Abstract
Background
Superparamagnetic iron oxide nanoparticles (SPIONs), such as ferumoxytol, are promising theranostic agents that can be assessed with MRI. Relaxation time mapping can provide reproducible and quantitative biomarkers of SPION distribution, suitable for longitudinal and cross‐individual studies. However, conventional approaches suffer from strong susceptibility artifacts, long echo times (TE), and prolonged scan times, which limit the accurate quantification of SPION biodistribution.
Purpose
To address the limitations of conventional approaches, this study aimed to develop a fast, B1 +‐corrected mapping protocol based on PETALUTE (a 3D dual‐echo ultrashort echo time MRI sequence with a petal‐like rosette k‐space trajectory) using a variable flip‐angle (VFA) acquisition for T1 mapping to assess ferumoxytol distribution.
Methods
Agarose phantoms containing 0–5000 µg/mL ferumoxytol were scanned on a preclinical 7T MRI system using PETALUTE and RARE‐VTR (rapid acquisition with relaxation enhancement and variable repetition time). PETALUTE T1 maps were computed from two VFA acquisitions (4° and 20°). Mean R1 values were correlated with ferumoxytol concentration to assess the method's reliability. For in vivo feasibility testing, mice bearing 4T1 mammary tumors and flank tumors were assigned to the control (n = 1) or ferumoxytol‐injected group (n = 2; 40 mg/kg i.v.). Abdominal MRI scans were performed 24 h post‐injection with both PETALUTE and RARE‐VTR. Regions of interest in the thigh muscle, mammary tumors, and flank tumors were analyzed to compare the estimated T1 and R1 values obtained with both methods.
Results
In the phantom study, PETALUTE preserved positive contrast for ferumoxytol at all concentrations, whereas the conventional RARE‐VTR sequence exhibited signal loss and hypointensity. For PETALUTE, there was a significant linear correlation between R1 and ferumoxytol concentration (R = 0.975, p < 0.01), while RARE‐VTR showed no significant correlation (R = 0.672, p = 0.144). In vivo, PETALUTE provided high‐resolution, non‐gated, whole‐abdominal images with short acquisition times (4 min 19 s). In ferumoxytol‐injected mice, flank tumors exhibited T1 shortening, consistent with the expected iron accumulation. The dual‐echo capabilities of PETALUTE facilitated observation of elevated T1 with preserved T2*‐weighted signal in one of the mammary tumors.
Conclusions
The proposed PETALUTE‐based T1 mapping enables fast and positive‐contrast ferumoxytol imaging with higher spatial coverage and more stable measurements across a wider concentration range than conventional RARE‐VTR T1 mapping.
Keywords: ferumoxytol, T1 mapping, UTE
1. INTRODUCTION
Nanoparticles (NPs) are a powerful platform for precision cancer therapy, enhancing the therapeutic index by concentrating agents in tumors while minimizing off‐target exposure. 1 , 2 In practice, NP uptake varies widely across patients and tumor types. 3 , 4 This variability makes the effective dose unpredictable, complicating treatment planning and clinical translation. Theranostic NPs, which combine therapeutic and imaging functions in a single platform, have therefore been used for the simultaneous treatment and visualization of NP delivery. 5 , 6 , 7 However, a noninvasive imaging modality is still required to confirm target engagement and enable personalized, response‐adaptive dosing strategies. 4
Magnetic resonance imaging (MRI) is well‐suited for theranostic NP tracking. It is non‐ionizing, widely available, and offers high spatial resolution with excellent soft‐tissue contrast. Superparamagnetic iron oxide nanoparticles (SPIONs), such as ferumoxytol, are MRI‐visible theranostic NPs owing to their intrinsic physical properties. 8 They shorten both the transverse (T2/T2*) and longitudinal relaxation times (T1) 9 , 10 , 11 appearing brighter on T1‐weighted MRI images and darker on T2/T2*‐weighted MRI images.
However, MRI signal intensity alone is not a reliable quantitative indicator of SPION, as it is inherently arbitrary. Quantitative MRI (qMRI) is a more reliable approach that offers reproducible measurements grounded in physics models of the MR signal, derived from multiple images acquired with varied parameters to capture tissue‐specific properties. 12 In particular, T1 and T2/T2* mapping yield voxel‐wise parametric maps that encode quantitative relaxation values, reflecting tissue composition and the local magnetic environment. 13 , 14 , 15 , 16 These relaxation mapping techniques have also been investigated for assessing SPION biodistribution 9 , 17 , 18 , 19 , 20 , 21 although several challenges remain in their application to SPION imaging.
T2/T2* mapping is sensitive to SPIONs, but its accuracy and efficiency decline as concentration increases. 9 At high concentrations, strong susceptibility artifacts cause the blooming effect and signal voids, degrading spatial localization and measurement accuracy. 20 In contrast, T1 mapping is more robust to susceptibility artifacts and can offer a higher signal‐to‐noise ratio (SNR) than T2 mapping, 20 but it still has essential limitations in SPION imaging. First, conventional T1 mapping methods, such as Inversion Recovery Spin‐Echo (IR‐SE) and Rapid Acquisition with Relaxation Enhancement with Variable TR (RARE‐VTR), utilize echo times (TE) longer than the ultrashort T2 of SPION‐rich regions, resulting in T2‐related signal loss and inaccurate T1 calculations. In addition, a long repetition time (TR) is usually required to achieve complete relaxation for T1 mapping, increasing susceptibility to motion artifacts and hindering clinical feasibility.
Ultra‐short echo time (UTE) MRI is a promising technique to address the challenges described above. 21 , 22 It can capture rapidly decaying signals and provide positive T1 contrast for SPIONs. 23 UTE can also be combined with a variable‐flip‐angle (VFA) approach to enable rapid T1 mapping for SPIONs assessment. 24 However, the UTE MRI signal can be challenging to interpret in vivo because the short‐T2* signal from SPION is often masked by the long‐T2* signal from water and fat (e.g., necrotic tissue) 22 , 25 or confounded by other short‐T2* components such as calcification or hemorrhagic byproducts. 26 Dual‐echo UTE methods have been developed to separate short‐ and long‐T2* components, but, to our knowledge, their applications have mostly been tissue‐specific rather than focused on SPION tracking. 27 , 28 A dual‐echo UTE T1 mapping approach needs to be adapted for SPION monitoring.
