Abstract
Phase change nanodroplets (PCND) consist of a liquid fluorocarbon core, which vaporizes into gas upon energy excitation, and a stabilizing shell. Through this phase-change property, PCND combine nanoparticle-like stability and pharmacokinetics, such as tissue extravasation and uptake by the mononuclear phagocyte system, with distinctive microbubble-like capabilities, namely ultrasound contrast enhancement and energy-triggered therapeutic action. PCND serve both as highly sensitive imaging agents, from molecular imaging to high-resolution vascular mapping, and as versatile therapeutics, used for example for thrombolysis, tissue ablation, biofilm removal, immunomodulation, and drug delivery. This review article provides a focused review of the biological behavior of PCND, including blood half-life, biodistribution, clearance pathways, and tumor accumulation, following a brief overview of the key physicochemical features that shape their in vivo fate. By comparing data across studies, we identify major inconsistencies in reported pharmacokinetics and highlight critical knowledge gaps, particularly regarding safety data. We further summarize advances in PCND-enabled imaging and therapy, and propose strategies to strengthen formulation design, biological evaluation, and stimulation protocols to support future clinical translation. Overall, PCND hold strong potential as a versatile theranostic platform. Continued systematic and clinically focused research is expected to accelerate their path toward clinical translation.
Key words: Nanodroplets, Phase change, Ultrasound, Biological behavior, Molecular imaging, Drug delivery, Theranostics
Graphical abstract
This review analyzes formulation-dependent pharmacokinetics and safety of phase change nanodroplets, identifies key inconsistencies and knowledge gaps, and proposes design and evaluation strategies to accelerate clinical translation in advanced applications.
1. Introduction
Ultrasound is used for many different purposes, ranging from synthetic chemistry to clinical imaging and therapy1. In medical ultrasound, the introduction of microbubbles has significantly improved diagnostic accuracy by providing information on microvascular blood flow in organs and diseased regions2. In addition, the oscillation of microbubbles during ultrasound exposure can induce cavitation, shear forces, and microstreaming, promoting transient opening of biological barriers such as tumor endothelium or the blood–brain barrier through a process known as sonopermeation, thereby enhancing drug accumulation in the targeted tissue. However, their relatively large size (1–10 μm), which prevents their extravasation into tissues and limits their circulation times to several minutes, has prompted interest in smaller nanoparticles as an alternative3. Nanoparticles, with sizes typically ranging from 5 to 200 nm, can circulate for longer times (hours to days) and accumulate in tumor tissues through mechanisms such as the enhanced permeation and retention (EPR) and the active transport and retention (ATR) effects4. However, most nanoparticles do not exhibit significant acoustic response, limiting their potential for ultrasound imaging and therapy5.
Therefore, phase change nanodroplets (PCND) have been developed as a hybrid solution, combining the prolonged circulation and tissue penetration of nanoparticles with the acoustic properties of microbubbles. This topic has seen rapidly growing interest in recent years (78 publications in 2023, 119 in 2024, and 136 in 2025, as of middle December 2025). Structurally, PCND have diameters of 50–1000 nm and consist of a liquid fluorocarbon core coated with a stabilizing shell. Upon external excitation, the liquid core can vaporize and form a gaseous microbubble that is significantly larger (up to 5 times in diameter)6 than the original PCND.
In this review, we first outline the physicochemical properties of PCND, as an essential basis to understand their working principle and functional performance. Next, we discuss the impact of their formulation design on their biological behavior regarding blood half-life, organ biodistribution, and clearance pathways, followed by a comparison of PCND with microbubbles and conventional nanoparticles in these aspects. Finally, we discuss the current limitations of PCND that hinder clinical translation and suggest strategies to overcome them.
2. Chemical composition of PCND and functionalization toolbox
The fluorocarbons used in PCND synthesis cover a wide boiling point range (from −36.7 to 145 °C) but remain in a metastable liquid state under storage conditions. This stability is reinforced by the shell, which imposes Laplace pressure around the liquid core7. Additionally, the homogeneous nucleation theory has been proposed to explain their exceptional stability, as the formation of a critical vapor nucleus requires overcoming a substantial free-energy barrier that prevents spontaneous vaporization8. Although PCND remain stable under ambient conditions, this metastability makes them highly responsive to external energy inputs. Under exposure to sufficient acoustic or radiation pressure (from ultrasound or laser pulses, respectively), these formulation characteristics ultimately determine the phase transition process known as acoustic or optical droplet vaporization (ADV or ODV)9, 10, 11. In ADV, the negative pressure phase of the ultrasound wave transiently reduces the total pressure acting on the PCND—consisting of ambient, Laplace, and acoustic components—below the vapor pressure of the fluorocarbon core, triggering a liquid-to-gas transition and generating mechanical forces12. In ODV, laser-induced heating of embedded absorbers raises the temperature of PCND above the vaporization threshold, producing microbubbles detectable by ultrasound13. Vaporized PCND primarily undergo sustained inertial cavitation14,15. This contrasts with microbubbles, where both stable and inertial cavitation mechanisms are involved in ultrasound imaging and therapy strategies16. Because the vaporization threshold of PCND depends on both external parameters (e.g., transducer frequency of ultrasound, wavelength of laser) and intrinsic formulation parameters (e.g., boiling point of fluorocarbon core, shell composition), understanding the physicochemical properties of PCND is important. This section therefore reports core and shell materials involved in PCND formulations (illustrated in Fig. 1A) and discusses how these components influence the properties of PCND. In addition, we provide examples of a typical PCND characterization workflow, including a transmission electron microscopy micrograph of a PCND17, optical micrographs captured before and after vaporization18, and representative ultrasound images10 (shown in Fig. 1B–D, respectively). These examples highlight the potential for standardizing the PCND characterization pipeline.
Figure 1.
Chemical composition and characterization of phase-change nanodroplets (PCND). (A) Schematic illustration of core and shell materials used in PCND designs. Representative fluorinated core compounds are shown as spheres, labeled with their boiling points and color-coded from light to dark blue to indicate increasing vaporization energy required for phase transition. Each core is encapsulated within a shell composed of surfactants, lipids, proteins, or polymers, which can further modulate the vaporization energy depending on their composition and interfacial properties. (B) Transmission electron microscopy image of a C6F14-core PCND with a polymer shell. Reproduced from Ref. 17, under CC-BY-NC-ND license. (C) Phase contrast microscopy images of PCND before and after ultrasound exposure, showing phase size changes. Reproduced from Ref. 18, under CC BY-4.0. (D) Contrast-mode ultrasound images showing laser-induced vaporization of PCND and enhanced echogenicity in an ex vivo spleen model. Reproduced from Ref. 10, under CC-BY-NC-ND license.
2.1. Fluorinated core
The selection of the fluorinated core material is critical for determining the stability and phase transition behavior of PCND. Fluorocarbons are small molecules composed of strong carbon–fluorine (C–F) bonds and exhibit low overall polarity due to their symmetrical structure, which contributes to bioinertness19 and hydrophobicity20. While short-chain fluorocarbons exhibit weak intermolecular forces and low boiling points, increasing molecular weight leads to higher boiling points due to stronger van der Waals interactions20. The most commonly used fluorocarbons in PCND synthesis include perfluoropropane (C3F8), perfluorobutane (C4F10), perfluoropentane (C5F12), and perfluorohexane (C6F14), with boiling points ranging from −39 °C to 56 °C. Table 1 summarizes fluorocarbons, used as core material, and the resulting PCND formulations. Rational selection of the fluorocarbon core enables fine-tuning of vaporization thresholds and optimization of ultrasound responsiveness14,21. Fluorocarbon blends (e.g., C4F10/C5F1222 or C5F12/C5H2F1023) allow modulation of vaporization thresholds and stability by altering intermolecular interactions and phase change dynamics, leading to improved control over ultrasound triggered activation. In addition to fluorocarbon, fluorinated ethers, such as C10F20O5, have emerged as promising core materials for PCND, offering a balance of low surface tension and suitable volatility that supports stable droplet formation and efficient ultrasound-triggered phase transition24.
Table 1.
Summary of PCND compositions based on the choice of fluorocarbon used for the core material, and the shell of different types.
| Core material | Molecular weight | Boiling point (°C) | Shell composition | Mean diameter | Ref. |
|---|---|---|---|---|---|
| C3F8 | 188.2 | −36.7 | Lipid | 100–200 nm | 35,36 |
| Lipid | 250–350 nm | 37 | |||
| C4F10 | 238.3 | −1.9 | Lipid | 100–200 nm | 38, 39, 40 |
| Lipid | 200–300 nm | 15,41,42 | |||
| Lipid | <1 μm | 43 | |||
| Lipid + surfactant | 300–400 nm | 44 | |||
| Lipid | 500–600 nm | 45 | |||
| Polymer | 700–800 nm | 46 | |||
| C5F12 | 288.03 | 28 | Lipid | <100 nm | 47 |
| Lipid | 100–200 nm | 48,49 | |||
| Lipid | 200–300 nm | 50 | |||
| Lipid | 300–400 nm | 51, 52, 53 | |||
| Lipid | 400–500 nm | 25,54 | |||
| Surfactant | 200–300 nm | 55 | |||
| Surfactant | 300–400 nm | 10 | |||
| Polymer | 100–200 nm | 56, 57, 58 | |||
| Polymer | 200–300 nm | 18 | |||
| Polymer | 300–400 nm | 34 | |||
| Polymer | 700–800 nm | 59,60 | |||
| C5H2F10 | 252.05 | 55 | Lipid | 300–400 nm | 54 |
| Polymer | Peak at 160,720 nm | 61 | |||
| C6F14 | 338.04 | 56 | Lipid | 100–200 nm | 14 |
| Lipid | 200–300 nm | 62 | |||
| Lipid | 300–400 nm | 63 | |||
| Surfactant | 100–200 nm | 64 | |||
| Surfactant | 300–400 nm | 65,66 | |||
| Polymer | 200–300 nm | 17 | |||
| Polymer | <100 nm | 60 | |||
| Hybrid (Lipid and polymer) | 200–300 nm | 67 | |||
| C10F20O5 | 580.07 | 145 | Polymer | 100–200 nm | 68 |
| Polymer | 200–300 nm | 24 | |||
| Polymer | 300–400 nm | 61 |
2.2. Shell materials
The shell material of PCND—typically composed of lipids, surfactants, proteins, or polymers—serves as a protective barrier that prevents rapid core dissolution and multiparticle coalescence (Fig. 1A).
Lipid-based shells, primarily made from phospholipids, such as 1,2-dipalmitoyl-sn-glycero-3–phosphocholine (DPPC) and 1,2-distearoyl-sn-glycero-3-phosphocholine (DSPC), offer excellent biocompatibility and shell elasticity25. These properties provide a favorable balance between stability and ultrasound responsiveness, making lipid shells well-suited for applications requiring efficient ultrasound-triggered vaporization.
Like lipids, synthetic amphiphilic surfactants possess a hydrophilic head and a hydrophobic tail, enabling them to self-assemble into thin monolayers26. In fluorosurfactants, the hydrocarbon tail is replaced by a fluorocarbon chain, which significantly increases interfacial tension (26–31 mN/m for fluorosurfactants vs. 7–16 mN/m for non-fluorosurfactants)27. This higher interfacial tension promotes uniform size distribution and results in smaller PCND diameters. However, a major limitation of PCND formulated with surfactants such as ZONYL-FSO, Span 20, or Pluronic F68 is their reduced stability compared to lipid-based PCND. Surfactant-coated PCND were shown to remain stable in the storage solution for only 30–60 min after synthesis and began to coalesce within 2–3 h, leading to increased PCND size28. In contrast, lipid-coated PCND remained stable for approximately 4 h and exhibited only minor size increases thereafter25.
Serum albumins are also commonly used as protein-based shell stabilizers for PCND. Human serum albumin (HSA), the most abundant protein in human plasma, has a physiological concentration of 35–50 mg/mL and a blood half-life of up to 19 days29. These properties have made HSA a key component in the formulation of early clinically approved microbubble formulations, such as air-filled Albunex® and fluorocarbon-filled successor, Optison®2. Similarly, bovine serum albumin (BSA) has been frequently used in PCND formulations30. BSA closely resembles HSA, sharing a similar molecular weight (∼66.5 kDa), disulfide bond pattern, and 76% structural identity. Its low cost makes BSA attractive for formulation development and experimental studies of PCND31. However, the non-human origin of BSA goes along with immunogenicity risks in clinical applications. Using HSA improves translational safety and regulatory acceptance, as previously demonstrated with Abraxane, a clinically approved cancer nanomedicine, which consists of HSA nanoparticles loaded with paclitaxel32.
Polymer-based shells, constructed from materials such as poly(lactic-co-glycolic acid) (PLGA), poly(caprolactone) (PCL), poly(lactic acid) (PLA) and polydopamine (PDA), offer enhanced rigidity and superior shelf-life stability at 4 °C compared to lipid-coated PCND33. In addition, amphiphilic block copolymers such as polyethylene glycol (PEG)-b-PCL, composed of hydrophobic and hydrophilic segments in a defined sequence, promote emulsification of the fluorocarbon core through hydrophobic interactions, while the hydrophilic segments provide steric stabilization in the aqueous phase34. However, their increased shell rigidity raises the vaporization threshold, requiring more intense ultrasound exposure for phase transition than their lipid- or protein-coated counterparts25,33.
2.3. Synthesis routes
The PCND synthesis toolbox includes various techniques, including homogenization, sonication, microfluidic techniques, condensation, extrusion, and spontaneous nucleation, with the choice of method depending on the specific formulation69. Emulsification-based methods—such as sonication and homogenization—are widely used for PCND synthesis due to their simplicity, accessibility of homogenizers, and scalability. Specifically, sonication employs high-frequency sound waves to disperse the fluorocarbon phase into the aqueous medium, while homogenization applies high shear forces to generate more uniformly sized PCND with improved monodispersity. While effective, these methods can lead to broad size distributions if process parameters are not carefully controlled70. In contrast, microfluidic synthesis enables precise control over PCND size and monodispersity by manipulating fluid flow within microscale channels53,71,72. While this method offers high reproducibility, its low throughput limits its applicability to research settings rather than large-scale production33. The condensation method first requires the synthesis of microbubbles with fluorocarbon gas cores, which are then condensed into submicron PCND by precisely lowering the ambient temperature and increasing the ambient pressure45,73. However, achieving a narrow size distribution at the nanoscale is challenging, as the process begins with highly polydisperse lipid-coated microbubbles45,74. Extrusion involves forcing an emulsion through a membrane with defined pore sizes, resulting in improved monodispersity compared to sonication and homogenization, which is particularly effective for achieving uniform size distributions75. Spontaneous nucleation methods, such as phase inversion and solvent evaporation, rely on physicochemical changes to induce droplet formation. While these approaches can yield highly uniform PCND, they require precise control over solvent removal and concentration gradients, making them more complex than emulsification-based techniques44,76. A comprehensive overview of synthesis techniques can be found in a review by Sheeran et al.75.
