Quantitative comparison of 68-keV monoenergetic, routine-dose, and low-dose 120-kVp CT for calcium-mimicking inserts in thoracic phantoms: effects of phantom size and reconstruction method

Yeu-Sheng Tyan 1,2,3, Ya-Yun Hsiao 4, Cheng-Chang Lu 2,5, Tou-Rong Chen 1,2,✉
PMCID: PMC13627631  PMID: 42814164

Abstract

Objective

68-keV monoenergetic images from dual-energy CT are often compared with conventional 120-kVp CT, but quantitative performance may depend on phantom size, reconstruction method, and dose settings. We performed a technical phantom-based comparison of 68-keV monoenergetic, routine-dose, and low-dose 120-kVp CT for calcium-mimicking inserts.

Materials and methods

In this phantom study, a thorax phantom with a cardiac calcification insert was scanned using routine-dose 120-kVp CT, low-dose 120-kVp CT, and dual-energy CT with 68-keV monoenergetic reconstruction. Small- and large-phantom conditions were created without and with an extension ring. Images were reconstructed at 2.5-mm slice thickness using filtered back projection (FBP), iterative reconstruction (IR), and deep-learning reconstruction (DL). Mean attenuation, background noise, contrast-to-noise ratio (CNR), and figure of merit (FOM, CNR²/CTDIvol) were summarized descriptively. Repeated scans and repeated region of interest redraws were not performed.

Results

For both 5-mm and 3-mm inserts, attenuation generally increased with hydroxyapatite (HA) density. The with-ring condition showed higher background noise and lower CNR. DL showed the most favorable descriptive quantitative results. Under most conditions, 68-keV imaging showed the highest CNR, followed by routine-dose and low-dose 120-kVp CT; however, it was also associated with the highest CTDIvol. When the dose was evaluated using FOM, the differences among the three strategies were smaller.

Conclusion

In this technical phantom study, 68-keV monoenergetic imaging showed a descriptively higher CNR than routine-dose and low-dose 120-kVp CT, but this was accompanied by a higher radiation dose. Dose-normalized performance was more comparable across strategies. Findings should be interpreted as descriptive condition-level comparisons rather than statistically tested superiority.

Key Points

Question How do 68-keV monoenergetic, routine-dose 120-kVp, and low-dose 120-kVp CT differ in conspicuity and dose-normalized performance for calcium-mimicking targets?

Findings 68-keV imaging generally yielded the highest CNR, but also the highest CTDIvol; when dose was considered using FOM, performance differences became smaller.

Relevance statement This phantom study provides a technical framework for calcium-focused CT protocol optimization by comparing conspicuity, radiation dose, and dose-normalized performance across monoenergetic and conventional acquisitions.

Graphical Abstract

graphic file with name 41747_2026_818_Figa_HTML.webp

Keywords: Calcium-mimicking insert, Deep-learning reconstruction, Dual-energy CT, Low-dose CT, Monoenergetic imaging

Background

Coronary and vascular calcifications are important imaging targets because their visibility can influence detection, characterization, and quantitative assessment. In CT, the conspicuity of calcified structures depends not only on lesion size and attenuation, but also on acquisition parameters, patient habitus, image noise, and reconstruction method. These factors become particularly relevant when small calcified targets are evaluated or when dose reduction is desired, because image quality and radiation exposure must be balanced carefully [1–3].

Dual-energy computed tomography (CT) enables the generation of monoenergetic images with attenuation characteristics different from those of conventional polychromatic 120-kVp CT. Among these, 68-keV monoenergetic images are frequently used in clinical practice and are often compared with standard 120-kVp CT. However, the relationship between 68-keV monoenergetic imaging and conventional 120-kVp CT cannot be fully described by visual similarity alone. Quantitative image performance may differ depending on the metric considered, including attenuation, background noise, and contrast-to-noise ratio (CNR), and may further depend on phantom size and reconstruction method [4–8].

Advances in CT reconstruction, especially iterative reconstruction (IR) and deep learning-based reconstruction, further complicate this comparison. Compared with filtered back projection (FBP), these methods may substantially reduce image noise and alter lesion conspicuity. Consequently, the relative performance of 68-keV monoenergetic, routine-dose 120-kVp, and low-dose 120-kVp CT should be interpreted in the context of both reconstruction algorithm and imaging task [9–14].

For calcium-focused imaging, comparison based on CNR alone may be incomplete because radiation dose also differs across acquisition strategies. A higher CNR does not necessarily indicate greater dose efficiency if the corresponding volume CT dose index (CTDIvol) is also higher. Therefore, dose-normalized assessment using a figure of merit (FOM) may provide a more balanced comparison among 68-keV monoenergetic, routine-dose, and low-dose 120-kVp CT. In addition, phantom size is particularly relevant because increased attenuation and scatter may degrade image quality and alter the trade-off between conspicuity and dose [4–6, 15–18].

