Abstract
Introduction
Multicenter radiomic studies are fundamentally constrained by inter-center feature variability arising from differences in scanner hardware, acquisition protocols, and reconstruction parameters. Existing harmonization methods, including ComBat, treat these sources of variability as undifferentiated batch noise and primarily correct feature distribution location, leaving scale-level heterogeneity unaddressed. This study proposes Category-ComBat with Distribution Coefficient (Category-ComBat-ω), a two-level harmonization framework for multicenter CT radiomics.
Methods
A total of 1,789 pulmonary nodule CT images were retrospectively collected from three public datasets (LIDC-IDRI, NSCLC-Radiomics, and Lung-PET-CT-Dx), with 93 radiomic features extracted per case. Category-ComBat reconstructs the ComBat framework by explicitly separating imaging-related variability into systematic components, including scanner type, acquisition protocol, and reconstruction kernel, and random components. Individualized least-squares estimation was used to estimate systematic biases, together with category-stratified mean normalization. A distribution coefficient ω ∈ (0, 1] was introduced as an orthogonal scale-adjustment mechanism to adaptively compress feature distribution spread without perturbing category-level means. Performance was evaluated using Wilcoxon tests, Q-Q plots, three-dimensional principal component analysis (PCA), probability density plots, coefficient of variation (COV), intraclass correlation coefficient (ICC), and downstream machine learning classification experiments.
Results
Category-ComBat effectively removed systematic inter-center variability while preserving benign-malignant discriminative separation. Introduction of ω progressively improved feature reproducibility and inter-center consistency as ω decreased. The best overall performance, evaluated using ICC, COV, and classification AUC, was observed for ω values between 0.5 and 0.9. Scale compression remained independent of location adjustment, confirming the orthogonality of the two-level design. Category-ComBat achieved the highest classification performance, with the mean AUC increasing from 0.84 for the Original features and 0.82 for ComBat to 0.89 after Category-ComBat harmonization.
Discussion
Category-ComBat-ω provides a principled approach to multicenter radiomic harmonization by addressing both systematic imaging-related confounding and scale heterogeneity. Its tunable distribution coefficient enables flexible adjustment of feature distribution spread while preserving category-level means, providing an adaptable framework for different multicenter radiomic study requirements.
Keywords: algorithm, category-ComBat, discriminative, multicenter harmonization, radiomics
Highlights
Category-ComBat decomposes imaging factors into systematic and random components for precise multicenter radiomic harmonization.
Removal of the systematic factor component improves distributional conformity and empirical Bayesian estimation accuracy.
A tunable distribution coefficient balances feature consistency with biological information preservation.
Downstream machine learning validation demonstrates preserved or improved classification performance after harmonization.
1. Introduction
Radiomics refers to the high-throughput extraction of quantitative imaging features from medical images—including morphological, intensity-based, and textural descriptors—and their systematic association with clinical endpoints such as pathological diagnosis, treatment response, and prognosis (1, 2). By converting routine radiological images into mineable data, radiomics offers the potential to augment clinical decision-making without additional invasive procedures (3, 4). Over the past decade, radiomic approaches have been applied across a wide spectrum of oncological contexts, including lung, breast, prostate, and hepatocellular carcinomas, demonstrating consistent associations with histopathological subtypes, genomic signatures, and survival outcomes (5–7).
The translational promise of radiomics, however, is fundamentally constrained by the sensitivity of radiomic features to acquisition and reconstruction conditions. Features extracted from CT images are known to vary substantially across scanner manufacturers, tube voltage and current settings, slice thickness, field of view, and reconstruction kernels (8, 9). When data are aggregated from multiple institutions—as is necessary for adequately powered multicenter studies—these technical sources of variability introduce systematic inter-center bias that is confounded with the biological signal of interest (10, 11). Without harmonization, cross-center radiomic models suffer from poor generalizability, inflated false discovery rates, and irreproducible findings (12, 13).
To address inter-center variability, several harmonization strategies have been proposed. Image-level standardization approaches, such as histogram normalization and intensity standardization, attempt to reduce acquisition-related differences prior to feature extraction (14). Feature-level methods operate downstream of extraction, applying statistical correction directly to the feature matrix. Among these, the ComBat algorithm—originally developed for genomic batch correction (15)—has been most widely adopted in radiomic harmonization (16, 17). ComBat employs an empirical Bayes framework to estimate and remove additive (location) and multiplicative (scale) batch effects under the assumption that batch effects are uniform across all observations within a center. Its application to radiomic features has demonstrated meaningful improvements in inter-scanner reproducibility as measured by Intraclass Correlation Coefficients (ICC) and Coefficient of Variation (COV) (11, 18).
Nevertheless, standard ComBat presents several limitations when applied to radiomic data. First, it treats the entire imaging center as a homogeneous batch, conflating qualitatively distinct sources of variability: systematic factors that are deterministically linked to scanner configuration and acquisition protocol, and random factors attributable to operator behavior, patient positioning, and physiological motion (19). This undifferentiated treatment prevents targeted correction of the dominant systematic components, risks over-correction of biologically meaningful variance, and may reduce the discriminative power of harmonized features for downstream classification or regression tasks (20). Second, standard ComBat corrects only the location of feature distributions—adjusting center-specific additive offsets—while leaving the scale structure of the harmonized distribution unchanged except through a global multiplicative term that is not sensitive to category-level distributional differences (21, 22).
These limitations motivate the development of Category-ComBat, an enhanced harmonization framework that restructures the ComBat model to explicitly separate systematic imaging-related variability from random variability within each center (23). By decomposing the batch effect into identifiable systematic components—scanner manufacturer, acquisition protocol, reconstruction kernel—and estimating their contributions via individualized least-squares within each imaging category, Category-ComBat achieves more precise removal of technically induced bias while preserving the category-level distributional structure required for classification tasks such as benign-malignant discrimination of pulmonary nodules. Mechanistically, the algorithm incorporates a systematic factor matrix and its corresponding regression vector into the data standardization process, enabling the model to explicitly fit and remove variations attributable to deterministic imaging-system factors. Separately, additive and multiplicative error terms are defined to capture and reduce the residual variability introduced by random factors such as operator behavior, patient positioning, and physiological motion. Validation against standard ComBat using Wilcoxon signed-rank tests, Q-Q plot analysis, and three-dimensional PCA demonstrated superior distributional alignment and better preservation of discriminative feature separation (24).
However, even after location-based harmonization by Category-ComBat, residual scale heterogeneity may persist across centers, reflecting differences in the dispersion of feature distributions that are not fully captured by additive or multiplicative correction of the mean (25). Feature distribution spread—as quantified by COV and the tails of inter-center density plots—has been shown to influence the stability and reproducibility of radiomic signatures independently of mean-level alignment (19, 26). This residual scale variability represents an orthogonal dimension of harmonization that the base Category-ComBat algorithm does not explicitly address.
