. 2026 Sep 17;26(18):5894. doi: 10.3390/s26185894

RF Power, B1+rms, and SAR Variation with RF Coils, Conductive Metallic Implants, and Ionic Solutions at 1.5T and 3T

David H Gultekin 1
PMCID: PMC13611460  PMID: 42817490

Abstract

Highlights

  • Access to magnetic resonance imaging (MRI) for patients with active and passive implants is restricted due to radiofrequency (RF) safety concerns. Ongoing research aims to evaluate the utility of RF power, time-averaged root-mean-square RF field (B1+rms), and specific absorption rate (SAR) for implant safety.

What are the main findings?

  • MRI is remotely sensitive to metallic wavelengths and ionic concentrations.

  • RF power is sensitive to the presence or absence of conductive metallic implants and ionic solutions, implant wavelengths, ionic concentrations, RF coils, and Tx/Rx combinations, whereas B1+rms is not sensitive.

What are the implications of the main findings?

  • RF power may be monitored because scanner-reported SAR may vary unpredictably with metallic and ionic conductivities.

  • Saturation bands increase RF power, B1+rms, and SAR, and they must be turned off when imaging patients with conductive metallic implants.

Abstract

Currently, the time-averaged root-mean-square RF field (B1+rms) and specific absorption rate (SAR) are being evaluated to monitor and control RF-induced heating near conductive metallic implants, such as deep brain stimulation (DBS) leads, during MRI. However, experimental methods to assess the relationship between RF power, B1+rms, and SAR are lacking for RF coils, RF pulse sequences, metallic implants, and ionic solutions. A method is developed to evaluate the variation in RF power, B1+rms, and SAR with RF coils, metallic implants, and ionic solutions using phantoms consisting of water (H2O) and sodium chloride (NaCl) with four ionic concentrations (0, 1, 2, 3%), four metallic wavelengths (0, λ/2, λ, 2λ), two RF coils (body, head), transmit/receive (Tx/Rx) combinations, and five RF pulse flip angles (30, 45, 60, 75, 90 degrees) in two B0 fields (1.5T and 3T). The scanner-reported RF power and SAR varied with RF pulse sequences, RF coils, Tx/Rx, metallic implants, and ionic solutions, whereas B1+rms varied only with RF pulse sequences. The RF power, B1+rms, and SAR relationship depends on RF pulse sequences, RF coils, Tx/Rx, metallic implant wavelengths, and ionic concentrations. SAR (whole-body, head) scaled with RF power by absorption ratios (α) variable with experimental conditions.

Keywords: RF power, B1+rms, SAR, RF coils, RF heating, temperature, metallic implants, ionic solutions

1. Introduction

There is a growing number of patients with conductive metallic implants and electronic devices that require magnetic resonance imaging (MRI) for diagnosis and therapy response monitoring. More than 4.6 million hip and knee procedures, including primary and revision arthroplasty procedures, were performed between 2012 and 2024 [1]. More than 400,000 patients in the US and about 1.7 million patients in the world receive cardiac implantable electronic devices (CIEDs) every year [2,3]. Up until now, it is estimated that more than 300,000 patients have received deep-brain stimulation (DBS) leads globally, and the number is increasing with time at a rate of about 12,000 per year [4]. The applications of electrical stimulation in neuroscience and neurosurgery have led to important advances in surgical treatments of neurological diseases and continue to grow in a wide range of neurological conditions [5,6,7,8,9,10,11,12,13].

However, conductive metallic implants may heat up significantly during the MRI and create safety risks for the patients [14,15,16,17,18]. The time-varying electromagnetic fields in the MRI may induce heating in the leads and potentially damage the cells in the vicinity of the leads implanted in the tissues [19,20,21,22,23]. Therefore, it is important to develop safety models and methods to predict and control the magnitude of heating in conductive implants during the MRI [24,25,26].

In general, most adverse effects reported about MRI involve thermal injury to the patients [27], through possible effects caused by electromagnetic field interactions [28,29], and catastrophic incidents in the presence of conductive metallic implant leads [30]. As a consequence, the potential applications of MRI are limited partly by considerations of radiofrequency (RF) safety that impose constraints on the design of RF coils and RF pulse sequences [31]. A current dilemma in MRI is how to optimize image quality and scan time while maintaining RF safety. Often, one factor may be improved at the expense of other factors.

There is a growing need and interest in assessing the utility of temperature rise (dT), specific absorption rate (SAR), and the time-averaged root-mean-square RF fields (B1+rms) to monitor RF absorption and thermal hazards in tissues, especially in the presence of conductive metallic implants such as DBS electrodes [32,33,34,35,36]. In a previous study, the relationship between RF-induced heating and temperature rise near a conductive metallic DBS lead, and the scanner-reported B1+rms and RF power for RF pulse sequences, RF coils, and transmit and receive (Tx/Rx) combinations was investigated at 3T [37]. It was demonstrated that both temperature rise and RF power varied with RF pulse sequences, RF coils, and Tx/Rx combinations, whereas B1+rms varied only with RF pulse sequences but not with RF coils and Tx/Rx combinations at 3T. The RF-induced heating and temperature rise were associated with RF power propagation and loss corresponding to the B1+rms for RF coils and Tx/Rx combinations at 3T. The study also showed the RF heating effects of RF coils and Tx/Rx combinations with RF adjustments, including B1 shimming at 3T [37,38,39,40].

However, transient temperature measurements are invasive, complicated, and take a much longer time, as the temperature increases during the heating period over the scan duration and decreases during the cooling period. Experiments on in vivo and ex vivo brain tissue, with and without conductive implants, show that the magnitude of RF-induced heating and temperature rise increases with frequency, power, and duration, necessitating relatively long cooling times between experiments [20,41]. There is also global heating and melting of the gel medium in a finite-size phantom with repeated experiments, especially with higher frequency, power, and time, causing thermal convection and distortion of the temperature profile, which may require longer cooling time to return to baseline after several RF-intensive experiments [20,41]. During the MRI, temperature changes (increases when RF pulses are on and decreases when RF pulses are off), while RF power and B1+rms stay constant.

In this study, the specific aim is to assess the relationship between the three scanner-reported noninvasive parameters, RF power, B1+rms, and SAR, varying with RF pulse sequences, RF coils, Tx/Rx combination, conductive metallic implant wavelengths, and ionic concentrations at 1.5T and 3T. This study also extends the previous study by comparing the effects of RF adjustments at 1.5T and 3T. This method can be used to compare experimental measurements with computational electromagnetic and thermal modeling to study RF coils, RF pulse sequences, and conductive metallic implants and media for RF-induced heating during MRI.

2. Materials and Methods

2.1. Theory

In nuclear magnetic resonance imaging, nuclear spins are placed in a static magnetic field (B0) and perturbed away from the equilibrium magnetization by an external transverse RF magnetic field (B1), and the resultant signal is received and processed spatially and temporally using a set of gradient coils and radiofrequency coils [42,43,44].

The relationship between the nuclear spin phase (θ) and the magnetic field (B1) can be derived from Larmor’s equation [42,43].

ω=γB1θ=ωtp=γB1tp (1)

where ω is the angular Larmor frequency (rads/s), γ is the angular gyromagnetic ratio (rads/s·T), B1 is an external transverse magnetic field (Tesla), and tp is the pulse length (s). A plot of B1 vs. θ yields a straight line with a slope proportional to 1/γtp, which can be measured experimentally.

The specific absorption rate in an ex vivo tissue or a substance in the absence of perfusion can be related to the electric field or the initial rate of temperature rise at a point in the body, as

SAR=σE2ρSAR=Cp∂T∂tt→0 (2)

where σ is the electrical conductivity (S/m), E is the root-mean-square (rms) electric field (V/m), ρ is the mass density (kg/m3), Cp is the specific heat (J/kg·K) in tissue, and (∂T/∂t)(t→0) is the initial temperature rise with time (K/s) [45].

The whole-body average SAR can be related to the average incident RF power (P) as

SARwb=PabsmP=mαSARwb (3)

where m is the patient’s weight (kg) and α = Pabs/P is a calculated dimensionless parameter called the absorption ratio, the ratio of the absorbed power (Pabs) to the incident power (P). The expected value of α is a fraction (α < 1, Pabs < P) that increases between 0 and 1 (0 < α < 1) with increasing frequency. A linear regression of P vs. SARwb yields a straight line with a slope proportional to m/α, which can be measured experimentally.

In the simplest coil element, the E field and the incident RF power (P) in a loop of wire with a radius (r) can be formulated using Equations (2) and (3) as

E=12ωrB1P=σmω2r2B124αρ (4)

where ω is the angular frequency, and B1 is the RF field [46,47,48,49]. A linear regression of P vs. the square of the B1 field yields a straight line with a slope proportional to σmω2r2/4αρ, which can be measured experimentally.

Further, by substituting Equation (1) into Equation (4) and scaling it with two dimensionless parameters, number of slices (NS) and duty factor (D), the absorbed RF power and incident RF power can be related to the flip angle (θ) as

Pabs=σmω2r2NSθ24ργ2tpTRP=σmω2r2NSθ24αργ2tpTR (5)

where α = Pabs/P and D = tp/TR is the ratio of pulse length (tp) to the repetition time (TR) [46,47,48,49]. These formulations are very simplified, and the exact relation for power deposition depends on the much more complicated electromagnetic fields in RF coils, implants, the direction of the B1 field, and the coordinate system in the tissue [28,50].

RF coils for MRI require a complicated electromagnetic design and optimization. However, the simplest element of a conductor loop and a capacitor (LC) still makes up the unit element of arrays with up to 128 LC loops [31]. The relationship between the RF power, magnetic field, and electric field in a coil depends on a specific coil design and a number of parameters [31,51,52]. However, the RF power and B1 field can be monitored for a coil or a combination of coils experimentally. Here, the variation in RF power, B1+rms, and SAR is evaluated in various experimental conditions by varying the metallic wavelengths and ionic concentrations using the commercially available RF coils and RF pulse sequences in two MR systems.

2.2. MRI Systems and RF Coils

Two MRI systems (Magnetom, 1.5T Aera and 3T Prisma, 64 MHz and 128 MHz, 1H NMR, proton nuclear magnetic resonance frequencies, both Syngo MR XA30A, Siemens Healthcare GmbH, Erlangen, Germany) were used in this study. In a previous study, five RF coil combinations, two Tx/Rx volume coils, a circularly polarized head coil (CP-H) and a body coil (BC), and three Rx-only phased-array surface coils, a 20-channel head and neck coil (20-H/N), a 32-channel head coil (32-H), and a 64-channel head and neck coil (64-H/N), were evaluated at 3T Prisma [37]. In this study, a widely used combination of RF coils, a BC Tx/Rx, and a 20-H/N Rx, in two Tx/Rx combinations, was used at both 1.5T and 3T to reduce the number of experiments. When the 20-H/N coil was turned off, the Tx/Rx was BC/BC, and when the 20-H/N coil was turned on, the Tx/Rx was BC/20-H/N. The RF polarization modes were circular polarization (CP) and elliptical polarization (EP). The 1.5T system had a no-tune BC Tx/Rx with 16 rungs, CP Tx, and 26.1 kW peak power, and the 3T system had a no-tune BC Tx/Rx with 32 rungs, multi-channel 2 (MC-2), CP and EP Tx, and 43.2 kW peak power (17.5 kW and 25.7 kW for channels 0 and 1). The BC specifications (maximum applied B1+ and B1+rms) were 27 μT and 6.2 μT for 1.5T, and 30 μT and 3.6 μT for 3T. The BC Tx field attenuation (3 dB and 10 dB) from the isocenter was 0.18 m and 0.38 m for 1.5T, and 0.15 m and 0.25 m for 3T.