Therefore, the purpose of this study was to perform a quantitative comparison of 68-keV monoenergetic, routine-dose 120-kVp, and low-dose 120-kVp CT for calcium-mimicking inserts in thoracic phantoms. Specifically, we evaluated the effects of phantom size (without ring versus with ring) and reconstruction method (FBP, IR, and deep-learning reconstructionn (DL)) on attenuation, background noise, CNR, and dose-normalized performance expressed as FOM. The primary quantitative analysis focused on 5-mm and 3-mm inserts, whereas 1-mm inserts were evaluated exploratorily, given their expected susceptibility to partial volume effects at a reconstructed slice thickness of 2.5 mm. The present work should be interpreted primarily as a technical phantom-based comparison relevant to calcium-focused protocol optimization rather than as direct clinical validation.

Materials and methods

Study design

This phantom-based experimental study compared the quantitative performance of 68-keV monoenergetic, routine-dose 120-kVp, and low-dose 120-kVp CT in thoracic phantom configurations of different effective sizes. Ethics approval was not required because no human participants, human data, or animal data were involved.

Phantom configuration and calcium-mimicking inserts

A QRM thorax phantom (QRM GmbH, Möhrendorf, Germany) was used to simulate thoracic anatomy. The phantom had external dimensions of approximately 200 × 300 mm and a height of 100 mm, with a central bore hole designed to accommodate inserts. A QRM cardiac calcification insert (CCI D100) was positioned in the central bore hole to simulate calcium-containing vascular structures. This configuration was defined as the small-phantom condition (without a ring).

To simulate a larger body habitus, a QRM extension ring (L size) was added around the thorax phantom. This configuration was defined as the large-phantom condition (with ring). The extension ring, composed of fat-equivalent material, increased the effective phantom cross-sectional area.

The CCI insert contained nine calcium-mimicking cylindrical inserts with three hydroxyapatite (HA) densities (800, 400, and 200 mg HA/cm³) and three nominal sizes (5, 3, and 1 mm in diameter, with corresponding lengths of 5, 3, and 1 mm, respectively). These inserts were used to simulate calcified vascular targets of different sizes and attenuation levels.

CT acquisition and reconstruction

CT scans were performed using a GE Revolution Apex scanner (GE HealthCare, Waukesha, WI, USA). Three acquisition strategies were evaluated: routine-dose 120-kVp CT, low-dose 120-kVp CT, and dual-energy CT with 68-keV monoenergetic reconstruction.

Dual-energy scanning was performed using rapid kV switching between 80 and 140 kVp, followed by monoenergetic image reconstruction at 68 keV. The 68-keV monoenergetic level was selected because it represented the predefined monoenergetic reconstruction setting used in the present workflow and is frequently discussed in relation to standard 120-kVp CT in prior literature. Detailed acquisition parameters, noise index (NI) settings, and CTDIvol values for each protocol and phantom condition are summarized in Table 1.

Table 1.

Scan parameters, noise index settings, and CTDIvol for routine-dose, low-dose, and dual-energy CT acquisitions in the small and large phantom configurations

Protocola Phantomb Tube voltage/mono-energy Smart mA Noise index Focal spotc CTDIvol (mGy)d
Low-dose Small 120 kVp 10–90 45.7 S 0.40
Low-dose Large 120 kVp 70–500 55.0 S 1.51
Routine-dose Small 120 kVp 20–650 43.1 S 2.39
Routine-dose Large 120 kVp 35–900 49.8 S 4.05
Dual-energy Small 68 keVe 145 38.2 L 4.57
Dual-energy Large 68 keV 405 38.2 L 9.93

a All protocols had a slice thickness of 2.5 mm, a pitch of 0.992:1, and a rotation time of 0.5 s

b The phantom without a ring represents the small configuration, whereas the phantom with a ring represents the large configuration

c S indicates a small focal spot, and L indicates a large focal spot

d FBP, IR, and DL were reconstructed from the same raw data acquisition; therefore, CTDIvol was identical across reconstruction methods within each protocol and phantom condition

e Dual-energy images were acquired using rapid kV switching between 80 and 140 kVp and subsequently reconstructed as 68-keV monoenergetic images

All images were reconstructed at a slice thickness of 2.5 mm using a standard kernel and three reconstruction methods: FBP, ASiR-V 50% IR, and TrueFidelity DLIR, medium strength (DL). All protocol groups were reconstructed at 2.5-mm slice thickness to maintain internal consistency across acquisition strategies and reconstruction methods within the present comparative framework. Because FBP, IR, and DL images were generated from the same raw data acquisition, CTDIvol was identical across reconstruction methods within each protocol and phantom condition [4, 5, 9, 13].

Image selection

For each acquisition protocol, phantom condition, and reconstruction method, two slices were selected from each image series. These represented the slices showing the highest conspicuity of the calcium-mimicking inserts and were referenced by their original DICOM filenames to ensure reproducibility and avoid ambiguity related to slice ordering. Slice selection and region of interest (ROI) center determination were performed by the same investigator and then applied consistently across the corresponding reconstructions from the same acquisition.

ROI analysis

ROI placement was performed using ImageJ (National Institutes of Health, Bethesda, MD, USA). For each selected slice, nine circular ROIs were placed at the centers of the calcium-mimicking inserts, corresponding to the three HA densities and three insert sizes. In addition, one large background ROI was placed in a visually homogeneous region of the insert background, with exclusion circles applied to avoid contamination by the calcification inserts and their immediate surroundings.

