Response to Letter to the Editor refers to ‘Upgrade procedures of pacemakers or implantable cardioverter-defibrillators require creative clinical decision making and careful planning that begins at the first implantation’ by Alexander H. Maass et al., https://doi.org/10.1093/eurospace/euag174.
We thank Maass et al. for their thoughtful comments on our consensus statement on device upgrades and downgrades1 and welcome the opportunity to clarify several points.
We agree that many of the principles underpinning decision-making for device upgrades are equally applicable to the management of lead dysfunction. In both settings, the balance between lead extraction, lead abandonment and implantation of additional hardware should be tailored, taking into account patient characteristics, venous access, infection risk and future treatment requirements.
With regard to lead selection, we recognize that both DF-1 and DF-4 systems have advantages and disadvantages.2 While future flexibility is an important consideration, it represents only one of many factors influencing lead choice. Decisions at the time of initial ICD implantation should primarily reflect the patient's contemporary clinical needs and the available evidence, rather than a future scenario that may never arise. We therefore do not believe that current evidence supports the routine use of DF-1 leads in all patients undergoing ICD implantation in anticipation of possible future downgrading. We acknowledge the increasing recognition that, although DF-4 systems have simplified ICD implantation and become the contemporary standard, this has come at the expense of reduced flexibility in selected clinical scenarios. Although DF-1 systems may more readily facilitate selected downgrade strategies, integrated DF-4 leads simplify implantation by reducing header connections and opportunities for connection errors, reduce header and generator bulk, and have become the contemporary standard with an established safety and reliability profile. Neither system is universally superior; rather, each offers distinct advantages depending on the clinical scenario, and lead selection should therefore remain individualized rather than driven by hypothetical future requirements.
We agree that, in many patients undergoing ICD deactivation, replacement with a like-for-like ICD generator and deactivation of tachyarrhythmia therapies represents the simplest—and often safest—strategy. In the elderly and frail population in whom this is typically considered, implanting an additional pacing lead or replacing a functioning DF-4 lead may expose patients to disproportionate procedural risk.
Regarding venous access, we agree that contralateral lead implantation and tunnelling should generally be avoided where possible, given the cumulative lead burden within the superior vena cava and the potential implications for future venous access and device infection. We also acknowledge the authors’ experience with brachial venoplasty and agree that it represents a valuable technique in experienced hands. Although published series report excellent procedural success, these results derive largely from experienced centres and may not be readily generalizable to all operators or institutions.3 However, whether establishing venous access before opening the device pocket confers a net procedural advantage during upgrade procedures remains uncertain, as existing leads may still be vulnerable to inadvertent damage during venous instrumentation and the risk of introducing infection is not eliminated even if the device pocket remains unopened. Furthermore, despite successful venoplasty, definitive axillary or subclavian access is still required and elastic recoil may occasionally limit subsequent venous puncture. We therefore believe that the optimal strategy should continue to be guided by patient anatomy, operator expertise and procedural objectives.
We thank the authors for their valuable comments, which reinforce the central principle of the consensus statement that management of patients undergoing device revision should remain individualized.
Contributor Information
Daniel Keene, National Heart and Lung Institute, Imperial College London, London, UK.
Nandita Kaza, National Heart and Lung Institute, Imperial College London, London, UK.
Haran Burri, Department of Cardiology, University Hospital of Geneva, Geneva, Switzerland.
Archana Rao, Department of Cardiology, Liverpool Heart and Chest Hospital, Liverpool, UK.
Jens Cosedis Nielsen, Department of Cardiology, Aarhus University, Aarhus, Denmark.
Funding
This work received no grant from any funding agency in the public, commercial, or not for profit sectors.
Data availability
No new data were generated or analysed in support of this research.
References
- 1.Keene D, Nielsen JC, Burri H, Chavez-Gutierrez CA, Deharo JC, Drossart Iet al. Cardiac implantable electronic device upgrades and downgrades: a clinical consensus statement of the European heart rhythm association (EHRA) of the ESC, the Asia Pacific heart rhythm association (APHRS), Canadian heart rhythm society (CHRS), heart rhythm society (HRS), and the Latin American heart rhythm society (LAHRS). Europace 2025;27:euaf252. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Sticherling C, Ellenbogen KA, Burri H. Stepping back for good reasons: a reappraisal of the DF-1 connector for defibrillator leads. Europace 2024;26:euae057. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Lipsic E, Daniels F, Groenveld HF, Rienstra M, Maass AH. When and how to perform venoplasty for lead placement. Heart Rhythm 2024;21:1923–8. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
No new data were generated or analysed in support of this research.