Abstract
Targeted protein degradation (TPD) is a therapeutic strategy that utilizes small molecules to induce the proximity-driven degradation of disease-causing proteins. Because the efficacy and selectivity of TPD compounds must be validated across thousands of proteins, high-throughput proteomics is essential for the rapid screening and characterization of these novel degraders. Here, we developed a 300 sample per day (SPD) LC-MS/MS method using the Orbitrap Astral mass spectrometer for ultra-high-throughput TPD compound screening. We identified close to 8,000 protein groups from a single cell line with a coefficient of variation (CV) of less than 10%, highlighting the deep proteome coverage and method reproducibility even at 300 SPD. This high degree of precision provides the statistical confidence to detect subtle, yet significant, changes in protein abundance that were previously challenging to quantify in high-throughput workflows. To evaluate the quantitation accuracy of this method, we further mixed the digests from two or three species at different ratios. Our three-proteome mixture results demonstrated highly accurate quantitation for proteins with both small and large fold changes. Moreover, our two-proteome mixture experiment, where 20 to 160 ng of yeast digest was spiked into 200 ng of HeLa digest, showed an R2 of 0.999 for the yeast proteome, underscoring the quantitation accuracy of the method. Utilizing this workflow, we studied dose-dependent protein degradation patterns induced by pomalidomide, iberdomide, and mezigdomide. Our results indicate that mezigdomide may possess enhanced efficacy in T cells by degrading additional proteins such as IKZF2, thereby boosting anti-cancer immunity. Together, we developed an ultra-high-throughput LC-MS/MS method with excellent proteome coverage and quantitation accuracy that is highly suitable for chemoproteomics screening of drug libraries.
Keywords: Quantitative proteomics, High-throughput screening, Targeted protein degradation, Chemoproteomics, Orbitrap Astral, Molecular glue
Graphical Abstract

Introduction
Mass spectrometry-based proteomics has become an indispensable tool for understanding complex biological systems and for driving drug discovery. In chemoproteomics and targeted-protein degradation (TPD) field, the ability to screen large libraries of compounds against cellular proteomes is critical for identifying and characterizing novel therapeutics. This necessitates the development of high-throughput analytical methods that do not sacrifice depth, precision, or quantitative accuracy 1–4. A key challenge has been the reliable detection of subtle protein abundance changes (e.g., 10–20% degradation), as these modest effects are often obscured by analytical variance in conventional high-speed methods.
Over the last five years, data-independent acquisition (DIA) has emerged as a powerful technique for quantitative proteomics with the maturation of data analysis algorithm and software 5–10, offering comprehensive peptide detection and reproducible quantification across large sample cohorts. Unlike data-dependent acquisition (DDA), DIA systematically fragments all ions within predefined mass-to-charge (m/z) windows, creating complex but information-rich datasets that can be retrospectively mined 8,11–14. However, the adoption of DIA for ultra-high-throughput screening has been historically limited by a trade-off between analytical speed, proteome coverage, and sensitivity. Shortening liquid chromatography (LC) gradients to increase sample throughput often leads to a significant loss in the number of identified and quantified proteins.
The recent introduction of the Orbitrap Astral mass spectrometer marks a significant advancement in addressing these challenges. This innovative platform combines a high-resolution Orbitrap mass analyzer with a novel, high-speed Astral mass analyzer, enabling unprecedented scan speeds (up to 200 Hz) with high resolution, sensitivity and a wide dynamic range 15,16. This unique architecture is exceptionally well-suited for DIA, as it allows for the rapid acquisition of high-quality fragment ion spectra even with the narrow chromatographic peaks produced by ultra-short LC gradients 17,18. This capability opens the door to achieving both high throughput and high quantitative fidelity simultaneously.
In this study, we leverage the capabilities of the Orbitrap Astral mass spectrometer to develop and validate an ultra-high-throughput LC-MS/MS workflow capable of analyzing 300 samples per day (SPD). We systematically benchmark the quantitative accuracy and precision of the method using standardized multi-proteome mixtures, with a particular focus on its ability to accurately quantify small fold changes against a complex background. We then demonstrate its suitability for deep and reproducible profiling of human cell lines. Finally, we apply this workflow to a real-world TPD study, profiling the dose-dependent effects of several immunomodulatory drugs (IMiDs) to showcase its power for high-throughput compound screening.
Methods
Cell Culture and Drug Treatment
Jurkat cell line was purchased from ATCC. Jurkat cells were cultured in RPMI-1640 (Corning, 10-040-CV) supplemented with 10% FBS (Corning, 35-011-CV) in the 37 °C incubator with 5% CO2. The day before treatment, 100 μL Jurkat cells were plated in a 96-well U-bottom plate with a density of 2 million/mL. On the following day, a serial dilution of degrader compounds in DMSO or equal volume of DMSO were added into each well, making the final compound concentrations as 8, 40, 200, 1000 nM. Two biological replicates were prepared for each compound concentration and DMSO-treated controls.
TPD Global Proteomics Sample Preparation
After 24 hours of degrader treatment, cells were centrifuged down, and the media was aspirated, followed by a 1x PBS wash. Cell pellets were lysed in 50 μL of 200 mM HEPES, 0.25% SDS at pH 8.5, with 1x Halt protease inhibitor cocktail (Thermo Scientific, 78440) and 10U/well benzonase (Sigma, E1014), shaking at 800 rpm at 37°C for 30 min. Reduction was performed by adding 5 μL of 50 mM DTT and shaking at 37°C for 15 min, followed by alkylation with 14 μL of 100 mM iodoacetamide at room temperature in the dark for 30 min. To each well, 100 μg of SP3 beads (a 1:1 mixture of E3 and E7 carboxylic magnetic beads from Cytiva) were added. Protein binding was initiated by adding 75 μL of 100% ethanol to each well and shaking at 800 rpm for 10 min at room temperature. The beads were washed twice on a magnetic rack with 200 μL of 80% ethanol. After the final wash, beads were briefly dried, and proteins were digested in 30 μL of 12.5 ng/μL trypsin/LysC (Promega, V5073) in 200 mM HEPES, shaking at 1,000 rpm at 37°C for 18 hours with the plate sealed.
