Machine learning quantitative OCT derived fibrosis indicators for corneal haze evaluation

Corentin Soubeiran 1,2, Maëlle Vilbert 1,2,3, Benjamin Memmi 3, Cristina Georgeon 3, Vincent Borderie 3, Anatole Chessel 1,#, Karsten Plamann 1,2,*,#
Editor: Helen Howard4
PMCID: PMC13690194  PMID: 42911951

Abstract

Photorefractive Keratectomy (PRK) is a widely used laser-assisted refractive surgical technique. In some cases, it leads to temporary subepithelial fibrosis linked to visual haze, but objective and quantitative tools to monitor these symptoms are lacking. We here present a machine learning-based method for the classification of corneal images from spectral domain optical coherence tomography images recorded in vivo on standard clinical devices and use it to extract quantitative fibrosis indicators. Because of the transient nature and relative rarity of significant post-PRK haze, we trained the model on corneas presenting Fuchs’ Endothelial Corneal Dystrophy, causing similar, but permanent, fibrotic changes, and applied it to images from patients who have undergone PRK surgery. Our study shows that the model output (probability of pathological classification) combined with OCT-derived quantitative fibrosis descriptors provides an interpretable and explainable indicator of corneal healing for post-operative follow-up.

Introduction

Photorefractive Keratectomy (PRK) is a widely performed laser-assisted surgical procedure for correcting refractive errors of the eye. While generally safe and effective, a notable complication is the development of corneal haze, a condition characterized by a loss of corneal transparency due to subepithelial fibrosis [1,2]. This haze can significantly impair visual quality, causing symptoms such as blurred vision, glare, and halos, which limit the success of the post-operative recovery [3]. The clinical management of corneal haze relies on careful monitoring of its progression and resolution over time. However, current evaluation methods are predominantly subjective, creating significant challenges for consistent and reliable patient care.

The standard clinical assessment of corneal haze is performed using slit-lamp biomicroscopy, where an ophthalmologist visually grades the severity of the haze based on rating scales [1]. This method is inherently subjective and suffers from high inter-observer and even intra-observer variability [4]. An ophthalmologist’s assessment can be influenced by factors such as their experience, the lighting conditions, and even the sequence of patients seen. To address this limitation, some quantitative techniques have been explored. Scheimpflug imaging, for instance, can provide a measure of corneal densitometry, which correlates with corneal scattering [2,4]. Some work aims to generate quantitative assement using slit-lamp photographs [5,6]. However, this technique has lower resolution than Optical Coherence Tomography (OCT) and does not provide physical parameters directly related to the tissue’s properties. Consequently, there is a pressing clinical need for an objective, quantitative, and reliable tool for the evaluation of corneal haze. Such a tool would enable clinicians to unequivocally attribute a patient’s visual complaints to haze, monitor its evolution objectively, and standardize treatment protocols.

Spectral Domain Optical Coherence Tomography (SD-OCT) is a non-invasive imaging modality that provides high-resolution, cross-sectional images of the cornea. It has become an indispensable tool in ophthalmology for diagnosing and managing a wide range of corneal pathologies. In the context of corneal haze, OCT images reveal a characteristic hyper-reflective band in the anterior stroma, just beneath the epithelium, which corresponds to the subepithelial fibrosis [7,8]. This morphological feature is a key indicator for clinicians when assessing haze. Interestingly, similar structural changes are observed in other corneal diseases, such as Fuchs’ Endothelial Corneal Dystrophy (FECD). In advanced stages, FECD can also lead to subepithelial fibrosis and stromal opacification, presenting with similar symptoms of blurred vision and glare [9]. The commonality of these structural alterations, particularly the anterior stromal fibrosis, suggests that a quantitative approach developed for one condition might be applicable to the other [8,10,11]. In our previous work, we developed algorithms for analyzing corneal transparency from SD-OCT images based on signal attenuation in the corneal stroma [12,13]. However, these methods did not specifically exploit the information contained within the subepithelial region, which is crucial for assessing both FECD-related fibrosis and post-PRK haze.

This study aims to address the lack of objective haze quantification by developing a machine learning-based method for the automatic classification of clinical OCT images using explanable quantifiers allowing the clinician to understand the decision process. Our primary objective is to classify corneas into two categories: “healthy” and “pathological” (i.e., showing signs of fibrosis). To achieve this, we leverage the morphological features of subepithelial fibrosis visible in OCT images. Given the transient nature and relative rarity of significant post-PRK haze, we adopted a strategy of training our model on a more readily available dataset of corneas with FECD, which presents with similar, but permanent, fibrotic changes. The trained model is then applied to images from post-PRK patients to evaluate corneal haze.

We hypothesize that the output of our machine learning model: the probability of an image being classified as pathological, can serve as a quantitative, interpretable, and explainable indicator of corneal fibrosis. This provides clinicians with a much-needed objective tool for monitoring corneal healing and managing patient care following PRK surgery. In this paper, we describe our method, from the extraction of quantitative fibrosis descriptors from OCT images to the training and validation of a classification model. We demonstrate how this approach can provide a reliable and measurable tool for the diagnosis and extendable follow-up of corneal haze in its transient stage.

Materials and methods

We adopted an automatic classification approach based on a random forest classifier applied to physics-based descriptors of fibrosis extracted from OCT images, as described in the following section. This method was selected over deep-learning approaches, such as convolutional neural networks operating directly on raw images, for two reasons. The random forest classifier achieved the best classification performance among the methods tested. It has also been reported to be competitive with deep learning methods on tabular data [14]. Additionaly and unlike deep-learning, this machine learning approach relies on interpretable physical parameters rather than learned image features as it can be done in deep-learning, enabling clinicians to understand the rationale behind each classification. Such interpretability is essential for clinical adoption, as practitioners are more likely to integrate quantitative descriptors into their practice when the classification process is transparent rather than perceived as a “black box.”

Morphological parameters of subepithelial fibrosis

Compared to a healthy cornea (Fig 1a), subepithelial fibrosis as observed in corneal haze [8,11] (Fig 1c) and some FECD cases [10] (Fig 1b) are characterized by three distinctive morphological features visible on OCT B-scans:

Fig 1. Comparison of SD-OCT images.

Fig 1

(a) a normal cornea, (b) a cornea presenting Fuchs endothelial corneal dystrophy; the zone of subepithelial fibrosis is indicated by the white arrow, (c) corneal haze, the red dotted rectangle delimitates specular reflection an artefact of SD-OCT signal not to consider, the arrow indicates subepithelial fibrosis. Images have been recorded using the clinical SD-OCT RTVue-XR Avanti.

  1. Sub-epithelial hyperreflectivity. Mature myofibroblasts and the disorganised deposition of type III collagen fibrils increase light back-scattering, producing a “hyperreflective” band in the Sub-epithelial / anterior stroma zone that is markedly brighter than the surrounding tissue [10,11] (Fig 1b–c).

  2. Blurred epithelium/stroma interface. The intense scattering within the fibrotic zone reduces the contrast at the epithelial/stromal boundary, making it difficult to delineate on the OCT cross-section.

  3. Altered reflectivity depth profile, stromal transparency. In a healthy cornea the reflectivity drops sharply beyond the epithelium, whereas in a fibrotic cornea the signal remains elevated well into the anterior stroma before decaying gradually.

These features have a notable practical consequence. Commercial OCT devices include proprietary layer-segmentation algorithms, for instance the corneal pachymetry mapping modules of instruments such as the Optovue RTVue or Zeiss Cirrus platforms, that automatically delineate corneal layers to produce thickness maps. Because these algorithms were designed and validated on healthy corneas, they misinterpret the hyperreflective fibrotic zone as part of the epithelium: the boundary detector locks onto the deeper edge of the fibrotic band rather than the true epithelial/stromal interface. This results in a systematic overestimation of epithelial thickness in the pachymetry maps, erroneously suggesting epithelial hyperplasia when the actual cause is stromal fibrosis [9]. Although this segmentation artefact provides an indirect clinical clue to the presence of corneal haze, it also illustrates the lack of dedicated quantitative tools: existing device algorithms detect an anomaly but misattribute it, offering clinicians no direct, reliable measurement of the fibrosis itself. Building on the three morphological features described above, we designed objective, physics-based parameters that each capture one aspect of the fibrotic signature in the OCT reflectivity profile across corneal depth (the extraction procedure is detailed in the following subsections). These quantifiers serve as input features for the automatic classification model. A central design criterion was to ensure that each parameter remains physically interpretable by clinicians, so that the classification output can be traced back to observable morphological changes in the cornea.

Standardisation of clinical images

The SD-OCT images are pre-processed following the pipeline described in full detail in [13] and available in GitHub repository [15], supplemented here with an explicit alignment step (final code available in GitHub repository [16]). For completeness and reproducibility, the eight sequential steps of the pipeline are described below. The processing pipeline allow to “flatten” the OCT images and correct artifacts, enabling the extraction of standardized depth profiles and morphological quantifiers as illustrated in Fig 2. All images have the same axial pixel size of 4.3μm, lateral pixel size varies between 5.85μm to 8.51μm depending on device recording mode (precision on device mode in “Databases and experimental design”). Images are considered on respective individual coordinate systems.

Fig 2. Extraction of morphological quantifiers associated with the epithelium/stroma interface performed on a healthy cornea OCT signal.

Fig 2

(a): OCT of a normal cornea acquired with clinical SD-OCT RTVue-XR. The red arrow indicate posterior stromal artefact in the central region. (b): The standardized flattened OCT image, limited to the central part of 6 mm (for an overall field-of-view of 8 mm during acquisition). The red arrow indicate the same posterior region that underwent PCA correction (c): 1.5 mm window and associated laterally averaged depth intensity profile. (d) Zoom on the two anterior intensity peaks and definition of several morphological quantifiers for the peak associated with the interface between Bowman’s layer and the corneal stroma: Depth, Peak intensity, Peak Width, and peak Area Ratio. (e) Additional quantifiers to this peak resulting from profile estimation: Sigma, MSE and COV. (f) The last two quantifiers: Alpha and Beta corresponding to stromal signal decay.

Step A: Exposure adjustment (optional). For over-exposed images, a histogram-based correction is applied: when the mean histogram value exceeds a threshold derived from typical clinical images, a constant offset is subtracted from all pixels to restore standard black levels in air and aqueous humour. This linear adjustment has no effect on the extracted quantifiers and is applied for visualisation purposes only, prior to any artefact detection in the case the graphical user interface is used for validation. The quantifier extraction pipeline itself operates on the raw image data, ensuring that all measurements are based on unaltered pixel values.

Step B: Central saturation artefact detection and removal. The column-averaged OCT intensity profile is computed and smoothed using a second-order Savitzky–Golay filter; its first derivative is then calculated. An abrupt lateral variation in the smoothed derivative identifies the specular reflection artefact at the corneal apex (illustrated in Fig 1c). Artefact boundaries are set at the lateral coordinates of the derivative extrema ±55μm, and the corresponding image columns are cropped. This cropping is necessary because the saturation artefact introduces a non-tissue-related saturated signal that would corrupt the depth-intensity profile. A graphical user interface prompts the user to validate or manually correct the automatic detection; in practice, for the large majority of good-quality clinical images, the automatic result is accepted without modification.

Step C: Anterior surface detection and A-scan alignment. To identify the anterior corneal surface from raw image (Fig 2a) the image is pre-filtered with a 2D-Gaussian filter (σ=2 pixels) to attenuate noise and high-spatial-frequency contributions from corneal scatterers. Local maxima corresponding to the air/tear film interface are detected column by column using a robust, SNR-dependent thresholding procedure described in [13]. If detection fails at image edges due to low SNR, the anterior surface axial position is linearly interpolated from adjacent columns. The resulting axial-coordinate vector is smoothed with a median filter (window: 15 pixels). Each A-scan is then individually aligned along the depth axis so that its detected air/tear film interface coincides with a common reference depth. As this alignment is hard to manually readjust when it fails due to low SNR or pathological morphology, we implemented an automatic quality control step to what is described in [13] using normalised cross-correlation. This second alignment is performed by normalised cross-correlation [17]. In this method each A-scan is compared to the central A-scan serving as reference: the cross-correlation between the two signals exhibits a peak corresponding to their relative axial offset. Parabolic interpolation around this peak is then applied to achieve sub-pixel estimation of the shift, which is subsequently used to align the A-scans, ensuring sub-pixel co-registration along their first interface. The alignment is displayed in the graphical user interface for validation, in practice all images of our databases were correctly aligned without need for manual correction after this second step leading to flat OCT images as illustrated in Fig 2b.

Step D: Corneal flattening and stromal region-of-interest segmentation. All subsequent depth-intensity profiles are therefore extracted along the local axial direction of the flattened image along a vertical axis. The stromal region of interest is defined over a 6 mm wide apex-centred lateral area (reduced by the width of any artefact cropping performed in Step B). This width encompasses the optically functional central cornea and coincides with the highest-SNR region of corneal SD-OCT images [13]. This 6 mm corresponds to maximum physiological pupil dilation [18], encompasses the optically functional cornea (the central 3.2 mm being sufficient for reading vision at ±10° visual field [19]), and coincides with the highest-SNR region in corneal SD-OCT images [20].

Steps E–F: Posterior stromal artefact correction by PCA. The posterior stromal artefact manifests as a structured hyperreflective intensity increase within the deep central stroma, of instrumental rather than tissue origin, confirmed by its absence in SD-OCT images of a morphologically realistic corneal phantom (Rowe Technical Design Inc.; see [13]). As shown in Fig 2a (red arrow), this artefact is spatially confined to the central region of the image and exhibits a characteristic lateral signature that can be isolated from the true tissue signal. Principal Component Analysis (PCA) is applied to correct this artefact on the flatenned cornea.

  • Step E: Axially, the selected region to apply PCA is limited to the stromal region, two depth (axial) margins are automatically excluded from the stromal region: +50μm beyond the epithelium/Bowman’s layer interface (high keratocyte density; non-exponential attenuation) and −30μm before the endothelium (endothelial–Descemet complex), as these zones exhibit attenuation behaviour inconsistent with the homogeneity assumption of the fitting model. These margins were set from known corneal histology and adjusted empirically to maximise measurement repeatability on normal corneas. A graphical user interface allows manual override of the segmentation when required.

  • Step F: The input is a matrix of size 20×1018, where each row contains the axially-averaged OCT intensity of one of 20 equal-thickness in-depth stromal sub-layers across the 1018 lateral pixel positions; the matrix is mean-centered but not scaled. A first PCA on the full matrix identifies two dominant components(PC1 and PC2). Across a normative dataset, PC1 and PC2 accounted for approximately 94% and 4% of the variance, respectively. They respectively capture the overall depth-dependent attenuation and the curvature-driven lateral intensity decrease for PC1, and the structured lateral signature of the posterior artefact for PC2 as reported in [13]. The lateral boundaries of the artefact are derived from local minima of the PC2 reconstruction around the central image region. A second PCA is then applied to the input matrix restricted to this artefact zone; its PC1 quantifies the artefact amplitude, from which a two-dimensional correction mask with negative values is constructed and subtracted from the stromal region. Any residual intensity distortion introduced by this numerical correction is reflected in the 95% confidence interval of the output quantifiers [13].

Once applied the flattened and artefact-corrected image is illustrated in Fig 2b.

Step G: Lateral normalisation. The corrected image is normalised by the smoothed anterior surface intensity signal (smoothing the detected curve using second-order Savitzky-Golay filter, sliding window: 115 pixels) to compensate for the peripheral intensity loss associated with corneal curvature geometry.