To resolve this limitation, the present study introduces a distribution coefficient ω as a tunable scale-adjustment mechanism appended to the Category-ComBat framework, yielding the integrated algorithm Category-ComBat with Distribution Coefficient (Category-ComBat-ω). The coefficient ω ∈ (0, 1] modulates the spread of harmonized feature distributions in a continuous and category-preserving manner: values approaching 0 compress the distribution toward the category mean, progressively reducing inter-subject dispersion and improving reproducibility metrics, while ω = 1 recovers the unscaled Category-ComBat output. Critically, ω operates orthogonally to the location adjustment performed by Category-ComBat, ensuring that category-level mean normalization and scale compression are independently controllable.
The primary objectives of this study are: (1) to present the unified theoretical formulation of Category-ComBat-ω; (2) to empirically validate the algorithm on a multicenter pulmonary nodule CT dataset; (3) to characterize the effect of ω on feature reproducibility (COV), inter-center consistency (ICC), and distributional alignment across the full range ω ∈ (0, 1]; and (4) to provide practical guidance on ω selection for harmonization pipelines with different reproducibility-information tradeoff requirements. The algorithm is implemented as an open-source module compatible with standard radiomic workflows.
2. Materials and methods
2.1. Algorithm design
2.1.1. Traditional ComBat model
The ComBat algorithm assuming that radiomics feature data includes data from centers , sample and feature , The ComBat model is modeled as follows (Equation 1):
| (1) |
Where, Yijf represents the value of the radiomics feature, αf is the overall measurement value of the feature, X is the design matrix of the radiomics feature samples, and βf is the regression coefficient of the vector X. The ϵijf error term can be assumed to follow a standard normal distribution with an expected value of zero and a variance of σ2. γif and δif are the additive and multiplicative error terms of the multicenter effect, respectively.
Although ComBat has been widely applied to radiomic feature harmonization, it was designed for genetic data and carries fundamental limitations when applied to imaging studies. The covariate matrix X does not distinguish between biologically relevant covariates and imaging-system–related systematic factors, such as scanner type, acquisition parameters, and reconstruction protocols. Consequently, ComBat cannot separately model systematic imaging variability and random measurement noise, resulting in incomplete harmonization and potential over-correction of biologically meaningful variance.
2.1.2. Category-ComBat model
To overcome these limitations, we propose the Category-ComBat algorithm, which reconstructs the ComBat framework by (1) introducing category-level mean normalization to preserve biological variability across diagnostic classes, and (2) explicitly separating systematic imaging factors from random factors. As summarized in Table 1, factors influencing radiomic features are classified into two types:
Table 1.
Factors in radiomics features and their classifications.
| Factor category | Factor name | |
|---|---|---|
| Random factor | Electromagnetic interference, vibration, Air cleanliness environment Autocorrelation, Agency site, operator, Power fluctuations, Patient motion, motion artifacts | |
| Systematic factor | Difference in equipment | Different detector systems, scanner type, Coil Sensitivity, Positional and Physiological, Changes During, Acquisition, Magnetic field changes in MRI |
| Differences in protocol | Voltage, Tube current, Exposure time, Slice thickness, Voxel size, Reconstruction kernel, Reconstruction protocol, Field of view, Detector Configuration | |
Systematic factors: deterministic, reproducible sources of variability arising from imaging equipment and acquisition protocols, including scanner type, detector configuration, tube voltage, tube current, slice thickness, voxel size, reconstruction kernel, and convolution protocol.
Random factors: stochastic, unpredictable sources of variability including electromagnetic interference, operator performance, patient motion, motion artifacts, vibration, and environmental conditions.
As shown in Figure 1, The model comprises three components:
Figure 1.

The design rationale of the Category-ComBat algorithm model.
Component 1 — Category-level mean (αfc): Rather than estimating a single global mean αf as in ComBat, Category-ComBat estimates a category-specific mean αfc for each diagnostic class c (e.g., benign vs. malignant). This preserves the biological distributional differences between classes throughout harmonization.
Component 2 — Systematic factor correction (Xr, Xs, βr, βs): Xr is the systematic factor design matrix encoding per-case imaging acquisition conditions (manufacturer, scanner model, slice thickness, tube current, tube voltage, exposure time, and convolution kernel), with βr as its individualized least-squares regression vector. Xs is the biological covariate preservation matrix with regression vector βs. The operator hstack[·] denotes horizontal matrix concatenation, enabling Xsβs and Xrβr to be handled independently without introducing confounding between biological and technical effects.
Component 3 — Random factor correction (γif, δif): The additive error term γif and multiplicative error term δif jointly model the residual stochastic variability ϵijfc attributable to random factors. It has been established that this additive–multiplicative error structure effectively captures random-factor–induced variation across diverse imaging settings.
2.1.3. Category-ComBat parameter estimation
The parameters , and were estimated using moment estimation and least squares. Normalized sample data Zijfc can be calculated by (Equation 2):
| (2) |
The standardized data Zijfc conforms to a normal distribution of . The parameter φif conforms to a normal distribution of , and the parameter conforms to an inverse gamma distribution of Inverse Gamma .
2.1.4. Category-ComBat multicenter effect adjustment
Based on the distribution assumptions described above, the empirical Bayesian parameter estimation method is employed to estimate the posterior distribution parameters and for Zijfc.
“Category Parameter” ω was introduced as adjustment parameter for scaling the sample data in the final data adjustment formula (Equation 3). The purpose of ω is to enhance the reproducibility of features.
| (3) |
The “Category Parameter” ω is a coefficient between 0 and 1. It proportionally compresses the distribution of the harmonized data with the center of the category position center αfc. As a result, it can preserve the reproducibility of the features without compromising their discriminative power.
2.2. Datasets
Three publicly available lung CT datasets from The Cancer Imaging Archive (TCIA; https://www.cancerimagingarchive.net/) were included in this study. No ethical concerns apply as all datasets are de-identified and publicly released.
2.2.1. Dataset 1 — LIDC-IDRI
This dataset includes 1,018 patients with benign or malignant lung nodules. All CT images are stored in DICOM format; nodule segmentation data are stored in XML format. Four experienced thoracic radiologists independently annotated each nodule in a blinded fashion, followed by cross-review of one another’s assessments. Malignancy likelihood was rated using a five-level scale: Highly Unlikely, Moderately Unlikely, Indeterminate, Moderately Suspicious, and Highly Suspicious.
2.2.2. Dataset 2 — NSCLC-radiomics
This dataset comprises 422 patients with pathologically confirmed non-small cell lung cancer (NSCLC); mean age 67.5 years (range 33–91), with 132 females and 290 males. Tumor segmentation was manually delineated by expert radiation oncologists based on fused PET-CT images (lung window: W1700/L−300; mediastinal window: W600/L40; PET window: W30000/L15000) and stored in RTSTRUCT format. Gross tumor volume (GTV) was defined from FDG PET-CT data.
