. 2026 Oct 9;21(10):e0358882. doi: 10.1371/journal.pone.0358882

Lower limb prosthetic users’ activity classification through real-world inertial sensing

Matthew Wassall 1,*, Malcolm Granat 2, Saeed Zahedi 3, Sibylle Thies 2
Editor: Gary Guerra4
PMCID: PMC13652245  PMID: 42853806

Abstract

The prescription of lower limb prosthetics relies heavily on the assigned K-level. K-level assessment has been criticised for subjectivity and variability, potentially leading to inappropriate component selection and poorer functional outcomes. This study presents an original methodology for objectively determining activity levels from inertial sensor data, developed through rigorous data collection on 20 amputees who walked on various types of outdoor terrain (flat ground, stairs, ramp, grass/uneven/unstable surfaces), with and without a walking aid. Advanced machine learning algorithms were applied to the real-world IMU data, to establish a validated classification model for daily activity outside the laboratory. A random forest classifier achieved a mean leave‑some‑out accuracy of 65.70% (SD 6.6%; 95% CI 47.3–84.1%) and cross‑validation accuracy of 85.65% (SD 0.79%; 95% CI 83.4–87.9%). Per‑class sensitivity ranged from approximately 44–78% in leave‑some‑out analyses, with lower performance for ramp ascending and descending. The difference between cross‑validation and leave‑some‑out accuracies indicates participant‑specific overfitting, so the leave‑some‑out accuracy is considered the current conservative estimate of performance on unseen users and the cross-validation accuracy an estimate of potential accuracy with a more robust dataset. The research subsequently incorporated a preliminary clinical evaluation of the approach on a subset of 3 amputees using a data collection protocol for unsupervised, longitudinal real-world monitoring, with preliminary insight into perceived clinical significance of outcomes ascertained through exploratory interviews with 4 prescribing health care professionals. The findings suggest potential benefits of real-world inertial sensing, supporting a move from primarily subjective clinical assessment toward more objective, data-driven activity classification. This research demonstrates that an evidence-based tool for enhancing K-level prescription could in future provide more appropriate component selection and improve long-term mobility for prosthetic users, but current accuracies and sample sizes indicate that the system should be viewed as a promising proof‑of‑concept rather than a definitive tool for changing K‑level prescription.

Introduction

It is estimated that 35–40 million people globally require prosthetics and orthotics [1], and the need for prosthesis in particular is on the rise, for example in 2005 there were 1.6 million people living with limb loss in the United States of America (USA) and it is projected that there will be 3.6 million by 2050 [2]. Activity levels of lower limb prosthesis patients are characterised by K-levels and the assigned K-level determines which prosthesis components the patient will be given [3]. K-levels were established by the US Health Care Financing Administration and span from K0 to K4 [3]. There is no set procedure to assess patients for classification of K-levels. K-level assessments variability and subjectivity can at times lead to inappropriate prosthetic prescriptions which could impact the user’s functional outcomes [4].

One solution to improve the reliability of K-level assessments could be the use of activity monitoring in the real-world, to provide clinicians with objective every-day data on activities which could support clinicians in making more consistent K‑level assessments and justify component selection, but this relationship has not yet been empirically demonstrated. Activity monitoring involves the use of inertial measurement units (IMUs) placed on the body that measure acceleration, rotation, and orientation. IMUs provide data which can then be used to characterise movements of the body. Activity monitors are increasingly used in human movement science, but they are not yet routinely employed in clinical practice to assess patients’ activity levels [5]. Notably, understanding what measures derived from IMU data could potentially be useful to support K-level assessments is key to supporting clinical decision-making.

As a first step, we conducted exploratory interviews with 6 clinical experts to identify issues with current K-level assessments and what measures would be desirable to help improve the reliability of the assessments [6]. These interviews revealed that the issues with the current assessments relate to the subjectivity and variability of the assessments, and that there is a clinical need for objective activity data. Key variables identified as most clinically useful were: step count, cadence and ability to change cadence, walking aid use, and ability to traverse different terrain types. Specific types of terrain of interest included flat ground, stairs, ramps, and uneven surfaces (e.g., grass, unstable gravel). It was also discovered that the issues mainly occur in decisions between K2 and K3 users.

Step count and cadence have been successfully assessed using a shank mounted inertia sensor for able-bodied and lower limb prosthetic users [7–11]. However, use of a walking aid in amputees has to-date not been classified, and terrain classification has only been attempted in healthy adults [12–15] and small groups of prosthetic users [16–19]. To classify terrain, wearable sensors have previously shown high accuracies when used on the lower limb prosthetic population [20]. Some previous studies have used sensors that would not be feasible for long-term real-world use [21–31]. A visual sensor, for example a depth camera or laser, would not be practical for real-world use as they need line of sight, thereby restricting what clothes a user could wear, but could be used to train a more practical sensor to reduce data processing. EMG sensors need to be attached to the user’s skin which is not practical for long-term real-world use, as this would require the prosthetic user to attach the electrodes. A few studies have shown that an IMU attached to a prosthesis can be used to classify certain terrain [32–34]. These studies only collected data from a small number of lower limb prosthetic users and did not incorporate real-world terrain. IMUs have also been shown to be able to classify terrain in larger healthy populations [35].

Machine learning offers the capability to classify human movement activities based on IMU data, but there is no conclusive agreement on the best algorithm to use or how to process the data, for example appropriate sampling rate or windowing method. Algorithms that have performed well in previous studies include Support Vector Machine (SVM) [16,19,36], Random Forest (RF) [37] and Long Short Term Memory neural networks (LSTM) [19,21,36]. Logistic Regression (LR) and K-Nearest Neighbours (KNN) have also previously been used and are considered simple models [38], hence could be used for benchmark comparisons. There is no conclusive evidence regarding which machine learning model would produce the highest terrain classification or walking aid recognition accuracies from wearable sensors, hence a comparison of these models is yet required for these purposes.

Similarly, there is currently no conclusion on window type and size: studies using time-based windows vary in window size from 4ms [39] to 8s [40], all producing similar accuracies. Some studies have produced high accuracies splitting the data into strides [12,41,42], but similar accuracies have also been reported in studies that have analysed individual datapoints [32,43]. Likewise, sampling rate which ranges from 20 Hz [36] to 500 Hz [28] in different studies produced similar results.

Sensor placement also varied in previous studies. To be minimally intrusive for a prosthetic user, a system would ideally be attached to the prosthesis and not the user’s body. Reported differences in knee and ankle kinematics across varying terrains indicate that an IMU positioned on the shank is likely to produce distinguishable signals [44]. Supporting this a single shank-mounted IMU can produce good terrain classification accuracies (0.90 F1-score) on able-bodied individuals [35], but this has not been investigated on lower limb prosthetic users. Good accuracies were also found with a trunk (0.84 F1-score) and thigh (0.89 F1-score) placed sensor with all sensors combined producing the highest accuracies (0.92 F1-score). This study used all attributes of the IMU in the models. Hence, comparison between prosthetic‑shank IMU with sensors on the thigh, contralateral shank and trunk is needed.

To assess accuracy of the classifiers, most studies used a cross-validation (CV) evaluation instead of a leave-one-out approach. This could have provided a less realistic real-world accuracy. The only previous study found that classified real-world terrain for lower limb prosthetic users using a wearable sensor (accelerometer) and assessed using a leave-one-out approach achieved a 56.68% mean accuracy, but this study had only recruited 4 prosthetic users and only classified between flat, stairs (ascending and descending) and slopes (ascending and descending) [45].

Walking aid recognition using wearable sensors has previously been investigated in able-bodied adults but only by using a wrist-worn accelerometer [46,47]. High accuracies have been achieved, but only in limited indoor environments so the performance on a variety of terrain has to-date not been investigated. Studies using gait-lab based 3D motion analysis have found differences in lower limb kinematics during strides with and without a walking aid on flat ground [48,49] and during stair climbing [50]. Considering that wearable sensors on the lower limbs are able to measure the kinematics, their use on the lower limbs should likewise be able to detect the differences between strides with and without a walking aid.

These studies show that inertial sensors and machine‑learning methods can classify human activities, but they use heterogeneous sensor placements, windowing strategies, feature sets and algorithms, often in small, laboratory‑based samples. For lower‑limb prosthesis users there is, to our knowledge, no validated, minimally obtrusive, single‑sensor system that can classify both terrain type and walking‑aid use under real‑world conditions. It is also unclear which combination of sensor placement, window definition, sampling density and IMU features offers the best trade‑off between classification performance and clinical feasibility.

The aim of this study was to investigate the use of a single prosthesis shank-mounted IMU for terrain classification and walking aid recognition in lower limb prosthetic users in the real-world and evaluate its potential clinical benefit. The objectives were:

  1. Compare machine learning algorithms for terrain classification and walking aid recognition. We hypothesised that algorithm performance would differ across models, but there is insufficient prior evidence to predict a single best‑performing algorithm; therefore, we treated this comparison as exploratory.

  2. To see how prosthetic shank IMU placement compares to other sensor locations (thigh, trunk and contralateral shank) and against using multiple sensors for terrain classification and walking aid recognition. We hypothesised that a single prosthetic‑shank IMU will achieve accuracies comparable to the other single sensors and a multi‑sensor configuration.

  3. Investigate how windowing method and sampling rates affect classification accuracies. We hypothesised that different windowing methods and sampling rates will lead to measurable differences in classification accuracies, but since existing evidence does not support a directional hypothesis, we treated this comparison as exploratory.

  4. Evaluate the impact of each feature of the IMU (accelerometer, gyroscope, magnetometer), and combinations of these, on classification accuracy. We hypothesised that including multiple features will improve or at least not degrade performance compared with single‑feature models, but there was no prior evidence to predict an optimal feature combination so we treated this as exploratory.

  5. Explore whether longitudinal real-world activity measures generated using the most promising data analysis approach are clinically meaningful to prescribing clinicians. Based on prior interviews with clinicians, we hypothesised that the real-world activity measures would be perceived as relevant to K‑level assessment.

Methods

Overall study design

To ensure clinical relevance of the sensor system, the associated research contained two phases: 1) development and training of a machine learning system to classify clinically relevant measures, and 2) clinical evaluation of the system by experts. The study had university and NHS IRAS ethical approval (4743 and 314743, respectively) and all participants provided informed consent.

Phase 1: System development and data collection

Participants and data collection.

Participants were recruited through prosthetic organisations, charities, academic institutes and prosthetic clinics across the UK. Inclusion criteria for participants included ability to walk using their prosthesis unaided on the tested terrain and experience with using a walking aid. As the issues with K-level assessments predominantly occur with K2 and K3 users, we used the following exclusion criteria: not regularly participating in active sport with their prosthesis (football, running, etc.) and not primarily using a wheelchair; we note that this could limit the generalisability of the findings to K1 and K4 users. Twenty lower limb prosthetic users (mean±SD age: 64.6 ± 11.8 years, 2 female, 18 male; 11 transtibial, 8 transfemoral, 1 bilateral amputee) were recruited, participant details are shown in Table 1. The participant demographics did not perfectly match those of UK lower limb prosthetic users: 10% of participants were female versus approximately 30% nationally, and 85% were older than 54 years compared with around 70% in the UK population, however, the distribution of amputation levels was similar (transtibial 55% vs 54%, transfemoral 40% vs 37%, bilateral 5% vs 6%), reflecting representativeness of the data in that regard. For pragmatic reasons we did not sample to exactly mirror national demographics, which may limit generalisability of the findings, particularly with respect to sex‑specific and age‑related patterns of activity.

Table 1. Participant information and tests conducted for each participant. “wi”: data collected with walking aid, “wo”: data collected without walking aid.
Participant no. Age Amputation level Sex Flat Grass Stairs Slope Unstable Uneven
1 85 TT M wi/wo wo wi/wo wi/wo wo wi/wo
2 63 TT M wi/wo wo wi/wo wi/wo wo wi/wo
3 74 TT M wi/wo wi/wo wi/wo
4 69 TT M wi/wo wi/wo wi/wo wi/wo wi/wo wi/wo
5 56 TT M wi/wo wi/wo wi/wo wi/wo
6 72 TT M wi/wo wi/wo wi/wo wi/wo wi/wo wi/wo
7 69 TT M wi/wo wi/wo wi/wo wi/wo wi/wo
8 33 TT F wi/wo wi/wo wi/wo
9 62 TT F wo wo wo
10 61 TT M wo wo wo
11 65 TT M wo wo wo wo
12 69 TF M wi/wo wi/wo wi/wo wi/wo wi/wo
13 80 TF M wo wo wo wo wo wo
14 53 TF M wi/wo wi/wo wi/wo wi/wo wi/wo wi/wo
15 59 TF M wi/wo wi/wo wi/wo wi/wo wi/wo wi/wo
16 64 TF M wi/wo wi/wo wi/wo wi/wo wi/wo
17 44 TF M wo wo wo wo
18 67 TF M wi wi wi wi
19 80 TF M wi/wo wo wi/wo
20 66 TT/TF M wi/wo wi/wo wi/wo wi/wo wi/wo

Each participant wore four Xsens Awinda IMUs (100 Hz sampling rate) placed on the anterior prosthetic shank, thigh, the contralateral shank (approximately at the midpoint of each limb segment, see Fig 1). A further IMU was placed on the trunk at the approximate location of the L4 vertebra as shown in Fig 1. Prosthetic shank placement varied depending on the participant’s prosthesis; however, the IMU was placed on the prosthesis pylon to ensure consistent orientation. For each IMU, raw tri‑axial acceleration, free acceleration, angular velocity and magnetic field signals were used. Cadence signals were also calculated from the step count. Free accelerations are the acceleration signals with the effect of gravity removed. The resultant acceleration and free acceleration signals were also calculated. Magnetic field strength was included because it provides complementary information on sensor heading and orientation relative to the environment, which may differ across variable terrain (e.g., slopes vs level ground). It is not known which signals are the most important for terrain classification and walking aid recognition. Sensors were mounted so that the local X, Y and Z axes were approximately aligned with the segment’s vertical, medio-lateral and anterior–posterior directions, respectively. The Xsens Awinda IMUs incorporate factory calibration for accelerometer, gyroscope, and magnetometer sensors; no additional user‑level calibration procedure was applied before each session.

Fig 1. IMU placement.

