Abstract
Purpose
Using computer-assisted equipment (CAE) in total hip arthroplasty (THA) has facilitated improved accuracy of acetabular cup placement, but the impact of aligning the femoral stem on post-operative patient satisfaction remains unclear. In this study, we evaluated the post-operative patient-reported outcome measures (PROMs) and identified factors related to femoral stem alignment associated with post-operative PROMs after THA.
Methods
This retrospective study included 200 patients (205 hips; mean age, 67.3 ± 9.8 years) undergoing THA using CAE for acetabular cup placement at our hospital. Alignment parameters of the femoral stem, including stem anteversion, change in stem anteversion, femoral offset, and coronal and sagittal stem angles, were evaluated using computed tomography scans and radiographs. Patients were categorised into the normal and Low&High alignment groups based on predefined, clinically acceptable ranges and deviation from native femoral anatomy. Patient satisfaction was assessed using a visual analogue scale and clinical outcomes were evaluated using the Japanese Orthopaedic Association Hip Disease Evaluation Questionnaire (JHEQ) at one year post-operatively.
Results
Patients in the normal alignment group demonstrated significantly higher JHEQ total and movement scores than those in the Low&High group. Through multivariate logistic regression analysis, we identified change in stem anteversion as an independent factor associated with high post-operative patient satisfaction.
Conclusion
Accurate femoral stem alignment, particularly minimizing deviations in stem anteversion from the native femoral anteversion, was significantly associated with improved post-operative PROMs after computer-assisted THA. These results highlight the importance of achieving accurate femoral stem alignment, even when acetabular cup placement is optimised using CAE.
Supplementary Information
The online version contains supplementary material available at https://doi.org/10.1007/s00264-026-06949-2.
Keywords: Computer-assisted surgery, Femoral anteversion, Femoral stem alignment, Patient satisfaction, Patient-reported outcome measures, Total hip arthroplasty
Introduction
Total hip arthroplasty (THA) is an effective surgical treatment for alleviating pain and restoring function in patients with hip disorders such as osteoarthritis, osteonecrosis of the femoral head, and rheumatoid arthritis [1]. Accurately positioning the implant is essential for achieving optimal clinical outcomes because malposition may lead to dislocation, impingement, discrepancy in leg length, prosthetic wear, and early revision surgery [2–4].
Recent advances in computer-assisted equipment (CAE) include robotic-assisted systems, computed tomography (CT)-based navigation, and portable augmented reality-based navigation. Such advances have significantly improved the accuracy of acetabular cup placement during THA by providing real-time intraoperative guidance [5, 6].
However, despite these technological advances and their increasing adoption, improvements in radiographic accuracy of cup placement do not necessarily translate into better post-operative patient-reported outcome measures (PROMs) [7, 8].
Kaneta et al. further demonstrated that, among different CAE modalities, the smallest cup placement errors were achieved with robotic-assisted systems. Yet this increased accuracy did not result in improved post-operative PROMs [9]. Thus, factors beyond acetabular positioning, particularly those related to the femoral component, may influence post-operative satisfaction.
Femoral stem alignment has been reported to play a crucial role in restoring hip biomechanics and ensuring joint stability [10]. Malalignment in stem anteversion, coronal or sagittal alignment, or offset may lead to altered joint kinematics, soft tissue imbalance, or impingement, ultimately affecting patient-perceived outcomes [11, 12]. However, although previous studies have investigated radiological and functional outcomes after THA, only a few have evaluated post-operative PROMs from the perspective of stem alignment and quantitatively examined the association between detailed stem alignment parameters and post-operative PROMs in computer-assisted THA.
Therefore, this study investigated the post-operative PROMs and the association between femoral stem alignment and post-operative PROMs under standardised, computer-assisted acetabular cup placement.
We hypothesised that femoral stem alignment has an impact on post-operative PROMs and that alignment closer to the native femoral anatomy, particularly minimal change in stem anteversion, would be associated with higher patient satisfaction in computer-assisted THA.
Materials and methods
Study participants
This retrospective single-institution study adhered to principles of the Declaration of Helsinki and was approved by the institutional review board of our hospital. Informed consent was obtained from all participants.
Patients who underwent primary THA using one of three CAE systems for acetabular cup placement at our hospital between May 2021 and August 2023, with at least 12 months follow-up, were included. The CAE systems included robotic-assisted (Mako; Stryker, Kalamazoo, MI, USA), CT-based navigation (CT-Hip System version 1.1, Stryker, Freiburg, Germany), and portable navigation (AR Hip; Zimmer Biomet Japan, Tokyo, Japan). Patients undergoing bilateral or revision THA, those with neurological, knee, or spinal diseases, contralateral severe hip osteoarthritis, arthroplasty, non-elective or oncological THA, or missing PROM data were excluded [5, 13]. Spinal disease was defined as any case of lumbar spinal stenosis, scoliosis, or vertebral compression fractures identified using whole-spine X-ray examinations, because these may influence PROMs and spino-pelvic mobility. All patients participated in our institution-wide comprehensive THA pathway program, including uniform protocols for perioperative care, institutional post-operative rehabilitation, and pain management.
Pre-operative planning
For all patients, pre-operative planning for both the acetabular and femoral components was performed using a CT-based three-dimensional template, with the supine functional pelvic plane used as the radiographic reference [14]. To improve reproducibility of the functional pelvic plane during CT scanning, a pillow was placed under the sacrum to align the pelvis in the supine position with the bilateral anterior superior iliac spines in the same horizontal plane. The inter-teardrop line was used as the mediolateral axis.
The target cup inclination was set at 40° and the anteversion angle was determined according to Widmer’s combined anteversion theory, adjusting it to 15° ± 5° based on individual patient anatomy [15]. The planned stem version was calculated to achieve the optimal combined anteversion and avoid impingement.
Surgical procedure
A single surgeon performed THA in all patients using the anterolateral supine approach. Based on a study by Kaneta et al., one of the three CAE systems was used to assist acetabular cup placement [9]. To mimimise selection bias, the type of CAE system used for each patient was randomly assigned.