2.2.3. Dataset 3 — lung-PET-CT-Dx
This dataset contains 355 cases of lung cancer (both small cell and non-small cell) confirmed by histology and cytology. Each case includes CT, PET, and fused PET-CT volumes. Lesion segmentation was performed using rectangular bounding boxes by experienced radiologists; one radiologist provided initial annotations, which were subsequently reviewed by four others.
Each dataset was treated as an independent imaging center, yielding a three-center study design.
2.3. Data preprocessing
As show in Figure 2, A total of 1,789 pulmonary nodule CT cases were initially collected from three publicly available datasets, including LIDC-IDRI (n = 1012), NSCLC-Radiomics (n = 422), and Lung-PET-CT-Dx (n = 355).
Figure 2.

Illustrates the complete inclusion and exclusion process for the study population.
To ensure complete image acquisition information for the proposed Category-ComBat harmonization algorithm, seven acquisition parameters were extracted from the DICOM headers of each case, including manufacturer, manufacturer model, slice thickness, tube current, tube voltage, exposure time, and convolution kernel. Cases with missing information for any of these parameters were excluded, resulting in the removal of 568 cases. Subsequently, 75 cases were excluded because of incomplete or inaccurate ROI delineation or image format inconsistencies.
For the LIDC-IDRI dataset, pulmonary nodules rated as Highly Unlikely or Moderately Unlikely were classified as benign, whereas nodules rated as Moderately Suspicious or Highly Suspicious were classified as malignant. Nodules with an Indeterminate malignancy rating were excluded. For patients presenting with multiple nodules, only one representative nodule was retained to avoid repeated sampling. Consequently, 337 LIDC-IDRI cases were excluded during this step.
As show in Table 2, After applying all inclusion and exclusion criteria, a total of 789 CT cases were included in this study, comprising 223 benign and 566 malignant pulmonary nodules. Specifically, the final dataset consisted of 223 benign and 150 malignant cases from LIDC-IDRI, 213 malignant cases from NSCLC-Radiomics, and 203 malignant cases from Lung-PET-CT-Dx.
Table 2.
Baseline characteristics of the included datasets.
| Characteristic | LIDC-IDRI | NSCLC-Radiomics | Lung-PET-CT-Dx |
|---|---|---|---|
| Cases included in this study | 373 | 213 | 203 |
| Benign | 223 | 0 | 0 |
| Malignant | 150 | 213 | 203 |
| Scanner manufacturers | GE Medical Systems, Siemens, Toshiba | Siemens | GE Medical Systems, Philips, Siemens |
| Slice thickness (mm) | 0.6–4.0 | N/A* | N/A* |
| Tube voltage (kVp) | 120–140 | 120–140 | 80–140 |
| Tube current (mA) | 40–581 | 50–461 | 100–454 |
* N/A indicates that the slice thickness information was not available in the corresponding dataset.
As show is Figure 3, The radiomics features were extracted from the region of interest using the open-source Python program Pyradiomics 3.0.1. Total of 93 features were extracted. These features were grouped based on their matrix type, including First orders, Glcm (Gray-Level Co-occurrence Matrix), Gldm (Gray-Level Dependence Matrix), Glrlm (Gray-Level Run Length Matrix), Glszm (Gray-Level Size Zone Matrix), and Ngtdm (Neighborhood Gray-Tone Difference Matrix). The extraction process complied with the IBSI Standardization Initiative (25).
Figure 3.

Flowchart of multicenter harmonization algorithm evaluation.
All CT image voxel values were linearly normalized to the range [0, 1]. Isotropic voxel reconstruction was applied with voxel spacing resampled to a uniform 1:1:1 ratio. Gray-level intensities were subsequently discretized into 25 distinct intensity bins. Nodule segmentations were derived directly from each dataset’s provided annotations; all data were converted to NIfTI (NII) format for downstream processing. All shape-related features were excluded from the Lung-PET-CT-Dx dataset, as all ROIs in that dataset are represented as rectangular bounding boxes.
2.4. Machine learning model building
The Least Absolute Shrinkage and Selection Operator (LASSO) was utilized to identify the most diagnostically relevant radiomics features. Subsequently, various machine learning models (logistic regression, random forest, support vector machine, and neural networks) were constructed to classify benign and malignant nodules. All implementations were conducted using Python 3.7 and Scikit-Learn 1.0.2.
To strictly prevent data leakage during multicenter harmonization, the dataset was partitioned into a 70% training set and a 30% independent test set prior to any harmonization steps. The Category-ComBat parameters (including the empirical Bayes estimates and category location centers) were fitted exclusively on the training set. The test set was subsequently transformed using these pre-estimated training parameters. For baseline comparisons, models were also trained and evaluated on pre-harmonized (raw) data and traditional ComBat-harmonized data.
Regarding model hyperparameters:
Logistic Regression: L1 regularization, ‘lbfgs’ optimizer, optimization tolerance of 0.01.
Random Forest: 100 trees, ‘Gini’ impurity criterion, minimum samples split of 2, maximum depth of 8.
SVM: RBF kernel, polynomial degree 3, regularization parameter (C) of 2, and ‘scale’ gamma.
Neural Network (MLP): Stochastic Gradient Descent (SGD) solver, L2 regularization, 5 hidden layers with 20 neurons each, learning rate of 0.01, and a maximum of 300 iterations.
2.5. Statistical analysis
All statistical analyses were conducted in Python 3.7 and R 4.1.3 (RStudio). The harmonization performance of Category-ComBat was compared with that of the traditional ComBat algorithm. Inter-center reproducibility of radiomic features before and after harmonization was evaluated using the Intraclass Correlation Coefficient (ICC) based on a two-way random-effects model with absolute agreement (ICC(2,1)). This model was selected because both imaging centers and subjects were considered random effects, making it appropriate for assessing reproducibility across multiple centers. Coefficient of Variation (COV) was calculated to evaluate the relative variability of each feature across centers:
where SD is the standard deviation and is the mean across centers.
Principal Component Analysis (PCA) was applied in three dimensions to visually assess the degree of distributional separation and overlap between centers before and after harmonization. Q-Q plots were used to evaluate the goodness-of-fit of the harmonized feature distributions to a theoretical normal distribution. Probability density distributions were plotted to visualize the effect of each ω value (0.8, 0.6, 0.4, 0.2) on the scale of feature distributions, enabling selection of the optimal distribution coefficient. Wilcoxon signed-rank tests were conducted to compare radiomic feature distributions between benign and malignant categories before and after harmonization. To control for multiple comparisons across the 93 radiomic features, P values from the Friedman tests performed after harmonization were adjusted using the Benjamini–Hochberg false discovery rate (FDR) procedure, with an adjusted P< 0.05 considered statistically significant. Model performance was quantified using accuracy, precision, recall, F1 score, and the Area Under the Receiver Operating Characteristic Curve (AUC).
