ABSTRACT
Background:
Conventional gravimetric uroflowmetry, although reliable, is restricted to clinical settings and may be influenced by inhibition and logistical constraints. The present study device is a gravity-driven volumetric uroflowmetry device that uses a calibrated multichamber collection system to derive standard uroflowmetry parameters without requiring electronics or external power. This study was done to assess the agreement between a novel portable uroflowmetry device (Pflow™) and conventional gravimetric uroflowmetry.
Materials and Methods:
This prospective two-center, comparative study was conducted over 12 months (July 2023–July 2024) across two tertiary urology centers and included 372 patients (292 males; mean age 50.3 ± 18.5 years). Measurements were recorded simultaneously using both the devices during a single voiding event. Parameters analyzed included voided volume, peak flow rate (Qmax), average flow rate (Qavg), and voiding time. Agreement was assessed using the Wilcoxon signed-rank test, Spearman correlation, intraclass correlation coefficient (ICC), and Bland–Altman analysis.
Results:
No statistically significant differences were observed between the two methods for voided volume (271 mL vs. 270 mL, P = 0.440), voiding time (40 s vs. 38 s, P = 0.204), Qmax (15 mL/s vs. 14.4 mL/s, P = 0.209), and Qavg (9 mL/s vs. 10 mL/s, P = 0.739) and strong correlations were observed (ρ = 0.799–0.963, P < 0.001). ICC demonstrated excellent agreement for voided volume (0.963) and good agreement for all measured flow parameters. Bland–Altman analysis showed minimal bias.
Conclusion:
Pflow™ demonstrates good agreement with conventional uroflowmetry and may serve as a simple adjunct for monitoring urinary flow in clinical settings.
INTRODUCTION
Uroflowmetry is a widely used noninvasive diagnostic tool for evaluating lower urinary tract dysfunction.[1,2,3,4] It provides objective assessment of voiding patterns and plays an important role in the management of lower urinary tract symptoms (LUTS).[1,2] Conventional gravimetric uroflowmetry, although reliable, is limited to clinical settings and may be affected by patient’s inhibitions and logistical constraints.
Home-based uroflowmetry offers a potential solution[5,6,7,8,9,10,11] by enabling assessment in a natural environment and facilitating repeated measurements. Several portable uroflowmetry systems have previously been described, including funnel-based devices, electronic home uroflowmeters, mobile acoustic systems, wearable ultrasound Doppler systems, and tele-home monitoring platforms. However, many of these systems rely on electronic sensors, mobile applications, wireless connectivity, batteries, or external power sources. The present study device is a portable gravity-driven volumetric uroflowmetry device that operates without electricity, electronic sensors, batteries, software, wireless connectivity, or internet access and uses a calibrated horizontally arranged multichamber fluid collection system with controlled outflow to reconstruct the uroflowmetry curve and derive standard uroflowmetry parameters.
This simplified design has the potential to measure all the important parameters of uroflowmetry, reduce the cost and improve accessibility, particularly in resource-limited settings. Therefore, this study evaluated the agreement between the study device and conventional gravimetric uroflowmetry using a simultaneous measurement design to minimize intraindividual variability.
MATERIALS AND METHODS
Ethical considerations
The study was approved by the Muljibhai Patel Society for Research in Nephro-Urology Ethics Committee (MPSRNUEC), MPUH, Nadiad, Gujarat, India (MPUH) (Approval No. EC/977/2023; approval granted on 8 September 2023). The study was conducted in accordance with the institutional ethical standards at both the participating institutes, namely MPUH, India, and Institute of Urology, Dhule. Written informed consent was obtained from all the participants before enrollment.
Study design and population
This prospective two center study was conducted over 12 months (July 2023–July 2024) across two tertiary urology centers. A total of 372 patients (292 males, 80 females; mean age 50.3 ± 18.5 years, included 300 from MPUH and 72 from IOU), were included in the study.
Patients were recruited from both the centers using uniform inclusion and exclusion criteria. A standardized study protocol was followed across both the institutions to ensure consistency in data collection, device usage, and measurement procedures.
