Abstract
Children are more radiosensitive than adults, making dose optimisation in paediatric computed tomography (CT) essential. Although Diagnostic Reference Levels (DRLs) are internationally recommended as optimisation tools, national DRLs for paediatric CT in Jordan remain limited and outdated. This study aimed to establish national DRLs for six paediatric CT protocols, evaluate inter-hospital and age-related dose variations, and compare results with international benchmarks. A retrospective multicentre study was conducted across six Jordanian hospitals between December 2023 and September 2025, including 3794 paediatric patients stratified into four age groups (< 1, 1–4, 5–10, and 11–18 years). Volumetric CT dose index (CTDIvol) and dose-length product (DLP) data were collected for six protocols: brain, chest, abdomen–pelvis, chest–abdomen–pelvis, sinuses, and contrast-enhanced neck soft tissue. DRLs were defined as the 75th percentile of institutional median CTDIvol and DLP values. Inter-hospital and age group variations were analysed, and univariable regression analyses assessed acquisition parameters associated with dose variation. Dose increased with patient age for trunk protocols, whereas brain, sinuses, and neck CT showed comparatively stable patterns. Substantial inter-hospital variability was observed across protocols, with institutional median CTDIvol differing markedly between centres, particularly for trunk examinations in younger age groups. In univariable regression analyses, all four acquisition parameters (kVp, mAs, pitch, and slice thickness) were significantly associated with CTDIvol and DLP (p ≤ 0.001), with kVp demonstrating the strongest association (R2 = 0.603 for CTDIvol; R2 = 0.630 for DLP). Compared with published international DRLs, Jordanian brain CT dose metrics were higher in multiple age groups; chest comparisons should be interpreted cautiously where international benchmarks are weight-banded. This study established national paediatric CT DRLs for Jordan, highlighting the need for standardised protocols and periodic DRL review to enhance radiation protection.
Keywords: Diagnostic reference levels, Paediatric computed tomography, Radiation dose, Dose optimisation, Radiation protection
Introduction
Computed tomography (CT) is widely used in clinical imaging due to its high spatial resolution and diagnostic efficacy. However, its contribution to collective radiation dose is substantial, particularly in paediatric populations, who are more radiosensitive than adults due to their developing tissues and longer life expectancy [1]. As a result, children are more vulnerable to radiation-induced effects, underscoring the importance of dose optimisation in paediatric CT imaging [2, 3].
Globally, the sharp increase in CT utilisation has been recognised as a major source of medical radiation exposure. According to the United Nations Scientific Committee on the Effects of Atomic Radiation, CT accounts for approximately 9.6% of all medical radiological examinations, yet contributes about 61.6% of the overall collective effective dose from medical radiological exposure [4].
Diagnostic Reference Levels (DRLs) are widely endorsed as an optimisation tool in radiological protection. While not regulatory limits, DRLs are typically set at the 75th percentile of dose distributions obtained from routine clinical practice. Their purpose is to identify unusually high doses and trigger local review of imaging protocols [5]. The International Commission on Radiological Protection (ICRP) recommends that DRLs be reviewed and updated at regular intervals, generally every three to five years, to reflect changes in technology, clinical practice, and patient demographics [5]. When implemented effectively, DRLs can support dose reduction through standardisation and optimisation of protocols. In the United Kingdom, a national paediatric CT dose audit proposed updated DRLs for head CT that were lower than previous UK NDRLs by approximately 18–41% (depending on age band), whereas the same audit reported insufficient data to propose national DRLs for abdomen–pelvis CT [6].
In Jordan, national data on paediatric CT radiation exposure remain scarce. A previous study by Rawashdeh et al. (2019) proposed local DRLs for four CT examination types: brain, chest, abdomen–pelvis, and chest–abdomen–pelvis across four hospitals. Their findings showed considerable variability in the volumetric CT dose index (CTDIvol) and Dose Length Product (DLP) values between centres, with up to a four-fold difference for some procedures. In general, DLP values in Jordan were relatively high compared to international benchmarks, indicating the need for more consistent and optimised scanning protocols [7]. More recently, Hamedat et al. (2025) conducted a multicentre national survey of paediatric head CT across eight Jordanian hospitals and proposed age-based national DRLs, reporting substantial inter-hospital variability and underscoring the need for dose standardisation [8].
The present study builds upon and substantially extends the earlier work, providing more recent data with broader national representation. While the previous study was limited to four hospitals and four CT examinations, the current study includes six major hospitals and six paediatric CT procedures, representing a markedly larger and more comprehensive dataset. This expansion enables a more robust assessment of variations in CTDIvol and DLP, while the proposed DRLs are benchmarked against internationally reported values to support optimisation and harmonisation of paediatric CT protocols in Jordan.
This study therefore aims to establish updated national DRLs for paediatric CT in Jordan, assess inter-hospital and age-related variations in CTDIvol and DLP, and compare the proposed DRLs with international benchmarks to support dose optimisation and protocol harmonisation.
Methods
Study design
This multicentre, retrospective study was conducted to propose updated national DRLs for paediatric CT examinations in Jordan. Data were collected from December 2023 to September 2025 across six hospitals strategically selected to represent different geographical regions: two from northern Jordan, two from the central region, and two from the south. The selection of hospitals was based on their geographic distribution, technological diversity, and commitment to quality assurance programmes. A purposive sampling strategy was adopted to ensure representativeness across Jordan’s healthcare regions (north, central, and south). In addition, hospitals were included based on the availability of sufficient paediatric CT data, as paediatric examinations were not routinely performed in all centres. This approach ensured that the participating hospitals reflected both regional diversity and practical feasibility for comprehensive data collection.
In accordance with the conditions of ethical approval obtained from the Jordanian Scientific Research Ethics Committee, the names of participating hospitals have been anonymised and are referred to as Hospitals A to F throughout the manuscript.
The CT equipment varied in manufacturer and model, reflecting the technological diversity of participating institutions. The distribution was as follows: two Philips Brilliance 64-slice scanners (Philips Healthcare, Best, The Netherlands) at Hospitals A and B; one Siemens Somatom Definition 128-slice scanner (Siemens Healthineers, Erlangen, Germany) at Hospital C; one GE Revolution EVO 64-slice scanner (GE Healthcare, Chicago, IL, USA) at Hospital D; one Toshiba Aquilion Dual 64-slice scanner (Canon Medical Systems, Otawara, Japan) at Hospital E; and one Philips Brilliance 128-slice scanner (Philips Healthcare, Best, The Netherlands) at Hospital F.
