. 2026 Sep 23;51(3):680–687. doi: 10.4103/jmp.jmp_178_26

A Lightweight Python-based Monte Carlo Framework for Visualizing Energy-dependent Dose Calibrator Response: A Nuclear Medicine Education Tool

Anil Kumar Pandey 1,✉, Shanshila Rayamajhi 1, Jagrati Chaudhary 1, Param Dev Sharma 1, Rakesh Kumar 1
PMCID: PMC13675019  PMID: 42891450

Abstract

Purpose:

Trainees often have limited understanding of radionuclide dose-calibrator physics because access to radioactive materials is restricted by radiation-safety requirements, and the complexity of established Monte Carlo (MC) platforms can present a barrier to early learning. We present a lightweight, Python-based MC simulation framework designed to visualize energy-dependent photon transport and ionization-chamber response for nuclear medicine education.

Materials and Methods:

A spherical gas-filled ionization chamber (radius 5 cm) was modelled. Photon free-path lengths were sampled from exponential attenuation distributions; deposited energy was assigned via a Beta-distributed stochastic transfer fraction (biased toward partial deposition). Ionization yield was estimated assuming 33.97 eV is required to produce one ion pair. Three clinically representative radionuclides – 99mTc (140 keV, μ = 0.030 cm−1), 131I (364 keV, μ = 0.015 cm−1), and 18F (511 keV, μ = 0.011 cm−1) – were each simulated with 100,000 independent photon histories.

Results:

Intrinsic efficiency decreased monotonically with photon energy: 13.65% (99mTc), 7.01% (131I), and 5.25% (18F). Median deposited energy increased from 37.16 keV to 95.87 keV and 132.07 keV; mean ion-pair yield rose from 1180, 3046, and 4254. Detector response increased nonlinearly from 161 to 214 and 223 ion pairs per emitted photon. All trends were consistent with exponential attenuation theory.

Conclusions:

The framework reproduces first-order ionization-chamber physics without specialist software or radioactive sources. Its transparency, rapid execution, and interactive visualizations make it well suited for classroom demonstrations and self-directed nuclear medicine physics training, while its explicit limitations underscore the gap between educational modelling and clinical-grade Monte Carlo simulation.

Keywords: 131I, 18F, 99mTc, dose calibrator, ionization chamber, Monte Carlo simulation, nuclear medicine education, photon transport

INTRODUCTION

Accurate activity measurement of radiopharmaceuticals is an essential aspect of safe nuclear medicine practice, directly influencing administered dose in both diagnostic scintigraphy and radionuclide therapy.[1,2] The ionization chamber-based radionuclide dose calibrator is the universal instrument for this purpose, yet its energy-dependent response mechanisms are poorly understood by many trainees. Dose calibrator readings depend not only on the disintegration rate but also on photon energy, source geometry, container material, sample volume, and detector attenuation characteristics.[3,4] In routine clinical practice, dose calibrators are operated using calibration factors (dial values), which often leads trainees to memorize these values without understanding the underlying physics.[5]

Monte Carlo (MC) methods are the benchmark for modelling radiation transport and detector response in medical physics. Platforms such as Geant4[5] and GATE[6] are capable of high-fidelity simulation of dose calibrator geometry, but their installation complexity, programming requirements, and computational demands render them impractical for routine undergraduate, postgraduate, or residency-level teaching. Simpler, purpose-built educational simulators have been advocated in the literature,[7] yet few specifically address ionization-chamber physics for nuclear medicine.

Simulation-based learning offers educational advantages beyond logistics: it permits unlimited repetition without radiation exposure, facilitates systematic parameter exploration, and renders otherwise invisible microscopic processes into observable, competency-aligned exercises consistent with modern medical physics education frameworks.[8] The IAEA has explicitly endorsed simulation tools for nuclear medicine training in resource-limited settings.[9]

We therefore developed a lightweight MC framework implemented in pure Python (NumPy only) that models photon emission, exponential attenuation, stochastic energy deposition, and ionization production within a simplified spherical ionization chamber. The framework requires no specialist software, runs on a standard laptop in seconds, and generates visualizations of photon transport and detector response for 99mTc, 131I, and 18F. The objectives of the study were to (i) demonstrate energy-dependent detector response in a physically coherent manner, (ii) quantify interaction probability, deposited energy, ion-pair yield, intrinsic efficiency, and energy efficiency across a clinically relevant energy range, and (iii) evaluate convergence and reproducibility of the framework as an educational proof of concept.

