Energy spectrum-based correction of washout rates in dual-isotope 123I-BMIPP/201Tl imaging

Kenichi Nakajima 1,2,✉, Shotoku Makida 3,4, Masato Okamura 5, Masa-aki Kawashiri 3, Takayuki Shibutani 6
PMCID: PMC13619768  PMID: 42217155

Abstract

Background

Myocardial metabolism and perfusion in patients with ischemic heart disease or cardiomyopathy can be simultaneously assessed using 123I-β-methyl-iodophenyl-pentadecanoic acid (BMIPP) and 201Thallium (Tl) dual-isotope images. However, cross-contaminated counts between these tracers complicate quantitative analyses, particularly of 123I-BMIPP washout rates (WRs). We developed a simplified method to determine WRs corrected for energy spectral crosstalk from dual-isotope images.

Methods

Our correction method was based on fixed energy-spectrum spill-in ratios (iRatio for 123I and tRatio for 201Tl) that were empirically determined from phantom and patient datasets. These ratios were applied to separate 123I and 201Tl counts in 38 studies of 36 patients with ischemic heart disease using dual-energy planar and single-photon emission computed tomography (SPECT) images. We calculated WRs before and after correction using early and late anterior planar images, as well as mean tomographic counts derived from polar maps. Thresholds of 5% (planar) and 10% (SPECT) were used to assess classification agreement between corrected and uncorrected WRs.

Results

The range of iRatios was 0.57–0.73 and that of tRatios was 0.10–0.12. Higher iRatio values were associated with greater body-weight patients, whereas tRatios were relatively stable regardless of body weight. The corrected 123I-BMIPP WR depends primarily on tRatio and, by formulation, is not directly affected by iRatio. Using a fixed tRatio of 0.12, the correction method revealed strong correlations between corrected and uncorrected WRs for planar imaging (R² = 0.969, p < 0.0001) and SPECT (R² = 0.964, p < 0.0001). Corrected 123I-BMIPP WRs were comparable to uncorrected values for both planar imaging (mean difference, − 1.4%) and SPECT (mean difference, 0.8%). Among the 38 studies, classification of tentative normal versus abnormal WRs remained concordant in 35 (92%) planar studies and 36 (95%) SPECT studies.

Conclusion

Our simplified energy spectrum–based method demonstrated that correction of 123I-BMIPP WRs is feasible under dual-isotope conditions. Although uncorrected WRs may provide approximate estimates, correction is important for reliable quantitative use of dual-isotope datasets, where spectral cross-talk limits accuracy. The proposed model-based approach improves the robustness and interpretability of WR assessment and enables practical, standardized quantitation.

Supplementary Information

The online version contains supplementary material available at https://doi.org/10.1007/s12149-026-02235-8.

Keywords: nuclear cardiology, fatty acid imaging, ischemic heart disease, cardiomyopathy, 123I-labeled radiopharmaceuticals

Introduction

Dual-isotope SPECT with 123I-β-methyl-iodophenyl pentadecanoic acid (123I-BMIPP) and 201Tl-chloride (201Tl) [1–3] has been used to simultaneously evaluate myocardial fatty acid metabolism and perfusion in patients with coronary artery disease and heart failure. Previous visual assessments of matched and mismatched uptake patterns between these tracers have provided critical insights into myocardial viability, especially during the early phase after acute ischemic events. Early mismatches between preserved perfusion (201Tl) and reduced fatty acid uptake (123I-BMIPP) are associated with a viable but ischemic myocardium and may predict functional recovery [4, 5]. Despite the clinical value of dual-tracer imaging for viability assessment, most applications remain qualitative, with limited consideration of quantitative parameters such as 123I-BMIPP washout rates (WRs).

Accurate quantitation of myocardial dual-isotope radiotracer clearance, typically expressed as WRs between early and late images, is important for diagnostic assessment and for understanding myocardial pathophysiology [6, 7]. However, the physical characteristics of these radionuclides complicate quantitative analysis. Whereas 123I emits a principal gamma photon at 159 keV, 201Tl emits characteristic mercury X-rays around 70 keV and additional gamma photons at 167 keV. The 167 keV component of 201Tl partially overlaps the 123I 159 keV photopeak, particularly when energy windows such as 15%–30% are relatively wide. Despite this crosstalk, practical and standardized correction strategies applicable to routine clinical imaging remain limited.

Washout rates of 123I-BMIPP images are key quantitative parameters for diagnosing triglyceride deposit cardiomyovasculopathy (TGCV), for which a WR threshold of 10% is a common diagnostic criterion [8, 9]. Although WRs can be calculated in various ways [10–12], the specific impact of dual-isotope acquisition on WR accuracy has not been systematically evaluated. This limitation is particularly relevant in routine dual-tracer imaging, where spectral overlap between 123I and 201Tl can bias measured WRs.

The primary aims of this study were to evaluate the feasibility of determining 123I-BMIPP WRs under dual-isotope acquisition conditions, to develop a simple energy spectrum–based method for separating tracer signals, and to quantify the bias introduced by dual-isotope imaging. We further propose a practical correction approach to estimate WRs for 123I-BMIPP and to explore its applicability to 201Tl using planar and single-photon emission computed tomography (SPECT) images.