The digested peptides were acidified with 1 μL/well of formic acid (FA) and further desalted and cleaned up with StageTips (CDS Analytical, 6091). Tips were moistened by adding 30 μL of 80% ACN/0.1% FA, centrifuging the assembly at 800 rpm for 1 min, and removing the waste liquid. Air was plunged through to remove excess liquid from the tips. The tips were then equilibrated by adding 30 μL of 0.1% FA and repeating the centrifugation and plunging steps. The acidified digests were loaded onto the StageTips, followed by centrifugation and plunging. The tips were washed twice with 30 μL of 0.1% FA and peptides were eluted with 100 μL of 80% ACN/0.1% FA into the 96-well PCR collection plate. The eluate plate was dried in a speed vac, and peptides were resuspended in 30 μL of 5% ACN/0.1% FA. Peptide concentration was measured with a Fluorescent Quantitative Peptide Assay (Pierce, 23290) and normalized across the plate before LC-MS/MS injection.
Generation of Proteome Mixtures
Human HeLa cell digest (Thermo Scientific, 88328), E. coli digest (Waters Corporation, 186003196), and yeast digest (Thermo Scientific, A47951) standards were dissolved in 5% ACN/0.1% FA as 10 ng/μL stock solutions. For three proteome mix samples, various volumes of digests were combined to make the final sample containing 2 μg total proteomes with the specified percentage of components. For two-proteome mix samples, 20, 40, 60, 80, 100, and 160 ng of yeast digest were spiked into a 200 ng HeLa digest background. Samples were Speed-Vac dried and reconstituted in 10 μL 5% ACN/0.1% FA immediately before loading for LC-MS/MS analysis.
Liquid Chromatography with tandem Mass Spectrometry (LC-MS/MS) Analysis
A total of 50 or 200 ng of the reconstituted peptides were loaded onto a PepMap Neo Trap Cartridge (Thermo Scientific, 174500) using a Vanquish Neo UHPLC system (Thermo Scientific) and separated by a 15-cm EASY-Spray PepMap column (Thermo scientific, ES906) for 60 or 180 SPD, or a 50-cm EASY-Spray PepMap Neo HPLC column (Thermo Scientific, ES75500PN) for 24 SPD analysis. For 300 SPD analysis, the peptides were resolved by an Aurora Rapid® 5×75 C18 UHPLC column or Aurora Rapid® 8×75 XT C18 UHPLC column (IonOpticks). Flow rate and gradients were optimized for individual column and throughput, and the parameters are detailed in Supplementary Table 1.
For DIA analysis, MS1 spectra were collected on the Orbitrap mass analyzer with a resolution of 240,000 and a scan range of 380–980 m/z and 580–780 m/z for 24, 60, 180 SPD and 300 SPD, respectively. The AGC was set as 500%, and the maximum injection time was set to 5 ms. Precursor ions were fragmented through HCD with a normalized collision energy (NCE) of 25%. MS2 spectra were acquired on the Astral mass analyzer with the parameters of each of the throughput listed in Supplementary Table 1.
Data Processing and Statistical Analysis
Data was processed by Spectronaut software (Biognosys, v19) using the directDIA approach. The spectra were searched against human, yeast or E. coli proteome downloaded from Uniprot. Enzyme was set as trypsin that cleaves the C-terminal of lysine or arginine except for those immediately after proline. Carbamidomethylating on cysteine was set as fixed modification; acetylation on protein N-terminal and oxidation on methionine was set as variable modification. The FDR control was set as 1% for PSM, peptide and protein group.
DIA data acquired using each chromatographic gradient were processed and analyzed separately in Spectronaut to ensure optimal performance under the specific chromatographic conditions of each gradient. Independent searches were performed to minimize retention time variability between gradients and to prevent potential inflation of peptide and protein identifications in the shorter-gradient datasets. This strategy enabled an accurate and unbiased evaluation of method performance for each chromatographic condition.
Normalization was performed using the human peptide signal, which represents the predominant proteome in both the two- and three-proteome mixture experiments. This approach is based on the assumption of common global proteomics setup that the majority of the proteome remains unchanged across conditions. Local normalization was applied using the algorithm implemented in Spectronaut, as described by Callister et al., to correct for systematic intensity differences between runs while preserving relative quantitative relationships. Additional details have been incorporated into the Methods section of the revised manuscript.
Proteomics data analysis was performed using a custom Python script (version 3.9.7). The protein quantification report in pivot format exported from Spectronaut or pg/pr_matrix from DIA-NN served as the input for the analysis. Data preprocessing involved filtering out common contaminants, and options for data imputation were employed to handle missing values in both treatment and reference groups, ensuring robust comparisons even with incomplete datasets.
The data points per peak are calculated by considering all identified precursors, with values reported based on the number of sampling points between the peak start and peak end (base-to-base) for each precursor.