MATERIALS AND METHODS

Study design

This was a methodological development and proof-of-concept computational study. No human participants, patient data, biological specimens, or physical radioactive sources were used; ethical approval and informed consent were therefore not required. The simulation used a fixed pseudo-random seed to produce the same result on different platforms.

Detector geometry

The dose calibrator was represented as a homogeneous, air-filled spherical ionization chamber of radius R = 5 cm, chosen as a computationally tractable analogue to a re-entrant well-chamber geometry. The maximum chord length available for photon transport is 2R = 10 cm. Geometric path lengths were sampled from a uniform distribution U (0, 2R). Although the true chord-length distribution for a sphere is not uniform,[10] this deliberate simplification preserves the conceptual relationship between geometry and photon escape probability while minimizing implementation complexity for learners. This assumption is explicitly flagged as an educational approximation. Because the large majority of emitted photons escape without interacting [Table 1], the overall path-length distribution is dominated by this uniform geometric-escape population; on the logarithmic path-length axis used in Figures 1–3a, this produces a distribution that rises toward the chamber diameter rather than decaying, since each logarithmic bin at larger path lengths spans a proportionally larger interval of the underlying uniform distribution.

Table 1.

Photon transport and interaction statistics (n=100,000 per radionuclide)

Parameter 99mTc 131I 18F
Photon energy (keV) 140 364 511
Attenuation coefficients (cm−1) 0.030 0.015 0.011
Mean path length (cm) 4.54 4.75 4.82
Interacting photons 13,655 7013 5249
Escaping photons 86,345 92,987 94,751
Interaction probability (%) 13.65 7.01 5.25
Escape probability (%) 86.34 92.99 94.75

Linear attenuation coefficients are for air at standard temperature and pressure, taken from the NIST XCOM database[11]. “Mean Path Length” is the mean of the simulated, geometry-bounded photon path within the 10-cm-diameter chamber (min (free_path, geom_path)), not the theoretical mean free path 1/µ of the exponential attenuation law

Figure 1.

Figure 1

Monte Carlo simulation results for 99mTc (140 keV), (a) Photon path length distribution (log scale), (b) Energy deposition spectrum for all emitted photons, including noninteracting photons (zero deposited energy), (c) Energy deposition spectrum for interacting photons only

Figure 3.

Figure 3

Monte Carlo simulation results for 18F (511 keV), (a) Photon path length distribution (log scale), (b) Energy deposition spectrum for all emitted photons, including noninteracting photons (zero deposited energy), (c) Energy deposition spectrum for interacting photons only

Radionuclides and attenuation coefficients

Three radionuclides covering the clinical photon energy range were selected: 99mTc (140 keV), 131I (364 keV), and 18F (511 keV annihilation photon).[11] Linear attenuation coefficients in air at these energies were assigned as 0.030, 0.015, and 0.011 cm−1, respectively, consistent with NIST XCOM tabulated values for air at standard temperature and pressure.[12] Monoenergetic emission was assumed; radionuclide-specific branching ratios and cascade photons were excluded to focus on first-order energy dependence.

Monte Carlo photon transport

Free path lengths (x) were sampled using inverse-transform sampling of the exponential attenuation distribution:

x = -ln(r₁)/μ

where r₁ ~ U (0,1) and μ is the energy-specific linear attenuation coefficient. An interaction was recorded when x ≤ Lgeom, where Lgeom ~ U (0, 2R) is the sampled geometric path length. Photons with x > Lgeom were classified as escaping without interaction. Each radionuclide was simulated with n = 100,000 independent photon histories. The overall MC simulation workflow is summarized in Figure 4.

Figure 4.