Step H: Depth-intensity profile extraction and quantifier averaging. For each lateral position of a sliding window of width 1.5 mm across the 6 mm region of interest, the OCT intensity values are averaged laterally within the window to yield a one-dimensional depth-intensity profile. This lateral averaging reduces noise while preserving the morphological information along the surface normal. Sliding window method corresponds to incrementally (pixel-wise) slide a working window, it can mathematically be seen as a convolution by a mean operator with lateral dimension 1.5 mm. The window width of 1.5 mm was selected as a compromise between noise reduction and spatial resolution. This parameter was determined a priori and applied uniformly to all samples. In healthy corneas, where steps A–-G were performed fully automatically without manual correction, the resulting depth–intensity profiles were observed to be consistent across the 6 mm region of interest for this window width.

The nine morphological quantifiers defined in the following subsection are computed for each window position. The final value of each quantifier Q is obtained as a weighted mean over all window positions:

Q^=∑kwkQk∑kwk,wk=1−12|xk|xmax, (1)

where xk is the lateral distance of window k from the corneal apex, and xmax=3 mm is the half-width of the region of interest. The weight wk therefore decreases linearly from 1 at the apex to 0.5 at the periphery. This weighting reduces the influence of peripheral windows, where curvature-induced SNR degradation and residual artefacts from the flattening procedure are most pronounced, while still retaining their contribution so as not to discard clinically relevant information.

The lateral averaged intensity profile of the central region is displayed in a final graphical user interface for validation, with the display of two landmarks: the first peak corresponding to the air/tear film interface and the second peak corresponding to the epithelial basement membrane/Bowman’s layer interface. In practice it would be laborious to evaluate the landmark position across the entire 6 mm region of interest, we capitalized on the use off the weighted mean of quantifiers and the extraction of 9 different quantifiers to mitigate the impact of potential landmark detection failures in some windows.

Definition of subepithelial fibrosis quantifiers

When examining corneal OCT images for signs of fibrosis, ophthalmologists focus on three key visual features in the anterior cornea. First, they search for hyperreflectivity in the subepithelial zone as a bright band immediately beneath the epithelium that signals the presence of activated myofibroblasts and disorganised collagen deposition [1,2]. Second, they evaluate the overall stromal transparency by assessing how dark (i.e., transparent) the stromal tissue appears on the OCT B-scan; in pathological cases, increased light scattering elevates the signal throughout the anterior stroma, reducing this darkness. Third, they examine the regularity of the anterior corneal surface, as irregular curvature may indicate prior PRK surgery and potential scarring. To objectively quantify these clinical observations, we designed physics-based descriptors that capture each aspect: peak intensity and width metrics characterize the subepithelial hyperreflective zone, exponential decay parameters quantify stromal transparency, and peak shape regularity reflects tissue homogeneity. The following quantifiers translate these visual assessments into measurable, reproducible parameters.

The morphological characterisation of subepithelial fibrosis is based on the identification of peaks of interest in the depth-averaged intensity profile of the corneal OCT signal: among others, the first peak corresponds to the air/tear film interface and the second to the epithelial basement membrane/Bowman’s layer interface (Fig 2a–c). This second peak constitutes our signal of interest because the excimer laser acts from this depth onwards for the corneal stroma remodelling, at about 100 μ m depth in the cornea. Therefore, at about this depth, the scarring reaction may lead to inflammation or even fibrosis, as shown in OCT by the deepening of the peak in question (Fig 1c).

The intensity peak associated with the epithelial basement membrane is identified via the two local minima around the peak (Fig 2c,d). It is defined by its location in depth and delimited by the distance between the axial position of the peak and the location of the two local minima (Fig 2d). A Gaussian regression is applied to the intensity profile corresponding to the peak. The Gaussian regression provides four quantifiers (Fig 2d,e):

  • Depth: below the air/lacrimal film interface (μm) to the estimated mean of the Gaussian regression. This parameter localises the primary scattering zone within the corneal stroma.

  • Sigma (standard deviation of the peak): quantifies the spread of the peak in depth (μm). In the case of fibrosis, it therefore characterizes its depth extent, reflecting how far the fibrotic reaction has propagated into the stroma from the epithelial basement membrane.

  • Peak width (distance between the two local minima): it also characterizes the spread of the peak in depth but by looking at local minima independently of peak shape, providing a complementary metric for fibrosis thickness or shape irregularities that is less sensitive to peak asymmetry.

  • Peak intensity (maximum intensity of the second peak compared to the maximum intensity of the first peak): it characterizes the relative intensity of the backscatter at the epithelium/stroma interface compared to the reflection at the air/lacrimal film interface. This ratio directly quantifies the hyperreflectivity observed clinically in fibrotic corneas and correlates with the density of light-scattering structures, primarily activated myofibroblasts and abnormal collagen within the anterior stroma [2].

Two parameters quantify the quality of the Gaussian regression (Fig 2e):

  • MSE (mean square error): characterizes the deviation between the regression and the actual values of the profile.

MSE=S−gσ^2―=1N∑i=1N(Si−gσ^(zi))2 (2)

with Si the actual intensity value of the profile at depth zi, gσ^(zi) the corresponding value of the Gaussian regression at this depth, and N the number of depth points in the profile. Note that •― denotes the mean over all depth points.

  • COV (covariance between the profile and the regression): quantifies the joint difference between the expectations of the profile and the Gaussian.

COV=cov(S,gσ^)=1N∑i=1N(Si−S―)(gσ^(zi)−gσ^―) (3)

where S― and gσ^― are the mean values of the profile and the Gaussian regression, respectively.

These last two parameters were defined following the observation of the images, and in particular, the observation that the Gaussian is sometimes poorly resolved for pathological corneas. High values of squared error or poor covariance can be a warning of pathology, as irregular peak morphology, especially asymmetry, reflects heterogeneous tissue organisation that deviates from the relatively uniform structure of healthy corneal stroma.

An exponential regression is then performed between the first peak and the endothelium to characterize the attenuation of the OCT intensity as it propagates through the stroma (Fig 2f). In healthy transparent cornea, light attenuation follows an exponential decay law due to scattering processes within the stromal tissue [12]. Deviations from this behaviour signal altered stromal transparency. Two additional quantifiers correspond to the fit variables of the regression function:

f(z)=αexp(−βz)+C (4)

with z the depth in the stroma, α the initial amplitude of the signal at the anterior stroma, β the decay rate of the signal with depth, and C a constant representing background noise. The two quantifiers are:

  • Alpha: the amplitude α, representing the initial scattering intensity within the anterior stroma.

  • Beta: the rate of decay β, which quantifies how rapidly light is attenuated through the stromal depth. A lower decay rate (smaller β) indicates reduced transparency, as more light is scattered and less penetrates deeply into the tissue, a hallmark of corneal opacification [12].

Finally, a last quantifier characterizes the optical contrast:

  • Area ratio (R) the ratio between the peak (epithelium/stroma interface or fibrosis) area (Fig 2d blue) and the epithelium/stroma intensity decay area (Fig 2d orange, named signal area for convinience), the bottom of the peak is generated by linear fitting between the two local minima on either side of the peak. This quantifier captures the overall contribution of the hyperreflective zone to the signal, reflecting both its intensity and extent, and thus serves as a metric of the fibrotic signature in the OCT profile.

Once these quantifiers are extracted for each image, they are used as input features for the classification model described in the next section. The interpretability of these quantifiers ensures that the model’s classification decisions can be traced back to specific morphological characteristics of the corneal tissue, facilitating clinical understanding and potential integration into practice.

The separation of quantifier distributions for a group of healthy corneas (166 OCT images) and a group of corneas with FECD (313 OCT images; inclusion criterion is selection for endothelial grafting, no stage of pathology specified) is statistically significant for seven of these quantifiers: Peak Width, Sigma, Depth, MSE, Area Ratio, Alpha and Peak intensity (S1 Fig.).

Processing time: The complete pipeline is implemented in Python available at [16]. Two steps involve optional graphical user interface validation: central saturation artefact detection (Step B) and stromal layer segmentation (Step D). In both cases, the automatic result is accepted without modification for the vast majority of good-quality images, and only brief training is required for non-expert users to operate the pipeline autonomously. The total processing time per image including preprocessing and quantifier extraction is on the order of 2–3 min with the graphical user interface interaction, and less than 30 seconds when the automatic results are accepted without modification on a standard workstation (Intel Core i7-8665U, 1.9 GHz, 16 GB DDR4 RAM).

Robustness to low-SNR conditions and landmark detection failures: The reliability of quantifier extraction depends on the correct identification of profile landmarks (first anterior peak, second peak, and its bounding local minima; see Fig 2). Images are acquired by professionals trained in OCT acquisition and meet established clinical quality standards. In addition to user-driven quality control via the graphical user interface for validating artefact removal and landmark detection, the pipeline integrates complementary mechanisms to preserve robustness when landmark identification is compromised by high noise levels or pathological morphologies:

  • The lateral averaging over the 1.5 mm window in Step H intrinsically increases the profile SNR before landmark detection, attenuating single-column noise spikes that could trigger false minima or maxima.

  • In the specific case of anterior surface detection (Step C), edge failures due to low SNR are handled by linear interpolation from reliably detected neighbouring columns.

  • The Gaussian regression quality metrics (MSE and COV; see following subsection) provide per-image indicators of landmark detection confidence, and unusually high values flag potentially unreliable estimates.

Databases and experimental design

Our primary objective is to develop an automatic classification method for clinical OCT images that can objectively identify corneal fibrosis, with the ultimate goal of quantifying post-PRK corneal haze. However, due to the transient nature of haze and the difficulty of clinical diagnosis in mild cases, insufficient validated clinical images exist to constitute a robust training database. We therefore adopted a transfer learning strategy: we trained and validated our classification model on FECD as in its advanced stages, presents similar subepithelial fibrosis localized in the anterior stroma. We subsequently applied the trained model to post-PRK patients to quantify corneal haze. This approach is justified by the morphological similarity between these conditions and the fact that FECD leads to permanent symptoms, enabling extended data collection with unambiguous clinical diagnosis.

All databases were collected between January 1, 2019, and January 1, 2023, and exported for research purposes in September 2023 (with patient or their legal guardian’s informed consent; see consent and ethical considerations section) from RTVue-XR Avanti clinical SD-OCT devices. The complete study design comprised three distinct cohorts, each serving a specific purpose (Table 1). Instrumentation and image acquisition protocols is identical with [13]: all images are exported as 8-bit grayscale B-scans with identical axial sampling (after manufacturer dewarping to physical space). The four acquisition modes reduce to two resolution groups:

Table 1. Databases used for the automatic classification exported from the RTVue-XR Avanti clinical SD-OCT. The pathologies of interest (corneal haze and Fuchs’ Dystrophy) have in common the development of subepithelial fibrosis in their advanced stages.

Database Label Patients Eyes Images Modalities Used for
Healthy Corneas Healthy 43 83 166 Line/CrossLine (83); Pachy/PachyWide (83) Train/Test
Fuchs Endothelial Corneal Dystrophy (FECD) Fuchs 47 48 313 Line/CrossLine (155); Pachy/PachyWide (158) Train/Test
Repeatability Healthy 5 10 (5x rep.) 50 Line (50) Test
FECD 3 5 (4–6 rep.) 26 Line/CrossLine (26)
Visual Haze Haze – 6 6 Line/CrossLine (3); Pachy/PachyWide (3) Test
Post-Surgery Follow-up Haze 5 9 37 (3–5 per eye) Line (37) Test
  • Line / CrossLine: identical sampling and geometry; 8 mm lateral FOV over 2.2 mm scan depth (4.3 μm/pixel axially; laterally 7.84 μm/pixel and 8.51 μm/pixel for Line and CrossLine, respectively), generated by the device from 30 averaged B-scans.

  • Pachymetry / PachymetryWide: identical sampling between them; radial acquisitions with 6 mm and 9 mm lateral FOV, respectively, over 1.8 mm scan depth, with few averaged B-scans (exact number unknown) (4.3 μm/pixel axially; laterally 5.88 μm/pixel and 5.85 μm/pixel for Pachymetry and PachymetryWide, respectively).

Training and validation cohort (FECD model development): This cohort consisted of 166 OCT images from healthy corneas and 313 images from patients with FECD (inclusion criterion: selection for endothelial grafting; no specific disease stage requirement). The nine morphological quantifiers described above were computed for each image. The total 479 images set has been splitted into 250 FECD, 133 healthy images for trainning and 63 FECD, 33 healthy for testing (80% train, 20% test). On the training set, we performed 5-fold cross-validation to train and select classifiers on FECD/healthy classification. Model performance metrics (accuracy, sensitivity, specificity) reported in the Results section derive from this cross-validation procedure. The test set was reserved for evaluating the final model performance on unseen data. We ensured that images from the same patient were not split between training and test sets to prevent data leakage and ensure an unbiased assessment of model generalization.

Repeatability cohort: To assess the reproducibility of our quantifiers and classification outputs across repeated acquisitions, we established an independent repeatability cohort comprising 10 healthy corneas and 6 corneas with FECD. These subjects underwent three consecutive OCT imaging sessions. Crucially, this cohort was entirely separate from the 479 images used for model training and validation, ensuring an unbiased assessment of measurement consistency.

Application cohort (haze quantification): Once the model was trained and validated on FECD data, we applied it (without any retraining or parameter adjustment) to OCT images from patients who had undergone PRK surgery and reporting visual haze symptoms. The model output for these images represents the probability of classification as pathological (i.e., presenting fibrosis similar to FECD).

Extension cohort (post-PRK follow-up): To further evaluate the clinical utility of our quantification method, we conducted a longitudinal follow-up of post-PRK patients. This extension cohort included 9 patients eyes with serial OCT imaging at multiple time points post-surgery. This to study the transient evolution of post surgical inflammation and fibrosis, and to evaluate how our quantifiers track the progression and resolution of corneal haze and fibrosis indicators over time, providing insights into the healing process and the potential for our quantifiers to monitor disease progression and resolution in a clinical setting.

Interpretability

As our classification method relies on physics-based parameters derived from the corneal OCT intensity signal rather than on abstract learned representations, feature importance analysis provides direct insight into the decision-making process of the model. To achieve this, we employed SHapley Additive exPlanations (SHAP) [21], a widely used framework for interpreting machine-learning predictions. SHAP is model agnostic, meaning it can be applied to any machine learning model regardless of its internal structure, including ensemble methods, neural networks, and linear models. SHAP quantifies the contribution of each input feature to a model’s output by computing Shapley values, a concept from cooperative game theory. These values represent the marginal contribution of a feature to the prediction when considered in all possible subsets of features. For a given feature i, its Shapley value ϕi is computed as (adapted from [21]):

ϕi=∑S⊆F⧵{i}|S|!(|N|−|S|−1)!|N|![f(S∪{i})−f(S)] (5)

where F is the set of all features, S is a subset of features excluding i, and f(S) is the model’s output based on S. In practice, SHAP approximates f(S) and f(S∪{i}) without retraining the model for every subset. Instead, it marginalizes the model’s predictions over the distribution of missing features. For tree-based models, the TreeExplainer variant computes these values efficiently by leveraging the tree structure to conditionally traverse decision paths and split probabilities. This approach ensures a fair allocation of the predicted outcome among all features, satisfying properties of consistency and additivity. SHAP provides both global interpretability by aggregating feature contributions across the dataset and local interpretability by explaining individual predictions. These capabilities allowed us to identify the most influential features driving the model’s decisions while ensuring transparency and alignment with the underlying physical phenomena.

