Abstract
Purpose
This study aimed to assess the potential enhancement of image quality in fast-spin-echo T2-weighted images (FSE-T2WI) with respiratory-gating compared to conventional FSE-T2WI and to clarify how abdominal wall motion affects uterine displacement and consequently impacts image quality.
Methods
One hundred and three women who underwent pelvic MRI using a 3T-MRI scanner were enrolled. FSE-T2WI with and without respiratory-gating was visually assessed for motion-related artifacts, and the image quality was categorized as poor, moderate, or excellent. The uterus was classified as type 1 (on the bladder), type 2 (not on the bladder and not retroflexed), or type 3 (uterus contacting the vertebrae or a retroflexed uterus). The relationships (1) between the abdominal wall and uterine displacements, (2) between the quality of FSE-T2WI with and without respiratory-gating and uterine displacement, and (3) between uterine displacement and types were investigated.
Results
Respiratory-gated FSE-T2WI achieved better overall image quality, with poor, moderate, and excellent ratings observed in 27, 44, and 32 patients, respectively, compared with 53, 34, and 16 patients for conventional FSE-T2WI. Improvement with respiratory-gating was observed in 43 patients, particularly in those with type 1 and type 2 uteri, whereas only 8 patients showed decreased image quality. A strong correlation was found between abdominal wall and uterine displacements (P < 0.001), indicating that uterine motion is largely driven by respiration. The amplitude of uterine displacement was significantly smaller in type 3 uteri (P < 0.05), consistent with fewer motion-related artifacts.
Conclusion
Respiratory-gated FSE-T2WI effectively reduces motion artifacts caused by respiratory-induced uterine displacement, particularly in type 1 and type 2 uteri where displacement is pronounced.
Keywords: female pelvis, magnetic resonance imaging, motion artifacts, respiratory gating, T2-weighted image
Introduction
In the field of gynecology, imaging diagnosis is valuable for localizing diseases and evaluating treatment efficacy.1–3 Fast spin echo-T2-weighted image (FSE-T2WI) is particularly important in imaging diagnosis because of its high contrast resolution.4,5 However, the uterus, surrounded by organs, such as the abdominal wall, bladder, and bowel, undergoes displacement over time, leading to motion-related artifacts,6,7 and clinical challenges may arise when motion-related artifacts complicate imaging diagnosis.
Previous studies have reported the usefulness of sequences that are resilient to body displacements and those acquired in shorter imaging durations compared to physiological displacements in mitigating motion artifacts.8–11 However, tradeoffs exist, such as the appearance of streak artifacts and a decrease in the SNR and contrast.8–11 Therefore, motion artifact countermeasures using these sequences have not yet provided a fundamental solution to image quality degradation. Moreover, it is unclear to what extent these sequences can effectively counteract uterine displacement.
Breathing-related uterine displacement is presumed to be the underlying mechanism of motion-related artifacts in female pelvic MRI. Previous studies have noted uterine displacement primarily in the head–foot direction rather than in the anteroposterior or left-right directions.12,13 In abdominal and cardiac MRI, motion-compensation techniques such as respiratory or electrocardiogram gating have been proven effective in minimizing motion artifacts. We therefore hypothesized that aligning interventions with the uterine displacement mechanism would reduce the incidence of non-diagnostic cases due to motion-related artifacts on FSE-T2WI of the female pelvis.
However, to date, there is a lack of literature regarding the efficacy of FSE-T2WI with respiratory gating (RESP-T2WI) for female pelvic MRI. Accordingly, this study aimed to determine whether RESP-T2WI enhances image quality compared to conventional FSE-T2WI (C-T2WI) by quantitatively evaluating uterine motion and its correlation with abdominal wall displacement across different uterine positions.
Materials and Methods
Ethics approval
This study was approved by the Research Ethics Committee of Fukushima Medical University (approval number: REC2024-131). All procedures involving human participants followed the ethical standards of the institutional and national research committees and the 1964 Declaration of Helsinki and its subsequent amendments or comparable ethical standards. The need for written informed consent was waived due to the retrospective nature of the study and data anonymization.
