Abstract
Introduction
Inferior vena cava (IVC) filters are widely used in the management of pulmonary embolism (PE), although they carry risks of significant mechanical complications.
Case presentation
We report the case of a patient with deep vein thrombosis and PE who developed an unexplained decrease in hemoglobin level following Tempofilter II filter placement and thrombolysis. Computed tomography angiography and digital subtraction angiography confirmed an abdominal aortic pseudoaneurysm caused by a penetrating filter strut. The filter was successfully retrieved, and the pseudoaneurysm was excluded via covered stent implantation, with favorable clinical outcomes.
Discussion
Although rare, IVC filters can cause aortic pseudoaneurysm, a potentially life-threatening complication. Risk factors include advanced age, aortic tortuosity, and design-related propensity for migration of the Tempofilter II filter. This case study underscores the importance of thorough preoperative vascular assessment, prompt CT evaluation in patients with postprocedural back pain, individualized endovascular management, and patient education on activity restrictions.
Conclusion
This case highlights the necessity of vigilant monitoring for rare but severe vascular complications following IVC filter placement.
Keywords: abdominal aorta injury, deep vein thrombosis, inferior vena cava filter, inferior vena cava perforation, pseudoaneurysm
1. Introduction
Inferior vena cava (IVC) filters are a well-established intervention for preventing pulmonary embolism (PE), a major cause of hospital mortality. Although filter-induced aortic injury is rare, it represents a severe procedural complication (1, 2). Herein, we report the case of an elderly patient who presented with acute left lower-limb pain, swelling, and PE, and subsequently underwent IVC filter placement followed by thrombolysis. Three days later, the patient developed back pain, a drop in hemoglobin levels, tachycardia, and hypotension secondary to IVC perforation with abdominal aortic pseudoaneurysm. Emergency covered stent repair and filter retrieval were successful. This case highlights the importance of standardized implantation practices, the early recognition of complications, and timely intervention to improve patient outcomes.
2. Case presentation
A 78-year-old woman presented with acute pain and swelling of the left lower limb that had persisted for 3 days, precipitated by prolonged sitting. Routine investigations revealed no evidence of thrombophilia, malignancy, or other provoking factors. Physical examination revealed pitting edema, mild erythema, and a slight increase in skin temperature of the left leg, without sensory or motor deficits. Limb circumferences measured at the ankle, 10 cm below the patella, and 15 cm above the patella were 23, 35, and 45 cm, respectively, on the left side, compared with 19, 31, and 38 cm on the right side (Supplementary Figure 1A). The left dorsalis pedis pulse was mildly diminished, whereas the right dorsalis pedis pulse was normal. Based on the aforementioned findings, there was no clinical evidence of phlegmasia cerulea dolens or acute limb ischemia. The patient had a 6-year history of hypertension but no history of diabetes mellitus, coronary heart disease, prior interventional procedures, smoking, or alcohol consumption. She also presented with chest tightness and cough, but no dyspnea or syncope. Doppler ultrasonography revealed extensive thrombosis involving the left iliac and femoropopliteal veins, while computed tomography angiography (CTA) demonstrated bilateral emboli in the main pulmonary arteries (Supplementary Figure 1B). These findings align with previous reports indicating that 70%–80% of PE cases originate from the lower extremity or pelvic veins, and that PE predominantly affects older adults (3).
Based on laboratory findings and echocardiographic assessments, and in the absence of hypotension or other signs of hemodynamic instability, the PE was classified as low risk. Given the extensive iliofemoral thrombosis, the patient's and family's strong preference for active treatment, economic considerations, and current deep vein thrombosis (DVT) guidelines (4, 5), catheter-directed thrombolysis with temporary IVC filter protection was selected. Through right internal jugular vein access, IVC venography was performed to exclude filling defects and identify the renal vein orifices (Supplementary Figure 2). A temporary IVC filter (Tempofilter II, B. Braun, Germany) was placed 1 cm below the renal veins, and a 5F thrombolytic catheter (Unifuse, AngioDynamics, USA) was inserted into the fibular vein (Supplementary Figures 3A,B). Postoperatively, 100,000 U of urokinase was administered by continuous infusion every 6 h, combined with 5,000 IU of low-molecular-weight heparin administered subcutaneously every 12 h. On postoperative day 3, the patient developed back pain, a marked decrease in hemoglobin level (from 117 g/L on day 1 to 60 g/L on day 3; Supplementary Figure 3C), tachycardia (from 90 to 116 bpm), and hypotension (from 133/75 to 93/62 mmHg). Notably, blood transfusion had been performed after stent implantation. Anticoagulation and thrombolytic therapy were discontinued immediately. Based on the patient's clinical symptoms and imaging findings demonstrating filter morphology (Supplementary Figures 4A,B), filter perforation with abdominal aortic injury was strongly suspected. CT subsequently confirmed IVC perforation with an abdominal aortic pseudoaneurysm (Figure 1).