Inclusion and exclusion criteria
Adult patients attending the participating urology outpatient clinics and undergoing uroflowmetry as a part of routine clinical evaluation, who were able to undergo a uroflowmetry test, were eligible for inclusion. Patients were required to provide informed consent and achieve a voided volume between 150 mL and 750 mL during the study measurement. The lower threshold of 150 mL was selected in accordance with the accepted uroflowmetry standards to ensure reliable interpretation of the flow parameters, whereas the upper limit of 750 mL corresponded to the maximum capacity of the study device.
Patients who were unable to complete the uroflowmetry, had incomplete recordings, declined participation, or produced voided volumes exceeding 750 mL were excluded from the analysis. Only one voiding episode per participant was included in the study.
Device description and working principle
The study device is a gravity-driven volumetric uroflowmetry device [Figure 1]. It is manufactured from medical-grade plastic (polypropylene) and measures approximately 20 cm × 20 cm × 20 cm. The device is intended as a low-cost disposable or limited-reuse uroflowmetry system. The estimated manufacturing cost is approximately Indian Rupees (INR) 240 per unit, substantially lower than that of conventional electronic uroflowmetry systems.
Figure 1.

Schematic representation of the study device demonstrating the multi-chamber collection system. Urine enters through the inlet and sequentially fills chambers via an overflow pathway. Each chamber has a calibrated outlet aperture (~1 mL/s), enabling gravity-driven controlled flow. The distribution of urine across chambers reflects temporal variation in urinary flow, allowing reconstruction of uroflowmetry parameters
The device consists of two horizontal rows of chambers. Urine is collected through a funnel and enters the upper row, where it passes sequentially through a series of cubical chambers arranged in an overflow pathway. Each chamber is fitted with a calibrated outlet aperture at its base, allowing a near-constant outflow of approximately 1 mL/s into a corresponding lower-row collection chamber. Because each chamber has a known volume, the distribution of urine across the lower collection chambers reflects temporal variations in the urinary flow during the voiding. The controlled outflow through the calibrated apertures enables proportional volume accumulation within the collection chambers over time. This distribution pattern remains consistent across a wide range of inflow rates encountered during a routine uroflowmetry, thereby permitting reconstruction of the flow parameters. Chamber volume data are subsequently processed using a predefined computational algorithm to derive standard uroflowmetry parameters, including voided volume, voiding time, peak flow rate (Qmax), and average flow rate (Qavg). The same algorithm was applied uniformly to all the measurements obtained during the study.
During device development, fluid flow characteristics were initially modelled using computer-aided design-based fluid dynamic simulations to determine appropriate chamber geometry and outlet aperture dimensions. Following manufacture, calibration testing was performed by passing known fluid volumes through individual chambers and measuring the volume discharged through the outlet over time. Aperture dimensions were adjusted iteratively until the desired outflow characteristics were achieved. This calibration process ensured consistent flow behavior across the operating range of the device.
Procedure
The study device was placed directly over the weight transducer of a conventional uroflowmeter, allowing simultaneous measurement during a single voiding event under identical physiological conditions [Figure 2]. The device urine collection pathway was arranged to permit unobstructed urine passage without active flow restriction, splashing, or external overflow before entering the study device and the conventional uroflowmeter simultaneously. The conventional uroflowmeter was zeroed before each measurement. Each participant served as his or her own control, as measurements from both the conventional uroflowmeter and the study device were obtained simultaneously during the same voiding episode. The conventional uroflowmeter generated and printed the reference uroflowmetry report, whereas chamber volume measurements obtained from the study device were photographed. The resulting images were analyzed using dedicated software to quantify the volume in each chamber, and these data were subsequently processed using a predefined computational algorithm to derive the corresponding uroflowmetry parameters for comparison.
Figure 2.

Experimental setup showing simultaneous uroflowmetry using the study device and a conventional gravimetric uroflowmeter. The study device is positioned directly over the weight transducer to ensure identical measurement conditions. The conventional uroflowmetry console generates the reference uroflowmetry recordings, while chamber-volume measurements obtained from the study device are subsequently processed using the predefined computational algorithm to derive uroflowmetry parameters
This paired study design eliminated intraindividual variability and enabled direct comparison between the two measurement methods under identical physiological conditions.