Figure 4

Flowchart of the Monte Carlo simulation used to model dose calibrator response for 99mTc, 131I, and 18F. Photon free paths and geometric path lengths are sampled to determine interaction probability, energy deposition, and ion-pair production

Energy deposition model

For interacting photons, deposited energy (E_deposited) was calculated as:

E_deposited = E_Υ × (f),(f) ∼ Beta (α,β)

For a random variable defined on the interval [0,1], f ∼ Beta (α,β) denotes the Beta probability distribution with density f(x; α; β) = xα-1 (1–x)β-1/B(α, β), where B(α,β) is the Beta function. We used α = 2 and β = 5, giving a mean fractional energy transfer of α/(α + β) ≈0.29and a mode near 0.2. These shape parameters were chosen heuristically, rather than being derived from the Klein–Nishina differential Compton-scattering cross-section, which would require sampling scattering angles and computing the corresponding energy transfer from Compton kinematics. Instead, the Beta distribution was selected as a simple phenomenological approximation that qualitatively biases the transferred energy fraction toward the small values characteristic of the predominance of partial energy transfer in Compton interactions within low-density media, while avoiding the complexity of a full physics-based simulation.[13]

where EΥ is the incident photon energy, and f is a stochastic energy-transfer fraction drawn from a Beta distribution constrained to [0,1], chosen to bias f toward lower values, approximating the predominance of partial energy transfer in Compton-dominated low-density media.[13] This representation does not implement the Klein–Nishina differential cross-section, secondary electron transport, multiple scatter, wall effects, or electrode effects, consistent with the educational scope of the model. Noninteracting photons were assigned Edeposited = 0.

Ionization yield and detector response

Ion-pair production per photon history was estimated as: N (ion pairs) = (Edeposited)⁄Wair using the ICRU-recommended mean ionization energy for air, Wair = 34 eV per ion pair.[14] Aggregate detector response (ion pairs per emitted photon) was defined as Response = (∑Nion)⁄Nphotons.

Secondary metrics included intrinsic efficiency, εint = Ninteractions/Nphotons, i.e., the fraction of emitted photons that interacted; and energy efficiency, ϵenergy = (∑Edeposited)/(Nphotons × Eγ), i.e., the fraction of the total energy carried by all emitted photons that was actually captured by the detector, the energy actually deposited relative to the maximum possible energy (Nphotons × Eγ) that would have been captured had every photon deposited its full energy. We also characterized how far photons travel (mean and median photon path length) and what shape the deposited-energy histogram has, its spread (standard deviation) and skewness (asymmetry of the distribution).

Mean Path Length [Table 1] is not the theoretical mean free path 1/μ of the exponential attenuation law; it is the mean of the actual (geometry-limited) distance travelled by simulated photons inside the finite 10-cm-diameter chamber, i.e., the mean of min (free_path, geom_path) as defined in the algorithm [Figure 4]. The median path length, the 50th-percentile travel distance before a photon either interacts or escapes, is reported alongside the mean because the two components of this minimum behave very differently: the geometric path length is bounded (U[0, 2R]), while the free-path sample is drawn from an unbounded exponential distribution, giving the overall path-length distribution a long right tail. The median is therefore a more robust indicator of the “typical” photon travel distance than the mean, which is more strongly influenced by the small number of long, near-grazing free paths that get truncated at the chamber boundary.

Convergence and quality control

Convergence, quality-control, and physical-consistency checks were performed to verify the reliability of the MC simulation. Convergence was assessed by monitoring the cumulative detector response (ion-pair yield) as a function of the number of simulated photon histories [Figure 5]. Quality-control procedures included verification of input parameter ranges, unit-consistency audits, and repeated simulation runs to ensure reproducibility. Physical consistency was evaluated by examining the dependence of interaction probability on photon path length and confirming agreement with the exponential attenuation behavior predicted by the Beer–Lambert law [Figures 6 and 7].

Figure 5.

Figure 5

Monte Carlo convergence analysis showing the running mean ion-pair yield per photon history as a function of the number of simulated histories. The estimate stabilizes with increasing histories, indicating convergence of the simulation results. The dashed line marks the final simulation size of 100,000 histories

Figure 6.