Fig 1

Participants walked over designated sections of various types of terrain (flat ground, stairs, ramp, grass, uneven ground, unstable gravel) with and without a walking aid in separate trials. The order of the terrain and walking aid condition was randomised for each participant. The walking aid used in the study was a walking stick. The only unstable terrain used in the data collection was gravel, assessed by the researcher to be a loose, small‑stone surface deep enough that the stones moved underfoot and visibly deformed with each step. A ramp was classified as a slope steeper than 1:20 [51]. Uneven ground was hard terrain with variable footing heights and/or angles that the researcher conducting the study deemed to be uneven, for example cobblestones, this may reduce reproducibility and this subjectivity is revisited in the Discussion as a potential contributor to misclassification. Not all the participants were assessed on every terrain, due to terrain conditions. Four participants did not feel comfortable using the walking aid, and one participant only felt comfortable conducting the data collection with a walking aid. Data collection conditions for each participant are displayed in Table 1. Depending on the length of the terrain available and the participant’s comfort on the terrain, the number of separate trials conducted on each terrain varied between 2 and 8. Fig 2 shows representative terrain used in the data collection. The total number of strides recorded on each terrain with and without a walking aid is shown in Table 2. The number of recorded strides per terrain class was imbalanced, reflecting the practical difficulty of collecting data on different types of terrain. Class imbalance may have contributed to poorer performance for under-represented or more heterogeneous classes, but the risk was considered limited in this dataset due to the marked biomechanical differences in traversing the different types of terrain. In the discussion individual class imbalances are discussed. Participants walked at their self-selected natural walking speed on each terrain, which increases ecological validity of the dataset. The total length of a single data collection session was less than 2 hours and participants were encouraged to rest between trials when needed. The data were filtered using a 5 Hz low-pass Butterworth filter, applied to all IMU signals. The cut‑off was selected based on a frequency analysis of resultant acceleration from a representative subset of strides across terrains and walking aid conditions, which showed the dominant gait‑related power to be below 4 Hz. A combination of per‑person normalised and non-normalised data were used in the models to stabilise training while preserving absolute between‑subject differences. Normalisation is known to improve optimisation and reduce sensitivity to scale, but purely person‑specific normalisation can attenuate clinically relevant between‑subject variation; including both representations was therefore chosen as a pragmatic compromise in this feasibility study [52,53]. To our knowledge, combining per‑person normalised and raw IMU channels within the same feature set has not been reported previously, but in preliminary analysis [6] this combination produced the most favourable balance of cross‑validation and leave‑some‑out accuracies, so it was adopted for the present models.

Fig 2. Representative terrain used in the data collection.

Fig 2

A) Uneven ground (cobblestones), B) Grass, C) Unstable gravel, D) Stairs.

Table 2. Number of strides recorded on each terrain with and without a walking aid.
Without walking aid With walking aid Total
Flat 1603 1348 2951
Grass 671 340 1011
Stairs ascending 412 288 700
Stairs descending 423 287 710
Ramp ascending 646 293 939
Ramp descending 711 331 1042
Uneven ground 858 560 1418
Unstable gravel 603 343 946
Total 5927 3790 9717

Model development.

A comparative analysis was performed to determine the optimal machine learning model for terrain classification and walking aid recognition. The algorithms tested were SVM, KNN, RF, and LSTM. For walking aid recognition, a LR classifier was also used as a baseline. The following model parameters were investigated to optimise performance:

  • Sensor Placement: Data from IMUs placed on the prosthetic shank, prosthetic leg thigh, contralateral shank and trunk were compared as well as data from all IMUs combined were also evaluated.

  • Windowing: Multiple time-based non-overlapping windows (0.1s to 2s) were evaluated alongside windows comprising a full stride versus individual datapoints. The individual datapoint windows were not tested using cross-validation as datapoints from the same stride would otherwise be split across train and validation sets, leading to severe information leakage and artificially high accuracies. The sampling rate or number of samples per stride was also examined.

  • Feature Selection: Various combinations of IMU features (acceleration, angular velocity, magnetic field strength, free acceleration, cadence) were evaluated.

To label and segment the data into stride-based windows, we required not only accurate stride detection but also sufficient temporal precision to identify a consistent point within each stride, ensuring that each window contained a complete stride. Preceding analysis [6] established that the rate of turn along the medio-lateral direction of the prosthetic shank produces an accuracy of 96.15% and a precision of 97.97%, and was hence used to detect strides in the data processing. Strides were detected during the mid-swing phase of gait, and each stride was labelled according to the terrain under the prosthetic foot at initial contact. Steps that transition between different types of terrain were not separately identified. Step count was determined using the same method as stride detection with an accuracy of 96.15% [6]. The time taken for each stride was used to calculate the cadence for individual strides, from which mean cadence and cadence variability were derived.

Participants included transtibial, transfemoral and one bilateral amputee, and all models were trained on the combined dataset without stratification by amputation level. Analyses were carried out that did not show significant differences in classification accuracy between transtibial and transfemoral subgroups [6], however, the present study cannot exclude that performance varies by amputation level and this remains a limitation.

For each algorithm a Bayesian hyperparameter optimiser was run to mitigate overfitting. The optimisation was run on a representative subset of the data to obtain reasonable default settings. The optimiser evaluated a loss function on held‑out data, seeking the hyperparameter combination that minimised this loss. This initial optimisation was performed once per algorithm using 100 Hz data with 0.2 s windows and stride‑based windows (100 datapoints per stride). The optimal values were very similar across window types, and where small differences (<5%) occurred the mean of these values were used.

Analysis was conducted sequentially so the condition or processing method that produced the highest accuracy with consideration also given to reducing the computational need from a step was then used for the sequential steps in the analysis. The initial condition used for the sensor placement analysis was 100 Hz data, split into windows that contained a whole stride split into 100 datapoints.

Performance evaluation.

Overall percentage accuracy was used as the primary optimisation metric because the target application is multi‑class terrain and walking aid recognition, where correct class assignment across all categories is clinically most relevant and directly comparable with prior studies [17,18,20,45]. Accuracy was calculated for each validation test set as the proportion of held-out stride windows correctly classified. Per-class performance is reported via confusion matrix for the final model, with precision, sensitivity and specificity reported. For each class, sensitivity was calculated as TP/(TP+FN), precision as TP/(TP+FP), and specificity as TN/(TN+FP), where TP, TN, FP and FN denote true positives, true negatives, false positives and false negatives, respectively; metrics were calculated using a one-versus-rest approach. Model accuracy was determined using both 5-fold CV and a more robust 5-group leave-some-out approach (LSO), which is similar to a leave-one-out approach, but participants are quasi-randomly assigned to groups to ensure a mix of prosthetic types and reduce computation need. The mean accuracy across the five test folds/groups are reported for both LSO and CV, with standard deviation (SD) calculated across fold/group accuracies and 95% confidence intervals (CI) calculated using the Student’s t distribution (±t0.975,4×SD/5). Before comparing model accuracies, normality was assessed for each set of values using the Shapiro–Wilk test. When all models for a given comparison (e.g., windowing method, IMU position, feature set) satisfied the normality assumption (p ≥ 0.05), a one‑way ANOVA (with F-value (F), degrees of freedom and p-value (p) reported) was used with Tukey’s HSD post‑hoc tests. Kruskal-Wallis (with Chi-square (X2), degrees of freedom and p values (p) reported) with Dunn’s post-hoc test was used for non-parametric data. Significance was accepted as p values lower than 0.05. For significant ANOVA effects we report partial eta‑squared as an effect size; for key pairwise post‑hoc contrasts we additionally report Cohen’s d.

Phase 2: Clinical evaluation of the system by experts

Real-world data were collected unsupervised over two weeks using a prosthesis shank-mounted IMU on 3 lower limb prosthetic users that had already participated in Phase 1. These participants also had their K-levels assessed by a clinician who had 8 years’ experience in conducting K-level assessments using standard clinical practice. The IMU data were processed using the optimal machine learning model developed in Phase 1. A report that documented daily measures for number of steps, number of steps on each type of terrain, mean cadence, steps above 20% mean cadence and number of steps taken with a walking aid and a separate summary of an average for each measure over the recoding period, along with a summary of the participants K-level assessment, was presented to 4 clinical experts (mean experience 21.5 years, 3 had participated in the preceding interviews that had provided knowledge in the clinical need [6]. Semi-structured interviews were then conducted with the clinical experts to obtain their evaluation of the system and the clinical relevance of the data it produced. Interviews were analysed using thematic analysis based on a framework approach described by Braun and Clarke [54]. Details of the interviews and thematic analysis can be found in supplementary file S1 File.

Results

The mean accuracies for all models tested for terrain classification and walking aid recognition are presented in Table 3, but the main findings are summarised in the text that follows.

Table 3. Mean percentage accuracy for each model analysed (standard deviation shown in brackets), for terrain classification and walking aid recognition.

Test Terrain classification Walking aid recognition
KNN RF SVM LSTM KNN RF SVM LSTM LR
LSO CV LSO CV LSO CV LSO CV LSO CV LSO CV LSO CV LSO CV LSO CV
IMU position
Prosthetic Shank 44.19

(3.92)
84.30 (0.63) 52.09 (3.48) 81.66 (0.72) 52.87 (6.40) 84.29 (0.44) 47.53 (3.75) 81.99 (0.40) 55.08 (3.84) 74.31 (0.80) 61.45 (8.33) 79.39 (0.82) 64.39 (7.15) 82.21 (0.49) 54.59 (7.12) 69.29 (1.26) 53.70 (6.66) 65.15 (0.43)
Contralateral Shank 43.52 (2.83) 83.32 (0.73) 53.56 (4.38) 80.60 (1.27) 52.41 (7.42) 82.11 (1.44) 46.62 (7.63) 82.20 (0.62) 52.44 (1.73) 75.22 (0.57) 61.43 (9.78) 79.11 (0.45) 62.58 (7.94) 83.97 (0.69) 56.55 (6.36) 70.79 (0.92) 55.35 (5.00) 65.76 (0.96)
Thigh 37.53 (2.64) 84.72 (0.90) 44.36 (6.38) 82.11 (0.82) 43.95 (6.61) 83.15 (0.88) 39.64 (4.65) 83.21 (0.90) 54.83 (2.65) 77.24 (0.83) 65.50 (12.71) 80.29 (1.08) 67.95 (13.53) 84.84 (0.68) 59.84 (6.56) 68.91 (0.73) 55.65 (10.90) 58.01 (0.76)
Trunk 43.51 (2.11) 81.67 (0.91) 54.30 (6.41) 79.90 (0.82) 53.22 (4.58) 81.36 (0.84) 50.31 (5.59) 81.48 (0.77) 53.48 (3.01) 76.63 (0.89) 60.68 (9.45) 79.41 (0.73) 60.61 (7.01) 83.45 (0.50) 56.49 (6.30) 71.43 (1.29) 48.43 (5.47) 65.24 (0.46)
All 44.73 (4.39) 82.68 (0.69) 52.00 (3.77) 84.92 (0.72) 49.08 (3.25) 87.32 (0.76) 50.06 (5.31) 81.61 (0.85) 51.40 (3.62) 78.21 (1.23) 66.50 (14.31) 81.71 (0.90) 67.45 (13.77) 75.04 (0.34) 64.45 (10.71) 65.61 (2.48) 48.99 (4.78) 53.19 (1.60)
Windowing method
0.1s 43.97 (3.92) 83.40 (0.18) 43.39 (2.00) 79.70 (0.13) 37.46 (2.97) 48.58 (0.47) 48.05 (3.73) 76.75 (0.44) 58.12 (6.99) 77.96 (0.26) 61.64 (8.29) 81.88 (0.31) 58.64 (8.60) 79.53 (0.16) 63.46 (8.06) 65.64 (0.09) 61.55 (8.63) 62.18 (1.24)
0.2s 39.99 (0.72) 77.68 (0.37) 43.04 (1.84) 74.32 (0.45) 41.56 (5.01) 51.50 (0.30) 44.31 (4.43) 54.94 (0.74) 57.88 (6.11) 73.55 (0.26) 61.81 (8.43) 76.96 (0.39) 59.01 (6.77) 77.19 (0.39) 54.85 (11.10) 55.69 (3.50) 61.68 (6.83) 61.05 (0.07)
0.5s 39.99 (0.72) 68.26 (0.64) 43.11 (2.01) 69.53 (0.60) 41.56 (5.00) 61.59 (0.26) 44.46 (2.59) 69.32 (0.41) 56.08 (6.93) 69.20 (0.59) 62.19 (8.99) 72.90 (0.54) 60.63 (7.98) 72.01 (0.78) 48.13 (4.79) 56.23 (2.34) 59.14 (7.33) 62.74 (1.09)
1s 36.49 (1.50) 63.01 (0.58) 43.01 (2.27) 62.79 (0.84) 43.24 (5.53) 64.54 (0.56) 39.40 (3.74) 67.15 (0.81) 53.89 (3.12) 77.78 (0.66) 61.55 (8.19) 74.96 (0.27) 62.68 (7.71) 78.61 (0.70) 50.91 (3.96) 72.35 (0.81) 61.79 (8.09) 66.24 (0.83)
1.5s 39.72 (1.37) 60.28 (1.53) 42.88 (1.53) 60.85 (0.63) 39.09 (5.91) 63.98 (1.38) 38.60 (3.10) 67.00 (0.77) 52.77 (1.94) 70.56 (1.08) 61.15 (6.46) 66.06 (1.22) 61.96 (7.86) 68.97 (0.73) 49.93 (1.70) 50.65 (1.18) 59.52 (6.76) 60.20 (2.75)
2s 36.35 (1.10) 56.70 (0.72) 42.11 (1.53) 56.08 (0.67) 37.53 (4.39) 57.72 (0.63) 37.03 (2.09) 61.93 (0.77) 51.99 (1.22) 64.40 (1.01) 61.64 (8.29) 68.81 (0.93) 61.46 (8.28) 65.87 (1.15) 50.73 (1.60) 50.32 (0.82) 57.48 (5.26) 61.09 (1.73)
Stride 44.19 (3.92) 84.30 (0.63) 52.09 (3.48) 81.66 (0.72) 52.87 (6.40) 84.29 (0.44) 47.53 (3.75) 81.99 (0.40) 55.08 (3.84) 74.31 (0.80) 61.81 (8.43) 79.39 (0.82) 64.39 (7.15) 82.21 (0.49) 53.70 (6.66) 65.15 (0.43) 54.59 (7.12) 69.29 (1.26)
Individual 44.42 (6.16) 46.52 (3.54) 32.72 (3.53) 48.23 (7.07) 59.52 (6.67) 62.19 (8.99) 62.65 (9.22) 54.39 (10.09) 60.43 (8.29)
Number of datapoints
10 45.45 (3.28) 85.27 (0.24) 54.40 (4.01) 82.59 (1.08) 52.10 (4.23) 77.31 (0.37) 45.58 (4.47) 80.78 (0.82) 55.97 (5.45) 74.85 (0.71) 60.07 (9.91) 78.71 (0.92) 59.61 (6.52) 83.10 (0.42) 53.61 (7.63) 70.86 (1.34) 57.02 (5.41) 66.66 (2.31)
20 44.55 (3.57) 85.02 (0.39) 53.13 (4.07) 82.32 (0.67) 51.78 (4.90) 79.95 (0.90) 47.35 (4.33) 81.65 (0.95) 55.17 (4.43) 74.79 (0.37) 62.07 (8.42) 79.60 (0.53) 60.22 (6.91) 83.32 (0.80) 54.64 (10.6) 73.25 (0.55) 55.96 (5.20) 67.75 (2.11)
50 44.31 (3.71) 84.34 (0.53) 53.25 (4.81) 81.62 (0.32) 52.10 (6.29) 82.10 (0.52) 47.39 (2.06) 82.23 (0.80) 54.69 (4.23) 74.60 (0.81) 61.99 (8.61) 79.25 (0.37) 62.74 (7.81) 83.28 (0.78) 50.90 (9.32) 71.78 (0.52) 55.57 (9.23) 67.39 (1.40)
100 44.19 (3.92) 84.30 (0.63) 52.09 (3.48) 81.66 (0.72) 52.87 (6.40) 84.29 (0.44) 47.53 (3.75) 81.99 (0.40) 55.08 (3.84) 74.31 (0.80) 61.45 (8.33) 79.39 (0.82) 64.39 (7.15) 82.21 (0.49) 53.70 (6.66) 65.15 (0.43) 54.59 (7.12) 69.29 (1.26)
Variables
Accelerations 36.80 (2.16) 76.46 (0.36) 44.12 (4.04) 76.26 (0.34) 42.14 (2.53) 58.96 (0.69) 40.36 (2.43) 71.60 (1.28) 54.81 (3.14) 74.96 (1.18) 65.04 (6.42) 77.84 (0.89) 59.76 (11.12) 78.94 (0.45) 59.58 (7.71) 72.36 (0.99) 56.56 (8.19) 66.94 (1.22)
Angular Velocity 29.47 (4.39) 73.66 (1.84) 36.32 (3.03) 74.07 (0.50) 29.03 (2.65) 48.26 (0.53) 32.86 (2.60) 62.75 (0.56) 57.40 (3.66) 75.48 (0.69) 55.68 (10.80) 76.80 (0.55) 58.39 (6.97) 76.54 (0.81) 56.98 (8.90) 68.23 (1.01) 55.47 (10.92) 63.34 (1.52)
Magnetic Field Strength 28.57 (3.14) 80.77 (0.43) 41.48 (4.58) 79.47 (0.65) 35.80 (2.69) 47.98 (0.58) 34.00 (6.30) 66.74 (1.25) 54.65 (4.74) 70.00 (1.27) 60.66 (7.52) 72.90 (0.42) 55.32 (6.29) 68.59 (0.99) 57.63 (6.03) 67.02 (1.47) 55.03 (11.07) 62.32 (1.34)
Free Accelerations 40.91 (2.29) 80.04 (0.63) 50.03 (4.04) 77.52 (0.25) 48.60 (3.73) 57.54 (1.09) 46.22 (3.51) 74.16 (0.75) 55.25 (3.37) 73.56 (0.66) 61.33 (6.32) 73.92 (0.44) 59.24 (6.39) 73.87 (0.42) 56.47 (8.80) 66.99 (0.69) 57.02 (5.02) 64.27 (1.95)
AVM 41.82 (4.18) 84.85 (0.85) 49.49 (4.07) 82.68 (1.13) 49.39 (4.12) 70.08 (1.04) 41.15 (5.87) 79.49 (1.73) 55.27 (4.38) 74.31 (0.41) 60.67 (10.44) 79.00 (0.55) 59.56 (9.76) 82.73 (0.68) 56.10 (9.59) 74.29 (1.01) 62.00 (5.70) 65.54 (0.94)
AVMF 44.87 (3.82) 85.67 (1.04) 54.18 (3.35) 83.18 (0.85) 52.28 (3.94) 74.44 (0.88) 43.32 (2.94) 80.18 (1.60) 55.63 (3.90) 74.30 (0.37) 60.58 (9.14) 78.16 (0.46) 59.63 (8.11) 82.01 (0.73) 54.75 (9.70) 74.44 (0.75) 55.58 (3.22) 64.14 (1.16)
AVMFC 45.46 (2.82) 85.61 (0.63) 54.36 (3.79) 82.83 (0.25) 52.83 (3.98) 75.17 (1.09) 46.94 (4.09) 80.59 (0.75) 55.51 (4.57) 75.49 (0.55) 60.97 (8.94) 78.54 (0.79) 59.55 (8.06) 82.90 (0.72) 54.70 (9.70) 74.46 (0.39) 59.92 (7.70) 63.36 (2.17)
All 45.45 (3.28) 85.27 (0.24) 54.40 (4.01) 82.59 (1.08) 52.10 (4.23) 77.31 (0.37) 45.58 (4.47) 80.78 (0.82) 55.97 (5.45) 74.85 (0.71) 60.07 (9.91) 78.71 (0.92) 59.61 (6.52) 83.10 (0.42) 53.61 (7.63) 70.86 (1.34) 57.02 (5.41) 66.66 (2.31)

