Abstract
Infants and toddlers are a challenging population upon which to perform magnetic resonance imaging (MRI) of the brain, both in research and clinical settings. Because of the large range in head size during the early years of development, paediatric neuro‐MRI requires a radiofrequency (RF) coil, or set of coils, that is tailored to head size to provide the highest image quality. Mitigating techniques must also be employed to reduce and correct for subject motion. This manuscript describes an RF coil with a tailored mechanical–electrical design that can adapt to the head size of 3‐month‐old infants to 3‐year‐old toddlers. The RF coil was designed with tight‐fitting coil elements to improve the signal‐to‐noise ratio (SNR) in comparison with commercially available adult head coils, while simultaneously aiding in immobilization. The coil was designed without visual obstruction to facilitate an unimpeded view of the child's face and the potential application of camera or motion‐tracking systems. Despite the lack of elements over the face, the paediatric coil produced higher SNR over most of the brain compared with adult coils, including more than twofold in the periphery. Acceleration rates of fourfold in each Cartesian direction could be achieved. High SNR allowed for short acquisition times through accelerated imaging protocols and reduced the probability of motion during a scan. Modification of the acquisition protocol, with immobilization of the head through the adjustable coil geometry, and subsequently being combined with a motion‐tracking system, provides a compelling platform for scanning paediatric populations without sedation and with improved image quality.
Keywords: magnetic resonance imaging, motion correction, neuroimaging, paediatric imaging, parallel imaging, radiofrequency coil
A radiofrequency coil is presented that adapts to the head size of 3‐month‐old infants to 3‐year‐old toddlers. The coil was designed without visual obstruction to facilitate an unimpeded view of the child's face and the potential application of camera or motion‐tracking systems. Modification of the acquisition protocol, with immobilization of the head through the adjustable geometry, and subsequently being combined with a motion‐tracking system, produces a compelling platform for scanning paediatric populations.

Abbreviations
- CAD
computer‐aided design
- EPI
echo‐planar imaging
- FOV
field of view
- IEC
International Electrotechnical Commission
- PIN
positive‐intrinsic‐negative
- RF
radiofrequency
- ROI
region of interest
- SENSE
sensitivity encoding
- SNR
signal‐to‐noise ratio
- TE
echo time
- TR
repetition time
1. INTRODUCTION
Magnetic resonance imaging (MRI) is becoming an increasingly prevalent technique for studying brain structure and function across the lifespan because of its noninvasive nature. This has made MRI particularly beneficial for studying brain development in the paediatric population in both clinical and research settings. 1 , 2
Infant and toddler populations are a challenging cohort to image because of their inability to remain still during an MRI examination. This difficulty is confounded with the relatively few vendor‐supplied and commercial options for radiofrequency (RF) coils tailored to infant and toddler populations compared with those available for adult imaging. When designed with the specific challenges of young paediatric neuroimaging in mind, the RF coil can aid in mitigating motion and improving image quality.
Two of the primary technical challenges when developing an RF coil for paediatric MRI of the brain are: (i) retaining high coil sensitivity (and therefore image signal‐to‐noise ratio [SNR]) over a range of head sizes; and (ii) mitigating head motion. Coil elements must be placed in close proximity to the subject to produce the highest sensitivity; however, this can be difficult to accomplish over the entire age range because of the variance in head size. Coils tailored to the paediatric age group of interest have been developed to address this problem, 3 including an entire suite of paediatric coils optimized for multiple age groups. 4 , 5 To negate the need for multiple coils, size‐adaptable designs have been developed that can provide a tight fit over a range of head sizes. 6 , 7 , 8
Head motion is the second ubiquitous challenge when imaging these age groups. Motion can lead to image blurring, ghosting, and distortions, 9 and is exacerbated in younger paediatric populations because of their predilection for moving when subjected to the stressors of being in an MRI scanner while awake. MRI scans are often acquired during natural sleep; however, infants may still move during sleep or awaken during sequences. In the clinical setting, sedation and anaesthesia are often administered to address motion, albeit with intrinsic risk to the patient.
