Abstract
JCp-1-NP is a commercially distributed pressed nano-powder pellet calcium carbonate reference material. It is widely used as a matrix-matched reference material in Laser Ablation-Inductively Coupled Plasma-Mass Spectrometry (LA-ICP-MS) analyses, yet its composition is not well-documented. In this study, we report the uncertainty of the pellet composition by analysing 20 element-to-Ca ratios measured in three different laboratories. The interlaboratory comparison dataset was outlier-corrected and provides a robust assessment of the composition and long-term stability of this nano-powder pressed pellet. In addition, a review of all publicly available compositional data using this reference material published so far is compiled. The dataset is available online at the Figshare repository.
Subject terms: Palaeoceanography, Element cycles, Marine chemistry, Geochemistry, Marine biology
Background & Summary
In the palaeoclimatology and biomineralisation community, “JCp-1” (GSJ CRM JCp-1)1 is the most used and best characterised reference material (RM)2. It is analysed either as powdered original GSJ (Geological Survey of Japan) material for solution-based bulk analysis, or as nanoparticulate pressed powder pellet made from reprocessed original GSJ material. To accomplish direct solid analysis of the GSJ powder and to ensure its analytical reproducibility, pressed nano-powder and binder-free pellets have been developed for LA-ICP-MS analysis (e.g.3). The commercially distributed JCp-1-NP (myStandards GmbH) is a nano-particulate pressed calcium carbonate powder pellet of the original JCp-1 RM, originating from the skeleton of the scleractinian coral Porites sp. It is assumed that the nano-particulate powder shows a different composition compared with the original material due to contamination with material added during the milling process (e.g.4,5). JCp-1-NP is an aragonitic powder designed for the determination of major and trace element mass fractions in biologically formed calcium carbonates and used to assess the analytical procedure and quality of LA-ICP-MS measurements. Although JCp-1(-NP) is currently the most used biogenic carbonate reference material, its multielement composition is not very well documented, underlining the importance of an interlaboratory comparison study.
In recent years, JCp-1-NP has been used as a quality control measure in several studies6–13. However, the online geological and environmental reference materials (GeoReM) database (https://georem.earth/;14) lists - until the publication of this dataset - only one publication for most elements for JCp-1-NP (GeorREM 113774). Due to the scarcity of published data regarding the trace element mass fractions in JCp-1-NP, there is a need for long-term measurements to assess the stability and homogeneity across different measurement conditions.
We present a compilation of all LA-ICP-MS measurements on a single JCp-1-NP pellet (myStandards GmbH, Kiel, Germany) with the batch number: 20170823-60. The size of this batch was in total 60 pellets. The measurements were conducted over three years at the University of Geneva (Switzerland). This work includes the pellet ablation characteristics under frequently used analytical conditions, which are important to reduce mass load of the plasma, avoid downhole fractionation and prevent contamination of the ablation cell. Further it contains a comparison of the composition of the same pellet acquired at two additional laboratories (ETH Zürich and University of Bern, Switzerland). A dataset of 20 element-to-Ca (E/Ca) ratios, ablated with varying laser parameters (spot and line analysis, different spot sizes, laser fluence, repetition rates) over three years, provide a long-term record of measurement uncertainty - comprising trueness and precision of the measurements15,16 - of major and trace elements in the carbonate RM. In addition, we summarised analytical LA-ICP-MS results from existing literature to present a dataset of published JCp-1-NP values against which our results are compared.
In summary, we characterised the chemical composition (Table 1) and the ablation efficiency of JCp-1-NP (Table 3). Regarding the ablation characteristics using two laser settings, the ablation depth was 100 to 240 nm per pulse (Table 3) and with that similar to the average aragonitic crystalline reference ablation of ≈230 nm per pulse at 2 J cm-217. Regarding the chemical characterisation, we did not observe any long-term temporal trends in measurement uncertainty for any measured E/Ca ratios across the different analytical sessions. Given the instrumental stability that was routinely examined by tuning to the same conditions, our results suggest that the reference material is stable under the chosen storage conditions (i.e., stored dark and dry in an airtight container filled with a desiccant at room temperature).
Table 2.
Results of the interlaboratory study of JCp-1-NP (continued).