KNN = K-Nearest Neighbours, RF = Random Forest, SVM = Support Vector Machine, LSTM = Long Short-Term Memory neural network, LR = Logistic Regression, AVM = Acceleration Angular Velocity + Magnetic Field Strength, AVMF = Acceleration + Angular Velocity + Magnetic Field Strength + Free acceleration, AVMFC = Acceleration + Angular Velocity + Magnetic Field Strength + Free acceleration + Cadence.

Terrain classification

Comparing IMU location it was found that there were no significant differences in classification accuracies between IMU positions for LSO results. For CV results IMU position had a significant effect for all the algorithms except for LSTM (KNN X2(4)=17.60 p = 0.002, RF X2(4)=17.25 p = 0.002 and SVM X2(4)=19.62 p < 0.001). The ‘all- combined’ produced significantly higher accuracies than using any IMU by itself; for RF models all-combined compared to prosthesis shank IMU the mean difference was 3.26% (p < 0.001, Cohen’s dz=2.70). A minimally obtrusive system would be prosthesis shank-mounted, and using this single IMU did not show significantly lower accuracies than ‘all-combined’ IMU or any other single IMU for any LSO or the KNN, SVM and LSTM CV models. Only data from the single prosthesis shank IMU were hence used for the rest of the analysis.

Windowing method had a significant effect on terrain‑classification accuracy for all four algorithms for LSO (KNN X2(7)=25.34 p < 0.001, RF X2(7)=16.32 p < 0.001, SVM X2(7)=19.08 p < 0.001 and LSTM X2(7)=20.39 p < 0.001) and CV tests (KNN X2(6)=33.33 p < 0.001, RF X2(6)=33.24 p < 0.001, SVM X2(6)=32.69 p < 0.001 and LSTM X2(6)=32.74 p < 0.001). The stride-based windows produced the highest accuracies for RF and SVM LSO tests (RF 52.09%, SVM 52.89%) and CV tests (RF 81.66%, SVM 84.29%). KNN and LSTM showed similar trends. On this basis, stride‑based windows were used in subsequent analyses. The number of datapoints per window showed a significant effect for only KNN and SVM CV results (KNN X2(3)=8.47 p = 0.0372, SVM X2(3)=17.86 p < 0.001). Post-hoc tests indicated that reducing the windows from 100 datapoints to 10 significantly reduced the SVM CV accuracy (mean difference 6.98%, p < 0.001, Cohen’s dz=8.13), whereas RF and KNN accuracies were less sensitive to sampling density. The 10 datapoint model was hence used for the rest of the analysis.

The features used in a model had significant effects for both LSO (KNN X2(8)=29.31 p < 0.001, RF X2(8)=27.01 p < 0.001, SVM X2(8)=29.82 p < 0.001 and LSTM X2(8)=21.05 p = 0.004) and CV (KNN X2(8)=34.29 p < 0.001, RF X2(8)=33.59 p < 0.001, SVM X2(8)=37.57 p < 0.001 and LSTM X2(8)=34.24 p < 0.001). Combining features produced higher LSO and CV accuracies compared to single feature models. Combining accelerations, angular velocity, magnetic field strength, free accelerations and cadence (AVMFC) data together produced the highest mean accuracy for KNN, SVM and LSTM LSO tests and produced the second highest for all CV tests. Combining accelerations, angular velocity, magnetic field strength and free accelerations did produce higher CV accuracies for KNN and RF but lower LSO accuracies for all tests, but none of these differences were significant. Hence AVMFC was chosen to be used in the successive analysis.

Based on the LSO accuracy, the RF model was selected as the final classifier for terrain classification. While the KNN model produced a slightly higher CV accuracy, the RF model achieved higher LSO accuracy. Because LSO withheld participant groups during training, it was considered a more conservative estimate of performance in unseen users than the CV estimate. An additional Bayesian hyperparameter optimisation was run on the final RF model (prosthetic shank IMU, 10 data point stride-based windows, AVMFC features) and the re-optimised model achieved a mean LSO accuracy across held-out participant groups of 56.89% (SD 5.5%; 95% CI 50.1–63.7%) and a mean CV accuracy of 83.64% (SD 0.8%; 95% CI 82.6–84.7%).

To investigate participant variability leave-one-participant-out (LOPO) analysis was run using the final RF model, Table 4. Mean accuracies between amputation levels were similar (TT 51.8 ± 14.6%, TF 50.4 ± 20.0%), and Welch’s t‑test showed no significant difference between groups (p = 0.86), suggesting that amputation level did not influence classification accuracy.

Table 4. LOPO classification accuracies.

Participant Amputation level Terrain classification accuracy Percentage of steps with a walking aid Walking aid recognition accuracy
P1 TT 71.71% 19.02% 54.34%
P2 TT 64.19% 25.22% 62.09%
P3 TT 40.34% 44.02% 22.32%
P4 TT 54.24% 49.64% 61.86%
P5 TT 28.77% 22.43% 89.62%
P6 TT 32.54% 50.93% 49.04%
P7 TT 67.55% 50.36% 53.98%
P8 TT 42.83% 60.17% 47.24%
P9 TT 44.87% 0.00% 59.40%
P10 TT 65.56% 0.00% 66.70%
P11 TT 63.31% 0.00% 82.32%
P12 TF 71.47% 50.77% 60.78%
P13 TF 33.81% 0.00% 51.80%
P14 TF 40.54% 52.06% 18.92%
P15 TF 24.48% 47.06% 41.15%
P16 TF 22.40% 43.10% 34.90%
P17 TF 62.61% 0.00% 58.32%
P18 TF 59.61% 100.00% 96.31%
P19 TF 70.18% 35.42% 56.96%
P20 TT 45.30% 54.01% 29.91%
TF 68.16% 55.23% 59.34%

Walking aid recognition

IMU location did not have a significant effect on LSO accuracies but did significantly affect CV accuracies (KNN X2(4)=18.35 p = 0.0011, RF X2(4)=12.87 p = 0.012, SVM X2(4)=20.53 p < 0.0001, LR X2(4)=15.49 p = 0.0038, LSTM X2(4)=19.14 p < 0.0001). The trunk IMU using a SVM model produced the highest LSO and CV accuracy, but there were no significant differences between IMU position for any of the LSO models. A minimally obtrusive system would be prosthesis shank-mounted. Using this single IMU was only significantly worse than the trunk IMU for the SVM and KNN models, but the difference was small (SVM 2.63% p = 0.034 Cohen’s dz=2.52, KNN 2.93% p = 0.026 Cohen’s dz=5.00). Data from the single prosthetic shank IMU only was therefore used for the rest of the analysis.

Windowing method had no significant effect on LSO but did significantly affect CV accuracies (KNN X2(6)=32.01 p < 0.001, RF X2(6)=33.16 p < 0.001, SVM X2(6)=32.98 p < 0.001, LR X2(6)=31.82 p < 0.001, LSTM X2(6)=25.13 p < 0.001). Stride-based windows produced the highest accuracies for the LSO and CV tests using a SVM model, and no stride-based window model produced accuracies that were significantly lower than any other windowing method. Stride-based windows were hence used for the sequential analysis. Reducing the windows to 10 datapoints did not significantly reduce the LSO or CV accuracies for any model. The 10 datapoint model produced the highest KNN and LSTM LSO accuracies and was therefore used for the rest of the analysis.

The features included in the models had no significant effect on the accuracies of any LSO model but did show a significant effect on CV accuracies (KNN X2(8)=35.51 p < 0.001, RF X2(8)=38.09 p < 0.001, SVM X2(8)=42.18 p < 0.001, LR X2(8)=18.50 p = 0.018, LSTM X2(8)=32.48 p < 0.001). Combining features generally produced higher CV accuracies compared to single feature models. Combining accelerations, gyroscope, magnetometer, free accelerations and cadence data together produced consistently high LSO and CV accuracies for all models and did not produce significantly lower accuracies than any other feature combination for any LSO or CV model, so only these features were used in successive analysis.

The SVM model provided the best performance for walking aid recognition. It had the highest CV accuracy and showed no significant difference from the other models in LSO accuracy, making it the most reliable choice. The final SVM model (prosthetic shank IMU, 10 data point stride-based windows, AVMFC features) was re-optimised and achieved a mean LSO accuracy across held-out participant groups of 61.56% (SD 16.1%; 95% CI 45.4–77.7%) and a mean CV accuracy of 82.90% (SD 2.1%; 95% CI 80.8–85.0%).

LOPO walking aid recognition accuracies showed a large inter participant variability, Table 4, but amputation level did not affect model accuracies (TT 56.6 ± 19.0%, TF 53.2 ± 21.4%; Welch’s t‑test p = 0.71). The proportion of steps performed with a walking aid was also not significantly associated with walking aid recognition accuracy (Pearson r=−0.10, p = 0.65; Spearman p=−0.29, p = 0.20), suggesting that uneven exposure to aided steps did not systematically bias performance.

Phase 2: Clinical evaluation

Detailed results from the thematic analysis can be found in supplementary file S1 File. All the interviewees thought the system was clinically relevant and when comparing the system data to the clinical assessment summaries they thought that the sensor data could have changed the K-level for some of the participants that they otherwise would not have assigned to them. They all thought the data would be very useful to have to add to their clinical assessment, to justify decisions, prompt in-depth questions on patients’ lives, and enable difficult conversations to be had objectively. The use of the data to prompt questions was a positive that the interviewees were very enthusiastic about, as it would allow them to get more information about their patients and, therefore, a greater understanding of their lives, which could help them ensure that their patients are getting the right prescription and support.

The ability to distinguish between different types of terrain was particularly appreciated, as this provided a new level of detail for assessing a patient's activity. The primary criticism of the system was distinguishing between different grass, uneven and unstable terrain. Some of the clinical experts suggested that combining these types of terrains would make the data easier to understand. Additional measures of loading through the prosthesis and wear-time were suggested to add to the clinical relevance of the system.

In summary, the clinical evaluation of the developed system was highly positive. The clinicians perceived the system as a useful tool for quantifying K-level assessments. The data provided by the system was deemed to be highly valuable, as it provided objective, real-world data that was previously unobtainable.