A cementless Taperloc Microplasty stem with a G7 acetabular cup (Zimmer Biomet) or an Accolade II femoral stem with a Trident II Tritanium cup (Stryker) was implanted in all patients. These were combined with a highly cross-linked polyethylene liner and a 32- or 36-mm delta ceramic head. The femoral stem was inserted according to the manufacturer’s recommendations, with careful attention given to achieving anatomical alignment in terms of anteversion, varus–valgus angle, flexion–extension angle, offset, and leg length.
Radiographic evaluation
At 3 months post-operatively, all patients underwent biplanar radiographic imaging of the hip, including a supine anteroposterior pelvic view in the neutral position and a cross-table lateral view [16]. Hip CT scans, extending from the anterior superior iliac spine to the distal femoral condyles through the knee joint, were obtained using a 320-row multidetector helical CT scanner (detector beam collimation: 80 × 0.5 mm2, 40 mm; slice thickness: 1.00 mm; slice interval: 1.00 mm; Aquilion ONE; Toshiba Medical Healthcare, Tochigi, Japan) [13].
Measurement of stem alignment
Stem alignment parameters were evaluated according to the recommendations of the International Society of Biomechanics [17]. The anteversion of the femoral stem was assessed using axial CT scans obtained in the supine position at 3 months post-operatively, covering the region from the pelvis to the distal femur [18]. Measurements of femoral offset [19], and coronal and sagittal stem angles [20] were obtained according to the descriptions illustrated in Fig. 1. Measurement reliability was excellent for all stem alignment parameters (Table S1).
Fig. 1.

Methods for measuring the alignment parameters of the femoral stem. a Stem anteversion was measured using axial computed tomography scans as the angle between the posterior condylar line (black line) and the axis of the stem neck (blue line). b Femoral offset was measured as the perpendicular distance from the centre of the femoral head to the axis of the femoral shaft on coronal radiographs; the femoral shaft axis is shown in black and the perpendicular offset distance in blue. c The coronal stem angle was measured as the angle between the femoral anatomical axis (black line) and femoral stem axis (blue line) on coronal radiographs. d The sagittal stem angle was measured as the angle between the femoral anatomical axis (black line) and femoral stem axis (blue line) using sagittal radiographs
Normal ranges were defined for each parameter based on previous studies and the clinically acceptable range of measurement or biomechanical influence previously reported [4, 13, 21–23]: stem anteversion, 10°–40° [13]; change in stem anteversion (post-operative minus pre-operative), ± 5° (positive for increased anteversion, negative for decreased) [23]; femoral offset, difference between the operated and contralateral side ± 3 mm [21]; coronal angle, ± 3° (positive for varus, negative for valgus) [13]; and sagittal angle, ± 3° (positive for flexion, negative for extension) [4]. These thresholds were based on previously published studies and represent clinically accepted ranges used to define appropriate implant alignment in total hip arthroplasty, enabling clinically meaningful group comparisons. For comparison of post-operative outcomes, patients were then classified into the Normal or Low&High groups based on whether the measured values fell within or outside the normal range.
Leg length discrepancy (LLD) was measured on an anteroposterior pelvic radiograph using the vertical distance from the inter-teardrop line to the most prominent aspect of the lesser trochanter on each side, according to a previously described method [24]. LLD was calculated as the difference between the operated and contralateral sides (positive values indicating lengthening and negative values indicating shortening).
Patient-reported outcomes
Clinical evaluation was conducted using the Japanese Orthopaedic Association Hip Disease Evaluation Questionnaire (JHEQ), a PROM used to assess the clinical condition of patients with hip disorders. The total JHEQ score ranges from 0 (worst) to 84 (best) points and includes three subscales: pain (0–28 points), movement (0–28 points), and mental (0–28 points). For each subscale, 0 points represents the worst outcome and 28 points represents the best [25]. Overall patient satisfaction was assessed using a visual analogue scale (VAS) for satisfaction, with 0 points representing the worst possible satisfaction and 100 points representing the best [26]. The JHEQ and VAS satisfaction were administered pre-operatively and at one year post-operatively.
Statistical analysis
We used the Mann–Whitney U test to compare each parameter between the two stem-alignment groups. For stem anteversion change, stratified into three groups, we used the Kruskal–Wallis followed by Dunn–Bonferroni post hoc tests. Categorical variables were compared using the chi-squared test. We performed logistic regression analysis to identify independent stem alignment factors associated with higher post-operative patient satisfaction scores assessed using the VAS satisfaction. Patient satisfaction was treated as a binary outcome, with a VAS score of ≥ 75 defined as satisfied (1) and a score of < 75 defined as unsatisfied (0), according to a previous study [27]. Odds ratios with 95% confidence intervals were calculated to estimate the strength of the association between stem alignment parameters and patient satisfaction. The explanatory variables included stem anteversion, change in stem anteversion, femoral offset difference, coronal and sagittal stem angles, and LLD to adjust for the potential influence of limb length restoration on patient satisfaction. Continuous variables are presented as mean ± standard deviation (range). Statistical significance was set at p < 0.05. Statistical analyses were performed using GraphPad Prism 10.0 (San Diego, CA, USA).
Results
Two hundred patients (205 hips) underwent primary THA using one of three CAE systems for cup placement (Fig. 2). Patient demographics, complications, and stem alignment parameters are summarised in Table 1. The mean age was 67.3 ± 9.8 years; 45 male and 160 female hips. The primary diagnoses were osteoarthritis (165 hips) and osteonecrosis of the femoral head (40 hips). The average change in stem anteversion was + 1.5° ± 6.7°, indicating a slight post-operative increase. The distribution of changes in stem anteversion is shown in Supplementary Fig. S1. Other stem alignment parameters are summarized in Table 1. No post-operative complications, including dislocation, infection, fracture, readmission, or revision surgery were observed.
Fig. 2.

Flowchart of patient screening and inclusion in the study. THA, total hip arthroplasty; PROMs, patient-reported outcome measures; CAE, computer-assisted equipment
Table 1.