3. Results
3.1. Friedman test
Friedman’s test results are summarized in Table 3. Prior to harmonization, 93 radiomic features exhibited statistically significant between-scanner differences (P< 0.05). Following Category-ComBat harmonization, significant differences were no longer detected in 88 of 93 features (FDR-adjusted P > 0.05), demonstrating that Category-ComBat effectively reduces multicenter effects in radiomic signatures.
Table 3.
Harmonization test for Friedman’s test between different scanners.
| Features names | Without category-combat | With category-combat (FDR P) |
|---|---|---|
| Firstorder Minimum | <0.05 | 0.0008 |
| Glcm Autocorrelation | <0.05 | 0.0214 |
| Glcm Cluster Prominence | <0.05 | 0.0187 |
| Glcm Cluster Shade | <0.05 | 0.0385 |
| Gldm High Gray Level Emphasis | <0.05 | 0.0442 |
| Others 88 Radiomics features | <0.05 | >0.05 |
Note: Bold text indicates the remaining 88 radiomics features not individually listed in the table.
3.2. Boxplot analysis
Figure 4 presents boxplots of representative GLCM features extracted from the LIDC-IDRI dataset acquired using GE Medical Systems and Siemens CT scanners. The original data distributions showed statistically significant between-category differences (P< 0.05). After ComBat harmonization, the position centroids of the two categories were aligned (P > 0.05), but this alignment simultaneously weakened the discriminative power of the features. Category-ComBat, by contrast, harmonized the between-scanner distributional differences while preserving the statistically significant between-category differences (P< 0.05). Additionally, the scatter distributions within boxplots indicate that Category-ComBat drives the feature distributions toward normality.
Figure 4.

Box plots of the distribution of Glcm Imc1 features extracted from nodules of two categories after harmonization, and the differences were tested using the Wilcoxon.
3.3. PCA analysis
Figure 5 presents PCA results on the three clinical datasets (LIDC-IDRI, NSCLC-Radiomics, and Lung-PET-CT-Dx). In the first row, the original feature distributions show that the three datasets cluster in clearly separated regions with differing dispersions, confirming substantial multicenter effects that preclude direct integration of unharmonized radiomic features.
Figure 5.

PCA plots of radiomics features of three different datasets and two categories of pulmonary nodules (benign and malignant) harmonized by ComBat and Category-ComBat.
After traditional ComBat harmonization, the cluster positions and dispersions of the three datasets became aligned, confirming ComBat’s harmonization capacity. However, this alignment risks removing clinically meaningful distributional differences between datasets with different patient compositions.
After Category-ComBat harmonization, features from the three datasets were aligned in the second principal component dimension with coordinated dispersion, while differences in cluster positions along the first principal component were preserved. Notably, LIDC-IDRI—which contains both benign and malignant nodules—maintained a distinct distributional cluster position relative to NSCLC-Radiomics and Lung-PET-CT-Dx, which contain exclusively malignant cases. This result demonstrates that Category-ComBat harmonizes technical between-center variability while retaining clinically relevant biological distributional differences.
In the second row, PCA plots stratified by nodule category confirm that the traditional ComBat algorithm erroneously eliminated radiomic feature differences between benign and malignant nodules after harmonization, whereas Category-ComBat preserved robust distributional separation between the two categories across all three datasets.
3.4. Coefficient of variation analysis
As show in Figure 6, to evaluate the harmonization performance of Category-ComBat in terms of feature consistency, COV was computed across all 93 radiomic features before and after harmonization. As shown in Figure 6 (left), the mean COV of Original_Data was 0.72, which decreased to 0.58 following traditional ComBat harmonization and further decreased to 0.51 after Category-ComBat harmonization. Statistical comparison revealed no significant difference between Original_Data and ComBat (p = 0.14), whereas the difference between Original_Data and Category-ComBat reached statistical significance (p = 0.00023), indicating that Category-ComBat achieves a more substantial reduction in inter-center feature variability than traditional ComBat.
Figure 6.

(A) the mean coefficient of variation (COV) of 93 radiomics features after applying ComBat and Category-ComBat harmonization methods, and the group differences were tested using the Wilcoxon test. (B) The number of radiomics features with COV values in different ranges after harmonization.
To further investigate the effect of the distribution coefficient ω on COV, we analyzed the number of features exhibiting a decreasing or increasing COV trend across five ω values (ω in {0.1, 0.3, 0.5, 0.7, 1.0}). As shown in Figure 6 (right), the number of features with decreased COV ranged from 84 (ω = 1.0) to 90 (ω = 0.1), while the number of features with increased COV ranged from 3 (ω = 0.1) to 9 (ω = 1.0). These results demonstrate that smaller values of ω yield greater improvements in feature consistency; however, given the risk of over-compression and biological information loss at extremely low ω values, a range of 0.5 ≤ ω ≤ 0.9 is recommended as a practical balance between harmonization effectiveness and the preservation of discriminative feature information. As shown in Supplementary Figure S2, the impact of different distribution coefficient (ω) values on feature reproducibility was investigated.
3.5. Intraclass correlation coefficient analysis
Figure 7A presents the distribution of ICC values across all 93 radiomic features before and after harmonization. Following Category-ComBat harmonization, ICC values increased significantly compared to the original unharmonized data (P< 0.05), indicating that Category-ComBat substantially enhances between-center feature consistency by removing systematic factor variability. The traditional ComBat algorithm, by contrast, produced minimal improvement in ICC values. The effects of different distribution coefficient (ω) values on the mean ICC are shown in Supplementary Figure S1.
Figure 7.

(A) Mean intraclass correlation coefficient (ICC(2,1), 95% confidence interval) of 93 radiomic features after ComBat and Category-ComBat harmonization. Group differences were assessed using the Wilcoxon signed-rank test. (B) The number of radiomics features with ICC(2,1) values in different ranges after harmonization.
Figure 7B categorizes features by ICC range (excellent: 0.90–1.00; good: 0.75–0.90; acceptable: 0.40–0.75; poor: 0.00–0.40). After Category-ComBat harmonization, the number of features with excellent ICC increased from 33 to 50. Conversely, after ComBat harmonization, the number of excellent-ICC features decreased to 15. These results confirm that Category-ComBat’s explicit removal of systematic factor variability yields superior feature consistency relative to traditional ComBat harmonization.