| Parameters | V/Ca | Mn/Ca | Co/Ca | Cu/Ca | Zn/Ca | Sr/Ca | Y/Ca | ||
|---|---|---|---|---|---|---|---|---|---|
| Laboratory 1 (University of Geneva) | Point | Sample size (n) | 89 | 210 | 194 | 223 | 198 | 229 | 226 |
| Median [μmol/mol] | 0.54 | 3.1 | 0.56 | 0.9 | 1.4 | 8667 | 0.42 | ||
| SD [μmol/mol] | 0.03 | 0.2 | 0.05 | 0.1 | 0.2 | 180 | 0.03 | ||
| SE [μmol/mol] | <0.01 | <0.1 | <0.01 | <0.1 | <0.1 | 12 | <0.01 | ||
| 2RSD [%] | 11.9 | 14.6 | 17.4 | 19.5 | 27.2 | 4.0 | 15.3 | ||
| LOQ [μmol/mol] | 0.01 | 0.1 | 0.02 | 0.1 | 0.2 | <1 | <0.01 | ||
| Line | Sample size (n) | 15 | 30 | 18 | 18 | 17 | 30 | 31 | |
| Median [μmol/mol] | 0.54 | 2.9 | 0.55 | 0.8 | 1.2 | 8736 | 0.42 | ||
| SD [μmol/mol] | 0.02 | 0.2 | 0.04 | <0.1 | 0.1 | 94 | 0.04 | ||
| SE [μmol/mol] | <0.01 | <0.1 | 0.01 | <0.1 | <0.1 | 17 | 0.01 | ||
| 2RSD [%] | 6.4 | 12.3 | 12.9 | 9.8 | 19.9 | 2.1 | 20.4 | ||
| LOQ [μmol/mol] | 0.01 | <0.1 | 0.01 | <0.1 | 0.1 | <1 | <0.01 | ||
| Laboratory 2 (ETH Zürich) | Point | Sample size (n) | 14 | 14 | 13 | 15 | 13 | 15 | 13 |
| Median [μmol/mol] | 0.52 | 3.1 | 0.61 | 0.9 | 1.6 | 8108 | 0.40 | ||
| SD [μmol/mol] | 0.02 | 0.2 | 0.04 | 0.1 | 0.2 | 177 | 0.02 | ||
| SE [μmol/mol] | 0.01 | <0.1 | 0.01 | <0.1 | 0.1 | 46 | <0.01 | ||
| 2RSD [%] | 9.2 | 11.1 | 12.8 | 11.9 | 24.3 | 4.4 | 8.2 | ||
| LOQ [μmol/mol] | 0.04 | 0.3 | 0.31 | 0.5 | 0.5 | <1 | <0.01 | ||
| Line | Sample size (n) | 14 | 14 | 2 | 6 | 15 | 13 | 14 | |
| Median [μmol/mol] | 0.54 | 3.0 | 1.08 | 1.1 | 1.8 | 8278 | 0.41 | ||
| SD [μmol/mol] | 0.01 | 0.2 | 0.05 | 0.2 | 0.4 | 65 | 0.02 | ||
| SE [μmol/mol] | <0.01 | <0.1 | 0.03 | 0.1 | 0.1 | 18 | 0.01 | ||
| 2RSD [%] | 4.2 | 11.3 | 8.8 | 37.6 | 37.9 | 1.6 | 9.8 | ||
| LOQ [μmol/mol] | 0.05 | 0.5 | 0.87 | 1.1 | 0.8 | <1 | 0.01 | ||
| Laboratory 3 (University of Bern) | Point | Sample size (n) | 19 | 19 | 17 | 18 | 19 | 18 | 17 |
| Median [μmol/mol] | 0.56 | 3.0 | 1.10 | 0.9 | 1.6 | 8404 | 0.48 | ||
| SD [μmol/mol] | 0.03 | 0.1 | 0.07 | 0.1 | 0.2 | 127 | 0.02 | ||
| SE [μmol/mol] | 0.01 | <0.1 | 0.02 | <0.1 | <0.1 | 30 | 0.01 | ||
| 2RSD [%] | 10.5 | 6.6 | 13.0 | 19.3 | 18.9 | 3.0 | 9.6 | ||
| LOQ [μmol/mol] | 0.01 | 0.2 | 0.01 | <0.1 | 0.4 | <1 | <0.01 | ||
| Literature values (Figshare table 619) | |||||||||
| Number of publications (n) | 3 | 7 | 1 | 5 | 4 | 8 | 4 | ||
| Reported range [μmol/mol] | 0.43–0.53 | 1.6–2.5 | 0.53 | 0.8–1.2 | 1.6–2 | 7653–8814 | 0.34–0.41 | ||
Table 1.
Results of the interlaboratory study of JCp-1-NP.
| Parameters | Li/Ca | B/Ca | Na/Ca | Mg/Ca | Al/Ca | K/Ca | Ti/Ca | ||
|---|---|---|---|---|---|---|---|---|---|
| Laboratory 1 (University of Geneva) | Point | Sample size (n) | 226 | 230 | 185 | 231 | 226 | 227 | 200 |
| Median [μmol/mol] | 9.5 | 403 | 19120 | 4335 | 1905 | 515 | 13 | ||
| SD [μmol/mol] | 0.8 | 46 | 1398 | 254 | 137 | 45 | 1 | ||
| SE [μmol/mol] | 0.1 | 3 | 103 | 17 | 9 | 3 | 0 | ||
| 2RSD [%] | 16.7 | 25.8 | 15.3 | 13.4 | 14.2 | 18.6 | 16.1 | ||
| LOQ [μmol/mol] | 1.2 | 5 | 4 | <1 | <1 | 3 | <1 | ||
| Line | Sample size (n) | 31 | 31 | 19 | 31 | 16 | 15 | 16 | |
| Median [μmol/mol] | 8.7 | 355 | 17313 | 4075 | 1915 | 485 | 14 | ||
| SD [μmol/mol] | 0.6 | 67 | 829 | 408 | 104 | 57 | 1 | ||
| SE [μmol/mol] | 0.1 | 12 | 190 | 73 | 26 | 15 | <1 | ||
| 2RSD [%] | 13.2 | 37.1 | 9.5 | 20.0 | 10.8 | 24.7 | 7.8 | ||
| LOQ [μmol/mol] | 0.4 | 2 | 3 | <1 | <1 | 1 | <1 | ||
| Laboratory 2 (ETH Zürich) | Point | Sample size (n) | 15 | 15 | 13 | 13 | 14 | 15 | 14 |
| Median [μmol/mol] | 8.9 | 552 | 18763 | 4082 | 1935 | 568 | 13 | ||
| SD [μmol/mol] | 0.8 | 57 | 647 | 111 | 87 | 38 | 1 | ||
| SE [μmol/mol] | 0.2 | 15 | 180 | 31 | 23 | 10 | <1 | ||
| 2RSD [%] | 18.6 | 21.3 | 6.9 | 5.4 | 9.0 | 13.3 | 15.0 | ||
| LOQ [μmol/mol] | 5.4 | 10 | 35 | 4 | 5 | 17 | <1 | ||
