A Discrete Hemodynamic Control System: Proof-of-Concept In-Vivo Study in Canine Heart Failure Using a Pulmonary Artery Catheter

Yasuyuki Kataoka 1,✉, Kazunori Uemura 2, Takuya Nishikawa 2, Yukiko Fukuda 1, Jon Peterson 1, Kei Sato 2, Hiroki Matsushita 2, Hidetaka Morita 2, Yuta Nakamura 2, Masahiro Otake 2, Kenta Ohba 2, Nana Hiraki 2, Keita Saku 2, Mitsuji Sampei 3, Joe Alexander 1, Kenji Sunagawa 4
PMCID: PMC13667294  PMID: 42877831

Abstract

Objective: Managing heart failure requires simultaneous regulation of mean arterial pressure, cardiac output, and left atrial pressure, which is challenging due to complex cardiovascular interactions and limited monitoring. This study presents a proof-of-concept in-vivo evaluation of a discrete two-degree-of-freedom hemodynamic control system designed to regulate these three hemodynamic variables through titration of cardiovascular interventions under pulmonary artery catheter monitoring. Results: Across five consecutive trials in three canine models of heart failure induced by tachycardia pacing, the cohort-averaged mean arterial pressure, cardiac output, and left atrial pressure were each within ±10% of their respective targets when evaluated using the final three intervention measurements, with absolute percentage errors decreasing from 16.8% to 9.0%, 23.4% to 3.8%, and 87.5% to 5.5%, respectively. In a more severely complicated heart failure model with coronary embolization, outcomes revealed control limitation primarily due to pharmacological resistance, highlighting the importance of choosing the appropriate target value. Conclusions: This proof-of-concept in-vivo study demonstrates the feasibility of implementing discrete two-degree-of-freedom hemodynamic control under pulmonary artery catheter-based monitoring and phlebotomy-mediated volume removal in anesthetized canine models of tachycardia-pacing-induced heart failure. Further studies are required to evaluate the effects of inter-subject variability and clinical generalizability.

Keywords: Hemodynamic control system, in-vivo validation, heart failure

I. Introduction

A. Clinical Challenges in Hemodynamic Management

In managing patients with heart failure (HF), major guidelines recommend optimizing mean arterial pressure (MAP), cardiac output (CO), and left atrial pressure (Inline graphic) to preserve organ perfusion and reduce pulmonary congestion [1], [2], [3]. Simultaneous control of MAP, CO, and Inline graphic remains difficult even for experienced clinicians due to complex cardiovascular interactions and limited monitoring.

Autonomous pharmacologic control systems of the three hemodynamic variables have been proposed, yet many depend on continuous measurements of the metrics through highly invasive sensors under thoracotomy, restricting their clinical applicability [4], [5], [6].

In intensive care or coronary care units (ICUs or CCUs) using a pulmonary artery catheter (PAC), CO and Inline graphic are usually measured intermittently by thermodilution technique and catheter balloon inflation, which reflects the pulmonary artery wedge pressure (PAWP), respectively [7], [8]. Noninvasive continuous CO monitoring methods exist, yet their accuracy remains limited [9].

B. Hemodynamnic Control System

Since CO and Inline graphic are discretely available in real clinical practice, we developed a discrete two-degree-of-freedom (2DoF) hemodynamic control framework [10], consisting of feedforward and feedback control. Unlike previous approaches relying on continuous hemodynamic monitoring under thoracotomy, the proposed framework is designed for PAC-based discrete measurements routinely available in clinical practice. Fig. 1 shows one implementation, where the input Inline graphic comprises five interventions, the state Inline graphic includes six cardiovascular parameters, and the output Inline graphic denotes five hemodynamic variables. A discrete step spans from a drug infusion update to hemodynamic stabilization. At each step, Inline graphic is updated once. Over the multiple steps, Inline graphic are regulated. Previous in-silico studies validated the system [10] (supplementary materials S-I & S-II). Translating it in vivo, however, raises critical concerns.

Figure 1.

Figure 1.

Overview of Discrete Two-Degree-of-Freedom Hemodynamic Control System.

