Abstract
Objective
Cochlear implant (CI) surgery aims to achieve atraumatic electrode array insertion to preserve residual hearing and optimize outcomes. Tip fold-over, a mechanical complication during insertion, can compromise stimulation effectiveness and may require reinsertion or revision surgery. This study investigates the incidence, influencing factors, and diagnostic approaches for tip fold-over across a large patient cohort implanted with various CI electrode array designs.
Methods
We conducted a retrospective analysis of adult patients who underwent CI surgery between 2013 and 2024 at the Department of Otorhinolaryngology–Head and Neck Surgery, Medical Center–University of Freiburg. Included electrode arrays were from Cochlear™ (Slim Straight [SSA], Slim Modiolar [SMA], Contour Advance [CA]), MED-EL (FlexSoft, Flex28, Flex26, Flex24), and Advanced Bionics (SlimJ, Mid Scala [MS]). Detailed reviews of surgical reports, intraoperative electrophysiological data, and postoperative imaging—evaluated independently by at least two reviewers—were performed. Diagnostic imaging modalities included cone beam computed and digital volume tomography (DVT).
Results
Among 1,575 implanted ears, 13 cases of tip fold-over were detected, yielding an overall incidence of 0.83%. Notably, 10 of these occurred with precurved, perimodiolar arrays (SMA and CA, Cochlear™; MS, Advanced Bionics), while only 3 were found in straight lateral wall designs (SSA, Cochlear™). No tip fold-over events were observed with any MED-EL electrode arrays or with the SlimJ electrode array from Advanced Bionics. Intraoperative Transimpedance matrix (TIM) analysis and intraoperative X-ray reliably detected tip fold-over in cases with normal anatomy.
Conclusion
Electrode array design significantly influences tip fold-over risk. A potentially higher incidence has been observed with perimodiolar electrode arrays compared with flexible straight arrays. The importance of intraoperative diagnostics is underscored in this study. TIM analysis and intraoperative X-ray imaging proved effective in detecting tip fold-over events in real-time, allowing for immediate corrective action and avoiding revision surgeries. This large-scale retrospective study confirms that tip fold-over remains a rare but significant complication in CI surgery, with a higher prevalence in precurved perimodiolar electrode arrays. This study emphasizes the need for ongoing vigilance across all CI systems.
Keywords: cochlear implant, electrode array design, Perimodiolar array, straight electrode array, tip fold over
Introduction
An increasing number of cochlear implant (CI) electrode array designs have been developed to achieve the goal of atraumatic insertion (1). Due to individual variability in cochlear size (2–4), CI manufacturers offer a range of electrode array designs with varying diameters and lengths (5, 6). Furthermore, two main principles of stimulation are represented based on the position of the electrode array: perimodiolar electrode arrays (PM) with pre-curved designs aiming to stimulate near the modiolus (1, 7), while lateral wall electrode arrays are designed to be positioned farther from the modiolus along the lateral wall to minimize trauma within the cochlea (8, 9, 53). The most common form of cochlear trauma due to electrode insertion is scalar translocation, i.e., dislocation from one scala to another (2, 10). The goal is still a complete insertion of the electrode array into the scala tympani (11, 12, 54), as this provides the best conditions for hearing rehabilitation (13). The risk of insertion trauma has decreased over time due to greater surgical experience (14) and the development of less traumatic electrode arrays (1, 15–18, 52). A complete scala tympani insertion also plays an important role in preserving residual hearing (19) and may influence the scalar position of the electrode array if reinsertion is required in the future (16, 18, 20).
As described by Dhanasingh and Jolly (8), another potential trauma during electrode array insertion is tip fold-over, where the tip folds back on itself, potentially causing cochlear trauma and mis-stimulation and consequently a reduction of insertion depth. This complication has been reported in multiple studies before for the Slim Modiolar electrode array (SMA; models 532/632), and the Contour Advance electrode array (CA; models CI24RECA, CI412/512/612) from Cochlear Limited (NSW, Australia) (8, 21–31). Furthermore, other studies have reported the occurrence of tip folds in MED-EL (Innsbruck, Austria) electrode arrays (32) and in electrode arrays from Advanced Bionics (Valencia, United States) (31).
The aim of this study is to evaluate the occurrence of tip fold-over in a large patient cohort, to assess whether individual or institutional learning curves can reduce the incidence of this complication, and to determine whether intraoperative versus postoperative diagnostics—such as electrophysiology, Transimpedance matrix (TIM) analysis, and imaging—are effective in identifying tip fold-over and enabling prompt reinsertion, thereby avoiding the need for revision surgery.
Methods
This is a retrospective analysis of adult patients who received CIs between 2013 and 2024 at the Department of Otorhinolaryngology–Head and Neck Surgery, Medical Center–University of Freiburg. Patients implanted with devices from the following manufacturers and electrode arrays were included: the Slim Straight electrode array (SSA; models 422/522/622), the Slim Modiolar electrode array (SMA; models 532/632), and the Contour Advance electrode array (CA; models CI24RECA, CI412/512/612) from Cochlear Limited (NSW, Australia); the MED-EL FlexSoft, Flex28, Flex26 and Flex24 arrays from MED-EL (Innsbruck, Austria); and the SlimJ and MidScala (MS) arrays from Advanced Bionics (Valencia, United States).
A thorough review was conducted of all surgical reports, intraoperative measurements, and follow-up documentation including pre- and postoperative residual hearing. Additionally, postoperative imaging was re-evaluated by at least two independent authors of this study. Imaging modalities included cone beam computed tomography (CBCT) with a DynaCT-equipped Axiom Artis dTA angiography unit (Siemens Co., Erlangen, Germany) until 2022, and digital volume tomography (DVT) using the NewTom 5G/GXL system (Hillus Medical Engineering KG) from 2022 onward, as previously described by our research group [(1, 2, 11, 17, 33); (5, 20, 22, 52)]. We excluded patients with clear signs of cochlear fibrosis or abnormal anatomy, such as incomplete partition type III, as previous studies have described differing insertion conditions in these cases (34). All included ears demonstrated morphological characteristics within the range previously described by Ketterer et al. (2), with distances A and B assessed according to the definitions established by Escudé et al. (35).