Study design and participants
This retrospective study involved patients who had female pelvic MRI examination between June 2023 and November 2024. Patients under 18 years of age, with a previous total hysterectomy, without records of T2WI with respiratory gating and cine images, and with inappropriate cine images were excluded (Fig. 1). The clinical diagnoses are summarized in Table 1.
Fig. 1.

Overview of the patient selection process.HASTE, half-Fourier-acquired single-shot turbo spin-echo; T2WI, T2-weighted images.
Table 1.
Clinical diagnoses in the study cohort
| Clinical diagnosis | Cases |
|---|---|
| Uterine fibroids | 41 (39.8%) |
| Benign ovarian lesion | 13 (12.6%) |
| Uterine cancer | 8 (7.8%) |
| Endometriosis | 4 (3.9%) |
| Adenomyosis | 2 (1.9%) |
| Ovarian cancer | 2 (1.9%) |
| Malignant tumor screening | 2 (1.9%) |
| Others | 31 (30.1%) |
MRI examination
All MRI examinations were performed using a 3T-MRI scanner (Skyra; Siemens, Erlangen, Germany) equipped with an 18-channel phased-array body coil. This study utilized C-T2WI, RESP-T2WI, and half-Fourier-acquired single-shot turbo spin-echo (HASTE) static images of female pelvis acquired during routine pelvic MRI scans. In addition, HASTE-cine images, comprising 32 frames captured at 1 fps in the sagittal plane, were acquired before FSE-T2WI and focused on the uterus to determine uterine body displacement. These cine images were used to measure the uterine and abdominal wall displacements. No antispasmodic medications, dietary preparations, or urinary preparations were administered prior to imaging. The imaging parameters for the C-T2WI, RESP-T2WI, HASTE-cine, and HASTE-static images are listed in Table 2.
Table 2.
Imaging parameters
| Pulse sequence | C-T2WI | RESP-T2WI | HASTE-cine | HASTE-static |
|---|---|---|---|---|
| Repetition time (ms) | 4000 | About 2213–7478 (3979 ± 1040) | 1000 | 3000 |
| Echo time (ms) | 99 | 99 | 76 | 96 |
| FOV (mm) | 260 | 260 | 256 | 260 |
| Matrix size | 384 × 307 | 320 × 256 | 256 × 256 | 320 × 256 |
| Phase encoding direction | Head–foot | Head–foot | Anteroposterior | Anteroposterior |
| Average | 2 | 2 | 1 | 1 |
| Slice thickness (mm) | 5 | 5 | 10 | 5 |
| Slice gap (mm) | 1 | 1 | — | 1 |
| Refocusing flip angle (°) | 160 | 160 | 150 | 160 |
| Echo train length | 14 | 15 | — | — |
| Parallel imaging factor | 2 | 3 | 2 | 3 |
| Phase partial Fourier | — | — | 5/8 | 5/8 |
| Imaging plane | Sagittal | Sagittal | Sagittal | Sagittal |
| Scan time (s) | 148 | About 168–471 (312 ± 62) | 32 | 79 |
C-T2WI, conventional T2-weighted images; HASTE, half-Fourier-acquired single-shot turbo spin echo; RESP-T2WI, respiratory-gated T2-weighted images.
Classification of uterine position
The uterine position was classified into 3 types based on FSE-T2WI, referring to the work of Tusboyama et al., as follows: type 1, uterus on the bladder; type 2, uterus not on the bladder and without retroflexion; and type 3, uterus contacting the vertebrae or a retroflexed uterus (Fig. 2).7
Fig. 2.

Example of different quality images for different uterine types.