2.4. Functionalization
Surface functionalization of the PCND shell plays a critical role in prolonging circulation time and enabling targeted delivery. This can be achieved through covalent modifications, such as ligand conjugation77, or through non-covalent interactions, including hydrophobic78,79, electrostatic80, and avidin-biotin binding77.
To prolong circulation, PEGylation is commonly applied by introducing PEG chains onto the PCND outer shell, forming a hydrophilic barrier that minimizes opsonization and reduces immune recognition during PCND circulation in the blood pool81. In addition, direct coating PCND with red blood cell membranes, which retain key surface proteins such as CD47, is also possible. This “self-marker” protein inhibits phagocytosis by macrophages, thereby extending circulation time78,79.
PCND can also be functionalized with various ligands, including small molecules (e.g., folic acid), peptides, and antibodies, to enable selective binding to disease-specific biomarkers (Table 2). Tumor-targeted strategies primarily fall into three categories: vascular targeting, direct cancer cell targeting, and hypoxic region targeting. PCND designed for vascular targeting focus on markers of tumor angiogenesis. For instance, functionalization with cyclic RGD peptides enables binding to αvβ3 integrin, a key angiogenic marker overexpressed on endothelial cells50,51,78,82.
Table 2.
Selected active targeting ligands involved in PCND formulations.
| Ligands/specific molecule | Specific antigens or receptor | Shell material | Indication | Ref. |
|---|---|---|---|---|
| Plaque-targeted PP1 peptide | Class A scavenger receptors | Polymer | Macrophages in the rupture-prone plaque | 86 |
| Fibrin-targeted peptide | Fibrin | Lipid | Thrombus | 88 |
| HA | CD44, the receptor for hyaluronate-mediated motility | Hybrid (lipid and polymer) | Hepatocellular carcinoma | 80 |
| Folic acid | Folic acid receptor | Polymer | Nasopharyngeal carcinoma | 18 |
| Glycyrrhetinic acid | Glycyrrhetinic acid receptors | Lipid | Human liver cancer tumor model | 84 |
| Antibody or Herceptin/anti-HER2/neu peptide | HER2 | Polymer | HER2-overexpressing breast cancer | 77,83 |
| cRGD | Integrin αvβ3 | Hybrid (lipid and RBCM) | Tumor angiogenesis | 50,51,78,82 |
| Bifidobacterium | N/A | Lipid | Hypoxic zone of tumor | 62 |
| VCAM -1-targeted peptide | VCAM-1 | Lipid | The vascular cells of biofilm-associated surgical site infection | 87 |
Abbreviations: cRGD, cyclic Arg-Gly-Asp peptide; HA, hyaluronic acid; HER2, human epidermal growth factor receptor 2; RBCM, red blood cell membrane; VCAM-1, vascular cell adhesion molecule-1; N/A, not available.
For direct cancer cell targeting, PCND can be conjugated with antibodies or peptides that recognize receptors overexpressed on tumor cells. For example, PCND modified with anti-HER2 antibodies or HER2-binding peptides can recognize and engage with HER2-overexpressing breast cancer cells, potentially enhancing molecular imaging and cargo delivery77,83. Similarly, hyaluronic acid (HA)-modified PCND target CD44 receptors, potentially enhancing the retention of the nanoformulation in the tumor area80. Folate-conjugated PCND target folate-receptors-overexpressing cancer cells and facilitate intracellular uptake in nasopharyngeal carcinoma18. In addition, glycyrrhetinic acid-modified PCND accumulate strongly in hepatocellular carcinoma cells and improve the cytotoxic efficacy of chemotherapeutic agents84. However, despite enhanced receptor engagement and cellular uptake, active targeting often compromises nanoparticle stability and blood circulation time. Ligand modification may accelerate clearance by the mononuclear phagocyte system and reduce tumor accumulation and therapeutic efficacy. Here, a meta-analysis suggests that, in many cases, actively targeted nanoparticles do not (or only marginally) outperform passive targeting in terms of tumor delivery85.
In hypoxic regions of solid tumors, the facultative anaerobic Bifidobacterium can colonize. Positively charged PCND can attach to the negatively charged surfaces of Bifidobacterium and tumor cell membranes via electrostatic interactions, enhancing deposition and retention within the tumor microenvironment62.
Beyond tumor targeting, several ligand-modified PCND formulations have shown promise in other disease areas. For instance, PCND targeting class A scavenger receptors can accumulate in rupture-prone atherosclerotic plaques, supporting their application in anti-atherosclerosis therapy86. Additionally, PCND functionalized with peptides targeting vascular cell adhesion molecule-1 (VCAM-1) can selectively bind to endothelial cells at sites of biofilm-associated surgical site infections. This targeting facilitates transcytosis across tight vascular barriers via VCAM-1-mediated pathways, allowing the nanocarriers to penetrate biofilms and enhance therapeutic efficacy87.
Altogether, these examples highlight the potential of ligand-modified PCND to improve imaging, drug delivery, and therapeutic efficacy across diverse disease models. The choice of functionalization strategy should be guided by the desired circulation profile and target specificity to ensure optimal performance in biomedical applications.
3. In vivo fate of phase change nanodroplets
To optimize the therapeutic efficacy of PCND, the biological effects induced by the change of PCND size during vaporization at the target site89 need to be comprehensively understood. Furthermore, their circulation time, biodistribution, and clearance mechanisms can determine their effectiveness as contrast agents and drug carriers. Techniques, such as gas chromatography–mass spectrometry (GC–MS), fluorescence imaging (FI), photoacoustic imaging, positron emission tomography (PET), magnetic resonance imaging (MRI), nuclear magnetic resonance spectroscopy, and optical imaging, enable precise tracking of labeled or unlabeled PCND and shed light on their interactions with the mononuclear phagocyte system (MPS) and clearance kinetics.
3.1. Blood circulation times
Following intravenous administration, PCND interact with plasma proteins, leading to opsonization and subsequent uptake by MPS cells90. Meanwhile, some PCND extravasate into tissues and are internalized by cells91. Cellular uptake and tissue extravasation, along with the subsequent clearance, are influenced by core material, shell composition, and particle size (Table 3).
Table 3.
The half-life of PCND is dependent on their composition.
| Core | Shell | Mean diameter | Half-life | Test method | Ref. |
|---|---|---|---|---|---|
| C3F8 | Lipid | 154 ± 64 nm | 3.7 ± 1.1 min | Ultrasound (kidney) | 35 |
| C4F10 | Lipid | 166 ± 59 nm | 10.8 ± 1.6 min | Ultrasound (kidney) | 35 |
| C4F10 | Lipid | 560 ± 10 nm | 19 ± 10 min | GC–MS | 45 |
| C4F10 | Lipid | 560 ± 10 nm | 38 ± 6 min | Ultrasound (vena cava) | 45 |
| C4F10 | Lipid | 560 ± 10 nm | 37 ± 9 min | Fluorescence intensity | 45 |
| C4F10 | Lipid | 155 ± 4 nm | 26.3 ± 16.0 min | Ultrasound (kidney) | 38 |
| C4F10 | Lipid | 124 ± 4 nm | 16.6 ± 9.8 min | Ultrasound (kidney) | 38 |
| C4F10 | Lipid | 164 ± 9 nm | 2.6 ± 0.8 min | Ultrasound (kidney) | 38 |
| C6F14 | Surfactant | ∼175 nm | 12 min | 18F–PET | 64 |
| C6F14 | Lipid | 205 ± 2 nm | 43.3 min | GC–MS | 92 |
| C10F20O5 | Protein | 160 nm | 300 min | GC–MS | 68 |
| C10F20O5 | Polymer | 275 ± 50 nm | 120 min | 19F–MRI | 24 |
| C10F20O5 | Polymer | – | 360 min | 19F–MRI | 24 |
| C10F20O5 | Polymer | ∼270 nm | 240–300 min | Fluorescence intensity | 79 |
| C10F20O5 | Lipid | ∼83 nm | 360–600 min | 19F–MRI | 93 |
Abbreviations: GC–MS, gas chromatography–mass spectrometry; PET, positron emission tomography; MRI, magnetic resonance imaging.
Note: The standard deviation was not reported in some studies (e.g., 24,64,68,92).
To quantitatively assess how physicochemical factors influence PCND circulation, we summarized literature-reported blood circulation half-life data (Table 3). PCND with the lowest boiling point core (C3F8) shows very short half-lives (3.7 min, n = 1)35, while C4F10–based formulations exhibit a broader range (2.6–38 min, n = 7, median = 19 min) (Fig. 2A)35,38,45. Two C6F14 PCND were reported to have intermediate half-lives (12–43 min; median = 27.7 min)64,92, whereas droplets with high-boiling-point C10F20O5 cores remained in circulation for 120–600 min (n = 5, median = 300 min)24,68,79,93. Overall, the increase in mean half-life from approximately 4 to 300 min shows that the circulation time is strongly dependent on the core boiling point. Partial overlap exists between C4F10 and C6F14 formulations, suggesting that factors beyond the core boiling point, such as shell composition or PCND size, also influence the circulation behavior. Among C4F10–based PCND, Yoo et al.38 reported that the lipid chain length significantly affected the half-life of PCND. Shorter chains (e.g., DPPC) result in short half-lives (2.6 ± 0.8 min), whereas longer chains (e.g., DBPC) prolong circulation to 26.3 ± 15.6 min by improving PCND stability and reducing dissolution. In C10F20O5 PCND, smaller particles (>100 nm) had longer half-lives (360–600 min)93, while larger particles (>200 nm) were more rapidly cleared from the blood (half-life 120–300 min) by the MPS, as shown in Fig. 2B24,79. In addition to these formulation-dependent factors, differences in analytical methodology contribute to apparent discrepancies. Melich et al.45 demonstrated that the detection techniques resulted in varying half-life measurements for the same PCND formulation. Specifically, half-lives measured by fluorescence intensity were longer than those measured by GC–MS. This discrepancy arises because fluorescence detection reflects the persistence of labeled shell components, whereas GC–MS quantifies the core fluorocarbon content.
Figure 2.
Blood circulation half-life of phase change nanodroplets (PCND) is jointly influenced by particle size and core boiling point. Data points represent literature-reported PCND blood circulation half-lives summarized in Table 3 (n = 15); SD values were not available for all entries. (A) PCND half-lives plotted against core boiling point; dot color represents particle size, where points are slightly displaced along the X-axis at boiling points of −1.9 °C and 145 °C for visualization purposes only. (B) PCND half-lives plotted against particle size; dot color represents the boiling point of the core material.
3.2. Organ biodistribution
Liver and spleen are the dominant sites of PCND accumulation due to MPS uptake11,34,45,87,94, 95, 96, 97 (Table 4). Despite this, particle size can strongly affect their biodistribution patterns. For example, PCND in the 100–200 nm size range display lower uptake by macrophages in the liver and spleen, and overall accumulations in these organs, than PCND with sizes below 100 nm96. Lung accumulation also varies with size, but in a non-linear manner: very small PCND (∼33 nm) have an extremely low lung accumulation whereas intermediate-sized particles (∼100 nm) have high lung uptake98. Larger particles (>200 nm) again have lower lung accumulation17,18,99. These trends highlight the complex interplay between particle size and organ-specific clearance mechanisms. The accumulation of PCND in the kidneys is moderate compared to these organs, about half as high as in the liver11,18,34,51,63,80,99.
Table 4.
Overview of biodistribution studies reported for PCND.
| PCND formulation (core + shell material) | Mean diameter | Imaging label | Distribution | Test method | Ref. |
|---|---|---|---|---|---|
| C4F10 + lipid | 560 ± 10 nm | Cy5.5 | Mainly in the liver and spleen | FI | 45 |
| C4F10 + polymer | ∼800 nm | – | Mainly in the liver and spleen; minor in the kidneys. | Ultrasound | 46 |
| C5F12 + lipid | 208 ± 23 nm | Cy5 | Mainly in the liver and spleen, kidneys; less in the lungs, intestine and heart (24 h) | FI | 87 |
| C5F12 + lipid | 318 ± 149 nm | IR800 | Mainly in the liver and spleen, less in the lungs, kidneys and heart (24 h) | FI | 34 |
| C5F12 + lipid | 245 ± 10 nm | DiR | Mainly in the liver and spleen; less in the lungs, kidneys and heart (24 h) | FI | 80 |
| C5F12 + polymer | 216 ± 63 nm | DiR | Mainly in the liver and spleen; less in the lungs, kidneys and heart (2 h) | FI | 18 |
| C5F12 + polymer | <100 nm | Poly epinephrine | Mainly in the liver and spleen (4 h) | FI | 96 |
| C5F12 + lipid | 330 ± 62 nm | Cy5.5 | Mainly in the liver, less in the kidneys, spleen, intestine and heart (24 h) | FI | 51 |
| C5F12 + polymer | 357 ± 15 nm | DiR | Mainly in the liver and spleen | FI | 25 |
| C5F12 + polymer | 409 ± 25 nm | DiR | Mainly in the liver and spleen | FI | 25 |
| C5F12 + lipid | <100 nm | DiR | Mainly in the liver and spleen (6 h and 24 h) | FI | 100 |
| C5F12 + lipid | 205 ± 17 nm | ICG | Mainly in the liver, less in the kidneys and lungs (24 h) | FI | 99 |
| C5F12 + lipid | 332 ± 14 nm | DiR | Mainly in the liver and spleen (2 h) | FI | 97 |
| C6F14 + surfactant | ∼266 nm | FITC | Mainly in the kidneys, less in liver; Both decreased over 24 h; minor in the heart, spleen and lungs | FI | 17 |
| C6F14 + surfactant | ∼175 nm | 18F | Liver accumulation decreased over 2 h; spleen peaked at 45–65 min, then cleared rapidly | 18F-PET | 64 |
| C6F14 + surfactant | 33 nm | Gadolinium | Mainly the in liver and spleen; less in the kidneys, heart and lungs | T1-MRI | 11 |
| C6F14 + lipid | 106 nm | IR780 | Mainly in the liver and lungs (24 h) | FI | 98 |
| C6F14 + lipid | 300 nm | DiR | Mainly in the liver and spleen; less in the lungs, kidneys and heart (48 h) | FI | 63 |
| C6F14 + polymer | 239 ± 6 nm | Ce6 | Mainly in the liver and spleen | FI | 101 |
Abbreviations: Ce6, chlorin e6, a fluorescent dye with emission at 668 nm; Cy5, cyanine 5, a fluorescent dye with emission at 670 nm; Cy5.5, cyanine 5.5, a fluorescent dye with emission at 695 nm; DiR, 1,1ʹ-dioctadecyl-3,3,3ʹ,3ʹ-tetramethylindotricarbocyanine iodide, a fluorescent dye with emission at 778 nm; FI, fluorescence imaging; FITC, fluorescein-5-isothiocyanate, a fluorescent dye with emission at 516 nm; IR780, infrared 780, a fluorescent dye with emission at 792 nm; IR800, infrared 800, a fluorescent dye with emission at 792 nm; ICG, indocyanine green, a fluorescent dye with emission at 813 nm; MRI, magnetic resonance imaging; PET, positron emission tomography.