ROI definitions were held constant within each scan condition and applied consistently across both selected slices. A custom Python-based workflow was then used to read the DICOM images directly, convert pixel values to Hounsfield units (HU) using the DICOM rescale slope and intercept, and extract ROI statistics [10, 11].

Quantitative measurements

For each insert ROI and background ROI, the following parameters were extracted: ROI area in mm² (area_mm²), mean attenuation in Hounsfield units (mean HU), and standard deviation (SD). Background noise was defined as the standard deviation of the background ROI. CNR was defined as the absolute difference between the mean attenuation of the insert ROI and that of the background ROI, divided by the standard deviation of the background ROI. To account for radiation dose, a FOM was calculated as:

FOM=CNR2/CTDIvol.

Thus, CNR was used to characterize lesion conspicuity, whereas FOM was used to evaluate dose-normalized quantitative performance.

Primary and exploratory analysis targets

The primary quantitative analysis focused on the 5-mm and 3-mm calcium-mimicking inserts because these targets were more consistently visualized and quantified on images reconstructed at 2.5-mm slice thickness. The 1-mm inserts were analyzed only as an exploratory target, because their extremely small size relative to the reconstructed slice thickness made them highly susceptible to partial volume effects and unstable attenuation measurements.

Comparative framework

Quantitative comparisons were performed across the three acquisition strategies: 68-keV monoenergetic CT, routine-dose 120-kVp CT, and low-dose 120-kVp CT. For each insert size and HA density, comparisons were stratified by phantom condition (without ring versus with ring) and reconstruction method (FBP, IR, DL).

The primary interpretation focused on the relationship among attenuation, background noise, CNR, CTDIvol, and FOM. In particular, the analysis was designed to determine whether higher lesion conspicuity was accompanied by higher radiation dose, and whether the observed differences among acquisition strategies should be interpreted in the context of protocol-dependent noise targets and dose settings.

Statistical analysis

Quantitative measurements were summarized descriptively for each acquisition protocol, phantom condition, reconstruction method, insert size, and HA density. For the main manuscript tables, values from the two selected slices were averaged for each condition. Repeated scans and repeated ROI redraws were not performed in the present study. Because this was a phantom-based technical study and the primary objective was to compare attenuation, noise, CNR, and FOM across imaging conditions, the primary analysis was descriptive [10, 19]. Accordingly, statements describing relative performance among acquisition strategies should be interpreted as descriptive comparisons rather than as formally tested statistical differences. The 5-mm and 3-mm inserts were treated as the primary analysis targets, whereas the 1-mm inserts were treated as exploratory due to their limited measurement stability.

Use of a large language model for assistance

Large language model assistance was used only for language refinement and manuscript organization. It was not used for study design, data generation, data analysis, statistical analysis, or scientific interpretation. All scientific content, analysis, interpretation, and final manuscript decisions remained the responsibility of the authors.

Results

Overall quantitative trends

Quantitative measurements were successfully obtained for all three acquisition protocols, both phantom configurations, and all three reconstruction methods. For the 5-mm and 3-mm calcium-mimicking inserts, measured attenuation generally increased with increasing hydroxyapatite (HA) density, supporting the consistency of ROI placement and analysis. In contrast, the 1-mm inserts showed substantially less stable behavior and were therefore treated as exploratory rather than primary quantitative targets.

Across nearly all conditions, the with-ring configuration showed higher background noise and lower CNR than the corresponding without-ring configuration. Reconstruction-dependent differences were also consistently observed, with DL generally providing the lowest background noise and the highest CNR, followed by IR, whereas FBP showed the highest noise and the lowest CNR [10].

Quantitative results for 5-mm inserts

The quantitative results for the 5-mm inserts are summarized in Table 2, and the corresponding CNR trends are illustrated in Fig. 1. For all three acquisition strategies, attenuation and CNR were highest for the 800 mg HA/cm³ inserts, intermediate for the 400 mg HA/cm³ inserts, and lowest for the 200 mg HA/cm³ inserts, indicating a density-dependent response across reconstruction methods.

Table 2.

Quantitative results for 5-mm calcium-mimicking inserts. Values represent the average of the two selected slices for each condition

