Abstract
Background
Accurate modelling and performance evaluation of gamma-ray scintillation detectors are required in single photon emission computed tomography (SPECT) systems to ensure reliable nuclear medicine diagnostics. Monte Carlo simulation is therefore a powerful computational technique used to accurately model the gamma-ray scintillation detectors and photon transport in medical imaging systems. Therefore, the objective of this study is to validate the SIMIND Monte Carlo simulation code using a self-developed peak-detection algorithm for clinical SPECT gamma cameras (Siemens E.Cam and GE Infinia), with a focus on energy-spectrum decomposition and image-profile analysis.
Methods
This study presents a dual-system evaluation of clinical SPECT gamma cameras—the Siemens E.Cam and GE Infinia systems using the SIMIND Monte Carlo simulation code. We assessed collimator and detector-specific SPECT performance metrics, including energy resolution, sensitivity, image evaluation, spatial resolutions, and spectral fidelity for a 99mTc point source and spatial resolution with Co-57 flood source. A custom algorithm was then developed to automatically decompose energy spectra and identify key spectral components, including the photopeak, Compton edge, iodine escape, and lead X-ray peaks. Quantitative methods, such as 2D and 3D image profile analysis and FWHM comparisons, were integrated to validate the simulation against experimental data.
Results
Energy resolution differences ≤ 4% and sensitivity variations < 1.2% were observed between simulated and experimental measurements. Spatial resolution analysis revealed a range of 3.5–4.8 mm for Siemens E.Cam (bar and pie phantoms) and ~ 1 cm for GE Infinia, highlighting system-specific imaging capabilities. A custom algorithm successfully identified key spectral components, including photopeaks, Compton edges, and lead X-ray peaks, with deviations < 3%. Image profile analysis revealed FWHM values of 4–4.5 pixels for both systems.
Conclusion
This work shows that SIMIND accurately reproduces the fundamental photon interaction characteristics including energy spectra, energy resolution, and sensitivity for both Siemens E.Cam and GE Infinia SPECT systems. However, for image-profile evaluation, good agreement was obtained for Siemens E.Cam, whereas the GE Infinia results revealed a system-specific discrepancy associated with the actual clinical condition of the installed camera. Nevertheless, when used appropriately, SIMIND provides a robust framework for optimizing quantitative SPECT protocols and guiding system design improvements, particularly for well-calibrated systems.
Graphical abstract