This study was approved by the Freiburg medical center Ethics Committee (approval number: 406/19; amendment numbers: 240275), and is registered in the Freiburg Clinical Trials Register (FRKS number: FRKS006332) and the German Clinical Trials Register (DRKS number: DRKS00037881).
Results
Distribution and influence of electrode array design
After applying the inclusion criteria, a total of 1,575 implanted ears were analyzed. These procedures were performed between 2013 and 2024 at the Department of Otorhinolaryngology–Head and Neck Surgery, Medical Center–University of Freiburg. Devices from three manufacturers were included in the analysis: Cochlear™, MED-EL, and Advanced Bionics (see Table 1). Within this cohort, 13 cases of tip fold-over were identified, corresponding to an overall tip fold-over incidence rate of 0.83%. The majority of these events (10 cases) were associated with precurved, perimodiolar electrodes, specifically the SMA, CA, and one case in the MS array (see Figure 1). This represents a tip fold-over rate of 1.59% among the 629 precurved perimodiolar electrode arrays implanted. Focusing on the Advanced Bionics MS electrode array specifically, only one tip fold-over event was observed among 114 implanted ears, yielding a tip fold-over rate of approximately 0.88%.
Table 1.
Distribution table of the included and analyzed manufacturers and electrode arrays (n total = 1,575).
| Manufacturer | Electrode array | n total |
|---|---|---|
| Cochlear TM | SSA | 553 |
| SMA | 159 | |
| CA | 356 | |
| MED EL | Flex24 | 31 |
| Flex26 | 76 | |
| Flex28 | 170 | |
| FlexSoft | 4 | |
| Advanced Bionics | MS | 114 |
| SlimJ | 112 |
Figure 1.

Postoperative cone-beam computed tomography demonstrating electrode array tip fold-over (a) Example of a tip fold-over involving a Mid-Scala (MS) array from Advanced Bionics (b) Example of a tip fold-over involving a Slim Straight Array (SSA) from Cochlear™ (c) Corresponding abnormal intraoperative TIM measurement, which clearly detected the tip fold of the SSA (b).
In contrast, only 3 cases of tip fold-over were detected among the 946 lateral wall (straight) electrode arrays, which include the Cochlear™ SSA, MED-EL Flex series, and Advanced Bionics SlimJ. The incidence rate in this group was significantly lower, at 0.32%. All three tip fold-over cases in this group involved lateral wall SSA arrays, corresponding to an incidence of approximately 0.54% (3 out of 553). No tip fold-over cases were detected with the Advanced Bionics SlimJ or any of the included MED-EL electrode arrays. Of the 13 tip fold-over events, 11 occurred during primary implantation, and 2 occurred during reimplantation procedures for device failure. Residual hearing, expressed as PTA-4, was present preoperatively in 8 of the 13 affected cases. Postoperatively, residual hearing deteriorated in 2 of these cases, while the remaining 5 cases lost residual hearing completely (see Table 2).
Table 2.
Clinical, audiological, and surgical characteristics of the 13 cases with tip fold-over.
| No. | Age | Year of surgery | Reimplantation | Otogenic diagnosis | Tip fold-over findings | Manufacturer | Electrode | New Implant required | Second electrode | Transimpedance matrix measurement (TIM) | X-ray | Preoperative residual hearing (PTA-4) | Postoperative residual hearing (PTA-4) | Surgeon experience |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| 1 | 55 | 2014 | No | PHL | TFP | AB | MS | No | 0 | NT | NP | 65 dB | 115 dB | JS |
| 2 | 60 | 2015 | No | PHL | TFP | Cochlear | CI 512 | No | 0 | NT | NP | 0 | 0 | SS |
| 3 | 41 | 2015 | No | PHL | TFP | Cochlear | CI 532 | No | 0 | NT | P | 110 dB | 0 | SS |
| 4 | 27 | 2015 | Yes | DF | TFI | Cochlear | CI 512 | No | 0 | NT | NP | 0 | 0 | SS |
| 5 | 77 | 2016 | Yes | DF | TFI | Cochlear | CI 512 | No | 0 | NT | NP | 0 | 0 | SS |
| 6 | 46 | 2020 | No | PHL | TFI | Cochlear | CI 632 | Yes | CI 612 | TAN | P | 100 dB | 0 | SS |
| 7 | 56 | 2021 | No | PHL | TFP | Cochlear | CI 622 | No | 0 | TAN | N | 100 dB | 100 dB | SS |
| 8 | 72 | 2021 | No | PHL | TFP | Cochlear | CI 622 | No | 0 | TAN | N | 95 dB | 0 | SS |
| 9 | 57 | 2022 | No | PHL | TFI | Cochlear | CI 622 | No | 0 | TAN | P | 110 dB | 0 | SS |
| 10 | 55 | 2022 | No | PHL | TFI | Cochlear | CI 632 | No | 0 | TAN | N | 80 dB | 110 dB | SS |
| 11 | 73 | 2023 | No | PHL | TFI | Cochlear | CI 632 | No | 0 | TAN | N | 0 | 0 | JS |
| 12 | 54 | 2023 | No | PHL | TFI | Cochlear | CI 632 | No | 0 | TAN | N | 75 dB | 0 | SS |
| 13 | 31 | 2023 | No | PHL | TFI | Cochlear | CI 632 | No | 0 | TAN | N | 0 | 0 | JS |
The table lists age at surgery, year of implantation, reimplantation status, otogenic diagnosis, mode of tip fold-over detection, manufacturer, electrode type, requirement for a new implant, secondary electrode used (if applicable), results of transimpedance matrix measurement (TIM), radiological findings (X-ray), pre- and postoperative residual hearing expressed as PTA-4 (pure tone average at 0.5, 1, 2, and 4 kHz), and surgeon experience.