Note: The timepoint means the time interval after PCND injection.
3.3. Clearance
Due to the lack of specific studies on the in vivo metabolism of PCND, their clearance pathways, which probably differ between the pre- and post-vaporized states, are inferred from fluorocarbon nanoemulsions and microbubbles (Fig. 3), which share similar materials and size-dependent behaviors. Prior to vaporization, the nanometer-sized PCND are internalized by monocytes in the blood or macrophages in the tissue. The degradation of the PCND shell has not been investigated well, but it should be comparable to that of other nanoparticles using the same materials. Therefore, shell degradation may occur through lysosomal enzymatic hydrolysis or other metabolic pathways, followed by the excretion of byproducts, such as monomers or small molecules102. For example, polymeric shells are cleaved by enzymes, whereas lipid shells are hydrolyzed by lysosomal acid lipase. Protein shells, such as albumin, are degraded into peptide fragments in lysosomes throughout the body, and kidney-filtered fragments are reabsorbed and hydrolyzed into amino acids before re-entering systemic metabolism103. The released fluorocarbon core is cleared via the lungs, and the gas is exhaled. How fast fluorocarbon is exhaled depends on the PCND degradation site. After systemic administration, most PCND are internalized by macrophages in MPS organs; fluorocarbon cores are released into the bloodstream by the phagocytic cells and transported by lipoprotein carriers due to their lipophilicity. More hydrophilic fluorocarbons are rapidly cleared via the lungs104, whereas less soluble fluorocarbons may temporarily accumulate in adipose tissue before slowly redistributing into the circulation for eventual elimination through the lungs90,105,106. Additionally, the uptake of circulating PCND by blood monocytes has been reported, which subsequently migrate to the pulmonary alveoli and contribute to fluorocarbon clearance. However, this clearance pathway is considered minor compared to uptake by macrophages in MPS organs105.
Figure 3.
Clearance pathways of Phase change nanodroplets (PCND). Pre- and post-vaporized PCND follow distinct clearance routes. PCND are taken up by the mononuclear phagocyte system (MPS), where the shell undergoes degradation, and the fluorocarbon core is either exhaled or redistributed based on solubility. After vaporization, the gaseous fluorocarbon is rapidly cleared through the lungs, while residual shell components are metabolized in major MPS organs. These routes are influenced by the physicochemical state and composition of the PCND.
After vaporization, PCND are transformed into micrometer-sized bubbles, changing their clearance dynamics. The vaporized fluorocarbon is then rapidly eliminated through pulmonary exhalation, a much faster process than the MPS-mediated clearance of non-vaporized ones. The clearance rate of fluorocarbon is influenced by their vapor pressure and molecular weight, with lower molecular weight fluorocarbon being excreted more rapidly90. The fate of the shell after vaporization is expected to resemble that of non-vaporized PCND (macrophage-mediated degradation and clearance through the liver, spleen, and renal pathways). These complex clearance pathways highlight the influence of PCND states, compositions, and physicochemical properties on their biodistribution and excretion, and represent a difficult task for fulfilling regulatory required kinetic tests for clinical translation.
3.4. Safety profiles
Although safety data for PCND are limited (as the field is still at the preclinical stage of development), the use of similar fluorocarbon-based materials allows cautious extrapolation from fluorocarbon nanoemulsion studies. In a Phase Ib/II stroke trial (n = 24), a C5F12 nanoemulsion was administered intravenously at 0.001, 0.002, and 0.0034 g/kg, with no dose-limiting toxicities and no significant differences in adverse events, such as cough, headache, hypertension, and musculoskeletal pain, compared to placebo107. Although not ultrasound-activated, the shared fluorocarbon chemistry provides a clinically relevant safety context. In line with this, preclinical toxicity studies of fluorocarbon nanoemulsions indicate a high safety margin, with lethal doses ranging from 25 to 54 g fluorocarbon/kg and no-adverse-effect levels between 2.7 and 9 g fluorocarbon/kg90, both well above typical application doses of PCND (lower than 0.2 g/kg62,98). However, unlike nanoemulsions, PCND are vaporization-enabled agents, and their behavior in vivo is influenced not only by chemical composition but also by physical phase transitions. In particular, low-boiling-point fluorocarbon cores exhibit higher volatility and are more prone to rapid phase change under physiological conditions, potentially resulting in distinct and more acute biological responses. This means the safety thresholds derived from nanoemulsion studies may not directly apply to PCND, especially those formulated with low-boiling-point components.
One notable vaporization-related phenomenon is increased pulmonary residual volume (IPRV), observed in some animal models following intravenous nanoemulsion injection. IPRV is characterized by impaired lung deflation due to air trapping and is thought to result from fluorocarbon vaporization across the alveolar–capillary barrier and interactions with pulmonary surfactants, leading to foam formation. Its severity appears to correlate with fluorocarbon volatility, with more volatile cores inducing stronger effects. Susceptibility also varies by species—rabbits, pigs, and monkeys are more sensitive (e.g., 8.1 g fluorocarbon/kg in rabbits, 5.4 g in pigs, and 8.1–10.8 g in monkeys), while dogs, rats, and mice show higher tolerance. Histological analysis often reveals only mild interstitial changes, and IPRV has not been reported in humans, possibly due to anatomical and physiological differences such as larger alveoli, higher transpulmonary pressures, and more effective clearance mechanisms90.
In addition to volatility-driven effects, the ADV process itself can also induce biological effects, which are primarily governed by the acoustic parameters that control vaporization dynamics. In rat kidneys, low-boiling-point PCND were well tolerated at moderate mechanical index (MI = 0.8–1.35), whereas activation at the clinical upper limit (MI = 1.9) induced acute tubular hemorrhage and reversible elevations in serum creatinine, with full histologic recovery within 2–4 weeks. Control experiments confirmed that ultrasound exposure alone, even at comparable or higher intensities, caused no renal damage108, confirming that tissue effects arise primarily from ADV and cavitation.
Immunogenicity of shell materials is another important safety consideration. BSA, commonly used in preclinical PCND formulations, is subject to regulatory scrutiny due to theoretical bovine spongiform encephalopathy-related risks109, and its intrinsic immunogenicity. Although elevated anti-BSA antibodies have been reported without associated clinical symptoms110, its non-human origin makes it unsuitable for repeated systemic dosing. In addition, avidin-biotin coupling systems, including the widely used streptavidin-biotin variant for ligand attachment, can be associated with substantial immunogenicity111. Even clinically approved excipients such as PEG can induce anti-PEG antibodies in a significant portion of the population, potentially accelerating clearance or causing infusion reactions112. These studies outline important aspects of immunogenicity; they leave open several fundamental questions regarding dose thresholds and long-term biological effects.
Regarding organ responses, histological examinations of the liver have shown mild sinusoidal dilatation and macrophage accumulation in granulomas, indicating a sub-clinical inflammatory response potentially related to PCND inflow and hepatic uptake46. Overall, animal studies report good in vivo tolerability of PCND at appropriate doses, with no evidence of microvascular obstruction, organ toxicity, hemolysis, or discomfort11,46,58.
Nonetheless, these studies characterize only acute and formulation-specific safety aspects, while leaving critical long-term questions unresolved. Systematic evaluations of maximum tolerated dose, organ-specific toxicity, and the effects of chronic exposure, repeated dosing, and long-term organ accumulation are still lacking, hindering the establishment of reliable safety margins for clinical translation.
3.5. Comparison with microbubbles and nanoparticles
Due to their size and phase change characteristics, PCND exhibit nanoparticle-like properties before vaporization and microbubble-like properties thereafter. Since both reference systems already have clinically implemented equivalents, comparative studies with both systems are crucial for elucidating their biological behavior and as a guide for clinical translation of PCND.
3.5.1. Comparison with microbubbles
Commercial microbubbles such as Sonovue® and Definity®, ranging from 1 to 10 μm in size, remain confined to the vasculature with minimal extravascular accumulation113, 114, 115. They remain in the bloodstream for only a few minutes, depending on the gas core composition and shell material: for instance, Sonovue® (sulfur hexafluoride gas, lipid shell) has a half-life of around 6 min116, while Definity® (C3F8 gas core, lipid shell) lasts 79 ± 25 s117. Because PCND are nanoscale prior to vaporization, their biodistribution and clearance differ significantly from microbubbles.
3.5.2. Comparison with nanoparticles: nanobubbles, liposomes, protein and polymeric agents
Nanoparticles (1–100(0) nm) are typically opsonized and cleared via the MPS, similar to PCND118. Nanobubbles (200–500 nm), which possess a gaseous core stabilized by a shell, have been developed for ultrasound imaging and drug delivery beyond the vascular compartment119. Nanobubbles typically have a circulation half-life of about 2 h120, which may exceed the half-life of PCND with low-boiling-point cores (C3F8, C4F10, C6F14; −36.7 °C to 56 °C) as PCND tend to vaporize, expand, and be cleared rapidly. In contrast, PCND with high-boiling-point cores (C10F20O5; 145 °C) display slower vaporization, therefore remaining in the nanoscale range and achieving prolonged circulation comparable to or longer than that of nanobubbles. Nanoparticles with similar shell materials—such as liposomes121, 122, 123, 124, 125, polymer-based nanoparticles126, and protein-shelled nanoparticles127—generally exhibit significantly longer circulation times than PCND. This discrepancy is primarily due to the instability of the liquid fluorocarbon core of PCND, which predisposes them to rapid clearance or vaporization after injection. Thus, low-boiling-point cores limit the stability of PCND in vivo, and they also pose challenges for storage stability, indicating that improving robustness without sacrificing acoustic responsiveness should be the focus of further translational research.
3.6. Tumor accumulation
Understanding the dynamics of PCND accumulation in tumors is essential for optimizing diagnostic and therapeutic interventions, particularly in determining the optimal time window for imaging and drug release. While most studies reported an increase in tumor accumulation within 1–12 h post-injection, often peaking 24 h or even 48 h post-injection18,56,62,96,100,128,129, some studies have observed a different pattern. For instance, Xavierselvan et al.99, Wang et al.17, and Hu et al.101, working with PCND with sizes around 200–300 nm, reported rapid tumor accumulation within the first few hours, followed by a gradual decline over 24 h. Among these patterns, the commonly reported strong tumor concentrations at 24–48 h appear inconsistent with the short systemic half-lives (<45 min) reported for most PCND formulations. One possible explanation for this discrepancy lies in what is being tracked in each type of study. Tumor accumulation studies (Table 5) typically monitor fluorescence signals from labeled shell components, whereas half-life assessments (Table 3) often focus on the clearance of the fluorocarbon core using methods like GC–MS. Since the shell and core may be processed and eliminated through distinct biological pathways, tracking shell components alone may overestimate the retention of intact PCND in tumor tissue, which calls for cautious interpretation. Moreover, quantitative FI inherently faces challenges, including tissue autofluorescence, light attenuation, and depth-dependent signal loss. Studies may also methodologically differ in image acquisition and data processing, such as background subtraction, region-of-interest normalization, or other calibration methods. These methodological differences can substantially affect the fluorescence signal trends and thus the conclusions about PCND tumor accumulation.
Table 5.
Tumor accumulation of PCND assessed in rodent models.
| PCND formulation (core + shell material) | Mean diameter | Imaging label | Tumor accumulation | Method | Ref. |
|---|---|---|---|---|---|
| C5F12 + lipid | 205 ± 17 nm | ICG | Peak accumulation at 1.5 h, decreasing to 24 h | FI | 99 |
| C5F12 + lipid | ∼159 nm | Epolight 3072 dye | Significant accumulation at 24 h | Photoacoustic imaging | 134 |
| C5F12 + lipid | 61 ± 3 nm | DiR | Increase from 2 to 6 h, plateau from 6 to 24 h, decrease from 24 to 192 h | FI | 100 |
| C5F12 + lipid | 61 ± 3 nm | N/A | Increase from 0 to 6 h, plateau from 6 to 24 h | Ultrasound | 100 |
| C5F12 + polymer | ∼180 nm | Poly-epinephrine | Constant accumulation from 0 to 4 h | Ultrasound | 96 |
| C5F12 + polymer | ∼180 nm | Poly-epinephrine | Strong accumulation from 0 to 12 h, little further accumulation from 12 to 24 h, maximum accumulation at 24 h | FI | 96 |
| C5F12 + polymer | 195 ± 21 nm | Cy7 | Significant accumulation at 8 h; gradual increase from 8 to 12 h, plateau from 12 to 24 h | FI | 56 |
| C5F12 + polymer | 216 ± 33 nm | DiR | Rapid accumulation from 0 to 1 h; gradual increase from 1 to 6 h; plateau from 6 to 24 h | FI | 18 |
| C6F14 + lipid | 106 nm | IR780 | Prominent tumor accumulation at 24 h | FI | 98 |
| C6F14 + lipid | 200 nm | IR780 | Maximum accumulation at 24 h | Ultrasound + FI | 128 |
| C6F14 + lipid | 280 ± 60 nm | DiR | Maximum accumulation at 48 h | FI | 62 |
| C6F14 + lipid | 209 ± 7 nm | Polypyrrole | Maximum accumulation at 24 h | Photoacoustic imaging | 129 |
| C6F14 + lipid | 230 ± 4 nm | IR780 | Maximum accumulation at 24 h | FI | 135 |
| C6F14 + surfactant | 266 nm | FITC | Maximum accumulation at 2 h, then decrease | FI | 17 |
| C6F14 + surfactant | 33 nm | Gadolinium | Constant increase from 0 to 6 h, decrease at 24 h | T1-weighted MRI | 11 |
| C6F14 + polymer | 239 ± 6 nm | Ce6 | Rapid increase from 0 to 4 h, gradual decrease from 4 to 24 h | Photoacoustic imaging | 101 |
| C6F14 + lipid | ∼300 nm | DiR | Increase from 0 to 2 h, plateau from 2 to 48 h | FI | 63 |
| C6F14 + RBCM | ∼529 nm | DiO | Increased within 24 h | FI | 78 |
Abbreviations: Ce6, chlorin e6, a fluorescent dye with emission at 668 nm; Cy7, cyanine 7, a fluorescent dye with emission at 779 nm; DiO, dialkylcarbocyanine, a fluorescent dye with emission at 501 nm; DiR, 1,1′-dioctadecyl-,3,3′,3′-tetramethylindotricarbocyanine iodide, a fluorescent dye with emission at 778 nm; FI, fluorescence imaging; FITC, fluorescein-5-isothiocyanate, a fluorescent dye with emission at 516 nm; ICG, indocyanine green, a fluorescent dye with emission at 813 nm; IR780, infrared 780, a fluorescent dye with emission at 792 nm; MRI, magnetic resonance imaging; RBCM, red blood cell membrane; NA, not available.