Figure 1.

Abdominal aortic pseudoaneurysm: imaging findings. CT images in four planes show that a filter strut penetrated the IVC and entered the adjacent abdominal aorta, resulting in the formation of a pseudoaneurysm (red arrow). (A) Axial image. (B) 3D reconstruction. (C) Coronal image. (D) Sagittal image.
Considering the recent anticoagulation and thrombolytic therapy, advanced age, and financial constraints, the patient underwent abdominal aortic covered stent implantation with concomitant IVC filter retrieval. The procedure was performed through a right femoral arterial approach using a 14F vascular sheath. Angiography showed marked migration of a filter strut with formation of an abdominal aortic pseudoaneurysm (Figure 2A). A covered stent (Cook, 58 mm × 24 mm) was then deployed; however, angiography demonstrated persistent contrast extravasation (Figure 2B). Consequently, a second covered stent (Cook, 58 mm × 24 mm) was placed distally as a bridging stent, with no residual contrast extravasation. The stents remained well clear of the superior mesenteric and renal arteries (Figure 2C). The filter was subsequently retrieved via a transjugular approach without confirmatory angiography; therefore, venous extravasation or arteriovenous fistula could not be excluded intraoperatively, representing a procedural limitation. After discharge, the patient received oral rivaroxaban 20 mg once daily for anticoagulation. Follow-up CT at 1 month demonstrated a well-positioned stent without contrast extravasation (Figure 2D). The absence of related clinical symptoms also suggested that no clinically significant IVC injury had occurred. The timeline of clinical events is detailed in Table 1.
Figure 2.

Endovascular management of the abdominal aortic pseudoaneurysm. (A) Angiography shows a filter strut adjacent to the abdominal aortic pseudoaneurysm (red arrow). (B) After deployment of the first covered stent, contrast extravasation persists, and the pseudoaneurysm remains visible (red arrow). (C) After deployment of a second covered stent, completion angiography confirms complete exclusion of the pseudoaneurysm without residual contrast extravasation (red arrow). (D) Postoperative follow-up confirms sustained pseudoaneurysm exclusion (red arrow).
Table 1.
The timeline of clinical events.
| Days | Clinical events |
|---|---|
| 1 | Laboratory tests (Hb 117 g/L); Tempofilter II placement with catheter-directed thrombolysis; postoperative urokinase 100,000 U q6h thrombolysis; low-molecular-weight heparin 5,000 U q12h. |
| 2 | Laboratory tests (Hb 112 g/L); continued anticoagulation and thrombolysis; gradual resolution of limb swelling. |
| 3 | Laboratory tests (Hb 60 g/L); patient complained of back pain; acute hemoglobin drop; CT revealed filter migration and perforation with abdominal aortic pseudoaneurysm; anticoagulation and thrombolysis discontinued; emergent endovascular repair with IVC filter retrieval and blood transfusion performed. |
| 4 | Hemoglobin stabilized; anticoagulation resumed. |
| 5–14 | Clinical condition stable; symptoms gradually resolved; no significant lower-limb swelling. |
| 15 | Discharged; prescribed long-term rivaroxaban 20 mg once daily. |
| 30 | Follow-up aortic CTA performed. |
3. Discussion
In this patient with lower-limb DVT and PE, temporary IVC filter implantation followed by catheter-directed thrombolysis was complicated by back pain and a marked decrease in hemoglobin level. CT and angiography confirmed that a filter strut had perforated the IVC and penetrated the adjacent abdominal aorta, resulting in pseudoaneurysm formation. This case highlights the importance of recognizing aortic injury as a rare but severe complication of IVC filter placement.