Discussion

This study has suggested that a single prosthesis-mounted IMU can be used to measure clinically relevant variables, specifically step-count, cadence and terrain classification, which could improve lower limb prosthesis prescriptions. The findings confirm a strong desire from clinicians to obtain this objective data to support and enhance K-level decisions.

The final RF model produced a LSO accuracy of 56.89%, Fig 3, but the model overpredicted for flat terrain (50.8% precision, 67.2% specificity and 83.4% sensitivity) but the precision of steps predicted as stair or ramp ascending/descending is much higher (stair ascending 81.8%, stair descending 77.5%, ramp ascending 73.5%, ramp descending 68.1%). Low precision values were observed in grass and uneven terrain classifications which were commonly misclassified as each other or flat terrain. The final RF CV model, Fig 4, also overclassified for flat terrain (73.8% precision, 86.5% specificity and 92.4% sensitivity) but produced high precision for all other terrain classifications (grass 87.8%, stair ascending 90.1%, stair descending 91.2%, ramp ascending 93.8%, ramp descending 89.7%, uneven 86.4%, unstable 91.1%). LOPO accuracies ranged from 22.4% to 71.7% indicating considerable inter‑individual variability. This indicates that while the classifier generalises reasonably, performance for some users is lower and may require user‑specific calibration or a larger more diverse training dataset. The clinical experts in Phase 2 suggested combining grass, uneven and unstable terrain categories into one uneven terrain category. In response to their comment, a RF model with these terrain categories was created for supplementary data analysis which then produced an accuracy increase to 65.70% (SD 6.6%; 95% CI 47.3–84.1%) for LSO and to 85.65% (SD 0.79%; 95% CI 83.4–87.9%) for CV. This increase in accuracy is purely down to the high accuracy in the combined uneven class (LSO precision 60.4%, sensitivity 77.9% and specificity 75.7%) with the accuracies from the other classes essentially unchanged, and per‑class sensitivity ranged from approximately 44–78%.

Fig 3. Confusion matrix for terrain classification RF leave-some-out.

Fig 3

Fig 4. Confusion matrix for terrain classification RF Cross-validation.

Fig 4

These accuracies are similar to a previous study [19] which achieved a leave-some-out accuracy of 56.68% and a cross-validation accuracy of 78.46%, but flat, grass, uneven and unstable terrain were all grouped as flat terrain. Using the same terrain groupings our model achieved a leave-some-out accuracy of 79.05% (4.03) and a cross-validation accuracy of 90.21% (0.78). These accuracies are comparable to other studies that classified limited terrain in laboratory conditions [17,18,42]. Collecting data on different variations of the same terrain category should make the model more robust but limits number of strides on each variation which could be lowering the accuracies. The variation in the lower limb prosthetic [13] could also be contributing to lower accuracies. Although the dataset contained data from more lower limb prosthetic users than had been collected in previous studies, a larger dataset would be needed to achieve high accuracies and robustness for clinical use. Furthermore, the data used for algorithm development were collected in real-world environments under study conditions (with the researcher present); it has been shown that there can be changes in gait in study conditions compared to free-living behaviour [14]. Training the models on data collected in free-living conditions would further improve the robustness of the accuracies. The cross‑validation accuracies being significantly higher than leave‑some‑out accuracies indicates participant‑specific overfitting could be present. No class‑weighted training or additional regularisation was applied to the models, which could be addressed in future work.

In this study a ramp was classified as a surface that was steeper than 1:20, the UK building regulation definition of a slope [51], but this meant any surface less steep was classified as flat. It would be unreasonable to assume that a participant’s gait only changes on slopes steeper than 1:20 and that gait then doesn’t continue to change with steeper slopes. Strides on slopes just steeper than 1:20 could be similar to strides on slopes just shallower than 1:20 which could be a source of misclassification in the model. Moreover, what constitutes “uneven terrain” was decided by the researcher conducting the study. Even with terrain that was considered very uneven there could be places where the foot placement is flat, but for this study it would have been classified as uneven. This could be a source of lower accuracies when classifying uneven terrain as seen in the confusion matrices of Figs 3 and 4. Measuring ankle movement instead of classifying terrain could offer the required clinically relevant data more accurately.

In this paper transition steps (steps moving the person from one terrain to another) were not identified. Some studies have found transition steps to be different [17,18,42], and therefore they could affect accuracies if not labelled correctly. There is also the question of when a stride should be classified as a certain terrain. A stride was detected during the swing phase and the terrain the foot lands on during the stride was labelled, however, no analysis was done to identify which parts of the gait are the most important for the classifiers and therefore where the transition between different gait patterns for different types of terrain occurs. Labelling transition steps could eliminate this, or another option that was not explored would be using stride-based windows which contain multiple strides.

It must be noted that the final SVM LSO walking aid recognition accuracy of 61.56% was low considering that 61.00% of the steps were conducted without a walking aid, therefore classifying all steps as without a walking aid would produce a similar accuracy to the final model. The CV accuracy of 82.90% is also too low for clinical use, as the 17.10% error could inaccurately change a clinician’s assessment of a patient and affect the prosthetic components they are prescribed. A walking aid-mounted IMU may be a solution to enhance classification accuracy of walking aid use, however, it would require an additional sensor to the shank-mounted IMU, and more if the patient used multiple walking aids.

The clinical evaluation demonstrated the clinical relevance and desire for the data that the system outputs. Load through the prosthesis was highlighted as an additional measure that could be clinically beneficial. There have been studies [16,18,38,39] that have incorporated load through a prosthesis with data of an IMU to classify limited terrain in controlled conditions, and these indicate that a measure of load through a prosthesis could improve terrain classification. No previous study has investigated walking aid recognition from load through a prosthesis, but it has been shown that users can offload 25% [55] of body weight when using a walking stick and 12% using a walking frame [56], hence load measurement may also enhance walking aid use classification.

The limitations of the current system’s accuracies and robustness indicate further research is needed before such a system could be used in clinical practice. However, this study demonstrates the potential clinical impact a system could have pending further development and a larger training dataset.

A formal test–retest reliability assessment for stride detection, cadence estimation or IMU attachment repeatability on the prosthesis pylon in this specific cohort was not performed. Instead, this research relied on a previously validated stride‑detection method from previous work and standardised mounting procedures [6]. Future studies should quantify how repeated donning of the IMU and day‑to‑day variability in placement affect classification accuracy. Phase‑2 of the study only involved 4 clinical experts analysing reports on 3 prosthetic users. The small participant sample should be considered and caution taken when generalising the findings. Moreover, clinicians responded to simulated reports rather than using the system to make assessment decisions, so we cannot determine whether, or how, the information would actually change K‑level assignments or component selection in routine practice. A controlled study designed to test whether the reports alter K‑level assignment or component selection is an important next step.

This research did not establish accuracy thresholds clinicians would consider sufficient for K‑level assessments or prescription decisions. How specific errors might influence clinical decisions was also not investigated. In Phase 2, clinicians commented that inaccuracies could be alleviated by combining system outcomes with current assessment methods rather than replacing these. Future work should include prospective studies in which clinicians use system outputs in real prescription workflows, to determine acceptable accuracy levels and identify which types of errors are clinically tolerable, also under consideration of risks to safety.

Conclusions

This study developed an original methodology for objectively determining activity levels from inertial sensor data. Through rigorous data collection and exhaustive algorithm design we were able to show that a single, prosthesis-mounted IMU has the potential to offer objective, real-world activity data to aid clinical decision-making for lower limb prosthetic users. The work was further substantiated by interviews which revealed that there is a desire from clinicians to obtain these data on their patients to improve and support K-level decisions. Further work could be conducted to incorporate an additional IMU on the walking aid and/or a measure of load through the prosthesis to improve walking aid use classification. A measure of load could furthermore be used to improve terrain classification and measure the stress on prosthesis components. The positive clinical evaluation alludes to the clinical significance of this technology to bridge the gap between subjective assessments and objective data for better patient outcomes in the continuously growing amputee population [1,2]. Future work needs to collect further critical data to pave the way to clinical deployment.

Supporting information

S1 File. Supplement with detailed qualitative interview results from “Lower limb prosthetic users’ activity classification through real-world inertial sensing”.

(PDF)

pone.0358882.s001.pdf (572.2KB, pdf)

Acknowledgments

I would like to express my gratitude to Sarah Kirkwood from the University of Salford and Dr Chantel Ostler from Portsmouth Hospitals University NHS Trust for their guidance in relation to the clinical relevance of the research and participant recruitment.

Data Availability

The minimal data set and accompanying MATLAB classification code generated during this study are available at Figshare via https://doi.org/10.6084/m9.figshare.32756211 and https://doi.org/10.6084/m9.figshare.32805296.

Funding Statement

The author(s) received no specific funding for this work.