Results

Training and validation (FECD model development)

We adopted a supervised two-class framework. The combined dataset of 166 healthy and 313 FECD images was split into a training set (80%, preserving class proportions) and a held-out test set (20%), ensuring that images from the same patient did not appear in both sets [22,23]. The nine morphological quantifiers described above serve as input features; the model output is the predicted probability of FECD classification, pFECD, which quantifies the degree of confidence in the classification.

Model selection was performed by 5-fold cross-validation on the training set: at each iteration, the model was trained on four folds and evaluated on the remaining fold. This procedure provides a robust estimate of generalisation performance while mitigating overfitting to any single data partition. We evaluated: linear SVM, radial-basis-function SVM, logistic regression, decision tree, random forest, and multilayer perceptron. The random forest achieved the highest cross-validation accuracy and was therefore retained for final evaluation on the test set (S1 Table.). As the training set is unbalanced, we used a class weight training for all thoses methods by adjusting class weights inversely proportional to class frequencies in the input data. This approach helps to mitigate bias towards the majority class and improve performance on the minority class [22]. Hyperparameters were optimized using grid search. The selected random forest model consists of 100 trees, with a minimum of 2 samples required to split an internal node (maximum depth selected according to this sample split). The Gini impurity criterion was used for measuring the quality of splits. The models were implemented using the scikit-learn library in Python [24]. Model selection was performed using 5-fold stratified cross-validation on the training set, based on the weighted F1-score, to compare candidate classifiers. The random forest model consistently achieved the best performance and was therefore selected. Following this selection, the final model was re-instantiated and retrained on the full training dataset using the chosen hyperparameters.

On the held-out test set, the random forest classifier attained a sensitivity (true positive rate) of 96.8% and a specificity (true negative rate) of 97.0%. The area under curve of the receiver operating characteristic curve (ROC AUC) is 0.9976 (Fig 3a–c). Only three images were misclassified; the ambiguous nature of these cases is examined in the Discussion. The distribution of pFECD across the test set comprising 33 healthy, 63 FECD, and 6 haze images, is shown in Fig 3a and reveals a clear separation in the x-axis (FECD probability) between the control and pathological groups (y-axis being the model-estimated probability of FECD represented by the violin curve for data distribution visualization), underscoring the discriminative power of the physics-based quantifiers and motivating a closer examination of the few borderline cases.

Fig 3. Random Forest Results.

Fig 3

(a) Distribution of predicted FECD probabilities (pFECD) for healthy corneas, FECD, and visual-haze images in the independent test set. The horizontal axis corresponds to the model-estimated probability of FECD represented by the violin curve. Healthy samples cluster near 0, whereas FECD and visual-haze samples cluster near 1 (healthy mean predicted probability 10.0% std: 13.2%; Fuchs’ Dystrophy mean predicted probability: 90.9% std: 14.2%; Haze mean predicted probability 96.0% std: 4.7%). (b) Receiver Operating Characteristic (ROC) Curve, Area Under the Curve (AUC) 0.9976. (c) Confusion matrix on test set.

Repeatability results

A repeatability study was performed to test the robustness of the classification. We tested the model on a series of 5 images acquired on the same day by the same orthoptist on the same patients, including 10 healthy subjects (10 corneas × 5 repetitions) and 3 patients with Fuchs’ Dystrophy (5 corneas × 5 repetitions). The results of the repeatability study are shown in Fig 4. The classification accuracy of the model is 100%, with a compact and statistically significant separation between the healthy and pathological groups. However, we denote a strong dispersion of results in the Healthy group (Fig 4a.) that do not correlate with the SNR index (Eq 6), nor with quantifiers dispersion as illustrated with peak width in Fig 4b suggesting that factors other than signal quality may contribute to the observed variability in the healthy group. Nevertheless, we noticed variation in the OCT signal itself in terms of signal intensity, texture or artefacts, as illustrated in S2 Fig. potentially coming from the OCT system recording process.

Fig 4. Repeatability results: study on 10 normal corneas (group control) and 5 corneas with Fuchs’ dystrophy (pathological group).

Fig 4

(a) Probabilities of dystrophy classification for the 5 images per corneas, respectively normal and pathological. (b) Repeatability of peak width quantifier.

SNR=10log10(PsignalPnoise) (6)

where Psignal is the power of the signal (mean intensity of the corneal region from the air/lacrymal filp to the 1.5 mm deep, as it is the region where quantifiers are calculated) and Pnoise is the power of the noise (standard deviation of the background region above and below the cornea). The SNR values for the images in the repeatability study ranged from 15 to 25 dB.

Transfer on the application cohort (haze quantification)

The trained model was applied to the 6 images from patients that underwent PRK surgery and reporting vizual haze (application cohort from Table 1). The predicted probabilities of FECD classification, pFECD, for these images are shown in Fig 3a (purple). Notably, all 6 haze images were assigned a high probability of being classified as pathological, with pFECD values ranging from 0.95 to 0.99, indicating that the model successfully identified the fibrotic features in these post-PRK corneas. This result supports our hypothesis that the morphological quantifiers derived from OCT images can serve as a reliable and interpretable indicator of corneal fibrosis, even when the model is trained on a different but related pathology (i.e., FECD). The successful transfer of the model to the haze cohort demonstrates its potential utility for objective quantification of corneal haze in clinical practice.

Extension to post-PRK surgery follow-up

To further evaluate the model’s applicability for monitoring corneal haze progression, we applied it to a longitudinal dataset of 9 corneas (from 5 patients) who underwent PRK surgery and were followed up at multiple time points post-operatively. The predicted probabilities of FECD classification, pFECD, were tracked over time for each patient (Fig 5a). We observed that pFECD values were elevated in the early post-operative period, consistent with the presence of subepithelial inflammation (normal after the surgery) often associated to a visual haze symptom for the patient, similarly to fibrosis. Over subsequent follow-up visits, a gradual decrease in pFECD was noted in most patients, reflecting a probable resolution of inflammation of fibrosis and return towards a more normal corneal state. We found it insightful to represent this follow-up relatively to the initial value of the probability and therefore represent potential improvement or degradation of such quantifier (Fig 5a). Comparing Fig 5a and Fig 5b, we can observe that patients with low initial pFECD values tend to maintain high probabilities over time, whereas patients with higher initial pFECD values show a more pronounced decrease, suggesting a potential correlation between the initial severity of fibrosis and the trajectory of recovery.

Fig 5. Haze post-surgery follow-up study (a) Predicted probabilities pFECD (left) and relative to initial pFECD probability (right) of FECD classification in Pre/Post-surgery follow-up for 9 patients.

Fig 5

(b) Pre/post-surgery follow-up of peak width. The size of the points is relative to the signal-to-noise ratio of the OCT image analyzed.

Those results illustrate after the surgery a quatitative simularity of corneal structure of patient after PRK surgery with the one of patients with FECD, and therefore the potential of our model to monitor the evolution of corneal haze in a quantitative and objective manner compared to conventional clinical assessment. It suggests patient-specific trajectories of recovery that could be further investigated in larger cohorts to identify potential prognostic factors and inform personalized management strategies for patients undergoing PRK surgery.

Interpretability using SHAP

By aggregating the absolute Shapley values across all instances, we derived a ranked feature importance profile (Fig 6a), highlighting the peak width as the most influential variable contributing to the model’s predictive accuracy. The SHAP summary plot (Fig 6a.) revealed nuanced insights into feature interactions: for instance, higher values of peak width, sigma, area ratio, and depth; and lower values of beta (reduced stromal intensity signal decay) are associated with higher SHAP values and, therefore, a higher FECD class probability prediction. Those parameters relating by construction to the 3 morphological parameters of subepithelial fibrosis: sub-epithelial hyper reflectivity (link to area ratio), blurred epithelium/stroma interface (link to peak width, sigma and depth), altered reflectivity depth profile, stromal transparency (link to beta).

Fig 6. Interpretability insights (a): Shapeley Additive explanation (SHAP) features importance displaying how each training samples (dots) influences the model output through lower SHAP values (tendency to healthy class) or higher SHAP values (tendency to FECD class).

Fig 6

Features are ordered from higher to lower impact (top to bottom). (b): Gini feature importance. (c) Example of SHAP analysis for True Healthy Cornea (top) and True Cornea with Fuchs’ Dystrophy (FECD). The contribution of each feature to output prediction is illustrated. (d): False Healthy cornea SHAP analysis. (e): False Fuchs’ Dystrophy (FECD) SHAP analysis.

The feature importance deduced by SHAP is confirmed by traditional impurity-based importance scores (e.g., Gini Impurity) (a higher score indicates that the feature contributes more to the predictive power of the model) (Fig 6b.). Additionally, SHAP provides a local explanation: for each prediction, it shows how each feature either increases (pushes toward) or decreases (pushes against) the predicted value compared to a baseline (e.g., the mean prediction or expected value). For instance, on a true healthy cornea, all features participate in the predicted probability of 0 (Fig 6c, top); conversely, on a true FECD cornea, all features participate in the predicted probability of 1 (Fig 6c, bottom). In those two examples, Peak width, Sigma, and area ratio are the main participating features. Additionally, such a study can be made on misclassified data to understand what leads to a wrong prediction (Fig 6d, e) and will be discussed later. Altogether, these pieces of information indicate the peak width as a potential good biomarker of sub-epithelium fibrosis. Training the Random Forest model only with the peak width parameter leads to a classification accuracy on the FECD/healthy dataset (evaluated by cross-validation) of 88.5% (std: 2.94%), similar accuracy is found for SVM or logistic models. The use of the additional 8 quantifiers allows us to reach a testing accuracy of 96.9%.

Discussion

Misclassifications

Although all haze images were classified in the FECD class without retraining, one healthy cornea and two corneas with FECD were misclassified in the test set. Notably, the model outputs probabilities close to 0.5 for these three images (Fig 3a; Fig 6d,e), deviating from the rest of the dataset predictions and indicating classification uncertainty. An extended SHAP analysis of these three misclassified images reveals, for the false FECD cornea, a potential conflict between peak width and sigma quantifiers (Fig 6d). Similarly, for the false healthy corneas, a conflict between peak width, sigma, and stromal signal parameters (alpha and beta) is appearing (Fig 6e). The depth-averaged intensity profiles for these cases do not demonstrate distinctive features that would explain the misclassification (S3 Fig.), suggesting that the classification ambiguity may arise from subtle interactions between multiple quantifiers rather than from obvious morphological anomalies.

Future work could address these borderline cases by introducing additional quantitative parameters sensitive to signal density, spatial disparity, or textural information, which might help identify interpretation biases that contribute to classification errors. Alternatively, convolutional neural networks [25,26], although less interpretable than our physics-based approach, might detect local fine-scale features not captured by our depth-averaged quantifiers. Additionally, explainability techniques such as Gradient-weighted Class Activation Mapping (Grad-CAM) [27] could highlight directly on OCT images the regions participating most in the classification, potentially revealing spatial patterns that complement our quantitative descriptors.

Post-operative follow-up after refractive surgery

According to clinical literature, transient inflammation induced haze following PRK surgery is typically expected within 2–6 weeks post-surgery, maturing between 1 and 3 months, and then regressing within a few weeks to 12 months [2,11]. Our longitudinal follow-up data are consistent with this temporal pattern: an increase in pFECD was observed between 1 and 3 months post-operatively, followed by either a reduction (improvement) or stagnation in 8 of 9 cases, with only one case showing no improvement (Results section and Fig 5a). Fig 7 illustrates the challenge of assessing mild corneal haze: visual distinction between normal images (pre-surgery, Fig 7b, and at Month 9, Fig 7f) and corneas with haze (at Month 1, Fig 7c) is difficult, highlighting the value of the quantitative pFECD indicator for objective post-operative monitoring. However, the sample size of this longitudinal cohort (9 corneas) is too small to draw definitive conclusions about the validity of this approach for fine-grained haze tracking over time. Extension to a larger and more temporally resolved cohort is necessary to validate the present approach. Previous densitometry studies have shown the persistence of stromal light scattering more than twenty years after surgery [28–30]: comparing our model estimates in similar long-term follow-up cases would provide valuable insight into the relationship between our physics-based quantifiers and established densitometric measures.

Fig 7. Example of surgery follow up.

Fig 7

(a): Evolution of Fuchs Dystrophy probability predicted by the model used as an indicator of fibrosis pre and post-surgery. (b–f) associated SD-OCT images recorded respectively before and 1,3,6,9 month after surgery.

Clinical utility as a diagnostic aid

The primary output of our model is the predicted probability pFECD, which should be considered as one aggregated quantitative indicator among others in a clinical diagnostic aid context. The probability pFECD reflects the degree of confidence in the classification: clear-cut values approaching 0 or 1 provide strong evidence for classification as healthy or FECD like symptoms, respectively, and can serve to confirm clinical findings obtained through conventional slit-lamp examination. Conversely, values near 0.5 or predictions that contradict expected clinical presentation should be interpreted as diagnostic warnings, prompting closer examination of the individual quantifiers and the underlying OCT images.

In such ambiguous cases, the individual physics-based quantifiers provide interpretable indicators that can guide clinical reasoning and make comparison point. For instance, the alpha and beta parameters offer information on stromal signal decay and transparency, mean square error (MSE) reflects the regularity or irregularity of the depth profile feature that are subjectively accessed in current clinical setting. SHAP analysis can help prioritise which features contributed most to a given prediction, as illustrated in Fig 6c to e. This interpretability distinguishes our approach from black-box machine-learning methods and facilitates clinical trust and adoption.

Although peak width emerged as the most influential single quantifier, achieving a cross-validation accuracy of 88.5% when used alone, this performance is insufficient for clinical deployment. While this parameter has high interpretability in terms of the axial extent of fibrosis and demonstrates strong discriminative power, a classification accuracy of 88.5% falls below the conventional 95% confidence threshold commonly required in medical statistics. By incorporating the full set of 9 quantifiers, our model attains a test set accuracy of 96.9%, exceeding this clinical acceptability threshold and confirming the added value of the multi-parameter approach for robust and reliable classification.

Conclusion

In this paper, we propose a robust approach for automatic classification of clinical SD-OCT images for the detection of subepithelial fibrosis. The model, based on 9 morphological quantifiers and trained on FECD, achieved a test set accuracy of 96.9%. We demonstrated how training on a common disease (FECD) with permanent fibrotic features enables effective transfer learning to a rarer, transient pathology (corneal haze) presenting similar morphological characteristics. The trained model successfully classified all post-PRK corneal images reporting visual haze with high confidence(high pFECD), supporting its utility for objective haze quantification.

The primary output of our model, the predicted probability pFECD, provides a single quantitative indicator that reflects the degree of confidence in the classification and the severity of fibrotic changes. Critically, this indicator remains interpretable: through SHAP analysis, we established that it is strongly related to the axial extent of fibrosis, as captured primarily by the peak width quantifier, which guarantees clinical transparency. The physics-based foundation of our quantifiers distinguishes this approach from non-interpretable machine-learning methods and facilitates its integration into clinical practice, especially in cases where a follow-up is necessary to monitor the evolution of an inflammation, fibrosis or haze over time.

Supporting information

S1 Fig. Distribution on training set of the 9 quantifiers for each condition.

Statistics are performed with Mann-Whitney U-test (no normality assumption) using Bonferonni correction of p-values (***: p-value<0.001/m; **; p-value<0.01/m; p-value<0.05/m; m = 9).

(TIFF)

pone.0328320.s001.tiff (48.1MB, tiff)
S1 Table. Cross-validation scores of classic machine learning classification models.

(PDF)

pone.0328320.s002.pdf (57.8KB, pdf)
S2 Fig. Raw Repeatability data example with zoom on the apex region.

(a) healthy cornea example, (b) FECD cornea example.