Methods

Data acquisition

Syringes containing 123I-BMIPP and 201Tl solutions (111 MBq each) were placed in a cylindrical acrylic phantom (diameter, 13 cm) designed to simulate the human neck and an upper insert to represent the thyroid gland. Energy spectra were acquired using a dual-head gamma camera (NM830, GE Healthcare, Tokyo, Japan) equipped with extended low-energy general-purpose (ELEGP) collimators. Energy spectra of 123I, 201Tl, and dual-tracer (123I + 201Tl) were acquired for 3 min. These datasets were used to characterize energy-spectrum crosstalk and to establish spill-in ratios for correcting dual-isotope images.

Patients

We acquired 201Tl and 123I-BMIPP images from 36 consecutive patients (38 studies; mean age, 69 ± 10 y; male, n = 22) with ischemic heart disease. Baseline diagnoses included acute myocardial infarction after percutaneous coronary intervention (n = 12; 14 studies), old myocardial infarction (n = 7), vasospastic angina (n = 2), myocardial infarction with non-obstructive coronary arteries (n = 2), cardiomyopathy (n = 2), heart failure (n = 1), and atypical or typical chest discomfort (n = 10). All data were anonymized before analysis, and identifiable patient information was neither accessed nor stored. The Ethics Committee at Kaga Medical Center approved this technical evaluation of dual-tracer correction using anonymized clinical data (Approval No.: study 2025 [R7]-7). The need for patient consent was waived due to the retrospective study design.

Planar and SPECT imaging

Planar and SPECT images were acquired from all patients at 20 min (early) and at 3 h (late) after intravenous injections of 201Tl and 123I-BMIPP (111 MBq each). Anterior planar images were acquired using a 128 × 128 matrix for 5 min per frame under a dual-energy acquisition protocol. Energy windows were centered at 71 keV (30%) for 201Tl and 159 keV (20%) for 123I. Subsequent SPECT images were acquired over a 180° arc. Projection data were acquired using a 64 × 64 matrix, and tomographic images were reconstructed using filtered back projection (FBP) with a Butterworth pre-filter (cutoff 0.45 cycles/cm, order 10) and ordered-subset expectation maximization (OSEM; 2 iterations, 10 subsets, Butterworth post-filter) algorithms. Energy spectra were extracted from quality control (QC) XML files of a subset of seven patients, at early and late time points, to confirm spectral consistency in vivo.

Creation and calculation of separate and combined energy spectra

Separate 123I and 201Tl energy spectra were acquired to simulate combined dual-radionuclide conditions. The relative intensities were adjusted to match approximate peak counts at 71 keV (201Tl) and 159 keV (123I) to mimic clinical acquisition conditions (Fig. 2A). The main energy windows were defined as ± 10% (20% total width) and ± 15% (30% total width) for 123I and 201Tl, centered at 159 and 71 keV, respectively, consistent with standard clinical protocols.

Fig. 2.

Fig. 2

Patient-derived energy spectra and dependence of spill-in ratios on body weight. Energy spectra from five patients weighing 43, 63, 68, 73 and 81 kg acquired under conditions of (A) dual-isotope (123I + 201Tl), (B) 123I, and (C) 201Tl. (D) Relationship between body weight and spill-in ratios (iRatio and tRatio) derived from energy spectra. Shaded areas in panels A and B indicate acquisition windows. 123I-BMIPP, 123I-β-methyl-iodophenyl-pentadecanoic acid

Counts extracted from the spectra are presented as C (radionuclide, window), where I and T denote 123I and 201Tl, and wI and wT denote the 123I and 201Tl windows, respectively. The following variables were defined as:

CI, wI, CI, wT represent 123I counts in the 123I and 201Tl windows, respectively; CT, wI, CT, wT represent 201Tl counts in 123I and 201Tl windows, respectively.

We then used these definitions to calculate spectral contamination “spill-in” ratios as: iRatio = CI, wT /CI, wI and tRatio = CT, wI /CT, wT, representing spill-in of 123I into the 201Tl window and of 201Tl into the 123I window, respectively (Fig. 1B).

Fig. 1.

Fig. 1

Energy spectra of 123I, 201Tl, and dual-isotope 123I/201Tl acquisition in phantom and clinical studies. (A) Energy spectra acquired from a phantom under single- and dual-isotope conditions (123I, 201Tl, and both 123I + 201Tl). (B) Definition of spill-in iRatio and tRatio based on counts in 201Tl and 123I energy windows. (C) Comparison of dual-isotope spectra between the phantom and a representative patient during early and late imaging phases. (D) Dual-isotope energy spectra from six studies of three patients, each acquired at early (E) and late (L) phases. Shaded areas in panels A and B indicate acquisition windows. 123I-BMIPP, 123I-β-methyl-iodophenyl-pentadecanoic acid; i, iodine; Pt, patient; t, thallium

Combined energy spectra generated from phantom data were compared with patient-derived spectra obtained from QC files to confirm the consistency and applicability of the simulated dual-tracer spectra.

Energy spectra under dual-nuclide conditions were analyzed in three patients (six studies) at early and late time points. In addition, five patients underwent sequential acquisitions under single-nuclide conditions (two patients with 123I first and three with 201Tl first). Spectra were acquired again after dual-isotope administration, and the individual 123I and 201Tl components were estimated by subtracting the corresponding single-nuclide spectra from the dual-nuclide spectra.

Separation of 201Tl and 123I counts

When counts were measured during dual-energy acquisition, each energy window represented a mixture of emissions from both radionuclides due to spectral overlap. Therefore, the measured counts in the 201Tl (Inline graphic) and the 123I (Inline graphic) windows under dual-tracer conditions were approximated as follows. A step-by-step mathematical derivation of the correction equations is provided in the Supplementary Material.

