ABSTRACT
Transesophageal echocardiography (TEE) is a cornerstone of cardiovascular diagnostics. A major advantage of TEE over transthoracic echo (TTE) is superior resolution of posterior structures such as the mitral valve and the left atrial appendage (LAA). TEE is invasive and requires the use of moderate and sometimes deeper levels of sedation. Anesthesia accounts for most TEE‐related risks including hemodynamic (hypotension, arrythmia) and respiratory compromise. Risks of anesthesia are amplified by the high prevalence of co‐morbidities in patients undergoing TEE. Anesthesia also requires specialized procedural space and skilled providers, limiting the availability of TEE, and increasing the overall cost of the procedure. Alternative TEE methods that minimize (or negate) the use of anesthesia would enhance workflow, expand availability, reduce risk, and may provide similar or more accurate information compared to conventional TEE. The nasal orifice is an alternative to the oral approach, and transnasal TEE can eliminate the need for anesthesia and expand availability of use, while minimizing risk of oropharyngeal injury. This review explores the emerging role of transnasal TEE as a viable alternative to the conventional oral approach and critically evaluates current technologies and clinical applications. Particular attention is given to unmet clinical needs, barriers to adoption, and future directions that may enable broader, safer, and more effective use of TEE in both procedural and non‐procedural settings.
Keywords: cardiac hemodynamics, cardiac imaging, transesophageal echo, transnasal
Transesophageal echocardiography (TEE) is a cornerstone of cardiovascular diagnostics. Alternative TEE methods that minimize the use of anesthesia would enhance workflow, expand availability, reduce risk, and may provide similar or more accurate information compared to conventional TEE (a). Transnasal TEE (b) can eliminate the need for anesthesia and expand availability of use, while minimizing risk of oropharyngeal injury.

1. Introduction
Transesophageal echocardiography (TEE) is an essential diagnostic tool in cardiology [1]. The main advantage over transthoracic echocardiography (TTE) is superior resolution of posterior cardiac structures, including the evaluation of the left atrial appendage (LAA) for thrombus [2, 3, 4] and in assessing the severity of mitral valve regurgitation (MR) [5, 6], given the elimination of imaging artifact from lung scatter which often degrades or obscures TTE images. More recently, the use of TEE has become standard practice in the diagnosis and management of structural heart disease to guide complex minimally invasive transcatheter procedures, owing to enhanced visualization of cardiac structures (in 2D and 3D) and intracardiac instrumentation. TEE has also become the go‐to imaging modality in the critical care setting when TTE imaging is sub‐optimal (recent sternotomy, bandages, pneumomediastinum, etc.) and there is concern for cardiac hemodynamic instability. TEE imaging allows for consistent visualization of cardiac structures and function, but also for assessment of real time hemodynamic parameters. Along these lines, continuous cardiac imaging has been shown to be effective for guiding management in critically ill patients [7]. However, TEE is invasive and requires the use of anesthesia prior to insertion of the TEE probe into the esophagus, which necessitates specialized space and skilled providers, increasing cost, patient risk, and limiting availability [8]. Up to 80% of TEEs results in at least mild oropharyngeal injury, with up to 40% resulting in complex lesions [9, 10]. Anesthesia accounts for most TEE‐related risks including hemodynamic (hypotension, arrythmia) and respiratory compromise; yet is often crucial for procedural success even with adequate topicalization [11, 12, 13]. The serious risks of anesthesia are amplified by the high prevalence of co‐morbidities in patients undergoing TEE, particularly in critically ill patients. Anesthesia‐induced hypotension can also confound the assessment of MR, which is highly sensitive to changes in blood pressure and afterload [14, 15, 16]. Alternative TEE methods that minimize (or negate) the use of anesthesia will enhance workflow, expand availability, reduce risk, and may provide similar or more accurate information compared to conventional TEE. Conversely, intracardiac echo (ICE) probes (inserted into the vascular space for use in invasive cardiac procedures) are only 3–4 mm in diameter; although ICE is not an ideal alternative to TEE due to increased invasiveness (requiring dedicated venous access), the need for specialized procedural space, and a highly skilled procedural team.
The nasal orifice is an alternative to the oral approach (Figure 1), and transnasal TEE, with the appropriate equipment, can eliminate the need for anesthesia, reduce risk of oropharyngeal injury, and allow for physiologic data in the unsedated patient. We define transnasal TEE as the insertion of an ultrasound transducer through the nasal passage into the nasopharynx and subsequently guided into the esophagus or stomach for cardiac visualization. Early investigations relied upon custom‐engineered miniature TEE probes (6.0–7.3 mm diameter). As the technique evolved, subsequent studies established the feasibility of utilizing standard pediatric TEE probes (approximately 7.0–8.0 mm) via a transnasal approach. Building upon these findings, several reports have since documented the ‘off‐label’ use of intracardiac echocardiography (ICE) catheters as de facto TEE probes [17, 18, 19, 20, 21, 22]. A recent comprehensive review delineates the relative strengths and limitations of these various probes within the context of structural heart disease (SHD) guidance [23, 24]. While that review does not specifically address the transnasal TEE approach, it provides a robust framework for understanding the performance characteristics for each probe class. The present discussion builds upon those comparisons by examining the specific utility of these technologies when deployed via the nasal route. Table 1 summarizes key characteristics of discussed cardiac probes, and Figure 2 shows the relative size of a standard TEE, pediatric (micro), and ICE probes.
FIGURE 1.

(a) Schematic demonstrating transesophageal echo (transoral), and (b) Transnasal transesophageal echo.
TABLE 1.
Comparison of the typical characteristics of echocardiography probes.
| Standard TEE | Pediatric (Micro) TEE | Intracardiac Echo (ICE) | |
|---|---|---|---|
| Physical specifications | |||
| Primary use | Adults (>20–25 kg) | Infants/Children (> 2.5 kg) | Interventional procedures |
| Anesthesia | GA, DS, MS | GA, DS, MS | LA |
| Probe diameter | Large (∼10–16 mm) | Medium to Small (∼5–8 mm) | Very Small (2.7–3.3 mm) |
| # Transducer elements | 2500+ | 32–48 | 64 (Acunav) vs 840 (Verisight, Nuvision) |
| Performance and clinical characteristics | |||
| Access route | Esophageal (Oral) | Esophageal (Oral) | Vascular (Femoral Vein) |
| Resolution | Gold standard | Good; smaller field of view | Good near‐field; limited far‐field |
| Key rdvantage | Comprehensive 3D/4D | Less traumatic; used in small infants | No general anesthesia; operator‐controlled |
| Key risk | Esophageal injury, aspiration | Probe heating | Vascular injury, pericardial effusion |
| Transnasal suitability | Too large, not feasible | Multiplane capable, safe (Especially when used with nasal conduit) | Monoplane, lower image quality, good safety profile in limited studies |
| Cost | Reusable (Moderate cost per use) | Reusable (Moderate cost per use) | High (Single‐use disposable probe) |
Abbreviations: DS, Deep Sedation; GA, General Anesthesia; MS, Moderate Sedation.