PHL, Progressive Hearing Loss; DF, Device Failure; TIM, Transimpedance Matrix Measurement; AB, Advanced Bionics; MS, Mid-Scala; JS, Junior Surgeon; SS, Senior Surgeon; N, Normal finding; P, Pathological finding; NP, No X-ray performed; NT, Prior to the implementation of TIM; TAN, First TIM abnormal, post-reimplantation TIM normal; TFI, Tip fold-over detected intraoperatively; TFP, Tip fold-over detected postoperatively.
Influence of intraoperative measurements
Intraoperative tip fold-over detection was documented in 8 cases (61.5%), while postoperative detection occurred in 5 cases (38.5%) and was detected with either CBCT or DVT (see Table 2). Intraoperatively detected tip fold-overs were managed within the same surgical session. In one case, repeated attempts to reinsert the SMA failed to correct the fold-over, necessitating the use of a second implant (CA), which was successfully inserted.
We performed TIM constantly in Cochlear TM electrode arrays following the introduction with surgical routine measurements in June 2019. TIM measurements were performed intraoperatively in 8 cases, all showing abnormal findings consistent with tip fold-over. The remaining cases occurred prior to the implementation of TIM. Both reimplantation procedures (see Table 2) were performed prior to the implementation of TIM.
Final radiological evaluation was performed with modified transorbital intraoperative X-ray imaging. Three examinations revealed abnormalities consistent with tip fold-over. In one notable case (see Table 2, case nr. 3), an abnormal intraoperative X-ray finding led to reinsertion of the electrode array, followed by a normal X-ray result. However, a postoperative cone beam CT demonstrated a tip fold-over. Importantly, this case occurred before the implementation of TIM measurements, which were therefore not available for intraoperative guidance. A crucial aspect to emphasize is the timing of intraoperative radiological imaging. Except for two cases, X-ray examinations were routinely performed at the end of surgery. When abnormal TIM findings suggested a possible tip fold-over, the electrode array was first removed and reinserted. After normalization of the TIM measurements, intraoperative X-ray imaging was performed at the end of the procedure to confirm the final electrode array position. This workflow allowed TIM to serve as the primary intraoperative screening tool for detecting abnormal insertion patterns while avoiding unnecessary additional radiation exposure from repeated intraoperative imaging. In the two exceptional cases, X-ray imaging was performed immediately after abnormal TIM findings, which confirmed tip fold-over and guided subsequent successful reinsertion of the electrode array. Detailed demographic, audiological, and surgical data for all 13 tip fold-over cases are provided in Table 2.
Influence of surgical experience
We additionally examined the potential influence of surgeon experience. Among the 13 cases involving tip fold-over, 11 procedures were performed by a senior surgeon and 2 by a junior surgeon. This distribution likely reflects the greater overall number of procedures conducted by senior surgeons within the cohort rather than an effect of surgical expertise per se. Specifically, Senior Surgeon A performed 703 cochlear implantations with 4 tip fold-over events (0.57%), Senior Surgeon B performed 554 implantations with 6 tip fold-over events (1.08%), and Senior Surgeon C performed 74 implantations with 1 tip fold-over event (1.35%). Among junior surgeons, Surgeon D performed 45 implantations with 1 tip fold-over event (2.22%), and Surgeon E performed 56 implantations with 1 tip fold-over event (1.79%), whereas the remaining five junior surgeons collectively performed 143 implantations without any tip fold-over events. A senior surgeon was defined as a surgeon who had more than 5 years of surgical CI experience. An analysis of the annual distribution of tip fold-over cases showed an initial decline following the first occurrences in 2014 and 2015. No events were documented between 2017 and 2019. From 2020 onward, however, a gradual increase in cases was noted, reaching a peak in 2023 (Figures 2 and 3). These findings suggest that tip fold-over remains a sporadic yet persistent complication over time.
Figure 2.

Annual distribution of tip fold-over cases according to electrode array type. The x-axis represents the year of implantation and the y-axis indicates the number of tip fold-over events identified in that year. Colors correspond to the implanted electrode array type: Mid-Scala (MS), Slim Modiolar Array (SMA), Contour Advance (CA), and Slim Straight Array (SSA).
Figure 3.

Annual distribution of total implantations and tip fold-over cases. The x-axis represents the year of implantation. The left y-axis shows the total number of tip fold-over events per year (all electrode types combined), while the right y-axis indicates the total number of cochlear implantations in adult patients performed annually.
Discussion
Distribution and influence of electrode array design
Tip fold-over is a well-known complication in CI surgery and is particularly influenced by electrode array design. Previous studies have shown that precurved, perimodiolar electrode arrays are associated with a higher incidence of tip fold-over compared to straight, lateral wall electrode arrays (26). The present study confirms these findings in one of the largest cohorts reported to date. We detected 13 cases of tip fold-over in 1575 implanted ears (0.83%): 10 occurred in perimodiolar electrode arrays and 3 in the straight electrode array of Cochlear (SSA). No tip fold-over events were observed in any of the evaluated MED-EL electrode arrays or in the SlimJ array of Advanced Bionics.
In a large retrospective review, Högerle et al. (26) reported a tip fold-over rate of approximately 5.3% for perimodiolar electrode arrays, compared with 0.5–1% for straight lateral wall electrodes. In our study, lower incidence rates were observed, with 1.59% in perimodiolar electrodes and 0.32% in straight electrode arrays. The increased risk associated with perimodiolar electrode arrays has been attributed to their inherent shape and mechanical properties, which can predispose the tip to buckle or rotate during insertion, particularly in anatomically challenging cochleae (36). Their fixed curling geometry may interact unpredictably with the variable anatomy of the cochlear modiolus, resulting in mechanical buckling or rotation of the array tip during insertion.
