Multidimensional T1−T2 Relaxation Imaging In vivo on a Portable 0.064 T MRI Scanner

Ella Wilczynski 1, Kulam Najmudeen Magdoom 1,2,3, Alexandru V Avram 1, Nathan H Williamson 1,2,3, Dan Benjamini 4, Silvina G Horovitz 5, Peter J Basser 1
PMCID: PMC13622502  PMID: 42818277

Abstract

Purpose:

To demonstrate the feasibility of whole-brain multidimensional T1−T2 relaxation imaging on a portable 0.064 T MRI scanner.

Methods:

A two-dimensional inversion-recovery, fast spin-echo acquisition was used to jointly encode T1 and T2 relaxation. Joint relaxation distributions were reconstructed voxel-wise using marginal-distribution constrained optimization (MADCO). The approach was evaluated in a quantitative relaxation MRI phantom and subsequently applied in vivo in a healthy volunteer. Joint distributions, marginal distributions, distribution-derived measures, and statistical dependencies between the two relaxation dimensions were examined.

Results:

Phantom joint relaxation estimates showed good overall agreement with mono-exponential relaxation measurements and previously reported values at 0.064 T. The in vivo acquisition provided whole-brain coverage over a broad range of relaxation weightings and tissue contrasts. Reconstructed joint T1−T2 distributions showed spatially organized features across the relaxation space, including broad and overlapping relaxation components that were not fully represented by single-value relaxation maps or by either 1D marginal distribution alone. Selected regions of the joint relaxation space produced component maps with distinct spatial patterns. Statistical analysis further showed a dependence between the two relaxation dimensions, supporting the presence of added information in the joint distribution that is lost when treating T1 and T2 separately.

Conclusion:

Whole-brain multidimensional T1−T2 relaxation imaging is clinically feasible on a portable 0.064 T MRI scanner. Joint relaxation distributions provide information beyond single-value relaxation mapping and may support further developments of quantitative multidimensional imaging at ultra-low field.

Keywords: ultra-low-field, portable MRI, multidimensional, T1−T2 relaxation, quantitative, joint distribution

Introduction

Portable ultra-low-field (ULF) MRI scanners (i.e., operating below 0.1 T) are emerging as accessible alternatives to conventional MRI scanners. They have lower infrastructure requirements and cost and can operate in point-of-care and resource-limited environments (1; 2; 3). These permanent-magnet scanners do not require cryogenic cooling or conventional shielded MRI infrastructure and have been used for neuroimaging in intensive-care, emergency, mobile, and geographically distributed settings (4; 5; 6). Recent studies have also demonstrated their potential for evaluating a range of intracranial pathologies (7; 8). These advantages may be particularly important where access to conventional MRI remains limited. For example, the Ultra-low-field Neuroimaging In The Young (UNITY) initiative is deploying 0.064 T MRI across multiple sites in low- and middle-income countries (LMICs) to study early brain development (9). As access to portable ULF MRI expands, there is also a need to extend the quantitative information available from these scanners beyond structural imaging alone.

Quantitative MRI (qMRI) goes beyond anatomical imaging by estimating tissue properties such as T1 and T2, which can provide information related to tissue composition and microstructure. A range of approaches have been developed for quantitative relaxation mapping, including conventional relaxometry and more highly encoded methods such as MR Fingerprinting (10; 11). Because relaxation times vary with magnetic field strength, measurements obtained at conventional field strengths do not necessarily translate directly to ULF. Relaxation properties therefore need to be characterized specifically at these lower field strengths (12). At ULF, previous studies have demonstrated in vivo quantitative relaxation measurements in both adult (13; 14; 15) and neonatal populations (16). More recent work has also demonstrated repeatable low-field T1 mapping across sites and portable MRI scanners using accelerated acquisition and reconstruction approaches (17). These studies establish the feasibility and value of quantitative relaxation mapping at ULF, however, conventional relaxation maps typically represent each voxel by a single estimated T1 or T2 value. Such scalar measures cannot describe the heterogeneous relaxation environments that may coexist within a voxel.

Rather than estimating a single voxel-averaged value, relaxation spectroscopic MRI aims to characterize the distribution of MR-sensitive parameters within each voxel, including T2 (18), T1 (19), and mean diffusivity (MD) (20). These distributions can reveal multiple relaxation or microstructural environments that are not represented by a single scalar measurement. Multidimensional MRI (mdMRI) extends this concept by jointly encoding multiple MR-sensitive parameters and reconstructing their joint distribution within each voxel (21; 22). In particular, joint T1−T2 distributions describe how relaxation components are related across the two dimensions. This relationship is lost when T1 and T2 are considered independently (23). Multidimensional approaches have demonstrated sensitivity to tissue microstructure and compartmental composition (24; 25; 26; 27; 28; 29; 21), and have been applied to investigate placental dysfunction (30), astrogliosis (31), diffuse axonal injury (32), and diffusivity-anisotropy domains in vivo (33; 34; 35).

Translating multidimensional MRI to portable ULF scanners presents additional technical challenges. Multidimensional relaxation imaging requires sampling a large jointly encoded parameter space, resulting in long acquisition times and increased susceptibility to noise, motion artifacts, and reconstruction instability. These challenges are amplified at ULF by reduced signal-to-noise ratio (SNR), limited gradient performance, and practical constraints on scan duration. Although multidimensional imaging has been demonstrated at conventional field strengths, its application to portable ULF scanners remains largely unexplored. To our knowledge, voxel-wise joint T1−T2 relaxation distributions have not previously been reported in vivo on a portable ULF MRI scanner.

In this work, we present a multidimensional T1−T2 relaxation imaging framework for a portable 0.064 T MRI scanner using an inversion-recovery fast spin-echo (IR-FSE) acquisition. Joint T1−T2 distributions were reconstructed on a voxel-wise basis using marginal-distribution constrained optimization (MADCO) (36; 21) and evaluated first in a quantitative relaxation MRI phantom and subsequently in vivo in a healthy volunteer. We also examined T1−T2 distribution-derived measures, spatial patterns within the relaxation spectrum, and statistical dependence between the two relaxation dimensions. This proof-of-concept study establishes the feasibility of whole-brain multidimensional relaxation imaging at ULF and provides a basis for future studies addressing specific biological and clinical questions.

Theory

Multidimensional Relaxation Modeling

Multidimensional nuclear magnetic resonance (NMR) is well established in chemical spectroscopy, porous media NMR, and structural biology (37). Its translation to MRI has been more difficult because multidimensional relaxometry requires large amounts of data and long scan times. Approaches such as compressed sensing (38) and MADCO (39) have substantially reduced these data requirements. These developments have made voxel-wise multidimensional relaxation measurements more practical while maintaining spectral accuracy and precision.

The relaxation-weighted signal acquired in each voxel, S(TI,TE,TR), can be modeled as an integral over the joint distribution of the longitudinal and transverse magnetic relaxation rates, R1=1T1, and R2=1T2, respectively, weighted by their corresponding relaxation kernels:

S(TI,TE,TR)=∬pR1,R21−2ηe−R1TI+e−R1TRe−R2TEdR1dR2 [1]

The parameter η represents the apparent inversion efficiency. The form of Eq. 1 resembles a multidimensional Laplace transform. In practice, the joint distribution is obtained by treating the problem as a Fredholm integral equation of the first kind, which is discretized and solved by applying additional constraints (40).

