. 2026 Oct 1;21(10):e0358746. doi: 10.1371/journal.pone.0358746

Assessment of epithelial barrier integrity in cellular monolayer models using low cost, clinically approved fluorescent contrast agents

Elena Monfort Sanchez 1,#, Aaron F Watson 1,#, Tafhima Haider 2, Despoina Chrysostomou 3, Maelle Storder 1, Nilanjan Mandal 1, Paul Kelly 2,3, Julian R Marchesi 4, James Avery 1,5, Claire D Bourke 2,6, Alex J Thompson 1,*
Editor: Mária A Deli7
PMCID: PMC13630259  PMID: 42821541

Abstract

Epithelial barriers serve as the interface between body tissue and the external environment, maintaining tissue homeostasis by regulating the transport of various cells, molecules and microbes between these compartments. In particular, the epithelial barrier of the gut is a key component in numerous physiological processes that collectively define human health, such as nutrient absorption, microbiota regulation and neuroimmune communication. As such, gut epithelial barrier dysfunction, also known as ‘leaky gut’ syndrome has been associated with a variety of different conditions such as inflammatory bowel disease, type II diabetes, HIV and undernutrition. A diverse range of in vitro models have therefore been developed to better understand epithelial barrier function in different disease contexts. To study gut permeability in vitro, typical measurements include transepithelial electrical resistance (TEER) and the use of fluorescent contrast agents such as FITC-dextran to assess barrier integrity and permeability, with microfluidic ‘gut-on-a-chip’ models exclusively using FITC-dextran. However, utilising FITC-dextran in this context has multiple drawbacks for translational gastroenterology research. Namely, FITC-dextran is expensive, not clinically approved, and there is no agreed molecular weight ‘cut-off’ to determine barrier damage. The molecular weight of FITC-dextran also does not mimic substances used for barrier assessment in clinical settings (e.g., lactulose) nor the molecules thought to drive enteropathy in vivo (e.g., LPS and other pathogen associated molecular patterns (PAMPs)). Therefore, the clinically approved fluorescent contrast agents fluorescein and methylene blue (which have molecular weights comparable to substances used in the clinic) were explored in cellular monolayer models of the gut epithelium as low-cost alternatives for fluorescent assessment of epithelial barrier integrity. Both fluorescein and methylene blue exhibited analogous behavior to FITC-dextran, indicating suitability for use in fluorescent epithelial barrier assays. Hence, wider adoption of fluorescein and methylene may provide cost savings for in vitro epithelial barrier assessment assays as well as opportunities for more advanced, coherent and translational assessment of epithelial barrier function in health and disease.

1. Introduction

The gut epithelial barrier separates the internal contents of the intestine from the rest of the body and is involved in several crucial physiological processes that underpin human health, such as nutrient absorption, microbiota regulation, neuroimmune communication, as well as active and passive immunity [1–3]. Disruption of the epithelial barrier occurs in multiple human diseases and can range from sub-clinical increases in epithelial permeability due to reductions in tight junction proteins between epithelial cells to full breaches in the barrier due to physical damage or severe immunopathology. Intestinal epithelial barrier dysfunction is associated with leakage of blood and tissue into the gut lumen and increased transport of microbial and dietary antigens from the lumen into the lamina propria, which can enter circulation to drive systemic inflammation [4]. As such, a broad range of chronic inflammatory diseases including, liver inflammation, celiac disease, type II diabetes, HIV and neurodegenerative disorders such as Parkinson’s and Alzheimer’s disease, are associated with a so-called ‘leaky gut’ phenotype [3–6]. Thus, various models in vitro have been developed and deployed to further understand the underlying mechanisms that contribute to gut barrier impairment in a variety of disease contexts [7].

Cell culture monolayers utilising the human colorectal adenocarcinoma cell line Caco-2, using a cell culture insert-based system, have been extensively reported in the literature as a 2D in vitro model to study colonic barrier dysfunction. As a result, Caco-2 monolayers grown on cell culture inserts is widely considered the ‘gold standard’ for transport and barrier studies [8]. The utility of this model is largely made possible by the simplicity of introducing a transepithelial electrical resistance (TEER) electrode to measure tight junction integrity to validate stable monolayer formation [8]. Alternatively, Madin-Darby canine kidney (MDCK) cells have also been used to generate epithelial monolayers [9]. Although MDCK cells are derived from the canine kidney they offer a robust, reproducible and physiologically relevant model that is often used as a complementary model alongside human intestinal epithelial cell lines such as Caco-2 and T84. MDCK cells can serve this purpose as they form well developed tight junctions and are highly polarized when in culture, forming distinct apical and basolateral compartments imperative for transport and barrier studies. Since MDCK cells can produce stable monolayers within 3–4 days (rather than 14–21 days in their Caco-2 counterpart), MDCK cells have been adopted to improve throughput of barrier integrity studies within short culture periods [10]. While both Caco-2 and MDCK cell lines produce stable epithelial monolayers, these models lack the cellular heterogeneity typically found in the native gut microenvironment. To overcome this, co-culture and triple culture insert-based models (e.g., incorporating M cells and goblet cells to produce a mucus layer) have been utilised in various studies [11–14].

Significant advances have been made in developing more physiologically relevant in vitro models of the human gut epithelial barrier, using 2- and 3-dimensional cultures of primary stem cells from human biopsy and resection tissue. These ‘miniguts’ or organoids can generate the cellular diversity that epithelial cell lines lack, such as intestinal stem cells, enteroendocrine cells and Paneth cells [15]. These organoids also can retain their crypt-villus organization and more accurately mimic tight junction integrity and mucus uniformity [15]. Gut organoids can also be derived from patient-specific stem cells making them highly useful for gut barrier studies tailored for personalized medicine [15]. However, while gut organoids are established methods for improving the cellular heterogeneity of in vitro models, they have a number of limitations such as high cost and large heterogeneity between donors. Organoid models also remain difficult to culture for experiments that necessitate a high throughput. Nonetheless, the advantages of employing gut organoids to study gut barrier physiology have led to their assembly within cell culture inserts and in more advanced microfluidic ‘gut-on-a-chip’ models [15,16]. Organoid assembly in a culture insert advances on the above monolayer systems and is compatible with current TEER electrodes. However, because gut-on-a-chip models are typically complex and miniaturized, they are usually not compatible with TEER electrodes (although there have been attempts to integrate TEER electrodes within microfluidic chips [17]).

To address this issue and to allow an alternative assessment of epithelial barrier integrity, FITC-dextran is also commonly used to measure barrier permeability in both cell culture insert and microfluidic models [18,19]. FITC-dextran consists of the fluorophore fluorescein conjugated to dextran molecules of specific molecular weights. By introducing FITC-dextran to the apical (upper insert chamber) or endothelial channel (in gut-on-a-chip systems) sides of the epithelial barrier and measuring the fluorescence intensity of samples collected from the basolateral (well in a cell culture plate) or epithelial channel (gut-on-a-chip) sides, it is thus possible to further assess epithelial barrier integrity (i.e., by measuring apical-to-basal / endothelial-to-epithelial transmission of FITC-dextran across the epithelial barrier). This fluorescent methodology allows FITC-dextran to be used for assessment of a range of barrier defects (by using dextrans of varying molecular weights).

Despite FITC-dextran being able to be used for the assessment of a range of barrier dysfunctions, there are numerous drawbacks of using FITC-dextran for testing in vitro. Namely, FITC-dextran is expensive, not clinically approved and there is no consensus agreement to a molecular weight cut-off to determine barrier damage. Additionally, the molecular weights of FITC-dextrans are typically several kDa or higher, which does not correspond with the molecules used for barrier integrity assessment in humans. For example, the lactulose-mannitol (L:M) test is commonly deployed for gut barrier assessment in vivo and in humans [20,21]. Lactulose and mannitol are sugars with molecular weights of 342 g mol-1 and 182 g mol-1 respectively, much lower than the molecular weight of the smallest FITC-dextrans (minimum molecular weight ≈ 4000 g mol-1). These molecular weight discrepancies suggest that the permeation of FITC-dextran may not be a physiologically relevant assessment of gut barrier integrity.

Hence, to investigate development or more advanced and more physiologically relevant gut barrier integrity assays, this exploratory study involved a preliminary investigation of the permeation of three fluorescent contrast agents – fluorescein, FITC-dextran (4 kDa) and methylene blue – across MDCK cell monolayers. Contrast agent permeation was assessed in a insert-based system under various severities of epithelial barrier damage to investigate the potential use of fluorescent contrast agents with lower molecular weights than FITC-dextran for measurements of gut barrier integrity. An MDCK insert-based monolayer model was chosen for this purpose as it allowed formation of well-developed tight junctions over short culture periods, which was well suited to a pilot study investigating multiple contrast agents. Importantly, fluorescein and methylene blue have molecular weights comparable to lactulose and can therefore be expected to provide gut barrier damage readouts at physiologically relevant levels. Furthermore, both fluorescein and methylene blue are clinically approved and so can potentially be used to assess epithelial barrier integrity in both laboratory and clinical environments. Indeed, fluorescein was utilized for in vivo assessment of human gut barrier function in several recent studies [22–27]. Finally, both fluorescein and methylene blue are considerably cheaper than FITC-dextran offering the opportunity to reduce the costs of fluorescent epithelial barrier function assays, especially those involving ‘gut-on-a-chip’ approaches where use of FITC-dextran is necessitated [28].

2. Materals and methods

2.1. Cell culture

MDCK cells (provided by Professor Julian R. Marchesi, Imperial College London) were maintained in complete DMEM which constituted DMEM (Thermo) supplemented with 10% FBS (v/v, Thermo), 1% GlutaMAX (v/v, Thermo) and 1% penicillin-streptomycin (v/v, Thermo) and were passaged upon reaching 80% confluency. Cells were incubated in a 37°C, 5% CO2 humidified cell culture incubator.

2.2. Measurement of transepithelial electrical resistance in MDCK monolayers

When cultured on cell culture inserts, MDCK cells can polarize into apical and basolateral domains within a short time frame (3−4 days) allowing for the generation of 2D monolayers to test epithelial barrier integrity. MDCK cells were seeded onto cell culture inserts (pore size: 0.4 μm) in 24-well plates at a density of 4x105 cells ml−1. Cells were maintained in complete DMEM medium and were used for experiments 4 days after seeding. Monolayer integrity was confirmed by TEER measurements taken with an EVOM3 Manual (World Precision Instruments). TEER was measured in Ohms cm2 and calculated as:

TEERmonolayer=(TEERsample−TEERblank)×A (1)

where TEERblank is the raw TEER value measured in a blank cell culture well (with no cells but containing complete DMEM), TEERsample is the raw TEER value measured in the sample (i.e., cell culture well with cells in complete cell culture medium), and A is the surface area of the cell culture insert (0.33 cm2).