Characteristics of patients and clinical outcomes
| Variable | Categorisation/Unit of measure | Value |
|---|---|---|
| Patient | (hips) | 205 |
| Age | (years) | 67.3 ± 9.8 (44 to 88) |
| Sex | (Male/Female) (n) | 45/160 |
| Primary disease | (OA/ON) (n) | 165/40 |
| Body mass index | (kg/m2) | 23.3 ± 3.5 (16.7 to 30.3) |
| Post-operative complications | ||
| Dislocation | (n) | 0 |
| Infection | (n) | 0 |
| Fracture | (n) | 0 |
| Readmission | (n) | 0 |
| Revision | (n) | 0 |
| Radiographic parameters | ||
| Stem anteversion | (°) | 23.0 ± 10.7 (2.0 to 56.6) |
| Change in stem anteversion | (°) | 1.5 ± 6.7 (−19.5 to 22.3) |
| Femoral offset | (mm) | 1.1 ± 3.1 (−6.8 to 9.5) |
| Coronal angle | (°) | 0.7 ± 1.5 (−5.0 to 4.5) |
| Sagittal angle | (°) | 2.8 ± 1.9 (−4.5 to 7.8) |
| Leg length discrepancy | (mm) | 0.0 ± 4.3 (−6.8 to 6.2) |
Stem anteversion: Angle between the stem neck axis and posterior femoral condylar line on axial computed tomography scans (°)
Change in stem anteversion: Post-operative minus pre-operative stem anteversion (°)
Femoral offset: Perpendicular distance from the centre of the femoral head to the femoral shaft axis, expressed as the difference between the operated and contralateral sides (mm)
Coronal stem angle: Angle between the femoral stem axis and femoral anatomical axis on coronal radiographs (°)
Sagittal stem angle: Angle between the femoral stem axis and femoral anatomical axis on sagittal radiographs (°)
Leg length discrepancy: Difference in the vertical distance from the inter-teardrop line to the most prominent aspect of the lesser trochanter between the operated and contralateral sides on anteroposterior pelvic radiographs (mm)
Data are shown as mean ± standard deviation (range)
OA, osteoarthritis; ON, osteonecrosis of the femoral head
PROMs significantly improved after surgery (Table S2). Patient satisfaction (VAS) increased from 19.2 ± 17.0 points pre-operatively to 90.7 ± 14.9 points at 1 year post-operatively. The total JHEQ score also improved significantly from before to after surgery, with all subscales also showing significant post-operative improvements.
Baseline characteristics of Normal vs Low&High groups showed no significant differences in age, sex, body mass index, or primary diagnosis for any alignment parameter (Table 2).
Table 2.
Characteristics of patients according to stem alignment
| Variable | Categorisation/Unit of measure | Normal group | Low&High group | p-Value |
|---|---|---|---|---|
| Stem anteversion | ||||
| Patient | (hips) | 176 | 29 | |
| Age | (years) | 67.3 ± 9.8 (44 to 88) | 67.5 ± 9.4 (44 to 88) | 0.908 |
| Sex | (Male/Female) (n) | 35/141 | 10/19 | 0.149 |
| Primary disease | (OA/ON) (n) | 144/32 | 21/8 | 0.391 |
| Body mass index | (kg/m2) | 23.2 ± 3.2 (16.7 to 30.3) | 24.0 ± 5.1 (18.2 to 30.1) | 0.840 |
| Change in stem anteversion | ||||
| Patient | (hips) | 133 | 72 | |
| Age | (years) | 66.1 ± 9.6 (44 to 88) | 69.5 ± 9.8 (44 to 88) | 0.069 |
| Sex | (Male/Female) (n) | 28/105 | 17/55 | 0.590 |
| Primary disease | (OA/ON) (n) | 109/24 | 56/16 | 0.454 |
| Body mass index | (kg/m2) | 23.0 ± 2.9 (18.2 to 30.3) | 22.9 ± 3.7 (16.7 to 29.6) | 0.716 |
| Femoral offset | ||||
| Patient | (hips) | 155 | 50 | |
| Age | (years) | 67.9 ± 10.0 (44 to 88) | 65.5 ± 8.8 (45 to 85) | 0.147 |
| Sex | (Male/Female) (n) | 40/115 | 5/45 | 0.079 |
| Primary disease | (OA/ON) (n) | 122/33 | 43/7 | 0.396 |
| Body mass index | (kg/m2) | 23.4 ± 3.5 (16.7 to 30.3) | 22.9 ± 3.5 (18.4 to 28.5) | 0.706 |
| Coronal angle | ||||
| Patient | (hips) | 190 | 15 | |
| Age | (years) | 67.1 ± 9.7 (44 to 88) | 70.1 ± 10.8 (51 to 85) | 0.256 |
| Sex | (Male/Female) (n) | 39/151 | 6/9 | 0.140 |
| Primary disease | (OA/ON) (n) | 152/38 | 13/2 | 0.788 |
| Body mass index | (kg/m2) | 23.9 ± 3.3 (16.7 to 30.3) | 26.8 ± 2.8 (24.8 to 30.3) | 0.059 |
| Sagittal angle | ||||
| Patient | (hips) | 130 | 75 | |
| Age | (years) | 68.0 ± 9.2 (45 to 87) | 66.2 ± 10.6 (44 to 88) | 0.247 |
| Sex | (Male/Female) (n) | 26/104 | 19/56 | 0.414 |
| Primary disease | (OA/ON) (n) | 106/24 | 59/16 | 0.597 |
| Body mass index | (kg/m2) | 23.2 ± 3.6 (16.7 to 30.3) | 23.6 ± 3.3 (18.2 to 29.6) | 0.662 |
Stem anteversion: Angle between the stem neck axis and the posterior femoral condylar line on axial compute tomography scans (°)
Change in stem anteversion: Post-operative minus pre-operative stem anteversion (°)
Femoral offset: Perpendicular distance from the centre of the femoral head to the femoral shaft axis, expressed as the difference between the operated and contralateral sides (mm)
Coronal stem angle: Angle between the femoral stem axis and femoral anatomical axis on coronal radiographs (°)
Sagittal stem angle: Angle between the femoral stem axis and femoral anatomical axis on sagittal radiographs (°)
Definitions for the Normal group: Stem anteversion, 10°–40°; change in stem anteversion, − 5° to + 5°; femoral offset difference, − 3 to + 3 mm; coronal stem angle, − 3° to + 3° (positive = varus); sagittal stem angle, − 3° to + 3° (positive = flexion)
The Low&High group included patients with values below or above the defined normal range
Data are shown as mean ± standard deviation (range)
OA, osteoarthritis; ON, osteonecrosis of the femoral head
Comparisons between the Normal and Low&High groups for each stem alignment parameter are presented in Table 3. All evaluated PROMs for stem anteversion and change in stem anteversion were significantly higher in the Normal group than in the Low&High group (all p < 0.05). For femoral offset, no significant between-group differences were observed for any PROMs. Regarding the coronal stem angle, patient satisfaction, total JHEQ score, and movement scores were significantly higher in the Normal group. For the sagittal stem angle, all PROMs were significantly higher in the Normal group (all p < 0.05). As an exploratory analysis, post hoc power analysis was conducted based on the observed differences in the patient VAS satisfaction scores between the Normal and Low&High groups for change in stem anteversion and sagittal stem angle. The mean VAS scores were 93.8 ± 12.2 points versus 84.8 ± 17.7 for stem anteversion change, and 93.6 ± 13.1 points versus 85.2 ± 16.7 points for sagittal stem angle. Using these values and the respective sample sizes, the calculated effect sizes were f = 0.314 for stem anteversion change and f = 0.289 for sagittal stem angle, yielding post hoc statistical powers of 0.970 and 0.951, respectively. These results indicate that the study had sufficient power to detect significant differences in patient satisfaction for these parameters.