First, measurement and estimation errors can affect control performance. Second, unavoidable discrepancies between the analytical model used in the system (Fig. 1) and the actual in-vivo physiology can also affect performance. The analytical model is grounded in physiological causality using a drug infusion model (Inline graphic) and circulatory model (Inline graphic) [10], [11]. The discrepancies may arise from model simplifications and parameter uncertainties, including variability in drug sensitivities.

To date, this system has not been tested in vivo for its ability to suppress these disturbances.

C. Study Objectives

We aimed to validate the system control in canine HF models under the real-world disturbances.

The main protocol included five consecutive experiments in three tachycardia-induced HF animals, where we examined how accurately the system controlled the three hemodynamic variables to within ±10% of their targets. The supplemental protocol included two experiments conducted in a single animal of tachycardia-induced HF complicated by coronary embolization, where we examined control limitations and their dependence on control targets.

In each experiment, five discrete drug interventions were applied to the animal to regulate the three metrics toward their targets.

II. Results

A representative experimental time trace is shown in Fig. 2, which illustrates five discrete drug interventions that span approximately one hour, with continuous measurements of MAP, Inline graphic, and PAP, alongside discrete measurements of cardiac index (CI) and Inline graphic via PAC. Let CI denote Inline graphic, which is Inline graphic:CO normalized to body surface area, providing a clinically relevant index for scaling CO across individuals. The hemodynamic variables MAP, CI, and Inline graphic gradually converge toward their target values, Inline graphic mmHg, Inline graphic L/min/mInline graphic, and Inline graphic mmHg. These fixed-point targets were selected to facilitate quantitative evaluation of control performance while remaining within clinically acceptable hemodynamic ranges.

Figure 2.

Figure 2.

Representative Experimental Data. Inline graphic:MAP, continuously monitored by arterial line. Inline graphic:CI, discretely measured via thermodilution using a PAC. Inline graphic:Inline graphic, discretely estimated from PAWP measurements obtained via PAC. Inline graphic:Inline graphic, continuously monitored via PAC. Inline graphic:PAP, continuously monitored via PAC. Note: Manual measurements (Inline graphic, Inline graphic) were performed after a 10-min observation. Due to manual procedures, the time intervals between discrete steps were not strictly uniform.

A. Main Protocol

The hemodynamic variables \bm y measured in the main protocol are summarized in Fig. 3. Based on cohort-averaged CI and Inline graphic in Fig. 3 (b), the baseline hemodynamics corresponded to Subset II in the Forrester Classification [12].

Figure 3.

Figure 3.

Results of hemodynamic variables Inline graphic in main protocol

The control performance metrics are indicated in Table 1. For evaluating the control performance, absolute percentage errors (APE [%] Inline graphic) were calculated for MAP, CI, and Inline graphic, allowing each convergence performance to be evaluated. Given the limited sample size, results are reported descriptively (mean ± standard deviation) without formal inferential statistical testing.

TABLE 1. Performance Results for the Main Protocol. APE: Absolute Percentage Error. Before: Baseline At k=0. After: Average of the Measurements Obtained At k=3–5. Inline graphic or Inline graphic: Extrema of Continuous Measurements. Inline graphic or Inline graphic: Extrema of Post-Intervention Measurements.

APE for MAP APE for CI APE for Inline graphic MAP(t) CI[k] Inline graphic HR(t)
EID before after before after before after minInline graphic maxInline graphic minInline graphic maxInline graphic minInline graphic maxInline graphic
exp1 4.1% 13.5% 20.4% 11.9% 79.3% 5.0% 59.7 93.1 3.1 13.5 96 122
exp2 20.2% 3.6% 22.9% 2.8% 125.0% 4.8% 64.3 107.0 2.4 11.4 117 149
exp3 11.4% 5.3% 33.7% 3.4% 47.2% 3.4% 66.3 88.9 3.3 10.2 121 149
exp4 18.5% 10.9% 8.0% 0.2% 111.1% 7.1% 73.9 106.8 3.1 10.1 70 155
exp5 29.8% 11.4% 32.1% 0.8% 75.1% 7.2% 69.9 110.3 2.9 11.4 96 127
Mean±SD 16.8±9.7% 9.0±4.2% 23.4±10.4% 3.8±4.7% 87.5±30.9% 5.5±1.6% 66.8±5.4 101.2±9.6 3.0±0.3 11.3±1.4 105.2±22.3 135.2±24.4

