Comparison and reproducibility of transthoracic bioimpedance and dual beam Doppler ultrasound measurement of cardiac function in healthy volunteers.

H W Ng 1, T Walley 1, Y Tsao 1, A M Breckenridge 1
PMCID: PMC1368518  PMID: 1777364

Abstract

1. We compared the ease of use and reproducibility of two noninvasive methods, transthoracic electrical bioimpedance (TEB) (BoMed NCCOM3-R7) and non-imaging dual beam Doppler ultrasound (Quantascope--Vital Science), in cardiac output (CO) and stroke volume (SV) measurement in healthy volunteers at rest and during physiological stress, both short term and from day to day. 2. The TEB method was easier to use and not dependent on the operator. The TEB method was more reproducible both in the short term and from day to day. Both devices were able to detect CO and SV changes under physiological stress, but were less reproducible day to day during exercise. 3. At supine rest, the within subject coefficient of variation between time point for TEB was 4.6% for CO and 6.1% for SV, and 7.9% for CO and 7.4% for SV with Doppler. 4. The results from each device showed a linear correlation coefficient (r) of 0.69 for CO (P less than 0.0005) and 0.64 for SV (P less than 0.0005). The correlation coefficient improved to 0.76 (P less than 0.0005) when only changes in CO and SV were considered. There was no systematic difference in the changes detected by the two methods, but the individual variation was wide.

Full text

PDF
275

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Appel P. L., Kram H. B., Mackabee J., Fleming A. W., Shoemaker W. C. Comparison of measurements of cardiac output by bioimpedance and thermodilution in severely ill surgical patients. Crit Care Med. 1986 Nov;14(11):933–935. doi: 10.1097/00003246-198611000-00004. [DOI] [PubMed] [Google Scholar]
  2. Bernstein D. P. A new stroke volume equation for thoracic electrical bioimpedance: theory and rationale. Crit Care Med. 1986 Oct;14(10):904–909. doi: 10.1097/00003246-198610000-00017. [DOI] [PubMed] [Google Scholar]
  3. Bernstein D. P. Continuous noninvasive real-time monitoring of stroke volume and cardiac output by thoracic electrical bioimpedance. Crit Care Med. 1986 Oct;14(10):898–901. doi: 10.1097/00003246-198610000-00015. [DOI] [PubMed] [Google Scholar]
  4. Bland J. M., Altman D. G. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986 Feb 8;1(8476):307–310. [PubMed] [Google Scholar]
  5. Hassan S., Turner P. Systolic time intervals: a review of the method in the non-invasive investigation of cardiac function in health, disease and clinical pharmacology. Postgrad Med J. 1983 Jul;59(693):423–434. doi: 10.1136/pgmj.59.693.423. [DOI] [PMC free article] [PubMed] [Google Scholar]
  6. Hottinger C. F., Meindl J. D. Blood flow measurement using the attenuation-compensated volume flowmeter. Ultrason Imaging. 1979 Jan;1(1):1–15. doi: 10.1177/016173467900100101. [DOI] [PubMed] [Google Scholar]
  7. Mattar J. A., Baruzzi A. C., Diament D., Szynkier R. T., de Felippe J., Jr, da Luz P. L., Auler J. O., Jr, Lage S., Pileggi F., Jatene A. A clinical comparison between cardiac output measured by thermodilution versus noninvasive thoracic electrical bioimpedance. Acute Care. 1986;12(1):58–60. [PubMed] [Google Scholar]
  8. Northridge D. B., Findlay I. N., Wilson J., Henderson E., Dargie H. J. Non-invasive determination of cardiac output by Doppler echocardiography and electrical bioimpedance. Br Heart J. 1990 Feb;63(2):93–97. doi: 10.1136/hrt.63.2.93. [DOI] [PMC free article] [PubMed] [Google Scholar]
  9. Shoemaker W. C., Appel P. L., Kram H. B., Nathan R. C., Thompson J. L. Multicomponent noninvasive physiologic monitoring of circulatory function. Crit Care Med. 1988 May;16(5):482–490. doi: 10.1097/00003246-198805000-00004. [DOI] [PubMed] [Google Scholar]
  10. Silke B., Verma S. P., Taylor S. H. Pharmacodynamic monitoring during acute intervention in ischaemic heart disease using a new echo-Doppler device. Br J Clin Pharmacol. 1990 Jun;29(6):741–747. doi: 10.1111/j.1365-2125.1990.tb03696.x. [DOI] [PMC free article] [PubMed] [Google Scholar]
  11. Taylor S. H., Silke B. Is the measurement of cardiac output useful in clinical practice? Br J Anaesth. 1988;60(8 Suppl 1):90S–98S. doi: 10.1093/bja/60.suppl_1.90s. [DOI] [PubMed] [Google Scholar]
  12. Wallenstein S., Zucker C. L., Fleiss J. L. Some statistical methods useful in circulation research. Circ Res. 1980 Jul;47(1):1–9. doi: 10.1161/01.res.47.1.1. [DOI] [PubMed] [Google Scholar]

Articles from British Journal of Clinical Pharmacology are provided here courtesy of British Pharmacological Society

RESOURCES