Abstract
Introduction
This study aimed to investigate the clinical feasibility of using an ultra-dilute, low-concentration iodine contrast agent in combination with dual-layer spectral computed tomography (CT) for pulmonary artery imaging.
Methods
A total of 143 patients who underwent CT pulmonary angiography (CTPA) between May 2024 and May 2025 were enrolled. After excluding 19 patients with suboptimal image quality due to technical factors, 124 patients were included in the final analysis (spectral group: n = 64; control group: n = 60). Patients were assigned to either a spectral group (100 kV, 160 mgI/mL contrast, 25 mL, 2.5 mL/s) or a control group (120 kV, 350 mgI/mL contrast, 35 mL, 4 mL/s). In the spectral group, images were reconstructed using conventional CT and virtual monoenergetic imaging at 40, 70, and 110 keV. The control group underwent conventional CT imaging only. Image quality, CT attenuation values, signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and radiation doses were assessed and compared between the two groups.
Results
In the spectral group, 40 keV monoenergetic images yielded the highest CT values, SNR, and CNR for all pulmonary artery branches. The control group demonstrated superior image quality on conventional CT and at higher monoenergetic levels (70 and 110 keV), with statistically significant differences (P < 0.05). However, no significant differences were observed between the 40 keV spectral images and the control group. Subjective image quality scores were significantly lower at 70 and 110 keV, but comparable at 40 keV (P > 0.05).
Discussion
The use of an ultra-low concentration iodine contrast agent (160 mgI/mL) combined with dual-layer spectral CT provides diagnostic image quality equivalent to that of conventional protocols, while reducing the iodine load by 67% (to 4 g per examination). This approach reduces iodine load while maintaining diagnostic image quality, although its effect on renal safety requires further clinical validation.
Conclusion
Dual-layer spectral CT combined with 40 keV virtual monoenergetic imaging allowed pulmonary artery imaging with markedly reduced iodine load while maintaining diagnostic image quality comparable to that of the conventional protocol.
Keywords: X-ray computed tomography, Spectral computed tomography, Contrast media, Pulmonary artery imaging, Radiation dose, Signal-to-noise ratio
1. INTRODUCTION
Pulmonary embolism (PE) is a life-threatening condition caused by the obstruction of the pulmonary artery or its branches by a thrombus. It is the third most common cardiovascular cause of death after myocardial infarction and ischemic stroke [1]. Delayed or missed diagnosis of PE may lead to adverse clinical outcomes and increased mortality [2]. Therefore, a timely and evidence-based diagnostic evaluation is essential in patients with suspected PE [3]. Computed tomography pulmonary angiography (CTPA) is widely used as the preferred imaging modality for suspected PE because of its non-invasive nature, high spatial resolution, and rapid acquisition [4-7]. However, conventional CTPA protocols require high-concentration iodine contrast agents (270–370 mgI/mL), rapid injection rates (up to 5 mL/s), and relatively large volumes (60–100 mL); therefore, optimizing contrast administration remains clinically relevant, particularly in the context of iodine exposure, renal safety, and hypersensitivity-related concerns [8-10]. Recent advances in dual-layer spectral CT and related reconstruction techniques enable the generation of virtual monoenergetic images (MonoE), significantly enhancing vascular visualization [11-14]. This technology offers the potential to reduce contrast agent dosage while maintaining diagnostic image quality. However, optimizing contrast agent use without compromising diagnostic reliability remains a key challenge in current CTPA research. Low-energy monoenergetic imaging can increase the CT attenuation of iodine, thereby compensating for reduced contrast concentration. Therefore, we hypothesized that combining an ultra-dilute low-concentration iodine contrast agent (160 mgI/mL) with 40 keV virtual monoenergetic imaging using dual-layer spectral CT could substantially reduce iodine intake while preserving diagnostic image quality. This study aimed to validate the clinical feasibility of this protocol [15-18].
1.1. Contributions of the Study
First application of an iodine contrast agent diluted to 160 mgI/mL (46% of conventional concentration) combined with dual-layer spectral CT for pulmonary artery imaging.
Achievement of a 67% reduction in iodine intake (to 4 g per examination) through multi-parameter optimization, including low concentration, low injection rate, low tube voltage, and spectral reconstruction.
Systematic comparison of image quality across different keV levels (40, 70, and 110 keV), identifying 40 keV as the optimal monoenergetic level.
Demonstration that the ultra-low iodine protocol provides diagnostic image quality comparable to that of conventional CTPA.
2. MATERIALS
2.1. Patients
This was a prospective, single-center, controlled study designed to evaluate the feasibility of an ultra-low iodine contrast protocol in dual-layer spectral CT pulmonary angiography. Patients were assigned to either the spectral group or the control group based on the examination period: the spectral group included patients from May to November 2024, and the control group included patients from December 2024 to May 2025. All image acquisitions and analyses followed standardized operating procedures.