To achieve this, we propose a T1 mapping protocol based on PETALUTE, a 3D dual‐echo UTE MRI sequence with a rosette k‐space trajectory, in which the sampling path traces repeated petal‐like loops in k‐space (Figure 1). 29 , 30 Compared with conventional UTE‐VFA methods, PETALUTE combines dual‐echo acquisition with efficient volumetric sampling, enabling collection of both ultrashort‐TE and longer‐TE signals within a single acquisition. This combination is particularly advantageous for SPION imaging because it improves sensitivity to SPION‐related signal while providing longer‐TE information that may help distinguish SPION‐related signal from long‐T2* components. In addition, the rosette k‐space trajectory supports high spatial resolution with broad volumetric coverage, enhancing the practicality.
FIGURE 1.

3D rosette k‐space trajectory used in PETALUTE. The sampling path traces petal‐like loops in k‐space, repeatedly passing through the center while covering peripheral regions in multiple directions. Blue and red curves indicate the first and second echoes, respectively.
In this study, we developed and evaluated a PETALUTE T1 mapping protocol that combines VFA acquisition 31 and B1 + map correction 32 for ferumoxytol assessment in vitro and in vivo. We hypothesize that PETALUTE will maintain a positive contrast of ferumoxytol while reducing susceptibility artifacts, even at high concentrations, and will enable acquisition of two contrasts with different TEs. To test this hypothesis, we compared PETALUTE T1 maps with conventional RARE‐VTR T1 maps across a range of ferumoxytol concentrations in phantoms. In addition, we assessed the in vivo feasibility of using PETALUTE T1 maps for evaluating ferumoxytol in a dual‐tumor mouse model.
2. MATERIALS AND METHODS
2.1. T1 map calculation
PETALUTE T1 mapping workflow is illustrated in Figure 2. We first generated apparent T1 maps using PETALUTE first echoes from the VFA acquisitions (4°, i.e., the Ernst angle, and 20°) using Equation 1, based on the spoiled gradient‐echo signal equation. 31 , 33
| (1) |
T1, app, represents the apparent T1 value; and S1, S2 are signal intensities from PETALUTE first echoes acquired with flip angles of 4° (ɑ1) and 20° (ɑ2), respectively.
| (2) |
FIGURE 2.

PETALUTE T1 mapping workflow. (a) PETALUTE image reconstructed using nonuniform fast Fourier transform method. (b) B1+ map generated by the variable flip angle method with short‐TR correction and smoothed using a Gaussian filter with σ = 3; color bar range from 0.5 to 0.8. (c) Apparent T1 map generated using the two flip‐angle PETALUTE images; color bar range from 0 to 6000 ms. (d) PETALUTE T1 map derived from the apparent T1 map corrected using the B1+ map; color bar range from 0 to 6000 ms.
Under a fully relaxed condition (TR ≥5T1), the ratio between the two signal intensities, r, can be expressed as in Equation 2, 34 where ɑ1, act was the actual flip angle.
The PETALUTE sequence, however, uses a very short TR that prevents complete longitudinal relaxation. Therefore, a short‐TR correction was applied to estimate the fully relaxed signals. 32
The actual flip angle ɑ1, act was then derived from Equation 3:
| (3) |
| (4) |
The corresponding voxel‐wise B1 + map was then computed as the ratio of the actual flip angle to nominal flip angle (Equation 4).
Finally, the B1 + map was applied to correct the apparent T1 map, producing the PETALUTE T1 map (Equation 5).
| (5) |
| (6) |
The RARE‐VTR T1 maps were computed by voxel‐wise nonlinear least‐squares fitting of a mono‐exponential recovery model (Equation 6) to the multi‐TR data.
2.2. Imaging protocol and reconstruction
Imaging protocol included vendor‐provided RARE‐VTR, Multi‐Echo Gradient‐Echo (MGE) and the proposed PETALUTE sequences. For the in vivo feasibility study, the MGE sequence was included to provide a reference T2* weighted image for assessment of tissue contrast and quantitative T2* analysis was not performed in this study. All imaging protocol parameters are summarized in Table 1. For PETALUTE, the reported acquisition time corresponds to a single flip‐angle acquisition. Since two flip angles were acquired for T1 estimation, the total acquisition time for one PETALUTE T1 map was 2 min 10 s in the phantom study and 8 min 38 s in the in vivo feasibility study. For RARE‐VTR and MGE, the reported acquisition time corresponds to the full multi‐TR protocol used for T1 estimation. All scans were performed on a 7T horizontal‐bore small animal MRI system (BioSpec 70/30; Bruker Instruments; maximum gradient strength: 660 mT/m; maximum slew rate: 4570 T/m/s).
TABLE 1.
Image protocol parameters for RARE‐VTR, MGE, and PETALUTE sequences used in phantom and in vivo acquisitions.
| RARE‐VTR a | MGE b | PETALUTE c | |
|---|---|---|---|
| Resolution | 0.234 × 0.234 × 1 mm3 | 0.234 × 0.234 × 1 mm3 | 0.2 × 0.2 × 0.2 mm3 |
| FOV d |
30 × 30 mm3 20 slices (thickness = 1 mm) |
30 × 30 mm3 20 slices (thickness = 1 mm) |
40 × 40 × 40 mm3 (Phantom); 60 × 60 × 60 mm3 (In‐vivo) |
| TE e | 20.3 ms | 2 ms, 6 ms, 10 ms, 14 ms, 18 ms, 22 ms, 26 ms, 30 ms, 34 ms, 38 ms |
0.016 ms (first echo) 3.2 ms (second echo) |
| TR f | 934.58 ms, 1788.44 ms, 2986.55 ms, 5011.78 ms, 15000.0 ms | 1500 ms | 7 ms |
| Flip angle | 90° | 30° | 4°, 20° |
| Acquisition time | 6 min 11 s | 6 min 24 s |
1 min 5 s/flip‐angle (Phantom) 4 min 19 s/flip‐angle (In‐vivo) |
RARE‐VTR = rapid acquisition with relaxation enhancement–variable TR.
MGE = multi‐gradient echo.
PETALUTE = 3D dual‐echo ultrashort echo time.
FOV = field of view.
TE = echo time.
TR = repetition time.