Notably, the tip fold-over rate in our study was not uniform across all three included precurved electrode arrays (SMA, CA, and MS). The MS array of Advanced Bionics exhibited only one tip fold-over event in 114 cases (0.88%).
A large retrospective review of 1,722 CIs—including perimodiolar, lateral wall, and MS electrode arrays—found an overall tip fold-over rate of 0.87% (24). While a temporal bone study evaluating the Advanced Bionics MS electrode demonstrated reliable, atraumatic insertions without any observed tip fold-over events in 20 insertions (37), a clinical study described the occurrence of tip fold for the MS electrode array before (31). In summary, tip fold-over is a recognized complication that appears to occur more frequently with precurved and perimodiolar electrode arrays.
Tip fold-over has been linked to transient increases in intracochlear pressure, particularly when the electrode twists or rotates (36). Additionally, Cochlear duct length has been identified as a predisposing factor, with longer cochlear ducts increasing the risk—especially for proximal fold-over events in slim modiolar electrodes (38). In contrast, straight lateral wall arrays are more flexible and conform to the cochlear curvature, resulting in a much lower propensity for tip fold-over (6, 39, 40). Overall, precurved perimodiolar arrays carry a substantially higher risk of tip fold-over than straight lateral wall arrays, and this difference has been consistently demonstrated across multiple device manufacturers and clinical settings (6, 24, 26, 38, 39).
Influence of intraoperative measurements
Intraoperative detection of tip fold-over is essential, as unrecognized tip fold-over can result in poor speech perception outcomes and other postoperative complications. While fluoroscopy has traditionally been used for this purpose, newer techniques such as TIM and capacitive sensing systems have demonstrated higher sensitivity and reliability for detecting tip fold-over events (41–43).
In our cohort, intraoperative TIM measurements were performed in 8 of the 13 cases with tip fold-over, and all 8 showed abnormal findings consistent with tip fold-over. This underscores the effectiveness of TIM in providing immediate intraoperative feedback, enabling correction before wound closure. Since June 2019, TIM has been routinely applied in all Cochlear TM CI procedures at our center, allowing for prompt detection and management of tip fold-over events during the same surgical session. Although TIM has demonstrated high sensitivity for detecting tip fold-over, false-positive findings may occur; therefore, TIM results should always be interpreted in conjunction with intraoperative imaging and the clinical context (44, 45). Especially in reimplantations due to possible fibrotic changes (16, 18, 20) and in malformations, e.g., incomplete partition type III (34).
Radiological evaluation further highlighted the limitations of conventional X-ray imaging in detecting fold-over. One particularly notable case in this study occurred before the implementation of TIM: an abnormal intraoperative X-ray prompted reinsertion. Following reinsertion, X-ray imaging showed no evidence of tip fold-over, yet postoperative cone-beam computed tomography (CBCT) still demonstrated a tip fold-over (case number 3; see Table 2).
Furthermore, in another patient (case number 8; see Table 2), TIM indicated a possible tip fold-over (Figure 1c). An X-ray was performed immediately afterward but showed no signs of tip fold-over, so the electrode array was left in place. However, 2 days later, a cone-beam computed tomography (CBCT) scan (Figure 1b) clearly demonstrated a tip fold-over, leading to revision surgery. These findings highlight the potential for false-negative results when relying solely on modified transorbital X-ray imaging, as a tip fold-over may not be detectable depending on its orientation relative to the X-ray projection.
Our findings support previous reports indicating that intraoperative electrophysiological monitoring is especially important for slim modiolar and other precurved electrode arrays, where tip fold-over occurs more frequently and may be missed by standard X-ray imaging alone (46, 47). Although intraoperative imaging such as CBCT or high-resolution fluoroscopy is less frequently required, it remains valuable in cases of challenging cochlear anatomy or when resistance is encountered during insertion (26, 34, 39).
Influence of surgical experience
In our cohort, of the 13 cases with tip fold-over, 11 procedures were performed by senior surgeons and 2 by junior surgeons. However, this distribution must be interpreted in the context of the substantially higher procedural volume among senior surgeons. Indeed, the calculated tip fold-over rates per surgeon remained low across all experience levels, ranging from 0.57 to 1.35% among senior surgeons and from 1.79 to 2.22% among junior surgeons demonstrating both individual and in-house learning curves. Furthermore, no tip fold-over events occurred in the remaining five junior surgeons, despite a combined total of 143 cochlear implant procedures. These findings suggest that tip fold-over may occur even in the hands of highly experienced CI surgeons and is likely influenced more strongly by electrode design, insertion mechanics, and individual cochlear anatomy than by surgical experience alone. Perimodiolar, precurved electrode arrays are designed with a predefined curvature and are typically inserted using a stylet or a sheath for the SMA. Upon removal of the stylet, the electrode array assumes its preformed shape and is intended to position itself close to the modiolus. However, this mechanism carries an inherent risk: if the electrode is released too early or too late, or if the individual cochlear anatomy—such as an increased cochlear duct length —does not match the fixed curvature of the electrode, the tip may buckle, resulting in a tip fold-over (38, 40, 48). It should also be noted that the SMA and the MS must be inserted using AOS technique, whereas the CA does not necessarily have to be inserted using AOS technique.
In contrast, straight lateral wall electrode arrays from MED-EL are not precurved but are highly flexible and passively follow the lateral wall of the cochlea during insertion. These arrays do not require a stylet and lack a predefined curvature that could conflict with individual cochlear anatomy. As a result, the risk of tip fold-over is considered to be very low in these electrode designs (26).