Both apical and basolateral compartments of the 24-well cell culture insert system were filled with complete DMEM (200 µl in apical chamber, 1 ml in basolateral chamber). These DMEM volumes provided approximately equal liquid heights in the apical and basolateral compartments and also ensured that the medium in the basolateral compartment was in contact with the semi-permeable insert. This meant that the pressures of the two liquid columns were balanced and allowed for nutrient exchange between chambers.

2.3. Preparation of Ethanol, EGTA and fluorescent tracer solutions

Ethanol and EGTA (egtazic acid) were used to induce varying levels of epithelial barrier damage. Ethanol solutions at concentrations of 0.5%, 1%, 2%, 3%, 4%, 5% and 6% (v/v) were prepared by diluting absolute (≈ 100%) ethanol in DI water to 70% and then subsequently diluting 70% ethanol down to respective ethanol concentrations using complete DMEM. Cells were treated with the resulting ethanol solutions. EGTA was directly diluted in DMEM and used at concentrations of 0.25 mM, 0.5 mM, 1 mM, 2 mM, 4 mM, 8mM and 16 mM.

Fluorescein (Sigma), FITC-dextran (Sigma) and methylene blue (Sigma) solutions (concentrations shown in Table 1) were prepared by dissolving powdered dyes in complete DMEM medium prior to their addition onto MDCK monolayers. As the same concentration of different fluorescent tracers may yield different fluorescent intensities (e.g., due to differences in the fluorescence quantum yield), we calculated individual concentrations for each fluorescent contrast agent that would correspond to approximately the same fluorescence intensity across tracers. This was performed (prior to experiments) by standardising calculations to 100 μg mL−1 of FITC-dextran, an acceptable concentration in FITC-dextran flux assays. We first calculated the mass of FITC in 100 μg of FITC-dextran (based on the molar masses of FITC and glucose and the approximate FITC:glucose molar ratio in FITC-dextran reported by the supplier) as 0.835 μg. This was then chosen as the mass of fluorescein to give approximately equal fluorescence intensity to a mass of 100 μg of FITC-dextran.

Table 1. Fluorescent contrast agent concentrations applied to cell monolayers (i.e., concentrations in apical chambers) for each contrast agent (fluorescein, FITC-dextran and methylene blue) and each experimental concentration used (shown as percentages of maximal concentrations). All concentrations shown are in μg ml-1.

Contrast agent Control 6.25% 25% 100%
FITC-dextran 0 6.25 25 100
Fluorescein 0 0.052 0.209 0.835
Methylene Blue 0 0.091 0.365 1.461

To calculate the equivalent mass for methylene blue, we then took the fluorescence quantum yield (ΦF) of fluorescein and methylene blue (as provided by the manufacturer: 0.91 for fluorescein, 0.52 for methylene blue) and calculated a correction factor (C) for methylene blue of C=ΦFMethylene Blue/ΦFFluorescein. This accounted for the lower fluorescence quantum yield of methylene blue (relative to fluorescein) and yielded a mass of methylene blue of 1.461 μg. These masses then determined the maximum concentrations of FITC-dextran, fluorescein and methylene blue to be used in experiments (i.e., 100 μg ml−1 FITC-dextran, 0.835 μg ml−1 fluorescein, and 1.461 μg ml-1 methylene blue). These concentrations are hereafter referred to as 100% concentrations (as they were the maximal concentrations applied to MDCK monolayers). Two dilutions were also prepared for each contrast agent at equivalent concentrations of 6.25% and 25%. MDCK experiments were performed at all three contrast agent concentrations (6.25%, 25% and 100%) and under control conditions (with pure DMEM applied to the apical chamber containing no fluorescent contrast agent; referred to as 0% contrast agent concentration). The actual concentrations of each contrast agent (in the apical chamber) for each condition are presented in Table 1.

2.4. Permeability assay

To assess translocation of fluorescent contrast agents across the epithelial barrier, fluorescein (332.21 g mol−1), methylene blue (319.85 g mol−1) and FITC-dextran (4000 g mol−1) were added (individually) into the apical compartment of the insert-based system, either alone or in combination with ethanol or EGTA. Following addition of fluorescent contrast agents, the apical compartment contained 200 µl of fluorescent tracer mixed with complete DMEM, and the basolateral chamber contained 1 ml DMEM. As above, we observed that these volumes led to approximately equal liquid heights in the two chambers ensuring that the pressure of the liquid columns were balanced.

We note that a horizontal plate shaker (or other method to avoid an unstirred water layer) was not used in this study. As the chosen fluorescent contrast agents (fluorescein, FITC-dextran and methylene blue) are all highly soluble in aqueous media, an unstirred water layer and/or use of a horizontal plate shaker would not be expected to significantly affect results. Therefore, methodological steps to provide mechanical stirring were not used here.

Baseline samples (100 µl) were taken from the apical and basolateral chambers prior to addition of contrast agents and topped back up with 100 µl of complete DMEM. Following the addition of fluorescent agents, samples (100 µl) were collected from both chambers at 3 hours after dye addition (and the experiment was terminated after sampling). To enable fluorescence measurements, samples were transferred to new multiwell plates and fluorescence intensity was measured using a CLARIOstar fluorescence plate reader (BMG Labtech). Excitation and emission wavelengths used for fluorescence measurements for the three contrast agents are shown in Table 2.

Table 2. | Excitation and emission wavelengths used for fluorescence measurements of each fluorescent contrast agent (performed using CLARIOstar plate reader).

Contrast agent Excitation (nm) Emission (nm)
FITC-dextran 490 520
Fluorescein 490 520
Methylene Blue 665 686

TEER measurements were also recorded at baseline and at 1.5 and 3 hours. To account for variability across experiments, changes in TEER with respect to baseline (ΔTEER) were calculated in all cases as:

ΔTEER=TEERmonolayer(t)−TEERmonolayer(t0) (2)

where TEERmonolayer(t) represents the absolute TEER value of the monolayer (calculated according to equation 1 above) at time t (i.e., at a measurement/sampling time point of 1.5 hours or 3 hours), and TEERmonolayer(t0) represents the absolute TEER value of the monolayer at baseline (i.e., t = 0 hours).

Statistical analysis was performed in GraphPad Prism v8 using the Shapiro-Wilk test to assess normality followed by two-way ANOVA to compare the fluorescence intensities observed under different conditions (i.e., at different contrast agent and disruptor concentrations). The experimental procedure is illustrated in Fig 1A.

Fig 1. Optimization of epithelial barrier disruptor concentrations | (A) Schematic of the experimental apparatus used to measure MDCK monolayer integrity in a cell culture insert-based system.

Fig 1

(B-C) Changes in TEER measured at 1.5 hours and 3 hours after the addition of different concentrations of (B) ethanol (v/v) and (C) EGTA. Two biological repeats (n = 2), one technical repeat (N = 1). Error bars shown on the figure represent the standard deviation.

3. Results

3.1. Confirmation of cell monolayer formation and optimization of epithelial barrier disruption

To study fluorescent contrast agent permeation across the epithelial barrier, a 2D cell culture insert system was deployed consisting of an MDCK monolayer cultured in the apical compartment of the insert-based system (Fig 1A). After 3 days, TEER measurements were taken to assess the establishment of MDCK monolayers prior to experimentation (Fig 1A). In this set of experiments monolayers that showed TEER readings greater than 1000 Ωcm2 were used. In this case TEER values of approximately 1500 Ωcm2 were observed in all wells, indicating the formation of monolayers with well-established tight junctions in all cases.

Ethanol and EGTA were applied to MDCK monolayers and TEER measurements were taken 1.5 hours and 3 hours post-exposure to optimise gut barrier disruption for downstream dye permeation experiments (Fig 1B,C). As expected, increasing the concentration of either ethanol or EGTA led to greater negative changes in TEER values (Fig 1B,C). Interestingly, the effect of ethanol after 3 hours is less pronounced than that observed after 1.5 hours (Fig 1B) and less pronounced than that observed with EGTA (Fig 1B,C). This difference was tentatively attributed to the evaporation of ethanol due to the small volumes used, leading to less barrier damage over time. The most dramatic change in gut barrier disruption was observed at 16 mM EGTA after 3 hours (Fig 1C).

3.2. Impact of Fluorescent contrast agents on epithelial barrier integrity

After validating which concentrations of barrier impairing agents effectively disrupt the MDCK monolayer, subsequent experiments explored the change in TEER value 3 hours after addition of fluorescein, FITC-dextran and methylene blue. These experiments were performed under control conditions (i.e., no barrier disruption) and in the presence of barrier disrupting agents (ethanol or EGTA) (Fig 2) to investigate any impact of the fluorescent contrast agents on epithelial barrier integrity. Concentrations of ethanol (v/v: 0.5%, 2% and 5%) and EGTA (0.5 mM, 2 mM and 8 mM) were chosen based on the experiments above (Fig 1 B,C) to provide a range of barrier damage severities.

Fig 2. Assessment of epithelial barrier integrity using TEER for different concentrations of barrier disruptors (ethanol – left, EGTA – right) and for different fluorescent contrast agents.

Fig 2

(A-F) Relative TEER changes (0-3 hours) for each fluorescent contrast agent with different levels of barrier disruption and different contrast agent concentrations. Control values reperesent TEER values for wells where no disruptor was applied. Two biological repeats (n = 2), one technical repeat (N = 1). Error bars represent standard deviations. Inset points shown alongside bars represent values from individual experiments.

As the severity of barrier damage increases (i.e., for increasing concentrations of ethanol and EGTA), a greater decrease in TEER value is observed for all fluorescent contrast agents and all contrast agent concentrations (Fig 2A-2F). This further demonstrates that higher concentrations of ethanol and EGTA induce greater degrees of epithelial barrier damage.

For fluorescein and FITC-dextran, the change in TEER value is approximately constant for all contrast agent concentrations (0–100%; see Fig 2A-2D). This data demonstrates that fluorescein and FITC-dextran (at the concentrations and incubations times used here) do not inherently damage the MDCK epithelial barrier (Fig 2 A-D). However, for methylene blue, TEER values appear to decrease for increasing methylene blue concentration (with no addition of ethanol/EGTA) (Fig 2E,2F). In addition a greater change in TEER value is observed at the highest methylene blue concentration (100%; 4.57 µM) for both barrier disrupting agents and for all disruptor concentrations (Fig 2,2F). This is likely to be the consequence of methylene blue toxicity on cell monolayers at higher concentrations [29]. Nonetheless, for all lower methylene blue concentrations tested here (25% and below; < 1.14 µM), changes in TEER values observed following addition of disruptors are approximately consistent across different dye concentrations and do not differ from those observed in control experiments (i.e., where no methylene blue was added; Fig 2E,2F).