Table 3.
Patient-reported outcome measures according to stem alignment
| Scale according to parameter | Unit of measure | Normal group | Low&High group | p-Value |
|---|---|---|---|---|
| Stem anteversion | ||||
| Patient satisfaction (VAS) score | (pts) | 91.5 ± 14.6 (70 to 100) | 86.3 ± 15.8 (50 to 100) | 0.037 |
| Total JHEQ score | (pts) | 69.9 ± 12.2 (50 to 84) | 58.2 ± 11.9 (46 to 82) | < 0.001 |
| Pain score | (pts) | 25.5 ± 3.8 (22 to 28) | 23.2 ± 4.8 (20 to 28) | 0.016 |
| Movement score | (pts) | 20.5 ± 5.9 (11 to 28) | 14.8 ± 6.3 (4 to 26) | < 0.001 |
| Mental score | (pts) | 23.9 ± 5.1 (13 to 28) | 20.2 ± 5.8 (11 to 28) | 0.001 |
| Change in stem anteversion | ||||
| Patient satisfaction (VAS) score | (pts) | 93.8 ± 12.2 (70 to 100) | 84.8 ± 17.7 (50 to 100) | < 0.001 |
| Total JHEQ score | (pts) | 73.1 ± 10.9 (51 to 84) | 58.8 ± 11.0 (46 to 84) | < 0.001 |
| Pain score | (pts) | 26.4 ± 3.2 (22 to 28) | 22.8 ± 4.4 (20 to 28) | < 0.001 |
| Movement score | (pts) | 21.6 ± 5.7 (12 to 28) | 16.1 ± 5.8 (4 to 28) | < 0.001 |
| Mental score | (pts) | 25.1 ± 4.2 (14 to 28) | 20.0 ± 5.6 (11 to 28) | < 0.001 |
| Femoral offset | ||||
| Patient satisfaction (VAS) score | (pts) | 91.2 ± 14.5 (60 to 100) | 89.3 ± 16.1 (50 to 100) | 0.510 |
| Total JHEQ score | (pts) | 68.4 ± 13.1 (52 to 84) | 67.5 ± 12.0 (46 to 84) | 0.471 |
| Pain score | (pts) | 25.2 ± 4.1 (20 to 28) | 25.1 ± 3.6 (21 to 28) | 0.555 |
| Movement score | (pts) | 19.6 ± 6.6 (12 to 28) | 19.8 ± 5.4 (4 to 28) | 0.813 |
| Mental score | (pts) | 23.6 ± 5.3 (13 to 28) | 22.6 ± 5.4 (11 to 28) | 0.189 |
| Coronal angle | ||||
| Patient satisfaction (VAS) score | (pts) | 91.5 ± 14.3 (60 to 100) | 81.5 ± 19.5 (50 to 100) | 0.017 |
| Total JHEQ score | (pts) | 69.0 ± 12.5 (51 to 84) | 58.6 ± 13.2 (46 to 77) | 0.005 |
| Pain score | (pts) | 25.3 ± 3.8 (21 to 28) | 23.4 ± 5.8 (20 to 28) | 0.173 |
| Movement score | (pts) | 20.1 ± 6.1 (12 to 28) | 14.5 ± 6.7 (4 to 23) | 0.003 |
| Mental score | (pts) | 23.6 ± 5.1 (11 to 28) | 20.8 ± 6.9 (11 to 28) | 0.126 |
| Sagittal angle | ||||
| Patient satisfaction (VAS) score | (pts) | 93.6 ± 13.1 (70 to 100) | 85.2 ± 16.7 (50 to 100) | < 0.001 |
| Total JHEQ score | (pts) | 72.5 ± 10.2 (54 to 84) | 61.2 ± 13.7 (46 to 84) | < 0.001 |
| Pain score | (pts) | 26.2 ± 3.0 (22 to 28) | 23.5 ± 4.8 (20 to 28) | < 0.001 |
| Movement score | (pts) | 21.1 ± 5.7 (12 to 28) | 17.3 ± 6.5 (4 to 28) | < 0.001 |
| Mental score | (pts) | 25.1 ± 4.0 (14 to 28) | 20.4 ± 5.9 (11 to 28) | < 0.001 |
Stem anteversion: Angle between the stem neck axis and the posterior femoral condylar line on axial computed tomography scans (°)
Change in stem anteversion: Post-operative minus pre-operative stem anteversion (°)
Femoral offset: Perpendicular distance from the centre of the femoral head to the femoral shaft axis, expressed as the difference between the operated and contralateral sides (mm)
Coronal stem angle: Angle between the femoral stem axis and femoral anatomical axis on coronal radiographs (°)
Sagittal stem angle: Angle between the femoral stem axis and femoral anatomical axis on sagittal radiographs (°)
Normal group definitions: Stem anteversion, 10°–40°; change in stem anteversion, − 5° to + 5°; femoral offset difference, − 3 to + 3 mm; coronal stem angle, − 3° to + 3° (positive = varus); sagittal stem angle, − 3° to + 3° (positive = flexion)
The Low&High group included patients with values below or above the defined normal range
Data are shown as mean ± standard deviation (range)
JHEQ, Japanese Orthopaedic Association Hip Disease Evaluation Questionnaire; pts, points; VAS, visual analogue scale
Using multivariate logistic regression analysis, change in stem anteversion and sagittal stem angle were identified as independent factors positively associated with patient satisfaction (Table 4), whereas LLD was not significantly associated with postoperative patient satisfaction (p = 0.973). The model showed good fit (Hosmer–Lemeshow test, p = 0.618) and discrimination (receiver operating characteristic curve analysis, area under the receiver operating characteristic curve = 0.793; p < 0.001), indicating good predictive performance.