| Line | Sample size (n) | 14 | 15 | 15 | 15 | 15 | 13 | 15 | |
| Median [μmol/mol] | 8.9 | 463 | 17530 | 4248 | 1788 | 494 | 13 | ||
| SD [μmol/mol] | 0.6 | 29 | 2173 | 160 | 253 | 76 | 1 | ||
| SE [μmol/mol] | 0.2 | 7 | 561 | 41 | 65 | 21 | <1 | ||
| 2RSD [%] | 12.7 | 12.1 | 23.2 | 7.5 | 27.3 | 28.9 | 14.7 | ||
| LOQ [μmol/mol] | 1.4 | 3 | 50 | 2 | 9 | 2 | 2 | ||
| Laboratory 3 (University of Bern) | Point | Sample size (n) | 19 | 19 | 19 | 19 | 19 | 19 | 19 |
| Median [μmol/mol] | 8.9 | 431 | 17454 | 3815 | 2005 | 334 | 15 | ||
| SD [μmol/mol] | 0.6 | 15 | 502 | 127 | 42 | 54 | 1 | ||
| SE [μmol/mol] | 0.1 | 3 | 115 | 29 | 10 | 12 | <1 | ||
| 2RSD [%] | 13.5 | 7.0 | 5.8 | 6.6 | 4.2 | 32.4 | 10.1 | ||
| LOQ [μmol/mol] | <0.1 | 1 | <1 | <1 | <1 | 11 | <1 | ||
| Literature values (Figshare table 619) | |||||||||
| Number of publications (n) | 2 | 4 | 4 | 8 | 4 | 2 | 2 | ||
| Reported range [μmol/mol] | 6.5–8.2 | 428–492 | 16843–20000 | 3315–4723 | 1406–1900 | 482–530 | 10–17 | ||
Table 3.
Results of the interlaboratory study of JCp-1-NP (continued).
| Parameters | Ba/Ca | La/Ca | Ce/Ca | Pb/Ca | Th/Ca | U/Ca | ||
|---|---|---|---|---|---|---|---|---|
| Laboratory 1 (University of Geneva) | Point | Sample size (n) | 225 | 158 | 160 | 227 | 83 | 232 |
| Median [μmol/mol] | 8.8 | 0.08 | 0.11 | 0.16 | 0.023 | 1.13 | ||
| SD [μmol/mol] | 0.6 | 0.01 | 0.01 | 0.01 | 0.002 | 0.05 | ||
| SE [μmol/mol] | <0.1 | <0.01 | <0.01 | <0.01 | <0.001 | <0.01 | ||
| 2RSD [%] | 14.1 | 23.3 | 25.4 | 16.7 | 20.4 | 9.5 | ||
| LOQ [μmol/mol] | <0.1 | <0.01 | <0.01 | <0.01 | <0.001 | <0.01 | ||
| Line | Sample size (n) | 31 | 3 | 3 | 28 | 3 | 31 | |
| Median [μmol/mol] | 8.3 | 0.07 | 0.12 | 0.15 | 0.022 | 1.09 | ||
| SD [μmol/mol] | 0.5 | <0.01 | 0.01 | 0.01 | 0.002 | 0.06 | ||
| SE [μmol/mol] | 0.1 | <0.01 | 0.01 | <0.01 | 0.001 | 0.01 | ||
| 2RSD [%] | 12.5 | 12.1 | 19.7 | 11.6 | 16.1 | 10.9 | ||
| LOQ [μmol/mol] | <0.1 | <0.01 | <0.01 | <0.01 | <0.001 | <0.01 | ||
| Laboratory 2 (ETH Zürich) | Point | Sample size (n) | 14 | 15 | 15 | 15 | 15 | 15 |
| Median [μmol/mol] | 8.7 | 0.07 | 0.11 | 0.18 | 0.023 | 1.24 | ||
| SD [μmol/mol] | 0.7 | 0.01 | 0.02 | 0.01 | 0.003 | 0.05 | ||
| SE [μmol/mol] | 0.2 | <0.01 | <0.01 | <0.01 | 0.001 | 0.01 | ||
| 2RSD [%] | 16.6 | 20.2 | 27.2 | 10.7 | 23.2 | 8.1 | ||
| LOQ [μmol/mol] | <0.1 | <0.01 | <0.01 | 0.01 | <0.001 | <0.01 | ||
| Line | Sample size (n) | 14 | 13 | 15 | 15 | 15 | 12 | |
| Median [μmol/mol] | 8.7 | 0.08 | 0.12 | 0.17 | 0.021 | 1.20 | ||
| SD [μmol/mol] | 0.4 | <0.01 | 0.02 | 0.01 | 0.002 | 0.03 | ||
| SE [μmol/mol] | 0.1 | <0.01 | <0.01 | <0.01 | 0.001 | 0.01 | ||
| 2RSD [%] | 10.3 | 10.0 | 31.6 | 12.8 | 19.0 | 5.8 | ||
| LOQ [μmol/mol] | <0.1 | <0.01 | <0.01 | <0.01 | <0.001 | <0.01 | ||
| Laboratory 3 (University of Bern) | Point | Sample size (n) | 18 | 18 | 17 | 19 | 19 | 18 |
| Median [μmol/mol] | 9.1 | 0.09 | 0.13 | 0.17 | 0.028 | 1.25 | ||
| SD [μmol/mol] | 0.3 | 0.01 | 0.01 | 0.02 | 0.003 | 0.04 | ||
| SE [μmol/mol] | 0.1 | <0.01 | <0.01 | <0.01 | 0.001 | 0.01 | ||
| 2RSD [%] | 6.3 | 20.8 | 16.0 | 19.2 | 21.5 | 6.1 | ||
| LOQ [μmol/mol] | <0.1 | <0.01 | <0.01 | 0.07 | <0.001 | <0.01 | ||
| Literature values (Figshare table 619) | ||||||||
| Number of publications (n) | 6 | NA | 3 | 6 | 4 | 4 | ||
| Reported range [μmol/mol] | 6.4–7.6 | NA | 0.06–0.09 | 0.13–0.16 | 0.010–0.021 | 0.98–1.11 | ||
The precision, expressed as two times the relative standard deviation (2RSD) calculated from the arithmetic mean, was generally <30% (2RSD). The interlaboratory comparison showed no systematic deviation between Laboratories 1, 2 and 3, with most E/Ca displaying a relative bias (RB), calculated as , of less than 10%. Compared to published data, our E/Ca ratios deviate by ≈23% ± 17% (RB) from those of Jochum et al.4, by ≈15% ± 30% (RB) from the manufacturer’s specified values, and on average by 12% ± 11% (RB) from other published literature values6–13.
