The impact of subsidence on hip function: A comparative study between Proximal Femoral Nail Anti-rotation-2 and Halifax Femoral Nail in unstable trochanteric fractures

Soumya Ranjan Nayak 1, Amit Rastogi 1, Shivam Sinha 1,*, Birju Manjhi 2, Rahul Rai 1, Sushil Gandhi 1
PMCID: PMC13625632  PMID: 42813734

Abstract

Background: Unstable trochanteric fractures pose a significant challenge in orthopedic surgery, particularly in elderly patients with poor bone quality. Achieving biomechanical stability while minimizing complications requires careful consideration of intramedullary nail design. This study aims to compare two commonly used implants – Proximal Femoral Nail Anti-rotation-2 (PFNA-2) and Halifax Femoral Nail – by evaluating their clinical, radiological, and functional outcomes, with a focus on surgical efficacy, complications, and functional recovery. Methods: A prospective, randomized clinical trial involving 60 elderly patients with AO/OTA 31A2 or 31A3 unstable trochanteric fractures was conducted over 1 year. Patients aged 60–80 years were randomized into two equal groups (n = 30 each) to receive either PFNA-2 or Halifax Nail fixation. Primary outcomes were evaluated using the Harris Hip Score (HHS) at 6 and 12 months postoperatively. Secondary outcome measures included neck-shaft angle, tip-apex distance, and screw positioning according to the Cleveland Index. Operative parameters, including surgical time, blood loss, and complications, were systematically recorded and analyzed. Results: Both implants demonstrated comparable efficacy regarding operative time (Halifax: 31.7 ± 5.5 vs. PFNA-2: 33.0 ± 7.9 min, p = 0.45) and intraoperative blood loss (Halifax: 89.5 ± 12.6 vs.PFNA-2: 97.7 ± 22.9 ml, p = 0.09). Halifax Nail achieved significantly superior screw positioning with 76.7% optimal center–center placement compared to 43.3% in the PFNA-2 group (p = 0.008). Subsidence was absent in the Halifax group but occurred in 6.7% of PFNA-2 cases. HHSs improved progressively in both groups, with Halifax showing marginally higher scores at 6 months (83.6 ± 4.3 vs.81.5 ± 6.4) and 12 months (87.4 ± 3.9 vs.85.8 ± 5.6), though differences were not statistically significant. Discussion: Both PFNA-2 and Halifax Nails are effective for managing unstable trochanteric fractures. The Halifax Nail’s innovative tri-wire technology™ offers superior rotational stability, optimal screw positioning, and demonstrates a trend towards reduced subsidence, making it particularly advantageous for patients with compromised bone quality. These findings support individualized implant selection based on patient factors and bone quality assessment.

Keywords: Unstable trochanteric fractures, Intramedullary nailing, Proximal Femoral Nail Anti-rotation-2, Halifax Femoral Nail, Harris Hip Score, Subsidence

Introduction

Hip fractures are among the most severe injuries in the elderly, leading to significant morbidity, functional decline, and increased mortality. Intertrochanteric fractures account for nearly half of all hip fractures in individuals over 65 and have higher mortality and poorer functional outcomes than femoral neck fractures [1]. Unstable intertrochanteric fractures require precise surgical interventions to prevent complications like excessive collapse [2]. These injuries have higher early mortality [3], and fewer than 50% of patients regain pre-fracture walking ability within a year [4].

Fractures are classified as stable or unstable based on morphology and comminution. Unstable types (e.g., AO/OTA 31A2, 31A3) involve comminution and reverse obliquity, increasing collapse and malalignment risks [5]. Intramedullary devices like the Proximal Femoral Nail Anti-rotation (PFNA) are preferred for minimal soft tissue damage and early weight-bearing [6]. While dynamic hip screws (DHS) remain alternatives, issues like implant irritation and longer surgical time persist [7].