Protocol Phantoms Recona Density mg HA/cm³ Mean HU Background noiseb CNR FOMc
Low-dose Small DL 800 682 20 32 2,812
Low-dose Small DL 400 337 20 15 655
Low-dose Small DL 200 200 20 8 169
Low-dose Small FBP 800 679 37 17 812
Low-dose Small FBP 400 327 37 8 180
Low-dose Small FBP 200 199 37 4 48
Low-dose Small IR 800 675 24 27 1,927
Low-dose Small IR 400 321 24 12 420
Low-dose Small IR 200 190 24 6 102
Low-dose Large DL 800 657 12 50 1,057
Low-dose Large DL 400 349 12 25 261
Low-dose Large DL 200 178 12 11 51
Low-dose Large FBP 800 668 26 24 248
Low-dose Large FBP 400 347 26 12 59
Low-dose Large FBP 200 184 26 5 13
Low-dose Large IR 800 666 17 37 558
Low-dose Large IR 400 346 17 18 132
Low-dose Large IR 200 175 17 8 26
Routine-dose Small DL 800 683 10 64 2,931
Routine-dose Small DL 400 395 10 35 830
Routine-dose Small DL 200 204 10 16 185
Routine-dose Small FBP 800 687 19 34 830
Routine-dose Small FBP 400 390 19 18 225
Routine-dose Small FBP 200 195 19 8 47
Routine-dose Small IR 800 685 12 53 2,013
Routine-dose Small IR 400 387 12 28 541
Routine-dose Small IR 200 191 12 12 107
Routine-dose Large DL 800 789 14 53 610
Routine-dose Large DL 400 422 14 27 158
Routine-dose Large DL 200 201 14 11 27
Routine-dose Large FBP 800 786 29 26 144
Routine-dose Large FBP 400 421 29 13 37
Routine-dose Large FBP 200 201 29 5 6
Routine-dose Large IR 800 786 18 41 367
Routine-dose Large IR 400 421 18 21 95
Routine-dose Large IR 200 196 18 8 15
Dual-energy Small DL 800 893 9 96 2,269
Dual-energy Small DL 400 451 9 46 547
Dual-energy Small DL 200 231 9 22 116
Dual-energy Small FBP 800 889 14 61 932
Dual-energy Small FBP 400 443 14 29 218
Dual-energy Small FBP 200 230 14 14 48
Dual-energy Small IR 800 890 10 83 1,721
Dual-energy Small IR 400 442 10 40 399
Dual-energy Small IR 200 228 10 19 86
Dual-energy Large DL 800 653 10 61 434
Dual-energy Large DL 400 453 10 41 171
Dual-energy Large DL 200 196 10 16 26
Dual-energy Large FBP 800 644 20 30 103
Dual-energy Large FBP 400 442 20 20 39
Dual-energy Large FBP 200 200 20 8 7
Dual-energy Large IR 800 639 13 47 257
Dual-energy Large IR 400 438 13 31 97
Dual-energy Large IR 200 194 13 12 15

a FBP, IR, and DL denote filtered back projection, ASiR-V 50% iterative reconstruction, and TrueFidelity DLIR (medium strength), respectively

b Background noise was defined as the standard deviation of the background ROI

c FOM was calculated as CNR²/CTDIvol

Fig. 1.

Fig. 1

Quantitative comparison of contrast-to-noise ratio (CNR) for 5-mm calcium-mimicking inserts. CNR values are shown for three hydroxyapatite densities (800, 400, and 200 mg HA/cm³) under routine-dose 120-kVp CT, low-dose 120-kVp CT, and dual-energy CT with 68-keV monoenergetic reconstruction. a–c correspond to the without-ring condition reconstructed with FBP, IR, and DL, respectively, whereas (d–f) correspond to the with-ring condition. For each acquisition condition, the plotted values represent the average of the two selected slices

In the without-ring condition, 68-keV imaging generally yielded the highest CNR, followed by routine-dose and low-dose 120-kVp CT. Among the three reconstruction methods, DL consistently produced the highest CNR and FOM, whereas FBP yielded the lowest FOM because of its higher background noise. For example, in the without-ring condition at 800 mg HA/cm³, CNR increased from approximately 34.0 with routine-dose FBP to 61.3 with 68-keV FBP, and from 63.9 with routine-dose DL to 95.7 with 68-keV DL. As shown in Fig. 1, this ranking was preserved across the three acquisition strategies.

In the with-ring condition, quantitative performance decreased for all protocols and reconstructions. Nevertheless, the same general pattern remained, with DL > IR > FBP, and with 68-keV imaging generally producing the highest CNR. However, this CNR advantage must be interpreted together with radiation dose, because CTDIvol was highest for the 68-keV acquisition, intermediate for routine-dose 120-kVp CT, and lowest for low-dose 120-kVp CT. Accordingly, although 68-keV imaging yielded the highest CNR, the corresponding differences in FOM were smaller than the CNR differences alone would suggest [15, 18].

Quantitative results for 3-mm inserts

The quantitative results for the 3-mm inserts are summarized in Table 3, and the corresponding CNR trends are illustrated in Fig. 2. As expected, measured attenuation and CNR were lower than those observed for the 5-mm inserts under otherwise matched conditions, reflecting the smaller target size and greater susceptibility to partial-volume effects.

Table 3.