This dataset was acquired by using the same commercially well distributed pellet over more than three years with different instrumental settings and in three different laboratories providing insights into the possible deviations arising from the different measurement conditions. The amount of data points together with the diversity of set ups in addition to a strictly uniform evaluation of the results provide a robust dataset for the carbonate reference material JCp-1-NP.
Methods
LA-ICP-MS analyses
Long-term study in Laboratory 1
Measurements were performed over three years (19.11.2021 to 27.09.2024) with a 193 nm ArF excimer laser ablation system (ESL 193 HE) coupled to an Agilent 8900 triple quadrupole inductively coupled plasma mass spectrometer (LA-ICP-MS) at the University of Geneva (Laboratory 1). During the evaluation period, spot as well as line scans were carried out on the pellet. The instrument was routinely tuned on NIST SRM 610 silicate glass18 while using a 40 μm circular spot, an ablation rate of 10 Hz, and a laser fluence of 7 J cm−2. A ThO/Th ratio (indicating oxide formation) of <0.3%, a mass 21/mass 42 ratio (doubly charged interferences) of <0.3% and a U+/Th+ ratio (element fractionation) of 1.00 to 1.05 was achieved in single quadrupole no-gas mode. The laser ablation cell was flushed with 800–850 ml min−1 He as carrier gas while Ar was used as the nebuliser gas. For signal smoothing the sampling gas was flowing through a commercial gas mixing device from ESI (Elemental Scientific Lasers).
The mass-to-charge ratios (Figshare Table 119) were selected for the least isobaric, polyatomic and multiple charged ion interferences in LA-ICP-MS analyses of carbonates. The selection was based on mass-to-charge ratio recommendations20 for low-mass resolution. A typical static spot analysis consisted of ≈30 s of background acquisition, followed by ≈45 s of signal acquisition and a washout phase using various laser ablation parameters. The dwell-times for the chosen mass-to-charge ratios are reported in Figshare Table 119. The quantification was achieved by using NIST 612 as a primary calibrant18. The quantification method was kept the same for all three laboratories.
Laboratory 2 and 3
For comparison purposes, the JCp-1-NP (myStandards GmbH, batch number: 20170823-60) was also analysed at ETH Zürich (Laboratory 2) and University of Bern (Laboratory 3) using different instruments and laser settings (Table 4). Laser parameters were kept constant during individual measurements. The dwell-times for the chosen mass-to-charge ratios for both laboratories are reported in Figshare Table 119.
Table 4.
Overview of the used instruments and their settings for all three laboratories. Value in parentheses shows median.
| Laboratory 1 | Laboratory 2 | Laboratory 3 | |
|---|---|---|---|
| ICP-MS | Agilent 8900 | Thermo Scientific ELEMENT XR | PerkinElmer Elan DRC-e |
| Laser | ESL 193 HE (Elemental Scientific), 193 nm ArF excimer laser | Resonetics RESOlution LR (ASI/Applied Spectra), 193 nm ArF excimer laser | COMPex-Pro (Lambda Physics),193 nm ArF excimer laser |
| Laser fluence | 2.0–8.0 (4.0) J cm−2 | 3.5 J cm−2 | ≈10 J cm−2 |
| Ablation chamber | TwoVol2, two-volume cell (Elemental Scientific) | S-155, two-volume cell (Laurin Technic) | custom-built 20 cm3 ablation cell |
| Spot size | ablation line and spot: 10–50 × 100–150 µm and 20–60 (60) µm respectively | ablation line and spot: 43 µm | spot: 90 µm |
| Repetition rate | 3–10 (10) Hz | 10 Hz | 10 Hz |
| Line scan speed | 5–20 (10) µm s−1 | 25 µm s−1 | — |
| spots/lines (n) | 88–232/3–31* | 15/15 | 19/0 |
*depending on the E/Ca. Refer to Table 1.