From the baseline to the average of the third to fifth intervention results, the cohort-averaged APEs of MAP, CI, and Inline graphic decreased from 16.8% to 9.0%, 23.4% to 3.8%, and 87.5% to 5.5%, respectively, indicating that each cohort-averaged variable was regulated to within ± 10% of its target. Achieving simultaneous regulation of MAP, CI, and Inline graphic across all five experiments would require relaxing the convergence criterion to ±13.5%.

During the transient response following treatment initiation, hypotension (MAP < 65 mmHg [13]) was observed in 2 of the 5 experiments. In contrast, low perfusion (CI < 2.2 L/min/mInline graphic [12]) and pulmonary congestion (Inline graphic mmHg [12]) were not observed after treatment initiation (supplementary materials S-III).

The changes in input Inline graphic and state Inline graphic in the main protocol are shown in Fig. 4. Despite uniform HF induction across subjects, different drug combinations and dosages were used.

Figure 4.

Figure 4.

Results of Interventions Inline graphic and Inline graphic in both main and supplemental protocol

B. Supplemental Protocol

The hemodynamic variables \bm y in the supplemental protocol are shown in Fig. 5. In both experiments, the hemodynamics at baseline was classified as Subset IV in the Forrester Classification, indicating more severe HF condition than the cases in the main protocol. The corresponding changes in input Inline graphic and state Inline graphic are shown in Fig. 4.

Figure 5.

Figure 5.

Results of hemodynamic variables Inline graphic in supplemental protocol:severe heart failure case

In exp 6, when target CI was set to 4.5L/min/mInline graphic, the system failed to achieve convergence within the 10% error range even though the system continuously increased inotropes Inline graphic:DOB. In contrast, in exp 7, when target CI was reset to 3.0L/min/mInline graphic, all three hemodynamic variables (MAP, CI, and Inline graphic) converged within 10% of their respective target values. Clearly in exp 6, the left ventricular contractility, Inline graphic:Inline graphic, remained suppressed and did not improve in response to the inotropes. These results suggest that further optimization, including appropriate selection of the target value, is required especially under complex pathophysiological conditions.

III. Discussion

This study presents the first in-vivo validation of the performance of the discrete 2DoF hemodynamic control system using PAC. With this clinically established but intermittent measurement modality of CI and Inline graphic, the system achieved simultaneous regulation of MAP, CI, and Inline graphic in a clinically likely setting, demonstrating a significant advance toward its clinical implementation.

A. Clinical Interpretation and Implications

1). Measurement Feasibility

By utilizing PAC, routinely available in ICUs and CCUs, the system worked effectively under realistic discrete hemodynamic measurements and cardiovascular parameter estimation methods [14], [15] (Supplementary Materials S-IV).

2). Robust Control Performance

Despite the control algorithm relying on a simplified analytical model with a provisional drug library, the system regulated hemodynamics within ±10% of targets across multiple interventions, demonstrating robustness in in-vivo conditions against model discrepancies, as well as measurement and estimation errors. These results validate feedback compensation around approximate drug-effect directions rather than the specific entries of the drug library matrix.

3). Interpretability and Transparency

The control strategy of the system is based on established cardiovascular causality, allowing prediction of intermediate physiological states and outputs (supplementary materials V). This enables identification of causes of control failure (e.g. less sensitivity of Inline graphic in exp 6), providing transparency, debuggability, and clinical accountability.

4). Potential for Treatment Recommendation Tool

By updating interventions only after hemodynamic assessment at each discrete step, the proposed framework is compatible with the clinical workflow in which treatment decisions are revised based on newly obtained hemodynamic information. Thus, the proposed system is intended to initially function as a clinical decision-support tool, allowing treatment recommendations to be incorporated into routine clinical assessments [16].