Figure 6

Physical consistency check showing interaction probability as a function of photon path length. Interaction probability increases with path length and approaches unity, consistent with expected photon attenuation behavior in matter

Figure 7.

Figure 7

Observed interaction fraction as a function of photon path-length bin. The fraction of photons undergoing interaction increases with path length, reflecting the greater probability of attenuation

Statistical uncertainty was quantified using the standard error of the mean, SEM = SD/√N, where is the sample standard deviation of the per-photon ion-pair yield and is the number of simulated histories; 95% confidence intervals were calculated as CI₉₅ = mean ± 1.96 × SEM, using the normal approximation justified by the large sample size (n = 100,000), consistent with established MC uncertainty-reporting practices.[15]

RESULTS

Photon transport and interaction statistics

Path-length distributions for all three radionuclides, plotted on a logarithmic path-length axis [Figures 1–3, panel a], showed a pronounced rise toward the chamber diameter (10 cm) rather than a short-path-dominated profile. This reflects the fact that the large majority of emitted photons [86%–95%, depending on radionuclide; Table 1] do not interact within the chamber and instead traverse their full geometric path to the chamber wall, which is uniformly distributed over [0, 2R]; on a logarithmic axis, this uniform population necessarily produces more counts per bin toward the upper end of the range, since each logarithmic bin spans a progressively larger interval of linear path length. Only the small interacting fraction is governed by the exponential free-path law and contributes counts at shorter path lengths. Mean path length increased slightly with energy, consistent with the lower attenuation coefficients (and correspondingly lower interaction probabilities) at higher photon energies, from 4.54 cm (99mTc) to 4.75 cm (131I) to 4.83 cm (18F) [Table 1].

The reported mean path length is the average distance actually travelled by photons inside the finite 10 cm-diameter chamber. For each photon, this distance is the smaller of the sampled free path and the available geometric path within the chamber (i.e., min [free path, geometric path]). It should not be confused with the theoretical mean free path (1/μ), which is 33.3 cm for 99mTc, 66.7 cm for 131I, and 90.9 cm for 18F – all much larger than the chamber diameter. Since most photons leave the chamber without interacting, they travel only up to the chamber boundary. Consequently, their path lengths are determined by the chamber geometry and have an average value of 5 cm. The simulated mean path lengths are slightly <5 cm, particularly for lower-energy photons, because these photons interact more frequently and therefore stop before reaching the chamber boundary.

Interaction probability fell from 13.65% for 99mTc to 7.01% for 131I and 5.25% for 18F (escape probabilities 86.34%, 92.99%, and 94.75%, respectively), consistent with Beer–Lambert predictions for these attenuation coefficients across a 10-cm chamber diameter [Table 1].

Energy deposition and ionization yield

When noninteracting photons were included, the deposited-energy distribution was strongly zero-inflated [Figures 1b and 2b, and 3b], with 86%–95% of emitted photons depositing no energy. Consequently, the median deposited energy calculated across all emitted photons was 0 keV for all three radionuclides [Table 2]. This metric differs from the deposited-energy distributions of interacting photons only [Figures 1c and 2c, and 3c], which, by construction, exclude zero-deposition events. As shown in Figures 1c and 2c, and 3c, the deposited-energy histograms for interacting photons were unimodal and markedly right-skewed for all three radionuclides.

Figure 2.

Figure 2

Monte Carlo simulation results for 131I (364 keV), (a) Photon path length distribution (log scale), (b) Energy deposition spectrum for all emitted photons, including noninteracting photons (zero deposited energy), (c) Energy deposition spectrum for interacting photons only

Table 2.

Energy deposition and ion-pair production

Parameter 99mTc 131I 18F
Median deposited energy, interacting photons only (keV) 37.16 95.87 132.07
Median deposited energy, all photons (keV) 0 0 0
SD deposited energy, interacting only (eV) 22,390.88 57,798.38 81,777.78
Skewness, deposited energy (interacting only) 0.59 0.58 0.64
Mean ion pairs per interacting photon 1180 3046 4254

SD: Standard deviation

Median deposited energy among interacting photons increased with photon energy: 37.16 keV (99mTc), 95.87 keV (131I), and 132.07 keV (18F) [Table 2]. Consistent with this trend, the mean ion-pair yield per interacting photon increased from 1180 to 3046 and 4254 ion pairs, illustrating that, despite fewer interactions, higher-energy photons produce substantially greater ionization per event.