IVC filter penetration is defined as extension of a filter strut >3 mm beyond the caval wall and may result in complications such as aortic pseudoaneurysm, arterial thrombosis, retroperitoneal hematoma, arteriovenous fistula, and aortic dissection. Abdominal pain is the most frequently reported symptom, occurring in 30.2% of cases (6). Although the mechanism of IVC filter penetration is not fully understood, proposed risk factors include filter geometry (conical filters with free struts), outward radial force, barb sharpness, strut stiffness, filter tilt >15°, small IVC diameter, and reduced caval wall strength (7). Conical filters have been associated with higher rates of fracture and penetration, while cylindrical or umbrella-like filters are associated with a higher incidence of IVC occlusion (8, 9). Additional contributing factors include abdominal trauma, cardiopulmonary motion, postural changes, and fluctuations in caval pressure (10, 11). The Tempofilter II is a tethered, barbless filter anchored to a subcutaneous cervical port via an indwelling jugular catheter. Physiological neck movements—including flexion, extension, and rotation—transmit traction through the catheter, allowing the filter to migrate up to 2 cm within the IVC. Accordingly, postoperative management includes 24 h of bed rest and restriction of vigorous neck movements, with flexion and extension limited to ≤30° during the first postoperative week to prevent excessive filter displacement. In this case, the Tempofilter II was selected primarily for economic reasons because the patient came from a rural area. The device can be retrieved at the bedside without specialized equipment (12), thereby reducing procedural costs and operator radiation exposure. However, these advantages are accompanied by tradeoffs, including filter migration related to cervical motion, increased nursing care burden, risk of catheter-related infection, and reduced patient comfort. From a technical perspective, the filter is easy to retrieve and aligns with current guidelines for the management of lower-extremity DVT. Nevertheless, the patient's advanced age and long-standing hypertension may have contributed to aortic sclerosis and reduced vascular compliance, resulting in aortic tortuosity and a reduced distance between the IVC and the abdominal aorta—anatomical features that may increase the risk of filter-related vascular injury. This case underscores the importance of considering preoperative aortic imaging, such as ultrasonography or CTA, in elderly patients to guide appropriate filter selection. It also emphasizes that prompt evaluation for filter-related injury in patients who develop symptoms such as back pain may facilitate timely diagnosis and definitive management.
Documentation of intra-abdominal aortic injury secondary to IVC filter is largely limited to case reports (13, 14). Kim et al. described a patient who underwent IVC filter removal, debridement, and reconstruction of the aortic and IVC. In another case, Jalili et al. performed open surgical repair with filter removal and reconstruction of the aorta using a Dacron patch. These reports emphasize that treatment strategies must be individualized according to clinical severity, institutional expertise, and patient comorbidities. In the present case, because of the patient's advanced age and hemorrhagic shock, an endovascular approach was selected for both filter retrieval and aortic repair. After considering the practicality and cost-effectiveness of abdominal endovascular devices, a CUFF stent was selected for abdominal aortic repair. Complete exclusion of the pseudoaneurysm was achieved using two CUFF stents. The endovascular approach effectively managed the complication while avoiding open surgery, and 1-month follow-up confirmed successful exclusion of the pseudoaneurysm.
However, several limitations should be acknowledged. First, preoperative aortic imaging was unavailable, precluding the assessment of preexisting aortic pathology (e.g., ulcer, dissection, or aneurysm), which could have contributed to underestimation of the lesion's complexity, including the possibility of multiple perforations and inadequate stent apposition in the tortuous aortic segment. Second, patient compliance and activity history (e.g., excessive neck movement, precipitating factors, or physiological motion related to diaphragmatic excursion or intestinal peristalsis) were not systematically documented, limiting the evaluation of the proposed mechanism of filter migration and vascular injury. Third, cavography was not performed after filter retrieval to exclude contrast extravasation from the IVC. Finally, long-term follow-up was not possible because the patient resided in a remote rural area with limited access to healthcare.
This case report, despite its limitations, provides practical clinical insights for clinicians and may inform future IVC filter design. In particular, it highlights the potential risk of vascular perforation associated with the device's permissible migration range, underscores the importance of the preoperative evaluation of perivascular anatomy in high-risk patients, supports prompt contrast-enhanced CT in patients who develop postprocedural symptoms such as back pain, and highlights the need for postoperative patient education and careful filter selection based on individual anatomical and clinical characteristics.
4. Conclusion
This case of Tempofilter II filter-induced aortic pseudoaneurysm suggests that the device's permissible migration range may contribute to the risk of vascular perforation, particularly in elderly patients with aortic tortuosity. It underscores the importance of the preoperative assessment of perivascular anatomy, prompt contrast-enhanced CT evaluation in patients who develop postprocedural back pain, meticulous patient education, and careful filter selection based on individual anatomical and clinical characteristics.
Acknowledgments
We thank the patient for granting permission to publish this clinical information.
Funding Statement
The author(s) declared that financial support was received for this work and/or its publication. This study was supported by research funding from the corresponding author at Chongqing Medical University, China (No. 2030220220018-X1-2993).
Footnotes
Edited by: Mafalda Massara, GOM Reggio Calabria, Italy
Reviewed by: Adham Alkurashi, George Washington University Hospital, United States
Khurram Khaliq Bhinder, Mayo Clinic, United States
Data availability statement
The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.
Ethics statement
Written informed consent was obtained from the individual(s) for the publication of any potentially identifiable images or data included in this article.