Inclusion criteria were: (1) clinically suspected PE; (2) no known allergy to iodinated contrast agents; and (3) signed informed consent. Exclusion criteria included: (1) confirmed hypersensitivity to iodinated contrast; (2) pregnancy or lactation; (3) altered mental status or severe anxiety precluding successful CTPA; (4) significant cardiac, hepatic, or renal dysfunction; and (5) poor image quality due to technical factors unrelated to the contrast protocol (e.g., motion artifacts, bolus tracking failure, equipment malfunction). Importantly, no cases were excluded solely due to low vascular enhancement, potentially related to the low-concentration contrast agent. After excluding 19 patients, 124 were included in the final analysis (spectral group: n = 64; control group: n = 60). The study flowchart is presented in Fig. (1).
Fig. (1).

Flowchart for patient inclusion criteria.
This study was approved by the Ethics Committee of Civil Aviation General Hospital (Approval No. 2024-L-K-123) and conducted in accordance with the Declaration of Helsinki.
2.2. Contrast Injection and CT Acquisition
All examinations were performed using a dual-layer spectral CT scanner (Spectral CT 7500, Philips Healthcare, Netherlands). Patients were positioned supine with arms elevated, and scanned during free breathing from the thoracic inlet to the costophrenic angle. Automated bolus tracking was applied with a trigger threshold of 90 HU at the main pulmonary artery.
Scanning parameters were as follows: tube voltage of 100 kV for the spectral group and 120 kV for the control group; automatic tube current modulation (DoseRight) with a ReduceDose setting of 18; collimation of 128 × 0.625 mm; slice thickness and interval of 1 mm; pitch of 1.178; rotation time of 0.27 s; post-injection delay of 1.5 s; and post-threshold delay of 3.4 s.
For the spectral group, iopamidol (320 mgI/mL) was diluted 1:1 with saline to achieve a final concentration of 160 mgI/mL, and 25 mL was injected at 2.5 mL/s, followed by 25 mL saline at the same rate. The control group received undiluted iopamidol (350 mgI/mL) at 35 mL and 3.5 mL/s, followed by 30 mL saline at the same rate.
The parameter differences between groups were intentional: (1) lower tube voltage in the spectral group was used to enhance iodine contrast at low energy levels, compensating for reduced iodine concentration; (2) the spectral group protocol (low volume, low flow rate, diluted concentration) was designed to test the feasibility of ultra-low iodine load, while the control group followed standard clinical practice; (3) automatic tube current modulation ensured comparable image noise and radiation doses between groups.
2.3. Image Post-Processing
Raw data were transferred to a dedicated workstation (IntelliSpace Portal V9, Philips Healthcare) for post-processing. Spectral-based imaging (SBI) was used to generate virtual monoenergetic images at 40, 70, and 110 keV.
2.4. Image Quality Assessment
2.4.1. Subjective Evaluation
Two experienced radiologists independently assessed image quality using a 5-point scale:
5: Excellent - clear vascular contours, good continuity, uniform density, no artifacts.
4: Good - generally clear contours, minor discontinuity in 1–2 segments, acceptable density uniformity, minimal artifacts.
3: Fair - somewhat unclear contours, discontinuity in 3–5 segments, acceptable density, visible artifacts.
2: Poor - poorly defined contours, discontinuity in >50% of segments, poor density uniformity, numerous artifacts.
1: Non-diagnostic - unclear contours, discontinuity in >60% of segments, poor density, severe artifacts.
2.4.2. Objective Evaluation
CT attenuation values (in HU) and standard deviations (SD) were measured in the main, lobar, segmental, and subsegmental pulmonary arteries, as well as in the paravertebral muscles. SNR and CNR were calculated as follows:
SNR = CT value / SD
CNR = (CT value of artery − CT value of muscle) / SD
Radiation dose parameters, including CT dose index volume (CTDIvol), dose-length product (DLP), and effective dose (ED), were recorded. ED was calculated as DLP × k, with k = 0.014 mSv/(mGy·cm).
2.5. Statistical Analysis
Statistical analyses were performed using SPSS version 27.0 and GraphPad Prism version 10.0. Normality was assessed using the Kolmogorov–Smirnov test. Normally distributed data were expressed as mean ± SD; non-normally distributed data were presented as median with interquartile range. Group comparisons of demographic and radiation dose data were performed using the t-test or Mann–Whitney U test as appropriate. Inter-rater agreement was evaluated using the intraclass correlation coefficient (ICC). Subjective and objective image quality parameters were compared using ANOVA. A P-value < 0.05 was considered statistically significant.
3. RESULTS
3.1. Patient Characteristics
The median age was 65 years (range: 56–72 years), and the mean BMI was 24.39 ± 3.67 kg/m2. No significant differences were observed between the spectral group (n = 64) and control group (n = 60) in terms of sex, age, BMI, or radiation dose parameters (CTDIvol, DLP, ED; all P > 0.05; Table 1).
Table 1.