3.6. Distribution assumption validation via Q-Q plots
As shown in formula 3, we assume that the standardized data Zijfc follows a normal distribution, . The mean φ of the standardized data Zijfc is assumed to follow a normal distribution, , and the variance δ2 is assumed to follow an inverse gamma distribution, . A Quantile-Quantile (Q-Q) plot is an important tool for assessing the conformity between the actual data distribution and its assumed theoretical distribution. The closer the distribution of actual data distribution is to the theoretical distribution, the more accurate the estimation of the posterior distribution parameters of the model.
As shown in Figure 8, the distribution of the standardized data Zijfc becomes closer to a normal distribution after removing the systematic factor component. This indicates that by removing components of variability attributed to systematic factors, the assumed distribution for standardizing data in the Category-ComBat algorithm aligns more closely with the actual data distribution. The systematic factor component represents specific factors influencing the consistency of CT image features. The traditional ComBat algorithm is unable to address this bias, resulting in a deficiency in removing the multicenter effects of radiomics features. To more accurately remove the multicenter effect, it is important to remove the affected portion of the systematic factor component.
Figure 8.

The QQ plot of the normalized data Zijfc and φ and δ2 prior distribution after ComBat and Category-ComBat harmonization.
The parameter assumption φ represents the mean of the standardized data Zijfc, and compared to the traditional ComBat algorithm, the φ estimated by the Category-ComBat algorithm is closer to a normal distribution. The parameter δ2 represents the variance of the standardized data Zijfc, and the δ2 estimated by the Category-ComBat algorithm is closer to an inverse gamma distribution.
Therefore, after removing the systematic factor component, the aforementioned distribution assumptions become more accurate. Through the process of empirical Bayesian parameter estimation, the posterior parameters and of the harmonized feature data are more precise. Overall, after removing the distribution of the systematic factor component, the model fits the real data distribution more closely. The Category-ComBat algorithm can better coordinate the multicenter effects in radiomic features caused by complex image conditions.
3.7. Machine learning classification evaluation
Table 4 presents the predictive performance of four distinct machine learning algorithms (Logistic Regression, Random Forest, SVM, and Neural Network) evaluated on the independent 30% hold-out test set. Models were trained and tested using three data configurations: original unharmonized data, traditional ComBat-harmonized data, and Category-ComBat-harmonized data.
Table 4.
Classification performance of four different machine learning models after ComBat coordination and Category-ComBat coordination.
| Model | Coordination method | Accuracy | Precision | Recall | F1 | AUC |
|---|---|---|---|---|---|---|
| Logistic regression | Original data | 0.88 (0.81-0.94) |
0.88 (0.83-0.92) |
0.92 (0.88-0.96) |
0.92 (0.88-0.96) |
0.89 (0.88-0.95) |
| ComBat | 0.72 (0.64-0.80) |
0.76 (0.71-0.82) |
0.92 (0.87-0.97) |
0.89 (0.81-0.97) |
0.75 (0.60-0.91) |
|
| Category-ComBat |
0.92
(0.88-0.96) |
0.97
(0.94-0.99) |
0.93
(0.87-0.98) |
0.94
(0.91-0.97) |
0.94
(0.91-0.97) |
|
| Random Forest |
Original data | 0.86 (0.82-0.91) |
0.91 (0.87-0.96) |
0.90 (0.84-0.96) |
0.90 (0.87-0.93) |
0.88 (0.77-0.95) |
| ComBat | 0.84 (0.78-0.90) |
0.89 (0.82-0.96) |
0.94 (0.84-0.96) |
0.89 (0.85-0.93) |
0.86 (0.77-0.94) |
|
| Category-ComBat |
0.97
(0.95-0.99) |
0.95
(0.93-0.97) |
0.98
(0.96-0.99) |
0.98
(0.96-0.99) |
0.97
(0.94-0.99) |
|
| SVM | Original data | 0.82 (0.87-0.88) |
0.85 (0.79-0.90) |
0.91 (0.85-0.97) |
0.93 (0.89-0.97) |
0.84 (0.78-0.95) |
| ComBat | 0.81 (0.74-0.88) |
0.83 (0.78-0.87) |
0.94 (0.89-0.99) |
0.88 (0.84-0.92) |
0.81 (0.70-0.93) |
|
| Category-ComBat |
0.97
(0.94-1.00) |
0.97
(0.95-1.00) |
0.98
(0.93-1.00) |
0.98
(0.90-1.00) |
0.96
(0.95-0.97) |
|
| Neural Network |
Original data | 0.83 (0.79-0.88) |
0.84 (0.81-0.88) |
0.94 (0.89-0.98) |
0.89 (0.86-0.92) |
0.85 (0.80-0.93) |
| ComBat | 0.71 (0.64-0.77) |
0.74 (0.70-0.78) |
0.81 (0.75-0.87) |
0.91 (0.87-0.96) |
0.75 (0.53-0.78) |
|
| Category-ComBat |
0.84
(0.79-0.89) |
0.84
(0.79-0.89) |
0.97
(0.94-1.00) |
0.90
(0.87-0.93) |
0.88
(0.78-0.97) |
Bold values indicate the best performance.
Notably, models trained on traditional ComBat-harmonized data generally exhibited a decrease in the Area Under the Curve (AUC) and other metrics compared to the original data. This confirms our hypothesis that traditional global harmonization inadvertently removes disease-specific discriminative signals alongside batch effects, thereby degrading classifier performance.
Conversely, the implementation of Category-ComBat substantially enhanced the predictive capacity across all machine learning architectures. By effectively eliminating scanner-induced noise while safeguarding diagnostic-specific feature locations, Category-ComBat provided the models with high-fidelity, generalizable signals. The Random Forest classifier achieved the most superior performance, yielding an AUC of 0.97 (95% CI: 0.94-0.99), an Accuracy of 0.97 (95% CI: 0.95-0.99), and an F1-score of 0.98 (95% CI: 0.96-0.99) on the independent test set.
The ROC curves Figure 9 further corroborate these findings, visibly demonstrating that Category-ComBat systematically out-performs both the unharmonized and traditional ComBat approaches. These results definitively validate the superiority of the Category-ComBat algorithm in robustly resolving multicenter batch effects while maximizing the clinical utility of radiomic biomarkers.
Figure 9.

ROC curves of four different machine learning models after ComBat coordination and Category-ComBat coordination.
4. Discussion
The reproducibility, discriminative power, and cross-center consistency of radiomic features are critical determinants of the robustness and generalizability of radiomics models. Recent studies have also emphasized that improving radiomic feature robustness is essential for reliable multicenter radiomics analysis, as imaging and reconstruction variability can substantially influence feature stability and downstream model performance (27). Recent studies have demonstrated that harmonization methods, particularly ComBat, can substantially improve the robustness of multicenter radiomic analysis and enhance downstream machine learning performance. Our proposed Category-ComBat further extends this concept by incorporating biological category information into the harmonization process, aiming to preserve discriminative biological variability while reducing technical inter-center effects (28).