(TIFF)

pone.0328320.s003.tiff (58.2MB, tiff)
S3 Fig. The 3 Misclassified corneas and corresponding SD-OCT raw signal (right top) associated standardized signal and depth profile (right) as a density map to represent signal disparities for the depth profile (one curve is on “column” or axial profile, total number of curves is 750 for a 6 mm region), μ is the mean of the curves and σ is the standard deviation of the curves.

(a) false Fuchs dystrophy cornea. (b–c) false Healthy cornea.

(TIFF)

pone.0328320.s004.tiff (66.4MB, tiff)

Data Availability

The minimal data set underlying the findings of this study: the values of the OCT-derived quantifiers for all corneas, the per-sample model predictions, and the values behind all reported means, standard deviations, statistical tests and figures, is fully anonymised and is available as Supporting Information (S1 Dataset) and in the public repository https://github.com/Coohrentiin/Subepithelial-Fibrosis-Quantifiers. The analysis code is publicly available in the same repository. The raw source data (individual SD-OCT corneal images from which the quantifiers were derived) cannot be shared publicly. These are patient-derived clinical images: the informed consent obtained from participants and the institutional approvals granted for the study did not provide for public release of the raw images, and their dissemination would compromise patient confidentiality. The images are held by the Centre Hospitalier National d’Ophtalmologie (CHNO) des Quinze-Vingts, Paris, France, and the study was approved by the Ethics Committee of the French Society of Ophthalmology (IRB #00008855). Researchers who meet the criteria for access to confidential clinical data may request the raw images from the Data Protection Officer of the CHNO des Quinze-Vingts digital health data warehouse (Entrepôt de Données de Santé), a non-author institutional contact reachable at dpo@15-20.fr, who is able to handle data access requests from researchers external to the hospital. Requests concerning the ethics approval may additionally be addressed to the Ethics Committee of the French Society of Ophthalmology (Société Française d’Ophtalmologie, 17 Villa d’Alésia, 75014 Paris, France; sfo@sfo.asso.fr), chaired by Dr Jean-Jacques Saragouss.

Funding Statement

This study was partially funded through the ‘Prix de la Banque Française des Yeux 2022’ obtained by Anatole Chessel. There was no additional external funding received for this study.

References

  • 1.Fantes FE, Hanna KD, Waring GOI, Pouliquen Y, Thompson KP, Savoldelli M. Archives of Ophthalmology. 1990;108(5):665–75. doi: 10.1001/archopht.1990.01070070051034 [DOI] [PubMed] [Google Scholar]
  • 2.Charpentier S, Keilani C, Maréchal M, Friang C, De Faria A, Froussart-Maille F, et al. Corneal haze post photorefractive keratectomy. J Fr Ophtalmol. 2021;44(9):1425–38. doi: 10.1016/j.jfo.2021.05.006 [DOI] [PubMed] [Google Scholar]
  • 3.Agnia E, Kriskasari G, Dharmawidiarini D, Napitupulu S. Late Onset Corneal Haze Post Photorefractive Keratectomy. Vision Science and Eye Health Journal. 2024;3(3):65–8. doi: 10.20473/vsehj.v3i3.2024.65-68 [DOI] [Google Scholar]
  • 4.Ní Dhubhghaill S, Rozema JJ, Jongenelen S, Ruiz Hidalgo I, Zakaria N, Tassignon M-J. Normative values for corneal densitometry analysis by Scheimpflug optical assessment. Invest Ophthalmol Vis Sci. 2014;55(1):162–8. doi: 10.1167/iovs.13-13236 [DOI] [PubMed] [Google Scholar]
  • 5.Maldonado MJ, Arnau V, Navea A, Martínez-Costa R, Mico FM, Cisneros AL, et al. Direct objective quantification of corneal haze after excimer laser photorefractive keratectomy for high myopia. Ophthalmology. 1996;103(11):1970–8. doi: 10.1016/s0161-6420(96)30400-4 [DOI] [PubMed] [Google Scholar]
  • 6.Soya K, Amano S, Oshika T. Quantification of simulated corneal haze by measuring back-scattered light. Ophthalmic Res. 2002;34(6):380–8. doi: 10.1159/000067046 [DOI] [PubMed] [Google Scholar]
  • 7.Montorio D, Cennamo G, Menna F, Donna P, Napolitano P, Breve MA, et al. Evaluation of corneal structures in myopic eyes more than twenty-two years after photorefractive keratectomy. J Biophotonics. 2020;13(10):e202000138. doi: 10.1002/jbio.202000138 [DOI] [PubMed] [Google Scholar]
  • 8.Weng T-H, Chang Y-M, Lin F-H, Weng Z-X, Wang T-W, Chen Y-H, et al. Investigation of corneal epithelial thickness and irregularity by optical coherence tomography after transepithelial photorefractive keratectomy. Clin Exp Optom. 2024;107(1):23–31. doi: 10.1080/08164622.2023.2197107 [DOI] [PubMed] [Google Scholar]
  • 9.Borderie V, Georgeon C. La dystrophie de Fuchs: une pathologie dont la compréhension et le traitement sont en plein essor. Journal Title Not Provided. 2023.
  • 10.Yasukura Y, Oie Y, Kawasaki R, Maeda N, Jhanji V, Nishida K. New severity grading system for Fuchs endothelial corneal dystrophy using anterior segment optical coherence tomography. Acta Ophthalmol. 2021;99(6):e914–21. doi: 10.1111/aos.14690 [DOI] [PubMed] [Google Scholar]
  • 11.Awwad ST, Bteich Y, Assaf JF, Ghosn A, Hafezi F, Torres-Netto E, et al. Prospective Objective Analysis of Corneal Haze Following Customized Transepithelial PRK Without Mitomycin C Combined With Accelerated Corneal Cross-Linking Versus Corneal Cross-Linking Alone. J Refract Surg. 2024;40(9):e583–94. doi: 10.3928/1081597X-20240715-03 [DOI] [PubMed] [Google Scholar]
  • 12.Bocheux R, Pernot P, Borderie V, Plamann K, Irsch K. Quantitative measures of corneal transparency, derived from objective analysis of depth-resolved corneal images, demonstrated with full-field optical coherence tomographic microscopy. PLoS One. 2019;14(8):e0221707. doi: 10.1371/journal.pone.0221707 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Vilbert M, Bocheux R, Georgeon C, Borderie V, Pernot P, Irsch K, et al. A new method for in vivo assessment of corneal transparency using spectral-domain OCT. PLoS One. 2023;18(10):e0291613. doi: 10.1371/journal.pone.0291613 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Grinsztajn L, Oyallon E, Varoquaux G. Why do tree-based models still outperform deep learning on tabular data?. arXiv. 2022. doi: 10.48550/arXiv.2207.08815 [DOI]
  • 15.Vilbert M. FitOCT-preprocessing. 2021.
  • 16.Soubeiran C. Subepithelial-Fibrosis-Quantifiers; 2026.
  • 17.Kaso A. Computation of the normalized cross-correlation by fast Fourier transform. PLoS One. 2018;13(9):e0203434. doi: 10.1371/journal.pone.0203434 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Winn B, Whitaker D, Elliott DB, Phillips NJ. Factors affecting light-adapted pupil size in normal human subjects. Invest Ophthalmol Vis Sci. 1994;35(3):1132–7. [PubMed] [Google Scholar]
  • 19.Leyland M. Validation of Orbscan II Posterior Corneal Curvature Measurement for Intraocular Lens Power Calculation. Eye. 2004;18(4):357–60. doi: 10.1038/sj.eye.6700659 [DOI] [PubMed] [Google Scholar]
  • 20.Larocca F, Chiu SJ, McNabb RP, Kuo AN, Izatt JA, Farsiu S. Robust automatic segmentation of corneal layer boundaries in SDOCT images using graph theory and dynamic programming. Biomed Opt Express. 2011;2(6):1524–38. doi: 10.1364/BOE.2.001524 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Lundberg S, Lee SI. A Unified Approach to Interpreting Model Predictions. 2017. doi: 10.48550/arXiv.1705.07874 [DOI]
  • 22.Haixiang G, Yijing L, Shang J, Mingyun G, Yuanyue H, Bing G. Learning from class-imbalanced data: Review of methods and applications. Expert Systems with Applications. 2017;73:220–39. doi: 10.1016/j.eswa.2016.12.035 [DOI] [Google Scholar]
  • 23.Lemaitre G, Nogueira F, Aridas CK. Imbalanced-Learn: A Python Toolbox to Tackle the Curse of Imbalanced Datasets in Machine Learning. arXiv. 2016. doi: 10.48550/arXiv.1609.06570 [DOI]
  • 24.Pedregosa F, Varoquaux G, Gramfort A, Michel V, Thirion B, Grisel O. Scikit-Learn: Machine Learning in Python. arXiv. 2018. doi: 10.48550/arXiv.1201.0490 [DOI]
  • 25.LeCun Y, Bengio Y, Hinton G. Deep learning. Nature. 2015;521(7553):436–44. doi: 10.1038/nature14539 [DOI] [PubMed] [Google Scholar]
  • 26.Han YS, Yoo J, Ye JC. Deep Residual Learning for Compressed Sensing CT Reconstruction via Persistent Homology Analysis. arXiv. 2016. doi: 10.48550/arXiv.1611.06391 [DOI]
  • 27.Selvaraju RR, Cogswell M, Das A, Vedantam R, Parikh D, Batra D. Grad-CAM: Visual Explanations from Deep Networks via Gradient-based Localization. International Journal of Computer Vision. 2020;128(2):336–59. doi: 10.1007/s11263-019-01228-7 [DOI] [Google Scholar]
  • 28.Cennamo G, Menna F, Sinisi F, Cennamo G, Breve MA, Napolitano P, et al. Twenty-Year Follow-Up of Excimer Laser Photorefractive Keratectomy: A Retrospective Observational Study. Ophthalmol Ther. 2020;9(4):917–27. doi: 10.1007/s40123-020-00281-7 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Poyales F, Garzón N, Mendicute J, Illarramendi I, Caro P, Jáñez O, et al. Corneal densitometry after photorefractive keratectomy, laser-assisted in situ keratomileusis, and small-incision lenticule extraction. Eye (Lond). 2017;31(12):1647–54. doi: 10.1038/eye.2017.107 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Netto MV, Mohan RR, Sinha S, Sharma A, Dupps W, Wilson SE. Stromal haze, myofibroblasts, and surface irregularity after PRK. Exp Eye Res. 2006;82(5):788–97. doi: 10.1016/j.exer.2005.09.021 [DOI] [PMC free article] [PubMed] [Google Scholar]

Decision Letter 0

Georgios Labiris

28 Dec 2025

PONE-D-25-32169

Explainable machine learning for post PKR surgery follow-up

PLOS One

Dear Dr. Soubeiran,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

Both reviewers identify substantial concerns regarding clarity, methodological transparency, and reproducibility that must be addressed before the manuscript can be considered for publication. For acceptance, the authors are required to clearly and consistently define the study objective across the title, abstract, and main text; explicitly describe the datasets, labeling strategy, and training/validation/testing design; and provide sufficient methodological detail (including image processing steps, feature definitions, mathematical formulations, and machine learning procedures) to allow independent replication, in line with PLOS ONE’s criteria for methodological rigor and transparency. Figures and tables that are currently unclear or incomplete must be revised or removed, and terminology should be made consistent throughout. Recommended revisions include expanding the clinical and technical background in the Introduction to improve accessibility for a broad readership and strengthening the literature context. Where reviewer comments overlap (e.g., on unclear methodology), these should be addressed comprehensively; where they diverge, priority should be given to ensuring reproducibility and clarity rather than perceived novelty or impact. The decision for Major Revision is based solely on the need to meet PLOS ONE’s standards for clarity, rigor, and reproducibility.

Please submit your revised manuscript by Feb 11 2026 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.

  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.

  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols.

We look forward to receiving your revised manuscript.

Kind regards,

Georgios Labiris, MD, PhD

Academic Editor

PLOS One

Journal Requirements:

When submitting your revision, we need you to address these additional requirements.

1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at

https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and

https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

2. Please note that PLOS One has specific guidelines on code sharing for submissions in which author-generated code underpins the findings in the manuscript. In these cases, we expect all author-generated code to be made available without restrictions upon publication of the work. Please review our guidelines at https://journals.plos.org/plosone/s/materials-and-software-sharing#loc-sharing-code and ensure that your code is shared in a way that follows best practice and facilitates reproducibility and reuse.

3. Thank you for stating the following in the Acknowledgments Section of your manuscript:

“This study was partially funded through the “Prix de la Banque Francaise des Yeux 2022" obtained by Anatole Chessel.”

We note that you have provided funding information that is not currently declared in your Funding Statement. However, funding information should not appear in the Acknowledgments section or other areas of your manuscript. We will only publish funding information present in the Funding Statement section of the online submission form.

Please remove any funding-related text from the manuscript and let us know how you would like to update your Funding Statement. Currently, your Funding Statement reads as follows:

“This study was partially funded through the “Prix de la Banque Francaise des Yeux 2022" obtained by Anatole Chessel.”

Please include your amended statements within your cover letter; we will change the online submission form on your behalf.

4. We note that the grant information you provided in the ‘Funding Information’ and ‘Financial Disclosure’ sections do not match.

When you resubmit, please ensure that you provide the correct grant numbers for the awards you received for your study in the ‘Funding Information’ section.

5. Thank you for stating in your Funding Statement:

“This study was partially funded through the “Prix de la Banque Francaise des Yeux 2022" obtained by Anatole Chessel.”

Please provide an amended statement that declares *all* the funding or sources of support (whether external or internal to your organization) received during this study, as detailed online in our guide for authors at

http://journals.plos.org/plosone/s/submit-now.  Please also include the statement “There was no additional external funding received for this study.” in your updated Funding Statement.

Please include your amended Funding Statement within your cover letter. We will change the online submission form on your behalf.

6. Thank you for stating the following financial disclosure:

“This study was partially funded through the “Prix de la Banque Francaise des Yeux 2022" obtained by Anatole Chessel.”

Please state what role the funders took in the study.  If the funders had no role, please state: "The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript."

If this statement is not correct you must amend it as needed.

Please include this amended Role of Funder statement in your cover letter; we will change the online submission form on your behalf.

7. We note that you have indicated that there are restrictions to data sharing for this study. PLOS only allows data to be available upon request if there are legal or ethical restrictions on sharing data publicly. For more information on unacceptable data access restrictions, please see http://journals.plos.org/plosone/s/data-availability#loc-unacceptable-data-access-restrictions.

Before we proceed with your manuscript, please address the following prompts:

a) If there are ethical or legal restrictions on sharing a de-identified data set, please explain them in detail (e.g., data contain potentially identifying or sensitive patient information, data are owned by a third-party organization, etc.) and who has imposed them (e.g., a Research Ethics Committee or Institutional Review Board, etc.). Please also provide contact information for a data access committee, ethics committee, or other institutional body to which data requests may be sent.

b) If there are no restrictions, please upload the minimal anonymized data set necessary to replicate your study findings to a stable, public repository and provide us with the relevant URLs, DOIs, or accession numbers. For a list of recommended repositories, please see

https://journals.plos.org/plosone/s/recommended-repositories. You also have the option of uploading the data as Supporting Information files, but we would recommend depositing data directly to a data repository if possible.

We will update your Data Availability statement on your behalf to reflect the information you provide.

8. Your ethics statement should only appear in the Methods section of your manuscript. If your ethics statement is written in any section besides the Methods, please move it to the Methods section and delete it from any other section. Please ensure that your ethics statement is included in your manuscript, as the ethics statement entered into the online submission form will not be published alongside your manuscript.