Quantitative results for 3-mm calcium-mimicking inserts. Values represent the average of the two selected slices for each condition

Protocol Phantoms Recona Density mg HA/cm³ Mean HU Background noiseb CNR FOMc
Low-dose Small DL 800 398 20 18 1,375
Low-dose Small DL 400 214 20 9 361
Low-dose Small DL 200 119 20 4 63
Low-dose Small FBP 800 396 37 10 397
Low-dose Small FBP 400 211 37 5 108
Low-dose Small FBP 200 122 37 2 21
Low-dose Small IR 800 388 24 15 939
Low-dose Small IR 400 204 24 7 236
Low-dose Small IR 200 114 24 3 40
Low-dose Large DL 800 262 12 18 134
Low-dose Large DL 400 162 12 10 39
Low-dose Large DL 200 78 12 3 3
Low-dose Large FBP 800 273 26 9 34
Low-dose Large FBP 400 172 26 5 11
Low-dose Large FBP 200 109 26 3 3
Low-dose Large IR 800 267 17 13 73
Low-dose Large IR 400 163 17 7 21
Low-dose Large IR 200 100 17 3 5
Routine-dose Small DL 800 410 10 37 1,081
Routine-dose Small DL 400 244 10 20 280
Routine-dose Small DL 200 135 10 9 67
Routine-dose Small FBP 800 416 19 20 311
Routine-dose Small FBP 400 248 19 11 82
Routine-dose Small FBP 200 135 19 5 18
Routine-dose Small IR 800 411 12 30 744
Routine-dose Small IR 400 239 12 16 182
Routine-dose Small IR 200 128 12 7 39
Routine-dose Large DL 800 423 14 27 172
Routine-dose Large DL 400 210 14 12 34
Routine-dose Large DL 200 104 14 4 4
Routine-dose Large FBP 800 423 29 13 41
Routine-dose Large FBP 400 213 29 6 9
Routine-dose Large FBP 200 134 29 3 2
Routine-dose Large IR 800 405 18 20 98
Routine-dose Large IR 400 202 18 9 20
Routine-dose Large IR 200 121 18 4 4
Dual-energy Small DL 800 503 9 52 672
Dual-energy Small DL 400 275 9 27 185
Dual-energy Small DL 200 149 9 13 39
Dual-energy Small FBP 800 506 14 34 282
Dual-energy Small FBP 400 271 14 17 74
Dual-energy Small FBP 200 148 14 8 16
Dual-energy Small IR 800 504 10 46 514
Dual-energy Small IR 400 268 10 23 132
Dual-energy Small IR 200 145 10 10 27
Dual-energy Large DL 800 366 10 33 197
Dual-energy Large DL 400 227 10 19 45
Dual-energy Large DL 200 118 10 8 11
Dual-energy Large FBP 800 370 20 16 48
Dual-energy Large FBP 400 229 20 9 11
Dual-energy Large FBP 200 124 20 4 3
Dual-energy Large IR 800 360 13 25 117
Dual-energy Large IR 400 219 13 14 26
Dual-energy Large IR 200 112 13 6 6

a FBP, IR, and DL denote filtered back projection, ASiR-V 50% iterative reconstruction, and TrueFidelity DLIR (medium strength), respectively

b Background noise was defined as the standard deviation of the background ROI

c FOM was calculated as CNR²/CTDIvol

Fig. 2.

Fig. 2

Quantitative comparison of contrast-to-noise ratio (CNR) for 3-mm calcium-mimicking inserts. CNR values are shown for three hydroxyapatite densities (800, 400, and 200 mg HA/cm³) under routine-dose 120-kVp CT, low-dose 120-kVp CT, and dual-energy CT with 68-keV monoenergetic reconstruction. a–c correspond to the without-ring condition reconstructed with FBP, IR, and DL, respectively, whereas (d–f) correspond to the with-ring condition. For each acquisition condition, the plotted values represent the average of the two selected slices

Despite this reduction, the 3-mm inserts still showed consistent density-dependent trends across most conditions. In both phantom configurations, DL again produced the most favorable quantitative performance, with higher CNR and FOM than IR and FBP. For example, in the without-ring condition at 800 mg HA/cm³, CNR increased from approximately 19.7 with routine-dose FBP to 33.8 with 68-keV FBP, and from 36.7 with routine-dose DL to 52.1 with 68-keV DL. In the with-ring condition, the corresponding CNR values were lower across all three acquisition strategies, illustrating the adverse effect of increased phantom size.

Similar to the 5-mm results, 68-keV imaging generally yielded the highest CNR for 3-mm inserts, whereas low-dose 120-kVp CT yielded the lowest. However, once CTDIvol was taken into account, the differences in FOM among 68-keV, routine-dose, and low-dose acquisitions became less pronounced. Thus, for the 3-mm inserts, the relative advantage of 68-keV imaging was more evident for CNR than for dose-normalized performance [15, 18].

Comparison of CNR, dose, and FOM among acquisition strategies

When the three acquisition strategies were compared directly, several consistent patterns emerged. First, the without-ring configuration provided lower background noise and higher CNR than the with-ring configuration across all protocols, demonstrating the expected adverse effect of increased phantom size. Second, DL reconstruction consistently improved CNR and FOM relative to both IR and FBP, regardless of acquisition protocol. Third, the ranking of CNR was generally: 68-keV > routine-dose 120-kVp > low-dose 120-kVp CT.

However, CTDIvol followed the same general order: 68-keV > routine-dose 120-kVp > low-dose 120-kVp CT.

As a result, the observed CNR advantage of 68-keV imaging was attenuated when performance was normalized to dose using FOM. These trends are visually summarized in Figs. 1 and 2 and numerically summarized in Tables 2 and 3. Taken together, the data indicate that 68-keV monoenergetic imaging generally provides the highest lesion conspicuity, but that the overall dose-normalized performance of the three acquisition strategies is more similar than CNR alone would imply.