For discrete relaxation rates R1,n and R2,n, let Sm=STIm,TEm,TRm denote the measured signal for the mth combination of TI,TE, and TR. The corresponding signal predicted by the discretized model is S0(Ψp)m:

Sm≈S0(Ψp)m, [2]
Ψmn=1−2ηe−R1,nTIm+e−R1,nTRme−R2,nTEm. [3]

Here, Ψmn is the m×n encoding kernel, and p is the discretized joint probability distribution whose nth element, pn, corresponds to the relaxation-rate pair R1,n,R2,n. The approximate equality accounts for the difference between the measured and predicted signals, which is minimized in the constrained optimization framework described below.

Constrained Estimation of the Joint Relaxation Distribution

Estimating multidimensional relaxation distributions is an ill-posed inverse problem, so some form of regularization and additional constraints are required. In MADCO, the solution is limited to positive relaxation rates with R1<R2. The joint distribution, pR1,R2, must also be nonnegative and normalized to one. Independently estimated one-dimensional marginal distributions further restrict the admissible joint R1−R2 parameter space (39).

The estimation of the joint distribution, pR1,R2, can be formulated as the constrained convex optimization problem

minpn,S0,κmaxSm−S0Ψmnpn22+λpn22,κ. [4]

In Eq. 4, the first term measures the difference between the measured and predicted signals. The ℓ2 term regularizes the joint distribution, with its contribution controlled by λ. The optimization is subject to the following physical, probabilistic, and marginal-distribution constraints:

pR1,n>R2,n=0,R1,n,R2,n>0,pn≥0,∑npn=1,Φi,klpl−pi,k′2≤κi,i=1,…,M. [5]

The first four constraints define a physically admissible, nonnegative, normalized joint probability distribution over the 2D relaxation-rate space. The final constraint requires agreement with the independently estimated marginal distributions within a tolerance κi. Here, Φi,kl maps the joint distribution to its ith marginal, and pi,k′ is the corresponding independently estimated marginal.

Methods

All experiments were performed on portable 0.064 T MRI scanners (Hyperfine SWOOP®, Guilford, CT, USA). A research agreement with the manufacturer allowed us to modify the pulse sequence timing and implement multidimensional contrast encoding. We used a two-dimensional IR-FSE acquisition to jointly encode longitudinal and transverse relaxation. The acquisition was repeated over multiple inversion time (TI) and repetition time (TR) combinations, with multiple echo times (TE) sampled within each FSE echo train. These relaxation-weighted images were then used to estimate the joint T1−T2 distributions. The acquisition framework is shown in Figure 1.

Figure 1:

Figure 1:

IR-FSE pulse sequence and multidimensional relaxation-encoding strategy. An adiabatic inversion pulse is followed after inversion time (TI) by a 90° excitation pulse and a train of 180° refocusing pulses separated by the echo spacing. Multiple echoes are acquired within each repetition, with each echo contributing one line in 3D k-space. The Lines Per Image (LPI) parameter determines the number of echoes contributing to each reconstructed echo image. The effective echo time (TE) is the interval between the 90° excitation pulse and the final echo contributing to a given echo image. Repetition time (TR) is defined between successive inversion pulses. The sequence block is repeated for multiple TI and TR combinations, and the full set of echo times is acquired for each combination to provide joint T1−T2 relaxation encoding.

Phantom Study

Data Acquisition

Phantom data were acquired using a commercially available quantitative MRI phantom (Model 137, CaliberMRI, Boulder, CO, USA), hereafter referred to as the UNITY phantom (41). The phantom contains tissue-mimicking compartments with varying concentrations of NiCl2 and MnCl2, covering a broad range of T1 and T2 values. It also includes an array of polyvinylpyrrolidone (PVP) compartments for diffusion calibration, geometric fiducials, and other quality-control features. A detailed description of the phantom design and characterization has been reported previously (41).

Data were acquired on a Hyperfine SWOOP portable 0.064 T MRI scanner (hardware version 1.8, software version 9.0.1). We used the two-dimensional IR-FSE acquisition described above and shown in Figure 1. The protocol included 32 inversion times ranging from 22 to 5000 ms and 80 echo images ranging from approximately 6.5 to 522 ms. The repetition time (TR) was 5700 ms. Images were acquired with an in-plane resolution of 2.0 × 2.0 mm2, a slice thickness of 4.17 mm, and a total acquisition time of approximately one week. A complete summary of the acquisition protocol is provided in Table 1.

Table 1:

Acquisition parameters for the phantom and in vivo multidimensional IR-FSE protocols.

Parameter Phantom In vivo
Hardware version 1.8 1.6
Software version 9.0.1 rc9.0.0 Beta1
Non-inversion scan Yes Yes (with extended 60 echoes)
TR (ms) 5700 1400
TI range (ms) 22 to 5000 18 to 1200
Number of TIs 32 9
Echo spacing (ms) 6.52 5.12
Echo images acquired 80 20
TE range (ms) 6.5 to 522 5.12 to 410
Lines per image 1 4
In-plane resolution (mm2) 2.0 × 2.0 2.5 × 2.5
Slice thickness (mm) 4 6
Scan time approximately one week approximately 100 min

Data Analysis

Phantom data were analyzed using mono-exponential and joint relaxation fitting. Each phantom compartment was designed to be chemically homogeneous, so we assumed a single relaxation component. T1 and T2 were estimated independently from the inversion recovery and multi-echo datasets using voxel-wise mono-exponential least-squares fitting:

S(TI)=S01−2e−TI/T1+e−TR/T1,S(TE)=S0e−TE/T2, [6]

where S0 represents the equilibrium signal amplitude. Only voxel-wise fits with a coefficient of determination R2>0.95 were retained for subsequent analysis.

Joint relaxation fitting used the complete multidimensional phantom dataset and the discrete forward model in Eq. 2, with the encoding kernel defined in Eq. 3. Under the single-component assumption, the joint relaxation-rate distribution for each compartment was modeled as a delta function, pR1,R2=δR1−R1′δR2−R2′. The relaxation times are then T1=1/R1′ and T2=1/R2′, and the multidimensional signal model reduces to:

S(TI,TE,TR)=S01−2e−TI/T1+e−TR/T1e−TE/T2. [7]

Goodness of fit for the mono-exponential T1 and T2 estimates obtained using Eqs. 6 was assessed using linear regression, coefficient of determination R2, and root-mean-square error (RMSE). For each phantom compartment, we calculated mean T1 and T2 values and their standard errors using regions of interest based on the phantom geometry.

Quantitative validation of the joint T1−T2 analysis focused on phantom compartments with relaxation properties spanning the range relevant to human brain tissue. We compared the joint relaxation estimates with the mono-exponential T1 and T2 estimates. Where available, we also compared them with previously published MRI and NMR measurements obtained at 0.064 T (41; 42). Agreement between the estimated and reference relaxation times was evaluated relative to the line of identity.

In Vivo Human Brain Imaging

Data Acquisition

In vivo MRI experiments were performed in one healthy adult volunteer (28-year-old male) under an Institutional Review Board (IRB)-approved research protocol (NCT06203626). Data were acquired on a Hyperfine SWOOP portable 0.064 T MRI system (hardware version 1.6, software version rc9.0.0_Beta1). We used the two-dimensional IR-FSE acquisition described above and shown in Figure 1.

The volunteer was scanned using nine inversion times ranging from 18 to 1200 ms. An additional acquisition without an inversion pulse was used to sample near-fully relaxed magnetization. TR was fixed at 1400 ms for all inversion times. Four lines per image (LPI = 4) were acquired. Multiple echoes within each FSE echo train therefore filled successive lines in 3D k-space to form an echo image, as shown in Figure 1. The echo spacing was 5.12 ms, with TE values extending to 409.6 ms. Images were acquired with an in-plane resolution of 2.5 × 2.5 mm2, a slice thickness of 5.8 mm, and a field of view (FOV) of 180 × 220 × 156 mm3. Complete acquisition parameters are provided in Table 1. The total acquisition time was approximately 100 minutes.