Even under the influence of anaesthesia, patient motion may persist, with the dominant direction of movement being toward the foot end of the bed 10 ; awake children also demonstrate a significant nodding movement. 10 Tailoring the coil housing to the shape of the participant's head, as well as employing an aforementioned adaptable coil design, can aid in mitigating this motion. These strategies become even more effective when combined with pulse‐sequence optimizations that reduce scan times, 11 such as parallel imaging 12 , 13 and compressed sensing 14 , 15 : these techniques benefit substantially from the increased SNR afforded by dedicated paediatric coils. For younger infants, feeding and swaddling can promote sleep to effectively reduce motion. 16 Nonetheless, residual motion persists—notably for older infants and particularly awake toddlers—that can degrade image quality and produce unusable images. Motion tracking and correction have therefore been introduced to address this residual motion, 17 , 18 , 19 , 20 , 21 with some techniques requiring the ability to view the child's face with a camera. 19 , 20
This article presents a tailored solution for imaging infants and toddlers aged 3 months to 3 years on a 3‐T MRI scanner. The children were scanned, either sleeping or awake, without the use of and potential risks of anaesthesia. A mechanically conforming RF coil was designed that can adapt to the range in head size of the infant and toddler age groups to simultaneously increase SNR and help immobilize the head: this adaptive topology was modelled after the design presented by Ghotra et al. 7 (for infants aged 1–18 months) with important modifications made to accommodate a wider age range. An open‐face configuration was employed to accommodate the potential application of a camera or motion‐tracking system, the consequences of which are evaluated in comparison with conventional adult coils. All computer‐aided design (CAD) files have been provided to add this solution to the limited repertoire of publicly available paediatric coil designs.
2. METHODS
2.1. Radiofrequency coil
2.1.1. Coil mechanicals
The coil housing was designed to accommodate children between the ages of 3 months and 3 years. The coil was split into four disparate sections to adapt to the variability in head size within the age range (Figure 1). The geometry for the inner surface of the housing was based on the average head shape of a 3‐year old (n = 36; provided by the Neurodevelopmental MRI Database 22 , 23 ) and allowed for approximately 6 mm of space for foam cushioning. The arch of the neck was included in the coil housing to allow closer proximity of coils and to create a ‘cup’ for the occiput of the head to provide better immobilization.
FIGURE 1.

(A) Photographs of the 32‐channel paediatric coil. (B) Lateral sections can slide laterally to accommodate varying head width and are fixated afterwards with custom cam locks. The coaxial cables of lateral elements are routed through the housing to the posterior section. Three interchangeable anterior sections were constructed to accommodate the large range in head size over the desired age range (3 months to 3 years). High‐density connectors (ODU‐USA Inc.) connect coaxial and DC cables from the anterior to the posterior sections, precluding the need for a cumbersome plug on the anterior sections. (C) The circuit schematic of a single receive element depicts the active and passive detuning networks. Coil elements are connected directly to low‐input‐impedance preamplifiers. CM, matching capacitor; CT, tuning capacitor; F, fuse.
The large posterior section of the design was stationary, while the left and right sections could slide laterally to create a housing width anywhere between 12.0 and 17.1 cm (excluding space for foam). Once the lateral sections were positioned as close as possible to the subject's head, they could be locked into place using custom‐built cam locks. This served two purposes: (i) to increase coil sensitivity by ensuring the closest possible proximity to the brain; and (ii) to mitigate head motion. This adaptive topology was based on the design presented by Ghotra et al. 7 (for infants aged 1–18 months); however, the older and wider age range desired for this study necessitated design modifications to accommodate a larger variance in the anterior–posterior dimension of the head.
The variance in head size in the anterior–posterior direction over the desired age range was addressed by constructing three interchangeable anterior sections based on the average head sizes of a 3‐month‐old, 1‐year‐old, and 3‐year‐old. 22 , 23 The anterior–posterior dimension of the housing was therefore either 16.5, 18.5, or 20.0 cm (Figure 2). No coil elements or mechanical structure were placed in front of the subject's face, inferior to the brow ridge, to accommodate the use of a camera or motion‐tracking system.
FIGURE 2.