References

  • 1.WHO. WHO standards for prosthetics and orthotics. WHO and USAID. 2017. [Google Scholar]
  • 2.Ziegler-Graham K, MacKenzie EJ, Ephraim PL, Travison TG, Brookmeyer R. Estimating the prevalence of limb loss in the United States: 2005 to 2050. Arch Phys Med Rehabil. 2008;89(3):422–9. doi: 10.1016/j.apmr.2007.11.005 [DOI] [PubMed] [Google Scholar]
  • 3.Amputee-Coalition. Do you know your k-level? An interview with interim president CEO Sue Stout about the importance of amputees knowing their k-levels. inMotion. 2013. [Google Scholar]
  • 4.Balk EM, Gazula A, Markozannes G, Kimmel HJ, Saldanha IJ, Resnik LJ. Lower limb prostheses: measurement instruments, comparison of component effects by subgroups, and long-term outcomes. Rockville (MD): Agency for Healthcare Research and Quality (US). 2018. [PubMed] [Google Scholar]
  • 5.Jamieson AG, Murray L, Buis A. The use of physical activity outcomes in rehabilitation interventions for lower limb amputees: A systematic review. Can Prosthet Orthot J. 2020;3(1):33931. doi: 10.33137/cpoj.v3i1.33931 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Wassall M. Advancing prosthetic prescriptions: developing a sensor-based system for objective activity assessment. University of Salford. 2025. [Google Scholar]
  • 7.Catalfamo P, Ghoussayni S, Ewins D. Gait event detection on level ground and incline walking using a rate gyroscope. Sensors (Basel). 2010;10(6):5683–702. doi: 10.3390/s100605683 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Simonetti E, Villa C, Bascou J, Vannozzi G, Bergamini E, Pillet H. Gait event detection using inertial measurement units in people with transfemoral amputation: a comparative study. Med Biol Eng Comput. 2020;58(3):461–70. doi: 10.1007/s11517-019-02098-4 [DOI] [PubMed] [Google Scholar]
  • 9.Gouda A, Andrysek J. Rules-based real-time gait event detection algorithm for lower-limb prosthesis users during level-ground and ramp walking. Sensors (Basel). 2022;22(22):8888. doi: 10.3390/s22228888 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Maqbool HF, Husman MAB, Awad MI, Abouhossein A, Mehryar P, Iqbal N, et al. Real-time gait event detection for lower limb amputees using a single wearable sensor. Annu Int Conf IEEE Eng Med Biol Soc. 2016;2016:5067–70. doi: 10.1109/EMBC.2016.7591866 [DOI] [PubMed] [Google Scholar]
  • 11.Gatti AA, Stratford PW, Brenneman EC, Maly MR. GT3X+ accelerometer placement affects the reliability of step-counts measured during running and pedal-revolution counts measured during bicycling. J Sports Sci. 2016;34(12):1168–75. doi: 10.1080/02640414.2015.1096018 [DOI] [PubMed] [Google Scholar]
  • 12.Madaoui L, Kerdjidj O, Kedir-Talha M. Design and implementation of IMU-based locomotion mode recognition system on Zynq SoC. Microprocessors and Microsystems. 2023;102:104927. doi: 10.1016/j.micpro.2023.104927 [DOI] [Google Scholar]
  • 13.Liu Y, An H, Ma H, Wei Q. Novel feature extraction and locomotion mode classification using intelligent lower-limb prosthesis. Machines. 2023;11(2):235. doi: 10.3390/machines11020235 [DOI] [Google Scholar]
  • 14.Papapicco V, Chen B, Munih M, Davalli A, Sacchetti R, Gruppioni E, et al. A Classification approach based on directed acyclic graph to predict locomotion activities with one inertial sensor on the thigh. IEEE Trans Med Robot Bionics. 2021;3(2):436–45. doi: 10.1109/tmrb.2021.3075096 [DOI] [Google Scholar]
  • 15.Vu HTT, Cao H-L, Dong D, Verstraten T, Geeroms J, Vanderborght B. Comparison of machine learning and deep learning-based methods for locomotion mode recognition using a single inertial measurement unit. Front Neurorobot. 2022;16:923164. doi: 10.3389/fnbot.2022.923164 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Gonzales-Huisa OA, Oshiro G, Abarca VE, Chavez-Echajaya JG, Elias DA. EMG and IMU data fusion for locomotion mode classification in transtibial amputees. Prosthesis. 2023;5(4):1232–56. doi: 10.3390/prosthesis5040085 [DOI] [Google Scholar]
  • 17.Bruinsma J, Carloni R. IMU-based deep neural networks: Prediction of locomotor and transition intentions of an osseointegrated transfemoral amputee. IEEE Trans Neural Syst Rehabil Eng. 2021;29:1079–88. doi: 10.1109/TNSRE.2021.3086843 [DOI] [PubMed] [Google Scholar]
  • 18.Zheng E, Wan J, Gao S, Wang Q. Adaptive locomotion transition recognition with wearable sensors for lower limb robotic prosthesis. IEEE/ASME Trans Mechatron. 2024;29(1):279–89. doi: 10.1109/tmech.2023.3278315 [DOI] [Google Scholar]
  • 19.Jamieson A. Development of a clinically-targeted human activity recognition system to aid the prosthetic rehabilitation of individuals with lower limb amputation in free living conditions. 2021. http://purl.org/coar/resource_type/c_db06 [DOI] [PMC free article] [PubMed]
  • 20.Labarrière F, Thomas E, Calistri L, Optasanu V, Gueugnon M, Ornetti P, et al. Machine learning approaches for activity recognition and/or activity prediction in locomotion assistive devices-a systematic review. Sensors (Basel). 2020;20(21):6345. doi: 10.3390/s20216345 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Wang D, Gu X, Yu H. A comparison of four neural networks algorithms on locomotion intention recognition of lower limb exoskeleton based on multi-source information. J Bionic Eng. 2023;21(1):224–35. doi: 10.1007/s42235-023-00435-w [DOI] [Google Scholar]
  • 22.Luo S, Shu X, Zhu H, Yu H. Early prediction of lower limb prostheses locomotion mode transition based on terrain recognition. IEEE Sensors J. 2023;23(22):27941–8. doi: 10.1109/jsen.2023.3320274 [DOI] [Google Scholar]
  • 23.Yin S, Chen X, Ma T, Wang Y, Guo Y, Leng Y. Environmental Recognition and Multimode Continuous-Phase Control for a Powered Transfemoral Prosthesis. In: 2023 IEEE International Conference on Development and Learning (ICDL); 2023 9–11 Nov. 2023. https://ieeexplore.ieee.org/document/1234567
  • 24.Zhong B, Silva RL da, Tran M, Huang H, Lobaton E. Efficient environmental context prediction for lower limb prostheses. IEEE Trans Syst Man Cybern, Syst. 2022;52(6):3980–94. doi: 10.1109/tsmc.2021.3084036 [DOI] [Google Scholar]
  • 25.Qian Y, Chen C, Xiong J, Wang Y, Leng Y, Yu H, et al. Terrain-adaptive exoskeleton control with predictive gait mode recognition: A pilot study during level walking and stair ascent. IEEE Trans Med Robot Bionics. 2024;6(1):281–91. doi: 10.1109/tmrb.2024.3349624 [DOI] [Google Scholar]
  • 26.Cheng S, Laubscher CA, Best TK, Gregg RD. Ambilateral activity recognition and continuous adaptation with a powered knee-ankle prosthesis. IEEE Trans Robot. 2025;41:2251–67. doi: 10.1109/tro.2025.3539206 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Gao S, Wang Y, Fang C, Xu L. A smart terrain identification technique based on electromyography, ground reaction force, and machine learning for lower limb rehabilitation. Appl Sci. 2020;10(8):2638. doi: 10.3390/app10082638 [DOI] [Google Scholar]
  • 28.Li J, Zhang J, Li K, Cao J, Li H. A multimodal framework based on deep belief network for human locomotion intent prediction. Biomed Eng Lett. 2024;14(3):559–69. doi: 10.1007/s13534-024-00351-w [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Son C-S, Kang W-S. Multivariate CNN model for human locomotion activity recognition with a wearable exoskeleton robot. Bioengineering (Basel). 2023;10(9):1082. doi: 10.3390/bioengineering10091082 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Wang X, Bai S, Li L, Li Y, Yu H. Enhancing Locomotion-Mode Recognition and Transition Prediction With (Bio)Mechanical Sensor Fusion for Intelligent Prosthetic Knees. IEEE J Biomed Health Inform. 2025;29(11):7955–68. doi: 10.1109/JBHI.2025.3583319 [DOI] [PubMed] [Google Scholar]
  • 31.Ahkami B, Kristoffersen MB, Ortiz-Catalan M. Real-time locomotion mode detection in individuals with transfemoral amputation and osseointegration. J Neuroeng Rehabil. 2025;22(1):142. doi: 10.1186/s12984-025-01672-2 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Gao F, Liu G, Liang F, Liao W-H. IMU-based locomotion mode identification for transtibial prostheses, orthoses, and exoskeletons. IEEE Trans Neural Syst Rehabil Eng. 2020;28(6):1334–43. doi: 10.1109/TNSRE.2020.2987155 [DOI] [PubMed] [Google Scholar]
  • 33.Mai J, Xu D, Li H, Zhang S, Tan J, Wang Q. Implementing a SoC-FPGA Based Acceleration System for On-Board SVM Training for Robotic Transtibial Prostheses. In: 2018 IEEE International Conference on Real-time Computing and Robotics (RCAR); 2018 1–5 Aug. 2018.
  • 34.Shell CE, Klute GK, Neptune RR. Identifying classifier input signals to predict a cross-slope during transtibial amputee walking. PLoS One. 2018;13(2):e0192950. doi: 10.1371/journal.pone.0192950 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Hu B, Li S, Chen Y, Kavi R, Coppola S. Applying deep neural networks and inertial measurement unit in recognizing irregular walking differences in the real world. Appl Ergon. 2021;96:103414. doi: 10.1016/j.apergo.2021.103414 [DOI] [PubMed] [Google Scholar]
  • 36.Lu Z, Narayan A, Yu H. A Deep Learning Based End-to-End Locomotion Mode Detection Method for Lower Limb Wearable Robot Control. In: 2020 IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS), 2020. 4091–7. 10.1109/iros45743.2020.9341183 [DOI]
  • 37.Lum PS, Shu L, Bochniewicz EM, Tran T, Chang LC, Barth J. Improving accelerometry-based measurement of functional use of the upper extremity after stroke: machine learning versus counts threshold method. Neurorehabil Neural Repair. 2020;34(12):1078–87. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38.Choudhary R, Gianey HK. Comprehensive review on supervised machine learning algorithms. In: International Conference on Machine Learning and Data Science (MLDS); 2017 14–15 Dec. 2017.
  • 39.Oh H-W, Hong Y-D. Divergent component of motion-based gait intention detection method using motion information from single leg. J Intell Robot Syst. 2023;107(4). doi: 10.1007/s10846-023-01843-0 [DOI] [Google Scholar]
  • 40.Chauhan P, Singh AK, Raghuwanshi NK. Classifying the walking pattern of humans on different surfaces using convolutional features and shallow machine learning classifiers. Proceedings of the Institution of Mechanical Engineers, Part C: J Mech En Sci. 2024;238(18):8943–54. doi: 10.1177/09544062241247956 [DOI] [Google Scholar]
  • 41.Qi Z, Song Q, Liu Y, Guo C. FSM-HSVM-based locomotion mode recognition for exoskeleton robot. Appl Sci. 2022;12(11):5483. doi: 10.3390/app12115483 [DOI] [Google Scholar]
  • 42.Marcos Mazon D, Groefsema M, Schomaker LRB, Carloni R. IMU-based classification of locomotion modes, transitions, and gait phases with convolutional recurrent neural networks. Sensors (Basel). 2022;22(22):8871. doi: 10.3390/s22228871 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 43.Shin D, Lee S, Hwang S. Locomotion mode recognition algorithm based on Gaussian mixture model using IMU sensors. Sensors. 2021;21(8):2785. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 44.Reznick E, Embry KR, Neuman R, Bolívar-Nieto E, Fey NP, Gregg RD. Lower-limb kinematics and kinetics during continuously varying human locomotion. Sci Data. 2021;8(1):282. doi: 10.1038/s41597-021-01057-9 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 45.Jamieson A, Murray L, Stankovic L, Stankovic V, Buis A. Human activity recognition of individuals with lower limb amputation in free-living conditions: A pilot study. Sensors (Basel). 2021;21(24):8377. doi: 10.3390/s21248377 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 46.Antos SA, Danilovich MK, Eisenstein AR, Gordon KE, Kording KP. Smartwatches can detect walker and cane use in older adults. Innov Aging. 2019;3(1):igz008. doi: 10.1093/geroni/igz008 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 47.Moder T, Wisiol K, Wieser M. Walking aid identification using wearables. Cham: Springer International Publishing. 2016. [Google Scholar]
  • 48.Tazawa M, Arii H, Ibe Y, Kobayashi H, Chikuda H, Wada N. Effects of walking with a cane on frontal plane hip joint loading in patients with late-stage unilateral hip osteoarthritis. Arch Rehabil Res Clin Transl. 2022;4(3):100209. doi: 10.1016/j.arrct.2022.100209 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 49.Zhang Y, Tao C, Zhang X, Guo J, Fan Y. Effects of cane use on the kinematic and kinetic of lower-extremity joints in inexperienced users. J Biomech. 2023;146:111426. doi: 10.1016/j.jbiomech.2022.111426 [DOI] [PubMed] [Google Scholar]
  • 50.Hsue B-J, Su F-C. Gait and kinematics of the trunk and lower extremities in stair ascent using quadricane in healthy subjects. Gait Posture. 2009;29(1):146–50. doi: 10.1016/j.gaitpost.2008.06.001 [DOI] [PubMed] [Google Scholar]
  • 51.HM-Goverment. Protection from falling, collision and impact: Approved Document K. 2013.
  • 52.Huang L, Qin J, Zhou Y, Zhu F, Liu L, Shao L. Normalization techniques in training DNNs: methodology, analysis and application. IEEE Trans Pattern Anal Mach Intell. 2023;45(8):10173–96. doi: 10.1109/TPAMI.2023.3250241 [DOI] [PubMed] [Google Scholar]
  • 53.Hwang S, Kim J, Yang S, Moon HJ, Cho KH, Youn I. Machine learning based abnormal gait classification with IMU considering joint impairment. Sensors. 2024;24(17):5571. doi: 10.3390/s24175571 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 54.Braun V, Clarke V. Using thematic analysis in psychology. Qualitative Res Psychology. 2006;3(2):77–101. doi: 10.1191/1478088706qp063oa [DOI] [Google Scholar]
  • 55.Youdas JW, Kotajarvi BJ, Padgett DJ, Kaufman KR. Partial weight-bearing gait using conventional assistive devices. Arch Phys Med Rehabil. 2005;86(3):394–8. doi: 10.1016/j.apmr.2004.03.026 [DOI] [PubMed] [Google Scholar]
  • 56.Thies SB, Bevan S, Wassall M, Shajan BK, Chowalloor L, Kenney L, et al. Evaluation of a novel biomechanics-informed walking frame, developed through a Knowledge Transfer Partnership between biomechanists and design engineers. BMC Geriatr. 2023;23(1):734. doi: 10.1186/s12877-023-04443-7 [DOI] [PMC free article] [PubMed] [Google Scholar]

Decision Letter 0

Gary Guerra

24 Feb 2026

Dear Dr. Wassall,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

Please submit your revised manuscript by Apr 10 2026 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

  • A letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.

  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.

  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols.

We look forward to receiving your revised manuscript.

Kind regards,

Gary Guerra, Ph.D.

Academic Editor

PLOS One

Journal Requirements:

When submitting your revision, we need you to address these additional requirements.

1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at

https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and

https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

2. Please note that PLOS One has specific guidelines on code sharing for submissions in which author-generated code underpins the findings in the manuscript. In these cases, we expect all author-generated code to be made available without restrictions upon publication of the work. Please review our guidelines at https://journals.plos.org/plosone/s/materials-and-software-sharing#loc-sharing-code and ensure that your code is shared in a way that follows best practice and facilitates reproducibility and reuse.

3. Thank you for uploading your study's underlying data set. Unfortunately, the repository you have noted in your Data Availability statement does not qualify as an acceptable data repository according to PLOS's standards.

At this time, please upload the minimal data set necessary to replicate your study's findings to a stable, public repository (such as figshare or Dryad) and provide us with the relevant URLs, DOIs, or accession numbers that may be used to access these data. For a list of recommended repositories and additional information on PLOS standards for data deposition, please see https://journals.plos.org/plosone/s/recommended-repositories.

4. We note that there is identifying data in the Supporting Information file “S1 File.pdf”. Due to the inclusion of these potentially identifying data, we have removed this file from your file inventory. Prior to sharing human research participant data, authors should consult with an ethics committee to ensure data are shared in accordance with participant consent and all applicable local laws.

Data sharing should never compromise participant privacy. It is therefore not appropriate to publicly share personally identifiable data on human research participants. The following are examples of data that should not be shared:

-Name, initials, physical address

-Ages more specific than whole numbers

-Internet protocol (IP) address

-Specific dates (birth dates, death dates, examination dates, etc.)

-Contact information such as phone number or email address

-Location data

-ID numbers that seem specific (long numbers, include initials, titled “Hospital ID”) rather than random (small numbers in numerical order)

Data that are not directly identifying may also be inappropriate to share, as in combination they can become identifying. For example, data collected from a small group of participants, vulnerable populations, or private groups should not be shared if they involve indirect identifiers (such as sex, ethnicity, location, etc.) that may risk the identification of study participants.

Additional guidance on preparing raw data for publication can be found in our Data Policy (https://journals.plos.org/plosone/s/data-availability#loc-human-research-participant-data-and-other-sensitive-data) and in the following article: http://www.bmj.com/content/340/bmj.c181.long.

Please remove or anonymize all personal information (<specific identifying information in file to be removed>), ensure that the data shared are in accordance with participant consent, and re-upload a fully anonymized data set. Please note that spreadsheet columns with personal information must be removed and not hidden as all hidden columns will appear in the published file.

5. We note that this data set consists of interview transcripts. Can you please confirm that all participants gave consent for interview transcript to be published?

If they DID provide consent for these transcripts to be published, please also confirm that the transcripts do not contain any potentially identifying information (or let us know if the participants consented to having their personal details published and made publicly available). We consider the following details to be identifying information:

- Names, nicknames, and initials

- Age more specific than round numbers

- GPS coordinates, physical addresses, IP addresses, email addresses

- Information in small sample sizes (e.g. 40 students from X class in X year at X university)

- Specific dates (e.g. visit dates, interview dates)

- ID numbers

Or, if the participants DID NOT provide consent for these transcripts to be published:

- Provide a de-identified version of the data or excerpts of interview responses

- Provide information regarding how these transcripts can be accessed by researchers who meet the criteria for access to confidential data, including:

a) the grounds for restriction

b) the name of the ethics committee, Institutional Review Board, or third-party organization that is imposing sharing restrictions on the data

c) a non-author, institutional point of contact that is able to field data access queries, in the interest of maintaining long-term data accessibility.

d) Any relevant data set names, URLs, DOIs, etc. that an independent researcher would need in order to request your minimal data set.

For further information on sharing data that contains sensitive participant information, please see: https://journals.plos.org/plosone/s/data-availability#loc-human-research-participant-data-and-other-sensitive-data

If there are ethical, legal, or third-party restrictions upon your dataset, you must provide all of the following details (https://journals.plos.org/plosone/s/data-availability#loc-acceptable-data-access-restrictions):

a. A complete description of the dataset

b. The nature of the restrictions upon the data (ethical, legal, or owned by a third party) and the reasoning behind them

c. The full name of the body imposing the restrictions upon your dataset (ethics committee, institution, data access committee, etc)

d. If the data are owned by a third party, confirmation of whether the authors received any special privileges in accessing the data that other researchers would not have

e. Direct, non-author contact information (preferably email) for the body imposing the restrictions upon the data, to which data access requests can be sent

6. If the reviewer comments include a recommendation to cite specific previously published works, please review and evaluate these publications to determine whether they are relevant and should be cited. There is no requirement to cite these works unless the editor has indicated otherwise.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

Reviewer #1: Partly

Reviewer #2: Yes

Reviewer #3: Yes

**********

2. Has the statistical analysis been performed appropriately and rigorously? -->?>

Reviewer #1: No

Reviewer #2: N/A

Reviewer #3: Yes

**********

3. Have the authors made all data underlying the findings in their manuscript fully available??>

The PLOS Data policy

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: Yes

**********

4. Is the manuscript presented in an intelligible fashion and written in standard English??>

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: Yes

**********

Reviewer #1: A. General comments

This manuscript addresses a clinically important and timely problem: the subjectivity inherent in K-level assignment for lower-limb prosthesis users. The authors propose an IMU-based machine-learning framework for real-world classification of terrain and walking-aid use, complemented by a small clinical evaluation. Real-world, multi-terrain data were collected from an amputee cohort (n = 20). Overall, the motivation is well-founded, the topic is relevant to clinical practice, and the dataset represents a valuable contribution.