For the three-proteome benchmarking experiments, quantitative accuracy was assessed using the log2Ratio. The measured log2Ratio for a specific protein was calculated as:
where represents the protein group intensity reported by Spectronaut after global normalization. The “expected values” (theoretical ratios) were derived from the known mass-based mixing ratios of the human (HeLa), yeast, and E. coli digests. For example, in the standard mixture (Figure 1B), the expected log2Ratio for yeast was calculated based on the 75 ng (Mixture A) and 150 ng (Mixture B) loading amounts, resulting in a theoretical log2(150/75) = 1.0. To maintain quantitative rigor, no imputation methods were used to fill missing values in the benchmarking datasets. For the calculation of log2Ratios in the three-proteome mixtures, a protein was required to have valid quantitative values in at least two out of three replicate injections within a group. Proteins failing to meet this criterion were excluded from the ratio analysis and corresponding plots. This conservative approach ensures that the reported accuracy and precision metrics accurately reflect the raw performance of the Orbitrap Astral MS without artificial smoothing.
Figure 1. Benchmarking Quantitative Accuracy Across Various Throughputs.

(A) Schematic overview of the liquid chromatography with tandem mass spectrometry (LC-MS/MS) workflow.
(B) Composition of the standard three-proteome mixture used to assess quantitative performance, with digests from human (HeLa), yeast, and E. coli mixed at defined ratios.
(C) A bar chart showing the number of protein groups for different gradient lengths.
(D) Box plots showing the measured log2Ratio of proteins from each proteome. The analysis was performed at throughputs of 24, 60, 180, and 300 SPD. The dashed lines indicate the theoretically expected ratios, demonstrating high quantitative accuracy across different gradient lengths.
(E) Bar charts showing the median of observed ratios across different species. The gray dashed lines indicate theoretical ratios defined by sample mixtures in Figure 1B.
For the two-proteome mixture experiment, protein intensities were first normalized to the 160 ng group to obtain relative fold changes. To assess linearity, the relative fold change for each quantified protein was calculated and subsequently averaged across all proteins. The spiked-in yeast digest amounts and the corresponding averaged fold changes (normalized to the 160 ng group) were then fitted using the linregress function in SciPy (Python) to determine the slope and y-intercept.
Differential abundance analysis was conducted to compare treatment groups against a reference group (e.g., DMSO). An R-limma model was implemented for statistical analysis, leveraging the limma library to identify significant changes. Proteins were considered significantly differentially expressed if they met specific fold-change (log2FC) and adjusted p-value thresholds, defined as log2FC < −1 and Padj < 0.01 for downregulation.
For visualization and pattern recognition, hierarchical clustering was performed on biologically significant proteins. Pairwise distances were computed, and Ward's method was used for hierarchy linkage. Volcano plots were generated to visualize the distribution of fold changes relative to adjusted p-values. Additionally, a comprehensive bubble dendrogram plot was developed to simultaneously display the magnitude of change (color scale representing log2FC) and statistical significance (bubble size representing −log10 Padj) across different treatment conditions, integrated with hierarchical clustering results.
Results
Benchmarking the Orbitrap Astral with Various Throughputs
To establish the performance of the Orbitrap Astral mass spectrometer for our high-throughput proteomics workflow (Figure 1A), we first evaluated its quantitative accuracy with both long and short gradients. We analyzed a standard three-proteome mixture (Human, Yeast, E. coli, Figure 1B) at different throughputs, ranging from 24 to 300 samples per day (SPD) (Table S1). The results demonstrated that the instrument is highly suitable for both ultra-high-throughput analysis and longer gradients intended for in-depth proteome coverage. As expected, the number of quantified proteins increased with longer active gradients (Figure 1C). Moreover, we observed a modest decrease in quantitative precision and accuracy as throughput increased from 24 to 300 SPD, characterized by a slight widening of the ratio distributions. However, even at the 300 SPD limit, the measured protein ratios for each species aligned closely with theoretical values (Figure 1D–E). The median numbers of data points per peak are 2–4 due to the specific scanning setup employed (Figure S1B). For this reason, we thoroughly evaluated the quantitative performance among all gradients. As expected, the protein coefficients of variation (CVs) increased with shorter gradient, however, the impact remained within acceptable limits for screening applications. Even for the E. coli proteome, which represents the lowest abundance fraction in our mixtures, the 300 SPD method maintained a median CV < 20%, while the human and yeast backgrounds were approximately 10% (Figure S1A). All these results together suggest that the high sensitivity and scan stability of the Astral analyzer partially compensate for lower sampling density, providing sufficient robustness for detecting significant biological perturbations in high-throughput drug discovery.
Assessing Quantitation in Extreme Ratio Three-Proteome Mixtures
To further challenge the quantitation capabilities of the workflow, we designed an “extreme ratio” three-proteome mixture. In this experiment, digests of E. coli and yeast were mixed in varying, inverse proportions against a constant background of 50% of HeLa digest. This simultaneously created samples with both large fold changes for the yeast proteome (from 0.5% to 20%, a 40-fold change) and subtle fold changes for the E. coli proteome (from 49.5% to 30%, a ~1.6-fold change) (Figure 2A). Even with the 300 SPD method, the workflow consistently identified 7,000–8,000 protein groups across all mixtures from a 200 ng total protein load (Figure 2B).
Figure 2. Accurate Quantification of Extreme Ratio Three-Proteome Mixtures.

(A) Schematic table detailing the composition of the five extreme-ratio mixtures, with varying amounts of yeast and E. coli digest against a constant HeLa background. The expected log2Ratio between samples is shown for both yeast and E. coli.
(B) Bar charts showing the number of protein groups in each mixture sample with a 200 ng loading amount.