Together, these results show that fluorescein and methylene blue (in addition to FITC-dextran) could be suitable fluorescent contrast agents for assessment of epithelial barrier integrity (albeit with methylene blue only being suitable at the lower concentrations used here) as their impact on epithelial barrier integrity is negligible compared to that of the barrier disrupting agents (ethanol and EGTA).

3.3. Fluorescein and methylene blue provide analogous readouts of epithelial barrier integrity to FITC-dextran

Permeation of fluorescent contrast agents across epithelial barriers was further explored to determine which dyes were able to effectively pass through MDCK monolayers under different severities of epithelial barrier damage. For all fluorescent contrast agents, the general trends observed are that fluorescence intensity increases with both increasing fluorescent dye concentration and increasing disruptor concentration (Fig 3).

Fig 3. Permeation of fluorescent contrast agents across MDCK cell monolayers under different severities of epithelial barrier damage.

Fig 3

(A-F) Fluorescence intensity recorded in basolateral compartments for each fluorescent contrast agent with different concentrations of ethanol (left) and EGTA (right) used to induce epithelial barrier damage. (A, B) fluorescein; (C,D) FITC-dextran; (E, F) methylene blue. Two biological repeats (n = 2), one technical repeat (N = 1). Data were statistically compared across different dye and disruptor concentrations using 2-way ANOVA. Asterisks indicate statistically significant differences with p < 0.05 (*), p < 0.01 (**), p < 0.001 (***) and p < 0.0001 (****). Error bars represent standard deviations. Inset points shown alongside bars represent values from individual experiments.

The Shaprio-Wilk test demonstrated normality for all but one condition (2 mM EGTA, FITC-Dextran). Therefore a two-way ANOVA was used for further statistical analysis. Importantly, fluorescence signals are lowest for the 0% contrast agent concentration in all cases (Fig 3), demonstrating that background signals (e.g., from fluorescent compounds in cell culture media) did not detrimentally affect our results. The most pronounced increases in fluorescence intensity are observed with 8 mM EGTA disruption (Fig 3B,D,F), in agreement with the results presented in Fig 1B,C (which show that 8 mM EGTA causes greater barrier disruption than any ethanol concentration at the 3 hour time point).

Interestingly, the trends observed for fluorescein (Fig 3A,B) are very similar to those observed for FITC-dextran (Fig 3C,D), but with higher fluorescence intensities measured using fluorescein (for all experimental conditions). This suggests that fluorescein can provide an in vitro readout of epithlial barrier damage in an analogous manner to FITC-dextran, but with a greater fraction of contrast agent crossing the epithelial barrier under all conditions. Interestingly, this may suggest that fluorescein could allow investigation of a wider range of barrier defects than FITC-dextran, by providing sensitivity to lower levels of barrier damage.

For methylene blue (Fig 3E,F), the trends observed are similar to those for both fluorescein and FITC-dextran (Fig 3A-3D) up to dye concentrations of 25%. At the highest methylene blue concentration (100%), the impact of ethanol and EGTA appears to be limited, with similar fluorescence intensities being measured at all disruptor concentrations (Fig 3E,F, 100% dye concentration). As discussed above, this was tentatively attributed to cellular toxicity of methylene blue at high concentrations as methylene blue has been shown to be toxic in a number of different cell lines [30,31.32]. In combination with ethanol or EGTA, the highest concentration of methylene blue appears to induce significant barrier disruption leading to observation of high fluorescence intensities for all disruptor concentrations. Despite this, at lower contrast agent concentrations (25% or below), methylene blue exhibited similar trends to those observed with FITC-dextran (compare Fig 3E,F against Fig 3C,D).

Hence, taken together, these results demonstrate that both fluorescein and methylene blue are promising candidates for monitoring of epithelial barrier integrity in vitro. Both dyes exhibit analogous permeation behaviour to FITC-dextran, which has been widely used for epithelial barrier assessment.

3.4. Permeation of all fluorescent contrast agents correlate with changes in TEER

Linear regression analysis was applied to the fluorescence intensities observed for fluorescein, FITC-dextran and methylene blue (at all contrast agent concentrations) to investigate the relationship between fluorescence intensity and change in TEER across different barrier disruption conditions (Fig 4; Supplementary Figures S1-S3 in S2 File). At a contrast agent concentration of 0%, correlation analysis demonstrates no change in fluorescence intensity despite changes in barrier integrity (as expected due to the lack of contrast agent; see Supplementary Figure S1 in S2 File). At the 6.25% contrast agent concentration, all fluorescent tracers show statistically significant correlations between fluorescence intensity and ΔTEER for both EGTA and ethanol disruption (Fig 4). Similar trends are observed for contrast agent concentrations of 25% and 100% (see Supplementary Figures S2 and S3 in S2 File). Together, this further validates methylene blue and fluorescein as potential markers of epithelial barrier integrity.

Fig 4. Regression analysis of fluorescence intensity and ΔTEER across various barrier disruptors at 6.25% dye concentration.

Fig 4

Fluorescence intensity plotted against ΔTEER under EGTA (left) and ethanol (right) disruption. (A, B) Fluorescein; (C, D) FITC-dextran; (E, F) methylene blue. Two biological repeats (n = 2), one technical repeat (N = 1). Data was statistically analysed using Spearman’s Rank test. Correlation coefficients (r) and p-values are shown in their corresponding graphs. *p < 0.05. Error bars represent standard deviations.

4. Discussion and conclusions

In this exploratory study using in vitro epithelial barrier models, it is reported that fluorescein and methylene blue provide analogous readouts of epithelial barrier integrity to FITC-dextran, which is commonly used for gut epithelial barrier assessment in cellular, organoid and animal models. While results should be viewed as preliminary due to the low number of biological repeats, these findings have a number of important implications.

First, this data presents the opportunity to significantly reduce costs of current fluorescent permeability assays that use FITC-dextran (i.e., due to the lower cost and higher fluorescence quantum yields of fluorescein and methylene blue compared to FITC-dextran). Second, these findings indicate the possibility to explore physiologically relevant gut barrier defect sizes by assessing epithelial barrier integrity using fluorescent contrast agents in vitro that more closely mimic the sizes of the substances used in vivo for clinical assessment of gut barrier function (such as the sugars lactulose and mannitol used in the L:M test). Furthermore, as fluorescein and methylene blue are clinically approved contrast agents, these results also suggest the potential to assess epithelial barrier function in both laboratory and clinical settings using the same agents. This exciting advance would permit more coherent investigation of gut barrier function in health and disease by allowing comparable measurements to be made in cells, tissues, animals and humans. Indeed, this work also further validates the use of fluorescein for assessment of gut barrier function in vivo, as previously reported [21–27], as it demonstrates greater transmission of fluorescein across epithelial barriers under greater degrees of barrier disruption.

Third, and finally, these results indicate potential to explore multiple barrier defect sizes simultaneously by combining multiple fluorescent contrast agents with differing emission profiles and molecular weights. In particular, the results in this paper indicate that methylene blue and FITC-dextran could be used simultaneously (due to their differing fluorescence emission profiles) to provide a multifaceted assessment of epithelial barrier integrity.

The key limitations of this study include the use of an insert-based MDCK monolayer model (one of the simplest available epithelial barrier models), the investigation of a small number of fluorescent contrast agents, limits to the scope of the experiments performed (e.g., measuring contrast agent permeation at a maximum of two time points, studying response to two barrier disruptors, etc.), and a low number of biological replicates (n = 2). As such, future work will now entail extending this study to investigate behaviour of a larger number of contrast agents in more complex gut barrier models and under a wider range of barrier degradation conditions.

Importantly, the aim of this study was to further evaluate whether fluorescein and methylene blue can serve as fluorescent tracers of paracellular permeability analogously to FITC-dextran in in vitro barrier assays. As such, this preliminary study focused on measuring changes in TEER and fluorescence intensity of samples collected from basolateral chambers in addition to investigating TEER-fluorescence correlations. The data therefore served to demonstrate the potential of fluorescent contrast agent permeation as a proxy for epithelial barrier integrity.

Previous studies using fluorescence for in vitro epithelial barrier assessment have reported measurements of apparent permeability (Papp), which provides a quantification of the mass of fluorescent contrast agent transported across the barrier [33–35]. Measurements of Papp enable comparison of data across studies but are still limited by variability in the integrity of the underlying monolayer (i.e., variability in baseline monolayer integrity will induce Papp variability across studies). In this study, we chose not to perform full Papp measurements (which require sample collection at multiple time points). Instead, simpler fluorescence intensity measurements at single time points were undertaken alongside TEER assessment to serve as proof-of-concept. While this does not enable direct comparison of fluorescent contrast agent transport against previous studies, it nonetheless allows preliminary demonstration of increased contrast agent transport across barriers under conditions of greater barrier damage.

This dual fluorescence-TEER approach has been used in previous studies, with significant inverse correlations having been reported between TEER and paracellular tracer flux in Caco-2 epithelial barrier models using probes including FITC-dextran and Lucifer Yellow [36–38]. Indeed, this dual approach has been applied in vitro under different contexts, for example to assess barrier changes induced by protein hydrolysates in the intestinal epithelium in Type 1 Diabetes [36]. Our pilot study followed a similar methodological approach to those investigations.

While FITC-dextran is one of the most widely used fluorescent tracers for cell monolayer permeability assessment, fluorescein has also previously been used for this purpose [39–41]. In this context, the comparable directional changes observed with all three fluorescent tracers (fluorescein, FITC-dextran and methylene blue) in relation to TEER serve to provide further validation of fluorescein as a permeability marker and proof-of-concept for methylene blue. To the best of our knowledge, methylene blue has not previously been reported as a validated fluorescent tracer for in vitro paracellular permeability assays in the wider literature. However, its behaviour has been directly compared with fluorescein in detecting tissue barrier leaks, where it performs similarly. For example, methylene blue and fluorescein have both been used to detect paracellular leaks in ex vivo porcine intestinal segments. Here, our preliminary data demonstrates similar behaviour in vitro between methylene blue and fluorescein further suggesting that methylene blue could be repurposed as a functional fluorescent contrast agent for in vitro studies investigating paracellular leakage.

A further limitation of the present study is that cytotoxicity was not directly assessed, and therefore the relative contributions of tight junction modulation versus epithelial cell loss to the observed decrease in TEER cannot be fully distinguished. However, epithelial barrier dysfunction in vivo is multifactorial and not solely driven by tight junction alterations. In conditions such as environmental enteropathy (EE) and inflammatory intestinal disorders, epithelial shedding and cell loss can also contribute to increased paracellular permeability [42,43]. In this context, the barrier disruption in the present model may reflect a combination of mechanisms including tight junction modulation and epithelial cell loss induced by ethanol or EGTA. While this may limit mechanistic resolution, it does not preclude functional assessment of paracellular permeability. Indeed, the primary aim of this study was to evaluate tracer flux under varying degrees of barrier disruption. The observed correlation between TEER reduction and increased tracer permeability supports the utility of this approach. Further studies incorporating direct measures of cell viability will nonetheless be important to further distinguish between the contributions of these mechanisms.

Importantly, however, the ethanol and EGTA concentrations used in this study were selected based on prior work demonstrating barrier modulation in epithelial monolayers without overt cytotoxicity. In particular, Ma et al. reported that exposure of filter-grown Caco-2 monolayers to ethanol concentrations of up to 10% (v/v) resulted in a dose-dependent decrease in TEER associated with disruption of the tight junction protein ZO-1, while lactate dehydrogenase (LDH) release assays indicated minimal cytotoxicity within this range [44]. On this basis, ethanol concentrations below 10% were selected in the present study to induce graded barrier disruption. Consistent with these findings, a concentration-dependent decrease in TEER was observed, with a slightly greater magnitude compared toobersvations made by Ma et al. (which may reflect differences between MDCK and Caco-2 models) [44].

Similarly, EGTA concentrations were chosen based on previous findings where Caco-2 monolayers were treated with EGTA concentrations ranging from 1–7 mM and exhibited up to a 60% decrease in TEER following treatment [43]. Interestingly, Panou et al. found that after 21 hours of recovery in cell culture medium TEER returned to baseline for EGTA concentrations ranging from 1–5 mM with ~90% recovery for Caco-2 monolayers treated with 7 mM EGTA [45]. This suggests that the lower EGTA concentrations used in this study (0.5 mM and 2 mM) are sufficient to disrupt tight junction integrity without inducing cytotoxicity, while the higher concentration (8 mM) is likely to cause only limited cytotoxicity. As such, all of the EGTA and ethanol concentrations used in this study are expected to induce tight junction disruption while causing only limited (or zero) cytotoxicity.

Overall, despite the limitations discussed above, this pilot study demonstrates that fluorescein and methylene blue – low-cost, clinically approved fluorescent contrast agents – can be used to assess epithelial barrier integrity in an analogous manner to FITC-dextran. This implies potential to dramatically reduce costs of in vitro experiments and to provide a range of exciting opportunities for more advanced and coherent investigation of epithelial barrier function in health and disease.

Supporting information

S1 File. Raw Data for Figures.

Raw data made available for the figures within this manuscript.

(XLSX)

pone.0358746.s001.xlsx (24.5KB, xlsx)
S2 File. Supplementary Figures.

Supplementary file containing supplementary figures S1-S3.

(DOCX)

pone.0358746.s002.docx (3.2MB, docx)

Acknowledgments

The authors acknowledge both financial support from the GI Tools Project (MRC; grant number – MR/V012452/1) and intellectual input from all members of the GI Tools Consortium, which helped guide and support protocol development, experimental design, data collection and data analysis. All members of the GI Tools Consortium are listed below. GI Tools Consortium: Beatrice Amadi, Rosemary Banda, Ellen Besa, Claire Bourke, Mutsa Bwakura-Dangarembizi, Ian Chisenga, Christine Edwards, Gary Frost, Isabel Garcia Perez, Mahek Jain, Leolin Katsidzira, Lydia Kazhila, Paul Kelly, Mirriam Kunaka, Kathryn Maitland, Nilanjan Mandal, Julian R Marchesi, Callum Melvin, Elena Monfort Sanchez, Douglas Morrison, Monica N Mweetwa, Mulima Mwiinga, Perpetual Ngalande, Tracy N Phiri, Joram M Posma, Ruari Robertson, Jose Ivan Serrano Contreras, Aaron Konzani Tembo, Alex J Thompson, James W Weatherill.

Data Availability

All relevant data are within the manuscript and its Supporting Information files.

Funding Statement

This article reports independent research supported by: the UK Medical Research Council (MRC; grant number – MR/V012452/1); the National Institute for Health Research (NIHR) Imperial Biomedical Research Centre (BRC); and the Imperial College London UKRI Impact Acceleration Account supported by MRC (grant number – MR/X502959/1) and Rosetrees Trust. J.A. acknowledges an Imperial College Research Fellowship. The views expressed in this publication are those of the authors and not necessarily those of the NHS, the NIHR or the Department of Health. The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.

References

  • 1.Di Tommaso N, Gasbarrini A, Ponziani FR. Intestinal barrier in human health and disease. Int J Environ Res Public Health. 2021;18(23):12836. doi: 10.3390/ijerph182312836 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Kesika P, Suganthy N, Sivamaruthi BS, Chaiyasut C. Role of gut-brain axis, gut microbial composition, and probiotic intervention in Alzheimer’s disease. Life Sci. 2021;264:118627. doi: 10.1016/j.lfs.2020.118627 [DOI] [PubMed] [Google Scholar]
  • 3.Mowat AM, Agace WW. Regional specialization within the intestinal immune system. Nat Rev Immunol. 2014;14(10):667–85. doi: 10.1038/nri3738 [DOI] [PubMed] [Google Scholar]
  • 4.Aho VTE, Houser MC, Pereira PAB, Chang J, Rudi K, Paulin L, et al. Relationships of gut microbiota, short-chain fatty acids, inflammation, and the gut barrier in Parkinson’s disease. Mol Neurodegener. 2021;16(1):6. doi: 10.1186/s13024-021-00427-6 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Paray BA, Albeshr MF, Jan AT, Rather IA. Leaky gut and autoimmunity: an intricate balance in individuals health and the diseased state. Int J Mol Sci. 2020;21(24):9770. doi: 10.3390/ijms21249770 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Tincati C, Douek DC, Marchetti G. Gut barrier structure, mucosal immunity and intestinal microbiota in the pathogenesis and treatment of HIV infection. AIDS Res Ther. 2016;13:19. doi: 10.1186/s12981-016-0103-1 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Haddad MJ, Sztupecki W, Delayre-Orthez C, Rhazi L, Barbezier N, Depeint F, et al. Complexification of in vitro models of intestinal barriers, a true challenge for a more accurate alternative approach. Int J Mol Sci. 2023;24(4):3595. doi: 10.3390/ijms24043595 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Lopez-Escalera S, Wellejus A. Evaluation of Caco-2 and human intestinal epithelial cells as in vitro models of colonic and small intestinal integrity. Biochem Biophys Rep. 2022;31:101314. doi: 10.1016/j.bbrep.2022.101314 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Forlano R, Martinez-Gili L, Takis P, Miguens-Blanco J, Liu T, Triantafyllou E, et al. Disruption of gut barrier integrity and host-microbiome interactions underlie MASLD severity in patients with type-2 diabetes mellitus. Gut Microbes. 2024;16(1):2304157. doi: 10.1080/19490976.2024.2304157 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Sánchez EM, Avery DJ, Darzi PA, Thompson AJ. Development of a compact fluorescence spectroscopy sensor for non-invasive monitoring gut function in Biophotonics Congress: Biomedical Optics 2022 (Translational, Microscopy, OCT, OTS, BRAIN), Technical Digest Series. 2022. TS2B.6.
  • 11.Carrière V, Lesuffleur T, Barbat A, Rousset M, Dussaulx E, Costet P, et al. Expression of cytochrome P-450 3A in HT29-MTX cells and Caco-2 clone TC7. FEBS Lett. 1994;355(3):247–50. doi: 10.1016/0014-5793(94)01199-0 [DOI] [PubMed] [Google Scholar]
  • 12.Yuan H, Chen CY, Chai GH, Du YZ, Hu FQ. Improved transport and absorption through gastrointestinal tract by PEGylated solid lipid nanoparticles. Molecular Pharmaceutics. 2013;10(5):1865–73. doi: 10.1021/mp300649z [DOI] [PubMed] [Google Scholar]
  • 13.Ponce de León-Rodríguez MDC, Guyot J-P, Laurent-Babot C. Intestinal in vitro cell culture models and their potential to study the effect of food components on intestinal inflammation. Crit Rev Food Sci Nutr. 2019;59(22):3648–66. doi: 10.1080/10408398.2018.1506734 [DOI] [PubMed] [Google Scholar]
  • 14.García-Rodríguez A, Vila L, Cortés C, Hernández A, Marcos R. Exploring the usefulness of the complex in vitro intestinal epithelial model Caco-2/HT29/Raji-B in nanotoxicology. Food Chem Toxicol. 2018;113:162–70. doi: 10.1016/j.fct.2018.01.042 [DOI] [PubMed] [Google Scholar]
  • 15.Almeqdadi M, Mana MD, Roper J, Yilmaz ÖH. Gut organoids: mini-tissues in culture to study intestinal physiology and disease. Am J Physiol Cell Physiol. 2019;317(3):C405–19. doi: 10.1152/ajpcell.00300.2017 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Bein A, Shin W, Jalili-Firoozinezhad S, Park MH, Sontheimer-Phelps A, Tovaglieri A, et al. Microfluidic organ-on-a-chip models of human intestine. Cell Mol Gastroenterol Hepatol. 2018;5(4):659–68. doi: 10.1016/j.jcmgh.2017.12.010 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Henry OYF, Villenave R, Cronce MJ, Leineweber WD, Benz MA, Ingber DE. Organs-on-chips with integrated electrodes for trans-epithelial electrical resistance (TEER) measurements of human epithelial barrier function. Lab Chip. 2017;17(13):2264–71. doi: 10.1039/c7lc00155j [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Mitrofanova O, Nikolaev M, Xu Q, Broguiere N, Cubela I, Camp JG, Bscheider M, Lutolf MP. Bioengineered human colon organoids with in vivo-like cellular complexity and function. Cell Stem Cell. 2024 Aug 1;31(8):1175–86.e7. doi: 10.1016/j.stem.2024.05.007 [DOI] [PubMed] [Google Scholar]
  • 19.Gleeson JP, Zhang SY, Subelzu N, Ling J, Nissley B, Ong W, et al. Head-to-head comparison of Caco-2 Transwell and gut-on-a-chip models for assessing oral peptide formulations. Molecular Pharmaceutics. 2024;21(8):3880–8. doi: 10.1021/acs.molpharmaceut.4c00210 [DOI] [PubMed] [Google Scholar]
  • 20.Denno DM, VanBuskirk K, Nelson ZC, Musser CA, Hay Burgess DC, Tarr PI. Use of the lactulose to mannitol ratio to evaluate childhood environmental enteric dysfunction: a systematic review. Clin Infect Dis. 2014;59(Suppl 4):S213-9. doi: 10.1093/cid/ciu541 [DOI] [PubMed] [Google Scholar]
  • 21.Gan J, Nazarian S, Teare J, Darzi A, Ashrafian H, Thompson AJ. A case for improved assessment of gut permeability: a meta-analysis quantifying the lactulose:mannitol ratio in coeliac and Crohn’s disease. BMC Gastroenterol. 2022;22(1):16. doi: 10.1186/s12876-021-02082-z [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Maurice J, Lett AM, Skinner C, Lim A, Richardson M, Thomas AP, et al. Transcutaneous fluorescence spectroscopy as a tool for non-invasive monitoring of gut function: first clinical experiences. Sci Rep. 2020;10(1):16169. doi: 10.1038/s41598-020-73149-2 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Sanchez EM, Avery J, Gan J, Qian J, Mandal N, Agarwal A, et al. Transcutaneous fluorescence spectroscopy: development and characterization of a compact, portable, and fiber-optic sensor. J Biomed Opt. 2024;29(2):027003. doi: 10.1117/1.JBO.29.2.027003 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Mbuki R, Chileya S, Thompson AJ, Kelly P, Kayamba V. Rapid testing of gut permeability using oral fluorescein and confocal laser endomicroscopy in Zambian adults. Trans R Soc Trop Med Hyg. 2021;115(10):1226–8. doi: 10.1093/trstmh/trab083 [DOI] [PubMed] [Google Scholar]
  • 25.Gan J, Monfort Sánchez E, Avery J, Barbouti O, Hoare J, Ashrafian H, et al. Non-invasive assessment of intestinal permeability in healthy volunteers using transcutaneous fluorescence spectroscopy. Methods Appl Fluoresc. 2022;10(4):10.1088/2050-6120/ac9513. doi: 10.1088/2050-6120/ac9513 [DOI] [PubMed] [Google Scholar]
  • 26.Gan J, Chen Q, Monfort Sanchez E, Mandal N, Xu J, Wang Z, et al. Non-invasive fluorescence sensing reveals changes in intestinal barrier function and gastric emptying rate in a first-in-human study of Crohn’s disease. Ther Adv Gastroenterol. 2025;18:17562848251361634. doi: 10.1177/17562848251361634 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Monfort Sanchez E, Mandal N, Mwiinga M, Banda R, Xu J, Mweetwa MN, et al. Non-invasive measurement of intestinal barrier function in environmental enteropathy using transcutaneous fluorescence sensing. Gut Science. 2026;1:e000011. doi: 10.1136/gutsci-2025-000011 [DOI] [Google Scholar]
  • 28.Srinivasan B, Kolli AR, Esch MB, Abaci HE, Shuler ML, Hickman JJ. TEER measurement techniques for in vitro barrier model systems. J Lab Autom. 2015;20(2):107–26. doi: 10.1177/2211068214561025 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Pakavathkumar P, Sharma G, Kaushal V, Foveau B, LeBlanc AC. Methylene Blue Inhibits Caspases by Oxidation of the Catalytic Cysteine. Scientific Reports. 2015;5:13730. doi: 10.1038/srep13730 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Campos Chaves Lamarque G, Cusicanqui Méndez DA, Arruda Matos A, José Dionísio T, Andrade Moreira Machado MA, Magalhães AC, et al. Cytotoxic effect and apoptosis pathways activated by methylene blue-mediated photodynamic therapy in fibroblasts. Photodiagnosis Photodyn Ther. 2020;29:101654. doi: 10.1016/j.pdpdt.2020.101654 [DOI] [PubMed] [Google Scholar]
  • 31.Lee YS, Wurster RD. Methylene blue induces cytotoxicity in human brain tumor cells. Cancer Lett. 1995;88(2):141–5. doi: 10.1016/0304-3835(94)03629-w [DOI] [PubMed] [Google Scholar]
  • 32.Turchiello RF, Oliveira CS, Fernandes AU, Gómez SL, Baptista MS. Methylene blue-mediated Photodynamic Therapy in human retinoblastoma cell lines. J Photochem Photobiol B. 2021;222:112260. doi: 10.1016/j.jphotobiol.2021.112260 [DOI] [PubMed] [Google Scholar]
  • 33.Biedermann F, Ghale G, Hennig A, Nau WM. Fluorescent artificial receptor-based membrane assay (FARMA) for spatiotemporally resolved monitoring of biomembrane permeability. Commun Biol. 2020;3(1):383. doi: 10.1038/s42003-020-1108-9 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Patient JD, Hajiali H, Harris K, Abrahamsson B, Tannergren C, White LJ, et al. Nanofibrous Scaffolds Support a 3D in vitro Permeability Model of the Human Intestinal Epithelium. Front Pharmacol. 2019;10:456. doi: 10.3389/fphar.2019.00456 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Zhao W, Han L, Bae Y, Manickam DS. Lucifer Yellow - A Robust Paracellular Permeability Marker in a Cell Model of the Human Blood-brain Barrier. J Vis Exp. 2019;150. doi: 10.3791/58900 [DOI] [PubMed] [Google Scholar]
  • 36.Beisl J, Varga E, Braun D, Warth B, Ehling-Schulz M, Del Favero G, et al. Assessing Mixture Effects of Cereulide and Deoxynivalenol on Intestinal Barrier Integrity and Uptake in Differentiated Human Caco-2 Cells. Toxins (Basel). 2021;13(3):189. doi: 10.3390/toxins13030189 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Nollevaux G, Devillé C, El Moualij B, Zorzi W, Deloyer P, Schneider Y-J, et al. Development of a serum-free co-culture of human intestinal epithelium cell-lines (Caco-2/HT29-5M21). BMC Cell Biol. 2006;7:20. doi: 10.1186/1471-2121-7-20 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38.Bharadiya V, Rong Y, Zhang Z, Lin R, Guerrerio AL, Tse CM, et al. Type 1 diabetes human enteroid studies reveal major changes in the intestinal epithelial compartment. Sci Rep. 2024;14(1):11911. doi: 10.1038/s41598-024-62282-x [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 39.Hellinger E, Veszelka S, Tóth AE, Walter F, Kittel A, Bakk ML, et al. Comparison of brain capillary endothelial cell-based and epithelial (MDCK-MDR1, Caco-2, and VB-Caco-2) cell-based surrogate blood-brain barrier penetration models. Eur J Pharm Biopharm. 2012;82(2):340–51. doi: 10.1016/j.ejpb.2012.07.020 [DOI] [PubMed] [Google Scholar]
  • 40.Giusti S, Sbrana T, La Marca M, Di Patria V, Martinucci V, Tirella A, et al. A novel dual-flow bioreactor simulates increased fluorescein permeability in epithelial tissue barriers. Biotechnol J. 2014;9(9):1175–84. doi: 10.1002/biot.201400004 [DOI] [PubMed] [Google Scholar]
  • 41.Clothier R, Starzec G, Pradel L, Baxter V, Jones M, Cox H, et al. The prediction of human skin responses by using the combined in vitro fluorescein leakage/Alamar Blue (resazurin) assay. Altern Lab Anim. 2002;30(5):493–504. doi: 10.1177/026119290203000503 [DOI] [PubMed] [Google Scholar]
  • 42.Kelly P, Besa E, Zyambo K, Louis-Auguste J, Lees J, Banda T, et al. Correction: Endomicroscopic and Transcriptomic Analysis of Impaired Barrier Function and Malabsorption in Environmental Enteropathy. PLoS Negl Trop Dis. 2022;16(6):e0010505. doi: 10.1371/journal.pntd.0010505 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 43.Hodges P, Tembo M, Kelly P. Intestinal Biopsies for the Evaluation of Environmental Enteropathy and Environmental Enteric Dysfunction. J Infect Dis. 2021;224(12 Suppl 2):S856–63. doi: 10.1093/infdis/jiab372 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 44.Ma TY, Nguyen D, Bui V, Nguyen H, Hoa N. Ethanol modulation of intestinal epithelial tight junction barrier. Am J Physiol. 1999;276(4):G965-74. doi: 10.1152/ajpgi.1999.276.4.G965 [DOI] [PubMed] [Google Scholar]
  • 45.Panou DA, Pedersen SF, Kristensen M, Nielsen HM. Epithelium dynamics differ in time and space when exposed to the permeation enhancers penetramax and EGTA. A head-to-head mechanistic comparison. Front Drug Deliv. 2023;3:1221628. doi: 10.3389/fddev.2023.1221628 [DOI] [PMC free article] [PubMed] [Google Scholar]

Decision Letter 0

Mária Deli

29 Nov 2025

Dear Dr. Watson,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

Please submit your revised manuscript by Jan 13 2026 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.

  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.

  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols.

We look forward to receiving your revised manuscript.

Kind regards,

Mária A. Deli, M.D., Ph.D.

Academic Editor

PLOS ONE

Journal Requirements:

When submitting your revision, we need you to address these additional requirements.

1.Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at

https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and

https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

2. We note that the grant information you provided in the ‘Funding Information’ and ‘Financial Disclosure’ sections do not match.

When you resubmit, please ensure that you provide the correct grant numbers for the awards you received for your study in the ‘Funding Information’ section.

3. Thank you for stating the following financial disclosure:

“This article reports independent research supported by: the UK Medical Research Council

(MRC; grant number – MR/V012452/1); the National Institute for Health Research (NIHR) Imperial

Biomedical Research Centre (BRC); and the Imperial College London UKRI Impact Acceleration

Account supported by MRC (grant number – MR/X502959/1) and Rosetrees Trust. J.A.

acknowledges an Imperial College Research Fellowship. The views expressed in this publication

are those of the authors and not necessarily those of the NHS, the NIHR or the Department of

Health.”

Please state what role the funders took in the study.  If the funders had no role, please state: "The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript."

If this statement is not correct you must amend it as needed.

Please include this amended Role of Funder statement in your cover letter; we will change the online submission form on your behalf.

4. Please note that funding information should not appear in any section or other areas of your manuscript. We will only publish funding information present in the Funding Statement section of the online submission form. Please remove any funding-related text from the manuscript.

5. Please include captions for your Supporting Information files at the end of your manuscript, and update any in-text citations to match accordingly. Please see our Supporting Information guidelines for more information: http://journals.plos.org/plosone/s/supporting-information

6. If the reviewer comments include a recommendation to cite specific previously published works, please review and evaluate these publications to determine whether they are relevant and should be cited. There is no requirement to cite these works unless the editor has indicated otherwise.

Additional Editor Comments:

Two experts gave detailed opinions on the manuscript. They agreed that the work has merits but both of them suggested amendments before acceptance. Please clarify and/or correct all the methodological points including repeat numbers (biological parallels or number of experiments?), TEER and Papp calculations

[Note: HTML markup is below. Please do not edit.]

Reviewer's Responses to Questions

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

Reviewer #1: No

Reviewer #2: Partly

**********

2. Has the statistical analysis been performed appropriately and rigorously? -->?>

Reviewer #1: No

Reviewer #2: No

**********

3. Have the authors made all data underlying the findings in their manuscript fully available??>

The PLOS Data policy

Reviewer #1: Yes

Reviewer #2: Yes

**********

4. Is the manuscript presented in an intelligible fashion and written in standard English??>

Reviewer #1: Yes

Reviewer #2: Yes

**********

Reviewer #1: This manuscript of Elena Monfort-Sanchez and Aaron F. Watson focuses on an important topic.

The gut barrier plays a crucial role in maintaining tissue homeostasis. Disruption of this barrier (commonly called “leaky gut”) is linked to several diseases, hence proper in vitro models and methods are crucial which are comparable with in vivo conditions. Authors conclude that fluorescein and methylene blue can effectively replace FITC-dextran in fluorescent assays of epithelial barrier function. However, the execution of the experiments, the interpretation of the results, and the conclusions drawn from them raise concerns and are not sufficiently convincing. Addressing these issues is crucial for the acceptance of the manuscript.

Minor comments:

1. Transwell is a brand name therefore it is written with capital first letter. If authors do not refer for the brand it is called cell culture insert or culture inserts. Please correct it where needed.

2. When presenting the results, please try to use identical scale divisions on the axes of graphs wherever possible, as this will make it easier to compare the results.

Major comments:

1. What is the reason for the difference in the number of biological repeats between Figures 2 and 3? Were the TEER measurements and the permeability assays not performed in the same experiment? My other concern is that 2 biological repeats and 1 technical repeat are not sufficient to draw reliable and robust conclusions. I would recommend repeating the experiments or including more biological replicates.

2. How can the fluorescence intensity data of this manuscript be compared with permeability data in the literature? It is suggested to calculate the Papp values (expressed as cm/s) to ensure comparability (e.g. Youdim et al, PMID: 14643163)

3. The discussion section is very short and does not compare the obtained results with literature data at all. I consider this as one of the weak points of the paper. Furthermore, fluorescein has been widely used by many research groups to assess the tightness of in vitro epithelial cell layers (e.g. Hellinger et al, PMID: 22906709)

Comments related to methods:

1. The authors should add the exact liquid volume used in the apical and basal compartment of the inserts. It is also important that the liquid levels in the two compartments are approximately at the same height to ensure that the pressure of the liquid columns is balanced.

2. It is not clear how ethanol treatment was performed. “Ethanol solutions at concentrations of 0.5%, 1%, 2%, 3%, 4%, 5% and 6% (v/v) were prepared by diluting absolute (≈ 100%) ethanol in DI water.” Why not in DMEM? Were the cells treated with DI water? Please clarify this.

3. Please describe how were the treatment concentrations of ethanol and EGTA determined? Were cytotoxicity tests (eg. MTT) performed beforehand to ensure that these concentrations do not cause cell death, but only the opening of the cell layer? Ethanol permeabilizes the plasma membrane of the cells, please add some sentences related to this effect in the Discussion.

4. “Fluorescein (Sigma), FITC-dextran (Sigma) and methylene blue (Sigma) stock solutions (concentrations shown in Table 1) were prepared by dissolving powdered dyes in sterilized DI water and diluting in complete DMEM medium prior to their addition onto MDCK monolayers.” Please describe the stock solution concentrations of the fluorescence dyes and the final concentration of DI water in cell treatment solutions.

5. “Contrast agent concentrations were chosen to give approximately equal fluorescence intensities for each agent at each concentration level as different fluorescent agents have different quantum yields.”

Was it calculated or measured prior the experiment? What are these concentrations based on? Please specify. Also please write the wavelengths to measure the fluorescent intensities. Please check the name of the instrument. Is it not Clariostar?

6. I have many concerns related to the permeability experiment.

6.1. It is recommended to use a horizontal plate shaker to avoid unstirred water layer (e.g. Youdim et al, PMID: 14643163)

6.2. Medium should be phenolred free, since it can interfere with fluorescence signal.

6.3. Please justify why 100 μL was taken out at 1.5h timepoint.

Every time you remove 100 µL from the donor, you change the donor concentration (and total donor volume). If donor concentration changes substantially, sink conditions are violated and you must correct for the changing donor concentration when computing permeability. Please clarify how was the permeability corrected for this.

Please calculate Papp values to reflect permeability. Without calculating Papp (which normalizes for concentration, area, and time), it’s difficult to directly compare results across experiments or with other studies.

7. The given raw TEER values are negative numbers. TEER values reflecting resistance can be zero but not negative numbers. If these data were calculated some other way, please provide the formula. Were cell-free inserts used for background TEER and also permeability measurements?

Comments related to results

Figure 1 and Figure 2. In my opinion, the interpretation of TEER results is not correct.

The Y-axis shows Ω×cm2, however, this is incorrect, as TEER values cannot be negative numbers. If the values were recalculated, please indicate the method used. I recommend presenting the values as a percentage of the control instead. If the values reflect relative changes this should be clearly described.

3.3 section

“The most pronounced increases in fluorescence intensity are observed with 8 mM EGTA disruption (Fig. 3B,D,F), in agreement with the results presented in Fig. 1B,C.” Only at figure 1C.

How was the fluorescent intensity calculated? How can these numbers be compared to the literature?

Reviewer #2: In the manuscript entitled “Assessment of epithelial barrier integrity in cellular monolayer models using low cost, clinically approved fluorescent contrast agents” the Authors describe the fluorescent method to measure the epithelial monolayer permeability using two markers: fluorescein and methylene blue. The idea of the study is reasonable and the results are promising. However, I have found some weak points that should be corrected before the manuscript is accepted for the publication in PLOS One.

Specific issues

Divide the “Methods” chapter into subsections.

Specify, what kind of TEER device was used with the necessary descriptions of the experimental conditions. How the TEER values were calculated?

What was the reason to use one-way ANOVA (parametric test) for the analysis? Was this decision supported by a normality test? If such a test was not performed, a non-parametric test (Kruskal-Wallis) should be used. The Authors have to perform the normality test to support the use of ANOVA or another parametric test for statistical analysis.

The bar charts are widely used in most scientific papers, but this is inappropriate in terms of statistics. Instead, a dot plot, boxes with whiskers, or the points showing the mean values with the error bars for standard deviation should be used. Please change the graph type in all the figures due to my suggestions.

The Authors have to specify if the error bars on the figures present the standard deviations (SD) or standard errors of the mean (SEM). From a substantive point of view the error bars should present the SD. If the error bars show the SEM, then it should be changed to SD.

Two biological repeats is not enough to perform a reliable statistical test. Three biological repeats is the absolute minimum. The Authors have to perform one more biological repeat wherever there are two biological repeats.

Editorial issues (page numbering applies to the entire Manuscript Draft document):

Correct the reference number 19 and 23. Number 19: “ACS Paper” – this information is not enough. Number 23: the reference is split into two paragraphs, thus merge these two paragraphs into one.

Page 11, line 5: change “However, While” to “However, while” (probably a typo).

English language comments

Page 10, third line in “Introduction”: to avoid unnecessary repetitions, change “and active and passive immunity” to “as well as active and passive immunity”.

The manuscript has to be resubmitted with the corrections according to the above suggestions, including the additional experiments to provide three biological repeats, and reconsidered again for publication.

**********

what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review?  For information about this choice, including consent withdrawal, please see our Privacy Policy

Reviewer #1: No

Reviewer #2: No

**********

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

To ensure your figures meet our technical requirements, please review our figure guidelines: https://journals.plos.org/plosone/s/figures

You may also use PLOS’s free figure tool, NAAS, to help you prepare publication quality figures: https://journals.plos.org/plosone/s/figures#loc-tools-for-figure-preparation.

NAAS will assess whether your figures meet our technical requirements by comparing each figure against our figure specifications.

PLoS One. 2026 Oct 1;21(10):e0358746. doi: 10.1371/journal.pone.0358746.r002

Author response to Decision Letter 1


1 Aug 2026

We thank the reviewers for taking the time to assess our manuscript and for their insightful comments. Please see below our responses to all of the reviewers’ comments. In this document, reviewer comments are shown in bold, black font and our responses are shown in green text. In our revised manuscript, all changes are highlighted using tracked changes.

Minor Comments:

1. Transwell is a brand name therefore it is written with capital first letter. If authors do not refer for the brand it is called cell culture insert or culture inserts. Please correct it where needed.

We thank the reviewer for highlighting this point. We have corrected all instances of the term ‘transwell’ to instead refer to ‘cell culture inserts.’

2. When presenting the results, please try to use identical scale divisions on the axes of graphs wherever possible, as this will make it easier to compare the results.

As suggested by the reviewer, we have adjusted y-axis scales to show equal scale divisions on all TEER graphs. For fluorescence graphs, we have chosen to retain different scales across graphs. This is because different fluorescent dyes and experimental conditions generate quite different fluorescence intensity values. Thus, plotting all graphs on identical scales makes some trends very hard to discern. For that reason, we have retained differing scales for fluorescence graphs to ensure that all experimental trends are easily observable and interpretable.

Major Comments (Results):

1. What is the reason for the difference in the number of biological repeats between Figures 2 and 3? Were the TEER measurements and the permeability assays not performed in the same experiment? My other concern is that 2 biological repeats and 1 technical repeat are not sufficient to draw reliable and robust conclusions. I would recommend repeating the experiments or including more biological replicates.

We thank the reviewing for highlighting this point. This was a typographical error within the original manuscript. TEER measurements and the permeability assays were indeed performed in the same experiments, and we performed 2 biological repeats and 1 technical repeat in all cases. This has been corrected in the updated manuscript.

We acknowledge that 2 biological repeats with 1 technical repeat are not ideal. However, due to a number of logistical considerations, we are unable to perform further experiments on this project (lead author has completed PhD, funding has expired, and author providing biological laboratory access has relocated). As such, we stress that the intention of our paper is to present a preliminary evaluation of the use of different yet potentially analogous fluorescent tracers to FITC-dextran for their suitability in gut transport assays. The intended interpretation of the data is to support preliminary methodological validation and tracer selection rather than making strong inferential/translational/mechanical claims. Although we only have two biological repeats, our results still capture meaningful, statistically significant trends that are appropriate for early-stage tracer selection and assay development. Hence, while preliminary, we believe the current data is suitable for publication and will help to enable future, larger scale studies to further guide choice of tracers and experimental parameters.

We have edited the text in numerous sections of the manuscript to acknowledge these points and to frame our results as “exploratory” and/or “proof-of-concept.”

2. How can the fluorescence intensity data of this manuscript be compared with permeability data in the literature? It is suggested to calculate the Papp values (expressed as cm/s) to ensure comparability (e.g. Youdim et al, PMID: 14643163)

While we agree that measurements of apparent permeability (Papp) would enable more direct comparison of our results against existing literature, we did not perform Papp measurements here. In addition, we are not able to calculate this retrospectively as we only recorded fluorescence data at a single time point in most experiments (accurate Papp measurements require multiple time points).

Despite this, we do not believe this detracts from the claims of our paper, which is focused on presenting proof-of-concept data supporting use of different/multiple fluorescent tracers for barrier function assessment. In that scenario, demonstration of higher fluorescence intensities at greater degrees of barrier disruption and inverse correlation between fluorescence intensity and TEER (which we have added to the updated manuscript) serve the purpose of our experiments As such, while calculation of Papp values would enhance cross-study comparability, it would not alter the results and trends observed in our pilot study showing TEER reduction and increased tracer flux at higher levels of barrier degradation.

Importantly, it is also noteworthy the Papp values are also dependent on the epithelial barrier itself (i.e. tighter barriers will naturally exhibit lower Papp values for any given fluorophore). This means that while Papp calculation would improve cross-study comparability, it would still have limitations in this regard (due to differences in baseline barrier conditions in different studies).

As such, we believe our results demonstrating changes in both TEER and fluorescence intensity with increasing barrier disruption (and inverse correlation between fluorescence intensity and TEER change) are appropriate for this investigation. These results clearly demonstrate that all fluorescent dyes investigated in this study pass MDCK epithelial barriers to greater degrees as barriers degrade. Thus, our exploratory data serves as proof-of-concept and supports further investigation of the use of multiple fluorescent contrast agents for assessment of epithelial barrier integrity.

We have added text into the Discussion & Conclusions section to address these points.

3. The discussion section is very short and does not compare the obtained results with literature data at all. I consider this as one of the weak points of the paper. Furthermore, fluorescein has been widely used by many research groups to assess the tightness of in vitro epithelial cell layers (e.g. Hellinger et al, PMID: 22906709)

As requested by the reviewer, we have incorporated additional text into the Discussion & Conclusions section to provide a more detailed comparison of our results against existing literature. As part of this, we acknowledged prior use of fluorescein and have cited the reference highlighted by the reviewer.

Major Comments (Methods):

1. The authors should add the exact liquid volume used in the apical and basal compartment of the inserts. It is also important that the liquid levels in the two compartments are approximately at the same height to ensure that the pressure of the liquid columns is balanced.

We have added the exact liquid volumes that were used in the apical and basolateral domains of the insert-based cell culture system. We have also noted that we observed these volumes led to equivalent heights of liquids in the apical and basolateral compartments (see Methods Sections 2.2 and 2.4).

2. It is not clear how ethanol treatment was performed. „Ethanol solutions at concentrations of 0.5%, 1%, 2%, 3%, 4%, 5% and 6% (v/v) were prepared by diluting absolute (≈ 100%) ethanol in DI water.” Why not in DMEM? Were the cells treated with DI water? Please clarify this.

We have clarified how the ethanol treatment was performed; we diluted absolute ethanol down to 70% using DI water and then diluted 70% ethanol down to the respective ethanol concentrations using complete DMEM. Cells were therefore effectively exposed to both ethanol and DI water in DMEM in all ethanol experiments, but with DMEM making up at least 90% (v/v) of the solutions in all cases. This approach was used to minimise evaporation of ethanol during the solution preparation process (by using a working ethanol concentration of 70% rather than 100%). We have clarified this process in the Methods section (see section 2.3).

3. Please describe how were the treatment concentrations of ethanol and EGTA determined? Were cytotoxicity tests (eg. MTT) performed beforehand to ensure that these concentrations do not cause cell death, but only the opening of the cell layer? Ethanol permeabilizes the plasma membrane of the cells, please add some sentences related to this effect in the Discussion.

We did not perform cytotoxicity tests such as the MTT assay as part of this pilot study. Instead, we chose EGTA and ethanol concentrations based on previous published work that would be expected to provide disruption of tight junctions without causing significant cytotoxicity. We have added additional text into the Discussion section to outline how the specified ethanol and EGTA concentrations were chosen. We have also specifically acknowledged the lack of cytotoxicity assays as a limitation of our study.

4. “Fluorescein (Sigma), FITC-dextran (Sigma) and methylene blue (Sigma) stock solutions (concentrations shown in Table 1) were prepared by dissolving powdered dyes in sterilized DI water and diluting in complete DMEM medium prior to their addition onto MDCK monolayers.” Please describe the stock solution concentrations of the fluorescence dyes and the final concentration of DI water in cell treatment solutions.

We have clarified that the fluorescent agents were not dissolved in DI water but rather were directly dissolved in complete DMEM. This was a mistake in our original manuscript and we thank the reviewer for highlighting this. As such, no DI water was applied to the cells as part of the fluorescent dye solutions. The concentration of the stock solutions and all dilutions are shown in Table 1 (see Methods section, section 2.3). The concentrations shown in Table 1 are the final concentrations applied to cells (i.e. the concentrations in the apical chamber). We have added further text into section 2.3 to clarify these points.

5. “Contrast agent concentrations were chosen to give approximately equal fluorescence intensities for each agent at each concentration level as different fluorescent agents have different quantum yields.” Was it calculated or measured prior the experiment? What are these concentrations based on? Please specify. Also please write the wavelengths to measure the fluorescent intensities. Please check the name of the instrument. Is it not Clariostar?

We have provided further detail to clarify our calculations of fluorescent contrast agent concentrations in the methods section (see section 2.3). As part of this, we have noted that these values were calculated prior to experiments (and then used in all subsequent experiments).

The name of the instrument has been corrected to ‘CLARIOstar’ (see section 2.4) – we thank the reviewer for highlighting this error. We have also listed the excitation and emission wavelengths used for each fluorescent contrast agent (see section 2.4 and Table 2), as requested by the reviewer.

6.1. It is recommended to use a horizontal plate shaker to avoid unstirred water layer (e.g. Youdim et al, PMID: 14643163)

A horizontal plate shaker was not used in the experiments presented here. As discussed above, we are not currently able to perform further experiments to incorporate this. Importantly, however, the unstirred water layer primarily affects highly lipophilic and rapidly permeating compounds. The fluorescent compounds used in this study are all highly soluble in aqueous media. As such, the unstirred water layer would be expected to have only a minimal effect on the permeation of the fluorescent agents used here (and so incorporating a horizontal plate shaker would not significantly alter our findings). Furthermore, as use of a horizontal plate shaker would act to speed up diffusion of fluorescent contrast agent in the apical chamber, any changes induced would only be expected to cause greater permeation of fluorescent tracers across the epithelial barrier (by accelerating transport towards the barrier). Thus, our data provides a sound proof-of-concept for translocation of all fluorescent agents across epithelial barriers and, if anything, our results may underestimate the degree of translocation in each case.

We have included additional text in the Methods section (section 2.4) to explain these points.

6.2. Medium should be phenol red free, since it can interfere with fluorescence signal.

We agree with the reviewer that use of phenol red free medium would have been optimal for this study. However, only standard DMEM (containing phenol red) was available to us in this pilot study. Importantly, phenol red interference is most problematic for low-intensity signals. Fluorescein and FITC-dextran exhibit very strong fluorescence at 490/520 nm excitation/emission, and methylene blue also has a strong absorbance that is distinct from phenol red background. Our results also show that fluorescence signals observed from our chosen fluorescent contrast agents are significantly higher than any background observed in the absence of exogenous fluorescent dyes – i.e. fluorescence signals from fluorescein, FITC-dextran and methylene blue are all high compared to any phenol red background (compare fluorescence intensities for 6.25%, 25% and 100% fluorescent dye concentrations against those for 0% concentration in Figure 3). This demonstrates that phenol red background signals did not detrimentally affect our results. We have added a sentence into the Results section (section 3.3) to make note of these points.

6.3. Please justify why 100 μL was taken out at 1.5h timepoint.

Every time you remove 100 µL from the donor, you change the donor concentration (and total donor volume). If donor concentration changes substantially, sink conditions are violated and you must correct for the changing donor concentration when computing permeability. Please clarify how was the permeability corrected for this. Please calculate Papp values to reflect permeability. Without calculating Papp (which normalizes for concentration, area, and time), it’s difficult to directly compare results across experiments or with other studies.

Samples for fluorescence measurement were only collected at 3 hours post-addition of fluorescent contrast agents. Baseline samples were collected prior to addition of contrast agent from both the apical and basolateral chambers, with the media topped back up by the same volume immediately after sample collection. Following the addition of fluorescent agents, 100 µl samples were collected from both chambers at a single time point (3 hours) after dye addition. Experiments were then terminated after sampling. As a result, it was not necessary to perform any correction to account for removal of fluorescent agent from the cell culture system. We have edited text in the Methods section to clarify our experimental protocol (see section 2.4) where we have also noted that TEER measurements were made at three time points (baseline, 1.5 hours and 3 hours). We thank the reviewer for highlighting this point, which was somewhat ambiguous in our original manuscript.

We also note again that the goal of our study was to show preliminary data demonstrating that fluorescein and methylene blue can be used to assess barrier function in an analogous manner to FITC-dextran. While we agree that measurements of apparent permeability can be useful in allowing comparison across studies, showing an inverse relationship between TEER and fluorescence is a widely accepted proxy in the literature and is appropriate to demonstrate proof-of-concept for fluorescent detection of barrier degradation. As discussed above, we have added text into the Discussion & Conclusions section to explain these points.

7. The given raw TEER values are negative numbers. TEER values reflecting resistance can be zero but not negative numbers. If these data were calculated some other way, please provide the formula. Were cell-free inserts used for background TEER and also permeability measurements?

figure 1 and Figure 2. In my opinion, the interpretation of TEER results is not correct. The Y-axis shows Ω×cm2, however, this is incorrect, as TEER values cannot be negative numbers. If the values were recalculated, please indicate the method used. I recommend presenting the values as a percentage of the control instead. If the values reflect relative changes this should be clearly described.

We thank the reviewer for highlighting this inconsistency. All TEER values presented in Figures 1 and 2 represent changes in TEER relative to baseline measurements (i.e. ΔTEER) and not absolute TEER values. As such, values are generally negative as we observed decreases in TEER following addition of disruptors EGTA and ethanol (although we note that some positive values are observed under control conditions where monolayers continued to grow between baseline and measurement/sampling time points). We have now added text into the Methods section to clearly describe how TEER values were calculated (both absolute values and ΔTEER values) – see sections 2.2 and 2.4, and equations (1) and (2).

3.3section

“The most pronounced increases in fluorescence intensity are observed with 8 mM EGTA disruption (Fig. 3B,D,F), in agreement with the results presented in Fig. 1B,C.” Only at figure1C. How was the fluorescent intensity calculated? How can these numbers be compared to the literature?

In this section (3.3), we have added text to the paragraph highlighted by the reviewer to more clearly explain the reference to Figure 1B,C (where we in fact compare against results for both ethanol (Fig. 1B) and EGTA disruption (Fig. 1C).

To measure fluorescence intensity, a 96-well plate containing the 100 µl samples was placed in a CLARIOstar fluorescence plate reader at the set excitation and emission wavelengths of the tracer (see Table 2, section 2.4) and the resulting fluorescence intensity values were plotted. Importantly, as discussed above, these numbers are not to be cross-compared with literature but rather to show internal consistency that higher fluorescence intensity is observed at greater levels of barrier disruption and that higher fluorescence intensity correlates with greater decreases in TEER.

We have amended the Methods text in section 2.4 to more clearly explain how fluorescence measurements were performed.

Divide the “Methods” chapter into subsections.

We have divided the Methods section into subsections as requested by the reviewer.

Specify, what kind of TEER device was used with the necessary descriptions of the experimental conditions. How were the TEER values calculated?

The TEER device used was the EVOM3 Manual by World Precision Instruments. To measure TEER values, we first measured absolute TEER values on both the samples (i.e. cell culture inserts with MDCK monolayers) and blank inserts containing cell culture medium only. The monolayer TEER value was then calculated by subtracting the blank value from the sample value (see equation (1), section 2.2). To assess changes in response to ethanol or EGTA disruption, we then calculated changes in TEER relative to baseline (see equation (2), section 2.4). The resulting ΔTEER values are presented in Figures 1 and 2.

We have included additional text in sections 2.2 and 2.4 of the Methods to further explain these points.

What was the reason to use one-way ANOVA (parametric test) for the analysis? Was this decision supported by a normality test? If such a test was not performed, a non-parametric test (Kruskal-Wallis) should be used. The Authors have to perform the normality test to support the use of ANOVA or another parametric test for statistical analysis.

We thank the reviewer for highlighting this point and we clarify that a two-way ANOVA was used for the analysis. The Shapiro-Wilk test was used to assess normality and all conditions passed normality but one (2 mM EGTA, FITC-Dextran). Given this result, and given the preliminary nature of our data (i.e. two measurements per condition), we chose to use two-way ANOVA for all further statistical analysis. While the size of our dataset means that all results should be considered exploratory at this stage (regardless of significance level), use of two-way ANOVA was nonetheless informative in guiding interpretation of results and was appropriate to the dataset under investigation (due to the size of the dataset and the observation of normality in all but one condition). This has been clarified in both the methods section 2.4 and the results section 3.3.

The bar charts are widely used in most scientific papers, but this is inappropriate in terms of statistics. Instead, a dot plot, boxes with whiskers, or the points showing the mean values with the error bars for standard deviation should be used. Please change the graph type in all the figures due to my suggestions.

Due to our relatively low number of experimental repeats, box plots are not informative in this case. Instead, we have retained bar charts but have also shown the individual data points alongside the bars in the results section 3.2 and 3,3 (figures 2 and 3). This provides all experimental data and so we hope that this suitably addresses the point raised by the reviewer.

The Authors have to specify if the error bars on the figures present the standard deviations (SD) or standard errors of the mean (SEM). From a substantive point of view the error bars should present the SD. If the error bars show the SEM, then it should be changed to SD.

We have now specified that the errors bars shown on the figures represent the standard deviation (see the figure legends of all figures). We thank the reviewer for highlighting this oversight.

Two biological repeats is not enough to perform a reliable statistical test. Three biological repeats is the absolute minimum. The Authors have to perform one more biological repeat wherever there are two biological repeats.

We agree with the reviewer that two biological repeats do not provide robust statistical data and our results should therefore be viewed as preliminary/exploratory. As discussed above, we are sadly unable to perform further experiments at present. Instead, we have addressed this point in a number of ways.

1. We have reframed the discussion within our manuscript in numerous locations to more clearly present our work as exploratory and to frame our conclusions as preliminary / proof-of-concept. We believe our data has value and is appropriate for publication when pitched at this level.

2. We have shown individual data points on Figures 1 and 2 to better visualise variability within our experiments.

3. We have performed further analyses to assess correlations between our fluorescence and TEER measurements. As for the standalone fluorescence plots, these analyses show statistically significant results, demonstrating correlations between fluorescence intensities and TEER values (see section 3.4, Figure 4 and Supplementary Figures). While these results should still be viewed as exploratory due to the low number of experimental repeats, this nonetheless further demonstrates that we have observed clear effects in this study (i.e. with effect sizes high enough to generate statistical significance even with a low number of measurements).

Together, we believe this addresses the reviewer’s point and justifies publication of our preliminary data at this stage. Importantly, this will help to support and enable further investigation of the use of multiple fluorescent agents for epithelial barrier assessment, which in turn will further validate our proof-of-concept results.

Editorial issues (page numbering applies to the entire Manuscript Draft document):

Correct the reference number 19 and 23. Number 19: “ACS Paper” – this information is not enough. Number 23: the reference is split into two paragraphs, thus merge these two paragraphs into one. Page 11, line 5: change “However, While” to “However, while” (probably typo). English language comments Page 10, third line in “Introduction”: to avoid unnecessary repetitions, change “and active and passive immunity” to “as well as active and passive immunity”.

We thank the reviewer for highlighting these points. We have corrected the editorial issues as requested.

Attachment

Submitted filename: Itemised Response to Reviewers PLOS One.docx

pone.0358746.s004.docx (34.9KB, docx)

Decision Letter 1

Mária Deli

7 Sep 2026

Assessment of epithelial barrier integrity in cellular monolayer models using low cost, clinically approved fluorescent contrast agents

PONE-D-25-52156R1

Dear Dr. Watson,

We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.

Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.

An invoice will be generated when your article is formally accepted. Please note, if your institution has a publishing partnership with PLOS and your article meets the relevant criteria, all or part of your publication costs will be covered. Please make sure your user information is up-to-date by logging into Editorial Manager at Editorial Manager® and clicking the ‘Update My Information' link at the top of the page. For questions related to billing, please contact billing support.

If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

Kind regards,

Mária A. Deli, M.D., Ph.D.

Academic Editor

PLOS One

Additional Editor Comments (optional):

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

Reviewer #1: All comments have been addressed

Reviewer #2: All comments have been addressed

**********

2. Is the manuscript technically sound, and do the data support the conclusions??>

Reviewer #1: Yes

Reviewer #2: Yes

**********

3. Has the statistical analysis been performed appropriately and rigorously? -->?>

Reviewer #1: Yes

Reviewer #2: Yes

**********

4. Have the authors made all data underlying the findings in their manuscript fully available??>

The PLOS Data policy

Reviewer #1: Yes

Reviewer #2: Yes

**********

5. Is the manuscript presented in an intelligible fashion and written in standard English??>

Reviewer #1: Yes

Reviewer #2: Yes

**********

Reviewer #1: (No Response)

Reviewer #2: The Authors have accurately addressed my suggestions. The manuscript no. PONE-D-25-52156R1 is ready for publication in PLOS One.

**********

what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review?  For information about this choice, including consent withdrawal, please see our Privacy Policy

Reviewer #1: No

Reviewer #2: Yes:  Radoslaw Bednarek

**********

Acceptance letter

Mária Deli

PONE-D-25-52156R1

PLOS One

Dear Dr. Watson,

I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS One. Congratulations! Your manuscript is now being handed over to our production team.

At this stage, our production department will prepare your paper for publication. This includes ensuring the following:

* All references, tables, and figures are properly cited

* All relevant supporting information is included in the manuscript submission,

* There are no issues that prevent the paper from being properly typeset

You will receive further instructions from the production team, including instructions on how to review your proof when it is ready. Please keep in mind that we are working through a large volume of accepted articles, so please give us a few days to review your paper and let you know the next and final steps.

Lastly, if your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

You will receive an invoice from PLOS for your publication fee after your manuscript has reached the completed accept phase. If you receive an email requesting payment before acceptance or for any other service, this may be a phishing scheme. Learn how to identify phishing emails and protect your accounts at https://explore.plos.org/phishing.

If we can help with anything else, please email us at customercare@plos.org.

Thank you for submitting your work to PLOS One and supporting open access.

Kind regards,

PLOS One Editorial Office Staff

on behalf of

Prof. Mária A. Deli

Academic Editor

PLOS One

Associated Data

    This section collects any data citations, data availability statements, or supplementary materials included in this article.

    Supplementary Materials

    S1 File. Raw Data for Figures.

    Raw data made available for the figures within this manuscript.

    (XLSX)

    pone.0358746.s001.xlsx (24.5KB, xlsx)
    S2 File. Supplementary Figures.

    Supplementary file containing supplementary figures S1-S3.

    (DOCX)

    pone.0358746.s002.docx (3.2MB, docx)
    Attachment

    Submitted filename: Itemised Response to Reviewers PLOS One.docx

    pone.0358746.s004.docx (34.9KB, docx)

    Data Availability Statement

    All relevant data are within the manuscript and its Supporting Information files.