Detector response and efficiency

Aggregate detector response increased nonlinearly with photon energy: 161 ion pairs per photon (99mTc), 214 (131I), and 223 (18F) [Table 3].

Table 3.

Detector response and efficiency metrics

Parameter 99mTc 131I 18F
Detector response (ion pairs/emitted photon) 161 214 223
SD (ion pairs/photon, all photons) 472.69 898.67 1097.11
SEM (response) 1.49 2.84 3.46
95% CI lower (response) 158.22 208.03 216.49
95% CI upper (response) 164.08 219.17 230.09
Intrinsic efficiency, εint (%) 13.65 7.01 5.25
Energy efficiency, ϵenergy (%) 3.91 1.99 1.48

SD: Standard deviation, CI: Confidence interval, SEM: Standard error of mean

Note that “Detector Response” [Table 3] and “Mean Ion Pairs per Photon” [Table 2] are different normalizations: the former is the aggregate ion-pair yield per emitted photon (interacting and escaping photons combined), while the latter is the mean yield per interacting photon only; approximately, Response ≈ εint × (mean ion pairs per interacting photon).

The steeper rise between 140 keV and 364 keV compared with the plateau toward 511 keV reflects saturation of the energy-deposition gain against continued decline in interaction probability. Intrinsic efficiency fell monotonically from 13.65% to 5.25%, and energy efficiency from 3.91% to 1.48%, confirming that increasing photon penetration progressively reduces both interaction likelihood and fractional energy capture [Table 3].

Convergence

The running mean ion-pair yield stabilized with increasing particle histories and showed minimal variation beyond approximately 100,000 histories [Figure 5], indicating adequate numerical convergence.

Figure 5 provides a direct visual check that the chosen sample size is sufficient for the reported summary statistics to be stable estimates rather than artifacts of an under-sampled simulation, which is the standard justification for the sample size used throughout this study.

Interaction probability increased monotonically with path length [Figure 6], consistent with Beer–Lambert attenuation theory. Deposited-energy spectra exhibited the expected positively skewed distribution with a predominance of noninteraction events [Figure 7]. Summary quality-control metrics are presented in Table 4, and it is derived from the same simulation run as Table 1.

Table 4.

Per radionuclide Monte Carlo interaction summary (quality control)

Radionuclide 99mTc 131I 18F
Simulated histories 100,000 100,000 100,000
Interaction events 13,655 7013 5249
Noninteraction events 86,345 92,987 94,751
Interaction fraction (%) 13.66 7.01 5.25
Mean ion pairs/history 161.2 213.6 223.3
SD ion pairs/history 472.69 898.67 1097.1

SD: Standard deviation

DISCUSSION

This study demonstrates that a compact Python script, requiring only NumPy, can faithfully replicate the first-order energy-dependent behavior of an ionization chamber radionuclide dose calibrator. The simulated trends, inverse relationship between photon energy and intrinsic efficiency, nonlinear increase in aggregate detector response, and the geometry-dominated path-length distributions, are consistent with well-established photon transport theory[13,16] and with the expected behavior of a simplified, nonscattering absorber geometry. The consistency lends physical credibility to the framework and supports its suitability as an educational reference.

The nonlinear dependence of detector response on photon energy [Table 3] reflects a fundamental trade-off in gas ionization detectors: higher-energy photons penetrate further and escape more readily (reducing εint from 13.65% to 5.25%), but those that do interact deposit more energy and generate proportionally more ion pairs (mean Nion rising from 1180 to 4254). The net effect – increasing response despite decreasing efficiency – is precisely the conceptual point that dial-value-based teaching obscures. Commercial dose calibrators handle this through nuclide-specific calibration settings that empirically account for these competing effects;[3] the simulation makes the underlying physics explicit.