2.3. Aqueous Solutions of NaCl-H2O

Aqueous solutions of sodium chloride (NaCl) and water (H2O) with four concentrations (C) of 0, 1, 2, and 3% NaCl by weight, corresponding to molar concentrations of 0, 171, 342, and 513 mM (millimole per liter), and specific gravities (SG) of 1.000, 1.005, 1.013, 1.020 kg/L (kilogram per liter), respectively, were used. These electrolytes with monovalent ions also correspond to 0, 171, 342, and 513 mEq/L (milliequivalent per liter) of Na+ (cations) and Cl− (anions). NaCl has a molecular weight of 58.44 g/mol (22.99 g/mol Na + 35.45 g/mol Cl) and a maximum solubility of 26.5% by weight or 4.53 M (mol/L) in H2O. The electrical conductivity (σ) of electrolytes varies with temperature, ionic concentration, and frequency [53,54]. Additionally, conductivity artifacts, where the RF signal decays from the perimeter to the center of the phantom, increase with ionic concentration [54]. For reference, σ values for the deionized water with 0, 1, 2, and 3% NaCl-H2O solutions were measured as 0.0016, 18.4, 34.1, and 48.5 mS/cm (milli Siemens per centimeter), respectively, using a conductivity meter (InLab 731-ISM, Mettler-Toledo, Columbus, OH 43240, USA). Measurement of the conductivity of electrolytes in the RF range requires a specialized method and costly instrumentation not readily available [53]. Here, MRI is also being tested as a remote sensor for RF coil loadings with different ionic concentrations at the RF frequencies of 64 MHz and 128 MHz.

For comparison, normal saline is 0.9% NaCl (9 g/L, 154 mM, 154 mEq/L Na+ and Cl−) and is used for the treatment of dehydration and fluid imbalance, and the hypertonic saline is 3% NaCl (30 g/L, 513 mM, 513 mEq/L Na+ and Cl−), and is used for treatment of hyponatremic encephalopathy, traumatic brain injury, and cerebral edema. The human serum has a normal range of 135 to 145 mEq/L Na+.

2.4. Phantoms

A phantom consisting of a conductive metallic wire immersed in a conductive ionic solution was used in this study. Since the implant materials, designs, and positions affect RF-induced heating, the same implant with different wavelengths was used to reduce the number of experiments [55,56,57,58]. Any conductive metallic wire, such as a DBS lead or any other wire, such as a 14-gauge (Ga) wire, can be used in this method, although with a different relationship between RF power, B1+rms, and SAR. A 14 Ga stranded wire with 64 strands, strand diameter of 0.2 ± 0.008 mm, copper-clad aluminum (CCA), 8% copper (Cu) cladding, 92% aluminum (Al) core, 4.1 μm Cu clad thickness, outside diameter (OD) of 1.85 mm, polyvinylchloride (PVC) insulation, average/minimum thickness of 0.425/0.375 mm, OD of 2.7 ± 0.1 mm, and an electrical resistance (R) of 14.1 Ohm/km) was used.

The penetration or skin depth (δ) of RF in a conductor is [59,60]

δ=πσfμrμ0−1/2 (6)

where σ is the electrical conductivity (S/m), f is the RF frequency (Hz), μr = μ/μ0 is the relative permeability of the material, the ratio of the permeability of the material (μ) to the permeability of free space (μ0).

The wavelength (λ) of RF in a medium is [61]

λ=λ0εr−1/2 (7)

where λ0 = c/f0 is the wavelength in the vacuum, the ratio of the speed of light in vacuum (c) to the frequency in vacuum (f0), and εr = ε/ε0 is the dielectric constant, the ratio of the permittivity of material (ε) to the permittivity of free space (ε0).

The electric permittivity and magnetic permeability of free space are related as

ε0=μ0c2−1 (8)

where ε0 = 8.854 × 10−12 F/m (Farad per meter), μ0 = 4π × 10−7 H/m (Henry per meter), and c = 299,792,458 m/s (meter per second) as in Equation (8) [62].

Using the electrical conductivity (σ) of 5.96 × 107 S/m for Cu and 3.77 × 107 S/m for Al, and μr of unity for Cu and Al in Equation (6), the RF skin depths at 64 MHz and 128 MHz were calculated as 8.15 μm and 5.76 μm for Cu and 10.3 μm and 7.25 μm for Al, respectively. The RF skin depth in both conductors at both frequencies is higher than the Cu clad thickness, making the conductivity of CCA wire a function of both conductors and both frequencies [63,64]. For reference, the conductivity of the CCA wire was calculated from its resistance as 3.5 × 107 S/m.

The metallic wire wavelengths (λ) of 0λ, λ/2, λ and 2λ corresponding to lengths (L) of 0, 26, 52 and 104 cm (1.5T) and 0, 13, 26 and 52 cm (3T) in water (εr = 81), as calculated by Equation (7), were used in combinations with the ionic concentrations (C) of 0, 1, 2 and 3% or 0, 171, 342, and 513 mM NaCl-H2O solutions.

As shown in Figure 1, the metallic wires of wavelengths of λ/2, λ and 2λ, corresponding to 13 cm (a), 26 cm (b), and 52 cm (c) at 3T or 26 cm (b), 52 cm (c) and 104 cm (d) at 1.5T, were attached at one end to the walls of poly methyl methacrylate (PMMA) plastic containers with dimensions of 22.86 cm, 11.125 cm, and 10.49 cm. A volume of 1.5L of each ionic solution with a height of 6.4 cm in the container was used, and the wire was positioned horizontally at 3.2 cm. The wire had a straight I-shape (a), a slightly curved C-shape (b), a U-shape (c), or a loop U-shape (d) in the phantom. The same metallic implant and trajectories were used in the phantoms when testing the ionic solutions, RF pulse sequences, and Tx/Rx combinations. The phantom with 0% NaCl and 0 λ or 0 cm implant had only deionized water and no wire.

Figure 1.

Figure 1

A phantom with conductive ionic solutions of NaCl-H2O, conductive metallic wire wavelengths of λ/2, λ, and 2λ corresponding to the metallic lengths of 13 cm (a), 26 cm (b), and 52 cm (c) at 3T or 26 cm (b), 52 cm (c), and 104 cm (d) at 1.5T.

2.5. RF Pulse Sequences

Five RF pulse sequences were prescribed to vary RF flip angles linearly as an independent parameter. A two-dimensional (2D) gradient echo (GRE) sequence with a variant of the fast low-angle shot (FLASH) sequence was modified with flip angles (θ) of 30, 45, 60, 75, and 90 degrees while keeping all other RF pulse parameters the same. The GRE sequences optimized with nominal parameters were taken from each of the scanners at 1.5T and 3T, respectively. The echo time (TE) was 4.76 ms and 2.46 ms, repetition time (TR) was 370 ms and 250 ms, slice thickness (z) was 5 mm and 4 mm, the total acquisition time (TA) was 69 and 77 s, pixel bandwidth (BW) was 189 Hz and 372 Hz at 1.5T and 3T, respectively. The number of slices (NS) was 3, the field of view (FOV) was 220 mm × 220 mm, and the matrix (m) size was 320 × 320. In the nominal sequences, an RF spatial saturation band (SB) of 50 mm with a 90-degree flip angle parallel to the axial slice was prescribed at 1.5T, but not at 3T. By default, the RF adjustment strategy was set to standard, B0 shimming was tune-up, B1 shimming was TrueForm, adjustment tolerance was auto, adjust with body coil was off, confirm frequency was never, and adjustment volume position was set to isocenter [65]. The RF adjustments, including B1 shimming (available at 3T but not at 1.5T), were performed by the scanner automatically [66]. Additionally, matched sequences with equal NS, SB, and TR were used to assess the difference between the nominal sequences at 1.5T and 3T. Since a previous study showed that RF power and the square of B1+rms scaled linearly with three flip angles (30, 60, 90 degrees) and two RF pulse types (fast and normal), only one RF pulse type (fast) was prescribed in this study [37].

These five RF sequences were repeated for two RF coil Tx/Rx combinations (BC/BC and BC/20-H/N), four concentrations (C) of aqueous solutions of NaCl-H2O (0, 1, 2 and 3%) and four wavelengths (λ) of wires (0λ, λ/2, λ and 2λ) corresponding to four lengths (L) of wires (0, 26, 52 and 104 cm at 1.5T, and 0, 13, 26 and 52 cm at 3T).

The scanner-reported RF power, B1+rms, and SAR for the whole body and head (SARwb and SARh) were recorded with high precision during the experiments. For consistency in SAR computation by the scanner’s algorithm, constant patient weight, height, and age (68 kg, 1.8 m, and 20 years) were used for all series. Additionally, transmit reference amplitude (TxRA) and transmit frequency (TxF) were recorded, and the absorption ratio (α) was computed from RF power and SARwb as α = m·SARwb/P for each of the experiments.

2.6. Experimental Design and Statistical Analysis

A full factorial design of experiments (DOE) with multiple factors and levels was used to study the variation of the dependent variables (P, B1+rms, SAR, TxRA, TxF, and α) with independent variables (θ, Tx/Rx, L, and C). The experiments consisted of 5 levels of θ (30, 45, 60, 75, and 90 degrees), 2 levels of Tx/Rx combinations (1 and 2), 4 levels of L (0λ, λ/2, λ and 2λ), and 4 levels of C (0, 1, 2 and 3%), and a total of 160 experiments (5 × 2 × 4 × 4) per B0 field and 320 experiments (160 × 2) for both B0 fields (1.5T and 3T), with each session spanning over several hours. Multiple replicates were not performed due to the high number of experiments. The linear regression analyses were performed for B1+rms vs. θ, P vs. the square of B1+rms, and P vs. SARwb, for Tx/Rx, L, and C, respectively.

A fractional factorial DOE was performed to assess the variation in RF power, B1+rms, and SAR with the number of slices, saturation bands, Tx/Rx, and repetition time using a standard phantom with an aqueous solution of 5 g/L NaCl and 3.75 g/L NiSO4·6H2O (Siemens, 5300 mL) without an implant at both 1.5T and 3T. Additionally, experiments were performed with the same standard phantom to assess the effects of saturation band thickness and slice thickness on P, B1+rms, and SAR using the same RF pulse sequences with three flip angles (30, 60, 90 degrees) and the Tx/Rx combination of BC/20-H/N.

3. Results

The scanner-reported time-averaged RF power (P), time-averaged root-mean-square RF field (B1+rms), and specific absorption rate (SAR) for the whole body (SARwb) and head (SARh) were evaluated as functions of flip angle (θ), RF coil Tx/Rx combinations (BC/BC and BC/20-H/N), metallic implant lengths (L), and ionic concentrations (C) of NaCl-H2O solutions. All linear regression and statistical analyses were based on the scanner-reported parameters.

The linear regression analyses were performed for B1+rms vs. θ, P vs. the square of B1+rms, and P vs. SARwb for 5 RF pulse sequences. The linear regression parameters (slope and intercept) and the coefficient of determination (R2) were computed for 160 experiments, comprising 32 linear fittings each at 1.5T and 3T, respectively. The slopes of dB1/dθ, dP/dB12, dP/dSwb, and the intercepts of B1(0), P(0), and P(0), along with the regression standard errors, are given in Table 1 and Table 2 for 1.5 and 3T, respectively. Here, RF power, B1+rms, SARwb, and SARh are denoted by P, B1, Swb, and Sh in the regression analysis.

Table 1.

Linear regression results (slope and intercept with standard errors) for B1+rms vs. θ, P vs. the square of B1+rms, and P vs. SARwb for RF coils, Tx/Rx, L, and C at 1.5T.