Note: The timepoint means the time interval after PCND injection.
PCND accumulate in solid tumors via the EPR and ATR effects79, as other nanoparticles do. The EPR effect allows drug carriers to accumulate in the tumor microenvironment (TME) due to the abnormal architecture and increased permeability of the tumor vasculature, as well as the disturbed lymphatic and venous drainage130, 131, 132. Additionally, nanocarriers under 200 nm can pass through the leaky tumor vasculature via transvascular transport, a process contributing to ATR in tumors. Larger nanoparticles face restricted transvascular transport130, and microbubbles cannot diffuse through leaky tumor vasculature113, 114, 115. Using real-time imaging, Rapoport et al.91 showed that PCND extravasated into the tumor tissue, which was characterized by a gradual loss of fluorescence within tumor blood vessels and a simultaneous accumulation of fluorescence in the tumor parenchyma. At the microscopic level, studies have provided insights into the spatial distribution of PCND in tumors over time. Cao et al.133 reported that C5F12 + lipid PCND (mean size 234 nm) gradually extravasated, showing minimal accumulation immediately after injection, with increased localization mainly inside and around blood vessels after 12 h—consistent with the observations by Wang et al.87 that PCND remained mainly perivascular at this time point—progressing to extravascular diffusion after 24 h. Extravasation of PCND via ATR can be assumed but has not been reported so far.
PCND exhibit a non-uniform distribution within tumor tissue91,136. To overcome this, low-intensity focused ultrasound (LIFU) is employed to enhance both vascular and interstitial permeability through sonopermeation18, thereby promoting EPR–mediated accumulation and enabling a more uniform distribution within tumors91. Furthermore, the cavitation effect caused by LIFU untightens the tumor stroma and drives PCND deeper into the tissue51.
4. Biomedical applications
4.1. Diagnostic studies
PCND enable efficient and targeted tumor imaging through vaporization-triggered contrast, supporting vascular mapping, molecular imaging, and therapy monitoring. After PCND (C4F10 + lipid) vaporization by focused ultrasound, the generated bubbles can be used to detect tumor-feeding arteries for subsequent tumor embolization44. PCND can also be engineered for molecular ultrasound imaging of tumors. For instance, HA- or FA-modified PCND (Fig. 4A–C) enabled vaporization and imaging at target-rich tumor sites, achieving signal enhancement 1 h after injection11,80. For monitoring proton beam radiotherapy, C4F10 + polymer PCND vaporization by the proton beam enabled the estimation of the proton range in vivo46.
Figure 4.
Biomedical applications of phase change nanodroplets (PCND). Panels A–E illustrate diagnostic applications, while panels F–K represent therapeutic application. (A) Representative B-mode (left) and contrast-mode ultrasound images of tumors in mice treated with hyaluronic acid (HA)/PCND + low-intensity focused ultrasound (LIFU), PCND + LIFU, or HA/PCND without LIFU. Prior to administration, all tumors exhibit low baseline echogenicity. One-hour post-injection, LIFU is applied to the HA/PCND + LIFU and PCND + LIFU groups. Only the HA/PCND + LIFU group shows marked post-trigger signal enhancement, indicating efficient accumulation and LIFU-triggered acoustic droplet vaporization. (B, C) Quantitative analysis of echo intensity in B-mode and contrast-mode. Results indicate that HA/PCND accumulated effectively in tumor tissue and underwent acoustic droplet vaporization upon LIFU activation, enhancing imaging signals. These findings demonstrate the potential of PCND for targeted molecular ultrasound imaging. Reproduced from Ref. 80, under CC BY-NC 4.0. Time-dependent thrombolytic efficacy of streptokinase-loaded PCND combined with ultrasound (7.5 MHz probe, mechanical index = 1.2) during sonothrombolysis. (D) Representative ultrasound images of sonothrombolysis progression. (E) Mean grayscale intensity reduction in the region of interest demonstrates enhanced clot dissolution with streptokinase-loaded PCND over 15 min. Reproduced from Ref. 65, under CC BY-NC 3.0. LIFU-triggered propofol release from intravascular PCND suppresses seizure activity in vivo, without direct brain penetration or nonspecific tissue effects. (F) The setup consists of a transducer integrated with a stereotactic positioning frame for precise targeting. Those red ellipses represent sonication focuses. (G) Propofol-loaded or blank PCND are administered intravenously, followed by LIFU (2 min, 20% duty cycle) at the seizure focus and real-time neural activity monitoring. (H) Seizure suppression is observed only in rats receiving propofol-loaded PCND + LIFU. Ex vivo/in vivo magnetic resonance imaging and histology (I–K) reveal no parenchymal damage, confirming the safety of this technique and distinguishing it from other ultrasound-mediated neuromodulation methods that rely on nonspecific brain tissue damage. Reproduced from Ref. 34. Copyright © 2017 American Chemical Society.
Beyond tumor imaging, PCND have shown diagnostic value in inflammatory diseases. Ramirez et al.43 used contrast-enhanced ultrasound to visualize PCND accumulation in pancreatic islets, correlating with immune infiltration and enabling early detection of presymptomatic type 1 diabetes. Furthermore, Riaz et al.58 demonstrated successful cardiac chamber opacification in the ventricles of both rats and dogs, supporting the use of C5F12 + polymer PCND in echocardiography. C4F10 + lipid PCND have also been used to image the microvasculature and microcirculation in the spinal cord76. However, in this context, Li et al.76 reported that conventional B-mode imaging failed to visualize the contrast of the PCND, and assessing PCND specific nonlinear response profiles and suppressing the background using harmonic imaging sequences became essential.
PCND have also been used for ultrasound localization microscopy. In vitro, C5F12 + lipid PCND enhanced the contrast in low-flow microvessels by generating intermittent, high-intensity signals upon vaporization. However, their transient signal limits sustained flow tracking47. To address this limitation, PCND were activated by short acoustic pulses and imaged at high frame rates to reconstruct vascular tracks. Using refined acquisition modes—including multi-angle plane-wave compounding and selective activation—in vivo studies in rabbit kidneys achieved significant improvements in spatial resolution (resolving vessels at 115 μm spacing) and reduced measurement times to sub-second intervals, enabling phase-specific and branch-selective vascular imaging137.
PCND also promote multimodal imaging. For instance, loading PCND with photoabsorbers like near-infrared dyes or gold nanoparticles enables photoacoustic imaging, allowing real-time tracking of PCND and monitoring of vaporization events10. This approach also supported lymph node imaging, as PCND can reach lymphatic tissue after intradermal or intravenous injection and provide sustained ultrasound and photoacoustic contrast for up to 72 h52. Additionally, by loading PCND with Fe3O4138 or Gadovist11, they can function as T2-or T1-weighted MRI contrast agents, respectively. Besides, fluorocarbons themselves contain 19F nuclei, which can be detected by 19F MRI with high specificity. This property makes them promising candidates for non-invasive cell tracking, but low-molecular-weight 19F tracers in PCND face challenges in vivo due to their rapid efflux139. Therefore, higher-molecular-weight fluorocarbon, such as C10F20O5, are preferred.
Collectively, these findings demonstrate the broad diagnostic applicability of PCND for assessing vasculature and EPR-related processes. However, it has also become clear that the success of diagnostic PCND approaches depends not only on their contrast-generating capacity but also on their in vivo kinetics. Furthermore, the authors believe that for molecular imaging of extravascular targets (e.g., at tumor cells), as shown for liposomes, micelles, and other nanoparticles140, the non-specific, EPR-based accumulation may generate a strong background uptake that will make the correct assessment of target receptor expression very difficult.
4.2. Therapy studies
4.2.1. Vascular lesions
PCND have shown promise in vascular treatments involving cavitation-induced mechanical sonothrombolysis. Their submicron size allows for deep clot penetration, resulting in faster and more efficient lysis compared to surface-restricted microbubbles54. Kim et al.39 demonstrated that PCND achieved approximately 140% higher thrombolysis efficiency than compositionally identical microbubbles under ultrasound in an aged bovine blood clot flow model. The enhanced efficacy was attributed to their predominantly intra-clot cavitation compared with microbubbles (internal-to-external clot ratio = 0.38 vs. 0.04). Extending this concept, streptokinase-loaded PCND were developed to enhance sonothrombolysis for treating deep vein thrombosis. In vitro studies showed a significant reduction in clot weight and hemoglobin content compared with controls (Fig. 4D and E)65. Advances in device integration have improved the clinical applicability of PCND. Goel et al.40 integrated PCND with a sub-megahertz forward-viewing intravascular ultrasound catheter, achieving localized sonothrombolysis of retracted clots under clinically safe acoustic conditions (0.9 MPa, MI ≤ 1.9). PCND-mediated treatment achieved markedly higher lysis efficiency than all control groups. Clot mass was reduced by ∼30% for PCND + ultrasound and ∼40% for PCND + thrombolytic agent + ultrasound, compared with only 9%–17% for microbubble- or thrombolytic agent-only treatments, underscoring its superior thrombolytic efficacy. In addition to improved efficacy, PCND-mediated thrombolysis also offers enhanced procedural safety. Guo et al.66 reported that introducing PCND into ultrasound-mediated thrombolysis resulted in smaller overall clot debris and a markedly reduced volume fraction of large fragments (>10 μm) than microbubbles, minimizing the risk of distal embolization. This effect was attributed to the ability of PCND to penetrate fibrin pores and vaporize locally within the thrombus, enabling fine disruption.
In vivo studies have demonstrated the broad therapeutic potential of PCND in thrombolysis. In porcine models, ultrasound-induced PCND cavitation effectively dissolved coronary microthrombi141 and deep venous thrombi37, yielding over fourfold improvement in clot resolution compared with thrombolytic therapy alone37. A strategy of endowing PCND with fibrin-targeting further enhanced clot affinity and ultrasound-mediated reperfusion in a rat hindlimb microvascular obstruction model88. Converging evidence from plaque studies, Gao et al.86 observed macrophage apoptosis and dispersal of activated platelets within atherosclerotic plaques during PCND cavitation. This suggests a multifaceted therapeutic mechanism involving both physical and biological modulation. Collectively, these findings underscore that PCND are capable of achieving safe, efficient, and targeted thrombolysis across multiple vascular contexts.
4.2.2. Tumor therapy
When activated by high-intensity focused ultrasound (HIFU), PCND can induce tissue ablation through both thermal and mechanical mechanisms. For example, in an in vitro study in tissue-mimicking phantoms, selective vaporization of PCND at the acoustic focus improved energy deposition and heating efficiency, producing larger focal lesions (53 vs. 1 mm3 in controls) while minimizing near-field overheating. In contrast, microbubbles caused substantial off-target surface heating, dissipating most energy away from the focus due to their interaction with the HIFU beam along the entire acoustic path. This focal confinement enabled faster, deeper, and more controlled ablation, potentially reducing treatment time by one-third while improving safety22. Furthermore, integration with multi-focus HIFU further amplified focal heating while keeping the pre-focal temperature rise below 6 °C, nearly doubling ablation volume and efficiency142.
PCND substantially reduce the pressure threshold required to generate histotripsy. Under short-pulse insonation, Glickstein et al.143 demonstrated that effective tissue fractionation could be achieved at pressures as low as 4.1 MPa (vs. ∼20 MPa conventionally), entirely within safe acoustic limits (MI < 1.9) in an ex vivo liver model. The presence of PCND amplifies local cavitation activity and mechanical stress, leading to efficient tissue disintegration within tumor sites62,22,142.
Beyond ultrasound, optically driven vaporization of Herceptin-targeted PCND using laser stimulation has been shown to disrupt cancer cell membranes and induce targeted cell death, offering an alternative strategy for image-guided, receptor-specific therapy83.
Acoustically activated PCND also modulate immunological responses. For example, Cao et al.133 found that at low intensities (0.5–1.0 W/cm2) the subtle tissue damage induced local immune cell infiltration. Higher intensities (1.5 W/cm2) led to rapid bubble collapse, causing extensive necrosis and stronger immune responses at the lesion margins. In addition, Li et al.50 found that LIFU-triggered PCND induced immunogenic cell death by promoting the release of necrosis-associated damage-associated molecular patterns release and calreticulin exposure. These effects enhanced dendritic cell maturation and downstream antitumor immunity. These studies suggest that PCND-mediated ablation can be used to enhance anti-tumor immunity.
In addition to direct ablation and immune modulation, PCND can inhibit tumor growth by selectively disrupting the tumor vasculature. Ultrasound activated antifibrinolytic drug-loaded PCND mediate a cavitation-induced endothelial damage and initiate the extrinsic coagulation cascade, leading to the selective thrombotic infarction of tumors, effectively starving rapidly proliferating tumor tissues78.
Furthermore, PCND can act as oxygen reservoirs due to the high solubility of oxygen in liquid fluorocarbon and help to alleviate tumor hypoxia, a major obstacle to many therapies. Accordingly, oxygen-carrying PCND have been shown to enhance multiple oxygen-dependent modalities, including radiotherapy79,98, photodynamic therapy99,101,128,144, photothermal therapy67,96,129, chemo-dynamic therapy56,96, and sonodynamic therapy42,50,135. These enhancements are generally attributed to increased generation and amplification of reactive oxygen species generation (e.g. 1O2 and OH)42,50,128,135, greater radiation-induced DNA damage79,98, and elevated oxidative stress and redox activity56,96 that collectively promote tumor cell death.