The path-length distributions [Figures 1–3, panel a] are dominated by the majority population of noninteracting, escaping photons (86%–95% across the three radionuclides; Table 1), whose geometric path lengths were sampled from U (0, 2R) as described in Methods. On the logarithmic path-length axis, this uniform population produces a distribution that rises toward the chamber diameter rather than decaying – a direct visual consequence of the deliberate geometric simplification adopted for this educational model, rather than of exponential (Beer–Lambert) attenuation per se. The exponential free-path law governs only the smaller interacting subset of photons and is not separately visualized in these panels; its effect is instead captured quantitatively through the interaction probabilities and mean ion-pair yields reported in Tables 1-3. The increasing mean overall path length from ⁹⁹ᵐTc (4.54 cm) to ¹⁸F (4.83 cm) reflects the correspondingly lower attenuation coefficients – and hence lower interaction probabilities – at higher photon energies, providing a quantitative bridge between abstract cross-section tables and observable detector behavior. We note that this geometry-dominated path-length behavior is a feature of the simplified U (0, 2R) chord-length approximation adopted here (Methods), and caution against over-interpreting it as directly analogous to the chord-length distributions reported in full-scale Geant4 models of re-entrant ionization chambers, which use the true (nonuniform) spherical or cylindrical chord-length geometry.[5]

The strongly zero-inflated deposited-energy distribution (median = 0 keV across all emitted photons, for all three radionuclides) illustrates that most emitted photons traverse the chamber without interaction. This counterintuitive finding often surprises trainees and motivates discussion of detector design optimizations such as pressurized chambers, high-Z fill gases, and re-entrant geometry, topics well covered in standard references. The Beta-distributed partial-energy-transfer model, while simpler than a full Klein–Nishina treatment, adequately captures the asymmetric nature of Compton-dominated partial deposition in low-density media and provides a foundation for discussing Compton kinematics in follow-up teaching.[17]

From a pedagogical standpoint, the framework addresses recognized deficiencies in nuclear medicine physics education. Simulation-based learning has been shown to improve conceptual retention and clinical reasoning in medical physics trainees.[18] The IAEA Technical Report Series and AAPM TG reports on dose calibrator quality assurance[2,9] provide the regulatory context within which this conceptual understanding becomes clinically actionable; the present tool bridges the gap between those guidelines and trainees’ experiential understanding.

Several limitations warrant explicit acknowledgement. First, the uniform geometric path-length approximation diverges from the true, nonuniform chord-length distribution of a sphere,[10] systematically overestimating the proportion of long-path photons. Second, the single-interaction approximation and the absence of Klein–Nishina angular kinematics, secondary electron transport, bremsstrahlung, and wall/electrode effects preclude quantitative agreement with physical dose calibrators.

Critically, the framework has not yet been validated against an independent MC code or physical measurements; the internal consistency checks reported here [Figures 6 and 7 and Table 1] confirm that the free-path sampling engine reproduces Beer–Lambert attenuation correctly, but this is not equivalent to external validation, and quantitative agreement with physical dose calibrators should not be assumed.

Third, monoenergetic emission does not capture complex spectra such as the 131I cascade or the variable positron endpoint energy relevant to 18F. Finally, air was used as the detector medium; commercial instruments employ pressurized argon or nitrogen, which alters the effective attenuation coefficients and W-value.[8,14] These deviations are intentional design choices to maximize educational transparency; instructors should contextualize them explicitly to avoid misconceptions.

These limitations, most importantly the absence of independent validation against an established MC code such as Geant4/GATE or against physical measurements, define the scope for subsequent, more advanced simulation work and are the focus of ongoing development.

CONCLUSIONS

A lightweight MC framework implemented in pure Python reproduces the energy-dependent photon transport and ionization response of a radionuclide dose calibrator for 99mTc, 131I, and 18F in a physically coherent and educationally transparent manner. The framework is executable on any standard computer without specialist software, generates interpretable visualizations of otherwise imperceptible detector physics, and exposes the competing mechanisms – photon penetration versus energy deposition – that govern ionization-chamber response and underpin clinical calibration settings. With explicit acknowledgement of its simplifying assumptions and its current lack of independent validation, the tool is well suited for integration into nuclear medicine physics curricula, supporting the development of analytical reasoning and quality-assurance competency in the next generation of nuclear medicine practitioners.