Comparison of basic clinical information for two groups.
| Variable | Total (n=124) | Spectrum Group (n=64) | Control Group (n=60) | Statistical value | P |
|---|---|---|---|---|---|
| Gender, n(%) | - | - | - | χ2=0.00 | 0.981 |
| Women | 58 (46, 77) | 30 (46, 88) | 28 (46, 67) | - | - |
| Man | 66 (53, 23) | 34 (53, 12) | 32 (53, 33) | - | - |
| Age (year) | 65(56-72) | 62 ± 15 | 66(57-72) | U=1882 | 0.849 |
| BMI (kg/m2) | 24.39 ± 3.67 | 24.59 ± 3.47 | 24.17 ± 3.89 | t=0.15 | 0.882 |
| CTDIvol (mGy) | 6.9(5.7-8) | 7.39 ± 2.02 | 6.65(5.55-7.65) | U=1629.5 | 0.146 |
| DLP (mGy·cm) | 264(218.5-308) | 277.81 ± 74.95 | 262(215-294) | U=1747.5 | 0.388 |
| ED (mSv) | 3.70(3.05-4.31) | 3.89 ± 1.05 | 3.67(3.01-4.12) | U=1751.5 | 0.399 |
3.2. Subjective Image Quality
Significant differences in subjective image quality were observed across energy levels in the spectral group (all P < 0.001). The 40 keV images received the highest scores (Rater A/B: 4.72 ± 0.60 / 4.67 ± 0.57), substantially higher than those at 70 keV (~2.27 ± 0.57) and 110 keV (~1.10 ± 0.52). Conventional CT images scored approximately 1.97 ± 0.76 between the 70 keV and 110 keV levels. No significant differences were found between the 40 keV spectral images and the control group (P = 0.12–0.57), indicating comparable clinical applicability. Inter-rater agreement was good for 70 keV (ICC = 0.834) and conventional CT (ICC = 0.835), and moderate for 40 keV (ICC = 0.610) and the control group (ICC = 0.657), suggesting more stable scoring at higher energy levels (Tables 2-4).
Table 2.
Image quality scores for spectral groups.
| Variable | Total(n=256) | 40keV(n= 64) | 70keV(n = 64) | 110keV (n = 64) | Traditional CT images (n = 64) | Statistical value | P |
|---|---|---|---|---|---|---|---|
| Reviewer A, Mean ± SD | 2.91 ± 1.56 | 4.72 ± 0.60 | 2.31 ± 0.56 | 1.09 ± 0.72 | 1.98 ± 0.72 | F=502.95 | <.001 |
| Reviewer B, Mean ± SD | 2.90±1.57 | 4.67±0.57 | 2.23±0.58 | 1.11±0.32 | 1.95±0.79 | F=505.32 | <.001 |
Table 4.
Consistency analysis between raters.
| Group | ICC | 95% confidence interval | Reliability level | ||
|---|---|---|---|---|---|
| Spectrum | 40keV | 0.610 | 0.430 | 0.743 | Medium |
| 70keV | 0.834 | 0.739 | 0.896 | Good | |
| 110keV | 0.746 | 0.614 | 0.838 | Medium | |
| Conventional CT images | 0.835 | 0.743 | 0.897 | Good | |
| Control | 0.657 | 0.492 | 0.776 | Medium | |
The ICC correlation coefficient less than 0.5 indicates poor consistency; 0.5 to 0.75 indicates moderate consistency; 0.75 to 0.9 indicates good consistency; greater than 0.9 indicates excellent consistency.
3.3. Objective Image Quality
Significant differences in objective image quality were observed across energy levels in the spectral group (all P < 0.001). The 40 keV images demonstrated the highest SNR and CNR across all pulmonary artery branches, including the main, lobar, segmental, and subsegmental arteries. Notably, at the subsegmental level, the 40 keV group achieved a CNR of 19.86 ± 9.87 and an SNR of 21.81 ± 9.99, significantly outperforming other energy levels (P < 0.001), indicating a distinct advantage for visualizing distal small vessels.
When comparing the 40 keV spectral group with the control group, no statistically significant differences were found in SNR or CNR for any pulmonary artery branch (P = 0.227–0.845), confirming that the ultra-low iodine protocol with 40 keV reconstruction yields objective image quality equivalent to that of conventional CTPA (Tables 5 and 6; Fig. 2). The distribution of SNR and CNR across imaging groups is shown in Figs. (3 and 4), respectively.
Table 5.
Comparison of objective image quality of spectral groups.
| Variable | Total (n = 256) | 40keV (n = 64) | 70keV(n = 64) | 110keV(n = 64) | Conventional CT images (n = 64) | P |
|---|---|---|---|---|---|---|
| Main pulmonary artery SNR | 13.21 ± 6.91 | 21.05 ± 9.05 | 11.6 ± 3.40 | 10.34±1.89 | 9.84 ± 3.35 | <.001 |
| Main pulmonary artery CNR | 9.99 ± 7.54 | 18.46 ± 9.54 | 9.13 ± 3.19 | 3.74 ± 1.73 | 8.64 ± 3.33 | <.001 |
| Pulmonary artery leaf SNR | 15.27 ± 10.22 | 20.79 ±12.43 | 15.4±7.05 | 11.23±10.5 | 13.64 ± 7.56 | <.001 |
| Pulmonary artery leaf CNR | 12.3 ± 11.07 | 27.38 ± 11.54 | 8.87 ± 4.30 | 4.14 ± 2.03 | 8.89 ± 4.19 | <.001 |
| Pulmonary segmental artery SNR | 14.79 ± 9.99 | 26.55 ±10.02 | 11.6± 7.43 | 10.8± 4.52 | 10.13 ± 6.29 | <.001 |
| Pulmonary segmental artery CNR | 12.44 ± 9.77 | 24.52 ± 10.40 | 9.13 ± 6.18 | 8.16 ± 3.41 | 7.95 ± 5.47 | <.001 |
| Lobar segmental artery CNR | 7.11 ± 11.55 | 19.86 ± 9.87 | 3.2 ± 10.20 | 2.71 ± 7.69 | 2.61 ± 7.66 | <.001 |
| Lobar segmental artery SNR | 9.76 ± 11.90 | 21.81 ± 9.99 | 5.8 ± 12.70 | 5.68 ± 7.73 | 5.94 ± 7.69 | <.001 |
Record the CNR and SNR values of the main pulmonary artery and branches in the comparative spectral group (40 keV, 70 keV, and 110 keV) and traditional CT images.