Counts in 123I window:

graphic file with name d33e595.gif 1.1

Counts in 201Tl window:

graphic file with name d33e605.gif 1.2

where Inline graphic is the true 123I count in the 123I window (Inline graphic) and Inline graphic is the true 201Tl count in the 201Tl window (Inline graphic). The spill-in ratios are defined as:

graphic file with name d33e636.gif

Based on these definitions, the dual-window counts can be expressed as:

graphic file with name d33e642.gif 2.1
graphic file with name d33e646.gif 2.2

Solving this linear system yields the isolated true counts

graphic file with name d33e652.gif 3.1
graphic file with name d33e656.gif 3.2

Calculation of apparent and true washout rates

We calculated WRs using early (E) and late (L) variables, namely Inline graphic, Inline graphic for 201Tl, and Inline graphic and Inline graphic for 123I-BMIPP, with decay correction applied to late counts. The apparent WRs (without spectral correction) for 123I (IWRapp) and 201Tl (TWRapp) were calculated as:

graphic file with name d33e702.gif 4.1
graphic file with name d33e706.gif 4.2

The decay correction factor (DCF) was defined as:

graphic file with name d33e712.gif 4.3

where HL is the physical half-life (13 h for 123I and 73 h for 201Tl), and Δt is the time interval between early and late scans (hours). After spill-in correction using the spectral separation model, the corrected WRs were calculated from the estimated separated counts:

graphic file with name d33e722.gif 5.1
graphic file with name d33e726.gif 5.2

These equations enabled direct comparisons of apparent WRs derived from dual-window counts and corrected WRs under clinical dual-isotope acquisition conditions.

Calculation of counts in clinical study

Myocardial counts from anterior images were measured for planar analysis. A circular region of interest (ROI) was placed over the cardiac region on 201Tl images acquired in the 71 keV window. The same ROI was then copied to the corresponding 123I-BMIPP image acquired in the 159 keV window to ensure spatial consistency. We calculated WRs for apparent (uncorrected) and spill-in corrected values using the spectral separation model described.

Reconstructed short-axis SPECT images were used to generate polar maps spanning the basal to apical regions of the myocardium. Images were processed using vendor-provided software (GE Healthcare, Tokyo, Japan) using FBP and OSEM reconstruction. The mean myocardial count derived from the polar map served as the representative value for WRs, which were calculated as:

graphic file with name d33e742.gif

where Cearly and Clate indicate mean polar map counts from the early and late scans, respectively. Relationships between planar myocardial counts and polar map counts, and between planar and SPECT-derived WRs, were evaluated by comparing these parameters.

Graphic display and statistical analysis

All quantitative data are presented as mean ± standard deviation (SD). Relationships between corrected and uncorrected values and between planar and SPECT-derived WRs were evaluated using correlation coefficients and linear regression analyses. Linear regression lines were displayed to facilitate comparisons of tracer separation performance and WR reproducibility. Statistical significance was defined as p < 0.05. Energy spectra, simulated dual-tracer spectral overlap, and scatter plots for WR comparisons were generated using the Wolfram Language version 14.3 (Wolfram Research, Inc., Champaign, IL, USA) and JMP v.18.2 (SAS Institute Inc., Cary, NC, USA).

Results

Energy spectra of 123I and 201Tl in phantom and patient studies

Energy spectra were acquired under single-tracer conditions in phantom experiments and clinical patient studies to characterize spectral overlap between 123I and 201Tl. Distinct photopeaks for 123I (159 keV) and 201Tl (≈ 71 keV) were observed in phantom spectra, confirming appropriate energy window settings. Based on the phantom spectra, the spill-in iRatio and tRatio were 0.33 and 0.10, respectively (Fig. 1B).

The overall spectral profiles remained consistent at early and late time points in clinical data derived from three patients (six studies). Shoulder counts slightly increased in patient spectra compared with the phantom (Fig. 1C), which likely reflected scatter and Compton contributions. Nonetheless, peak positions and shapes remained stable among all six studies (Fig. 1D). These findings supported the reproducibility of energy distribution in vivo.

Scatter fractions were higher in energy spectra obtained under dual-nuclide conditions for 123I and 201Tl in patients weighing 43, 63, 68, 73, and 81 kg (Fig. 2) than in phantom spectra. The scatter contribution of 201Tl in the intermediate energy range between the 201Tl and 123I peaks increased with body weight, resulting in a weight-dependent increase in spill-in from 123I into the 201Tl window. The iRatio range was accordingly 0.57–0.73. In contrast, spill-in from 201Tl into the 123I window was relatively stable, with a tRatio of 0.10–0.12. Figure 2D shows that tRatio remained approximately constant (~ 0.12), whereas iRatio increased with body weight. The relationship between the iRatio and body weight was approximated using the following empirical model (body weight < 81 kg):

graphic file with name d33e825.gif 6

However, as shown in Fig. 2, the fit was not perfect, and the equation should be interpreted as an approximation rather than a strict predictive model. Because the corrected 123I-BMIPP washout rate (IWRestimated) depended primarily on the tRatio (Eq. 5.1), the observed stability of tRatio supported the use of a fixed spill-in ratio for subsequent dual-tracer corrections. Accordingly, in subsequent calculations, we used a fixed tRatio of 0.12 and a representative iRatio (median 0.64). These empirically determined ratios, derived from clinical data, were applied in all subsequent analyses to separate tracer counts and calculate corrected WRs in planar and SPECT images under dual-isotope conditions.