At each site, a clinical coordinator with expertise in CT imaging was assigned to facilitate data collection. Standardised and pre-validated data collection forms were used across all hospitals to ensure consistency and reliability.
The study focused on six commonly performed CT examinations in paediatric imaging: brain, chest, abdomen–pelvis, chest–abdomen–pelvis, sinuses, and neck soft tissue. All examinations were acquired without intravenous contrast, except for neck soft tissue, which was performed with contrast. Examination selection was based on frequency of use and clinical relevance across the participating hospitals.
Dose data and analysis
Radiation dose information was extracted directly from the Picture Archiving and Communication Systems (PACS) and Radiology Information Systems (RIS) of all participating hospitals using a standardised data-collection template. Extracted variables included patient age (years), sex, CT protocol, and scanner acquisition parameters (kilovoltage peak (kVp), tube current time product (mAs), pitch, field of view (FOV), and slice thickness), together with the corresponding dose indices (CTDIvol and DLP). All data were anonymised prior to analysis. Patients were stratified into four age categories (< 1 year, 1–4 years, 5–10 years, and 11–18 years). International paediatric DRL guidance recommends stratifying patients using discrete age bands and/or body-size surrogates (commonly weight bands for trunk examinations). European guidance commonly applies age points up to 15 years (0, 1, 5, 10 and 15 years) and recommends weight bands for thorax/abdomen CT, while age bands are frequently used for head CT [9]. In the present retrospective multicentre dataset, patient weight (and other size surrogates) was not consistently recorded across all participating hospitals; therefore, age-based grouping was adopted to ensure consistent stratification across centres and sufficient numbers per category. The four age categories (< 1, 1–4, 5–10 and 11–18 years) approximate commonly used international paediatric groupings (< 1, 1– < 5, 5– < 10, 10– < 15 years) while reflecting local paediatric service coverage up to 18 years [9].
For each hospital, CT protocol, and age group, the hospital's typical value was defined as the median CTDIvol and DLP across eligible patients. Proposed national DRLs were then defined as the 75th percentile of these hospital-specific values across participating centres for each protocol and age group.
Inclusion and exclusion criteria: Examinations were included if (i) patient age at scanning was ≤ 18 years; (ii) the CT was performed between December 2023 and September 2025; (iii) the RIS/PACS examination label matched one of the six predefined protocols (brain, chest, abdomen–pelvis, chest–abdomen–pelvis, sinuses, and contrast-enhanced neck soft tissue); and (iv) CTDIvol and DLP were available. Records were excluded if dose indices were missing or the examination did not match the predefined protocol categories. Where multiple eligible CT examinations were identified for the same patient within the study period, only the first eligible examination was retained to avoid repeated measures. Patient weight, clinical indication, and sedation status were not consistently available across centres and were therefore not used for stratification.
Prior to analysis, extracted dose indices and acquisition parameters were screened for completeness and plausibility using standardised variable definitions across sites. Outlying or implausible values were flagged and verified against the original PACS record and/or scanner dose report to confirm protocol classification and dose indices, and records with incomplete dose data were excluded.
All statistical analyses were performed using IBM SPSS Statistics (version 31; IBM Corp., Armonk, NY, USA), and figures were generated using GraphPad Prism (version 10; GraphPad Software, San Diego, CA, USA). As dose indices were not assumed to follow a normal distribution, non-parametric hypothesis testing was applied. Inter-group differences in CTDIvol and DLP were assessed using the Kruskal–Wallis H test (independent samples), with statistical significance defined as p < 0.05 (two-sided). For each CT protocol (and analysed separately for CTDIvol and DLP), two sets of Kruskal–Wallis tests were undertaken: (i) inter-hospital variation within each age group (Hospitals A–F compared for each age category); and (ii) inter-age-group variation within each hospital (four age categories compared within each hospital). As the Kruskal–Wallis test indicates whether at least one group differs, results were interpreted primarily from group medians and interquartile ranges; pairwise post-hoc testing was not performed. To investigate technical contributors to dose variation, univariable linear regression analyses were performed in SPSS with CTDIvol and DLP as dependent variables and kVp, mAs, pitch, and slice thickness as individual predictors. Model performance was summarised using the coefficient of determination (R2), and statistical significance was assessed at p < 0.05.
To contextualise the proposed Jordanian DRLs, the derived 75th percentile CTDIvol and DLP values were compared descriptively with published paediatric CT DRLs from Switzerland, Germany, South Korea, and the United Kingdom, acknowledging differences in methodology, scanner technology, and year of data collection.
Dose metrics
Two standard dosimetric indicators were considered:
- CTDIvol (mGy): This metric reflects the average radiation output delivered within the scanned volume and is commonly used to characterise scanner performance during helical CT. CTDIvol is derived by dividing the weighted CT dose index (CTDIw) by the pitch, as shown in Eq. (1) (AAPM Report No. 96 [10]):
where CTDIw denotes the weighted CT dose index, and pitch is defined as the table travel per gantry rotation divided by the total collimated beam width [10].
1
This parameter reflects the overall radiation burden to the patient by accounting for both the scanner output (CTDIvol) and the length of the body region covered [10].
Results
Patient and examination characteristics
A total of 3794 paediatric patients were included in this study, of whom 56.7% were male (n = 2152) and 43.3% were female (n = 1642). The sample was stratified into four age groups: < 1 year (29.5%), 1–4 years (21.5%), 5–10 years (23.1%), and 11–18 years (25.9%). The distribution of patients across the six participating hospitals was generally balanced, ranging from 12.0% to 23.9% per centre. Details of the patient demographics and hospital contributions are summarised in Table 1.
Table 1.
Distribution of patients by hospital, sex, and age group
| Variable | Category | Frequency (%) |
|---|---|---|
| Hospital | A | 640 (16.9%) |
| B | 636 (16.8%) | |
| C | 643 (16.9%) | |
| D | 516 (13.6%) | |
| E | 905 (23.9%) | |
| F | 454 (12.0%) | |
| Sex | Male | 2152 (56.7%) |
| Female | 1642 (43.3%) | |
| Age Group | < 1 year | 1118 (29.5%) |
| 1–4 years | 817 (21.5%) | |
| 5–10 years | 876 (23.1%) | |
| 11–18 years | 983 (25.9%) |
The age distribution by hospital is shown in Table 2, while Table 3 summarises the number of CT examinations performed across anatomical regions and age groups. This distribution supports the representativeness of the dataset and ensures sufficient coverage to establish reliable DRLs.