Table 6.
Comparison of objective image quality between the spectral group and the control group.
| Variable | Total(n = 124) | Spectrum 40 keV(n=64) | Control(n=60) | P |
|---|---|---|---|---|
| Main pulmonary artery SNR, | 20.55 ± 7.89 | 21.05 ± 9.05 | 20.00 ± 6.47 | 0.457 |
| Main pulmonary artery CNR | 18.20 ± 8.23 | 18.46 ± 9.54 | 17.93 ± 6.62 | 0.719 |
| Pulmonary artery leaf SNR | 20.62 ± 10.26 | 20.79 ± 12.43 | 20.44 ± 7.37 | 0.845 |
| Pulmonary artery leaf CNR | 26.36 ± 9.77 | 27.38 ± 11.54 | 25.28 ± 7.38 | 0.227 |
| Pulmonary segmental artery SNR | 25.89 ± 8.96 | 26.55 ± 10.02 | 25.18 ± 7.69 | 0.395 |
| Pulmonary segmental artery CNR | 23.54 ± 9.19 | 24.52 ± 10.40 | 22.50 ± 7.64 | 0.219 |
| Lobar segmental artery CNR | 19.52 ± 7.95 | 19.86 ± 9.87 | 19.16 ± 5.24 | 0.619 |
| Lobar segmental artery SNR | 21.20 ± 8.22 | 21.81 ± 9.99 | 20.56 ± 5.85 | 0.396 |
Fig. (2).

Shows the measurement methods of attenuation values (AV) and noise (SD) for the branches of the pulmonary artery using CT imaging at different energy levels. The left side represents the control group with traditional CT imaging, while the right side shows spectral imaging at 40 keV, 70 keV, and 110 keV, respectively. The ROIs (Region of interest) marked by red circles are located at the main trunk of the pulmonary artery, the lobar arteries, the segmental arteries, and the subsegmental arteries.
Fig. (3).

Box plots of the lobar pulmonary arteries SNR (A), main pulmonary artery SNR (B), segmental arteries SNR (C), and subsegmental arteries SNR (D) measured in the control group (conventional CT) and the spectral group at 40 keV, 70 keV, 110 keV, and conventional CT images. For SNR of pulmonary arteries at all levels, the 40 keV images yield significantly higher values compared to 70 keV and 110 keV images (P < 0.001), with no significant differences between the 40 keV spectral group and the control group.
Note: ***P < 0.001; ns = not significant (P > 0.05). The x-axis indicates the imaging groups, and the y-axis indicates the SNR values.
Fig. (4).

Box plots showing the lobar pulmonary arteries CNR (A), main pulmonary artery CNR (B), segmental arteries CNR (C), and subsegmental arteries CNR (D) in the control group (conventional CT images) and the spectral CT group at 40 keV, 70 keV, and 110 keV. For CNR of pulmonary arteries at all levels, the 40 keV images yield significantly higher values compared to 70 keV and 110 keV images (P < 0.001), with no significant differences between the 40 keV spectral group and the control group.
Note: ***P < 0.001; ns = not significant (P > 0.05). The x-axis indicates the imaging groups, and the y-axis indicates the CNR values.
4. DISCUSSION
This study demonstrates that an ultra-low iodine CTPA protocol using 160 mgI/mL contrast agent (25 mL, 2.5 mL/s) combined with dual-layer spectral CT and 40 keV virtual monoenergetic imaging achieves diagnostic image quality comparable to that of conventional CTPA, while reducing the iodine load by 67% (to 4 g per examination). Because renal outcomes were not assessed in the present study, this finding should be interpreted as evidence of image-quality feasibility and iodine-load reduction rather than direct evidence of reduced contrast-associated kidney injury. This distinction is clinically relevant because acute kidney injury after CTPA in patients with acute pulmonary embolism has been reported and may be influenced by hemodynamic severity and right ventricular dysfunction [19].
The spectral group employed a lower tube voltage (100 kV vs. 120 kV), which enhances iodine contrast and improves CNR. Despite the voltage reduction, radiation doses (CTDIvol, DLP, ED) were comparable between groups, likely due to the adaptive tube current modulation (DoseRight), which maintained consistent image noise levels. Similar findings have been reported in coronary and CTA dual-low imaging studies [20-25].