Calculation of estimated 201Tl and 123I counts

True radionuclide-specific counts were estimated using the fixed spill-in ratios and solving the simultaneous equations described in the Methods section. This correction enabled the separation of dual-tracer contributions within each energy window. Table 1 shows the count separation results derived from early and late planar and SPECT images of a representative patient. After applying the spill-in ratios, tracer-specific counts were obtained for both planar and SPECT datasets. The difference between apparent and corrected 123I WRs was small in this patient (planar and SPECT images: 18% vs. 17% and 21% vs. 20%, respectively).

Table 1.

Examples of 123I and 201Tl counts separation from dual-window measurements

201Tl (71 keV, 30%) 123I (159 keV, 20%)
Spill-in ratios
iRatio of 123I to 201Tl window 0.64 -
tRatio of 201Tl to 123I window - 0.12
Planar study: anterior images
Early/late counts 265/183 218/153
Early/late corrected counts 136/92 202/142
Apparent WR (uncorrected, %) 29% 18%
Corrected WR (spill-in corrected, %) 30% 17%
SPECT study: polar maps
Early/late counts 794/497 713/480
Early/late corrected counts 366/206 669/455
Apparent WR (uncorrected, %) 36% 21%
Corrected WR (spill-in corrected, %) 42% 20%

WR, washout rate

Comparison of corrected and uncorrected washout rates

The corrected and uncorrected WRs for 123I-BMIPP (p < 0.0001) were strongly correlated after applying spectral separation to correct cross-contamination (Fig. 3A). Bland–Altman analysis showed a small negative bias (corrected – uncorrected), with a mean difference of -1.4% ± (SE) 0.3%. In contrast, the 201Tl WRs showed only moderate correlation between corrected and uncorrected data, and the corrected WRs of 201Tl were unstable (R2 = 0.708, p < 0.0001; Fig. 3B). Even when a body weight -dependent iRatio was applied (Eq. 6), the correlation remained similar (R2 = 0.684, p < 0.0001), indicating no substantial improvement with the use of a fixed iRatio (0.64).

Fig. 3.

Fig. 3

Comparison of uncorrected and corrected planar WRs. Scatter plots show relationships between uncorrected and corrected WRs for (A) 123I-BMIPP and (B) 201Tl, with linear regression lines. Right panels show the corresponding Bland–Altman plots. Dashed lines in panel A indicate the 5% WR threshold used for classification. 123I-BMIPP, 123I-β-methyl-iodophenyl-pentadecanoic acid; WR(s), washout rate(s)

Because of this instability, further analyses focused on 123I-BMIPP SPECT-derived counts. The comparison of planar and SPECT-derived myocardial counts further supported the validity of the correction approach. Counts derived from anterior planar images and averaged from SPECT polar maps closely correlated; the relationships between SPECT and planar counts were as follows: R2 = 0.862; SPECT = -224 + 4.21 × planar count for 123I-BMIPP (Fig. 4A), and R2= 0.801; SPECT = -157 + 3.12 × planar count for 201Tl (p < 0.0001 for all).

Fig. 4.

Fig. 4

Relationships between planar and SPECT counts and WRs for 123I-BMIPP. (A) Correlation between anterior planar myocardial counts and mean polar map counts derived from SPECT images. (B) Comparison of BMIPP counts from SPECT images reconstructed using FBP and OSEM. (C) Right panel shows the relationship between uncorrected and corrected SPECT BMIPP WRs, with the corresponding Bland–Altman plot. Dashed lines in panel C indicate the reference 10% WR threshold. 123I-BMIPP, FBP, filtered back projection; OSEM, ordered-subset expectation maximization, WRs, washout rates

Correlations between SPECT counts corrected by FBP and OSEM reconstruction were excellent (R2 = 0.996, Fig. 4B) and remained high when WR values derived from SPECT were compared before and after correction (R2 = 0.964; p < 0.0001; Fig. 4C). Bland–Altman plots revealed minimal differences between corrected and uncorrected WRs (0.8%, p = 0.129).

Threshold-based discrepancy analysis

We assessed the potential clinical impact of spectral cross-talk on WR classification using the threshold criteria of 5% (planar) and 10% (SPECT) to classify 123I-BMIPP washout as normal or abnormal [10]. Classification agreement was evaluated before and after spill-in correction.

Corrected and uncorrected planar WR values resulted in discordant classification in only 3 (8%) of 38 studies. All discrepancies occurred near the 5% cutoff, indicating that crosstalk correction may influence interpretation primarily in borderline cases. A comparison of corrected planar and SPECT WRs based on their respective thresholds revealed discordant classification in 3 (8%) of 38 studies, again clustered near the threshold values.

Discussion

We developed a practical reproducible method for correcting spectral cross-talk between 123I and 201Tl in dual-isotope imaging. Fixed spill-in ratios derived from energy spectra (iRatio for spill-in of 123I into the 201Tl window and tRatio for spill-in of 201Tl into the 123I window) allow estimation of tracer-specific counts and corrected WRs that could be calculated without requiring additional hardware or repeated patient-specific energy spectral acquisition. This approach is particularly relevant in clinical settings where accurate 123I-BMIPP WR quantitation is required, such as TGCV evaluations and myocardial metabolic impairment assessments after ischemia. The simplicity and robustness of our proposed method support wider clinical implementation of dual-tracer protocols.