B. Main Protocol

1). MAP Control

During control, MAP often remained below Inline graphic, likely due to physiological effects not captured by the model. In most trials (except exp 5), phlebotomy-induced SBV reduction appeared to trigger the Bezold–Jarisch reflex [17], reducing HR, CI, and consequently MAP. Although this reflex was not modeled, its effect was compensated for by MIMO feedback control.

For example, to compensate for the reduction in MAP, increasing Inline graphic or SBV could adversely affect CI or Inline graphic, respectively, whereas increasing Inline graphic improves both MAP and CI without worsening Inline graphic.

Representative examples are exp 1 (k=1 → 2) and exp 4 (k=2 → 3) shown in Fig. 4. In exp 1, the controller selected DOB and PHL, whereas in exp 4 it selected DOB, SNP, and HES. These distinct interventions illustrate how the MIMO feedback controller selected different therapeutic strategies according to the hemodynamic condition.

Remaining challenges include the system's inability to respond during transients, allowing short hypotensive episodes. To address this, we proposed a hybrid hemodynamic control strategy that continuously prioritizes MAP while adjusting CO and Inline graphic discretely [18].

2). CI Control

CI was affected by HR reduction. In exp 4, HR dropped from 122.4 to 87.95 bpm between k=1 and k=2, reducing CI from 4.34 to 3.05 L/min/mInline graphic. The system responded with large adjustments in DOB, SNP, and HES at the next step, restoring CI near the target at k=3 despite only partial recovery of HR.

Measurement noise and rounding errors caused variability. In thermodilution, the mean standard deviation of the three lowest-variance samples was 0.125 L/min/mInline graphic (Inline graphic% of Inline graphic). The device's one-decimal display of CO introduced rounding errors up to ±0.11 L/min/mInline graphic (Inline graphic% of Inline graphic). Despite these measurement errors, low perfusion (CI ≤ 2.2 L/min/mInline graphic) was avoided in all cases in the main protocol.

3). Inline graphic Control

According to the circulatory equilibrium framework [11], Inline graphic can be lowered by reducing SBV and enhancing the cardiac function curve. In this study, interventions explicitly incorporating this principle demonstrated highly precise Inline graphic control.

Averaging two PAWP measurements caused measurement variability. Across all experiments, the mean standard deviation was 0.32 mmHg (Inline graphic% of Inline graphic = 10 mmHg). Despite these measurement errors, pulmonary edema (Inline graphic mmHg) was avoided in all cases in the main protocol.

Phlebotomy was used instead of diuretics to reduce experimental variability and validate the concept. In clinical practice, diuretic effects depend on renal function and transrenal pressure gradient. Future studies should incorporate diuretics and their variability.

C. Supplemental Protocol

1). Strategies in Limited Pharmacological Therapy

In exp 6, cardiac contractility did not improve despite increased dose of inotrope, indicating drug resistance. The new target was manually selected based on the observed pharmacological limitation in exp 6. In future systems, poor drug responsiveness and reachable hemodynamic changes should ideally be identified proactively to support adaptive target setting. In exp 7, control was achieved by lowering the CI target; however, higher Inline graphic target settings may also contribute to higher CI according to the Frank-Starling principle.

Overall, the results highlight the importance of adaptive target setting. Future supervisory logic may assess target feasibility, identify unreachable objectives, and dynamically redefine target regions according to the patient's physiological condition and therapeutic priorities.

2). Strategies to Avoid Coronary Hypoperfusion

In the supplemental protocol, the initial MAP was low. The current system treats MAP, CI, and Inline graphic with equal priority, using SNP to reduce afterload for CI improvement and using phlebotomy to decrease SBV for Inline graphic improvement. While both interventions are reasonable for CI and Inline graphic, these decrease MAP, risking coronary hypoperfusion. Experimentally, coronary arterial pressure below 67.0 mmHg reduces Inline graphic [19], further worsening CI and complicating hemodynamic control. For clinical application, the system will need to be fine-tuned to prioritize MAP, and target trajectories for Inline graphic should be designed to avoid abrupt reductions.