Compared with previous research, the contrast concentration reduction in this study is more substantial. Kristiansen et al. [26] reduced contrast concentration to 240 mgI/mL, whereas we achieved 160 mgI/mL-a 33% further reduction. In terms of iodine intake, conventional CTPA protocols typically involve 12–15 g of iodine per examination. Through multi-parameter optimization, we reduced this to 4 g, which is 30–50% lower than the low-dose protocol reported by Lee et al. [27] (approximately 6–8 g). Moreover, unlike studies that focused solely on reducing contrast volume, we simultaneously optimized concentration, injection rate, tube voltage, and spectral reconstruction, demonstrating the advantage of a multi-dimensional approach [28-30].
The 40 keV monoenergetic images provided the highest vascular attenuation, effectively compensating for the reduced iodine concentration. Prior CTA studies have shown that lowering tube voltage from 120 kV to 100 kV can substantially increase iodine attenuation [31]. In our study, the combination of low tube voltage and 40 keV reconstruction yielded significantly improved CT values compared with conventional images, with no compromise in subjective or objective quality [32-35]. Notably, distal vessel enhancement was better preserved with low-concentration contrast, consistent with findings by Kristiansen et al. [26]. Recent pulmonary CTA and related contrast-enhanced CT studies have also shown that advanced detector technology, low-energy VMI, and reconstruction- or AI-based enhancement techniques can preserve or improve image quality while supporting dose or contrast-media reduction [36-39].
This study has several limitations. First, it was a single-center study with a relatively small sample size, which may introduce selection bias and limit statistical power, particularly for assessing rare adverse events. Stratified analysis by renal function was not performed, although patients with renal impairment may benefit most from low-iodine protocols. Second, we did not monitor post-scan renal function, as the primary aim was to evaluate image quality rather than clinical outcomes; therefore, the impact of low-concentration contrast on CIN incidence could not be assessed. Third, the generalizability of this protocol may be limited to institutions equipped with spectral CT, and standardized protocols across centers are needed. For obese patients (BMI > 30 kg/m2), parameter adjustments may be necessary to maintain image quality [28,40]. Finally, while Xue L et al. [29] reported no significant increase in mild adverse events in low-contrast spectral pulmonary artery CT angiography, larger studies are needed to determine whether combining low concentration with reduced flow rate further lowers adverse event rates.
CONCLUSION
In conclusion, dual-layer spectral CT combined with 40 keV virtual monoenergetic imaging allowed CTPA to be performed with markedly reduced iodine load while maintaining subjective and objective image quality comparable to that of the conventional protocol. This protocol may be considered for selected patients when iodine dose reduction is desirable, but its impact on renal safety and broader clinical outcomes requires further prospective validation.
Table 3.
Comparison of subjective image quality scores between the spectral group and the control group.
| Variable | Spectrum Group 40 keV(n = 64) | Control Group (n = 60) | Statistical value | P |
|---|---|---|---|---|
| Reviewer A, Mean ± SD | 4.72 ± 0.60 | 4.55 ± 0.59 | t=1.57 | 0.12 |
| Reviewer B, Mean ± SD | 4.67 ± 0.57 | 4.62 ± 0.52 | t=0.56 | 0.57 |
ACKNOWLEDGEMENTS
Declared none.
LIST OF ABBREVIATIONS
- AV
Attenuation value
- BMI
Body mass index
- CIN
Contrast-induced nephropathy
- CNR
Contrast-to-noise ratio
- CT
Computed tomography
- CTDIvol
Volume CT dose index
- CTPA
Computed tomography pulmonary angiography
- DLP
Dose-length product
- ED
Effective dose
- HU
Hounsfield unit
- ICC
Intraclass correlation coefficient
- MonoE
Monoenergetic imaging
- PE
Pulmonary embolism
- ROI
Region of interest
- SBI
Spectral-based imaging
- SD
Standard deviation
- SNR
Signal-to-noise ratio
- VMI
Virtual monoenergetic imaging
AUTHORS’ CONTRIBUTIONS
The authors confirm their contributions to the paper as follows: Y.S., Z.H.: Study conception and design; T.Y., W.Y.: Data collection; Y.S., Z.H., W.J.: Analysis and interpretation; Y.S., Q.J.: Manuscript drafting. All authors reviewed and approved the final version.
ETHICS APPROVAL AND CONSENT TO PARTICIPATE
This study was approved by the Ethics Committee of Civil Aviation General Hospital, China (Approval No. 2024-L-K-123).
HUMAN AND ANIMAL RIGHTS
All procedures performed in studies involving human participants were in accordance with the ethical standards of institutional and/or research committees and with the 1975 Declaration of Helsinki, as revised in 2013.
CONSENT FOR PUBLICATION
Informed consent was obtained from all patients.
STANDARDS OF REPORTING
STROBE guidelines were followed.
AVAILABILITY OF DATA AND MATERIALS
The data and supportive information is available within the article.
FUNDING
None.
CONFLICT OF INTEREST
The authors declare no conflict of interest, financial or otherwise.
