Abstract
Highly cross-linked polyethylene (HXLPE) has become the standard acetabular liner material in total hip arthroplasty (THA) because it has markedly reduced wear-related osteolysis and aseptic loosening. However, cross-linking also reduces resistance to fatigue and crack propagation, which may lead to failure in areas of concentrated stress. Acetabular liner failure remains rare and most commonly occurs following trauma.
We report an atraumatic early catastrophic failure of an HXLPE acetabular liner in a 62-year-old woman, 20 months after primary THA at an outside institution. She presented with groin and lateral hip pain, joint squeaking, and an inability to flex the hip past 90 degrees. Radiographs demonstrated eccentric positioning of the femoral head within the acetabular shell and displacement of the acetabular locking ring. During revision, the liner was found fractured into multiple fragments with surrounding metallosis. An eccentrically placed acetabular shell screw protruded into the inner surface of the shell, and the location of the screw head corresponded with backside liner wear. The rim screw was removed and the liner was exchanged. At telephone follow-up over eight years later, the patient reported ambulating without difficulty and denied pain or subsequent revision.
This case describes a preventable, technique-dependent mechanism of early HXLPE liner failure and supports routinely confirming flush seating of the acetabular screw before liner insertion during THA.
Keywords: acetabular liner failure, acetabular screw protrusion, hip joint, polyethylene liner, revision arthroplasty
Introduction
Total hip arthroplasty (THA) is a highly effective surgical procedure for the management of various hip pathologies including end-stage hip osteoarthritis, with excellent patient reported outcomes and long-term implant survivorship. Historically, non-cross-linked ultra-high-molecular-weight polyethylene (UHMWPE) was favored for the acetabular liner component in THA due to its low friction and wear-resistant properties [1,2]. However, the most common cause of late revision THA with a metal or ceramic head on a conventional UHMWPE liner was aseptic loosening driven by osteolysis resulting from particulate wear debris [3,4].
The development of highly cross-linked polyethylene (HXLPE) liners has markedly reduced THA revision rates related to osteolysis and aseptic loosening [5-8]. HXLPE is manufactured utilizing sequential irradiation and annealing processes that enhance wear resistance by forming carbon-carbon bonds between polymer chains and reducing oxidation and free radical formation [9]. Although these manufacturing steps improve wear performance, they may also reduce resistance to fatigue crack propagation and fracture under concentrated or repetitive stress [10,11]. Accordingly, while catastrophic liner failure remains uncommon, accounting for less than 1% of early revision THAs [12], early mechanical failure has been reported in high-stress regions such as the liner rim, locking mechanism, or areas exposed to impingement or localized point-loading [13,14].
HXLPE liners are widely used in modern THA. Therefore, understanding the risk of fatigue failure and minimizing technical factors that concentrate stress on the material are important. Previously reported failure mechanisms have largely been related to the implant or component malposition, which may be difficult for the surgeon to address during the index operation. In this case, the stress riser resulted from a single modifiable technical step, the seating of an acetabular shell screw. This mechanism has been described with a ceramic liner [15] but is rarely reported with HXLPE. This report defines the angular tolerance needed to avoid this failure and identifies a simple intraoperative check to confirm appropriate screw seating.
Case presentation
Clinical findings and diagnostic imaging
A 62-year-old female with a medical history of fibromyalgia, lower back pain, and BMI of 39.7 was referred to our institution in 2015 with complaints of groin and lateral hip pain following a primary left THA performed at an outside hospital 20 months earlier. Three months prior to being seen at our institution, she experienced an acute worsening of pain without trauma, characterized by sharp pain and an inability to bear weight. She was placed on bed rest for one month by the direction of an outside surgeon and later resumed ambulation with assistance. Upon presentation to our clinic, the patient described a constant, sharp pain that began acutely after flexing her hip. She also reported associated squeaking in the joint and an inability to flex her hip past 90 degrees.
On physical examination, the patient was unable to flex her left hip past 90 degrees, with pain limiting both internal and external rotation. Her gait was antalgic, and she had a positive Trendelenburg sign. Radiographs of the left hip showed eccentric positioning of the femoral head within the acetabular shell, with cephalad displacement of the prosthetic femoral head and osteolysis at the lateral aspect of the acetabular component (Figures 1, 2). The femoral component was in a neutral position. Additionally, radiographs showed displacement of the acetabular locking ring. Failure or dislodgement of the polyethylene acetabular liner was suspected, and surgical revision was planned.