Keywords: SIMIND Monte Carlo simulation, Scintillation detectors, Gamma camera, SPECT, 99mTc, FWHM
Introduction
Single photon emission computed tomography (SPECT) is a nuclear medicine imaging technique that generates three-dimensional functional images by detecting gamma rays emitted from radiopharmaceuticals administered to patients. A gamma camera rotates around the patient during a SPECT scan, capturing projection data from multiple angles. These projections are then reconstructed into cross-sectional images, enabling physicians to assess physiological processes such as metabolic activity or blood flow. However, optimizing SPECT systems for clinical use requires rigorous performance evaluation of scintillation detectors. Therefore, a regular detector-specific characteristic performance is required to analyze the system's behavior.
Nuclear medicine imaging depends on precise and quantitative models to accurately represent the interactions between photons and biological tissues. These interactions are governed by the radiative transport equation (RTE), a mathematical framework that describes how radiation propagates and scatters within matter. Unfortunately, analytical solutions to the RTE are limited to basic domains, including infinite slabs, semi-infinite media, and infinite media; for complex or random media, numerical computations are necessary. Therefore, the Monte Carlo method solves the RTE by simulating the paths of many photons (or quanta of energy) as they propagate through the medium, experiencing scattering, absorption, and possibly emission. Each photon's path is determined by random sampling from probability distributions that represent the physical processes described by the RTE. Monte Carlo simulation is therefore a powerful computational technique used to model the behavior of photons in medical imaging systems. It is especially useful in scenarios where analytical solutions are difficult or impossible to obtain due to the complexity of the system [1].
Over the years, several Monte Carlo codes have been developed for the simulation of radiation transport, such as EGS [2, 3], EGSnrc [4], Geant4 [5], PENELOPE [6], SuperMC [7, 8]. These codes are capable of simulating particle transport and their interactions with matter under almost all circumstances. However, modeling nuclear medicine facilities such as SPECT and PET (Positron Emission Tomography) is more convenient in dedicated Monte Carlo codes, which include SIMIND [9], GATE [10], and SimSET [11]. SIMIND mainly focused on simulating standard clinical SPECT systems, with particular emphasis on accurate modeling of collimator-detector interactions and photon transport. It can be easily modified to handle different types of calculations and measurements related to SPECT imaging, such as studying the effects of patient motion, photon attenuation, and scatter on image quality.
A limited number of studies [12–21] have been done for the simulation of the SPECT system through the Monte Carlo code SIMIND [9]. Performance evaluation of the Siemens E.Cam system through SIMIND was done by Bahreyni Toossi et al. [12]. A similar analysis was performed by Rodrigues et al. [16], where GEANT4 has been used as a simulation toolkit. However, a gap exists in comprehensive energy spectrum analysis, including quantitative characterization of secondary peaks (e.g., Compton edge, iodine escape, lead X-rays), critical for spectral fidelity. This study introduces a custom peak identification algorithm (“Energy spectrum analysis and peak characterization algorithm” section) for detailed energy spectrum decomposition, enabling the automatic identification and classification of photon interaction peaks, including photopeaks, backscatter, Compton edges, and iodine escape peaks. This approach rigorously validates simulated spectra against experiments, addressing a key limitation.
In this work, we simulated the performance of Siemens’ E.Cam dual-head gamma camera using the SIMIND Monte Carlo program [9], installed in our nuclear medicine center, to evaluate the collimator and detector-specific performance in terms of energy resolution, sensitivity, image evaluation, and spatial resolution using 99mTc point source. A new method of image evaluation is proposed by drawing image profiles and comparing the corresponding FWHM. Further, a comprehensive analysis of measured and simulated energy spectra is carried out to compare all peaks that appeared in the spectrum due to photon interactions with matter.
Methods and materials
The Siemens E.Cam dual-head gamma camera (Siemens Medical Solutions, Hoffman Estates, USA) and the GE Infinia gamma camera were modeled in SIMIND Monte Carlo code (v6.1.2), incorporating manufacturer-specified detector geometries and material properties. Experimental and simulated performance metrics, including energy resolution, sensitivity, spatial resolution, and image evaluation, were assessed for both Siemens E.Cam and GE Infinia systems using standardized phantoms (bar, pie, and cylindrical) and radioactive sources (99mTc point source, 57Co flood source) simulations.
Experimental setup
To evaluate energy resolution, sensitivity, and image quality, a point source was prepared in the hot lab using a small piece of cotton and a syringe cap (serving as a cylindrical phantom). The point source is an isotropic 99mTc source emitting 140 keV photons. Two gamma camera systems, Siemens E.Cam and GE Infinia, were then used to carry out the performance analysis. The Siemens E.Cam dual-head gamma camera features two gamma-ray detector units, each integrating a removable Low Energy All Purpose (LEAP) collimator optimized for 140 keV photons, coupled to a thallium-activated sodium iodide (NaI(Tl)) scintillation crystal (9.5 mm thickness, 40,000 photons/MeV light yield) with peak emission at 415 nm. The optical system consists of a Pyrex light guide (n = 1.47) that efficiently channels scintillation photons to a 59-channel photomultiplier tube (PMT) array (Hamamatsu R6231-100), achieving 9.9% energy resolution at 140 keV. This detector configuration enables precise spatial and energetic characterization of gamma emissions through coordinated signal readout from both detector heads.
The same performance analysis tests have also been done on the GE Infinia dual-head camera to ensure the versatility of the SIMIND Monte Carlo code. The GE Infinia system also incorporates two detector assemblies, each equipped with a modular Low Energy General Purpose (LEGP) collimator. The LEGP collimator consists of parallel hexagonal channels (1.9 mm pore diameter, 35 mm length, 0.20 mm septal thickness), optimized for imaging low-energy γ-emitters like 99mTc (photopeak 140 keV). This honeycomb structure provides a geometric efficiency of 1.2 × 10–4 while maintaining 90% septal penetration rejection at 140 keV.
Monte Carlo simulation
The Monte Carlo code SIMIND was used to model both gamma camera systems to carry out their performance evaluation, as shown in Fig. 1. SIMIND employs a photon transport algorithm that tracks each gamma ray through successive interactions, including Compton scattering (modeled using the Klein–Nishina differential cross-section), photoelectric absorption (with characteristic X-ray emission), and coherent scattering. For NaI(Tl) detectors, SIMIND also simulates scintillation light generation, light transport through the crystal and light guide, and photomultiplier tube (PMT) signal formation.
Fig. 1.