9. Please remove your figures from within your manuscript file, leaving only the individual TIFF/EPS image files, uploaded separately. These will be automatically included in the reviewers’ PDF.

10. If the reviewer comments include a recommendation to cite specific previously published works, please review and evaluate these publications to determine whether they are relevant and should be cited. There is no requirement to cite these works unless the editor has indicated otherwise.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Partly

Reviewer #2: Partly

2. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

Reviewer #2: I Don't Know

3. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: No

Reviewer #2: No

4. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: No

Reviewer #2: Yes

5. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: This manuscript deals with an important clinical issue. It proposes hand-crafted features extracted from OCT images and uses classical deep learning techniques to identify clinically abnormal corneal conditions. However, serious issues exist, which need to be addressed before reconsidering for publication.

One major issue of this manuscript is the lack of clarity about the aim and the objective. The title is very unclear as “machine learning for PRK follow up”.

In Line 149 - 150 it is stated “Our final goal is an automatic classification of clinical OCT images into two categories: "healthy corneas" and "corneal haze"”. This statement, if it holds, it should be made very early in the manuscript.

In the abstract it is stated “we trained the model on corneas presenting Fuchs’ dystrophy, causing similar, but permanent, fibrosis symptoms, and applied it to images from patients who have undergone PRK surgery”. Immediately after that is follows “… the model output (probability of Fuchs’ dystrophy classification) …”

Then, results are presented using 5-cross validation on the FECD images. In Table 1 several databases are mentioned without explaining the training set and the test set from which results are presented.

Moreover, in Fig. 5 repeatability results are presented from 10 healthy and 5 Fuchs retinas, without explaining if these are different from the 166 healthy, 313 Fuch retinas, or the remaining Haze retinas of Table 5.

Thus it is unclear to the reader what is the aim of the work.

The authors should modify the title and rewrite the abstract to make it more specific. Following the introduction, a short subsection entitled “Objective and contributions” (or similar) should be added describing exactly that, without any ambiguity. Table 2 should be redesigned to clarify all the above.

Fig. 4a is completely unclear. The “distribution” axis appears to have no meaning. The blue and red areas are not explained. Please either describe the figure extensively in the text, explaining every aspect, or remove it from the manuscript, since in its current form it only creates questions to the readers.

Subsection 3.2, Lines 84 - 95 “the processing chain described in Ref. 2. This pipeline is supplemented with an additional alignment procedure relying on normalised cross-correlation”

The authors must include the exact steps that are followed, both for [2], as well as for the alignment procedure and the rest of the steps, since the above are non-trivial image processing tasks. If this is too technical for the journal, then some form of additional material should be used. In its current form, the statement is not acceptable for publication, since the methodology is obscure and researchers cannot dublicate the procedures.

Caption of Fig. 2: “red square” --> “red rectangle”

Subsection 3.3

The descriptions are very brief, without definitions and clear steps. Just to mention some confusing points:

There is no clear description how the image in Fig. 3b is generated from Fig.3a

The “depth-averaged profile” is calculated along the Y-axis, or along the normal vector to the corneal surface ? Please define explicitly.

Every step of the process after definition of the profile is based on clearly identifiable landmarks of the profile (first anterior maximum, 2nd maximum, minimum before and after the 2nd maximum, etc). It seems from the manuscript that if these landmarks are not correctly detected, the remaining process will fail. The authors should discuss how will their proposed system compensate for depths with increased SNR, where the profile maybe very noisy preventing the detection of these curve landmarks.

In Line 134 the exponential regressor has a constant term C, which is missing from Fig. 3f.

In Fig. 3e the MSE is wrong (summation over the samples is missing, division be the number of samples).

As a general comment, mathematical quantities should be rigorously defined in the manuscript in numbered Equations, with every symbol explained. Please remove the incomplete, hard-to-read mathematical definitions from the figures.

Line 110: “An interpolation is performed to improve the sampling in this area …”:

Once again, please define clearly every aspect of the step.

Define the calculation of “Area ratio”. The relevant lines 141-143 are completely confusing.

Subsection 4.3

The purpose of the subsection is unclear. Is there any connection with the proposed system ? Please clarify

Fig. 6

Please define the calculation of SNR of the OCT image, the effect it has in the prediction. Also provide the numerical values.

Line 236, 260 “FD class” is the same as “FECD” class ? If yes, please replace.

Line 236 why is “additional training” mentioned ? Has it been used or discussed somewhere else in the manuscript ?

Fig. S3, right column: please explain in the text how are the figures generated, from how many images and how are μ, σ calculated and from which specific images in each figure

Reviewer #2: Soubeiran et al have presented a machine learning based approach to classify healthy corneas vs those with Fuchs’ dystrophy. The study setup seems appropriate and the results are promising.

However, as a general PLOS One reader unfamiliar with this medical field, I was unable to fully follow the manuscript and found the description of both the disease and the methodology lacking for a comprehensive review. I have described both my major and minor concerns below:

1. As PLOS One caters to a diverse audience, it would be helpful if the authors presented more information about the medical condition and the problem motivation. I appreciate the need to keep the manuscript succinct, and do not expect a full background of every medical term. However using simpler language, describing the most relevant terms, and emphasizing the motivation of this study would make the manuscript more broadly accessible.

2. The methodologies are not described sufficiently for a comprehensive review. For example, data cleaning methodologies are only subjectively described. Feature ranges are missing. Image labeling methodology is missing. Hyperparameter selection is missing.

3. Line 15 - Suggest including the name of the machine learning algorithm in the title, abstract, and Introduction. ML is too broad a term to describe the authors’ specific work.

4. Line 28 - It is not clear why the authors have started the Introduction with a reference to their own previous work. While both works relate to Fuchs, the previous work has not included any ML. I would suggest starting the Introduction with the specific problem being solved in this work, and referencing previous work in a later paragraph.

5. Introduction - There is almost no literature review in Introduction. The authors have claimed that they are not aware of quantitative tools but they have not provided sufficient data to support their statement. Additionally, even if quantitative tools are not available, surely there would be other work related to Fuchs that motivates the authors’ work and highlights why haze diagnosis is required. Other work will also serve as a comparison for how the authors’ ML model improves upon existing evaluation performance.

6. Line 49 - What are the missing physical parameters?

7. Fig 1 - What are Ref 3 and Ref 6?

8. Line 56 - Please specify the algorithm instead of broad term like “automatic classification approach”.

9. Line 60 - The sentence is unclear.

10. Line 70 - The sentence is unclear.

11. Line 74 - What segmentation algorithms are being referred to? If there are segmentation algorithms applied here, how does this relate to authors’ pervious statement about missing quantitative tools?

12. Line 75 - What is the issue with “overestimation of the epithelial layer thickness”?

13. Line 78 - Since the “marked morphological features” are an essential input in the authors’ model, it is insufficient to refer to their description in another reference. I would suggest including a description in this manuscript.

14. Line 85 - What is Ref 2?

15. Line 85 - What is the “additional alignment procedure”?

16. Line 87 - How much time and effort is required to detect the “central saturation artefact”? If its detection is essential to leveraging the authors’ methodology, would the image analysis already be done, thus making their approach extraneous?

17. Section 3.2 - Would these image processing capabilities be available to physicians who intend to use the authors’ methodology?

18. Line 90 - Please include more description and reference for PCA.

19. Line 90 - How are low intensity regions defined and removed?

20. Line 91 - How is 1.5 mm chosen?

21. Line 93 - What is the functional form of the weighting function? How is the weight range chosen?

22. Fig 3 - “peak Area Ratio” is not described in Fig 3d.

23. Line 119 - What is being defined in item 4?

24. Line 143 - What is meant by “available simultaneously”?

25. Line 153 - It would be helpful if the authors provided evidence about how the classification model applied to FECD images justifies the model’s application for corneal haze detection.

26. Line 155 - Are the 9 quantifiers the features for the ML models?

27. Line 164 - How would the doctors label the images? Would they use any of the 9 quantifiers to make their determination?

28. Line 169 - “very precisely” is a subjective term, and not descriptive in this context.

29. Line 170 - How did the authors train SVM, determine its hyperparameters, and experiment with reducing overfitting?

30. Line 172 - Does this mean that no label separation was done for implementation of SVM? How can the models performances be compared if they were trained with different datasets?

31. Line 177 - How did the authors select 50% as the probability threshold, especially because the training dataset is unbalanced?

32. Line 184 - After cross-validation, since there will be 5 trained models, which one of the models was applied to the test set?

33. Line 186 - The sentence structure makes it seem that the high TPR and FPR at a single threshold contributes to the high ROC, which is inaccurate since ROC is generated over the entire threshold range.

34. Line 218 - Suggest changing AI to ML to remain consistent with the rest of the manuscript.

6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

Reviewer #2: No

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

To ensure your figures meet our technical requirements, please review our figure guidelines: https://journals.plos.org/plosone/s/figures

You may also use PLOS’s free figure tool, NAAS, to help you prepare publication quality figures: https://journals.plos.org/plosone/s/figures#loc-tools-for-figure-preparation.

NAAS will assess whether your figures meet our technical requirements by comparing each figure against our figure specifications.

PLoS One. 2026 Oct 22;21(10):e0328320. doi: 10.1371/journal.pone.0328320.r002

Author response to Decision Letter 1


5 Mar 2026

We sincerely thank Reviewers for their detailed, constructive comments. We have carefully revised the manuscript in response to all points raised. Below, we provide a point-by-point response. All changes are indicated in the revised manuscript; revised line numbers refer to the resubmitted version. Detailed answers are formatted into the "Response to Reviewer" document.

Reviewer #1:

1. Comment: One major issue of this manuscript is the lack of clarity about the aim and the objective. The title is very unclear as “machine learning for PRK follow up”.

In Line 149 - 150 it is stated “Our final goal is an automatic classification of clinical OCT images into two categories: "healthy corneas" and "corneal haze"”. This statement, if it holds, it should be made very early in the manuscript.

In the abstract it is stated “we trained the model on corneas presenting Fuchs’ dystrophy, causing similar, but permanent, fibrosis symptoms, and applied it to images from patients who have undergone PRK surgery”. Immediately after that is follows “… the model output (probability of Fuchs’ dystrophy classification) …”

Then, results are presented using 5-cross validation on the FECD images. In Table 1 several databases are mentioned without explaining the training set and the test set from which results are presented.

Moreover, in Fig. 5 repeatability results are presented from 10 healthy and 5 Fuchs retinas, without explaining if these are different from the 166 healthy, 313 Fuch retinas, or the remaining Haze retinas of Table 5.

Thus it is unclear to the reader what is the aim of the work.

The authors should modify the title and rewrite the abstract to make it more specific. Following the introduction, a short subsection entitled “Objective and contributions” (or similar) should be added describing exactly that, without any ambiguity. Table 2 should be redesigned to clarify all the above.

a. Answer: We understand that the initial version of the manuscript did not sufficiently clarify the aim, workflow, and dataset usage, which could lead to confusion regarding the objective and evaluation strategy. In the revised manuscript, we clarified the objective: The title has been refined to better reflect the scope of the study: derive fibrosis indicators. We believe we now explicitly state that the primary goal is the automatic classification of corneal OCT images into healthy vs. fibrotic (haze-like) corneas, using physics-based quantifiers and a machine learning model. We explain the rationale of training on FECD images (as a proxy for fibrosis) and applying the model to PRK follow-up data.

b. In details, we added in the abstract a precision about the transient nature of the fibrosis and the link with Fuchs’ Dystrophy (and shared symptoms), and clarified the use of OCT-derived quantifiers. As the reviewer reported confusion and a structural problem in the introduction. We performed a major revision of the introduction. We now start by the source of the corneal Haze the PKR surgery, describing its clinical symptoms and presenting the current means of evaluation (revision l1-24). We explain why OCT may be more reliable for quantitative assessment (revision l26-29). We then describe the transient nature of the fibrosis occurring in Haze and why Fuch’s Dystrophy is interesting, as it shares similar symptoms but is permanent and therefore suitable to create a bigger database to access such symptoms (revision l30-42). We conclude the introduction by a clear statement of our aim (address the lack of objective haze quantification) and hypothesis. To remove ambiguity, we introduced a dedicated paragraph at the end of the Introduction (revision l43–53) explicitly stating the objective and hypothesis of the study. This paragraph clearly distinguishes: the training/validation task (classification of FECD vs. healthy corneas), from the application task (transfer to PRK haze assessment).

c. We also clarified the use of datasets throughout the manuscript. In the revised Materials and Methods and Results sections, we explicitly describe: the composition of each dataset (healthy, FECD, PRK), the role of each dataset (training, validation, independent testing), and the cross-validation strategy (revision l376–383, Table 1 updated accordingly). By defining each cohort successively we explain what they address (revision l372-404) and kept the same order in results (revision l332-516).

d. Regarding the confusion on specific figures: For Fig. 5 (repeatability), we now explicitly state that the subset “10 healthy subjects (10 corneas × 5 repetitions) and 3 patients with Fuchs’ Dystrophy (5 corneas × 5 repetitions)” (revision l472-473), we believe the the new structuration explicit the independent subset used exclusively for repeatability analysis, distinct from the training and validation cohorts. We clarified these distinctions in the figure caption and corresponding text, and Table 1 (and related tables) has been redesigned to clearly indicate dataset splits, sample sizes, and their respective roles (training, validation, testing,), ensuring full transparency and reproducibility.

e. We understand that this comment was broader then the listed points so we added other modifications such as clarifying the processing pipeline (Steps A-H), the quantifiers construction (and their link to clinical insight), precised the processing time (l327-335), robustness (l337-354).

2. Comment: Fig. 4a is completely unclear. The “distribution” axis appears to have no meaning. The blue and red areas are not explained. Please either describe the figure extensively in the text, explaining every aspect, or remove it from the manuscript, since in its current form it only creates questions to the readers.

a. Answer: Thank you for this remark. We renamed the axis (“probability density”), and the colors are detailed in the legend.

3. Comment: Subsection 3.2, Lines 84 - 95 “the processing chain described in Ref. 2. This pipeline is supplemented with an additional alignment procedure relying on normalised cross-correlation”

The authors must include the exact steps that are followed, both for [2], as well as for the alignment procedure and the rest of the steps, since the above are non-trivial image processing tasks. If this is too technical for the journal, then some form of additional material should be used. In its current form, the statement is not acceptable for publication, since the methodology is obscure and researchers cannot dublicate the procedures.

a. Answer: We agree. In the revised version, we have addressed this concern by explicitly detailing the complete processing chain within the manuscript (Section Materials and Methods, revision l112–220). All steps (A to H) are now described, including preprocessing, artefact detection and removal, anterior surface detection, A-scan alignment, stromal segmentation, PCA-based artefact correction, lateral normalization, and depth-intensity profile extraction. In particular, the alignment procedure previously referenced has now been fully described (revision l150). To further improve clarity and reproducibility, we ensured that all algorithmic parameters (e.g., filter sizes, thresholds, window widths) are specified in the text, and we provide mathematical formulations where applicable. As a result, the methodology is now self-contained within the manuscript and does not rely on external references for critical steps.

4. Comment: Caption of Fig. 2: “red square” --> “red rectangle”

a. Answer: Done

5. Comment: Subsection 3.3 The descriptions are very brief, without definitions and clear steps. Just to mention some confusing points:

There is no clear description how the image in Fig. 3b is generated from Fig.3a

The “depth-averaged profile” is calculated along the Y-axis, or along the normal vector to the corneal surface ? Please define explicitly.