The analyses at ETH Zürich (Switzerland) were performed on 22.11.2024 using a RESOlution LR (ASI/Applied Spectra) 193 nm ArF excimer laser system coupled to an Element XR (Thermo Scientific) sector field ICP-MS. Ablation was executed in a dual-volume, fast-washout S-155 ablation cell (Laurin Technic) and fluxed with a carrier gas (≈0.25 l min−1 He) followed by a make-up gas (≈1 l min−1 Ar and 2 ml min−1 N2 to increase the sensitivity). Before introduction into the ICP-MS, the ablated aerosol was homogenised by flushing through an in-house squid device. The instrument was optimised for high sensitivity on Pb, Th and U masses whilst keeping production of oxides (248ThO+/232ThO+ ≤0.25%) low and the U/Th ratio at ≈1 (on NIST SRM 612 glass). For the analysis of JCp-1-NP, a spot size of 43 µm with a laser repetition rate of 10 Hz and a laser energy density on the surface of ≈3.5 J cm−2 was used. These ablation parameters were applied to static spot and line scan acquisitions (25 µm s−1 line scan speed). Before each line scan acquisition, the sample surface was cleansed by a fast scanning (100 µm s−1) pre-ablation line (51 µm spot size). Measurements for each line scan consisted of 19 s of background acquisition followed by 40 s of sample ablation. Before each static spot analysis, the sample surface was cleansed by three pre-ablation pulses (51 µm spot size). Measurements for static spot analysis consisted of a 40 s background acquisition followed by 40 s of sample ablation. Per peak 100 samples were recorded with a mass window of 10. A standard bracketing procedure with NIST 612 was applied to quantify the sample results and a synthetic glass GSD-1G used for quality control21.
The analysis at the University of Bern (Switzerland) was performed on 17.12.2024 using a GeoLas (Lambda Physik) pulsed 193 nm ArF excimer laser system coupled with a PerkinElmer ELAN DRC-e inductively coupled plasma quadrupole mass spectrometer. Helium was employed as carrier gas at a flow rate of 1.0 l min−1, supplemented with 8 ml min−1 H2 gas before entering the custom-built 18 cm3 ablation cell (e.g.22). Before introduction into the mass spectrometer, 0.60 l min−1 of Ar was added as nebuliser gas to the carrier gas mixture. Following thermal stabilisation of the system, the mass spectrometer was tuned on a NIST SRM 612 glass to maximise signal intensities across the entire mass range while maintaining a ThO/Th ratio below 0.5% and a Th/U sensitivity ratio between 0.97 and 1.00. Tuning was done with a circular laser beam of 64 µm diameter, at a repetition rate of 10 Hz and a fluence of 10 J cm−2. The same repetition rate and fluence were used during the analysis of JCp-1-NP. Each spot analysis consisted of a 30-second background measurement, followed by 40 seconds of ablation using a 90 µm beam diameter, and a 40-second washout period. A total of 19 measurements of JCp-1-NP were performed, bracketed by six analyses each of NIST SRM 612 and GSD-1G, with three analyses conducted before and three after the sample sequence for each standard.
Data processing
Signal processing for all three laboratory datasets was carried out in the same way using RStudio23 with R version 4.2.024. 43Ca was used as the internal standard element for signal normalisation. Signal and background were selected automatically based on changes in the acquired Ca counts using change point functions25,26. Elemental data were normalised to an assumed Ca content of 40.04 wt.% in pure aragonite to avoid any uncertainties resulting from the composition of the powdered material. At the beginning and end of each sequence, the synthetic glass standard NIST SRM 61218 was analysed. Mass fractions were calculated as element-to-Ca ratios with NIST SRM 612 for external calibration. Element-specific limit of detections (LOD)27 were calculated as:
| 1 |
With the standard deviation of the background counts, S the sensitivity, determined on the calibration material NIST SRM 612, the background counts, and the analyte counts. If measured mass fractions fell below the calculated LODs, those values were discarded in the data processing. Elemental mass fractions are presented as E/Ca ratios in µmol mol−1.
The results of this study were rounded according to the element-specific long-term uncertainty . The significant digits were calculated as:
| 2 |
Outlier detection
Three outlier detection methods were applied to the obtained E/Ca dataset: two parametric approaches assuming normal distribution (Grubbs’ test for larger sample sizes and Dixon’s Q test for smaller sample sizes) and a non-parametric method (modified z-score). Parametric tests are generally robust for larger sample sizes (n > 30–40), even with data not normally distributed27. Due to the small sample size for Laboratory 2 (n = 15 for spot and line measurements) and Laboratory 3 (n = 19), we present the results of both parametric and non-parametric tests.
Parametric tests: Grubbs’ test and Dixon’s Q test
Grubbs’28 and Dixon’s29 outlier detection tests were performed using PAST (PAleontological STatistics) software v.5.230. The Grubbs’ test statistic is defined as
| 3 |
where is the individual sample, is the sample mean, and the standard deviation. Outliers were identified iteratively, when
| 4 |
with the sample size, and is the critical value of the t-distribution at a significance level of with -2 degrees of freedom. For smaller sample sizes (N ≤ 30), Dixon’s Q test was performed by calculating the test statistic value of the sorted data according to:
| 5 |
| 6 |
with xn being the n’th value in the ordered row of datapoints. The value x1 is regarded as an outlier if the r22 is larger than the critical value for N, the number of observations and the level of significance α (here = 0.05).