Visual assessment
Sagittal C-T2WI and RESP-T2WI were individually assessed by 2 experienced radiologists with 13 and 9 years of experience, respectively. To ensure consistency, the evaluation criteria were standardized before assessment, and HASTE-static images were used as reference images for identifying and grading motion-related artifacts. Visual assessment focuses on artifacts such as ghosting and motion blurring. Ghosting artifacts were rated on a 4-point scale according to the method outlined by Ishikawa et al.12: 0 = no apparent ghosts, 1 = visible but non-disturbing ghosts, 2 = visible and slightly disturbing ghosts, and 3 = ghosts interfering with the diagnostic interpretation. Motion blur was assessed on a 4-point scale: 0, sharply defined organ margins with clear internal structures, 1, minimal motion blur of margins and/or internal structures for diagnosis, 2, moderate motion blur that could affect diagnosis; and 3, marked motion blur (non-diagnostic). The overall image quality of each FSE-T2WI was categorized as excellent, moderate, or poor, based on the combined scores of ghosting artifacts and motion blur: excellent (0 points for both parameters), poor (3 points for either parameter), and moderate (1 or 2 points for either parameter) (Fig. 3). Discrepancies between the evaluators were resolved by discussion to reach a consensus. Subsequently, the rates of concordant and discordant visual assessments between C-T2WI and RESP-T2WI were calculated. Cases exhibiting improved or deteriorated image quality on RESP-T2WI relative to C-T2WI were identified and further analyzed. In addition, the association between visual assessment results and uterine position type was investigated.
Fig. 3.

Example of different quality images for T2WI of the female pelvis. From left to right, the images are of poor, moderate, and excellent quality.T2WI, T2-weighted images.
Displacement and amplitude
Dataset
The initial frame of the HASTE-cine image dataset was omitted from the analysis because of its non-steady-state longitudinal magnetization, leading to a distinct signal intensity compared with subsequent frames. The initial frame was referred to as a discarded frame. The first frame in the dataset, excluding the discarded frame, was defined as the reference image.
Measurement of displacement of the uterus and abdominal wall
The displacements of the uterus (head–foot direction) and abdominal wall (anteroposterior direction) were calculated pixel-by-pixel from the reference image in the data set onwards (Fig. 4). Uterine displacement was determined using the “imregdemons” function in MATLAB 2022b (MathWorks, Natick, MA, USA). This function uses the Demons algorithm for non-rigid body registration to generate a displacement vector deforming one image based on density differences and gradients from the displacement of 2 images.14,15 Subsequent images in the dataset were registered to the initial image, with ROIs for uterine and abdominal wall displacements defined on the first image (Fig. 5). The ROI for the abdominal wall was defined over the height range from the pubic bone to the uterus. These ROIs were then overlaid onto the displacement maps, and the displacements of the uterus and abdominal wall were quantified over a 30 second period.
Fig. 4.

Calculation of displacement maps. The top figure shows an explanation of registration using HASTE-cine images. The HASTE images from the third one onward represent displacement relative to the second image (the fixed image), which serves as the reference. The bottom figures (a) and (b) show the displacement maps in the anteroposterior and head–foot directions, respectively. The pixel values of these maps indicate the displacement (mm). In both maps, black indicates displacement from the posterior to the anterior direction in (a) and from the foot to the head direction in (b). White indicates displacement from anterior to posterior in (a) and from head to foot in (b). The gray regions represent zero displacement (see gray scale bar).HASTE, half-Fourier-acquired single-shot turbo spin-echo.
Fig. 5.

ROIs setting. ROIs in the abdominal wall and uterine body. The abdominal wall and uterus of the HASTE-cine were manually traced.HASTE, half-Fourier-acquired single-shot turbo spin-echo.
Measurement of the amplitude of uterine displacement
Positive and negative peaks were identified from the time-series data of the uterine displacement. The amplitude of uterine displacement was determined as the difference between the adjacent positive and negative peaks.12 Previous studies have established the primary direction of uterine body displacement as head–foot;12,13 therefore, the maximum amplitude of uterine displacement was calculated in this direction. The maximum amplitude signifies the maximum range of uterine displacement. Peaks were detected using the “findpeaks” function in MATLAB 2022b. Subsequently, the maximum amplitude values, classified based on visual assessment of C-T2WI and RESP-T2WI, were plotted for each group.