D. Other Limitations

1). Uniform Drug Library

Pharmacological responses are subject to inter- and intra-individual variability as well as nonlinear dose–response characteristics. Real-time identification of the drug library is desirable for achieving more individualized and accurate hemodynamic control [20]. For clinical translation, the canine-derived drug library will require allometric scaling and calibration using human physiological and pharmacological data.

2). Phlebotomy Instead of Diuretics

Decreasing SBV was precisely achieved via phlebotomy as a proof-of-concept study, although this approach is not used in standard HF treatment. In clinical practice, variability in diuretic responsiveness may compromise SBV and consequently Inline graphic regulation. Future in-vivo studies should therefore incorporate diuretic therapy.

3). Convergence Speed

One disadvantage of the discrete control compared with continuous feedback control is its slower convergence. Therefore, assessing the number of intervention updates required for effective hemodynamic control is critical. However, because phlebotomy rather than diuretics was used in the present experiments, drawing definitive conclusions from this study is not warranted.

4). Convergence Performance

The clinical goals for optimizing MAP, CI, and Inline graphic are range-based, and the present results meet these criteria. Implementing more complex tasks in in-vivo conditions, including minimization of myocardial oxygen consumption (Inline graphic) [10], [20], may require the system to achieve more precise hemodynamic regulation, making fine-tuning indispensable.

Future extensions may explicitly incorporate physiological reflexes, such as the Bezold–Jarisch reflex, into the predictive model. However, whether increased physiological fidelity improves control performance without compromising robustness remains an open question.

5). Sample Size and Diversity

As a proof-of-concept study, the limited sample size precludes assessment of inter-subject variability and generalizability. Larger studies are needed to evaluate these aspects.

IV. Conclusion

This study presents the first in vivo validation of a discrete two-degree-of-freedom hemodynamic control system regulating MAP, CO, and Inline graphic by titrating the cardiovascular drugs under the pulmonary artery catheter monitoring. In tachycardia-induced HF canine models, cohort-averaged hemodynamic variables were regulated to each within ±10% of their respective targets after five interventions despite real-world uncertainties, including pharmacological variability, model discrepancies, and PAC-related measurement noise. Control performance varied in a more complex HF model with coronary embolization, highlighting the need for tuning such as target value selection. These findings provide initial in-vivo evidence supporting the feasibility of closed-loop hemodynamic control in anesthetized tachycardia-pacing canine HF under PAC monitoring and phlebotomy-mediated volume removal. Future work should assess generalizability and incorporate real-world variability, including diuretic response.

Materials and Methods

A. In-Vivo Experiments

1). Main & Supplemental Protocols

In the main protocol, the objective was to evaluate control performance in simultaneously regulating MAP, CI, and Inline graphic, with target values of 85mmHg, 4.5L/min/mInline graphic, and 10 mmHg, respectively, over five intervention updates. Five consecutive experiments were conducted in three canine models of tachycardia-induced HF.

In the supplemental protocol, the objective was to demonstrate the limitations of the control system and pharmacological intervention, and their dependence on the target values. Two consecutive experiments were conducted in a canine model of severe HF induced by tachycardia and coronary embolization.

A total of seven experiments were conducted in four canines as described in Table 2.

TABLE 2. Experiment Setups; EID = Experimental ID, AID = Animal ID, Inline graphic = Target CI Value.
EID mark protocol AID init. Subset Inline graphic
exp 1 ∘ main a Subset I 4.5
exp 2 ∘ main b Subset II 4.5
exp 3 ∘ main b Subset I 4.5
exp 4 ∘ main c Subset II 4.5
exp 5 ∘ main c Subset I 4.5
exp 6 ▿ supplement d Subset IV 4.5
exp 7 △ supplement d Subset IV 3.0

2). Animal Preparation

Animal housing and experimental procedures followed the ARRIVE guidelines [21] and the Guiding Principles for the Care and Use of Animals in the Field of Physiological Sciences [22] approved by the Physiological Society of Japan. All protocols were reviewed and approved by the Animal Subjects Committee at the National Cerebral and Cardiovascular Center (Permission numbers:24020).