Every step of the process after definition of the profile is based on clearly identifiable landmarks of the profile (first anterior maximum, 2nd maximum, minimum before and after the 2nd maximum, etc). It seems from the manuscript that if these landmarks are not correctly detected, the remaining process will fail. The authors should discuss how will their proposed system compensate for depths with increased SNR, where the profile maybe very noisy preventing the detection of these curve landmarks.

a. Answer: Thank you for this comment we therefore reworked this section 3.3, which now corresponds to “Definition of subepithelial fibrosis quantifiers” section l229, additionally to the pipeline details we added in Step A-H. The definitions of quantifiers is now clearer and we mention their meaning. Additionally, regarding the implementation itself, we added details and “Processing time” (l327). We also want to mention that the code will be available upon publication for the reader as required by PlosOne.

b. Step C now explain the figure 2 (replacing figure 3) a to b. We understand that this remark was wider than just this point, and we therefore meticulously detailed the other steps and quantifiers.

c. Regarding landmark detection, we addressed this point in a new paragrapher “Robustness to low-SNR conditions and landmark detection failures” l337

6. Comment: In Line 134 the exponential regressor has a constant term C, which is missing from Fig. 3f.

a. Answer: Added

7. Comment: In Fig. 3e the MSE is wrong (summation over the samples is missing, division be the number of samples).

As a general comment, mathematical quantities should be rigorously defined in the manuscript in numbered Equations, with every symbol explained. Please remove the incomplete, hard-to-read mathematical definitions from the figures.

a. Answer: Thank you, the MSE was displaying the upper bar that is the mathematical symbol of the mean that includes the mentioned division. However, we agree that more mathematical formalism would clarify the manuscript, we added all the formulas, including the MSE (equation 2 l279-280) that clarify the bar notation and explexitely mention it at l282. We understand the fact that MSE and COV were hard to read in figure 2 (replacing fig3), but as we want this figure to display the 9 quantifiers, we keep it by putting it in a box and coloring the 9 quantifiers in red.

8. Comment: Line 110: “An interpolation is performed to improve the sampling in this area …”:

Once again, please define clearly every aspect of the step.

Define the calculation of “Area ratio”. The relevant lines 141-143 are completely confusing.

a. Answer: We now describe better in Step H (l202-228) the average profile extraction. The area ratio is also now explicated in the figure 2 (replacing fig 3) and in the text (l309)

9. Comment: Subsection 4.3 The purpose of the subsection is unclear. Is there any connection with the proposed system ? Please clarify

a. Answer: As now we describe the cohort and their purpose from the method section we believe this aspect has been clarified (l398-404 and l512-590 for the results). We clarified for instance with l401:“This to study the transient evolution of post surgical inflammation and fibrosis, and to evaluate how our quantifiers track the progression and resolution of corneal haze and fibrosis indicators over time, providing insights into the healing process and the potential for our quantifiers to monitor disease progression and resolution in a clinical setting.“, and l521“It suggests patient-specific trajectories of recovery that could be further investigated in larger cohorts to identify potential prognostic factors and inform personalized management strategies for patients undergoing PRK surgery.“. We believe these results have a place here since Haze is transient.

10. Comment: Fig. 6 Please define the calculation of SNR of the OCT image, the effect it has in the prediction. Also provide the numerical values.

a. Answer: Thank you, equation 6 has been added and described l483-487, legend edited snr values scale on each figure have the same scale, but we agree it needs to be in the legend for each one.

11. Comment: Line 236, 260 “FD class” is the same as “FECD” class ? If yes, please replace.

Line 236 why is “additional training” mentioned ? Has it been used or discussed somewhere else in the manuscript ?

a. Answer: Thank you, FD has been edited to FECD in the revision

b. This meant that once the model was trained on healthy/FECD corneas we presented it with Haze cornea (and not re-train it on those images) it was a methodological point. Has this was unclear, we detailed each cohort more clearly in the material and methods section, and present in the same order in the results section. For this specific point this result in exposed now in revision l488-499, and the corresponding sentence edited to “The trained model was applied to the 6 images from patients that underwent PRK surgery and reporting vizual haze (application cohort from Table 1).” At l489

12. Comment: Fig. S3, right column: please explain in the text how are the figures generated, from how many images and how are μ, σ calculated and from which specific images in each figure

a. Answer: We included this remark in the revision by defining those parameters and detailing the image generation.

Reviewer 2:

1. Comment: As PLOS One caters to a diverse audience, it would be helpful if the authors presented more information about the medical condition and the problem motivation. I appreciate the need to keep the manuscript succinct, and do not expect a full background of every medical term. However using simpler language, describing the most relevant terms, and emphasizing the motivation of this study would make the manuscript more broadly accessible.

a. Answer: We reshaped the introduction to introduce more medical background (revision l1-24), we refined the relevance of OCT (revisionl26-42) and clarified the aim (revision l43-53) to make the revised manuscript more accessible in accordance with this comment.

2. Comment: The methodologies are not described sufficiently for a comprehensive review. For example, data cleaning methodologies are only subjectively described. Feature ranges are missing. Image labeling methodology is missing. Hyperparameter selection is missing.

a. Answer: Thank you for those remarks. Overall we tried to clarify in the revision each step of the process to answer this concern. Specifically to those mentioned points: we now give a description of the methodology for the pipeline (Steps A to H, revision l118-220) we specified range values of algorithm varible; the clinical parameters used for labelling are now described (revision l76-90); hyperparameter selection precised in revision l455-459; we detail the clinical meaning of our quantifiers, their equation, range of the quantifiers are available in S1 fig; in to answer this concern.

b. We also understand that this point was more global than just mentioned point; this is we rewrote different sections. The introduction is to make our objective clearer and give more context on the clinical assessment, the automatic approach of the literature, and the missing parts. We reworked the material and methods section to clarify the methodology and the datasets. Specifically, we also detailed more extensively each cohort so the results section is more reproducible.

3. Comment: Line 15 - Suggest including the name of the machine learning algorithm in the title, abstract, and Introduction. ML is too broad a term to describe the authors’ specific work.

a. Answer: We refined the title while keeping “machine learning”, although machine learning is indeed a broad term, it still refers to a specific type of algorithm that we specifically tested in this paper and referenced both in the manuscript text (revision l448-450) and in S1 Table. However, to add some nuance on the machine learning vs. deep learning in revision l70-71.

4. Comment: Line 28 - It is not clear why the authors have started the Introduction with a reference to their own previous work. While both works relate to Fuchs, the previous work has not included any ML. I would suggest starting the Introduction with the specific problem being solved in this work, and referencing previous work in a later paragraph.

a. Answer: We agree with this remarker and changed the introduction to start by the source of the corneal Haze the PKR surgery, described it’s clinical symptoms and presented the current means of evaluation. We explain why OCT may be more reliable for quantitative assessment. We then describe the transient nature of the fibrosis occurring in Haze and why Fuch’s Dystrophy is interesting as it shares similar symptoms but permanently and is therefore suitable to create a bigger database to access such symptoms. We conclude the introduction by a clear statement of our aim (address the lack of objective haze quantification) and hypothesis.

5. Comment: Introduction - There is almost no literature review in Introduction. The authors have claimed that they are not aware of quantitative tools but they have not provided sufficient data to support their statement. Additionally, even if quantitative tools are not available, surely there would be other work related to Fuchs that motivates the authors’ work and highlights why haze diagnosis is required. Other work will also serve as a comparison for how the authors’ ML model improves upon existing evaluation performance.

a. Answer: Similarly, we addressed this remark by refactoring the introduction, adding a deeper literature review and clinical context.

6. Comment: Line 49 - What are the missing physical parameters?

a. Answer: The idea was to mention that current means of evaluation are not based on any physical parameters and are referred as “presence of haze”. We can’t list what is missing but can assess that there are none. However, we re-worked this sentence by extending what the ophthalmologist is looking for (revision l14-19) and explained the difference between Slit-lamp (subjective that can’t provide “values” or physical parameter) and OCT (that provide a signal linked to the cornea absorption) (revision l19-21). Finally, we conclude on why there is a need for quantitative/objective assessment (revision l24-25).

7. Comment: Fig 1 - What are Ref 3 and Ref 6?

a. Answer: References are now fixed in the revision (Ref3 = revision file ref [1], Ref 6 = revision file ref [2])

8. Comment: Line 56 - Please specify the algorithm instead of broad term like “automatic classification approach”.

a. Answer: We added a mention to Random Forest in revision l64 corresponding to this sentence in the initial submission.

9. Comment: Line 60 - The sentence is unclear.

a. Answer: We tried to clarify this in revision l70-75 corresponding to this sentence in the initial submission by detailing the difference between deeplearning extracting image features that can’t be interpreted by “human” and features created by humans (interpretable) passed to machine learning for classification.

10. Comment: Line 70 - The sentence is unclear.

a. Answer: We added a better description of the “pipeline” in the revision and deeply explained what we are performing in Step B (revision l127-134) to make this clearer.

11. Comment: Line 74 - What segmentation algorithms are being referred to? If there are segmentation algorithms applied here, how does this relate to authors’ pervious statement about missing quantitative tools?

a. Answer: As a big part of this paragraph was unclear for the reviewer, we re-worked it. For this specific point, we now detail in revision l92-95 that the commercial OCT device has segmentation tools that we detailed. But we mentioned their limitations and are not using them. We think that this new paragrapher from l91to 110 gives a better overview.

12. Comment: Line 75 - What is the issue with “overestimation of the epithelial layer thickness”?

a. Answer: Similarly to the previous answer, we are now detailing l94 this point.

13. Comment: Line 78 - Since the “marked morphological features” are an essential input in the authors’ model, it is insufficient to refer to their description in another reference. I would suggest including a description in this manuscript.

a. Answer: We now include this description in step B detailed in revision l127-137.

14. Comment: Line 85 - What is Ref 2?

a. Answer: References are now fixed in the revision (Ref 2 = [13] in revision)

15. Comment: Line 85 - What is the “additional alignment procedure”?

a. Answer: Now described in Step C, specifically in revision l148 to 155.

16. Comment: Line 87 - How much time and effort is required to detect the “central saturation artefact”? If its detection is essential to leveraging the authors’ methodology, would the image analysis already be done, thus making their approach extraneous?

a. Answer: Now detailed in Step B (l128-136), concerning the time and effort we added a “Processing time” section in revision l327-334.

17. Comment: Section 3.2 - Would these image processing capabilities be available to physicians who intend to use the authors’ methodology?

a. Answer: Code is available at publication but we now provide a detailed pipeline (Step A-H) and processing time information specifically l333.

18. Comment: Line 90 - Please include more description and reference for PCA.

a. Answer: Yes, now in E-F l164-195.

19. Comment: Line 90 - How are low intensity regions defined and removed?

a. Answer: Now in Step D with a precision in footnote p5.

20. Comment: Line 91 - How is 1.5 mm chosen?

a. Answer: Now detailed in revision l203 and 207: “The window width of 1.5mm was selected as a compromise between noise reduction and spatial resolution, on healthy corneas where steps A-G steps was performed automatically without need for manual correction, we observed that the depth-intensity profiles extracted was constant across the 6~mm region of interest for this window width.”

21. Comment: Line 93 - What is the functional form of the weighting function? How is the weight range chosen?

a. Answer: Now detailed in revision l217-219:”This weighting reduces the influence of peripheral windows, where curvature-induced SNR degradation and residual artefacts from the flattening procedure are most pronounced, while still retaining their contribution so as not to discard clinically relevant information”. We also precised the weight formula in equation 1). We also mention in l224-228: “In practice, it would be laborious to evaluate the landmark position across the entire 6mm region of interest; we capitalized on the use off the weighted mean of quantifiers and the extraction of 9 different quantifiers to mitigate the impact of potential landmark detection failures in some windows.”

22. Comment: Fig 3 - “peak Area Ratio” is not described in Fig 3d.

a. Answer: We now include it and named it R.

23. Comment: Line 119 - What is being defined in item 4?

a. Answer: We now start each physical parameter by its name

24. Comment: Line 143 - What is meant by “available simultaneously”?

a. Answer: We meant that once the profile is extracted, we can extract those 9 quantifiers. But we agree it was confusing, we removed this sentence, transitioning with “Once these quantifiers are extracted for each image” in revision l316.

25. Comment: Line 153 - It would be helpful if the authors provided evidence about how the classification model applied to FECD images justifies the model’s application for corneal haze detection.

a. Answer: Thank you for this remark. We now give a clear clinical point in revision l80 to 90 to detail what changes in case of fibrosis before detailing the quantifier and detailing for each one what they are “designed” to detect/measure in this context. Then, when presenting the FECD you are mentioning in this question, we detailed: “due to the transient nature of haze and the difficulty of clinical diagnosis in mild cases, insufficient validated clinical images exist to constitute a robust training database. We therefore adopted a transfer learning strategy: we trained and validated our classification model on FECD as in its advanced stages, presents similar subepithelial fibrosis localized in the anterior stroma.” (revision l358-362). Note also that now in the introduction (revision l35-38) we are citing 3 papers on that topic: “presenting with similar symptoms of blurred vision and glare [9]. The commonality of these structural alterations, particularly the anterior stromal fibrosis, suggests that a quantitative approach developed for one condition might be applicable to the other [8, 10, 11]”

26. Comment: Line 155 - Are the 9 quantifiers the features for the ML models?

a. Answer: Yes we now explicitly mention it from the beginning l64 “classification approach based on a random forest classifier applied to physics-based descriptors of fibrosis extracted from OCT images, as described in the following section.“ and after detailing the quantifiers at l316 “Once these quantifiers are extracted for each image, they are used as input features for the classification model described in the next section”

27. Comment: Line 164 - How would the doctors label the images? Would they use any of the 9 quantifiers to make their determination?

a. Answer: We now mention this explicitly in revision l80-90 by explaining what the doctor use to characterize those images. Then when detailing the quantifier, we explain their link to clinical observation (why we choose/buid them) for instance l273 “his ratio directly quantifies the hyperreflectivity observed clinically in fibrotic corneas and correlates with the density of light-scattering structures, primarily activated myofibroblasts and abnormal collagen within the anterior stroma [2]”

28. Comment: Line 169 - “very precisely” is a subjective term, and not descriptive in this context.

a. Answer: We agree, removed

29. Comment: Line 170 - How did the authors train SVM, determine its hyperparameters, and experiment with reducing overfitting?

a. Answer: We now describe Hyperparametrization in l455-459.

30. Comment: Line 172 - Does this mean that no label separation was done for implementation of SVM? How can the models performances be compared if they were trained with different datasets?

a. Answer: Maybe it was unclear but a classic methodology in the literature if sometime to train SVM in a single class (as the cited reference [20]) (for instance healthy) and then present a new dataset not from the original distribution (for instance pathological corneas FECD or Haze) to detect if it fall “within” or “without” the training distribution. In our case, it did not work, and we mention it here. Has it seemed unclear, we reworked this sentence at l433: “in which only the healthy cornea class is modelled in detail and unknown images are classified according to their deviation from this reference distribution.”

31. Comment: Line 177 - How did the authors select 50% as the probability threshold, especially because the training dataset is unbalanced?

a. Answer: We understand that the mention of the 50% was unclear, additionally in fact this information is not used in the rest of the paper as it was an arbitrary value, we removed this mention as it does not affect any result nor conclusion. We consider this probability as a continuous value that “quantifies the degree of confidence in the classification” (revision l444)

32. Comment: Line 184 - After cross-validation, since there will be 5 trained models, which one of the models was applied to the test set?

a. Answer: We now explicitly mention this in l460 “On the held-out test set, the random forest classifier attained an accuracy “. Additionally, in Table 1 we noted the use of each set, and detailed the exact methodology in revision l376-383.