Aera 1.5T B1 vs. θ P vs. B12 P vs. SARwb
No. RF Coils
Tx/Rx (L, C)
dB1/dθ
[nT/deg]
B1 (0)
[nT]
dP/dB12
[W/μT2]
P (0)
[W]
dP/dSwb
[kg/α]
P (0)
[W]
1 BC/BC (0 cm, 0%) 3.5079 ± 0.2423 839.2774 ± 15.4182 8.9448 ± 0.0038 −0.0072 ± 0.0042 462.2415 ± 0.1941 −0.0072 ± 0.0042
2 BC/20-H/N (0 cm, 0%) 3.5079 ± 0.2423 839.2774 ± 15.4182 8.9448 ± 0.0038 −0.0072 ± 0.0042 462.2415 ± 0.1941 −0.0072 ± 0.0042
3 BC/BC (0 cm, 1%) 3.4941 ± 0.2413 835.9820 ± 15.3576 10.6169 ± 0.0092 0.0071 ± 0.0102 516.9831 ± 0.4459 0.0071 ± 0.0102
4 BC/20-H/N (0 cm, 1%) 3.4941 ± 0.2413 835.9820 ± 15.3576 10.6169 ± 0.0092 0.0071 ± 0.0102 516.9831 ± 0.4459 0.0071 ± 0.0102
5 BC/BC (0 cm, 2%) 3.4951 ± 0.2414 836.2311 ± 15.3622 14.5225 ± 0.0108 −0.0084 ± 0.0120 699.8340 ± 0.5193 −0.0084 ± 0.0120
6 BC/20-H/N (0 cm, 2%) 3.4951 ± 0.2414 836.2311 ± 15.3622 14.5225 ± 0.0108 −0.0084 ± 0.0120 699.8340 ± 0.5193 −0.0084 ± 0.0120
7 BC/BC (0 cm, 3%) 3.4965 ± 0.2415 836.5594 ± 15.3683 22.9795 ± 0.0045 −0.0113 ± 0.0050 849.6723 ± 0.1650 −0.0113 ± 0.0050
8 BC/20-H/N (0 cm, 3%) 3.4965 ± 0.2415 836.5594 ± 15.3683 22.9795 ± 0.0045 −0.0113 ± 0.0050 849.6723 ± 0.1650 −0.0113 ± 0.0050
9 BC/BC (26 cm, 0%) 3.5026 ± 0.2419 838.0267 ± 15.3952 8.4311 ± 0.0071 0.0036 ± 0.0079 433.1065 ± 0.3639 0.0036 ± 0.0079
10 BC/20-H/N (26 cm, 0%) 3.5026 ± 0.2419 838.0267 ± 15.3952 8.4311 ± 0.0071 0.0036 ± 0.0079 433.1065 ± 0.3639 0.0036 ± 0.0079
11 BC/BC (26 cm, 1%) 3.4981 ± 0.2416 836.9313 ± 15.3751 11.0085 ± 0.0074 0.0071 ± 0.0083 531.5672 ± 0.3584 0.0071 ± 0.0083
12 BC/20-H/N (26 cm, 1%) 3.4981 ± 0.2416 836.9313 ± 15.3751 11.0085 ± 0.0074 0.0071 ± 0.0083 531.5672 ± 0.3584 0.0071 ± 0.0083
13 BC/BC (26 cm, 2%) 3.4954 ± 0.2414 836.3007 ± 15.3635 15.9331 ± 0.0000 −0.0123 ± 0.0000 761.7769 ± 0.0000 −0.0123 ± 0.0000
14 BC/20-H/N (26 cm, 2%) 3.4954 ± 0.2414 836.3007 ± 15.3635 15.9331 ± 0.0000 −0.0123 ± 0.0000 761.7769 ± 0.0000 −0.0123 ± 0.0000
15 BC/BC (26 cm, 3%) 3.4999 ± 0.2417 837.3756 ± 15.3833 21.0570 ± 0.0055 −0.0050 ± 0.0062 911.6638 ± 0.2394 −0.0050 ± 0.0062
16 BC/20-H/N (26 cm, 3%) 3.4999 ± 0.2417 837.3756 ± 15.3833 21.0570 ± 0.0055 −0.0050 ± 0.0062 911.6638 ± 0.2394 −0.0050 ± 0.0062
17 BC/BC (52 cm, 0%) 3.4970 ± 0.2415 836.6827 ± 15.3706 20.5929 ± 0.0084 0.0003 ± 0.0094 933.1221 ± 0.3820 0.0003 ± 0.0094
18 BC/20-H/N (52 cm, 0%) 3.4970 ± 0.2415 836.6827 ± 15.3706 20.5929 ± 0.0084 0.0003 ± 0.0094 933.1221 ± 0.3820 0.0003 ± 0.0094
19 BC/BC (52 cm, 1%) 3.4971 ± 0.2415 836.7101 ± 15.3710 13.7038 ± 0.0113 −0.0074 ± 0.0125 676.6085 ± 0.5554 −0.0074 ± 0.0125
20 BC/20-H/N (52 cm, 1%) 3.4971 ± 0.2415 836.7101 ± 15.3710 13.7038 ± 0.0113 −0.0074 ± 0.0125 676.6085 ± 0.5554 −0.0074 ± 0.0125
21 BC/BC (52 cm, 2%) 3.4966 ± 0.2415 836.5728 ± 15.3684 18.0586 ± 0.0108 −0.0028 ± 0.0120 881.7400 ± 0.5257 −0.0028 ± 0.0120
22 BC/20-H/N (52 cm, 2%) 3.4966 ± 0.2415 836.5728 ± 15.3684 18.0586 ± 0.0108 −0.0028 ± 0.0120 881.7400 ± 0.5257 −0.0028 ± 0.0120
23 BC/BC (52 cm, 3%) 3.4944 ± 0.2413 836.0545 ± 15.3590 21.4671 ± 0.0085 −0.0165 ± 0.0094 896.9432 ± 0.3538 −0.0165 ± 0.0094
24 BC/20-H/N (52 cm, 3%) 3.4944 ± 0.2413 836.0545 ± 15.3590 21.4671 ± 0.0085 −0.0165 ± 0.0094 896.9432 ± 0.3538 −0.0165 ± 0.0094
25 BC/BC (104 cm, 0%) 3.5023 ± 0.2419 837.9548 ± 15.3939 12.8766 ± 0.0055 0.0109 ± 0.0062 663.8650 ± 0.2847 0.0109 ± 0.0062
26 BC/20-H/N (104 cm, 0%) 3.5023 ± 0.2419 837.9548 ± 15.3939 12.8766 ± 0.0055 0.0109 ± 0.0062 663.8650 ± 0.2847 0.0109 ± 0.0062
27 BC/BC (104 cm, 1%) 3.4937 ± 0.2413 835.8888 ± 15.3560 14.5720 ± 0.0061 0.0022 ± 0.0068 723.3868 ± 0.3034 0.0022 ± 0.0068
28 BC/20-H/N (104 cm, 1%) 3.4937 ± 0.2413 835.8888 ± 15.3560 14.5720 ± 0.0061 0.0022 ± 0.0068 723.3868 ± 0.3034 0.0022 ± 0.0068
29 BC/BC (104 cm, 2%) 3.4948 ± 0.2414 836.1600 ± 15.3610 21.8251 ± 0.0118 0.0040 ± 0.0131 882.8685 ± 0.4766 0.0040 ± 0.0131
30 BC/20-H/N (104 cm, 2%) 3.4948 ± 0.2414 836.1600 ± 15.3610 21.8251 ± 0.0118 0.0040 ± 0.0131 882.8685 ± 0.4766 0.0040 ± 0.0131
31 BC/BC (104 cm, 3%) 3.4958 ± 0.2414 836.3860 ± 15.3650 23.3877 ± 0.0111 −0.0118 ± 0.0123 841.0524 ± 0.3987 −0.0117 ± 0.0123
32 BC/20-H/N (104 cm, 3%) 3.4958 ± 0.2414 836.3860 ± 15.3650 23.3877 ± 0.0111 −0.0118 ± 0.0123 841.0524 ± 0.3987 −0.0117 ± 0.0123

Table 2.

Linear regression results (slope and intercept with standard errors) for B1+rms vs. θ, P vs. the square of B1+rms, and P vs. SARwb for RF coils, Tx/Rx, L, and C at 3T.