Construction of a gamma camera in the SIMIND Monte Carlo code
A computer equipped with AMD Ryzen 7 4700U 2.00 GHz (8 CPUs) with Radeon(TM) Graphics and 8 GB memory is used for the simulation of the SPECT system. The SIMIND Monte Carlo simulation code employs a dual-program architecture for radiation detector modeling. The CHANGE module provides a graphical interface for configuring the gamma camera's physical parameters, including source geometry, collimator specifications (hole pattern, septal thickness), and detector properties (scintillator thickness, material composition). These user-defined parameters are stored in standardized input files. The SIMIND then processes these inputs using a photon transport algorithm that accounts for Compton scattering, photoelectric absorption, and characteristic X-ray emission within the detector system.
As documented in Table 1, all simulations incorporated manufacturer-specified parameters for both the Siemens E.Cam and GE Infinia systems, ensuring physical accuracy in modeling the radiation detection from photon interaction to signal generation.
Table 1.
Parameters of Siemens E.Cam and GE Infinia defined in the SIMIND Monte Carlo code.
| Parameters | Specifications | ||
|---|---|---|---|
| Siemens E.Cam | GE Infinia | ||
| Source | Sources |
(i) Tc-99m Point source (ii) 57Co Flood source (5 mCi) for spatial resolution |
|
| Phantom | Cylindrical Phantom for point source, Bar phantom, Pie phantom | ||
| Source activity | 51.06 MBq | 24.79 MBq | |
| Number of counts | 128 million counts | ||
| Gamma-ray energy | 140 keV | ||
| Detector | Crystal | NaI(Tl) | |
| Collimator | Collimator type | Low Energy All Purpose (LEAP) | Low Energy General Purpose (LEGP) |
| Hole shape | Hex | ||
| No. of holes | 90000 | 56560 | |
| Hole length | 24.1 mm | 35 mm | |
| Septal thickness | 0.2 mm | 0.2 mm | |
| Hole diameter | 1.45 mm | 1.9 mm | |
| Image | Pixel size |
0.39 mm 128x128 |
|
| Matrix size | |||
| Camera Parameters | Source to detector height | 25 cm | 25 cm |
| Back-scattering (BK) material | Pyrex | Pyrex | |
| Thickness of BK material | 6 cm | 6 cm | |
| Intrinsic spatial resolution | 3.9 mm | 3.9 mm | |
| Energy resolution | 9.90% | 9.80% | |
| Energy window | 15% | 10% | |
Performance evaluation of gamma-ray scintillation detectors
The performance of gamma-ray scintillation detectors was evaluated through four critical metrics: energy spectrum analysis (characterizing photopeaks and secondary features), sensitivity quantification, spatial resolution assessment using specialized phantoms, and image evaluation via 2D and 3D profile comparisons. Both experimental measurements and SIMIND Monte Carlo simulations were conducted in parallel for the Siemens E.Cam and GE Infinia systems, enabling direct validation of detector models against clinical performance data under identical conditions.
Energy spectrum analysis and peak characterization algorithm
For energy resolution measurements, a lower activity of 3.7 MBq was used to minimize pulse pile-up effects, while higher activities (51.06 MBq) were employed for sensitivity and image quality assessments to achieve adequate count statistics within clinical acquisition times. Energy resolution was measured using a 3.7 MBq 99mTc point source centered in the detector field of view at 25 cm from the NaI(Tl) crystal surface. For the Siemens E.Cam system, energy spectra were acquired under two configurations: (1) with the LEAP collimator (1.45 mm hexagonal holes, 24.1 mm length) to assess extrinsic resolution, and (2) without collimation to determine intrinsic detector performance. Each measurement collected 1 million photopeak events (140 keV ± 10%) to ensure statistical significance (< 2% Poisson noise). The GE Infinia system was similarly evaluated using its LEGP collimator (1.9 mm holes, 35 mm length), maintaining identical source geometry for comparative analysis. Monte Carlo simulations in SIMIND replicated these experimental conditions, including exact collimator geometries and the 25 cm source-to-detector distance.
For a 99mTc source (140 keV gamma rays) interacting with a NaI(Tl) scintillation crystal, the photopeak represents full energy deposition, the Compton edge arises from inelastic scattering, the iodine escape peak (112 keV) results from characteristic X-ray emission following photoelectric absorption in iodine, and lead X-ray peaks (~ 72–75 keV) originate from collimator interactions. To quantitatively analyze the energy spectra acquired through experimental measurements and SIMIND Monte Carlo simulations, a custom algorithm was developed to identify and characterize key photopeaks and secondary spectral features (Compton edge, back-scatter peak, characteristic Lead X-ray peak, and Iodine escape peak). Accurate energy spectrum analysis is critical for validating Monte Carlo detector models because different photon interaction mechanisms produce distinct spectral features. This analysis was critical for determining the energy resolution, full width at half maximum (FWHM), and spectral fidelity of both gamma camera systems.
(a) Calculation of energy resolution and FWHM
The energy resolution (R) is given as the ratio of the full width at half maximum (FWHM) of the photopeak to its peak energy Epeak, expressed as a percentage.
![]() |
1 |
For 99mTc, the nominal peak energy is 140 keV. To compute FWHM reliably from discrete energy spectra, the following sequential procedure was applied identically to both experimental and simulated data.
(1) Smoothing: The raw spectrum often contains statistical fluctuations due to Poisson noise. Therefore, prior to peak identification, smoothing filters are applied to reduce statistical fluctuations in measured spectra [23–25]. To enhance peak visibility and stability of calculations, a two-sided moving average filter was applied. For each bin i, the smoothed count value Ssmooth(i) was calculated by averaging the counts in the interval from i−w to i+w, where w is the half-window size of the filter and S(j) denotes the original spectrum count in bin j:
![]() |
2 |
(2) Peak detection: After smoothing, a peak was identified as a local maximum satisfying the condition given in equation (3).
![]() |
3 |
Only peaks with heights exceeding a certain threshold (≥ 5% of the maximum spectrum intensity) were selected for further analysis. The energy corresponding to a detected peak at index p is assigned as the peak energy, Epeak = E(p), where E(p) is the calibrated energy value corresponding to bin p. The 140 keV photopeak was selected for FWHM and energy-resolution calculation.
(3) Half-maximum positions: In order to calculate FWHM, the two energy values EL (left half-maximum energy) and ER (the right half-maximum energy) are calculated where the peak amplitude Ssmooth(p) of the smoothed spectrum falls to half its maximum height. Let H = Ssmooth(p) be the peak height, then:
![]() |
4 |
Because the spectrum is discrete, linear interpolation between adjacent bins gives:
![]() |
5 |
where
is the nearest bin to the peak on the left side for which
, and
. Thus EL is obtained by interpolation between bins
and
. Similarly, the right half-maximum energy ER is
![]() |
6 |
where
is the nearest bin to the peak on the right side for which
, and
. Thus ER is obtained by interpolation between bins
and
.
(4) FWHM and energy resolution: The full width at half maximum, as shown in Fig. 2, is then calculated as
![]() |
7 |
Fig. 2.

