ABSTRACT
This review traces the first 20 years of Orbitrap mass spectrometry as a mainstream high‑resolution and accurate‑mass (HR/AM) technology. It outlines the historical development of the Orbitrap analyzer, the evolution of major instrument families, and the key technological innovations that enabled its widespread adoption. Particular emphasis is placed on hybrid and Tribrid architectures, democratization of HR/AM through benchtop platforms, extension to high‑mass and native analysis, and integration with ion mobility, advanced fragmentation methods, and emerging applications such as structural biology, isotope ratio measurements, and space research. The review concludes with a perspective on future directions and the anticipated role of Orbitrap instrumentation as the core workhorse in analytical laboratories worldwide.
Abbreviations
- ADMET
absorption, distribution, metabolism, excretion, and toxicity
- AGC
automatic gain control
- AIF
all‑ion fragmentation
- AP
atmospheric pressure
- APPI
atmospheric‑pressure photoIonization
- ASAP
atmospheric solids analysis probe
- Astral
ASymmetric TRAck Lossless (analyzer)
- CCS
collision cross‐section(s)
- CDMS
charge detection mass spectrometry
- CI
chemical Ionization
- CID
collision‑induced dissociation
- CryoEM
cryogenic electron microscopy
- DART
direct analysis in real time
- DDA
data‑dependent acquisition
- DESI
desorption electrospray ionization
- DIA
data‑independent acquisition
- DMT
direct mass technology
- ECD
electron‐capture dissociation
- EDR
enhanced dynamic range
- eFT
Enhanced Fourier Transform
- EI
electron impact
- ETciD
electron‑transfer/collision‑induced dissociation
- ETD
electron transfer dissociation
- EThcD
electron‑transfer/higher‑energy collision dissociation
- FAIMS
field asymmetric waveform ion mobility spectrometry
- FSDM
phase‑constrained signal deconvolution method
- FT ICR
fourier transform ion cyclotron resonance
- GC/MS
gas chromatography–mass spectrometry
- HCD
higher‑energy collision dissociation
- HCTT
high‐capacity transfer tube
- HR/AM
high resolution and accurate mass
- ICD
independent charge detector
- ICP
inductively coupled plasma
- IMS
ion mobility spectrometry
- IRM
ion routing multipole
- IRMPD
infrared multiphoton dissociation
- LC
liquid chromatography
- LDTD
laser diode thermal desorption
- LIFDI
liquid injection field desorption ionization
- LS‐APGD
liquid sampling–atmospheric‐pressure glow discharge
- MALDESI
matrix‐assisted laser desorption electrospray ionization
- MALDI
matrix‑assisted laser desorption/ionization
- MAM
multi‑attribute method
- MION
multi‐scheme chemical ionization
- MS
mass spectrometry or mass spectrometer
- OLED
organic light‐emitting diode(s)
- OTMS
orbitrap mass spectrometry
- PCB
printed circuit board
- PFAS
per‐ and polyfluoroalkyl substances
- PRM
parallel reaction monitoring
- PTCR
proton transfer charge reduction
- PTM
post‐translational modification
- QqQ
triple quadrupole
- REMPI
resonance‑enhanced multiphoton ionization
- RF
radiofrequency
- ROI
region(s) of interest
- RTS
real‐time search
- SICRIT
soft ionization by chemical reaction in transfer
- SID
surface‐induced dissociation
- SIM
selected ion monitoring
- SLIM
structures for lossless ion manipulations
- SPS
synchronous precursor selection
- S/N
signal‑to‑noise
- TMT
tandem mass tag(s)
- TOF
time‐of‐flight
- TRL
technology readiness level
- UHPLC
ultra‑high performance liquid chromatography
- UHV
ultra‑high vacuum
- UVPD
ultraviolet photodissociation
1. Introduction
This year marks the 20th anniversary of Orbitrap™ Mass Spectrometry (OTMS) as a mainstream technology for high‑resolution, accurate‑mass (HR/AM) analysis. Anniversaries invite both celebration and reflection, and this review aims to examine where OTMS originated, where it stands today, and where it may develop in the future.
Two decades already constitute a meaningful historical timescale in our rapidly evolving world. It is therefore fitting to borrow an analytical framework from one of my hobbies—the study of civilizations.
“The Anna Karenina principle” was introduced by Jared Diamond (Diamond 1994) and later generalized in his book on human societies (Diamond 1997). The name of the principle derives from the famous first sentence of Tolstoy's great novel “Anna Karenina” (Tolstoy 2001): “Happy families are all alike; every unhappy family is unhappy in its own way.”
As I always struggled with understanding this sentence, Diamond's interpretation immediately caught my attention: for a marriage to succeed, a wide range of conditions must be simultaneously satisfied‐mutual attraction, agreement about money, child discipline, religion, in‐laws, and other vital issues. Failure in any one of those dimensions can doom the marriage even if it has all the other ingredients.
This concept generalizes well to many areas of life. People often seek simple, single‑factor explanations for success, but in reality, numerous potential failure points must all be avoided. On the pessimistic side, the principle implies that “no single property guarantees success, but many guarantee failure” (Diamond 1997).
Innovation follows a similar pattern: “Successful innovations are all alike; every failed innovation fails in its own way.” This principle comes as a very important complement to the known expression attributed to R. W. Emerson: “If you build a better mousetrap the world will beat a path to your door” (Emerson 1882). In reality, a breakthrough requires much more than a superior idea (Bornmann and Marx 2012). It demands a motivated and capable development team, supportive leadership, clear performance advantages, strong manufacturing practices, effective marketing, and the ability to sell and service the product.
All this was certainly true in the Orbitrap story. Each Orbitrap instrument got a good blend of all these factors, although each time slightly or considerably different. Most importantly, it has attracted and involved a lot of good and talented individuals. It is to these people that I dedicate this review of Orbitrap instrumentation!
Given this emphasis, the review focuses on innovations in instrumentation and the first demonstrations of new instruments, rather than attempting to cover the immense universe of Orbitrap applications. This universe keeps exponentially growing as reflected by a current publication rate exceeding 60 papers per day. Although broad reviews have appeared previously (Eliuk and Makarov 2015; Hecht et al. 2019; Deschamps et al. 2023), the field has grown beyond 160,000 Orbitrap‐based publications, making comprehensive coverage in a single review no longer feasible.
2. History Prior to the Commercial Launch
The long and complicated journey towards OTMS started around 30 years ago in a fertile land of blue‐sky ideas and discussions as vividly described in (Makarov 2013; Makarov 2019). This work was fueled by necessity to secure the future of the small startup HD Technologies Ltd that became my home, and by ambition to create a next‐generation mass analyzer that could combine all the strengths of the most advanced analyzers of the time such as Fourier transform Ion Cyclotron Resonance (FT ICR), time‐of‐flight (TOF) and radiofrequency (RF) ion traps (Brunée 1987).
In view of the unparalleled mass accuracy and resolution of FT ICR, image‑current detection was chosen as the fundamental operating principle. The large superconducting magnet, however, was replaced with a combination of dynamic ion trapping in Kingdon's electrostatic trap (Kingdon 1923; Knight 1981) and an electrostatic field known from energy analyzers Korsunskii and Basakutsa (1958) and TOF instruments (Gall et al. 1986). This refined geometry enabled axially harmonic orbital motion with oscillation frequency independent of ion energy and initial position (Makarov 1999).
Named the “Orbitrap mass analyzer” (Makarov 2000), the device utilized three electrodes—one central trapping electrode and two outer detection electrodes—and offered the highest oscillation frequency and best resistance to space‑charge effects for its size and voltage. Upon acquisition of HD Technologies in January 2000, Orbitrap became a trademark of Thermo Fisher Scientific Inc.
Maturation of this scientific curiosity into a practical Orbitrap analyzer required several more significant scientific and technological advances (Makarov et al. 2013). Among them were development and stable production of ultra‐precise Orbitrap electrodes, ultra‐stable high‐voltage power supplies, ultra‐high vacuum (UHV) systems and many others. The most crucial advance, however, was the separation of the ion source from the analyzer using an RF‑only storage device acting as a pseudo‑source (Makarov et al. 2002; Hardman and Makarov 2003; Hu et al. 2005). This ultimately resulted in the development of a novel device bent like a letter C and hence named the C‐trap (Makarov et al. 2005). It was interfaced to the electrostatic trap via novel, differentially pumped planar ion optics that also included vertical shift of ion trajectory to block the line of sight.
This new Orbitrap analyzer (Figure 1) was first introduced to a wider scientific community in 2005 as a part of the commercial LTQ Orbitrap hybrid instrument. This system married the sensitivity and ease of use of a linear ion trap with the ultra‑high resolution, mass accuracy, and dynamic range characteristic of FTMS with both analyzers operating in parallel (Makarov, Denisov, Kholomeev, et al. 2006). This launch opened access to this technology for the analytical community and therefore marked the entry of OTMS into mainstream mass spectrometry.
Figure 1.

Schematic of the Orbitrap analyzer (ion optics and differential pumping not shown). Ions are injected into the RF‐only quadrupole of the C‐trap along its curved axis, where they undergo collisional cooling with bath gas and are accumulated along the axis, compressed toward the center by voltages applied to the end electrodes. At the beginning of each acquisition cycle, ions are released from the C‐trap as short packets by switching off the RF and applying a pulsed transverse electric field. These ion packets enter the analyzer while a voltage ramp is applied to the central electrode. During this ramp, the ions spread into rotating rings around the central electrode and are radially compressed until they reach an optimal orbit diameter midway between the central and outer electrodes. Because the ions are injected off the equatorial plane, the electrostatic field drives them toward the equator, initiating coherent axial oscillations along the central electrode. These axial oscillations induce image currents on the split outer electrodes, which are detected by a differential amplifier, digitized, and Fourier transformed to produce the frequency spectrum and the corresponding m/z spectrum. (Reproduced with permission of Thermo Fisher Scientific). [Color figure can be viewed at wileyonlinelibrary.com]
3. Technology Development of LTQ Orbitrap Family of Instruments
LTQ Orbitrap instrument experienced enthusiastic acceptance initially by proteomic (Olsen et al. 2005; Scigelova and Makarov 2006) and antidoping community (Thevis et al. 2005) as they were the first to appreciate the great gain in specificity enabled by superior Orbitrap mass accuracy, resolution and sensitivity in combination with MSn capability. Other fields like Absorption, Distribution, Metabolism, Excretion, and Toxicity (ADMET) studies and metabolomics followed soon (Muenster and Makarov 2006, Lim et al. 2007; Scigelova and Makarov 2009). Typically, analysis was performed using Data‑dependent Acquisition (DDA), with overview Full MS spectrum acquired by Orbitrap analyzer and selected MS/MS spectra acquired using linear trap analyzer. This success created growing demand for further technology innovations that fueled growth of Orbitrap mass spectrometry. Figure 2 presents an overview of this growth over the entire 2 decades, as shown by the timeline on the left.
Figure 2.

Evolution tree of the Orbitrap MS over the last two decades with time as the vertical dimension. The horizontal axis classifies instruments according to their analytical capabilities: MS‐only, MS2, and MSn. Key innovations in the Orbitrap analyzer itself are shown in bold, while developments in complementary technologies are indicated in italics. Instrument families are color‐coded, each red circle represents the start of a new instrument family while each diamond represents a subsequent product launch, sometimes encompassing multiple models: (1) LTQ Orbitrap MS followed by LTQ Orbitrap XL and Discovery (1a), LTQ Orbitrap MALDI (1b), Orbitrap Velos (1c), Orbitrap Elite and Velos Pro MS (1d). (2) Exactive MS followed by Q Exactive (2b), Exactive Plus (2c), Q Exactive Plus (2d), Q Exactive HF (2e), Q Exactive GC (2f), Exactive GC (2g), Q Exactive Focus (2h), Q Exactive HF‐X (2i), Q Exactive UHMR MS (2j), DMT option for Q Exactive UHMR MS (2k). (3) Orbitrap Fusion MS followed by Orbitrap Fusion Lumos (3a), Orbitrap ID‐X (3b) Orbitrap Eclipse (3c), Orbitrap IQ‐X (3d), Orbitrap Ascend MS (3e). (4) Orbitrap Exploris 480 MS followed by Orbitrap Exploris 120 and 240 (4a), Orbitrap Exploris GC and GC240 (4b), Orbitrap Exploris MX (4c), Orbitrap Exploris EFOX MS (4d). (5) Orbitrap Astral followed by Orbitrap Astral Zoom MS (5a). (6) Orbitrap Excedion Pro MS. (Reproduced with permission of Thermo Fisher Scientific). [Color figure can be viewed at wileyonlinelibrary.com]
The entire evolution tree is grouped according to analytical capabilities of the instruments, starting from MS‐only analysis, followed by MS/MS capabilities and finally MSn. Let us examine this development tree in greater detail in Sections 3, 4, 5, 6, 7, 8 as part of an exercise in instrumentation archaeology!
Following the launch of the LTQ Orbitrap MS in 2005, higher‐energy collisional dissociation (HCD) (Olsen et al. 2007) and the asymmetric C‐trap (Makarov et al. 2008) were introduced in 2007 as a part of LTQ Orbitrap XL MS. HCD in an RF‐only gas‐filled multipole enabled improved analysis of such important posttranslational modification (PTM) as phosphorylation. In parallel, a lower performance LTQ Orbitrap Discovery MS was launched, for the first time attempting to segment the market into low‐ and high‐tiers.
Behind these dry facts hides an emotional story of successes and failures in the Bremen factory. Unknown to the general public, the unexpected skyrocketing of demand exposed the lack of automation in instrument testing and reduction of Orbitrap manufacturing yield with increasing volumes, sometimes resulting in production stops to search for the root‐cause. This overloaded the small project team, with the pace of its growth not even closely following the expansion of work. This, in turn, has led to taking shortcuts during new product development carried out by the same team in parallel to production support. This revealed itself later, when a significant number of HCD cells had to be recalled in 2008 and replaced by a more elaborate and robust design.
In 2008, a dedicated LTQ Orbitrap instrument with Matrix‐Assisted Laser Desorption and Ionization (MALDI) ion source was unveiled (Strupat et al. 2009), featuring collisional cooling of MALDI ions. For many years it remained the ultimate MALDI imaging workhorse (Eliuk and Makarov 2015), offering a unique combination of high mass accuracy, resolution, dynamic range and MSn capabilities. It also enabled, for the first time, the concept of acquiring a full mass spectrum from a single laser shot. It was also used for other applications, for example for biomass analysis using laser desorption (Smith and Lee 2010).
In 2009, an integrated C‐trap and HCD cell, back‐end electron transfer dissociation (ETD) (Syka et al. 2004; McAlister et al. 2008), dual pressure linear trap and a more sensitive atmosphere‐to‐vacuum interface became part of the LTQ Orbitrap Velos™ system (Olsen et al. 2009). The integrated C‐trap and HCD cell reduced ion losses on transfer and improved sensitivity, but most of the innovation came from the brand‐new LTQ Velos front end (Pekar Second et al. 2009) which also offered a back‐end ETD option that became indispensable for analysis of labile PTMs and proteins. Sensitivity was drastically increased by the introduction of a stacked‐ring ion guide (S‐lens), a simplified ion funnel without an axial field that uses a supersonic gas jet from a round‐bore transfer tube to propel ions through the lens.
There is no surprise that rapid switching of demand almost entirely to this new instrument brought some new challenges. While the S‐lens provided a 5–10× increase in ion current and thus sensitivity, it also led to a proportional acceleration of ion‐optics contamination. It turned out that there was simply not sufficient time during the development process to test all possible consequences of this effect. More importantly, the quality of some ion‐optical elements slightly deteriorated upon scaling up—stealthily enough not to be noticed during factory testing but clearly sufficient to cause charging after months of continuous operation under high sample load. This caused quite a lot of trouble for both the development and service teams as many instruments had to be upgraded later in the field with new solutions to reduce contamination.
As Orbitrap MS gained increased adoption, new applications continued to appear. One of the most important became analysis of intact proteins such as antibodies (Bondarenko et al. 2009). For the first time detection of individual multiply charged protein ions was demonstrated (Makarov and Denisov 2009), paving the way to Orbitrap charge‐detection MS (CDMS) of the 2020s.
Having LTQ mass spectrometer as the front‐end, all instruments of this family also were naturally able to operate with different front‐end options as described in Section 10.
The pinnacle of the first‐generation Orbitrap family became the Orbitrap Elite™ instrument (Figure 3) (Michalski, Damoc, Lange, et al. 2011). It incorporated two key innovations, first being a novel high‐field Orbitrap design as shown on Figure 4 alongside the original version. It almost doubled the acquisition speed and increased sensitivity. Another key innovation, enhanced Fourier transform (eFT™) processing combined so‐called absorption mode with magnitude mode and provided another doubling of resolving power (Lange et al. 2014), Overall, this last and ultimate member of the LTQ Orbitrap family quadrupled resolution per unit time compared to previous instruments. It also incorporated a significantly higher‐temperature bakeout of the Orbitrap analyzer, enabling improved ultra‐high vacuum conditions and thereby facilitating its increasing use in top‐down protein analysis (Fornelli et al. 2012). This marked the beginning of a long journey toward subsequent breakthroughs in high‐mass Orbitrap analysis.