Data Analysis

Prior to spectral reconstruction, the MR images were processed in the following order: 1) random noise was suppressed using a Marchenko-Pastur principal component analysis (MP-PCA) algorithm (43) implemented in the DIPY software environment (44); 2) inter-scan subject motion, odd-even echo errors, and eddy-current-induced geometric distortions were reduced by registering all volumes to the first echo acquisition using a 3D affine transformation (45; 46) implemented in the FSL software environment (47); and 3) signal polarity was restored for magnitude images using the algorithm described in (48) to account for the inversion.

Mono-exponential T1 and T2 maps were generated using the signal models defined in Eq. 6. T1 was estimated from the minimum-TE inversion-recovery images and T2 from the multi-echo non-inversion acquisition. Representative white matter, gray matter, and cerebrospinal fluid (CSF) voxels were selected for comparison with previously reported relaxation values.

Voxel-wise joint relaxation distributions were reconstructed from the in vivo dataset using MADCO and the ℓ2-regularized optimization and constraints defined in Eqs. 4 and 5. The joint distribution, pR1,R2, was represented on a logarithmically spaced 21 × 21 grid of relaxation rates, corresponding to T1 and T2 ranges of approximately 18–5,600 ms and 10–1,000 ms, respectively. The solution was required to be nonnegative and normalized, with the marginal distributions restricting the admissible R1−R2 parameter space according to Eq. 5. This space included regions encompassing 95% of the marginal probability, with adaptive gridding used to increase sampling density where needed and reduce it elsewhere. The regularization parameter λ was determined independently in each voxel using the S-curve method. The apparent inversion efficiency, η, was estimated voxel-wise and incorporated into the longitudinal relaxation kernel used for the joint reconstruction (Figure 6b).

Figure 6:

Figure 6:

Full spatial representation of the reconstructed joint relaxation distribution and apparent inversion efficiency. (a) Voxel-wise spatial maps of the reconstructed pT1,T2 distribution displayed over the complete 21 × 21 relaxation grid. Each map represents the reconstructed spectral probability associated with the indicated combination of T1 and T2. The spatial T1 marginal distributions, obtained by summation over T2, are shown across the top, and the corresponding T2 marginal distributions, obtained by summation over T1, are shown along the left side. (b) Voxel-wise apparent inversion efficiency (η) across five representative axial slices, displayed on a common scale from 0 to 1.

For visualization and comparison with mono-exponential measurements, the reconstructed spectra were converted to relaxation-time coordinates using T1=1/R1 and T2=1/R2. The spectral bins were reordered in ascending T1 and T2 without changing their reconstructed amplitudes. The one-dimensional marginal distributions were obtained by integration (i.e., summation) of the joint distribution over the complementary relaxation-rate dimension:

pR1=∑R2pR1,R2,pR2=∑R1pR1,R2. [8]

Mean relaxation maps, T1 and T2, were calculated from the marginal distributions together with the normalized distribution width, defined as the coefficient of variation (CV), σT1/T1 and σT2/T2.

Spectral component fractions were calculated by dividing the two-dimensional relaxation space into selected regions. We chose these regions manually and heuristically by inspecting spatial patterns across the complete reconstructed joint relaxation space. Contiguous boundaries on the discrete relaxation grid were then used to group bins with related spatial characteristics. We also examined the resulting component maps for spatial coherence. For each voxel, the component fractions were calculated as

Fk=∑(i,j)∈Ωkpij∑k∑(i,j)∈Ωkpij, [9]

where pij denotes the reconstructed probability at grid location (i,j),Ωk denotes the region corresponding to component k, and Fk represents its fractional contribution. By construction, ∑kFk=1 within each voxel.

We calculated voxel-wise measures of statistical dependence to examine information in the joint R1−R2 distribution that is not captured by the individual relaxation dimensions. Covariance and Pearson correlation describe linear dependence between R1 and R2. Mutual information (MI) was included as a measure of statistical dependence that does not assume a linear relationship.

These statistical measures were computed directly from the reconstructed joint distribution, pR1,R2. Although reconstruction was performed on a logarithmically spaced grid, all statistical moments were calculated using the physical relaxivity values R1 and R2. In each voxel, the distribution was normalized such that ∑i,jpR1,i,R2,j=1.

Covariance was used to quantify the joint linear variation between the two relaxivities:

CovR1,R2=∑i,jpR1,i,R2,jR1,i−R‾1R2,j−R‾2, [10]

where R‾1=∑i,jpR1,i,R2,jR1,i and R‾2=∑i,jpR1,i,R2,jR2,j are the mean relaxivities used to center the distribution.

Pearson correlation was used to normalize the covariance by the standard deviations of the corresponding marginal relaxivity distributions:

ρR1,R2=CovR1,R2σR1σR2. [11]

MI was used to quantify statistical dependence between R1 and R2 beyond linear correlation. The marginal distributions were defined as pR1,i=∑jpR1,i,R2,j and pR2,j=∑ipR1,i,R2,j, and MI was calculated as

IR1,R2=∑i,j:pij>0pijlnpijpR1,ipR2,j, [12]

where pij=pR1,i,R2,j. The natural logarithm was used, so MI is reported in nats. MI is zero when the joint distribution is separable, pR1,R2=pR1pR2, whereas values greater than zero indicate statistical dependence between the two relaxation dimensions.

All statistical measures were calculated from the same normalized, nonnegative pR1,R2 distribution within the brain mask. Voxels with zero total spectral probability or invalid numerical values were excluded, and correlation was calculated only when both marginal variances were positive. No S0 weighting, additional bin-width weighting, or Jacobian transformation was applied because the calculations were performed directly on the native relaxivity reconstruction grid.

Results

Phantom Validation

The joint T1−T2 fitting approach was evaluated using the UNITY phantom. Figure 2a shows three slices containing the primary quantitative MnCl2, PVP, and NiCl2 arrays. The joint T1 maps showed well-defined, spatially uniform compartments across the three phantom slices, with clear variation in relaxation time across the MnCl2 and NiCl2 concentration series and between the selected PVP compartments. The corresponding joint T2 maps demonstrated the same overall compartmental organization, although with greater within-compartment spatial variability, particularly in compartments with longer T2 values.

Figure 2:

Figure 2:

Phantom validation of joint T1−T2 relaxation estimates at 0.064 T. (a) Phantom schematics and corresponding joint T1 and T2 maps for the three acquired slices containing MnCl2, PVP, and NiCl2 compartments. Dashed circles indicate compartments with relaxation properties spanning the range relevant to biological tissues and selected for quantitative comparison. (b) Joint T1 and T2 estimates compared with mono-exponential relaxometry (top) and previously published MRI measurements of the same phantom acquired at 0.064 T (41) (bottom). Points are colored according to phantom composition: MnCl2 (red), PVP (yellow), and NiCl2 (green). Dashed lines indicate the line of identity. Compartment-wise mean values, standard errors, and additional published NMR measurements at 0.064 T (42) are provided in the Supplementary Material.

Figure 2b compares the joint relaxation estimates with those obtained using mono-exponential T1 and T2 relaxometry. Overall, the two approaches agreed closely across the selected compartments. For example, in the 0.1353 mM MnCl2 compartment, mono-exponential fitting yielded T1=352.7±6.2ms and T2=173.9±1.1ms, compared with joint-distribution estimates of T1=344.0±5.3ms and T2=176.0±0.7ms. Similarly, for the 50% PVP outer compartment, the corresponding estimates were T1=302.2±6.1ms and 302.3 ± 6.1 ms, respectively, while the T2 estimates were 259.3 ± 6.0 ms and 289.6 ± 6.1 ms. In the 11.3mMNiCl2 compartment, mono-exponential estimates of T1=131.9±0.5 ms and T2=127.0±0.8ms were compared with joint estimates of T1=133.4±0.5ms and T2=122.5±0.7ms. Joint estimates were also compared with previously published MRI measurements of the same phantom at 0.064 T (41); agreement was generally stronger for T1, while larger differences were observed for some T2 compartments. Complete compartment-wise measurements are provided in Supplementary Table S1.