Schematic illustration of the interpolation method used to determine the left and right half-maximum energies, EL and ER, from a smoothed energy spectrum
Finally, the energy resolution R of the 140 keV photopeak was calculated using Eq. (1). This value provides a quantitative assessment of the detector’s ability to resolve monoenergetic gamma rays. The interpolation procedure used to determine the left and right half-maximum energies, EL and ER, corresponding to Eqs. (5) and (6), is illustrated schematically in Fig. 2.
(b) Peak classification
Each peak was classified based on its central energy Epeak using known interaction signatures for 99mTc. Table 2 summarizes the characteristic spectral features observed when 140 keV 99mTc γ-rays interact with NaI(Tl), including the photopeak, Compton edge, backscatter peak, iodine escape peak, and lead X-ray peaks, which collectively enable detector performance evaluation.
Table 2.
Energy calculation of different peaks in the energy spectrum of 99mTc
| Peaks in energy spectrum | Calculation |
|---|---|
| Photo-peak | 140 keV |
| Compton edge | ![]() |
| Back-scatter peak |
= 90.442 keV |
| Characteristic Lead X-ray peak | 72 keV |
| Iodine escape peak |
= 112 keV |
The classification was determined by evaluating:
![]() |
8 |
This algorithm was implemented in C++ and applied identically to both experimental and simulated spectra, ensuring uniformity in performance validation across both modalities. It enabled high-fidelity peak characterization critical for detector benchmarking and image quality analysis in SPECT systems. The pseudocode of this algorithm is given in “Pseudocode for energy spectrum analysis” Appendix section.
Sensitivity
The sensitivity of the gamma-ray detector was determined as
![]() |
9 |
Where Ndet is the total counts in the photopeak window (126–154 keV), t is acquisition time (in seconds), and A is the source activity (MBq).
Spatial resolution
Spatial resolution of the Siemens E.Cam dual-head gamma camera and GE Infinia system was measured by placing a flood source of 57Co (5 mCi) on the top of the bar phantom. The bar phantom is placed on the top of the collimator, while simulations employed a QC Pie Phantom to qualitatively assess resolution limits.
Image evaluation
The acquisition parameters for experiments were a 128 × 128 matrix, 0.39 mm pixel size, and approximately 1 million counts/projection. Image and corresponding 2-D and 3-D profiles were obtained with collimators through both gamma camera systems and compared with their simulated image profiles and corresponding FWHM.
Results
Energy resolution
The simulated and measured energy spectra produced by Siemens E.Cam dual-head gamma camera with and without LEAP collimator for 99mTc point source (γ -ray energy of 140 keV) are shown in Fig. 3. An excellent agreement is obtained between the simulated and the measured energy spectra for both cases.
Fig. 3.

Comparison of the measured and simulated energy spectra for a 99mTc point source from Siemens E.Cam a with LEAP collimator, and b without collimator
For the 140 keV photopeak, the experimental FWHM values were 14.58 keV (both with and without LEAP collimator), corresponding to relative energy resolutions of 10.42%. Monte Carlo simulations yielded comparable results, with FWHM values of 14.05 keV (10.03%) and 14.11 keV (10.08%) for collimated and uncollimated configurations, respectively. This ≤ 0.4% deviation between simulated and experimental resolutions validates the accuracy of our detector modeling approach. Notably, both results align well with the manufacturer's specified resolution [25] of 9.9% at 140 keV, confirming proper system calibration and simulation parameterization.
The simulated and experimentally determined values of FWHM, energy resolution, and energies of other photon interactions of a 99mTc point source acquired through Siemens E.Cam are given in Table 3. Percentage differences are calculated as
![]() |
10 |
Table 3.
Results from the energy spectrum of a 99mTc point source acquired through Siemens E.Cam
| Parameters | Simulated | Measured | % Difference | |||
|---|---|---|---|---|---|---|
| With LEAP Collimator (Extrinsic) | Without LEAP Collimator (Intrinsic) | With LEAP Collimator (Extrinsic) | Without LEAP Collimator (Intrinsic) | Extrinsic | Intrinsic | |
| FWHM (keV) | 14.05 | 14.11 | 14.58 | 14.58 | 3.6 | 3.2 |
| Energy resolution (%) | 10.03 | 10.08 | 10.42 | 10.42 | 3.7 | 3.3 |
| Compton edge (keV) | 47.2 | 47 | 47 | 47 | 0.4 | 0 |
| Pb X-Ray peak (keV) | 74.9 | – | 75 | – | 0.1 | – |
| Back-scatter peak (keV) | 88.8 | 92 | 90.3 | 90 | 1.7 | 2.2 |
| Iodine escape peak (keV) | 113 | 111.9 | 112 | 111.5 | 0.9 | 0.4 |
| Photo-peak (keV) | 140.95 | 140.8 | 140 | 140 | 0.68 | 0.57 |
The GE Infinia system also demonstrated strong agreement between measured and simulated performance metrics, as shown in Fig. 4. Experimental measurements yielded an FWHM of 13.44 keV for the 140 keV photopeak, corresponding to an energy resolution of 9.80%. Monte Carlo simulations predicted slightly broader spectral resolution, with FWHM = 13.84 keV (9.60% resolution). This close correspondence validates the accuracy of our Monte Carlo model in replicating the LEGP collimator's photon detection characteristics.
Fig. 4.