Figure 3.

Schematic layout and picture of a linear trap‐Orbitrap hybrid MS: (a) LTQ Orbitrap MS, (b) Orbitrap Elite MS with ETD option (reproduced with permission of Thermo Fisher Scientific). [Color figure can be viewed at wileyonlinelibrary.com]
Figure 4.

Cut‐outs of a standard (top) and a high‐field (bottom) Orbitrap analyzer (reproduced with permission of Thermo Fisher Scientific). [Color figure can be viewed at wileyonlinelibrary.com]
4. Tribrid Family of Instruments as the Next Generation MSn System
A major breakthrough along the MSn line came in 2013 with the launch of the Orbitrap Fusion™ Tribrid™ MS (Senko et al. 2013). Here a high‐field Orbitrap analyzer was combined with a segmented hyperbolic quadrupole mass filter and a dual‐pressure linear trap (Figure 5). Based on lessons learned from the LTQ Orbitrap platform, the Orbitrap analyzer in this instrument was moved to precede the linear trap, with HCD cell extending its role to a bi‐directional ion transfer and becoming an Ion Routing Multipole (IRM). The Orbitrap operating voltage was increased from 3.5 kV in Orbitrap Elite MS to 5 kV, further increasing the speed of analysis for a given resolution.
Figure 5.

Schematic layout of a Tribrid MS: (a) Orbitrap Fusion and Orbitrap ID‐X MS; (b) Orbitrap Ascend MS (reproduced with permission of Thermo Fisher Scientific). [Color figure can be viewed at wileyonlinelibrary.com]
This instrument also featured front‐end internal calibration source (Easy‐IC™ source) that could be also used for ETD (Easy‐ETD™) fragmentation. To maximize peptide/protein sequence coverage, ETD could be also combined with CID in the linear trap to yield so‐called ETciD as well as with HCD in the IRM to result in EThcD (Frese et al. 2012). This in turn enabled new applications such as top‐down sequencing (Fornelli et al. 2017) and discrimination of leucine and isoleucine in peptide sequencing (Lebedev et al. 2014).
Improved instrument control enabled truly parallel operation of all three analyzers (dynamic scan management) and new modes of operation, such as Synchronous Precursor Selection (SPS) by multinotch isolation in the linear trap. This mode became the basis for Tandem Mass Tag (TMT) MS3 workflow to overcome the issue of ratio compression (McAlister et al. 2014).
In 2015, Orbitrap Fusion Lumos™ MS pioneered the next generation atmosphere‐to‐vacuum interface. There the round‐bore transfer tube and S‐lens of Orbitrap Fusion MS was superseded by a high‐capacity transfer tube (HCTT) of slit‐like cross‐section and ion funnel. Resulting 4x increase in incoming gas flow became the decisive factor for the dramatic increase of sensitivity. The instrument also featured more elaborate pressure control in IRM, improving top‐down analysis.
Since 2017, Tribrid instruments became capable to support options of 2nd generation Field Asymmetric Waveform Ion Mobility Spectrometry (FAIMS Pro™ system) (Pfammatter et al. 2018; Melani et al. 2019; Goswami et al. 2025), Ultra‐Violet Photo‐Dissociation (UVPD) (Fornelli et al. 2020) and 1 Million mass resolution (1 M) (Denisov et al. 2012).
In 2018, a variant of the original Orbitrap Fusion was launched as Orbitrap ID‐X™ MS for small molecule analysis. It has introduced AcquireX™ data acquisition mode for automated iterative DDA analysis in metabolome profiling.
Orbitrap Eclipse™ became the next highest performance Orbitrap MS in 2019. The arsenal of fragmentation methods was further enhanced by adding Proton Transfer Charge Reduction (PTCR) using negatively charged reagent in Easy‐ETD (Huguet et al. 2019). This method became an increasingly powerful method for untangling complexity of overlapping multiply charged fragments.
Larger diameter hyperbolic‐rod quadrupole enabled narrower mass filtering and higher resistance to contamination from the deposited ion beam.
TMT and other DDA methods were improved by Real‐Time Search (RTS), in which on‐the‐fly spectral searching allowed to save time on spectra of insufficient quality (Erickson et al. 2019). Specifically for TMT, a new phase‐constrained signal deconvolution method (FSDM) (Grinfeld et al. 2017) was applied to double the speed of reporter ion measurement.
The increase of mass range of isolation and analysis has enabled the innovative method of NativeOmics, wherein MSn of protein complexes made it possible for the first time to sequence also endogenous lipids deep inside those complexes (Gault et al. 2020).
The next‐generation instrument for small molecule analysis, the Orbitrap IQ‐X™ MS, was introduced in 2021. Developed on the Orbitrap Fusion Lumos platform, it featured an Auto‐Ready ion source that allows fully automated calibration using a parallel source independent of the LC inlet.
Orbitrap Ascend MS was launched in 2022, and it signified a drastic redesign of the instrument architecture, adding an additional IRM in front of the C‐trap as a buffer for ion storage prior to MSn experiments and an Auto‐Ready ion source for automatic system calibration. The mass range of linear trap isolation was extended to 8000 m/z and Orbitrap detection‐ to 16000 m/z. The instrument also used a longer ion funnel to improve analysis of labile compounds, from phosphopeptides to metabolites (He et al. 2023).
Since 2024, this very versatile model has been offered in three fit‐for‐purpose Editions such as multiomics, structural biology and biopharma. For the latter two, quadrupole electronics was reworked to enable high‐mass isolation up to 8000 m/z, further extending instrument applications in Native MS. New methods such as DIA‐PTCR for top‐down mass spectrometry allowed to detect up to an order of magnitude more proteoforms (Schachner et al. 2024).
This ever‐ongoing Tribrid evolution enables analysis of analytes and complex mixtures which even recently have been beyond the reach of mass spectrometry as witnessed by deep analysis of membrane protein complexes (Lutomski et al. 2025; Schachner et al. 2024).
5. (Q) Exactive Family of Instruments as the First Milestone in Democratizing HR/AM Analysis
5.1. Toward Democratization of HR/AM
Although Orbitrap MS rather unexpectedly became the top tier of mass‐spectrometric instrumentation and, by the late 2000s, largely displaced the previous crown bearer FT ICR MS, it was always clear to us that the true destiny of Orbitrap technology was the democratization of HR/AM analysis.
For this reason, already 1 year after the launch of the very first LTQ Orbitrap instrument, active research began along two very different paths: the first—to make Orbitrap‐based instrumentation more affordable and easier to use, and the second—to develop an analyzer that could eventually surpass the Orbitrap as the next performance leader.
The first direction led to the (Q) Exactive™ and later the Orbitrap Exploris™ and Excedion™ product lines, while the second ultimately resulted in the emergence of the Astral™ platform in 2023.
Naturally, the first direction received the highest priority because market demand for accessible HR/AM instruments already existed. However, addressing this need required a complete redesign of the earlier Orbitrap technology. In the run‐up to the LTQ Orbitrap launch, so many aspects of the system were still unknown that even reasonably predictable components (e.g., differential pumping, thermal isolation and stabilization, ultra‐high vacuum sealings) were deliberately over‐engineered to avoid surprises during production. As we embarked on designing a cost‐efficient Orbitrap instrument, we had to re‐work and optimize the entire instrument layout.
It became apparent that the new instrument had to be benchtop, and that dedicated multistage turbopumps would be required to replace the numerous pumps used in the LTQ Orbitrap MS. The first instrument in the Exactive series included only the Orbitrap analyzer for MS‐only operation, although its mechanical layout intentionally left sufficient room for the later addition of a quadrupole mass filter. The main challenges were largely engineering‐related: we needed to demonstrate that the Orbitrap analyzer could operate reliably in an aluminum—rather than a stainless‐steel chamber, without ConFlat seals and without water‐cooling for temperature stabilization and bakeout, and that it could still detect ions in the presence of numerous electrical and mechanical interferences inside a tightly packed benchtop enclosure. These challenges were successfully resolved by a small R&D team working closely with vacuum pump vendors, although several unexpected side effects of the new layout forced us to abandon a few configurations as technological dead ends. Later, as instrument production scaled up, numerous new issues with electronic boards emerged and required multiple redesigns—changes that eventually proved indispensable for the ramp‐up to the next‐generation Orbitrap hybrids.
5.2. Main Instruments of the Q Exactive LC Family
The Exactive MS was announced as early as 2008, and shipments began in 2009 (Bateman et al. 2009) (Figure 2). In the absence of LTQ detection, it relied on Orbitrap detection for Automatic Gain Control (AGC) (Stafford et al. 1992), that is, to regulate the filling time of the C‐trap. Because the instrument acquired only full‐scan MS spectra without mass selection (Figure 6), its sole option for structural elucidation was All‐Ion Fragmentation (AIF) (Geiger et al. 2010).
Figure 6.

Schematic layout of (a) a stand‐alone Orbitrap instrument, the Exactive Plus MS; (b) quadrupole‐Orbitrap hybrid mass spectrometer, the Q Exactive MS (reproduced with permission of Thermo Fisher Scientific). [Color figure can be viewed at wileyonlinelibrary.com]
This instrument took advantage of having just a single ultra‐precise voltage to run the analyzer: as this voltage needed switching on every ion injection, it was now switched continuously between −5 and +5 kV power supplies, both being ultra‐precise and temperature‐stabilized via Peltier elements. Keeping both supplies constantly “warm” meant that polarity could be alternated from one scan to the next simply by adjusting the switching timing, without sacrificing part‐per‐million external mass accuracy. This unique capability became widely adopted in various small‐molecule screening applications (Moulard et al. 2011).
The next‐generation instrument, launched in 2011 in parallel with Orbitrap Elite MS, was the Q Exactive quadrupole–Orbitrap hybrid (Michalski, Damoc, Hauschild, et al. 2011). To ensure efficient filling of MS2 spectra, AGC was enhanced by adding an independent ion‐current measurement device in the form of an electrometer located after the HCD cell. This allowed to include into the fill time prediction, a substantial part of ion current in complex mixtures that fell below detection threshold in the Orbitrap spectra. Externally, the instrument closely resembled the Exactive, and it soon became a routine workhorse in thousands of laboratories worldwide.
Three years later, a dedicated edition of the Q Exactive MS for environmental, food safety, and toxicology applications was introduced as the Q Exactive Focus MS (Simirgiotis et al. 2016; Wang et al. 2017). Meanwhile, the Exactive MS itself was upgraded and released as the Exactive Plus MS (Rochat et al. 2012; Berube et al. 2012).
As the Q Exactive platform proved both more affordable and more productive in DDA experiments than the LTQ Orbitrap family (Kelstrup et al. 2012), it rapidly became a true workhorse for HR/AM analysis. This widespread adoption drove continual performance improvements. In 2013, the Q Exactive Plus MS introduced an improved hyperbolic quadrupole mass filter with pre‐ and postfilters, preceded by a new ion guide that utilized Printed Circuit Boards (PCBs) to generate an axial field. This configuration doubled MS/MS sensitivity and substantially extended intervals between required cleanings when analyzing complex “‐omics” mixtures. The instrument also offered higher mass resolution and an intact‐protein mode.
In 2014, the family was further extended with an ultra‐high‐field Orbitrap analyzer, similar to the one used in the Tribrid series, resulting in the Q Exactive HF MS (Scheltema et al. 2014). Then, in 2017, the system was further enhanced with a high‐capacity transfer tube and an ion funnel design similar to those on the Orbitrap Fusion Lumos MS, yielding the Q Exactive HF‐X MS (Kelstrup, Bekker‐Jensen, et al. 2018). These instruments were optimized for fast separations by Ultra‑high Performance Liquid Chromatography (UHPLC) as well as for new methods of data‐independent acquisition (DIA) (Panchaud et al. 2009) and parallel reaction monitoring (PRM) (Peterson et al. 2012). Both DIA and PRM relied on pushing the limits of high resolution and accurate mass of panoramic MS/MS spectra.
This parade of product names and milestones may give the impression of an uninterrupted march of success, but it does not adequately reflect the enormous amount of work behind the scenes—across R&D, marketing, product support, and the global sales organization. Although the democratization of HR/MS was ultimately successful, progress was repeatedly challenged by unexpected complications arising from the “chaotic” side of Mother Nature and the inherent complexity of high‐tech manufacturing.
Two particularly devastating incidents are especially memorable. The first resulted from a small, unannounced cost‐saving change introduced by a turbopump vendor in the rotor bearing design—after which UHV pumps began failing after only 1–2 years of operation. It took years to identify and correct the root cause across thousands of instruments worldwide, delivering a significant reputational blow to the brand.
The second incident arose from a similarly unannounced modification made by the supplier of the C‐trap end‐lenses in response to tightening environmental regulations. Metal‐coated ceramic lenses that had functioned flawlessly since the early LTQ Orbitrap days began to exhibit microscopic metallization defects, which charged up after weeks of continuous exposure to the intense ion beams in the Q Exactive HF‐X MS. Because this effect was easily masked by other potential sources of charging, it proved exceptionally difficult to trace and ultimately required the recall of many tens of instruments, followed by extensive field upgrades to repair the remaining systems.
Looking back, we would have spared no effort to avoid these failures, particularly in light of the substantial damage and disruption they caused for our valued and trusting users and their research!
5.3. Orbitrap MS for High‐Mass Analysis
As understanding of the motion of intact proteins within the Orbitrap analyzer deepened (Scigelova and Makarov 2009), Orbitrap instruments continued to expand their accessible mass range toward increasingly large macromolecular assemblies. A major breakthrough emerged through collaboration with the Prof. Heck's group (Rose et al. 2012; Rosati et al. 2012), which demonstrated for the first time high‐sensitivity, high‐resolution native analysis of large intact protein complexes such as GroEL. This work led to the development of the Exactive EMR instrument series, featuring Orbitrap detection extended beyond 20,000 m/z (Maple et al. 2014).
Subsequent improvements focused on widening the mass range of quadrupole isolation (Belov et al. 2013; Gault et al. 2016; Dyachenko et al. 2015), extending the mass range of the ion optics by employing lower‐frequency RF waveforms (Snijder et al. 2014), and introducing in‐source trapping to enhance desolvation and fragmentation (Van de Waterbeemd et al. 2017; Fort, Van de Waterbeemd, et al. 2018). These advancements culminated in the 2018 release of the Q Exactive UHMR MS, the final instrument in the mighty Q Exactive family.
High‐mass analysis was further advanced with the introduction of individual‐ion MS signal processing approaches (Kafader, Melani, et al. 2019; Kafader, Beu, et al. 2019; Kafader et al. 2020; Wörner et al. 2020), which were incorporated in 2022 into the Direct Mass Technology (DMT) option (Strasser et al. 2023). This represented the first commercial implementation of Charge Detection Mass Spectrometry (CDMS), previously available only in research‐grade instruments (Fuerstenau and Benner 1995; Contino et al. 2013). Compared with the linear electrostatic traps traditionally used for CDMS, the Orbitrap analyzer offers exceptionally high space‐charge capacity and an ion frequency that is independent of ion energy, enabling simultaneous CDMS measurement of hundreds, and sometimes thousands, of individual ions within a single acquisition.
6. Exploris and Excedion Instruments Supersede the Q Exactive Hybrid Family
6.1. Next Step in Democratization of HR/AM
Building on the experience gained with the Q Exactive series, it became possible to commit to a new ambitious goal of making the hybrid quadrupole–Orbitrap architecture as amenable to mass production and broad analytical utilization as the most mass‐produced tandem mass spectrometers, that is, triple quadrupoles. This effort evolved into the Orbitrap Exploris MS platform, guided by the following design principles (Hauschild et al. 2020):
More efficient use of physical space and improved serviceability, achieved largely by reducing the size of the pumping system and ion optics. A novel six‐port turbopump was a key innovation, enabling a two‐fold reduction in instrument size. Initially deemed impossible, such a pump became the first stone in the foundation of the Exploris platform, with first experiments starting as early as 2010 at Pfeiffer Vacuum GmbH.
Performance flexibility enabled by modular hardware and software components, with increased commonality shared with Tribrid instruments. This work began in earnest in the mid‐2010 s as Tribrid platform started to mature.
Greater use of electronics‐inspired technologies in ion‐optical design, particularly the implementation of PCB‐based RF ion guides.
Harmonization of interfaces across instrument families, facilitating workflow sharing and easier transfer of analytical methods. Because it uses the same ion source as the Tribrid MS platform, the Exploris family supports the same options, including the FAIMS Pro interface and the EASY‐IC internal lock‐mass ion source.
The C‐trap and Orbitrap analyzer were merged into a single monobloc assembly to improve robustness against noise and mechanical misalignment, with symmetrically supported Orbitrap electrodes providing enhanced noise cancellation.