Category-ComBat algorithm extends the traditional ComBat algorithm by introducing two key innovations: (1) the categorization of imaging condition factors into systematic and random components, enabling personalized and targeted correction of their respective effects on radiomic features; and (2) a feature distribution coefficient (ω) that adjusts the scale of feature distributions to further improve reproducibility and consistency while preserving discriminative information.
The traditional ComBat algorithm was originally developed to address batch effects in genomic data (15). In the field of medical imaging, however, the sources of inter-center variability are considerably more complex, encompassing acquisition parameters (scanner model, tube voltage, tube current, scanning protocol), reconstruction parameters (reconstruction kernel, reconstruction field of view), and environmental factors (vibration, motion artifacts, operator variation) (29). These heterogeneous imaging conditions represent a major obstacle to the effective multicenter harmonization of radiomic features. In contrast to traditional ComBat algorithms, including traditional ComBat, B-ComBat algorithm (30), M-ComBat algorithm (31), ComBat-seq algorithm (32) and other ComBat-derived algorithm (33) — Category-ComBat explicitly partitions these sources of variability. Acquisition and reconstruction parameters that exert systematic, reproducible effects on features are classified as systematic factors, while environmental disturbances and operator-related variability that introduce stochastic variation are classified as random factors.
To mitigate these adverse effects, Category-ComBat introduces a systematic factor matrix Xr, which captures the personalized imaging conditions for each case during feature extraction. By estimating the regression vector via the matrix Xr, the algorithm removes the systematic factor component Xrβr in a case-specific manner, enabling more accurate elimination of deterministic inter-center bias. For random factors, additive and multiplicative error terms model their stochastic effects, an approach that has been demonstrated to effectively capture random-factor-induced variation (15). This integrated design robustly addresses the full spectrum of imaging condition variability, improving both the modeling accuracy of the algorithm and the consistency and robustness of the harmonized features.
We collected a total of 1789 lung nodule cases from three publicly available datasets — LIDC-IDRI, NSCLC-Radiomics, and Lung-PET-CT-Dx — encompassing both benign and malignant nodules across three imaging centers. Three-dimensional PCA visualization confirmed that multicenter effects in radiomic features were effectively removed by Category-ComBat while discriminative differences between categories were preserved, a property not consistently achieved by the traditional ComBat algorithm, which often compromises feature discriminability during harmonization (34). Intraclass correlation coefficient (ICC) analysis further demonstrated that after categorizing and removing the systematic factor component, Category-ComBat significantly reduced inter-center variability in radiomic features, enhancing both their consistency and robustness.
This study used the reference segmentations provided by the original public datasets and did not perform additional manual delineation. Therefore, inter- and intra-observer segmentation reproducibility was not evaluated, which should be considered a limitation of the present study. Future work using repeated manual segmentations will allow assessment of the combined effects of segmentation variability and feature harmonization.
Similar to the approach adopted by Johnson et al. (15). to evaluate the fit of the traditional ComBat model using Q-Q plots, we employed Q-Q plots to assess the influence of the systematic factor component Xrβr on the distributional assumptions underlying Category-ComBat. The results demonstrated that the empirical distribution of the standardized data Zijfc aligned substantially more closely with the assumed normal distribution after removing the systematic factor component. Correspondingly, the prior mean parameter φ exhibited better agreement with its assumed normal prior, and the prior variance parameter δ2 more closely conformed to its assumed inverse gamma prior. These improvements validate the distributional assumptions of the model and provide a more accurate basis for empirical Bayesian parameter estimation, ultimately yielding more precise posterior parameters and and enhancing the overall harmonization precision of the algorithm.
To further characterize the role of the systematic factor component, we performed Wilcoxon tests on the Xrβr values of 93 radiomic features and analyzed the resulting p-value distributions between benign and malignant lung nodule categories. A p-value distribution skewed toward zero indicates that the apparent discriminative ability of certain features is not intrinsic to the biological signal, but rather a spurious consequence of systematic imaging variability. Such detrimental variation undermines feature robustness and leads to features that appear effective within a single center but fail to generalize across institutions — a critical limitation that Category-ComBat is specifically designed to address.
Beyond systematic factor removal, this study introduces the distribution coefficient ω as an additional tuning parameter to further optimize harmonization performance. The coefficient ω adjusts the scale of feature distributions after harmonization: when ω = 1, the feature distribution range is preserved without compression; as ω decreases below 1, the feature distributions are proportionally compressed. Analysis of coefficient of variation (COV) and ICC values across 93 features demonstrated that Category-ComBat at ω = 1 already slightly outperforms traditional ComBat, and that further reductions in ω progressively enhance feature reproducibility and consistency. Nevertheless, ω functions as a double-edged sword: excessively low values compress feature distributions to a degree that may destroy biologically meaningful variation, impairing the discriminative capacity of the radiomic signature and potentially degrading the classification performance of downstream radiomics models. Based on the experimental results of the datasets evaluated in this study, a weighting coefficient in the range of 0.5 ≤ ω ≤ 0.9 provided a favorable balance between harmonization and preservation of biological variability. This recommendation is empirical and may depend on dataset characteristics such as scanner heterogeneity, sample composition, and imaging protocols. Further validation on additional multicenter datasets is warranted before generalizing this recommendation.
Unlike PCA visualization and Wilcoxon analysis, downstream machine learning performance provides direct evidence of discriminative information preservation. The consistently improved AUC observed after Category-ComBat further supports that the proposed harmonization preserves biologically relevant information while reducing technical variability.
Category-ComBat has the potential to facilitate the integration of medical images from different institutions into a harmonized, consistent multicenter radiomic feature repository, providing researchers with more robust disease biomarkers and a more reliable foundation for clinical radiomics applications. However, several limitations should be acknowledged. First, the categorization of imaging condition factors into systematic and random components warrants more detailed and comprehensive validation across diverse imaging protocols and scanner types. Second, additional independent datasets and experimental conditions are needed to further validate the generalizability and effectiveness of the algorithm. Third, the impact of Category-ComBat harmonization on downstream machine learning model performance — including classification accuracy, AUC, and feature selection stability — requires systematic investigation.
Although the present study employed three publicly available datasets to maximize imaging diversity, these datasets differ in disease composition, lesion annotation strategy, segmentation methodology, and imaging workflow. Consequently, biological heterogeneity cannot be completely separated from technical imaging variability. Therefore, the current study should be interpreted as a proof-of-concept evaluation of Category-ComBat under heterogeneous real-world radiomics datasets rather than a definitive validation using clinically matched multicenter cohorts. Future studies will evaluate the proposed algorithm on prospectively collected multicenter datasets with identical clinical tasks, standardized segmentation protocols, and harmonized acquisition criteria to further isolate scanner-related variability.