Sagittal and axial cross‐sections of the three coil configurations: for children younger than 3 months of age, aged 3 months to 1 year, and 1–3 years. The representative heads approximate the average size of a 3‐month‐old, 1‐year‐old, and 3‐year‐old, 22 , 23 with dimensions provided along the anterior–posterior and left–right directions. The adjustability of the lateral and anterior sections allows for consistent loading across the age range.
All CAD drawings have been made openly available to facilitate reproducibility: https://doi.org/10.17605/OSF.IO/6EZ4Q.
2.1.2. Coil circuitry
Each receive‐coil configuration (i.e., using one of three possible anterior sections) was comprised of 32 loop elements arranged in a soccer‐ball geometry 24 with adjacent elements within each section overlapped to reduce inductive coupling. 25 Fourteen elements were placed on the posterior section, six elements on the anterior section, and six elements on each of the right and left sections. Elements were tuned to 123.2 MHz and matched to 200 + j50 Ω (i.e., the optimal impedance to minimize the noise figure of preamplifiers) when loaded with a head‐shaped phantom of a 2‐year‐old filled with 50‐mM sodium chloride and water. A parallel‐resonant, active‐detuning circuit with a high‐power PIN diode (Macom MA4P7435NM‐1091 T) was placed near the input of each receive element. An additional PIN diode (biased through an RF choke and resistor) was placed across the preamplifier input to further isolate the preamplifier during transmission. A parallel‐resonant, passive‐detuning circuit 26 was added to each element as a secondary detuning method and to enhance detuning during transmission. As an extra precaution, a fast‐blow fuse (315‐mA limit; Conrac) was added to the daughterboard of the preamplifier. A circuit schematic of a receive element is provided in Figure 1C.
Most preamplifiers were located directly at the input of their respective coil element. Preamplifiers (Siemens; distributed by Stark Contrast, Erlangen, Germany) had a low input impedance to facilitate preamplifier decoupling 25 and an integrated cable trap between the first and second amplification stage. Two elements on each lateral section were connected to their respective preamplifier, located in the posterior section of the housing, through 8.9‐cm‐long flexible coaxial cables. Coaxial cables from the lateral elements were routed to the posterior section through a feed‐through that would move with the lateral housing. Coaxial cables from the anterior sections were routed to the posterior section through high‐density coaxial connectors built into the coil housing (ODU‐MAC product line). From the posterior section, each multiconductor cable was wound into a large choke balun before exiting the housing: this prevented common‐mode currents on the coaxial‐cable grounding shields. Preamplifier decoupling and active detuning of each receive element were measured using standard double‐probe techniques, as described by Keil et al. 4 The coupling between overlapping receive elements, S ij, was measured from the preamplifier inputs.
2.2. Imaging
All imaging was performed at the Centre for Functional and Metabolic Mapping at the University of Western Ontario. MRI data collection was performed on a human, whole‐body Siemens Prisma Fit scanner (Siemens Healthineers AG, Erlangen, Germany) operating with a 3‐T main magnetic field and an XR80/200 gradient coil (maximum gradient strength: 80 mT/m, maximum slew rate: 200 T/m/s). The system is equipped with 128 receiver channels, of which 32 were utilized: plugs were interfaced to Tim coil interface legacy adaptors (part numbers: 14460282 and 14418519) to allow operation on the Siemens Prisma hardware platform.
All human imaging was performed in accordance with the human subject research ethics board (HSREB) at the University of Western Ontario (REB #115440). All imaging was performed with the informed consent of a parent or legal guardian. Rigorous testing of both coil and patient safety was performed before the scanning of human subjects to ensure robust operation and subject safety. Testing followed regulatory guidelines specified by the International Electrotechnical Commission (IEC), 27 , 28 summarised by Hoffmann et al., 29 and described in a previous study by the authors. 30 All children were scanned with hearing protection (ear plugs), while infants also wore ear muffs (MiniMuffs).
2.3. Coil performance
The coil performance was evaluated by assessing image SNR, noise level, noise correlation, and geometry factor. The performance was compared with two vendor‐supplied commercial coils intended for adult imaging (Siemens Healthineers AG): (i) a 20‐channel head and neck coil (routinely employed for paediatric neuroimaging within our facility because of its large dimensions [approximately 26.8 cm {anterior–posterior} by 22.8 cm {left–right}]) that can accommodate an infant's shoulders; and (ii) a 32‐channel head‐only coil (approximate dimensions: 23.0 cm [anterior–posterior] by 19.6 cm [left–right]), chosen because it had the same number of coil elements as the paediatric coil.