Figure 1. Anterior posterior radiograph of the left hip showing eccentric positioning of the femoral head implant within the acetabular shell, and a displaced RingLoc mechanism (red arrow) around the femoral stem.

Figure 2. Frog leg lateral radiograph of the left hip confirming displacement of the RingLoc mechanism (red arrow) without evidence of periprosthetic fracture.

Records from the outside hospital indicated the primary THA was performed via a posterior lateral approach. The prior non-cemented implants consisted of a femoral stem (size 6 x 132) with a cobalt-chromium head (36 mm, +3 mm offset), a RingLoc acetabular shell with an HXLPE liner, and a single acetabular shell screw (Zimmer Biomet, USA) placed in the posterior superior quadrant.
Surgical intervention
A posterior lateral approach was used with the patient in lateral decubitus position. Intraoperatively, synovialized membrane was found in the subcutaneous tissues superficial to the deep fascial layer, suggesting prior seroma formation. A dark, black fluid, consistent with metallosis, was aspirated from the joint space, and the periacetabular synovial tissues were stained with metallosis.
The HXLPE acetabular shell liner was fractured into multiple pieces (Figure 3) with the RingLoc metal mechanism (Zimmer Biomet, USA) displaced into the surrounding soft tissues. The acetabular dome screw was found to be eccentrically placed in the shell hole, protruding into the inner surface of the acetabular shell. Eccentric wear present on the backside of the liner in continuum with the margins of the fracture fragments corresponded with the location of the prominent screw. The screw was removed, and both the acetabular and femoral components were confirmed to be well-fixed both visually and radiographically. The acetabular component was appropriately positioned regarding inclination and anteversion. The femoral component was appropriately positioned regarding version.
Figure 3. (a) Fragmented HXLPE acetabular shell liner with 3 large fragments and the RingLoc mechanism; (b) Close-up image showing significant wear and metallosis staining.

Due to the well-fixed and well-positioned existing components, a constrained liner was selected for revision. A series 23 constrained liner, RingLoc mechanism, and 36 mm neutral Freedom head (Zimmer Biomet, USA) were placed. The components were confirmed to be well-seated, and the hip was concentrically reduced. Intraoperative radiographs confirmed reduction and appropriate placement of the components. A post-operative radiograph showed anatomic alignment of the implants without evidence of hardware complications (Figure 4). The patient was given instructions to bear weight as tolerated using a front wheeled walker for six weeks. She ambulated with the physical therapist prior to being discharged home.
Figure 4. Post-operative portable anterior posterior pelvic radiograph showing removal of the acetabular screw, concentric positioning of the femoral head within the acetabular shell and a replaced RingLoc mechanism. No periprosthetic fracture was observed.

Follow-up and outcomes
At the two-week follow-up, physical examination revealed a well healing skin incision, and radiographs did not show signs of loosening or failure. The patient was able to flex her hip to 90° without pain. At six weeks, the patient continued to use a front wheeled walker, and a repeat pelvic radiograph was unchanged without signs of osteolysis, loosening or implant failure (Figure 5). Over eight years later, at an out-of-state hospital, the patient received a surgical evaluation for pain in her contralateral hip. At this time, the patient followed up with our clinic over the phone and reported ambulating well on her left leg. She denied having any pain in her left leg and denied having any complications or revision surgery on her left hip since her six-week follow-up appointment.
Figure 5. Anterior posterior pelvic radiograph at six-week follow-up, without evidence of periprosthetic fracture or dislocation.

Discussion
Overall revision rates have ranged from 0 to 8.6%, while the revision rates due to polyethylene wear or osteolysis range from 0 to 0.01% [16]. The development and widespread utilization of HXLPE liners has contributed to the reduction in revision rates [17,18]. However, the improved wear properties of HXLPE come with a tradeoff of reduced fatigue and fracture resistance [17,19]. Significant advancements in material science and manufacturing including vitamin E treatment, irradiation methods, and post-radiation thermal treatments, have helped address these materials characteristics, but continued post-market surveillance and evaluation of long-term data is needed to evaluate their superiority [17].