An independent charge detector (ICD) module was placed upstream of the C‐trap to increase AGC speed and to free the rear of the HCD cell for future extensions, effectively transforming it into an ion‐routing multipole similar to that used in Tribrid systems. This design choice later proved crucial for interfacing the Exploris platform with the Astral back end.
In almost every field of hardware development, establishment of a new platform is associated with a much higher effort, risk and overall uncertainty than any, however ambitious, extension of an existing platform. This appeared to be especially true for the Exploris platform as development took longer than expected and priorities of development were adjusted “on the go” in response to changes in business priorities.
6.2. Main Instruments of Next‐Generation Orbitrap Hybrids
The high‐end Orbitrap Exploris 480 MS, featuring an HCTT/ion‐funnel interface (Figure 7), was launched in 2019 for ‐omics applications [Bekker‐Jensen et al. 2020]. Its model number reflects the maximum resolving‐power setting. Like the Orbitrap Eclipse MS, it incorporated the FSDM method, which doubles the speed of TMT reporter‐ion analysis (Grinfeld et al. 2017; Kelstrup, Aizikov, et al. 2018). With the Biopharma option, it is optimized for implementing different variants of multi‑attribute method (MAM) in biopharmaceutical analysis.
Figure 7.

Schematic layout of an Orbitrap Exploris hybrid MS: (a) Orbitrap Exploris 480 MS, (b) Orbitrap Exploris 120 and 240 MS, (c) Orbitrap Exploris GC and GC 240 MS (reproduced with permission of Thermo Fisher Scientific). [Color figure can be viewed at wileyonlinelibrary.com]
The following year, the universal Orbitrap Exploris 240 MS was introduced (Knizner et al. 2022), together with the Orbitrap Exploris 120 MS (Monaghan et al. 2021). These instruments featured a more traditional round‐bore transfer tube/S‐lens interface and maximum resolution settings reflected by the model number. They were optimized for quantitative environmental, food‐safety, toxicology, and other small‐molecule applications. This successful dual launch, at the height of the COVID‐19 lockdown, was made possible only by the timely implementation of a comprehensive remote instrument control system in the preceding years.
In 2022, the product line was complemented by the MS‐only Orbitrap Exploris MX MS, designed specifically for quality control in biopharmaceutical manufacturing. Another version of this MS‐only instrument was launched in 2025 for environmental, forensic and toxicology analysis under the name Orbitrap Exploris EFOX MS.
In 2025, the Orbitrap Exploris 480 MS was superseded by the Orbitrap Excedion Pro MS. This model added a more robust sprayer, a gentler ion‐funnel interface similar to that used in the Orbitrap Ascend MS, and—most importantly—Easy‐ETD capability together with EThcD support (Kessler et al. 2025). It also introduced the Enhanced Dynamic Range (EDR) Full‐MS scanning mode for the Orbitrap analyzer, extending the method first evaluated in (Meier et al. 2018).
6.3. Isotope Ratio Measurements
When issues with isotope accuracy were identified in early Orbitrap spectra in the late 2000s (Kaufmann and Walker 2012), dedicated internal research was initiated and accompanied by continuous improvements in Orbitrap hardware quality and signal‐processing methods. As accuracy improved, a productive collaboration was established with Prof. Eiler's group at Caltech, aiming to achieve the pro‐mille‐level isotope‐ratio precision required for the majority of applications. This goal was eventually met for intact organic molecules analyzed using both Gas Chromatography (GC) (Eiler et al. 2017; Hofmann et al. 2020; Zeichner et al. 2022) and LC separations (Mueller et al. 2021).
Building on this success, the Orbitrap Exploris Isotope Solutions package was introduced in 2022. It enables direct measurement and calculation of isotope ratios on any Exploris MS from the relative abundances of Isotopologues in an intact compound or its fragments (Hilkert et al. 2021). The solution includes a dual‐inlet unit and a flow‐injection valve, opening a new field of application for Orbitrap MS in isotope‐ratio measurement—an area previously dominated by magnetic‐sector instruments with multiple collectors.
Independently of this work, collaboration with Prof. Zubarev's group at the Karolinska Institut led to the development of a proteomics‐compatible method for isotopic‐ratio determination based on immonium fragments with baseline‐resolved isotopologues (Gharibi, Chernobrovkin, Saei, et al. 2022; Gharibi, Chernobrovkin, Eriksson, et al. 2022).
7. Astral Analyzer Enhances and Complements Orbitrap Performance
The ASymmetric TRAck Lossless (Astral) analyzer (Stewart et al. 2023) represents the culmination of many years of development that began in parallel with the Exactive series of instruments. Its most distinctive feature is a multi‐reflection analyzer formed by two opposing planar mirrors that are tilted relative to one another (Figure 8), with an ion foil positioned between them (Grinfeld et al. 2024). Ions from a quasi‐continuous beam are stored—and, if desired, fragmented—in an ion processor derived from the C‐trap but redesigned to provide much tighter spatial and temporal focusing (Stewart, Grinfeld, Wagner, et al. 2024). When ions are pulsed into the mirror assembly through prism‐based ion optics, the combined action of the mirror tilt and the ion foil reverses the ion trajectories as they bounce repeatedly between the mirrors, ultimately focusing them onto a novel high‐dynamic‐range detector. The entire system is gridless and delivers exceptionally high‐fidelity spatial and temporal focusing, making it effectively “lossless,” with ion losses far lower than those observed in any previous mass analyzer.
Figure 8.

Schematic layout of an Orbitrap Astral MS (reproduced with permission of Thermo Fisher Scientific). [Color figure can be viewed at wileyonlinelibrary.com]
The Astral analyzer achieves single‐ion sensitivity at mass resolutions up to 100,000 and ppm‐level mass accuracy, while operating at acquisition speeds that initially reached 200 spectra per second and now extend up to 270 spectra per second. Each spectrum is fully compatible with modern data‐analysis workflows, as it typically contains thousands of ions.
In the Orbitrap Astral MS, the complementary strengths of the Orbitrap and Astral analyzers are exploited to their fullest extent. The Orbitrap analyzer, offering the highest ultimate resolution and external mass accuracy, is dedicated to Full MS acquisition, whereas the Astral analyzer—with its extraordinary speed and sensitivity—is used for MS/MS spectra. Together, they form an almost ideal envelope of analyzer performance, a synergy reflected in the numerous groundbreaking achievements reported in ‐omics research (Guzman et al. 2024; Serrano et al. 2024; Lancaster et al. 2024; Petrosius et al. 2025].
In 2025 the second model, Orbitrap Astral Zoom MS (Hsu et al. 2025) was launched, offering higher acquisition rate and multipass operation for even higher resolution (Stewart, Grinfeld, Giannakopulos, et al. 2024). It is entirely plausible that, over time, the Astral analyzer may diverge from the Orbitrap lineage and establish a lineage of its own!
8. 10 Years of Orbitrap GC/MS
The 20th anniversary of Orbitrap MS coincides with another important milestone: the 10‐year anniversary of the first serial GC–Orbitrap product. As in many developments, this instrument emerged from a highly productive collaboration—this time with Prof. Coon's group at the University of Wisconsin–Madison. While developing a back‐end ETD option for the LTQ Orbitrap MS that made use of a standard electron impact (EI) source from a GC–quadrupole instrument (McAlister et al. 2008), the group revealed an opportunity to apply this configuration directly to conventional GC analytes (Peterson et al. 2010). The concept was later extended to the Q Exactive instrument (Peterson et al. 2014), demonstrating the need for the newest high‐field Orbitrap analyzer to secure a sufficient number of data points across a characteristically narrow GC peak.
Building on this foundation, the first GC/Q‐Exactive instrument with EI and CI ion sources was introduced in 2015 (Weidt et al. 2016), followed a year later by a simpler MS‐only GC/Exactive model. The combination of very high sensitivity, mass accuracy, and resolution enabled detection limits that were comparable even in Full‐MS mode to those achieved with triple‐quadrupole (QqQ) instruments operating in their most sensitive single‐reaction‐monitoring (SRM) mode (Belarbi et al. 2021).
As the (Q) Exactive family was succeeded by the Exploris line, the GC‐Orbitrap platform followed the same evolution. In 2021, the earlier instruments were replaced by the MS‐only Orbitrap Exploris GC and the hybrid Orbitrap Exploris 240 GC, which now serve as the next‐generation GC–Orbitrap systems.
It is noteworthy that GC/MS was the first field to see the entry of a new vendor, Anyeep, in 2024 (Suzhou Anyi Spectrum 2025), with the so‐called Cassitrap analyzer, which matches the standard Orbitrap geometry.
9. Understanding the Orbitrap Analyzer
Ultra‐high resolving power in the Orbitrap analyzer became achievable in practice only when a deeper understanding of nonlinear, space‐charge–dominated ion oscillations in a perturbed quadratic potential was combined with manufacturing techniques capable of producing electrode geometries with accuracies in the high‐nanometer range.
Although ion motion in the Orbitrap field appears to follow the simple equations of 18th‐century celestial mechanics (Makarov 2000), real ion packets exhibit far more complex behavior, requiring analytical tools from 21st century physics of complex systems (Bolotskikh et al. 2011; Kharchenko et al. 2012; Skoblin et al. 2015). It was found that space‐charge interactions between ion packets depend strongly on even slight deviations of the electric field from ideality, leading to phenomena such as coalescence of adjacent intense peaks (Gorshkov et al. 2012; Kaufmann and Walker 2018; Grinfeld et al. 2019). Optimal performance is achieved when the trap incorporates a specific micrometer‐scale nonlinearity in its field profile (Grinfeld et al. 2015; Grinfeld et al. 2019). As space charge pushes ion rings to larger oscillation amplitudes, carefully shaping the field in these outer regions compensates for the expansion, enabling linear corrections to space‐charge influence on ion frequency.
Space‐charge effects are even more pronounced in the C‐trap, as shown by both simulations (Grinfeld et al. 2011) and experimental studies (Gorshkov et al. 2010; Kaufmann et al. 2010). These effects can be mitigated through intelligent C‐trap filling strategies (Kaufmann and Walker 2016; Meier et al. 2018; Panepinto et al. 2025), ultimately culminating in the EDR scan introduced with the Orbitrap Excedion Pro MS.
As manufacturing tolerances advanced into the high‐nanometer regime, progressively higher resolving powers could be achieved at increasingly high coalescence thresholds—first up to one million (Denisov et al. 2012), and later to two million and beyond (Denisov et al. 2021; Deslignière et al. 2024). This progress has allowed Orbitrap MS to expand into essentially all application areas historically addressed by FT ICR MS, such as petroleomics (Zhurov et al. 2013; Schmidt et al. 2018).
Although the Orbitrap analyzer is typically used solely for the detection of injected ions, several additional operating modes are also possible. As demonstrated in (Perry et al. 2008) and references therein, the analyzer can support high‐fidelity de‐excitation and re‐excitation of specific ions. While still largely unexplored, such modes may become valuable in the future, possibly for the analysis of very large macromolecular assemblies reaching the so‐called “plateau of stability” (Wörner et al. 2022).
Although the Orbitrap analyzer represents a specific axially symmetric realization of the broader class of Cassini fields (Köster 2009), its rotational symmetry makes it much simpler and therefore much more practical compared to other members of this field's family. Compared with other electrostatic trap geometries, such as linear or toroidal traps (Makarov 2010), the Orbitrap design provides the highest oscillation frequency—and therefore the highest resolving power for a given transient duration—at a given size and ion energy. Although signal processing is often regarded as independent from hardware development, in Orbitrap MS the two are deeply intertwined. One of the key factors is the importance of a virtual common starting point for ion trajectories as a side effect of rapid extraction from the C‐trap. This allows the initial phase for each m/z to be precisely defined and calibrated, enabling a two‐fold increase in resolution via enhanced Fourier transform (eFT) processing (Lange et al. 2014) and a further doubling beyond the Fourier limit through FSDM (Grinfeld et al. 2017; Kelstrup, Aizikov, et al. 2018; Steigerwald et al. 2024). Both methods also feature improved peak shape when dead time at the beginning of the transient is minimized. This required a refinement of preamplifier electronics and replacement of dielectric Orbitrap spacers to reduce dielectric losses. Signal processing of spectra with significant Coulomb interactions or with high in‐spectrum dynamic range has led to the development of sophisticated models for Orbitrap noise (Keenan et al. 2025).
Dedicated hardware and software from Spectroswiss (Nagornov et al. 2020; Leach et al. 2024) have enabled the user community to access raw transients directly and experiment with alternative signal‐processing methods. These tools also allow commercial instruments to operate in special modes that extend transient durations beyond standard settings (Nagornov et al. 2020), supporting experiments with ultra‐long transients (Deslignière et al. 2024). When ion masses within a narrow range are known a priori, least‐squares fitting can further improve resolving power (Aushev et al. 2014; Kozhinov et al. 2023).
Collision cross‐sections (CCS) of analyte ions can be extracted directly from Orbitrap transients alongside m/z by analyzing the exponential decay of ion signals caused by gas‐phase collisions within the analyzer. Early work (Sanders et al. 2018) established the fundamental “hard‐sphere” model linking transient damping to CCS. Subsequent studies refined the data‐processing framework, extended applicability to native proteins, and introduced “soft‐sphere” or energy‐dependent models that improve accuracy for low‐charge or low‐energy ions (James et al. 2022; James et al. 2023). The method now applies to analytes ranging from small molecules (Ni and Arevalo 2024) to large proteins, with good correlation to low‐field ion‐mobility CCS measurements when pressure and ion energy are well controlled and calibration against an internal standard is used. Current efforts aim to extend CCS determination to individual‐ion signal processing.
10. Orbitrap MS Outside of Mainstream
10.1. Introduction
As mainstream commercial Orbitrap MS evolved into a powerful evolution tree, an entire Amazon jungle of different third party solutions sprouted around it. Many of these developments introduced alternative approaches to ion generation and fragmentation, tailored to specific applications and market needs. With their number now reaching into the hundreds, this review necessarily concentrates on those options and extensions that have had the greatest impact or achieved broad recognition. Regrettably, the sheer diversity of these contributions precludes detailed treatment of each. Instead, a more structured, taxonomic overview is adopted—an approach that can only partially capture the richness and creativity of this highly dynamic and diverse landscape, which ultimately reflects the remarkable versatility of mass spectrometry as an analytical technique.
10.2. Front‐End Options and Ion Sources
Already the very first Orbitrap MS family benefited from a wide range of plug‐in commercial front‐end technologies originally developed for LTQ MS including:
Field Asymmetric waveform Ion Mobility Spectrometry (FAIMS) (Saba et al. 2009) from Thermo Fisher Scientific,
Atmospheric pressure (AP) matrix assisted laser desorption and ionization (MALDI) source (McEwen et al. 2010), commercialized by Masstech Inc.
Desorption electrospray ionization (DESI) [(Hu et al. 2006), commercialized by Prosolia Inc.
Direct analysis in real time (DART) (Romão et al. 2016), commercialized by IonSense Inc.
Laser diode thermal desorption (LDTD) (Munoz et al. 2015) from Phytronix Technologies Inc.
Atmospheric pressure photo ionization (APPI) (Núñez et al. 2012) from Syagen Inc.
Advion nanomate for high‐throughput infusion experiments, for example in lipidomics (Schuhmann et al. 2012).
These ion sources interfaced directly with the standard electrospray hardware port of the instrument.
Later, additional plug‐in ion sources became commercially available for hybrid and Tribrid Orbitrap systems, such as:
Atmospheric solids analysis probe (ASAP) (McEwen and Gutteridge 2007), (Jagerdeo and Wriston 2017) later commercialized by ASAP-MS.com
APPI GC interface developed in collaboration with Prof. Benter's group at the Wuppertal University and later commercialized by Mascom GmbH (Kersten et al. 2016).
Soft ionization by chemical reaction in transfer (SICRIT) (Basham et al. 2022)
Multiple ionization methods under the umbrella of “inlet ionization” (Trimpin et al. 2024), originating from (McEwen et al. 2010, Pagnotti et al. 2011) and later commercialized by MSTM Inc.
High‐resolution atmospheric pressure (AP) MALDI from Transmit GmbH (Kompauer et al. 2017).
Multisprayer chip by Newomics Inc. (Chen et al. 2018).
Multi‐scheme chemical ionization (MION) by Karsa Oy (Partovi et al. 2023).
Secondary electrospray ionization source for breath analysis commercialized by Fossilion (Singh et al. 2018, Singh et al. 2019) following initial research in Zenobi's group (García‐Gómez et al. 2015, Gaugg et al. 2016).
Laser ablation electrospray ionization DP‐1000 source marketed for several years by Protea Biosciences (Beach et al. 2016). Some plug‐in sources remain research‐grade and used by multiple groups in custom‐built configurations, although not yet commercialized:
Liquid sampling–atmospheric pressure glow discharge (LS‐APGD) for combined atomic and molecular ionization (Quarles et al. 2012, Kenneth Marcus 2023).