Despite these limitations, Category-ComBat demonstrates clear advantages in improving the consistency and robustness of multicenter radiomic features, and future work aimed at addressing these limitations will further enhance its practical reliability and clinical applicability.
5. Conclusions
In this study, we proposed Category-ComBat, a multicenter radiomic feature harmonization algorithm that extends traditional ComBat by explicitly decomposing imaging condition factors into systematic and random components. By introducing the factor matrix Xr to model case-specific systematic effects, the algorithm enables targeted removal of the systematic factor component , improving the conformity of standardized data Zijfc to the distributional assumptions of the model and yielding more precise empirical Bayesian posterior parameter estimates and . Furthermore, the introduction of the distribution coefficient ω provides a tunable mechanism to enhance feature reproducibility and consistency, with a recommended range of for multicenter pulmonary nodule radiomics studies. Experimental results demonstrated that Category-ComBat effectively reduced inter-center variability while preserving discriminative information, outperforming traditional ComBat in terms of ICC and COV across 93 radiomic features. These findings suggest that Category-ComBat offers a robust and flexible solution for multicenter radiomic harmonization, with broad potential for application in large-scale clinical radiomics research. Furthermore, downstream machine learning evaluation demonstrated that Category-ComBat maintained or improved benign–malignant classification performance after harmonization, providing direct evidence that the proposed framework preserves biologically meaningful discriminative information while reducing technical imaging variability.
Acknowledgments
The authors thank all participants and colleagues who contributed to this study.
Funding Statement
The author(s) declared that financial support was received for this work and/or its publication. The study was supported in by Chengde Medical College Project (202423,202311), Hebei Provincial Department of Education and is part of the Hebei Provincial Higher Education Scientific Research Program (QN2026704). Supported by Hebei Natural Science Foundation (H2026406068).
Footnotes
Edited by: Michele Avanzo, Aviano Oncology Reference Center (IRCCS), Italy
Reviewed by: Haitham Al-Mubarak, University of Glasgow, United Kingdom
Pooja Dwivedi, Advanced Centre for Treatment, Research and Education in Cancer (ACTREC), India
Data availability statement
Publicly available datasets were analyzed in this study. This data can be found here: cancerimagingarchive.net.
Ethics statement
This study utilizes exclusively publicly available datasets and does not involve any restricted or sensitive content. The studies were conducted in accordance with the local legislation and institutional requirements. Written informed consent for participation was not required from the participants or the participants’ legal guardians/next of kin because this study utilizes exclusively publicly available datasets and does not involve any restricted or sensitive content.
Author contributions
XZ: Conceptualization, Methodology, Validation, Writing – original draft, Writing – review & editing. XW: Data curation, Resources, Validation, Writing – review & editing. WW: Data curation, Writing – review & editing. LL: Supervision, Writing – review & editing. MM: Supervision, Writing – review & editing. XD: Conceptualization, Data curation, Supervision, Writing – review & editing. YL: Conceptualization, Project administration, Supervision, Writing – review & editing.
Conflict of interest
The author(s) declared that this work was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Generative AI statement
The author(s) declared that generative AI was not used in the creation of this manuscript.
Any alternative text (alt text) provided alongside figures in this article has been generated by Frontiers with the support of artificial intelligence and reasonable efforts have been made to ensure accuracy, including review by the authors wherever possible. If you identify any issues, please contact us.
Publisher’s note
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.
Supplementary material
The Supplementary Material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fonc.2026.1896994/full#supplementary-material
References
- 1. Ma Z, Men Y, Liu Y, Bao Y, Liu Q, Yang X, et al. Preoperative CT-based radiomic prognostic index to predict the benefit of postoperative radiotherapy in patients with non-small cell lung cancer: a multicenter study. Cancer Imaging. (2024) 24(1):61. doi: 10.21203/rs.3.rs-4113664/v1 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2. Liu Y, Wang Y, Wang X, Xue L, Zhang H, Ma Z, et al. MR radiomics predicts pathological complete response of esophageal squamous cell carcinoma after neoadjuvant chemoradiotherapy: a multicenter study. Cancer Imaging. (2024) 24(1):16. doi: 10.1186/s40644-024-00659-x [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3. Lambin P, Leijenaar RTH, Deist TM, Peerlings J, de Jong EEC, van Timmeren J, et al. Radiomics: the bridge between medical imaging and personalized medicine. Nat Rev Clin Oncol. (2017) 14(12):749–762. doi: 10.1038/nrclinonc.2017.141 [DOI] [PubMed] [Google Scholar]
- 4. Gillies RJ, Kinahan PE, Hricak HJR. Radiomics: images are more than pictures, they are data. (2016) 278:563–77. doi: 10.1148/radiol.2015151169 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5. Wu X, Dai W. Beyond imaging: integrating radiomics, genomics, and multi-omics for precision breast cancer management. (2025) 17:3408. doi: 10.3390/cancers17213408 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6. Wu Z, Ouyang S, Gao J, Hu J, Guo Q, Liu D, et al. Role of radiomics-based multiomics panel in the microenvironment and prognosis of hepatocellular carcinoma. Acad Radiol. (2025) 32(4):1961–1970. doi: 10.1016/j.acra.2024.12.039 [DOI] [PubMed] [Google Scholar]
- 7. Luciano N, Orlandella FM, Braile M, Cavaliere C, Aiello M, Franzese M, et al. Association of radiomic features with genomic signatures in thyroid cancer: a systematic review. J Transl Med. (2024) 22(1):1088. doi: 10.1186/s12967-024-05896-z [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8. Trojani V, Bassi MC, Verzellesi L, Bertolini M. Impact of preprocessing parameters in medical imaging-based radiomic studies: a systematic review. Cancers. (2024) 16(15):2668. doi: 10.3390/cancers16152668 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9. Zhang H, Lu T, Wang L, Xing Y, Hu Y, Xu Z, et al. Robustness of radiomics within photon-counting detector CT: impact of acquisition and reconstruction factors. Eur Radiol. (2025) 35(8):4661–4673. doi: 10.1007/s00330-025-11374-x [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10. Wennmann M, Rotkopf LT, Bauer F, Hielscher T, Kächele J, Mai EK, et al. Reproducible radiomics features from multi-MRI-scanner test-retest-study: influence on performance and generalizability of models. J Magn Reson Imaging. (2025) 61(2):676–686. doi: 10.1002/jmri.29442 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11. Felicetti F, Lamonaca F, Carnì DL, Costanzo S. Image quality standardization in radiomics: a systematic review of artifacts, variability, and feature stability. Sensors. (2026) 26(3):1039. doi: 10.3390/s26031039 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12. Da-Ano R, Visvikis D, Hatt M. Harmonization strategies for multicenter radiomics investigations. Phys Med Biol. (2020) 65(24):24TR02. doi: 10.1088/1361-6560/aba798 [DOI] [PubMed] [Google Scholar]