RARE‐VTR and MGE images were reconstructed on‐scanner automatically using Bruker ParaVision 6.0.1. PETALUTE data were processed in MATLAB (202Xa, MathWorks, USA). The nonuniform fast Fourier transform (NUFFT) method was used to reconstruct the PETALUTE images using BART (Berkeley Advanced Reconstruction Toolbox; nufft for NUFFT reconstruction, ecalib for ESPIRiT coil‐sensitivity estimation, and pics for iterative CS‐SENSE) and MIRT (Michigan Image Reconstruction Toolbox; Gmri and ir_mri_density_comp to define the non‐Cartesian encoding operator and density‐compensation weights).
2.3. Phantom study
Seven phantoms at nominal ferumoxytol concentrations of 0, 100, 250, 750, 1000, 1250, and 5000 µg/mL were prepared using 4.2% agarose (w/v) and ferumoxytol (Feraheme, AMAG Pharmaceuticals). Ferumoxytol was mixed into the heated agarose solution under continuous stirring. The ferumoxytol‐agarose mixture was allowed to cool at room temperature until gelation. After solidification, a small piece of the ferumoxytol‐containing agarose was removed and coated with a thin layer of water‐resistant polyurethane adhesive to maintain its integrity. This piece was then embedded in a 1 mL microcentrifuge tube prefilled with 4.2% agarose (w/v) to form the final phantom. This design produced a compact phantom that remained compatible with the size constraints of the MRI coil, while the ferumoxytol‐containing agarose exhibited positive contrast relative to the surrounding agarose. After cooling to room temperature, all the phantoms were stored in a refrigerated environment. They were allowed to warm at room temperature for two hours before scanning.
All phantoms were scanned using a 23‐mm‐diameter Tx/Rx 1H RF volume coil (Bruker BioSpin, Ettlingen, Germany). PETALUTE images were acquired with a total of 9,216 petals, each petal containing 206 sampling points with 103 sampling points per echo. RARE‐VTR and PETALUTE T1 maps were calculated as previously described. Regions of interest (ROIs) were defined in 3D Slicer 35 around the ferumoxytol (Figure 3). A fixed ROI size of 45 voxels (0.36 mm3) was used for all phantoms to maintain consistency. For each concentration, the mean and standard deviation of T1 and R1 were calculated within the ROI.
FIGURE 3.

Reconstructed images of phantoms using RARE‐VTR and PETALUTE.
Phantoms containing various concentrations of ferumoxytol (0 – 5000 ppm) appear as bright regions in PETALUTE first echo (flip angle = 4°) and dark regions in PETALUTE second echo (flip angle = 4°) and RARE‐VTR images (TR = 934.58 ms). The ROI is outlined in blue.
The relaxation rate R1, defined as the reciprocal of T1, has an approximately linear relationship with the concentration of ferumoxytol, as described in Equation 7. 19
| (7) |
where C is the ferumoxytol concentration (µg/mL), R1, 0 is the intrinsic R1 of the agarose solution (s−1), and a is the slope (mL·µg−1·s−1).
The calibration curve was established by linear regression analysis of R1 versus ferumoxytol concentration up to 1250 µg/mL. A 5000 µg/mL ferumoxytol phantom was included to assess the upper practical limit of the current experimental setup.
2.4. In vivo feasibility study
Female BALB/c mice (8–10 weeks old) were obtained from Envigo (Lafayette, Indiana) and housed in the local facility under pathogen‐free conditions. All experimental procedures were reviewed and approved by the local institution in accordance with ARRIVE guidelines.
4T1 cells were obtained from American Type Culture Collection (ATCC) and cultured in RPMI‐1640 medium containing 10% FBS and 1% penicillin/streptomycin. All cultures were incubated at 37°C under humidified conditions with 5% CO2 and 95% air.
Three female BALB/c mice bearing 4T1 mammary and flank tumors (dual tumor model) were used in this experiment. To establish the model, mice were subcutaneously injected with a single‐cell suspension of 1 × 105 4T1 tumor cells in 50 µL of phosphate‐buffered saline (PBS) into the fourth right mammary fat (designated as primary tumor), and 5 × 104 4T1 cells in 50 µL of PBS into the left upper thigh (designated as flank tumor). Mice were divided into control (n = 1) and ferumoxytol‐injected (n = 2) groups. Fourteen days after tumor implantation, the ferumoxytol group received a single dose of 40 mg/kg ferumoxytol administered via the tail vein.
MRI scans were conducted 24 h after injection. All mice were scanned using a 7T MRI system with a 40‐mm‐diameter Tx/Rx 1H RF volume coil (Bruker BioSpin, Ettlingen, Germany), and anesthesia was maintained at 2% isoflurane in oxygen throughout the imaging process. PETALUTE images were acquired using 36,864 petals with 206 sampling points per petal (103 sampling points per echo). RARE‐VTR images were also obtained for comparison.
ROIs were manually delineated in the thigh muscle (as a reference region), mammary tumors, and flank tumors using 3D Slicer. 35 For the thigh muscle reference region, the ROI was placed in the mid‐thigh to avoid the adjacent bone, and intramuscular vessels were excluded from analysis to minimize vascular partial‐volume effects (Figure 6). Quantitative analyses were performed in MATLAB. For each ROI, the mean and standard deviation of T1 and R1 were calculated from both PETALUTE and RARE‐VTR T1 maps to enable comparative evaluation.
FIGURE 6.

PETALUTE and RARE‐VTR images and corresponding T1 maps from the control mouse (Mouse 47) and the ferumoxytol‐injected mice (Mouse 48 and Mouse 49). PETALUTE images are first echo acquired with a flip angle of 4°, while RARE‐VTR images are acquired with TR = 934.58 ms. ROIs used for quantitative analysis are overlaid on the RARE‐VTR anatomical images: thigh muscle (red), flank tumor (yellow), and mammary tumor (green). The thigh muscle ROI for Mouse 47 and Mouse 49 and the mammary tumor ROI for Mouse 48 were located on different slices and are therefore not shown. Color maps show T1 values in milliseconds, with a scale range of 0–6000 ms.