Combination of 123I-labeled radiopharmaceuticals and 201Tl

Dual-isotope imaging with 123I-labeled radiopharmaceuticals and 201Tl has been widely applied in Japan to evaluate ischemic heart disease, as 123I-BMIPP determines myocardial fatty acid metabolism and 201Tl defines cardiac perfusion. The high sensitivity of 123I-BMIPP allows the detection of metabolic abnormalities in acute coronary syndrome and spasm-related ischemia [13, 14]. During the acute phase particularly after successful coronary intervention, mismatch patterns characterized by preserved 201Tl uptake and reduced 123I-BMIPP uptake indicate viable, but stunned myocardium, and may predict subsequent functional recovery. Similarly, BMIPP defects in vasospastic angina often correspond to vasospastic territories and normalize together with recovered left ventricular contractility after appropriate therapy [15, 16].

The Japanese Circulation Society (JCS) guidelines support the application of 123I-BMIPP SPECT with Class I or IIa indications for diagnosing myocardial ischemia, evaluating vasospastic angina, and assessing myocardial damage in patients undergoing hemodialysis [17]. Although separate acquisitions for metabolism and perfusion are feasible, simultaneous dual-isotope imaging with 123I-BMIPP and 201Tl remains clinically attractive because it enables efficient assessment of mismatch patterns in time-sensitive settings.

Subtraction-based approaches and multi-window correction methods have been applied to isolate tracer activity to address photon energy overlap between 123I and 201Tl. While these methods support visual interpretation, their impact on quantitative indices such as WRs has not been systematically validated. We addressed this gap by quantifying the influence of spectral crosstalk and correction on WR measurements in phantom and clinical datasets.

Dual-isotope imaging has also been applied to combinations of 123I-meta-iodobenzylguanidine (MIBG) and 201Tl, enabling assessment of perfusion–innervation mismatch. This mismatch has clinical value for risk stratification in heart failure, particularly for predicting sudden cardiac death and ventricular arrhythmias [18, 19]. Therefore, the proposed correction framework may be applicable to other 123I/201Tl dual-isotope protocols such as 123I-MIBG studies, although radionuclide- and system-specific determination of spill-in ratios would be required.

Late 123I-BMIPP images and washout rates

Late-phase 123I-BMIPP images are typically acquired 3 h after a radiolabeled tracer is injected into a patient. This approach was initially explored during early clinical investigations in the 1990s but was not widely adopted into routine practice. This was largely due to the predominant clinical emphasis on detecting mismatch patterns between fatty acid metabolism (123I-BMIPP) and perfusion [1, 20, 21], rather than focusing on quantitative indices such as WRs. Consequently, clinical use of late BMIPP imaging gradually declined in most institutions.

However, interest in WR quantitation has been revitalized with the recognition of TGCV as a distinct clinical entity [9, 22–24]. A viable myocardium rapidly takes up and incorporates 123I-BMIPP into the intracellular triglyceride pool, where it is subsequently cleared over time. Therefore, WRs provide a quantitative measure of the dynamic status of myocardial lipid metabolism.

Proposed correction method using energy-spectrum analysis

Our proposed approach assumes that spill-in ratios between radionuclides remain sufficiently stable under clinical imaging conditions to allow mathematical decomposition of measured counts using fixed coefficients. This assumption is supported by the finding that overall energy-spectral patterns are reproducible in vivo. In this study, the tRatio (spill-in of 201Tl into the 123I window) remained relatively stable among patients, whereas iRatio (spill-in of 123I into the 201Tl window) varied with body habitus. This variability is likely attributable to differences in scatter conditions, which are influenced not only by body weight but also by stature and attenuation geometry, as shown by 201Tl spectral variation (Fig. 2B). Although alternative contamination indices such as CT, wI/CI, wI or CI, wT/CT, wT ratios could also be defined, they were not adopted because they depend on relative administered activities and myocardial 201Tl and 123I-BMIPP uptake, which may substantially vary among patients. In contrast, the spill-in ratios derived from energy spectra in our model were less sensitive to inter-patient variability in tracer uptake.

Spill-in ratios were derived from clinical energy spectra rather than phantom measurements. The apparent 201Tl peak detected around 71 keV is a composite, rather than a single discrete photon energy, reflecting characteristic mercury X-rays (Kα = 70.8 keV, 73.7%; Kβ = 80.3 keV, 20.4%) as well as additional characteristic X-rays generated by interactions with the lead collimator (Pb Kα ≈ 75 keV; Pb Kβ ≈ 85 keV). Therefore, peak positions and spectral shapes may differ slightly between phantom and patient conditions. However, because spill-in ratios were consistently derived from clinical data, these variations are considered unlikely to significantly affect correction performance.

Optimal iRatios and tRatios might differ among imaging systems and protocols, including detector type (Anger vs. cadmium-zinc-telluride [CZT]), collimator design, energy window settings, and reconstruction parameters. Although agreement between FBP and OSEM reconstructions was excellent, newer CZT cameras with improved energy resolution might further reduce spectral overlap compared with conventional Anger systems, and may improve quantitative accuracy in dual-isotope imaging, including protocol combinations of 123I-BMIPP with perfusion tracers [25, 26].

The purpose of this study was not to define universal spill-in ratios, but rather to provide a systematic and clinically practical correction framework. The method can be locally optimized at any institution using simple phantom measurements and a few representative patient spectra. This spill-in ratio–based approach allows dual-isotope protocols to retain both qualitative and quantitative utility, including more accurate WR calculation, even in the presence of overlapping radionuclide energy spectra.