Drug delivery is another application field for PCND. Drugs (e.g., doxorubicin or quercetin) loaded in PCND have shown higher serum concentrations, prolonged circulation half-lives, reduced clearance, and greater tumor accumulation (over 3-fold increases 5 h post i.v.) than in their free form, primarily due to the EPR effect and reduced off-target clearance25,145. Moreover, encapsulating drugs like doxorubicin in PCND can improve hemocompatibility, likely by reducing myelosuppression and hemolysis of red blood cells60. In addition to improved pharmacokinetics and accumulation, ultrasound plays a pivotal role in optimizing the intratumoral distribution and therapeutic efficacy of PCND-delivered drugs. PCND + LIFU enhanced vascular permeability and promoted deeper and more uniform drug distribution within the TME146. It also enhanced drug penetration into poorly perfused tumor cores49. In another publication, it was shown that the extravasation distance of doxorubicin from the vessel wall increased from 12 μm to over 100 μm upon ultrasound exposure146. Correspondingly, LIFU significantly elevated intratumoral doxorubicin concentrations, confirming its role in facilitating local drug uptake and triggering PCND release60. The concept was also successful when PCND were stimulated by a laser147. Furthermore, ultrasound-induced PCND cavitation promoted transient permeabilization of cellular and nuclear membranes, facilitating intracellular and intranuclear drug delivery61. Besides improving therapeutic efficacy, PCND ultrasound minimized off-target toxicity in cardiac, hepatic, and renal tissues58,97.
Building on their potential for drug delivery, PCND have been engineered to encapsulate chemotherapeutic agents like paclitaxel24,59,136, docetaxel11,25,49,57,60,129,148, or 10-hydroxycamptothecin80,138. These formulations effectively reduced tumor volume, suppressed proliferation, and promoted apoptosis. In a triple-negative breast cancer model, paclitaxel- and doxorubicin-loaded PCND improved drug delivery efficiency and delayed tumor progression97. Additionally, hydroxychloroquine-loaded PCND, an autophagy inhibitor, suppressed tumor growth by modulating the autophagic process in tumor cells17. In addition, combination with anti-PD-L1 antibodies has expanded PCND’s utility in immunotherapy, improving immune activation and tumor suppression50,100,135. PCND have also been explored for nucleic acid delivery, including plasmid DNA for gene transfection77 and microRNA-122, a key regulator of fatty-acid metabolism whose downregulation contributes to the development of hepatocellular carcinoma149.
4.2.3. Disorders of central nervous system
In central nervous system applications, PCND combined with LIFU can induce localized neuromodulation, likely mediated by cavitation-generated mechanical forces transmitted from cerebral vasculature to surrounding brain tissue, activating mechanosensitive pathways150. They also confine the action of BBB-permeable drugs to millimeter-scale brain regions. For example, LIFU-triggered release of propofol from C5F12–cored PCND occurred only within the focal vascular bed, enabling focal drug entry during first-pass perfusion and resulting in seizure suppression without inducing off-target anesthesia (Fig. 4F–K)34. PCND can also be operated in regimes where BBB opening is desired, with the magnitude and spatial extent tuned by the core composition. Owing to the low vaporization threshold of C3F8–cored PCND (MI = 0.24), ADV can be achieved at modest pressures (MI = 0.3–0.5) that are compatible with prolonged and large-volume sonication, enabling broad BBB opening. In contrast, C4F10-cored PCND require higher vaporization pressures (MI = 1.2), exceeding the pressure tolerated by brain tissue for prolonged sonication (MI ≤ 0.7–0.9). Consequently, high-pressure exposure (MI = 1.2) must be restricted to brief, focal bursts to trigger ADV, followed by lower-pressure pulses (MI = 0.5) to sustain locally generated bubble oscillations, resulting in a highly localized BBB opening36.
4.2.4. Biofilm
Bacterial biofilms are responsible for many chronic and relapsing infections. They respond poorly to antibiotics because they contain drug-tolerant persister cells (subpopulations of bacteria that enter a dormant state, surviving antibiotic treatments, even if they are not genetically resistant) and a biofilm matrix that serves as a physical barrier to antimicrobial penetration151. To address this, Durham et al.152 used PCND combined with ultrasound to disrupt methicillin-resistant Staphylococcus aureus biofilms and improve drug penetration, thereby enhancing the effectiveness of antibiotics against persister cells. Choi et al.153. further demonstrated that PCND-mediated delivery increased antibiotics accumulation in bacterial membranes and cytoplasm by more than tenfold, reducing the concentration needed for complete persister clearance. This strategy has been adapted to clinically relevant biofilm models. A VCAM-1-targeted PCND was developed for biofilm-associated surgical site infections, where VCAM-1-mediated transcytosis enables PCND to cross the inflamed endothelium and extravasate into infected tissue surrounding the biofilm. Upon ultrasound activation, cavitation, generated by PCND, produces strong mechanical forces that propagate into the adjacent biofilm. These forces disrupt the biofilm extracellular polymeric substance, a microbially secreted extracellular macromolecular matrix that forms the structural scaffold of the biofilm, thereby enhancing drug penetration87. Similarly, Dang et al.48 developed PCND loaded with chlorhexidine or antibiotics to disrupt mature Enterococcus faecalis biofilms in a human tooth model, offering a potential strategy for root canal disinfection. Together, these findings highlight the strong translational potential of PCND-assisted sonotherapy for recalcitrant biofilm infections, although its efficacy and safety in complex in vivo environments remain to be fully validated.
4.3. Clinical translation status
Although no registered clinical trials currently investigate PCND in patients, several early translational efforts indicate that PCND-based systems are progressing toward clinical evaluation. Examples include pre-clinical programs for Microvascular Therapeutics for brain tumor ablation (MVT-101) and ultrasound-mediated thrombolysis (MVT-201)154,155.
In parallel, fluorocarbon nanoemulsions with similar submicron dimensions and fluorocarbon cores as PCND have entered early phase human studies. NuvoX Therapeutics is currently conducting three clinical programs using its oxygen-carrying C5F12 nanoemulsion (NVX-108 for glioblastoma156, a Phase Ib/II trial in acute ischemic stroke107, and the PROVEN Phase IIb trial for large-vessel-occlusion stroke157). While these nanoemulsions are not activated by ultrasound and therefore do not constitute PCND in the strict sense, their reproducible manufacturing, intravenous safety, and regulatory progress provide encouraging precedents for the potential clinical translation of PCND.
5. Conclusions
PCND represent a highly versatile platform for ultrasound-mediated diagnostics and therapy. Their nanoscale dimensions, biocompatibility, and triggerable vaporization offer a suite of capabilities ranging from image-guided drug delivery and neuromodulation to oxygenation modulation and transient disruption of biological barriers. Despite encouraging preclinical progress, three key challenges continue to hinder the clinical translation of PCND: (1) inconsistent data on pharmacokinetics, (2) confliction and incomplete safety evaluations, and (3) limited vaporization control under physiological conditions. These challenges largely stem from formulation-dependent properties, for example the complex interplay between core boiling point, shell composition, and droplet size, which together affect both pharmacokinetics and vaporization behavior. As a result, reported circulation half-lives and tumor accumulation profiles diverge across studies, a variability further amplified by different imaging protocols and data analysis methods. Addressing these challenges will require both rational formulation strategies to improve systemic stability and on-demand responsiveness, and the development of standardized pharmacokinetic assessment methods to ensure cross-study comparability. Specifically, actionable steps include establishing application-specific formulation guidelines that account for fluorocarbon core and shell properties as well as adopting fixed pharmacokinetic sampling time points across studies to reduce methodological variability in imaging and data analysis.
Safety remains another critical knowledge gap. In particular, low-boiling-point formulations show greater sensitivity to ambient conditions and may induce distinct biological responses, such as pulmonary effects, not observed with more stable nanoemulsions. This underscores that existing safety thresholds derived from studies on nanoemulsions may not directly apply to volatile PCND formulations. Although preclinical data suggest a generally good safety profile at adjusted doses, systematic evaluation of dose limits, organ-specific toxicity, and long-term safety are still lacking, especially for low-boiling-point cores. This gap must be addressed to ensure safe clinical translation.
Beyond biological and safety considerations, formulation-driven characteristics also present practical manufacturing challenges. The need for precise control of droplet size distribution, encapsulated fluorocarbon content, and shell poses challenges for upscaling, while the metastable nature of superheated cores raises concerns regarding shelf life and handling. Ensuring batch-to-batch reproducibility and establishing GMP-compatible production methods will therefore be essential. In parallel, robust and reproducible ultrasound protocols, tuned to the activation thresholds and vaporization dynamics of each formulation, are required to achieve consistent therapeutic outcomes.
Looking forward, PCND systems have the potential to drive a new generation of precision ultrasound theranostics. Their unique phase-change properties and multifunctionality set them apart from other nanomaterials. Beyond oncology, their applications may extend to targeted neurostimulation, immune modulation, and treatment of vascular or inflammatory diseases. However, translating these promising PCND into clinical practice will require overcoming significant technical and translational challenges through thoughtful design, rigorous evaluation, and coordinated interdisciplinary efforts.
Author contributions
Xiaoyu Wang: data curation, writing–original draft, writing–review & editing, conceptualization. Roman Barmin: writing–original draft, writi–review & editing, conceptualization; Roger Molto Pallares: writing–original draft, writing–review & editing, conceptualization. Anne Rix: writing–original draft, writing–review & editing, conceptualization. Fabian Kiessling: writing–original draft, writing–review & editing, conceptualization, supervision, funding acquisition.
Declaration of generative AI in scientific writing
The authors declare that no generative artificial intelligence (AI) or AI-assisted technologies were used in the writing, editing, or preparation of this manuscript.
Conflicts of interest
The authors declare no conflicts of interest.
Acknowledgments
The authors acknowledge funding support from the DFG Research Training Group RTG2375 (project number 331065168), the China Scholarship Council (CSC:202208230040) and the Leibniz ScienceCampus ACTISONO, supported by the Leibniz Association (grant number W89/2023).
Footnotes
Peer review under the responsibility of Chinese Pharmaceutical Association and Institute of Materia Medica, Chinese Academy of Medical Sciences.
References
- 1.Jiao H., Mao Q., Razzaq N., Ankri R., Cui J. Ultrasound technology assisted colloidal nanocrystal synthesis and biomedical applications. Ultrason Sonochem. 2024;103 doi: 10.1016/j.ultsonch.2024.106798. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Paefgen V., Doleschel D., Kiessling F. Evolution of contrast agents for ultrasound imaging and ultrasound-mediated drug delivery. Front Pharmacol. 2015;6:197. doi: 10.3389/fphar.2015.00197. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Mannaris C., Bau L., Grundy M., Gray M., Lea-Banks H., Seth A., et al. Microbubbles, nanodroplets and gas-stabilizing solid particles for ultrasound-mediated extravasation of unencapsulated drugs: an exposure parameter optimization study. Ultrasound Med Biol. 2019;45:954–967. doi: 10.1016/j.ultrasmedbio.2018.10.033. [DOI] [PubMed] [Google Scholar]
- 4.Dasgupta A., Sofias A.M., Kiessling F., Lammers T. Nanoparticle delivery to tumours: from EPR and ATR mechanisms to clinical impact. Nat Rev Bioeng. 2024;2:714–716. doi: 10.1038/s44222-024-00203-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Barmin R.A., Moosavifar M., Dasgupta A., Herrmann A., Kiessling F., Pallares R.M., et al. Polymeric materials for ultrasound imaging and therapy. Chem Sci. 2023;14:11941–11954. doi: 10.1039/d3sc04339h. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Wong Z.Z., Kripfgans O.D., Qamar A., Fowlkes J.B., Bull J.L. Bubble evolution in acoustic droplet vaporization at physiological temperature via ultra-high speed imaging. Soft Matter. 2011;7:4009–4016. [Google Scholar]
- 7.Pitt W.G., Singh R.N., Perez K.X., Husseini G.A., Jack D.R. Phase transitions of perfluorocarbon nanoemulsion induced with ultrasound: a mathematical model. Ultrason Sonochem. 2014;21:879–891. doi: 10.1016/j.ultsonch.2013.08.005. [DOI] [PubMed] [Google Scholar]