MassSpec Pen ambient sampling device for in‐surgery diagnostics (Zhang et al. 2017).
SpiderMass source for ambient analysis of biological tissues ablated by a fiber laser (Ogrinc et al. 2019).
AP matrix‐assisted laser desorption electrospray ionization (MALDESI) (Barry et al. 2014).
nanoDESI (Roach et al. 2010).
Photoinduced thermal desorption on an atomic force microscope platform coupled with mass spectrometry for multimodal imaging (Lorenz et al. 2019).
These lists are almost certainly incomplete, as numerous research groups worldwide continue developing specialized setups for specific applications.
A number of unique ion sources require deep modifications to the Orbitrap instrument itself, including:
Dual ESI/MALDI (Belov et al. 2017) and MALDI‐2 (Niehaus et al. 2019) sources operating at sub‐mbar pressures, marketed by Spectroglyph LLC.
Liquid injection field desorption ionization (LIFDI) (Braunschweig et al. 2017; Légaré et al. 2018) commercialized by Linden CMS GmbH
ConDuct interface for increasing ion transmission from atmospheric pressure (Krutchinsky et al. 2015).
Resonance‐enhanced multiphoton ionization (REMPI) with ionization inside the C‐trap (Kösling et al. 2021).
Inductively coupled plasma (ICP) source (described only in conference reports)
Direct laser ionization in vacuum without using the C‐trap from a surface, either without (Makarov et al. 2013) or with matrix assistance (Grechnikov et al. 2014).
The following subsections describe techniques and applications that required similarly extensive modifications to Orbitrap instruments.
10.3. OrbiSIMS MS
The commercialization of the OrbiSIMS instrument by IonTOF GmbH represents one of the most successful examples of adapting Orbitrap MS to a completely new type of analysis. In OrbiSIMS, secondary ions are generated by bombardment of a flat sample with primary Bi‐ or Ar‐cluster ion beams under high‐vacuum conditions (Passarelli et al. 2017). The Orbitrap analyzer provides high‐resolution MS and MS/MS, complementing a fast but lower‐resolution TOF analyzer.
The concerted operation of both analyzers enables imaging at both macroscopic and microscopic (submicron) scales, while delivering excellent mass resolution and MS/MS sensitivity for selected regions of interest (ROI).
Such instruments have become powerful tools for biological analysis (Zhang et al. 2020) as well as for the study of organic‐based microelectronic devices such as organic light‐emitting diodes (OLED) circuits (Trindade et al. 2023).
10.4. Ion Mobility on Orbitrap MS
As ion mobility expanded rapidly over the past 25 years, numerous efforts were made to interface different ion‐mobility techniques with Orbitrap MS despite substantial differences in their timescales of separation. In addition to FAIMS option described above, several notable implementations include:
LC‐compatible atmospheric‐pressure ion mobility spectrometry (IMS) drift tube with single and double gating (Hagan et al. 2017) by Excellims,
AP drift tube (Keelor et al. 2017) developed by Prof. Clowers,
Low‐pressure drift tube (Ibrahim et al. 2016) with signal averaging, single or double multiplexing IMS.
Fourier transform ion mobility‐orbitrap mass spectrometer for native mass spectrometry (Poltash et al. 2018).
Structures for lossless ion manipulations (SLIM) (Hollerbach et al. 2024; Deng et al. 2025).
High‐resolution planar FAIMS (Baird et al. 2019), later commercialized by Heartland Mobility LLC.
Although the relatively long (10s–100s ms) cycle time of image‐current detection in Orbitrap MS is not well suited to the millisecond‐scale peaks produced by conventional drift tubes, ion mobility information may nevertheless be retained, for example, through the use of linked‐scan approaches as described in (Makarov 2025).
10.5. Light Spectroscopy and Photofragmentation
The trapping nature of Orbitrap instruments makes them naturally well suited for photofragmentation, for example, in a RF‐only collision cell. Also, it enables even light‐based spectroscopy across a wide range of wavelengths, from vacuum ultraviolet to the far‐infrared ends of the spectrum.
Infrared multiphoton dissociation (IRMPD) was the first photofragmentation technique applied to Orbitrap MS (Vasicek et al. 2011). It was later used to improve peptide sequence coverage by providing supplemental activation for ETD (Riley et al. 2017), and it has also been applied to the desolvation and sequencing of large protein complexes (Lutomski et al. 2023).
UVPD was initially implemented using excimer lasers (Shaw et al. 2013), first for top‐down protein analysis and later for a broad spectrum of applications, ranging from lipids (O'Brien et al. 2014) to large protein complexes under native conditions (Fort et al. 2016; Brodbelt et al. 2019). UVPD using solid‐state lasers was subsequently introduced (Halim et al. 2015) and became the basis for the commercial UVPD option on Orbitrap Tribrid MS systems (Fornelli et al. 2020).
While many groups implemented fixed‐wavelength photofragmentation across different parts of the spectrum (Brodbelt et al. 2019), far fewer were able to pursue wavelength‐scanning approaches. This latter strategy enabled action spectroscopy (Kopysov et al. 2015; Kopysov et al. 2016; Zviagin et al. 2025), in which fragment‐ion spectra reflect the internal structure of the trapped ions. Because fragmentation efficiency depends strongly on chromophore environment, the method provides remarkable structural specificity, enabling separation and quantification of even co‐eluting isomers of the same molecule. Its applicability has been extended to weakly absorbing species such as lipids by adding fluorophores that form complexes with the analytes (Saparbaev et al. 2021).
10.6. Orbitrap Analyzer for Space Research
As LTQ Orbitrap MS was applied to laboratory analogues of Titan aerosols (Pernot et al. 2010; Gautier et al. 2014), it became evident that the Orbitrap analyzer held significant potential for space research‐particularly for missions involving in‐situ sampling of extraterrestrial materials (Cottin et al. 2010). This realization led to the formation of the CosmOrbitrap consortium, which launched a long‐term effort to develop lightweight, low‐power Orbitrap‐based instruments and progressively raise their Technology Readiness Level (TRL) (Briois et al. 2016; Arevalo et al. 2018).
Current work focuses on the most pragmatic aspect of space exploration: the analysis of lunar surface materials for minerals and organics to support future lunar habitats (Willhite et al. 2020). A long‐term strategic aim is the eventual deployment of Orbitrap analyzers for in‐situ astrobiology missions, enabling the high‐resolution characterization of complex organic signatures beyond Earth (Arevalo et al. 2023).
10.7. Emerging Fragmentation Techniques Integrated With Orbitrap Analysis
The past decade has brought significant advances in fragmentation technologies which turned out to be synergetic with the high resolution and sensitivity of Orbitrap MS.
It has been demonstrated that positively charged biological ions can be activated using beams of high‐energy cations (Chingin et al. 2014). The Electron‐Capture Dissociation (ECD) option developed by eMSion Inc. has proven valuable for protein fragmentation (Shaw et al. 2018; Fort, Cramer, et al. 2018) and for charge reduction of large protein complexes up to intact viruses (Le Huray et al. 2024; Peris‐Díaz et al. 2025).
Surface‐induced dissociation (SID) (VanAernum et al. 2019) has emerged as a powerful method for analyzing noncovalent complexes of biological significance, with Orbitrap detection enabling improved quantification of ligand binding and resolution of complex isotopic envelopes.
Combining Orbitrap MS with the Omnitrap platform (Papanastasiou et al. 2022) represents another step toward “freedom of fragmentation.” This configuration enables a combinatorial explosion of MSn fragmentation schemes, spanning a wide range of electron energies and optionally incorporating IR and UV lasers (Smyrnakis et al. 2023). Notably, recent work showed that only distinct narrow ranges of electron energies improve sequence coverage most effectively. In another set of bottom‐up experiments, fine‐tuned electron energies enabled detailed sequencing of biantennary glycopeptides (Li et al. 2024) and even singly charged glycans (Wei et al. 2023). The utility of Omnitrap–Orbitrap integration for top‐down MS was further demonstrated for an endogenous antibody light chain (Garcia et al. 2025).
10.8. Preparative Orbitrap MS
Recent years have seen the rise of a powerful new emerging application for the long‐established technique of soft‐landing ions onto surfaces: in‐vacuo preparation of samples for cryogenic electron microscopy (CryoEM).
Initial demonstrations by Prof. S. Rauschenbach's group (Fremdling et al. 2022; Esser et al. 2022) were later complemented by work from Prof. J. Coon, who focused on depositing ions onto a matrix (Westphall et al. 2022; Salome et al. 2022; Lee et al. 2023). Embedding ions in ice (Esser et al. 2022) under cryogenic conditions (Westphall et al. 2023) proved instrumental for improving CryoEM resolution (Esser et al. 2024). Furthermore, laser re‐melting has shown promise for restoring the native structure of protein complexes after deposition (Mertz et al. 2025), enabling accurate determination of their 3D structures by CryoEM.
Soft‐landing setup have also shown utility for depositing analytes onto other types of sensors, such as nanoelectromechanical systems (Neumann et al. 2024).
10.9. “Paper Instruments”
A plethora of Orbitrap‐based instrument concepts has been proposed—typically, in patents—but still remain on paper for various practical reasons:
Multielectrode Orbitrap devices such as (Makarov 2007, Li and Mordehai 2010) offer dynamic voltage control across multiple electrodes with simpler geometries such as circular rings. However, they still require the same ultra‐high mechanical precision as conventional Orbitrap electrodes, making fabrication and alignment probably even more challenging due to their quantity and the number of very accurate voltage supplies.
Orbitrap instruments with “slow” injection (Makarov 1999; Doroshenko and Misharin 2014; Makarov 2017): while avoiding fast RF switching, this approach requires broadband excitation similar to FT‐ICR and introduces additional cycle‐time overhead, significantly reducing its practical attractiveness.
Multiple detection electrodes (Makarov, Denisov, Jung, et al. 2006; Makarov 2012, Jarrold and Todd 2022) could enable detection higher‐frequency and hence higher throughput detection (e.g., via enhanced third harmonics of the axial frequency) or improved sensitivity by detecting image current on both inner and outer electrodes. In practice, however, implementation is difficult due to requirements for extremely small gaps, ultra‐high geometric precision, and/or floating ultra‐sensitive preamplifiers at time‐dependent high voltages.
Multi‐Orbitrap design (Makarov and Horning 2008) allows to increase throughput of analysis roughly proportionally to the number of analyzers operating in parallel. However, this benefit comes with nearly linear increases in mechanical and electronic complexity and cost, with only modest savings from shared vacuum systems and power supplies. Notably, the value of operating two high‐resolution analyzers in parallel has since been demonstrated in the Orbitrap Astral platform, where one Orbitrap performs slower ultra‐high‐resolution acquisition (potentially including EDR), while the second analyzer focuses exclusively on high‐speed MS/MS acquisition. Overall, each of these concepts is fundamentally feasible, and some may yet prove valuable as enabling technologies and engineering solutions continue to evolve.
11. Future of Orbitrap MS
With the Astral analyzer now representing the current pinnacle of performance due to its superior acquisition rate and sensitivity, Orbitrap technology is ready to become the major workhorse of high‐resolution mass spectrometry. This evolution is expected to unfold along several major directions.
The first major thrust will involve the democratization of Orbitrap technology toward mass‐detector–class instruments, as already demonstrated with the Orbitrap Exploris MX and EFOX systems. This trend is anticipated to extend further into mobile and field‐deployable platforms, including applications such as intraoperative analysis with the MassSpec Pen.
The second major thrust involves the deeper integration of Orbitrap analyzers with complementary technologies, including:
New instrument configurations, continuing the trend seen in Tribrid, Orbitrap Astral, OrbiSIMS MS.
Integration with emerging fragmentation methods, particularly those involving electrons and photons.
Support for non‐MS modalities such as ion mobility, spectroscopy, preparative MS, and so forth.
Deployment within specialized instruments for clinical diagnostics, isotope‐ratio measurements, structural biology, de novo sequencing, per‐/poly‐fluoroalkyl substances (PFAS) analysis, and more
Fit‐for‐purpose interfaces to specialized ion sources, such as ICP and photoionization sources
Both of these trends share the focus on cost‐optimized mechanical design, contamination‐resistant ion optics, remote and online diagnostics, and automatic calibration.
In addition, numerous opportunities remain for improving Orbitrap performance. A major avenue is the extension of DMT to new applications across both higher and lower m/z ranges, enabled by advances in signal‐processing techniques.
New Orbitrap designs could further enhance performance—whether optimized specifically for DMT or implemented in multitrap architectures (Makarov and Horning 2008). Ultra‐long transients offer the potential to reach extreme resolving powers in the multimillion range.
Further improvements in operational methods are expected as well. For example, EDR scanning may continue to increase the depth and dynamic range of screening workflows.
Novel signal‐processing methods such as FSDM (Kelstrup, Aizikov, et al. 2018; Steigerwald et al. 2024) can double resolving power for the same transient duration, while transient‐fitting approaches hold promise for improving isotope‐ratio measurements, structural analysis, and CCS determinations (James et al. 2022; James et al. 2023; Sanders et al. 2018).
As Orbitrap technology matures, future advances are likely to arise less from purely hardware‐driven improvements and increasingly from the integration of sophisticated software and artificial intelligence (AI).
Intelligent control algorithms can enable adaptive data acquisition, dynamically optimizing resolution, ion injection, and fragmentation methods in real time. AI‐enhanced signal processing may further improve effective resolution, mass accuracy, intensity accuracy and dynamic range by mitigating noise and space‐charge effects. In parallel, predictive diagnostics and automated calibration can enhance robustness and facilitate remote, autonomous operation.
Ultimately, the integration of high‐fidelity hardware with intelligent software layers may shift Orbitrap mass spectrometry toward becoming an integral part of self‐optimizing, experiment‐adaptive analytical systems. As a part of such integration, it is conceivable that for certain well‐defined and routine analytical tasks, simpler configurations (e.g., MS‐only platforms) could replace more complex tandem MS systems.
12. Conclusion
Orbitrap mass spectrometry has advanced enormously over the past 20 years and continues to evolve. During this period, the performance frontiers of Orbitrap technology have expanded enormously across nearly every analytical dimension. Limits of quantitation below 1 attomole, acquisition speeds reaching 70 spectra per second, mass resolutions exceeding 2 million, isotope‐ratio precision better than 0.1%, and sub‐ppm mass accuracy have become hallmarks of the technology, enabling highly confident molecular identification.
Orbitrap technology also continues to redefine the boundaries of protein analysis. The latest instruments can isotopically resolve intact antibodies and determine the charge states of individual ions across an exceptionally broad mass range, from small peptides to ions with masses in the multi‐megadalton regime.
This level of performance has led to widespread global deployment, with a five‐digit number of Orbitrap instruments operating worldwide around the clock across nearly all domains of mass spectrometry, contributing to the scientific literature at a rate of two to three publications per hour.
By continuing to advance Orbitrap technology and integrating it more deeply into analytical workflows, our goal is to bring high‐resolution mass spectrometry to every laboratory, ultimately contributing to a safer, cleaner, and healthier world!
Author Contributions
Alexander Makarov: conceptualization, writing – original draft, writing – review and editing, investigation, visualization.
Acknowledgments
I would like to express my deepest gratitude to all colleagues who have contributed to the success of Orbitrap mass spectrometry, each of whom has invested a part of their talent, soul, and professional life to its development.
I am also deeply grateful to Stevan Horning, Andreas Wieghaus, Peter Parlett, Albert Heck, Karl Mechtler, and Richard Cole for their thoughtful notes, careful reading, and constructive edits.