Exploratory results for 1-mm inserts

The 1-mm inserts showed considerably less stable quantitative behavior than the 5-mm and 3-mm inserts. In several conditions, measured attenuation values did not follow the expected monotonic ranking with increasing HA density, particularly in the larger phantom configuration. This instability is likely attributable to severe through-plane partial-volume averaging at the 2.5-mm slice thickness, together with the extremely small physical size of the target [19].

Because of this limited stability, the 1-mm results were not considered sufficiently robust for primary quantitative comparisons. Instead, they were treated as exploratory observations supporting the interpretation that 2.5-mm reconstructed images are not optimal for reliable quantitative assessment of calcification-mimicking targets of this size.

Summary of main findings

Overall, the quantitative analysis showed that 68-keV monoenergetic imaging generally provided the highest CNR for 5-mm and 3-mm inserts, whereas low-dose 120-kVp CT generally provided the lowest. Quantitative performance was consistently influenced by phantom size and reconstruction method, with the with-ring condition reducing CNR and FOM and DL reconstruction providing the most favorable results. However, because CTDIvol was also highest for the 68-keV acquisition, FOM values among 68-keV, routine-dose 120-kVp, and low-dose 120-kVp CT were more comparable than the CNR results alone would suggest.

Discussion

The present study compared dual-energy CT-derived 68-keV monoenergetic images with routine-dose and low-dose 120-kVp CT for calcium-mimicking inserts in thoracic phantom configurations of different effective sizes. The main findings can be summarized as follows. First, for the 5-mm and 3-mm inserts, 68-keV imaging generally produced the highest CNR, followed by routine-dose and low-dose 120-kVp CT. Second, the observed CNR advantage of 68-keV imaging was accompanied by higher radiation dose, such that FOM values across the three acquisition strategies were more similar than the CNR values alone would suggest. Third, phantom size strongly affected quantitative performance, with the with-ring condition consistently increasing background noise and reducing CNR and FOM. Fourth, the reconstruction method had a marked effect on quantitative performance, with DL generally providing the most favorable results. Finally, the 1-mm inserts showed limited quantitative stability at 2.5-mm slice thickness and were not suitable as primary analysis targets [4–6].

One motivation for this study was the common assumption that 68-keV monoenergetic images approximate conventional 120-kVp CT. The present results suggest that this formulation is incomplete. In our data, 68-keV imaging was not merely similar to routine-dose 120-kVp CT; rather, it generally produced higher CNR for both 5-mm and 3-mm inserts. Accordingly, the relationship between 68-keV and conventional 120-kVp CT is better understood as a trade-off between lesion conspicuity and radiation dose rather than as simple visual equivalence. The present work should therefore be interpreted as a technical phantom-based comparison relevant to calcium-focused protocol optimization rather than as direct clinical validation. This interpretation is consistent with the concept that monoenergetic image behavior is influenced not only by effective energy, but also by acquisition noise and reconstruction characteristics [4, 6, 15].

The comparison with low-dose 120-kVp CT further supports this interpretation. Low-dose CT consistently produced lower CTDIvol, but also lower CNR. Routine-dose 120-kVp CT occupied an intermediate position between low-dose 120-kVp CT and 68-keV monoenergetic imaging for both dose and CNR. This pattern indicates that the three acquisition strategies form a dose-performance continuum rather than a binary comparison. In this framework, 68-keV imaging offers improved conspicuity, but at a higher dose cost, whereas low-dose 120-kVp CT reduces dose at the expense of lower lesion conspicuity [15, 17, 18].

The observed differences in CNR among 68-keV monoenergetic, routine-dose, and low-dose 120-kVp CT should also be interpreted in the context of protocol design. In the present study, the dual-energy protocol had the lowest noise index (NI) and the highest CTDIvol, whereas the low-dose protocol had the highest NI and the lowest CTDIvol. Accordingly, the descriptively higher CNR observed with 68-keV imaging cannot be attributed solely to energy-dependent image characteristics, but should also be understood in relation to protocol-dependent noise targets and dose settings. This suggests that the relative ranking of noise and CNR across acquisition strategies may be protocol-dependent rather than fixed.

A key implication of the present data is that CNR alone does not fully capture the relative performance of the three acquisition strategies. Because CTDIvol was highest for the 68-keV acquisition, the observed CNR advantage of 68-keV imaging in Figs. 1 and 2 should be interpreted alongside the FOM results in Tables 2 and 3. Once the dose was taken into account, the differences among 68-keV, routine-dose, and low-dose CT became smaller. Thus, the present findings do not indicate an overall advantage of 68-keV imaging across all performance metrics; rather, they suggest that 68-keV imaging provides higher CNR, whereas the dose-normalized efficiency of the three strategies is more comparable.

The reconstruction-dependent findings were also consistent across protocols. DL generally yielded lower background noise and higher CNR than IR and FBP, and it also produced the most favorable FOM values under most conditions. This observation is consistent with previous reports indicating that advanced reconstruction methods can improve image quality by reducing noise while preserving relevant signal characteristics [9, 11, 13]. In practical terms, this means that conclusions regarding the relative performance of 68-keV and 120-kVp CT should always be interpreted in the context of the reconstruction method.