FIGURE 2.

Size comparison of (a) Philips Verisight ICE Probe (3 mm tip width), (b) Philips S8 Pediatric probe (7.5 mm tip width), (c) Philips X8‐2T TEE probe (16.9 mm tip width).
In this review, we highlight the existing literature on transnasal TEE, critically evaluate current technologies and clinical applications, and explore emerging innovations that may further redefine the role of esophageal echocardiography. Particular attention is given to unmet clinical needs, barriers to adoption, and future directions that may enable broader, safer, and more effective use of TEE in both procedural and non‐procedural settings.
2. Methods
To conduct this review, we conducted a comprehensive literature search across major medical databases, including PubMed and Scopus, to identify relevant studies on transnasal echocardiography. The search strategy utilized a combination of keywords such as “transnasal echo”, “transnasal echocardiography”, “nasal echo”, “nasal echocardiography”. Peer‐reviewed articles were evaluated by three different clinicians to determine which articles were appropriate for inclusion. The initial search yielded a total of 70 records. After removing 21 duplicates, 49 records remained for title and abstract screening. During the initial screening, 23 articles were excluded as they did not involve transnasal TEE. The remaining 26 articles were assessed for eligibility. A further nine articles were excluded based on predefined criteria: four were editorials or meeting abstracts, two were published in non‐peer‐reviewed journals, two focused on pediatric populations, one was a non‐clinical publication, one article could not be accessed. A total of 17 articles were included in this scoping review. Data was extracted from the articles by the authors and included type of study, size of study, probe utilized (size, and capabilities), success of nasal intubation, patient comfort, adverse reactions, and image quality.
2.1. Transnasal TEE Using Mini TEE Probes
The earliest report of transnasal TEE was established by Spencer et al. in 1997 [25]. Their seminal work introduced a newly developed mini monoplane TEE probe with a 32‐element transducer capable of pulse wave and color Doppler. The probe's imaging head measured 7.3 mm in maximal width. This initial study aimed to demonstrate transnasal TEE feasibility (safety/tolerability), adequate image quality, and a new‐use modality as a continuous monitor for LV function (leave‐in’ probe). The study included a comparison arm which utilized a traditional pediatric multiplane TEE probe (9.1 mm width). A total of 128 patients were enrolled, all of whom received local anesthesia to both the nasal cavity and oropharynx. Early in the study, minor mucosal bleeding prompted the investigators to introduce a modified soft plastic nasal trumpet as a conduit for probe insertion; this modification, implemented in the final 37 patients, effectively mitigated further bleeding. Transnasal intubation was generally considered easy, with and without the nasal trumpet, although eight attempts were unsuccessful, most commonly due to difficulty advancing the probe through the proximal nasal passage. Patient tolerance was comparable between transnasal and oral insertion and did not differ significantly between moderate and deep sedation. Image quality of the transnasal probe was considered adequate for assessment of cardiac structure and function but was significantly limited by a lack of multiplane capability. Notably, eight patients in the intensive care unit had the transnasal probe stabilized in the transgastric short‐axis view for up to 2 h for continuous cardiac monitoring. This study was pivotal in demonstrating the feasibility of transnasal TEE. 2 years later, the authors reported similar findings in a larger multicenter cohort using the same custom miniaturized monoplane probe, further reinforcing the feasibility of this approach [26]. In 1998 [27], Spencer and colleagues extended this concept by performing proof‐of‐principle transnasal stress echocardiography in 15 healthy volunteers. Using the same 7.3 mm monoplane probe inserted transnasally via a nasal trumpet. The probe was positioned in the stomach to obtain transgastric LV short‐axis views at the level of the papillary muscles. Participants exercised for a mean duration of 10 ± 2.3 min, with most reporting minimal or mild nasopharyngeal discomfort. Although limited to healthy individuals, this study demonstrated the feasibility of unsedated, exercise‐based transnasal TEE.
In 1999, Greim et al. [28], evaluated the utility of the same miniaturized monoplane probe for transnasal LV assessment in critically ill patients with suspected cardiac dysfunction. In total, 42 mechanically ventilated patients underwent transnasal TEE (with conventional TEE serving as a comparator). Image quality for LV short‐axis and four‐chamber long‐axis views did not differ significantly between transnasal and conventional monoplane TEE. However, the authors concluded that the absence of multiplane transnasal probes limited the role of nasal TEE for comprehensive examinations at that time.
In 2009, Fukuda et al. [29], assessed the feasibility of transnasal TEE for detection of LAA thrombus in patients with atrial fibrillation or prior stroke. A monoplane microprobe (Aloka UST‐5293S; 6.0 mm width) was inserted transnasally without sedation following local nasal anesthesia. Prior to probe insertion, a soft plastic conduit was placed in the nasal cavity. The insertion of the TEE probe was successful in 52 (84%) of 62 patients. Eight patients had unsuitable nasal anatomy, and two were intolerant of probe insertion. Despite 56% of patients receiving warfarin therapy, only two cases of mild, self‐limiting epistaxis were observed. Patient‐reported tolerability was high, with 72% reporting no discomfort during nasal insertion. Transnasal TEE identified a potential intracardiac embolic source in 14 patients, including LAA thrombus (n = 10), spontaneous echocardiographic contrast (n = 2), patent foramen ovale (PFO; n = 1), and mobile aortic plaque (n = 1). Limitations included the absence of a conventional multiplane TEE comparator and reliance on early‐generation monoplane technology.