Conflicts of interest

There are no conflicts of interest.

Funding Statement

Nil.

REFERENCES

  • 1.Early PJ, Sodee DB. Principles and Practice of Nuclear Medicine. 2nd ed. St. Louis: Mosby; 1995. [Google Scholar]
  • 2.International Atomic Energy Agency. Radiation Protection in Nuclear Medicine. Safety Reports Series No. 40. Vienna: International Atomic Energy Agency; 2004. [Google Scholar]
  • 3.Zanzonico P. Routine quality control of clinical nuclear medicine instrumentation: A brief review. J Nucl Med. 2008;49:1114–31. doi: 10.2967/jnumed.107.050203. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Andreo P. Monte Carlo techniques in medical radiation physics. Phys Med Biol. 1991;36:861–920. doi: 10.1088/0031-9155/36/7/001. [DOI] [PubMed] [Google Scholar]
  • 5.Agostinelli S, Allison J, Amako K, Apostolakis J, Araújo H, Arce P, et al. GEANT4—a simulation toolkit. Nucl Instrum Methods Phys Res A. 2003;506:250–303. doi: 10.1016/S0168-9002(03)01368-8. [Google Scholar]
  • 6.Jan S, Santin G, Strul D, Staelens S, Assié K, Autret D, et al. GATE: A simulation toolkit for PET and SPECT. Phys Med Biol. 2004;49:4543–61. doi: 10.1088/0031-9155/49/19/007. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Ljungberg M, Strand SE, King MA, editors. Monte Carlo Calculations in Nuclear Medicine: Applications in Diagnostic Imaging. 2nd ed. Boca Raton: CRC Press; 2012. [Google Scholar]
  • 8.Cherry SR, Sorenson JA, Phelps ME. Physics in Nuclear Medicine. 4th ed. Philadelphia: Saunders/Elsevier; 2012. [Google Scholar]
  • 9.International Atomic Energy Agency. Nuclear Medicine Physics: A Handbook for Teachers and Students. Vienna: International Atomic Energy Agency; 2014. [Google Scholar]
  • 10.Coleman R. Random paths through convex bodies. J Appl Probab. 1969;6:430–41. [Google Scholar]
  • 11.Kondev FG, Wang M, Huang WJ, Naimi S, Audi G. The NUBASE2020 evaluation of nuclear physics properties. Chin Phys C. 2021;45:030001. [Google Scholar]
  • 12.Hubbell JH, Seltzer SM. Tables of X-Ray Mass Attenuation Coefficients and Mass Energy-Absorption Coefficients. Version 1.4. Gaithersburg, MD: NIST; 2004. [Google Scholar]
  • 13.Attix FH. Introduction to Radiological Physics and Radiation Dosimetry. New York: Wiley-Interscience; 1986. [Google Scholar]
  • 14.International Commission on Radiation Units and Measurements. Average Energy Required to Produce an Ion Pair. ICRU Report 31. Bethesda, MD: ICRU; 1979. [Google Scholar]
  • 15.Lux I, Koblinger L. Monte Carlo Particle Transport Methods: Neutron and Photon Calculations. Boca Raton: CRC Press; 1991. [Google Scholar]
  • 16.Knoll GF. Radiation Detection and Measurement. 4th ed. Hoboken: Wiley; 2010. [Google Scholar]
  • 17.Biggs S, Jennings M, Swerdloff S, Chlap P, Lane D, Rembish J, et al. PyMedPhys: A community effort to develop an open-source medical physics library. J Open Source Softw. 2022;7:4555. [Google Scholar]
  • 18.McGaghie WC, Issenberg SB, Cohen ER, Barsuk JH, Wayne DB. Does simulation-based medical education with deliberate practice yield better results than traditional clinical education? A meta-analytic comparative review of the evidence. Acad Med. 2011;86:706–11. doi: 10.1097/ACM.0b013e318217e119. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Journal of Medical Physics are provided here courtesy of Wolters Kluwer -- Medknow Publications