3. RESULT
3.1. Phantom study
Figure 3 presents reconstructed images of all phantoms acquired using RARE‐VTR and PETALUTE, demonstrating the positive contrast of ferumoxytol produced by the PETALUTE first echo and the negative contrast by the second echo. It should be noted that the 5000 µg/mL phantom showed degraded ferumoxytol‐to‐agarose contrast and reduced reliability of PETALUTE R1 quantification at this concentration (Tables S1–S3 and Figures S2, S3). In contrast, the RARE‐VTR images showed signal loss and hypointensity at every tested ferumoxytol concentration.
To evaluate the concentration‐dependent R1 response, the relationship between mean R1 values and ferumoxytol concentration was analyzed (Figure 4). A significant positive linear correlation was observed for PETALUTE‐based mean R1 (R = 0.975, p < 0.01). In contrast, the RARE‐VTR‐based mean R1 did not show significant correlation (R = 0.672, p = 0.144). Additionally, PETALUTE‐based R1 showed lower standard deviations (0.002–0.006 s− 1) than those of RARE‐VTR (0.008–0.255 s− 1) across the full concentration range, indicating improved precision.
FIGURE 4.

Linear regression of mean R1 versus ferumoxytol concentration for PETALUTE (circles) and RARE‐VTR (triangles). Error bars represent the standard deviation within each ROI. The R values were 0.975 for PETALUTE and 0.672 for RARE‐VTR.
3.2. In vivo feasibility study
Reconstructed PETALUTE images covered the whole abdominal region and provided visualization of anatomical features (Figure 5). Notably, structures such as the kidneys and livers, which were poorly defined in the RARE‐VTR images due to a limited field of view, were distinguishable on PETALUTE images. In addition, while RARE‐VTR images displayed a hypointense region within the flank tumor, PETALUTE showed slightly higher signal intensity near the tumor center (Figure 6).
FIGURE 5.

Representative coronal reformatted PETALUTE images from a control mouse (Mouse 47) showing the first echo and second echo (FA = 4°). The figure illustrates the full abdominal coverage by the PETALUTE acquisition.
ROIs within the thigh muscle, flank tumors, and mammary tumors allowed the determination of T1 values from PETALUTE and RARE‐VTR. The overall mean T1 and R1 values for each mouse and the respective ROI are reported in Table 2. For both T1 mapping methods, muscle tissue mean T1 values remained relatively stable across mice. For RARE‐VTR, the mean muscle T1 values for the ferumoxytol‐injected mice were within 6% of the Control mouse T1 values. For PETALUTE, the mean muscle T1 values for ferumoxytol‐injected mice were within 4% of the Control mouse T1 values. In the flank tumor, both T1 mapping methods yielded lower mean T1 values in ferumoxytol‐injected mice than in the Control mouse. The T1 reduction in the flank tumor in ferumoxytol‐injected mice was approximately 10%–12% with RARE‐VTR and 4%–7% with PETALUTE. In the mammary tumor, T1 changes differed across the two T1 mapping methods. With RARE‐VTR, both the ferumoxytol‐injected mice exhibited an approximately 9%–11% T1 reduction relative to the Control mouse. PETALUTE, however, revealed divergent responses: relative to the Control mouse, the mean T1 increased by approximately 42% in FMX‐1 but decreased by approximately 4% in FMX‐2.
TABLE 2.
Mean and standard deviation (SD) of T1 and R1 values within the ROIs of thigh muscle, flank tumors and mammary tumors in three mice.
| ID a | ROI | RARE‐VTR b R1 mean ± SD (s‐1) | PETALUTE c R1 mean ± SD (s‐1) | RARE‐VTR T1 mean ± SD (ms) | PETALUTE T1 mean ± SD (ms) |
|---|---|---|---|---|---|
| Control | Thigh Muscle | 0.242 ± 0.023 | 0.247 ± 0.015 | 4164 ± 392 | 4063 ± 249 |
| Flank Tumor | 0.260 ± 0.012 | 0.298 ± 0.043 | 3850 ± 184 | 3424 ± 464 | |
| Mammary Tumor | 0.265 ± 0.029 | 0.374 ± 0.063 | 3821 ± 396 | 2745 ± 443 | |
| FMX‐1 | Thigh Muscle | 0.258 ± 0.029 | 0.240 ± 0.021 | 3919 ± 427 | 4203 ± 357 |
| Flank Tumor | 0.295 ± 0.037 | 0.318 ± 0.042 | 3438 ± 389 | 3198 ± 416 | |
| Mammary Tumor | 0.302 ± 0.073 | 0.265 ± 0.048 | 3472 ± 691 | 3894 ± 667 | |
| FMX‐2 | Thigh Muscle | 0.239 ± 0.029 | 0.251 ± 0.020 | 4252 ± 545 | 4002 ± 322 |
| Flank Tumor | 0.306 ± 0.060 | 0.307 ± 0.030 | 3369 ± 556 | 3286 ± 341 | |
| Mammary Tumor | 0.302 ± 0.060 | 0.386 ± 0.049 | 3401 ± 500 | 2632 ± 330 |
Control = control mouse; FMX‐1 and FMX‐2 = ferumoxytol‐injected mice.
RARE‐VTR = rapid acquisition with relaxation enhancement–variable TR.
PETALUTE = 3D dual‐echo ultrashort echo time.
4. DISCUSSION
We developed a B1 +‐corrected PETALUTE‐VFA T1 mapping protocol that enables high‐resolution T1 mapping for ferumoxytol biodistribution assessment. This new protocol improved the linearity between R1 and ferumoxytol concentration, produced positive contrast at high ferumoxytol concentrations, and expanded spatial coverage, thereby overcoming the poor image quality and signal losses that limit conventional T1 mapping for SPION imaging.
