Abstract
Background
Subcutaneous implantable cardioverter defibrillators (S-ICD) are extensively used in the primary and secondary prevention of sudden cardiac death. Ventricular fibrillation (VF) induction during implantation remains standard to confirm device functionality.
Objective
To compare electrocardiographic (ECG) and technical variables between induced and spontaneous ventricular arrhythmias in S-ICD recipients.
Methods
A retrospective observational study was performed on 10 consecutive patients with S-ICD. Variables included episode duration, sensing time, charge time, and shock impedance. For each patient, the implant induced VF episode was compared to at least one spontaneous event during follow-up.
Results
The mean age was 36.4 years (range: 22–61); 2 patients were male. All patients received intermuscular S-ICD implants, with the lead placed in the left parasternal position in 7 patients and right parasternal in 3. Episode duration and sensing time were consistently shorter in spontaneous events, and shock impedance was lower. Charge time, however, was longer in spontaneous events (by 1.4 to 1.8 s).
Conclusions
In this case series, spontaneous ventricular arrhythmic events in S-ICD patients demonstrated shorter sensing and therapy delivery times compared with induced ventricular fibrillation episodes. A measurable Post-Induction Residual Time (PRT) was identified in all induced episodes and was absent during spontaneous ventricular arrhythmias, suggesting that 50-Hz-induced VF may not fully reproduce spontaneous arrhythmic events from the perspective of device sensing performance. These findings support the hypothesis that part of the delay observed during implant testing may be related to the induction process itself rather than to the intrinsic sensing capabilities of the S-ICD. Further studies are warranted to confirm these observations and to evaluate alternative VF induction strategies in future device platforms. Future integration of S-ICD systems with leadless pacing technologies may provide alternative ventricular fibrillation induction strategies, including T-wave shock delivery during either sensed intrinsic rhythm or leadless-paced rhythm. Such approaches could potentially avoid the occurrence of Post-Induction Residual Time (PRT) and may provide a model that more closely reproduces spontaneous ventricular fibrillation than current 50-Hz induction techniques.
Keywords: ICD (implantable cardioverter-defibrillator), sensing, S-ICD, sudden cardiac death, ventricular fibrillation
Introduction
The subcutaneous implantable cardioverter-defibrillator (S-ICD) is an established alternative to transvenous systems for selected patients requiring defibrillation therapy. A key step in the implant procedure involves inducing ventricular fibrillation (VF) to validate the device's detection and therapeutic response. However, it remains unclear whether induced arrhythmias accurately reflect the system's performance during spontaneous episodes (1–5).
This study aimed to analyze the differences in detection times for induced arrhythmias compared with spontaneous episode detection times, as well as variables such as shock impedance and charging time, in a series of patients with stored subcutaneous ECG data.
Methods
An observational case series was conducted in 10 consecutive patients who received a Boston Scientific EMBLEM MRI S-ICD for secondary prevention of sudden cardiac death. Implantation was intermuscular in all cases. Data were obtained from clinical records and stored device information. Stored electrocardiographic recordings from induced and spontaneous ventricular arrhythmias were reviewed in all patients.
The main variables analyzed included total episode duration, sensing time (from arrhythmia onset to device detection), charging time, and shock impedance. In addition, we evaluated a novel parameter termed Post-Induction Residual Time (PRT). Post-Induction Residual Time (PRT) was defined as the interval between termination of the 50-Hz ventricular fibrillation induction burst and the first valid ventricular sensing marker recognized by the S-ICD detection algorithm. PRT was measured in all induced episodes and was not observed in spontaneous ventricular arrhythmias.
Statistical analysis
Descriptive statistics were performed using IBM SPSS Statistics v26 and GraphPad Prism v9. Continuous variables were expressed as mean ± standard deviation (SD), categorical data are expressed as numbers (percentage), and paired comparisons between induced and spontaneous events were analyzed using the Wilcoxon signed-rank test. A p-value <0.05 was considered statistically significant.
Results
The analysis included patients with idiopathic VF (60%), idiopathic dilated cardiomyopathy (30%), and Brugada syndrome (10%), with one patient having a history of previous infection. (Table 1).
Table 1.