Table 2.
Distribution of paediatric patients by hospital and age group
| Hospital | < 1 year | 1–4 years | 5–10 years | 11–18 years | Total |
|---|---|---|---|---|---|
| A | 160 | 160 | 160 | 160 | 640 |
| B | 182 | 126 | 157 | 171 | 636 |
| C | 161 | 167 | 154 | 161 | 643 |
| D | 232 | 93 | 94 | 97 | 516 |
| E | 279 | 164 | 197 | 265 | 905 |
| F | 104 | 107 | 114 | 129 | 454 |
| Total | 1118 | 817 | 876 | 983 | 3794 |
Table 3.
Distribution of CT examinations by anatomical region and age group
| Examination | < 1 year | 1–4 years | 5–10 years | 11–18 years | Total |
|---|---|---|---|---|---|
| Brain | 441 | 309 | 299 | 371 | 1420 |
| Chest | 179 | 104 | 96 | 103 | 482 |
| Neck Soft Tissue (w/c) | 123 | 116 | 112 | 111 | 462 |
| Sinuses | 117 | 114 | 124 | 117 | 472 |
| Abdomen–Pelvis | 136 | 88 | 122 | 140 | 486 |
| Chest–Abdomen–Pelvis | 122 | 86 | 123 | 141 | 472 |
| Total | 1118 | 817 | 876 | 983 | 3794 |
*(w/c): with contrast media
Dose metrics and proposed DRLs
The proposed DRLs for each CT examination were defined as the 75th percentile of the distribution of institutional (hospital-level) median CTDIvol and DLP values across the six participating hospitals for each protocol and age group. Accordingly, Table 4 reports the 25th percentile, median, and 75th percentile of these institutional medians, with the 75th percentile representing the proposed national DRL. Pooled patient-level dose distributions are presented separately in Figs. 1 and 2 to illustrate within-group variability and outliers.
Table 4.
Quartiles of institutional (hospital-level) median CTDIvol (mGy) and DLP (mGy∙cm) values across six hospitals, by protocol and age group (75th percentile = proposed national DRL)
| CT protocol | Age group | CTDIvol (mGy) | DLP (mGy∙cm) | ||||
|---|---|---|---|---|---|---|---|
| 25th | Median | 75th | 25th | Median | 75th | ||
| Brain | < 1 year | 31.82 | 37.51 | 47.88 | 572.94 | 683.27 | 836.41 |
| 1–4 years | 38.16 | 50.93 | 54.79 | 726.05 | 875.06 | 1011.05 | |
| 5–10 years | 38.82 | 51.09 | 65.02 | 820.60 | 1009.50 | 1123.07 | |
| 11–18 years | 40.27 | 51.70 | 60.21 | 743.80 | 1061.80 | 1189.70 | |
| Chest | < 1 year | 3.79 | 5.17 | 5.63 | 74.00 | 86.00 | 126.00 |
| 1–4 years | 3.00 | 3.90 | 5.63 | 76.10 | 117.55 | 207.45 | |
| 5–10 years | 3.67 | 4.79 | 6.40 | 122.39 | 185.05 | 261.95 | |
| 11–18 years | 3.95 | 5.30 | 7.34 | 144.40 | 198.90 | 275.90 | |
| Neck Soft Tissue (w/c) | < 1 year | 3.48 | 4.60 | 6.11 | 74.80 | 129.40 | 176.30 |
| 1–4 years | 3.58 | 4.51 | 5.81 | 74.17 | 130.80 | 179.55 | |
| 5–10 years | 3.68 | 5.21 | 6.69 | 80.90 | 134.60 | 201.90 | |
| 11–18 years | 3.62 | 4.97 | 6.14 | 83.00 | 132.50 | 189.80 | |
| Sinuses | < 1 year | 0.99 | 1.90 | 2.86 | 17.35 | 38.80 | 63.25 |
| 1–4 years | 1.38 | 2.00 | 2.70 | 17.35 | 35.35 | 60.55 | |
| 5–10 years | 1.29 | 2.10 | 2.79 | 17.57 | 38.00 | 56.80 | |
| 11–18 years | 1.30 | 1.88 | 2.83 | 14.85 | 35.80 | 63.80 | |
| Abdomen–Pelvis | < 1 year | 4.06 | 6.11 | 8.88 | 105.07 | 246.35 | 345.10 |
| 1–4 years | 4.57 | 6.22 | 8.02 | 186.60 | 289.90 | 373.83 | |
| 5–10 years | 5.36 | 7.46 | 9.70 | 264.85 | 370.95 | 540.92 | |
| 11–18 years | 6.60 | 9.52 | 16.12 | 316.55 | 538.95 | 689.27 | |
| Chest–Abdomen–Pelvis | < 1 year | 5.49 | 7.50 | 12.65 | 274.72 | 437.60 | 571.32 |
| 1–4 years | 6.04 | 7.48 | 9.52 | 306.98 | 451.40 | 630.92 | |
| 5–10 years | 6.84 | 9.32 | 11.61 | 365.95 | 592.10 | 728.50 | |
| 11–18 years | 6.89 | 9.80 | 16.12 | 474.90 | 677.20 | 875.45 | |
Fig. 1.

Box plots of CTDIvol (mGy) for six paediatric CT examinations by age group, combining data from all hospitals. The plots display the median, interquartile range, and full range of CTDIvol values for each protocol. Whiskers represent the minimum and maximum values; outlying values were reviewed against source PACS/RIS records prior to analysis
Fig. 2.

Box plots of DLP (mGy·cm) for six paediatric CT examinations by age group, combining data from all hospitals. The plots display the median, interquartile range, and full range of DLP values for each protocol. Whiskers represent the minimum and maximum values; outlying values were reviewed against source PACS/RIS records prior to analysis
Notable increases in radiation dose with patient age were observed for most examinations. For example, in abdomen–pelvis CT, the institutional median CTDIvol increased from 6.11 mGy in infants (< 1 year) to 9.52 mGy in adolescents (11–18 years), while the corresponding institutional median DLP values rose from 246.35 to 538.95 mGy·cm. Similarly, for brain CT, the median of institutional median CTDIvol increased from 37.51 mGy in infants (< 1 year) to 51.70 mGy in adolescents (11–18 years), with a corresponding increase in the median institutional DLP from 683.27 to 1061.80 mGy·cm.