PETALUTE T1 mapping establishes a reliable calibration curve for ferumoxytol assessment up to 1250 µg/mL, whereas the conventional RARE‐VTR method did not yield a comparable linear response over this range. The improved performance of PETALUTE T1 mapping in the phantom study is attributed to its ultrashort echo time, which captures signal from iron before substantial T2*‐related dephasing and therefore yields higher SNR and reduced variability in estimated T1 values. Consistent with this, the PETALUTE‐based R1 standard deviation remained relatively low across the concentration range, whereas the RARE‐VTR‐based R1 standard deviation increased substantially with ferumoxytol concentration (Figure 4), reaching approximately 40 times that of PETALUTE at 1250 µg/mL. This smaller concentration‐dependent variability of PETALUTE‐based R1 further supports the lower susceptibility of PETALUTE to T2*‐related signal loss at high ferumoxytol concentrations. In addition, reduced susceptibility effects at lower magnetic field strengths are anticipated to extend the usable linear range of the calibration curve, suggesting that the limits reported here at 7T are likely conservative compared with low‐field ferumoxytol assessment methods. 36
Another factor affecting comparability between PETALUTE and RARE‐VTR is spatial resolution. PETALUTE provides high isotropic resolution, whereas RARE‐VTR is anisotropic (Table 1), particularly because of its thicker slice dimension. The larger voxel volume in RARE‐VTR increases susceptibility to partial‐volume effects, such that the measured T1 may reflect an average of heterogeneous ferumoxytol concentrations within a single voxel. In contrast, the isotropic resolution of PETALUTE reduces partial‐volume averaging and improves spatial specificity of the measured ferumoxytol distribution. Therefore, differences between PETALUTE and RARE‐VTR measurements may reflect not only sequence‐specific T1 estimation behavior, but also differing sensitivity to partial‐volume effects.
A discrepancy of about 12% was observed in the baseline agarose phantom R1 values (0 µg/mL ferumoxytol) between the two methods (RARE‐VTR: 0.270 s− 1; PETALUTE: 0.237 s− 1). This difference likely comes from systematic bias caused by different T1 mapping methods. RARE‐VTR and PETALUTE use different acquisition strategies, signal models, and fitting methods. These differences can lead to offsets in the measured R1 values. The discrepancy may also come from residual bias in the short‐TR VFA framework, including imperfect B1 + correction and incomplete spoiling. 37 Spatial resolution may also play a role, but partial‐volume effects are less likely to be a major cause in this relatively homogeneous phantom. Similar systematic differences between VFA‐based T1 mapping and reference relaxation‐based methods have been reported previously. 38 Despite this baseline offset, PETALUTE still showed excellent linearity up to 1250 µg/mL. Notably, the Bland‐Altman analysis indicates a residual proportional bias between the PETALUTE and RARE‐VTR methods after B1 + correction (Figure S4). Therefore, PETALUTE and RARE‐VTR T1 mapping should not be considered interchangeable quantitative T1 methods at this stage. Rather, PETALUTE is more suitable for relative or trend‐based assessment of ferumoxytol‐induced T1 change.
The linear R1 behavior observed in homogeneous phantoms does not establish quantitative validity in the biologically heterogeneous environment of in vivo tissue. Unlike the controlled phantom environment, in vivo tissue contains variable water content, necrosis, inflammation, hemorrhage, and spatially heterogeneous ferumoxytol distribution, all of which can influence the measured T1. Therefore, in vivo T1 should be interpreted as a ferumoxytol‐sensitive marker rather than a direct concentration readout. Nevertheless, the phantom calibration curve provides a useful reference for the expected relationship between R1 change and ferumoxytol accumulation, enabling relative quantitative assessment when comparing baseline and post‐injection R1 values. 39
The in vivo scanning demonstrated the feasibility of using the PETALUTE T1 mapping protocol to assess ferumoxytol distribution in rodent tumors. Given that no substantial accumulation of ferumoxytol was expected in muscle tissue, the muscle served as a reference region. As expected, we observed minimal differences in T1 values in this region between the control and ferumoxytol‐injected groups, with mean muscle T1 values within 6% for RARE‐VTR and within 4% for PETALUTE relative to the Control mouse. In tumors, accumulation of ferumoxytol would be expected to lead to T1 shortening, including decreased T1 values and increased T1‐weighted MRI signal intensity. This behavior was observed in the flank tumors of ferumoxytol‐injected mice in our study, which exhibited lower T1 values with both methods and higher intratumoral signal intensity in the PETALUTE first echo. This finding is consistent with the preferential accumulation of ferumoxytol within tumor tissue, consistent with prior reports of tumor‐specific ferumoxytol uptake that manifests as decreased T1 values. 40 The greater T1 reduction observed with RARE‐VTR is likely due to its longer TE (20 ms), which leads to substantial T2 dephasing in the presence of ferumoxytol. This residual T2 weighting may reduce the effective MRI signal in RARE‐VTR images, causing hypointensity and exaggerating the apparent T1‐shortening effect. In contrast, the ultrashort TE of PETALUTE (0.016 ms) minimizes T2* decay, enabling acquisition of a more direct T1‐weighted signal before substantial dephasing occurs.
An unexpected result was observed in the in vivo study. One of the ferumoxytol‐injected mice (FMX‐1) exhibited a nearly 42% increase in mean PETALUTE T1 in the mammary tumor. To investigate this unexpected finding, multimodal images were examined (Figure 7). In the FMX‐1 mammary tumor, the combination of hypointensity in the T1‐weighted image (indicated by solid arrow in Figure 7) and persistent signal in the T2*‐weighted image (indicated by dashed arrow in Figure 7) is suggestive of necrosis‐like change based on prior reports. 41 However, this interpretation remains provisional. Similar imaging features may also arise from other biological processes, including inflammation, 42 hemorrhage, 43 or other long‐T2* components. Importantly, the larger tumor volume in the FMX‐1 mammary tumor (Control: 152 mm3, FMX‐1: 210 mm3, FMX‐2: 189 mm3) may have increased the likelihood of necrosis, which dominated the MR signal response. 41
FIGURE 7.

Multimodal images of the control mouse (Mouse 47) and the ferumoxytol‐injected mice (Mouse 48 and Mouse 49) focusing on the mammary tumor. Shown are PETALUTE first echo, PETALUTE second echo, T1‐weighted (RARE‐VTR, TE = 20 ms, TR = 933 ms), T2*‐weighted (MGE, TE = 6 ms, TR = 1500 ms), PETALUTE T1 map, and RARE‐VTR T1 map images. Mammary tumors are contoured in red. Dashed arrows indicate T1 hypointensity with persistent T2* signal indicated in Mouse 48 while solid arrows indicate T1 hypointensity with T2* signal loss in Mouse 49. Mouse 47 (control) showed uniform intratumor intensity. Color maps display T1 values in milliseconds, with a scale range of 0–6000 ms.