Early mini‐TEE probes were characterized by several inherent technical limitations that constrained their diagnostic utility. These included a reduced element count, and monoplane imaging capabilities that precluded multiplane assessment of complex cardiac structures. Furthermore, these early‐generation probes lacked three‐dimensional (3D) reconstruction capabilities. Despite these hardware restrictions, the clinical application of mini‐TEE represented a significant procedural milestone, establishing feasibility of performing diagnostic TEE in completely unsedated patients. Literatures describing the use of mini‐TEE probes for transnasal TEE are summarized in Table 2.
TABLE 2.
Summary of publications of transnasal TEE using mini TEE probes.
| Author | Year | Study Type | No. | Probe (head width) | Plane View | Type of Anesthesia | Use of Nasal Trumpet/Conduit | Success of Nasal Intubation | Adverse Outcomes | Patient Comfort | Image Quality (Rated by Clinicians) |
|---|---|---|---|---|---|---|---|---|---|---|---|
| Spencer et al. [25]. | 1997 | Prospective Study | 128 | HP mini TEE (7.3 mm) | Mono | Local nare/oropharynx and conscious sedation | Yes (for the last 37 patients) | 81% (88% when using the nasal conduit) | 1.6% (1 patient) with epistaxis requiring intervention | 84% Good, 11% acceptable, 5% poor | Acceptable/good in all cases |
| Spencer et al. [27]. | 1998 | Prospective Study | 15 | HP mini TEE (7.3 mm) | Mono | Local nare | Yes | 87% | None | 85% minimal to mild discomfort | Acceptable/good in all cases |
| Spencer et al. [26]. | 1999 | Prospective Study | 194 | HP mini TEE (7.3 mm) | Mono | Local nare/oropharynx ± conscious sedation | Yes | 88% | 12% mild epistaxis | 96% excellent/good, 4% poor | 97% acceptable |
| Greim et al. [28]. | 1999 | Prospective Study | 42 | HP mini TEE (7.3 mm) | Mono | Local nare/oropharynx ± conscious sedation | No | 90% | 31% epistaxis, 1 patient required treatment | Overall good, but patients sedated and intubated | Adequate but limited compared to multiplane probe |
| Greim et al. [30]. | 2001 | Prospective Study | 57 | HP mini TEE (7.3 mm) | Mono | Local nare/oropharynx + conscious sedation | No | 98% | None | Overall good | Acceptable/good in all cases |
| Zimmerman et al. [31]. | 2003 | Prospective Study | 45 | HP mini TEE (7.3 mm) | Mono | Local nare/oropharynx ± conscious sedation | No | 96% | None | Overall good (1 aborted due to discomfort) | 73% good, 23% moderate |
| Fukuda et al. [29]. | 2009 | Prospective Study | 62 | Aloka UST‐5293S (6.0 mm) | Mono | Local nare/oropharynx | Yes | 84% | 3% (2 patients) mild epistaxis | 72% no pain/discomfort | Not quantified |
2.2. Transnasal TEE Using Pediatric TEE Probes
The transition to pediatric micro‐TEE probes represented an important technological advancement in transnasal imaging. While these transducers typically maintained a similar element density to their “mini‐probe” predecessors (ranging from 32–48 elements), they introduced multiplane imaging capabilities. Although these probes remained limited by a lack of volumetric (3D) reconstruction, the integration of an adjustable imaging matrix permitted a more comprehensive assessment of complex cardiac morphology. This shift from monoplane to multiplane arrays significantly enhanced the clinical utility of transnasal TEE by allowing for standard cross sectional views and more precise valvular and structural characterization. Literatures describing the use of pediatric TEE probes for transnasal TEE are summarized in Table 3.
TABLE 3.
Summary of publications of transnasal TEE using pediatric (Micro) TEE probes.
| Author | Year | Study type | No. | Probe (head width) | Plane view | Type of anesthesia |
Use of nasal trumpet/ Conduit |
Success of nasal intubation | Adverse outcomes | Patient comfort | Image quality (Rated by clinicians) |
|---|---|---|---|---|---|---|---|---|---|---|---|
| Klettas et al. [32]. | 2017 | Prospective study | 24 | Philips S8‐3T (7.5 mm) | Multiplane | Local nare/oropharynx ± conscious sedation | Yes | 100% | 8% (2 patients) hypoxia (secondary to sedation) | 100% good | Good to very good in all cases |
| Wang et al. [35]. | 2019 | Case report | 1 | Philips S8‐3T (7.5 mm) | Multiplane | Unclear | No | 100% (single case) | None | Well tolerated | Good |
| Sun et al. [33]. | 2020 | Prospective study | 34 | Philips S8‐3T (7.5 mm) | Multiplane | Local nare | No | 100% | 7% (2 patients) mild epistaxis | Well tolerated | Good (66.7%) and acceptable (33.3%) |
| Lanki et al. [34]. | 2024 | Retrospective/Prospective study | 364 | Philips S8‐3T (7.5 mm) | Multiplane | Local nare/oropharynx ± conscious sedation | No | 98% | 5.9% (10 patients) epistaxis (4 patients required treatment) | Not mentioned | 88% good or acceptable, but all cases adequate for study |
Klettas et al. [32], were the first to perform transnasal TEE with a multiplane capable probe. They utilized the Philips S8‐3T pediatric micro‐TEE probe. The Philips S8‐3T is a TEE microprobe originally designed for pediatric use, with extended operating frequency range of 3–8 MHz and physical dimensions at the tip width of 7.5 mm. The microprobe has multiplane capability and is equipped with steerable PW Doppler, CW Doppler, color Doppler, M‐mode, and harmonic imaging. The study examined whether transnasal TEE could provide sufficient image quality and patient tolerability to avoid general anesthesia during structural heart interventions. The study included 24 patients who underwent one of a transfemoral aortic valve implantation (TAVI), transfemoral mitral valve implantation (TMVI), atrial septal defect (ASD), or PFO closure. All patients underwent both transnasal TEE as well as conventional TEE. Although overall image quality was inferior to that of standard TEE, likely reflecting the substantial disparity in transducer element count compared with contemporary adult probes, anatomic delineation was sufficient for procedural guidance in all cases. No nasal bleeding was reported. The authors concluded that transnasal TEE represents a promising alternative imaging modality for interventional procedures, particularly as a means to avoid general anesthesia.