The specifications of the four beagle dogs (Kitayama Labes Ltd, Gifu, Japan) [female/male = 1/3, body weight 10 ± 1 kg, age 1 years]. For all, HF was induced by pacing the heart at 240 in bpm starting 28 days before the experiment. On the experimental day, animals were anesthetized, intubated, and mechanically ventilated with continuous isoflurane inhalation. Isoflurane concentration was manually adjusted to ensure MAP ≤ 110 only once before the experiment to prevent unintended emergence from anesthesia. Animals were placed supine. A surface ECG was recorded to monitor HR. Arterial pressure was measured via a 22 G catheter inserted into the right radial artery and connected to a pressure transducer (DX-200, Nihon Kohden, Tokyo, Japan). An 8-Fr catheter was placed in the right femoral vein for drug/fluid administration and another in the left femoral vein for phlebotomy. A 8-Fr catheter was inserted into the left internal jugular vein, through which a 5-Fr PAC (Edwards LifeSciences) was introduced. Zero pressure was referenced to the junction of the vena cava and right atrium [23]. One out of the four animals underwent coronary embolization, causing partial occlusion of a coronary artery and inducing acute myocardial infarction for the supplemental protocol. Trans-thoracic ultrasound echo-cardiography revealed that left ventricular ejection fraction decreased from 59.0 ± 4.1 % before the pacing to 34.0 ± 3.5 % at the experiment (supplementary materials S-VI), indicating that the pacing with/without the coronary embolization successfully induced HF. The complete setup is shown in Fig. 6.

Figure 6.

Figure 6.

Experiment Setup for Canine Model.

For repeated experiments in the same animal, drug infusions were discontinued and a washout/stabilization period of at least 1 h was allowed. Blood removed by phlebotomy was reinfused before the subsequent experiment. Because hemodynamic conditions were not expected to return to identical baseline values, repeated experiments were treated as distinct control scenarios while following the principles of the 3Rs (Replacement, Reduction, and Refinement) in animal research.

3). Measurement and Estimation

This setup allowed continuous measurement of Inline graphic:MAP(t), Inline graphic:Inline graphic, Inline graphic:PAP(t) and Inline graphic:HR(t). Using PAC, Inline graphic:CO[k] by thermodilution with 5ml cold saline bolus injection and Inline graphic:Inline graphic by PAWP are obtained discretely.

The states Inline graphic were estimated based on these measurements Inline graphic. The systemic and pulmonary resistances, Inline graphic:Inline graphic and Inline graphic:Inline graphic were discretely computed by basic vascular resistance model. The left and right ventricular elastances, Inline graphic:Inline graphic and Inline graphic:Inline graphic, were estimated using inverse end-systolic pressure volume relationship estimation method [14]. The stressed blood volume, Inline graphic:SBV, was estimated using the venous return surface [15]. (supplementary materials S-IV)

4). Intervention

The inputs Inline graphic computed by the 2DoF hemodynamic control system were administered as follows. The inputs Inline graphic:DOB, Inline graphic:NE, and Inline graphic:SNP were delivered via an electronically controlled syringe pump (TerufusionInline graphic, TERUMO, Japan) at constant rates in μg/min/kg through a right femoral vein catheter. Fluid infusion for volume elevation Inline graphic:HES and phlebotomy for volume reduction Inline graphic:PHL were performed manually via the femoral vein at approximately 30 mL/min. Phlebotomy was chosen instead of diuretics for its rapid onset, precision, and independence from renal function, enabling rigorous evaluation of the control system under idealized conditions.

The 2DoF control scheme was applied with the parameter settings in two phases. (supplementary materials S-I & S-II) At the initial step (k=0 to k=1), substantial hemodynamic changes were induced while avoiding overshoot and undershoot using feedforward control, with 70% of the current-to-target gap set as an intermediate goal. In the subsequent phase (k ≥ 1), hemodynamics were gradually driven toward the target values using feedback control, as changes were smaller. Drug-specific upper dose limits are imposed as optimization constraints (S-15), and infeasibility indicates that the target is unreachable within the predefined dose limits.