(C) Quantitative performance of proteome mixtures with 200 ng loading amount and 300 SPD method. In each comparison, the scatter plot shows measured log2FoldChange versus log2Intensity for pairwise comparisons between mixtures (MA plots, top left). Proteins for S. cerevisiae (red), E. coli (blue), and H. sapiens (gray) are shown. The density plot (top right) illustrates the distribution of observed fold-changes, with the peak maxima values annotated. Residual analysis is presented in the bottom panels: a scatter plot of residuals (Observed − Expected log2FC) versus log2Intensity (bottom left) and a violin plot showing the distribution of residuals for each organism (bottom right). Three technical replicates were performed.
(D) Distribution of Protein Coefficient of Variation (CV) across five extreme 3-proteome mixtures analyzed with 200 ng loading amount and 300 SPD method. The boxplots display the distribution of CVs for proteins attributed to S. cerevisiae (red), E. coli (blue), and H. sapiens (gray) in each mix. The box represents the interquartile range (IQR) with the median; whiskers extend to 1.5 times the IQR.
The workflow showed great linearity in quantifying these mixtures. Remarkably, the subtle fold-changes of the E. coli proteome were quantified with high accuracy, closely tracking the expected theoretical ratios (Figure 2C, blue dots). As expected, some ratio compression was observed for the large-fold-change yeast proteome, particularly for lower abundance proteins (Figure 2C, red dots). The performance was also validated with even lower sample loading amounts (50 ng in Figure S3A). To further evaluate method precision across varying abundances, we analyzed the CV for each organism in the extreme-ratio mixtures (Figure 2D). While the human and E. coli proteins showed consistently low variation with a median CV < 10%, the yeast proteins exhibited clear abundance-dependent trends. In Mix 1, where the yeast component is as low as 1 ng (0.5% of the total protein load), we observed an increased spread in quantified ratios. This indicates that these low-abundance proteins are reaching the boundary conditions of the 300 SPD analysis, a precision limit that is further detailed in the boxplots provided in Figure S8.
To benchmark this performance against established high-resolution acquisitions, we compared the 300 SPD results with a longer 60 SPD gradient. We observed improvements in quantitative accuracy with the longer gradient, characterized by reduced ratio compression in high-fold-change comparisons (Figure 2C vs Figure S2, and Figure S3A vs S3B). Importantly, both gradients maintained excellent accuracy for the smaller E. coli ratios, indicating that the significant gain in throughput with the 300 SPD method comes with only a minor trade-off in accuracy for large fold-changes. Collectively, these 3-proteome mixture data suggest that while the 300 SPD method is highly precise, longer gradients and higher loading amounts can further enhance quantification accuracy, likely due to increased peptide identifications supporting protein-level quantification. We also observed less ratio compression when comparing the fold changes on the peptide level (Figure S4).
All these observation highlights the ability of LFQ-DIA on Orbitrap Astral mass spectrometer to maintain quantitative accuracy over a wide dynamic range, a critical feature for identifying significant biological changes in complex samples. Crucially, this experiment demonstrates the method's strength in confidently quantifying small, precise changes even when other proteins in the sample are changing dramatically.
Development and Validation of a 300 SPD Method Using an 8-cm Column
Building on the initial benchmarks, we further developed an optimized ultra-high-throughput method capable of analyzing 300 SPD (Table S1). To achieve this, we utilized a recently released 8-cm IonOpticks Aurora Rapid column, which is compatible with commercial heating devices, ensuring greater chromatographic reproducibility through temperature control. Despite the very short gradient time, this method achieved an impressive 87% usage of the total MS acquisition time, maximizing instrument efficiency (Figure 3A).
Figure 3. In-depth and Precise Proteome Analysis at 300 SPD.

(A) A representative total ion chromatogram from a 200 ng HeLa and K562 digest analyzed with the 4-minute gradient on an 8-cm column.
(B) The number of protein and peptide groups identified from HeLa and K562 cells (n=5, mean ± SD).
(C) Violin plots showing the coefficient of variation (CV) of quantified protein groups percentages from five replicate injections for each cell line, with medians labeled.
(D) Principal component analysis (PCA) plot of quantified proteins, showing clear separation between the HeLa and K562 cell lines.
(E) Bar charts showing the quantitative results of cell line specific markers, including CA1 and HBG2 for K562 cells, and KRT8 and 18 for HeLa cells.
To evaluate the proteome coverage and quantitative precision of this new method, we analyzed 200 ng digests of two different human cell lines, HeLa and K562. The workflow identified 7,944 protein groups in HeLa and 7,583 protein groups in K562, along with over 60,000 and 57,000 peptide groups, respectively, indicating excellent proteome depth for screening applications (Figure 3B). Furthermore, the median CV for protein quantification was approximately 9% for both cell lines, demonstrating high precision (Figure 3C). This high reproducibility is the foundation for confidently detecting small biological perturbations. A principal component analysis (PCA) showed clear and distinct clustering of the two cell lines, confirming that the method is sensitive enough to distinguish between different biological samples (Figure 3D).
To further confirm data quality, we examined cell line specific markers. K562 cells are erythroleukemic and are expected to exhibit increased expression of erythroid enzymes and hemoglobins. This was confirmed by the elevated expression of CA1 and HBG2 in K562 samples. In contrast, HeLa cells are adherent epithelial cells that require higher levels of keratin proteins, which should not be considered as background contamination in the HeLa dataset. Consistent with this, we observed substantially higher expression of KRT8 and KRT18 in HeLa compared to K562 cells (Figure 3E).
Demonstrating Quantitative Linearity with a Two-Proteome Mixture
After establishing the depth and precision of the 8 cm column 300 SPD method, we assessed its quantitative linearity. To evaluate this, we conducted a classic three-proteome mixture experiment, in which human, yeast, and E. coli digests were combined at defined ratios (Figure 1B). The medians of the observed proteome ratios closely matched the theoretical ratios and exhibited tight coefficients of variation (CVs), even under high-throughput conditions. These results demonstrate the robustness and suitability of the workflow for high-throughput proteomic analysis (Figure S5).