Statistical analysis
Data were analyzed using IBM SPSS Statistics, Version 31.0 (IBM Corp., Armonk, NY, USA). Wilcoxon test, Spearman correlation, intraclass correlation coefficient (ICC), and Bland–Altman analysis were used. P < 0.05 was considered significant. The authors confirm the availability of, and the access to, all the original data reported in this study. The authors have not used artificial intelligence (AI) and AI-assisted technologies, including large language learning models, in the preparation of the manuscript. The authors confirm that they have reviewed the manuscript and take full responsibility for the contents in the article.
RESULTS
A total of 372 patients were analyzed (mean age 50.34 ± 18.52 years). Baseline demographic characteristics of the study population are summarized in Table 1. Most participants were male (78.2%), and the largest age subgroup was between 61 and 70 years.
Table 1.
Baseline demographic characteristics of the study population
| Characteristic | Value, n (%) |
|---|---|
| Total patients, n | 372 |
| Age (years), mean±SD | 50.34±18.52 |
| Sex | |
| Male | 292 (78.2) |
| Female | 80 (21.8) |
| Age 18–20 years | 21 (5.6) |
| Age 21–30 years | 47 (12.6) |
| Age 31–40 years | 57 (15.3) |
| Age 41–50 years | 49 (13.2) |
| Age 51–60 years | 66 (17.7) |
| Age 61–70 years | 75 (20.2) |
| Age 71–80 years | 47 (12.6) |
| Age >80 years | 10 (2.7) |
SD=Standard deviation
No significant differences were observed between the two methods [Table 2]:
Table 2.
Comparison of uroflowmetry variables between methods
| Variable | Conventional (median, IQR) | Pflow™ (median, IQR) | Z statistic | P | Significance |
|---|---|---|---|---|---|
| Voided volume (mL) | 271 (190–392) | 270 (195–400) | 9999.5 | 0.440 | NS |
| Void time (s) | 40 (20–63) | 38 (23–63) | 4850.0 | 0.204 | NS |
| Qmax (mL/s) | 15 (9–17) | 14.4 (9.6–17) | 12124.5 | 0.209 | NS |
| Qavg (mL/s) | 9 (6–11.7) | 10 (6–11) | 6889.0 | 0.739 | NS |
*Wilcoxon signed-rank test, *Statistically NS (P>0.05). Q1=Lower quartile and Q3=Upper quartile. Qmax=Peak flow rate, Qavg=Average flow rate, IQR=Inter quartile range, NS=Not significant
Voided Volume: 271 versus 270 mL (P = 0.440)
Voiding Time: 40 versus 38 s (P = 0.204)
Qmax: 15 versus 14.4 mL/s (P = 0.209)
Qavg: 9 versus 10 mL/s (P = 0.739).
Strong correlations were observed (ρ = 0.799–0.963, P < 0.001) between conventional uroflowmetry and the study device.
The Spearman rho correlation highlighted a strong positive relationship between all the variables: Table 3.
Table 3.
Spearman correlation analysis
| Variable | ρ (rho) | P | Interpretation |
|---|---|---|---|
| Voided volume | 0.963 | <0.001 | Very strong correlation |
| Void time | 0.870 | <0.001 | Strong correlation |
| Qmax | 0.799 | <0.001 | Strong correlation |
| Qavg | 0.813 | <0.001 | Strong correlation |
Qmax=Peak flow rate, Qavg=Average flow rate
ICC 2,1 showed excellent agreement for volume (0.963) and good agreement for flow parameters: Table 4.
Table 4.
Intraclass correlation coefficients
| Variable | ICC value | Interpretation |
|---|---|---|
| Voided volume | 0.963 | Excellent agreement |
| Void time | 0.841 | Good agreement |
| Qmax | 0.793 | Good agreement |
| Qavg | 0.745 | Moderate–good agreement |
Qmax=Peak flow rate, Qavg=Average flow rate, ICC=Intraclass correlation coefficient
Bland–Altman analysis demonstrated minimal bias and 95% limits of agreement for all the parameters [Figure 3 and Table 5].