- 8.Mountford P.A., Borden M.A. On the thermodynamics and kinetics of superheated fluorocarbon phase-change agents. Adv Colloid Interface Sci. 2016;237:15–27. doi: 10.1016/j.cis.2016.08.007. [DOI] [PubMed] [Google Scholar]
- 9.Loskutova K., Grishenkov D., Ghorbani M. Review on acoustic droplet vaporization in ultrasound diagnostics and therapeutics. Biomed Res Int. 2019;2019 doi: 10.1155/2019/9480193. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Cao Y., Dumani D.S., Hallam K.A., Emelianov S.Y., Ran H. Real-time monitoring of NIR-triggered drug release from phase-changeable nanodroplets by photoacoustic/ultrasound imaging. Photoacoustics. 2023;30 doi: 10.1016/j.pacs.2023.100474. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Maghsoudinia F., Tavakoli M.B., Samani R.K., Hejazi S.H., Sobhani T., Mehradnia F., et al. Folic acid-functionalized gadolinium-loaded phase transition nanodroplets for dual-modal ultrasound/magnetic resonance imaging of hepatocellular carcinoma. Talanta. 2021;228 doi: 10.1016/j.talanta.2021.122245. [DOI] [PubMed] [Google Scholar]
- 12.Qin D., Zou Q., Lei S., Wang W., Li Z. Predicting initial nucleation events occurred in a metastable nanodroplet during acoustic droplet vaporization. Ultrason Sonochem. 2021;75 doi: 10.1016/j.ultsonch.2021.105608. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Strohm E., Rui M., Gorelikov I., Matsuura N., Kolios M. Vaporization of perfluorocarbon droplets using optical irradiation. Biomed Opt Express. 2011;2:1432–1442. doi: 10.1364/BOE.2.001432. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Zhang W., Metzger H., Vlatakis S., Claxton A., Carbajal M.A., Fung L.F., et al. Characterising the chemical and physical properties of phase-change nanodroplets. Ultrason Sonochem. 2023;97 doi: 10.1016/j.ultsonch.2023.106445. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Burgess M.T., Porter T.M. Acoustic cavitation-mediated delivery of small interfering ribonucleic acids with phase-shift nano-emulsions. Ultrasound Med Biol. 2015;41:2191–2201. doi: 10.1016/j.ultrasmedbio.2015.04.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Hernot S., Klibanov A.L. Microbubbles in ultrasound-triggered drug and gene delivery. Adv Drug Deliv Rev. 2008;60:1153–1166. doi: 10.1016/j.addr.2008.03.005. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Wang X., Shang M., Sun X., Guo L., Xiao S., Shi D., et al. Dual-responsive nanodroplets combined with ultrasound-targeted microbubble destruction suppress tumor growth and metastasis via autophagy blockade. J Control Release. 2022;343:66–77. doi: 10.1016/j.jconrel.2022.01.009. [DOI] [PubMed] [Google Scholar]
- 18.Yang D., Chen Q., Zhang M., Feng G., Sun D., Lin L., et al. Drug-loaded acoustic nanodroplet for dual-imaging guided highly efficient chemotherapy against nasopharyngeal carcinoma. Int J Nanomed. 2022;17:4879–4894. doi: 10.2147/IJN.S377514. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.O’Hagan D. Understanding organofluorine chemistry. An introduction to the C–F bond. Chem Soc Rev. 2008;37:308–319. doi: 10.1039/b711844a. [DOI] [PubMed] [Google Scholar]
- 20.Miller M.A., Sletten E.M. Perfluorocarbons in chemical biology. Chembiochem. 2020;21:3451–3462. doi: 10.1002/cbic.202000297. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Sheeran P.S., Dayton P.A. Improving the performance of phase-change perfluorocarbon droplets for medical ultrasonography: current progress, challenges, and prospects. Scientifica (Cairo) 2014;2014 doi: 10.1155/2014/579684. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Phillips L.C., Puett C., Sheeran P.S., Wilson Miller G., Matsunaga T.O., Dayton P.A. Phase-shift perfluorocarbon agents enhance high intensity focused ultrasound thermal delivery with reduced near-field heating. J Acoust Soc Am. 2013;134:1473–1482. doi: 10.1121/1.4812866. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Kawabata K.I., Sugita N., Yoshikawa H., Azuma T., Umemura S.I. Nanoparticles with multiple perfluorocarbons for controllable ultrasonically induced phase shifting. Jpn J Appl Phys. 2005;44:4548. [Google Scholar]
- 24.Rapoport N., Nam K.H., Gupta R., Gao Z., Mohan P., Payne A., et al. Ultrasound-mediated tumor imaging and nanotherapy using drug loaded, block copolymer stabilized perfluorocarbon nanoemulsions. J Control Release. 2011;153:4–15. doi: 10.1016/j.jconrel.2011.01.022. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Cao Y., Chen Y., Yu T., Guo Y., Liu F., Yao Y., et al. Drug release from phase-changeable nanodroplets triggered by low-intensity focused ultrasound. Theranostics. 2018;8:1327–1339. doi: 10.7150/thno.21492. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Jahan R., Bodratti A.M., Tsianou M., Alexandridis P. Biosurfactants, natural alternatives to synthetic surfactants: physicochemical properties and applications. Adv Colloid Interface Sci. 2020;275 doi: 10.1016/j.cis.2019.102061. [DOI] [PubMed] [Google Scholar]
- 27.Astafyeva K., Somaglino L., Desgranges S., Berti R., Patinote C., Langevin D., et al. Perfluorocarbon nanodroplets stabilized by fluorinated surfactants: characterization and potentiality as theranostic agents. J Mater Chem B. 2015;3:2892–2907. doi: 10.1039/c4tb01578a. [DOI] [PubMed] [Google Scholar]
- 28.Singh R., Husseini G.A., Pitt W.G. Phase transitions of nanoemulsions using ultrasound: experimental observations. Ultrason Sonochem. 2012;19:1120–1125. doi: 10.1016/j.ultsonch.2012.02.005. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Tao H.Y., Wang R.Q., Sheng W.J., Zhen Y.S. The development of human serum albumin-based drugs and relevant fusion proteins for cancer therapy. Int J Biol Macromol. 2021;187:24–34. doi: 10.1016/j.ijbiomac.2021.07.080. [DOI] [PubMed] [Google Scholar]
- 30.Maksimova E.A., Nozdriukhin D., Kalva S.K., Lyu S., Lafci B., Augath M.A., et al. Multilayer polymer shell perfluoropentane nanodroplets for multimodal ultrasound, magnetic resonance, and optoacoustic imaging. Laser Photon Rev. 2023;17 [Google Scholar]
- 31.Hu H., Quintana J., Weissleder R., Parangi S., Miller M. Deciphering albumin-directed drug delivery by imaging. Adv Drug Deliv Rev. 2022;185 doi: 10.1016/j.addr.2022.114237. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Pallares R.M., Barmin R.A., Wang A., Kiessling F., Lammers T. Clinical cancer nanomedicines. J Control Release. 2025;385 doi: 10.1016/j.jconrel.2025.113991. [DOI] [PubMed] [Google Scholar]
- 33.Melich R., Bussat P., Morici L., Vivien A., Gaud E., Bettinger T., et al. Microfluidic preparation of various perfluorocarbon nanodroplets: characterization and determination of acoustic droplet vaporization (ADV) threshold. Int J Pharm. 2020;587:119651. doi: 10.1016/j.ijpharm.2020.119651. [DOI] [PubMed] [Google Scholar]
- 34.Airan R.D., Meyer R.A., Ellens N.P., Rhodes K.R., Farahani K., Pomper M.G., et al. Noninvasive targeted transcranial neuromodulation via focused ultrasound gated drug release from nanoemulsions. Nano Lett. 2017;17:652–659. doi: 10.1021/acs.nanolett.6b03517. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Sheeran P.S., Rojas J.D., Puett C., Hjelmquist J., Arena C.B., Dayton P.A. Contrast-enhanced ultrasound imaging and in vivo circulatory kinetics with low-boiling-point nanoscale phase-change perfluorocarbon agents. Ultrasound Med Biol. 2015;41:814–831. doi: 10.1016/j.ultrasmedbio.2014.10.020. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Dauba A., Spitzlei C., Bautista K.J.B., Jourdain L., Selingue E., VanTreeck K.E., et al. Low-boiling-point perfluorocarbon nanodroplets for adaptable ultrasound-induced blood-brain barrier opening. J Control Release. 2024;376:441–456. doi: 10.1016/j.jconrel.2024.10.023. [DOI] [PubMed] [Google Scholar]
- 37.Lusk J., Kim J., Urtz M., Woodhead G., Meuillet E.J., Ue C. Novel phase shift microbubbles, mvt-101, enhance sonothrombolysis in a porcine model of deep vein thrombosis. Nanotheranostics. 2025;9:289–298. doi: 10.7150/ntno.120358. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Yoo K., Walker W.R., Williams R., Tremblay-Darveau C., Burns P.N., Sheeran P.S. Impact of Encapsulation on in vitro and in vivo performance of volatile nanoscale phase-shift perfluorocarbon droplets. Ultrasound Med Biol. 2018;44:1836–1852. doi: 10.1016/j.ultrasmedbio.2018.04.015. [DOI] [PubMed] [Google Scholar]
- 39.Kim J., DeRuiter R.M., Goel L., Xu Z., Jiang X., Dayton P.A. A comparison of sonothrombolysis in aged clots between low-boiling-point phase-change nanodroplets and microbubbles of the same composition. Ultrasound Med Biol. 2020;46:3059–3068. doi: 10.1016/j.ultrasmedbio.2020.07.008. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Goel L., Wu H., Zhang B., Kim J., Dayton P.A., Xu Z., et al. Nanodroplet-mediated catheter-directed sonothrombolysis of retracted blood clots. Microsyst Nanoeng. 2021;7:3. doi: 10.1038/s41378-020-00228-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.Lea-Banks H., Hynynen K. Sub-millimetre precision of drug delivery in the brain from ultrasound-triggered nanodroplets. J Control Release. 2021;338:731–741. doi: 10.1016/j.jconrel.2021.09.014. [DOI] [PubMed] [Google Scholar]
- 42.Lea-Banks H., Wu S.K., Lee H., Hynynen K. Ultrasound-triggered oxygen-loaded nanodroplets enhance and monitor cerebral damage from sonodynamic therapy. Nanotheranostics. 2022;6:376–387. doi: 10.7150/ntno.71946. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43.Ramirez D.G., Ciccaglione M., Upadhyay A.K., Pham V.T., Borden M.A., Benninger R.K.P. Detecting insulitis in type 1 diabetes with ultrasound phase-change contrast agents. Proc Natl Acad Sci U S A. 2021;118 doi: 10.1073/pnas.2022523118. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44.An J., Zhang J., Dong F., Yin J., Feng F., Guo W., et al. Arterial labeling ultrasound subtraction angiography (ALUSA) based on acoustic phase-change nanodroplets. Small. 2022;18 doi: 10.1002/smll.202105989. [DOI] [PubMed] [Google Scholar]
- 45.Melich R., Emmel P., Vivien A., Sechaud F., Mandaroux C., Mhedhbi S., et al. In vitro and in vivo behavioral evaluation of condensed lipid-coated perfluorocarbon nanodroplets. Ultrasound Med Biol. 2024;50:1010–1019. doi: 10.1016/j.ultrasmedbio.2024.03.009. [DOI] [PubMed] [Google Scholar]
- 46.Carlier B., Heymans S.V., Collado-Lara G., Musetta L., Ingram M., Toumia Y., et al. Phase-change ultrasound contrast agents for proton range verification: towards an in vivo application. Phys Med Biol. 2024;69 doi: 10.1088/1361-6560/ad7e76. [DOI] [PubMed] [Google Scholar]
- 47.Burgess M.T., Aliabouzar M., Aguilar C., Fabiilli M.L., Ketterling J.A. Slow-flow ultrasound localization microscopy using recondensation of perfluoropentane nanodroplets. Ultrasound Med Biol. 2022;48:743–759. doi: 10.1016/j.ultrasmedbio.2021.12.007. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 48.Dang J., Zhu M., Dong F., Zhong R., Liu Z., Fang J., et al. Ultrasound-activated nanodroplet disruption of the enterococcus faecalis biofilm in dental root canal. ACS Appl Bio Mater. 2022;5:2135–2142. doi: 10.1021/acsabm.1c01031. [DOI] [PubMed] [Google Scholar]
- 49.VanOsdol J., Ektate K., Ramasamy S., Maples D., Collins W., Malayer J., et al. Sequential HIFU heating and nanobubble encapsulation provide efficient drug penetration from stealth and temperature sensitive liposomes in colon cancer. J Control Release. 2017;247:55–63. doi: 10.1016/j.jconrel.2016.12.033. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 50.Li M., Zhu Y., Yang C., Yang M., Ran H., Zhu Y., et al. Acoustic triggered nanobomb for US imaging guided sonodynamic therapy and activating antitumor immunity. Drug Deliv. 2022;29:2177–2189. doi: 10.1080/10717544.2022.2095058. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 51.Chen J., Feng L., Jin P., Shen J., Lu J., Song Y., et al. Cavitation assisted endoplasmic reticulum targeted sonodynamic droplets to enhanced anti-PD-L1 immunotherapy in pancreatic cancer. J Nanobiotechnol. 2022;20:283. doi: 10.1186/s12951-022-01459-w. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 52.Santiesteban D.Y., Hallam K.A., Yarmoska S.K., Emelianov S.Y. Color-coded perfluorocarbon nanodroplets for multiplexed ultrasound and photoacoustic imaging. Nano Res. 2019;12:741–747. doi: 10.1007/s12274-019-2279-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 53.Martz T.D., Bardin D., Sheeran P.S., Lee A.P., Dayton P.A. Microfluidic generation of acoustically active nanodroplets. Small. 2012;8:1876–1879. doi: 10.1002/smll.201102418. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 54.Kim J., Bautista K.J.B., Deruiter R.M., Goel L., Jiang X., Xu Z., et al. An analysis of sonothrombolysis and cavitation for retracted and unretracted clots using microbubbles versus low-boiling-point nanodroplets. IEEE Trans Ultrason Ferroelectrics Freq Control. 2022;69:711–719. doi: 10.1109/TUFFC.2021.3137125. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 55.Williams R., Wright C., Cherin E., Reznik N., Lee M., Gorelikov I., et al. Characterization of submicron phase-change perfluorocarbon droplets for extravascular ultrasound imaging of cancer. Ultrasound Med Biol. 2013;39:475–489. doi: 10.1016/j.ultrasmedbio.2012.10.004. [DOI] [PubMed] [Google Scholar]
- 56.Xu X., Wang Y., Yan D., Ren C., Cai Y., Liao S., et al. Enhanced Fe(II)-artemisinin-mediated chemodynamic therapy with efficient Fe(III)/Fe(II) conversion circulation for cancer treatment. Biomater Sci. 2024;12:5856–5869. doi: 10.1039/d4bm01095g. [DOI] [PubMed] [Google Scholar]
- 57.Rapoport N., Gao Z., Kennedy A. Multifunctional nanoparticles for combining ultrasonic tumor imaging and targeted chemotherapy. J Natl Cancer Inst. 2007;99:1095–1106. doi: 10.1093/jnci/djm043. [DOI] [PubMed] [Google Scholar]