Statistical analysis
Unless stated otherwise, values are presented as mean ± standard deviation. First, to assess the reliability of visual assessments between the 2 radiologists, the inter-observer agreement (κ) in the visual assessment of the image quality of FSE-T2WI was calculated. The agreement rates were categorized as follows: moderate (0.41–0.60), substantial (0.61–0.80), and near-perfect agreement (≥ 0.81).16 Before performing group or correlation analyses, data distribution and variance characteristics were examined to select appropriate statistical tests. To investigate whether the motion of the uterus corresponded with that of the abdominal wall, Spearman’s rank correlation coefficient was computed. Absolute values were used to assess the strength of the correlation. The associations between the amplitude of uterine displacement and both image quality and uterine type were evaluated using the Kruskal–Wallis test with Steel–Dwass post-hoc analysis. All statistical analyses were performed using R statistical analysis software for Windows version 4.3.1 (R: Language and environment for statistical computing; R Foundation for Statistical Computing, Vienna, Austria).17 Significance was defined as P < 0.05.
Results
Participants
The analysis included 103 women aged 20–84 years (mean: 45.8 ± 13.1 years).
Classification of uterine position
The uterine position was classified as type 1, type 2, and type 3 in 75 (72.8%), 19 (18.4%), and 9 (8.7%) patients, respectively.
Visual assessment
Image quality of C-T2WI was categorized as poor, moderate, and excellent in 53 (51.5%; type 1 = 45, type 2 = 7, and type 3 = 1 cases), 34 (33.0%; type 1 = 24, type 2 = 9, and type 3 = 1 cases), and 16 (15.5%; type 1 = 6, type 2 = 3, and type 3 = 7 cases) patients, respectively. Conversely, image quality of RESP-T2WI was categorized as poor, moderate, and excellent in 27 (26.2%; type 1 = 22, type 2 = 4, and type 3 = 1 cases), 44 (42.7%; type 1 = 37, type 2 = 5, and type 3 = 2 cases), and 32 (31.1%; type 1 = 16, type 2 = 10, and type 3 = 6 cases) patients, respectively (Fig. 6). The inter-observer agreement κ was 0.63 (P < 0.001). Discrepancies between the 2 radiologists were observed in 81 cases (39.3%). The concordant and discordant results between C-T2WI and RESP-T2WI were 52 (50.5%) and 51 (49.5%), respectively. Cases with higher and lower visual assessment of RESP-T2WI compared to C-T2WI were 43 (41.7%; type 1 = 31, type 2 = 10, and type 3 = 2 cases) and 8 (7.8%; type 1 = 4, type 2 = 2, and type 3 = 2 cases) cases, respectively (Table 3).
Fig. 6.

Comparison of image quality of C-T2WI and RESP-T2WI. The figure illustrates the number of cases for each image quality group in C-T2WI and RESP-T2WI. Yellow bars indicate C-T2WI, while blue bars indicate RESP-T2WI.C-T2WI, conventional T2-weighted images; RESP-T2WI, T2-weighted images with respiratory-gating.
Table 3.
Results of the visual assessment
| RESP-T2WI | |||||
|---|---|---|---|---|---|
| Poor | Moderate | Excellent | Total | ||
| C-T2WI | Poor | 23 | 22 | 8 | 53 |
| Moderate | 3 | 18 | 13 | 34 | |
| Excellent | 1 | 4 | 11 | 16 | |
| Total | 27 | 44 | 32 | 103 | |
C-T2WI, conventional T2 weighted images; RESP-T2WI, respiratory-gated T2 weighted images.
Correlation between abdominal wall displacement and uterine displacement
The correlation coefficient between the abdominal wall and uterine displacement was 0.73 ± 0.22 (P < 0.05). The abdominal wall displacements for type 1, type 2, and type 3 uteri were 2.75 ± 2.21, 3.51 ± 2.26, and 1.92 ± 0.67 mm, respectively, and uterine displacements of each type were 1.41 ± 0.96, 0.96 ± 0.62, and 0.46 ± 0.24 mm, respectively. Nine cases showed a weak correlation (range: 0.02–0.32) between abdominal wall and uterine displacements, without a significant difference (range: 0.08–0.89) in correlation strength among cases. Four (3 of type 1 and 1 of type 2) of these 9 cases did not show an improvement in the quality of RESP-T2WI, and 1 of the 4 cases displayed degradation of the visual assessment from moderate to poor. The remaining 5 cases (3 cases from type 1 and 2 cases from type 2) showed improvement in the quality of RESP-T2WI (2 cases, poor to excellent; 2 cases, moderate to excellent; 1 case, poor to moderate). In addition, 7 of 8 cases degraded image quality in RESP-T2WI showed strong correlation (range: 0.56–0.93) between abdominal wall and uterine displacements, with a significant difference (all P < 0.01).