Baseline demographic characteristics of included participants.
| Baseline demographic characteristics of included participants | |||||
|---|---|---|---|---|---|
| Variable | n (or mean) | % | IC 95%b | mean std | |
| Sex | |||||
| Male | 2 | 20% | 3% | 56% | ——— |
| Female | 8 | 80% | 44% | 97% | ——— |
| age | 36,4 | ——— | 26,12 | 46,67 | 4,54 |
| Diagnostic | |||||
| IVF | 6 | 60% | 26% | 88% | ——— |
| IDC | 3 | 30% | 7% | 65% | ——— |
| Brugada | 1 | 10% | 0% | 45% | ——— |
| Medication | |||||
| amiodaronea | 3 | 30% | 7% | 65% | ——— |
| no drugs | 7 | 70% | 35% | 93% | ——— |
| Indication | |||||
| secondary | 10 | 100% | 69% | 100% | ——— |
| Implant type | |||||
| intermuscular | 10 | 100% | 69% | 100% | ——— |
Patients with IDC.
Confidence intervals under Clopper Pearson for all variables, except age (t-interval for the mean).
The mean age was 36.4 years (range: 22–61), and 2 patients were male. The average implantation time was 59.3 min (range: 38–75 min).
Lead position was left parasternal in 7 cases and right parasternal in 3 cases. Vector configurations included 2/3 sensing vectors in 7 patients and 3/3 in 3 patients. Overall, a measurable post-induction interval was present in all induced episodes and absent during spontaneous ventricular arrhythmias.
The main variables analyzed included total episode duration, sensing time (from onset to detection), charging time, and shock impedance.
Total event duration: Decreased on spontaneous episodes by 2.94 s (95% CI [−4.967; −0.913]) (Figure 1).
Figure 1.

With a p-value of less than 1%, a significant difference is observed in the total mean times of induced and spontaneous events (spontaneous events are statistically smaller).
Sensing time: Decreased on spontaneous VF episodes by 3.1 s (95% CI [−4.824; −1.376]).
Charging time: Longer during spontaneous episodes by an average of 0.71 s, although the difference was not statistically significant.
Shock impedance: Lower during spontaneous events, suggesting improved current delivery dynamics; however, impedance did not show a significant difference (Figure 2).
Figure 2.

The figure graph presents a comparison of differences in four key variables.
These findings suggest improved device efficiency, with significant reductions in both total detection and sensing times during spontaneous episodes. However, differences in episode timing may also be influenced by factors beyond sensing/detection alone, including the context and therapy parameters applied during spontaneous events versus induction testing.
Discussion
This preliminary analysis suggests that spontaneous ventricular arrhythmias may be detected and treated more efficiently by S-ICD than induced episodes. The observed post-induction interval after 50-Hz induction indicates that VF induced by an alternating-current protocol may not fully reproduce spontaneous VF from the perspective of device sensing dynamics. Importantly, this observation should not be interpreted as a device malfunction; rather, it reflects an outcome shaped by the induction procedure and the shared sensing/detection context across induced cases. (Figure 3).
Figure 3.

(A) VF event induced with 50 Hz with a post-induction time (PRT) of 1.6 s and a total episode time of 21.6 s (B) Spontaneous event in the same patient performing daily tasks, which, being an episode of spontaneous arrhythmia, does not have a PRT.
A key mechanistic interpretation is that the 50-Hz induction process may transiently affect sensing recovery, creating a brief interval before stable ventricular sensing is re-established. Consistent with this, the interval was not present during spontaneous ventricular arrhythmias, supporting the hypothesis that it is related to induction conditions rather than intrinsic differences in spontaneous VF physiology.
These results have methodological implications for how induction findings are extrapolated to real-world sensing/detection performance. This limitation should be considered alongside prior work describing the constraints of VF induction protocols and their relationship to device performance metrics (6, 7). In particular, evidence from the PRAETORIAN-DFT trial highlights that VF testing and induction strategies are tightly linked to how performance endpoints are operationalized and should be interpreted within the context of the testing paradigm (8). Previous S-ICD studies and registries have demonstrated the clinical performance of the system across different patient populations and clinical settings (9–12); however, these studies were not specifically designed to compare sensing dynamics between induced and spontaneous ventricular arrhythmias.