To further confirm the quantitation accuracy of higher dynamic range, we subsequently performed a two-proteome mixture experiment where increasing amounts of yeast digest (from 20 ng to 160 ng) were spiked into a constant background of 200 ng of HeLa digest, resulting in a series of samples with various loading amounts (Figure 4A). By comparing the measured protein intensities for the yeast proteome across these samples, we observed a strong linear correlation between the signal and the amount of spiked-in yeast digest. The linear regression analysis yielded an R2 value of 0.999, underscoring the outstanding quantitative accuracy and linearity of the 300 SPD method even under different loading amounts (Figure 4B).
Figure 4. Excellent Quantitative Linearity of the Ultra-High-Throughput Workflow.

(A) Schematic diagram of the two-proteome mixture experiment, where 20, 40, 60, 80, 100, and 160 ng of yeast digest were spiked into a 200 ng HeLa digest background.
(B) A plot of the relative fold change of the yeast proteome versus the amount of yeast digest added. Boxes show the three quartile values of the distribution. The whiskers extend to points that lie within 1.5 interquartile ranges (IQRs) of the lower and upper quartile. The blue dots represent the average ratios relative to the 160-ng group. The data points show a clear linear trend, with a linear regression fit demonstrating an R2 of 0.999.
Application in TPD: Profiling Dose-Dependent Effects of CELMoDs
With full confidence in the developed high-throughput method, we applied it to a real-world drug screening scenario: studying the proteome-wide effects of cereblon E3 ligase modulators (CELMoDs). CELMoDs are a class of small-molecule degraders, often referred to as 'molecular glues,' that recruit the CRBN E3 ubiquitin ligase to specific neosubstrate proteins, leading to their subsequent proteasomal degradation. We treated human Jurkat T cells for 24 hours with four different concentrations (8, 40, 200, and 1000 nM) of three immunomodulatory drugs (IMiDs): pomalidomide, iberdomide, and mezigdomide (Figure 5A).
Figure 5. Dose-Dependent Protein Degradation Induced by Pomalidomide, Iberdomide, and Mezigdomide.

(A) Schematic diagram showing experimental design. Cells were treated with various concentrations of IMiDs (pomalidomide, iberdomide, and mezigdomide) for 24 hrs, followed by KingFisher-assisted SP3 magnetic beads sample preparation and LC-MS/MS analysis with 8-cm column and 300 SPD method.
(B) Bubble plot showing the dose-dependent protein-groups quantification by pomalidomide, iberdomide, and mezigdomide treatment in human Jurkat cells. The bubble color denotes the relative degradation magnitude, while bubble size reflects statistical significance based on adjusted p-values. Grey highlighting on the protein name indicates this is a G-loop containing protein 4.
(C) Volcano plots illustrating the differential protein level analysis in Jurkat cells treated with 1000 nM of pomalidomide, iberdomide, or mezigdomide compared to DMSO-treated controls. (2 biological replicates by 2 technical replicates)
The analysis revealed distinct, dose-dependent protein degradation patterns for each compound (Figure 5B). As expected, all three drugs induced the degradation of known cereblon neosubstrates like IKZF1 and ZFP91 (Figure 5B, C), with mezigdomide showing the highest potency at lower concentrations. Interestingly, our data revealed that mezigdomide also led to the significant degradation of additional proteins, most notably IKZF2. Mezigdomide was designed as an ultra-potent IKZF1/3 degrader, yet it exhibits a uniquely robust but unexplained immunomodulatory profile. Since IKZF2 is a critical transcription factor for maintaining regulatory T cell (Treg) suppressive functions, its degradation provides a highly novel and necessary mechanistic rationale for mezigdomide’s enhanced T cell costimulation and unique clinical efficacy 19. Our finding suggests that mezigdomide’s therapeutic efficacy may be significantly augmented by this broader substrate profile.
These findings demonstrate the power of our ultra-high-throughput workflow for chemoproteomics-based drug screening. Even at extreme throughput, the system retains the sensitivity necessary to identify nuanced, dose-dependent neosubstrate degradation and discover novel drug mechanisms. Detailed volcano plots for all tested concentrations are provided in Figure S6.
Discussion
In this study, we successfully developed and validated an ultra-high-throughput proteomics workflow capable of analyzing 300 samples per day using the Orbitrap Astral mass spectrometer. The method delivers deep proteome coverage, identifying nearly 8,000 protein groups from human cell lines with excellent quantitative precision, making it highly suitable for large-scale applications such as compound screening for targeted protein degradation. Our results from the TPD application underscore this, where we not only confirmed known targets of IMiDs but also identified novel dose-dependent degradation patterns, such as for IKZF2 by mezigdomide, providing new biological insights.
A key aspect of validating any quantitative workflow is understanding its limitations, particularly when pushing the boundaries of throughput. Our analysis of the extreme three-proteome mixture was designed specifically to challenge the method's performance across a wide dynamic range, covering the fold change from ~2.5% to ~95%. With the 300 SPD method, we observed some ratio compression, a phenomenon where the measured fold change is smaller than the true biological change. This effect was most pronounced for proteins undergoing large fold changes and was particularly evident for lower-intensity proteins. This is a well-documented challenge in shotgun proteomics especially in isobaric labeling quantification techniques like tandem mass tag (TMT), often attributed to factors like co-isolation of interfering precursor ions, especially in complex samples analyzed with very short chromatographic gradients 20. However, for DIA, this issue is substantially mitigated because quantification relies on multiple fragment ions. Modern DIA analysis software can effectively distinguish and quantify target peptides by matching their specific, high-quality fragment ion spectra from the Astral analyzer, even in the presence of co-eluting interferences at the precursor level. Nevertheless, another layer of complexity arises for low-abundance proteins, as their quantification often relies on a small number of identified peptides—sometimes only one or two. This can introduce a degree of stochasticity, making their quantification inherently less robust than that of more abundant proteins represented by multiple peptides.