In chest CT, the institutional median CTDIvol increased moderately with age (from 5.17 to 5.30 mGy), whereas DLP showed a more notable increase (from 86.00 to 198.90 mGy cm). The chest–abdomen–pelvis protocol recorded the highest overall DLP values among all examinations, with a median DLP of 437.60 mGy·cm in infants and 677.20 mGy·cm in adolescents. Median CTDIvol for chest–abdomen–pelvis increased from 7.50 to 9.80 mGy.
In contrast, dose values for neck soft tissue and sinuses CT remained relatively stable across age groups. The proposed DRLs (75th percentile values) for sinuses CT were consistent, ranging from 2.70 to 2.86 mGy for CTDIvol and from 56.80 to 63.80 mGy·cm for DLP.
These findings support the establishment of national DRLs for six commonly performed paediatric CT protocols in Jordan and highlight the importance of age-specific dose optimisation, particularly in higher-dose examinations such as brain, abdomen–pelvis, and chest–abdomen–pelvis.
Dose variation
Inter-hospital (Hospitals A–F) and within-hospital (between age categories within the same hospital) comparisons, assessed using Kruskal–Wallis tests, demonstrated statistically significant variability in CTDIvol and DLP for several protocols. For clarity, inter-hospital p-values refer to comparisons across Hospitals A–F performed separately within each age group, whereas within-hospital p-values refer to comparisons across age groups performed separately within each hospital.
Overall, the largest between-hospital differences were observed for brain, chest, abdomen–pelvis, and chest–abdomen–pelvis examinations.
For brain CT, inter-hospital differences in CTDIvol and DLP were statistically significant in every age group (all p < 0.001). Among infants (< 1 year), the institutional median CTDIvol ranged from 28.24 mGy (Hospital D) to 58.80 mGy (Hospital F), and the institutional median DLP ranged from 399.7 mGy·cm (Hospital D) to 1030.85 mGy·cm (Hospital A). Within-hospital age-group comparisons indicated significant differences across age categories at Hospitals B, D, E and F (p < 0.001), whereas age-related differences were not statistically significant at Hospitals A and C (CTDIvol: p = 0.298 and p = 0.632; DLP: p = 0.544 and p = 0.333). This corresponds to an approximately twofold difference between hospitals in infant brain CTDIvol and an approximately 2.6-fold difference in infant brain DLP, indicating clinically meaningful variation in typical practice.
In abdomen–pelvis CT, inter-hospital variation was statistically significant for both CTDIvol and DLP in every age group (all p < 0.001). For adolescents (11–18 years), institutional median CTDIvol values ranged from 5.80 mGy (Hospital F) to 16.13 mGy (Hospital B), while institutional median DLP values ranged from 247.3 mGy·cm (Hospital F) to 767.0 mGy·cm (Hospital B). Within-hospital age-group comparisons were significant at Hospitals A, B, D, E and F (p ≤ 0.002), but not at Hospital C (CTDIvol: p = 0.733; DLP: p = 0.164), indicating inconsistent age adaptation between centres.
For chest CT, inter-hospital variation was statistically significant across age groups for CTDIvol (p = 0.006) and for DLP (all p < 0.001). Hospital F reported the highest institutional medians across multiple age categories (e.g., infants: CTDIvol 30.48 mGy; DLP 553.10 mGy·cm). Within-hospital age-group comparisons were statistically significant for CTDIvol at all hospitals (p ≤ 0.025). For DLP, age-group differences were statistically significant at Hospitals A, B, C, E and F (p ≤ 0.007) but not at Hospital D (p = 0.380).
Chest–abdomen–pelvis CT exhibited substantial between-hospital differences, with statistically significant inter-hospital variation for both CTDIvol and DLP in every age group (all p < 0.001). In infants (< 1 year), institutional median CTDIvol ranged from 3.06 mGy (Hospital E) to 60.33 mGy (Hospital F), while institutional median DLP ranged from 69.68 mGy·cm (Hospital E) to 587.6 mGy·cm (Hospital C). Within-hospital age-group comparisons were statistically significant for CTDIvol in four hospitals (A, B, D and F; p ≤ 0.014) and for DLP in five hospitals (A, B, D, E and F; p ≤ 0.044), with Hospitals C and E showing no statistically significant age-group differences in CTDIvol, and Hospital C also showing no statistically significant age-group differences in DLP. The infant chest–abdomen–pelvis protocol showed the greatest between-centre spread, with an approximately 20-fold difference in institutional median CTDIvol across hospitals, reinforcing the need to standardise protocol in smaller children.
For neck soft tissue CT (with contrast), inter-hospital differences were not statistically significant across age groups for CTDIvol or DLP (p = 0.192 and p = 0.694), and within-hospital age-group comparisons were also not statistically significant (all p > 0.05). This suggests comparatively consistent protocol implementation for this examination across participating centres.
For sinuses CT, inter-hospital variation was not statistically significant for CTDIvol or DLP (p = 0.632 and p = 0.901, respectively), and within-hospital age-group comparisons were also not statistically significant (all p > 0.05). Dose values for this protocol were comparatively low relative to body CT examinations.
Collectively, these findings indicate marked between-centre variability for several high-dose protocols, reinforcing the need for protocol harmonisation and ongoing dose optimisation.
To further illustrate the variation in dose by age group in the pooled national dataset, Figs. 1 and 2 present box plots of CTDIvol and DLP values, respectively, across the four paediatric age categories for each CT protocol. These plots demonstrate age-related increases in DLP for body protocols (e.g., chest, chest–abdomen–pelvis, and abdomen–pelvis), with greater variability observed in body examinations. As DLP incorporates both CTDIvol and scan length (Eq. 2), age-related increases in DLP may reflect changes in scan range and/or scanner output.
Acquisition parameters and predictors
Substantial variation was observed across hospitals, reflecting differences in scanner models, protocol design, and radiologist preferences. For example, the kVp ranged from 76 to 140 kVp across sites, while mAs values spanned a wide range (e.g., 10–392 mAs for brain CT at Hospital D). Similarly, pitch values varied from 0.2 to 1.7, and slice thickness ranged from 1.0 to 7.0 mm, depending on the protocol and hospital.
Notably, even for the same examination, inter-institutional variability in scan settings was pronounced. For example, abdomen–pelvis CT protocols ranged from 1.0 to 7.0 mm in slice thickness, 42–583 mAs, and pitch values from 0.2 to 1.2, as presented in Table 5.