5). Control Experiment Procedure

Discrete recordings were performed in the following sequence:

  • 1)

    10-s averages of MAP, Inline graphic, PAP, and HR;

  • 2)

    10-s average of PAWP to estimate Inline graphic;

  • 3)

    Four thermodilution measurements of CO, with the three most consistent values averaged;

  • 4)

    10-s average of PAWP to estimate Inline graphic;

  • 5)

    10-s averages of MAP, Inline graphic, PAP, and HR;

  • 6)

    Averaging these values for Inline graphic at the step;

  • 7)

    Estimation of the cardiovascular state Inline graphic at the step;

  • 8)

    Administration of the optimal input Inline graphic at the step;

  • 9)

    10-min waiting period for physiological stabilization;

  • 10)

    Steps 1–9 were repeated until step k=5.

Post-thermodilution measurements were performed to account for changes in SBV from four 5 mL cold-saline injections and to calculate the average hemodynamics before and after the 20 mL infusion. MAP, Inline graphic, PAP, PAWP, and HR were recorded using LabChart (ADInstruments Ltd., Dunedin, New Zealand), while CO measured by thermodilution was manually entered into the same tool.

Appendix

Supplementary materials provided online include:

  • I.

    2DoF Hemodynamic Control System

  • II.

    Parameter values of the model and the controller

  • III.

    Experimental data of CI and Inline graphic

  • IV.

    Discrete CV Parameters Inline graphic estimation methods

  • V.

    Visualization tool for hemodynamic control

  • VI.

    Left Ventricular Ejection Fraction before and after HF induction

along with a nomenclature.

Supplementary Materials
supp1-3723319.pdf (65KB, pdf)
supp2-3723319.pdf (235.3KB, pdf)