The observation of reduced ratio compression at the peptide level (Figure S4) likely reflects a population-level statistical effect. Global peptide distributions are dominated by high-intensity peptides from abundant proteins, which exhibit high quantitative accuracy. In this aggregate view, the minor ratio compression occurring in low-intensity peptides is essentially masked by the high-fidelity signals of the broader peptide population. However, at the protein level, low-abundance proteins lack this redundancy and must be quantified using only these susceptible peptides, thereby making the compression more apparent. To further investigate the impact of peptide redundancy on quantitative accuracy, we analyzed the residual error of protein ratios relative to their peptide counts (Figure S7). As expected, proteins quantified via only one or two peptides showed significantly higher variance and ratio compression compared to proteins with high peptide redundancy.
Significantly, we demonstrated that this ratio compression could be effectively mitigated by adjusting the analytical method. By employing a longer gradient (60 SPD), we observed improvement in quantitative accuracy and a reduction in ratio compression for the same extreme proteome mixture sample. This highlights the flexibility and strength of the Orbitrap Astral MS platform. The ultra-fast 300 SPD method is ideal for primary screening, where the goal is to rapidly identify the most potent hits and significant changes. For subsequent validation or more detailed follow-up studies requiring higher quantitative fidelity for low-abundance targets, switching to a moderately longer gradient provides a straightforward solution without compromising the core benefits of the platform.
Furthermore, the exceptional sensitivity of the Astral analyzer is particularly powerful for resolving small fold changes. This capability is invaluable in early-stage drug discovery campaigns, where initial compound hits may only induce small, yet statistically significant, changes in protein abundance. The robustness of this approach was validated under stringent, blind-test conditions. The three-proteome mixture experiment was conducted as a blinded evaluation where the sample ratios, prepared by BCM, were unknown to Thermo Fisher Scientific scientists during data acquisition and analysis. The subsequent accurate determination of the proteome ratios against the ground truth demonstrated the platform's reliability and convinced us that it meets the highly demanding requirements for drug discovery workflows in chemoproteomics 21–23 and targeted protein degradation 1,2,4,24,25.
In conclusion, the Orbitrap Astral mass spectrometer enables a paradigm shift in high-throughput proteomics. The workflow presented here offers a powerful, scalable, and adaptable solution for drug discovery and other large-scale proteomics studies. While ratio compression of low-intensity proteins is a consideration in ultra-fast methods, we have shown that it is a manageable variable, allowing researchers to strike the optimal balance between throughput and quantitative depth to meet diverse experimental needs.
Supplementary Material
Supporting Information
The following supporting information is available free of charge at ACS website.
Figure S1. Quantitative matrixes of three-proteome mixtures at different sampling throughput.
Figure S2. Quantitative performance of proteome mixtures with 200 ng loading amount and 60 SPD method.
Figure S3. Accurate quantification of LFQ-DIA workflow on low loading sample (50 ng).
Figure S4. Peptide group quantification of LFQ-DIA workflow on high loading sample (200 ng).
Figure S5. Accurate quantification linearity of the 8 cm column LFQ-DIA workflow.
Figure S6. Volcano plots illustrating the differential protein level analysis in IMiDs-treated Jurkat cells.
Figure S7. Impact of Peptide Redundancy on Quantitative Accuracy.
Figure S8. Distribution of Protein and Peptide Quantification Coefficient of Variation (CV) across five extreme 3-proteome mixtures or TPD samples.
Figure S9. Protein and Peptide Identification Counts across Extreme Proteome Mixtures.
Supplementary Table S1. LCMS parameters for 24, 60, 180, 300 SPD.xlsx
Acknowledgement
The research was supported in part by the Michael E. DeBakey, M.D., Professorship in Pharmacology and the seed funding for Center for NextGen Therapeutics.
Footnotes
Competing Interest Statement
J.W. is the co-founder of Chemical Biology Probes LLC. J.W. has stock ownership in CoRegen Inc and serves as a consultant for this company. J.W. is a co-founder of Fortitude Biomedicines, Inc. and holds equity interest in this company. Y.Y., S.M., and Q.S. are employees of Thermo Fisher Scientific, Inc.
Medical Ethics Statement
This research did not involve human or animal participants.
Data and code availability
All the raw data from Orbitrap Astral has been deposited to MassIVE database under accession: MSV000099486.
All the codes for Astral DIA data analysis have been deposited at https://github.com/Hanfeng-Lin/Astral_DIA_analysis.