Table 5.
Ranges of acquisition parameters across CT protocols and hospitals
| CT examination | kVp | mAs | Pitch factor | FOV (mm) | Slice thickness |
|---|---|---|---|---|---|
| Hospital A | |||||
| Brain | 100–120 | 201–397 | 0.901–1.200 | 120–199 | 2.5–3.0 |
| Chest | 80–120 | 61–198 | 0.901–1.19 | 161–239 | 2.0–2.5 |
| Neck Soft Tissue (w/c) | 80–120 | 106–344 | 0.291–0.902 | 140–199 | 1–1.5 |
| Sinuses | 80–120 | 20–99 | 0.903–1.19 | 120–198 | 2.0–3.0 |
| Abdomen–Pelvis | 80–120 | 101–348 | 0.295–0.904 | 160–247 | 2.5–3.0 |
| Chest–Abdomen–Pelvis | 80–120 | 249–399 | 0.904–1.198 | 183–278 | 2.0–3.0 |
| Hospital B | |||||
| Brain | 100–140 | 100–448 | 0.2–1.01 | 123–586 | 1.0–5.0 |
| Chest | 80–120 | 26–399 | 0.55–1.20 | 115–450 | 1.0–5.0 |
| Neck Soft Tissue (w/c) | 80–120 | 102–275 | 0.902–1.094 | 121–184 | 1.0–5.0 |
| Sinuses | 80–120 | 21–111 | 1.01–1.452 | 126–188 | 2.0–3.0 |
| Abdomen–Pelvis | 76–120 | 103–339 | 0.2–1.2 | 123–586 | 1.0–5.0 |
| Chest–Abdomen–Pelvis | 100–120 | 110–400 | 0.55–1.2 | 122–770 | 1.0–5.0 |
| Hospital C | |||||
| Brain | 100–120 | 182–379 | 0.907–1.199 | 130–189 | 2.5–3.0 |
| Chest | 80–120 | 70–179 | 0.911–1.19 | 190–277 | 2.0–3.0 |
| Neck Soft Tissue (w/c) | 80–120 | 90–318 | 0.901–1.20 | 140–199 | 1.0–1.50 |
| Sinuses | 80–120 | 25–89 | 0.902–1.19 | 130–189 | 2.0–3.0 |
| Abdomen–Pelvis | 80–120 | 122–319 | 0.908–1.198 | 170–257 | 2.50–3.0 |
| Chest–Abdomen–Pelvis | 100–120 | 160–347 | 0.905–1.196 | 190–277 | 2.0–3.0 |
| Hospital D | |||||
| Brain | 80–120 | 10–392 | 0.50–1.70 | 230–333 | 1.0–7.0 |
| Chest | 100–120 | 10–165 | 0.55–1.20 | 80–420 | 1.0–5.0 |
| Neck Soft Tissue (w/c) | 80–100 | 82–159 | 0.903–1.09 | 141–199 | 1.0–1.50 |
| Sinuses | 100–120 | 10–165 | 0.55–1.20 | 80–420 | 1.0–5.0 |
| Abdomen–Pelvis | 80–140 | 42–281 | 0.55–1.20 | 98–546 | 1.0–7.0 |
| Chest–Abdomen–Pelvis | 80–120 | 17–361 | 0.55–1.20 | 80–675 | 1.0–5.0 |
| Hospital E | |||||
| Brain | 100–120 | 182–379 | 0.907–1.19 | 130–189 | 2.5–3.0 |
| Chest | 80–120 | 70–179 | 0.91–1.199 | 160–227 | 2.0–2.5 |
| Neck Soft Tissue (w/c) | 80–120 | 90–318 | 0.901–1.20 | 140–199 | 1.0–1.50 |
| Sinuses | 80–120 | 130–189 | 0.902–1.191 | 130–189 | 2.0–3.0 |
| Abdomen–Pelvis | 80–120 | 122–319 | 0.908–1.198 | 170–257 | 2.5–3.0 |
| Chest–Abdomen–Pelvis | 100–120 | 160–347 | 0.905–1.196 | 190–277 | 3.0–3.0 |
| Hospital F | |||||
| Brain | 100–120 | 37–280 | 0.5–1.5 | 105–451 | 1.0–5.0 |
| Chest | 100–120 | 110–365 | 0.5–1.00 | 200–723 | 1.0–5.0 |
| Neck Soft Tissue (w/c) | 100–120 | 200–349 | 0.901–1.09 | 140–199 | 1.0–1.5 |
| Sinuses | 100–120 | 70–119 | 1.20–1.5 | 140–199 | 2.0–3.0 |
| Abdomen–Pelvis | 100–120 | 30–320 | 0.5–1.00 | 192–580 | 2.0–5.0 |
| Chest–Abdomen–Pelvis | 100–120 | 30–320 | 0.5–1.00 | 192–580 | 2.0–5.0 |
All four acquisition parameters were significantly associated with dose metrics (p ≤ 0.001). The R2 from univariable regression analyses are presented in Table 6.
Table 6.
Univariable linear regression (R2) between acquisition parameters (kVp, mAs, pitch, slice thickness) and CTDIvol/DLP
| Predictor | R2 | P-value | |
|---|---|---|---|
| CTDIvol | mAs | 0.473 | < 0.001 |
| Pitch | 0.561 | < 0.001 | |
| Slice thickness | 0.597 | < 0.001 | |
| kVp | 0.603 | < 0.001 | |
| DLP | mAs | 0.562 | < 0.001 |
| Pitch | 0.616 | < 0.001 | |
| Slice thickness | 0.626 | 0.001 | |
| kVp | 0.630 | 0.001 |
Among the predictors, kVp demonstrated the strongest relationship with both CTDIvol (R2 = 0.603) and DLP (R2 = 0.630), followed closely by slice thickness and pitch. While mAs had a weaker association, it remained a statistically significant predictor.
These findings indicate that variability in acquisition protocols was associated with variation in patient dose indices (or contributed to variation and support the need for national standardisation of acquisition parameters. Optimising kVp selection, limiting excessive slice thickness, and avoiding unnecessarily low pitch values may help reduce dose without compromising diagnostic quality.