References

  • [1].McDonagh T. A. et al. , “2021 ESC guidelines for the diagnosis and treatment of acute and chronic heart failure: Developed by the task force for the diagnosis and treatment of acute and chronic heart failure of the European society of cardiology (ESC) with the special contribution of the heart failure association (HFA) of the ESC,” Eur. Heart J., vol. 42, no. 36, pp. 3599–3726, 2021. [DOI] [PubMed] [Google Scholar]
  • [2].Heidenreich P. A. et al. , “2022 AHA/ACC/HFSA guideline for the management of heart failure: A report of the american college of cardiology/american heart association joint committee on clinical practice guidelines,” Circulation, vol. 145, no. 18, pp. e895–e1032, 2022. [DOI] [PubMed] [Google Scholar]
  • [3].Kitai T. et al. , “JCS/JHFS 2025 guideline on diagnosis and treatment of heart failure,” J. Cardiac Failure, vol. 31, pp. 1164–1322, 2025. [DOI] [PubMed] [Google Scholar]
  • [4].Uemura K. et al. , “Automated drug delivery system to control systemic arterial pressure, cardiac output, and left heart filling pressure in acute decompensated heart failure,” J. Appl. Physiol., vol. 100, no. 4, pp. 1278–1286, 2006. [DOI] [PubMed] [Google Scholar]
  • [5].Uemura K., Sunagawa K., and Sugimachi M., “Computationally managed bradycardia improved cardiac energetics while restoring normal hemodynamics in heart failure,” Ann. Biomed. Eng., vol. 37, no. 1, 2008, Art. no. 82. [DOI] [PubMed] [Google Scholar]
  • [6].Nishikawa T., Uemura K., Hayama Y., Kawada T., Saku K., and Sugimachi M., “Development of an automated closed-loop β-blocker delivery system to stably reduce myocardial oxygen consumption without inducing circulatory collapse in a canine heart failure model: A proof of concept study,” J. Clin. Monit. Comput., vol. 36, pp. 849–860, 2022. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [7].Connolly D. C., Kirklin J. W., and Wood E. H., “The relationship between pulmonary artery wedge pressure and left atrial pressure in man,” Circulation Res., vol. 2, no. 5, pp. 434–440, 1954. [DOI] [PubMed] [Google Scholar]
  • [8].Kadosh B. S. et al. , “Pulmonary artery catheter use and mortality in the cardiac intensive care unit,” JACC, Heart Failure, vol. 11, no. 8, pp. 903–914, 2023. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [9].Ruste M., Jacquet-Lagrèze M., and Fellahi J.-L., “Advantages and limitations of noninvasive devices for cardiac output monitoring: A literature review,” Curr. Opin. Crit. Care, vol. 29, no. 3, pp. 259–267, 2023. [DOI] [PubMed] [Google Scholar]
  • [10].Kataoka Y. et al. , “A discrete hemodynamic control framework: Proof-of-concept study for autonomous drug therapy in acute heart failure,” IEEE Trans. Biomed. Eng., vol. 73, no. 10, pp. 3442–3454, 2026. [DOI] [PubMed] [Google Scholar]
  • [11].Uemura K. et al. , “A novel framework of circulatory equilibrium,” Amer. J. Physiol.-Heart Circulatory Physiol., vol. 286, no. 6, pp. H2376–H2385, 2004. [DOI] [PubMed] [Google Scholar]
  • [12].J. S., Forrester, G. A., Diamond, and H. J. C. S., “Correlative classification of clinical and hemodynamic function after acute myocardial infarction,” Amer. J. Cardiol., vol. 39, no. 2, pp. 137–145, 1977. [DOI] [PubMed] [Google Scholar]
  • [13].Dupont M., Mullens W., Finucan M., Taylor D. O., Starling R. C., and Tang W. W., “Determinants of dynamic changes in serum creatinine in acute decompensated heart failure: The importance of blood pressure reduction during treatment,” Eur. J. Heart Failure, vol. 15, no. 4, pp. 433–440, 2013. [DOI] [PubMed] [Google Scholar]
  • [14].Kataoka Y. et al. , “Inverse ESPVR estimation with singularity avoidance via constrained edpvr parameter optimization,” in Proc. 45th Annu. Int. Conf. IEEE Eng. Med. Biol. Soc., 2023, pp. 1–6. [DOI] [PubMed] [Google Scholar]
  • [15].Uemura K. et al. , “Prediction of circulatory equilibrium in response to changes in stressed blood volume,” Amer. J. Physiol.-Heart Circulatory Physiol., vol. 289, no. 1, pp. H301–H307, 2005. [DOI] [PubMed] [Google Scholar]
  • [16].Bitterman D. S., Aerts H. J. W. L., and Mak R. H., “Approaching autonomy in medical artificial intelligence,” Lancet Digit. Health, vol. 2, no. 9, pp. e447–e449, 2020. [DOI] [PubMed] [Google Scholar]
  • [17].Mark A. L., “The Bezold-Jarisch reflex revisited: Clinical implications of inhibitory reflexes originating in the heart,” J. Amer. College Cardiol., vol. 1, no. 1, pp. 90–102, 1983. [DOI] [PubMed] [Google Scholar]
  • [18].Kataoka Y. et al. , “Hybrid hemodynamic control system: A clinically feasible design for stabilizing total perfusion in shock,” in Proc. Amer. Control Conf., 2025, pp. 2035–2040. [Google Scholar]
  • [19].Sunagawa K., Maughan W. L., Friesinger G., Guzman P., Chang M. S., and Sagawa K., “Effects of coronary arterial pressure on left venticular end-systolic pressure-volume relation of isolated canine heart,” Circulation Res., vol. 50, pp. 727–734, 1982. [DOI] [PubMed] [Google Scholar]
  • [20].Kataoka Y. et al. , “Patient-specific optimal hemodynamic control with minimal myocardial oxygen consumption,” IFAC-PapersOnLine, vol. 58, no. 18, pp. 47–52, 2024. [Google Scholar]
  • [21].Kilkenny C., Browne W. J., Cuthill I. C., Emerson M., and Altman D. G., “Improving bioscience research reporting: The ARRIVE guidelines for reporting animal research,” Animals, vol. 4, no. 1, pp. 35–44, 2014. [DOI] [PubMed] [Google Scholar]
  • [22].“Guiding principles for the care and use of animals in the field of physiological sciences,” J. Physiol. Soc. Jpn., vol. 64, no. 7/8, pp. 143–146, 2002. [PubMed] [Google Scholar]
  • [23].Selzer A., “Circulatory physiology-cardiac output and its regulation ed 2,” Arch. Intern. Med., vol. 134, no. 2, pp. 390–391, 1974. [Google Scholar]