Despite the small sample size, these findings warrant further investigation in larger prospective studies. Future modular systems integrating leadless pacing technologies may enable alternative VF induction strategies, including T-wave shock induction, potentially reducing or avoiding the post-induction interval and providing a model that more closely resembles spontaneous ventricular fibrillation.
Conclusions
In this case series, spontaneous ventricular arrhythmic events in S-ICD patients demonstrated shorter sensing and therapy delivery times compared with induced ventricular fibrillation episodes. A measurable post-induction interval (PRT) was identified in all induced episodes and was absent during spontaneous ventricular arrhythmias, suggesting that 50-Hz-induced VF may not fully reproduce spontaneous arrhythmic events from the perspective of device sensing dynamics.
These findings support the hypothesis that part of the delay observed during implant testing may be related to the induction process itself rather than intrinsic sensing capabilities of the S-ICD. Further studies are warranted to confirm these observations and to evaluate alternative VF induction strategies in future device platforms.
Future integration of S-ICD systems with leadless pacing technologies may provide alternative VF induction strategies, including T-wave shock delivery on either sensed intrinsic rhythm or leadless-paced rhythm. Such approaches could potentially avoid the post-induction interval and may provide a model that more closely resembles spontaneous ventricular fibrillation than current 50-Hz induction techniques.
Funding Statement
The author(s) declared that financial support was not received for this work and/or its publication.
Footnotes
Edited by: Sergio Dubner, Sanatorio De Los Arcos, Argentina
Reviewed by: Yoshiyasu Aizawa, Nippon Medical School, Japan
Federico Migliore, University of Padua, Italy
Oswaldo Gutierrez, University of Costa Rica, Costa Rica
Data availability statement
The original contributions presented in the study are included in the article/Supplementary Material, further inquiries can be directed to the corresponding author.
Ethics statement
Ethical approval was not required for the study involving humans in accordance with the local legislation and institutional requirements. Written informed consent to participate in this study was not required from the participants or the participants’ legal guardians/next of kin in accordance with the national legislation and the institutional requirements. Written informed consent was obtained from the individual(s) for the publication of any potentially identifiable images or data included in this article.
Author contributions
CL: Conceptualization, Formal analysis, Investigation, Methodology, Project administration, Validation, Writing – original draft, Writing – review & editing. IM: Conceptualization, Data curation, Investigation, Methodology, Supervision, Validation, Writing – original draft, Writing – review & editing. MB: Data curation, Formal analysis, Investigation, Methodology, Project administration, Supervision, Writing – review & editing. LT: Conceptualization, Data curation, Formal analysis, Investigation, Methodology, Project administration, Supervision, Validation, Writing – review & editing. EA: Data curation, Supervision, Validation, Writing – review & editing. MA: Investigation, Supervision, Validation, Writing – review & editing. JA: Data curation, Supervision, Validation, Writing – review & editing. JC: Data curation, Supervision, Validation, Writing – review & editing.
Conflict of interest
Carlos Lopez has served as a Technical Consultant to Boston Scientific Argentina. Ignacio Mondragon and Leandro Tomas have served as proctors for Boston Scientific in the Southern Cone region. These relationships were not involved in the study design, data collection, analysis, interpretation, or manuscript preparation. The remaining authors declared that this work was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Generative AI statement
The author(s) declared that generative AI was not used in the creation of this manuscript.
Any alternative text (alt text) provided alongside figures in this article has been generated by Frontiers with the support of artificial intelligence and reasonable efforts have been made to ensure accuracy, including review by the authors wherever possible. If you identify any issues, please contact us.
Publisher's note
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.