Figure 3.

(a) Bland Altman: Voided Volume, (b) Bland Altman: Void Time, (c) Bland Altman: Qmax, (d) Bland Altman: Qavg
Table 5.
Bland–Altman analysis of agreement between conventional and Pflow™ devices
| Variable | Mean difference (bias) | SD of differences | 95% limits of agreement |
|---|---|---|---|
| Voided volume | −1.9 (mL) | 32.7 | −65.9 to+62.1 (mL) |
| Void time | +0.5 (s) | 10.4 | −19.9 to+21.0 (s) |
| Qmax | +0.6 (mL/s) | 3.2 | −5.8 to+6.9 (mL/s) |
| Qavg | +0.04 (mL/s) | 2.8 | −5.5 to+5.6 (mL/s) |
Qmax=Peak flow rate, Qavg=Average flow rate, SD=Standard deviation
DISCUSSION
This two-center study demonstrates that the study device provided measurements comparable to the conventional uroflowmetry across key parameters. The simultaneous measurement design strengthens validity by eliminating intravoid variability.
The two-center design enhances the generalizability of the findings across different clinical settings and reduces centre-specific bias.
Various portable uroflowmetry systems have been described in the literature.[6,7,8,9,10,11,12,13,14,15,16,17,18,19] Early nonelectronic systems included funnel-based peak flow devices[8,14] described by Smith and Currie, as well as categorized flow measurement systems developed by Pel and Van Mastrigt. Subsequent devices such as Peakometer and CaptiFlow™[9,10] attempted simplified home-based flow estimation. More recent electronic and digital systems include Da Capo™, mobile acoustic uroflowmetry applications, wearable ultrasound Doppler systems, and tele-home uroflowmetry platforms.[11,12,13,14,15,16,17,18,19] Several investigators have reported strong agreement between portable uroflowmetry systems and conventional gravimetric uroflowmetry.[6,15,16,17,18,19] De La Rosette et al. demonstrated improved reliability using portable home uroflowmetry systems,[6] while Lee et al.[15] and Lee et al.[16] showed that mobile acoustic uroflowmetry provided voided volume and flow rate measurements comparable to the standard uroflowmetry. Kato et al. similarly reported no major differences between clinic-based and home based uroflowmetry measurements using the portable “Freeflow” device.[17] The findings of the present study are consistent with these reports and further extend the concept by validating a gravity-driven volumetric system. The principal novelty of the present device lies in its completely passive design. Unlike most contemporary portable uroflowmetry systems that depend on sensors, acoustic recording, ultrasound technology, batteries, mobile applications, or wireless data transmission, the study device uses a calibrated multichamber fluid collection system to generate multiple uroflowmetry parameters without requiring on-site batteries, electronics, internet connectivity, or dedicated hardware. This simplified architecture may offer advantages in terms of affordability, portability, disposability, ease of use, and applicability in low-resource settings while maintaining good agreement with the conventional gravimetric uroflowmetry.
Conventional weight-based (gravimetric) uroflowmetry remains the established reference standard for uroflow assessment and offers automated flow-curve generation with high measurement accuracy. However, these systems generally require dedicated equipment, electrical power, maintenance, trained personnel, and attendance at a healthcare facility. In contrast, the present portable volumetric device uses controlled fluid outflow and chamber-based volume accumulation. While weight-based systems may provide greater automation and data integration, volumetric systems may offer potential advantages in portability, affordability, disposability, ease of transport, and applicability in resource-limited settings. The two approaches should therefore be viewed as complementary rather than competing technologies, with portable volumetric devices potentially expanding access to uroflowmetry outside conventional hospital environments.