Non-parametric test: Modified z-score
The modified z-score was calculated using R Version 4.2.024 with
| 7 |
where is the sample median, is the median absolute deviation, calculated as . Outliers are identified when |z| > 3.5. Trueness was evaluated using the median of samples with outliers removed via the modified z-score, and precision for JCp-1-NP is likewise reported based on the outlier-fitted dataset. Results of the Grubbs’ and Dixon’s Q test are summarised in Figshare Table 519.
Ablation characteristics
To document the ablation behaviour of the nano-powder pressed pellet a set of six ablation spots was created under typical ablation conditions (Fig. 1). Three different numbers of laser pulses per spot (n = 5, 10 and 20) were combined with two different laser fluences (2.5 and 3.5 J cm−2). The surface topography of the ablated area was imaged with a white light interferometer - a Bruker Contour GT – at the University of Lausanne, Switzerland. The software package Vision 64 from Bruker was used to export the surface morphology data. Subsequently, the data were further evaluated with MATLAB to extract an average crater depth. The arithmetical mean of three transects per ablation pit was used to calculate the depth. Each transect was background corrected with the rim values, and about 30 to 40 data points were averaged for each transect for the depth in the pit. The results are summarised in Table 5 and Fig. 1.
Fig. 1.

Topography of the pressed nano-powder pellet JCp-1-NP after laser ablation with a set of six different parameter combinations. The white color represents areas where no reliable estimates of the topography could be calculated. A: 5 pulses, 3.5 J cm−2 B: 10 pulses, 3.5 J cm−2 C: 20 pulses, 3.5 J cm−2 D: 5 pulses, 2.5 J cm−2 E: 10 pulses, 2.5 J cm−2 F: 20 pulses, 2.5 J cm−2 (for data see Table 3).
Table 5.
Ablation efficiency of the aragonitic pressed nano-powder pellet JCp-1-NP under different laser ablation settings.
| Ablation setting | Laser pulses (n) | Fluence [J cm−2] | Crater depth ± SD [µm] | Ablation efficiency ± SD [nm pulse−1] |
|---|---|---|---|---|
| A | 5 | 3.5 | 0.7 ± 0.2 | 146 ± 29 |
| B | 10 | 3.5 | 2.4 ± 0.1 | 243 ± 10 |
| C | 20 | 3.5 | 4.6 ± 0.1 | 229 ± 7 |
| D | 5 | 2.5 | 0.5 ± 0.1 | 96 ± 11 |
| E | 10 | 2.5 | 1.2 ± 0.1 | 118 ± 2 |
| F | 20 | 2.5 | 3.5 ± 0.1 | 173 ± 5 |
Data Records
Tables 1–3 show a summary of the raw median concentrations of all three laboratories for line and spot ablation of JCp-1-NP including the numbers of measurement, the analytical precision, LOQ, and the concentration range reported in literature. The LA-ICP-MS setup of all three laboratories is provided in Table 4 and the laser ablation efficiency of laboratory 1 is provided in Table 5. Figshare Table 119 contains the chosen mass-to-charge ratios of all three labs and the corresponding dwell times used. In Figshare Table 219 all raw concentration data points are provided that were used in the data descriptor. While Figshare Table 319 provides the outlier filtered data using the z-score method, Figshare Table 419 shows the Dixons Q filtered data. A summary of the filtered concentration data of all three laboratories is summarised in Figshare table 619. A compilation of all published concentration values of JCp-1-NP is provided in Figshare table 619 while Figshare table 719 contains the relative deviation of the concentrations results of laboratory 1 to published values. The data tables are provided separately as.csv files in addition to a single.xlsx file containing all tables of the data descriptor in the Figshare repository19.
Technical Validation
Measurement uncertainty and internal consistency were assessed on the nano-powder pressed pellet JCp-1-NP (myStandards GmbH, batch number: 20170823-60). Precision (2RSD) was calculated from long-term measurements at Laboratory 1. Trueness is expressed as relative bias (%)16, and was determined by comparing the outlier-removed median values to results from the two other labs (Laboratory 2 and Laboratory 3), and to published reference. These reference values were grouped into (i) GeoReM listed values or manufacturer-provided data, and (ii) values reported in studies explicitly using a JCp-1-NP pellet.
Long-term consistency of measurements
We observe no long-term temporal trends in relative bias or precision for any measured E/Ca ratios across the different sessions, indicating that the reference material is stable over time and not prone to changes or alteration while being stored in an air-tight container over silica beads to ensure a dry environment. The precision of combined line and spot analysis in Laboratory 1 was <30% (2RSD) per element (Fig. 2a). The median precision across all elements was around ≈17 ± 6% (2RSD). We can confirm the finding that the precision of elements within the natural RM is dependent on their abundance within the RM4, with lower abundance elements showing a higher uncertainty (Fig. 2a).
Fig. 2.