Table 4.
Multivariate logistic regression analysis of factors associated with patient satisfaction
| Factor | Odds ratio | 95% CI | p-Value |
|---|---|---|---|
| Stem anteversion | 0.24 | 0.03—1.17 | 0.080 |
| Change in stem anteversion | 8.43 | 2.72—32.26 | < 0.001 |
| Femoral offset | 0.87 | 0.24—2.85 | 0.825 |
| Coronal angle | 0.85 | 0.04—7.14 | 0.892 |
| Sagittal angle | 3.36 | 1.11—10.81 | 0.032 |
| Leg length discrepancy | 1.00 | 0.87–1.15 | 0.973 |
Stem anteversion: Angle between the stem neck axis and posterior femoral condylar line on axial computed tomography scans (°)
Change in stem anteversion: Post-operative minus pre-operative stem anteversion (°)
Femoral offset: Perpendicular distance from the centre of the femoral head to the femoral shaft axis, expressed as the difference between the operated and contralateral sides (mm)
Coronal stem angle: Angle between the femoral stem axis and femoral anatomical axis on coronal radiographs (°)
Sagittal stem angle: Angle between the femoral stem axis and femoral anatomical axis on sagittal radiographs (°)
Leg length discrepancy: Difference in the vertical distance from the inter-teardrop line to the most prominent aspect of the lesser trochanter between the operated and contralateral sides on anteroposterior pelvic radiographs (mm)
Patient satisfaction was assessed using a visual analogue scale and treated as a binary outcome (satisfied = 1, unsatisfied = 0)
CI confidence interval
Exploratory analysis stratified patients by native femoral anteversion; post-operative changes in stem anteversion were further categorised. Across all pre-operative anteversion groups, patients with minimal changes in stem anteversion demonstrated significantly higher PROMs (Table S3).
Discussion
We investigated the association between femoral stem alignment and post-operative patient satisfaction after THA performed using CAE for accurate acetabular cup placement. The main findings were that (1) smaller changes in stem anteversion from pre- to post-operative measurements and (2) appropriate sagittal stem alignment were significantly associated with higher patient satisfaction and better PROMs. Using multivariate analysis, we identified these two parameters as independent predictors of patient satisfaction. These results highlight the importance of achieving accurate femoral stem alignment, even when acetabular cup placement is optimised using CAE.
Previous studies have primarily focused on the acetabular component, demonstrating that accurate cup placement reduces complications such as dislocation and impingement [2, 3]. However, Shoji et al. reported that implant-related factors beyond cup positioning influence PROMs [13]. Our observation that patient satisfaction was more strongly related to femoral stem alignment than to other perioperative factors supports this view. Our findings further demonstrated that the change in stem anteversion from pre- to post-operative measurement, rather than the absolute anteversion value alone, was a particularly important determinant of post-operative satisfaction. This finding is consistent with previous reports indicating that native femoral version is predictive of achieved stem version in cementless THA, highlighting the importance of preserving native femoral anatomy when possible [28].
From a biomechanical perspective, excessive deviation of stem anteversion may alter the combined anteversion of the hip joint, potentially leading to impingement or instability [29, 30]. However, recent studies have shown that dislocation may still occur even when target combined anteversion is achieved, indicating that patient-specific femoral morphology and stem alignment remain important considerations [31]. Similarly, inappropriate sagittal stem alignment may affect soft tissue tension, hip flexion mechanics, and load transfer through the femur, possibly contributing to patient discomfort or reduced satisfaction [4, 13]. Our findings suggest that maintaining stem anteversion close to the native femoral version and minimising sagittal malalignment are crucial for restoring normal joint kinematics and achieving favourable PROMs. However, accurately controlling stem version in cementless THA remains challenging, as stem orientation may be influenced by implant design and patient-specific femoral morphology [32]. Within the context of our findings, maintaining stem anteversion within ± 5° of the native femoral version and avoiding sagittal malalignment beyond ± 3° may represent practical alignment targets for optimising early patient satisfaction.
Femoral offset plays an important role in improving abductor muscle function and stabilising gait [33]. However, we found no significant between-group differences in PROMs according to femoral offset. Similarly, LLD was not independently associated with patient satisfaction after adjustment for stem alignment parameters. One possible reason for these consistent findings is that all procedures were performed using the anterolateral supine approach, in which fluoroscopy is used intra-operatively to optimise both femoral offset and LLD [34]. In addition, the use of CAE and fluoroscopic guidance may have reduced variability in post-operative LLD across patients, thereby limiting the presence of clinically meaningful discrepancies large enough to significantly influence PROMs [35].
Although we found no significant associations between femoral offset and early post-operative PROMs, long-term consequences of initial component alignment, including the accuracy of cup placement and orientation of the femoral stem (anteversion, sagittal angle, and offset), could still substantially influence joint function and the risk of revision, effects not captured by early PROMs [36].
Therefore, long-term follow-up studies focusing on stem alignment are warranted. PROMs may also be influenced by factors such as post-operative rehabilitation and patient expectations [37–39].