Our case report details the catastrophic failure of an HXLPE liner due to eccentric and prominent placement of an acetabular shell rim screw and highlights the importance of surgical technique in THA to promote survivorship. The screw trajectory must be within a pre-specified range which varies based on the acetabular shell and screw choice. For the RingLoc acetabular shell and rim screw in this case, the allowed range is +10 to -10 degrees from the trajectory of the screw hole of the acetabular shell. This leaves the orthopedic surgeon with a 20-degree target for screw placement (Figure 6). Deviations outside this range will lead to the screw head sitting prominently above the inner surface of the acetabular shell. In this case, the backside of the HXLPE liner was subjected to localized stress due to the prominence of the screw head. A similar early-failure mechanism was observed by Lee et al. in a ceramic-on-ceramic head-liner combination due to prominent acetabular screw placement [15]. Backside wear may also be implicated in higher rates of periacetabular osteolysis observed in THA utilizing acetabular screws compared to cementless components fitted without screws [20].
Figure 6. Surgical technique diagram for acetabular rim screw placement, annotated with the –10 to +10 range for the RingLoc implant system.

The figure has been created by Dr. David Dallas-Orr on Microsoft PowerPoint.
In summary, avoidance of acetabular screw malposition can be accomplished through adherence to appropriate surgical technique. We recommend utilization of a variable angle drill instead of a fixed angle drill to avoid screw angulation outside of the accepted margins. Similarly, the drill guide must be fully seated in the shell’s screw counterbore to create a pilot hole within the limits of angulation trajectory. Additionally, prior to insertion of the liner, we recommend confirming flush positioning of the acetabular screw with visual and tactile inspection.
Conclusions
In conclusion, this case report demonstrates a rare occurrence of an atraumatic failure of a HXLPE acetabular cup liner due to wear and localized stress from a prominent acetabular rim screw. Additionally, it contributes to the literature for the evaluation of the failure mechanisms of HXLPE liners. Lastly, it emphasizes the importance of proper surgical technique during the placement of acetabular screws.
Disclosures
Human subjects: Informed consent for treatment and open access publication was obtained or waived by all participants in this study.
Conflicts of interest: In compliance with the ICMJE uniform disclosure form, all authors declare the following:
Payment/services info: All authors have declared that no financial support was received from any organization for the submitted work.
Financial relationships: All authors have declared that they have no financial relationships at present or within the previous three years with any organizations that might have an interest in the submitted work.
Other relationships: All authors have declared that there are no other relationships or activities that could appear to have influenced the submitted work.
Author Contributions
Concept and design: David Dallas-Orr, Gregory J. Harbison, Samuel K. Simister, Shannon Tse, John Meehan
Acquisition, analysis, or interpretation of data: David Dallas-Orr, Gregory J. Harbison
Drafting of the manuscript: David Dallas-Orr, Gregory J. Harbison
Critical review of the manuscript for important intellectual content: David Dallas-Orr, Gregory J. Harbison, Samuel K. Simister, Shannon Tse, John Meehan
Supervision: John Meehan
References
- 1.John Charnley’s assistant: Harry Craven and the Charnley hip. Metcalfe RG, Porter M. Biomed Mater Devices. 2024;2:1156–1174. [Google Scholar]
- 2.Polyethylene in total hip arthroplasty: half a century in the limelight. García-Rey E, García-Cimbrelo E. J Orthop Traumatol. 2010;11:67–72. doi: 10.1007/s10195-010-0091-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Failure analysis on retrieved ultra high molecular weight polyethylene (UHMWPE) acetabular cups. Burger ND, de Vaal PL, Meyer JP. Eng Fail Anal. 2007;14:1329–1345. [Google Scholar]
- 4.Osteolysis as it pertains to total hip arthroplasty. Kurcz B, Lyons J, Sayeed Z, Anoushiravani AA, Iorio R. Orthop Clin North Am. 2018;49:419–435. doi: 10.1016/j.ocl.2018.06.001. [DOI] [PubMed] [Google Scholar]