REFERENCES
- 1.Konstantinides S.V., Meyer G. The 2019 ESC guidelines on the diagnosis and management of acute pulmonary embolism. Eur. Heart J. 2019;40(42):3453–3455. doi: 10.1093/eurheartj/ehz726. [DOI] [PubMed] [Google Scholar]
- 2.Keller K., Hobohm L., Ebner M., Kresoja K.P., Münzel T., Konstantinides S.V., Lankeit M. Trends in thrombolytic treatment and outcomes of acute pulmonary embolism in Germany. Eur. Heart J. 2020;41(4):522–529. doi: 10.1093/eurheartj/ehz236. [DOI] [PubMed] [Google Scholar]
- 3.Kow R. Y., Razali K., Low C. L., et al. Bibliometric analysis of diagnostic yield of CT Pulmonary Angiogram (CTPA) in the diagnosis of Pulmonary Embolism (PE). Cureus. 2023;15(7):e41979. doi: 10.7759/cureus.41979. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Melo R.H., Gioli-Pereira L., Lourenço I.D., Da Hora Passos R., Bernardo A.T., Volpicelli G. Diagnostic accuracy of multi-organ point-of-care ultrasound for pulmonary embolism in critically ill patients: a systematic review and meta-analysis. Crit. Care. 2025;29(1):162. doi: 10.1186/s13054-025-05359-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.de Jong C.M.M., Kroft L.J.M., van Mens T.E., Huisman M.V., Stöger J.L., Klok F.A. Modern imaging of acute pulmonary embolism. Thromb. Res. 2024;238:105–116. doi: 10.1016/j.thromres.2024.04.016. [DOI] [PubMed] [Google Scholar]
- 6.Batra K., Kay F.U., Sibley R.C., Peshock R.M. Imaging of acute pulmonary embolism: An update. Radiol. Clin. North Am. 2025;63(2):207–222. doi: 10.1016/j.rcl.2024.08.003. [DOI] [PubMed] [Google Scholar]
- 7.Diaz-Lorenzo I., Alonso-Burgos A., Friera Reyes A., Pacios Blanco R.E., de Benavides Bernaldo de Quiros M.C., Gallardo Madueño G. Current role of CT pulmonary angiography in pulmonary embolism: A state-of-the-art review. J. Imaging. 2024;10(12):323. doi: 10.3390/jimaging10120323. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Campion D., Ponzo P., Risso A., Caropreso P., Caviglia G.P., Sanavia T., Frigo F., Bonetto S., Giovo I., Rizzo M., Martini S., Bugianesi E., Mengozzi G., Marzano A., Manca A., Saracco G.M., Alessandria C. A prospective, multicenter, three-cohort study evaluating contrast-induced acute kidney injury (CI-AKI) in patients with cirrhosis. J. Hepatol. 2024;80(1):62–72. doi: 10.1016/j.jhep.2023.10.010. [DOI] [PubMed] [Google Scholar]
- 9.Ahn K.M., Ahn Y.H., Cho M.K., Kang D.Y., Lee S.Y., Kang H.R. Validation of practical pathway in patients with anaphylaxis to low osmolar contrast media: A retrospective cohort study. J. Allergy Clin. Immunol. Pract. 2022;10(10):2685–2692.e2. doi: 10.1016/j.jaip.2022.03.030. [DOI] [PubMed] [Google Scholar]
- 10.Ganigara M., Priya S., Nagpal P. Optimizing contrast administration in pediatric cardiovascular CT: Safety and technique. J. Cardiovasc. Comput. Tomogr. 2025;19(6):736–745. doi: 10.1016/j.jcct.2025.09.008. [DOI] [PubMed] [Google Scholar]
- 11.Yang L., Sun J., Li J., Peng Y. Dual-energy spectral CT imaging of pulmonary embolism with Mycoplasma pneumoniae pneumonia in children. Radiol. Med. (Torino) 2022;127(2):154–161. doi: 10.1007/s11547-021-01442-9. [DOI] [PubMed] [Google Scholar]
- 12.Sauter A.P., Shapira N., Kopp F.K., Aichele J., Bodden J., Knipfer A., Rummeny E.J., Noël P.B. CTPA with a conventional CT at 100 kVp vs. a spectral-detector CT at 120 kVp: Comparison of radiation exposure, diagnostic performance and image quality. Eur. J. Radiol. Open. 2020;7:100234. doi: 10.1016/j.ejro.2020.100234. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Ethiraju V., Varatharajaperumal R., Arunachalam V.K., KuruVambath A., Punniyakotti R.S., Rajasekaran S., Mehta P., Cherian M. Role of virtual monoenergetic images in the assessment of vessel enhancement in segmental level in third-generation dual-source dual-energy CT pulmonary angiography: A prospective study. Indian J. Radiol. Imaging. 2025;35(2):254–262. doi: 10.1055/s-0044-1788575. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Qiu S.M., Zhang H., Liu Z.X., Zhang L., Meng Y.K., Sun X.N., Xie L.X., Zhang Y.C., Wang H., Xu K. The application value of deep learning image reconstruction on improving image quality and evaluating the Qanadli embolism index of dual low-dose CT pulmonary angiography. Zhonghua Yi Xue Za Zhi. 2023;103(19):1477–1482. doi: 10.3760/cma.j.cn112137-20230313-00392. [DOI] [PubMed] [Google Scholar]