B. Major comments

1. Some wording in the abstract, for example, “significant shift” (line 35) and “crucial, evidence-based” (lines 36-37), appears overstated given the evaluation design and modest performance. The authors should temper these claims and ensure that similarly strong language is appropriately qualified throughout the manuscript. In addition, given the results, the work is better framed as a feasibility and methods paper, not as a clinically ready tool.

2. The introduction interweaves clinical motivation, technical limitations, and prior qualitative findings, but does not adequately synthesize these elements into a clearly articulated research question. As written, repeated statements indicating “no conclusive evidence” read as a descriptive list rather than a coherent, focused knowledge gap. Although five objectives are presented, explicit, testable hypotheses are not clearly defined. The authors should restructure this section to more clearly articulate the central research gap and align the objectives with corresponding hypotheses.

3. Additional detail on the terrain conditions is needed to improve clarity and reproducibility. Including representative photographs for each terrain is strongly recommended.

4. It is unclear whether terrain and walking-aid conditions were randomized. The authors should clarify this in the methods.

5. The manuscript states that IMU data were filtered with a 5 Hz-pass Butterworth filter. This is commonly low for gait classification and may remove meaningful gait dynamics. The authors should justify why 5 Hz was used. Was this applied to all IMU channels?

6. IMU features are derived from three axes (X, Y, Z), but it is unclear how these components were used. The authors should specify how multi-axis data were incorporated into the features and define how each IMU axis maps to anatomical directions.

7. The authors should justify the inclusion of each model parameter. For example, why IMU placement on the trunk is expected to provide information distinct from thigh placement, and why magnetic field strength is hypothesized to contribute to the model.

8. Model hyperparameters and optimization process should be specified.

9. The results of the normality tests underlying the ANOVA are not reported.

10. What is the dependent variable in ANOVA? The authors should also report F-values and effect sizes.

11. Figures 2 and 3 show heavy misclassification. It is recommended to provide per-class metrics in the main text rather than just matrices.

12. Some interview methodology details are missing, such as the interview guide, coding procedure, and inter-rater reliability.

13. The manuscript states that clinicians reviewed the IMU-derived reports, but there is no quantitative comparison to clinician K-level ratings. Did clinicians change their K-level assessment? Did IMU data increase confidence?

14. LSO accuracies remain low, barely exceeding baselines. While CV is higher, the gap suggests overfitting, potentially due to inter-subject variability in prosthetics or terrain definitions. Subgroup analyses are recommended to potentially reveal biases, e.g., do higher-level amputees show different patterns? The authors should also critically discuss why the model generalizes poorly to new individuals and how this might impact clinical adoption.

15. While 20 participants are reasonable for an initial ML study, the gender balance is heavily skewed. The authors should discuss how this might limit the generalizability of the findings to female prosthetic users.

16. The literature review could be strengthened by adding more key points, for example, recent prosthetic-specific IMU classification studies and terrain classification using deep learning architectures.

C. Minor comments

1. Clarity would be improved by explicitly specifying the final models in the abstract.

2. Were there any exclusive criteria applied?

3. Key demographic characteristics are missing, including time since amputation and current K-level.

4. “Other shank” (line 149) could be renamed, for example, “intact limb shank” for clarity.

5. “Unstable gravel” (line 157) should be defined more precisely.

6. Did all participants attempt all terrains? Any missing data?

7. The type of walking aid is not reported.

8. There is no description of IMU calibration.

Reviewer #2: Review – PLOS ONE (PONE-D-26-02026)

Reviewer Comments

Thank you for considering me reviewing this manuscript “"Lower limb prosthetic users’ activity classification through real-world inertial sensing". Please see my comments below:

Abstract

1. The abstract frames the system as a crucial tool for improving prescriptions, yet LSO accuracies are modest (~56–62%). These risks overstating clinical applicability.

2. The difference between CV and LSO should be explained briefly; many readers equate CV with generalizable accuracy.

3. Phase-2 clinical evaluation (n=3) is very small; claims should be toned down.

4. Sensitivity/specificity or per-class performance would be more informative than overall accuracy.

Introduction

1. The introduction sometimes implies that objective data will directly improve prescription, but this link is not empirically tested.

2. Prior studies with similar LSO accuracies should be contextualized more critically.

3. The rationale for why shank placement should be sufficient could be elaborated biomechanically.

Methods

Participants

1. Only 2 females. Severe sex imbalance; discuss implications.

2. Broad mix of amputation levels; unclear whether models generalize equally.

3. Exclusion of high-activity users may bias K-level relevance.

Data collection

1. Terrain definitions rely partly on researcher judgement; reproducibility issue.

2. Number of strides per class is imbalanced.

3. Clarify whether walking speed was controlled or natural.

Signal processing

1. More justification is needed for the 5 Hz filter.

2. Person-specific normalization may reduce generalizability; please discuss.

Model development

1. Hyperparameter tuning process is not sufficiently described.

2. Clarify how overfitting was prevented.

3. Why was accuracy the only optimization metric?

Validation

1. Clarify: normality checks, multiple comparison correction, and effect sizes.

Results

1. Very large tables are difficult to interpret; consider summarizing key findings.

2. Confidence intervals might be clearer than SD.

3. Improvement after merging terrain classes appears post-hoc.

4. Per-participant variability would be useful.

5. n=3 users, 4 clinicians: Limited generalizability.

6. Experts reacted to reports, not real prescription decisions.

Discussion

1. What accuracy threshold clinicians would require,

2. How errors might change reimbursement or safety.

Conclusions

1. Currently it’s too strong. The system shows potential, but further development and larger datasets are essential before clinical deployment.

Language and terminology

1. Optima vs optimal

2. Some figure cross-references appear broken.

3. Walking aid vs walking-aid

Recommendation

Major revision. High clinical relevance, but interpretation and reporting must better align with the achieved performance.

Reviewer #3: This work has some clinical relevance which directly targets a known problem (subjective/variable K-level assessment leading to poor prescriptions) with ecologically valid data collection (outdoor/real terrains, walking aid variations, researcher-present but aiming for free-living extension). This work also includes clinician input and addresses key variables (step count, cadence/variability, terrain, aid use).

Limited details on feature engineering (exact time/frequency-domain stats per window), normalization/filtering impacts (5 Hz low-pass Butterworth), outlier/missing data handling.

Appears to be no external hold-out validation Individual datapoint windows not fully tested with CV due to leakage risk. No test-retest reliability for stride detection/cadence in this cohort or IMU attachment repeatability on prosthesis pylon.

Stride labeling based on foot-contact terrain without analysis of swing/stance contributions or optimal labeling point

It might help to refine terrain definitions (e.g., objective slope/roughness metrics, ankle kinematics as alternative); explicitly label/handle transitions; analyze feature importance. Or at least explain them in a bit more detail for readers.

What were the exact features extracted per window (e.g., mean, SD, RMS, frequency stats)? How were hyperparameters tuned, and was there risk of overfitting in the step-wise optimization?

How were strides precisely labeled for terrain (e.g., heel-strike foot placement)? Were transition strides identified/excluded? Was stride detection validated on this dataset?

Overall, this is an ambitious, clinically motivated study with strong real-world data collection and open data practices.

**********

what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy

Reviewer #1: No

Reviewer #2: Yes:  Dr. Huthaifa Atallah

Reviewer #3: No

**********

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

To ensure your figures meet our technical requirements, please review our figure guidelines: https://journals.plos.org/plosone/s/figures

You may also use PLOS’s free figure tool, NAAS, to help you prepare publication quality figures: https://journals.plos.org/plosone/s/figures#loc-tools-for-figure-preparation.

NAAS will assess whether your figures meet our technical requirements by comparing each figure against our figure specifications.

Attachment

Submitted filename: renamed_1dc7b.pdf

pone.0358882.s002.pdf (157.8KB, pdf)
PLoS One. 2026 Oct 9;21(10):e0358882. doi: 10.1371/journal.pone.0358882.r002

Author response to Decision Letter 1


15 Jul 2026

Reviewer 1

1. Some wording in the abstract, for example, “significant shift” (line 35) and “crucial, evidence-based” (lines 36-37), appears overstated given the evaluation design and modest performance. The authors should temper these claims and ensure that similarly strong language is appropriately qualified throughout the manuscript. In addition, given the results, the work is better framed as a feasibility and methods paper, not as a clinically ready tool.

Response: We agree and have tempered the language throughout the manuscript to better align with the observed performance and feasibility‑focus of the study. The abstract and Conclusion now describe the system as an exploratory feasibility and methods approach rather than a clinically ready tool. Abstract lines 38, 40, 43-47, Conclusion lines 601, 608, 611-612.

2. The introduction interweaves clinical motivation, technical limitations, and prior qualitative findings, but does not adequately synthesize these elements into a clearly articulated research question. As written, repeated statements indicating “no conclusive evidence” read as a descriptive list rather than a coherent, focused knowledge gap. Although five objectives are presented, explicit, testable hypotheses are not clearly defined. The authors should restructure this section to more clearly articulate the central research gap and align the objectives with corresponding hypotheses.

Response: we agree and now comment that it remains unclear which combination of sensor placement, window definition, sampling density and IMU features offers the best trade off between classification performance and clinical feasibility and identify this as the central research gap our study addresses. We have also added hypothesis for each objective. Introduction lines 141-148, 153-175

3. Additional detail on the terrain conditions is needed to improve clarity and reproducibility. Including representative photographs for each terrain is strongly recommended.

Response: We thank the reviewer for this helpful suggestion. We have expanded the description of each terrain in the Methods to specify how flat, ramp, stairs, uneven ground and unstable gravel were operationally defined, to improve clarity and reproducibility. In addition, we have added representative photographs illustrating each terrain type as Fig 2. Methods lines 232-235, 244

4. It is unclear whether terrain and walking-aid conditions were randomized. The authors should clarify this in the methods.

Response: We agree that this should be clarified. We have added a statement to the Methods specifying that the order of terrain conditions and the presence or absence of a walking aid were randomised for each participant to minimise order effects. methods lines 230-231

5. The manuscript states that IMU data were filtered with a 5 Hz-pass Butterworth filter. This is commonly low for gait classification and may remove meaningful gait dynamics. The authors should justify why 5 Hz was used. Was this applied to all IMU channels?

Response: We agree that this required clarification. In the revised Methods we now state that a 5 Hz low‑pass Butterworth filter was applied to all IMU channels (acceleration, free‑acceleration, angular velocity and magnetic field). The 5 Hz cut‑off was chosen based on a frequency analysis of resultant acceleration from a representative subset of strides across terrains and walking‑aid conditions, which showed that the dominant gait‑related power was concentrated below approximately 4 Hz, so the filter was expected to preserve relevant gait dynamics while attenuating higher‑frequency noise. See Methods lines 255-258.

6. IMU features are derived from three axes (X, Y, Z), but it is unclear how these components were used. The authors should specify how multi-axis data were incorporated into the features and define how each IMU axis maps to anatomical directions.

Response: We agree and have clarified both the axis definitions and how multi‑axis data were used. In the revised Methods we now state that the IMUs were mounted so that the local X, Y and Z axes were approximately aligned with the segment’s vertical (longitudinal), medio‑lateral and antero‑posterior directions, respectively, for all sensor locations. We also specify that for each channel the tri‑axial signals (X, Y, Z) were included separately in the models, and that we additionally computed the resultant magnitude for acceleration and free‑acceleration as axis‑invariant features. See Methods lines 213-223

7. The authors should justify the inclusion of each model parameter. For example, why IMU placement on the trunk is expected to provide information distinct from thigh placement, and why magnetic field strength is hypothesized to contribute to the model.

Response: We agree and have expanded the Introduction to clarify the rationale for each model parameter. We now explain that shank, thigh and trunk placements are expected to capture partly distinct information because segment kinematics differ across the lower limb and trunk when negotiating different terrains and using a walking aid. In addition, we justify inclusion of magnetic field strength by noting that it provides complementary information on sensor heading and orientation relative to the environment, which can differ between terrains such as slopes versus level ground and between stair ascent and descent. Introduction lines 116-124, 141-148, Methods line 217-221

8. Model hyperparameters and optimization process should be specified.

Response: We agree and have expanded the Methods to describe the hyperparameter optimisation procedure. For each algorithm, an automatic Bayesian hyperparameter optimiser was run once on a representative subset of the data to obtain default settings, using held‑out validation loss to select the best combination. The resulting optimal values were very similar across window types, and where paired optima differed by less than 5% of the larger value we used their mean as the final setting. See Methods lines 307-314

9. The results of the normality tests underlying the ANOVA are not reported.

Response: We agree that the normality checks should be described more clearly. In the revised Methods, we now state that for each set of model accuracies we assessed normality using the Shapiro–Wilk test. When the normality assumption was not violated (p≥0.05), we applied ANOVA with Tukey’s HSD post‑hoc tests; when it was violated (p<0.05), we instead used Kruskal–Wallis tests with Dunn’s post‑hoc comparisons. We have not listed every Shapiro–Wilk statistic in the Results because this would add substantial length without changing interpretation; instead, the revised Methods define the decision rule and the Results report the corresponding test statistics for each comparison. Methods lines 328-337

10. What is the dependent variable in ANOVA? The authors should also report F-values and effect sizes.

Response: We agree that this required clarification. In the revised Methods we now state explicitly that the dependent variable in all ANOVA/Kruskal–Wallis tests was model classification accuracy, with factors such as algorithm, sensor placement, window type and sampling rate as independent variables. We have also added the corresponding F‑values (or chi‑squared statistics for non‑parametric tests) and effect sizes: partial eta‑squared for omnibus ANOVA effects and Cohen’s d for key pairwise post‑hoc comparisons. methods line 328-337 results lines 368-370, 373, 379-382, 383-389, 391-394, 408-409, 419-420, 425-426, 429-430, 438-440, 450-451.

11. Figures 2 and 3 show heavy misclassification. It is recommended to provide per-class metrics in the main text rather than just matrices.

Response: We agree that per‑class metrics improve interpretability. In the revised Results, we now report per‑class sensitivities for the models in the main text, alongside the confusion matrices, so that readers can see which classes are most frequently misclassified. results lines 489, 494-497, 506

12. Some interview methodology details are missing, such as the interview guide, coding procedure, and inter-rater reliability.

Response: We agree and coding procedure and interview guide added to supporting document. Methods lines 353-354

13. The manuscript states that clinicians reviewed the IMU-derived reports, but there is no quantitative comparison to clinician K-level ratings. Did clinicians change their K-level assessment? Did IMU data increase confidence?