References
- Arevalo, R. , Selliez L., Briois C., et al. 2018. “An Orbitrap‐Based Laser Desorption/Ablation Mass Spectrometer Designed for Spaceflight.” Rapid Communications in Mass Spectrometry 32: 1875–1886. [DOI] [PubMed] [Google Scholar]
- Arevalo, R. , Willhite L., Bardyn A., et al. 2023. “Laser Desorption Mass Spectrometry With an Orbitrap Analyser for In Situ Astrobiology.” Nature Astronomy 7: 359–365. [Google Scholar]
- Aushev, T. , Kozhinov A. N., and Tsybin Y. O.. 2014. “Least‐Squares Fitting of Time‐Domain Signals for Fourier Transform Mass Spectrometry.” Journal of the American Society for Mass Spectrometry 25, no. 7: 1263–1273. [DOI] [PubMed] [Google Scholar]
- Baird, M. A. , Shliaha P. V., Anderson G. A., et al. 2019. “High‐Resolution Differential Ion Mobility Separations/Orbitrap Mass Spectrometry Without Buffer Gas Limitations.” Analytical Chemistry 91, no. 10: 6918–6925. [DOI] [PubMed] [Google Scholar]
- Barry, J. A. , Robichaud G., Bokhart M. T., et al. 2014. “Mapping Antiretroviral Drugs in Tissue by IR‐MALDESI MSI Coupled to the Q Exactive and Comparison With LC‐MS/MS SRM Assay.” Journal of the American Society for Mass Spectrometry 25, no. 12: 2038–2047. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Basham, V. , Hancock T., McKendrick J., Tessarolo N., and Wicking C.. 2022. “Detailed Chemical Analysis of a Fully Formulated Oil Using Dielectric Barrier Discharge Ionisation–Mass Spectrometry.” Rapid Communications in Mass Spectrometry 36, no. 14: e9320. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Bateman, K. P. , Kellmann M., Muenster H., Papp R., and Taylor L.. 2009. “Quantitative‐Qualitative Data Acquisition Using a Benchtop Orbitrap Mass Spectrometer.” Journal of the American Society for Mass Spectrometry 20, no. 8: 1441–1450. [DOI] [PubMed] [Google Scholar]
- Beach, D. G. , Walsh C. M., Cantrell P., et al. 2016. “Laser Ablation Electrospray Ionization High‐Resolution Mass Spectrometry for Regulatory Screening of Domoic Acid in Shellfish.” Rapid Communications in Mass Spectrometry 30, no. 22: 2379–2387. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Bekker‐Jensen, D. B. , Martínez‐Val A., Steigerwald S., et al. 2020. “A Compact Quadrupole‐Orbitrap Mass Spectrometer With FAIMS Interface Improves Proteome Coverage in Short LC Gradients.” Molecular & Cellular Proteomics 19, no. 4: 716–729. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Belarbi, S. , Vivier M., Zaghouani W., De Sloovere A. D., Agasse‐Peulon V., and Cardinael P.. 2021. “Comparison of New Approach of GC‐HRMS (Q‐Orbitrap) to GC–MS/MS (Triple‐Quadrupole) in Analyzing the Pesticide Residues and Contaminants in Complex Food Matrices.” Food Chemistry 359: 129932. [DOI] [PubMed] [Google Scholar]
- Belov, M. E. , Damoc E., Denisov E., et al. 2013. “From Protein Complexes to Subunit Backbone Fragments: A Multi‐Stage Approach to Native Mass Spectrometry.” Analytical Chemistry 85, no. 23: 11163–11173. [DOI] [PubMed] [Google Scholar]
- Belov, M. E. , Ellis S. R., Dilillo M., et al. 2017. “Design and Performance of a Novel Interface for Combined Matrix‐Assisted Laser Desorption Ionization at Elevated Pressure and Electrospray Ionization With Orbitrap Mass Spectrometry.” Analytical Chemistry 89, no. 14: 7493–7501. [DOI] [PubMed] [Google Scholar]
- Berube, M. , DiBussolo J., Lafontaine C., and Shi Y.. 2012. “Analysis of 25‐Hydroxyvitamin D in Serum Using an Automated Online Sample Preparation Technique With a High‐Resolution Benchtop Orbitrap Mass Spectrometer.” Clinical Biochemistry 13, no. 45: 1101. [Google Scholar]
- Bolotskikh, P. A. , Grinfeld D. E., Makarov A. A., and Monastyrskiy M. A.. 2011. “Coulomb Dynamics of Ion Bunches in Multi‐Reflection Electrostatic Traps.” Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 645: 146–152. [Google Scholar]
- Bondarenko, P. V. , Second T. P., Zabrouskov V., Makarov A. A., and Zhang Z.. 2009. “Mass Measurement and Top‐Down HPLC/MS Analysis of Intact Monoclonal Antibodies on a Hybrid Linear Quadrupole Ion Trap–Orbitrap Mass Spectrometer.” Journal of the American Society for Mass Spectrometry 20: 1415–1424. [DOI] [PubMed] [Google Scholar]
- Bornmann, L. , and Marx W.. 2012. “The Anna Karenina Principle: A Way of Thinking About Success in Science.” Journal of the American Society for Information Science and Technology 63, no. 10: 2037–2051. [Google Scholar]
- Braunschweig, H. , Krummenacher I., Légaré M. A., Matler A., Radacki K., and Ye Q.. 2017. “Main‐Group Metallomimetics: Transition Metal‐Like Photolytic CO Substitution at Boron.” Journal of the American Chemical Society 139, no. 5: 1802–1805. [DOI] [PubMed] [Google Scholar]
- Briois, C. , Thissen R., Thirkell L., et al. 2016. “Orbitrap Mass Analyser for In Situ Characterisation of Planetary Environments: Performance Evaluation of a Laboratory Prototype.” Planetary & Space Science 131: 33–45. [Google Scholar]
- Brodbelt, J. S. , Morrison L. J., and Santos I.. 2019. “Ultraviolet Photodissociation Mass Spectrometry for Analysis of Biological Molecules.” Chemical Reviews 120, no. 7: 3328–3380. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Brunée, C. 1987. “The Ideal Mass Analyzer: Fact or Fiction?” International Journal of Mass Spectrometry and Ion Processes 79, no. 3: R10. [Google Scholar]
- Chen, Y. , Mao P., and Wang D.. 2018. “Quantitation of Intact Proteins in Human Plasma Using Top‐Down Parallel Reaction Monitoring‐MS.” Analytical Chemistry 90, no. 18: 10650–10653. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Chingin, K. , Makarov A., Denisov E., Rebrov O., and Zubarev R. A.. 2014. “Fragmentation of Positively Charged Biological Ions Activated With a Beam of High‐Energy Cations.” Analytical Chemistry 86, no. 1: 372–379. [DOI] [PubMed] [Google Scholar]
- Contino, N. C. , Pierson E. E., Keifer D. Z., and Jarrold M. F.. 2013. “Charge Detection Mass Spectrometry With Resolved Charge States.” Journal of the American Society for Mass Spectrometry 24, no. 1: 101–108. [DOI] [PubMed] [Google Scholar]
- Cottin, H. , Briois C., Engrand C., et al. 2010. “Ilma Ion Laser Mass Analyser for in Situ Characterization of a Near Earth Object (Neo) for the Marco Polo Mission.” Origins of Life and Evolution of the Biosphere 40: 571–572. [Google Scholar]
- Deng, L. , Adamson B., McKenzie‐Coe A., et al. 2025. “Enhancing Sensitivity in Low‐Load Proteomics Orbitrap Workflows via SLIM Integration.” Analytical Chemistry 97, no. 10: 12613–12622. [DOI] [PubMed] [Google Scholar]
- Denisov, E. , Damoc E., Lange O., and Makarov A.. 2012. “Orbitrap Mass Spectrometry With Resolving Powers above 500,000 and 1,000,000 on Chromatographic Time Scale.” International Journal of Mass Spectrometry 325–327: 80–85. [Google Scholar]
- Denisov, E. , Damoc E., and Makarov A.. 2021. “Exploring Frontiers of Orbitrap Performance for Long Transients.” International Journal of Mass Spectrometry 466: 116607. [Google Scholar]
- Deschamps, E. , Calabrese V., Schmitz I., Hubert‐Roux M., Castagnos D., and Afonso C.. 2023. “Advances in Ultra‐High‐Resolution Mass Spectrometry for Pharmaceutical Analysis.” Molecules 28, no. 5: 2061. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Deslignière, E. , Yin V. C., Ebberink E. H. T. M., et al. 2024. “Ultralong Transients Enhance Sensitivity and Resolution in Orbitrap‐Based Single‐Ion Mass Spectrometry.” Nature Methods 21: 619–622. [DOI] [PubMed] [Google Scholar]
- Diamond, J. 1994. “Zebras and the Anna Karenina Principle.” Natural History 103, no. 9: 4–10. [Google Scholar]
- Diamond, J. 1997. Guns, Germs, and Steel: The Fates of Human Societies. W. W. Norton & Company. [Google Scholar]
- Doroshenko V. M., and Misharin A.. 2014. Electrostatic Orbital Trap Mass Spectrometer, US patent US8796619.
- Dyachenko, A. , Wang G., Belov M., et al. 2015. “Tandem Native Mass Spectrometry on Antibody–Drug Conjugates and Submillion Da Antibody–Antigen Protein Assemblies on an Orbitrap EMR Equipped With a High‐Mass Quadrupole Mass Selector.” Analytical Chemistry 87, no. 12: 6095–6102. [DOI] [PubMed] [Google Scholar]
- Eiler, J. , Cesar J., Chimiak L., et al. 2017. “Analysis of Molecular Isotopic Structures at High Precision and Accuracy by Orbitrap Mass Spectrometry.” International Journal of Mass Spectrometry 422: 126–142. [Google Scholar]
- Eliuk, S. , and Makarov A.. 2015. “Evolution of Orbitrap Mass Spectrometry Instrumentation.” Annual Review of Analytical Chemistry 8: 61–80. [DOI] [PubMed] [Google Scholar]
- Emerson, R. W. (Attributed). 1882. “The value of good work. The Atlanta Constitution.” Quoted in Quote Investigator. (2015, March 24). If You Build a Better Mousetrap the World Will Beat a Path to Your Door. https://quoteinvestigator.com/2015/03/24/mousetrap/.
- Erickson, B. K. , Mintseris J., Schweppe D. K., et al. 2019. “Active Instrument Engagement Combined With a Real‐Time Database Search for Improved Performance of Sample Multiplexing Workflows.” Journal of Proteome Research 18, no. 3: 1299–1306. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Esser, T. K. , Böhning J., Fremdling P., et al. 2022. “Mass‐Selective and Ice‐Free Electron Cryomicroscopy Protein Sample Preparation via Native Electrospray Ion‐Beam Deposition.” PNAS Nexus 1, no. 4: pgac153. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Esser, T. K. , Böhning J., Önür A., et al. 2024. “Cryo‐EM of Soft‐Landed β‐Galactosidase: Gas‐Phase and Native Structures Are Remarkably Similar.” Science Advances 10: eadl4628. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Fornelli, L. , Ayoub D., Aizikov K., et al. 2017. “Top‐Down Analysis of Immunoglobulin G Isotypes 1 and 2 With Electron Transfer Dissociation on a High‐Field Orbitrap Mass Spectrometer.” Journal of Proteomics 159: 67–76. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Fornelli, L. , Damoc E., Thomas P. M., et al. 2012. “Analysis of Intact Monoclonal Antibody IgG1 by Electron Transfer Dissociation Orbitrap FTMS.” Molecular & Cellular Proteomics 11: 1758–1767. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Fornelli, L. , Srzentić K., Toby T. K., et al. 2020. “Thorough Performance Evaluation of 213 nm Ultraviolet Photodissociation for Top‐Down Proteomics.” Molecular & Cellular Proteomics 19, no. 2: 405–420. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Fort, K. L. , Cramer C. N., Voinov V. G., et al. 2018. “Exploring ECD on a Benchtop Q Exactive Orbitrap Mass Spectrometer.” J. Prot. Res. 17, no. 2: 926–933. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Fort, K. L. , Dyachenko A., Potel C. M., et al. 2016. “Implementation of Ultraviolet Photodissociation on a Benchtop Q Exactive Mass Spectrometer and Its Application to Phosphoproteomics.” Analytical Chemistry 88, no. 4: 2303–2310. [DOI] [PubMed] [Google Scholar]
- Fort, K. L. , Van de Waterbeemd M., Boll D., et al. 2018. “Expanding the Structural Analysis Capabilities on an Orbitrap‐Based Mass Spectrometer for Large Macromolecular Complexes.” Analyst 143: 100–105. [DOI] [PubMed] [Google Scholar]
- Fremdling, P. , Esser T. K., Saha B., et al. 2022. “A Preparative Mass Spectrometer to Deposit Intact Large Native Protein Complexes.” ACS Nano 16, no. 9: 14443–14455. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Frese, C. K. , Altelaar A. F. M., van den Toorn H., et al. 2012. “Toward Full Peptide Sequence Coverage by Dual Fragmentation Combining Electron‐Transfer and Higher‐Energy Collision Dissociation Tandem Mass Spectrometry.” Analytical Chemistry 84, no. 22: 9668–9673. [DOI] [PubMed] [Google Scholar]
- Fuerstenau, S. D. , and Benner W. H.. 1995. “Molecular Weight Determination of Megadalton DNA Electrospray Ions Using Charge Detection Time‐of‐Flight Mass Spectrometry.” Rapid Communications in Mass Spectrometry 9, no. 15: 1528–1538. [DOI] [PubMed] [Google Scholar]
- Gall, L. N. , Golikov Y. K., Aleksandrov M. L., Pechalina Y. E., and Holin N. A. 1986. “USSR Inventor's Certificate #1247973.”