Comparison of the measured and simulated energy spectra for a 99mTc point source from GE Infinia with LEGP collimator
The simulated and experimentally determined values of FWHM, energy resolution, and energies of other photon interactions of a 99mTc point source acquired through GE Infinia are given in Table 4.
Table 4.
Results from the energy spectrum of 99mTc point source acquired through GE Infinia system
| FWHM (keV) | Energy resolution (%) | Compton edge (keV) | Pb X-Ray peak (keV) | Back-scatter peak (keV) | Iodine escape peak (keV) | Photo-peak (keV) | |
|---|---|---|---|---|---|---|---|
| Simulated | 13.84 | 9.6 | 47.02 | 75 | 89 | 112.3 | 141.1 |
| Measured | 13.44 | 9.8 | 47 | 74.02 | 90.1 | 112 | 140 |
| % Difference | 2.97 | 2.04 | 0.04 | 1.3 | 1.22 | 0.27 | 0.79 |
Sensitivity
Sensitivity measurements demonstrated close agreement between experimental and simulated values for both detector systems. The Siemens E.Cam showed experimental sensitivity of 136.83 cps/MBq compared to simulated values of 138.35 cps/MBq, representing a 1.1% difference. Similarly, the GE Infinia exhibited experimental sensitivity of 107.67 cps/MBq versus simulated 108.91 cps/MBq, representing a 1.15% difference.
Image evaluation
Image evaluation of siemens E.Cam
The acquired images from the Siemens E.Cam system and their corresponding spatial profiles are presented in Figs. 5, 6 and 7.
Fig. 5.

Images of 99mTc point source obtained by the a simulation of Siemens E.Cam and through the b experiment
Fig. 6.

a Horizontal image profiles for simulated and actual images, b vertical image profiles for simulated and actual (measured) images
Fig. 7.

a simulated 3D profiles of the image of 99mTc point source, b measured 3D profile of the image of the 99mTc point source.
Figure 6 demonstrates strong agreement between the simulated and experimental image profiles, with all profiles exhibiting a full width at half maximum (FWHM) within the range of 4–4.5 pixels. This close correspondence is further supported by the 3D image profiles shown in Fig. 7, where the simulated and measured data show minimal divergence in spatial response.
Image evaluation of GE Infinia
Figures 8, 9 and 10 present the acquired images from the GE Infinia system and their corresponding spatial profiles.
Fig. 8.

Images of 99mTc point source obtained by the a simulation of GE Infinia and through the b experiment
Fig. 9.

a Horizontal image profiles for simulated and actual images, b vertical image profiles for simulated and measured images
Fig. 10.

a simulated 3D profiles of the image of 99mTc point source, b 3D profile of the measured image of 99mTc point source
As evident in Fig. 8b, the experimental data exhibit sub-optimal uniformity and linearity, resulting in broadened point-source profiles without a distinct sharp peak. While the simulated profiles (Figs. 9, 10) align well with theoretical expectations, they replicate the observed “star effect” artifact due to the inherent limitations of the clinical GE Infinia system, as confirmed by the measured image quality in Fig. 8b. The differences in image profiles, as shown in Figs. 9, 10, acquired through simulation and experiment, are because of the poor image quality of the GE Infinia installed in our nuclear medicine center, and this can be verified from Fig. 8b.
Spatial resolution
Spatial resolution of siemens E.Cam
The spatial resolution of the Siemens E.Cam dual-head gamma camera was measured by placing a 5 mCi 5⁷Co flood source on top of a bar phantom, which was positioned directly on the collimator. In the simulation, the spatial resolution was assessed using a QC Pie Phantom. The resulting images from both the experiment and simulation are shown in Fig. 11.
Fig. 11.

a Actual bar phantom image obtained from Siemens E.Cam, b Simulated pie phantom image
Notably, in the Pie Phantom, all rods are distinctly visible, with the smallest rod diameter being 4.8 mm, suggesting that the spatial resolution of the Siemens E.Cam could be as fine as 4.8 mm. The bar phantom image further supports this, indicating a resolution of approximately 3.5 mm or slightly higher. Based on these findings, it can be concluded that the spatial resolution of the Siemens E.Cam installed at our nuclear medicine facility ranges between 3.5 and 4.8 mm.
Spatial resolution of GE Infinia
The spatial resolution of the GE Infinia dual-head gamma camera was measured using the same method as that of the Siemens E.Cam. A bar phantom was placed on the collimator for the experimental measurement, while the spatial resolution in the simulation was qualitatively assessed using a QC Pie Phantom. The resulting images from both the experiment and simulation are presented in Fig. 12.
Fig. 12.