References
- (1).Steger M; Nishiguchi G; Wu Q; Schwalb B; Shashikadze B; McGowan K; Actis M; Aggarwal A; Shi Z; Price J; Mayasundari A; Yang L; Bednarz AH; Machata S; Graef T; Bartoschek D; Demichev V; Ohmayer U; Yang J; Daub H; Rankovic Z Unbiased Mapping of Cereblon Neosubstrate Landscape by High-Throughput Proteomics. Nat Commun 2025, 16 (1), 7773. 10.1038/s41467-025-62829-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (2).Baek K; Metivier RJ; Roy Burman SS; Bushman JW; Yoon H; Lumpkin RJ; Ryan JK; Abeja DM; Lakshminarayan M; Yue H; Ojeda S; Xiong Y; Che J; Verano AL; Schmoker AM; Gray NS; Donovan KA; Fischer ES Unveiling the Hidden Interactome of CRBN Molecular Glues. Nat Commun 2025, 16 (1), 6831. 10.1038/s41467-025-62099-w. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (3).Donovan KA; Ferguson FM; Bushman JW; Eleuteri NA; Bhunia D; Ryu S; Tan L; Shi K; Yue H; Liu X; Dobrovolsky D; Jiang B; Wang J; Hao M; You I; Teng M; Liang Y; Hatcher J; Li Z; Manz TD; Groendyke B; Hu W; Nam Y; Sengupta S; Cho H; Shin I; Agius MP; Ghobrial IM; Ma MW; Che J; Buhrlage SJ; Sim T; Gray NS; Fischer ES Mapping the Degradable Kinome Provides a Resource for Expedited Degrader Development. Cell 2020, 183 (6), 1714–1731.e10. 10.1016/j.cell.2020.10.038. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (4).Petzold G; Gainza P; Annunziato S; Lamberto I; Trenh P; McAllister LA; DeMarco B; Schwander L; Bunker RD; Zlotosch M; SriRamaratnam R; Gilberto S; Langousis G; Donckele EJ; Quan C; Strande V; De Donatis GM; Alabi SB; Alers J; Matysik M; Staehly C; Dubois A; Osmont A; Garskovas M; Lyon D; Wiedmer L; Oleinikovas V; Lieberherr R; Rubin NT; Lam DT; Lucas X; Liardo E; Widlund NI; Ritzén A; Caceres RM; Vigil D; Tsai J; Wallace O; Peluso M; Sadok A; Tiedt R; Paterson AM; Zarayskiy V; Fasching B; Bonenfant D; Warmuth M; Castle JC; Townson SA Mining the CRBN Target Space Redefines Rules for Molecular Glue–Induced Neosubstrate Recognition. Science 2025, 389 (6755), eadt6736. 10.1126/science.adt6736. [DOI] [PubMed] [Google Scholar]
- (5).Demichev V; Messner CB; Vernardis SI; Lilley KS; Ralser M DIA-NN: Neural Networks and Interference Correction Enable Deep Proteome Coverage in High Throughput. Nat Methods 2020, 17 (1), 41–44. 10.1038/s41592-019-0638-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (6).Demichev V; Szyrwiel L; Yu F; Teo GC; Rosenberger G; Niewienda A; Ludwig D; Decker J; Kaspar-Schoenefeld S; Lilley KS; Mülleder M; Nesvizhskii AI; Ralser M Dia-PASEF Data Analysis Using FragPipe and DIA-NN for Deep Proteomics of Low Sample Amounts. Nat Commun 2022, 13 (1), 3944. 10.1038/s41467-022-31492-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (7).Kistner F; Grossmann JL; Sinn LR; Demichev V QuantUMS: Uncertainty Minimisation Enables Confident Quantification in Proteomics. bioRxiv June 24, 2023, p 2023.06.20.545604. 10.1101/2023.06.20.545604. [DOI] [Google Scholar]
- (8).Kitata RB; Yang J-C; Chen Y-J Advances in Data-Independent Acquisition Mass Spectrometry towards Comprehensive Digital Proteome Landscape. Mass Spectrometry Reviews 2023, 42 (6), 2324–2348. 10.1002/mas.21781. [DOI] [PubMed] [Google Scholar]
- (9).Kohler D; Staniak M; Yu F; Nesvizhskii AI; Vitek O An MSstats Workflow for Detecting Differentially Abundant Proteins in Large-Scale Data-Independent Acquisition Mass Spectrometry Experiments with FragPipe Processing. Nat Protoc 2024, 19 (10), 2915–2938. 10.1038/s41596-024-01000-3. [DOI] [PubMed] [Google Scholar]
- (10).Cox J Prediction of Peptide Mass Spectral Libraries with Machine Learning. Nat Biotechnol 2023, 41 (1), 33–43. 10.1038/s41587-022-01424-w. [DOI] [PubMed] [Google Scholar]
- (11).Bruderer R; Bernhardt OM; Gandhi T; Xuan Y; Sondermann J; Schmidt M; Gomez-Varela D; Reiter L Optimization of Experimental Parameters in Data-Independent Mass Spectrometry Significantly Increases Depth and Reproducibility of Results. Molecular & Cellular Proteomics 2017, 16 (12), 2296–2309. 10.1074/mcp.RA117.000314. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (12).Ludwig C; Gillet L; Rosenberger G; Amon S; Collins BC; Aebersold R Data-independent Acquisition-based swath - ms for Quantitative Proteomics: A Tutorial. Molecular Systems Biology 2018, 14 (8), e8126. 10.15252/msb.20178126. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (13).Vowinckel J; Zelezniak A; Bruderer R; Mülleder M; Reiter L; Ralser M Cost-Effective Generation of Precise Label-Free Quantitative Proteomes in High-Throughput by microLC and Data-Independent Acquisition. Sci Rep 2018, 8 (1), 4346. 10.1038/s41598-018-22610-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (14).Hu A; Noble WS; Wolf-Yadlin A Technical Advances in Proteomics: New Developments in Data-Independent Acquisition. F1000Research March 31, 2016. 