The device may have potential future applications in tele-urology and remote patient monitoring wherein the report would be generated by the central processing unit after the patient has shared the pictures of the study unit after use. However, these applications were not evaluated in the present study, which was conducted entirely under supervised clinical conditions. Future studies are required to assess the device’s performance, usability, and patient compliance in real-world home-monitoring environments. If validated, such applications could facilitate longitudinal assessment of LUTS, postoperative follow-up, and treatment response monitoring while reducing the need for repeated hospital visits, and potentially supporting use in primary-care settings, rural healthcare facilities, and regions with limited access to conventional uroflowmetry equipment.
The present study additionally demonstrated strong positive correlation across all uroflowmetry variables, with Spearman correlation coefficients ranging from 0.799 to 0.963. Intraclass correlation analysis revealed excellent agreement for voided volume and good agreement for flow parameters.
Bland–Altman analysis showed minimal systematic bias and a narrow 95% limits of agreement, indicating strong concordance between the two methods. Qmax demonstrated slightly greater variability than voided volume, as reflected by the wider limits of agreement observed on Bland–Altman analysis. Despite this variability, the observed mean bias for Qmax was small (+0.6 mL/s), and agreement remained comparable to that reported for other portable uroflowmetry devices. Therefore, while individual Qmax values should always be interpreted within the broader clinical context, the study device demonstrated acceptable agreement with the conventional uroflowmetry for the assessment of peak flow. The simultaneous measurement design also minimized potential variability arising from separate voiding events and allowed direct comparison under identical physiological conditions. Although participants represented a broad age spectrum and both sexes, subgroup-specific agreement analyses were beyond the scope of the present study and warrant future investigation.
Limitations
The study was limited to two centers with supervised measurements, and the device was not evaluated in unsupervised home-monitoring conditions. Therefore, conclusions regarding home use and patient-operated deployment should be considered preliminary. Bench repeatability of the device was assessed during development; however, repeatability in patients was not specifically evaluated in the present study, as uroflowmetry parameters are known to exhibit physiological variability even within the same individual across repeated voids. Long-term performance was also not evaluated because the study device is intended for disposable or limited-reuse applications rather than prolonged repeated use. The population was predominantly male, and a formal subgroup analyses according to age and gender were not performed. Future studies should evaluate agreement across specific demographic groups to further establish generalizability.
CONCLUSION
The study device demonstrates good agreement with conventional uroflowmetry under supervised clinical conditions and may serve as a low-cost alternative method for uroflow assessment in settings where conventional electronic uroflowmetry is unavailable. Further large-scale multicenter studies, including evaluation in home-monitoring environments, are required before broader implementation can be recommended.
Conflicts of interest
There are no conflicts of interest.
Funding Statement
Nil.
REFERENCES
- 1.Coyne KS, Sexton CC, Thompson CL, Milsom I, Irwin D, Kopp ZS, et al. The prevalence of lower urinary tract symptoms (LUTS) in the USA, the UK and Sweden: Results from the epidemiology of LUTS (EpiLUTS) study. BJU Int. 2009;104:352–60. doi: 10.1111/j.1464-410X.2009.08427.x. [DOI] [PubMed] [Google Scholar]
