial tachycardia | 1 (2.6%) | 1 (14.3%) | 1 (4.8%) | 0 (0.0%) | 0.444* |
| 0.412** | |||||
| 1.000*** | |||||
| Atrioventricular nodal re-entrant tachycardia | 1 (2.6%) | 0 (0.0%) | 0 (0.0%) | 1 (10.0%) | 1.000* |
| 1.000** | |||||
| 0.323*** | |||||
| Accessory pathway | 5 (13.2%) | 0 (0.0%) | 2 (9.5%) | 3 (30.0%) | 1.000* |
| 0.228** | |||||
| 0.296*** | |||||
| Ventricular ectopic beats | 3 (7.9%) | 3 (42.9%) | 0 (0.0%) | 0 (0.0%) | 0.011* |
| 0.051** | |||||
| 1.000*** | |||||
| Procedural outcomes | |||||
| Procedural time [min] (IQR) | 232.5 (186.3-281.3) | 175.0 (157.5-225.0) | 225.0 (186.3-281.3) | 270.0 (235.0-285.0) | 0.183* |
| 0.023** | |||||
| 0.135*** | |||||
| Radiation dose [mGym2] (IQR) | 2.4 (1.3-4.4) | 2.1 (1.4-2.6) | 2.3 (1.3-3.9) | 3.4 (1.5-6.0) | 0.698* |
| 0.329** | |||||
| 0.322*** | |||||
| Periprocedural major complications | 2 (5.3%) | 0 (0.0%) | 2 (9.5%) | 0 (0.0%) | 1.000* |
| 1.000** | |||||
| 1.000*** | |||||
| Femoral vein access | 34 (89.5%) | 7 (100.0%) | 19 (90.5%) | 8 (80.0%) | 1.000* |
| 0.485** | |||||
| 0.577*** | |||||
| Femoral artery access | 17 (44.7%) | 0 (0.0%) | 10 (47.6%) | 7 (70.0%) | 0.030* |
| 0.010** | |||||
| 0.280*** | |||||
| RF time [min] (IQR) | 16.1 (10.4-21.9) | 16.4 (8.6-16.7) | 16.6 (12.7-25.7) | 12.7 (5.4-17.2) | 0.222* |
| 0.740** | |||||
| 0.263*** | |||||
| Follow-up outcomes | |||||
| Clinical arrhythmia cessation during procedure | 31 (81.6%) | 5 (71.4%) | 19 (90.5%) | 7 (70.0%) | 0.253* |
| 1.000** | |||||
| 0.296*** | |||||
| Arrhythmia-free during follow-up | 31 (81.6%) | 3 (42.9%) | 20 (95.2%) | 8 (80.0%) | 0.008* |
| 0.162** | |||||
| 0.237*** | |||||
Clinical acute arrhythmia cessation was observed after 31 procedures (81.6%). The distribution of arrhythmia recurrence over the years of the study period was comparable (Supplementary Material, Fig. S2). Periprocedural complications were noted in 2 cases (5.3%) – pericardial effusion in one patient, treated conservatively, and pharyngeal bleeding following periprocedural transoesophageal echocardiography (TEE) in another.
Follow-up data were obtained for 36 procedures. The median time from ablation to follow-up was 5.4 months (IQR 3.4–9.3). Arrhythmia recurrence was observed in 7 cases (18.4%). Partial success, defined as a significant reduction in clinical arrhythmia without complete resolution, was achieved in one patient, who had corrected d-TGA using a Mustard procedure.
In the most common CHD correction in our cohort, Mustard/Senning for d-TGA, all patients had IART. Arrhythmogenic substrate localisation was achieved using a transaortic approach in 10 patients, while the remaining patient with mechanical valve in tricuspid position underwent a standard transbaffle puncture approach under transoesophageal echocardiographic guidance [9].
Except for one patient with partial success at follow-up, no recurrences were observed. One patient with a double-outlet left ventricle and an implanted pulmonary homograft, who underwent atrial tachycardia ablation, died nine months after the procedure due to an unknown cause.
4. Discussion
Arrhythmias can have potentially fatal consequences in patients with congenital heart disease, as they often arise as part of the natural progression of the condition. In such patients, catheter ablation is indicated or should be considered [10]. In conventional manual procedures, anatomical access challenges may complicate this process and lead to delays in performing the ablation especially in patients after Mustard or Senning procedures for d-TGA. To our knowledge, this study represents the largest cohort of patients with CHD to undergo ablation using the new Genesis RMN system, which is more compact and faster than previous RMN systems.
Our analysis identified several significant findings.
-
1.
Ablation for arrhythmias in patients with CHD using the Genesis RMN system was safe and feasible.
-
2.
The use of RMN for catheter ablation in patients after Mustard or Senning correction allowed avoidance of the transbaffle puncture approach and the associated risk of residual leakage, with a very good procedural success rate, which confirms previous reports [11].