- Garcia, C. , Khristenko N. A., Fu T., et al. 2025. “Top‐Down Mass Spectrometry of a Clinical Antibody Light Chain Using the Omnitrap‐Orbitrap‐Booster Platform.” Journal of the American Society for Mass Spectrometry. [DOI] [PubMed]
- García‐Gómez, D. , Martínez‐Lozano sinues P., Barrios‐Collado C., Vidal‐De‐Miguel G., Gaugg M., and Zenobi R.. 2015. “Identification of 2‐Alkenals, 4‐Hydroxy‐2‐Alkenals, and 4‐Hydroxy‐2, 6‐Alkadienals in Exhaled Breath Condensate by UHPLC‐HRMS and in Breath by Real‐Time HRMS.” Analytical Chemistry 87, no. 5: 3087–3093. [DOI] [PubMed] [Google Scholar]
- Gaugg, M. T. , Gomez D. G., Barrios‐Collado C., et al. 2016. “Expanding Metabolite Coverage of Real‐Time Breath Analysis by Coupling a Universal Secondary Electrospray Ionization Source and High Resolution Mass Spectrometry—A Pilot Study on Tobacco Smokers.” Journal of Breath Research 10, no. 1: 016010. [DOI] [PubMed] [Google Scholar]
- Gault, J. , Donlan J. A. C., Liko I., et al. 2016. “High‐Resolution Mass Spectrometry of Small Molecules Bound to Membrane Proteins.” Nature Methods 13: 333–336. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Gault, J. , Liko I., Landreh M., et al. 2020. “Combining Native and ‘Omics’ Mass Spectrometry to Identify Endogenous Ligands Bound to Membrane Proteins.” Nature Methods 17, no. 5: 505–508. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Gautier, T. , Carrasco N., Schmitz‐Afonso I., et al. 2014. “Nitrogen Incorporation in Titan's Tholins Inferred by High Resolution Orbitrap Mass Spectrometry and Gas Chromatography–Mass Spectrometry.” Earth & Planetary Science Lett 404: 33–42. [Google Scholar]
- Geiger, T. , Cox J., and Mann M.. 2010. “Proteomics on an Orbitrap Benchtop Mass Spectrometer Using All‐Ion Fragmentation.” Molecular & Cellular Proteomics 9, no. 10: 2252–2261. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Gharibi, H. , Chernobrovkin A. L., Eriksson G., et al. 2022. “Abnormal (Hydroxy)Proline Deuterium Content Redefines Hydrogen Chemical Mass.” Journal of the American Chemical Society 144, no. 6: 2484–2487. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Gharibi, H. , Chernobrovkin A. L., Saei A. A., et al. 2022. “Proteomics‐Compatible Fourier Transform Isotopic Ratio Mass Spectrometry of Polypeptides.” Analytical Chemistry 94, no. 43: 15048–15056. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Gorshkov, M. V. , Fornelli L., and Tsybin Y. O.. 2012. “Observation of Ion Coalescence in Orbitrap Fourier Transform Mass Spectrometry.” Rapid Communications in Mass Spectrometry 26, no. 15: 1711–1717. [DOI] [PubMed] [Google Scholar]
- Gorshkov, M. V. , Good D. M., Lyutvinskiy Y., Yang H., and Zubarev R. A.. 2010. “Calibration Function for the Orbitrap FTMS Accounting for the Space Charge Effect.” Journal of the American Society for Mass Spectrometry 21, no. 11: 1846–1851. [DOI] [PubMed] [Google Scholar]
- Goswami, P. , Cesare J., Rekowski M. J., Clark Z., Thornton J., and Washburn M. P.. 2025. “Analysis of FAIMS for the Study of Affinity‐Purified Protein Complexes Using the Orbitrap Ascend Tribrid Mass Spectrometer.” Molecular Omics 21: 303–314. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Grechnikov, A. , Nikiforov S., Strupat K., and Makarov A.. 2014. “Determination of Rhenium and Osmium Complexes by Surface‐Assisted Laser Desorption/Ionization Coupled to Orbitrap Mass Analyzer.” Analytical and Bioanalytical Chemistry 406, no. 6: 3019–3023. [DOI] [PubMed] [Google Scholar]
- Grinfeld, D. , Aizikov K., Kreutzmann A., Damoc E., and Makarov A.. 2017. “Phase‐Constrained Spectrum Deconvolution for Fourier Transform Mass Spectrometry.” Analytical Chemistry 89, no. 2: 1202–1211. [DOI] [PubMed] [Google Scholar]
- Grinfeld, D. , Monastyrskiy M., and Makarov A.. 2015. “Control of Aberration and Space‐Charge Effects in the Orbitrap Mass Analyzer.” Supplement, Microscopy and Microanalysis 21, no. S4: 176–181. [Google Scholar]
- Grinfeld, D. , Stewart H., Balschun W., Skoblin M., Hock C., and Makarov A.. 2024. “Multi‐Reflection Astral Mass Spectrometer With Isochronous Drift in Elongated Ion Mirrors.” Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 1060: 169017. [Google Scholar]
- Grinfeld, D. , Stewart H., Skoblin M., Denisov E., Monastyrsky M., and Makarov A.. 2019. “Space‐Charge Dynamics in Orbitrap Mass Spectrometers.” International Journal of Modern Physics A 34: 1942007. [Google Scholar]
- Grinfeld, D. E. , Kopaev I. A., Makarov A. A., and Monastyrskiy M. A.. 2011. “Equilibrium Ion Distribution Modeling in RF Ion Traps and Guides With Regard to Coulomb Effects.” Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 645: 141–145. [Google Scholar]
- Guzman, U. H. , Martinez‐Val A., Ye Z., et al. 2024. “Ultra‐Fast Label‐Free Quantification and Comprehensive Proteome Coverage With Narrow‐Window Data‐Independent Acquisition.” Nature Biotechnology 42: 1855–1866. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Hagan, N. , Goldberg I., Graichen A., et al. 2017. “Ion Mobility Spectrometry‐High Resolution LTQ‐Orbitrap Mass Spectrometry for Analysis of Homemade Explosives.” Journal of the American Society for Mass Spectrometry 28, no. 8: 1531–1539. [DOI] [PubMed] [Google Scholar]
- Halim, M. A. , Girod M., MacAleese L., Lemoine J., Antoine R., and Dugourd P.. 2015. “213 nm Ultraviolet Photodissociation on Peptide Anions: Radical‐Directed Fragmentation Patterns.” Journal of the American Society for Mass Spectrometry 27, no. 3: 474–486. [DOI] [PubMed] [Google Scholar]
- Hardman, M. , and Makarov A. A.. 2003. “Interfacing the Orbitrap Mass Analyzer to an Electrospray Ion Source.” Analytical Chemistry 75: 1699–1705. [DOI] [PubMed] [Google Scholar]
- Hauschild, J.‐P. , Peterson A. C., Couzijn E., et al. 2020. “A Novel Family of Quadrupole‐Orbitrap Mass Spectrometers for a Broad Range of Analytical Applications.” Preprints: 2020060111. 10.20944/preprints202006.0111.v1. [DOI] [Google Scholar]
- He, Y. , Shishkova E., Peters‐Clarke T. M., et al. 2023. “Evaluation of the Orbitrap Ascend Tribrid Mass Spectrometer for Shotgun Proteomics.” Analytical Chemistry 95, no. 28: 10655–10663. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Hecht, E. S. , Scigelova M., Eliuk S., and Makarov A.. 2019. “Fundamentals and Advances of Orbitrap Mass Spectrometry.” In Encyclopedia of Analytical Chemistry, edited by Meyers R. A.. Wiley. [Google Scholar]
- Hilkert, A. , Böhlke J. K., Mroczkowski S. J., et al. 2021. “Exploring the Potential of Electrospray‐Orbitrap for Stable Isotope Analysis Using Nitrate as a Model.” Analytical Chemistry 93, no. 26: 9139–9148. [DOI] [PubMed] [Google Scholar]
- Hofmann, A. E. , Chimiak L., Dallas B., et al. 2020. “Using Orbitrap Mass Spectrometry to Assess the Isotopic Compositions of Individual Compounds in Mixtures.” International Journal of Mass Spectrometry 457: 116410. [Google Scholar]
- Hollerbach, A. L. , Ibrahim Y. M., Lin V. S., et al. 2024. “Identification of Unique Fragmentation Patterns of Fentanyl Analog Protomers Using Structures for Lossless Ion Manipulations Ion Mobility‐Orbitrap Mass Spectrometry.” Journal of the American Society for Mass Spectrometry 35, no. 4: 793–803. [DOI] [PubMed] [Google Scholar]
- Hsu, C. , Shulman N., Stewart H. I., et al. 2025. “Evaluation of a Prototype Orbitrap Astral Zoom Mass Spectrometer for Quantitative Proteomics—Beyond Identification Lists.” Journal of Proteome Research ASAP article. [DOI] [PMC free article] [PubMed]
- Hu, Q. , Noll R. J., Li H., Makarov A., Hardman M., and Graham Cooks R.. 2005. “The Orbitrap: A New Mass Spectrometer.” Journal of Mass Spectrometry 40: 430–443. [DOI] [PubMed] [Google Scholar]
- Hu, Q. , Talaty N., Noll R. J., and Cooks R. G.. 2006. “Desorption Electrospray Ionization Using an Orbitrap Mass Spectrometer: Exact Mass Measurements on Drugs and Peptides.” Rapid Communications in Mass Spectrometry 20, no. 22: 3403–3408. [DOI] [PubMed] [Google Scholar]
- Huguet, R. , Mullen C., Srzentić K., et al. 2019. “Proton Transfer Charge Reduction Enables High‐Throughput Top‐Down Analysis of Large Proteoforms.” Analytical Chemistry 91, no. 24: 15732–15739. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Ibrahim, Y. M. , Garimella S. V. B., Prost S. A., et al. 2016. “Development of an Ion Mobility Spectrometry‐Orbitrap Mass Spectrometer Platform.” Analytical Chemistry 88, no. 24: 12152–12160. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Jagerdeo, E. , and Wriston A.. 2017. “Rapid Analysis of Forensic‐Related Samples Using Two Ambient Ionization Techniques Coupled to High‐Resolution Mass Spectrometers.” Rapid Communications in Mass Spectrometry 31, no. 9: 782–790. [DOI] [PubMed] [Google Scholar]
- James, V. K. , Sanders J. D., Aizikov K., et al. 2023. “Expanding Orbitrap Collision Cross‐Section Measurements to Native Protein Applications Through Kinetic Energy and Signal Decay Analysis.” Analytical Chemistry 95, no. 26: 7656–7664. [DOI] [PubMed] [Google Scholar]
- James, V. K. , Sanders J. D., and Brodbelt J. S.. 2022. “Advancing Orbitrap Measurements of Collision Cross Sections.” Journal of the American Society for Mass Spectrometry 33, no. 12: 2147–2157. [Google Scholar]
- Jarrold, M. F. , and Todd A. R. 2022. “Orbitrap for Single Particle Mass Spectrometry.” US Patents 11,495,449.
- Kafader, J. O. , Beu S. C., Early B. P., et al. 2019. “STORI Plots Enable Accurate Tracking of Individual Ion Signals.” Journal of the American Society for Mass Spectrometry 30, no. 11: 2200–2203. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kafader, J. O. , Melani R. D., Durbin K. R., et al. 2020. “Multiplexed Mass Spectrometry of Individual Ions Improves Measurement of Proteoforms and Their Complexes.” Nature Methods 17: 391–394. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kafader, J. O. , Melani R. D., Senko M. W., Makarov A. A., Kelleher N. L., and Compton P. D.. 2019. “Measurement of Individual Ions Sharply Increases the Resolution of Orbitrap Mass Spectra of Proteins.” Analytical Chemistry 91, no. 4: 2776–2783. [DOI] [PubMed] [Google Scholar]
- Kaufmann, A. , and Walker S.. 2012. “Accuracy of Relative Isotopic Abundance and Mass Measurements in a Single‐Stage Orbitrap Mass Spectrometer.” Rapid Communications in Mass Spectrometry 26, no. 9: 1081–1090. [DOI] [PubMed] [Google Scholar]
- Kaufmann, A. , and Walker S.. 2016. “Extension of the Q Orbitrap Intrascan Dynamic Range by Using a Dedicated Customized Scan.” Rapid Communications in Mass Spectrometry 30, no. 8: 1087–1095. [DOI] [PubMed] [Google Scholar]
- Kaufmann, A. , and Walker S.. 2018. “Coalescence and Self‐Bunching Observed in Commercial High‐Resolution Mass Spectrometry Instrumentation.” Rapid Communications in Mass Spectrometry 32, no. 6: 503–515. [DOI] [PubMed] [Google Scholar]
- Kaufmann, A. , Widmer M., and Maden K.. 2010. “Post‐Interface Signal Suppression, a Phenomenon Observed in a Single‐Stage Orbitrap Mass Spectrometer Coupled to an Electrospray Interfaced Liquid Chromatograph.” Rapid Communications in Mass Spectrometry 24, no. 14: 2162–2170. [DOI] [PubMed] [Google Scholar]
- Keelor, J. D. , Zambrzycki S., Li A., Clowers B. H., and Fernández F. M.. 2017. “Atmospheric Pressure Drift Tube Ion Mobility–Orbitrap Mass Spectrometry: Initial Performance Characterization.” Analytical Chemistry 89, no. 21: 11301–11309. [DOI] [PubMed] [Google Scholar]
- Keenan, M. R. , Trindade G. F., Pirkl A., et al. 2025. “Orbitrap Noise Structure and Method for Noise Unbiased Multivariate Analysis.” Nature Communications 16, no. 1: 6398. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kelstrup, C. D. , Aizikov K., Batth T. S., et al. 2018. “Limits for Resolving Isobaric Tandem Mass Tag Reporter Ions Using Phase Constrained Spectrum Deconvolution.” Journal of Proteome Research 17, no. 11: 4008–4016. [DOI] [PubMed] [Google Scholar]
- Kelstrup, C. D. , Bekker‐Jensen D. B., Arrey T. N., Hogrebe A., Harder A., and Olsen J. V.. 2018. “Performance Evaluation of the Q Exactive HF‐X for Shotgun Proteomics.” J. Prot. Res. 17, no. 1: 727–738. [DOI] [PubMed] [Google Scholar]
- Kelstrup, C. D. , Young C., Lavallee R., Nielsen M. L., and Olsen J. V.. 2012. “Optimized Fast and Sensitive Acquisition Methods for Shotgun Proteomics on a Quadrupole Orbitrap Mass Spectrometer.” Journal of Proteome Research 11, no. 6: 3487–3497. [DOI] [PubMed] [Google Scholar]
- Kenneth Marcus, R. , Hoegg E. D., Hall K. A., Williams T. J., and Koppenaal D. W.. 2023. “Combined Atomic and Molecular (CAM) Ionization With the Liquid Sampling‐Atmospheric Pressure Glow Discharge Microplasma.” Mass Spectrometry Reviews 42, no. 2: 652–673. [DOI] [PubMed] [Google Scholar]
- Kersten, H. , Kroll K., Haberer K., et al. 2016. “Design Study of an Atmospheric Pressure Photoionization Interface for GC‐MS.” Journal of the American Society for Mass Spectrometry 27, no. 4: 607–614. [DOI] [PubMed] [Google Scholar]
- Kessler, A. L. , Fort K., Resemann H., et al. 2025. “Increased ETHcD Efficiency on the Hybrid Orbitrap Excedion Pro Mass Analyzer Extends the Depth in Identification and Sequence Coverage of HLA Class I Immunopeptidomes.” bioRxiv 2025: 2025. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kharchenko, A. , Vladimirov G., Heeren R. M. A., and Nikolaev E. N.. 2012. “Performance of Orbitrap Mass Analyzer at Various Space Charge and Non‐Ideal Field Conditions: Simulation Approach.” Journal of the American Society for Mass Spectrometry 23, no. 5: 977–987. [DOI] [PubMed] [Google Scholar]
- Kingdon, K. H. 1923. “A Method for the Neutralization of Electron Space Charge by Positive Ionization at Very Low Gas Pressures.” Physical Review 21: 408–418. [Google Scholar]
- Knight, R. D. 1981. “Storage of Ions From Laser‐Produced Plasmas.” Applied Physics Letters 38: 221–223. [Google Scholar]
- Knizner, K. T. , Bagley M. C., Garrard K. P., et al. 2022. “Optimized C‐Trap Timing of an Orbitrap 240 Mass Spectrometer for High‐Throughput Screening and Native MS by IR‐MALDESI.” Journal of the American Society for Mass Spectrometry 33, no. 2: 328–334. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kompauer, M. , Heiles S., and Spengler B.. 2017. “Autofocusing Maldi Mass Spectrometry Imaging of Tissue Sections and 3D Chemical Topography of Nonflat Surfaces.” Nature Methods 14, no. 12: 1156–1158. [DOI] [PubMed] [Google Scholar]
- Kopysov, V. , Makarov A., and Boyarkin O. V.. 2015. “Colors for Molecular Masses: Fusion of Spectroscopy and Mass Spectrometry for Identification of Biomolecules.” Analytical Chemistry 87, no. 9: 4607–4611. [DOI] [PubMed] [Google Scholar]
- Kopysov, V. , Makarov A., and Boyarkin O. V.. 2016. “Non‐Statistical UV Fragmentation of Gas‐Phase Peptides Reveals Conformers and Their Structural Features.” The Journal of Physical Chemistry Letters 7, no. 6: 1067–1071. [DOI] [PubMed] [Google Scholar]
- Korsunskii, M. I. , and Basakutsa V. A.. 1958. “A Study of the Ion‐Optical Properties of a Sector‐Shaped Electrostatic Field of the Difference Type.” Soviet Physics. Technical Physics 3: 1396–1409. [Google Scholar]
- Kösling, P. , Rüger C. P., Schade J., et al. 2021. “Vacuum Laser Photoionization Inside the C‐Trap of an Orbitrap Mass Spectrometer: Resonance‐Enhanced Multiphoton Ionization High‐Resolution Mass Spectrometry.” Analytical Chemistry 93, no. 27: 9418–9427. [DOI] [PubMed] [Google Scholar]
- Köster, C. 2009. “The Concept of Electrostatic Non‐Orbital Harmonic Ion Trapping.” International Journal of Mass Spectrometry 287, no. 1–3: 114–118. [Google Scholar]
- Kozhinov, A. N. , Johnson A., Nagornov K. O., et al. 2023. “Super‐Resolution Mass Spectrometry Enables Rapid, Accurate, and Highly Multiplexed Proteomics at the MS2 Level.” Analytical Chemistry 95, no. 7: 3712–3719. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Krutchinsky, A. N. , Padovan J. C., Cohen H., and Chait B. T.. 2015. “Maximizing Ion Transmission From Atmospheric Pressure Into the Vacuum of Mass Spectrometers With a Novel Electrospray Interface.” Journal of the American Society for Mass Spectrometry 26, no. 4: 649–658. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Lancaster, N. M. , Sinitcyn P., Forny P., et al. 2024. “Fast and Deep Phosphoproteome Analysis With the Orbitrap Astral Mass Spectrometer.” Nature Communications 15, no. 1: 7016. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Lange, O. , Damoc E., Wieghaus A., and Makarov A.. 2014. “Enhanced Fourier Transform for Orbitrap Mass Spectrometry.” International Journal of Mass Spectrometry 369: 16–22. [Google Scholar]
- Le Huray, K. I. P. , Wörner T. P., Moreira T., et al. 2024. “To 200,000 m/z and Beyond: Native Electron Capture Charge Reduction Mass Spectrometry Deconvolves Heterogeneous Signals in Large Biopharmaceutical Analytes.” ACS Central Science 10, no. 8: 1548–1561. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Leach, F. E. , Nagornov K. O., Kozhinov A. N., and Tsybin Y. O.. 2024. “External Data Systems Enable Enhanced (and Sustainable) Fourier Transform Mass Spectrometry Imaging for Legacy Hybrid Linear Ion Trap–Orbitrap Platforms.” Journal of the American Society for Mass Spectrometry 35, no. 11: 2690–2698. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Lebedev, A. T. , Damoc E., Makarov A. A., and Samgina T. Y.. 2014. “Discrimination of Leucine and Isoleucine in Peptides Sequencing With Orbitrap Fusion Mass Spectrometer.” Analytical Chemistry 86, no. 14: 7017–7022. [DOI] [PubMed] [Google Scholar]
- Lee, K. W. , Salome A. Z., Westphall M. S., Grant T., and Coon J. J.. 2023. “Onto Grid Purification and 3D Reconstruction of Protein Complexes Using Matrix‐Landing Native Mass Spectrometry.” Journal of Proteome Research 22, no. 3: 851–856. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Légaré, M. A. , Bélanger‐Chabot G., Dewhurst R. D., et al. 2018. “Nitrogen Fixation and Reduction at Boron.” Science 359, no. 6378: 896–900. [DOI] [PubMed] [Google Scholar]
- Li, G. , and Mordehai A. 2010. “Mass Spectrometer and Electric Field Source for Mass Spectrometer.” US Patent 7,755,040.