Response: We agree that a quantitative comparison with clinician K‑level ratings would be valuable. In this study, clinicians reviewed IMU‑derived reports and provided qualitative feedback, but they were not asked to assign or revise K‑levels, and we did not measure changes in K‑level decisions or rating confidence. We have clarified in the Results/Discussion that clinicians responded to the reports rather than making formal K‑level assessments, and that the present findings therefore cannot determine whether, or how, the system would change K‑level assignments or increase clinicians’ confidence. Results lines 582-588

14. LSO accuracies remain low, barely exceeding baselines. While CV is higher, the gap suggests overfitting, potentially due to inter-subject variability in prosthetics or terrain definitions. Subgroup analyses are recommended to potentially reveal biases, e.g., do higher-level amputees show different patterns? The authors should also critically discuss why the model generalizes poorly to new individuals and how this might impact clinical adoption.

Response: We agree that the gap between cross‑validation and LSO accuracies indicates limited generalisation to new individuals, likely due in part to inter‑subject variability in prostheses, gait patterns and terrain exposure. We have added a leave‑one‑participant‑out analysis and expanded the Discussion to examine participant‑level variability, potential biases and implications for clinical adoption. Results line 410-416, 452-457, discussion lines 497-500, 527-530

15. While 20 participants are reasonable for an initial ML study, the gender balance is heavily skewed. The authors should discuss how this might limit the generalizability of the findings to female prosthetic users.

Response: We agree that the skewed sex distribution is an important limitation. In the revised Methods we now compare our sample to UK lower‑limb prosthetic user demographics, noting that only 10% of participants were female versus approximately 30% nationally, and that this mismatch may limit the generalisability of our findings, particularly to female prosthetic users. Methods lines 197-204

16. The literature review could be strengthened by adding more key points, for example, recent prosthetic-specific IMU classification studies and terrain classification using deep learning architectures.

Response: We agree and have expanded the literature review to include additional prosthetic‑specific IMU classification studies and recent work on terrain classification. Introduction lines 85-95, 120-124, 127-131

C. Minor comments

1. Clarity would be improved by explicitly specifying the final models in the abstract.

Response: We agree and clarity of final model added to results. Results lines 407, 449

2. Were there any exclusive criteria applied?

Response: We agree and exclusion criteria added to methods. methods lines 192-194

3. Key demographic characteristics are missing, including time since amputation and current K-level.

Response: We agree and further explanation of demographic characteristics added to methods. We did not record K-level because many participants were unaware of their classification. Similarly, time since amputation was not recorded. Instead, we prioritised functional ability (e.g., the ability to independently traverse various terrains) as a more reliable and direct inclusion criterion for this study. We have added further explanation of these demographics to the Methods. See Methods lines 196-207

4. “Other shank” (line 149) could be renamed, for example, “intact limb shank” for clarity.

Response: We agree and “other shank” has been changed to “contralateral shank” throughout manuscript.

5. “Unstable gravel” (line 157) should be defined more precisely.

Response: We agree and explanation of gravel added to methods. methods lines 232-234

6. Did all participants attempt all terrains? Any missing data?

Response: We agree that this should be clarified. We have added Table 1, which summarises which terrain and walking‑aid conditions each participant completed and therefore makes clear that not all participants attempted every terrain. We also note in the Methods that these un-attempted conditions represent missing data for those terrain classes and explain that models were trained on the available strides only. See Methods lines 238-242.

7. The type of walking aid is not reported.

Response: Thank you for the suggestion and “The walking aid used in the study was a walking stick” has been added to methods. See Methods line 231.

8. There is no description of IMU calibration.

Response: We agree and IMU calibration explanation added to methods. See Methods lines 223-225.

Reviewer 2

Abstract

1. The abstract frames the system as a crucial tool for improving prescriptions, yet LSO accuracies are modest (~56–62%). These risks overstating clinical applicability.

Response: We agree and language toned down. Abstract lines 38-47

2. The difference between CV and LSO should be explained briefly; many readers equate CV with generalizable accuracy.

Response: sentence added to explain the accuracy difference. Abstract lines 31-35 “The difference between cross‑validation and leave‑some‑out accuracies indicates participant‑specific overfitting, so the leave-some-out accuracy is considered the current conservative estimate of performance on unseen users and the cross-validation accuracy an estimate on potential accuracy with a more robust dataset”

3. Phase-2 clinical evaluation (n=3) is very small; claims should be toned down.

Response: Language toned down. Abstract lines 38-47

4. Sensitivity/specificity or per-class performance would be more informative than overall accuracy.

Response: per-class sensitivity added to abstract. Abstract lines 29-31 “Per‑class sensitivities ranged from approximately 44–78% in leave‑some‑out analyses, with lower performance for ramp ascending and descending.”

Introduction

1. The introduction sometimes implies that objective data will directly improve prescription, but this link is not empirically tested.

Response: Clarification added to explain potential impact but current evidence is not explored. Introduction line 59-63 “One solution to improve the reliability of K level assessments could be the use of activity monitoring in the real world, to provide clinicians with objective every-day data on activities which could support clinicians in making more consistent K level assessments and justify component selection, but this relationship has not yet been empirically demonstrated.”

2. Prior studies with similar LSO accuracies should be contextualized more critically.

Response: Added analysis of previous study that used a LSO approach. Introduction lines 127-131 “The only previous study found that classified terrain for lower limb prosthetic users using a wearable sensor and assessed using a leave-one-out approach achieved a 56.68% mean accuracy, but this study had only recruited 4 prosthetic users and only classified between flat, stairs (ascending and descending) and slopes (ascending and descending)”.

3. The rationale for why shank placement should be sufficient could be elaborated biomechanically.

Response: rationale added to why it was believed shank IMU placement would produce different IMU signals. Introduction lines 116-120 “Reported differences in knee and ankle kinematics across varying terrains indicate that an IMU positioned on the shank is likely to produce distinguishable signals (44). Confirming this a single shank-mounted IMU can produce good terrain classification accuracies (0.90 F1-score) on able bodied individuals (35), but this has not been investigated on lower limb prosthetic users”

Methods

Participants

1. Only 2 females. Severe sex imbalance; discuss implications.

Response: explanation of demographics against recorded UK lower limb prosthetic user demographics added and statement about how this could affect the generalisations of findings. Methods lines 197-204 “The participant demographics did not perfectly match those of UK lower limb prosthetic users: 10% of participants were female versus approximately 30% nationally, and 85% were older than 54 years compared with around 70% in the UK population, however, the distribution of amputation levels was similar (transtibial 55% vs 54%, transfemoral 40% vs 37%, bilateral 5% vs 6%), reflecting representativeness of the data in that regard. For pragmatic reasons we did not sample to exactly mirror national demographics, which may limit generalisability of the findings, particularly with respect to sex‑specific and age‑related patterns of activity.”

2. Broad mix of amputation levels; unclear whether models generalize equally.

Response: The distribution of amputation level was broadly similar to UK service demographics, supporting relevance to the target population; however, the sample remains too small to conclude that model performance generalises equally across amputation levels.

A statement was also added to methods that states preliminary analysis that showed no significant difference to classification accuracy between amputation level. Methods line 301-306“Participants included transtibial, transfemoral and one bilateral amputee, and all models were trained on the combined dataset without stratification by amputation level. Analyses were carried out that did not show significant differences in classification accuracy between transtibial and transfemoral subgroups (6), however, the present study cannot exclude that performance varies by amputation level and this remains a limitation.”

3. Exclusion of high-activity users may bias K-level relevance.

Response: The research was aimed at K2/K3 users, which was identified in the pre-study clinical interviews. This context has now been added to the introduction and clarification added to inclusion and exclusion criteria. Introduction lines 78-79 “It was also discovered that the issues mainly occur in decisions between K2 and K3 users.” Methods lines 190-194 “As the issues with K level assessments predominantly occur with K2 and K3 users, we used the following exclusion criteria:, not regularly participating in active sport with their prosthesis (football, running, etc.) and not primarily using a wheelchair,; we note that this could limit the generalisability of the findings to K1 and K4 users.”

Data collection

1. Terrain definitions rely partly on researcher judgement; reproducibility issue.

Response: The Methods now define the terrain categories more precisely and explicitly note where researcher judgement was used, particularly for uneven ground. This is stated in Methods, lines 231-235. We also signpost that this subjectivity is considered in the Discussion, lines 531-542, as a potential contributor to misclassification.

2. Number of strides per class is imbalanced.

Response: clarification about decision on stride per class imbalance added to methods. Methods lines 246-251 “The number of recorded strides per terrain class was imbalanced, reflecting the practical difficulty of collecting data on different terrain. Class imbalance may have contributed to poorer performance for under-represented or more heterogeneous classes, but it was considered low for the data set due to the marked biomechanical differences in traversing the different terrains. In the discussion individual class imbalances are discussed”. (Discussion lines 488-508, 527-530)

3. Clarify whether walking speed was controlled or natural

Response: statement added to methods to explain that participants walked at self-selected walking speed. Methods lines 251-253 “Participants walked at their self-selected natural walking speed on each terrain, which increases ecological validity of the dataset.”

Signal processing

1. More justification is needed for the 5 Hz filter.

Response: clarification on why 5Hz was used for the filter was added to the methods. Methods lines 255-258“The cut‑off was selected based on a frequency analysis of resultant acceleration from a representative subset of strides across terrains and walking aid conditions, which showed the dominant gait‑related power below 4 Hz.”

2. Person-specific normalization may reduce generalizability; please discuss.

Response: clarification added to methods of why normalization method was chosen and reference to preliminary analysis which identified the best normalization methods. Methods 258-268 “A combination of per‑person normalised and non-normalised data were used in the models to stabilise training while preserving absolute between‑subject differences. Normalisation is known to improve optimisation and reduce sensitivity to scale, but purely person‑specific normalisation can attenuate clinically relevant between‑subject variation; including both representations was therefore chosen as a pragmatic compromise in this feasibility study (52, 53). To our knowledge, combining per person normalised and raw IMU channels within the same feature set has not been reported previously, but in preliminary analysis (6) this combination produced the most favourable balance of cross validation and leave some out accuracies, so it was adopted for the present models.

Model development

1. Hyperparameter tuning process is not sufficiently described.

Response: clarification on hyperparameter tuning added to methods. Methods lines 307-314 “For each algorithm a Bayesian hyperparameter optimiser was run to mitigate overfitting. The optimisation was run on a representative subset of the data to obtain reasonable default settings. The optimiser evaluated a loss function on held‑out data, seeking the hyperparameter combination that minimised this loss. This initial optimisation was performed once per algorithm using 100 Hz data with 0.2 s windows and stride‑based windows (100 datapoints per stride). The optimal values were very similar across window types, and the mean of these values where small differences (<5%) occurred were used.”

2. Clarify how overfitting was prevented.

Response: We did apply a Bayesian hyperparameter optimiser to mitigate overfitting, as detailed in the Methods. However, we acknowledge overfitting could still be a limitation with a statement added to the discussion. Sensitivity and specificity of final model presented via confusion matrix and per-participant performance added to results, which presents any overfitting that occurred. Discussion lines 527-530 “The cross‑validation accuracies being significantly higher than leave‑some‑out accuracies indicates participant‑specific overfitting could be present. No class‑weighted training or additional regularisation was applied to the models, which could be addressed in future work.”

3. Why was accuracy the only optimization metric?

Response: statement added to methods that states that the primary optimisation metric was overall accuracy because the target is correct class assignment across all categories and to be comparable to prior studies. Methods lines 320-323 “Overall percentage accuracy was used as the primary optimisation metric because the target application is multi‑class terrain and walking aid recognition, where correct class assignment across all categories is clinically most relevant and directly comparable with prior studies”.

Validation

1. Clarify: normality checks, multiple comparison correction, and effect sizes.

Response: statement added to methods to clarify normality checks, multiple comparison and effect size. Effect size results also added to results. Methods lines 328-337 “Before comparing model accuracies, normality was assessed for each set of values using the Shapiro–Wilk test. When all models for a given comparison (e.g. windowing method, IMU position, feature set) satisfied the normality assumption (ρ≥0.05), a one‑way ANOVA (with F-value (F), degrees of freedom and ρ-value (ρ) reported) was used with Tukey’s HSD post‑hoc tests. Kruskal-Wallis (with Chi-square (X2), degrees of freedom and ρ values (ρ) reported) with Dunn’s post-hoc test was used for non-parametric data. Significance was accepted as ρ values lower than 0.05. For significant ANOVA effects we report partial eta‑squared as an effect size; for key pairwise post‑hoc contrasts we additionally report Cohen’s d.”

Results

1. Very large tables are difficult to interpret; consider summarizing key findings.

Response: The full table has been retained because it provides transparency across models and processing choices, but the Results text has been revised so that readers are not required to interpret the table unaided. A summary sentence now introduces the table and states that the principal findings are summarised in the following results paragraphs. Clarity added to Results lines 357-359 “The mean accuracies for all models tested for terrain classification and walking aid recognition are presented in Table 2, but the main findings are summarised in the text that follows.”

2. Confidence intervals might be clearer than SD.

Response: confidence intervals added for key results. Results lines 408-409,450-451, discussion lines 504-505.

3. Improvement after merging terrain classes appears post-hoc.

Response: clarification added to results. Discussion lines 500-505 “The clinical experts in Phase 2 suggested combining grass, uneven and unstable terrain categories into one uneven terrain category. In response to their comment, a RF model with these terrain categories was created for supplementary data analysis which then produced an accuracy increase to 65.70% (SD 6.6%; 95% CI 47.3–84.1%) for LSO and to 85.65% (SD 0.79%; 95% CI 83.4-87.9%) for CV.”

4. Per-participant variability would be useful.

Response: per-participant performance added to results. Results lines 410-416 “To investigate participant variability leave-one-participant-out (LOPO) analysis was run using the final RF model, Table 3. Mean accuracies between amputation levels were similar (TT 51.8±14.6%, TF 50.4±20.0%), and Welch’s t‑test showed no significant difference between groups (p=0.86), suggesting that amputation level did not influence classification accuracy.” Table 3 and lines 452-457 “LOPO walking aid recognition accuracies showed a large inter participant variability, Table 3, but amputation level did not affect model accuracies (TT 56.6±19.0%, TF 53.2±21.4%; Welch’s t‑test p=0.71). The proportion of steps performed with a walking aid was also not significantly associated with walking aid recognition accuracy (Pearson r=−0.10, p=0.65; Spearman ρ=−0.29, p=0.20), suggesting that uneven exposure to aided steps did not systematically bias performance.”

5. n=3 users, 4 clinicians: Limited generalizability. 6. Experts reacted to reports, not real prescription decisions.