33. Comment: Line 186 - The sentence structure makes it seem that the high TPR and FPR at a single threshold contributes to the high ROC, which is inaccurate since ROC is generated over the entire threshold range.

a. Answer: We agree it is indeed the reverse ROC taken at one value (threshold) that gives the sensitivity/specificity. We edited the sentence accordingly, removing the causality now in revision l460-463.

34. Comment: Line 218 - Suggest changing AI to ML to remain consistent with the rest of the manuscript.

a. Answer: We strongly agree to replace this part. We also clarified why we specifically use ML in revision l64-75.

Attachment

Submitted filename: Response to Reviewers.docx

pone.0328320.s006.docx (33.9KB, docx)

Decision Letter 1

Georgios Labiris

12 Apr 2026

Dear Dr. Soubeiran,

The revised manuscript shows clear improvement in clarity and structure; however, both reviewers still identify important issues that must be addressed before acceptance. In particular, the authors should (i) substantially improve figure quality and ensure all figures are publication-ready and correctly integrated into the main text, (ii) provide detailed clarification of the methodological pipeline with specific attention to potential data leakage, cohort structure, and preprocessing steps, and (iii) justify or revise the analytical choices that may not be standard or fully supported (e.g., PCA variance assumptions, SVM-based reference distribution, and accuracy reporting in imbalanced settings). Reviewer 2’s concerns regarding possible data leakage and the unexpectedly high model performance are particularly critical and must be explicitly addressed with additional methodological detail or reanalysis where necessary. Please provide a clearer explanation and corrections to the confusing figure elements (especially Fig. 3a), and improve the interpretability of the results. Minor linguistic and formatting issues can be corrected during revision, but do not affect the decision.

plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

  • A letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.

  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.

  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols.

As the corresponding author, your ORCID iD is verified in the submission system and will appear in the published article. PLOS supports the use of ORCID, and we encourage all coauthors to register for an ORCID iD and use it as well. Please encourage your coauthors to verify their ORCID iD within the submission system before final acceptance, as unverified ORCID iDs will not appear in the published article. Only  the individual author can complete the verification step; PLOS staff cannot  verify ORCID iDs on behalf of authors.

We look forward to receiving your revised manuscript.

Kind regards,

Georgios Labiris, MD, PhD

Academic Editor

PLOS One

Journal Requirements:

If the reviewer comments include a recommendation to cite specific previously published works, please review and evaluate these publications to determine whether they are relevant and should be cited. There is no requirement to cite these works unless the editor has indicated otherwise.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

Reviewer #1: (No Response)

Reviewer #2: All comments have been addressed

**********

2. Is the manuscript technically sound, and do the data support the conclusions??>

Reviewer #1: Yes

Reviewer #2: Partly

**********

3. Has the statistical analysis been performed appropriately and rigorously? -->?>

Reviewer #1: Yes

Reviewer #2: I Don't Know

**********

4. Have the authors made all data underlying the findings in their manuscript fully available??>

The PLOS Data policy

Reviewer #1: No

Reviewer #2: Yes

**********

5. Is the manuscript presented in an intelligible fashion and written in standard English??>

Reviewer #1: Yes

Reviewer #2: Yes

**********

Reviewer #1: The authors have made substantial revisions to the manuscript, which is now more easily comprehendable.

The main general comment is that the quality of all figures is far below standard, thus, many of them are hardly readable. They should be reproduced with higher dpi.

Ideally they should be included in the relevant body of the text, rather than put at the end of the whole document.

I appreciate the description of the steps. Just few clarifications:

Line 140: Step C: is the Gaussian 1D (for each A-scan), or 2D ? If the latter case, is the σ same for both dimensions ? (are the pixels square) ?

Step E. Are the two margins (50μm and 30μm) set manually ?

Step F: The variance of the 1st and 2nd PCA component are set to 94% and 4%, however there seems to be no reason for this to hold in the general case.

Step G: Is the filtering performed in 1D ?

Step H: is the 1.5mm-width window placed in a overlapping manner, or is it applied 4 times to cover the 6mm width ?

Fig.3(a) needs to be explained in the text, because the authors draw important conclusions from it, about the quality of the classifier. The shape is not self-explanatory, the Y axis label is still confusing (does not remind of any probability density function) and the different colors that seem to contain are not visible. Its title is also too general.

Line 341-342 needs rewriting

In Fig. 5b the relevant probability spans [-1,1], whereas in (a) the range is [0,1]

Reviewer #2: I would like to thank the authors for a thorough review of the reviewers’ comments, and a major re-write of the manuscript in order to address them. The manuscript is a lot clearer throughout and the Introduction helps establish the problem motivation well.

After another review of the manuscript, I have some concerns regarding possible data leakage and the results being close to perfect, which is hardly seen in the real world. I would appreciate it if the authors could address the following minor and major comments:

1. Line 69: Sentence structure

2. Line 70: *Additionally*

3. Line 70, 72: *deep-learning*

4. Line 152: Can the authors include a summary of how their normalized cross-correlation method enables a second re-alignment?

5. Line 154: *alignment*

6. Line 184: *mean-centered*

7. Line 208: The statement indicates that the steps A-G would be applied differently to healthy and unhealthy corneas. This could introduce data leakage, and data leakage can potentially nullify later analyses that rely on these steps. Can the authors include in the manuscript how they ensured that that no data leakage occurred during their analyses?

8. Line 301: symbol *alpha*

9. Line 340: Sentence structure

10. Line 382: Can the authors provide data on how many patients were the images collected from, and how many images belonged to the same patient?

11. Line 434: Can the authors clarify whether “unknown” images are a different image dataset than the “FECD” images described previously?

12. Line 433-438: While using SVM to create a reference distribution using single label data can work in some scenarios, it’s not a widely used method, compared to let’s say, Gaussian Discriminative Analysis. As such, this section does not seem to contribute critical information and I would suggest removing it and only focusing on the two-label training.

13. Line 452: *training*

14. Line 460: Accuracy is not a valid metric for imbalanced datasets, so I would suggest removing it.

15. Line 460: Which RF model out of the 5 cross-validated models was used to evaluate the test set?

16. Line 462: I am concerned about the extremely high ROC and predictive power of the model. The results are “too good to be true” and indicate that this is essentially a trivial problem. Since the authors have previously stated that it is not easy to use the images to detect FECD and haze, while at the same time, a random forest model can achieve an ROC of 99.8%, there appears to be a mismatch. I would encourage the authors to carefully review their data generation process to identify and eliminate any “hidden” sources of data leakage.

17. Fig 3a: I have not previously come across this plot format for classification results, and therefore am not able to understand it. It will be helpful to explain the plot in more detail for interpretability.

18. Paragraph 500: It’s unclear how the probability associates with real patient outcomes in the longitudinal study. There does not appear to be any medical diagnosis to establish “ground truth”, which can be compared with the model predictions.

19. Paragraph 524: Since a primary contribution of the manuscript is “interpretability” of the models, I would suggest improving this section by expanding how the Shap values relate to the attributes of the disease, and whether the high importance features align with the expectations of the medical professionals.

20. All figures: Can the authors ensure that the figures in the final publication are high quality as the ones in the manuscript are very low-resolution and difficult to read, though it may be an artifact of journal post-processing?

**********

what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review?  For information about this choice, including consent withdrawal, please see our Privacy Policy

Reviewer #1: No

Reviewer #2: No

**********

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

To ensure your figures meet our technical requirements, please review our figure guidelines: https://journals.plos.org/plosone/s/figures

You may also use PLOS’s free figure tool, NAAS, to help you prepare publication quality figures: https://journals.plos.org/plosone/s/figures#loc-tools-for-figure-preparation.

NAAS will assess whether your figures meet our technical requirements by comparing each figure against our figure specifications.

PLoS One. 2026 Oct 22;21(10):e0328320. doi: 10.1371/journal.pone.0328320.r004

Author response to Decision Letter 2


16 Jul 2026

We would like to thank the reviewer for their detailed and informative comments. We addressed all comments as listed below:

Reviewer #1:

1. Comment 1 reviewer 1:The main general comment is that the quality of all figures is far below standard, thus, many of them are hardly readable. They should be reproduced with higher dpi. Ideally they should be included in the relevant body of the text, rather than put at the end of the whole document.

o Answer 1 (to reviewer 1): We fully agree on the importance of figure quality. The apparent low resolution is an artefact of the journal’s compilation workflow: following the editorial policy, the figures are not embedded in the manuscript file but are uploaded as individual high-resolution TIFF files (>600 dpi), which the editorial system inserts automatically at full resolution in the reviewer PDF. In the present resubmission, per the editorial office’s request, the figures have accordingly been removed from within the manuscript file, leaving only the captions in place, and are provided as separate TIFF image files. We remain happy to adjust figure placement or format according to the journal’s production requirements.

2. Comment 2 reviewer 1:Line 140: Step C: is the Gaussian 1D (for each A-scan), or 2D ? If the latter case, is the σ same for both dimensions ? (are the pixels square) ?

o Answer 2 (to reviewer 1): 2D Gaussian is performed; lateral/axial pixel size varies from 1 to 2 depending on the recording mode of the device (Line, Pachy, Cross, and Pachywide) images; axial pixel size is consistent across modes (4.3um), all modes are considered laterally in a physical distance coordinate system (in mm). To clarify this, we added line 118 (“Standardisation of clinical images” first paragraph): “Images have the same axial pixel size of 4.3um, lateral pixel size varies between 5.85um to 8.51um depending on device recording mode. Images are considered on respective individual coordinate systems.”; We also specified the acquisition mode and its repartition in “Databases and experimental design.”; Added: “2D Gaussian”

3. Comment 3 reviewer 1:Step E. Are the two margins (50μm and 30μm) set manually ?

o Answer 3 (to reviewer 1): Automatically validated by the user. Edited l174: “Automatically excluded”

4. Comment 4 reviewer 1:Step F: The variance of the 1st and 2nd PCA component are set to 94% and 4%, however there seems to be no reason for this to hold in the general case.

o Answer 4 (to reviewer 1): The 94% and 4% values reported for PC1 and PC2 are not preset parameters, but empirically observed values computed by running the PCA on each individual image of the normative dataset (n = 83 normal corneas), following flattening and segmentation of the stroma into 20 equal-thickness sub-layers. The method was developed and iteratively refined on a larger set of normal corneal SD-OCT images prior to this study, providing empirical confidence in the stability of the component interpretation across normal corneas acquired with the same device and protocol. Component selection is not based on these specific percentages, but on the physical interpretation of each component – which is grounded in the known physics of the SD-OCT corneal signal. In this strongly constrained data structure (same device, same protocol, homogeneous normal stroma), PC1 captures the dominant depth-dependent attenuation trend, and PC2 captures the first structured lateral deviation from it, corresponding to the posterior stromal artifact. This interpretation was validated consistently across all images in the dataset. Indirect validation is provided by the fitting results: after correction, stromal attenuation profiles are well described by a mono-exponential decay model (Birge ratio Br < 10 for all normal corneas, IDR: 1.8–4.1), as expected for homogeneous tissue. This would not hold if the PCA-based correction introduced systematic distortions in the depth-attenuation profile. We acknowledge that individual variability around the reported variance values was not documented – these should be understood as representative values observed across the normative dataset, not as fixed algorithmic thresholds.

o We therefore propose the edit by changing the sentence “A first PCA on the full matrix identifies two dominant components: PC1 (94% of variance), capturing the overall depth-dependent attenuation and the curvature-driven lateral intensity decrease, and PC2 (4% of variance), isolating the structured lateral signature of the posterior artefact.” to “A first PCA on the full matrix identifies two dominant components(PC1 and PC2). Across a normative dataset, PC1 and PC2 accounted for approximately 94% and 4% of the variance, respectively. They respectively capture the overall depth-dependent attenuation and the curvature-driven lateral intensity decrease for PC1, and the structured lateral signature of the posterior artefact for PC2 as reported in […].”

5. Comment 5 reviewer 1:Step G: Is the filtering performed in 1D ?

o Answer 5 (to reviewer 1): We are smoothing a curve corresponding to the anterior surface intensity signal. To clarify this: edited l198 to “The corrected image is normalised by the smoothed anterior surface intensity signal (smoothing the detected curve using second-order Savitzky-Golay filter, sliding window: 115 pixels) to compensate for the peripheral intensity loss associated with corneal curvature geometry.”

6. Comment 6 reviewer 1: Step H: is the 1.5mm-width window placed in a overlapping manner, or is it applied 4 times to cover the 6mm width ?

o Answer 6 (to reviewer 1): The 1.5mm-width window is in indeed applied in an overlapping manner (“moving window”), to clarify this we added a footnote for “sliding window” : “sliding window method corresponds to incrementally (pixel-wise) slide a working window, it can mathematically be seen as a convolution by a mean operator with lateral dimension 1.5mm.”

7. Comment 7 reviewer 1: Fig.3(a) needs to be explained in the text, because the authors draw important conclusions from it, about the quality of the classifier. The shape is not self-explanatory, the Y axis label is still confusing (does not remind of any probability density function) and the different colors that seem to contain are not visible. Its title is also too general.

o Answer 7 (to reviewer 1): the Y-axis often represents kernel density estimate (KDE) width, sometimes called “density” or “Probability density”. We propose to edit panel a legend from : “distribution of predicted fibrosis probabilities ($p_{\mathrm{FECD}}$) (Healthy mean predicted probability 10.0\% std: 13.2\%; Fuchs' Dystrophy mean predicted probability: 90.9\% std: 14.2\%; Haze mean predicted probability 96.0\% std: 4.7\%).” To “Distribution of predicted FECD probabilities ($p_{\mathrm{FECD}}$) for healthy corneas, FECD, and visual-haze images in the independent test set. The horizontal axis corresponds to the model-estimated probability of FECD represented by the violin curve. Healthy samples cluster near 0, whereas FECD and visual-haze samples cluster near 1 (healthy mean predicted probability 10.0\% std: 13.2\%; Fuchs' Dystrophy mean predicted probability: 90.9\% std: 14.2\%; Haze mean predicted probability 96.0\% std: 4.7\%).”

o Additionally in the text we edited “reveals a clear separation between the control and pathological groups” to “reveals a clear separation in the x-axis (FECD probability) between the control and pathological groups (y-axis being the model-estimated probability of FECD represented by the violin curve for data distribution visualization).”

o We renamed the y-axis to “Kernel density estimate (a.u.)” for clarity

8. Comment 8 reviewer 1: Line 341-342 needs rewriting

o Answer 8 (to reviewer 1): Text was: “In addition to user quality control of the automatic detection of the artefact removal and these landmarks via the graphical user interface, the pipeline incorporates several complementary mechanisms to maintain robustness when these landmarks are difficult to detect due to elevated noise or pathological morphology”

o Proposed version: “In addition to user-driven quality control via the graphical user interface for validating artefact removal and landmark detection, the pipeline integrates complementary mechanisms to preserve robustness when landmark identification is compromised by high noise levels or pathological morphologies:”

9. Comment 9 reviewer 1: In Fig. 5b the relevant probability spans [-1,1], whereas in (a) the range is [0,1]

o Answer 9 (to reviewer 1): Indeed, right panel represents the value relative to the pre-surgery value, as left panel span initial values from 0 to 1, relatively to initial value it can span from -1 to 1. In the figure caption “Predicted probabilities pFECD (left) and relative to initial pFECD probability (right) of FECD classification in Pre/Post-surgery follow-up for 9 patients.”

o It was also precised line 510 “We found it insightful to represent this follow-up relatively to the initial value of the probability and therefore represent potential improvement or degradation of such quantifier” and line 512 “Comparing Fig 5a and Fig 5b, we can observe that patients with low initial pFECD values tend to maintain high probabilities over time, whereas patients with higher initial pFECD values show a more pronounced decrease, suggesting a potential correlation between the initial severity of fibrosis and the trajectory of recovery.”