In Vivo Multidimensional Relaxation Imaging

Representative relaxation-weighted images from the multidimensional acquisition are shown in Figure 3. They illustrate both the whole-brain coverage and the range of image contrast obtained with the acquisition. Consecutive axial slices acquired at a single TI−TE combination show coverage across the imaging volume (Figure 3a), with the highlighted slice selected for subsequent visualization.

Figure 3:

Figure 3:

Representative in vivo multidimensional relaxation-weighted images acquired using the IR-FSE sequence. (a) Consecutive axial slices demonstrating whole-brain coverage for a representative TI−TE encoding. The slice outlined in blue was selected for subsequent visualization. (b) Multiple relaxation-weighted images from the selected slice are acquired over representative combinations of inversion time (TI) and echo time (TE), demonstrating the range of image contrast typically generated by the multidimensional acquisition. Images are displayed using a common grayscale intensity scale.

For this slice, varying TI and TE produced a broad range of relaxation-weighted signals (Figure 3b), from high signal to progressively attenuated signal approaching the noise level. A range of intermediate tissue contrasts is also visible. Together, these images show that the acquisition samples relaxation weightings across the useful signal range of the multidimensional encoding space.

Mono-exponential voxel-wise T1 and T2 maps are shown in Figure 4. These maps show the spatial variation in relaxation times across the brain and serve as a scalar reference for the multidimensional relaxation distributions presented below.

Figure 4:

Figure 4:

Mono-exponential quantitative relaxation mapping. A representative relaxation-weighted anatomical image is shown with the white matter (WM) (red), gray matter (GM) (green), and cerebrospinal fluid (CSF) (blue) voxels selected for subsequent evaluation, together with voxel-wise mono-exponential T1 and T2 maps.

Representative white matter (WM) and gray matter (GM) voxels were selected at locations comparable to those used for low-field relaxation measurements by Jordanova et al. (13). Monoexponential T1 estimates were 240 ms in WM and 346 ms in GM, while the corresponding T2 estimates were 83 ms and 156 ms, respectively. These values were broadly consistent with previously reported relaxation measurements acquired on similar 0.064 T MRI scanners (13; 17).

Figure 5a shows the joint T1−T2 relaxation distributions and corresponding marginals for the white matter (WM), gray matter (GM), and cerebrospinal fluid (CSF) voxels identified in Figure 4. The WM and GM spectra were concentrated at shorter and intermediate relaxation times, respectively, while also extending toward longer relaxation components. The CSF distribution was considerably broader and extended toward substantially longer T1 and T2 values. Projection of the joint distribution along each relaxation dimension yielded the corresponding T1 and T2 marginal distributions.

Figure 5:

Figure 5:

Joint relaxation distributions and distribution-derived summary statistics. (a) Representative joint T1−T2 relaxation distributions for white matter (WM), gray matter (GM), and cerebrospinal fluid (CSF) voxels identified in Figure 4. The corresponding T1 and T2 marginal distributions, obtained by projection of the joint distribution along the complementary relaxation dimension, are shown above and to the right of each joint distribution, respectively. (b) Voxel-wise mean relaxation times, T1 and T2, and coefficients of variation, CVT1 and CVT2, derived from the corresponding marginal distributions. The WM and GM voxels appear to have partial volume (CSF) contamination, which is expected at these voxel sizes.

For the selected voxels, distribution-derived relaxation times are reported as ⟨T⟩±σT, where σT describes the width of the corresponding marginal distribution. The T1±σT1 values were 287 ± 198 ms in WM, 396 ± 206 ms in GM, and 1947 ± 626 ms in CSF; the corresponding T2±σT2 values were 51 ± 33 ms, 86 ± 61 ms, and 412 ± 281 ms, respectively.

The marginal distributions were further characterized by their mean relaxation time and coefficient of variation, as shown in Figure 5b. The T1 and T2 maps retain the spatial relaxation contrast while accounting for the full marginal distribution. The coefficient-of-variation maps show how the relative width of these distributions varies across the brain.

Figure 6a shows the complete 21×21pR1,R2 reconstruction displayed spatially across the corresponding T1−T2 relaxation space. Each location in the grid therefore represents the spatial distribution of the reconstructed spectral probability associated with a particular combination of T1 and T2. Summation of the joint distribution along the complementary relaxation dimension yields the corresponding marginal distributions, shown as spatial T1 marginal maps across the top of the grid and T2 marginal maps along the left side. The full display can therefore be viewed alongside the marginals to see what is retained, and what is lost, when the joint distribution is projected onto either relaxation dimension alone.

Neighboring spectral bins frequently exhibit similar anatomical contrast, forming contiguous regions of related spatial patterns that vary gradually across the relaxation space and partially overlap with adjacent regions. These patterns are most apparent at intermediate relaxation times. Much less spectral signal is seen near the shortest and longest relaxation times covered by the reconstruction. The apparent inversion efficiency, η, which was estimated voxel-wise as part of the multidimensional reconstruction, was close to unity throughout the brain and across the representative slices shown in Figure 6b.

To explore this spatial organization further, Figure 7a shows a compact representation of selected T1−T2 bins spanning the relaxation space. Five regions were defined heuristically based on the spatial patterns observed across the full distribution and are indicated by the colored boundaries in the compact representation. For each voxel, the reconstructed probability within the bins assigned to each region was summed and normalized across the five selected regions, producing the component fraction maps shown in Figure 7b. The resulting component maps show distinct spatial patterns across the brain, with different regions of the joint relaxation spectrum contributing different spatial information.

Figure 7:

Figure 7:

Illustrative decomposition of selected regions within the joint T1−T2 relaxation space. (a) Compact representation of selected spectral bins spanning the reconstructed relaxation space. Colored boundaries indicate five regions defined heuristically based on spatial patterns observed across the complete joint distribution shown in Figure 6. (b) Corresponding component fraction maps obtained by summing the reconstructed probability within the bins assigned to each region and normalizing across the five selected regions. Colors identifying the five regions in (a) correspond to the component labels in (b).

Statistical Characterization

Voxel-wise mutual information (MI), covariance, and Pearson correlation coefficient maps derived from the joint pR1,R2 distributions are shown in Figure 8. The three measures exhibited distinct spatial variation across the brain.

Figure 8:

Figure 8:

Voxel-wise statistical characterization of the reconstructed joint pR1,R2 distributions. A representative relaxation-weighted anatomical image is shown for spatial reference, followed by maps of mutual information (MI), covariance, and Pearson correlation between R1 and R2. MI is reported in nats and quantifies statistical dependence between the two relaxation dimensions, while covariance and Pearson correlation characterize linear dependence. Statistical measures were calculated from the normalized, nonnegative joint and marginal probability distributions within the brain mask. The Pearson correlation map is displayed over the range 0–1 for visualization, correlation values were not constrained to be positive.

MI was positive in all valid voxels, with a median value of 0.297, indicating statistical dependence between R1 and R2 throughout the brain volume. The median covariance was 1.29 × 10−5, while the median Pearson correlation coefficient was 0.254, with positive correlations observed in 98.4% of admissible voxels. The three measures therefore capture different aspects of the relationship between R1 and R2, including dependence that is not limited to a linear association.