Performance metrics were acquired with a phantom that approximated the size of a 2‐year‐old's head and was filled with 50‐mM sodium chloride and water. The largest anterior section was required to accommodate the head phantom. Image SNR maps were derived from gradient‐echo images, with and without RF transmission, using a sum‐of‐squares reconstruction 25 , 31 (matrix size: 96 × 96, field of view [FOV]: 192 × 192 mm, number of slices: 96, slice thickness: 2 mm, TE/TR: 5/875 ms, flip angle: 20°, bandwidth [BW]: 250 Hz/pixel, number of averages: 4). The noise‐only acquisition was additionally used to calculate the noise correlation matrix and the normalized noise level of each channel. Geometry‐factor maps were created by retrospectively undersampling the fully sampled k‐space data and reconstructing using sensitivity encoding (SENSE). 13 All calculations were performed in Matlab (MathWorks, Natick, MA, USA).
To assess the benefits of having adjustable lateral sections and multiple anterior sections, the SNR was measured of a phantom that approximated the head size of a 3‐month‐old. The SNR was compared when the coil was placed in its largest configuration (i.e., lateral sections as wide as possible and using the largest anterior section) versus when in its smallest configuration (i.e., lateral elements placed tight to the phantom and using the smallest anterior section).
The abovementioned SNR comparisons were performed in Matlab using region of interest (ROI) analyses. The mean SNR was calculated within 2D ROIs in central, frontal, peripheral, lateral, and superior regions. SNR ratios were then computed for each ROI.
Image quality was assessed by acquiring T1‐weighted, T2‐weighted, and echo‐planar images of a 3‐month‐old infant and a 1‐year‐old infant. Where acquisition parameters were altered between the two subjects, parameters for the 1‐year‐old have been provided in brackets:
T1‐weighted, gradient‐recalled echo. Matrix size: 192 × 192, FOV: 192 × 192 mm, number of slices: 112 (144), slice thickness: 1 mm, TE/TR: 6.1/23 ms (6.3/20 ms), flip angle: 19°, BW: 120 Hz/pixel, partial Fourier: 6/8. For the 1‐year‐old infant, the acquisition was accelerated with a GRAPPA 12 factor of twofold along the anterior–posterior direction.
T2‐weighted, turbo‐spin‐echo. Matrix size: 192 × 156, FOV: 192 × 156 mm, number of slices: 115 (128), slice thickness: 1 mm, TE/TR: 140/6,150 ms (140/7500 ms), refocusing flip angle: 120°, echo‐train length: 19, BW: 190 Hz/pixel, GRAPPA acceleration rate: 2 (left–right).
Echo‐planar image. Matrix: 78 × 78, FOV: 192 × 192 mm, number of slices: 40 (44), slice thickness: 2.5 mm, TE/TR: 45/1000 ms, BW: 1490 Hz/pixel (1645 Hz/pixel), flip angle: 40°, partial Fourier: 6/8. For the 1‐year‐old infant, the acquisition was accelerated with a multiband 32 factor of fourfold along the superior–inferior (slice) direction.
3. RESULTS
3.1. Coil performance
Active detuning provided a minimum isolation of 34 dB during transmission. The reference voltage required by the scanner's body transmit coil, measured with and without the receive coil present, differed by less than 2.5% (for all three coil configurations; i.e., different anterior sections). Actual flip‐angle maps showed no regions of increased flip angle because of the presence of the receive coil.
RF and gradient heating tests produced negligible changes in the temperature of the coil housing and components, with less than a 3% reduction in image SNR postheating.
The mean geometric decoupling between adjacent elements was −16 dB (worst case: −10 dB). Preamplifier decoupling had a mean of −26 dB across all elements (worst case: −20 dB). The mean and maximum noise correlation was 9.5% and 59%, respectively (Figure 3). By contrast, the adult head/neck and head‐only coils had higher mean noise correlations (17% and 25%, respectively) due to being insufficiently loaded by the ‘2‐year‐old’ phantom. The normalized noise level of individual channels ranged from 0.76‐ to 1.46‐fold, compared with from 0.81‐ to 1.29‐fold and from 0.71‐ to 1.67‐fold for the adult head/neck and head‐only coils, respectively.