The observed ablation efficacy of approximately 80% in this selected cohort of CHD patients with difficult-to-access arrhythmogenic substrates or anticipated catheter instability underscores the value of RMN for managing complex arrhythmias in CHD [12]. Moderate CHD was the most common severity observed in the population studied. Right ventricular dysfunction was commonly observed on echocardiography while heart failure, although frequent in these patients, was mildly symptomatic (NYHA class I) in the majority of cases.
CHD makes catheter ablation challenging, but previous studies evaluating efficacy of manual ablation [[13], [14], [15]], in light of our results, suggest that RMN provides a promising method for arrhythmia treatment in hearts with congenital defects. Whereas conventional manual ablation often necessitates baffle puncture to access the pulmonary venous atrium, RMN may obviate the need for baffle puncture, thereby avoiding the risks of residual baffle leaks, hemodynamically significant shunting, and paradoxical embolism [16].
The two patients who experienced complications did not require additional procedures. One with throat hematoma was related to TEE and is not directly related with the RMN system used in this group of patients. In one case, the throat hematoma was related to TEE and was not directly associated with the RMN system used in this group of patients.
Akca et al. [17] previously described their results of ablation in the group of patients with CHD using the Niobe Magnetic Navigation System (Stereotaxis Inc., St. Louis, MO, USA). We obtained a longer procedural time, which may have been due to additional patient preparation time included in the overall procedure duration, whereas the total fluoroscopy time was shorter than in those studies.
A recently published meta-analysis about RMN in congenital heart diseases reported freedom from arrhythmia recurrence of 84.5% during a mean follow-up of 24.3 months, with the procedural complication rate of 3.5%, not attributable to RMN technology [18]. Another report revealed no significant difference in outcomes of RMN-guided ablation between acquired and congenital heart disease population [19].
The limitations of this study include the small sample size, which limits the statistical power of the results, the single-centre design, absence of a comparator group without CHD, and the relatively short follow-up period.
5. Conclusion
RMN is a valuable tool for catheter ablation of arrhythmias in patients with congenital heart disease and complex anatomy. Further studies with larger cohort size and greater number of events are needed to assess procedural efficacy and safety, especially for subgroup comparisons in a heterogenous CHD population.
Data availability statement
The data that support the findings of this study are available on request from the corresponding author (JM).
Sources of funding
Manuscript preparation was supported during Harvard Medical School's Polish Clinical Scholars Research Training Program, organized by the Agencja Badan Medycznych (ABM, English: Medical Research Agency, Warsaw, Poland).
CRediT authorship contribution statement
Łukasz Szumowski: Conceptualization, Funding acquisition, Project administration, Supervision. Michał Orczykowski: Methodology, Validation, Visualization, Writing – review & editing. Jakub Malinowski: Data curation, Formal analysis, Investigation, Resources, Writing – original draft. Piotr Urbanek: Data curation, Investigation. Robert Bodalski: Data curation, Investigation. Andrzej Hasiec: Data curation, Investigation. Grzegorz Warmiński: Data curation, Investigation. Filip Machaj: Data curation. Magdalena Lipczyńska: Data curation, Investigation, Writing – review & editing. Ewa Warchoł-Celińska: Supervision, Writing – review & editing. Mirosław Kowalski: Investigation, Validation. Maciej Sterliński: Supervision. Maria Bilińska: Investigation, Validation.
Declaration of competing interest
The authors declare the following financial interests/personal relationships which may be considered as potential competing interests: Lukasz Szumowski reports financial support was provided by Medical Research Agency, Warsaw, Poland. If there are other authors, they declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.
Footnotes
Supplementary data to this article can be found at https://doi.org/10.1016/j.ijcchd.2026.100697.
Appendix A. Supplementary data
The following is the Supplementary data to this article:
References
- 1.McLeod C.J., Warnes C. Recognition and management of arrhythmias in adult congenital heart disease. Curr Opin Cardiol. 2016;31(1):117–123. doi: 10.1097/HCO.0000000000000251. [DOI] [PubMed] [Google Scholar]
- 2.Hebe J. Role of catheter and surgical ablation in congenital heart disease. Cardiol Clin. 2002;20(3):469–486. doi: 10.1016/s0733-8651(02)00020-6. [DOI] [PubMed] [Google Scholar]
- 3.Van Hare G.F. Challenges in ablation of complex congenital heart disease. Card Electrophysiol Clin. 2019;11(4):711–718. doi: 10.1016/j.ccep.2019.08.002. [DOI] [PubMed] [Google Scholar]