Sun et al. [33], subsequently used the same S8‐3T microprobe in 30 patients undergoing non‐sedated LAA occlusion. All participants tolerated the transnasal probe with local anesthesia alone. Two cases of mild, self‐resolving epistaxis were reported. Image quality was rated as good in 66.7% and acceptable in 33.3% of cases by independent echocardiographers. All devices were successfully implanted, and follow‐up TEE at a median of 55 days demonstrated no significant residual leaks or device‐related thrombus.
In the most recent and largest trial to date, Lanki et al. [34], evaluated 336 patients undergoing PFO closure at Helsinki University Hospital from 2003–2021. Conventional TEE under general anesthesia was used until November 2018 (n = 167) after which nasal TEE was performed using the Philips S8‐3T (n = 169). Among patients undergoing transnasal TEE, epistaxis occurred in nine cases (5.3%), with four patients (2.4%) requiring intervention. One patient experienced prolonged nasal congestion lasting two months. Image quality was rated as acceptable to good in 88% of cases. Direct comparison with conventional TEE was not performed, as image quality, even when suboptimal, was sufficient for procedural success. Overall, transnasal TEE was successfully performed in 97.2% of cases.
2.3. Transnasal TEE Using ICE Probes
Intracardiac echocardiography (ICE) was introduced into clinical practice in 2000 (Acuson AcuNav 10F) primarily for endovascular guidance during invasive electrophysiological and structural procedures. These catheters, typically ranging from 8F–10F (2.7–3.3 mm) in diameter, feature a 64‐element phased‐array transducer capable of four‐way steering and full Doppler functionality. Operating at frequencies between 5.5 and 10 MHz, ICE provides a 90° sector view with a penetration depth of up to 16 cm. While ICE facilitates complex structural interventions without the need for general anesthesia, its conventional use is limited by its inherent invasiveness requiring dedicated large‐bore venous access, as well as the necessity for a specialized procedural suite and a highly skilled operator. However, an alternative, less invasive approach to both standard TEE and ICE, is the off‐label use of intracardiac echocardiography (ICE) probes transnasally as a de‐facto TEE probe. Several studies have looked at this and are summarized below. Literatures describing the use of ICE probes for transnasal TEE are summarized in Table 4.
TABLE 4.
Summary of publications of transnasal TEE using ICE probes.
| Author | Year | Study type | No. | Probe (head width) | Plane view | Type of anesthesia | Use of nasal Trumpet/Conduit | Success of nasal intubation | Adverse outcomes | Patient comfort | Image quality (Rated by clinicians) |
|---|---|---|---|---|---|---|---|---|---|---|---|
| Tallarico et al.17 | 2005 | Case report | 1 | Siemens AcuNav (3.3 mm) | Mono | None | No | 100% (single case) | None | Well tolerated | Adequate |
| Tallarico et al. [18]. | 2005 | Case report | 1 | Siemens AcuNav (3.3 mm) | Mono | None | No | 100% (single case) | None | Well tolerated | Adequate |
| Tallarico et al. [19]. | 2006 | Case report | 1 | Siemens AcuNav (3.3 mm) | Mono | None | No | 100% (single case) | None | Well tolerated | Adequate |
| Greco et al. [21]. | 2008 | Prospective study | 20 | Siemens AcuNav (3.3 mm) | Mono | None | No | 100% | None | Well tolerated | Acceptable/good in all cases |
| Schuster et al. [22]. | 2010 | Case report | 2 | Siemens AcuNav (3.3 mm) | mono | None | No | 100% (two cases) | None | Well tolerated | Adequate |
| Mitchell‐Heggs et al. [20]. | 2010 | Prospective study | 79 | Siemens AcuNav (3.3 mm) | mono | None | No | 100% | None | Well tolerated | Satisfactory in all cases |
In 2006, Tallarico et al. [19], reported a case of transnasal TEE using a 10‐French (3.3 mm) Siemens AcuNav ICE catheter (monoplane, 5.5–10 MHz frequency range, 50 transducer elements) in a 42‐year‐old man with recurrent cryptogenic stroke. Conventional TEE identified an atrial septal aneurysm and PFO, and the patient subsequently underwent percutaneous PFO closure. To avoid general anesthesia and additional vascular access, the ICE probe was inserted transnasally without topical or general anesthesia. Image quality was sufficient to visualize the atrial septal anatomy, detect right‐to‐left shunting, and guide device deployment. Follow‐up TEE confirmed appropriate device positioning, and no procedural complications were reported.
Greco et al. [21], expanded on this approach in 2008 arguing that PFO closure is often unnecessarily prolonged by conventional TEE and the requirement for general anesthesia. They employed the same 10‐French AcuNav ICE catheter in 20 patients with ischemic stroke and PFO confirmed on prior TEE. All patients were receiving dual antiplatelet therapy. The probe was inserted transnasally without sedation, and patients remained awake and communicative throughout the procedure. All procedures were completed successfully, with high patient satisfaction and good tolerability. Similarly, Schuster et al. [22], described two additional cases in which transnasal TEE using the AcuNav ICE probe facilitated atrial fibrillation radiofrequency ablation. Both procedures were completed successfully with acceptable procedural durations.
The largest experience with ICE‐based transnasal TEE was reported by Mitchell‐Heggs et al. [20] in 2010, involving 79 patients undergoing atrial fibrillation ablation. All patients were on uninterrupted anticoagulation and had undergone conventional pre‐procedural TEE. A 10‐French AcuNav probe was inserted transnasally without sedation, and imaging focused on the atrial septum during transseptal puncture. The probe was withdrawn after successful left atrial access. All patients tolerated the procedure well, and image quality was considered adequate in every case. Notably, transnasal TEE identified three PFOs that had been missed on prior conventional TEE. The authors concluded that this technique offers a safe, well‐tolerated, sedation‐free, and potentially lower‐cost alternative to conventional TEE.
Despite demonstrating proof of concept, these studies were limited by suboptimal image quality due to outdated ICE probe technology, inconsistent probe‐esophageal contact [21], difficulty maneuvering the ICE probe following insertion [21], and traumatic bleeding [36]. Therefore, further technological innovation is required to overcome these limitations and fully realize the potential of transnasal TEE.
3. Discussion and Future Directions
A summary of published reports evaluating transnasal transesophageal TEE is provided in Tables 2, 3, 4, corresponding to mini probes, pediatric micro probes, and ICE probes, respectively. Collectively, these studies suggest that transnasal TEE represents a feasible alternative imaging modality in several clinical scenarios.