Corrected versus uncorrected washout rates

This study showed that 123I-BMIPP WRs corrected for spectral crosstalk in dual-isotope imaging strongly and consistently correlated with uncorrected values. Corrected WRs derived from planar images were slightly lower on average than apparent WRs, while maintaining a high correlation, supporting the internal consistency of the correction approach. A similarly strong correlation was observed for SPECT-derived WRs. Thus, uncorrected WRs may provide reasonable estimates in many clinical situations. However, in cases near diagnostic thresholds (e.g., TGCV vs. normal), even small differences may influence classification. Therefore, the correction method is valuable for improving diagnostic accuracy and reducing the risk of misclassification in borderline cases. In contrast, 201Tl WRs were not sufficiently stable when using simple fixed spill-in ratios and were therefore not further evaluated for quantitative WR analyses. Moreover, variability in iRatio could not be sufficiently compensated by simple body weight–based adjustment.

In clinical settings such as acute coronary syndrome, angina pectoris, and myocardial infarction, WRs near 0% might suggest significantly impaired myocardial fatty acid clearance and possible TGCV. By reducing bias due to spectral overlap, our proposed correction method improves confidence in WR interpretation and enhances the reproducibility of quantitative assessment using dual-isotope 123I-BMIPP/201Tl protocols.

Washout rates in TGCV

Defective intracellular lipolysis in TGCV results in low 123I-BMIPP clearance (nearly zero), indicating severely impaired fatty acid utilization and metabolic trapping [11, 23]. A 123I-BMIPP WR of < 10% has been proposed as part of the diagnostic criteria for TGCV when supported by other characteristic findings such as myocardial triglyceride accumulation on computed tomography or magnetic resonance images, along with angiographic evidence of diffuse concentric coronary narrowing [9, 24, 27]. These elements form the basis of the TGCV diagnostic algorithm established through Japanese clinical research.

The Japanese Society of Nuclear Cardiology (JSNC) and the Japanese Society of Nuclear Medicine (JSNM) have recommended single-isotope acquisition for 123I-BMIPP WR analysis to support accurate WR measurements [11]. Although dual-isotope imaging is clinically attractive for simultaneous metabolism and perfusion assessment, it has not been recommended for WR quantitation because spectral cross-talk between 123I and 201Tl can introduce substantial bias in tracer separation and WR calculations.

Nevertheless, some institutions have applied dual-isotope 123I-BMIPP and 201Tl protocols during the early phase of TGCV research, before correction methods were standardized. The present findings provide a reproducible mathematical framework for separating 123I and 201Tl counts using empirically derived energy spectra, which allows estimation of 123I-BMIPP WRs under dual-isotope conditions. This approach may improve the interpretability of archived dual-isotope datasets and also support quantitative analysis in settings where dual-isotope acquisition is used for clinical efficiency.

To assess potential clinical impact, we applied previously proposed cutoff values for abnormal WR (5% for planar imaging and 10% for SPECT) [10]. Classification discrepancies between corrected and uncorrected WR values were uncommon when the 10% threshold was used. However, misclassification may occur primarily near decision thresholds, indicating that correction may influence interpretation mainly in borderline cases. Moreover, when dual-nuclide studies are performed in patients with coronary artery disease or heart failure, simultaneous calculation of corrected WRs may obviate the need for an additional 123I-BMIPP study. The proposed method also provides a practical means to re-evaluate previously acquired datasets, thereby preserving their clinical and research utility, particularly in the context of TGCV.

Finally, although several studies in this cohort had 123I-BMIPP WRs < 10%, TGCV cannot be inferred from WR alone. Diagnostic confirmation requires multimodal imaging and clinical correlation. Because clinical experience with 123I-BMIPP WR remains limited, further evaluation across diverse clinical and pathophysiological conditions is warranted.

Limitations and future directions

Several limitations should be acknowledged. This study, conducted at a single-center, obtained all energy spectra, spill-in ratios, and validation analyses using a single gamma camera system with fixed acquisition and reconstruction protocols. Therefore, generalizability to other institutions, camera models, and imaging conditions remains to be established.

Our proposed correction for 123I-BMIPP WR relies primarily on tRatio in both planar and SPECT imaging. Because SPECT reconstruction introduces depth-dependent attenuation and scatter effects, spill-in might vary spatially within the myocardium. Although agreement between FBP and OSEM reconstructions was excellent, other reconstruction approaches (e.g., resolution recovery, attenuation correction, or more sophisticated scatter correction models) might differently influence spill-in.

In addition, 201Tl WR could not be reliably quantified using fixed iRatio values, and iRatio variability related to body habitus was not evaluated in patients weighing > 81 kg. Although iRatio was associated with body weight, scatter within the body is influenced by multiple factors beyond body weight alone, which may contribute to its instability. This may explain why body weight–based adjustment did not sufficiently improve the reliability of 201Tl WR. In contrast, tRatio (spill-in from 201Tl into the 123I window) was relatively stable and can be used for calculating 123I-BMIPP WRs.

For a more rigorous comparison, data from both single-isotope and dual-isotope studies would be desirable. However, in clinical practice, it was not feasible to perform repeated acquisitions using both protocols in the same patients because of ethical and logistical constraints, including additional radiation exposure.

Although patient spectra were generally consistent in our cohort, spectral characteristics may differ among institutions depending on detector type (Anger vs. CZT), collimator design, energy resolution, and energy window settings. These factors might affect the applicability of fixed spill-in ratios when implemented under various clinical conditions.

Finally, while the present study demonstrates the feasibility of a practical correction framework, further investigation is required to validate and optimize the approach across diverse imaging systems and patient populations. Multicenter studies are needed to establish system-specific correction factors and to develop practical recommendations for routine clinical implementation of dual-isotope protocols.