References
- 1.Al-Khatib SM, Stevenson WG, Ackerman MJ, Bryant WJ, Callans DJ, Curtis AB, et al. 2017 AHA/ACC/HRS guideline for management of patients with ventricular arrhythmias and the prevention of sudden cardiac death. J Am Coll Cardiol. (2018) 72(14):e91–e220. 10.1016/j.jacc.2017.10.054 [DOI] [PubMed] [Google Scholar]
- 2.Knops RE, Olde Nordkamp LRA, Delnoy P-PHM, Boersma LVA, Kuschyk J, El-Chami MF, et al. Subcutaneous or transvenous defibrillator therapy. N Engl J Med. (2020) 383(6):526–36. 10.1056/NEJMoa1915932 [DOI] [PubMed] [Google Scholar]
- 3.Burke MC, Gold MR, Knight BP, Barr CS, Theuns DAMJ, Boersma LVA, et al. Safety and efficacy of the totally subcutaneous implantable defibrillator: 2-year results from a pooled analysis of the IDE study and EFFORTLESS registry. J Am Coll Cardiol. (2015) 65(16):1605–15. 10.1016/j.jacc.2015.02.047 [DOI] [PubMed] [Google Scholar]
- 4.Gold MR, Aasbo JD, El-Chami MF, Niebauer M, Herre J, Prutkin JM, et al. Subcutaneous implantable cardioverter-defibrillator post-approval study: clinical characteristics and perioperative results. Heart Rhythm. (2017) 14(10):1456–63. 10.1016/j.hrthm.2017.05.016 [DOI] [PubMed] [Google Scholar]
- 5.Theuns DAMJ, Brouwer TF, Jones PW. Prospective blinded evaluation of detection performance of a subcutaneous implantable cardioverter-defibrillator. Heart Rhythm. (2012) 9(10):1727–34. 10.1016/j.hrthm.2012.07.012 [DOI] [PubMed] [Google Scholar]
- 6.Migliore F, Biffi M, Viani S, Pittorru R, Francia P, Pieragnoli P, et al. Modern subcutaneous implantable defibrillator therapy in patients with cardiomyopathies and channelopathies: data from a large multicentre registry. Europace. (2023) 25(9):euad239. 10.1093/europace/euad239 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Migliore F, Ottaviano L, Arestia A, Nigro G, Dello Russo A, Viani S, et al. Subcutaneous implantable defibrillator therapy in patients with brugada syndrome: data from a large multicenter registry. JACC Clin Electrophysiol. (2025) 11(7):1572–82. 10.1016/j.jacep.2025.03.003 [DOI] [PubMed] [Google Scholar]
- 8.Knops RE, Marquie C, Nordbeck P, Tilz RR, Dijkshoorn LA, Quast A-FBE, et al. Subcutaneous defibrillator implantation with or without defibrillation test: the primary results of the randomized PRAETORIAN-DFT trial. Circulation. (2026) 154(6):520–31. 10.1161/CIRCULATIONAHA.126.080638 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Boersma LV, El-Chami MF, Bongiorni MG, Burke MC, Knops RE, Aasbo JD, et al. Understanding outcomes with the EMBLEM S-ICD in primary prevention patients with low EF study (UNTOUCHED): clinical characteristics and perioperative results. Heart Rhythm. (2019) 16(11):1636–44. 10.1016/j.hrthm.2019.04.048 [DOI] [PubMed] [Google Scholar]
- 10.Gold MR, Lambiase PD, El-Chami MF, Knops RE, Aasbo JD, Bongiorni MG, et al. Primary results from the understanding outcomes with the S-ICD in primary prevention patients with low ejection fraction (UNTOUCHED) trial. Circulation. (2021) 143(1):7–17. 10.1161/CIRCULATIONAHA.120.048728 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Boersma L, Barr C, Knops R, Theuns D, Eckardt L, Neuzil P, et al. Implant and midterm outcomes of the subcutaneous implantable cardioverter-defibrillator registry: the EFFORTLESS study. J Am Coll Cardiol. (2017) 70(7):830–41. 10.1016/j.jacc.2017.06.040 [DOI] [PubMed] [Google Scholar]
- 12.Lambiase PD, Barr C, Theuns DAMJ, Knops R, Neuzil P, Johansen JB, et al. Worldwide experience with a totally subcutaneous implantable defibrillator: early results from the EFFORTLESS S-ICD registry. Eur Heart J. (2014) 35(25):1657–65. 10.1093/eurheartj/ehu112 [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
The original contributions presented in the study are included in the article/Supplementary Material, further inquiries can be directed to the corresponding author.