- 15.Bellizzi A., Bezzina P., Zarb F. Low dose CTPA using a low kV technique combined with high IR: A clinical study. Radiography. 2023;29(4):738–744. doi: 10.1016/j.radi.2023.05.001. [DOI] [PubMed] [Google Scholar]
- 16.Aldosari S., Sun Z. A systematic review of double low-dose CT pulmonary angiography in pulmonary embolism. Curr. Med. Imaging. 2019;15(5):453–460. doi: 10.2174/1573405614666180813120619. [DOI] [PubMed] [Google Scholar]
- 17.Krasniqi L.Ø., Pietersen P.I., Kristiansen C.H., et al. Dual-energy CT pulmonary angiography with reduced contrast media volume-a randomized controlled trial. J. Med. Imaging Radiat. Sci. 2026;55(3):432–440. doi: 10.1016/j.jmir.2025.11.006. [DOI] [PubMed] [Google Scholar]
- 18.Borkowski P., Ng E., Vicaretti M., Young N. Use of ultra-low contrast dose CT aortography for the management of aortic aneurysmal disease. J. Cardiothorac. Surg. 2024;19(1):582. doi: 10.1186/s13019-024-03099-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Choi B., Heo S., Mcdonald J.S., Choi S.H., Choi W.M., Lee J.B., Lee E.A., Park S.H., Seol S., Gan S., Park B., Choi H.J., Kim B.J., Rhee S.Y., Hong S.B., Kim K.H., Lee Y.H., Kim S.S., Park R.W. Risk of contrast-induced acute kidney injury in computed tomography: A 16 institutional retrospective cohort study. Invest. Radiol. 2025;60(6):376–386. doi: 10.1097/RLI.0000000000001141. [DOI] [PubMed] [Google Scholar]
- 20.Yurt A., Özsoykal İ., Obuz F. Effects of the use of automatic tube current modulation on patient dose and image quality in computed tomography. Mol. Imaging Radionucl. Ther. 2019;28(3):96–103. doi: 10.4274/mirt.galenos.2019.83723. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Kapoor A., Mahajan G., Kapoor A. Evaluation of low-kVp low-volume iodinated contrast protocol for coronary CT angiography using retrospective ECG gating. Indian J. Radiol. Imaging. 2021;31(4):910–916. doi: 10.1055/s-0041-1741102. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Euler A., Taslimi T., Eberhard M., Kobe A., Reeve K., Zimmermann A., Krauss A., Gutjahr R., Schmidt B., Alkadhi H. Computed tomography angiography of the aorta-optimization of automatic tube voltage selection settings to reduce radiation dose or contrast medium in a prospective randomized trial. Invest. Radiol. 2021;56(5):283–291. doi: 10.1097/RLI.0000000000000740. [DOI] [PubMed] [Google Scholar]
- 23.Rajiah P.S., Dunning C.A.S., Rajendran K., Tandon Y.K., Ahmed Z., Larson N.B., Collins J.D., Thorne J., Williamson E., Fletcher J.G., McCollough C., Leng S. High-pitch multienergy coronary CT angiography in dual-source photon-counting detector CT scanner at low iodinated contrast dose. Invest. Radiol. 2023;58(9):681–690. doi: 10.1097/RLI.0000000000000961. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Yi Y., Zhao X.M., Wu R.Z., Wang Y., Vembar M., Jin Z.Y., Wang Y.N. Low dose and low contrast medium coronary CT angiography using dual-layer spectral detector CT. Int. Heart J. 2019;60(3):608–617. doi: 10.1536/ihj.18-340. [DOI] [PubMed] [Google Scholar]
- 25.Sun J., Li H., Gao J., Li J., Li M., Zhou Z., Peng Y. Performance evaluation of a deep learning image reconstruction (DLIR) algorithm in “double low” chest CTA in children: a feasibility study. Radiol. Med. (Torino) 2021;126(9):1181–1188. doi: 10.1007/s11547-021-01384-2. [DOI] [PubMed] [Google Scholar]
- 26.Kristiansen C.H., Thomas O., Nyquist A.B., Sanderud A., Boavida J., Geitung J.T., Tran T.T., Lauritzen P.M. A randomised controlled trial comparing three clinical administration strategies in spectral detector CT pulmonary angiography with low contrast medium dose. Eur. Radiol. 2025;35(8):4649–4660. doi: 10.1007/s00330-025-11420-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Lee W.C., Poon J.K., Siah J.J.H., Chong M.C., Lai C. Feasibility of low contrast volume and low injection flow rate in CT pulmonary angiography. J. Med. Imaging Radiat. Sci. 2025;56(1):101349. doi: 10.1016/j.jmir.2023.11.009. [DOI] [PubMed] [Google Scholar]