FIGURE 3.

Noise‐correlation matrices of the paediatric coil, 20‐channel adult head/neck coil, and 32‐channel adult head‐only coil. The paediatric coil produces a lower mean noise correlation (9.5% compared with 17%/25% for the adult head/neck and head‐only coils, respectively), due to being optimized for a smaller load.
The paediatric coil produced higher SNR throughout the majority of the brain relative to the adult coils, in large part because of the smaller dimensions of the paediatric coil (Figure 4). The relative improvement between the coils was spatially varying (see Table 1), with the largest gains (> twofold) in the periphery (cortex) of the brain. Superior, frontal, and lateral regions demonstrated relative improvements from 1.14‐ to 1.98‐fold. In the paediatric coil, there were no coil elements placed in front of the face (unlike for the larger adult head/neck coil): this design topology was adopted to allow a motion‐tracking camera to view the entire face, an otherwise difficult task to accomplish when using a small, tight‐fitting coil; however, this also resulted in lower SNR in the centre of the brain (0.73‐fold). By contrast, when compared with the adult head coil—which has fewer receive elements above the face—the paediatric coil produced a 1.04‐fold SNR improvement in the central region.
FIGURE 4.

(A) Photographs of the ‘2‐year‐old’ head phantom within the respective coils depicts the realistic positioning of a 2‐year‐old subject. (B) The image SNR of the head phantom, as acquired with the paediatric coil, adult head/neck coil, and adult head‐only coils, shows spatially varying SNR, mainly attributable to the difference in coil size. (C) The resultant ratio between image SNR attained with the paediatric coil versus the adult coils demonstrates large increases in the periphery. The inset provides the ROIs over which the mean SNR was calculated (see Table 1). ROI, region of interest; SNR, signal‐to‐noise ratio.
TABLE 1.
SNR ratio between the paediatric coil and the adult head/neck and head‐only coils for the ROIs illustrated in Figure 4.
| SNRpaediatric/SNRadult | ||
|---|---|---|
| ROI | Adult head/neck coil | Adult head coil |
| central | 0.73 | 1.04 |
| frontal | 1.14 | 1.40 |
| peripheral | 2.00 | 2.32 |
| lateral | 1.34 | 1.58 |
| superior | 1.89 | 1.98 |
Abbreviations: ROI, region of interest; SNR, signal‐to‐noise ratio.
The SNR of a ‘3‐month‐old’ head phantom when the coil was in its tightest configuration versus when in its largest configuration showed a 1.71‐fold increase in the frontal region, a 1.69‐fold increase in the lateral regions, and a 1.15‐fold increase in the centre of the head (Figure 5). Negligible differences were present in the posterior region of the coil, as the posterior section is nonadjustable.
FIGURE 5.

(A) SNR of a ‘3‐month‐old’ head phantom when in the smallest coil configuration and when in the largest coil configuration. (B) The SNR ratio between the two configurations shows a 1.71‐fold increase in the frontal region, a 1.69‐fold increase in the lateral regions, and a 1.15‐fold increase in the central region when employing the tightest coil configuration. The inset provides the ROIs over which SNR ratios were calculated. config, configuration; ROI, region of interest; SNR, signal‐to‐noise ratio.
The paediatric coil produced substantially lower geometry factors across all acceleration rates compared with the adult coils (Figure 6). The paediatric coil could achieve acceleration rates of fourfold along each Cartesian direction before a substantial degradation in accelerated SNR would occur, whereas the two adult coils could only achieve threefold acceleration rates with comparable geometry factors. For the paediatric coil, acceleration along the left–right axis produced the lowest geometry factors for acceleration rates up to fivefold.
FIGURE 6.

Inverse geometry factor maps of the ‘2‐year‐old’ head phantom obtained with the 32‐channel paediatric coil, 20‐channel adult head/neck coil, and 32‐channel adult head‐only coil with acceleration rates of three, four, and five along the superior–inferior, left–right, and anterior–posterior directions. The paediatric coil utilized its largest anterior section to accommodate the head phantom. The mean geometry factor is provided beneath each respective map. The paediatric coil demonstrates markedly lower geometry factors across all acceleration rates due to its smaller geometry and therefore more spatially independent receive profiles.