Author contributions
HZ: Writing – original draft. YW: Data curation, Visualization, Formal analysis, Writing – review & editing, Methodology, Investigation. ZT: Writing – review & editing. MK: Writing – review & editing. FW: Supervision, Writing – review & editing. FG: Writing – review & editing, Conceptualization, Funding acquisition.
Conflict of interest
The author(s) declared that this work was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Generative AI statement
The author(s) declared that generative AI was not used in the creation of this manuscript.
Any alternative text (alt text) provided alongside figures in this article has been generated by Frontiers with the support of artificial intelligence and reasonable efforts have been made to ensure accuracy, including review by the authors wherever possible. If you identify any issues, please contact us.
Publisher's note
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.
Supplementary material
The Supplementary Material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fsurg.2026.1895441/full#supplementary-material
References
- 1.Brugger M, Gaa J, Bergmann K. Disseminated embolization of an inferior vena cava filter with penetration of the aorta. Eur Heart J Cardiovasc Imaging. (2024) 25:e253. 10.1093/ehjci/jeae124 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Lee J, Roche-Nagle G. Permanent IVC filter strut penetration into an abdominal aortic aneurysm. BMJ Case Rep. (2021) 14:e241962. 10.1136/bcr-2021-241962 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Glazier CR, Baciewicz FA. Epidemiology, etiology, and pathophysiology of pulmonary embolism. Int J Angiol. (2024) 33:76–81. 10.1055/s-0044-1785487 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Vascular Surgery Group, Society of Surgery, Chinese Medical Association. Guidelines for the diagnosis and treatment of deep venous thrombosis (3rd edition). Chin J Gen Surg. (2017) 32:807. 10.3760/cma.j.issn.1007-631X.2017.09.032 [DOI] [Google Scholar]
- 5.Kaufman JA, Barnes GD, Chaer RA, Cuschieri J, Eberhardt RT, Johnson MS, et al. Society of Interventional Radiology clinical practice guideline for inferior vena cava filters in the treatment of patients with venous thromboembolic disease. J Vasc Interv Radiol. (2020) 31:1529–44. 10.1016/j.jvir.2020.06.014 [DOI] [PubMed] [Google Scholar]
- 6.Bai Y, Wang K, Jia Z. A systematic review of cases of aortic penetration by inferior vena cava filters. Abdom Radiol. (2025) 51:1035–45. 10.1007/s00261-025-05131-3 [DOI] [PubMed] [Google Scholar]
- 7.Hudák M, Koščo M, Rašiová M. Successful endovascular retrieval of an inferior vena cava filter penetrating into aorta. J Cardiol Cases. (2023) 28:239–41. 10.1016/j.jccase.2023.08.001 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Tian X, Liu J, Li J, Liu X. Case report: endoluminal removal of a retrievable conical inferior vena cava filter with a ruptured retraction hook attached to the wall. Front Surg. (2022) 9:985060. 10.3389/fsurg.2022.985060 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Deso S, Idakoji I, Kuo W. Evidence-Based evaluation of Inferior vena Cava filter complications based on filter type. Semin Intervent Radiol. (2016) 33:93–100. 10.1055/s-0036-1583208 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Jia Z, Wu A, Tam M, Spain J, McKinney JM, Wang W. Caval penetration by inferior vena Cava filters: a systematic literature review of clinical significance and management. Circulation. (2015) 132:944–52. 10.1161/CIRCULATIONAHA.115.016468 [DOI] [PubMed] [Google Scholar]
- 11.Kossa A, McMeeking RM. The effect of an implanted filter on valsalva-compression and respiratory-compression of the inferior vena Cava. J Elast. (2021) 145:383–408. 10.1007/s10659-021-09850-8 [DOI] [Google Scholar]
- 12.Yim NY, Chang NK, Lim JH, Kim JK. Retrograde tempofilter IITM placement within the superior vena cava in a patient with acute upper extremity deep venous thrombosis: the filter stands on its head. Korean J Radiol. (2011) 12:140. 10.3348/kjr.2011.12.1.140 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Jalili J, Vaseghi S, Baastani Khajeh M, Haji Kamanj Olia A. Penetration of inferior vena cava (IVC) filter into abdominal aorta and repairing by Dacron patch arthroplasty: a case report. Radiol Case Rep. (2025) 20:5184–8. 10.1016/j.radcr.2025.06.104 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Kim KW, Jung HS. Management of an infected infra-renal aortic pseudoaneurysm caused by penetration of an inferior vena cava filter: a case report. Ann Vasc Surg Brief Rep Innov. (2023) 3:100237. 10.1016/j.avsurg.2023.100237 [DOI] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.