Challenges in managing unstable fractures include age, osteoporosis, comorbidities, and poor bone quality, increasing fixation difficulty, and implant failure [8, 9]. Surgery is standard, offering better outcomes and early mobilization [10], though complication rates remain high [11]. Osteoporosis exacerbates these risks, which PFNA and cement augmentation help mitigate [12, 13].

Intramedullary implants improve stability, reduce surgical time and blood loss, and enable early mobilization [14–16]. The PFNA-2’s helical blade compacts cancellous bone, enhancing stability and reducing cut-out risk [12, 17]. However, complications like blade migration arise from poor placement and parameters like tip-apex distance (TAD) >27 mm and neck-shaft angle (NSA) <128° [18, 19]. The Halifax Nail, with tri-wire technology™, promises enhanced rotational stability, with preliminary data suggesting superior biomechanics [20].

This prospective, randomized study compares PFNA-2 and Halifax Nail in managing unstable trochanteric fractures, evaluating operative time, blood loss, radiographic results, functional recovery, and complications.

Materials and methods

This prospective, randomized clinical trial was conducted over 1 year at the Department of Orthopedics in a tertiary care hospital comparing PFNA-2 and Halifax Femoral Nail efficacy in treating unstable trochanteric fractures. Written informed consent was collected from all participants. Sample size was calculated based on the primary outcome measure, the Harris Hip Score (HHS) at 12 months. Assuming a standard deviation of 6, a minimum clinically important difference of 5 points in HHS, 80% power, and α of 0.05, the required sample size was 23 patients per group. Considering possible attrition, 30 patients were enrolled in each group, resulting in a total sample size of 60 patients.

Sixty patients aged 60–80 years with AO/OTA 31A2 or 31A3 fractures, eligible for surgery within 7 days, were randomized into two groups (n = 30 each) using computer-generated randomization.

Exclusion criteria included pathological fractures, prior proximal femur surgeries, fractures older than 1 week, and medical contraindications. Preoperative evaluation involved history-taking, physical exams, radiological assessments, and routine lab tests. Patients were prepared with fasting, antiseptic cleansing, and prophylactic antibiotics.

Surgical techniques for both implants followed similar steps involving patient positioning, fracture reduction, and fluoroscopic guidance. PFNA-2 used a helical blade, while Halifax Nail employed a tri-wire technology™ for enhanced rotational stability. Postoperative care included pain management, mobilization, imaging, and wound care, with regular follow-ups assessing pain, mobility, and healing.

Outcome measures included HHS, NSA, TAD, screw positioning, operative time, blood loss, and complications. Allocation concealment was ensured using sequentially numbered, sealed, opaque envelopes prepared by an independent staff member not involved in patient recruitment or outcome assessment. The envelopes were opened in the operating room after patient enrolment. Postoperative clinical and radiological assessments, including HHS evaluation and radiographic measurements, were performed by an independent assessor who was blinded to group allocation. Statistical analysis was performed using SPSS version 25.0. Descriptive statistics were expressed as mean ± standard deviation for continuous variables and frequencies with percentages for categorical variables. Independent t-tests compared continuous variables between groups, while chi-square or Fisher’s exact tests were employed for categorical variables. A p-value <0.05 was considered statistically significant.

Results

Demographic and clinical data

Age distribution

Sixty patients with a mean age of 71.5 ± 5.1 years (range: 63–80), median 72 years (IQR: 67.8–76) were included. 51.7% (31 patients) were aged 71–80, and 48.3% (29 patients) were 61–70 years. No significant age difference existed between groups: Halifax (71.3 ± 5.2) vs. PFNA-2 (71.6 ± 5.1), p = 0.82.

Gender distribution

The cohort had equal gender distribution (50% each), with 15 males and 15 females per group. No significant gender difference was observed (p = 1.00).

Mode of injury

Slip and fall was the most common cause (70%, 42 patients), followed by falls from stairs (10%), bike-related injuries (5%), and falls from chairs, hit by an auto, or bull hits (3.3% each). Injury modes were similarly distributed between groups (p = 0.79).