Meticulous, slow, and controlled insertion remains the most effective strategy to minimize tip fold-over, although the optimal technique varies according to the type of electrode array. For straight lateral wall arrays, which inherently carry a low risk of tip fold-over, a slow, continuous insertion along the lateral wall—ideally via the round window or extended round window—is recommended. Careful handling to avoid abrupt movements is essential to further reduce the risk of malposition (26, 39, 49, 50).
For precurved perimodiolar arrays, the insertion technique is more complex and requires additional precautions. Correct alignment to the modiolus, precise handling of the stylet in stylet-based systems, and a prolonged, controlled insertion are essential to prevent tip folding (6, 38, 51). Some studies described, that longer cochlear duct lengths (>36 mm) increase the risk of tip fold-over, emphasizing the importance of preoperative cochlear duct measurements and appropriate electrode selection (38). Intraoperative imaging or electrophysiological monitoring, such as TIM, provides immediate feedback to detect and correct malposition.
In summary, precurved perimodiolar arrays require heightened intraoperative vigilance, careful technique, and, ideally, real-time monitoring to reduce tip fold-over, while straight lateral wall arrays benefit primarily from gentle insertion under standard surgical precautions (26, 39).
In our cohort, tip fold-over occurred in 13 cases, predominantly affecting precurved perimodiolar arrays. Temporal analysis showed early cases in 2014–2015, a period without events from 2017–2019, and a resurgence from 2020, peaking in 2023. Although infrequent, these findings highlight that tip fold-over remains a sporadic but persistent complication, emphasizing the importance of meticulous intraoperative technique, surgeon experience, intraoperative measurements, and electrode array design.
Nevertheless, several limitations of this study should be acknowledged. First, this was a single-center study conducted at a large CI center. Therefore, the transferability of the surgical expertise and the reported findings to other centers, particularly smaller CI centers, may be limited. Second, the number of electrode arrays investigated varied considerably between electrode array types, which may have introduced a potential selection bias. However, nearly all currently available electrode arrays were included and analyzed, which may also be considered a strength of this study, as it provides a comprehensive overview across different electrode designs. Furthermore, the low incidence of tip fold-over limits the statistical power of the study. In particular, the occurrence of only a single tip fold-over with the MS electrode array does not allow meaningful conclusions regarding potential learning curves. Similarly, the limited number of cases does not permit robust conclusions regarding functional outcomes, such as preservation of residual hearing.
Conclusion
The design of electrode arrays plays a critical role in influencing the risk of tip fold-over. Perimodiolar electrode arrays have been associated with a higher incidence of this complication compared to flexible straight arrays. This study highlights the pivotal importance of intraoperative diagnostics. In our cohort, intraoperative TIM analysis combined with radiographic imaging enabled reliable real-time detection of tip fold-over and immediate correction, thereby reducing the need for revision surgery. Based on these findings, TIM should be considered the primary intraoperative screening tool, with abnormal findings prompting immediate reinsertion and subsequent radiological confirmation of correct electrode placement. Advanced imaging techniques such as intraoperative cone-beam CT or digital volume tomography may provide additional diagnostic certainty in complex cases. Findings from this large-scale retrospective analysis confirm that tip fold-over, while rare, represents a clinically significant complication in CI, with a notably higher prevalence in precurved perimodiolar electrode arrays. Furthermore, this investigation also reports tip fold-over in the Advanced Bionics MS array, underscoring the necessity for continued vigilance across all cochlear implant systems. These findings should encourage all manufacturers to develop similar measurement tools to enable intraoperative diagnostics of electrode array position.
Acknowledgments
The authors thank the Fördergesellschaft “Taube Kinder lernen Hören e. V.,” which has supported the cochlear implant rehabilitation center in Freiburg for the past several years.
Funding Statement
The author(s) declared that financial support was not received for this work and/or its publication.
Footnotes
Edited by: Brian John McKinnon, Meritus Health, United States
Reviewed by: Badr Eldin Mostafa, Ain Shams University, Egypt
Michael Canfarotta, WellStar Medical Group, United States
Data availability statement
The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.
Ethics statement
The studies involving humans were approved by the Ethics Committee of Albert-Ludwigs University Freiburg (Approval number: 406/19; Amendment number: 240275). The studies were conducted in accordance with the local legislation and institutional requirements. Written informed consent for participation was not required from the participants or the participants’ legal guardians/next of kin due to the retrospective nature of the study.
Author contributions
GM: Investigation, Software, Writing – review & editing, Validation, Resources, Methodology, Funding acquisition, Data curation, Project administration, Conceptualization, Supervision, Writing – original draft, Visualization, Formal analysis. JS: Data curation, Writing – review & editing, Methodology. LF: Conceptualization, Visualization, Investigation, Writing – review & editing. CY: Writing – review & editing, Methodology, Formal analysis, Data curation. AA: Resources, Funding acquisition, Formal analysis, Visualization, Project administration, Validation, Writing – review & editing. SA: Funding acquisition, Writing – review & editing, Project administration, Methodology, Formal analysis, Resources, Supervision. RB: Data curation, Project administration, Writing – review & editing, Funding acquisition, Investigation, Resources. MK: Methodology, Writing – original draft, Supervision, Formal analysis, Writing – review & editing, Software, Visualization, Data curation, Project administration, Resources, Validation, Funding acquisition, Investigation, Conceptualization.
Conflict of interest
The author(s) declared that this work was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
The author MK declared that they were an editorial board member of Frontiers, at the time of submission. This had no impact on the peer review process and the final decision.
Generative AI statement
The author(s) declared that Generative AI was not used in the creation of this manuscript.
Any alternative text (alt text) provided alongside figures in this article has been generated by Frontiers with the support of artificial intelligence and reasonable efforts have been made to ensure accuracy, including review by the authors wherever possible. If you identify any issues, please contact us.