T1‐weighted, T2‐weighted, and echo‐planar images of a 3‐month‐old infant and a 1‐year‐old infant are provided in Figure 7. Although motion between participants in this age range is highly dependent on their ability to sleep during the scan, no motion artifacts (ghosting or blurring) were present in these images.
FIGURE 7.

Representative axial slices of T1‐weighted, T2‐weighted, and echo‐planar images (EPI) of a 3‐month‐old infant and a 1‐year‐old infant. No motion artifacts (ghosting or blurring) are visible in the images.
4. DISCUSSION
The paediatric coil was designed to facilitate the use of a camera or motion‐tracking system that requires a clear view of the subject's face while inside the bore of the MRI scanner. 19 This was accomplished by eliminating elements and housing that would otherwise be in front of the face (below the brow ridge). When compared with two vendor‐provided adult coils, the paediatric coil produced superior SNR throughout most of the head: this can be primarily attributed to the smaller coil elements and the proximity of these coils to the head through an adjustable design. Despite the large SNR gains in the periphery of the brain (greater than twofold), a relatively small decrease in SNR was observed in the deeper regions of the head relative to the adult head/neck coil. This was in part due to the lack of coil elements in front of the face; however, this was deemed an acceptable compromise to facilitate the necessary monitoring of the child using cameras and/or motion‐tracking systems that could alleviate large decreases in SNR caused by subject motion. The SNR ratio to the adult head‐only coil was generally consistent with gains reported by Ghotra et al. 7 when compared with a dedicated neonatal coil (for children aged 1–18 months).
The SNR comparison in this study was performed on a phantom equivalent to the size of a 2‐year‐old's head, to determine the lower bound in potential SNR gains of the paediatric coil. Because the paediatric coil is adjustable in size, SNR gains are expected to be even higher when imaging smaller head sizes compared with the fixed geometry of conventional adult coils.
Scans of the ‘3‐month‐old’ head phantom in the coil's tightest configuration produced a substantial improvement in SNR in the frontal region (1.71‐fold) and lateral regions (1.69‐fold) compared with scanning in the coil's largest configuration. This demonstrates the importance of being able to substantially vary the coil size in both the left–right and anterior–posterior dimensions, not just in comparison with an adult but over the 3‐month (infant) to 3‐year (toddler) age range.
The paediatric coil produced superior geometry factors compared with the adult coils, despite the open‐face design. This is due to the closer proximity of its coil elements and therefore the higher spatial independence of the commensurate sensitivity profiles. When combined with high SNR (as described above), the result is the ability to employ high acceleration rates to reduce the scanning time, thus reducing the probability of motion occurring during the acquisition.
Adaptable coil designs have an additional benefit of requiring only a single coil to provide optimal performance over a given age range. The adaptive design described in this study was modelled after the design presented by Ghotra et al., 7 with important modifications to accommodate a wider and older age range (infant to toddler). The publicly available CAD files can now be easily adopted and adapted to promote the development of paediatric‐dedicated coils.
Modification of the acquisition protocol, with immobilization of the head through the adjustable coil geometry, and subsequently being combined with a motion‐tracking system, produces a compelling platform for scanning paediatric populations without sedation and with improved image quality.
ACKNOWLEDGEMENTS
The authors would like to thank Peter Zeman for fabrication assistance. This work was supported by the Canada Foundation for Innovation; Canada First Research Excellence Fund to BrainsCAN; Brain Canada Platform Support Grant; and the Molly Towell Perinatal Research Foundation.