Preoperative and operative parameters

Injury-to-surgery interval

The mean injury-to-surgery interval was 3.5 ± 1.2 days, median 3 days (range: 2–7). Most surgeries (41.7%) were performed within 3 days, followed by 4 days (23.3%), 2 days (16.7%), and 5 days (15%). The Halifax group had a slightly shorter mean interval (3.4 ± 0.9 vs. 3.7 ± 1.4 days), but this was not statistically significant (p = 0.34).

Blood loss

Average intraoperative blood loss was 93.6 ± 18.8 ml, median 90 ml (range: 50–200). Most patients (53.3%) had blood loss of 81–90 ml, while 36.7% had 91–100 ml. The Halifax group exhibited slightly lower mean blood loss (89.5 ± 12.6 vs. 97.7 ± 22.9 ml), but the difference was not statistically significant (p = 0.09).

Operative time

Mean operative time was 32.3 ± 6.8 minutes, median 32 min (range: 20–60). Most cases (46.7%) were completed within 20–30 min, followed by 31–40 min (41.7%). The Halifax group had a mean operative time of 31.7 ± 5.5 min vs. the PFNA-2 group, 33.0 ± 7.9 min. The difference was not statistically significant (p = 0.45).

Radiological outcomes

NSA

Mean postoperative NSA difference was 5.2 ± 2.2 degrees, median 4 degrees (range: 2–11). Most patients (51.7%) exhibited a difference of 2–4 degrees, followed by 38.3% with 5–8 degrees, and 10% with 9–11 degrees. No significant differences between groups: Halifax 5.0 ± 2.2 vs. PFNA-2 5.3 ± 2.1 degrees (p = 0.64).

TAD

Mean TAD was 22.2 ± 2.2 mm, median 22 mm (range: 15–26). Most cases (60%) fell within a 19–23 mm range, while 33.3% had TADs of 24–26 mm, and 6.7% had 15–18 mm. Halifax group: 22.4 ± 1.8 mm vs. PFNA-2 group: 22.0 ± 2.6 mm. Differences were not statistically significant (p = 0.46).

Screw position (Cleveland Index)

Optimal center-center (CC) positioning was achieved in 60% (36 cases) of the cohort. Remaining 40% (24 cases) had a center-inferior (CI) position. The Halifax group had a significantly higher proportion of CC positions (76.7%) compared to the PFNA-2 group (43.3%, p = 0.008). Conversely, the PFNA-2 group had a higher proportion of CI positions (56.7%) vs. Halifax (23.3%), as shown in Table 1 and Figure 1A.

Table 1.

Comparison of screw position between Halifax vs.PFNA2.

Screw position Halifax (n = 30) PFNA2 (n = 30) Total p-value
CC 23 (76.67%) 13 (43.33%) 36 (60%)
CI 7 (23.33%) 17 (56.67%) 24 (40%) 0.008†
Total 30 (100%) 30 (100%) 60 (100%)

†Chi-square test.

Figure 1.

Figure 1

(A) Comparison of screw position between Halifax vs.PFNA2. (B): Comparison of Harris Hip Score between Halifax and PFNA2. (C): Comparison of subsidence between Halifax vs.PFNA2.

Functional outcomes

HHS assessed functional recovery at 6 and 12 months postoperatively. At 6 months, mean HHS was 82.6 ± 5.5, median 84 (range: 66–90). The majority (65%) achieved scores 81–90, while 28.3% scored 71–80, and 6.7% scored 61–70. The Halifax group had slightly higher mean HHS (83.6 ± 4.3 vs. 81.5 ± 6.4), though not statistically significant (p = 0.14).

At 12 months, mean HHS improved to 86.6 ± 4.9, median 88 (range: 68–92). Most patients (75%) scored 81–90, with 13.3% achieving 91–95, 10% scoring 71–80, and 1.7% scoring 61–70. The Halifax group demonstrated slightly higher mean scores (87.4 ± 3.9 vs. 85.8 ± 5.6), but this was not statistically significant (p = 0.22).