10.12688/f1000research.7042.1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (15).Heil LR; Damoc E; Arrey TN; Pashkova A; Denisov E; Petzoldt J; Peterson AC; Hsu C; Searle BC; Shulman N; Riffle M; Connolly B; MacLean BX; Remes PM; Senko MW; Stewart HI; Hock C; Makarov AA; Hermanson D; Zabrouskov V; Wu CC; MacCoss MJ Evaluating the Performance of the Astral Mass Analyzer for Quantitative Proteomics Using Data-Independent Acquisition. J. Proteome Res. 2023, 22 (10), 3290–3300. 10.1021/acs.jproteome.3c00357. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (16).Stewart HI; Grinfeld D; Giannakopulos A; Petzoldt J; Shanley T; Garland M; Denisov E; Peterson AC; Damoc E; Zeller M; Arrey TN; Pashkova A; Renuse S; Hakimi A; Kühn A; Biel M; Kreutzmann A; Hagedorn B; Colonius I; Schütz A; Stefes A; Dwivedi A; Mourad D; Hoek M; Reitemeier B; Cochems P; Kholomeev A; Ostermann R; Quiring G; Ochmann M; Möhring S; Wagner A; Petker A; Kanngiesser S; Wiedemeyer M; Balschun W; Hermanson D; Zabrouskov V; Makarov AA; Hock C Parallelized Acquisition of Orbitrap and Astral Analyzers Enables High-Throughput Quantitative Analysis. Anal. Chem. 2023, 95 (42), 15656–15664. 10.1021/acs.analchem.3c02856. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (17).Guzman UH; Martinez-Val A; Ye Z; Damoc E; Arrey TN; Pashkova A; Renuse S; Denisov E; Petzoldt J; Peterson AC; Harking F; Østergaard O; Rydbirk R; Aznar S; Stewart H; Xuan Y; Hermanson D; Horning S; Hock C; Makarov A; Zabrouskov V; Olsen JV Ultra-Fast Label-Free Quantification and Comprehensive Proteome Coverage with Narrow-Window Data-Independent Acquisition. Nat Biotechnol 2024, 42 (12), 1855–1866. 10.1038/s41587-023-02099-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (18).Lancaster NM; Sinitcyn P; Forny P; Peters-Clarke TM; Fecher C; Smith AJ; Shishkova E; Arrey TN; Pashkova A; Robinson ML; Arp N; Fan J; Hansen J; Galmozzi A; Serrano LR; Rojas J; Gasch AP; Westphall MS; Stewart H; Hock C; Damoc E; Pagliarini DJ; Zabrouskov V; Coon JJ Fast and Deep Phosphoproteome Analysis with the Orbitrap Astral Mass Spectrometer. Nat Commun 2024, 15 (1), 7016. 10.1038/s41467-024-51274-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (19).Bonazzi S; d’Hennezel E; Beckwith REJ; Xu L; Fazal A; Magracheva A; Ramesh R; Cernijenko A; Antonakos B; Bhang HC; Caro RG; Cobb JS; Ornelas E; Ma X; Wartchow CA; Clifton MC; Forseth RR; Fortnam BH; Lu H; Csibi A; Tullai J; Carbonneau S; Thomsen NM; Larrow J; Chie-Leon B; Hainzl D; Gu Y; Lu D; Meyer MJ; Alexander D; Kinyamu-Akunda J; Sabatos-Peyton CA; Dales NA; Zécri FJ; Jain RK; Shulok J; Wang YK; Briner K; Porter JA; Tallarico JA; Engelman JA; Dranoff G; Bradner JE; Visser M; Solomon JM Discovery and Characterization of a Selective IKZF2 Glue Degrader for Cancer Immunotherapy. Cell Chemical Biology 2023, 30 (3), 235–247.e12. 10.1016/j.chembiol.2023.02.005. [DOI] [PubMed] [Google Scholar]
- (20).Ting L; Rad R; Gygi SP; Haas W MS3 Eliminates Ratio Distortion in Isobaric Labeling-Based Multiplexed Quantitative Proteomics. Nat Methods 2011, 8 (11), 937–940. 10.1038/nmeth.1714. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (21).Lin H; Yang B; Ding L; Yang Y-Y; Holt MV; Jung SY; Zhang B; Wang MC; Wang J COOKIE-Pro: Covalent Inhibitor Binding Kinetics Profiling on the Proteome Scale. Nat Commun 2025, 16 (1), 8373. 10.1038/s41467-025-63491-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (22).Lucas SCC; Milbradt AG; Blackwell JH; Bonomo S; Brierley A; Cassar DJ; Freeman J; Hadfield TE; Morrill LA; Riemens R; Sarda S; Schiesser S; Wiktelius D; Ahmed S; Bostock MJ; Börjesson U; De Fusco C; Guerot C; Hargreaves D; Hewitt S; Lamb ML; Su N; Whatling R; Wheeler M; Kettle JG Design of a Lead-Like Cysteine-Targeting Covalent Library and the Identification of Hits to Cys55 of Bfl-1. J. Med. Chem. 2024. 10.1021/acs.jmedchem.4c00781. [DOI] [PubMed] [Google Scholar]
- (23).Boike L; Henning NJ; Nomura DK Advances in Covalent Drug Discovery. Nat Rev Drug Discov 2022, 21 (12), 881–898. 10.1038/s41573-022-00542-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (24).Guo W-H; Qi X; Yu X; Liu Y; Chung C-I; Bai F; Lin X; Lu D; Wang L; Chen J; Su LH; Nomie KJ; Li F; Wang MC; Shu X; Onuchic JN; Woyach JA; Wang ML; Wang J Enhancing Intracellular Accumulation and Target Engagement of PROTACs with Reversible Covalent Chemistry. Nat Commun 2020, 11 (1), 4268. 10.1038/s41467-020-17997-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (25).Yu X; Lu D; Qi X; Paudel RR; Lin H; Holloman BL; Jin F; Xu L; Ding L; Peng W; Wang MC; Chen X; Wang J Development of a RIPK1 Degrader to Enhance Antitumor Immunity. Nat Commun 2024, 15 (1), 10683. 10.1038/s41467-024-55006-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data