Prisma 3T B1 vs. θ P vs. B12 P vs. SARwb
No. RF Coils
Tx/Rx (L, C)
dB1/dθ
[nT/deg]
B1 (0)
[nT]
dP/dB12
[W/μT2]
P (0)
[W]
dP/dSwb
[kg/α]
P (0)
[W]
1 BC/BC (0 cm, 0%) 9.4412 ± 0.0000 0.0000 ± 0.0000 3.2715 ± 0.0026 −0.0010 ± 0.0011 76.6099 ± 0.0610 −0.0010 ± 0.0011
2 BC/20-H/N (0 cm, 0%) 9.4173 ± 0.0000 0.0000 ± 0.0000 3.4138 ± 0.0029 −0.0021 ± 0.0012 79.9424 ± 0.0686 −0.0021 ± 0.0012
3 BC/BC (0 cm, 1%) 9.4790 ± 0.0000 0.0000 ± 0.0000 5.5152 ± 0.0038 0.0050 ± 0.0016 110.5703 ± 0.0759 0.0050 ± 0.0016
4 BC/20-H/N (0 cm, 1%) 9.4187 ± 0.0000 0.0000 ± 0.0000 7.8605 ± 0.0049 −0.0001 ± 0.0021 109.3546 ± 0.0685 −0.0001 ± 0.0021
5 BC/BC (0 cm, 2%) 9.5024 ± 0.0000 0.0000 ± 0.0000 7.5683 ± 0.0058 0.0043 ± 0.0025 110.5145 ± 0.0840 0.0043 ± 0.0025
6 BC/20-H/N (0 cm, 2%) 9.4298 ± 0.0000 0.0000 ± 0.0000 15.9123 ± 0.0079 −0.0006 ± 0.0034 88.5539 ± 0.0441 −0.0006 ± 0.0034
7 BC/BC (0 cm, 3%) 9.5023 ± 0.0000 0.0000 ± 0.0000 7.7695 ± 0.0070 0.0008 ± 0.0030 111.1972 ± 0.1007 0.0008 ± 0.0030
8 BC/20-H/N (0 cm, 3%) 9.4345 ± 0.0000 0.0000 ± 0.0000 19.7675 ± 0.0045 −0.0040 ± 0.0019 75.7073 ± 0.0171 −0.0040 ± 0.0019
9 BC/BC (13 cm, 0%) 9.4090 ± 0.0000 0.0000 ± 0.0000 3.0122 ± 0.0000 0.0000 ± 0.0000 70.5385 ± 0.0000 0.0000 ± 0.0000
10 BC/20-H/N (13 cm, 0%) 9.4211 ± 0.0000 0.0000 ± 0.0000 3.0986 ± 0.0029 0.0021 ± 0.0012 72.5624 ± 0.0685 0.0021 ± 0.0012
11 BC/BC (13 cm, 1%) 9.4699 ± 0.0000 0.0000 ± 0.0000 5.6690 ± 0.0026 0.0010 ± 0.0011 110.5300 ± 0.0505 0.0010 ± 0.0011
12 BC/20-H/N (13 cm, 1%) 9.4216 ± 0.0000 0.0000 ± 0.0000 9.3126 ± 0.0055 −0.0039 ± 0.0024 109.1341 ± 0.0649 −0.0039 ± 0.0024
13 BC/BC (13 cm, 2%) 9.5155 ± 0.0000 0.0000 ± 0.0000 7.1181 ± 0.0061 −0.0023 ± 0.0026 110.8938 ± 0.0943 −0.0023 ± 0.0026
14 BC/20-H/N (13 cm, 2%) 9.4465 ± 0.0000 0.0000 ± 0.0000 17.7112 ± 0.0026 −0.0010 ± 0.0011 81.2992 ± 0.0119 −0.0010 ± 0.0011
15 BC/BC (13 cm, 3%) 9.4835 ± 0.0000 0.0000 ± 0.0000 8.6972 ± 0.0055 −0.0039 ± 0.0024 111.0424 ± 0.0697 −0.0039 ± 0.0024
16 BC/20-H/N (13 cm, 3%) 9.4424 ± 0.0000 0.0000 ± 0.0000 27.0554 ± 0.0076 0.0005 ± 0.0033 69.3702 ± 0.0195 0.0005 ± 0.0033
17 BC/BC (26 cm, 0%) 9.4294 ± 0.0000 0.0000 ± 0.0000 1.5187 ± 0.0056 −0.0029 ± 0.0024 35.5636 ± 0.1306 −0.0029 ± 0.0024
18 BC/20-H/N (26 cm, 0%) 9.3764 ± 0.0000 0.0000 ± 0.0000 1.2947 ± 0.0026 −0.0010 ± 0.0011 30.3195 ± 0.0618 −0.0010 ± 0.0011
19 BC/BC (26 cm, 1%) 9.4882 ± 0.0000 0.0000 ± 0.0000 6.0481 ± 0.0030 −0.0011 ± 0.0013 111.6514 ± 0.0562 −0.0011 ± 0.0013
20 BC/20-H/N (26 cm, 1%) 9.4626 ± 0.0000 0.0000 ± 0.0000 12.0841 ± 0.0094 −0.0002 ± 0.0040 107.6392 ± 0.0836 −0.0002 ± 0.0040
21 BC/BC (26 cm, 2%) 9.5022 ± 0.0000 0.0000 ± 0.0000 6.8777 ± 0.0061 0.0030 ± 0.0027 110.9341 ± 0.0989 0.0030 ± 0.0027
22 BC/20-H/N (26 cm, 2%) 9.4645 ± 0.0000 0.0000 ± 0.0000 24.4226 ± 0.0051 0.0018 ± 0.0022 71.0778 ± 0.0148 0.0018 ± 0.0022
23 BC/BC (26 cm, 3%) 9.5049 ± 0.0000 0.0000 ± 0.0000 6.4879 ± 0.0029 0.0021 ± 0.0012 111.2169 ± 0.0493 0.0021 ± 0.0012
24 BC/20-H/N (26 cm, 3%) 8.5705 ± 0.5031 39.2144 ± 32.0185 37.1368 ± 0.0061 −0.0011 ± 0.0024 65.2964 ± 0.0106 −0.0011 ± 0.0024
25 BC/BC (52 cm, 0%) 9.4751 ± 0.0000 0.0000 ± 0.0000 3.6826 ± 0.0026 0.0010 ± 0.0011 86.2377 ± 0.0605 0.0010 ± 0.0011
26 BC/20-H/N (52 cm, 0%) 9.4154 ± 0.0000 0.0000 ± 0.0000 4.1841 ± 0.0095 −0.0002 ± 0.0040 97.9804 ± 0.2221 −0.0002 ± 0.0040
27 BC/BC (52 cm, 1%) 9.4833 ± 0.0000 0.0000 ± 0.0000 5.9493 ± 0.0055 −0.0029 ± 0.0024 112.2265 ± 0.1040 −0.0029 ± 0.0024
28 BC/20-H/N (52 cm, 1%) 9.4394 ± 0.0000 0.0000 ± 0.0000 10.4000 ± 0.0038 −0.0050 ± 0.0016 108.2694 ± 0.0398 −0.0050 ± 0.0016
29 BC/BC (52 cm, 2%) 9.4984 ± 0.0000 0.0000 ± 0.0000 7.6349 ± 0.0061 0.0023 ± 0.0026 111.3332 ± 0.0886 0.0023 ± 0.0026
30 BC/20-H/N (52 cm, 2%) 9.4276 ± 0.0000 0.0000 ± 0.0000 19.6711 ± 0.0039 −0.0028 ± 0.0017 75.7476 ± 0.0150 −0.0028 ± 0.0017
31 BC/BC (52 cm, 3%) 9.4731 ± 0.0000 0.0000 ± 0.0000 9.3410 ± 0.0082 −0.0010 ± 0.0035 116.2077 ± 0.1020 −0.0010 ± 0.0035
32 BC/20-H/N (52 cm, 3%) 9.4726 ± 0.0000 0.0000 ± 0.0000 23.9288 ± 0.0049 0.0001 ± 0.0021 71.2071 ± 0.0145 0.0001 ± 0.0021

The results (slopes, intercepts, and standard errors) for Tx/Rx combinations were identical in Table 1 but not in Table 2 due to the absence and presence of B1 shimming at 1.5T and 3T, respectively.

B1+rms varied linearly with θ, proportional to Equation (1), whereas P and SAR varied linearly with the square of B1+rms and the square of θ, proportional to Equations (4) and (5). The dB1/dθ did not change with any of the experiments as a function of Tx/Rx, L, and C, while dP/dB12 and dP/dSw changed with experiments as a function of Tx/Rx (3T), L, and C, as given in Table 1 and Table 2. The difference in dB1/dθ, 3.5 nT/deg vs. 9.5 nT/deg in Table 1 and Table 2 was due to the differences in the nominal RF pulse sequences at 1.5T (50 mm SB and TR of 370 ms) and at 3T (no SB and TR of 250 ms), respectively. In Table 1, the intercept of B1(0) = 839 nT for water with no implant, corresponding to a flip angle of zero degrees, is the B1+rms used for the SB in the nominal sequence at 1.5T for all the experiments.

The R2 values of linear regression of B1+rms vs. θ were less than 1 (R2 < 1) in Table 1 because of the prescription of SB at 1.5T. The R2 values of linear regression were 1 for all fittings in Table 2 because of the absence of SB at 3T, except for one: the regression of B1+rms vs. θ (L = 26 cm, C = 3% and Tx/Rx = BC/20-H/N). The B1+rms for θ = 90° was less than usual (0.7855 μT vs. 0.8525 μT), and a regression of B1+rms of 0.7855 μT vs. θ for this experiment calculated a θ = 83° instead of 90°, which can be confirmed by the linearity in the plot of P vs. B12 for this experiment at 3T (red closed markers). The R2 values of linear regressions of P vs. B12 and P vs. SARwb in Table 1 and Table 2 were all equal to 1 (R2 = 1) at 1.5T and 3T, respectively.

Table 3 shows the slopes and intercepts of dB1/dθ and B1(0) for B1 vs. θ, dP/dB12 and P(0) for P vs. B12, and the R2 for the experiments with matched RF sequences (NS, SB, Tx/Rx, and TR) at 1.5T and 3T, respectively. The slopes and intercepts, along with the R2, were computed by fitting B1 vs. θ and P vs. B12 by varying NS, SB, Tx/Rx, and TR. The NS, SB, and TR prescribed in the RF pulse sequences had effects on the B1+rms, P, and SAR at both 1.5T and 3T.

Table 3.

Linear regression results (slope and intercept with R2) for B1+rms vs. θ and P vs. the square of B1+rms as a function of NS, SB, Tx/Rx, and TR for a standard phantom with an aqueous solution of 5 g/L NaCl and 3.75 g/L NiSO4·6H2O without an implant at 1.5T and 3T.

B0 = 1.5T
B1 vs. θ P vs. B12
NS SB Tx/Rx TR
[ms]
dB1/dθ
[nT/deg]
B1 (0)
[nT]
R2 dP/dB12
[W/μT2]
P (0)
[W]
R2
1 1 2 250 2.4824 593.9244 0.9859 9.2459 0.0017 1.0000
1 1 2 370 2.0405 488.2029 0.9859 9.3522 −0.0090 1.0000
3 1 2 250 4.2996 1028.7073 0.9859 9.2561 −0.0141 1.0000
3 1 2 370 3.5343 845.5922 0.9859 9.3251 0.0056 1.0000
1 2 2 250 5.5634 0.0000 1.0000 9.2459 0.0010 1.0000
1 2 2 370 4.5731 0.0000 1.0000 9.3002 0.0043 1.0000
3 2 2 250 9.6361 0.0000 1.0000 9.2561 −0.0028 1.0000
3 2 2 370 7.9208 0.0000 1.0000 9.3154 0.0050 1.0000
3 1 1 370 3.5332 845.3501 0.9859 9.3207 0.0085 1.0000
3 2 1 370 7.9185 0.0000 1.0000 9.3305 0.0006 1.0000
B0 = 3T
B1 vs. θ P vs. B12
NS SB Tx/Rx TR
[ms]
dB1/dθ
[nT/deg]
B1 (0)
[nT]
R2 dP/dB12
[W/μT2]
P (0)
[W]
R2
1 1 2 250 2.8633 462.4106 0.9881 7.7708 0.0034 1.0000
1 1 2 370 2.3537 380.0992 0.9881 7.8077 0.0064 1.0000
3 1 2 250 4.9595 800.9186 0.9881 9.1130 −1.4165 0.9867
3 1 2 370 4.0766 658.3511 0.9881 7.8374 −0.0033 1.0000
1 2 2 250 5.5254 0.0000 1.0000 7.8220 −0.0032 1.0000
1 2 2 370 4.5418 0.0000 1.0000 7.8143 0.0032 1.0000
3 2 2 250 9.5703 0.0000 1.0000 7.7945 −0.0033 1.0000
3 2 2 370 7.8667 0.0000 1.0000 7.8374 0.0017 1.0000
3 1 1 250 4.9723 802.9972 0.9881 7.9491 0.0007 1.0000
3 2 1 250 9.5951 0.0000 1.0000 7.9543 −0.0019 1.0000

SB (1: on, 2: off), Tx/Rx (1: BC/BC, 2: BC/20-H/N).

As shown in Table 3, the B1+rms varied linearly with θ proportional to Equation (1), independent of the magnetic field strength (B0), when the same pulse sequence parameters were used. The dB1/dθ was equal at 1.5T and 3T, independent of the B0 field, when SB was off, but not equal when SB was on. The variation in B1+rms vs. θ was linear (R2 = 1) when SB was off, but nonlinear (R2 < 1) or quadratic (R2 = 1) when SB was on. The slope dB1/dθ and the intercept B1(0) varied distinctly when SB was on or off. The intercept corresponding to B1+rms for zero-degree flip angle was zero, B1(0) = 0, when SB was off and nonzero, B1(0) ≠ 0, when SB was on. The slope, dB1/dθ, and intercept, B1(0), had a linear relationship when SB was on. The slope, dB1/dθ, decreased, and the intercept, B1(0), increased when SB was on.

RF power varied linearly with the square of B1+rms and the square of θ, consistent with Equations (4) and (5), but with different slopes for each of the experiments as a function of Tx/Rx (3T), metallic wavelengths, and ionic concentrations, as in Table 1 and Table 2.

The plots of P vs. the square of B1+rms for NaCl-H2O solutions with NaCl concentrations of 0% (cyan squares, ☐), 1% (blue triangles, Δ), 2% (green diamonds, ♢) and 3% (red circles, 〇) with metallic implant wavelengths of 0 (a), λ/2 (b), λ (c) and 2λ (d) corresponding to the lengths (L) of 0 cm (a), 26 cm (b), 52 cm (c) and 104 cm (d) at 1.5T, and 0 cm (a), 13 cm (b), 26 cm (c) and 52 cm (d) at 3T for Tx/Rx combinations of BC/BC (open markers) and BC/20-H/N (closed markers) with the linear regression lines (dotted lines) and linear regression coefficients (legends) are given in Figure 2 and Figure 3, respectively.

Figure 2.

Figure 2

RF power vs. square of B1+rms for NaCl-H2O solutions with four conductive ionic concentrations, 0% (☐), 1% (Δ), 2% (♢) and 3% (〇), four conductive metallic implant wavelengths, 0λ (a), λ/2 (b), λ (c), and 2λ (d), corresponding to four lengths, 0 cm (a), 26 cm (b), 52 cm (c) and 104 cm (d), and two Tx/Rx combinations, BC/BC (open markers) and BC/20-H/N (closed markers) at 1.5T. Note that closed markers are overlaid on open markers at 1.5T.