a Actual bar phantom image obtained from GE Infinia, b simulated pie phantom image
In the Pie Phantom, all rods are clearly visible, with the smallest rod diameter being 4.8 mm, suggesting that the spatial resolution of the GE Infinia could be as fine as 4.8 mm. However, the bar phantom image yielded disappointing results, indicating a resolution of approximately 1 cm, which corresponds to the separation between the four quadrants. Thus, the spatial resolution of the GE Infinia installed at our nuclear medicine facility appears to be around 1 cm.
Discussion
The findings validate that the SIMIND Monte Carlo algorithm can accurately model the energy-spectral and sensitivity characteristics of clinical gamma camera systems (Siemens E.Cam and GE Infinia) under standardized testing conditions. However, the image profile discrepancies observed for the GE Infinia system are attributed to detector non-uniformity and electronic noise in the clinical installation, demonstrate that SIMIND models an idealized system.
Energy resolution: The measured and simulated energy resolutions for both systems exhibited close agreement, with discrepancies of ≤ 3.7% for Siemens E.Cam with LEAP collimator, and ≤ 3.3% for Siemens E.Cam without collimator. For the GE Infinia system with LEGP collimator, the difference between simulation and experiment was approximately 2.0%. This close agreement demonstrates that the SIMIND Monte Carlo model accurately reproduces both intrinsic and extrinsic detector energy resolution characteristics, including scintillation photon generation, photon transport in the NaI(Tl) crystal, and photomultiplier signal response.
To further evaluate the accuracy of the intrinsic detector modeling, the obtained energy resolution values were compared with previously published Monte Carlo studies of Siemens gamma cameras [12, 13]. Table 5 summarizes the intrinsic energy resolution values reported in the literature [12, 13] for Tc-99 m (140 keV), together with the present study results and manufacturer’s specified data [25], demonstrating that all values fall within the expected 9.5–10.5% range for clinical SPECT systems. The simulated and measured intrinsic energy resolution obtained in the present study is 10.08 and 10.42%, respectively, which is in good agreement with previously reported values 10.01% and 9.68% by Bahreyni Toossi et al. [12] and 9.68% and 9.54% by Morphis et al. [13] for Siemens gamma camera systems. The slight differences between studies may arise from variations in detector configuration (Siemens E.Cam vs Siemens Symbia T16 [12, 13]), source size, and the specific energy-resolution modeling approach implemented in the SIMIND code.
Table 5.
Comparison of intrinsic energy resolution of Siemens gamma cameras for Tc-99m (140 keV)
In addition to intrinsic detector performance, extrinsic energy resolution was also compared with previously reported results obtained using different collimator configurations. Table 6 summarizes the simulated and measured extrinsic energy resolutions for Tc-99 m reported in the present study and in earlier SIMIND-based gamma camera simulations [12, 13]. The comparison includes different collimator types, including LEGP, LEAP, and LEHR (Low Energy High Resolution), which influence photon penetration and scatter within the detector system. Despite these differences, the results remain consistent with manufacturer-specified extrinsic detector performance of approximately 9–10% at 140 keV.
Table 6.
Comparison of extrinsic energy resolution of GE Infinia and Siemens gamma cameras for Tc-99m (140 keV)
Figure 13 shows the comparison between simulated and measured energy resolution values at 140 keV (⁹⁹ᵐTc) for different SPECT systems and collimator configurations reported in this study and previous Monte Carlo validation studies, as shown in Tables 5 and 6. The proximity of the data points to the reference line indicates good agreement between simulated and experimentally measured detector energy resolutions across different studies.
Fig. 13.

Comparison between simulated and measured energy resolution values at 140 keV (⁹⁹ᵐTc) for different SPECT systems and collimator configurations reported in this study and previous Monte Carlo validation studies
Energy spectra: The comparison of simulated and measured energy spectra also confirms the accuracy of the detector model. The proposed peak detection algorithm successfully identified all principal spectral features produced by 99mTc photon interactions in NaI(Tl), including the photopeak (140 keV), Compton edge (~ 47 keV), Pb characteristic X-ray peak (~ 72–75 keV), backscatter peak (~ 90 keV), and iodine escape peak (~ 112 keV). The differences between simulated and measured peak positions were below 3% for both systems, demonstrating good spectral fidelity.
A comparison with previously published simulated spectra of the Siemens E.Cam system reveals that the primary emphasis has been placed on photopeak modeling and detector performance metrics such as sensitivity and spatial resolution [12, 13]. In particular, the backscatter peak (~ 90 keV) and iodine escape peak (~ 112 keV) are not clearly separated in the simulated spectra reported in previous SIMIND-based analyses [12]. In contrast, the present study explicitly identifies and quantitatively characterizes these secondary peaks using the proposed automated peak identification algorithm.
Sensitivity: Sensitivity comparison analyses indicated discrepancies of around 1.1% for Siemens E.Cam and 1.15% for GE Infinia, both of which fall within the bounds of acceptable experimental errors. The sensitivity measure is significantly affected by parameters like collimator geometry and detector material, which were precisely incorporated into the simulation according to manufacturer specifications. The < 2% overall variation between measured and simulated sensitivities validates the accuracy of our detector modeling through the Monte Carlo model while highlighting areas for future refinement in signal processing simulation.