Relationship between amplitude of uterine displacement and quality of FSE-T2WI
Figure 7 shows the correlation between the maximum amplitude of uterine displacement and the quality of FSE-T2WIs. In the C-T2WI, the maximum amplitudes of uterine displacement for the poor, moderate, and excellent image quality groups were 2.35 ± 1.35, 1.61 ± 0.79, and 1.34 ± 0.61 mm, respectively (P < 0.001). Significantly different values were observed between the poor and moderate image quality groups (P = 0.022), as well as between the poor and excellent image quality groups (P < 0.001). No significant differences were observed between the moderate and excellent groups (P = 0.050).
Fig. 7.

Relationship between quality of T2WI and maximum amplitude of uterus. (a) and (b) Relationship between the maximum amplitude of uterine displacement and quality of C-T2WI and RESP-T2WI, respectively. Figure (b) shows an increase in the maximum amplitudes of uterine displacement in the moderate and excellent groups compared with those of C-T2WI (*P < 0.05, **P < 0.01). C-T2WI, conventional T2-weighted images; n.s, not significant; RESP-T2WI, T2-weighted images with respiratory-gating; T2WI, T2-weighted images.
Similarly, in RESP-T2WI, the maximum amplitudes of uterine displacement for the poor, moderate, and excellent image quality groups were 2.24 ± 1.56, 2.18 ± 1.33, and 1.48 ± 0.74 mm, respectively (P = 0.084). No significant differences were observed between the groups (P = 0.084).
Relationship between amplitude of uterine displacement and uterine types
Figure 8 presents the relationship between the maximum amplitude of uterine displacement and the uterine type. The maximum amplitudes of the uterine displacement for type 1, type 2, and type 3 were 2.21 ± 1.38, 1.61 ± 0.77, and 0.86 ± 0.44 mm, respectively (P < 0.001). Significantly different values were observed between type 1 and type 3 (P < 0.001) and type 2 and type 3 (P = 0.017). No significant difference was observed between type 1 and type 2.
Fig. 8.

Relationship between amplitude of uterine displacement and uterine types. The maximum amplitudes of uterine displacement for type 1 were larger (P < 0.001), while those for type 3 were smallest (P < 0.05).
Discussion
A strong correlation was observed between abdominal wall and uterine displacements, indicating that uterine displacement was affected by respiration. Consequently, respiratory gating was particularly effective in patients with type 1 and type 2 uteri. The results of this study showed that the quality of RESP-T2WI was superior to that of C-T2WI of the female pelvis, because the incidence of poor image quality group for RESP-T2WI was half that of C-T2WI, whereas the incidence of excellent image quality group for RESP-T2WI was double.
The correlation coefficient between abdominal wall displacement (anteroposterior direction) and uterine displacement (head–foot direction) was high at 0.73 ± 0.22. This correlation suggested that uterine displacement in the head–foot direction is caused by respiration. During expiration, the abdominal wall is displaced from the anterior to the posterior direction, while the diaphragm is displaced from the foot to the head direction simultaneously. This series of displacements can be transmitted to the bladder and to type 1 and type 2 uteri via bowel and adipose tissue that are located superiorly (Supplementary Movies 1 and 2). Consequently, the uterus would be displaced in accordance with respiration. This finding supports the utility of RESP-T2WI.