In this study, CAE was used to assist with placing the acetabular cup to ensure consistent positioning; however, no navigation or robotic systems were used to implant the femoral stem. Despite the use of CAE for acetabular cup placement, femoral stem positioning remained dependent on surgical technique and femoral anatomy. Future studies should focus on pre-operative prediction of stem alignment deviation and on surgical strategies to achieve stem orientation within an optimal range. Considering that the stem alignment parameters were significant predictors of patient satisfaction, future advancements in robotic-assisted THA systems, such as the Mako platform, may help improve the accuracy of both cup and stem alignment [40]. This may further enhance patient-specific hip reconstruction and post-operative satisfaction.
This study had several limitations. First, it was a single-centre, retrospective study, which may have limited the external validity of the findings and introduced the potential for selection bias. Second, all procedures were performed by experienced surgeons via the anterolateral supine approach, and patients with spinal disease were excluded to minimise the influence of spino-pelvic imbalance on PROMs. Therefore, the generalisability of our findings to less experienced surgeons, different surgical approaches, or patients with hip–spine syndrome may be limited [41, 42]. Third, detailed intra-operative data, such as subtle adjustments during stem trialling or variations in bone quality, were unavailable; this may have influenced post-operative outcomes. In addition, stem anteversion was measured using two-dimensional CT slices rather than three-dimensional reconstruction techniques, which may be influenced by scanner settings and measurement methodology [43]. Finally, the follow-up period was limited to early post-operative outcomes, and the one-year results may not reflect long-term mechanical stability or late complications such as stress shielding or implant loosening related to stem alignment [44, 45]. Despite these limitations, the findings provide valuable insights into the association between femoral stem alignment and early PROMs.
In conclusion, in this study, we demonstrated that a minimal change in stem anteversion and appropriate sagittal stem alignment were independently associated with greater post-operative patient satisfaction following THA. These findings emphasise the importance of precise femoral alignment to complement accurate acetabular placement, suggesting that future technologies enabling improved control of stem positioning may further enhance patient-reported outcomes after THA.
Supplementary Information
Below is the link to the electronic supplementary material.
Acknowledgements
We would like to thank Editage (www.editage.com) for English language editing.
Author contribution
All authors contributed to the study conception and design. Material preparation, data collection, and analysis were performed by Hiroki Kaneta, Takeshi Shoji, Shinichi Ueki, Hiroyuki Morita, Yosuke Kozuma, Naoto Nakayama, and Shuhei Yamasaki. The first draft of the manuscript was written by Hiroki Kaneta and all authors commented on previous versions of the manuscript. All authors read and approved of the final manuscript.
Funding
Open Access funding provided by Hiroshima University.
Data availability
The datasets during and/or analyzed during the current study available from the corresponding author on reasonable request.
Declarations
Competing interests
The authors declare no competing interests.
Footnotes
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Contributor Information
Hiroki Kaneta, Email: kyokui090014@gmail.com.
Takeshi Shoji, Email: readyma@hiroshima-u.ac.jp.
References
- 1.Learmonth ID, Young C, Rorabeck C (2007) The operation of the century: total hip replacement. Lancet 370(9597):1508–1519. 10.1016/S0140-6736(07)60457-7 [DOI] [PubMed] [Google Scholar]
- 2.Lewinnek GE, Lewis JL, Tarr R, Compere CL, Zimmerman JR (1978) Dislocations after total hip-replacement arthroplasties. J Bone Joint Surg Am 60(2):217–220. 10.2106/00004623-197860020-00014 [DOI] [PubMed] [Google Scholar]
- 3.Shoji T, Yamasaki T, Izumi S, Murakami H, Mifuji K, Sawa M, Yasunaga Y, Adachi N, Ochi M (2017) Factors affecting the potential for posterior bony impingement after total hip arthroplasty. Bone Joint J 99-B(9):1140–1146. 10.1302/0301-620X.99B9.BJJ-2016-1078.R2 [DOI] [PubMed] [Google Scholar]
- 4.Yoon JY, Seo WY, Kim HJ, Yoo JJ (2022) The relationship between femoral stem tilt and stem length in total hip arthroplasty: a retrospective case-control study. Clin Orthop Surg 14(2):184–190. 10.4055/cios21042 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Ando W, Takao M, Hamada H, Uemura K, Sugano N (2021) Comparison of the accuracy of the cup position and orientation in total hip arthroplasty for osteoarthritis secondary to developmental dysplasia of the hip between the Mako robotic arm-assisted system and computed tomography-based navigation. Int Orthop 45(7):1719–1725. 10.1007/s00264-021-05015-3 [DOI] [PubMed] [Google Scholar]
- 6.Hayashi S, Kuroda Y, Nakano N, Matsumoto T, Kamenaga T, Tsubosaka M, Kuroda R (2024) Accuracy of portable navigation during THA in patients with severe developmental dysplasia of hip. Arch Orthop Trauma Surg 144(5):2429–2435. 10.1007/s00402-024-05338-x [DOI] [PubMed] [Google Scholar]
- 7.Emara AK, Zhou G, Klika AK, Koroukian SM, Schiltz NK, Higuera-Rueda CA, Molloy RM, Piuzzi NS (2021) Is there increased value in robotic arm-assisted total hip arthroplasty?: a nationwide outcomes, trends, and projections analysis of 4,699,894 cases. Bone Joint J 103-B(9):1488–1496. 10.1302/0301-620X.103B9.BJJ-2020-2411.R1 [DOI] [PubMed] [Google Scholar]