Figure 3.

Figure 3

RF power vs. square of B1+rms for NaCl-H2O solutions with four conductive ionic concentrations, 0% (☐), 1% (Δ), 2% (♢) and 3% (〇), four conductive metallic implant wavelengths, 0λ (a), λ/2 (b), λ (c), and 2λ (d), corresponding to four lengths, 0 cm (a), 13 cm (b), 26 cm (c), and 52 cm (d), and two Tx/Rx combinations, BC/BC (open markers) and BC/20-H/N (closed markers) at 3T. Note that closed markers are not overlaid on open markers at 3T.

RF power varied linearly with whole-body SAR proportional to Equation (3) but with different slopes (m/α) and absorption ratios (α) for each of the experiments as a function of Tx/Rx (3T), metallic wavelengths, and ionic concentrations.

The plots of P vs. the SARwb for NaCl-H2O solutions with NaCl concentrations of 0% (cyan squares, ☐), 1% (blue triangles, Δ), 2% (green diamonds, ♢) and 3% (red circles, 〇) with metallic implant wavelengths of 0λ (a), λ/2 (b), λ (c) and 2λ (d) corresponding to the lengths (L) of 0 cm (a), 26 cm (b), 52 cm (c) and 104 cm (d) at 1.5T, and 0 cm (a), 13 cm (b), 26 cm (c) and 52 cm (d) at 3T for Tx/Rx combinations of BC/BC (open markers) and BC/20-H/N (closed markers) with the linear regression lines (dotted lines) and linear regression coefficients (legends) are given in Figure 4 and Figure 5, respectively.

Figure 4.

Figure 4

RF power vs. whole-body SAR for NaCl-H2O solutions with four conductive ionic concentrations, 0% (☐), 1% (Δ), 2% (♢) and 3% (〇), four conductive metallic implant wavelengths, 0λ (a), λ/2 (b), λ (c), and 2λ (d), corresponding to four lengths, 0 cm (a), 26 cm (b), 52 cm (c) and 104 cm (d), and two Tx/Rx combinations, BC/BC (open markers) and BC/20-H/N (closed markers) at 1.5T. Note that closed markers are overlaid on open markers at 1.5T.

Figure 5.

Figure 5

RF power vs. whole-body SAR for NaCl-H2O solutions with four conductive ionic concentrations, 0% (☐), 1% (Δ), 2% (♢) and 3% (〇), four conductive metallic implant wavelengths, 0λ (a), λ/2 (b), λ (c), and 2λ (d), corresponding to four lengths, 0 cm (a), 13 cm (b), 26 cm (c), and 52 cm (d), and two Tx/Rx combinations, BC/BC (open markers) and BC/20-H/N (closed markers) at 3T. Note that closed markers are not overlaid on open markers at 3T.

In Figure 2, Figure 3, Figure 4 and Figure 5, the plots with no metallic implants (L = 0 cm, upper left plots, a) show the effects of the conductive ionic concentrations (0, 1, 2, and 3% NaCl-H2O) and the plots with no ionic solutions (0% NaCl-H2O, cyan lines and square markers ☐) show the effects of the conductive metallic implant wavelengths of 0λ (a), λ/2 (b), λ (c) and 2λ (d) (corresponding to lengths of 0 cm (a), 26 cm (b), 52 cm (c), and 104 cm (d) at 1.5T and 0 cm (a), 13 cm (b), 26 cm (c), and 52 cm (d) at 3T) on the RF power, B1+rms and SARwb for RF coil Tx/Rx combinations of BC/BC (open markers) and BC/20-H/N (closed markers), at 1.5T and 3T, respectively.

Note that in Figure 2 and Figure 4, corresponding to 1.5T, there were no differences in Tx/Rx combinations (closed markers were overlaid on the open markers), whereas in Figure 3 and Figure 5, corresponding to 3T, there were differences in Tx/Rx combinations (closed markers were not overlaid on the open markers) in plots of P vs. B12 and P vs. SARwb, respectively. This difference is due to the absence and presence of B1 shimming at 1.5T and 3T, respectively.

The slopes of dB1/dθ varied slightly from 3.49 to 3.51 nT/deg at 1.5T and from 9.38 to 9.52 nT/deg at 3T across all experiments with varying metallic wavelengths and ionic concentrations (Table 1 and Table 2).

The slopes of dP/dB12 varied from 8.4311 W/μT2 (Tx/Rx = 1 and 2, λ/2 = 26 cm, C = 0%) to 23.3877 W/μT2 (Tx/Rx = 1 and 2, 2λ = 104 cm, C = 3%) at 1.5T and from 1.2947 W/μT2 (Tx/Rx = 2, λ = 26 cm, C = 0%) to 37.1368 W/μT2 (Tx/Rx = 2, λ = 26 cm, C = 3%) at 3T with Tx/Rx, metallic wavelengths and ionic concentrations for all experiments (Table 1 and Table 2, and Figure 2 and Figure 3).

The slopes of dP/dSwb, equal to the ratio of patient weight to absorption ratio (m/α) proportional to Equation (3), varied from 433.1065 kg/α (Tx/Rx = 1 and 2, λ/2 = 26 cm, C = 0%) to 933.1221 kg/α (Tx/Rx = 1 and 2, λ = 52 cm, C = 0%) corresponding to α = 0.157 to α = 0.073 at 1.5T and from 30.3195 kg/α (Tx/Rx = 2, λ = 26 cm, C = 0%) to 116.2077 kg/α (Tx/Rx = 1, 2λ = 52 cm, C = 3%) corresponding to α = 2.243 to α = 0.585 at 3T, varying with Tx/Rx, metallic wavelengths, and ionic concentrations for all experiments (Table 1 and Table 2, and Figure 4 and Figure 5). The larger α values close to unity (α ≤ 1) indicate overestimation, and the outliers larger than unity (α > 1) indicate artifacts in overestimation of the whole-body SAR vs. average RF power at 3T, varying with RF adjustments, metallic wavelengths, and ionic concentrations.

The slopes of linear regressions of P vs. SAR for head and whole-body, dP/dSh and dP/Swb scaled with each other linearly, for the two Tx/Rx combinations of BC/BC and BC/20-H/N, with ratios of 0.1257 and 0.1257 at 1.5T and 0.1798 and 0.2223 at 3T, which were simply equal to the reciprocals of the ratios of SARh and SARwb, with 7.9554 and 7.9554 at 1.5T and 5.5605 and 4.4977 at 3T, respectively, for all experiments varying with metallic implant wavelengths and ionic concentrations. There was no difference in Tx/Rx at 1.5T, but there was a difference at 3T in the SAR ratio. This was due to the differences in RF adjustments and B1 shimming at 1.5T and 3T, respectively.

Using the descriptive statistics, the mean plus and minus one standard deviation (m ± sd) calculated for all 160 experiments at each B0 field without replicates was 1.0467 ± 0.0750 μT and 0.5669 ± 0.2007 μT (B1+rms), 17.8827 ± 6.1112 W and 3.7476 ± 4.2007 W (RF power), 0.0241 ± 0.0045 W/kg and 0.0451 ± 0.0596 W/kg (SARwb), 421.6163 ± 66.6073 V and 360.4076 ± 132.7007 V (TxRA), 0.0991 ± 0.0265 and 0.8270 ± 0.3613 (α) at 1.5T and 3T, respectively. Further tests with multiple replications may be necessary across MRI systems, RF coils, and implants.

The variation in dependent variables (RF power, B1+rms, whole-body specific absorption rate, transmit reference amplitude, transmit frequency, and absorption ratio) was evaluated as a function of independent variables (flip angle, RF coil transmit and receive combinations, metallic wavelengths, and ionic concentrations) for 1.5T and 3T.

The boxplots of dependent variables of P (W), B1+rms (μT), SARwb (W/kg), TxRA (V), TxF (MHz), and α (rows, from top to bottom, respectively) vs. independent variables of θ (30, 45, 60, 75, 90 degrees), Tx/Rx (1 and 2), L (0, 26, 52, and 104 cm at 1.5T, and 0, 13, 26, and 52 cm at 3T), and C (0, 1, 2, and 3%) (columns, from left to right, respectively) with statistics (minimum, first quartile, median, third quartile, maximum, and outliers) are given in Figure 6 and Figure 7 for 1.5T and 3T, respectively.

Figure 6.

Figure 6

Boxplots of dependent variables of P (W), B1+rms (μT), SARwb (W/kg), TxRA (V), TxF (MHz) and α (rows, from top to bottom, respectively) vs. independent variables of θ (30, 45, 60, 75, 90°), Tx/Rx (1: BC/BC, 2: BC/20-H/N), L (0, 26, 52, 104 cm) and C (0, 1, 2, 3%) (columns, from left to right, respectively) with statistics (minimum, first quartile, median, third quartile, maximum, and outliers) at 1.5T.

Figure 7.

Figure 7

Boxplots of dependent variables of P (W), B1+rms (μT), SARwb (W/kg), TxRA (V), TxF (MHz) and α (rows, from top to bottom, respectively) vs. independent variables of θ (30, 45, 60, 75, 90°), Tx/Rx (1: BC/BC, 2: BC/20-H/N), L (0, 13, 26, 52 cm) and C (0, 1, 2, 3%) (columns, from left to right, respectively) with statistics (minimum, first quartile, median, third quartile, maximum, and outliers) at 3T.

P and B1 both increased with SB while maintaining the P/B12 nearly constant in a standard phantom without an implant, as in Table 3. The ratios of P and B1+rms with and without SB, P(SB)/P and B1(SB)/B1, for five RF pulse flip angles (30, 45, 60, 75, 90 degrees) and equal sequence parameters (NS = 3, TR = 250 ms, Tx/Rx = 2) were 16.47, 7.84, 4.85, 3.47, 2.71 and 4.05, 2.80, 2.20, 1.86, 1.65 at 1.5T, and 11.31, 5.58, 3.56, 2.64, 2.31, and 3.36, 2.36, 1.89, 1.63, 1.46 at 3T, respectively. Note that the ratio of P(SB)/P scales with the square of the ratio of B1(SB)/B1 at both 1.5T and 3T.

P and B12 varied linearly with NS and 1/TR as in Equation (5). Both P and B1 increased with NS and decreased with TR while maintaining a close ratio of P/B12 in a standard phantom without an implant when turning the SB on or off at both 1.5T and 3T, as in Table 3. For example, the dP/dB12 values for five RF pulse flip angles and equal sequence parameters (NS = 1, TR = 250 ms, Tx/Rx = 2) with and without SB (1st and 5th rows) were 9.24 and 9.24 W/μT2 at 1.5T, and 7.77 and 7.82 W/μT2 at 3T, respectively.

B1+rms vs. θ varied with SB thickness in a step function where dB1/dθ took one value, 9.6 nT/deg, for an SB thickness of 0 mm, and one value, 6.6 nT/deg, for SB thicknesses of 3, 4, and 6 mm, and then another value, 4.9 nT/deg, for SB thicknesses of 12, 25, and 50 mm. The slice thickness (z), varied from 4 mm to 16 mm (4, 8, and 16 mm), did not have any effect on P, B1+rms, and SAR, and the linear regression coefficients, dB1/dθ, dP/dB12, and dP/dSwb, for three flip angles (30, 60, and 90 degrees) at 1.5T and 3T.

P and B1+rms increased with the thickness of SB in a step function while maintaining a close ratio of P/B12 at 1.5T and 3T in a standard phantom without an implant. The ratios of power and B1+rms with and without SB, P(SB)/P and B1(SB)/B1, for three flip angles (30, 60, 90 degrees) were 5.02, 2.00, 1.44, and 2.23, 1.41, 1.20 for SB thicknesses of 3, 4, and 6 mm, and 11.32, 3.57, 2.14, and 3.36, 1.89, 1.46 for SB thicknesses of 12, 25, and 50 mm, respectively. The dP/dB12 were 8.06, 8.04, and 8.06 W/μT2 for SB thicknesses of 0 mm, and SB thicknesses of 3, 4, and 6 mm, and SB thicknesses of 12, 25, and 50 mm, respectively.