First, transnasal TEE may be particularly useful when conventional TEE is contraindicated. Contraindications may arise from anatomic limitations (oral lesions, severe trismus), patient intolerance or preference, or an elevated risk associated with sedation or general anesthesia. In such cases, the use of a smaller‐caliber probe inserted via the nasal route may provide diagnostic information comparable to conventional TEE. Recently, the DIONISO study (Intracardiac Echocardiographic Probe Used via Transesophageal to Guide LAA Occlusion) evaluated the feasibility of using the Acuson AcuNav 8F ICE probe, inserted transorally, to provide procedural guidance during LAA occlusion (LAAO) [37]. This study confirmed that a micro‐caliber transducer can provide sufficient visualization to support routine LAAO interventions. Building on this, a 2025 pooled analysis by Laterra et al. integrated data from the FLXiEST and DIONISO registries, utilizing 1:1 propensity score matching to compare the efficacy of gold‐standard TEE against transesophageal ICE (TE‐ICE) [38]. Their results demonstrated equivalent procedural success rates between the two modalities. These findings suggest that TE‐ICE can effectively reduce the clinical dependence on general anesthesia without compromising outcomes. Perhaps most significantly, the study validated that the significantly smaller footprint of an ICE probe does not preclude the high‐fidelity imaging required for complex structural heart interventions when imaging from the esophagus.
Second, transnasal TEE may be advantageous when rapid or serial assessment of cardiac structures is required and TTE is non‐diagnostic. This is commonly encountered in patients with recent sternotomy, severe obesity, advanced pulmonary disease, trauma, or in critically ill patients. In postoperative and intensive care unit (ICU) settings, TEE has demonstrated utility in evaluating hemodynamic instability, either independently or in conjunction with invasive hemodynamic monitoring [39, 40, 41, 42, 43]. However, conventional TEE is associated with procedural risks, and although major complications are uncommon, they can include significant bleeding, esophageal perforation [9], and hemodynamic compromise [9, 11], risks that may be amplified in critically ill patients requiring repeated examinations.
In an effort to improve safety and usability in this population, IMACOR (New York, NY, USA) [7] developed a transoral monoplane microprobe (ClariTEE), measuring 5.5 mm at the imaging head, which has received U.S. Food and Drug Administration approval for indwelling use of up to 72 h for hemodynamic monitoring. Following insertion, the probe handle can be detached, allowing the probe to remain in situ for intermittent, on‐demand imaging while the ultrasound system is used for other patients. This approach represents an innovative alternative to full‐size or pediatric TEE probes and has demonstrated clinical utility in the ICU setting [44]. However, inherent technological limitations, most notably the absence of multiplane imaging, restrict its broader application as a replacement for conventional TEE in perioperative, outpatient, and non‐ICU inpatient settings.
The concept of a “leave‐in” TEE probe is nonetheless compelling and aligns well with the needs of critically ill patients, particularly in those with poor transthoracic windows (i.e.post‐sternotomy). Continued advancements in probe technology, including multiplane capability, three‐dimensional imaging, and improved spatial resolution, could substantially expand the clinical utility of this approach.
Unsedated transnasal TEE allows patients to remain physiologically active during imaging. This enables dynamic assessment of mitral valve function, including exercise‐based interrogation (e.g., supine bicycle testing), which may reveal clinically significant mitral regurgitation or stenosis that could be underestimated in sedated patients under resting conditions [14, 15, 16].
Another important application involves evaluation for patent foramen ovale (PFO) in patients with cryptogenic stroke. Approximately 300,000 individuals experience cryptogenic stroke annually, and up to 40% are found to have a PFO [45, 46]. Conventional TEE is frequently used for PFO assessment [47]; however, sedation and anesthesia reduce intracardiac pressure gradients and preclude effective performance of a Valsalva maneuver, contributing to missed or misclassified diagnoses in up to 10% of cases [48]. In contrast, transnasal TEE permits evaluation in an awake, unsedated patient who can actively perform provocative maneuvers, thereby improving diagnostic sensitivity.
Recent advances in probe technology further support the feasibility of transnasal TEE. Both Philips (X11‐4t) and GE (9VT‐D) have introduced updated pediatric TEE probes with more than 2000 transducer elements and imaging capabilities approaching those of adult probes. Nevertheless, with imaging head widths of approximately 11 mm, these probes are unlikely to be suitable for transnasal insertion despite potentially reducing sedation requirements when used transorally.
Conversely, several manufacturers including Siemens (Volume ICE), Philips (VeriSight), and Johnson & Johnson Biosense Webster (NuVision), have developed U.S. Food and Drug Administration approved, three‐dimensional intracardiac echocardiography (ICE) catheters with diameters ranging from 3.0–4.2 mm. The VeriSight (3.0 mm) and NuVision (3.3 mm) systems, in particular, offer multiplane imaging and functionality that closely resemble conventional TEE. Adaptation of these probes for transnasal transesophageal use could allow for comprehensive, sedation‐free TEE studies with minimal patient discomfort. However, widespread adoption will require addressing economic and logistical barriers, as current 3D ICE probes are costly and designed for single‐use intravascular applications.
From a technical standpoint, miniaturized probes now possess many of the imaging capabilities as standard adult TEE systems. For transnasal TEE to gain broader acceptance, patient comfort must remain a central consideration. Evidence from the endoscopy literature consistently demonstrates superior tolerance and patient satisfaction with transnasal compared with transoral endoscopy, in both sedated and unsedated settings [49, 50, 51]. Smaller‐diameter probes, particularly those less than 5.3 mm, are associated with higher success rates and easier nasal insertion than slightly larger devices [52]. Additionally, as demonstrated in several studies reviewed herein, the use of a protective transnasal conduit or nasal trumpet may reduce mucosal trauma, improve patient comfort during probe manipulation, and potentially make a leave‐in probe more feasible.
4. Limitations
This review is constrained by the limited body of available clinical evidence, much of which reflects earlier technological constraints of available ultrasound probes. Most included studies were feasibility‐focused, with few direct head‐to‐head comparisons against conventional TEE and no randomized study designs, with most studies lacking advanced statistical methods such as propensity score matching or inverse probability weighting. Additionally, a substantial proportion of the literature predates 2010, likely reflecting a period of technological stagnation, as the Philips S8‐3t microprobe, introduced in 2004, lacked a true successor until the X11‐4t received FDA approval in early 2024. Similarly, advancements in intracardiac echocardiography lagged, with the original AcuNav ICE probes (introduced in 2000) only recently being supplanted by newer‐generation devices capable of multiplane and three‐dimensional imaging.