- 5.Cross-linked polyethylene for total hip arthroplasty markedly reduces revision surgery at 16 years. de Steiger R, Lorimer M, Graves SE. J Bone Joint Surg Am. 2018;100:1281–1288. doi: 10.2106/JBJS.17.01221. [DOI] [PubMed] [Google Scholar]
- 6.The wear rate and survivorship in total hip arthroplasty using a third-generation ceramic head on a conventional polyethylene liner: a minimum of 15-year follow-up. Shim BJ, Park SJ, Park CH. Hip Pelvis. 2022;34:115–121. doi: 10.5371/hp.2022.34.2.115. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Otto Aufranc Award: crosslinking reduces tha wear, osteolysis, and revision rates at 15-year followup compared with noncrosslinked polyethylene. Hopper RH Jr, Ho H, Sritulanondha S, Williams AC, Engh CA Jr. Clin Orthop Relat Res. 2018;476:279–290. doi: 10.1007/s11999.0000000000000036. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Incidence of osteolysis and aseptic loosening following Metal-on-Highly Cross-Linked Polyethylene hip arthroplasty: a systematic review of studies with up to 15-year follow-up. Prock-Gibbs H, Pumilia CA, Meckmongkol T, Lovejoy J, Mumith A, Coathup M. J Bone Joint Surg Am. 2021;103:728–740. doi: 10.2106/JBJS.20.01086. [DOI] [PubMed] [Google Scholar]
- 9.A novel method of cross-linking ultra-high-molecular-weight polyethylene to improve wear, reduce oxidation, and retain mechanical properties. Recipient of the 1999 HAP Paul Award. Muratoglu OK, Bragdon CR, O'Connor DO, Jasty M, Harris WH. J Arthroplasty. 2001;16:149–160. doi: 10.1054/arth.2001.20540. [DOI] [PubMed] [Google Scholar]
- 10.Clinical trade-offs in cross-linked ultrahigh-molecular-weight polyethylene used in total joint arthroplasty. Pruitt LA, Ansari F, Kury M, et al. J Biomed Mater Res B Appl Biomater. 2013;101:476–484. doi: 10.1002/jbm.b.32887. [DOI] [PubMed] [Google Scholar]
- 11.Tradeoffs amongst fatigue, wear, and oxidation resistance of cross-linked ultra-high molecular weight polyethylene. Atwood SA, Van Citters DW, Patten EW, Furmanski J, Ries MD, Pruitt LA. J Mech Behav Biomed Mater. 2011;4:1033–1045. doi: 10.1016/j.jmbbm.2011.03.012. [DOI] [PubMed] [Google Scholar]
- 12.Early failures in total hip arthroplasty -- a changing paradigm. Melvin JS, Karthikeyan T, Cope R, Fehring TK. J Arthroplasty. 2014;29:1285–1288. doi: 10.1016/j.arth.2013.12.024. [DOI] [PubMed] [Google Scholar]
- 13.Unexpected failure of highly cross-linked polyethylene acetabular liner. Waewsawangwong W, Goodman SB. J Arthroplasty. 2012;27:323–324. doi: 10.1016/j.arth.2011.04.010. [DOI] [PubMed] [Google Scholar]
- 14.Fracture of a cross-linked polyethylene liner: a multifactorial issue. Blumenfeld TJ, McKellop HA, Schmalzried TP, Billi F. J Arthroplasty. 2011;26:666–668. doi: 10.1016/j.arth.2010.07.009. [DOI] [PubMed] [Google Scholar]
- 15.Acetabular screw head-induced ceramic acetabular liner fracture in cementless ceramic-on-ceramic total hip arthroplasty. Lee SC, Jung KA, Nam CH, Kim TH, Ahn NK, Hwang SH. Orthopedics. 2010;33:353–355. doi: 10.3928/01477447-20100329-30. [DOI] [PubMed] [Google Scholar]
- 16.Wear, osteolysis, and aseptic loosening following total hip arthroplasty in young patients with highly cross-linked polyethylene: a review of studies with a follow-up of over 15 years. Deans CF, Buckner BC, Garvin KL. J Clin Med. 2023;12:6615. doi: 10.3390/jcm12206615. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.What have we learned from 20 years of using highly crosslinked PE in total hip arthroplasty? Langlois J, Hamadouche M. Orthop Traumatol Surg Res. 2023;109:103457. doi: 10.1016/j.otsr.2022.103457. [DOI] [PubMed] [Google Scholar]
- 18.Deformation, yielding, fracture and fatigue behavior of conventional and highly cross-linked ultra high molecular weight polyethylene. Pruitt LA. Biomaterials. 2005;26:905–915. doi: 10.1016/j.biomaterials.2004.03.022. [DOI] [PubMed] [Google Scholar]
- 19.Medel FJ, Furmanski J. UHMWPE Biomaterials Handbook. Oxford: William Andrew; 2016. Fatigue and fracture of UHMWPE; pp. 693–720. [Google Scholar]
- 20.The use of acetabular screws in total hip arthroplasty and its influence on wear and periacetabular osteolysis in the long-term follow-up. Brodt S, Bischoff K, Schulze M, Nowack D, Roth A, Matziolis G. Int Orthop. 2022;46:717–722. doi: 10.1007/s00264-021-05219-7. [DOI] [PMC free article] [PubMed] [Google Scholar]