- 28.Fang X., Tong X., Liu Y., et al. Individualized scan protocol based on body mass index in dual-energy computed tomography pulmonary angiography to reduce radiation and contrast doses with improved image quality. Med. Sci. Monit. 2023;29:e939234. doi: 10.12659/MSM.939234. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Xue L., Zhong Q., Xu N., Zheng Y., Liu Y. Investigation in image quality and immediate patient safety using pre-dual-flow injection for low-contrast dose spectral pulmonary artery CT angiography. Eur. J. Radiol. Open. 2024;12:100571. doi: 10.1016/j.ejro.2024.100571. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Overhoff D., Jost G., McDermott M., Wichtmann B., Schoenberg S.O., Pietsch H., Attenberger U. Contrast saline mixture dualflow injection protocols for low-kilovolt computed tomography angiography: A systematic phantom and animal study. Invest. Radiol. 2020;55(12):785–791. doi: 10.1097/RLI.0000000000000706. [DOI] [PubMed] [Google Scholar]
- 31.Lenfant M., Comby P.O., Guillen K., Galissot F., Haioun K., Thay A., Chevallier O., Ricolfi F., Loffroy R. Deep learning-based reconstruction vs. iterative reconstruction for quality of low-dose head-and-neck CT angiography with different tube-voltage protocols in emergency-department patients. Diagnostics (Basel) 2022;12(5):1287. doi: 10.3390/diagnostics12051287. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Noda Y., Ando T., Kaga T., Yamda N., Seko T., Ishihara T., Kawai N., Miyoshi T., Ito A., Naruse T., Hyodo F., Kato H., Kambadakone A.R., Matsuo M. Pancreatic cancer detection with dual-energy CT: diagnostic performance of 40 keV and 70 keV virtual monoenergetic images. Radiol. Med. (Torino) 2024;129(5):677–686. doi: 10.1007/s11547-024-01806-x. [DOI] [PubMed] [Google Scholar]
- 33.Zhou X., Cui M., Liu Y., Wu Y., Hu D., Zhai D., Qin M., Shen J., Ju S., Fan G., Cai W. Low dose iodinated contrast material and radiation for virtual monochromatic imaging in craniocervical dual-layer spectral detector computed tomography angiography: A prospective and randomized study. Acad. Radiol. 2024;31(6):2501–2510. doi: 10.1016/j.acra.2023.12.004. [DOI] [PubMed] [Google Scholar]
- 34.Leng Y., Liu Y., Li S., Wang X., Deng W., Wang Y., Gong L. Feasibility of using 8 mL of iodinated contrast media in cerebral computed tomographic angiography with a dual-layer spectral detector. Quant. Imaging Med. Surg. 2024;14(1):514–526. doi: 10.21037/qims-23-914. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Gliner-Ron M., Sosna J., Leichter I., Goldberg S.N., Shaham D., Cohen D., Malul Y., Romman Z., Lev-Cohain N. Evaluation of the pulmonary arteries on CTPA with dual energy CT: Objective analysis and subjective preferences in a multireader study. J. Thorac. Imaging. 2024;39(4):201–207. doi: 10.1097/RTI.0000000000000782. [DOI] [PubMed] [Google Scholar]
- 36.Pannenbecker P., Heidenreich J.F., Grunz J.P., Huflage H., Gruschwitz P., Patzer T.S., Feldle P., Bley T.A., Petritsch B. Image quality and radiation dose of CTPA with iodine maps: A prospective randomized study of high-pitch mode photon-counting detector CT versus energy-integrating detector CT. AJR Am. J. Roentgenol. 2024;222(2):e2330154. doi: 10.2214/AJR.23.30154. [DOI] [PubMed] [Google Scholar]
- 37.Yalon M., Hoodeshenas S., Chan A., Horst K.K., Crum I., Thorne J.E., Lee Y.S., Yu L., McCollough C.H., Fletcher J.G., Rajiah P.S. Improved pulmonary artery evaluation using high-pitch photon-counting CT compared to high-pitch conventional or routine-pitch conventional dual-energy CT. J. Comput. Assist. Tomogr. 2024;48(6):897–905. doi: 10.1097/RCT.0000000000001645. [DOI] [PubMed] [Google Scholar]
- 38.Haubold J., Jost G., Theysohn J.M., Ludwig J.M., Li Y., Kleesiek J., Schaarschmidt B.M., Forsting M., Nensa F., Pietsch H., Hosch R. Contrast media reduction in computed tomography with deep learning using a generative adversarial network in an experimental animal study. Invest. Radiol. 2022;57(10):696–703. doi: 10.1097/RLI.0000000000000875. [DOI] [PubMed] [Google Scholar]
- 39.Takata T., Yamada K., Yamamoto M., Kondo H. Generative deep-learning-model based contrast enhancement for digital subtraction angiography using a text-conditioned image-to-image model. Comput. Biol. Med. 2025;195:110598. doi: 10.1016/j.compbiomed.2025.110598. [DOI] [PubMed] [Google Scholar]
- 40.Gohmann R.F., Gottschling S., Seitz P., Temiz B., Krieghoff C., Lücke C., Horn M., Gutberlet M. 3D-segmentation and characterization of visceral and abdominal subcutaneous adipose tissue on CT: influence of contrast medium and contrast phase. Quant. Imaging Med. Surg. 2021;11(2):697–705. doi: 10.21037/qims-20-907. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
The data and supportive information is available within the article.