Based on the phantom study (Figure 3), ferumoxytol accumulation would be expected to produce hyperintensity on the PETALUTE first echo and hypointensity on the second echo. Therefore, the hyperintensity region observed in the PETALUTE second echo of FMX‐1 (Figure 7) is not consistent with ferumoxytol accumulation alone and is more plausibly attributed to a long T2 component. Overall, these observations suggest that the hyperintensity observed in the FMX‐1 tumor across both PETALUTE echoes may reflect a mixture of ferumoxytol‐related signal and tissue changes. The resulting PETALUTE T1 map revealed elevated T1 values in these regions, whereas the RARE‐VTR T1 map was not sensitive to this effect (Figure 7). Similar intratumoral hyperintensity was also observed in both PETALUTE echoes of FMX‐2, though without measurable T1 elevation. This may reflect mild or early‐stage tissue alteration that was insufficient to alter the overall MR signal behavior.
The different behavior between mammary and flank tumors may also reflect systematic biological differences related to tumor location. Mammary tumors and flank tumors differ in vascularization, immune‐cell infiltration, and necrotic tendency, all of which may influence ferumoxytol delivery, retention, and the resulting T1 behavior. In the current setting, PETALUTE may be particularly sensitive to intratumoral biological heterogeneity. This sensitivity could be advantageous if it reflects biologically relevant variation, but it also complicates the interpretation of ferumoxytol‐related T1 alteration in this small exploratory cohort. Further histological studies will be required to determine whether these findings correspond to necrosis, inflammation, hemorrhage, or a combination of these processes. Given the small sample size, the mammary tumor finding in FMX‐1 should be interpreted cautiously as an individual response rather than evidence of a group‐level response.
A practical difference between PETALUTE and RARE‐VTR in the in vivo feasibility study is motion sensitivity. RARE‐VTR uses conventional Cartesian k‐space sampling and is more susceptible to respiration‐related motion because of its long acquisition times. Although respiratory triggering can reduce breathing artifacts, it further prolongs the scan. In contrast, PETALUTE uses a non‐Cartesian sampling strategy that is less prone to artifacts from respiration and cardiac motion and provides efficient k‐space coverage for undersampled imaging. 44 This enables abdominal imaging without respiratory gating within a practical scan time. However, the relative sensitivity of PETALUTE and RARE‐VTR to physiological motion was not systematically evaluated in the present study and should be examined in future work.
This study has several limitations that should be acknowledged. First, PETALUTE images appear slightly blurrier than those acquired with RARE‐VTR (Figure 6). This is primarily because we deliberately used a high undersampling factor together with a CS‐SENSE reconstruction to enable faster, non‐gated whole abdominal T1 mapping. For the target nominal resolution (matrix = 256 × 256 × 256, r = 128), the Nyquist criterion corresponds to approximately petals (). In comparison, our protocol uses 36,864, which is 18% of the Nyquist sampling density. This level of undersampling reduces image sharpness and fine edge detail, leading to some loss of anatomic definition relative to RARE‐VTR. However, previous work has shown that PETALUTE estimates remain stable even at higher undersampling level (18,192 petals) reconstructed with the same matrix size (256 × 256 × 256), so substantial degradation of values is not expected at the sampling density used here. 45 If higher anatomical clarity is desired, the protocol can be adapted by increasing the sampling density.
Second, although the PETALUTE T1 mapping approach extends the reliable linear range of ferumoxytol concentrations, it remains limited at very high concentrations. At 5000 µg/mL, the susceptibility of ferumoxytol at 7T results in a local frequency shift of approximately 20 kHz and an estimated T2* of about 0.01 ms. 24 These large off‐resonance effects and extremely short T2* lead to considerable signal attenuation and k‐space misregistration that severely degrade mapping accuracy. Under these conditions, the current rosette k‐space sampling and reconstruction framework is no longer sufficient to preserve reliable T1 information, and additional off‐resonance correction would likely be required. Therefore, 5000 µg/mL is reported here as an upper practical limit of the PETALUTE mapping approach for ferumoxytol assessment.
Finally, the in vivo experiments should be interpreted as an initial feasibility demonstration and proof‐of‐concept rather than a comprehensive method validation, given the small sample size. In addition, differences in tumor volume among the studied mice may have influenced the observed imaging features and biological interpretation. Future studies with larger animal cohorts are needed to statistically evaluate measurement accuracy and biological variability in ferumoxytol uptake using PETALUTE mapping.
Clinical translation of the present 7T results will require additional validation at 3T and 1.5T, since field strength affects relaxivity, susceptibility behavior, and T1 mapping performance. 18 Accordingly, the present T1 mapping protocol will need to be re‐established for each clinical field strength rather than assumed to transfer directly from 7T. In addition, ferumoxytol clinical translation will require careful dose optimization, monitored administration, and safety studies. A reasonable clinic migration pathway would therefore be field‐specific calibration and reproducibility testing at 3T and 1.5T, followed by prospective clinical feasibility studies.
We also note that the present study did not include a dedicated quantitative T2* analysis, simulation study, or multi‐echo relaxation modeling to directly disentangle the T1 and T2* contributions in SPION‐rich regions. Further evaluation of PETALUTE's performance will be carried out using simulations to understand how varying SPION concentration, off‐resonance, and T2* shortening affect mapping precision. Future studies will also focus on validating in vivo findings with histological evidence using both T1 and T2* mapping.
5. CONCLUSION
In summary, the PETALUTE mapping protocol provides more reliable ferumoxytol assessment compared with RARE‐VTR. Uniquely, this protocol enables high‐resolution, non‐gated whole‐abdominal images and dual‐echo imaging within a single scan. This multimodal capability makes PETALUTE particularly valuable for monitoring treatment with SPION‐based therapies.
FUNDING INFORMATION
Support was provided by Eli Lilly and Company. The authors also acknowledge support from NIH grant P30 CA023168 (Purdue University Institute for Cancer Research) and from Wellcome Trust Collaborative Award 223131/Z/21/Z (UE and SW).
CONFLICT OF INTEREST STATEMENT
The authors declare no conflicts of interest.
Supporting information
Supporting Information
ACKNOWLEDGMENTS
The authors declare no additional acknowledgments.
DATA AVAILABILITY STATEMENT
The data that support the findings of this study are available from the corresponding author upon reasonable request.