Discussion

This study demonstrates the feasibility of whole-brain multidimensional T1−T2 relaxation imaging on a portable 0.064 T MRI scanner. We combined a multidimensional IR-FSE acquisition with MADCO reconstruction to obtain joint relaxation distributions in vivo within practically achievable scan times. The reconstructed distributions were examined alongside mono-exponential relaxation measurements, distribution-derived metrics, and measures of statistical dependence within the joint relaxation space. To our knowledge, this is the first in vivo demonstration of multidimensional T1−T2 distribution imaging on a portable ULF MRI scanner. This work is intended as a proof of concept and a starting point for studies addressing more specific biological or clinical questions.

The phantom experiments provide quantitative validation of the multidimensional acquisition and reconstruction framework across a broad range of relaxation properties (Figure 2). Joint T1 estimates agreed well with both mono-exponential fitting and previously reported measurements at 0.064 T. The T2 estimates were more variable, particularly in compartments with longer T2, which may reflect limitations of the multi-echo FSE acquisition. The phantom protocol used LPI = 1 and therefore allowed sampling at short echo times, but the finite echo-train duration limited the longest achievable TE and the characterization of slowly decaying components. Repeated refocusing pulses can also generate stimulated-echo pathways and affect the measured signal evolution when refocusing is imperfect (49; 50). Limited control over RF phase cycling and residual odd-even echo variation on the scanner may have further contributed to the variability in T2. Even with these limitations, the joint estimates were generally consistent with the mono-exponential and previously reported relaxation measurements, supporting the use of the approach for in vivo experiments.

The reconstructed relaxation distributions contain information that is not captured by mono-exponential relaxation mapping. The WM and GM mono-exponential estimates (Figure 4) were broadly consistent with previous measurements at 0.064 T (13; 17) and provide a useful reference for interpreting the distributions. However, a mono-exponential fit assigns a single relaxation value to each voxel. The reconstructed spectra show a much broader range of relaxation components at the same locations (Figure 5a). WM and GM distributions were centered at shorter and intermediate relaxation times but often extended toward longer components, while the CSF distribution was particularly broad and shifted toward longer T1 and T2. Some of the broader or long-relaxation contributions in nominal WM and GM voxels are likely related to partial-volume effects, given the relatively large voxel dimensions and 5.8 mm slice thickness of the in vivo acquisition. CSF is also difficult to characterize with mono-exponential fitting because the maximum TI of 1200 ms provides limited direct encoding of its slowly recovering longitudinal magnetization. The multidimensional reconstruction instead combines the acquired TI,TE, and TR weightings with the MADCO constraints to estimate probability over a broader joint relaxation domain. It is encouraging that long-relaxation contributions are visible in CSF, but the present data cannot establish their quantitative accuracy outside the most strongly encoded relaxation range.

The full joint relaxation distribution shows clear spatial organization across the two-dimensional relaxation space (Figure 6). The marginal distributions retain much of the relaxation-dependent spatial contrast, but they necessarily collapse one dimension of the joint distribution. This becomes apparent when looking across the full T1−T2 space. Neighboring spectral bins often have related spatial patterns and form contiguous regions that change gradually across the distribution. At the same time, bins that contribute to similar locations in one marginal can look quite different depending on their position along the other relaxation dimension. This information cannot be recovered by looking at the marginal distributions alone. The full two-dimensional distribution therefore gives a more complete picture of how the relaxation components are organized and related.

The comparatively low spectral signal toward the shortest and longest relaxation-time limits reflects the relaxation range most strongly encoded by the present acquisition. The sensitivity of the multidimensional measurement to different regions of the relaxation spectrum is therefore expected to depend on the acquisition parameters used to sample the joint space. Future implementations could instead choose the TI,TE,TR, and echo-train sampling based on the region of the relaxation spectrum that is most relevant to a particular application. This would shift sensitivity toward selected relaxation components, rather than trying to sample the entire relaxation space equally.

The component analysis qualitatively explores this spatial organization by grouping selected regions of the joint relaxation space (Figure 7). In the compact display, bins with similar values along one relaxation dimension can have quite different spatial patterns depending on their position along the other dimension (Figure 7a). Grouping some of these regions produces component fraction maps with distinct spatial patterns (Figure 7b). The five components shown here should not be interpreted as definitive tissue compartments or as a systematic segmentation of the relaxation spectrum. They are an illustrative, heuristic decomposition of the information present in the joint space. Some components show patterns that are consistent with known anatomical or relaxation characteristics, including contributions at shorter relaxation times and a long-relaxation component with prominent signal in CSF-containing regions, but we do not assign specific biological identities to them here. Intermediate components may also include contributions from several tissue environments and partial-volume effects, particularly at the spatial resolution of the current ULF acquisition. Future studies will define and validate spectral regions for specific questions, including myelin-sensitive components, tissue transitions, aging-related changes, or pathological alterations, rather than using the illustrative regions selected here.

The statistical analysis gives another way to look at the relationship between the two relaxation dimensions (Figure 8). Mutual information (MI) was positive throughout the analyzed brain, indicating that R1 and R2 were not statistically independent. This matters because, if the two dimensions were independent, the joint distribution could be represented by the product of its marginals and there would be little additional information in reconstructing the full distribution, obviating the extra scans required to estimate it. The positive MI values therefore support what is also visible in the spatial maps: the joint distribution contains information that is lost when the two relaxation dimensions are considered separately. MI may also be useful for assessing joint encoding in other multidimensional MRI approaches, including the measurement of distributions that combine relaxation and diffusion dimensions (21; 51).

The observed dependence between the two relaxation dimensions may also have biological relevance. T1 and T2 arise from different relaxation mechanisms, but both are influenced by properties of the local tissue environment, including water mobility, macromolecular content, lipid composition, and chemical exchange. Consequently, changes in tissue composition or microstructure may alter not only the individual relaxation distributions but also their relationship within the joint T1−T2 space. The statistical measures shown in Figure 8 may therefore be useful for studying how these relationships change with tissue composition, microstructure, or pathology.

Limitations

A primary limitation of the present implementation is the acquisition time, with the in vivo protocol requiring approximately 100 minutes. Further acceleration will be needed for routine clinical imaging. The range of relaxation times directly encoded by the acquisition is also limited. In the phantom, the finite echo-train duration restricted sampling at long TE, while the maximum TI of 1200 ms in the in vivo acquisition provided limited direct encoding of long-T1 components such as CSF. Components outside the most strongly encoded relaxation range should therefore be interpreted cautiously. Spatial resolution also contributes to partial-volume effects; the relatively large voxel dimensions, including a 5.8 mm slice thickness, likely contribute to the broad distributions and overlapping spectral components seen in Figures 5 and 7. Finally, the current scanner platform provides limited user-level control over several acquisition and reconstruction parameters, including RF and gradient pulse specifications, phase cycling, and aspects of noise correction and image reconstruction. Greater control over these parameters would allow more systematic optimization of multidimensional encoding and effects such as stimulated-echo pathways and residual odd-even echo variation.

The scope of the present study is also limited. The in vivo experiment was performed in a single healthy subject and was not designed to assess intersubject variability or reproducibility. Larger studies using standardized acquisition protocols will be needed to address these questions. The spectral component analysis was exploratory, with regions selected manually from spatial and spectral patterns in the reconstructed distributions (Figure 7). These regions were not independently validated as specific tissue or microstructural compartments. More systematic definition and validation of spectral components will therefore be needed before they can be interpreted as quantitative biological markers.