REFERENCES
- 1. Copeland A, Silver E, Korja R, et al. Infant and child MRI: A review of scanning procedures. Front Neurosci. 2021;15:666020. doi: 10.3389/fnins.2021.666020 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2. Lenroot RK, Giedd JN. Brain development in children and adolescents: insights from anatomical magnetic resonance imaging. Neurosci Biobehav Rev. 2006;30(6):718‐729. doi: 10.1016/j.neubiorev.2006.06.001 [DOI] [PubMed] [Google Scholar]
- 3. Hughes EJ, Winchman T, Padormo F, et al. A dedicated neonatal brain imaging system. Magn Reson Med. 2017;78(2):794‐804. doi: 10.1002/mrm.26462 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4. Keil B, Alagappan V, Mareyam A, et al. Size‐optimized 32‐channel brain arrays for 3 T pediatric imaging. Magn Reson Med. 2011;66(6):1777‐1787. doi: 10.1002/mrm.22961 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5. Mareyam A, Xu D, Polimeni JR, et al. Brain arrays for neonatal and premature neonatal imaging at 3T. In: Proceedings of the 21st Annual Meeting of ISMRM; 2013:4386. [Google Scholar]
- 6. Lopez Rios N, Foias A, Lodygensky G, Dehaes M, Cohen‐Adad J. Size‐adaptable 13‐channel receive array for brain MRI in human neonates at 3 T. NMR Biomed. 2018;31(8):e3944. doi: 10.1002/nbm.3944 [DOI] [PubMed] [Google Scholar]
- 7. Ghotra A, Kosakowski HL, Takahashi A, et al. A size‐adaptive 32‐channel array coil for awake infant neuroimaging at 3 Tesla MRI. Magn Reson Med. 2021;86(3):1773‐1785. doi: 10.1002/mrm.28791 [DOI] [PubMed] [Google Scholar]
- 8. Banerjee M, Follante CK, Zemskov A, et al. Optimized Pediatric Suite with head array adjustable for patients 0–5 yrs of age. In: Proceedings of the 22nd Annual Meeting of ISMRM; 2014:1768. [Google Scholar]
- 9. Zaitsev M, Maclaren J, Herbst M. Motion artifacts in MRI: A complex problem with many partial solutions. J Magn Reson Imaging. 2015;42(4):887‐901. doi: 10.1002/jmri.24850 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10. Eichhorn H, Vascan AV, Nørgaard M, et al. Characterisation of children's head motion for Magnetic Resonance Imaging with and without general anaesthesia. PsyArXiv. Published online June. 2021. doi: 10.31234/osf.io/qvfkb [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11. Jaimes C, Gee MS. Strategies to minimize sedation in pediatric body magnetic resonance imaging. Pediatr Radiol. 2016;46(6):916‐927. doi: 10.1007/s00247-016-3613-z [DOI] [PubMed] [Google Scholar]
- 12. Griswold MA, Jakob PM, Heidemann RM, et al. Generalized autocalibrating partially parallel acquisitions (GRAPPA). Magn Reson Med. 2002;47(6):1202‐1210. doi: 10.1002/mrm.10171 [DOI] [PubMed] [Google Scholar]
- 13. Pruessmann KP, Weiger M, Scheidegger MB, Boesiger P. SENSE: sensitivity encoding for fast MRI. Magn Reson Med. 1999;42(5):952‐962. doi: 10.1002/(SICI)1522-2594(199911)42:5<952::AID-MRM16>3.0.CO;2-S [DOI] [PubMed] [Google Scholar]
- 14. Lustig M, Donoho D, Pauly JM. Sparse MRI: The application of compressed sensing for rapid MR imaging. Magn Reson Med. 2007;58(6):1182‐1195. doi: 10.1002/mrm.21391 [DOI] [PubMed] [Google Scholar]
- 15. Vasanawala SS, Alley MT, Hargreaves BA, Barth RA, Pauly JM, Lustig M. Improved pediatric MR imaging with compressed sensing. Radiology. 2010;256(2):607‐616. doi: 10.1148/radiol.10091218 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16. Antonov NK, Ruzal‐Shapiro CB, Morel KD, et al. Feed and wrap MRI technique in infants. Clin Pediatr. 2017;56(12):1095‐1103. doi: 10.1177/0009922816677806 [DOI] [PubMed] [Google Scholar]
- 17. van der Kouwe AJW, Benner T, Dale AM. Real‐time rigid body motion correction and shimming using cloverleaf navigators. Magn Reson Med. 2006;56(5):1019‐1032. doi: 10.1002/mrm.21038 [DOI] [PubMed] [Google Scholar]