4. Discussion

In a previous study, the relationship between RF-induced heating, RF power, and B1+rms was shown to vary with RF coils and Tx/Rx combinations: two Tx/Rx volume coils (a body coil and a head coil) and three Rx-only phased-array coils (a 20-channel head and neck, a 32-channel head, and a 64-channel head and neck) in combination with a Tx/Rx body coil at one B0 field (3T) [37]. The temperature rise, when stratifying the data for individual RF coils and Tx/Rx combinations, had a high linear relation with both RF power and the square of B1+rms at 3T. However, when combining the data for all RF coils and Tx/Rx combinations, the temperature rise had a better linear relation with RF power (R2 = 0.9232) than the square of B1+rms (R2 = 0.4945) at 3T, indicating the importance of the specifications of RF coils and Tx/Rx combinations.

In this study, only the most widely used Tx/Rx combinations of RF coils (a body coil and a 20-channel head and neck coil) were used at two B0 fields (1.5T and 3T) to drastically reduce the number of experiments. Each additional RF coil would add 160 experiments for each B0 field and 320 experiments for both B0 fields.

Here, the relationship between the scanner-reported RF power, B1+rms, and SAR is evaluated as a function of RF pulse sequences, RF coil Tx/Rx combinations, metallic wavelengths, and ionic concentrations. The RF power and SAR varied with RF pulse sequences, RF coil Tx/Rx combinations, metallic wavelengths, and ionic concentrations, whereas B1+rms varied with RF pulse sequences only, but not with Tx/Rx combinations, metallic wavelengths, and ionic concentrations.

The combination of body coil and head coil for transmit and receive had a significantly large effect on RF power, specific absorption rate, transmit reference amplitude, transmit frequency, and absorption ratio due to B1 shimming, which was available at 3T but not at 1.5T. The RF power and SAR for B1+rms corresponding to a prescribed flip angle change significantly with the RF adjustment strategies and B1 shimming algorithms [65].

The RF transmit field (B1) is described by three components: the positive component (B1+), the negative component (B1−), and the magnitude of both positive and negative components (|B1|). The effective field that flips the spins in the rotating frame of reference in synchronization with the precession of the spins is the B1+ field, and it is defined as the time-averaged root-mean-square of the spatially averaged field (B1+rms) in the central axial slab [66]. The scanner adjusts the RF power in the transmit coil during the pre-scan mode until the desired B1+ field is reached for the prescribed flip angle [67,68,69]. As part of RF adjustments, the B1 shimming (B1 and E field) process adjusts the RF power per RF coil Tx/Rx combinations and RF polarization while maintaining the B1+ field required for the prescribed flip angle [70,71,72]. This changes the relationship between the RF power and B1+rms depending on the RF coils, Tx/Rx combinations, and experimental conditions [37,38,39,40]. As a result, RF power changes, but B1 remains nearly the same for the Tx/Rx combinations, metallic wavelengths, and ionic concentrations.

As shown in Table 1 and Table 2, the dB1/dθ did not change, whereas dP/dB12 and dP/dSw changed with the experimental parameters such as Tx/Rx (3T), metallic wavelengths, and ionic concentrations. The slopes of dP/dB12 and dP/dSw varied with conductive metallic wavelengths and ionic concentrations at both 1.5T and 3T, but varied markedly with Tx/Rx combinations at 3T due to B1 shimming. This can be seen by comparing the dP/dB12 in Figure 2 and Figure 3 and the dP/dSw in Figure 4 and Figure 5 for 1.5T and 3T, respectively.

B1+rms varied linearly with θ, proportional to Equation (1), independent of the B0 field, Tx/Rx, L, and C when using the same RF pulse sequences. Initially, the differences in the average dB1/dθ were observed between the two B0 fields as 3.5 nT/deg at 1.5T and 9.5 nT/deg at 3T (Table 1 and Table 2). This was further investigated with matched RF sequences (equal NS, SB, and TR) and it was found that the difference was due to differences in the prescription of SB and TR in the nominal RF pulse sequences at 1.5T and 3T (Table 3). Note the effects of SB and TR of 370 ms at 1.5T vs. no SB and the TR of 250 ms at 3T, as shown in Table 1, Table 2 and Table 3 at 1.5T and 3T. When SB was on, the slope dB1/dθ decreased, but the intercept B1(0) increased at both 1.5T and 3T. Without SB, dB1/dθ was equal at both B0 fields (1.5T and 3T) at equal or scaled TR (Table 3). The square of the ratios of dB1/dθ scaled with the ratios of 1/TR; for example, (9.6/7.9)2 equals (370/250), as shown in Table 3. P varied linearly with B12 and 1/TR, proportional to Equations (4) and (5) at both 1.5T and 3T. The dB1/dθ did not change with Tx/Rx combinations, but it changed with NS, SB, and TR. Since the B1+rms in RF pulse sequences with SB is not sensitive to the presence or absence of metallic implants, ionic solutions, implant wavelengths, or ionic concentrations, care must be taken to avoid SB because SB increases both P and B1+rms significantly.

Since SAR is derived from RF power, their ratio can yield the absorption ratio as α = m·SARwb/P from Equation (3), or α = m/(dP/dSwb) from Table 1 and Table 2, which can reveal information about SAR computing algorithms of various scanners. A clear distinction must be made between measured and scanner-reported SAR when calculating the α because the scanner-reported SAR may yield α values greater than unity (α > 1, Pabs > P) that are not physically possible. The actual SAR and α can be calculated by measuring the electric field or the initial temperature rate for incident P as in Equations (2) and (3), respectively. The scanner-reported head and whole-body SAR scaled linearly with each other across all experiments, depending on the Tx/Rx combinations. Even though the scanners maintained constant SAR ratios throughout the experiments, a 68 kg patient weight and a small phantom in the head coil might affect the SAR computation. However, using the same phantom size in all experiments shows the variation of parameters with conductive implants and ionic solutions.

As shown in Figure 6 and Figure 7, RF power and SAR changed with flip angles, Tx/Rx (3T), implant wavelength, and ionic concentration, whereas B1+rms changed only with flip angles but not with Tx/Rx, implant wavelength, and ionic concentration. The transmit reference amplitude, transmit frequency, and absorption ratio did not change with flip angle but changed with Tx/Rx (3T), implant wavelength, and ionic concentration. There were more scattered data and outliers observed at 3T than at 1.5T, mainly due to the RF adjustments and B1 shimming effects present at 3T but not at 1.5T.

While RF power and SAR both vary with the experimental conditions, the ratio between RF power and SAR is highly variable and unpredictable with the experimental conditions. Also, scanner-reported SAR computation for the RF pulse sequences may vary with vendor-dependent algorithms and software versions. These experiments show that the most sensitive parameter appears to be the RF power, as it varies significantly with the experimental conditions for a given B1+rms, which varies only with the RF pulse sequence but not the experimental conditions. RF power is most responsive to the presence or absence of conductive metallic implants and ionic solutions, metallic wavelengths, and ionic concentrations. However, RF power also varies with RF coils, RF pulse sequences, RF adjustment strategies (B1 shimming), as well as patient size and weight. This presents a complex problem for the implementation of RF power in safety standards and technical specifications such as ISO/TS 10974 across MRI systems, implants, and patients [33].

The relationship between RF power, B1+rms, and SAR depends on the RF pulse sequences, RF coils, Tx/Rx combinations, RF adjustments, and B1 shimming strategies, the presence or absence of metallic implants and ionic solutions, implant wavelengths, and ionic concentrations. Experimental measurements are consistent with the simple theoretical formulations through the instrumentation proportionality constants, which can be determined for each of the RF coils and RF pulse sequences experimentally.

The limitation of this study is that two MRI systems (Siemens Aera and Prisma) with the same software versions (Syngo MR XA30A), two RF coil Tx/Rx combinations (BC/BC and BC/20-H/N), a simple metallic conductor (CCA wire), and a simple conductive phantom (H2O-NaCl solutions) were used to study the relations between RF power, B1+rms, and SAR. Further tests may be necessary to verify the results across MRI systems, RF coils, implants, and patients.

This experimental method provides a means to simultaneously evaluate the relationship between the scanner-reported RF power, B1+rms, and SAR varying with RF pulse sequences, RF coils, Tx/Rx combinations, metallic implant wavelengths, and ionic concentrations during MRI. This method can be used with temperature measurements to evaluate specifically each of the MRI systems, RF coils, RF pulse sequences, conductive metallic implants, and tissue-equivalent conductive ionic solutions to optimize parameters for implant safety.

5. Conclusions

The scanner-reported parameters provide noninvasive means and complementary measurements that can be improved and combined to improve RF safety. B1+rms is insensitive to the presence or absence of conductive metallic implants and ionic solutions, metallic wavelengths, ionic concentrations, RF coils, and Tx/Rx combinations. B1+rms alone may not be sufficient to monitor and control RF safety. B1+rms varies with flip angle, independent of the B0 field, RF coils, Tx/Rx combinations, metallic wavelengths, and ionic concentrations using the same RF pulse sequences. The scanner-reported SAR scales linearly with the scanner-reported average RF power and the square of flip angles, but with different absorption ratios varying with experimental conditions, and remains unpredictable. The scanner-reported average RF power is responsive to the experimental conditions and remains the most sensitive parameter to consider in relation to temperature and improving SAR computation as complementary to B1+rms to monitor and control RF safety. The scanner-reported RF power is sensitive to RF adjustments, metallic wavelengths, and ionic concentrations. Since the RF power was highly associated with RF heating and temperature rise of different magnitudes varying with individual RF coils and B1 shimming in a previous study, the response in RF power with experimental conditions may be considered and further studied in addition to B1+rms to improve the development of standards and specifications for implant safety [37,38,39,40]. Because these three scanner-reported parameters, RF power, B1+rms, and SAR, are noninvasive and readily available on the scanners, understanding the fundamental relationships between them and the temperature across the MRI systems, implants, and patients can help improve implant safety during MRI.

Prescribing RF saturation bands in RF pulse sequences increases RF power, B1+rms, and SAR by orders of magnitude, especially when using low-power RF pulse sequences; therefore, saturation bands must be turned off when imaging patients with conductive metallic implants. The RF power, B1+rms, and SAR increase linearly with increasing number of slices and decrease with increasing repetition time.

Acknowledgments

This work was supported and performed at the Rockefeller Neuroscience Institute, Department of Neuroradiology, School of Medicine, West Virginia University, Morgantown, WV 26506, USA.

Abbreviations

The following abbreviations are used in this manuscript:

MRI Magnetic Resonance Imaging
RF Radiofrequency
E Root-mean-square electric field, Volts/m, or V/m
B1+rms Root-mean-square magnetic field, micro-Tesla, nano-Tesla or μT, nT
B0 Magnetic field strength, Tesla or T
γ Gyromagnetic ratio, rads/s·T
P RF power, Watts or kilowatts, W or kW
Tx/Rx Transmit and receive
BC Body Coil
20-H/N 20 channel Head and Neck coil
m Patient weight, kg
SAR Specific absorption rate, W/kg
SARwb Whole body SAR, W/kg
SARh Head SAR, W/kg
H2O-NaCl Water and sodium chloride
σ Electrical conductivity, Siemens per meter or S/m
θ or FA Flip angle, rads or degrees
λ Wavelength, cm
L Length, cm
C Concentration, mM or %
α Calculated absorption ratio, dimensionless
D Duty cycle, dimensionless
ω Angular frequency, rads/s
CIED Cardiac implantable electronic devices
DBS Deep brain stimulation
tp Pulse length, milliseconds or ms
TR Repetition time, milliseconds or ms
NS Number of slices, dimensionless
SB Saturation bands, millimeters or mm
TxRA Transmit reference amplitude, Volts or V
TxF Transmit frequency, Hertz or Mega Hertz, Hz or MHz

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

The original contributions presented in this study are included in the article. Further inquiries can be directed to the corresponding author.