- 2.Parsons JK. Benign prostatic hyperplasia and male lower urinary tract symptoms: Epidemiology and risk factors. Curr Bladder Dysfunct Rep. 2010;5:212–8. doi: 10.1007/s11884-010-0067-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Drake MJ, Doumouchtsis SK, Hashim H, Gammie A. Fundamentals of urodynamic practice, based on international continence society good urodynamic practices recommendations. Neurourol Urodyn. 2018;37:S50–60. doi: 10.1002/nau.23773. [DOI] [PubMed] [Google Scholar]
- 4.Gratzke C, Bachmann A, Descazeaud A, Drake MJ, Madersbacher S, Mamoulakis C, et al. EAU guidelines on the assessment of non-neurogenic male lower urinary tract symptoms including benign prostatic obstruction. Eur Urol. 2015;67:1099–109. doi: 10.1016/j.eururo.2014.12.038. [DOI] [PubMed] [Google Scholar]
- 5.Sonke GS, Kiemeney LA, Verbeek AL, Debruyne FM, de la Rosette JJ. A method for estimating within-patient variability in maximal urinary flow rate adjusted for voided volume. Urology. 2002;59:368–72. doi: 10.1016/s0090-4295(01)01547-3. [DOI] [PubMed] [Google Scholar]
- 6.De La Rosette JJ, Witjes WP, Debruyne FM, Kersten PL, Wijkstra H. Improved reliability of uroflowmetry investigations: Results of a portable home-based uroflowmetry study. Br J Urol. 1996;78:385–90. doi: 10.1046/j.1464-410x.1996.00115.x. [DOI] [PubMed] [Google Scholar]
- 7.Bray A, Griffiths C, Drinnan M, Pickard R. Methods and value of home uroflowmetry in the assessment of men with lower urinary tract symptoms: A literature review. Neurourol Urodyn. 2012;31:7–12. doi: 10.1002/nau.21197. [DOI] [PubMed] [Google Scholar]
- 8.Hansen MV, Zdanowski A. The use of a simple home flow test as a quality indicator for male patients treated for lower urinary tract symptoms suggestive of bladder outlet obstruction. Eur Urol. 1997;32:34–8. [PubMed] [Google Scholar]
- 9.Schwartz BF, Soderdahl DW, Thrasher JB. Home flow rates in evaluation of lower urinary tract symptoms in men. Tech Urol. 1998;4:15–7. [PubMed] [Google Scholar]
- 10.Pridgeon S, Harding C, Newton D, Pickard R. Clinical evaluation of a simple uroflowmeter for categorization of maximum urinary flow rate. Indian J Urol. 2007;23:114–8. doi: 10.4103/0970-1591.32058. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Jørgensen JB, Jacobsen HL, Bagi P, Hvarnes H, Colstrup H. Home uroflowmetry by means of the Da Capo home uroflowmeter. Eur Urol. 1998;33:64–8. doi: 10.1159/000019534. [DOI] [PubMed] [Google Scholar]
- 12.Chun K, Kim SJ, Cho ST. Noninvasive medical tools for evaluating voiding pattern in real life. Int Neurourol J. 2017;21:S10–6. doi: 10.5213/inj.1734860.430. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Rattanasomrerk S, Honglertsakul C, Gesprasert G, Chantakanakakorn N, Naiyanetr P. Development of Tele-Home Uroflowmetry System with NB-IoT Implementation. Proc Biomed Eng Int Conf. 2019:p. 1–5. [Google Scholar]
- 14.Devrim F, Dincel N. Comparison of conventional uroflowmetry and new smart, self-directed outpatient uroflowmetry in the evaluation of lower urinary system dysfunction in children. Izmir Behcet Uz Cocuk Hast Derg. 2020;10:170–6. [Google Scholar]
- 15.Lee DG, Gerber J, Bhatia V, Janzen N, Austin PF, Koh CJ, et al. A prospective comparative study of mobile acoustic uroflowmetry and conventional uroflowmetry. Int Neurourol J. 2021;25:355–63. doi: 10.5213/inj.2142154.077. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Lee YJ, Kim MM, Song SH, Lee S. A novel mobile acoustic uroflowmetry: Comparison with contemporary uroflowmetry. Int Neurourol J. 2021;25:150–6. doi: 10.5213/inj.2040250.125. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Kato S, Watanabe H, Yamasue K. Freeflow: The novel portable uroflowmeter can help to realize practical urinary conditions at home. Low Urin Tract Symptoms. 2022;14:208–13. doi: 10.1111/luts.12426. [DOI] [PubMed] [Google Scholar]
- 18.Pandolfo SD, Crauso F, Aveta A, Cilio S, Barone B, Napolitano L, et al. A novel low-cost uroflowmetry for patient telemonitoring. Int J Environ Res Public Health. 2023;20:3287. doi: 10.3390/ijerph20043287. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Rangganata E, Rahardjo HE, Rasyid N, Widia F, Murwantara IM, Yugopuspito P, et al. Comparative study of mobile acoustic uroflowmetry and conventional uroflowmetry: A systematic review and meta analysis. Neurourol Urodyn. 2024;43:694–702. doi: 10.1002/nau.25405. [DOI] [PubMed] [Google Scholar]