Quality control assessment of the chemical composition of RM JCp-1-NP (myStandards GmbH, batch number: 20170823-60). (a) Interlaboratory comparison of chemical composition determined in Laboratory 1, 2 and 3; concentrations are plotted on a logarithmic scale. (b) Comparison of Laboratory 1 results with reference values reported by Jochum et al.4 and values provided by the manufacturer (myStandards GmbH); concentrations are shown on a logarithmic scale. (c) Analytical performance of Laboratory 1 results expressed as precision (2RSD) and trueness (relative bias [%]) relative to Laboratory 2 and 3. (d) Relative bias of Laboratory 1 results compared to GeoReM and manufacturer reference values. (e) Relative bias (RB) assessment (%) of Laboratory 1 values relative to a compilation of published LA-ICP-MS data (provided in Figshare table 619).
Interlaboratory comparison
To compare the results from three participating laboratories, the relative bias (RB) was calculated as the deviation of the mass fractions measured at the collaborating laboratories 2 and 3 from the measured values obtained at Laboratory 1. The result of the interlaboratory comparison showed no systematical deviation between Laboratory 2 and Laboratory 3, with most E/Ca displaying a relative bias of <10% (Fig. 2a). While results from Laboratory 2 exhibited a median deviation of ≈4% ± 5% (RB), the acquired mass fractions from Laboratory 3 deviated by approximately 8% ± 14% (RB) from results obtained from Laboratory 1.
Comparison to the GeoReM listed values and published values by the RM manufacturer
The comparison of Laboratory 1 data with LA-ICP-MS data by Jochum et al.4 (GeoReM values), and with published LA-ICP-MS values provided by the manufacturer revealed non-uniform deviations from the reported values (Fig. 2b). On average, E/Ca ratios reported by Jochum et al.4 differed by ≈23% ± 17% (RB), while Laboratory 1 values deviated from the manufacturer’s by ≈15% ± 30% (RB). High relative deviations from measured to reported values (>40% RB) were observed for Li/Ca, Mn/Ca, and Ce/Ca when compared to the GeoReM values, and for Mn/Ca, Ce/Ca and Th/Ca relative to the manufacturer’s values. However, it is important to note that Jochum et al.4 used different analytical conditions and did not analyse a nano-powder pellet produced by myStandards GmbH, but used a custom-pressed JCp-1-NP of their own. Contamination during the milling process could have altered the composition of the pellet (e.g.4,5) resulting in a composition different to the material of this study.
Comparison to other published reference values
A compilation of eight publications that conducted LA-ICP-MS analyses on nano-powdered JCp-1 revealed large variability in reported values (Fig. 2c). While our values deviated from published values on average by 12% ± 11% (RB), element-specific deviations between studies spanned up to two orders of magnitude (e.g., the relative bias in Al/Ca ranges from 0.3% to 20%). Most of the measured E/Ca ratios agreed (<20% RB) with previously published data (Fig. 2c). Although nano-powdered pellets show an improved homogeneity4 the trace element variability in the relative deviation to the reference of JCp-1-NP caused by chemical heterogeneity within coral skeletons31 and with that to the pellet cannot not be fully excluded. Further, counting statistics that are related to the mass fraction of the element of interest in the pellet which leads to lower observed measurement uncertainties for elements that are present in higher mass fraction.
Acknowledgements
S.F., F.F., J.V. and T.S. were all supported by the SNSF Eccellenza project of TS (No. 194204). SF was additionally funded by the Swiss National Science Foundation in project RESILIENCE (200020_201106). The authors would like to thank Dieter Garbe-Schönberg and Thomas Pettke for insightful discussions and valuable feedback.
Author contributions
S.F. and F.F. designed the study, analysed the data, and wrote the manuscript. S.F., F.F., J.V., V.P., S.S., A.T. and L.T. conducted the measurements. T.E.S. and E.S. provided funding for the study. All authors reviewed and revised the manuscript.
Funding
Open access funding provided by University of Geneva.
Data availability
The dataset is available at Figshare (10.6084/m9.figshare.29682554)19.
Code availability
The R code is available at Figshare (10.6084/m9.figshare.29682554)19.
Competing interests
The authors declare no competing interests.
Footnotes
Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
These authors contributed equally: Sebastian Flöter, Frank Förster.
Contributor Information
Sebastian Flöter, Email: sebastian.floeter@unige.ch.
Frank Förster, Email: frank.foerster@unige.ch.
References
- 1.Okai, T. et al. Preparation of a New Geological Survey of Japan Geochemical Reference Material: Coral JCp‐1. Geostandards Newsletter26(1), 95–99 (2007). [Google Scholar]
- 2.Hathorne, E. C. et al. Interlaboratory study for coral Sr/Ca and other element/Ca ratio measurements. Geochemistry, Geophysics, Geosystems14(9), 3730–3750 (2013). [Google Scholar]
- 3.Garbe-Schönberg, D. & Müller, S. Nano-particulate pressed powder tablets for LA-ICP-MS. J. Anal. At. Spectrom.29(6), 990–1000 (2014). [Google Scholar]
- 4.Jochum, K. P. et al. Nano‐Powdered Calcium Carbonate Reference Materials: Significant Progress for Microanalysis? Geostandards and Geoanalytical Research43(4), 595–609 (2019). [Google Scholar]
- 5.Xu, J. D. et al. Simultaneous Bulk‐Rock Elemental and Boron Isotope Ratio Measurement Using LA‐ICP‐MS on Silicate Micro‐Particulate Pellets. Geostandards and Geoanalytical Research48(4), 807–822 (2024). [Google Scholar]
- 6.Boer, W. et al. New Calcium Carbonate Nano‐particulate Pressed Powder Pellet (NFHS‐2‐NP) for LA‐ICP‐OES, LA‐(MC)‐ICP‐MS and µXRF. Geostandards and Geoanalytical Research46(3), 411–432 (2022). [Google Scholar]
- 7.de Winter, N.J. et al. Tracing timing of growth in cultured molluscs using strontium spiking. Frontiers in Marine Science, 10 (2023).