Both groups showed progressive improvement over time. The Halifax group exhibited marginally better outcomes at both time points, yet differences were not statistically significant, indicating comparable effectiveness as shown in Table 2 and Figure 1B.

Table 2.

Comparison of Harris Hip Score between Halifax and PFNA2.

Harris Hip Score Halifax (n = 30) PFNA2 (n = 30) Total p-value
At 6 months
61–70 1 (3.33%) 3 (10%) 4 (6.67%)
71–80 6 (20%) 11 (36.67%) 17 (28.33%) 0.217*
81–90 23 (76.67%) 16 (53.33%) 39 (65%)
Mean ± SD 83.6 ± 4.3 81.5 ± 6.43 82.55 ± 5.52
Median (25th–75th percentile) 84 (81.25–86.75) 83 (79.25–87.5) 84 (80–87) 0.143‡
Range 70–90 66–90 66–90
At 12 months
61–70 0 (0%) 1 (3.33%) 1 (1.67%)
71–80 2 (6.67%) 4 (13.33%) 6 (10%) 0.613*
81–90 23 (76.67%) 22 (73.33%) 45 (75%)
91–95 5 (16.67%) 3 (10%) 8 (13.33%)
Mean ± SD 87.37 ± 3.93 85.8 ± 5.63 86.58 ± 4.88
Median (25th–75th percentile) 88 (86–90) 88 (84–90) 88 (84.75–90) 0.217‡
Range 76–92 68–91 68–92

‡Independent t test.

*Fisher’s exact test.

Complications

Complications were minimal. Subsidence, defined as the lateral migration of the lag screw within the femoral head, was absent in the Halifax group but present in 2 patients (6.7%) in the PFNA-2 group; however, this difference was not statistically significant (p = 0.49) as shown in Table 3 and Figure 1C. The Halifax Nail’s innovative design offers superior screw positioning and demonstrated no subsidence in this cohort, suggesting it may be advantageous for patients with compromised bone quality. No cases of surgical site infection were reported in either group, indicating effective perioperative care and implant compatibility. No instances of non-union were noted in either group during follow-up, highlighting the effectiveness of both devices in achieving proper fracture healing.

Table 3.

Comparison of subsidence between Halifax and PFNA2.

Subsidence Halifax (n = 30) PFNA2 (n = 30) Total p-value
No 30 (100%) 28 (93.33%) 58 (96.67%)
Yes 0 (0%) 2 (6.67%) 2 (3.33%) 0.492*
Total 30 (100%) 30 (100%) 60 (100%)

*Fisher’s exact test.

Representative radiographic examples of pre-operative fracture patterns, immediate post-operative fixation with both implant types, subsidence complications, and complete construct assessment are illustrated in Figures 2 and 3.

Figure 2.

Figure 2

(A): PFNA-2 Pre-operative AP pelvis showing right intertrochanteric fracture. (B): PFNA-2 Post-operative radiographs (AP and lateral) demonstrating helical blade. (C): Halifax Nail Pre-operative AP pelvis showing right intertrochanteric fracture. (D): Halifax Nail Post-operative radiographs (AP and lateral) demonstrating screw via which tri-wire technology™ is employed.

Figure 3.

Figure 3

Complication of PFNA-2 fixation, AP and lateral radiographs demonstrating subsidence.

Discussion

The management of unstable intertrochanteric fractures remains challenging, particularly among elderly patients with osteoporosis. The Halifax Femoral Nail and PFNA-2 represent advanced solutions with unique biomechanical innovations aimed at enhancing stability, minimizing complications, and improving recovery.

Both groups showed comparable age and gender distributions. The mean age (Halifax: 71.3 ± 5.2 years; PFNA-2: 71.6 ± 5.1 years) aligns with the known prevalence of intertrochanteric fractures in the elderly population. The balanced gender representation reflects broader trends reported in similar studies, where fractures are equally common among men and women over 70 years. The primary mode of injury – slip-and-fall accidents – accounted for 70% of cases, consistent with low-energy trauma typically associated with such fractures. These observations align with findings by Simmermacher et al. and Singh and Bhartiya [12, 21].