Publisher’s note
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.
References
- 1.Ketterer MC, Aschendorff A, Arndt S, Beck R. Electrode array design determines scalar position, dislocation rate and angle and postoperative speech perception. Eur Arch Otorrinolaringol. (2022a) 279:4257–67. doi: 10.1007/s00405-021-07160-2, [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Ketterer MC, Aschendorff A, Arndt S, Hassepass F, Wesarg T, Laszig R, et al. The influence of cochlear morphology on the final electrode array position. Eur Arch Otorrinolaringol. (2018) 275:385–94. doi: 10.1007/s00405-017-4842-y, [DOI] [PubMed] [Google Scholar]
- 3.van der Marel KS, Briaire JJ, Wolterbeek R, Snel-Bongers J, Verbist BM, Frijns JH. Diversity in cochlear morphology and its influence on cochlear implant electrode position. Ear Hear. (2014) 35:e9–e20. doi: 10.1097/01.aud.0000436256.06395.63, [DOI] [PubMed] [Google Scholar]
- 4.Würfel W, Lanfermann H, Lenarz T, Majdani O. Cochlear length determination using cone beam computed tomography in a clinical setting. Hear Res. (2014) 316:65–72. doi: 10.1016/j.heares.2014.07.013 [DOI] [PubMed] [Google Scholar]
- 5.Beck RL, Aschendorff A, Arndt S, Hildenbrand T, Ketterer MC. Is the position of the basal-Most electrode depending on electrode Array design and influencing postoperative speech perception? A retrospective analysis of 495 ears. Otol Neurotol. (2024a) 45:502–6. doi: 10.1097/MAO.0000000000004165, [DOI] [PubMed] [Google Scholar]
- 6.Dhanasingh A, Jolly C. An overview of cochlear implant electrode array designs. Hear Res. (2017) 356:93–103. doi: 10.1016/j.heares.2017.10.005, [DOI] [PubMed] [Google Scholar]
- 7.Holden LK, Finley CC, Firszt JB, Holden TA, Brenner C, Potts LG, et al. Factors affecting open-set word recognition in adults with cochlear implants. Ear Hear. (2013) 34:342–60. doi: 10.1097/AUD.0b013e3182741aa7, [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Dhanasingh A, Jolly C. Review on cochlear implant electrode array tip fold-over and scalar deviation. J. Otology. (2019) 14:94–100. doi: 10.1016/j.joto.2019.01.002, [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Eshraghi AA, Yang NW, Balkany TJ. Comparative study of cochlear damage with three perimodiolar electrode designs. Laryngoscope. (2003) 113:415–9. doi: 10.1097/00005537-200303000-00005, [DOI] [PubMed] [Google Scholar]
- 10.Bas E, Dinh CT, Garnham C, Polak M, Van De Water TR. Conservation of hearing and protection of hair cells in cochlear implant patients with residual hearing. Anat Rec. (2012) 295:1909–27. doi: 10.1002/ar.22574, [DOI] [PubMed] [Google Scholar]
- 11.Aschendorff A, Kromeier J, Klenzner T, Laszig R. Quality control after insertion of the nucleus contour and contour advance electrode in adults. Ear Hear. (2007) 28:75S–9S. doi: 10.1097/AUD.0b013e318031542e, [DOI] [PubMed] [Google Scholar]
- 12.Finley CC, Holden TA, Holden LK, Whiting BR, Chole RA, Neely GJ, et al. Role of electrode placement as a contributor to variability in cochlear implant outcomes. Otol Neurotol. (2008) 29:920–8. doi: 10.1097/MAO.0b013e318184f492, [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Aschendorff A, Arndt S, Kröger S, Wesarg T, Ketterer MC, Kirchem P, et al. Quality of cochlear implant rehabilitation under COVID-19 conditions. HNO. (2021) 69:1–6. doi: 10.1007/s00106-020-00923-z. 33034674, [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Aschendorff A, Klenzner T, Arndt S, Beck R, Schild C, Roddiger L, et al. Insertion results for contour and contour advance electrodes: are there individual learning curves? HNO. (2011) 59:448–52. doi: 10.1007/s00106-011-2319-7, [DOI] [PubMed] [Google Scholar]
- 15.Everad F, Beck RL, Aschendorff A, Rauch AK, Fries L, Arndt S, et al. Are tinnitus burden and tinnitus exacerbation after cochlear implantation influenced by insertion technique, array dislocation, and intracochlear trauma? Front Neurol. (2024) 15:1477259. doi: 10.3389/fneur.2024.1477259, [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Ketterer MC, Shiraliyev K, Arndt S, Aschendorff A, Beck R. Implantation and reimplantation: epidemiology, etiology and pathogenesis over the last 30 years. Eur Arch Otorrinolaringol. (2024a) 281:4095–102. doi: 10.1007/s00405-024-08568-2, [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Ketterer MC, Aschendorff A, Arndt S, Beck R. Correction to: electrode array design determines scalar position, dislocation rate and angle and postoperative speech perception. Eur Arch Otorrinolaringol. (2022b) 279:4269–70. doi: 10.1007/s00405-021-07215-4, [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Ketterer MC, Brückerhoff K, Arndt S, Beck R, Aschendorff A. Insertion of a second electrode array-a rare complication of CI reimplantation. HNO. (2024b) 72:63–5. doi: 10.1007/s00106-023-01364-0. 37943372, [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Fries L, Everad F, Beck RL, Aschendorff A, Arndt S, Ketterer MC. The influence of CI electrode array design on the preservation of residual hearing. Front Neurol. (2025) 16:1599369. doi: 10.3389/fneur.2025.1599369, [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Beck R, Shiraliyev K, Arndt S, Rauch AK, Aschendorff A, Hassepass F, et al. Scalar position, dislocation analysis and outcome in CI reimplantation due to device failure. Eur Arch Otorrinolaringol. (2022) 279:4853–9. doi: 10.1007/s00405-022-07315-9. 35226182, [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Aschendorff A, Briggs R, Brademann G, Helbig S, Hornung J, Lenarz T, et al. Clinical investigation of the nucleus slim Modiolar electrode. Audiol. Neurotol. (2017) 22:169–79. doi: 10.1159/000480345, [DOI] [PubMed] [Google Scholar]