International comparison of DRLs
As presented in Tables 7 and 8, dose values showed considerable inter-country variability, particularly for high-dose protocols such as brain and abdomen–pelvis CT. For example, the CTDIvol for brain CT in the 5–10 years group was 65.02 mGy in Jordan, compared to 40 mGy in Switzerland and 44.1 mGy in South Korea. Similarly, the abdomen–pelvis DLP for the 11–18 years group was 689.27 mGy·cm in Jordan, compared with 780 mGy·cm reported for South Korea; however, UK national paediatric NDRLs do not provide paediatric abdomen–pelvis CT values, and the German paediatric DRWs report CTDIvol with indicative scan lengths (i.e., paediatric DLP is not explicitly listed) [6, 11, 12].
Table 7.
International comparison of 75th percentile CTDIvol (mGy) values for paediatric CT examinations across countries and age groups. The bracketed numbers (e.g., [8, 9]) indicate the reference sources from which the DRL values for each country were obtained
| CT examination | Age group | Switzerland [13] | Germany [12] | South Korea [14] | United Kingdom [6, 11] | Jordan (current study) |
|---|---|---|---|---|---|---|
| Brain | < 1 year | 20 | 25 | 39.1 | 19 | 47.88 |
| 1–4 years | 30 | 34 | 41.7 | 24 | 54.79 | |
| 5–10 years | 40 | 40 | 44.1 | 36 | 65.02 | |
| 11–18 years | – | 45* | 55.3 | 46 − 47* | 60.21 | |
| Chest | < 1 year | 5 | 1.0 − 1.5* | 5.7 | Weight-banded | 5.63 |
| 1–4 years | 8 | 2.0 | 6.8 | Weight-banded | 5.63 | |
| 5–10 years | 10 | 3.0 | 9.3 | Weight-banded | 6.40 | |
| 11–18 years | – | 5.0* | 13.8 | Weight-banded | 7.34 | |
| Neck Soft Tissue (w/c) | < 1 year | – | – | – | – | 6.11 |
| 1–4 years | – | – | – | – | 5.81 | |
| 5–10 years | – | – | – | – | 6.69 | |
| 11–18 years | – | – | – | – | 6.14 | |
| Sinuses | < 1 year | – | – | – | – | 2.86 |
| 1–4 years | – | – | – | – | 2.70 | |
| 5–10 years | – | – | – | – | 2.79 | |
| 11–18 years | – | – | – | – | 2.83 | |
| Abdomen–Pelvis | < 1 year | – | – | 7.5 | – | 8.88 |
| 1–4 years | – | – | 8.9 | – | 8.02 | |
| 5–10 years | – | 4.0 | 12.4 | – | 9.70 | |
| 11–18 years | – | 7.0 | 16.6 | – | 16.12 | |
| Chest–Abdomen–Pelvis | < 1 year | – | – | – | – | 12.65 |
| 1–4 years | – | – | – | – | 9.52 | |
| 5–10 years | – | – | – | – | 11.61 | |
| 11–18 years | – | – | – | – | 16.12 |
*UK values are from UKHSA National Diagnostic Reference Levels (NDRLs) (adopted 2022). Paediatric head NDRLs are age-banded (0– < 1, 1– < 5, 5– < 10, 10– < 15 years); for 15– < 18 years, adult head CT values are applied. Paediatric chest NDRLs are weight-banded (0– < 5 kg, 5– < 15 kg, 15– < 30 kg, 30– < 50 kg, 50– < 80 kg); therefore, direct comparison to age-banded results is limited. Germany's values are from the Bundesamt für Strahlenschutz (BfS) diagnostic reference values (DRW) (dated 17 Nov 2022; published 11 Jan 2023). Paediatric CT DRWs are reported as CTDIvol with indicative scan length; DLP is not reported for paediatric CT in this publication. Age/weight categories extend to 10– < 15 years; older adolescents may be classified as adult DRWs
Table 8.
International comparison of 75th percentile DLP (mGy·cm) values for paediatric CT examinations across countries and age groups
| CT examination | Age group | Switzerland [13] | Germany [12] | South Korea [14] | United Kingdom [6, 11] | Jordan (current study) |
|---|---|---|---|---|---|---|
| Brain | < 1 year | 270 | – | 545 | 290 | 836.41 |
| 1–4 years | 420 | – | 508 | 420 | 1011.05 | |
| 5–10 years | 560 | – | 798 | 570 | 1123.07 | |
| 11–18 years | – | – | 947 | 690 − 790* | 1189.70 | |
| Chest | < 1 year | 110 | – | 53 | Weight-banded | 126.00 |
| 1–4 years | 200 | – | 121 | Weight-banded | 207.45 | |
| 5–10 years | 220 | – | 160 | Weight-banded | 261.95 | |
| 11–18 years | – | – | 473 | Weight-banded | 275.90 | |
| Neck Soft Tissue (w/c) | < 1 year | – | – | – | – | 176.30 |
| 1–4 years | – | – | – | – | 179.55 | |
| 5–10 years | – | – | – | – | 201.90 | |
| 11–18 years | – | – | – | – | 189.80 | |
| Sinuses | < 1 year | – | – | – | – | 63.25 |
| 1–4 years | – | – | – | – | 60.55 | |
| 5–10 years | – | – | – | – | 56.80 | |
| 11–18 years | – | – | – | – | 63.80 | |
| Abdomen–Pelvis | < 1 year | – | – | 196 | – | 345.10 |
| 1–4 years | – | – | 338 | – | 373.83 | |
| 5–10 years | – | – | 513 | – | 540.92 | |
| 11–18 years | – | – | 780 | – | 689.27 | |
| Chest–Abdomen–Pelvis | < 1 year | – | – | – | – | 571.32 |
| 1–4 years | – | – | – | – | 630.92 | |
| 5–10 years | – | – | – | – | 728.50 | |
| 11–18 years | – | – | – | – | 875.45 |
*See the footnote to Table 7 for details regarding UK and German DRL categorisation and comparison limitations (age- versus weight-banding, and paediatric DLP reporting)
These findings underscore the dose variability across nations, likely influenced by differences in scanner models, imaging protocols, and national dose optimisation policies. While Jordan’s proposed DRLs for several protocols fall within the international range, higher values were observed for brain CT compared with Switzerland and the UK; chest CT comparisons with the UK should be interpreted cautiously because UK paediatric chest NDRLs are weight‑banded rather than age‑banded.