Image evaluation: The analysis of 2D and 3D image profiles for Siemens E.Cam demonstrated a strong agreement between simulation and experimental results, with FWHM values ranging from 4 to 4.5 pixels. This signifies that spatial response attributes, encompassing intrinsic resolution and collimator blur, are accurately represented in the simulation.
However, in the GE Infinia, wider experimental profiles and diminished peak sharpness were noted, indicating inherent constraints of the particular system implemented at the site, including compromised detector uniformity and electronic noise.
Spatial resolution: The Siemens E.Cam system exhibited spatial resolution between 3.5 mm (bar phantom) and 4.8 mm (pie phantom), aligning with the anticipated performance of an adequately operating LEAP-collimated system. In contrast, the GE Infinia system demonstrated markedly worse experimental resolution (~ 1 cm), which may be attributable to detector age or inadequate calibration.
A noticeable difference was observed between the simulated and experimentally measured image profiles for the GE Infinia system. While the SIMIND Monte Carlo simulation produced well-defined point-source profiles consistent with theoretical expectations, the experimentally acquired images exhibited broadened peaks and reduced uniformity.
The discrepancy between the simulated and measured image profiles of GE Infinia is mainly attributed to the actual operating condition of the GE Infinia system installed at our facility, rather than to the SIMIND Monte Carlo model. SIMIND models an ideal and properly calibrated SPECT system based on the defined physical and geometrical parameters. Therefore, if the clinical detector has local hardware, calibration, or uniformity-related problems, these effects may not be fully reproduced unless they are specifically characterized and introduced into the simulation model. In clinical gamma cameras, factors such as detector non-uniformity, photomultiplier tube (PMT) gain variations, electronic noise, and imperfect energy or uniformity calibration can degrade the spatial response function and introduce artifacts such as star patterns or profile broadening. Importantly, despite these spatial differences, the energy spectrum characteristics, energy resolution, and sensitivity measurements for the GE Infinia system remained in close agreement with the Monte Carlo simulations. This confirms that the SIMIND-based detector model accurately reproduces the fundamental photon interaction processes in the detector while also demonstrating the capability of the proposed methodology to identify performance deviations in clinical imaging systems.
Energy spectrum: The self-developed peak detection algorithm effectively recognized and categorized all principal spectral components (photopeak, Compton edge, iodine escape, Pb X-rays, backscatter peaks), exhibiting less than 3% difference between simulated and experimental peak positions for both systems. Furthermore, this also validates accurate modeling of photon interactions in NaI(Tl) crystals by the SIMIND Monte Carlo code.
Conclusion
This study presents a comprehensive Monte Carlo validation of two clinical gamma camera systems (Siemens E.Cam and GE Infinia) using the SIMIND Monte Carlo code. The results demonstrate that SIMIND can accurately reproduce the energy-spectrum characteristics, energy resolution, characteristic peak positions, and sensitivity of both systems under idealized detector-modeling conditions. For image-profile evaluation, good agreement was obtained for Siemens E.Cam, whereas the GE Infinia results revealed a system-specific discrepancy associated with the actual clinical condition of the installed camera.
Furthermore, the integration of the peak detection and classification algorithm for energy-spectrum analysis, together with sensitivity, spatial resolution, and image-profile evaluations, provides a useful methodology for SPECT performance characterization. Importantly, this framework not only validates the agreement between simulation and measurement for well-modeled detector responses but also helps identify deviations between ideal Monte Carlo predictions and the actual performance of clinical gamma camera systems. In addition, a comparison with previously published Monte Carlo validation studies demonstrates that the obtained energy resolution and spectral characteristics are consistent with reported values for clinical NaI(Tl)-based SPECT systems, further confirming the reliability of the proposed simulation and analysis framework. These findings confirm that the simulated system is suitable for further research, encompassing modifications, planning, and the development of a SPECT system to enhance image quality. This validation represents an essential step before the simulation can be used in more complicated studies of optimizing performance characteristics.
Acknowledgements
The authors would like to show their great appreciation to other members of the School of Safety and Management Engineering, School of Computer Science and Engineering, Hunan Institute of Technology, for their support and contribution to this research.
Abbreviations
- MC
Monte Carlo
- SPECT
Single photon emission computed tomography
- FWHM
Full width at half maximum
- PMT
Photomultiplier tube
- NaI(Tl)
Thallium-activated sodium iodide
- RTE
Radiative transport equation
- cps/MBq
Counts per second per megabecquerel
- LEAP
Low energy all purpose
- LEGP
Low energy general purpose
- LEHR
Low energy high resolution
Appendix
A1: Pseudocode for energy spectrum analysis