- 22.Beck R, Aschendorff A, Arndt S, Hildenbrand T, Rauch AK, Ketterer MC. Evaluation of insertion quality of a slim perimodiolar electrode array. Eur Arch Otorrinolaringol. (2024b) 281:1215–20. doi: 10.1007/s00405-023-08212-5, [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Cosetti MK, Troob SH, Latzman JM, Shapiro WH, Roland JT, Jr, Waltzman SB. An evidence-based algorithm for intraoperative monitoring during cochlear implantation. Otol Neurotol. (2012) 33:169–76. doi: 10.1097/MAO.0b013e3182423175, [DOI] [PubMed] [Google Scholar]
- 24.Gabrielpillai J, Burck I, Baumann U, Stöver T, Helbig S. Incidence of tip fold-over during cochlear implantation. Otol Neurotol. (2018) 39:1115–21. doi: 10.1097/MAO.0000000000001915 [DOI] [PubMed] [Google Scholar]
- 25.Grolman W, Maat A, Verdam F, Simis Y, Carelsen B, Freling N, et al. Spread of excitation measurements for the detection of electrode array foldovers: a prospective study comparing 3-dimensional rotational x-ray and intraoperative spread of excitation measurements. Otol Neurotol. (2009) 30:27–33. doi: 10.1097/mao.0b013e31818f57ab, [DOI] [PubMed] [Google Scholar]
- 26.Högerle C, Englhard A, Simon F, Grüninger I, Mlynski R, Hempel JM, et al. Cochlear implant electrode tip fold-over: our experience with long and flexible electrode. Otol Neurotol. (2022) 43:64–71. doi: 10.1097/MAO.0000000000003362, [DOI] [PubMed] [Google Scholar]
- 27.Jia H, Torres R, Nguyen Y, De Seta D, Ferrary E, Wu H, et al. Intraoperative Conebeam CT for assessment of Intracochlear positioning of electrode arrays in adult recipients of Cochlear implants. AJNR Am J Neuroradiol. (2018) 39:768–74. doi: 10.3174/ajnr.A5567. 29472297, [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.McJunkin JL, Durakovic N, Herzog J, Buchman CA. Early outcomes with a slim, Modiolar Cochlear implant electrode Array. Otol Neurotol. (2018) 39:e28–33. doi: 10.1097/MAO.0000000000001652. 29227449, [DOI] [PubMed] [Google Scholar]
- 29.Sabban D, Parodi M, Blanchard M, Ettienne V, Rouillon I, Loundon N. Intra-cochlear electrode tip fold-over. Cochlear Implants Int. (2018) 19:225–9. doi: 10.1080/14670100.2018.1427823, [DOI] [PubMed] [Google Scholar]
- 30.Sipari S, Iso-Mustajärvi M, Löppönen H, Dietz A. The insertion results of a mid-scala electrode assessed by MRI and CBCT image fusion. Otol Neurotol. (2018) 39:e1019–25. doi: 10.1097/MAO.0000000000002045, [DOI] [PubMed] [Google Scholar]
- 31.Zuniga MG, Rivas A, Hedley-Williams A, Gifford RH, Dwyer R, Dawant BM, et al. Tip fold-over in Cochlear implantation: case series. Otol Neurotol. (2017) 38:199–206. doi: 10.1097/MAO.0000000000001283, [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Widmann G, Dejaco D, Luger A, Schmutzhard J. Pre- and post-operative imaging of cochlear implants: a pictorial review. Insights Imaging. (2020) 11:93. doi: 10.1186/s13244-020-00902-6, [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Aschendorff A, Kubalek R, Turowski B, Zanella F, Hochmuth A, Schumacher M, et al. Quality control after cochlear implant surgery by means of rotational tomography. Otol Neurotol. (2005) 26:34–7. doi: 10.1097/00129492-200501000-00007, [DOI] [PubMed] [Google Scholar]
- 34.Alballaa A, Aschendorff A, Arndt S, Hildenbrand T, Becker C, Hassepass F, et al. Incomplete partition type III revisited-long-term results following cochlear implant. HNO. (2020) 68:25–32. doi: 10.1007/s00106-019-00732-z, [DOI] [PubMed] [Google Scholar]
- 35.Escudé B, James C, Deguine O, Cochard N, Eter E, Fraysse B. The size of the cochlea and predictions of insertion depth angles for cochlear implant electrodes. Audiol Neurootol. (2006) 11:27–33. doi: 10.1159/000095611, [DOI] [PubMed] [Google Scholar]
- 36.Chabuz CA, Banakis Hartl RM, Rodriguez K, Gonzalez J, Cass SP, Greene NT. Characterization of tip fold-over using fluoroscopy and intracochlear pressure in cadaver specimens. Laryngoscope. (2025) 135:1795–802. doi: 10.1002/lary.31977, [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Hassepass F, Bulla S, Maier W, Laszig R, Arndt S, Beck R, et al. The new mid-Scala electrode Array: a radiologic and histologic study in human temporal bones. Otol Neurotol. (2014) 35:1415–20. doi: 10.1097/MAO.0000000000000412, [DOI] [PubMed] [Google Scholar]