- 8.Sato K, Sato A, Okuda N, Masaaki M, Koga H (2023) A propensity score-matched comparison between Mako robotic arm-assisted system and conventional technique in total hip arthroplasty for patients with osteoarthritis secondary to developmental dysplasia of the hip. Arch Orthop Trauma Surg 143(5):2755–2761. 10.1007/s00402-022-04524-z [DOI] [PubMed] [Google Scholar]
- 9.Kaneta H, Shoji T, Ueki S, Morita H, Kozuma Y, Adachi N (2025) Evaluation of cup placement accuracy in computer assisted total hip arthroplasty. Arch Orthop Trauma Surg 145(1):191. 10.1007/s00402-025-05797-w [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Haversath M, Busch A, Jager M, Tassemeier T, Brandenburger D, Serong S (2019) The “critical trochanter angle”: a predictor for stem alignment in total hip arthroplasty. J Orthop Surg Res 14(1):165. 10.1186/s13018-019-1206-x [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Chen Z, Li B, Chen K, Feng J, Wang Y, Liu Z, He C (2021) Malalignment and distal contact of short tapered stems could be associated with postoperative thigh pain in primary total hip arthroplasty. J Orthop Surg Res 16(1):67. 10.1186/s13018-021-02215-w [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Shoji T, Yasunaga Y, Yamasaki T, Izumi S, Hachisuka S, Ochi M (2015) Low femoral antetorsion and total hip arthroplasty: a risk factor. Int Orthop 39(1):7–12. 10.1007/s00264-014-2452-5 [DOI] [PubMed] [Google Scholar]
- 13.Shoji T, Inoue T, Kato Y, Fujiwara Y, Sumii J, Shozen H, Adachi N (2022) Associations between implant alignment or position and patient-reported outcomes after total hip arthroplasty. Clin Biomech (Bristol, Avon) 97:105701. 10.1016/j.clinbiomech.2022.105701 [DOI] [PubMed] [Google Scholar]
- 14.Miki H, Yamanashi W, Nishii T, Sato Y, Yoshikawa H, Sugano N (2007) Anatomic hip range of motion after implantation during total hip arthroplasty as measured by a navigation system. J Arthroplasty 22(7):946–952. 10.1016/j.arth.2007.02.004 [DOI] [PubMed] [Google Scholar]
- 15.Widmer KH, Zurfluh B (2004) Compliant positioning of total hip components for optimal range of motion. J Orthop Res 22(4):815–821. 10.1016/j.orthres.2003.11.001 [DOI] [PubMed] [Google Scholar]
- 16.Kusma M, Bachelier F, Schneider G, Dienst M (2006) Radiological examination of the hip. Analysis of standard projections. Orthopade 35(1):16–21. 10.1007/s00132-005-0889-3 [DOI] [PubMed] [Google Scholar]
- 17.Wu G, Siegler S, Allard P, Kirtley C, Leardini A, Rosenbaum D, Whittle M, D’Lima DD, Cristofolini L, Witte H, Schmid O, Stokes I, Standardization and Terminology Committee of the International Society of Biomechanics (2002) ISB recommendation on definitions of joint coordinate system of various joints for the reporting of human joint motion–part I: ankle, hip, and spine. International Society of Biomechanics. J Biomech 35(4):543–548. 10.1016/s0021-9290(01)00222-6 [DOI] [PubMed] [Google Scholar]
- 18.Murphy SB, Simon SR, Kijewski PK, Wilkinson RH, Griscom NT (1987) Femoral anteversion. J Bone Joint Surg Am 69(8):1169–1176. 10.2106/00004623-198769080-00010 [DOI] [PubMed] [Google Scholar]
- 19.Oommen AT (2024) Offset restoration in total hip arthroplasty: important: a current review. World J Orthop 15(8):696–703. 10.5312/wjo.v15.i8.696 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Khalily C, Lester DK (2002) Results of a tapered cementless femoral stem implanted in varus. J Arthroplasty 17(4):463–466. 10.1054/arth.2002.32171 [DOI] [PubMed] [Google Scholar]
- 21.Hu X, Zheng N, Chen Y, Dai K, Dimitriou D, Li H, Tsai TY (2021) Optimizing the femoral offset for restoring physiological hip muscle function in patients with total hip arthroplasty. Front Bioeng Biotechnol 9:645019. 10.3389/fbioe.2021.645019 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Mahmood SS, Mukka SS, Crnalic S, Wretenberg P, Sayed-Noor AS (2016) Association between changes in global femoral offset after total hip arthroplasty and function, quality of life, and abductor muscle strength. A prospective cohort study of 222 patients. Acta Orthop 87(1):36–41. 10.3109/17453674.2015.1091955 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Satilmis AB, Cengiz T, Ulker A, Mutlu T (2025) The impact of femoral anteversion correction on clinical outcomes in total hip arthroplasty for adult developmental dysplasia of the hip. J Clin Med 14(9). 10.3390/jcm14093207 [DOI] [PMC free article] [PubMed]
- 24.Meermans G, Malik A, Witt J, Haddad F (2011) Preoperative radiographic assessment of limb-length discrepancy in total hip arthroplasty. Clin Orthop Relat Res 469(6):1677–1682. 10.1007/s11999-010-1588-x [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Matsumoto T, Kaneuji A, Hiejima Y, Sugiyama H, Akiyama H, Atsumi T, Ishii M, Izumi K, Ichiseki T, Ito H, Okawa T, Ohzono K, Otsuka H, Kishida S, Kobayashi S, Sawaguchi T, Sugano N, Nakajima I, Nakamura S, Hasegawa Y, Fukuda K, Fujii G, Mawatari T, Mori S, Yasunaga Y, Yamaguchi M (2012) Japanese Orthopaedic Association Hip Disease Evaluation Questionnaire (JHEQ): a patient-based evaluation tool for hip-joint disease. The Subcommittee on Hip Disease Evaluation of the Clinical Outcome Committee of the Japanese Orthopaedic Association. J Orthop Sci 17(1):25–38. 10.1007/s00776-011-0166-8 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Chandrasekaran S, Gui C, Walsh JP, Lodhia P, Suarez-Ahedo C, Domb BG (2017) Correlation between changes in visual analog scale and patient-reported outcome scores and patient satisfaction after hip arthroscopic surgery. Orthop J Sports Med 5(9):2325967117724772. 10.1177/2325967117724772 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Brokelman RB, Haverkamp D, van Loon C, Hol A, van Kampen A, Veth R (2012) The validation of the visual analogue scale for patient satisfaction after total hip arthroplasty. Eur Orthop Traumatol 3(2):101–105. 10.1007/s12570-012-0100-3 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Hirata M, Nakashima Y, Itokawa T, Ohishi M, Sato T, Akiyama M, Hara D, Iwamoto Y (2014) Influencing factors for the increased stem version compared to the native femur in cementless total hip arthroplasty. Int Orthop 38(7):1341–1346. 10.1007/s00264-014-2289-y [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Dennis DA, Bryman JA, Smith GH, Pierrepont JW, Jennings JM, Rajpura A, Plaskos C (2025) How do changes in femoral anteversion impact femoral rotation and anterior offset after total hip arthroplasty? J Arthroplasty 40(1):152-159 e151. 10.1016/j.arth.2024.07.027 [DOI] [PubMed] [Google Scholar]