- 18. Zaitsev M, Dold C, Sakas G, Hennig J, Speck O. Magnetic resonance imaging of freely moving objects: prospective real‐time motion correction using an external optical motion tracking system. Neuroimage. 2006;31(3):1038‐1050. doi: 10.1016/j.neuroimage.2006.01.039 [DOI] [PubMed] [Google Scholar]
- 19. Slipsager JM, Ellegaard AH, Glimberg SL, et al. Markerless motion tracking and correction for PET, MRI, and simultaneous PET/MRI. PLoS ONE. 2019;14(4):e0215524. doi: 10.1371/journal.pone.0215524 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20. Frost R, Wighton P, Karahanoğlu FI, et al. Markerless high‐frequency prospective motion correction for neuroanatomical MRI. Magn Reson Med. 2019;82(1):126‐144. doi: 10.1002/mrm.27705 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21. Slipsager JM, Glimberg SL, Højgaard L, et al. Comparison of prospective and retrospective motion correction in 3D‐encoded neuroanatomical MRI. Magn Reson Med. 2022;87(2):629‐645. doi: 10.1002/mrm.28991 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22. Richards JE, Sanchez C, Phillips‐Meek M, Xie W. A database of age‐appropriate average MRI templates. Neuroimage. 2016;124(Pt B):1254‐1259. doi: 10.1016/j.neuroimage.2015.04.055 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23. Richards JE, Xie W. Brains for all the ages: structural neurodevelopment in infants and children from a life‐span perspective. Adv Child Dev Behav. 2015;48:1‐52. doi: 10.1016/bs.acdb.2014.11.001 [DOI] [PubMed] [Google Scholar]
- 24. Wiggins GC, Triantafyllou C, Potthast A, Reykowski A, Nittka M, Wald LL. 32‐channel 3 Tesla receive‐only phased‐array head coil with soccer‐ball element geometry. Magn Reson Med. 2006;56(1):216‐223. doi: 10.1002/mrm.20925 [DOI] [PubMed] [Google Scholar]
- 25. Roemer PB, Edelstein WA, Hayes CE, Souza SP, Mueller OM. The NMR phased array. Magn Reson Med. 1990;16(2):192‐225. doi: 10.1002/mrm.1910160203 [DOI] [PubMed] [Google Scholar]
- 26. Burl M, Zou MX. Transmit mode coil detuning for MRI systems. US Patent 6,850,067. February 1, 2005. [Google Scholar]
- 27. International Electrotechnical Commission . Medical Electrical Equipment – Part 1: General Requirements for Basic Safety and Essential Performance Amendment 1. International Electrotechnical Commission; 2012. [Google Scholar]
- 28. International Electrotechnical Commission . Medical Electrical Equipment – Part 2–33: Particular Requirements for the Basic Safety and Essential Performance of Magnetic Resonance Equipment for Medical Diagnosis. Edition 3.1 ed. International Electrotechnical Commission; 2013. [Google Scholar]
- 29. Hoffmann J, Henning A, Giapitzakis IA, et al. Safety testing and operational procedures for self‐developed radiofrequency coils. NMR Biomed. 2016;29(9):1131‐1144. doi: 10.1002/nbm.3290 [DOI] [PubMed] [Google Scholar]
- 30. Gilbert KM, Klassen LM, Mashkovtsev A, Zeman P, Menon RS, Gati JS. Radiofrequency coil for routine ultra‐high‐field imaging with an unobstructed visual field. NMR Biomed. 2021;34(3):e4457. doi: 10.1002/nbm.4457 [DOI] [PubMed] [Google Scholar]
- 31. Kellman P, McVeigh ER. Image reconstruction in SNR units: a general method for SNR measurement. Magn Reson Med. 2005;54(6):1439‐1447. doi: 10.1002/mrm.20713 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32. Moeller S, Yacoub E, Olman CA, et al. Multiband multislice GE‐EPI at 7 Tesla, with 16‐fold acceleration using partial parallel imaging with application to high spatial and temporal whole‐brain fMRI. Magn Reson Med. 2010;63(5):1144‐1153. doi: 10.1002/mrm.22361 [DOI] [PMC free article] [PubMed] [Google Scholar]