Both devices demonstrated efficient surgical timelines. Halifax Nail showed a slightly shorter injury-to-surgery interval (3.4 ± 0.9 vs. 3.7 ± 1.4 days), though not statistically significant, indicating a marginal advantage in preoperative planning. Similarly, intraoperative blood loss was marginally lower in the Halifax group (89.5 ± 12.6 vs. 97.7 ± 22.9 ml). These differences may reflect Halifax Nail’s streamlined design, reducing surgical trauma. These trends are consistent with studies by Senthilnathan et al. [22].

Halifax Nail demonstrated a slight advantage in operative time, requiring an average of 31.7 ± 5.5 vs. 33.0 ± 7.9 min for PFNA-2. Although not statistically significant, this highlights Halifax Nail’s ease of insertion. Zou et al. similarly reported longer operative times for PFNA-2 due to helical blade design requiring precise positioning [23].

Radiological outcomes highlighted device performance differences. Both maintained alignment effectively, with comparable mean NSA differences postoperatively (Halifax: 5.0 ± 2.2°; PFNA-2: 5.3 ± 2.1°). This consistency reflects biomechanical stability provided by both devices, corroborated by Knobe et al., who emphasized the importance of implant design in preserving alignment [19]. TAD showed similarly optimal values in both groups, with Halifax achieving slightly higher mean TAD (22.4 ± 1.8 mm) than PFNA-2 (22.0 ± 2.6 mm). Studies by Yam et al. stress the importance of maintaining TAD under 25 mm to minimize screw cut-out risk, achieved by both devices [18].

Halifax Nail demonstrated clear superiority in screw placement, assessed by the Cleveland Index. The higher CC rate with the Halifax Nail likely reflects implant biomechanics rather than surgeon preference. PFNA-2 blades are often placed CI to improve purchase in the denser inferior cortex and reduce migration risk without a locking mechanism. In contrast, the Halifax lag screw is secured with a locking set screw and requires central alignment for its tri-wire technology™, thereby favoring CC placement. Optimal CC positioning was achieved in 76.7% of Halifax cases vs. 43.3% of PFNA-2 cases, a statistically significant difference (p = 0.008). This finding is consistent with Halifax Nail’s innovative tri-wire technology™, enhancing precision during implantation. By contrast, PFNA-2’s helical blade design has been associated with higher rates of suboptimal positioning, contributing to complications such as rotational instability and screw migration, as noted by Simmermacher et al. [12].

Functional outcomes, measured using the HHS, showed significant improvements in both groups over time. At 6 and 12 months, the Halifax group consistently achieved slightly higher mean HHS scores than PFNA-2, reflecting a trend toward superior functional recovery. While differences were not statistically significant, they align with findings by Schipper et al., who reported similarly effective functional outcomes for PFNA-2 and other intramedullary devices [24].

Complication rates provided further insights into relative advantages. Subsidence, defined as the lateral migration of the lag screw, was absent in the Halifax group but observed in 6.7% of PFNA-2 cases. This highlights Halifax Nail’s advantage in preventing screw migration, likely due to the patented set screw mechanism, and hence demonstrating a potential for reduced subsidence. By contrast, PFNA-2’s helical blade, while effective in achieving cancellous bone compression, has been associated with higher migration and subsidence risks, particularly in osteoporotic patients. These trends align with studies by Kammerlander et al., who identified subsidence as a common limitation of PFNA-2 in patients with compromised bone quality [17].

Limitations

Small sample size, single-center study.

Conclusion

Both PFNA-2 and Halifax Nail are effective for managing unstable trochanteric fractures. The Halifax Nail’s innovative design offers superior screw positioning and demonstrates a reduced trend towards subsidence, making it advantageous for patients with compromised bone quality. The Halifax Nail achieved significantly superior CC screw positioning according to the Cleveland Index, which may contribute to improved biomechanical stability. These findings support individualized implant selection based on patient factors and bone quality assessment. Further research is needed to confirm these findings with longer follow-up periods.