- 58.Riaz R., Shafiq S., Fatima M., Siddique M.A., Shah S., Abbas S.R. Contrast efficacy of novel phase convertible nanodroplets for safe CEUS imaging. Sci Rep. 2024;14 doi: 10.1038/s41598-024-66163-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 59.Rapoport N.Y., Kennedy A.M., Shea J.E., Scaife C.L., Nam K.H. Controlled and targeted tumor chemotherapy by ultrasound-activated nanoemulsions/microbubbles. J Control Release. 2009;138:268–276. doi: 10.1016/j.jconrel.2009.05.026. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 60.Baghbani F., Chegeni M., Moztarzadeh F., Mohandesi J.A., Mokhtari-Dizaji M. Ultrasonic nanotherapy of breast cancer using novel ultrasound-responsive alginate-shelled perfluorohexane nanodroplets: in vitro and in vivo evaluation. Mater Sci Eng C. 2017;77:698–707. doi: 10.1016/j.msec.2017.02.017. [DOI] [PubMed] [Google Scholar]
- 61.Mohan P., Rapoport N. Doxorubicin as a molecular nanotheranostic agent: effect of doxorubicin encapsulation in micelles or nanoemulsions on the ultrasound-mediated intracellular delivery and nuclear trafficking. Mol Pharm. 2010;7:1959–1973. doi: 10.1021/mp100269f. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 62.Gao X., Zou W., Jiang B., Xu D., Luo Y., Xiong J., et al. Experimental study of retention on the combination of bifidobacterium with high-intensity focused ultrasound (HIFU) synergistic substance in tumor tissues. Sci Rep. 2019;9:6423. doi: 10.1038/s41598-019-42832-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 63.Yin C., Wang G., Zhang Q., Li Z., Dong T., Li Q., et al. Ultrasound nanodroplets loaded with Siglec-G siRNA and Fe3O4 activate macrophages and enhance phagocytosis for immunotherapy of triple-negative breast cancer. J Nanobiotechnol. 2024;22:773. doi: 10.1186/s12951-024-03051-w. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 64.Amir N., Green D., Kent J., Xiang Y., Gorelikov I., Sei M., et al. 18F-Labeled perfluorocarbon droplets for positron emission tomography imaging. Nucl Med Biol. 2017;54:27–33. doi: 10.1016/j.nucmedbio.2017.07.001. [DOI] [PubMed] [Google Scholar]
- 65.Masood U., Riaz R., Shah S.U., Majeed A.I., Abbas S.R. Contrast enhanced sonothrombolysis using streptokinase loaded phase change nano-droplets for potential treatment of deep venous thrombosis. RSC Adv. 2022;12:26665–26672. doi: 10.1039/d2ra04467f. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 66.Guo S., Guo X., Wang X., Zhou D., Du X., Han M., et al. Reduced clot debris size in sonothrombolysis assisted with phase-change nanodroplets. Ultrason Sonochem. 2019;54:183–191. doi: 10.1016/j.ultsonch.2019.02.001. [DOI] [PubMed] [Google Scholar]
- 67.Huang C., Qin Y., Wu S., Yu Q., Mei L., Zhang L., et al. Temperature-responsive “nano-to-micro” transformed polymersomes for enhanced ultrasound/fluorescence dual imaging-guided tumor phototherapy. Nano Lett. 2024;24:9561–9568. doi: 10.1021/acs.nanolett.4c02137. [DOI] [PubMed] [Google Scholar]
- 68.Song X., Feng L., Liang C., Yang K., Liu Z. Ultrasound triggered tumor oxygenation with oxygen-shuttle nanoperfluorocarbon to overcome hypoxia-associated resistance in cancer therapies. Nano Lett. 2016;16:6145–6153. doi: 10.1021/acs.nanolett.6b02365. [DOI] [PubMed] [Google Scholar]
- 69.Shakya G., Cattaneo M., Guerriero G., Prasanna A., Fiorini S., Supponen O. Ultrasound-responsive microbubbles and nanodroplets: a pathway to targeted drug delivery. Adv Drug Deliv Rev. 2024;206 doi: 10.1016/j.addr.2023.115178. [DOI] [PubMed] [Google Scholar]
- 70.Lea-Banks H., O’Reilly M.A., Hynynen K. Ultrasound-responsive droplets for therapy: a review. J Control Release. 2019;293:144–154. doi: 10.1016/j.jconrel.2018.11.028. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 71.Martz T.D., Sheeran P.S., Bardin D., Lee A.P., Dayton P.A. Precision manufacture of phase-change perfluorocarbon droplets using microfluidics. Ultrasound Med Biol. 2011;37:1952–1957. doi: 10.1016/j.ultrasmedbio.2011.08.012. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 72.Bardin D., Martz T.D., Sheeran P.S., Shih R., Dayton P.A., Lee A.P. High-speed, clinical-scale microfluidic generation of stable phase-change droplets for gas embolotherapy. Lab Chip. 2011;11:3990–3998. doi: 10.1039/c1lc20615j. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 73.Sheeran P.S., Luois S., Dayton P.A., Matsunaga T.O. Formulation and acoustic studies of a new phase-shift agent for diagnostic and therapeutic ultrasound. Langmuir. 2011;27:10412–10420. doi: 10.1021/la2013705. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 74.Ferri S., Wu Q., Grazia A.D., Polydorou A., May J.P., Stride E., et al. Tailoring the size of ultrasound responsive lipid-shelled nanodroplets by varying production parameters and environmental conditions. Ultrason Sonochem. 2021;73 doi: 10.1016/j.ultsonch.2021.105482. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 75.Sheeran P.S., Matsuura N., Borden M.A., Williams R., Matsunaga T.O., Burns P.N., et al. Methods of generating submicrometer phase-shift perfluorocarbon droplets for applications in medical ultrasonography. IEEE Trans Ultrason Ferroelectrics Freq Control. 2017;64:252–263. doi: 10.1109/TUFFC.2016.2619685. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 76.Li D.S., Schneewind S., Bruce M., Khaing Z., O’Donnell M., Pozzo L. Spontaneous nucleation of stable perfluorocarbon emulsions for ultrasound contrast agents. Nano Lett. 2019;19:173–181. doi: 10.1021/acs.nanolett.8b03585. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 77.Gao D., Gao J., Xu M., Cao Z., Zhou L., Li Y., et al. Targeted ultrasound-triggered phase transition nanodroplets for Her2-overexpressing breast cancer diagnosis and gene transfection. Mol Pharm. 2017;14:984–998. doi: 10.1021/acs.molpharmaceut.6b00761. [DOI] [PubMed] [Google Scholar]
- 78.Xie H., Li W., Liu H., Chen Y., Ma M., Wang Y., et al. Erythrocyte membrane-coated invisible acoustic-sensitive nanoparticle for inducing tumor thrombotic infarction by precisely damaging tumor vascular endothelium. Small. 2022;18 doi: 10.1002/smll.202201933. [DOI] [PubMed] [Google Scholar]
- 79.Gao M., Liang C., Song X., Chen Q., Jin Q., Wang C., et al. Erythrocyte-membrane-enveloped perfluorocarbon as nanoscale artificial red blood cells to relieve tumor hypoxia and enhance cancer radiotherapy. Adv Mater. 2017;29 doi: 10.1002/adma.201701429. [DOI] [PubMed] [Google Scholar]
- 80.Zhao H., Wu M., Zhu L., Tian Y., Wu M., Li Y., et al. Cell-penetrating peptide-modified targeted drug-loaded phase-transformation lipid nanoparticles combined with low-intensity focused ultrasound for precision theranostics against hepatocellular carcinoma. Theranostics. 2018;8:1892–1910. doi: 10.7150/thno.22386. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 81.Suk J.S., Xu Q., Kim N., Hanes J., Ensign L.M. PEGylation as a strategy for improving nanoparticle-based drug and gene delivery. Adv Drug Deliv Rev. 2016;99:28–51. doi: 10.1016/j.addr.2015.09.012. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 82.Sheeran P.S., Streeter J.E., Mullin L.B., Matsunaga T.O., Dayton P.A. Toward ultrasound molecular imaging with phase-change contrast agents: an in vitro proof of principle. Ultrasound Med Biol. 2013;39:893–902. doi: 10.1016/j.ultrasmedbio.2012.11.017. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 83.Wang Y., Fadhel M.N., Hysi E., Pastenak M., Sathiyamoorthy K., Kolios M.C. In vivo spectroscopic photoacoustic imaging and laser-induced nanoparticle vaporization for anti-HER2 breast cancer. J Biophot. 2021;14 doi: 10.1002/jbio.202100099. [DOI] [PubMed] [Google Scholar]
- 84.Zhang P., Cao Y., Chen H., Zhou B., Hu W., Zhang L. Preparation and evaluation of glycyrrhetinic acid-modified and honokiol-loaded acoustic nanodroplets for targeted tumor imaging and therapy with low-boiling-point phase-change perfluorocarbon. J Mater Chem B. 2017;5:5845–5853. doi: 10.1039/c7tb01215b. [DOI] [PubMed] [Google Scholar]
- 85.Chen Q., Yuan L., Chou W.C., Cheng Y.H., He C., Monteiro-Riviere N.A., et al. Meta-analysis of nanoparticle distribution in tumors and major organs in tumor-bearing mice. ACS Nano. 2023;17:19810–19831. doi: 10.1021/acsnano.3c04037. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 86.Gao B., Xu J., Zhou J., Zhang H., Yang R., Wang H., et al. Multifunctional pathology-mapping theranostic nanoplatforms for US/MR imaging and ultrasound therapy of atherosclerosis. Nanoscale. 2021;13:8623–8638. doi: 10.1039/d1nr01096d. [DOI] [PubMed] [Google Scholar]
- 87.Wang G., Zhang C., Huang Z., Chen J., Chen H., Lin T., et al. Transcytosable and ultrasound-activated liposome enables deep penetration of biofilm for surgical site infection management. Adv Mater. 2025;37 doi: 10.1002/adma.202411092. [DOI] [PubMed] [Google Scholar]
- 88.Mohammed S.A., Amjad M.W., Acosta M.F., Chen X., Lavery L., Hanrahan D., et al. Fibrin-targeted phase shift microbubbles for the treatment of microvascular obstruction. Nanotheranostics. 2024;8:33–47. doi: 10.7150/ntno.85092. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 89.Chandan R., Mehta S., Banerjee R. Ultrasound-responsive carriers for therapeutic applications. ACS Biomater Sci Eng. 2020;6:4731–4747. doi: 10.1021/acsbiomaterials.9b01979. [DOI] [PubMed] [Google Scholar]
- 90.Flaim S.F. Pharmacokinetics and side effects of perfluorocarbon-based blood substitutes. Artif Cells Blood Substit Immobil Biotechnol. 1994;22:1043–1054. doi: 10.3109/10731199409138801. [DOI] [PubMed] [Google Scholar]
- 91.Rapoport N., Gupta R., Kim Y.S., O’Neill B.E. Polymeric micelles and nanoemulsions as tumor-targeted drug carriers: insight through intravital imaging. J Control Release. 2015;206:153–160. doi: 10.1016/j.jconrel.2015.03.010. [DOI] [PubMed] [Google Scholar]
- 92.Abdalkader R., Unga J., Yamashita F., Maruyama K., Hashida M. Evaluation of the theranostic potential of perfluorohexane-based acoustic nanodroplets. Biol Pharm Bull. 2019;42:2038–2044. doi: 10.1248/bpb.b19-00525. [DOI] [PubMed] [Google Scholar]
- 93.Becker K., Ding Z., Bouvain P., Koshy J., Massold T., Kleimann P., et al. Inflammatory stimuli impact on cellular uptake and biodistribution of perfluorocarbon nanoemulsions. J Leukoc Biol. 2025;117 doi: 10.1093/jleuko/qiae199. qiae199. [DOI] [PubMed] [Google Scholar]
- 94.Hoogendijk E., Swider E., Staal A.H.J., White P.B., Riessen N.K.V., Glasser G., et al. Continuous-flow production of perfluorocarbon-loaded polymeric nanoparticles: from the bench to clinic. ACS Appl Mater Interfaces. 2020;12:49335–49345. doi: 10.1021/acsami.0c12020. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 95.Du M., Wang T., Feng R., Zeng P., Chen Z. Establishment of ultrasound-responsive SonoBacteriaBot for targeted drug delivery and controlled release. Front Bioeng Biotechnol. 2023;11 doi: 10.3389/fbioe.2023.1144963. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 96.Lee K.K., Park K.W., Lee S.C., Lee C.S. Perfluorocarbon-polyepinephrine core-shell nanoparticles as a near-infrared light activatable theranostic platform for bimodal imaging-guided photothermal/chemodynamic synergistic cancer therapy. Theranostics. 2025;15:1077–1093. doi: 10.7150/thno.102743. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 97.Spatarelu C.P., Jandhyala S., Luke G.P. Dual-drug loaded ultrasound-responsive nanodroplets for on-demand combination chemotherapy. Ultrasonics. 2023;133 doi: 10.1016/j.ultras.2023.107056. [DOI] [PubMed] [Google Scholar]
- 98.Xu L., Qiu X., Zhang Y., Cao K., Zhao X., Wu J., et al. Liposome encapsulated perfluorohexane enhances radiotherapy in mice without additional oxygen supply. J Transl Med. 2016;14:268. doi: 10.1186/s12967-016-1033-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 99.Xavierselvan M., Cook J., Duong J., Diaz N., Homan K., Mallidi S. Photoacoustic nanodroplets for oxygen enhanced photodynamic therapy of cancer. Photoacoustics. 2022;25 doi: 10.1016/j.pacs.2021.100306. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 100.Tang Y., Shen Q., Lin P., Chen Z., Fan D., Zhuo M., et al. aPD-L1-facilitated theranostic and tumor microenvironment remodeling of pancreatic cancer via docetaxel-loaded phase-transformation nanoparticles triggered by low-intensity pulsed ultrasound. J Nanobiotechnol. 2025;23:48. doi: 10.1186/s12951-025-03105-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 101.Hu D., Zhong L., Wang M., Li H., Qu Y., Liu Q., et al. Perfluorocarbon-loaded and redox-activatable photosensitizing agent with oxygen supply for enhancement of fluorescence/photoacoustic imaging guided tumor photodynamic therapy. Adv Funct Mater. 2019;29 [Google Scholar]
- 102.Liu C.G., Han Y.H., Kankala R.K., Wang S.B., Chen A.Z. Subcellular performance of nanoparticles in cancer therapy. Int J Nanomed. 2020;15:675–704. doi: 10.2147/IJN.S226186. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 103.Zhao F., Wang J., Zhang Y., Hu J., Li C., Liu S., et al. In vivo fate of targeted drug delivery carriers. Int J Nanomed. 2024;19:6895–6929. doi: 10.2147/IJN.S465959. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 104.Spahn D.R. Blood substitutes. Artificial oxygen carriers: perfluorocarbon emulsions. Crit Care. 1999;3:R93–R97. doi: 10.1186/cc364. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 105.Tsuda Y., Yamanouchi K., Yokoyama K., Suyama T., Watanabe M., Ohyanagi H., et al. Discussion and considerations for the excretion mechanism of perfluorochemical emulsion. Biomater Artif Cells Artif Organs. 1988;16:473–483. doi: 10.3109/10731198809132602. [DOI] [PubMed] [Google Scholar]