Phantom size had a clear and systematic effect. The with-ring configuration was associated with degraded quantitative performance, as reflected by increased background noise and reduced CNR and FOM. This finding is physically plausible, given the increased attenuation and scatter introduced by the extension ring. As illustrated in Figs. 1 and 2, the penalty associated with increased phantom size was evident across all three reconstruction methods, although DL preserved CNR more effectively than IR and FBP. These results indicate that the balance between conspicuity and dose becomes less favorable as effective phantom size increases [4, 5, 10].

The 1-mm inserts require separate consideration. Their behavior was substantially less stable than that of the 5-mm and 3-mm inserts, and attenuation ranking across HA densities was not consistently preserved. This instability is most likely due to severe partial volume effects, as the nominal 1-mm insert length is markedly smaller than the 2.5-mm reconstructed slice thickness. These findings indicate that reliable quantitative assessment of targets of this size would likely require thinner reconstructions and/or a dedicated high-resolution protocol [10, 19, 20].

This study has several limitations. First, it was a phantom-based investigation conducted on a single CT platform using a single set of reconstruction parameters; therefore, the results may be vendor- and system-specific. Second, the primary analysis was descriptive and focused on quantitative image metrics rather than observer-based detection performance, and neither repeated scans nor repeated ROI redraws were performed. Third, only one monoenergetic level, 68 keV, was evaluated in detail, although other keV levels may also be relevant for optimization. Fourth, the acquisition strategies were not dose-matched, making it difficult to isolate the effect of monoenergetic imaging itself from the influence of radiation dose and noise target settings. Fifth, the exclusive use of 2.5-mm reconstructions likely increased partial-volume effects for the smaller inserts and limits direct extrapolation to applications in which thinner reconstructions are routinely used. Finally, the 1-mm inserts were not suitable for robust primary analysis at the present reconstruction thickness.

Despite these limitations, the study provides a structured comparison among routine-dose 120-kVp CT, low-dose 120-kVp CT, and dual-energy CT with 68-keV monoenergetic reconstruction under controlled phantom conditions. The inclusion of two phantom sizes and three reconstruction methods, together with dose-normalized assessment using FOM, strengthens the technical relevance of the findings and provides a useful framework for understanding how monoenergetic imaging behaves relative to conventional acquisitions in calcium-focused CT tasks. Rather than asking only whether 68-keV imaging is similar to routine-dose 120-kVp CT, the present data suggest that the more relevant question is how lesion conspicuity, radiation dose, and dose-normalized performance should be balanced across acquisition strategies. Viewed in this way, 68-keV monoenergetic imaging can provide higher CNR, whereas routine-dose and low-dose 120-kVp CT remain competitive when performance is evaluated in terms of dose.

For 5-mm and 3-mm calcium-mimicking inserts, 68-keV monoenergetic imaging generally provided the highest CNR, followed by routine-dose and low-dose 120-kVp CT. However, this advantage was accompanied by higher CTDIvol. When dose was accounted for using FOM, the quantitative performance of the three acquisition strategies became more comparable. Phantom size and reconstruction method substantially influenced this trade-off, with the with-ring condition reducing performance and DL yielding the most favorable overall results. In contrast, 1-mm insert measurements were unstable at 2.5-mm slice thickness and should be interpreted as exploratory. Because this was a phantom-based technical study, clinical validation remains necessary before these findings can be translated into routine calcium-focused CT tasks or patient-specific protocol selection. Further studies across additional monoenergetic energy levels, CT vendors, reconstruction platforms, and patient-based datasets are warranted to confirm the generalizability and clinical relevance of these descriptive findings.

Acknowledgements

The authors used a large language model only for language refinement and manuscript organization. All study design, data analysis, interpretation, and final manuscript content remained entirely the responsibility of the authors.

Abbreviations

CNR

Contrast-to-noise ratio

CT

Computed tomography

CTDIvol

Volume CT dose index

DL

Deep-learning reconstruction

FBP

Filtered back projection

FOM

Figure of merit

HA

Hydroxyapatite

IR

Iterative reconstruction

NI

Noise index

ROI

Region of interest

Author contributions

YST and TRC conceived and designed the study. YST, YYH, and CCL performed data collection and analysis. YST and TRC drafted the manuscript. YYH and CCL critically revised the manuscript. All authors read and approved the final manuscript.

Funding

This research was supported by CSMU’s grant CSH-2019-C-007.

Data availability

The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request.