- Li, R. , Xia C., Wu S., et al. 2024. “Direct and Detailed Site‐Specific Glycopeptide Characterization by Higher‐Energy Electron‐Activated Dissociation Tandem Mass Spectrometry.” Analytical Chemistry 96, no. 3: 1251–1258. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Lim, H. K. , Chen J., Sensenhauser C., Cook K., and Subrahmanyam V.. 2007. “Metabolite Identification by Data‐Dependent Accurate Mass Spectrometric Analysis at Resolving Power of 60 000 in External Calibration Mode Using an LTQ/Orbitrap.” Rapid Communications in Mass Spectrometry 21, no. 12: 1821–1832. [DOI] [PubMed] [Google Scholar]
- Lorenz, M. , Wagner R., Proksch R., and Ovchinnikova O. S.. 2019. “Photoinduced Thermal Desorption on an Atomic Force Microscope Platform Coupled With Mass Spectrometry for Multimodal Imaging.” Microscopy and Microanalysis 25, no. S2: 1064–1065. [Google Scholar]
- Lutomski, C. A. , Bennett J. L., El‐Baba T. J., et al. 2025. “Defining Proteoform‐Specific Interactions for Drug Targeting in a Native Cell Signalling Environment.” Nature Chemistry 17, no. 2: 204–214. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Lutomski, C. A. , El‐Baba T. J., Hinkle J. D., et al. 2023. “Infrared Multiphoton Dissociation Enables Top‐Down Characterization of Membrane Protein Complexes and G Protein‐Coupled Receptors.” Angewandte Chemie International Edition 135, no. 36: e202305694. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Makarov, A. 1999. “Mass Spectrometer.” US Patent 5, no. 886: 346. [Google Scholar]
- Makarov, A. 2000. “Electrostatic Axially Harmonic Orbital Trapping: A High‐Performance Technique of Mass Analysis.” Analytical Chemistry 72, no. 6: 1156–1162. [DOI] [PubMed] [Google Scholar]
- Makarov, A. 2007. “Multi‐Electrode Ion Trap.” WO Pat. 2007/000587.
- Makarov, A. 2010. “Theory and Practice of the Orbitrap Mass Analyzer.” In Practical Aspects of Trapped Ion Mass Spectrometry (Vol. 4), edited by March R. E. and Todd J. F. J., CRC Press. [Google Scholar]
- Makarov, A. 2012. “Ion Detection.” WO Patent 2012/080352.
- Makarov, A. 2013. “Orbitrap Against All Odds.” Analytical Sciences 1, no. 10: 26–33. [Google Scholar]
- Makarov, A. 2017. “Method and Apparatus for Ion Injection Into an Electrostatic Trap.” US Patent 9,831,079.
- Makarov, A. 2019. “Orbitrap Journey: Taming the Ion Rings.” Nature Communications 10: 3743. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Makarov, A. 2025. “Acquisition of Ion Mobility‐ Mass to Charge Ratio Data. US Patent Application 2025/0062112.
- Makarov, A. , and Denisov E.. 2009. “Dynamics of Ions of Intact Proteins in the Orbitrap Mass Analyzer.” Journal of the American Society for Mass Spectrometry 20: 1486–1495. [DOI] [PubMed] [Google Scholar]
- Makarov, A. , Denisov E., Jung G., Balschun W., and Horning S. 2006. “Improvements in an Electrostatic Trap.” WO Patent 2006/129109.
- Makarov, A. , Denisov E., and Kholomeev A. 2005. “RF Power Supply for a Mass Spectrometer.” WO Paent. 2005/124821.
- Makarov, A. , Denisov E., Kholomeev A., et al. 2006. “Performance Evaluation of a Novel Hybrid Linear Ion Trap/Orbitrap Mass Spectrometer.” Analytical Chemistry 78, no. 7: 2113–2120. [DOI] [PubMed] [Google Scholar]
- Makarov, A. , Grechnikov A., Nikiforov S., Tyutyunnik O., and Denisov E.. 2013. “The Orbitrap Mass Analyzer With Direct Ion Injection Interfaced to a Laser Desorption/Ionization Ion Source.” Journal of Analytical Chemistry 68, no. 14: 1165–1169. [Google Scholar]
- Makarov, A. , Hardman M., Schwartz J., and Senko M. 2002. “Mass Spectrometry Method and Apparatus.” WO Patent 2002/078046.
- Makarov, A. , and Horning S. 2008. “Parallel Mass Analysis.” WO Patent 2008/080604.
- Makarov, A. , Monastyrski M., and Grinfeld D. 2008. “Ion Trap.” WO Patent 2008/081334.
- Maple, H. J. , Scheibner O., Baumert M., et al. 2014. “Application of the Exactive Plus EMR for Automated Protein–Ligand Screening by Non‐Covalent Mass Spectrometry.” Rapid Communications in Mass Spectrometry 28, no. 13: 1561–1568. [DOI] [PubMed] [Google Scholar]
- McAlister, G. C. , Berggren W. T., Griep‐Raming J., et al. 2008. “A Proteomics Grade Electron Transfer Dissociation‐Enabled Hybrid Linear Ion Trap–Orbitrap Mass Spectrometer.” Journal of Proteome Research 7: 3127–3136. [DOI] [PMC free article] [PubMed] [Google Scholar]
- McAlister, G. C. , Nusinow D. P., Jedrychowski M. P., et al. 2014. “Multinotch MS3 Enables Accurate, Sensitive, and Multiplexed Detection of Differential Expression Across Cancer Cell Line Proteomes.” Analytical Chemistry 86, no. 14: 7150–7158. [DOI] [PMC free article] [PubMed] [Google Scholar]
- McEwen, C. , and Gutteridge S.. 2007. “Analysis of the Inhibition of the Ergosterol Pathway in Fungi Using the Atmospheric Solids Analysis Probe (ASAP) Method.” Journal of the American Society for Mass Spectrometry 18, no. 7: 1274–1278. [DOI] [PubMed] [Google Scholar]
- McEwen, C. N. , Larsen B. S., and Trimpin S.. 2010. “Laserspray Ionization on a Commercial Atmospheric Pressure‐MALDI Mass Spectrometer Ion Source: Selecting Singly or Multiply Charged Ions.” Analytical Chemistry 82, no. 12: 4998–5001. [DOI] [PubMed] [Google Scholar]
- Meier, F. , Geyer P. E., Virreira Winter S., Cox J., and Mann M.. 2018. “Boxcar Acquisition Method Enables Single‐Shot Proteomics at a Depth of 10,000 Proteins in 100 Minutes.” Nature Methods 15, no. 6: 440–448. [DOI] [PubMed] [Google Scholar]
- Melani, R. D. , Srzentić K., Gerbasi V. R., et al. 2019. “Direct Measurement of Light and Heavy Antibody Chains Using Ion Mobility and Middle‐Down Mass Spectrometry.” mAbs 11, no. 8: 1351–1357. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Mertz, K. L. , Jordahl D., Hemme C. A., et al. 2025. “Laser‐Induced Rehydration of Cryo‐Landed Proteins Restores Native Structure.” Molecular & Cellular Proteomics 24, no. 6: 100987. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Michalski, A. , Damoc E., Hauschild J. P., et al. 2011. “Mass Spectrometry‐Based Proteomics Using Q Exactive, a High‐Performance Benchtop Quadrupole Orbitrap Mass Spectrometer.” Molecular & Cellular Proteomics 10, no. 9: M111.011015. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Michalski, A. , Damoc E., Lange O., et al. 2011. “Ultra High Resolution Linear Ion Trap Orbitrap Mass Spectrometer (Orbitrap Elite) Facilitates Top‐Down LC MS/MS and Versatile Peptide Fragmentation Modes.” Molecular & Cellular Proteomics 11, no. 3: O111.013698. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Monaghan, J. , Jaeger A., Agua A. R., et al. 2021. “A Direct Mass Spectrometry Method for the Rapid Analysis of Ubiquitous Tire‐Derived Toxin N‐(1, 3‐Dimethylbutyl)‐N′‐Phenyl‐p‐Phenylenediamine Quinone (6‐PPDQ).” Environmental Science & Technology Letters 8, no. 12: 1051–1056. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Moulard, Y. , Bailly‐Chouriberry L., Boyer S., Garcia P., Popot M. A., and Bonnaire Y.. 2011. “Use of Benchtop Exactive High Resolution and High Mass Accuracy Orbitrap Mass Spectrometer for Screening in Horse Doping Control.” Analytica Chimica Acta 700, no. 1–2: 126–136. [DOI] [PubMed] [Google Scholar]
- Mueller, E. P. , Sessions A. L., Sauer P. E., Weiss G. M., and Eiler J. M.. 2021. “Simultaneous, High‐Precision Measurements of δ2H and δ13C in Nanomole Quantities of Acetate Using Electrospray Ionization‐Quadrupole‐Orbitrap Mass Spectrometry.” Analytical Chemistry 94, no. 2: 1092–1100. [DOI] [PubMed] [Google Scholar]
- Muenster, H. , and Makarov A.. 2006. “Breakthrough Technology for Bottom‐Up Proteomics and Small Molecule Research.” American Laboratory 38, no. 14: 32. [Google Scholar]
- Munoz, G. , Vo Duy S., Budzinski H., Labadie P., Liu J., and Sauvé S.. 2015. “Quantitative Analysis of Poly‐and Perfluoroalkyl Compounds in Water Matrices Using High Resolution Mass Spectrometry: Optimization for a Laser Diode Thermal Desorption Method.” Analytica Chimica Acta 881: 98–106. [DOI] [PubMed] [Google Scholar]
- Nagornov, K. O. , Zennegg M., Kozhinov A. N., Tsybin Y. O., and Bleiner D.. 2020. “Trace‐Level Persistent Organic Pollutant Analysis With Gas‐Chromatography Orbitrap Mass Spectrometry—Enhanced Performance by Complementary Acquisition and Processing of Time‐Domain Data.” Journal of the American Society for Mass Spectrometry 31, no. 2: 257–266. [DOI] [PubMed] [Google Scholar]
- Neumann, A. P. , Sage E., Boll D., et al. 2024. “A Hybrid Orbitrap‐Nanoelectromechanical Systems Approach for the Analysis of Individual, Intact Proteins in Real Time.” Angewandte Chemie International Edition 63: e202317064. [DOI] [PubMed] [Google Scholar]
- Ni, Z. , and Arevalo R.. 2024. “Collision Cross‐Section Measurements of Small Molecules via Transient Decay Profiles Observed in Orbitrap Mass Analyzers.” Rapid Communications in Mass Spectrometry 38, no. 20: e9887. [DOI] [PubMed] [Google Scholar]
- Niehaus, M. , Soltwisch J., Belov M. E., and Dreisewerd K.. 2019. “Transmission‐Mode MALDI‐2 Mass Spectrometry Imaging of Cells and Tissues at Subcellular Resolution.” Nature Methods 16, no. 9: 925–931. [DOI] [PubMed] [Google Scholar]
- Núñez, O. , Gallart‐Ayala H., Martins C. P. B., Moyano E., and Galceran M. T.. 2012. “Atmospheric Pressure Photoionization Mass Spectrometry of Fullerenes.” Analytical Chemistry 84, no. 12: 5316–5326. [DOI] [PubMed] [Google Scholar]
- O'Brien, J. P. , Needham B. D., Henderson J. C., Nowicki E. M., Trent M. S., and Brodbelt J. S.. 2014. “193 nm Ultraviolet Photodissociation Mass Spectrometry for the Structural Elucidation of Lipid A Compounds in Complex Mixtures.” Analytical Chemistry 86, no. 4: 2138–2145. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Ogrinc, N. , Saudemont P., Balog J., et al. 2019. “Water‐Assisted Laser Desorption/Ionization Mass Spectrometry for Minimally Invasive In Vivo and Real‐Time Surface Analysis Using Spidermass.” Nature Protocols 14, no. 11: 3162–3182. [DOI] [PubMed] [Google Scholar]
- Olsen, J. V. , de Godoy L. M. F., Li G., et al. 2005. “Parts Per Million Mass Accuracy on an Orbitrap Mass Spectrometer via Lock Mass Injection into a C‑Trap.” Molecular & Cellular Proteomics 4, no. 12: 2010–2021. [DOI] [PubMed] [Google Scholar]
- Olsen, J. V. , Macek B., Lange O., Makarov A., Horning S., and Mann M.. 2007. “Higher‑Energy C‑Trap Dissociation for Peptide Modification Analysis.” Nature Methods 4, no. 9: 709–712. [DOI] [PubMed] [Google Scholar]
- Olsen, J. V. , Schwartz J. C., Griep‐Raming J., et al. 2009. “A Dual Pressure Linear Ion Trap—Orbitrap Instrument With Very High Sequencing Speed.” Molecular & Cellular Proteomics 8: 2759–2769. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Pagnotti, V. S. , Chubatyi N. D., and McEwen C. N.. 2011. “Solvent Assisted Inlet Ionization: An Ultrasensitive New Liquid Introduction Ionization Method for Mass Spectrometry.” Analytical Chemistry 83, no. 11: 3981–3985. [DOI] [PubMed] [Google Scholar]
- Panchaud, A. , Scherl A., Shaffer S. A., et al. 2009. “Precursor Acquisition Independent From Ion Count: How to Dive Deeper into the Proteomics Ocean.” Analytical Chemistry 81, no. 15: 6481–6488. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Panepinto, M. C. , Minicozzi F., and Cifani P.. 2025. “High Dynamic Range Peptide Mass Spectrometry Using Segmented Precursor Ion Accumulation.” Analytical Chemistry 97: 23409–23418. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Papanastasiou, D. , Kounadis D., Lekkas A., et al. 2022. “The Omnitrap Platform: A Versatile Segmented Linear Ion Trap for Multidimensional Multiple‐Stage Tandem Mass Spectrometry.” Journal of the American Society for Mass Spectrometry 33, no. 10: 1990–2007. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Partovi, F. , Mikkilä J., Iyer S., et al. 2023. “Pesticide Residue Fast Screening Using Thermal Desorption Multi‐Scheme Chemical Ionization Mass Spectrometry (TD‐MION MS) With Selective Chemical Ionization.” ACS Omega 8, no. 29: 25749–25757. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Passarelli, M. K. , Pirkl A., Moellers R., et al. 2017. “The 3D OrbiSIMS—Label‐Free Metabolic Imaging With Subcellular Lateral Resolution and High Mass‐Resolving Power.” Nature Methods 14: 1175–1183. [DOI] [PubMed] [Google Scholar]
- Pekar Second, T. , Blethrow J. D., Schwartz J. C., et al. 2009. “Dual‐Pressure Linear Ion Trap Mass Spectrometer Improving the Analysis of Complex Protein Mixtures.” Analytical Chemistry 81, no. 18: 7757–7765. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Peris‐Díaz, M. D. , Barendregt A., Wörner T. P., et al. 2025. “Electron Capture‐Induced Charge Reduction Benefits the Recording of Ultralong Transients in Orbitrap‐Based Individual‐Ion Mass Spectrometry.” Analytical Chemistry 97, no. 22: 11653–11660. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Pernot, P. , Carrasco N., Thissen R., and Schmitz‐Afonso I.. 2010. “Tholinomics—Chemical Analysis of Nitrogen‐Rich Polymers.” Analytical Chemistry 82, no. 4: 1371–1380. [DOI] [PubMed] [Google Scholar]
- Perry, R. H. , Cooks R. G., and Noll R. J.. 2008. “Orbitrap Mass Spectrometry: Instrumentation, Ion Motion and Applications.” Mass Spectrometry Reviews 27, no. 6: 661–699. [DOI] [PubMed] [Google Scholar]
- Peterson, A. C. , Hauschild J. P., Quarmby S. T., et al. 2014. “Development of a GC/Quadrupole‐Orbitrap Mass Spectrometer, Part I: Design and Characterization.” Analytical Chemistry 86, no. 20: 10036–10043. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Peterson, A. C. , McAlister G. C., Quarmby S. T., Griep‐Raming J., and Coon J. J.. 2010. “Development and Characterization of a GC‐Enabled QLT‐Orbitrap for High‐Resolution and High‐Mass Accuracy GC/MS.” Analytical Chemistry 82, no. 20: 8618–8628. [DOI] [PubMed] [Google Scholar]
- Peterson, A. C. , Russell J. D., Bailey D. J., Westphall M. S., and Coon J. J.. 2012. “Parallel Reaction Monitoring for High Resolution and High Mass Accuracy Quantitative, Targeted Proteomics.” Molecular & Cellular Proteomics 11, no. 11: 1475–1488. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Petrosius, V. , Aragon‐Fernandez P., Arrey T. N., et al. 2025. “Quantitative Label‐Free Single‐Cell Proteomics on the Orbitrap Astral MS.” Molecular & Cellular Proteomics 24: 100982. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Pfammatter, S. , Bonneil E., McManus F. P., et al. 2018. “A Novel Differential Ion Mobility Device Expands the Depth of Proteome Coverage and the Sensitivity of Multiplex Proteomic Measurements.” Molecular & Cellular Proteomics 17, no. 10: 2051–2067. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Poltash, M. L. , McCabe J. W., Shirzadeh M., Laganowsky A., Clowers B. H., and Russell D. H.. 2018. “Fourier Transform‐Ion Mobility‐Orbitrap Mass Spectrometer: A Next‐Generation Instrument for Native Mass Spectrometry.” Analytical Chemistry 90, no. 17: 10472–10478. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Quarles, C. D. , Carado A. J., Barinaga C. J., Koppenaal D. W., and Marcus R. K.. 2012. “Liquid Sampling–Atmospheric Pressure Glow Discharge (LS‐APGD) Ionization Source for Elemental Mass Spectrometry: Preliminary Parametric Evaluation and Figures of Merit.” Analytical and Bioanalytical Chemistry 402, no. 1: 261–268. [DOI] [PubMed] [Google Scholar]