Author contributions
ZP and MAW designed the study, performed experiments and simulations for the 99mTc source; IS and FW conducted experimental studies for the 57Co flood source and validated the results; MAW and MS developed the peak detection algorithm and analyzed the spectral features. All authors wrote and reviewed the manuscript.
Funding
This work was supported by the Hunan Provincial Natural Science Foundation Project(No.2025JJ70160), Scientific Fund of Hunan Provincial Education Department (No.23A0629), Hunan Institute of Technology Provincial Applied Characteristic Discipline Open Project (Grant No. KF24014), Hunan Provincial Department of Education Key Teaching Reform Project (202401001561), and the College Students’ Innovation and Entrepreneurship Plan Projects of Hunan Province (Grant No. S202411528012, S202411528018S, S202411528108S, S202411528170, S202411528032X).
Data availability
The original contributions presented in this study are included in the article/supplementary material. Further inquiries can be directed to the corresponding author(s).
Declarations
Competing interests
The authors declare no competing interests.
Ethics approval and consent to participate
Not applicable.
Consent for publication
Not applicable.
Footnotes
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Rogers DW. Fifty years of Monte Carlo simulations for medical physics. Phys Med Biol. 2006;51(13):R287-301. 10.1088/0031-9155/51/13/R17. [DOI] [PubMed] [Google Scholar]
- 2.Nelson WR, Hirayama H, Rogers DWO. The EGS4 code system; SLAC report 265. Stanford: Stanford Linear Accelerator Center; 1985. [Google Scholar]
- 3.Hirayama H, Namito Y, Bielajew AF, Wilderman SJ, Nelson WR. The EGS5 code system; SLAC report 730. Stanford: Stanford Linear Accelerator Center; 2005. [Google Scholar]
- 4.Kawrakow I, Rogers DWO. The EGSnrc code system: Monte Carlo simulation of electron and photon transport; NRC report PIRS-701. National Research Council of Canada, Ottawa; 2001.
- 5.Agostinelli S, Allison J, Amako K, Apostolakis J, Araujo H, Arce P, et al. Geant4—a simulation toolkit. Nucl Instrum Methods Phys Res A. 2003;506:250–303. 10.1016/S0168-9002(03)01368-8. [DOI] [Google Scholar]
- 6.Salvat F, Fernández-Varea JM, Baró J, Sempau J. PENELOPE: a code system for Monte Carlo simulation of electron and photon transport. Paris: OECD/NEA; 2001. [Google Scholar]
- 7.Wu Y, Song J, Zheng H, Sun G, Hao L, Long P, et al. CAD-based Monte Carlo program for integrated simulation of nuclear systems. SuperMC Ann Nucl Energy. 2015;82:161–8. 10.1016/j.anucene.2014.08.058. [DOI] [Google Scholar]
- 8.Wasaye MA, Wang H, He P. An algorithm for Monte Carlo simulation of bremsstrahlung emission by electrons. Nucl Sci Tech. 2017;28:71. 10.1007/s41365-017-0218-7. [DOI] [Google Scholar]
- 9.Ljungberg M, Strand SE. A Monte Carlo program for the simulation of scintillation camera characteristics. Comput Methods Programs Biomed. 1989;29(4):257–72. 10.1016/0169-2607(89)90111-9. [DOI] [PubMed] [Google Scholar]
- 10.Jan S, Santin G, Strul D, Staelens S, Assié K, Autret D, et al. GATE: a simulation toolkit for PET and SPECT. Phys Med Biol. 2004;49:4543–61. 10.1088/0031-9155/49/19/007. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.R. L. Harrison, S. D. Vannoy, D. R. Haynor, S. B. Gillispie, M. S. Kaplan and T. K. Lewellen, "Preliminary Experience With The Photon History Generator Module Of A Public-domain Simulation System For Emission Tomography," 1993 IEEE Conference Record Nuclear Science Symposium and Medical Imaging Conference, San Francisco, CA, USA, 1993, pp. 1154–1158, 10.1109/NSSMIC.1993.701828 [DOI]
- 12.Bahreyni Toossi MT, Islamian JP, Momennezhad M, Ljungberg M, Naseri SH. SIMIND Monte Carlo simulation of a single photon emission CT. J Med Phys. 2010;35:42–7. 10.4103/0971-6203.55967. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Morphis M, van Staden JA, du Raan H, Ljungberg M. Validation of a SIMIND Monte Carlo modeled gamma camera for Iodine-123 and Iodine-131 imaging. Heliyon. 2021;7:e07196. 10.1016/j.heliyon.2021.e07196. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Ramonaheng K, van Staden JA, du Raan H. Validation of a Monte Carlo modelled gamma camera for Lutetium-177 imaging. Appl Radiat Isot. 2020;163:109200. 10.1016/j.apradiso.2020.109200. [DOI] [PubMed] [Google Scholar]
- 15.Roshan HR, Mahmoudian B, Gharepapagh E, Azarm A, Islamian JP. Collimator and energy window optimization for ⁹⁰Y bremsstrahlung SPECT imaging: A SIMIND Monte Carlo study. Appl Radiat Isot. 2016;108:124–8. 10.1016/j.apradiso.2015.12.041. [DOI] [PubMed] [Google Scholar]
- 16.Rodrigues S, Tomé B, Abreu MC, Santos N, Mendes PR, Peralta L. Monte Carlo simulation and experimental characterization of a dual-head gamma camera. J Instrum. 2008;3:P05001. 10.48550/arXiv.0711.2577. [DOI] [Google Scholar]
- 17.Ejeh JE, van Staden JA, du Raan H. Validation of SIMIND Monte Carlo simulation software for modelling a siemens symbia T SPECT scintillation camera. In: Lhotska L, Sukupova L, Lacković I, Ibbott GS, editors. World congress on medical physics and biomedical engineering 2018. IFMBE proceedings, vol. 681. Singapore: Springer; 2019. p. 602–6. 10.1007/978-981-10-9035-6_106. [DOI] [Google Scholar]
- 18.Yasumoto Y, Daisaki H, Sato M. Validation of the SIMIND simulation code using the myocardial phantom HL. Biomed Phys Eng Express. 2024;11(1):015029. 10.1088/2057-1976/ad960d. [DOI] [PubMed] [Google Scholar]
- 19.Di Domenico G, Di Biaso S, Longo L, et al. Validation of 99mTc and 177Lu quantification parameters for a Monte Carlo modelled gamma camera. EJNMMI Phys. 2023;10:27. 10.1186/s40658-023-00547-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Pieters H, van Staden JA, du Plessis FCP, du Raan H. Validation of a Monte Carlo simulated cardiac phantom for planar and SPECT studies. Physica Med. 2023;111(102617):102617. 10.1016/j.ejmp.2023.102617. [DOI] [PubMed] [Google Scholar]
- 21.Roth D, Larsson E, Ljungberg M, et al. Monte Carlo modelling of a compact CZT-based gamma camera with application to 177Lu imaging. EJNMMI Phys. 2022;9:35. 10.1186/s40658-022-00463-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Gilmore GR. Computer analysis of gamma-ray spectra. In: Practical gamma-ray spectrometry. 2nd ed. London: Wiley; 2008. p. 183–203. 10.1002/9780470861981.ch9. [DOI] [Google Scholar]
- 23.Burr T, Hamada M. Radio-isotope identification algorithms for NaI γ spectra. Algorithms. 2009;2(1):339–60. 10.3390/a2010339. [DOI] [Google Scholar]
- 24.Likar A, Vidmar T. A peak-search method based on spectrum convolution. J Phys D Appl Phys. 2003;36(15):1903–9. 10.1088/0022-3727/36/15/323. [DOI] [Google Scholar]
- 25.Data sheet of Siemens E.Cam dual-head gamma camera. 1992. p. 1–12.
- 26.Product Data - Release 2.5 Infinia® 3/8” or 1” free-geometry dual-detector cameras optional Hawkeye® Hybrid NM/CT. 2006. p. 1–12.
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
The original contributions presented in this study are included in the article/supplementary material. Further inquiries can be directed to the corresponding author(s).