Reviewer #2:

1. Comment 1 reviewer 2: Line 69: Sentence structure

o Answer 1 (to reviewer 2): Sentence was : “It is also repported as a competitive compare to deep learning on tabular data” we edit it to “It has also been reported to be competitive with deep learning methods on tabular data”

2. Comment 2 reviewer 2: Line 70: *Additionally*:

o Answer 2 (to reviewer 2): corrected thank you

3. Comment 3 reviewer 2: Line 70, 72: *deep-learning*.

o Answer 3 (to reviewer 2): corrected thank you

4. Comment 4 reviewer 2: Line 152: Can the authors include a summary of how their normalized cross-correlation method enables a second re-alignment?

o Answer 4 (to reviewer 2): Yes, we propose to complete this sentence by: “This second alignment is performed by normalised cross-correlation [17]. Each A-scan is compared to the central A-scan serving as reference: the cross-correlation between the two signals exhibits a peak corresponding to their relative axial offset. Parabolic interpolation around this peak is then applied to achieve sub-pixel estimation of the shift, which is subsequently used to align the A-scans, ensuring sub-pixel co-registration along their first interface.”

5. Comment 5 reviewer 2: Line 154: *alignment*:

o Answer 5 (to reviewer 2): corrected thank you

6. Comment 6 reviewer 2: Line 184: *mean-centered*:

o Answer 6 (to reviewer 2): corrected thank you

7. Comment 7 reviewer 2: Line 208: The statement indicates that the steps A-G would be applied differently to healthy and unhealthy corneas. This could introduce data leakage, and data leakage can potentially nullify later analyses that rely on these steps. Can the authors include in the manuscript how they ensured that that no data leakage occurred during their analyses?

o Answer 7 (to reviewer 2): The sentence is : “The window width of 1.5 mm was selected as a compromise between noise reduction and spatial resolution, on healthy corneas where steps A–G steps was performed automatically without need for manual correction, we observed that the depth-intensity profiles extracted was constant across the 6 mm region of interest for this window width.”

o We believe the confusion comes from “The window width of 1.5 mm was selected […] on healthy corneas”, we didn’t mean to insinuate there was a difference on the processing, and there is in fact no difference in this process, all images a processed similarly. The sentence here missed a “break” : The window width of 1.5 mm was selected as a compromise […]. On healthy corneas”,

o We can reformulate for the sake of clarity: “The window width of 1.5 mm was selected as a compromise between noise reduction and spatial resolution. This parameter was determined a priori and applied uniformly to all samples. In healthy corneas, where steps A–G were performed fully automatically without manual correction, the resulting depth–intensity profiles were observed to be consistent across the 6 mm region of interest for this window width.”

o Image processing and feature extraction were conducted after preprocessing using the same procedure for all samples.

8. Comment 8 reviewer 2: Line 301: symbol *alpha*

o Answer 8 (to reviewer 2): corrected thank you

9. Comment 9 reviewer 2: Line 340: Sentence structure:

o Answer 9 (to reviewer 2): changing sentence “First images are acquired by professionals trained in OCT acquisition and are match with the clinical quality standards” to “images are acquired by professionals trained in OCT acquisition and meet established clinical quality standards”

10. Comment 10 reviewer 2: Line 382: Can the authors provide data on how many patients were the images collected from, and how many images belonged to the same patient?

o Answer 10 (to reviewer 2): Thank you for this comment; table 1 edited, we also added a column for the acquisition mode and a detail on pixel size, as it was also asked in another comment.

11. Comment 11 reviewer 2: Line 434: Can the authors clarify whether “unknown” images are a different image dataset than the “FECD” images described previously?

o Answer 11 In this approach, during training there is no other image than healthy images: unknown images mean image the have never be seen including potential FECD images. Considering next reviewer comment we agreed on removing this part.

12. Comment 12 reviewer 2: Line 433-438: While using SVM to create a reference distribution using single label data can work in some scenarios, it’s not a widely used method, compared to let’s say, Gaussian Discriminative Analysis. As such, this section does not seem to contribute critical information and I would suggest removing it and only focusing on the two-label training.

o Answer 12 (to reviewer 2): We agreed on removing this part.

13. Comment 13 reviewer 2: Line 452: *training*:

o Answer 13 (to reviewer 2): corrected thank you

14. Comment 14 reviewer 2: Line 460: Accuracy is not a valid metric for imbalanced datasets, so I would suggest removing it.

o Answer 14 (to reviewer 2): we agreed

15. Comment 15 reviewer 2: Line 460: Which RF model out of the 5 cross-validated models was used to evaluate the test set?

o Answer 15 (to reviewer 2): Cross-validation was used on the training set to compare candidate models and estimate their performance. The selected model (Random Forest) was then re-instantiated and retrained on the full training dataset using the chosen hyperparameters. This final model was subsequently evaluated on the independent test set.

o We propose the clarification: “[…] Python [25]. Model selection was performed using 5-fold stratified cross-validation on the training set, based on the weighted F1-score, to compare candidate classifiers. The random forest model consistently achieved the best performance and was therefore selected. Following this selection, the final model was re-instantiated and retrained on the full training dataset using the chosen hyperparameters. On the held-out test set […]”

16. Comment 16 reviewer 2: Line 462: I am concerned about the extremely high ROC and predictive power of the model. The results are “too good to be true” and indicate that this is essentially a trivial problem. Since the authors have previously stated that it is not easy to use the images to detect FECD and haze, while at the same time, a random forest model can achieve an ROC of 99.8%, there appears to be a mismatch. I would encourage the authors to carefully review their data generation process to identify and eliminate any “hidden” sources of data leakage.

o Answer 16 (to reviewer 2): We understand the reviewer's comment, and carefully investigated for an eventual leakage, however images are processed similarly for each condition, devices, and practitioners are shared for each condition. We also want to highlight that we trained and tested the model on a set of images and further tested it as presented in the manuscript on repeatability analysis for both condition-varying practitioners. In our introduction, we emphasized the need for quantitative, objective, and reproducible, not hard we focus firstly on the need for objective, quantitative and reproducible analysis. It is now accepted that machine learning can accurately reproduce human judgement in many cases when done carefully, as we tried to do here. This is still a proof of principle, however, and evaluation on larger, multicentric datasets acquired by diverse acquisition devices, protocols, and practitioners may raise additional challenges

17. Comment 17 reviewer 2: Fig 3a: I have not previously come across this plot format for classification results, and therefore am not able to understand it. It will be helpful to explain the plot in more detail for interpretability.

o Answer 17 (to reviewer 2): the Y-axis often represents kernel density estimate (KDE) width, sometimes called “density” or “Probability density”. We propose to edit panel a legend from : “distribution of predicted fibrosis probabilities ($p_{\mathrm{FECD}}$) (Healthy mean predicted probability 10.0\% std: 13.2\%; Fuchs' Dystrophy mean predicted probability: 90.9\% std: 14.2\%; Haze mean predicted probability 96.0\% std: 4.7\%).” To “Distribution of predicted FECD probabilities ($p_{\mathrm{FECD}}$) for healthy corneas, FECD, and visual-haze images in the independent test set. The horizontal axis corresponds to the model-estimated probability of FECD represented by the violin curve. Healthy samples cluster near 0, whereas FECD and visual-haze samples cluster near 1 (healthy mean predicted probability 10.0\% std: 13.2\%; Fuchs' Dystrophy mean predicted probability: 90.9\% std: 14.2\%; Haze mean predicted probability 96.0\% std: 4.7\%).”

o Additionally in the text we edited “reveals a clear separation between the control and pathological groups” to “reveals a clear separation in the x-axis (FECD probability) between the control and pathological groups (y-axis being the model-estimated probability of FECD represented by the violin curve for data distribution visualization).”

o We renamed the y-axis to “Kernel density estimate (a.u.)” for clarity

18. Comment 18 reviewer 2: Paragraph 500: It’s unclear how the probability associates with real patient outcomes in the longitudinal study. There does not appear to be any medical diagnosis to establish “ground truth”, which can be compared with the model predictions.

o Answer 18 (to reviewer 2): This is an exploratory analysis, investigating further uses of the proposed model. It will need to be strengthened by a proper longitudinal study where a clinical baseline would be constructed for a proper clinical evaluation. As we are exploring the potential application of the automatic and reproducible quantification we are proposing.

19. Comment 19 reviewer 2: Paragraph 524: Since a primary contribution of the manuscript is “interpretability” of the models, I would suggest improving this section by expanding how the Shap values relate to the attributes of the disease and whether the high-importance features align with the expectations of the medical professionals.

o Answer 19 (to reviewer 2): The nine quantifiers have been build to probe the specific features described the 3 morphological feature described in “Morphological parameters of subepithelial fibrosis”. We propose to edit the l528 : « higher values of peak width, sigma, area ratio, and depth; and lower values of beta (reduced stromal intensity signal decay) are associated with higher SHAP values” by adding: “Those parameters relating by construction to the 3 morphological parameters of subepithelial fibrosis : sub-epithelial hyper reflectivity (link to area ratio), blurred epithelium/stroma interface (link to peak width, sigma and depth), altered reflectivity depth profile, stromal transparency (link to beta).”

20. Comment 20 reviewer 2: All figures: Can the authors ensure that the figures in the final publication are high quality as the ones in the manuscript are very low-resolution and difficult to read, though it may be an artifact of journal post-processing?

o Answer 20 (to reviewer 2): Indeed, the pixelated appearance is an artefact of the journal’s post-processing: following the editorial policy, images are removed from the manuscript file and uploaded individually as high-resolution TIFF files (>600 dpi), which the editorial system inserts automatically at full resolution in the compiled PDF. In this resubmission, per the editorial office’s request, the figures have been removed from within the manuscript file (captions retained) and are provided as separate TIFF image files, ensuring that the versions appearing in the final publication are of full quality.

Attachment

Submitted filename: Responses_To_Reviewers_3.docx

pone.0328320.s007.docx (25.4KB, docx)

Decision Letter 2

Helen Howard

27 Sep 2026

Machine learning quantitative OCT derived fibrosis indicators for corneal haze evaluation

PONE-D-25-32169R2

Dear Dr. Soubeiran,

We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.

Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.

An invoice will be generated when your article is formally accepted. Please note, if your institution has a publishing partnership with PLOS and your article meets the relevant criteria, all or part of your publication costs will be covered. Please make sure your user information is up-to-date by logging into Editorial Manager at Editorial Manager® and clicking the ‘Update My Information' link at the top of the page. For questions related to billing, please contact billing support.

If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

Kind regards,

Helen Howard

Staff Editor

PLOS One

Additional Editor Comments (optional):

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

Reviewer #2: All comments have been addressed

**********

2. Is the manuscript technically sound, and do the data support the conclusions??>

Reviewer #2: Yes

**********

3. Has the statistical analysis been performed appropriately and rigorously? -->?>

Reviewer #2: Yes

**********

4. Have the authors made all data underlying the findings in their manuscript fully available??>

The PLOS Data policy

Reviewer #2: No

**********

5. Is the manuscript presented in an intelligible fashion and written in standard English??>

Reviewer #2: Yes

**********

Reviewer #2: (No Response)

**********

what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review?  For information about this choice, including consent withdrawal, please see our Privacy Policy

Reviewer #2: No

**********

Acceptance letter

Helen Howard

PONE-D-25-32169R2

PLOS One

Dear Dr. Soubeiran,

I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS One. Congratulations! Your manuscript is now being handed over to our production team.

At this stage, our production department will prepare your paper for publication. This includes ensuring the following:

* All references, tables, and figures are properly cited

* All relevant supporting information is included in the manuscript submission,

* There are no issues that prevent the paper from being properly typeset

You will receive further instructions from the production team, including instructions on how to review your proof when it is ready. Please keep in mind that we are working through a large volume of accepted articles, so please give us a few days to review your paper and let you know the next and final steps.

Lastly, if your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

You will receive an invoice from PLOS for your publication fee after your manuscript has reached the completed accept phase. If you receive an email requesting payment before acceptance or for any other service, this may be a phishing scheme. Learn how to identify phishing emails and protect your accounts at https://explore.plos.org/phishing.

If we can help with anything else, please email us at customercare@plos.org.

Thank you for submitting your work to PLOS One and supporting open access.

Kind regards,

PLOS One Editorial Office Staff

on behalf of

Dr Helen Howard

Staff Editor

PLOS One

Associated Data

    This section collects any data citations, data availability statements, or supplementary materials included in this article.

    Supplementary Materials

    S1 Fig. Distribution on training set of the 9 quantifiers for each condition.

    Statistics are performed with Mann-Whitney U-test (no normality assumption) using Bonferonni correction of p-values (***: p-value<0.001/m; **; p-value<0.01/m; p-value<0.05/m; m = 9).

    (TIFF)

    pone.0328320.s001.tiff (48.1MB, tiff)
    S1 Table. Cross-validation scores of classic machine learning classification models.

    (PDF)

    pone.0328320.s002.pdf (57.8KB, pdf)
    S2 Fig. Raw Repeatability data example with zoom on the apex region.

    (a) healthy cornea example, (b) FECD cornea example.

    (TIFF)

    pone.0328320.s003.tiff (58.2MB, tiff)
    S3 Fig. The 3 Misclassified corneas and corresponding SD-OCT raw signal (right top) associated standardized signal and depth profile (right) as a density map to represent signal disparities for the depth profile (one curve is on “column” or axial profile, total number of curves is 750 for a 6 mm region), μ is the mean of the curves and σ is the standard deviation of the curves.

    (a) false Fuchs dystrophy cornea. (b–c) false Healthy cornea.

    (TIFF)

    pone.0328320.s004.tiff (66.4MB, tiff)
    Attachment

    Submitted filename: Response to Reviewers.docx

    pone.0328320.s006.docx (33.9KB, docx)
    Attachment

    Submitted filename: Responses_To_Reviewers_3.docx

    pone.0328320.s007.docx (25.4KB, docx)

    Data Availability Statement

    The minimal data set underlying the findings of this study: the values of the OCT-derived quantifiers for all corneas, the per-sample model predictions, and the values behind all reported means, standard deviations, statistical tests and figures, is fully anonymised and is available as Supporting Information (S1 Dataset) and in the public repository https://github.com/Coohrentiin/Subepithelial-Fibrosis-Quantifiers. The analysis code is publicly available in the same repository. The raw source data (individual SD-OCT corneal images from which the quantifiers were derived) cannot be shared publicly. These are patient-derived clinical images: the informed consent obtained from participants and the institutional approvals granted for the study did not provide for public release of the raw images, and their dissemination would compromise patient confidentiality. The images are held by the Centre Hospitalier National d’Ophtalmologie (CHNO) des Quinze-Vingts, Paris, France, and the study was approved by the Ethics Committee of the French Society of Ophthalmology (IRB #00008855). Researchers who meet the criteria for access to confidential clinical data may request the raw images from the Data Protection Officer of the CHNO des Quinze-Vingts digital health data warehouse (Entrepôt de Données de Santé), a non-author institutional contact reachable at dpo@15-20.fr, who is able to handle data access requests from researchers external to the hospital. Requests concerning the ethics approval may additionally be addressed to the Ethics Committee of the French Society of Ophthalmology (Société Française d’Ophtalmologie, 17 Villa d’Alésia, 75014 Paris, France; sfo@sfo.asso.fr), chaired by Dr Jean-Jacques Saragouss.