Conflicts of Interest

The author declares no conflicts of interest.

Funding Statement

This research received no external funding.

Footnotes

Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

References

  • 1.AAOS . American Joint Replacement Registry (AJRR): 2025 Annual Report. American Academy of Orthopaedic Surgeons (AAOS); Rosemont, IL, USA: 2025. [Google Scholar]
  • 2.Al-Khatib S.M. Cardiac Implantable Electronic Devices. N. Engl. J. Med. 2024;390:442–454. doi: 10.1056/NEJMra2308353. [DOI] [PubMed] [Google Scholar]
  • 3.Akinyele B., Marine J.E., Love C., Crawford T.C., Chrispin J., Vlay S.C., Spragg D.D., Eagle K.A., Berger R.D., Calkins H., et al. Unregulated Online Sales of Cardiac Implantable Electronic Devices in the United States: A Six-Month Assessment. Heart Rhythm O2. 2020;1:235–238. doi: 10.1016/j.hroo.2020.06.007. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Martinez-Nunez A.E., Rozell C.J., Little S., Tan H., Schmidt S.L., Grill W.M., Pajic M., Turner D.A., de Hemptinne C., Machado A., et al. Proceedings of the 12th Annual Deep Brain Stimulation Think Tank: Cutting Edge Technology Meets Novel Applications. Front. Hum. Neurosci. 2025;19:1544994. doi: 10.3389/fnhum.2025.1544994. Correction in Front. Hum. Neurosci. 2025, 19, 1612584. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Rezai A.R., Kopell B.H., Gross R.E., Vitek J.L., Sharan A.D., Limousin P., Benabid A.-L. Deep Brain Stimulation for Parkinson’s Disease: Surgical Issues. Mov. Disord. 2006;21:S197–S218. doi: 10.1002/mds.20956. [DOI] [PubMed] [Google Scholar]
  • 6.Youngerman B.E., Chan A.K., Mikell C.B., McKhann G.M., Sheth S.A. A Decade of Emerging Indications: Deep Brain Stimulation in the United States. J. Neurosurg. 2016;125:461–471. doi: 10.3171/2015.7.JNS142599. [DOI] [PubMed] [Google Scholar]
  • 7.Lozano A.M., Lipsman N., Bergman H., Brown P., Chabardes S., Chang J.W., Matthews K., McIntyre C.C., Schlaepfer T.E., Schulder M., et al. Deep Brain Stimulation: Current Challenges and Future Directions. Nat. Rev. Neurol. 2019;15:148–160. doi: 10.1038/s41582-018-0128-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Brandmeir N.J., Murray A., Cheyuo C., Ferari C., Rezai A.R. Deep Brain Stimulation for Multiple Sclerosis Tremor: A Meta-Analysis. Neuromodulation. 2020;23:463–468. doi: 10.1111/ner.13063. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Ranjan M., Ranjan N., Deogaonkar M., Rezai A. Deep Brain Stimulation for Refractory Depression, Obsessive-Compulsive Disorder and Addiction. Neurol. India. 2020;68:S282–S287. doi: 10.4103/0028-3886.302459. [DOI] [PubMed] [Google Scholar]
  • 10.Liu C.Y., Russin J., Adelson D.P., Jenkins A., Hilmi O., Brown B., Lega B., Whitworth T., Bhattacharyya D., Schwartz T.H., et al. Vagus Nerve Stimulation Paired with Rehabilitation for Stroke: Implantation Experience from the VNS-REHAB Trial. J. Clin. Neurosci. 2022;105:122–128. doi: 10.1016/j.jocn.2022.09.013. [DOI] [PubMed] [Google Scholar]
  • 11.Mahoney J.J., 3rd, Koch-Gallup N., Scarisbrick D.M., Berry J.H., Rezai A.R. Deep Brain Stimulation for Psychiatric Disorders and Behavioral/Cognitive-Related Indications: Review of the Literature and Implications for Treatment. J. Neurol. Sci. 2022;437:120253. doi: 10.1016/j.jns.2022.120253. [DOI] [PubMed] [Google Scholar]
  • 12.Peltola J., Colon A.J., Pimentel J., Coenen V.A., Gil-Nagel A., Gonçalves Ferreira A., Lehtimäki K., Ryvlin P., Taylor R.S., Ackermans L., et al. Deep Brain Stimulation of the Anterior Nucleus of the Thalamus in Drug-Resistant Epilepsy in the MORE Multicenter Patient Registry. Neurology. 2023;100:e1852–e1865. doi: 10.1212/WNL.0000000000206887. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Rezai A.R., Mahoney J.J., Ranjan M., Haut M.W., Zheng W., Lander L.R., Berry J.H., Farmer D.L., Marton J.L., Tirumalai P., et al. Safety and Feasibility Clinical Trial of Nucleus Accumbens Deep Brain Stimulation for Treatment-Refractory Opioid Use Disorder. J. Neurosurg. 2024;140:231–239. doi: 10.3171/2023.4.JNS23114. [DOI] [PubMed] [Google Scholar]
  • 14.Rezai A.R., Finelli D., Rugieri P., Tkach J., Nyenhuis J.A., Shellock F.G. Neurostimulators: Potential for Excessive Heating of Deep Brain Stimulation Electrodes during Magnetic Resonance Imaging. J. Magn. Reson. Imaging. 2001;14:488–489. doi: 10.1002/jmri.1212. [DOI] [PubMed] [Google Scholar]
  • 15.Kim D., Collins J.D., White J.A., Hanneman K., Lee D.C., Patel A.R., Hu P., Litt H., Weinsaft J.W., Davids R., et al. SCMR Expert Consensus Statement for Cardiovascular Magnetic Resonance of Patients with a Cardiac Implantable Electronic Device. J. Cardiovasc. Magn. Reson. 2024;26:100995. doi: 10.1016/j.jocmr.2024.100995. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Russo R.J., Gakenheimer-Smith L., Birgersdotter-Green U.M., Han J.K., Krahn A.D., Larsen T.R., Litt H.I., Liu C.F., Nazarian S., Woodard P.K., et al. HRS Call-to-Action: Improved MRI Access for Patients with Cardiovascular Implantable Electronic Devices. Heart Rhythm. 2025;22:e821–e836. doi: 10.1016/j.hrthm.2025.04.028. [DOI] [PubMed] [Google Scholar]
  • 17.Seo Y., Wang Z.J. Measurement and Evaluation of Specific Absorption Rate and Temperature Elevation Caused by an Artificial Hip Joint during MRI Scanning. Sci. Rep. 2021;11:1134. doi: 10.1038/s41598-020-80828-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Shellock F.G., Karacozoff A.M. Reference Manual for Magnetic Resonance Safety, Implants, and Devices. Biomedical Research Publishing Group; Los Angeles, CA, USA: 2020. [Google Scholar]
  • 19.Rezai A.R., Finelli D., Nyenhuis J.A., Hrdlicka G., Tkach J., Sharan A., Rugieri P., Stypulkowski P.H., Shellock F.G. Neurostimulation Systems for Deep Brain Stimulation: In Vitro Evaluation of Magnetic Resonance Imaging-Related Heating at 1.5 Tesla. J. Magn. Reson. Imaging. 2002;15:241–250. doi: 10.1002/jmri.10069. [DOI] [PubMed] [Google Scholar]
  • 20.Shrivastava D., Abosch A., Hanson T., Tian J., Gupte A., Iaizzo P.A., Vaughan J.T. Effect of the Extracranial Deep Brain Stimulation Lead on Radiofrequency Heating at 9.4 Tesla (400.2 MHz) J. Magn. Reson. Imaging. 2010;32:600–607. doi: 10.1002/jmri.22292. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Golestanirad L., Kirsch J., Bonmassar G., Downs S., Elahi B., Martin A., Iacono M.-I., Angelone L.M., Keil B., Wald L.L., et al. RF-Induced Heating in Tissue near Bilateral DBS Implants during MRI at 1.5 T and 3T: The Role of Surgical Lead Management. NeuroImage. 2019;184:566–576. doi: 10.1016/j.neuroimage.2018.09.034. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Boutet A., Chow C.T., Narang K., Elias G.J.B., Neudorfer C., Germann J., Ranjan M., Loh A., Martin A.J., Kucharczyk W., et al. Improving Safety of MRI in Patients with Deep Brain Stimulation Devices. Radiology. 2020;296:250–262. doi: 10.1148/radiol.2020192291. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Winter L., Seifert F., Zilberti L., Murbach M., Ittermann B. MRI-Related Heating of Implants and Devices: A Review. J. Magn. Reson. Imaging. 2021;53:1646–1665. doi: 10.1002/jmri.27194. [DOI] [PubMed] [Google Scholar]
  • 24.Abbott . MRI Procedure Information, Abbott Medical MR Conditional Deep Brain Stimulation Systems, Clinician’s Manual. Abbott Neuromodulation; Austin, TX, USA: 2022. [(accessed on 5 September 2025)]. Available online: www.neuromodulation.abbott.com. [Google Scholar]
  • 25.Medtronic . MRI Guidelines for Medtronic Deep Brain Stimulation Systems 2020-06-01. Medtronic, Inc.; Minneapolis, MN, USA: 2020. [Google Scholar]
  • 26.Smith & Nephew . MRI Safety Information and Parameters for Smith & Nephew Orthopaedics AG Hip Implants. Smith & Nephew Orthopaedics AG; Baar, Switzerland: 2023. [Google Scholar]
  • 27.Delfino J.G., Krainak D.M., Flesher S.A., Miller D.L. MRI-Related FDA Adverse Event Reports: A 10-Yr Review. Med. Phys. 2019;46:5562–5571. doi: 10.1002/mp.13768. [DOI] [PubMed] [Google Scholar]
  • 28.Bottomley P.A., Andrew E.R. RF Magnetic Field Penetration, Phase Shift and Power Dissipation in Biological Tissue: Implications for NMR Imaging. Phys. Med. Biol. 1978;23:630–643. doi: 10.1088/0031-9155/23/4/006. [DOI] [PubMed] [Google Scholar]
  • 29.Bottomley P.A. Turning up the Heat on MRI. J. Am. Coll. Radiol. 2008;5:853–855. doi: 10.1016/j.jacr.2008.04.003. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Henderson J.M., Tkach J., Phillips M., Baker K., Shellock F.G., Rezai A.R. Permanent Neurological Deficit Related to Magnetic Resonance Imaging in a Patient with Implanted Deep Brain Stimulation Electrodes for Parkinson’s Disease: Case Report. Neurosurgery. 2005;57:E1063; discussion E1063. doi: 10.1227/01.neu.0000180810.16964.3e. [DOI] [PubMed] [Google Scholar]
  • 31.Vaughan J.T. eMagRes. John Wiley & Sons; Hoboken, NJ, USA: 2020. RF Coils and MR Safety; pp. 101–112. [Google Scholar]
  • 32.ISO Implants for Surgery—Active Implantable Medical Devices—Part 3: Implantable Neurostimulators. International Organization for Standardization; Geneva, Switzerland: 2017. [Google Scholar]
  • 33.Assessment of the Safety of Magnetic Resonance Imaging for Patients with an Active Implantable Medical Device. International Organization for Standardization; Geneva, Switzerland: 2018. [Google Scholar]
  • 34.ICNIRP Guidelines for Limiting Exposure to Electromagnetic Fields (100 kHz to 300 GHz) Health Phys. 2020;118:483–524. doi: 10.1097/HP.0000000000001210. [DOI] [PubMed] [Google Scholar]