Response: paragraph added to discussion that explains the limitation of the study and that the results should not be generalised to how the system could affect clinical outcomes. Discussion lines 582-588 “Phase‑2 of the study only involved 4 clinical experts analysing reports on 3 prosthetic users. The small participant population should be considered and caution taken when generalising the findings. Moreover, clinicians responded to simulated reports rather than using the system to make assessments decisions, so we cannot determine whether, or how, the information would actually change K‑level assignments or component selection in routine practice.”

Discussion

1. What accuracy threshold clinicians would require,

2. How errors might change reimbursement or safety.

Response: paragraph added to discussion that comments on the limitation that the study did not establish acceptable accuracy thresholds or how errors could affect clinical decision making. Discussion lines 589-596 “This research did not establish accuracy thresholds clinicians would consider sufficient for K‑level assessments or prescription decisions. How specific errors might influence clinical decisions was also not investigated. In Phase 2, clinicians commented that inaccuracies could be alleviated by combining system outcomes with current assessment methods rather than replacing these. Future work should include prospective studies in which clinicians use system outputs in real prescription workflows, to determine acceptable accuracy levels and identify which types of errors are clinically tolerable, also under consideration of risks to safety.”

Conclusions

1. Currently it’s too strong. The system shows potential, but further development and larger datasets are essential before clinical deployment.

Response: language toned down in conclusion to better reflect the outcomes of the study and limitations. Conclusion line 601, 608-612.

Reviewer 3

Limited details on feature engineering (exact time/frequency-domain stats per window), normalization/filtering impacts (5 Hz low-pass Butterworth), outlier/missing data handling.

Response: We have expanded the Signal processing and Feature description subsections to clarify these points. We clarify that all IMU channels (acceleration, angular velocity, magnetic field and free acceleration) were filtered using a 5 Hz low pass Butterworth filter and justify this cut off based on frequency analysis of representative strides across terrains. See Methods line 213-225, 255-264

Appears to be no external hold-out validation Individual datapoint windows not fully tested with CV due to leakage risk. No test-retest reliability for stride detection/cadence in this cohort or IMU attachment repeatability on prosthesis pylon.

Response: No external hold out validation was conducted instead, LSO analysis was conducted and further LOPO analysis has been added to results. These analyses validate the models ability to classify unseen participants. We did not perform a new test–retest reliability study for stride detection, cadence or IMU attachment repeatability on the prosthesis pylon in this cohort. Instead, we used a previously validated stride‑detection and cadence algorithm based on the prosthetic‑shank medio‑lateral angular velocity, which in earlier work with lower‑limb prosthesis users achieved high detection accuracy and precision, and we standardised IMU mounting on the prosthesis pylon. We now acknowledge in the Discussion that we did not quantify day‑to‑day variability in IMU placement or the test–retest reliability of derived measures in this sample, and identify this as an important area for future work. We intentionally did not apply cross‑validation to individual‑datapoint windows, because datapoints from the same stride would otherwise be split across training and validation sets, causing severe information leakage and artificially inflated accuracies. We have clarified this in the Methods. Methods lines 282-285, Results 410-416, 452-457 Discussion line 579-581

Stride labeling based on foot-contact terrain without analysis of swing/stance contributions or optimal labeling point.

Response: We thank the reviewer for this comment. In the revised Methods we now clarify that each stride was labelled according to the terrain under the prosthetic foot at initial contact (heel strike), so that the label reflects the terrain over which the majority of the stance phase occurs. We did not perform a separate analysis of swing‑ versus stance‑phase signal contributions or alternative labelling points (e.g. mid‑stance or mid‑swing). See Methods lines 294-297.

It might help to refine terrain definitions (e.g., objective slope/roughness metrics, ankle kinematics as alternative); explicitly label/handle transitions; analyse feature importance. Or at least explain them in a bit more detail for readers.

Response: In the Methods, we now more precisely define each terrain class as used in the study (flat ground, stairs ascending/descending, ramps steeper than 1:20, uneven hard surfaces such as cobblestones, and unstable gravel defined as a loose, small‑stone surface that visibly deformed underfoot with each step). We acknowledge that roughness was determined pragmatically by the experimenter rather than via objective surface metrics and now discuss this subjectivity as a potential contributor to misclassification in the Discussion. We also clarify that strides were labelled according to the terrain under the prosthetic foot at initial contact, and that transition strides were not treated differently. Feature importance was not conduced but in model analysis feature sets were analysed. Methods lines 231-238 294-297.

What were the exact features extracted per window (e.g., mean, SD, RMS, frequency stats)? How were hyperparameters tuned, and was there risk of overfitting in the step-wise optimization?

Response: features were not extracted from IMU signals apart from cadence and resultant accelerations. Hyperparameter tuning explanation has been added to methods. Bayesian hyperparameter optimisation was run not step-wise, this has now been explained in methods. Methods line 213-223, 307-314.

How were strides precisely labelled for terrain (e.g., heel-strike foot placement)? Were transition strides identified/excluded? Was stride detection validated on this dataset?

Response: We agree that these details should be made explicit and have clarified them in the Methods. Strides were detected using prosthetic‑shank medio‑lateral angular velocity, and each stride was labelled according to the terrain under the prosthetic foot at heel strike. Transition strides were not separately identified. See Methods lines 294-297.

Attachment

Submitted filename: Response to Reviewers.docx

pone.0358882.s003.docx (36.7KB, docx)

Decision Letter 1

Gary Guerra

10 Aug 2026

Dear Dr. Wassall,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

Please submit your revised manuscript by Sep 24 2026 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

  • A letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.

  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.

  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols.

As the corresponding author, your ORCID iD is verified in the submission system and will appear in the published article. PLOS supports the use of ORCID, and we encourage all coauthors to register for an ORCID iD and use it as well. Please encourage your coauthors to verify their ORCID iD within the submission system before final acceptance, as unverified ORCID iDs will not appear in the published article. Only the individual author can complete the verification step; PLOS staff cannot verify ORCID iDs on behalf of authors.

We look forward to receiving your revised manuscript.

Kind regards,

Gary Guerra, Ph.D.

Academic Editor

PLOS One

Journal Requirements:

If the reviewer comments include a recommendation to cite specific previously published works, please review and evaluate these publications to determine whether they are relevant and should be cited. There is no requirement to cite these works unless the editor has indicated otherwise.

Please review your reference list to ensure that it is complete and correct. If you have cited papers that have been retracted, please include the rationale for doing so in the manuscript text, or remove these references and replace them with relevant current references. Any changes to the reference list should be mentioned in the rebuttal letter that accompanies your revised manuscript. If you need to cite a retracted article, indicate the article’s retracted status in the References list and also include a citation and full reference for the retraction notice.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

Reviewer #1: (No Response)

Reviewer #3: All comments have been addressed

**********

2. Is the manuscript technically sound, and do the data support the conclusions??>

Reviewer #1: Partly

Reviewer #3: Yes

**********

3. Has the statistical analysis been performed appropriately and rigorously? -->?>

Reviewer #1: Yes

Reviewer #3: Yes

**********

4. Have the authors made all data underlying the findings in their manuscript fully available??>

The PLOS Data policy

Reviewer #1: Yes

Reviewer #3: Yes

**********

5. Is the manuscript presented in an intelligible fashion and written in standard English??>

Reviewer #1: No

Reviewer #3: Yes

**********

Reviewer #1: The authors have substantially revised the manuscript in response to the previous comments. However, several concerns still need to be addressed before the manuscript can be considered further.

1. The values reported in the text are inconsistent with those shown in Figures 3 and 4. The authors should carefully verify all performance metrics and ensure consistency between the figures and the manuscript text. For example, in Line 508, according to Figure 3, 83.4% represents sensitivity, not specificity, whereas 50.8% represents precision, not sensitivity. The authors may find the following references useful for the appropriate calculation, application, and interpretation of these performance metrics: 10.1016/j.ipm.2009.03.002 and https://doi.org/10.3390/bioengineering11121250.

2. The authors should clarify how the accuracy values reported in Lines 424–425 were calculated.

3. Figure 2 has no panel key. Please add a key mapping each panel to its terrain type.

4. The authors should conduct a thorough typographical and grammatical proofread of the manuscript. Examples include “prothesis” in Line 220 and “recorded” in Line 255.

Reviewer #3: The manuscript addresses an important and relevant topic and is well designed and clearly presented. The methods are appropriate, the analyses are sound, and the conclusions are supported by the data. I have no major concerns regarding the manuscript and recommend its acceptance for publication.

**********

what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy

Reviewer #1: No

Reviewer #3: No

**********

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

To ensure your figures meet our technical requirements, please review our figure guidelines: https://journals.plos.org/plosone/s/figures

You may also use PLOS’s free figure tool, NAAS, to help you prepare publication quality figures: https://journals.plos.org/plosone/s/figures#loc-tools-for-figure-preparation.

NAAS will assess whether your figures meet our technical requirements by comparing each figure against our figure specifications.

PLoS One. 2026 Oct 9;21(10):e0358882. doi: 10.1371/journal.pone.0358882.r004

Author response to Decision Letter 2


24 Aug 2026

1. The values reported in the text are inconsistent with those shown in Figures 3 and 4. The authors should carefully verify all performance metrics and ensure consistency between the figures and the manuscript text. For example, in Line 508, according to Figure 3, 83.4% represents sensitivity, not specificity, whereas 50.8% represents precision, not sensitivity. The authors may find the following references useful for the appropriate calculation, application, and interpretation of these performance metrics: 10.1016/j.ipm.2009.03.002 and https://doi.org/10.3390/bioengineering11121250.

Response: We thank the reviewer for identifying this inaccuracy in our reporting. Precision, specificity and sensitivity is now correctly reported throughout the manuscript (Discussion lines 509, 510, 512, 515, 516, 528, 529, 530). We also added a sentence in the methods to state how these were calculated, Methods lines 334-338 “Per-class performance is reported via confusion matrix for the final model, with precision, sensitivity and specificity reported. For each class, sensitivity was calculated as TP/(TP+FN), precision as TP/(TP+FP), and specificity as TN/(TN+FP), where TP, TN, FP and FN denote true positives, true negatives, false positives and false negatives, respectively; metrics were calculated using a one-versus-rest approach”.

2. The authors should clarify how the accuracy values reported in Lines 424–425 were calculated.

Response: We thank the reviewer for identifying that the calculation and aggregation of the reported accuracy values was insufficiently described. We have revised the Methods and Results to clarify that accuracy was calculated at the stride-window level as the proportion of held-out windows correctly classified (Methods line 332-333 “Accuracy was calculated for each validation test set as the proportion of held-out stride windows correctly classified”). We now specify how mean accuracy, SD and 95% confidence intervals were calculated across validation iterations (Methods line 342-345 “The mean accuracy across the five test folds/groups are reported for both LSO and CV, with standard deviation (SD) calculated across fold/group accuracies and 95% confidence intervals (CI) calculated using the Student’s t distribution (±t_0.975,4×SD/√5)”). The reported values have been revised to add clarity to this (results lines 425-428 “the re-optimised model achieved a mean LSO accuracy across held-out participant groups of 56.89% (SD 5.5%; 95% CI 50.1–63.7%) and a mean CV accuracy of 83.64% (SD 0.8%; 95% CI 82.6–84.7%)” and lines 469-470 “and achieved a mean LSO accuracy across held-out participant groups of 61.56% (SD 16.1%; 95% CI 45.4-77.7%) and a mean CV accuracy of 82.90% (SD 2.1%; 95% CI 80.8-85.0%)”).

3. Figure 2 has no panel key. Please add a key mapping each panel to its terrain type.

Response: We thank the reviewer for identifying that Figure 2 does not have a panel key. A,B,C,D added to figure and key added to figure legend, Methods tine 274-275 “A) Uneven ground (cobble stones), B) Grass, C) Unstable gravel, D) Stairs.”.

4. The authors should conduct a thorough typographical and grammatical proofread of the manuscript. Examples include “prothesis” in Line 220 and “recorded” in Line 255.

Response: We thank the reviewer for identifying the errors in the manuscript. The manuscript has now been checked with mistakes corrected throughout the manuscript.

Attachment

Submitted filename: response_to_reviewers_auresp_2.docx

pone.0358882.s004.docx (16.1KB, docx)

Decision Letter 2

Gary Guerra

7 Sep 2026

Lower limb prosthetic users’ activity classification through real-world inertial sensing

PONE-D-26-02026R2

Dear Dr. Wassall

We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.

Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.

An invoice will be generated when your article is formally accepted. Please note, if your institution has a publishing partnership with PLOS and your article meets the relevant criteria, all or part of your publication costs will be covered. Please make sure your user information is up-to-date by logging into Editorial Manager at Editorial Manager® and clicking the ‘Update My Information' link at the top of the page. For questions related to billing, please contact billing support.

If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

Kind regards,

Gary Guerra, Ph.D.

Academic Editor

PLOS One

Acceptance letter

Gary Guerra

PONE-D-26-02026R2

PLOS One

Dear Dr. Wassall,

I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS One. Congratulations! Your manuscript is now being handed over to our production team.

At this stage, our production department will prepare your paper for publication. This includes ensuring the following:

* All references, tables, and figures are properly cited

* All relevant supporting information is included in the manuscript submission,

* There are no issues that prevent the paper from being properly typeset

You will receive further instructions from the production team, including instructions on how to review your proof when it is ready. Please keep in mind that we are working through a large volume of accepted articles, so please give us a few days to review your paper and let you know the next and final steps.

Lastly, if your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

You will receive an invoice from PLOS for your publication fee after your manuscript has reached the completed accept phase. If you receive an email requesting payment before acceptance or for any other service, this may be a phishing scheme. Learn how to identify phishing emails and protect your accounts at https://explore.plos.org/phishing.

If we can help with anything else, please email us at customercare@plos.org.

Thank you for submitting your work to PLOS One and supporting open access.

Kind regards,

PLOS One Editorial Office Staff

on behalf of

Dr. Gary Guerra

Academic Editor

PLOS One

Associated Data

    This section collects any data citations, data availability statements, or supplementary materials included in this article.

    Supplementary Materials

    S1 File. Supplement with detailed qualitative interview results from “Lower limb prosthetic users’ activity classification through real-world inertial sensing”.

    (PDF)

    pone.0358882.s001.pdf (572.2KB, pdf)
    Attachment

    Submitted filename: renamed_1dc7b.pdf

    pone.0358882.s002.pdf (157.8KB, pdf)
    Attachment

    Submitted filename: Response to Reviewers.docx

    pone.0358882.s003.docx (36.7KB, docx)
    Attachment

    Submitted filename: response_to_reviewers_auresp_2.docx

    pone.0358882.s004.docx (16.1KB, docx)

    Data Availability Statement

    The minimal data set and accompanying MATLAB classification code generated during this study are available at Figshare via https://doi.org/10.6084/m9.figshare.32756211 and https://doi.org/10.6084/m9.figshare.32805296.


    Articles from PLOS One are provided here courtesy of PLOS