- 13. Guiot J, Vaidyanathan A, Deprez L, Zerka F, Danthine D, Frix AN, et al. A review in radiomics: making personalized medicine a reality via routine imaging. Med Res Rev. (2022) 42(1):426–440. doi: 10.1002/med.21846 [DOI] [PubMed] [Google Scholar]
- 14. Carré A, Klausner G, Edjlali M, Lerousseau M, Briend-Diop J, Sun R, et al. Standardization of brain MR images across machines and protocols: bridging the gap for MRI-based radiomics. Sci Rep. (2020) 10(1):1–15. doi: 10.1038/s41598-020-69298-z [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15. Johnson WE, Li C, Rabinovic A. Adjusting batch effects in microarray expression data using empirical Bayes methods. Biostatistics. (2007) 8:118–27. doi: 10.1093/biostatistics/kxj037 [DOI] [PubMed] [Google Scholar]
- 16. Jia W, Li H, Ali R, Shanbhogue KP, Masch WR, Aslam A, et al. Investigation of ComBat harmonization on radiomic and deep features from multi-center abdominal MRI data. J Digit Imaging. (2025) 38(2):1016–1027. doi: 10.1007/s10278-024-01253-0 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17. D'Anna A, Stella G, Gueli AM, Marino C, Pulvirenti A. Mitigating interobserver variability in radiomics with ComBat: a feasibility study. J Imaging. (2024) 10(11):270. doi: 10.3390/jimaging10110270 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18. Lee SB, Hong Y, Cho YJ, Jeong D, Lee J, Choi JW, et al. Enhancing radiomics reproducibility: deep learning-based harmonization of abdominal computed tomography (CT) images. Bioengineering. (2024) 11(12):1212. doi: 10.3390/bioengineering11121212 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19. Priya S, Reutzel A, Dalla Pria OAF, Goetz S, Pham HT, Alatoum A, et al. Addressing inter-reconstruction variability in multi-energy myocardial CT radiomics: the benefits of ComBat harmonization. Eur J Radiol. (2025) 183:111891. doi: 10.1016/j.ejrad.2024.111891 [DOI] [PubMed] [Google Scholar]
- 20. Leithner D, Nevin RB, Gibbs P, Weber M, Otazo R, Vargas HA, et al. ComBat harmonization for MRI radiomics: impact on nonbinary tissue classification by machine learning. Invest Radiol. (2023) 58(9):697–701. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21. Jodoin P-M, Edde M, Girard G, Dumais F, Theaud G, Dumont M, et al. Challenges and best practices when using ComBat to harmonize diffusion MRI data. Sci Rep. (2025) 15(1):41508. doi: 10.1038/s41598-025-25400-x [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22. Horng H, Singh A, Yousefi B, Cohen EA, Haghighi B, Katz S, et al. Improved generalized ComBat methods for harmonization of radiomic features. Sci Rep. (2022) 12(1):19009. doi: 10.1038/s41598-022-23328-0 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23. Hu F, Lucas A, Chen AA, Coleman K, Horng H, Ng RW, et al. DeepComBat: a statistically motivated, hyperparameter-robust, deep learning approach to harmonization of neuroimaging data. Hum Brain Mapp. (2024) 45(11):e26708. doi: 10.1101/2023.04.24.537396 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24. Mali SA, Rad NM, Woodruff HC, Depeursinge A, Andrearczyk V, Lambin P, et al. Harmonizing CT scanner acquisition variability in an anthropomorphic phantom: a comparative study of image-level and feature-level harmonization using GAN, ComBat, and their combination. PLoS One. (2025) 20(5):e0322365. doi: 10.1371/journal.pone.0322365 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25. Zwanenburg A, Vallières M, Abdalah MA, Aerts HJ, Andrearczyk V, Apte A, et al. The image biomarker standardization initiative: standardized quantitative radiomics for high-throughput image-based phenotyping. Radiology. (2020) 295(2):328–338. doi: 10.1148/radiol.2020191145 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26. Hosseini SA, Shiri I, Hajianfar G, Ghafarian P, Karam MB, Ay MR. The impact of preprocessing on the PET-CT radiomics features in non-small cell lung cancer. Front Biomed Technol. (2021) 8(4):261–272. doi: 10.18502/fbt.v8i4.7754 [DOI] [Google Scholar]
- 27. Dwivedi P, Barage S, Jha AK, Choudhury S, Rangarajan V. Robustness of (18)F-FDG PET radiomic features in lung cancer: impact of advanced reconstruction algorithm. J Nucl Med Technol. (2025) 53(1):50–56. doi: 10.2967/jnmt.124.268252 [DOI] [PubMed] [Google Scholar]
- 28. Dwivedi P, Barage S, Singh R, Jha A, Choudhury S, Agrawal A, et al. PET-based radiomic analysis in multicentre lung cancer study and impact of feature domain harmonization. Phys Eng Sci Med. (2025) 48(4):1841–1851. doi: 10.1007/s13246-025-01625-y [DOI] [PubMed] [Google Scholar]
- 29. Mali SA, Ibrahim A, Woodruff HC, Andrearczyk V, Müller H, Primakov S, et al. Making radiomics more reproducible across scanner and imaging protocol variations: a review of harmonization methods. J Pers Med. (2021) 11(9):842. doi: 10.3390/jpm11090842 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30. Da-Ano R, Masson I, Lucia F, Doré M, Robin P, Alfieri J, et al. Performance comparison of modified ComBat for harmonization of radiomic features for multicenter studies. Sci Rep. (2020) 10(1):10248. doi: 10.1038/s41598-020-66110-w [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31. Stein CK, Qu P, Epstein J, Buros A, Rosenthal A, Crowley J, et al. Removing batch effects from purified plasma cell gene expression microarrays with modified ComBat. BMC Bioinformatics. (2015) 16(1):1–9. doi: 10.1186/s12859-015-0478-3 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32. Zhang Y, Parmigiani G, Johnson WE. ComBat-seq: batch effect adjustment for RNA-seq count data. NAR Genom Bioinform. (2020) 2(3):lqaa078. doi: 10.1093/nargab/lqaa078 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33. Stanzione A, Cuocolo R, Ugga L, Verde F, Romeo V, Brunetti A, et al. Oncologic imaging and radiomics: a walkthrough review of methodological challenges. Cancers. (2022) 14(19):4871. doi: 10.3390/cancers14194871 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34. Zhang X, Iqbal bin Saripan M, Wu Y, Wang Z, Wen D, Cao Z, et al. The impact of the ComBat method on radiomics feature compensation and analysis of scanners from different manufacturers. BMC Med Imaging. (2024) 24(1):137. doi: 10.1186/s12880-024-01306-4 [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
Publicly available datasets were analyzed in this study. This data can be found here: cancerimagingarchive.net.