Conclusion

This study demonstrated that correcting spectral crosstalk between 123I and 201Tl in dual-isotope imaging using fixed, energy spectrum–derived spill-in ratios is feasible and robust. The proposed method enables practical WR quantitation under dual-tracer acquisition conditions without requiring patient-specific spectral measurements. This approach supports reliable quantitative assessment, particularly in threshold-based interpretation, and may facilitate broader clinical use of dual-isotope protocols for myocardial ischemia and TGCV.

Electronic Supplementary Material

Below is the link to the electronic supplementary material.

Supplementary Material 1 (135.9KB, pdf)

Acknowledgements

We appreciate the editorial assistance of Norma Foster (English Express, North Vancouver, BC, Canada).

Abbreviations

BMIPP

β-methyl-iodophenyl pentadecanoic acid

CZT

Cadmium-zinc-telluride

FBP

Filtered back projection

OSEM

Ordered-subset expectation maximization

SPECT

Single-photon emission computed tomography

TGCV

Triglyceride deposit cardiomyovasculopathy

WR

Washout rate

Author contributions

K. Nakajima contributed to the initial study conception and design, and performed the statistical analysis. Data collection and analysis were performed by K. Nakajima and M. Okamura. Critical comments on methodology were provided by K. Shibutani. Clinical diagnosis was made by S. Makida and M. Kawashiri. All authors have read and approved the final manuscript.

Funding

Open Access funding provided by Kanazawa University. This study was funded by JSPS Grants-in-Aid for Scientific Research (C) in Japan (Grant Number 23K07203; Principal Investigator: K. Nakajima).

Data availability

The datasets generated during and/or analyzed during the current study are available from the corresponding author on reasonable request.

Declarations

Conflict of interest

K. Nakajima worked in an endowed department funded by PDRadiopharma, Tokyo, Japan, Nihon Medi-Physics, Tokyo, Japan, and Siemens Healthcare, Tokyo, Japan, until March 2024.

Ethics approval

The Ethics Committee at Kaga Medical Center approved this study (Approval No. 2025 [R7]-7). The requirement for patient consent was waived due to the retrospective study design.