Future Directions

Future work should focus on improving both the acquisition efficiency and the biological specificity of multidimensional relaxation imaging at ULF. Shorter acquisition times would improve practical feasibility, while the relaxation encoding could be broadened or targeted toward specific regions of the spectrum depending on the application. Further acceleration and development of denoising strategies, including approaches that make use of both magnitude and phase information (52), may also increase the efficiency and robustness of multidimensional imaging at ULF. Greater control over RF and gradient waveforms, potentially through open source sequence development frameworks such as Pulseq (53), would also enable further optimization of the multidimensional acquisition.

A second priority is the systematic identification and validation of biologically meaningful spectral features. The short-T1 and short-T2 regions observed within the joint relaxation space may be particularly relevant to future investigation of myelin-sensitive components, while other spectral regions could be defined to address specific biological or clinical questions. Larger studies will also be needed to characterize normative multidimensional relaxation distributions and their variability across populations. The framework could also be extended to additional encoding dimensions, particularly diffusion. This could enable higher-dimensional distributions such as pT1,T2,MD (21) or related joint relaxation-diffusion representations (51), with the potential to increase the specificity of multidimensional MRI at ULF.

Broader Implications

Extending quantitative and multidimensional MRI methods to portable ULF scanners could broaden the information available from accessible neuroimaging beyond anatomical contrast alone. The present whole-brain joint T1−T2 demonstration is an initial step toward bringing multidimensional tissue characterization to lower-cost, portable ULF MRI scanners.

Conclusions

This work demonstrates the feasibility of whole-brain multidimensional T1−T2 relaxation imaging on a portable 0.064 T MRI scanner. Joint relaxation distributions were reconstructed using an IR-FSE acquisition and MADCO, with quantitative phantom measurements used for validation before applying the approach in vivo. The joint distributions contained information beyond the single relaxation values obtained from mono-exponential mapping. This included the marginal distributions and their derived measures, spatial patterns across the joint relaxation space, and measurable dependence between the two relaxation dimensions. To our knowledge, this is the first in vivo demonstration of multidimensional T1−T2 distribution imaging on a portable ULF MRI scanner. Our results support further development of multidimensional MRI for biological and clinical applications.

Supplementary Material

Supplement 1
media-1.pdf (110.2KB, pdf)

Acknowledgments

This study was supported by the Intramural Research Program (IRP) of the Eunice Kennedy Shriver National Institute of Child Health and Human Development under award 1-ZIA-HD008971-07, the National Institute of Neurological Diseases and Stroke Intramural Research Program under award ZIC-NS009450, and the National Institute on Aging. This work was also funded by the Department of War in the Military Traumatic Brain Injury Initiative (MTBI2) under award, HU0001-22-2-0058. This work utilized computational resources of the NIH HPC Biowulf cluster (http://hpc.nih.gov). The authors have no conflicts of interest to declare. The views, information or content, and conclusions presented do not necessarily represent the official position or policy of, nor should any official endorsement be inferred on the part of, the Uniformed Services University, the Department of War, the U.S. Government, or the Henry M. Jackson Foundation for the Advancement of Military Medicine, Inc. The contributions of the NIH author(s) are considered Works of the United States Government. The findings and conclusions presented in this paper are those of the author(s) and do not necessarily reflect the views of the NIH or the U.S. Department of Health and Human Services. The authors wish to thank Sean Deoni of the Gates Foundation and the UNITY consortium created under the auspices of the Gates Foundation for their continued leadership and support in advancing ULF MRI developments and applications in LMICs.