5. Conclusion
In conclusion, transnasal TEE represents a viable and complimentary addition to conventional TEE in patients with anatomic or hemodynamic contraindications, in those at elevated risk from sedation, and in clinical scenarios that may benefit from repeated or continuous cardiac assessment. While it is not intended to replace conventional TEE as a primary diagnostic tool, its development addresses specific unmet needs in specialized cardiac care scenarios. Optimization of a miniaturized, high‐resolution probe, whether an adapted pediatric TEE or ICE‐derived technology, capable of transnasal esophageal insertion has the potential to eliminate the need for sedation, reduce procedural risk, expand access, and provide more physiologically relevant diagnostic information. Further investigation is needed to evaluate the feasibility, performance, and cost‐effectiveness of emerging ultrasound probe technologies in this evolving clinical application.
Funding
This work was supported by the National Heart, Lung, and Blood Institute [R33HL171790‐02].
Ethics Statement
Not applicable.
Patient/Participant Consent statement
No patient or participant consent was required for this review article.
Conflicts of Interest
The authors declare no conflicts of interest. https://doi.org/10.13039/100000050
References
- 1. Otto C. M., Textbook of Clinical Echocardiography E‐book (Elsevier, 2016). [Google Scholar]
- 2. Zabalgoitia M., Halperin J. L., Pearce L. A., et al., “Transesophageal Echocardiographic Correlates of Clinical Risk of Thromboembolism in Nonvalvular Atrial Fibrillation. Stroke Prevention in Atrial Fibrillation III Investigators,” Journal of the American College of Cardiology 31 (1998): 1622–1626, 10.1016/S0735-1097(98)00146-6. [DOI] [PubMed] [Google Scholar]
- 3. Manning W. J., Weintraub R. M., Waksmonski C. A., et al., “Accuracy of Transesophageal Echocardiography for Identifying Left Atrial Thrombi. A Prospective, Intraoperative Study,” Annals of Internal Medicine 123 (1995): 817–822, 10.7326/0003-4819-123-11-199512010-00001. [DOI] [PubMed] [Google Scholar]
- 4. Hwang J. J., Chen J. J., Lin S. C., et al., “Diagnostic Accuracy of Transesophageal Echocardiography for Detecting Left Atrial Thrombi in Patients With Rheumatic Heart Disease Having Undergone Mitral Valve Operations,” American Journal of Cardiology 72 (1993): 677–681. [DOI] [PubMed] [Google Scholar]
- 5. Min S., Song J., Kim Y., Park H., et al., “Discrepancy Between Mitral Valve Areas Measured by Two‐Dimensional Planimetry and Three‐Dimensional Transoesophageal Echocardiography in Patients With Mitral Stenosis,” Heart 99 (2013): 253–258, 10.1136/heartjnl-2012-302742. [DOI] [PubMed] [Google Scholar]
- 6. Shanks M., Siebelink H. J., Delgado V., et al., “Quantitative Assessment of Mitral Regurgitation: Comparison Between Three‐Dimensional Transesophageal Echocardiography and Magnetic Resonance Imaging,” Circulation: Cardiovascular Imaging 3 (2010): 694–700. [DOI] [PubMed] [Google Scholar]
- 7. Cioccari L., Baur H., Berger D., et al., “Hemodynamic Assessment of Critically Ill Patients Using a Miniaturized Transesophageal Echocardiography Probe,” Critical Care 17 (2013): R121, 10.1186/cc12793. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8. Hahn R. T., Abraham T., Adams M. S., et al., “Guidelines for Performing a Comprehensive Transesophageal Echocardiographic Examination: Recommendations From the American Society of Echocardiography and the Society of Cardiovascular Anesthesiologists,” Journal of the American Society of Echocardiography 26 (2013): 921–964, 10.1016/j.echo.2013.07.009. [DOI] [PubMed] [Google Scholar]
- 9. Hilberath J. N., Oakes D. A., Shernan S. K., et al., “Safety of Transesophageal Echocardiography,” Journal of the American Society of Echocardiography 23, no. 11 (2010): 1115–1127, 10.1016/j.echo.2010.08.013. [DOI] [PubMed] [Google Scholar]
- 10. Freitas‐Ferraz Afonso B., Mathieu B., Rosaire V., et al., “Safety of Transesophageal Echocardiography to Guide Structural Cardiac Interventions,” Journal of the American College of Cardiology 75, no. 25 (2020): 3164–3173, 10.1016/j.jacc.2020.04.069. [DOI] [PubMed] [Google Scholar]
- 11. Mathur S. K. and Singh P., “Transoesophageal Echocardiography Related Complications,” Indian Journal of Anaesthesia 53 (2009): 567–574. [PMC free article] [PubMed] [Google Scholar]
- 12. Ferson D., Thakar D., Swafford J., et al., “Use of Deep Intravenous Sedation With Propofol and the Laryngeal Mask Airway During Transesophageal Echocardiography,” Journal of Cardiothoracic and Vascular Anesthesia 17 (2003): 443–446, 10.1016/S1053-0770(03)00147-2. [DOI] [PubMed] [Google Scholar]
- 13. Freitas T. M., David C., Almeida A. G., et al., “Cardiovascular and Respiratory Safety of Sedation Strategies Used in Transesophageal Echocardiography: a Systematic Review Incorporating Network Meta‐Analysis,” Journal of Cardiothoracic and Vascular Anesthesia 36 (2022): 4129–4140, 10.1053/j.jvca.2022.07.003. [DOI] [PubMed] [Google Scholar]
- 14. Grewal K. S., Malkowski M. J., Piracha A. R., et al., “Effect of General Anesthesia on the Severity of Mitral Regurgitation by Transesophageal Echocardiography,” American Journal of Cardiology 85 (2000): 199–203, 10.1016/S0002-9149(99)00644-X. [DOI] [PubMed] [Google Scholar]