The maximum amplitude of uterine displacement was the highest in the poor image quality group of C-T2WI. Our results are consistent with our previous study.12 Moreover, the maximum amplitudes of type 1 and 2 uteri were significantly higher than those of type 3 uteri. Therefore, in cases of type 1 and type 2 uteri, motion-related artifacts would be prone to occur. In type 1 cases, RESP-T2WI is particularly useful for reducing motion-related artifacts. This is because the type 1 uterus may be sensitive to abdominal wall displacement due to respiration because the uterus is close to the abdominal wall. In type 2 uteri, respiratory gating was also effective, despite the uterus being far away from the abdominal wall and not located on the bladder. This is because the displacement of the intestine or adipose tissue surrounding the uterus, in accordance with respiration, may be transmitted to the uterus. In contrast, respiratory gating may be less effective in type 3 uteri because the uterus is located farther from the abdominal wall and its displacement amplitude tends to be smaller, which inherently result in fewer motion-related artifacts. This finding is consistent with a previous report by Tsuboyama, et al., which showed that the uterine type in contact with the vertebrae was not a risk factor for non-diagnostic quality.7
Some cases showed no noticeable improvement across all uterine types, suggesting that the effect of respiratory gating may vary among individuals. Cases with lower visual assessment in RESP-T2WI compared to C-T2WI may have been affected by factors such as sudden rectal displacement, gating mismatch, or unstable breathing during imaging. In these instances, unstable breathing can lead to inconsistent scan times and degraded image quality. Similarly, a lower correlation between abdominal wall and uterine displacements, potentially resulting from bowel displacement or bladder dilation, may diminish the effectiveness of respiratory gating. In such cases, abdominal wall movement no longer serves as a reliable surrogate for uterine motion, leading to the observed image degradation.
Regarding the reliability of image quality evaluation, our inter-observer agreement κ was equal to or higher than those reported in prior studies. Previous studies have compared the quality of uterine FSE-T2WI with those of BLADE in Siemens MRI scanners, which is the same as periodically rotated overlapping parallel lines with enhanced reconstruction (PROPELLER),18 and HASTE, with inter-observer agreement κ values for artifacts of 0.65 and 0.51, respectively.8,11 Therefore, the visual assessment in the present study can be considered justified.
We anticipate that employing RESP-T2WI could decrease the need for re-imaging and enhance image quality, despite its longer scan time (approximately 5 mins 11s on average). In the future, the integration of deep learning reconstruction into RESP-T2WI will reduce imaging time. Other measures of motion-related artifacts include sequences that are resilient to body displacements, such as PROPELLER18 and HASTE. Although the time extension is not as long as that of respiratory gating, there are some limitations, such as streak artifacts or a lower SNR.8,11
This study had several limitations. First, uterine and abdominal wall displacements during HASTE-cine may vary from those during FSE-T2WI due to differences in bowel peristalsis and bladder dilation. Second, this study did not consider age or disease status, which could have potentially affected the results. Third, the imaging time for RESP-T2WI is generally longer than for C-T2WI and can be further prolonged depending on the patient’s respiratory cycle. However, a single respiratory-gated imaging would be preferable to re-imaging of C-T2WI in terms of total examination time and T2WI quality, as the latter does not reliably improve image quality.
Despite these limitations, our results show that RESP-T2WI of the female pelvis can effectively reduce the number of non-diagnostic cases.
Conclusion
This study demonstrated the effectiveness of RESP-T2WI in maintaining image quality of the female pelvis, even with substantial uterine displacement. While C-T2WI showed a significant decrease in image quality with increasing uterine displacement amplitude, RESP-T2WI was unaffected by displacement magnitude. Consequently, in clinical settings, RESP-T2WI can be considered a viable approach for decreasing the number of non-diagnostic cases.
Footnotes
Conflicts of Interest: The authors declare that they have no conflict of interest.
Supplementary Materials
An example of cine images of type 1 uteri. The movie shows that type 1 uterus displaces in the head–foot direction in coordination with the superiorly located bowel and adipose tissue.
An example of cine images of type 2 uteri. The movie shows that type 2 uterus displaces in the head–foot direction in coordination with the superiorly located bowel and adipose tissue.