Declarations

Ethics approval and consent to participate

Not applicable. This was a phantom-based experimental study and did not involve human participants, human data, or animals.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Footnotes

Publisher’s Note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.Agatston AS, Janowitz WR, Hildner FJ, Zusmer NR, Viamonte M Jr, Detrano R (1990) Quantification of coronary artery calcium using ultrafast computed tomography. J Am Coll Cardiol 15:827–832. 10.1016/0735-1097(90)90282-T [DOI] [PubMed] [Google Scholar]
  • 2.Hecht HS, Blaha MJ, Kazerooni EA et al (2018) CAC-DRS: Coronary Artery Calcium Data and Reporting System. An expert consensus document of the Society of Cardiovascular Computed Tomography (SCCT). J Cardiovasc Comput Tomogr 12:185–191. 10.1016/j.jcct.2018.03.008 [DOI] [PubMed] [Google Scholar]
  • 3.Dzaye O, Dudum R, Reiter-Brennan C et al (2019) Coronary artery calcium scoring for individualized cardiovascular risk estimation in important patient subpopulations after the 2019 AHA/ACC primary prevention guidelines. Prog Cardiovasc Dis 62:423–430. 10.1016/j.pcad.2019.10.007 [DOI] [PubMed] [Google Scholar]
  • 4.McCollough CH, Leng S, Yu L, Fletcher JG (2015) Dual- and multi-energy CT: principles, technical approaches, and clinical applications. Radiology 276:637–653. 10.1148/radiol.2015142631 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.McCollough CH, Boedeker K, Cody D et al (2020) Principles and applications of multienergy CT: report of AAPM task group 291. Med Phys 47:e881–e912. 10.1002/mp.14157 [DOI] [PubMed] [Google Scholar]
  • 6.Albrecht MH, Vogl TJ, Martin SS et al (2019) Review of clinical applications for virtual monoenergetic dual-energy CT. Radiographics 39:1033–1049. 10.1148/radiol.2019182297 [DOI] [PubMed] [Google Scholar]
  • 7.Rajiah P, Parakh A, Kay F, Baruah D, Kambadakone A (2020) Update on multienergy CT: physics, principles, and applications. Radiographics 40:1284–1308. 10.1148/rg.2020200038 [DOI] [PubMed] [Google Scholar]
  • 8.Hamid S, Nasir MU, So A, Andrews G, Nicolaou S, Qamar SR (2021) Clinical applications of dual-energy CT. Korean J Radiol 22:970–982. 10.3348/kjr.2020.0996 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Willemink MJ, de Jong PA, Leiner T et al (2013) Iterative reconstruction techniques for computed tomography part 1: technical principles. Eur Radiol 23:1623–1631. 10.1007/s00330-012-2765-y [DOI] [PubMed] [Google Scholar]
  • 10.Ehman EC, Yu L, Manduca A et al (2014) Methods for clinical evaluation of noise reduction techniques in abdominopelvic CT. Radiographics 34:849–862. 10.1148/rg.344135128 [DOI] [PubMed] [Google Scholar]
  • 11.Greffier J, Frandon J, Larbi A, Beregi JP, Pereira F (2020) CT iterative reconstruction algorithms: a task-based image quality assessment. Eur Radiol 30:487–500. 10.1007/s00330-019-06359-6 [DOI] [PubMed] [Google Scholar]
  • 12.Tran DN, Healy C, Schwarze U, Fleischmann D (2009) Dual-energy CT discrimination of iodine and calcium: experimental results and implications for lower extremity CT angiography. Acad Radiol 16:160–171. 10.1016/j.acra.2008.09.004 [DOI] [PubMed] [Google Scholar]
  • 13.Koetzier LR, Mastrodicasa D, Szczykutowicz TP et al (2023) Deep learning image reconstruction for CT. Radiology 306: e221257. 10.1148/radiol.221257 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Brady SL, Trout AT, Sieren JP et al (2023) Implementation of AI image reconstruction in CT—how is it different from iterative reconstruction and what dose reductions can be achieved? Br J Radiol 96: 20220915. 10.1259/bjr.20220915 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Albrecht MH, Vogl TJ, Martin SS et al (2016) Comprehensive comparison of virtual monoenergetic and linearly blended images in third-generation dual-source dual-energy computed tomography angiography of the thorax and abdomen. Invest Radiol 51:582–590. 10.1097/RLI.0000000000000272 [DOI] [PubMed] [Google Scholar]
  • 16.Kawashima H, Ichikawa K, Hanaoka S, Matsubara K, Takata T (2018) Relationship between size-specific dose estimates and image quality in computed tomography depending on patient size. J Appl Clin Med Phys 19:246–251. 10.1002/acm2.12340 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Stiller W (2018) Basics of iterative reconstruction methods in computed tomography: a vendor-independent overview. Eur J Radiol 109:147–154. 10.1016/j.ejrad.2018.10.025 [DOI] [PubMed] [Google Scholar]
  • 18.Cester D, Eberhard M, Alkadhi H, Euler A (2022) Virtual monoenergetic images from dual-energy CT: systematic assessment of task-based image quality performance. Quant Imaging Med Surg 12:1378–1394. 10.21037/qims-21-477 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Samei E, Richard S (2015) Assessment of the dose reduction potential of a model-based iterative reconstruction algorithm using a task-based performance metrology. Med Phys 42:314–323. 10.1118/1.4903899 [DOI] [PubMed] [Google Scholar]
  • 20.Leipsic J, LaBounty TM, Heilbron B et al (2010) Adaptive statistical iterative reconstruction: assessment of image noise and image quality in coronary CT angiography. AJR Am J Roentgenol 195:649–654. 10.2214/AJR.10.4285 [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request.