References

  • 1.Cooley CZ, McDaniel PC, Stockmann JP, Srinivas SA, Cauley SF, Śliwiak M, Sappo CR, Vaughn CF, Guerin B, Rosen MS, Lev MH, Wald LL. 2021. A portable scanner for magnetic resonance imaging of the brain. Nature Biomedical Engineering 5:229–239. doi: 10.1038/s41551-020-00641-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Mazurek MH, Cahn BA, Yuen MM, Prabhat AM, Chavva IR, Shah JT, Crawford AL, Welch EB, Rothberg J, Sacolick L, Poole M, Wira C, Matouk CC, Ward A, Timario N, Leasure A, Beekman R, Peng TJ, Witsch J, Antonios JP, Falcone GJ, Gobeske KT, Petersen N, Schindler J, Sansing L, Gilmore EJ, Hwang DY, Kim JA, Malhotra A, Sze G, Rosen MS, Kimberly WT, Sheth KN. 2021. Portable, bedside, low-field magnetic resonance imaging for evaluation of intracerebral hemorrhage. Nature Communications 2021 12:1 12:5119–. doi: 10.1038/s41467-021-25441-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Basser P. 2022. Detection of stroke by portable, low-field mri: A milestone in medical imaging. Science Advances 8:eabp9307. doi: 10.1126/sciadv.abp9307. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Balaji S, Wiley N, Poorman ME, Kolind SH. 2024. Low-field mri for use in neurological diseases. Current Opinion in Neurology 37:381–391. doi: 10.1097/wco.0000000000001282. [DOI] [PubMed] [Google Scholar]
  • 5.Yuen MM, Prabhat AM, Mazurek MH, Chavva IR, Crawford A, Cahn BA, Beekman R, Kim JA, Gobeske KT, Petersen NH, Falcone GJ, Gilmore EJ, Hwang DY, Jasne AS, Amin H, Sharma R, Matouk C, Ward A, Schindler J, Sansing L, de Havenon A, Aydin A, Wira C, Sze G, Rosen MS, Kimberly WT, Sheth KN. 2022. Portable, low-field magnetic resonance imaging enables highly accessible and dynamic bedside evaluation of ischemic stroke. Science Advances 8. doi: 10.1126/sciadv.abm3952. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Chetcuti K, Chilingulo C, Goyal MS, Vidal L, O'Brien NF, Postels DG, Seydel KB, Taylor TE. 2022. Implementation of a low-field portable mri scanner in a resource-constrained environment: Our experience in malawi. American Journal of Neuroradiology 43:670–674. doi: 10.3174/ajnr.A7494. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Arnold TC, Tu D, Okar SV, Nair G, By S, Kawatra KD, Robert-Fitzgerald TE, Desiderio LM, Schindler MK, Shinohara RT, Reich DS, Stein JM. 2022. Sensitivity of portable low-field magnetic resonance imaging for multiple sclerosis lesions. NeuroImage: Clinical 35:103101. doi: 10.1016/j.nicl.2022.103101. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Kazimuddin HF, Pathakamuri A, Yi J, Soldatos A, Turtzo LC, Latour LL, Chittiboina P, Brown DA, Reich DS, Horovitz SG. 2026. Gadolinium-enhanced portable ultra-low-field mri for evaluating various intracranial pathologies. American Journal of Neuroradiology 47:1927–1932. doi: 10.3174/ajnr.A9249. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Abate F, Adu-Amankwah A, Ae-Ngibise KA, Agbokey F, Agyemang VA, Agyemang CT, Akgun C, Ametepe J, Arichi T, Asante KP, Balaji S, Baljer L, Basser PJ, Beauchemin J, Bennallick C, Berhane Y, Boateng-Mensah Y, Bourke NJ, Bradford L, Bruchhage MM, Lorente RC, Cawley P, Cercignani M, Sa VD, de Canha A, de Navarro N, D C ID, Delarosa J, Donald KA, Dvorak A, Edwards AD, Field D, Frail H, Freeman B, George T, Gholam J, Guerrero-Gonzalez J, Hajnal JV, Haque R, Hollander W, Hoodbhoy Z, Huentelman M, Jafri SK, Jones DK, Joubert F, Karaulanov T, Kasaro MP, Knackstedt S, Kolind S, Koshy B, Kravitz R, Lafayette SL, Lee AC, Lena B, Lepore N, Linguraru M, Ljungberg E, Lockart Z, Loth E, Mannam P, Masemola KM, Moran R, Murphy D, Nakwa FL, Nankabirwa V, Nelson CA, North K, Nyame S, Halloran RO, O'Muircheartaigh J, Oakley BF, Odendaal H, Ongeti CM, Onyango D, Oppong SA, Padormo F, Parvez D, Paus T, Pepper MS, Phiri KS, Poorman M, Ringshaw JE, Rogers J, Rutherford M, Sabir H, Sacolick L, Seal M, Sekoli ML, Shama T, Siddiqui K, Sindano N, Spelke MB, Springer PE, Suleman FE, Sundgren PC, Teixeira R, Terekegn W, Traughber M, Tuuli MG, van Rensburg J, Váša F, Velaphi S, Velasco P, Viljoen IM, Vokhiwa M, Webb A, Weiant C, Wiley N, Wintermark P, Yibetal K, Deoni SC, Williams SC. 2024. Unity: A low-field magnetic resonance neuroimaging initiative to characterize neurodevelopment in low and middle-income settings. Developmental Cognitive Neuroscience 69:101397. doi: 10.1016/j.dcn.2024.101397. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Ma D, Gulani V, Seiberlich N, Liu K, Sunshine JL, Duerk JL, Griswold MA. 2013. Magnetic resonance fingerprinting. Nature 495:187–192. doi: 10.1038/nature11971. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Filo S, Shtangel O, Salamon N, Kol A, Weisinger B, Shifman S, Mezer AA. 2019. Disentangling molecular alterations from water-content changes in the aging human brain using quantitative mri. Nature Communications 10:3403. doi: 10.1038/s41467-019-11319-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Rooney WD, Johnson G, Li X, Cohen ER, Kim SG, Ugurbil K, Springer CS. 2007. Magnetic field and tissue dependencies of human brain longitudinal 1h2o relaxation in vivo. Magnetic Resonance in Medicine 57:308–318. doi: 10.1002/mrm.21122. [DOI] [PubMed] [Google Scholar]
  • 13.Jordanova KV, Martin MN, Ogier SE, Poorman ME, Keenan KE. 2023. In vivo quantitative mri: T1 and t2 measurements of the human brain at 0.064 t. Magnetic Resonance Materials in Physics, Biology and Medicine 36:487–498. doi: 10.1007/s10334-023-01095-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.O'Reilly T, Webb AG. 2022. In vivo t1 and t2 relaxation time maps of brain tissue, skeletal muscle, and lipid measured in healthy volunteers at 50 mt. Magnetic Resonance in Medicine 87:884–895. doi: 10.1002/mrm.29009. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Deoni SC, O'Muircheartaigh J, Ljungberg E, Huentelman M, Williams SC. 2022. Simultaneous high-resolution t2-weighted imaging and quantitative t2 mapping at low magnetic field strengths using a multiple te and multi-orientation acquisition approach. Magnetic Resonance in Medicine 88:1273–1281. doi: 10.1002/mrm.29273. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Padormo F, Cawley P, Dillon L, Hughes E, Almalbis J, Robinson J, Maggioni A, Botella MDLF, Cromb D, Price A, Arlinghaus L, Pitts J, Luo T, Zhang D, Deoni SC, Williams S, Malik S, O'Muircheartaigh J, Counsell SJ, Rutherford M, Arichi T, Edwards AD, Hajnal JV. 2023. In vivo t1 mapping of neonatal brain tissue at 64 mt. Magnetic Resonance in Medicine 89:1016–1025. doi: 10.1002/mrm.29509. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Lena B, Padormo F, Teixeira RPA, Bennallick C, Gholam J, van den Broek R, Lafayette SL, Vavasour I, Cercignani M, Jones DK, Kolind S, Hajnal J, Bourke N, Dong Y, Hollander WJ, Karaulanov T, Deoni SC, Williams SC, Sundgren PC, Webb AG, Ljungberg E. 2025. Repeatability and reproducibility of rapid t1 mapping of brain tissues at 64 mt: A multicentre study. Imaging Neuroscience 3. doi: 10.1162/imag.a.916. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Whittall KP, MacKay AL, Graeb DA, Nugent RA, Li DK, Paty DW. 1997. In vivo measurement of t2 distributions and water contents in normal human brain. Magnetic Resonance in Medicine 37:34–43. doi: 10.1002/mrm.1910370107. [DOI] [PubMed] [Google Scholar]
  • 19.Labadie C, Lee JH, Rooney WD, Jarchow S, Aubert-Frécon M, Springer CSJ, Möller HE. 2014. Myelin water mapping by spatially regularized longitudinal relaxographic imaging at high magnetic fields. Magnetic Resonance in Medicine 71:375–387. doi: 10.1002/mrm.24670. [DOI] [PubMed] [Google Scholar]
  • 20.Avram AV, Sarlls JE, Basser PJ. 2019. Measuring non-parametric distributions of intravoxel mean diffusivities using a clinical mri scanner. NeuroImage 185:255–262. doi: 10.1016/j.neuroimage.2018.10.030. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Benjamini D, Bouhrara M, Komlosh ME, Iacono D, Perl DP, Brody DL, Basser PJ. 2021. Multidimensional mri for characterization of subtle axonal injury accelerated using an adaptive nonlocal multispectral filter. Frontiers in Physics 9:737374. doi: 10.3389/fphy.2021.737374. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Slator PJ, Palombo M, Miller KL, Westin CF, Laun F, Kim D, Haldar JP, Benjamini D, Lemberskiy G, de Almeida Martins JP, Hutter J. 2021. Combined diffusion-relaxometry microstructure imaging: Current status and future prospects. Magnetic Resonance in Medicine 86:2987–3011. doi: 10.1002/mrm.28963. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.English AE, Whittall KP, Joy ML, Henkelman RM. 1991. Quantitative two-dimensional time correlation relaxometry. Magnetic Resonance in Medicine 22:425–434. doi: 10.1002/mrm.1910220250. [DOI] [PubMed] [Google Scholar]