Discussion
This multicentre study proposed updated national DRLs for six paediatric CT examinations in Jordan, encompassing brain, chest, abdomen–pelvis, chest–abdomen–pelvis, sinuses, and neck soft tissue protocols. By including 3794 patients across six hospitals, this work provides one of the largest multicentre datasets reported in Jordan and provides age-specific DRLs that reflect typical practice across the country. The results demonstrated clear dose escalation with patient age in body examinations, while brain and sinus CT protocols showed more stable trends. Importantly, the proposed DRLs highlight both areas of alignment with international standards and protocols where further optimisation is required, particularly in brain and chest imaging [5]. A recently published national multicentre survey focusing specifically on paediatric head CT in Jordan also reported substantial inter-hospital variability in CTDIvol and DLP and highlighted the need for national dose standardisation; our findings for brain CT are consistent with this overall national concern, while the present work extends the scope by establishing DRLs across six routine paediatric CT protocols and evaluating technical predictors of dose variation [8].
Consistent with this, inter-hospital dose variability was pronounced, especially for high-dose examinations such as brain, abdomen–pelvis, and chest–abdomen–pelvis CT. In some instances, median CTDIvol values differed more than threefold between hospitals, underscoring the lack of harmonisation in protocol design and scanner operation. Although dose increases with age were expected due to larger body sizes and longer scan ranges, the degree of variation between hospitals suggested inconsistent adjustment of exposure parameters. This finding emphasises the need for stronger national guidance and quality assurance frameworks to reduce disparities and ensure equitable radiation protection for paediatric patients [15].
Analysis of acquisition parameters confirmed that variability in scan settings contributed significantly to dose differences. Substantial variation in kVp, mAs, pitch, and slice thickness was observed, with regression analysis identifying kVp, slice thickness, and pitch as the strongest predictors of CTDIvol and DLP. Hospitals employing low pitch values and larger slice thicknesses often recorded higher dose levels. These results reinforce the importance of tailored paediatric protocols and the consistent use of dose-saving technologies (e.g., AEC and iterative reconstruction) are recognised optimisation tools; however, their utilisation could not be assessed in this dataset [16].
Comparison with international data revealed that Jordanian DRLs for abdomen–pelvis and chest–abdomen–pelvis CT were generally comparable with those reported in Switzerland, Germany, South Korea, and the United Kingdom [13, 14, 17, 18]. However, DRLs for brain CT were consistently higher than international benchmarks, exceeding values from Switzerland and South Korea by 10–20 mGy in some age groups. Similarly, chest CT doses, particularly in younger children, were elevated compared with international values. These discrepancies are likely attributable to differences in scan length, operator practices, and incomplete adaptation of protocols for smaller patients [19]. The findings highlight the need for closer alignment with international optimisation strategies and reinforce the ICRP recommendation for regular review of DRLs every three to five years [5].
Overall, the establishment of national DRLs provides a critical tool for improving radiation protection in paediatric CT. These benchmarks allow institutions to assess their performance against national norms, identify protocols where dose levels are excessive, and initiate local reviews [20]. In particular, elevated brain and chest CT values warrant targeted interventions, while the observed inter-hospital variability suggests that harmonisation of acquisition protocols should be prioritised. Training radiologists and technologists in paediatric-specific dose optimisation, together with routine audits and quality assurance, will be essential for sustaining improvements [21].
The strengths of this study include its large sample size, multicentre design, and inclusion of six different CT protocols across four age groups. The integration of regression analysis provided new insights into the technical determinants of dose variation, supporting practical recommendations for optimisation. Nonetheless, certain limitations should be noted. Effective dose and image quality were not assessed, which limited direct evaluation of clinical risk–benefit balances. Patient weight, clinical indication, and sedation status were not consistently available from PACS/RIS across all centres; therefore, analyses were age-based and did not control for these factors. Most contrast-enhanced examinations were not included because of insufficient patient numbers, which reduces the generalisability of findings to such protocols. Differences in scanner technology and availability of dose reduction features across hospitals may also have influenced results.
The proposed national paediatric DRLs provide benchmark values that can be implemented in routine quality assurance programmes. Centres with typical doses exceeding the proposed DRLs should undertake local protocol review (e.g., scan range, kVp selection, pitch, and reconstruction settings) to identify remediable contributors to elevated CTDIvol and/or DLP. Periodic re-audit is recommended to monitor the impact of optimisation and to support national harmonisation of paediatric CT practice.
International comparisons were limited by variations in methodology, sample size, and year of data collection, which complicates direct equivalence of DRLs across countries [22]. Image quality and diagnostic adequacy were not assessed; therefore, dose comparisons are not directly linked to clinical performance. Further limitation is that patient stratification was age-based rather than weight- or size-based. For trunk CT examinations, patient weight (or other size surrogates) may better reflect attenuation and protocol selection, and age bands can increase within-group variability and limit comparability with DRL systems that use weight bands. Future national DRL updates should incorporate routinely recorded patient weight and/or size surrogates (e.g., SSDE) to enable weight- and/or size-stratified DRLs for body CT protocols [9].
Conclusion and future directions
This multicentre study established national DRLs for paediatric CT in Jordan, providing benchmarks that reflect current clinical practice and identify areas for further optimisation. While many values were comparable with international standards, elevated doses in brain and chest CT examinations, along with notable inter-hospital variability, underscore the need for protocol harmonisation and stronger adherence to paediatric-specific optimisation strategies. These findings offer an evidence base to support national quality assurance programmes and ongoing initiatives aimed at improving radiation protection in paediatric imaging.
Future research should incorporate effective dose estimation and image quality assessment to balance dose optimisation with diagnostic adequacy. Expanding DRL surveys to include smaller regional and private hospitals would enhance the representativeness of national benchmarks. The development of a centralised dose registry, supported by automated dose-tracking systems, would facilitate continuous data collection and periodic DRL review in accordance with ICRP recommendations. In parallel, greater emphasis on education and training for radiologists and technologists, together with the adoption of advanced dose-reduction technologies, will be essential for achieving sustained improvements in paediatric CT practice nationwide.
Author contributions
Abdel-Baset Bani Yaseen: conceptualisation, Literature Review, Data Curation, Formal Analysis, Writing, and Original Draft. Jamie Trapp: methodology, Supervision, Writing, Review, and Editing. Davide Fontanarosa: supervision, Analysis, Writing, Review and Editing, Project Administration.
Funding
Open Access funding enabled and organized by CAUL and its Member Institutions. This research received no external funding.
Data availability
The datasets used and/or analysed during the current study are available from the corresponding author on reasonable request.
Declarations
Conflict of interest
The authors have no relevant financial or non-financial interests to disclose.