  • 35.IEC Medical Electrical Equipment—Part 2-33: Particular Requirements for the Basic Safety and Essential Performance of Magnetic Resonance Equipment for Medical Diagnosis. IEC; Geneva, Switzerland: 2022. [Google Scholar]
  • 36.FDA . Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment, Guidance for Industry and Food and Drug Administration Staff. U.S. Department of Health and Human Services, Food and Drug Administration, Center for Devices and Radiological Health; Silver Spring, MD, USA: 2021. [Google Scholar]
  • 37.Gultekin D.H., Vaughan J.T., Shrivastava D. Experimental Evaluation of Relationship Between Radiofrequency Heating Near Implanted Conductive Devices, Scanner-Reported B1+rms, and Transmit Power. IEEE J. Microw. 2025;5:518–525. doi: 10.1109/JMW.2025.3550087. [DOI] [Google Scholar]
  • 38.Steckner M., Conroy M., Edmonson J., Goren T. Comments on “Experimental Evaluation of Relationship Between Radiofrequency Heating Near Implanted Conductive Devices, Scanner-Reported B1+rms, and Transmit Power”. IEEE J. Microw. 2026;6:489–490. doi: 10.1109/JMW.2026.3666625. [DOI] [Google Scholar]
  • 39.Gultekin D.H., Vaughan J.T., Shrivastava D. Reply to Comments on ‘Experimental Evaluation of Relationship Between Radiofrequency Heating Near Implanted Conductive Devices, Scanner-Reported B1+rms, and Transmit Power’. IEEE J. Microw. 2026;6:491–492. doi: 10.1109/JMW.2026.3666608. [DOI] [Google Scholar]
  • 40.Steckner M., Conroy M., Edmonson J., Goren T. Final Response to Authors of “Experimental Evaluation of Relationship Between Radiofrequency Heating Near Implanted Conductive Devices, Scanner-Reported B1+rms, and Transmit Power”. IEEE J. Microw. 2026;6:493–494. doi: 10.1109/JMW.2026.3666607. [DOI] [Google Scholar]
  • 41.Gultekin D.H., Siegel P.H. Absorption of 5G Radiation in Brain Tissue as a Function of Frequency, Power and Time. IEEE Access. 2020;8:115593–115612. doi: 10.1109/ACCESS.2020.3002183. [DOI] [Google Scholar]
  • 42.Larmor J. On the Theory of the Magnetic Influence on Spectra; and on the Radiation from Moving Ions. Lond. Edinb. Dublin Philos. Mag. J. Sci. 1897;44:503–512. doi: 10.1080/14786449708621095. [DOI] [Google Scholar]
  • 43.Bloch F. Nuclear Induction. Phys. Rev. 1946;70:460–474. doi: 10.1103/PhysRev.70.460. [DOI] [Google Scholar]
  • 44.Bottomley P.A., Edelstein W.A., Leue W.M., Hart H.R., Schenck J.F., Redington R.W. Head and Body Imaging by Hydrogen Nuclear Magnetic Resonance. Magn. Reson. Imaging. 1982;1:69–74. doi: 10.1016/0730-725X(82)90221-1. [DOI] [PubMed] [Google Scholar]
  • 45.IEEE Standard for Safety Levels with Respect to Human Exposure to Electric, Magnetic, and Electromagnetic Fields, 0 Hz to 300 GHz. IEEE; New York, NY, USA: 2019. pp. 1–312. [DOI] [Google Scholar]
  • 46.Bottomley P.A., Edelstein W.A. Power Deposition in Whole-Body NMR Imaging. Med. Phys. 1981;8:510–512. doi: 10.1118/1.595000. [DOI] [PubMed] [Google Scholar]
  • 47.Bottomley P.A., Redington R.W., Edelstein W.A., Schenck J.F. Estimating Radiofrequency Power Deposition in Body NMR Imaging. Magn. Reson. Med. 1985;2:336–349. doi: 10.1002/mrm.1910020404. [DOI] [PubMed] [Google Scholar]
  • 48.Röschmann P. Radiofrequency Penetration and Absorption in the Human Body: Limitations to High-Field Whole-Body Nuclear Magnetic Resonance Imaging. Med. Phys. 1987;14:922–931. doi: 10.1118/1.595995. [DOI] [PubMed] [Google Scholar]
  • 49.Schaefer D.J. Safety Aspects of Radiofrequency Power Deposition in Magnetic Resonance. Magn. Reson. Imaging Clin. N. Am. 1998;6:775–789. doi: 10.1016/s1064-9689(21)00481-5. [DOI] [PubMed] [Google Scholar]
  • 50.Bottomley P.A., Roemer P.B. Homogeneous Tissue Model Estimates of RF Power Deposition in Human NMR Studies. Local Elevations Predicted in Surface Coil Decoupling. Ann. N. Y. Acad. Sci. 1992;649:144–159. doi: 10.1111/j.1749-6632.1992.tb49604.x. [DOI] [PubMed] [Google Scholar]
  • 51.Vaughan J.T.J. Radio Frequency Coils Are the “lenses” of the MRI System. NMR BioMed. 2009;22:907. doi: 10.1002/nbm.1459. [DOI] [PubMed] [Google Scholar]
  • 52.Vaughan J.T., Jr., Lagore R. eMagRes. John Wiley & Sons; Hoboken, NJ, USA: 2020. RF Coil Construction for MRI; pp. 159–172. [Google Scholar]
  • 53.Giovannetti G., Frijia F., Menichetti L., Hartwig V., Viti V., Landini L. An Efficient Method for Electrical Conductivity Measurement in the RF Range. Concepts Magn. Reson. Part B Magn. Reson. Eng. 2010;37B:160–166. doi: 10.1002/cmr.b.20165. [DOI] [Google Scholar]
  • 54.Yang Q.X., Wang J., Zhang X., Collins C.M., Smith M.B., Liu H., Zhu X.-H., Vaughan J.T., Ugurbil K., Chen W. Analysis of Wave Behavior in Lossy Dielectric Samples at High Field. Magn. Reson. Med. 2002;47:982–989. doi: 10.1002/mrm.10137. [DOI] [PubMed] [Google Scholar]
  • 55.Bassen H., Kainz W., Mendoza G., Kellom T. MRI-Induced Heating of Selected Thin Wire Metallic Implants—Laboratory and Computational Studies—Findings and New Questions Raised. Minim. Invasive Ther. Allied Technol. 2006;15:76–84. doi: 10.1080/13645700600640931. [DOI] [PubMed] [Google Scholar]
  • 56.Bottomley P.A., Kumar A., Edelstein W.A., Allen J.M., Karmarkar P.V. Designing Passive MRI-Safe Implantable Conducting Leads with Electrodes. Med. Phys. 2010;37:3828–3843. doi: 10.1118/1.3439590. [DOI] [PubMed] [Google Scholar]
  • 57.Sonmez M., Saikus C.E., Bell J.A., Franson D.N., Halabi M., Faranesh A.Z., Ozturk C., Lederman R.J., Kocaturk O. MRI Active Guidewire with an Embedded Temperature Probe and Providing a Distinct Tip Signal to Enhance Clinical Safety. J. Cardiovasc. Magn. Reson. 2012;14:38. doi: 10.1186/1532-429X-14-38. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 58.Kazemivalipour E., Keil B., Vali A., Rajan S., Elahi B., Atalar E., Wald L.L., Rosenow J., Pilitsis J., Golestanirad L. Reconfigurable MRI Technology for Low-SAR Imaging of Deep Brain Stimulation at 3T: Application in Bilateral Leads, Fully-Implanted Systems, and Surgically Modified Lead Trajectories. NeuroImage. 2019;199:18–29. doi: 10.1016/j.neuroimage.2019.05.015. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 59.Giovannetti G., Flori A., Frijia F. Conductor Losses in Radiofrequency Coils for Magnetic Resonance below 3T: Estimation Methods and Minimization Strategies. Sensors. 2023;23:5586. doi: 10.3390/s23125586. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 60.Barta R., Volotovskyy V., Wachowicz K., Fallone B.G., De Zanche N. How Thin Can You Go? Performance of Thin Copper and Aluminum RF Coil Conductors. Magn. Reson. Med. 2021;85:2327–2333. doi: 10.1002/mrm.28540. [DOI] [PubMed] [Google Scholar]
  • 61.Webb A.G. Dielectric Materials in Magnetic Resonance. Concepts Magn. Reson. Part A. 2011;38A:148–184. doi: 10.1002/cmr.a.20219. [DOI] [Google Scholar]
  • 62.Mohr P.J., Newell D.B., Taylor B.N., Tiesinga E. CODATA Recommended Values of the Fundamental Physical Constants: 2022. Rev. Mod. Phys. 2025;97:025002. doi: 10.1103/RevModPhys.97.025002. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 63.Wojda R., Galigekere V.P., Pries J., Onar O. Proceedings of the 2020 IEEE Transportation Electrification Conference & Expo (ITEC) IEEE; New York, NY, USA: 2020. Copper-Clad Aluminum Windings as an Alternative Conductor for High-Power Electric Vehicle Wireless Charging; pp. 1197–1200. [Google Scholar]
  • 64.Giovannetti G. Full-Wave Simulation of Radio Frequency Resonators for MRI: Strip and Wire Conductors Inductance Estimation. Rev. Sci. Instrum. 2025;96:121401. doi: 10.1063/5.0268009. [DOI] [PubMed] [Google Scholar]
  • 65.Siemens . Operator Manual—TimTX TrueForm, TimTX TrueShape, Syngo MR XA30. Siemens Healthcare GmbH; Erlangen, Germany: 2020. [Google Scholar]
  • 66.Siemens . Operator Manual—Scanning, Syngo MR XA30. Siemens Healthcare GmbH; Erlangen, Germany: 2020. [Google Scholar]
  • 67.Collins C.M. Radiofrequency Field Calculations for High Field MRI. In: Robitaille P.-M., Berliner L., editors. Ultra High Field Magnetic Resonance Imaging. Springer; Boston, MA, USA: 2006. pp. 209–248. [Google Scholar]
  • 68.Ibrahim T.S., Tang L. Insight into RF Power Requirements and B1 Field Homogeneity for Human MRI via Rigorous FDTD Approach. J. Magn. Reson. Imaging. 2007;25:1235–1247. doi: 10.1002/jmri.20919. [DOI] [PubMed] [Google Scholar]
  • 69.Yao A., Li Z., Ma M. Impact of MRI RF Coil Design on the RF-Induced Heating of Medical Implants: Fixed B1+rms Exposure versus Normal Operating Mode. Phys. Med. Biol. 2024;69:055021. doi: 10.1088/1361-6560/ad2714. [DOI] [PubMed] [Google Scholar]
  • 70.Glover G.H., Hayes C.E., Pelc N.J., Edelstein W.A., Mueller O.M., Hart H.R., Hardy C.J., O’Donnell M., Barber W.D. Comparison of Linear and Circular Polarization for Magnetic Resonance Imaging. J. Magn. Reson. (1969) 1985;64:255–270. doi: 10.1016/0022-2364(85)90349-X. [DOI] [Google Scholar]
  • 71.Eryaman Y., Akin B., Atalar E. Reduction of Implant RF Heating through Modification of Transmit Coil Electric Field. Magn. Reson. Med. 2011;65:1305–1313. doi: 10.1002/mrm.22724. [DOI] [PubMed] [Google Scholar]
  • 72.Murbach M., Neufeld E., Cabot E., Zastrow E., Córcoles J., Kainz W., Kuster N. Virtual Population-Based Assessment of the Impact of 3 Tesla Radiofrequency Shimming and Thermoregulation on Safety and B1+ Uniformity. Magn. Reson. Med. 2016;76:986–997. doi: 10.1002/mrm.25986. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

The original contributions presented in this study are included in the article. Further inquiries can be directed to the corresponding author.