  • 24.Kim D, Doyle EK, Wisnowski JL, Kim JH, Haldar JP. 2017. Diffusion-relaxation correlation spectroscopic imaging: A multidimensional approach for probing microstructure. Magnetic Resonance in Medicine 78:2236–2249. doi: 10.1002/mrm.26629. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Barsoum S, Latimer CS, Nolan AL, Barrett A, Chang K, Troncoso JC, Keene CD, Benjamini D. 2025. Multidimensional mri reveals cortical astrogliosis linked to dementia in alzheimer's disease. Brain Communications 7:fcaf245. doi: 10.1093/braincomms/fcaf245. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Kundu S, Barsoum S, Ariza J, Nolan AL, Latimer CS, Keene CD, Basser PJ, Benjamini D. 2023. Mapping the individual human cortex using multidimensional mri and unsupervised learning. Brain Communications 5. doi: 10.1093/braincomms/fcad258. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Park JS, Manninen E, Yang Y, Benjamini D. 2026. Informed dictionary-guided monte carlo inversion for robust and reproducible multidimensional mri. Magnetic Resonance in Medicine 95:2947–2962. doi: 10.1002/mrm.70228. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Avram AV, Sarlls JE, Basser PJ. 2021. Whole-brain imaging of subvoxel t1-diffusion correlation spectra in human subjects. Frontiers in Neuroscience 15:671465. doi: 10.3389/fnins.2021.671465. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Magdoom KN, Komlosh ME, Saleem K, Gasbarra D, Basser PJ. 2022. High resolution ex vivo diffusion tensor distribution mri of neural tissue. Frontiers in Physics 10:807000. doi: 10.3389/fphy.2022.807000. [DOI] [Google Scholar]
  • 30.Slator PJ, Hutter J, Palombo M, Jackson LH, Ho A, Panagiotaki E, Chappell LC, Rutherford MA, Hajnal JV, Alexander DC. 2019. Combined diffusion-relaxometry mri to identify dysfunction in the human placenta. Magnetic Resonance in Medicine 82:95–106. doi: 10.1002/mrm.27733. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Benjamini D, Priemer DS, Perl DP, Brody DL, Basser PJ. 2023. Mapping astrogliosis in the individual human brain using multidimensional mri. Brain 146:1212–1226. doi: 10.1093/brain/awac298. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Benjamini D, Iacono D, Komlosh ME, Perl DP, Brody DL, Basser PJ. 2021. Diffuse axonal injury has a characteristic multidimensional mri signature in the human brain. Brain 144:800–816. doi: 10.1093/brain/awaa447. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.de Almeida Martins JP, Tax CM, Szczepankiewicz F, Jones DK, Westin CF, Topgaard D. 2020. Transferring principles of solid-state and laplace nmr to the field of in vivo brain mri. Magnetic Resonance 1:27–43. doi: 10.5194/mr-1-27-2020. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Martin J, Reymbaut A, Schmidt M, Doerfler A, Uder M, Laun FB, Topgaard D. 2021. Non-parametric d-r1-r2 distribution mri of the living human brain. NeuroImage 245:118753. doi: 10.1016/j.neuroimage.2021.118753. [DOI] [PubMed] [Google Scholar]
  • 35.Johnson JT, Irfanoglu MO, Manninen E, Ross TJ, Yang Y, Laun FB, Martin J, Topgaard D, Benjamini D. 2024. In vivo disentanglement of diffusion frequency-dependence, tensor shape, and relaxation using multidimensional mri. Human Brain Mapping 45:e26697. doi: 10.1002/hbm.26697. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Bai R, Benjamini D, Cheng J, Basser PJ. 2016. Fast, accurate 2d-mr relaxation exchange spectroscopy (rexsy): Beyond compressed sensing. Journal of Chemical Physics 145:154202. doi: 10.1063/1.4964144. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Callaghan PT. 2011. Translational dynamics and magnetic resonance: Principles of pulsed gradient spin echo NMR. Oxford University Press. [Google Scholar]
  • 38.Bai R, Cloninger A, Czaja W, Basser PJ. 2015. Efficient 2d mri relaxometry using compressed sensing. Journal of Magnetic Resonance 255:88–99. doi: 10.1016/j.jmr.2015.04.002. [DOI] [PubMed] [Google Scholar]
  • 39.Benjamini D, Basser PJ. 2016. Use of marginal distributions constrained optimization (madco) for accelerated 2d mri relaxometry and diffusometry. Journal of Magnetic Resonance 271:40–45. doi: 10.1016/j.jmr.2016.08.004. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40.Venkataramanan L, Song YQ, Hürlimann MD. 2002. Solving fredholm integrals of the first kind with tensor product structure in 2 and 2.5 dimensions. IEEE Transactions on Signal Processing 50:1017–1026. doi: 10.1109/78.995059. [DOI] [Google Scholar]
  • 41.Ljungberg E, Padormo F, Poorman M, Clemensson P, Bourke N, Evans JC, Gholam J, Vavasour I, Kollind SH, Lafayette SL, Bennallick C, Donald KA, Bradford LE, Lena B, Vokhiwa M, Shama T, Siew J, Sekoli L, van Rensburg J, Pepper MS, Khan A, Madhwani A, Banda FA, Mwila ML, Cassidy AR, Moabi K, Sephi D, Boakye RA, Ae-Ngibise KA, Asante KP, Hollander WJ, Karaulanov T, Williams SC, Deoni S. 2025. Characterization of portable ultra-low field mri scanners for multi-center structural neuroimaging. Human Brain Mapping 46:e70217. doi: 10.1002/hbm.70217. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 42.Martin MN, Jordanova KV, Kos AB, Russek SE, Keenan KE, Stupic KF. 2023. Relaxation measurements of an mri system phantom at low magnetic field strengths. Magnetic Resonance Materials in Physics, Biology and Medicine 36:477–485. doi: 10.1007/s10334-023-01086-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 43.Veraart J, Novikov DS, Christiaens D, Ades-Aron B, Sijbers J, Fieremans E. 2016. Denoising of diffusion mri using random matrix theory. NeuroImage 142:394–406. doi: 10.1016/j.neuroimage.2016.08.016. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 44.Garyfallidis E, Brett M, Amirbekian B, Rokem A, van der Walt S, Descoteaux M, Nimmo-Smith I, Contributors D. 2014. Dipy, a library for the analysis of diffusion mri data. Frontiers in Neuroinformatics 8:8. doi: 10.3389/fninf.2014.00008. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 45.Jenkinson M, Smith S. 2001. A global optimisation method for robust affine registration of brain images. Medical Image Analysis 5:143–156. doi: 10.1016/s1361-8415(01)00036-6. [DOI] [PubMed] [Google Scholar]
  • 46.Jenkinson M, Bannister P, Brady M, Smith S. 2002. Improved optimization for the robust and accurate linear registration and motion correction of brain images. NeuroImage 17:825–841. doi: 10.1006/nimg.2002.1132. [DOI] [PubMed] [Google Scholar]
  • 47.Smith SM, Jenkinson M, Woolrich MW, Beckmann CF, Behrens TE, Johansen-Berg H, Bannister PR, Luca MD, Drobnjak I, Flitney DE, Niazy RK, Saunders J, Vickers J, Zhang Y, Stefano ND, Brady JM, Matthews PM. 2004. Advances in functional and structural mr image analysis and implementation as fsl. NeuroImage 23:S208–S219. doi: 10.1016/j.neuroimage.2004.07.051. [DOI] [PubMed] [Google Scholar]
  • 48.Bakker CJ, de Graaf CN, van Dijk P. 1984. Restoration of signal polarity in a set of inversion recovery nmr images. IEEE Transactions on Medical Imaging 3:197–202. doi: 10.1109/tmi.1984.4307681. [DOI] [PubMed] [Google Scholar]
  • 49.Song YQ. 2002. Categories of coherence pathways for the cpmg sequence. Journal of Magnetic Resonance 157:82–91. doi: 10.1006/jmre.2002.2577. [DOI] [PubMed] [Google Scholar]
  • 50.Le Roux P. 2002. Non-CPMG Fast Spin Echo with Full Signal. Journal of Magnetic Resonance 155:278–292. doi: 10.1006/JMRE.2002.2523. [DOI] [PubMed] [Google Scholar]
  • 51.Magdoom KN, Sarlls JE, Basser PJ. 2026. Towards mesoscopic human brain imaging using non-parametric diffusion tensor distribution (dtd) mri. bioRxiv doi: 10.64898/2026.04.12.718025. [DOI] [Google Scholar]
  • 52.Bai R, Koay CG, Hutchinson E, Basser PJ. 2014. A framework for accurate determination of the t2 distribution from multiple echo magnitude mri images. Journal of Magnetic Resonance 244:53–63. doi: 10.1016/j.jmr.2014.04.016. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 53.Layton KJ, Kroboth S, Jia F, Littin S, Yu H, Leupold J, Nielsen JF, Stöcker T, Zaitsev M. 2017. Pulseq: A rapid and hardware-independent pulse sequence prototyping framework. Magnetic Resonance in Medicine 77:1544–1552. doi: 10.1002/mrm.26235. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supplement 1
media-1.pdf (110.2KB, pdf)

Articles from bioRxiv are provided here courtesy of Cold Spring Harbor Laboratory Preprints

RESOURCES