- Riley, N. M. , Westphall M. S., Hebert A. S., and Coon J. J.. 2017. “Implementation of Activated Ion Electron Transfer Dissociation on a Quadrupole‐Orbitrap‐Linear Ion Trap Hybrid Mass Spectrometer.” Analytical Chemistry 89, no. 12: 6358–6366. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Roach, P. J. , Laskin J., and Laskin A.. 2010. “Nanospray Desorption Electrospray Ionization: An Ambient Method for Liquid‐Extraction Surface Sampling in Mass Spectrometry.” Analyst 135, no. 9: 2233–2236. [DOI] [PubMed] [Google Scholar]
- Rochat, B. , Kottelat E., and McMullen J.. 2012. “The Future Key Role of LC–High‐Resolution‐MS Analyses in Clinical Laboratories: A Focus on Quantification.” Bioanalysis 4, no. 24: 2939–2958. [DOI] [PubMed] [Google Scholar]
- Romão, W. , Tose L. V., Vaz B. G., et al. 2016. “Petroleomics by Direct Analysis in Real Time–Mass Spectrometry.” Journal of the American Society for Mass Spectrometry 27: 182–185. [DOI] [PubMed] [Google Scholar]
- Rosati, S. , Rose R. J., Thompson N. J., et al. 2012. “Exploring an Orbitrap Analyzer for the Characterization of Intact Antibodies by Native Mass Spectrometry.” Angewandte Chemie International Edition 51, no. 52: 12992–12996. [DOI] [PubMed] [Google Scholar]
- Rose, R. J. , Damoc E., Denisov E., Makarov A., and Heck A. J. R.. 2012. “High‐Sensitivity Orbitrap Mass Analysis of Intact Macromolecular Assemblies.” Nature Methods 9: 1084–1086. [DOI] [PubMed] [Google Scholar]
- Saba, J. , Bonneil E., Pomiès C., Eng K., and Thibault P.. 2009. “Enhanced Sensitivity in Proteomics Experiments Using Faims Coupled With a Hybrid Linear Ion Trap/Orbitrap Mass Spectrometer.” Journal of Proteome Research 8, no. 7: 3355–3366. [DOI] [PubMed] [Google Scholar]
- Salome, A. Z. , Lee K. W., Grant T., Westphall M. S., and Coon J. J.. 2022. “Matrix‐Landing Mass Spectrometry for Electron Microscopy Imaging of Native Protein Complexes.” Analytical Chemistry 94, no. 50: 17616–17624. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Sanders, J. D. , Grinfeld D., Aizikov K., Makarov A., Holden D. D., and Brodbelt J. S.. 2018. “Determination of Collision Cross‐Sections of Protein Ions in an Orbitrap Mass Analyzer.” Analytical Chemistry 90, no. 9: 5896–5902. [DOI] [PubMed] [Google Scholar]
- Saparbaev, E. , Yamaletdinov R., and Boyarkin O. V.. 2021. “Identification of Isomeric Lipids by UV Spectroscopy of Noncovalent Complexes With Aromatic Molecules.” Analytical Chemistry 93, no. 38: 12822–12826. [DOI] [PubMed] [Google Scholar]
- Schachner, L. F. , Mullen C., Phung W., et al. 2024. “Exposing the Molecular Heterogeneity of Glycosylated Biotherapeutics.” Nature Communications 15, no. 1: 3259. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Scheltema, R. A. , Hauschild J. P., Lange O., et al. 2014. “The Q Exactive HF, a Benchtop Mass Spectrometer With a Pre‐Filter, High‐Performance Quadrupole and an Ultra‐High Field Orbitrap Analyzer.” Molecular & Cellular Proteomics 13: 3698–3708. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Schmidt, E. M. , Pudenzi M. A., Santos J. M., et al. 2018. “Petroleomics via Orbitrap Mass Spectrometry With Resolving Power above 1,000,000 at m/z 200.” RSC Advances 8: 6183–6191. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Schuhmann, K. , Almeida R., Baumert M., Herzog R., Bornstein S. R., and Shevchenko A.. 2012. “Shotgun Lipidomics on a LTQ Orbitrap Mass Spectrometer by Successive Switching Between Acquisition Polarity Modes.” Journal of Mass Spectrometry 47, no. 1: 96–104. [DOI] [PubMed] [Google Scholar]
- Scigelova, M. , and Makarov A.. 2006. “Orbitrap Mass Analyzer: Overview and Applications in Proteomics.” Proteomics 6: 16–21. [DOI] [PubMed] [Google Scholar]
- Scigelova, M. , and Makarov A.. 2009. “Advances in Bioanalytical LC–MS Using the Orbitrap Mass Analyzer.” Bioanalysis 1, no. 4: 741–754. [DOI] [PubMed] [Google Scholar]
- Senko, M. W. , Remes P. M., Canterbury J. D., et al. 2013. “Novel Parallelized Quadrupole/Linear Ion Trap/Orbitrap Tribrid Mass Spectrometer Improving Proteome Coverage and Peptide Identification Rates.” Analytical Chemistry 85, no. 24: 11710–11714. [DOI] [PubMed] [Google Scholar]
- Serrano, L. R. , Peters‐Clarke T. M., Arrey T. N., et al. 2024. “The One Hour Human Proteome.” Molecular & Cellular Proteomics 23: 100760. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Shaw, J. B. , Li W., Holden D. D., et al. 2013. “Complete Protein Characterization Using Top‐Down Mass Spectrometry and Ultraviolet Photodissociation.” Journal of the American Chemical Society 135, no. 34: 12646–12651. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Shaw, J. B. , Malhan N., Vasil'ev Y. V., et al. 2018. “Sequencing Grade Tandem Mass Spectrometry for Top–Down Proteomics Using Hybrid Electron Capture Dissociation Methods in a Benchtop Orbitrap Mass Spectrometer.” Analytical Chemistry 90, no. 18: 10819–10827. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Simirgiotis, M. , Quispe C., Bórquez J., Areche C., and Sepúlveda B.. 2016. “Fast Detection of Phenolic Compounds in Extracts of Easter Pears (Pyrus communis) From the Atacama Desert by Ultrahigh‐Performance Liquid Chromatography and Mass Spectrometry (UHPLC–Q/Orbitrap/MS/MS).” Molecules 21, no. 1: 92. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Singh, K. D. , Del Miguel G. V., Gaugg M. T., et al. 2018. “Translating Secondary Electrospray Ionization–High‐Resolution Mass Spectrometry to the Clinical Environment.” Journal of Breath Research 12, no. 2: 027113. [DOI] [PubMed] [Google Scholar]
- Singh, K. D. , Tancev G., Decrue F., et al. 2019. “Standardization Procedures for Real‐Time Breath Analysis by Secondary Electrospray Ionization High‐Resolution Mass Spectrometry.” Analytical and Bioanalytical Chemistry 411, no. 19: 4883–4898. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Skoblin, M. G. , Kopaev I. A., Greenfield D. E., Makarov A. A., Monastyrskiy M. A., and Alimpiev S. S.. 2015. “Influence of Coulomb Effects on the Resolving Power of Multi‐Reflection Mass‐Spectrometer Systems.” Quantum Electronics 45, no. 12: 1171–1178. [Google Scholar]
- Smith, E. A. , and Lee Y. J.. 2010. “Petroleomic Analysis of Bio‐Oils From the Fast Pyrolysis of Biomass: Laser Desorption Ionization‐Linear Ion Trap‐Orbitrap Mass Spectrometry Approach.” Energy & Fuels 24: 5190–5198. [Google Scholar]
- Smyrnakis, A. , Levin N., Kosmopoulou M., et al. 2023. “Characterization of an Omnitrap‐Orbitrap Platform Equipped With Infrared Multiphoton Dissociation, Ultraviolet Photodissociation, and Electron Capture Dissociation for the Analysis of Peptides and Proteins.” Analytical Chemistry 95, no. 32: 12039–12046. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Snijder, J. , Van de Waterbeemd M., Damoc E., et al. 2014. “Defining the Stoichiometry and Cargo Load of Viral and Bacterial Nanoparticles by Orbitrap Mass Spectrometry.” Journal of the American Chemical Society 136, no. 20: 7295–7299. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Stafford, Jr G. C., Taylor D. M., and Bradshaw S. C.. 1992. “Method of Increasing the Dynamic Range and Sensitivity of a Quadrupole Ion Trap Mass Spectrometer.” US Patent US 5, no. 107: 109. [Google Scholar]
- Steigerwald, S. , Sinha A., Fort K. L., et al. 2024. “Full Mass Range ΦSDM Orbitrap Mass Spectrometry for DIA Proteome Analysis.” Molecular & Cellular Proteomics 23, no. 2: 100713. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Stewart, H. , Grinfeld D., Giannakopulos A., Makarov A. A., and Hock C.. 2024. “Proof of Principle for Enhanced Resolution Multi‐Pass Methods for the Astral Analyzer.” International Journal of Mass Spectrometry 498: 117203. [Google Scholar]
- Stewart, H. , Grinfeld D., Wagner A., et al. 2024. “A Conjoined Rectilinear Collision Cell and Pulsed Extraction Ion Trap With Auxiliary DC Electrodes.” Journal of the American Society for Mass Spectrometry 35, no. 1: 74–81. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Stewart, H. I. , Grinfeld D., Giannakopulos A., et al. 2023. “Parallelized Acquisition of Orbitrap and Astral Analyzers Enables High‐Throughput Quantitative Analysis.” Analytical Chemistry 95, no. 42: 15656–15664. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Strasser, L. , Füssl F., Morgan T. E., Carillo S., and Bones J.. 2023. “Exploring Charge‐Detection Mass Spectrometry on Chromatographic Time Scales.” Analytical Chemistry 95, no. 40: 15118–15124. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Strupat, K. , Kovtoun V., Bui H., Viner R., Stafford G., and Horning S.. 2009. “MALDI Produced Ions Inspected With a Linear Ion Trap‐Orbitrap Hybrid Mass Analyzer.” Journal of the American Society for Mass Spectrometry 20, no. 8: 1451–1463. [Google Scholar]
- Suzhou Anyi Spectrum . 2025. “ANYEEP GC Cassitrap 120K Quadrupole Fourier Transform Electrostatic Trap Gas Mass Coupled Instrument.” https://zhuanlan.zhihu.com/p/1982409022406693938 and http://www.anyeep.com/chanpin/83.html.
- Syka, J. E. P. , Coon J. J., Schroeder M. J., Shabanowitz J., and Hunt D. F.. 2004. “Peptide and Protein Sequence Analysis by Electron Transfer Dissociation Mass Spectrometry.” Proceedings of the National Academy of Sciences 101, no. 26: 9528–9533. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Thevis, M. , Makarov A. A., Horning S., and Schänzer W.. 2005. “Mass Spectrometry of Stanozolol and Its Analogues Using Electrospray Ionization and Collision‐Induced Dissociation With Quadrupole‐Linear Ion Trap and Linear Ion Trap‐Orbitrap Hybrid Mass Analyzers.” Rapid Communications in Mass Spectrometry 19, no. 22: 3369–3378. [DOI] [PubMed] [Google Scholar]
- Tolstoy, L. 2001. “Anna Karenina (Original Work Published 1875–1877) (R. P. L. Volokhonsky, Trans.). Viking Penguin.
- Trimpin, S. , Yenchick F. S., Lee C., et al. 2024. “New Processes for Ionizing Nonvolatile Compounds in Mass Spectrometry: The Road of Discovery to Current State‐of‐the‐Art.” Journal of the American Society for Mass Spectrometry 35, no. 12: 2753–2784. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Trindade, G. F. , Sul S., Kim J., et al. 2023. “Direct Identification of Interfacial Degradation in Blue OLEDs Using Nanoscale Chemical Depth Profiling.” Nature Communications 14, no. 1: 8066. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Van de Waterbeemd, M. , Fort K. L., Boll D., et al. 2017. “High‐Fidelity Mass Analysis Unveils Heterogeneity in Intact Ribosomal Particles.” Nature Methods 14: 283–286. [DOI] [PubMed] [Google Scholar]
- VanAernum, Z. L. , Gilbert J. D., Belov M. E., Makarov A. A., Horning S. R., and Wysocki V. H.. 2019. “Surface‐Induced Dissociation of Noncovalent Protein Complexes in an Extended Mass Range Orbitrap Mass Spectrometer.” Analytical Chemistry 91, no. 5: 3611–3618. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Vasicek, L. A. , Ledvina A. R., Shaw J., et al. 2011. “Implementing Photodissociation in an Orbitrap Mass Spectrometer.” Journal of the American Society for Mass Spectrometry 22, no. 6: 1105–1108. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Wang, J. , Jia Z., Zhang Z., et al. 2017. “Analysis of Chemical Constituents of Melastoma Dodecandrum Lour. by UPLC‐ESI‐Q‐Exactive Focus‐MS/MS.” Molecules 22, no. 3: 476. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Wei, J. , Papanastasiou D., Kosmopoulou M., et al. 2023. “De Novo Glycan Sequencing by Electronic Excitation Dissociation MS 2‐guided MS 3 Analysis on an Omnitrap‐Orbitrap Hybrid Instrument.” Chemical Science 14, no. 24: 6695–6704. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Weidt, S. , Haggarty J., Kean R., et al. 2016. “A Novel Targeted/Untargeted GC‐Orbitrap Metabolomics Methodology Applied to Candida Albicans and Staphylococcus aureus Biofilms.” Metabolomics 12, no. 12: 189. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Westphall, M. S. , Lee K. W., Hemme C., et al. 2023. “Cryogenic Soft Landing Improves Structural Preservation of Protein Complexes.” Analytical Chemistry 95: 15094–15101. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Westphall, M. S. , Lee K. W., Salome A. Z., Lodge J. M., Grant T., and Coon J. J.. 2022. “Three‐Dimensional Structure Determination of Protein Complexes Using Matrix‐Landing Mass Spectrometry.” Nature Communications 13, no. 1: 2276. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Willhite, L. N. , Southard A. E., Bardyn A., et al. 2020. “CosmOrbitrap Consortium. Characterization of Regolith and Trace Economic Resources (CRATER) Instrument: Integration Into the Artemis Program.” LPI Contrib 2241: 5104. [Google Scholar]
- Wörner, T. P. , Aizikov K., Snijder J., Fort K. L., Makarov A. A., and Heck A. J. R.. 2022. “Frequency Chasing of Individual Megadalton Ions in an Orbitrap Analyser Improves Precision of Analysis in Single‐Molecule Mass Spectrometry.” Nature Chemistry 14, no. 5: 515–522. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Wörner, T. P. , Snijder J., Bennett A., Agbandje‐McKenna M., Makarov A. A., and Heck A. J. R.. 2020. “Resolving Heterogeneous Macromolecular Assemblies by Orbitrap‐Based Single‐Particle Charge Detection Mass Spectrometry.” Nat Meth 17: 395–398. [DOI] [PubMed] [Google Scholar]
- Zeichner, S. S. , Wilkes E. B., Hofmann A. E., et al. 2022. “Methods and Limitations of Stable Isotope Measurements via Direct Elution of Chromatographic Peaks Using Gas Chromatography‐Orbitrap Mass Spectrometry.” International Journal of Mass Spectrometry 477: 116848. [Google Scholar]
- Zhang, J. , Brown J., Scurr D. J., et al. 2020. “Cryo‐OrbiSIMS for 3D Molecular Imaging of a Bacterial Biofilm in Its Native State.” Analytical Chemistry 92, no. 13: 9008–9015. [DOI] [PubMed] [Google Scholar]
- Zhang, J. , Rector J., Lin J. Q., et al. 2017. “Nondestructive Tissue Analysis for Ex Vivo and In Vivo Cancer Diagnosis Using a Handheld Mass Spectrometry System.” Science Translational Medicine 9, no. 406: eaan3968. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Zhurov, K. O. , Kozhinov A. N., and Tsybin Y. O.. 2013. “Evaluation of High‐Field Orbitrap Fourier Transform Mass Spectrometer for Petroleomics.” Energy & Fuels 27, no. 6: 2974–2983. [Google Scholar]
- Zviagin, A. , Kopysov V., Kozlovskii V., Chernyshev D., Makarov A., and Boyarkin O. V.. 2025. “Identification of Isomeric Metabolites Using Cold Ion Spectroscopy Add‐On to Orbitrap‐Based Mass Spectrometer.” Analytical Chemistry 97, no. 27: 14097–14102. [DOI] [PubMed] [Google Scholar]