- 106.Yokoyama K., Yamanouchi K., Murashima R. Excretion of perfluorochemicals after intravenous injection of their emulsion. Chem Pharm Bull (Tokyo) 1975;23:1368–1373. doi: 10.1248/cpb.23.1368. [DOI] [PubMed] [Google Scholar]
- 107.Culp W.C., Onteddu S.S., Brown A., Nalleballe K., Sharma R., Skinner R.D., et al. Dodecafluoropentane emulsion in acute ischemic stroke: a phase Ib/II randomized and controlled dose-escalation trial. J Vasc Intervent Radiol. 2019;30 doi: 10.1016/j.jvir.2019.04.020. 1244–50. e1. [DOI] [PubMed] [Google Scholar]
- 108.Nyankima A.G., Rojas J.D., Cianciolo R., Johnson K.A., Dayton P.A. In vivo assessment of the potential for renal bio-effects from the vaporization of perfluorocarbon phase-change contrast agents. Ultrasound Med Biol. 2018;44:368–376. doi: 10.1016/j.ultrasmedbio.2017.10.016. [DOI] [PubMed] [Google Scholar]
- 109.Asher D.M. Bovine sera used in the manufacture of biologicals: current concerns and policies of the U.S. Food and Drug Administration regarding the transmissible spongiform encephalopathies. Dev Biol Stand. 1999;99:41–44. [PubMed] [Google Scholar]
- 110.Mogues T., Li J., Coburn J., Kuter D.J. IgG antibodies against bovine serum albumin in humans—their prevalence and response to exposure to bovine serum albumin. J Immunol Methods. 2005;300:1–11. doi: 10.1016/j.jim.2005.01.022. [DOI] [PubMed] [Google Scholar]
- 111.Dundas C.M., Demonte D., Park S. Streptavidin-biotin technology: improvements and innovations in chemical and biological applications. Appl Microbiol Biotechnol. 2013;97:9343–9353. doi: 10.1007/s00253-013-5232-z. [DOI] [PubMed] [Google Scholar]
- 112.Fu S., Zhu X., Huang F., Chen X. Anti-PEG antibodies and their biological impact on pegylated drugs: challenges and strategies for optimization. Pharmaceutics. 2025;17:1074. doi: 10.3390/pharmaceutics17081074. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 113.Boissenot T., Bordat A., Fattal E., Tsapis N. Ultrasound-triggered drug delivery for cancer treatment using drug delivery systems: from theoretical considerations to practical applications. J Control Release. 2016;241:144–163. doi: 10.1016/j.jconrel.2016.09.026. [DOI] [PubMed] [Google Scholar]
- 114.Pellow C., Goertz D.E., Zheng G. Breaking free from vascular confinement: status and prospects for submicron ultrasound contrast agents. Wiley Interdiscip Rev Nanomed Nanobiotechnol. 2018;10 doi: 10.1002/wnan.1502. [DOI] [PubMed] [Google Scholar]
- 115.Gorce J.M., Arditi M., Schneider M. Influence of bubble size distribution on the echogenicity of ultrasound contrast agents: a study of SonoVue. Investig Radiol. 2000;35:661–671. doi: 10.1097/00004424-200011000-00003. [DOI] [PubMed] [Google Scholar]
- 116.Schneider M. Characteristics of SonoVuetrade mark. Echocardiography. 1999;16:743–746. doi: 10.1111/j.1540-8175.1999.tb00144.x. [DOI] [PubMed] [Google Scholar]
- 117.McMahon D., Lassus A., Gaud E., Jeannot V., Hynynen K. Microbubble formulation influences inflammatory response to focused ultrasound exposure in the brain. Sci Rep. 2020;10 doi: 10.1038/s41598-020-78657-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 118.Ngo W., Ahmed S., Blackadar C., Bussin B., Ji Q., Mladjenovic S.M., et al. Why nanoparticles prefer liver macrophage cell uptake in vivo. Adv Drug Deliv Rev. 2022;185 doi: 10.1016/j.addr.2022.114238. [DOI] [PubMed] [Google Scholar]
- 119.Cooley M.B., Abenojar E.C., Wegierak D., Sen Gupta A., Kolios M.C., Exner A.A. Characterization of the interaction of nanobubble ultrasound contrast agents with human blood components. Bioact Mater. 2023;19:642–652. doi: 10.1016/j.bioactmat.2022.05.001. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 120.Choi S.Y., Kim Y.S., Seo Y.J., Yang J., Choi K.S. Gas-filled phospholipid nanoparticles conjugated with gadolinium play a role as a potential theragnostics for MR-guided HIFU ablation. PLoS One. 2012;7 doi: 10.1371/journal.pone.0034333. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 121.Papahadjopoulos D., Allen T.M., Gabizon A., Mayhew E., Matthay K., Huang S.K., et al. Sterically stabilized liposomes: improvements in pharmacokinetics and antitumor therapeutic efficacy. Proc Natl Acad Sci U S A. 1991;88:11460–11464. doi: 10.1073/pnas.88.24.11460. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 122.Cao Z., Zhang L., Jiang S. Superhydrophilic zwitterionic polymers stabilize liposomes. Langmuir. 2012;28:11625–11632. doi: 10.1021/la302433a. [DOI] [PubMed] [Google Scholar]
- 123.Lee W.C., Chang C.H., Huang C.M., Wu Y.T., Chen L.C., Ho C.L., et al. Correlation between radioactivity and chemotherapeutics of the 111In-VNB-liposome in pharmacokinetics and biodistribution in rats. Int J Nanomed. 2012;7:683–692. doi: 10.2147/IJN.S28279. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 124.Merian J., De Souza R., Dou Y., Ekdawi S.N., Ravenelle F., Allen C. Development of a liposome formulation for improved biodistribution and tumor accumulation of pentamidine for oncology applications. Int J Pharm. 2015;488:154–164. doi: 10.1016/j.ijpharm.2015.04.060. [DOI] [PubMed] [Google Scholar]
- 125.Lin W., Goldberg R., Klein J. Poly-phosphocholination of liposomes leads to highly-extended retention time in mice joints. J Mater Chem B. 2022;10:2820–2827. doi: 10.1039/d1tb02346b. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 126.Lupi M., Colombo C., Frapolli R., Ferrari R., Sitia L., Dragoni L., et al. A biodistribution study of PEGylated PCL-based nanoparticles in C57BL/6 mice bearing B16/F10 melanoma. Nanotechnology. 2014;25 doi: 10.1088/0957-4484/25/33/335706. [DOI] [PubMed] [Google Scholar]
- 127.Vishnu P., Roy V. Nab-paclitaxel: a novel formulation of taxane for treatment of breast cancer. Wom Health. 2010;6:495–506. doi: 10.2217/whe.10.42. [DOI] [PubMed] [Google Scholar]
- 128.Cheng Y., Cheng H., Jiang C., Qiu X., Wang K., Huan W., et al. Perfluorocarbon nanoparticles enhance reactive oxygen levels and tumour growth inhibition in photodynamic therapy. Nat Commun. 2015;6:8785. doi: 10.1038/ncomms9785. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 129.Yang Q., Li P., Ran H., Wan J., Chen H., Chen H., et al. Polypyrrole-coated phase-change liquid perfluorocarbon nanoparticles for the visualized photothermal-chemotherapy of breast cancer. Acta Biomater. 2019;90:337–349. doi: 10.1016/j.actbio.2019.03.056. [DOI] [PubMed] [Google Scholar]
- 130.Golombek S.K., May J.N., Theek B., Appold L., Drude N., Kiessling F., et al. Tumor targeting via EPR: strategies to enhance patient responses. Adv Drug Deliv Rev. 2018;130:17–38. doi: 10.1016/j.addr.2018.07.007. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 131.Vagena I.-A., Malapani C., Gatou M.-A., Lagopati N., Pavlatou E.A. Enhancement of EPR effect for passive tumor targeting: current status and future perspectives. Appl Sci. 2025;15:3189. [Google Scholar]
- 132.Fang J., Islam W., Maeda H. Exploiting the dynamics of the EPR effect and strategies to improve the therapeutic effects of nanomedicines by using EPR effect enhancers. Adv Drug Deliv Rev. 2020;157:142–160. doi: 10.1016/j.addr.2020.06.005. [DOI] [PubMed] [Google Scholar]
- 133.Cao J., Wang L., Li J., Song M., Zheng Y., He X., et al. Bubble cloud-mediated cavitation for tumor mechanical ablation and effector immune cell deployment. Ultrason Sonochem. 2025;115 doi: 10.1016/j.ultsonch.2025.107296. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 134.Yarmoska S.K., Yoon H., Emelianov S.Y. Lipid shell composition plays a critical role in the stable size reduction of perfluorocarbon nanodroplets. Ultrasound Med Biol. 2019;45:1489–1499. doi: 10.1016/j.ultrasmedbio.2019.02.009. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 135.Huang T., Wu W., Wu J., Tan Y., Zhang M., Long H., et al. Perfluorocarbon nanodrug induced oxygen self-enriching sonodynamic therapy improves cancer immunotherapy after insufficient radiofrequency ablation. Front Immunol. 2023;14 doi: 10.3389/fimmu.2023.1124152. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 136.Rapoport N., Kennedy A.M., Shea J.E., Scaife C.L., Nam K.H. Ultrasonic nanotherapy of pancreatic cancer: lessons from ultrasound imaging. Mol Pharm. 2010;7:22–31. doi: 10.1021/mp900128x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 137.Riemer K., Toulemonde M., Yan J., Lerendegui M., Stride E., Weinberg P.D., et al. Fast and selective super-resolution ultrasound in vivo with acoustically activated nanodroplets. IEEE Trans Med Imag. 2023;42:1056–1067. doi: 10.1109/TMI.2022.3223554. [DOI] [PubMed] [Google Scholar]
- 138.Liu J., Xu F., Huang J., Xu J., Liu Y., Yao Y., et al. Low-intensity focused ultrasound (LIFU)-activated nanodroplets as a theranostic agent for noninvasive cancer molecular imaging and drug delivery. Biomater Sci. 2018;6:2838–2849. doi: 10.1039/c8bm00726h. [DOI] [PubMed] [Google Scholar]
- 139.Chen J., Pal P., Ahrens E.T. Systems engineering approach towards sensitive cellular fluorine-19 MRI. NMR Biomed. 2025;38 doi: 10.1002/nbm.5298. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 140.Kiessling F., Mertens M.E., Grimm J., Lammers T. Nanoparticles for imaging: top or flop?. Radiology. 2014;273:10–28. doi: 10.1148/radiol.14131520. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 141.Jiang N., Wang Z., Deng Q., Zhou Y., Cao S., Zhou Q., et al. Low-intensity focused ultrasound guided dodecafluoropentane-loaded acoustic phase-change nanoparticles for treatment of porcine coronary microthromboembolism. Int J Cardiol. 2023;371:1–9. doi: 10.1016/j.ijcard.2022.09.078. [DOI] [PubMed] [Google Scholar]
- 142.Singh A., Nyankima A.G., Anthony Phipps M., Chaplin V., Dayton P.A., Caskey C. Improving the heating efficiency of high intensity focused ultrasound ablation through the use of phase change nanodroplets and multifocus sonication. Phys Med Biol. 2020;65 doi: 10.1088/1361-6560/ab9559. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 143.Glickstein B., Levron M., Shitrit S., Aronovich R., Feng Y., Ilovitsh T. Nanodroplet-mediated low-energy mechanical ultrasound surgery. Ultrasound Med Biol. 2022;48:1229–1239. doi: 10.1016/j.ultrasmedbio.2022.02.018. [DOI] [PubMed] [Google Scholar]
- 144.Zhang L., Yu Y., Ding K., Ji C., Zhang D., Liang P., et al. Tumor microenvironment ameliorative and adaptive nanoparticles with photothermal-to-photodynamic switch for cancer phototherapy. Biomaterials. 2025;313 doi: 10.1016/j.biomaterials.2024.122771. [DOI] [PubMed] [Google Scholar]
- 145.Chang L.W., Hou M.L., Hung S.H., Lin L.C., Tsai T.H. Pharmacokinetics of quercetin-loaded nanodroplets with ultrasound activation and their use for bioimaging. Int J Nanomed. 2015;10:3031–3042. doi: 10.2147/IJN.S78983. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 146.Ho Y.J., Yeh C.K. Concurrent anti-vascular therapy and chemotherapy in solid tumors using drug-loaded acoustic nanodroplet vaporization. Acta Biomater. 2017;49:472–485. doi: 10.1016/j.actbio.2016.11.018. [DOI] [PubMed] [Google Scholar]
- 147.Sheng D., Deng L., Li P., Wang Z., Zhang Q. Perfluorocarbon nanodroplets with deep tumor penetration and controlled drug delivery for ultrasound/fluorescence imaging guided breast cancer therapy. ACS Biomater Sci Eng. 2021;7:605–616. doi: 10.1021/acsbiomaterials.0c01333. [DOI] [PubMed] [Google Scholar]
- 148.Dong F., An J., Guo W., Dang J., Huang S., Feng F., et al. Programmable ultrasound imaging guided theranostic nanodroplet destruction for precision therapy of breast cancer. Ultrason Sonochem. 2024;105 doi: 10.1016/j.ultsonch.2024.106854. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 149.Guo H., Xu M., Cao Z., Li W., Chen L., Xie X., et al. Ultrasound-assisted miR-122–loaded polymeric nanodroplets for hepatocellular carcinoma gene therapy. Mol Pharm. 2020;17:541–553. doi: 10.1021/acs.molpharmaceut.9b00983. [DOI] [PubMed] [Google Scholar]
- 150.Lea-Banks H., Meng Y., Wu S.K., Belhadjhamida R., Hamani C., Hynynen K. Ultrasound-sensitive nanodroplets achieve targeted neuromodulation. J Control Release. 2021;332:30–39. doi: 10.1016/j.jconrel.2021.02.010. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 151.Uruen C., Chopo-Escuin G., Tommassen J., Mainar-Jaime R.C., Arenas J. Biofilms as promoters of bacterial antibiotic resistance and tolerance. Antibiotics (Basel) 2020;10:3. doi: 10.3390/antibiotics10010003. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 152.Durham P.G., Sidders A.E., Beam J.E., Kedziora K.M., Dayton P.A., Conlon B.P., et al. Harnessing ultrasound-stimulated phase change contrast agents to improve antibiotic efficacy against methicillin-resistant Staphylococcus aureus biofilms. Biofilm. 2021;3 doi: 10.1016/j.bioflm.2021.100049. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 153.Choi V., Carugo D., Stride E. Repurposing antimicrobials with ultrasound-triggered nanoscale systems for targeted biofilm drug delivery. NPJ Antimicrob Resist. 2025;3:22. doi: 10.1038/s44259-025-00086-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 154.PatSnap. MVT-101 Synapse Drug database. https://synapse.patsnap.com/drug/0a68901fb43ace6708c29abf6f2961ef Published 02 Apr, 2022. Updated 08 May 2025. Accessed 30 Oct 2025. Available from:
- 155.MVT Pharma, Transforming ultrasound from imaging to cure. MVT Pharma Website. Accessed 30 October 2025. Available from: https://www.mvtpharma.com.
- 156.NuvOx LLC The effects of NVX-108 as a radiation sensitizer in glioblastoma. ClinicalTrials.gov. https://clinicaltrials.gov/study/NCT02189109 Published 14 Jul, 2019. Updated 28 Feb, 2019. Accessed 24 Nov 2025. Available from:
- 157.NuvOx LLC NanO2 for large vessel occlusion stroke. ClinicalTrials.gov. https://clinicaltrials.gov/study/NCT05389748?term=nuvox&rank=1 Published 25 May, 2022. Updated 25 May, 2022. Accessed 24 Nov 2025. Available from: