Improved visualization of malpositioned middle ear prostheses using photon-counting detector CT: a cadaver study

Maryam Sadeghian 1,2, John C Benson 3, Paul J Farnsworth 4, John I Lane 5, Kevin J Treb 6, Karl R Khandalavala 7, Eric E Babajanian 8, Michael R Bruesewitz 9, Joseph R Swicklik 10, Fang-Shu Ou 11, Cynthia H McCollough 12,13, Matthew L Carlson 14, Shuai Leng 15,16,✉
PMCID: PMC13633848  PMID: 42829685

Abstract

Background

Evaluation of middle ear prostheses requires high-resolution imaging to assess position and structural integrity.

Purpose

To assess the diagnostic performance of photon-counting-detector (PCD)-CT for middle ear prosthesis imaging in cadaveric specimens.

Materials and Methods

Nine different middle ear prostheses were implanted in 5 fresh-frozen whole-head specimens, both in correct anatomic position and displaced to mimic malposition. Specimens were scanned on PCD-CT (NAEOTOM Alpha, Siemens) and energy-integrating-detector (EID)-CT (SOMATOM Force, Siemens), reconstructed at 0.2 mm and 0.4 mm, respectively. Eighteen cases (9 prostheses, 2 positions) per scanner were evaluated by 2 neuroradiologists (36 readings), blinded to scanner and position, who scored prosthesis position, reader confidence, and overall image quality. Agreement with the reference standard and between readers was compared.

Results

PCD-CT showed higher diagnostic confidence compared with EID-CT, while accuracy and inter-reader agreement did not demonstrate significant differences between scanners. Correct readings, PCD-CT versus EID-CT: lateral contact, 31/36 (86.1%; 95% CI, 58.3-94.4) vs 29/36 (80.6%; 95% CI, 55.6-91.7), P = .500; medial contact, 30/36 (83.3%; 95% CI, 55.6-94.4) vs 26/36 (72.2%; 95% CI, 50.0-83.3), P = .250. PCD-CT identified all cases of complete contact loss (score 3), while EID-CT missed 1 reading per contact. Inter-reader agreement may be higher with PCD-CT vs EID-CT (lateral: κw: 0.804 [95% CI, 0.578-0.941] vs 0.571 [95% CI, 0.302-0.778], P = .250; medial: 0.684 [95% CI, 0.191-0.934] vs 0.481 [95% CI, 0.024-0.835], P = .250). Reader confidence was significantly higher with PCD-CT; mean paired differences for Readers 1 and 2 were 9.4 (95% CI, 1.8-17.1; Holm-adjusted P = .019) and 23.1 points (95% CI, 13.4-32.7; P < .001) for lateral contact, and 16.4 (95% CI, 10.0-22.8; P < .001) and 17.8 points (95% CI, 6.8-28.8; P = .007) for medial contact. Readers preferred PCD-CT image quality in 16/18 (88.9%; 95% CI, 65.3-98.6) and 17/18 cases (94.4%; 95% CI, 72.7-99.9), respectively.

Conclusion

PCD-CT was associated with significantly higher reader confidence and preferred image quality, without significant differences in diagnostic accuracy or inter-reader agreement. These findings suggest that PCD-CT is a promising modality for postoperative middle ear evaluations and warrants further investigations.

Keywords: photon-counting detector-CT, energy-integrating detector-CT, middle ear prostheses, temporal bone, conductive hearing loss


Abbreviations 

PCD = Photon-counting-detector; EID = Energy-integrating-detector; FOV = Field-of-view; CTDIvol = Volume CT dose index; PORP = Partial ossicular replacement prosthesis; TORP = Total ossicular replacement prosthesis; κw = Linearly weighted kappa

Summary 

In a small sample of cadaveric specimens, photon-counting detector CT compared favorably to energy-integrating detector CT in detection of malpositioned middle ear prostheses, diagnostic agreement, and reader confidence.

Key Results 

  • Assessment of prostheses contacts in 9 different prostheses implanted in 5 cadaveric heads demonstrated agreement with the reference standard: 86.1% for PCD-CT vs 80.6% for EID-CT for lateral contact (P = .500), and 83.3% vs 72.2% for medial contact (P = .250).

  • Inter-reader agreement was similar between PCD-CT and EID-CT (P = .250).

  • Diagnostic confidence was higher with PCD-CT (all P ≤ .019).

  • PCD-CT image quality was preferred in 88.9% (95% CI, 65.3%-98.6%) of cases.

Introduction

Ossicular chain reconstruction with metallic-based prostheses is commonly used to improve conductive hearing loss among patients with ossicular chain erosion, discontinuity, or fixation caused by chronic middle ear disease, otosclerosis, trauma, or congenital abnormalities.1,2 However, prosthesis malposition or displacement is a well-recognized postoperative complication.3,4 A review of 80 published articles on titanium partial and total ossicular replacement prostheses reported a mean extrusion or dislocation rate of 5.2% (range 0%-35%).5 Moreover, within the revision stapes surgery literature, prosthesis dislocation has been reported as the most frequent intraoperative finding, accounting for approximately 56% of revision cases in a recent multicenter study.4 Postoperative imaging plays a critical role in patient monitoring, especially in cases of complications or recurrent/persistent conductive hearing loss.6,7 In clinical practice, CT imaging is commonly performed to evaluate prosthesis position relative to remnant ossicles, tympanic membrane, and the inner ear.6–8

A major challenge of imaging these prostheses is that the small size of the structures requires high spatial resolution. Conventional energy-integrating detector (EID)-CT faces limitations when imaging small prostheses within the complex anatomy of the temporal bone, particularly regarding contact with adjacent ossicles or the tympanic membrane.9,10 Lower spatial resolution and partial-volume effects restrict its ability to identify fine structures, creating diagnostic challenges that can affect clinical decision-making.6,8 Over the past decade, photon-counting detector (PCD)-CT has emerged as a promising technology with several key advantages over EID-CT, including higher spatial resolution, lower noise, better contrast-to-noise ratio (CNR), increased dose efficiency, and reduced blooming artifact and beam-hardening.11–18

Initial studies have shown that PCD-CT improves visualization of the temporal bone and ear anatomy, offering superior spatial resolution and lower radiation dose compared to conventional EID-CT scanners.19–23 While these studies primarily focused on anatomical visualization, image quality, and dose reduction, there are limited data specifically regarding PCD-CT imaging of ossicular prostheses. Although case examples and a recent in vivo patient comparison have shown improved visualization of prostheses,20,23–25 these have characterized image quality and spatial resolution rather than diagnostic accuracy against a known reference standard. A systematic evaluation of diagnostic performance for prosthesis position and complications has been lacking because patient studies cannot independently verify true prosthesis position or perform identical repeat scans across different CT systems, while phantom studies could not replicate real-world clinical challenges such as prosthesis interaction with surrounding structures.13,25,26

To address this gap, a model replicating real clinical practice with a reliable reference standard is needed. Fresh-frozen whole human head cadavers address these methodological limitations and serve as a valuable model for accurate assessment of middle ear prostheses. Cadaveric models offer both anatomical realism and the rigorous reference standard verification necessary for diagnostic accuracy studies.21,26 The aim of this study was to evaluate the visualization and diagnostic performance of PCD-CT compared to EID-CT for ossicular and stapes prostheses in fresh-frozen human cadaveric heads.

Materials and methods

Specimen preparation

This study was performed with the approval of our institution’s biospecimen committee. Five fresh-frozen whole human head specimens were obtained through the institution’s anatomy laboratory, selected based on availability and not screened for otologic pathology before inclusion. No gross middle ear abnormalities were identified during surgical exploration or subsequent imaging. Five stapes prostheses, 2 total ossicular replacement prostheses (TORP), and 2 partial ossicular prostheses (PORP) were evaluated. The prostheses were made of titanium, platinum, hydroxyapatite, and composite combinations (detailed in Table 1).

Table 1.

Characteristics of implanted prostheses.

Prosthesis type Name Length Reference number Material Type Shaft diameter/head size Picture
1 Stapes Robinson Bucket Handle 3.75 mm 420-375 Titanium Standard 0.4 mm For image description, please refer to the figure legend and surrounding text
2 Stapes Lippy Bucket Handle 4.50 mm 421-450 Titanium Notched 0.4 mm For image description, please refer to the figure legend and surrounding text
3 Stapes Eclipse Piston 4.0 mm 467-400 Nitinol/Flpl Piston 0.5 mm For image description, please refer to the figure legend and surrounding text
4 Stapes MEITI Piston 3.75 mm 418-375 Platinum/Titanium Piston 0.5 mm For image description, please refer to the figure legend and surrounding text
5 Stapes Eclipse Piston 4.50 mm 467-450 Nitinol/Flpl Piston 0.5 mm For image description, please refer to the figure legend and surrounding text
6 TORP Centered ALTO total 2.50-7.00 mm (adjustable) 604 Titanium – 3.00 mm For image description, please refer to the figure legend and surrounding text
7 TORP Wildcat Pro 3.00 mm 709-300 HA/Titanium Centered 1.00 × 1.75 mm (distal end) For image description, please refer to the figure legend and surrounding text
8 PORP Precise centered partial 3.50 mm 765-350 Titanium Centered
  • 0.2 mm (Shaft)

  • 1.45 mm (Cup)

For image description, please refer to the figure legend and surrounding text
9 PORP Centered ALTO partial 1.50 to 5.60 mm (Adjustable) 655 Titanium –
  • 3.00 mm (Head)

  • 1.45 mm (Cup)

For image description, please refer to the figure legend and surrounding text

Abbreviations: Flpl = fluoroplastic; HA = hydroxyapatite; PORP = partial ossicular replacement prosthesis; TORP = total ossicular replacement prosthesis.

Each prosthesis was first implanted in its correct anatomical position by fellowship-trained otologic surgeons and then intentionally displaced to mimic common types of prosthesis malposition, thereby providing an image series with correct and displaced conditions for assessment.

CT image acquisition and reconstruction

All specimens were scanned on a PCD-CT scanner (NAEOTOM Alpha, Siemens). For comparison, the same specimens were also scanned on an EID-CT scanner from the same manufacturer (SOMATOM Force, Siemens). Clinical temporal bone protocols were used on both scanners with matched tube potential (120 kV), volume CT dose index (CTDIvol), and field-of-view (FOV, 80 mm) (parameters summarized in Table 2). Images were reconstructed at the thinnest available slice thickness for each scanner: 0.2 mm for PCD-CT and 0.4 mm for EID-CT, using clinical kernels (Hr84 for PCD-CT and Ur77 for EID-CT).

Table 2.

CT acquisition and reconstruction parameters.

PCD-CT a EID-CTb
Scanner model PCD-CT (NAEOTOM Alpha, Siemens) Third-generation dual source (SOMATOM Force, Siemens)
Tube potential (kV) 120 120
CTDIvol (mGy) Matched between 2 scanners for each specimen. Varies among specimens (automatically adjusted by scanner based on size), range: 48.8-64.2.
Scan mode Spiral Spiral
Pitch 0.5 0.35
Rotation time (s) 1 1
Collimation (mm) 120 × 0.2 64 × 0.6
FOV (mm) 80 80
Slice thickness (mm) 0.2 0.4
Increment (mm) 0.1 0.3
Reconstruction mode/strength QIR/3 ADMIRE/3
Reconstruction kernel Hr84 Ur77
a

Photon-counting-detector.

b

Energy-integrating-detector.

Image evaluation

Eighteen image series (9 prostheses × 2 positions) were independently evaluated by 2 board-certified, fellowship-trained neuroradiologists, with 5 and 7 years of post-fellowship experience. Each scanner had 9 image series with prostheses in the correct position and 9 with prostheses displaced. To minimize bias, readers were blinded to scanner type and prosthesis positioning (correct vs displaced). For each case, PCD-CT and EID-CT image series were presented side-by-side in a paired-comparison format. The left-right position of each scanner type was randomized to prevent readers from identifying which image corresponded to which scanner. The presentation order of specimens was also randomized. Readers independently evaluated each image series for each specimen using a standardized assessment protocol. For each image, prosthesis position was scored for lateral and medial contact using a 3-point ordinal scale: 1 = good contact, 2 = tangential contact, 3 = no contact. For each contact assessment, readers provided their diagnostic confidence using a continuous scale from 0 to 100, where 0 indicated no confidence and 100 indicated complete confidence in their contact classification. Finally, readers evaluated overall image quality in the paired comparisons using a 5-point Likert scale (1 = EID significantly better, 2 = EID slightly better, 3 = no difference between scanners, 4 = PCD slightly better, 5 = PCD significantly better). All image evaluations were performed on a diagnostic workstation with standard bone window settings (width/level: 4000/1000 HU) and controlled ambient lighting conditions to ensure consistency.

Reference standard

The reference standard for contact classification was established by an experienced neuroradiologist (35 years of post-fellowship experience) and an otologic surgeon (14 years of experience, including fellowship training) using intraoperative microscopic photographs obtained during prosthesis placement. PCD-CT images were reviewed in conjunction to confirm that no displacement of the prosthesis had occurred between surgery and scanning. Each case was assessed by the reference-standard panel by consensus, and each contact was classified using the same 3-point scale.

Statistical analysis

The primary objectives were to assess the accuracy of contact assessment and reader confidence, with secondary analyses evaluating overall image quality. Exact agreement was defined as equality between the reader-assigned score and the reference standard score. Agreement between individual reader scores and the reference standard was quantified using linearly weighted kappa (κw), reported per reader and pooled across both readers. Inter-reader agreement for contact scoring was assessed using linearly weighted kappa (κw). Kappa values were interpreted according to the criteria of Landis and Koch27: ≤0.20, slight; 0.21 to <0.41, fair; 0.41 to <0.61, moderate; 0.61 to <0.81, substantial; and ≥0.81, almost perfect agreement. Ninety-five percent confidence levels (CIs) for κw estimations and pooled agreement proportions were obtained using bias-corrected and accelerated cluster bootstrap resampling with 10 000 iterations, with the prosthesis as the clustering unit. Reader-specific proportions were reported with Clopper-Pearson exact 95% CIs for descriptive purposes. Scanner differences in reader-specific agreement were assessed using the exact McNemar test, while pooled agreement and inter-reader agreement were tested with exact prosthesis-level permutation tests. Reader diagnostic confidence was compared between scanner types for each reader using a 2-sided paired t-test, with 95% CIs reported for paired mean differences. Wilcoxon signed-rank tests were performed as sensitivity analyses. Overall image quality ratings were compared with neutral score 3 using Wilcoxon signed-rank tests. The Holm method was applied to adjust for multiple comparisons within the accuracy and confidence analyses. Reader-specific McNemar, paired t-test, and Wilcoxon analyses preserved within-series pairing but did not additionally account for prosthesis or specimen level clustering and were therefore considered exploratory. A P value <.05 was considered statistically significant. All analyses were performed using Python (version 3.12).

Results

PCD-CT was preferred for prosthesis visualization and provided significantly higher diagnostic confidence than EID-CT for both readers (Figures 1-4). Radiation dose was matched between the 2 scanners for each specimen, with CTDIvol ranging from 48.8 to 64.2 mGy across specimens depending on specimen size (Table 2). Figure 1 represents a comparative visualization of PCD-CT versus EID-CT images from a sample stapes prosthesis. Figures 2 and 3 compare PCD-CT and EID-CT for assessing stapes prosthesis positioning. As shown in Figure 2, in anatomically correct placement, PCD-CT provided better delineation of the stapes hook embracing the lenticular process of incus. In the dislocation case shown in Figure 3, both modalities demonstrated loss of contact, and both readers assigned a score of 3 for lateral contact. Figure 4 represents a case where the reference standard categorized medial contact as good contact (score 1).

Figure 1.

Four CT images compare visualization of a stapes prosthesis using PCD-CT (A, C) and EID-CT (B, D) in axial-oblique and sagittal-oblique planes. The prosthesis is lateralized outside the oval window and displaced posteriorly toward the pyramidal eminence, demonstrating loss of lateral contact.

Comparison of prosthesis visualization between PCD-CT (photon-counting-detector CT) and EID-CT (energy-integrating-detector CT) imaging. Representative images of the prosthesis acquired using PCD-CT (A, C) and EID-CT (B, D), shown in axial-oblique planes in the first row and sagittal-obliques in the second row. The images show a 4.0-mm Eclipse Piston stapes prosthesis (solid arrows). The prosthesis is lateralized outside of the oval window (dashed arrow) and displaced posteriorly toward the pyramidal eminence. The reference standard classified lateral contact as score 3 (contact loss), and both readers assigned a contact score of 3 on both PCD-CT and EID-CT. Reader confidence was 100% for PCD-CT and 90% for EID-CT for reader 1 and 100% for PCD-CT and 60% for EID-CT for reader 2.

Figure 2.

Four CT images compare visualization of a stapes prosthesis using PCD-CT (A, C) and EID-CT (B, D) in axial-oblique and coronal-oblique planes. The prosthesis hook is positioned around the long process of the incus, demonstrating good lateral contact.

Comparison of PCD-CT (A, C) and EID-CT (B, D) in the assessment of stapes prosthesis positioning. Reformatted axial-oblique (A, B) and coronal-oblique (C, D) images show a 4.5-mm Eclipse Piston stapes prosthesis (solid straight arrows). Curved arrows indicate the prosthesis hook around the long process of the incus, which is represented by dashed arrows. The reference standard classified the lateral contact as good contact (score 1). Reader confidence was 100% for PCD-CT and 80% for EID-CT for reader 1 and 80% for PCD-CT and 40% for EID-CT for reader 2.

Figure 3.

Four coronal-oblique CT images compare visualization of a dislocated stapes prosthesis using PCD-CT (A, C) and EID-CT (B, D). The prosthesis hook is separated from the long process of the incus, demonstrating loss of lateral contact.

Visualization of dislocated stapes prosthesis shown in Figure 2. Coronal-oblique images from PCD-CT (A, C) and EID-CT (B, D) show the same 4.5-mm Eclipse Piston stapes prosthesis (marked with solid arrows), with no contact between the prosthesis hook and the long process of the incus (shown by dashed arrows). The reference standard classified lateral contact as contact loss (score 3), and both readers assigned a score of 3 on both PCD-CT and EID-CT. Reader confidence was 100% for PCD-CT for both readers, compared to 80% for reader 1 and 70% for reader 2 on EID-CT.

Figure 4.

Four CT images compare medial contact of a partial ossicular replacement prosthesis using PCD-CT (A, C) and EID-CT (B, D) in axial and coronal planes. The prosthesis demonstrates good medial contact on PCD-CT, whereas the contact appears tangential on EID-CT.

Comparison of medial prosthesis contact on PCD-CT and EID-CT. The first row shows reformatted axial images from PCD-CT (A) and EID-CT (B), and the second row shows coronal images for PCD-CT (C) and EID-CT (D). The images represent a 3.5-mm precise centered partial ossicular replacement prosthesis (PORP; solid arrows). Dashed arrows indicate the medial contact. The reference standard classified medial contact as good contact (score 1). Both readers assigned scores of 1 on PCD-CT but considered the contact tangential and assigned a score of 2 on EID-CT. Reader confidence was 75% for PCD-CT and 50% for EID-CT for reader 1, and 80% for PCD-CT and 50% for EID-CT for reader 2.

In lateral contact assessment, PCD-CT achieved agreement with the reference standard in 31 out of 36 readings (86.1%; 95% CI, 58.3-94.4; κw = 0.830 [95% CI, 0.496-0.964]), while EID-CT achieved agreement in 29 out of 36 readings (80.6%; 95% CI, 55.6-91.7; κw = 0.714 [95% CI, 0.400-0.861]). The difference in pooled agreement was not statistically significant (P = .500). For medial contact assessment, PCD-CT achieved agreement with the reference standard in 30 out of 36 readings (83.3%; 95% CI, 55.6-94.4; κw = 0.761 [95% CI, 0.335-0.964]), whereas EID-CT achieved it in 26 out of 36 readings (72.2%; 95% CI, 50.0-83.3; κw = 0.605 [95% CI, 0.308-0.855]). The difference was not statistically significant (P = .250).

Individual reader accuracy was numerically higher with PCD-CT for most contact assessments, though no difference reached statistical significance. For lateral contact assessment, Reader 1 achieved 94.4% accuracy (17/18; 95% CI, 72.7-99.9) with both modalities (κw = 0.929 [95% CI, 0.581-1.000 for PCD-CT, and 95% CI, 0.576-1.000 for EID-CT]). No discordant pairs were observed, so the McNemar comparison was not informative. Reader 2’s accuracy was higher with PCD-CT (14/18; 77.8%; 95% CI, 52.4-93.6; κw = 0.741 [95% CI, 0.400-0.933]) than with EID-CT (12/18; 66.7%; 95% CI, 41.0%-86.7%; κw = 0.500 [95% CI, 0.100-0.767]; McNemar: unadjusted P = .500; Holm-adjusted P = 1.000). For medial contact assessment, Reader 1’s accuracy increased from 66.7% (12/18; 95% CI, 41.0-86.7) with EID-CT (κw = 0.518 [95% CI, 0.145-0.902]) to 77.8% (14/18; 95% CI, 52.4-93.6) with PCD-CT (κw = 0.690 [95% CI, 0.169-1.000]; McNemar: unadjusted P = .500; Holm-adjusted P = 1.000), and Reader 2’s accuracy improved from 77.8% (14/18; 95% CI, 52.4-93.6; κw = 0.690 [95% CI, 0.368-0.929]) to 88.9% (16/18; 95% CI, 65.3-98.6; κw = 0.836 [95% CI, 0.407-1.000]; McNemar: unadjusted P = .500; Holm-adjusted P = 1.000).

PCD-CT was numerically better at detecting contact loss (score 3). For lateral contact, both readers correctly identified all 3 cases (6/6 readings) of contact loss with PCD-CT, but using EID-CT 1 of the readers missed 1 case. For medial contact, both readers correctly identified both cases (4/4 readings) with PCD-CT, while with EID-CT, 1 of the readers again missed 1 case. Due to the limited number of score 3 cases, these findings were analyzed descriptively, and no formal comparisons were conducted. Analysis of discordant ratings revealed that discrepancies occurred primarily in distinguishing between good contact (score 1) and tangential contact (score 2). Detailed ratings from each reader, compared to the reference standard, are presented in Table 3. Patterns of agreement and discordance for both readers are further summarized using confusion matrices arranged by scanner and contact type, as presented in Table 4.

Table 3.

Confusion matrices comparing reader-assigned scores with the reference standard for lateral and medial prosthesis contact across 18 cases.

Contact Type Reference standard Reader 1
Reader 2
PCD-CTa
EID-CTb
PCD-CT
EID-CT
Score 1 Score 2 Score 3 Score 1 Score 2 Score 3 Score 1 Score 2 Score 3 Score 1 Score 2 Score 3
Lateral Score 1 9 1 0 9 1 0 8 2 0 7 3 0
Score 2 0 5 0 0 5 0 0 3 2 1 3 1
Score 3 0 0 3 0 0 3 0 0 3 1 0 2
Total cases 18 (PCD-CT) 18 (EID-CT) 18 (PCD-CT) 18 (EID-CT)
Medial Score 1 8 4 0 7 5 0 10 2 0 8 4 0
Score 2 0 4 0 0 4 0 0 4 0 0 4 0
Score 3 0 0 2 0 1 1 0 0 2 0 0 2
Total cases 18 (PCD-CT) 18 (EID-CT) 18 (PCD-CT) 18 (EID-CT)

Diagonal cells indicate agreement with the reference standard, while off-diagonal cells represent discordant ratings. Scoring: 1 = good contact, 2 = tangential contact, 3 = no contact.

a

Photon-counting-detector.

b

Energy-integrating-detector.

Table 4.

Confusion matrices comparing reader-assigned scores with the reference standard and showing patterns of agreement and discordance for lateral and medial contact, according to scanner type, photon-counting-detector (PCD)-CT, and energy-integrating-detector (EID)-CT, with results from both readers pooled.

Contact type Reference standard PCD-CT (both readers)
EID-CT (both readers)
Score 1 Score 2 Score 3 Score 1 Score 2 Score 3
Lateral Score 1 17 3 0 16 4 0
Score 2 0 8 2 1 8 1
Score 3 0 0 6 1 0 5
Total 36 Readings 36 Readings
Medial Score 1 18 6 0 15 9 0
Score 2 0 8 0 0 8 0
Score 3 0 0 4 0 1 3
Total 36 Readings 36 Readings

Diagonal cells indicate agreement with the reference standard, while off-diagonal cells represent discordant ratings. Scoring: 1 = good contact, 2 = tangential contact, 3 = no contact.

Inter-reader agreement for contact scoring was numerically higher with PCD-CT than with EID-CT. For lateral contact, agreement was substantial with PCD-CT (κw = 0.804; 95% CI, 0.578-0.941) and moderate with EID-CT (κw = 0.571; 95% CI, 0.302-0.778), although the difference was not statistically significant (Δκw = 0.233; P = .250). For medial contact, agreement was substantial with PCD-CT (κw = 0.684; 95% CI, 0.191-0.934) and moderate with EID-CT (κw = 0.481; 95% CI, 0.024-0.835), with no statistically significant differences demonstrated between the 2 scanners (Δκw = 0.203; P = .250) (Figure 5).

Figure 5.

Bar graph showing inter-reader agreement for lateral and medial contact assessment with PCD-CT and EID-CT. Weighted kappa values are 0.804 and 0.571 for lateral contact with PCD-CT and EID-CT, respectively, and 0.684 and 0.481 for medial contact. A dashed horizontal line at κw = 0.61 marks the threshold for substantial agreement. PCD-CT values are above this threshold for both contact assessments, whereas EID-CT values are below it.

Linearly weighted kappa coefficients (κw) for inter-reader agreement in lateral and medial contact assessment with PCD-CT and EID-CT. For lateral contact, κw was 0.804 (95% CI, 0.578-0.941) for PCD-CT and 0.571 (95% CI, 0.302-0.778) for EID-CT. For medial contact, κw was 0.684 (95% CI, 0.191-0.934) for PCD-CT and 0.481 (95% CI, 0.024-0.835) for EID-CT. The dashed line shows the threshold for substantial agreement (κw = 0.61).

Reader confidence was significantly higher with PCD-CT compared to EID-CT for both lateral and medial contact assessments. For lateral contact, Reader 1 reported a mean confidence of 88.9 for PCD-CT compared to 79.4 for EID-CT, corresponding to a mean paired difference of 9.4 points (95% CI, 1.8-17.1; Holm-adjusted P = .019), and Reader 2 reported 86.9 compared to 63.9, corresponding to a mean difference of 23.1 points (95% CI, 13.4-32.7; Holm-adjusted P < .001). For medial contact, readers again reported higher confidence for PCD-CT. Reader 1 reported mean scores of 92.5 for PCD-CT compared to 76.1 for EID-CT, corresponding to a mean difference of 16.4 points (95% CI, 10.0-22.8; Holm-adjusted P < .001). Reader 2 reported mean scores of 93.3 and 75.6, corresponding to a mean paired difference of 17.8 points (95% CI, 6.8-28.8; Holm-adjusted P = .007). Wilcoxon signed-rank sensitivity analyses produced consistent results (all P ≤ .027).

For overall image quality, PCD-CT was also rated superior to EID-CT by both readers in paired comparisons. Reader 1 rated PCD-CT significantly better (score 5) in 16 out of 18 cases (88.9%; 95% CI, 65.3-98.6) and EID-CT slightly better (score 2) in 2 cases (11.1%; 95% CI, 1.4-34.7). Reader 2 rated PCD-CT significantly better (score 5) in 13 cases (72.2%; 95% CI, 46.5-90.3), PCD-CT slightly better (score 4) in 4 cases (22.2%; 95% CI, 6.4-47.6), and no difference (score 3) in 1 case (5.6%; 95% CI, 0.1-27.3); overall, Reader 2 preferred PCD-CT in 17 out of 18 cases (94.4%; 95% CI, 72.7-99.9). For both readers, ratings were significantly higher than the neutral score 3, showing a consistent preference for PCD-CT (both P < .001). The cases where EID-CT was preferred or rated equivalent to PCD-CT were those with severe metal artifacts, where both imaging techniques showed substantial artifacts but neither provided clearly superior visualization. In cases without metal artifacts, PCD-CT was uniformly preferred due to its superior spatial resolution and anatomical detail.

Discussion

This study assessed the visualization and diagnostic performance of PCD-CT compared to EID-CT for the evaluation of ossicular and stapes prostheses in cadaveric human heads. PCD-CT was consistently preferred for visualization of prosthesis positioning and adjacent anatomy. Given the small cohort size, the reader study demonstrated no statistically significant differences in assessing prosthesis contact points between the 2 technologies and in inter-reader agreement. However, diagnostic confidence was significantly higher on PCD-CT images, highlighting the potential advantages of PCD-CT technology.

The higher resolution and diagnostic confidence observed with PCD-CT compared with conventional EID-CT scanners may be explained by fundamental differences in detector design. PCD systems directly convert X-ray photons into electronic signals with a semiconductor, whereas EID systems use a 2-step process in which X-ray photons are first converted into visible light by a scintillator and then into electrical signal by photodiodes. This direct conversion design eliminates the reflective septa required between EID detector elements, allowing smaller detector pixels and improving both spatial resolution and geometric dose efficiency.11,21,28,29 In addition, individual photon counting and energy thresholding in PCD systems can suppress low-energy electronic noise and improve contrast-to-noise ratio.14,30 These advantages are more relevant in middle ear imaging for depicting small, high-contrast osseous and metallic structures, such as middle ear prostheses and their interactions with surrounding temporal bone anatomy.

Our findings are consistent with prior work demonstrating improved visualization of temporal bone anatomy and ossicular structures with PCD-CT.11,19,20,31 Hermans et al19 demonstrated that PCD-CT visualization of temporal bone anatomy was rated significantly higher than EID-CT across 71 exams (36 EID-CT vs 35 PCD-CT exams) at 22% lower radiation exposure. Benson et al20 reported an average 31% lower radiation dose with PCD-CT compared with EID-CT. In prior phantom- and patient-based work, results showed that PCD-CT provides superior spatial resolution and clearer visualization of ossicular prostheses compared to EID-CT.25,26 This study builds upon those findings by suggesting that these technical advantages may improve visualization and diagnostic confidence in a simulated surgical model, thereby bridging the gap between technical image quality assessment and clinically relevant interpretation.

These results have several clinical implications. Accurate postoperative assessment of ossicular prostheses is critical when surgical outcomes are unsatisfactory or complications are suspected. When postoperative hearing is suboptimal and imaging is inconclusive, the surgeon may recommend middle ear exploration and definitive intervention is determined intraoperatively. Reliance on surgical exploration for diagnosis and treatment underscores the limitations of conventional imaging. In a multicenter cohort of 308 revision procedures, preoperative CT could not identify the cause of failure in 46.1% of cases, corresponding to a sensitivity of 53.9%.4 Furthermore, some explorations are non-informative. Among 129 revision stapes surgeries, 6% revealed normal anatomy and apparently satisfactory prosthesis placement, with no structural abnormality to account for the deficit.32 Such procedures expose patients to the risks of revision surgery without a demonstrable benefit. As imaging can determine whether revision surgery is necessary, higher diagnostic certainty can lead to better management, potentially decreasing unnecessary revision surgery or avoiding unnecessary delay in revision surgery when indicated.

This study has several limitations. First, cadaveric specimens do not fully replicate in vivo conditions, as factors like motion, fluid, or postoperative changes were not represented in this model and may affect imaging performance. Second, the sample size and range of prosthesis types were narrow, potentially limiting the generalizability of the findings. The small number of prostheses and specimens also resulted in wide confidence intervals and limited statistical power. Therefore, these findings should be considered exploratory. While reader assessments of higher image quality and diagnostic confidence are promising, larger sample sizes will be required to determine whether PCD-CT provides superior diagnostic performance. Third, although readers were blinded to scanner type and the paired images were presented in random order, experienced readers may have recognized differences in image appearance, potentially introducing residual reader bias. The controlled design, paired comparisons with the same specimens, the blinded reader assessments, and availability of a reference standard help to mitigate these limitations. Future research to validate the findings in postoperative patient cohorts is needed to demonstrate that the technical advantages observed here translate into better clinical outcomes.

Conclusions

PCD-CT provided significantly greater confidence and preferred image quality for evaluating the position and contact points of middle ear prostheses compared to EID-CT under matched radiation dose. These exploratory findings support further development of PCD-CT as a potentially better modality for postoperative assessment of middle ear prostheses.

Supplementary Material

umag041_Supplementary_Data

Acknowledgments

This work was supported by an NIH grant (R01DC022228), a Mayo Clinic Radiology Research Internal Grant, and Mayo Clinic Graduate School of Biomedical Sciences. The authors thank Grace Medical for providing the prostheses used in this study. The authors also thank Kevin Kimlinger for his assistance in the manuscript submission.

Contributor Information

Maryam Sadeghian, Department of Radiology, Mayo Clinic, Rochester, MN 55905, United States; Mayo Clinic Graduate School of Biomedical Sciences, Mayo Clinic, Rochester, MN 55905, United States.

John C Benson, Department of Radiology, Mayo Clinic, Rochester, MN 55905, United States.

Paul J Farnsworth, Department of Radiology, Mayo Clinic, Rochester, MN 55905, United States.

John I Lane, Department of Radiology, Mayo Clinic, Rochester, MN 55905, United States.

Kevin J Treb, Department of Radiology, Mayo Clinic, Rochester, MN 55905, United States.

Karl R Khandalavala, Department of Otolaryngology-Head and Neck Surgery, Mayo Clinic, Rochester, MN 55905, United States.

Eric E Babajanian, Department of Otolaryngology-Head and Neck Surgery, Mayo Clinic, Rochester, MN 55905, United States.

Michael R Bruesewitz, Department of Radiology, Mayo Clinic, Rochester, MN 55905, United States.

Joseph R Swicklik, Department of Radiology, Mayo Clinic, Rochester, MN 55905, United States.

Fang-Shu Ou, Division of Quantitative Health Sciences, Mayo Clinic, Rochester, MN 55905, United States.

Cynthia H McCollough, Department of Radiology, Mayo Clinic, Rochester, MN 55905, United States; Mayo Clinic Graduate School of Biomedical Sciences, Mayo Clinic, Rochester, MN 55905, United States.

Matthew L Carlson, Department of Otolaryngology-Head and Neck Surgery, Mayo Clinic, Rochester, MN 55905, United States.

Shuai Leng, Department of Radiology, Mayo Clinic, Rochester, MN 55905, United States; Mayo Clinic Graduate School of Biomedical Sciences, Mayo Clinic, Rochester, MN 55905, United States.

Author contributions

Maryam Sadeghian (Conceptualization, Data curation, Formal analysis, Methodology, Validation, Visualization, Writing—original draft, Writing—review & editing), John Benson (Conceptualization, Funding acquisition, Investigation, Methodology, Resources, Visualization, Writing—review & editing), Paul Farnsworth (Conceptualization, Funding acquisition, Investigation, Methodology, Resources, Visualization, Writing—review & editing), John Lane (Conceptualization, Funding acquisition, Investigation, Methodology, Resources, Writing—review & editing), Kevin Treb (Methodology, Writing—review & editing), Karl Khandalavala (Methodology, Writing—review & editing), Eric Babajanian (Methodology, Writing—review & editing), Michael Bruesewitz (Methodology, Writing—review & editing), Joseph Swicklik (Methodology, Writing—review & editing), Fang-Shu Ou (Formal analysis, Writing—review & editing), Cynthia McCollough (Funding acquisition, Investigation, Methodology, Writing—review & editing), Matthew Carlson (Conceptualization, Funding acquisition, Investigation, Methodology, Resources, Writing—review & editing), and Shuai Leng (Conceptualization, Funding acquisition, Investigation, Methodology, Project administration, Resources, Software, Supervision, Validation, Writing—original draft, Writing—review & editing)

Supplementary material

Supplementary materialis available at Radiology Advances online.

Funding

Research reported in this work was supported by the National Institutes of Health under award number R01DC022228. The content is solely the responsibility of the authors and does not necessarily represent the official views of the National Institute of Health. Research reported in this work was also supported by the Mayo Clinic CT Clinical Innovation Center and the Mayo Clinic Graduate School of Biomedical Sciences.

Conflicts of interest

Please see ICMJE form(s) for author conflicts of interest. These have been provided as supplementary materials. C.H.M. is the recipient of a research grant to the institution from Siemens Healthineers. The other authors have no relevant conflicts of interest to disclose.

Data availability

The data underlying this article will be made publicly available in accordance with NIH data sharing requirements. Data will be deposited in a publicly accessible repository upon completion of embargo period; the repository name and access link will be provided upon publication.

Reference

  • 1. Volandri G, Di Puccio F, Forte P, Manetti S.  Model-oriented review and multi-body simulation of the ossicular chain of the human middle ear. Med Eng Phys. 2012;34(9):1339-1355. 10.1016/j.medengphy.2012.02.011 [DOI] [PubMed] [Google Scholar]
  • 2. Jia H, Zhang J, Ma K, Qiao X, Ren L, Shi X.  Application of convolutional neural networks in medical images: a bibliometric analysis. Quant Imaging Med Surg. 2024;14(5):3501-3518. 10.21037/qims-23-1600 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3. Tomasoni M, Borsetto D, Canzi P, et al.  Indications, surgical strategies and hearing outcomes of revision stapes surgery for otosclerosis: a systematic review and meta-analysis on 2602 patients. Otolaryngol Head Neck Surg. 2025;172(1):50-62. 10.1002/ohn.974 [DOI] [PubMed] [Google Scholar]
  • 4. Bruschini L, Lazzerini F, Capobianco S, et al.  Stapes revision surgery: intraoperative findings and audiological results. A multicentric study. Acta Otorhinolaryngol Ital. 2025;45(Suppl 1):S61-S70. 10.14639/0392-100X-suppl.1_3-45-2025-A1339 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5. Canzi P, Carlotto E, Bruschini L, et al.  Extrusion and dislocation in titanium middle ear prostheses: a literature review. Brain Sci. 2023;13(10):1476. 10.3390/brainsci13101476 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6. Kosling S, Bootz F.  CT and MR imaging after middle ear surgery. Eur J Radiol. 2001;40(2):113-118. 10.1016/s0720-048x(01)00379-5 [DOI] [PubMed] [Google Scholar]
  • 7. Ali HI, Khater NH.  Otosclerosis and complications of stapedectomy: CT and MRI correlation. Alexandria J Med. 2019;54(3):197-201. 10.1016/j.ajme.2017.07.001 [DOI] [Google Scholar]
  • 8. Tames H, Padula M, Sarpi MO, et al.  Postoperative imaging of the temporal bone. Radiographics. 2021;41(3):858-875. 10.1148/rg.2021200126 [DOI] [PubMed] [Google Scholar]
  • 9. Stone JA, Mukherji SK, Jewett BS, Carrasco VN, Castillo M.  CT evaluation of prosthetic ossicular reconstruction procedures: what the otologist needs to know. Radiographics. 2000;20(3):593-605. 10.1148/radiographics.20.3.g00ma03593 [DOI] [PubMed] [Google Scholar]
  • 10. Pickuth D, Brandt S, Berghaus A, Spielmann RP, Heywang-Kobrunner SH.  Vertigo after stapes surgery: the role of high resolution CT. Br J Radiol. 2000;73(873):1021-1023. 10.1259/bjr.73.873.11064662 [DOI] [PubMed] [Google Scholar]
  • 11. Rajendran K, Voss BA, Zhou W, et al.  Dose reduction for sinus and temporal bone imaging using photon-counting detector CT with an additional tin filter. Invest Radiol. 2020;55(2):91-100. 10.1097/RLI.0000000000000614 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12. Sartoretti T, Wildberger JE, Flohr T, Alkadhi H.  Photon-counting detector CT: early clinical experience review. Br J Radiol. 2023;96(1147):20220544. 10.1259/bjr.20220544 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13. Rajendran K, Petersilka M, Henning A, et al.  First clinical photon-counting detector CT system: technical evaluation. Radiology. 2022;303(1):130-138. 10.1148/radiol.212579 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14. Leng S, Bruesewitz M, Tao S, et al.  Photon-counting detector CT: system design and clinical applications of an emerging technology. Radiographics. 2019;39(3):729-743. 10.1148/rg.2019180115 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15. McCollough CH, Rajiah P, Bois JP, et al.  Comparison of photon-counting detector and energy-integrating detector CT for visual estimation of coronary percent luminal stenosis. Radiology. 2023;309(3):e230853. 10.1148/radiol.230853 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16. McCollough CH, Winfree TN, Melka EF, Rajendran K, Carter RE, Leng S.  Photon-counting detector computed tomography versus energy-integrating detector computed tomography for coronary artery calcium quantitation. J Comput Assist Tomogr. 2024;48(2):212-216. 10.1097/RCT.0000000000001554 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17. Leng S, Rajendran K, Gong H, et al.  150-Mum spatial resolution using photon-counting detector computed tomography technology: technical performance and first patient images. Invest Radiol. 2018;53(11):655-662. 10.1097/RLI.0000000000000488 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18. Willemink MJ, Grist TM.  Counting photons: the next era for CT imaging?  Radiology. 2022;303(1):139-140. 10.1148/radiol.213203 [DOI] [PubMed] [Google Scholar]
  • 19. Hermans R, Boomgaert L, Cockmartin L, Binst J, De Stefanis R, Bosmans H.  Photon-counting CT allows better visualization of temporal bone structures in comparison with current generation multi-detector CT. Insights Imaging. 2023;14(1):112. 10.1186/s13244-023-01467-w [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20. Benson JC, Rajendran K, Lane JI, et al.  A new frontier in temporal bone imaging: photon-counting detector CT demonstrates superior visualization of critical anatomic structures at reduced radiation dose. AJNR Am J Neuroradiol. 2022;43(4):579-584. 10.3174/ajnr.A7452 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21. Zhou W, Lane JI, Carlson ML, et al.  Comparison of a photon-counting-detector CT with an energy-integrating-detector CT for temporal bone imaging: a cadaveric study. AJNR Am J Neuroradiol. 2018;39(9):1733-1738. 10.3174/ajnr.A5768 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22. Macielak RJ, Benson JC, Lane JI, Carlson ML, Leng S.  Photon-counting detector CT for temporal bone imaging: up to three times the resolution at half the radiation dose. Otol Neurotol. 2022;43(10):e1205-e1207. 10.1097/MAO.0000000000003682 [DOI] [PubMed] [Google Scholar]
  • 23. Flohr T, Schmidt B.  Technical basics and clinical benefits of photon-counting CT. Invest Radiol. 2023;58(7):441-450. 10.1097/RLI.0000000000000980 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24. McCollough CH, Rajendran K, Baffour FI, et al.  Clinical applications of photon counting detector CT. Eur Radiol. 2023;33(8):5309-5320. 10.1007/s00330-023-09596-y [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25. Dutta A, Farnsworth PJ, Leng S, et al.  Comparative analysis of middle ear prostheses using photon-counting detector CT versus conventional CT. AJNR Am J Neuroradiol. 2026;47(4):1026-1030. 10.3174/ajnr.A9075 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26. Sadeghian M, Benson JC, Farnsworth PJ, et al.  Impact of spatial resolution on CT imaging of middle ear prostheses: comparison of photon-counting and energy-integrating detectors. AJNR Am J Neuroradiol. 2026;47(5):1341-1348. 10.3174/ajnr.A9103 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27. Landis JR, Koch GG.  The measurement of observer agreement for categorical data. Biometrics. 1977;33(1):159-174. [PubMed] [Google Scholar]
  • 28. McCollough CH, Rajendran K, Leng S, et al.  The technical development of photon-counting detector CT. Eur Radiol. 2023;33(8):5321-5330. 10.1007/s00330-023-09545-9 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29. Leng S, Yu Z, Halaweish A, et al.  Dose-efficient ultrahigh-resolution scan mode using a photon counting detector computed tomography system. J Med Imaging (Bellingham). 2016;3(4):043504. 10.1117/1.JMI.3.4.043504 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30. Flohr T, Petersilka M, Henning A, Ulzheimer S, Ferda J, Schmidt B.  Photon-counting CT review. Phys Med. 2020;79:126-136. 10.1016/j.ejmp.2020.10.030 [DOI] [PubMed] [Google Scholar]
  • 31. Rao D, Murray JV, Agarwal AK, Sandhu SJ, Rhyner PA.  Comprehensive review of inner ear anatomy on photon-counting CT. AJNR Am J Neuroradiol. 2025;46(5):862-867. 10.3174/ajnr.A8410 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32. Luryi AL, Schettino A, Michaelides EM, Babu S, Bojrab DI, Schutt CA.  Revision stapes surgery: hearing symptoms and associations with intraoperative findings and outcomes. Otolaryngol Head Neck Surg. 2022;167(2):350-355. 10.1177/01945998211062074 [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

umag041_Supplementary_Data

Data Availability Statement

The data underlying this article will be made publicly available in accordance with NIH data sharing requirements. Data will be deposited in a publicly accessible repository upon completion of embargo period; the repository name and access link will be provided upon publication.


Articles from Radiology Advances are provided here courtesy of Oxford University Press

RESOURCES