- 38.Oh H, Kim Y, Lee SJ, Carandang M, Song JJ, Choi BY. Two discrete types of tip fold-over in Cochlear implantation using slim Modiolar electrodes: influence of Cochlear duct length on tip fold-over. Otol Neurotol. (2023) 44:1011–4. doi: 10.1097/MAO.0000000000004022. 37733985, [DOI] [PubMed] [Google Scholar]
- 39.Dhanasingh A, Nielsen SB, Beal F, Schilp S, Hessler R, Jolly C, et al. Cochlear implant electrode design for safe and effective treatment. Front Neurol. (2024) 15:1348439. doi: 10.3389/fneur.2024.1348439, [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Pietsch M, Schurzig D, Salcher R, Warnecke A, Erfurt P, Lenarz T, et al. Variations in microanatomy of the human modiolus require individualized cochlear implantation. Sci Rep. (2022) 12:5047. doi: 10.1038/s41598-022-08731-x, [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.Hou L, Du X, Boulgouris NV, Hafeez N, Coulson C, Irving R, et al. A capacitive cochlear implant electrode array sensing system to discriminate fold-over pattern. J Speech Lang Hear Res. (2021) 64:3685–96. doi: 10.1044/2021_JSLHR-21-00067, [DOI] [PubMed] [Google Scholar]
- 42.Klabbers TM, Huinck WJ, Mylanus EAM. Comparison between Transimpedance matrix (TIM) measurement and X-ray fluoroscopy for intraoperative electrode array tip fold-over detection. Otol Neurotol. (2021a) 42:e1457–63. doi: 10.1097/MAO.0000000000003290, [DOI] [PubMed] [Google Scholar]
- 43.Klabbers TM, Huinck WJ, Heutink F, Verbist BM, Mylanus EAM. Transimpedance matrix (TIM) measurement for the detection of intraoperative electrode tip fold-over using the slim Modiolar electrode: a proof of concept study. Otol Neurotol. (2021b) 42:e124–9. doi: 10.1097/MAO.0000000000002875, [DOI] [PubMed] [Google Scholar]
- 44.Arambula AM, Piper R, Murray G, Mowry SE, Rivas A, Semaan M, et al. Sensitivity and specificity of intraoperative TransImpedance matrix recordings compared with X-ray imaging in detecting Perimodiolar Cochlear implant tip Foldovers: a multicenter study. Otol Neurotol. (2025) 46:627–33. doi: 10.1097/MAO.0000000000004502, [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45.Hoppe U, Brademann G, Stöver T, Ramos de Miguel A, Cowan R, Manrique M, et al. Evaluation of a transimpedance matrix algorithm to detect anomalous cochlear implant electrode position. Audiol Neurootol. (2022) 27:347–55. doi: 10.1159/000523784, [DOI] [PubMed] [Google Scholar]
- 46.Ayas M, Muzaffar J, Borsetto D, Eitutis S, Phillips V, Tam YC, et al. A scoping review on the clinical effectiveness of trans-impedance matrix (TIM) measurements in detecting extracochlear electrodes and tip fold-overs in Cochlear LTD devices. PLoS One. (2024) 19:e0299597. doi: 10.1371/journal.pone.0299597, [DOI] [PMC free article] [PubMed] [Google Scholar]
- 47.Inoue T, Tona Y, Okano T, Omori K, Yamamoto N. Detection of tip fold-over of the slim Modiolar electrode using intraoperative mobile cone-beam computed tomography. Otol. Head Neck Surg. (2023) 169:105–11. doi: 10.1002/ohn.240, [DOI] [PubMed] [Google Scholar]
- 48.Choi Y, Joo HA, Kang WS, Ahn JH, Chung JW, Park HJ. Anatomical relationship of Cochlear basal turn and facial nerve in patients with tip fold-over during Cochlear implantation using slim Modiolar electrode. Otol Neurotol. (2025) 46:1242–6. [DOI] [PubMed] [Google Scholar]
- 49.Areias B, Parente MPL, Gentil F, Natal Jorge RM. Finite element modelling of the surgical procedure for placement of a straight electrode array: mechanical and clinical consequences. J Biomech. (2021) 129:110812. doi: 10.1016/j.jbiomech.2021.110812, [DOI] [PubMed] [Google Scholar]
- 50.Jwair S, Versnel H, Stokroos RJ, Thomeer HGXM. The effect of the surgical approach and cochlear implant electrode on the structural integrity of the cochlea in human temporal bones. Sci Rep. (2022) 12:17068. doi: 10.1038/s41598-022-21399-7, [DOI] [PMC free article] [PubMed] [Google Scholar]
- 51.van der Jagt AMA, Briaire JJ, Boehringer S, Verbist BM, Frijns JHM. Prolonged insertion time reduces translocation rate of a Precurved electrode Array in Cochlear implantation. Otol Neurotol. (2022) 43:e427–34. doi: 10.1097/MAO.0000000000003499, [DOI] [PubMed] [Google Scholar]
- 52.Ketterer MC, Rauch AK, Beck RL, Jakob TF, Fries L, Aschendorff A, et al. The influence of electrode array design, scalar dislocation and insertion technique on postoperative vertigo in CI surgery – a prospective study. Eur Arch Otorrinolaringol. (2025) 282:2367–72. doi: 10.1007/s00405-024-09147-1, [DOI] [PMC free article] [PubMed] [Google Scholar]
- 53.Ketterer MC, Aschendorff A, Arndt S, Speck I, Rauch AK, Beck R, et al. Radiological evaluation of a new straight electrode array compared to its precursors. Eur Arch Otorhinolaryngol. (2021) 278:3707–3714. doi: 10.1007/s00405-020-06434-5, [DOI] [PMC free article] [PubMed] [Google Scholar]
- 54.Beck RL, Aschendorff A, Arndt S, Ketterer MC. Multivariate analysis of the exact effects of scalar position and insertion angle on speech discrimination. Front Neurol. (2025) 16:1569100. doi: 10.3389/fneur.2025.1569100, [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.