References
- 1.Kinkel K, Forstner R, Danza FM, et al. Staging of endometrial cancer with MRI: Guidelines of the European Society of Urogenital Imaging. Eur Radiol 2009; 19:1565–1574. [DOI] [PubMed] [Google Scholar]
- 2.Frei KA, Kinkel K. Staging endometrial cancer: Role of magnetic resonance imaging. J Magn Reson Imaging 2001; 13:850–855. [DOI] [PubMed] [Google Scholar]
- 3.Nougaret S, Horta M, Sala E, et al. Endometrial cancer MRI staging: Updated Guidelines of the European Society of Urogenital Radiology. Eur Radiol 2019; 29:792–805. [DOI] [PubMed] [Google Scholar]
- 4.Togashi K, Nakai A, Sugimura K. Anatomy and physiology of the female pelvis: MR imaging revisited. J Magn Reson Imaging 2001; 13:842–849. [DOI] [PubMed] [Google Scholar]
- 5.Hricak H. MRI of the female pelvis: A review. AJR Am J Roentgenol 1986; 146:1115–1122. [DOI] [PubMed] [Google Scholar]
- 6.Zand KR, Reinhold C, Haider MA, Nakai A, Rohoman L, Maheshwari S. Artifacts and pitfalls in MR imaging of the pelvis. J Magn Reson Imaging 2007; 26:480–497. [DOI] [PubMed] [Google Scholar]
- 7.Tsuboyama T, Takei O, Okada A, Wada K, Kuriyama K. Effect of uterine position and intrapelvic motions on the image quality of 3D T2-weighted MRI of the uterus: Can short prescans predict the non-diagnostic image quality? Eur J Radiol 2020; 130:109186. [DOI] [PubMed] [Google Scholar]
- 8.Fujimoto K, Koyama T, Tamai K, Morisawa N, Okada T, Togashi K. BLADE acquisition method improves T2-weighted MR images of the female pelvis compared with a standard fast spin-echo sequence. Eur J Radiol 2011; 80:796–801. [DOI] [PubMed] [Google Scholar]
- 9.Lane BF, Vandermeer FQ, Oz RC, Irwin EW, McMillan AB, Wong-You-Cheong JJ. Comparison of sagittal T2-weighted BLADE and fast spin-echo MRI of the female pelvis for motion artifact and lesion detection. AJR Am J Roentgenol 2011; 197:W307–13. [DOI] [PubMed] [Google Scholar]
- 10.Froehlich JM, Metens T, Chilla B, Hauser N, Klarhoefer M, Kubik-Huch RA. Should less motion sensitive T2-weighted BLADE TSE replace Cartesian TSE for female pelvic MRI? Insights Imaging 2012; 3:611–618. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Kataoka M, Kido A, Koyama T, et al. MRI of the female pelvis at 3T compared to 1.5T: Evaluation on high-resolution T2-weighted and HASTE images. J Magn Reson Imaging 2007; 25:527–534. [DOI] [PubMed] [Google Scholar]
- 12.Ishikawa H, Monma M, Ishimori Y, et al. Impact of uterine displacement on T2-weighted image quality in the female pelvic MRI. Radiol Phys Technol 2025; 18:127–135. [DOI] [PubMed] [Google Scholar]
- 13.Martin T, Janzen C, Li X, et al. Characterization of uterine motion in early gestation using MRI-based motion tracking. Diagnostics (Basel) 2020; 10:840. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Thirion JP. Image matching as a diffusion process: An analogy with Maxwell’s demons. Med Image Anal 1998; 2:243–260. [DOI] [PubMed] [Google Scholar]
- 15.Vercauteren T, Pennec X, Perchant A, Ayache N. Diffeomorphic demons: Efficient non-parametric image registration. Neuroimage 2009; 45:S61–S72. [DOI] [PubMed] [Google Scholar]
- 16.Landis JR, Koch GG. The measurement of observer agreement for categorical data. Biometrics 1977; 33:159–174. [PubMed] [Google Scholar]
- 17.Bates D, Mächler M, Bolker B, Walker S. Fitting linear mixed-effects models using lme4. J Stat Softw 2015; 67:1–48. [Google Scholar]
- 18.Pipe JG. Motion correction with PROPELLER MRI: Application to head motion and free-breathing cardiac imaging. Magn Reson Med 1999; 42:963–969. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
An example of cine images of type 1 uteri. The movie shows that type 1 uterus displaces in the head–foot direction in coordination with the superiorly located bowel and adipose tissue.
An example of cine images of type 2 uteri. The movie shows that type 2 uterus displaces in the head–foot direction in coordination with the superiorly located bowel and adipose tissue.