Footnotes

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.Hosokawa R, Nohara R, Fujibayashi Y, Okuda K, Ogino M, Hirai T, et al. Myocardial metabolism of 123I-BMIPP in a canine model with ischemia: implications of perfusion-metabolism mismatch on SPECT images in patients with ischemic heart disease. J Nucl Med. 1999;40:471–8. [PubMed] [Google Scholar]
  • 2.Nakamura A, Momose M, Kondo C, Nakajima T, Kusakabe K, Hagiwara N. Ability of 201Tl and 123I-BMIPP mismatch to diagnose myocardial ischemia in patients with suspected coronary artery disease. Ann Nucl Med. 2009;23:793–8. [DOI] [PubMed] [Google Scholar]
  • 3.Akutsu Y, Yamanaka H, Nishimura H, Hamazaki Y, Kaneko K, Kodama Y, et al. Thallium-201 and I-123 beta-methyl iodophenyl-pentadecanoic acid dual isotope single photon emission computed tomography for evaluating reperfusion injury after successful reperfusion therapy. Clin Nucl Med. 2009;34:213–5. [DOI] [PubMed] [Google Scholar]
  • 4.Kawamoto M, Tamaki N, Yonekura Y, Tadamura E, Fujibayashi Y, Magata Y, et al. Combined study with I-123 fatty acid and thallium-201 to assess ischemic myocardium: comparison with thallium redistribution and glucose metabolism. Ann Nucl Med. 1994;8:47–54. [DOI] [PubMed] [Google Scholar]
  • 5.Yasugi N, Koyanagi S, Ohzono K, Sakai K, Matsumoto T, Sako S, et al. Comparative study of dobutamine stress echocardiography and dual single-photon emission computed tomography (thallium-201 and I-123 BMIPP) for assessing myocardial viability after acute myocardial infarction. Circ J. 2002;66:1132–8. [DOI] [PubMed] [Google Scholar]
  • 6.Fujibayashi Y, Nohara R, Hosokawa R, Okuda K, Yonekura Y, Tamaki N, et al. Metabolism and kinetics of iodine-123-BMIPP in canine myocardium. J Nucl Med. 1996;37:757–61. [PubMed] [Google Scholar]
  • 7.Nohara R, Hosokawa R, Hirai T, Okuda K, Ogino M, Fujibayashi Y, et al. Basic kinetics of 15-(p-iodophenyl)-3-R,S-methylpentadecanoic acid (BMIPP) in canine myocardium. Int J Card Imaging. 1999;15:11–20. [DOI] [PubMed] [Google Scholar]
  • 8.Hirano K, Higashi M, Miyauchi H, Takagi A, Ikeda Y, Nakano Y, et al. Increased Washout of 123I-BMIPP in Triglyceride Deposit Cardiomyovasculopathy (TGCV) with Severe Coronary Stenosis. Ann Nucl Cardiol. 2019;5:47–9. [Google Scholar]
  • 9.Kobayashi K, Sakata Y, Miyauchi H, Ikeda Y, Nagasawa Y, Nakajima K, et al. The Diagnostic Criteria 2020 for Triglyceride Deposit Cardiomyovasculopathy. Ann Nucl Cardiol. 2020;6:99–104. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Chen Z, Nakajima K, Hirano K, Kamiya T, Yoshida S, Saito S, et al. Methods of calculating 123I-beta-methyl-P-iodophenyl-pentadecanoic acid washout rates in triglyceride deposit cardiomyovasculopathy. Ann Nucl Med. 2022;36:986–97. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Nakajima K, Miyauchi H, Hirano KI, Fujimoto S, Kawahito M, Iimori T, et al. Practice recommendation for measuring washout rates in (123)I-BMIPP fatty acid images. Ann Nucl Med. 2024;38:1–8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Miyauchi H, Ono R, Iimori T, Sawada K, Hoshi K, Hirano K, et al. Modified Algorithm Using Total Count for Calculating Myocardial Washout Rate in Single-Photon Emission Computerized Tomography. Ann Nucl Cardiol. 2022;9:19–25. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Kurisu S, Fujiwara H. Assessing a Myocardial Area at Risk in Non-ST Elevation Acute Myocardial Infarction Without Wall Motion Abnormalities Using Cardiac Magnetic Resonance and Radionuclide Imaging. Cureus. 2024;16:e55125. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Kawai Y, Tsukamoto E, Nozaki Y, Kishino K, Kohya T, Tamaki N. Use of 123I-BMIPP single-photon emission tomography to estimate areas at risk following successful revascularization in patients with acute myocardial infarction. Eur J Nucl Med. 1998;25:1390–5. [DOI] [PubMed] [Google Scholar]
  • 15.Yoneyama K, Akashi YJ, Kida K, Ashikaga K, Musha H, Suzuki K, et al. Metabolic planar imaging using 123I-beta-methyl-iodophenyl pentadecanoic acid identifies myocardial ischemic memory after intracoronary acetylcholine provocation tests in patients with vasospastic angina. Int Heart J. 2014;55:113–8. [DOI] [PubMed] [Google Scholar]
  • 16.Nakajima K, Shimizu K, Taki J, Uetani Y, Konishi S, Tonami N, et al. Utility of iodine-123-BMIPP in the diagnosis and follow-up of vasospastic angina. J Nucl Med. 1995;36:1934–40. [PubMed] [Google Scholar]
  • 17.Yamagishi M, Tamaki N, Akasaka T, Ikeda T, Ueshima K, Uemura S, et al. JCS 2018 Guideline on Diagnosis of Chronic Coronary Heart Diseases. Circ J. 2021;85:402–572. [DOI] [PubMed] [Google Scholar]
  • 18.Sazonova SI, Atabekov TA, Batalov RE, Mishkina AI, Varlamova JV, Zavadovsky KV, et al. Prediction of appropriate ICD therapy in patients with ischemic heart failure. J Nucl Cardiol. 2022;29:680–91. [DOI] [PubMed] [Google Scholar]
  • 19.McGhie AI, Corbett JR, Akers MS, Kulkarni P, Sills MN, Kremers M, et al. Regional cardiac adrenergic function using I-123 meta-iodobenzylguanidine tomographic imaging after acute myocardial infarction. Am J Cardiol. 1991;67:236–42. [DOI] [PubMed] [Google Scholar]
  • 20.Iida H, Hanaoka H, Okada Y, Kiyokawa T, Takakuwa Y, Yamada H, et al. A low perfusion-metabolic mismatch in 99mTl and 123I-BMIPP scintigraphy predicts poor prognosis in systemic sclerosis patients with asymptomatic cardiac involvement. Int J Rheum Dis. 2019;22:1008–15. [DOI] [PubMed] [Google Scholar]
  • 21.Morishima M, Kiriyama T, Miyagi Y, Otsuka T, Fukushima Y, Kumita SI, et al. Serial change in perfusion-metabolism mismatch after coronary artery bypass grafting. Ann Nucl Med. 2022;36:244–54. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Hirano K, Ikeda Y, Zaima N, Sakata Y, Matsumiya G. Triglyceride deposit cardiomyovasculopathy. N Engl J Med. 2008;359:2396–8. [DOI] [PubMed] [Google Scholar]
  • 23.Hirano K, Ikeda Y, Sugimura K, Sakata Y. Cardiomyocyte steatosis and defective washout of iodine-123-beta-methyl iodophenyl-pentadecanoic acid in genetic deficiency of adipose triglyceride lipase. Eur Heart J. 2015;36:580. [DOI] [PubMed] [Google Scholar]
  • 24.Nakajima K. Triglyceride deposit cardiomyovasculopathy: how to recognise a new disease entity. Heart. 2021;107:93–5. [DOI] [PubMed] [Google Scholar]
  • 25.Yamada Y, Nakano S, Gatate Y, Okano N, Muramatsu T, Nishimura S, et al. Feasibility of simultaneous 99mTc-tetrofosmin and 123I-BMIPP dual-tracer imaging with cadmium-zinc-telluride detectors in patients undergoing primary coronary intervention for acute myocardial infarction. J Nucl Cardiol. 2021;28:187–95. [DOI] [PubMed] [Google Scholar]
  • 26.Shibutani T, Onoguchi M, Yoneyama H, Konishi T, Nakajima K. Simultaneous dual-isotope SPECT imaging using a cardiac cadmium-zinc-telluride camera with various Technetium-99 m to Iodine-123 ratios. EJNMMI Rep. 2025;9:36. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Aoshima C, Fujimoto S, Kudo A, Kawaguchi YO, Takamura K, Matsue Y, et al. Clinical significance of 123I-BMIPP washout rate in patients with uncertain chronic heart failure. Eur J Nucl Med Mol Imaging. 2022;49:3129–39. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supplementary Material 1 (135.9KB, pdf)

Data Availability Statement

The datasets generated during and/or analyzed during the current study are available from the corresponding author on reasonable request.


Articles from Annals of Nuclear Medicine are provided here courtesy of Springer