Ethical approval
This study received ethical clearance from both Australian and Jordanian ethics bodies. Approval was granted by the Queensland University of Technology (QUT) Human Research Ethics Advisory Team (Approval No. 7324), in line with the principles outlined in the National Statement on Ethical Conduct in Human Research (2007, updated 2018). A waiver of consent for retrospective access to clinical imaging data was approved under Section 2.3.10 of the National Statement. Ethical approval was also obtained from the Scientific Research Ethics Committee in Jordan (Approval No. SREC/22112167), which authorised the study's conduct at participating centres.
Consent to participate
Not applicable.
Consent for publication
Not applicable.
Footnotes
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Scholz R (1994) On the sensitivity of children to radiation. Med Glob Surviv 1:38–44 [Google Scholar]
- 2.International Commission on Radiological Protection (2007) The 2007 recommendations of the international commission on radiological protection. Ann ICRP 37(2–4):1–332. 10.1016/j.icrp.2007.10.003 [DOI] [PubMed] [Google Scholar]
- 3.Strauss KJ, Frush DP, Goske MJ (2015) Image Gently campaign: making a world of difference. Med Phys Int 3(2):94–108 [Google Scholar]
- 4.United Nations Scientific Committee on the Effects of Atomic Radiation (UNSCEAR). Sources, Effects and Risks of Ionizing Radiation: UNSCEAR 2020/2021 Report to the General Assembly, with Scientific Annexes. Volume I: Scientific Annex A—Evaluation of medical exposure to ionizing radiation. New York: United Nations; 2022. ISBN: 978‑92‑1‑139206‑7.
- 5.Vañó E, Miller DL, Martin CJ, Rehani MM, Kang K, Rosenstein M et al (2017) Diagnostic reference levels in medical imaging: ICRP publication 135. Ann ICRP 46(1):1–144. 10.1177/0146645317717209 [DOI] [PubMed] [Google Scholar]
- 6.Worrall M, Holubinka M, Havariyoun G, Hodgson K, Edyvean S, Holroyd J et al (2022) Analysis and results from a UK national dose audit of paediatric CT examinations. Br J Radiol 95(1129):20210796. 10.1259/bjr.20210796 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Rawashdeh M, Abdelrahman M, Zaitoun M, Saade C, Alewaidat H, McEntee MF (2019) Diagnostic reference levels for paediatric CT in Jordan. J Radiol Prot 39(4):1060–1073 [DOI] [PubMed] [Google Scholar]
- 8.Hamedat SF, Md Radzi Y, Salman MD, Al‑ Shipli MA (2025) Evaluation of paediatric head CT radiation dose in Jordan: a new national diagnostic reference level survey. Radiography 31(6):103158. 10.1016/j.radi.2025.103158 [DOI] [PubMed] [Google Scholar]
- 9.European Commission (2018) Radiation protection No. 185: European guidelines on diagnostic reference levels for paediatric imaging. Publications Office of the European Union, Luxembourg [Google Scholar]
- 10.American Association of Physicists in Medicine (2008) The measurement, reporting, and management of radiation dose in CT. AAPM report No. 96. AAPM, College Park (MD) [Google Scholar]
- 11.UK Health Security Agency. National Diagnostic Reference Levels (NDRLs) from 11 December 2025 [Internet]. GOV.UK; 2025 [cited 2026 Feb 24]. Available from: https://www.gov.uk/government/publications/diagnostic-radiology-national-diagnostic-reference-levels-ndrls/ndrl.
- 12.Bundesamt für Strahlenschutz. Bekanntmachung der aktualisierten diagnostischen Referenzwerte für diagnostische und interventionelle Röntgenanwendungen (vom 17. November 2022). Bundesanzeiger. 2023 Jan 11. [DOI] [PubMed]
- 13.Verdun FR, Gutierrez D, Vader JP, Aroua A, Alamo‑ Maestre LT, Bochud F et al (2008) CT radiation dose in children: a survey to establish age‑based diagnostic reference levels in Switzerland. Eur Radiol 18(9):1980–1986 [DOI] [PubMed] [Google Scholar]
- 14.Hwang JY, Do KH, Yang DH, Cho YA, Yoon HK, Lee JS et al (2015) A survey of pediatric CT protocols and radiation doses in South Korean hospitals to optimize the radiation dose for pediatric CT scanning. Medicine (Baltimore) 94(50):e2146. 10.1097/MD.0000000000002146 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Adelodun MO, Anyanwu EC (2024) Global standards in radiation safety: a comparative analysis of healthcare regulations. Int J Eng Inventions 13(9):216–227 [Google Scholar]
- 16.Sackey TA (2022) Paediatric patients dose optimisation and risk assessment in computed tomography examination [thesis]. University of Cape Coast, Cape Coast (Ghana) [Google Scholar]
- 17.Pokora R, Krille L, Dreger S, Lee C, Günster C, Zeeb H et al (2016) Computed tomography in Germany: results and insights from a cohort study and health insurance data (AOK). Dtsch Arztebl Int 113(43):721–728. 10.3238/arztebl.2016.0721 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Shrimpton PC, Hillier MC, Lewis MA, Dunn M (2006) National survey of doses from CT in the UK: 2003. Br J Radiol 79(948):968–980. 10.1259/bjr/93277434 [DOI] [PubMed] [Google Scholar]
- 19.Stecker MS, Balter S, Towbin RB, Miller DL, Vañó E, Bartal G et al (2009) Guidelines for patient radiation dose management. J Vasc Interv Radiol 20(7 Suppl):S263–S273. 10.1016/j.jvir.2009.04.037 [DOI] [PubMed] [Google Scholar]
- 20.Demb J, Chu P, Nelson T, Hall D, Seibert A, Lamba R et al (2017) Optimizing radiation doses for computed tomography across institutions: dose auditing and best practices. JAMA Intern Med 177(6):810–817. 10.1001/jamainternmed.2017.0445 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Papadakis A, Stratakis J, Tsapaki V. (2019) Optimisation of examinations performed on paediatric patients. In: Damilakis J, editor. Radiation Dose Management of Pregnant Patients, Pregnant Staff and Paediatric Patients: Diagnostic and Interventional Radiology. Bristol (UK): IOP Publishing; p. 7‑1–7‑32.
- 22.Rehani MM (2015) Limitations of diagnostic reference level (DRL) and introduction of acceptable quality dose (AQD). Br J Radiol 88(1048):20140344. 10.1259/bjr.20140344 [DOI] [PMC free article] [PubMed] [Google Scholar]