- 15. Alachkar M. N., Kirschfink A., Grebe J., et al., “General Anesthesia Leads to Underestimation of Regurgitation Severity in Patients With Secondary Mitral Regurgitation Undergoing Transcatheter Mitral Valve Repair,” Journal of Cardiothoracic and Vascular Anesthesia 36 (2022): 974–982, 10.1053/j.jvca.2021.10.024. [DOI] [PubMed] [Google Scholar]
- 16. Chin J., Lee E., Choi D., et al., “The Effect of Depth of Anesthesia on the Severity of Mitral Regurgitation as Measured by Transesophageal Echocardiography,” Journal of Cardiothoracic and Vascular Anesthesia 26 (2012): 994–998, 10.1053/j.jvca.2012.05.012. [DOI] [PubMed] [Google Scholar]
- 17. Tallarico D., Chiavari P. A., Campolongo G., et al., “Transesophageal Echocardiography Through Nasal Way Diagnoses Bicuspid Aortic Valve,” Echocardiography 22 (2005): 149–150, 10.1111/j.0742-2822.2005.04046.x. [DOI] [PubMed] [Google Scholar]
- 18. Tallarico D., Chiavari P. A., Mollo P., et al., “Mitral Paraprosthetic Leak Diagnosed by Transesophageal Echocardiography Through Nasal Way,” European Journal of Echocardiography 6 (2005): 291–293, 10.1016/j.euje.2004.10.009. [DOI] [PubMed] [Google Scholar]
- 19. Tallarico D., Chiavari P. A., Mollo P., et al., “Transesophageal Echocardiography Through Nasal Way as a Guide to Percutaneous Closure of Patent Foramen Ovale,” Echocardiography 23 (2006): 790–792, 10.1111/j.1540-8175.2006.00313.x. [DOI] [PubMed] [Google Scholar]
- 20. Mitchell‐Heggs L., Lellouche N., Deal L., et al., “Transseptal Puncture Using Minimally Invasive Echocardiography During Atrial Fibrillation Ablation,” Europace: European Pacing, Arrhythmias, and Cardiac Electrophysiology: Journal of the Working Groups on Cardiac Pacing, Arrhythmias, and Cardiac Cellular Electrophysiology of the European Society of Cardiology 12 (2010): 1435–1438, 10.1093/europace/euq297. [DOI] [PubMed] [Google Scholar]
- 21. Greco C., Chiavari P. A., Campolongo G., et al., “Transnasal Transesophageal Echocardiography: a New Approach for the PFO Occlusion in Awake Patients,” Catheterization and Cardiovascular Interventions 72 (2008): 538–541, 10.1002/ccd.21665. [DOI] [PubMed] [Google Scholar]
- 22. Schuster P., Chen J., and Hoff P. I., “TEE Time? ICETEE time! IntraCardiac Echocardiography probe used for TransoEsophageal Echocardiography,” journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology 12 (2010): 1787–1788, Europace. [DOI] [PubMed] [Google Scholar]
- 23. Laterra G., Scalia L., Strazzieri O., et al., “Left Atrial Appendage Occlusion in the Era of Minimalist Approaches: Anesthesia and Imaging Considerations,” Journal of Clinical Medicine 15, no. 9 (2026): 3396–3410, 10.3390/jcm15093396. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24. Laterra G., Dattilo G., Correale M., et al., “Imaging Modality to Guide Left Atrial Appendage Closure: Current Status and Future Perspectives,” Journal of Clinical Medicine 12 (2023): 3756–3765, 10.3390/jcm12113756. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25. Spencer K. T., Krauss D., Thurn J., et al., “Transnasal Transesophageal Echocardiography,” Journal of the American Society of Echocardiography 10 (1997): 728–737, 10.1016/S0894-7317(97)70116-0. [DOI] [PubMed] [Google Scholar]
- 26. Spencer K. T., Goldman M., Cholley B., et al., “Multicenter Experience Using a New Prototype Transnasal Transesophageal Echocardiography Probe,” Echocardiography 16 (1999): 811–817, 10.1111/j.1540-8175.1999.tb00133.x. [DOI] [PubMed] [Google Scholar]
- 27. Spencer K. T., Thurn J., Bednarz J., et al., “Transnasal Transesophageal Stress Echocardiography,” Echocardiography 15 (1998): 467–472. [DOI] [PubMed] [Google Scholar]
- 28. Greim C. A., Brederlau J., Kraus I., et al., “Transnasal Transesophageal Echocardiography: a Modified Application Mode for Cardiac Examination in Ventilated Patients,” Anesthesia and Analgesia 88 (1999): 306–311. [DOI] [PubMed] [Google Scholar]
- 29. Fukuda S., Shimada K., Kawasaki T., et al., “Transnasal Transesophageal Echocardiography in the Detection of Left Atrial Thrombus,” Journal of Cardiology 54 (2009): 425–431, 10.1016/j.jjcc.2009.07.002. [DOI] [PubMed] [Google Scholar]
- 30. Greim C. A., Trautner H., Krämer K., et al., “The Detection of Interatrial Flow Patency in Awake and Anesthetized Patients: a Comparative Study Using Transnasal Transesophageal Echocardiography,” Anesthesia and Analgesia 92 (2001): 1111–1116, 10.1097/00000539-200105000-00006. [DOI] [PubMed] [Google Scholar]
- 31. Zimmermann P., Greim C., Trautner H., et al., “Echocardiographic Monitoring During Induction of General Anesthesia With a Miniaturized Esophageal Probe,” Anesthesia and Analgesia 96 (2003): 21–27, 10.1213/00000539-200301000-00006. [DOI] [PubMed] [Google Scholar]
- 32. Klettas D., Alcock E., Dworakowski R., et al., “Is Transnasal TEE Imaging a Viable Alternative to Conventional TEE During Structural Cardiac Interventions to Avoid General Anaesthesia? A Pilot Comparison Study of Image Quality,” Echo research and practice 4 (2017): 1–7, 10.1530/ERP-16-0029. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33. Sun W., Wang B., He L., et al., “Feasibility and Safety of Transnasal Transesophageal Echocardiography Guiding Left Atrial Appendage Closure Without General Anesthesia,” JACC Cardiovasc Imaging 13 (2020): 2043–2044, 10.1016/j.jcmg.2020.03.005. [DOI] [PubMed] [Google Scholar]