Cite this article as: Nayak SR, Rastogi A, Sinha S, Manjhi B, Rai R & Gandhi S (2026) The impact of subsidence on hip function: A comparative study between Proximal Femoral Nail Anti-rotation-2 and Halifax Femoral Nail in unstable trochanteric fractures. SICOT-J 12, 58. https://doi.org/10.1051/sicotj/2026027.

Funding

This research did not receive any specific funding.

Conflicts of interest

The authors declare that they have no relevant financial or non-financial interests to report.

Data availability statement

Data from the study is available on request from the corresponding author due to privacy restrictions.

Author contribution statement

Author 1: Concepts, Design, Definition of intellectual content, Literature search, Clinical studies, Experimental studies, Data acquisition, Data analysis, Statistical analysis.

Author 2: Concepts, Design, Definition of intellectual content, Literature search, Clinical studies, Experimental studies, Data acquisition, Data analysis, Statistical analysis.

Author 3: Concepts, Design, Definition of intellectual content, Literature search, Clinical studies, Experimental studies, Data acquisition, Data analysis, Statistical analysis.

Author 4: Data acquisition and review.

Author 5: Clinical studies, Data acquisition, Data analysis, Statistical analysis.

Author 6: Clinical studies, Data acquisition, Data analysis, Statistical analysis.

Ethics approval

This study received ethical approval from the Institutional Ethics Committee [2023/EC/6105].

Informed consent

Written informed consent was obtained from all participants.