- 8.Nambiar, R. et al. Accurate measurement of K/Ca in low‐[K] carbonate samples using laser‐ablation sector‐field inductively coupled plasma mass spectrometry. Rapid Communications in Mass Spectrometry, 38(5) (2024). [DOI] [PubMed]
- 9.Schmidt, S. et al. Incorporation of Dissolved Heavy Metals Into the Skeleton of Porites Corals Based on Multi‐Element Culturing Experiments. Geochemistry, Geophysics, Geosystems 25(9) (2024).
- 10.McCormick, C. A. et al. Basin scale evolution of zebra textures in fault‐controlled, hydrothermal dolomite bodies: Insights from the Western Canadian Sedimentary Basin. Basin Research35(5), 2010–2039 (2023). [Google Scholar]
- 11.Reuter, M. et al. Giant clam (Tridacna) distribution in the Gulf of Oman in relation to past and future climate. Sci Rep12(1), 16506 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Nambiar, R. et al. Controls on potassium incorporation in foraminifera and other marine calcifying organisms. Geochimica et Cosmochimica Acta351, 125–138 (2023). [Google Scholar]
- 13.Schmidt, S. et al. Heavy metal uptake of nearshore benthic foraminifera during multi-metal culturing experiments. Biogeosciences19(3), 629–664 (2022). [Google Scholar]
- 14.Jochum, K. P. et al. GeoReM: A New Geochemical Database for Reference Materials and Isotopic Standards. Geostandards and Geoanalytical Research29(3), 333–338 (2005). [Google Scholar]
- 15.JCGM, The international vocabulary of metrology—basic and general conceptsand associated terms (VIM) (2012).
- 16.Menditto, A., Patriarca, M. & Magnusson, B. Understanding the meaning of accuracy, trueness and precision. Accreditation and Quality Assurance12(1), 45–47 (2006). [Google Scholar]
- 17.Guillong, M. et al. Evaluating the reliability of U–Pb laser ablation inductively coupled plasma mass spectrometry (LA-ICP-MS) carbonate geochronology: matrix issues and a potential calcite validation reference material. Geochronology2(1), 155–167 (2020). [Google Scholar]
- 18.Jochum, K. P. et al. Determination of Reference Values for NIST SRM 610–617 Glasses Following ISO Guidelines. Geostandards and Geoanalytical Research35(4), 397–429 (2011). [Google Scholar]
- 19.Flöter, S. et al. Data and R code for ‘JCp-1-NP: New and compiled LA-ICP-MS elemental compositional data of the nano-powdered reference material’. Figshare10.6084/m9.figshare.29682554 (2026). [DOI] [PubMed]
- 20.Jochum, K. P. et al. Accurate trace element analysis of speleothems and biogenic calcium carbonates by LA-ICP-MS. Chemical Geology318-319, 31–44 (2012). [Google Scholar]
- 21.Guillong, M. et al. Preliminary Characterisation of New Glass Reference Materials (GSA-1G, GSC-1G, GSD-1G and GSE-1G) by Laser Ablation-Inductively Coupled Plasma-Mass Spectrometry Using 193 nm, 213 nm and 266 nm Wavelengths. Geostandards and Geoanalytical Research29(3), 315–331 (2005). [Google Scholar]
- 22.Peters, D. & Pettke, T. Evaluation of Major to Ultra Trace Element Bulk Rock Chemical Analysis of Nanoparticulate Pressed Powder Pellets by LA‐ICP‐MS. Geostandards and Geoanalytical Research41(1), 5–28 (2016). [Google Scholar]
- 23.Team, R., RStudio: Integrated Development Environment for R. 2025, RStudio, PBC: Boston, MA.
- 24.Team, R.C., A language and environment for statistical computing. 2022, R Foundation for Statistical Computing, Vienna, Austria.
- 25.Killick, R. & Eckley, I. A. changepoint: An R Package for Changepoint Analysis. Journal of Statistical Software58(3), 1–19 (2014). [Google Scholar]
- 26.Killick, R., K. Haynes, and I.A. Eckley, changepoint: An R package for changepoint analysis. 2022.
- 27.Ghasemi, A. & Zahediasl, S. Normality tests for statistical analysis: a guide for non-statisticians. Int J Endocrinol Metab10(2), 486–9 (2012). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Grubbs, F. E. Sample Criteria for Testing Outlying Observations. The Annals of Mathematical Statistics21(1), 27–58 (1950). [Google Scholar]
- 29.Dean, R. B. & Dixon, W. J. Simplified statistics for small numbers of observations. Analytical chemistry23(4), 636–638 (1951). [Google Scholar]
- 30.Hammer, O., Harper, D. & Ryan, P. PAST: Paleontological Statistics Software Package for Education and Data Analysis. Palaeontologia Electronica4, 1–9 (2001). [Google Scholar]
- 31.Meibom, A. et al. Compositional variations at ultra-structure length scales in coral skeleton. Geochimica et Cosmochimica Acta72(6), 1555–1569 (2008). [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Citations
Data Availability Statement
The dataset is available at Figshare (10.6084/m9.figshare.29682554)19.
The R code is available at Figshare (10.6084/m9.figshare.29682554)19.