- 34. Lanki J., Tuohinen S., Simonen P., et al., “Comparative Evaluation of Microprobe versus Conventional Transoesophageal Echocardiography for PFO Closure Guidance,” Open Heart 11 (2024): e002502, 10.1136/openhrt-002502. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35. Wang B., Zhang L., Sun W., et al., “Transnasal Transesophageal Echocardiography Guidance for Percutaneous Left Atrial Appendage Closure,” Annals of Thoracic Surgery 108 (2019): e161–e164, 10.1016/j.athoracsur.2019.01.039. [DOI] [PubMed] [Google Scholar]
- 36. Couture E., Darmon A., Stein G., et al., “P996Early experience of Transnasal Micro‐transesophageal Echocardiography to Guide Left Atrium Appendage Closure Under Conscious Sedation,” European Heart Journal 40 (2019): 591, 10.1093/eurheartj/ehz747.0589. [DOI] [Google Scholar]
- 37. Laterra G., Sacchetta G., Artale C., et al., “Intracardiac Echocardiographic Probe Used via Transesophageal to Guide Left Atrial Appendage Occlusion: the DIONISO Study,” JACC Cardiovascular interventions 17 (2024): 1855–1857, 10.1016/j.jcin.2024.05.042. [DOI] [PubMed] [Google Scholar]
- 38. Laterra G., Sacchetta G., Barrano G., et al., “Intracardiac Echocardiography via the Transesophageal Route versus Transesophageal Echocardiography for Guiding Left Atrial Appendage Occlusion,” Cardiovascular Revascularization Medicine 86 (2025): 19–25, 19:S1553‐8389(25)00469‐00475, 10.1016/j.carrev.2025.09.006. [DOI] [PubMed] [Google Scholar]
- 39. Vieillard‐Baron A., Slama M., Cholley B., et al., “Echocardiography in the Intensive Care Unit: from Evolution to Revolution?,” Intensive Care Medicine 34 (2008): 243–249, 10.1007/s00134-007-0923-5. [DOI] [PubMed] [Google Scholar]
- 40. Etchecopar‐Chevreuil C., François B., Clavel M., et al., “Cardiac Morphological and Functional Changes During Early Septic Shock: a Transesophageal Echocardiographic Study,” Intensive Care Medicine 34 (2008): 250–256, 10.1007/s00134-007-0929-z. [DOI] [PubMed] [Google Scholar]
- 41. Vignon P., Mentec H., Terré S., et al., “Diagnostic Accuracy and Therapeutic Impact of Transthoracic and Transesophageal Echocardiography in Mechanically Ventilated Patients in the ICU,” Chest 106 (1994): 1829–1834, 10.1378/chest.106.6.1829. [DOI] [PubMed] [Google Scholar]
- 42. Slama M. A., Novara A., Van de Putte P., et al., “Diagnostic and Therapeutic Implications of Transesophageal Echocardiography in Medical ICU Patients With Unexplained Shock, Hypoxemia, or Suspected Endocarditis,” Intensive Care Medicine 22 (1996): 916–922, 10.1007/BF02044116. [DOI] [PubMed] [Google Scholar]
- 43. Prager R., Bowdridge J., Pratte M., et al., “Indications, Clinical Impact, and Complications of Critical Care Transesophageal Echocardiography: a Scoping Review,” Journal of Intensive Care Medicine 38 (2023): 245–272, 10.1177/08850666221115348. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44. Vieillard‐Baron A., Slama M., Mayo P., et al., “A Pilot Study on Safety and Clinical Utility of a Single‐use 72‐Hour Indwelling Transesophageal Echocardiography Probe,” Intensive Care Medicine 39 (2013): 629–635, 10.1007/s00134-012-2797-4. [DOI] [PubMed] [Google Scholar]
- 45. Kottoor S. J. and Arora R. R., “Cryptogenic Stroke: To Close a Patent Foramen Ovale or Not to Close?,” Journal of central nervous system disease 10 (2018): 1179573518819476, 10.1177/1179573518819476. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 46. Mojadidi M. K., Zaman M. O., Elgendy I. Y., et al., “Cryptogenic Stroke and Patent Foramen Ovale,” Journal of the American College of Cardiology 71 (2018): 1035–1043, 10.1016/j.jacc.2017.12.059. [DOI] [PubMed] [Google Scholar]
- 47. Stafford M. B., Bagley J. E., and DiGiacinto D., “Comparison of Transthoracic Echocardiography, Transesophageal Echocardiography, and Transcranial Doppler in the Detection of Patent Foramen Ovale as the Etiology for Cryptogenic Stroke,” Journal of Diagnostic Medical Sonography 35 (2019): 127–133, 10.1177/8756479318816983. [DOI] [Google Scholar]
- 48. Souteyrand G., Motreff P., Lusson J., et al., “Comparison of Transthoracic Echocardiography Using Second Harmonic Imaging, Transcranial Doppler and Transesophageal Echocardiography for the Detection of Patent Foramen Ovale in Stroke Patients,” European Journal of Echocardiography 7 (2006): 147–154, 10.1016/j.euje.2005.04.007. [DOI] [PubMed] [Google Scholar]
- 49. Thota P. N., Zuccaro G. J., Vargo J. J., et al., “A Randomized Prospective Trial Comparing Unsedated Esophagoscopy via Transnasal and Transoral Routes Using a 4‐mm Video Endoscope With Conventional Endoscopy With Sedation,” Endoscopy 37 (2005): 559–565, 10.1055/s-2005-861476. [DOI] [PubMed] [Google Scholar]
- 50. Trevisani L., Cifalà V., Sartori S., et al., “Unsedated Ultrathin Upper Endoscopy Is Better Than Conventional Endoscopy in Routine Outpatient Gastroenterology Practice: A Randomized Trial,” World Journal of Gastroenterology 13 (2007): 906–911, 10.3748/wjg.v13.i6.906. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 51. Stroppa I., Grasso E., Paoluzi O. A., et al., “Unsedated Transnasal versus Transoral Sedated Upper Gastrointestinal Endoscopy: a One‐Series Prospective Study on Safety and Patient Acceptability,” Digestive and Liver Disease 40 (2008): 767–775, 10.1016/j.dld.2008.02.033. [DOI] [PubMed] [Google Scholar]
- 52. Craig A., Hanlon J., Dent J., et al., “A Comparison of Transnasal and Transoral Endoscopy With Small‐diameter Endoscopes in Unsedated Patients,” Gastrointestinal Endoscopy 49 (1999): 292–296, 10.1016/S0016-5107(99)70003-9. [DOI] [PubMed] [Google Scholar]