REFERENCES
- 1. Mitchell MJ, Billingsley MM, Haley RM, Wechsler ME, Peppas NA, Langer R. Engineering precision nanoparticles for drug delivery. Nat Rev Drug Discov. 2021;20(2):101–124. doi: 10.1038/s41573-020-0090-8 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2. Blanco E, Shen H, Ferrari M. Principles of nanoparticle design for overcoming biological barriers to drug delivery. Nat Biotechnol. 2015;33(9):941–951. doi: 10.1038/nbt.3330 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3. Wilhelm S, Tavares AJ, Dai Q, et al. Analysis of nanoparticle delivery to tumours. Nat Rev Mater. 2016;1(5):16014. doi: 10.1038/natrevmats.2016.14 [DOI] [Google Scholar]
- 4. Ng TSC, Garlin MA, Weissleder R, Miller MA. Improving nanotherapy delivery and action through image‐guided systems pharmacology. Theranostics. 2020;10(3):968–997. doi: 10.7150/thno.37215 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5. Chang DY, Speth JP, Scarpelli ML. Evaluating the theranostic potential of ferumoxytol when combined with radiotherapy in a mammary dual tumor mouse model. Med Phys. 2025;52(7):e17888. doi: 10.1002/mp.17888 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6. Dennahy IS, Han Z, MacCuaig WM, et al. Nanotheranostics for image‐guided cancer treatment. Pharmaceutics. 2022;14(5):917. doi: 10.3390/pharmaceutics14050917 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7. Fan D, Cao Y, Cao M, Wang Y, Cao Y, Gong T. Nanomedicine in cancer therapy. Sig Transduct Target Ther. 2023;8(1):293. doi: 10.1038/s41392-023-01536-y [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8. Lapusan R, Borlan R, Focsan M. Advancing MRI with magnetic nanoparticles: a comprehensive review of translational research and clinical trials. Nanoscale Adv. 2024;6(9):2234–2259. doi: 10.1039/D3NA01064C [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9. Girard OM, Du J, Agemy L, et al. Optimization of iron oxide nanoparticle detection using ultrashort echo time pulse sequences: comparison of T 1, T 2*, and synergistic T 1 − T 2* contrast mechanisms: iONP detection using UTE sequences. Magn Reson Med. 2011;65(6):1649–1660. doi: 10.1002/mrm.22755 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10. Jeon M, Halbert MV, Stephen ZR, Zhang M. Iron oxide nanoparticles as T1 contrast agents for magnetic resonance imaging: fundamentals, challenges, applications, and prospectives. Adv Mater. 2021;33(23):1906539. doi: 10.1002/adma.201906539 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11. Rahman M. Magnetic resonance imaging and iron‐oxide nanoparticles in the era of personalized medicine. Nanotheranostics. 2023;7(4):424–449. doi: 10.7150/ntno.86467 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12. Weiskopf N, Edwards LJ, Helms G, Mohammadi S, Kirilina E. Quantitative magnetic resonance imaging of brain anatomy and in vivo histology. Nat Rev Phys. 2021;3(8):570–588. doi: 10.1038/s42254-021-00326-1 [DOI] [Google Scholar]
- 13. Bottomley PA, Foster TH, Argersinger RE, Pfeifer LM. A review of normal tissue hydrogen NMR relaxation times and relaxation mechanisms from 1–100 MHz: dependence on tissue type, NMR frequency, temperature, species, excision, and age. Med Phys. 1984;11(4):425–448. doi: 10.1118/1.595535 [DOI] [PubMed] [Google Scholar]
- 14. Messroghli DR, Moon JC, Ferreira VM, et al. Clinical recommendations for cardiovascular magnetic resonance mapping of T1, T2, T2* and extracellular volume: a consensus statement by the society for cardiovascular magnetic resonance (SCMR) endorsed by the European association for cardiovascular imaging (EACVI). J Cardiovasc Magn Reson. 2017;19(1):75. doi: 10.1186/s12968-017-0389-8 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15. Kolb H, Absinta M, Beck ES, et al. T MRI differentiates remyelinated from demyelinated multiple sclerosis lesions. Ann Neurol. 2021;90(4):612–626. doi: 10.1002/ana.26194 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16. Eck BL, Yang M, Elias JJ, et al. Quantitative MRI for evaluation of musculoskeletal disease: cartilage and muscle composition, joint inflammation, and biomechanics in osteoarthritis. Invest Radiol. 2023;58(1):60. doi: 10.1097/RLI.0000000000000909 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17. Zhang J, Ring HL, Hurley KR, et al. Quantification and biodistribution of iron oxide nanoparticles in the primary clearance organs of mice using T1 contrast for heating. Magn Reson Med. 2017;78(2):702–712. doi: 10.1002/mrm.26394 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18. Simon GH, Bauer J, Saborovski O, et al. T1 and T2 relaxivity of intracellular and extracellular USPIO at 1.5T and 3T clinical MR scanning. Eur Radiol. 2006;16(3):738–745. doi: 10.1007/s00330-005-0031-2 [DOI] [PubMed] [Google Scholar]
- 19. Zhang L, Zhong X, Wang L, et al. T1‐weighted ultrashort echo time method for positive contrast imaging of magnetic nanoparticles and cancer cells bound with the targeted nanoparticles. J Magn Reson Imaging. 2011;33(1):194–202. doi: 10.1002/jmri.22412 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20. Korchinski DJ, Taha M, Yang R, Nathoo N, Dunn JF. Iron oxide as an MRI contrast agent for cell tracking: supplementary issue. Magn Reson Insights. 2015;8s1: MRI.S23557. doi: 10.4137/MRI.S23557 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21. Liu W, Dahnke H, Rahmer J, Jordan EK, Frank JA. Ultrashort T2* relaxometry for quantitation of highly concentrated superparamagnetic iron oxide (SPIO) nanoparticle labeled cells. Magn Reson Med. 2009;61(4):761–766. doi: 10.1002/mrm.21923 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22. Du J, Ma G, Li S, et al. Ultrashort echo time (UTE) magnetic resonance imaging of the short T2 components in white matter of the brain using a clinical 3T scanner. Neuroimage. 