References

  • 1.Haentjens P, Autier P, Barette M, et al. (2007) Survival and functional outcome according to hip fracture type: a one-year prospective cohort study in elderly women with an intertrochanteric or femoral neck fracture. Bone 41, 958–964. [DOI] [PubMed] [Google Scholar]
  • 2.Lindskog DM, Baumgaertner MR (2004) Unstable intertrochanteric hip fractures in the elderly. J Am Acad Orthop Surg 12, 179–190. [DOI] [PubMed] [Google Scholar]
  • 3.Fox KM, Magaziner J, Hebel JR, et al. (1999) Intertrochanteric versus femoral neck hip fractures: differential characteristics, treatment, and sequelae. J Gerontol Ser A Biol Sci Med Sci 54, M635–M640. [DOI] [PubMed] [Google Scholar]
  • 4.Hammer AJ (1988) Intertrochanteric fractures in an elderly white South African population. S Afr Med J 74, 124–126. [PubMed] [Google Scholar]
  • 5.Chan G, Hughes K, Barakat A, et al. (2021) Inter- and intra-observer reliability of the new AO/OTA classification of proximal femur fractures. Injury 52, 1434–1437. [DOI] [PubMed] [Google Scholar]
  • 6.Tunio ZH, Jhatiyal RA, Azeem Akhund M, Abbasi MK, Ali SM, Jokhio MF (2021) AO/OTA 31A2 and 31A3 intertrochanteric fractures managed by proximal femoral nail antirotation (PFNA). PJMHS 15, 2616–2618. [Google Scholar]
  • 7.Fu C-W, Chen J-Y, Liu Y-C, et al. (2020) Dynamic hip screw with trochanter-stabilizing plate compared with proximal femoral nail antirotation as a treatment for unstable AO/OTA 31-A2 and 31-A3 intertrochanteric fractures. BioMed Res Int 2020, 1–7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Barrios C, Walheim G, Broström LA, et al. (1993) Walking ability after internal fixation of trochanteric hip fractures with Ender nails or sliding screw plate: a comparative study of gait. Clin Orthop Relat Res 1993, 187–192. [PubMed] [Google Scholar]
  • 9.Siddiqui YS, Khan AQ, Asif N, et al. (2019) Modes of failure of proximal femoral nail (PFN) in unstable trochanteric fractures. MOJOR, 11. [Google Scholar]
  • 10.Favreau H, Ehlinger M, Adam P, et al. (2020) Total hip arthroplasty with exclusive use of dual-mobility cup after failure of internal fixation in trochanteric fracture. Orthop Traumatol Surg Res 106, 645–649. [DOI] [PubMed] [Google Scholar]
  • 11.Gogna P, Singh R, Rohilla R, et al. (2013) Radiological and functional outcome in unstable, osteoporotic trochanteric fractures stabilized with dynamic helical hip system. Strat Trauma Limb Reconst 8, 117–122. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Simmermacher RKJ, Ljungqvist J, Bail H, et al. (2008) The new proximal femoral nail antirotation (PFNA®) in daily practice: results of a multicentre clinical study. Injury 39, 932–939. [DOI] [PubMed] [Google Scholar]
  • 13.Chandan RK, Verma D, Meena RC, et al. (2016) Evaluation of osteoporotic trochanteric fractures treated with cement-augmented dynamic hip screw. Int J Res Orthop 2, 147. [Google Scholar]
  • 14.Makridis K, Georgoussis M, Mandalos V, et al. (2009) A prospective randomised study about two intramedullary devices for treating trochanteric fractures. Injury 40, S12. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Gupta SKV, Valisetti VS (2015) Comparative study between dynamic hip screw vs proximal femoral nailing in inter-trochanteric fractures of the femur in adults. Int J Orthop Sci 1, 07–11. [Google Scholar]
  • 16.Grønhaug KML, Dybvik E, Matre K, et al. (2022) Intramedullary nail versus sliding hip screw for stable and unstable trochanteric and subtrochanteric fractures: 17,341 patients from the Norwegian Hip Fracture Register. Bone Joint J 104-B, 274–282. [DOI] [PubMed] [Google Scholar]
  • 17.Kammerlander C, Hem ES, Klopfer T, et al. (2018) Cement augmentation of the proximal femoral nail antirotation (PFNA) – a multicentre randomized controlled trial. Injury 49, 1436–1444. [DOI] [PubMed] [Google Scholar]
  • 18.Yam M, Chawla A, Kwek E (2017) Rewriting the tip apex distance for the proximal femoral nail anti-rotation. Injury 48 1843–1847. [DOI] [PubMed] [Google Scholar]
  • 19.Knobe M, Nagel P, Maier K-J, et al. (2016) Rotationally stable screw-anchor with locked trochanteric stabilizing plate versus proximal femoral nail antirotation in the treatment of AO/OTA 31A2.2 fracture: a biomechanical evaluation. J Orthop Trauma 30, e12–e18. [DOI] [PubMed] [Google Scholar]
  • 20.She Z, Yang F, Zhang S, et al. (2023) A novel intramedullary nail design of intertrochanteric fracture fixation improved by proximal femoral nail antirotation. Comput Meth Biomech Biomed Eng 2023, 1–11. [DOI] [PubMed] [Google Scholar]
  • 21.Singh S, Bhartiya RK (2021) A comparative study of unstable intertrochanteric fracture by proximal femoral nail versus proximal femoral nail anti-rotation among adult patients. Gal Int J Health Sci Res 6(1), 27–33. [Google Scholar]
  • 22.Senthilnathan D, Prabhakar D, Raj D, et al. (2019) Our experience of managing intertrochanteric fractures in osteoporotic patients with trochanteric femoral nail (TFN) and proximal femoral anti-rotation augmentation nail (PFNA2). Int J Orthop Sci 2019, 287–292. [Google Scholar]
  • 23.Zou J, Xu Y, Yang H (2009) A comparison of proximal femoral nail antirotation and dynamic hip screw devices in trochanteric fractures. J Int Med Res 37, 1057–1064. [DOI] [PubMed] [Google Scholar]
  • 24.Schipper IB, Steyerberg EW, Castelein RM, et al. (2004) Treatment of unstable trochanteric fractures: Randomised comparison of the Gamma nail and the proximal femoral nail. J Bone Joint Surg Brit 86-B, 86–94. [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

Data from the study is available on request from the corresponding author due to privacy restrictions.


Articles from SICOT-J are provided here courtesy of EDP Sciences