- 4.Orczykowski M., Bak M., Kaczmarek K., et al. Factors influencing contact force in robotic magnetic navigation ablation. J Cardiovasc Electrophysiol. 2025;36(4):855–862. doi: 10.1111/jce.16597. [DOI] [PubMed] [Google Scholar]
- 5.Noten A.M.E., Szili-Torok T., Ernst S., et al. Best practices in robotic magnetic navigation-guided catheter ablation of cardiac arrhythmias, a position paper of the society for cardiac robotic navigation. Front Cardiovasc Med. 2024;11 doi: 10.3389/fcvm.2024.1431396. Published 2024 Sep. 27. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Da Costa A., Lafond P., Romeyer-Bouchard C., et al. Remote magnetic navigation and arrhythmia ablation. Arch Cardiovasc Dis. 2012;105(8–9):446–453. doi: 10.1016/j.acvd.2012.02.009. [DOI] [PubMed] [Google Scholar]
- 7.Suman-Horduna I., Babu-Narayan S.V., Ernst S. Remote navigation for complex arrhythmia. Arrhythmia Electrophysiol Rev. 2013;2(1):53–58. doi: 10.15420/aer.2013.2.1.53. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Khairy P., Van Hare G.F., Balaji S., et al. PACES/HRS expert consensus statement on the recognition and management of arrhythmias in adult congenital heart disease: developed in partnership between the pediatric and congenital electrophysiology society (PACES) and the heart rhythm society (HRS). Endorsed by the governing bodies of PACES, HRS, the American college of cardiology (ACC), the American heart association (AHA), the European Heart Rhythm Association (EHRA), the Canadian Heart Rhythm Society (CHRS), and the international Society for adult congenital heart disease (ISACHD) Heart Rhythm. 2014;11(10):e102–e165. doi: 10.1016/j.hrthm.2014.05.009. [DOI] [PubMed] [Google Scholar]
- 9.Orczykowski M., Lipczyńska M., Hasiec A., et al. Remote magnetic navigation catheter ablation for atrial tachycardia in a patient with d-transposition of the great arteries, atrial switch, and a mechanical valve in the tricuspid position. Kardiol Pol. 2024;82(5):560–561. doi: 10.33963/v.phj.100180. [DOI] [PubMed] [Google Scholar]
- 10.Baumgartner H., De Backer J., Babu-Narayan S.V., et al. ESC guidelines for the management of adult congenital heart disease. Eur Heart J. 2020;42(6):563–645. doi: 10.1093/eurheartj/ehaa554. 2021. [DOI] [PubMed] [Google Scholar]
- 11.Ernst S., Babu-Narayan S.V., Keegan J., et al. Remote-controlled magnetic navigation and ablation with 3D image integration as an alternative approach in patients with intra-atrial baffle anatomy. Circ Arrhythm Electrophysiol. 2012;5(1):131–139. doi: 10.1161/CIRCEP.111.962993. [DOI] [PubMed] [Google Scholar]
- 12.Waldmann V., Duthoit G., Pasquié J.L., et al. Catheter ablation in congenital heart diseases: a French nationwide study. Eur Heart J. May 21, 2025 doi: 10.1093/eurheartj/ehaf343. Published online. [DOI] [PubMed] [Google Scholar]
- 13.Liang J.J., Frankel D.S., Parikh V., et al. Safety and outcomes of catheter ablation for atrial fibrillation in adults with congenital heart disease: a multicenter registry study. Heart Rhythm. 2019;16(6):846–852. doi: 10.1016/j.hrthm.2018.12.024. [DOI] [PubMed] [Google Scholar]
- 14.de Groot N.M., Schalij M.J., Zeppenfeld K., Blom N.A., Van der Velde E.T., Van der Wall E.E. Voltage and activation mapping: how the recording technique affects the outcome of catheter ablation procedures in patients with congenital heart disease. Circulation. 2003;108(17):2099–2106. doi: 10.1161/01.CIR.0000092893.11893.38. [DOI] [PubMed] [Google Scholar]
- 15.Drago F., Russo M.S., Marazzi R., Salerno-Uriarte J.A., Silvetti M.S., De Ponti R. Atrial tachycardias in patients with congenital heart disease: a minimally invasive simplified approach in the use of three-dimensional electroanatomic mapping. Europace. 2011;13(5):689–695. doi: 10.1093/europace/eur027. [DOI] [PubMed] [Google Scholar]
- 16.Krause U., Backhoff D., Klehs S., Schneider H.E., Paul T. Transbaffle catheter ablation of atrial re-entrant tachycardia within the pulmonary venous atrium in adult patients with congenital heart disease. Europace. 2016;18(7):1055–1060. doi: 10.1093/europace/euv295. [DOI] [PubMed] [Google Scholar]
- 17.Akca F., Bauernfeind T., Witsenburg M., et al. Acute and long-term outcomes of catheter ablation using remote magnetic navigation in patients with congenital heart disease. Am J Cardiol. 2012;110(3):409–414. doi: 10.1016/j.amjcard.2012.03.040. [DOI] [PubMed] [Google Scholar]
- 18.Vô C., Bartoletti S., Benali K., et al. Robotic magnetic-guided catheter ablation in patients with congenital heart disease: a systematic review and pooled analysis. Expert Rev Cardiovasc Ther. 2023;21(3):227–236. doi: 10.1080/14779072.2023.2184798. [DOI] [PubMed] [Google Scholar]
- 19.Liu X.Y., Jacobsen P.K., Pehrson S., Chen X. Catheter ablation of incisional atrial tachycardia using remote magnetic navigation in patients after heart surgery: comparison between acquired and congenital heart disease. Europace. 2018;20(suppl_2):ii33–ii39. doi: 10.1093/europace/euy005. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
The data that support the findings of this study are available on request from the corresponding author (JM).