- 30.Qiu J, Ke X, Chen S, Zhao L, Wu F, Yang G, Zhang L (2020) Risk factors for iliopsoas impingement after total hip arthroplasty using a collared femoral prosthesis. J Orthop Surg Res 15(1):267. 10.1186/s13018-020-01787-3 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Hernandez A, Lakhani K, Nunez JH, Mimendia I, Pons A, Barro V (2021) Can we trust combined anteversion and Lewinnek safe zone to avoid hip prosthesis dislocation? J Clin Orthop Trauma 21:101562. 10.1016/j.jcot.2021.101562 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Nakahara I, Kyo T, Kuroda Y, Miki H (2021) Does difference in stem design affect accuracy of stem alignment in total hip arthroplasty with a CT-based navigation system? J Artif Organs 24(1):74–81. 10.1007/s10047-020-01196-7 [DOI] [PubMed] [Google Scholar]
- 33.Shapira J, Chen SL, Rosinsky PJ, Maldonado DR, Meghpara M, Lall AC, Domb BG (2020) The effect of postoperative femoral offset on outcomes after hip arthroplasty: a systematic review. J Orthop 22:5–11. 10.1016/j.jor.2020.03.034 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Gililland JM, Anderson LA, Boffeli SL, Pelt CE, Peters CL, Kubiak EN (2012) A fluoroscopic grid in supine total hip arthroplasty: improving cup position, limb length, and hip offset. J Arthroplasty 27(8 Suppl):111–116. 10.1016/j.arth.2012.03.027 [DOI] [PubMed] [Google Scholar]
- 35.Blum CL, Luzzi AJ, Frederick JS, Cooper HJ, Shah RP, Tatka J, Geller JA, Herndon CL (2024) Intraoperative fluoroscopy decreases magnitude and incidence of leg-length discrepancy following total hip arthroplasty. Arthroplast Today 30:101492. 10.1016/j.artd.2024.101492 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Sai Sathikumar A, Jacob G, Thomas AB, Varghese J, Menon V (2023) Acetabular cup positioning in primary routine total hip arthroplasty-a review of current concepts and technologies. Arthroplasty (Sunnyvale Calif) 5(1):59. 10.1186/s42836-023-00213-3 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Benedetti MG, Cavazzuti L, Amabile M, Tassinari E, Valente G, Zanotti G, Vaienti E, Orsini S, Mariani E, Taddei F (2021) Abductor muscle strengthening in THA patients operated with minimally-invasive anterolateral approach for developmental hip dysplasia. Hip Int 31(1):66–74. 10.1177/1120700019877174 [DOI] [PubMed] [Google Scholar]
- 38.Pinskiy M, Lubovsky O, Kalichman L (2021) The effect of a preoperative physical therapy education program on short-term outcomes of patients undergoing elective total hip arthroplasty: a controlled prospective clinical trial. Acta Orthop Traumatol Turc 55(4):306–310. 10.5152/j.aott.2021.20108 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Shao Z, Bi S (2022) Patient satisfaction after total hip arthroplasty: influencing factors. Front Surg 9:1043508. 10.3389/fsurg.2022.1043508 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Domb BG, Chandrasekaran S, Gui C, Yuen LC, Lodhia P, Suarez C (2017) Can stem version consistently correct native femoral version using robotic guidance in total hip arthroplasty? Surg Technol Int 31:389–395 [PubMed] [Google Scholar]
- 41.Peters RM, van Beers L, van Steenbergen LN, Wolkenfelt J, Ettema HB, Ten Have B, Rijk PC, Stevens M, Bulstra SK, Poolman RW, Zijlstra WP (2018) Similar superior patient-reported outcome measures for anterior and posterolateral approaches after total hip arthroplasty: postoperative patient-reported outcome measure improvement after 3 months in 12,774 primary total hip arthroplasties using the anterior, anterolateral, straight lateral, or posterolateral approach. J Arthroplasty 33(6):1786–1793. 10.1016/j.arth.2018.01.055 [DOI] [PubMed] [Google Scholar]
- 42.Suhardi VJ, Chiu YF, Sculco PK, Gonzalez Della Valle A (2021) Accuracy of acetabular cup placement positively correlates with level of training. Int Orthop 45(11):2797–2804. 10.1007/s00264-021-05165-4 [DOI] [PubMed] [Google Scholar]
- 43.Xiao H, Yibulayimu S, Zhao C, Sang Y, Chen Y, Ge Y, Sun Q, Ming Y, Bei M, Zhu G, Song Y, Wang Y, Wu X (2025) Rethinking femoral neck anteversion assessment: a novel automated 3D CT method compared to traditional manual techniques. BMC Musculoskelet Disord 26(1):466. 10.1186/s12891-025-08697-7 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44.Schmitz PP, van Susante JLC, Hol A, Brokelman R, van Loon CJM (2019) No decline in high patient satisfaction after total hip arthroplasty at long-term follow-up. Eur J Orthop Surg Traumatol 29(1):91–95. 10.1007/s00590-018-2243-6 [DOI] [PubMed] [Google Scholar]
- 45.Streit MR, Innmann MM, Merle C, Bruckner T, Aldinger PR, Gotterbarm T (2013) Long-term (20- to 25-year) results of an uncemented tapered titanium femoral component and factors affecting survivorship. Clin Orthop Relat Res 471(10):3262–3269. 10.1007/s11999-013-3033-4 [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
The datasets during and/or analyzed during the current study available from the corresponding author on reasonable request.