2014;87:32–41. doi: 10.1016/j.neuroimage.2013.10.053 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23. Ring HL, Gao Z, Sharma A, et al. Imaging the distribution of iron oxide nanoparticles in hypothermic perfused tissues. Magn Reson Med. 2020;83(5):1750–1759. doi: 10.1002/mrm.28123 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24. Liu S, Brisset JC, Hu J, Haacke EM, Ge Y. Susceptibility weighted imaging and quantitative susceptibility mapping of the cerebral vasculature using ferumoxytol. J Magn Reson Imaging. 2018;47(3):621–633. doi: 10.1002/jmri.25809 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25. Larson PEZ, Gurney PT, Nayak K, Gold GE, Pauly JM, Nishimura DG. Designing long‐T2 suppression pulses for ultrashort echo time imaging. Magn Reson Med. 2006;56(1):94–103. doi: 10.1002/mrm.20926 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26. Gatehouse PD, Bydder GM. Magnetic resonance imaging of short T2 components in tissue. Clin Radiol. 2003;58(1):1–19. doi: 10.1053/crad.2003.1157 [DOI] [PubMed] [Google Scholar]
- 27. Ma YJ, Searleman AC, Jang H, et al. Whole‐brain myelin imaging using 3D double‐echo sliding inversion recovery ultrashort echo time (DESIRE UTE) MRI. Radiology. 2020;294(2):362–374. doi: 10.1148/radiol.2019190911 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28. Jang H, Ma Y, Carl M, Jerban S, Chang EY, Du J. Ultrashort echo time cones double echo steady state (UTE‐Cones‐DESS) for rapid morphological imaging of short T2 tissues. Magn Reson Med. 2021;86(2):881–892. doi: 10.1002/mrm.28769 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29. Shen X, Özen AC, Monsivais H, et al. High‐resolution 3D ultra‐short echo time MRI with rosette k‐space pattern for brain iron content mapping. J Trace Elem Med Biol. 2023;77:127146. doi: 10.1016/j.jtemb.2023.127146 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30. Shen X, Özen AC, Sunjar A, et al. Ultra‐short T2 components imaging of the whole brain using 3D dual‐echo UTE MRI with rosette k‐space pattern. Magn Reson Med. 2023;89(2):508–521. doi: 10.1002/mrm.29451 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31. Helms G, Dathe H, Dechent P. Quantitative FLASH MRI at 3T using a rational approximation of the Ernst equation. Magn Reson Med. 2008;59(3):667–672. doi: 10.1002/mrm.21542 [DOI] [PubMed] [Google Scholar]
- 32. Ishimori Y, Shimanuki T, Kobayashi T, Monma M. Fast B1 mapping based on double‐angle method with T1 correction using standard pulse sequence. J Med Phys. 2022;47(1):93–98. doi: 10.4103/jmp.jmp_78_21 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33. Dathe H, Helms G. Exact algebraization of the signal equation of spoiled gradient echo MRI. Phys Med Biol. 2010;55(15):4231. doi: 10.1088/0031-9155/55/15/003 [DOI] [PubMed] [Google Scholar]
- 34. Insko EK, Bolinger L. Mapping of the radiofrequency field. J. Magn. Reson. Ser. A. 1993;103(1):82–85. doi: 10.1006/jmra.1993.1133 [DOI] [Google Scholar]
- 35. Pieper S, Halle M, Kikinis R, 3D Slicer. In: 2004 2nd IEEE International Symposium on Biomedical Imaging: Nano to Macro (IEEE Cat No. 04EX821) . 2004:632–635. 1. doi: 10.1109/ISBI.2004.1398617 [DOI] [Google Scholar]
- 36. Gillis P, Koenig SH. Transverse relaxation of solvent protons induced by magnetized spheres: application to ferritin, erythrocytes, and magnetite. Magn. Reson. Med. 1987;5(4):323–345. doi: 10.1002/mrm.1910050404 [DOI] [PubMed] [Google Scholar]
- 37. Preibisch C, Deichmann R. Influence of RF spoiling on the stability and accuracy of T1 mapping based on spoiled FLASH with varying flip angles. Magn. Reson. Med. 2009;61(1):125–135. doi: 10.1002/mrm.21776 [DOI] [PubMed] [Google Scholar]
- 38. Keenan KE, Gimbutas Z, Dienstfrey A, et al. Multi‐site, multi‐platform comparison of MRI T1 measurement using the system phantom. PLOS ONE. 2021;16(6):e0252966. doi: 10.1371/journal.pone.0252966 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39. Zhu A, Reeder SB, Johnson KM, et al. Quantitative ferumoxytol‐enhanced MRI in pregnancy: a feasibility study in the nonhuman primate. Magn Reson Imaging. 2020;65:100–108. doi: 10.1016/j.mri.2019.10.006 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40. Feng Q, Xu X, Wei C, et al. The dynamic interactions between nanoparticles and acrophages impact their fate in brain tumors. Small. 2021;17(49):2103600. doi: 10.1002/smll.202103600 [DOI] [PubMed] [Google Scholar]
- 41. Anderson MA, Knipp DE, Noda Y, et al. MRI‐based tumor necrosis depiction in pancreatic ductal adenocarcinoma: can it predict tumor aggressiveness?. Cancers. 2023;15(8):2313. doi: 10.3390/cancers15082313 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42. Wan Q, Peng H, Lyu J, et al. Water specific MRI T1 mapping for evaluating liver inflammation activity grades in rats with methionine‐choline‐deficient diet‐induced nonalcoholic fatty liver disease. J Magn Reson Imaging. 2022;56(5):1429–1436. doi: 10.1002/jmri.28131 [DOI] [PubMed] [Google Scholar]
- 43. Bradley WG. MR appearance of hemorrhage in the brain. Radiology. 1993;189(1):15–26. doi: 10.1148/radiology.189.1.8372185 [DOI] [PubMed] [Google Scholar]
- 44. Wright KL, Hamilton JI, Griswold MA, Gulani V, Seiberlich N. Non‐cartesian parallel imaging reconstruction. J Magn Reson Imaging. 2014;40(5):1022–1040. doi: 10.1002/jmri.24521 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45. Lipka A, Sawiak S, Shen X, et al. Accelerated Preclinical UHF Abdominal T1 Mapping Using Novel Rosette Ultrashort Echo Time (PETALUTE) . ISMRM Annual Meeting, 2024. doi: 10.58530/2024/0575 [DOI]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Supporting Information
Data Availability Statement
The data that support the findings of this study are available from the corresponding author upon reasonable request.